PORTABLE GAS ANALYZER
NOx SO2 CO CO2 O2 CH4
Portable Gas Analyzer
PG-300 Series
Laboratory-precise analyses, anywhere.
External Dimensions (mm)
PG-300 Series Analyzer Unit (Side guards excluded)
265 510
265
Bulletin:HRE-2879A Printed in Japan TH-G(SK)00
300 520
265
PG-300 Series Analyzer Unit (Side guards included)
Outlet
DetectorMeasurement cell
Light sourceInlet(Sample gas or Reference gas)
■ Cross-Flow Modulation advanced efficiency of NDIR analysis
In PG-300, Cross-Flow Modulation is applied to SO2, CO, and new CH4 analyzer for Non-Dispersive Infrared Absorption (NDIR) method. With Cross-Flow Modulation NDIR method, sample gas and reference gas flow into a single measurement cell switching one by one, and it brings about advantages that no optical adjustment is required, the zero point is kept stable, and the sample cell remains clean and it reduces span drift. The equipments will be kept safe for a long time as well. Cross-Flow Modulation Chemiluminescence detection method is already introduced for NOx analyzerin previous model and has the same effects as aforesaid analyzers.
■Capable of measuring methane (CH4) for expanded options
Improving on previous models, the new PG-300 is equipped with a methane (CH4) analyzer that is ideally suited for many current and emerging applications such as biomass combustion.
■Ethernet communication facilitates data management*1
Standard Ethernet interface for connection to LAN environments, enabling real-time data import.
Measurement So Easy It's Almost InstinctiveLaboratory-level precision in a portable unit for real-world measurements in the field.
PORTABLE GAS ANALYZER
NOx SO2 CO CO2 O2 CH4
The New Possibilities of Gas Analysis Begin with "Precision Mobility"For situations when you can only take measurements in the field, but you want the same precision that you get in the laboratory: Horiba presents the PG-300 Portable Gas Analyzer. The PG-300 offers the same accuracy and reliability of laboratory measurements in a portable unit that can measure five crucial components in the field. It offers a faster response time than existing models and yet is 20% lighter. Warm-up time has also been cut in half to facilitate mobile measurement. The PG-300 also has a touch screen for easy operation and a new design that protects the unit from shocks and vibrations — features that enhance its usefulness in the field. The PG-300 is the analyzer of the future — but it's here today, ready to meet the need for increasingly precise measurements with the mobility of on-site measurement capability.
1 2
The PG-300 achieves measurement performance equal to laboratory equipment in a highly portable package. The Cross-Flow Modulation type analyzers improve reliability and the addition of CH4 analyzer greatly expands applicability. With only half the warm-up time over the previous generation PG, operational performance has dramatically increased. A new timer function has been added for saving preparation time and quick start.
■Collecting data over LAN network*1
Once the network connection such as LAN has been set up, data can be uploaded while you are staying at the office or the laboratory, a distance away from where PG-300 is placed.
■Warm-up time has been cut in half, greatly reducing the instrument’s ready-to-measure time
Previous models required an hour of warm-up time. The PG-300 has been reduced to 30 minutes on the PG-300, greatly reducing the time required for measurement preparation.
■Timer function enables automatic instrument start and sleep modes
For example, setting the PG-300’s automatic start time 30 minutes ahead of when measurements are needed eliminates your need to wait for the instrument to warm up; it will be ready when you are. There is also a sleep mode that reduces power use when the unit is idle.
■Reduced response time for SO2 analyzer
The response time of the SO2 analyzer is faster than on previous models, increasing the overall measurement performance.
Timer is activated
Measurement begins
Warm-up
*1 Requires separate software.
unctions
Field×LabTo provide complete support for measurements in the field, the PG-300 body has been made up to 20% lighter than previous models. Side guards* are available to prevent from unexpectedimpacts during transport.Designed in this way for easy and safe transport, the PG-300 provides full support for measurements in the field.*Please see the back of the brochure.
ugged ightweight esign
Lighter than existing models to make transport easy.
Advanced measurement needs met with advanced functions.
●Expansion of Cross-Flow Modulation type detector ●Capable of measuring methane (CH4) ●Shorter warm-up time ●Timer function ●Ethernet compatible ●Capable of remote operation
Cross-Flow Modulation NDIR detector
Outlet
DetectorMeasurement cell
Light sourceInlet(Sample gas or Reference gas)
■ Cross-Flow Modulation advanced efficiency of NDIR analysis
In PG-300, Cross-Flow Modulation is applied to SO2, CO, and new CH4 analyzer for Non-Dispersive Infrared Absorption (NDIR) method. With Cross-Flow Modulation NDIR method, sample gas and reference gas flow into a single measurement cell switching one by one, and it brings about advantages that no optical adjustment is required, the zero point is kept stable, and the sample cell remains clean and it reduces span drift. The equipments will be kept safe for a long time as well. Cross-Flow Modulation Chemiluminescence detection method is already introduced for NOx analyzerin previous model and has the same effects as aforesaid analyzers.
■Capable of measuring methane (CH4) for expanded options
Improving on previous models, the new PG-300 is equipped with a methane (CH4) analyzer that is ideally suited for many current and emerging applications such as biomass combustion.
■Ethernet communication facilitates data management*1
Standard Ethernet interface for connection to LAN environments, enabling real-time data import.
Measurement So Easy It's Almost InstinctiveLaboratory-level precision in a portable unit for real-world measurements in the field.
PORTABLE GAS ANALYZER
NOx SO2 CO CO2 O2 CH4
The New Possibilities of Gas Analysis Begin with "Precision Mobility"For situations when you can only take measurements in the field, but you want the same precision that you get in the laboratory: Horiba presents the PG-300 Portable Gas Analyzer. The PG-300 offers the same accuracy and reliability of laboratory measurements in a portable unit that can measure five crucial components in the field. It offers a faster response time than existing models and yet is 20% lighter. Warm-up time has also been cut in half to facilitate mobile measurement. The PG-300 also has a touch screen for easy operation and a new design that protects the unit from shocks and vibrations — features that enhance its usefulness in the field. The PG-300 is the analyzer of the future — but it's here today, ready to meet the need for increasingly precise measurements with the mobility of on-site measurement capability.
1 2
The PG-300 achieves measurement performance equal to laboratory equipment in a highly portable package. The Cross-Flow Modulation type analyzers improve reliability and the addition of CH4 analyzer greatly expands applicability. With only half the warm-up time over the previous generation PG, operational performance has dramatically increased. A new timer function has been added for saving preparation time and quick start.
■Collecting data over LAN network*1
Once the network connection such as LAN has been set up, data can be uploaded while you are staying at the office or the laboratory, a distance away from where PG-300 is placed.
■Warm-up time has been cut in half, greatly reducing the instrument’s ready-to-measure time
Previous models required an hour of warm-up time. The PG-300 has been reduced to 30 minutes on the PG-300, greatly reducing the time required for measurement preparation.
■Timer function enables automatic instrument start and sleep modes
For example, setting the PG-300’s automatic start time 30 minutes ahead of when measurements are needed eliminates your need to wait for the instrument to warm up; it will be ready when you are. There is also a sleep mode that reduces power use when the unit is idle.
■Reduced response time for SO2 analyzer
The response time of the SO2 analyzer is faster than on previous models, increasing the overall measurement performance.
Timer is activated
Measurement begins
Warm-up
*1 Requires separate software.
unctions
Field×LabTo provide complete support for measurements in the field, the PG-300 body has been made up to 20% lighter than previous models. Side guards* are available to prevent from unexpectedimpacts during transport.Designed in this way for easy and safe transport, the PG-300 provides full support for measurements in the field.*Please see the back of the brochure.
ugged ightweight esign
Lighter than existing models to make transport easy.
Advanced measurement needs met with advanced functions.
●Expansion of Cross-Flow Modulation type detector ●Capable of measuring methane (CH4) ●Shorter warm-up time ●Timer function ●Ethernet compatible ●Capable of remote operation
Cross-Flow Modulation NDIR detector
■Equipped with an SD memory card slot to enable data to be saved immediately
SD memory card slot accessed from the front of the instrument enables necessary data to be saved on the spot in the universal CSV format.
■Screen capture function enables data to be saved immediately as a bitmap image onto the SD memory card.
No paper or pen required - simply touch the SCREEN CAPTURE icon and a screen shot is stored in memory.
■On screen guidance function allows you to confirm review operating procedures instantly
The simple guidance function provides assistance when you forget how to perform an operation. You can review regular operational procedures or important points right on the screen.
Simple, intuitive operation makes on-site measurement easy. The PG-300 has a highly visible and easy to operate LCD color touch screen. Data is readily saved on an SD memory card for easy transfer to a PC. The unit is equipped with a screen capture function as a standard feature, enabling necessary data to be saved on the spot. There is also an intuitive on screen guidance function, when the operator’s manual is not at hand.
Operation is simple and intuitive, making it easy to perform measurements in the laboratory or the field.
■LCD touch screen improves ease of operation
All operations, including calibration, measurement and saving on-screen data, can be performed on the touch screen. The high visibility color display makes it easy to check the status.
■Easy real time analysis using the color trend graph
There is a convenient color trend graph function, enabling gas component trends as a function of time to be confirmed at a glance.
The SD card slot is located on the front of the unit for easy access.
Current date and time
Measurement gas flow rate
Current measurement range
Measurements for each gas component
Name of measuredcomponent
Status and messagedisplay area
POWER icon
MENU icon
SCREEN CAPTURE icon
SD CARD icon
WARNING icon (red)INFORMATION icon (yellow)
GUIDE icon
PAGE FORWARD / BACK icons
Note: Calibration requires separately purchased calibration gas and pressure regulator.When you press the GUIDE button... ... guidance appropriate for the currently displayed screen appears.
[Sample display screens] [Color trend graph] [Calibration screen]
●Unit status is clearly displayed on the LEDs on the front of the unit.
●Easy-to-operate unit yields precision analysis results. ●The touch screen on the front makes operation easy.
3 4
asy peration Analyzer Specifications
Standard Accessories Replacement parts
Analyzers except SO2 analyzer : 45 sec. or less(From sample inlet, response time setting of electrical system : 10 sec.)
SO2 analyzer : 180 sec. or less(From sample inlet, response time setting of electrical system : 10 sec.)
Moving average selectable (10 or 30 sec.)
±1.0% of Full scale / day (For SO2 analyzer only : ±2.0% of Full scale / day
CH4 : 0-2000/5000 ppm CO : 0-2000/5000 ppm CO2 : 0-5/10/20 vol% O2 : 0-5/10/25 vol%
±1.0% of Full scale
±1.0% of Full scale / day
Model
Components Measured
Analysis Principle
Ranges
Repeatability
Linearity
Drift
Response Time (T90)
Sample Gas Flow Rate
Display
Output
Warm-up Time
Data Saving
Ambient Temperature
Ambient Humidity
Power
Power Consumption
Dimensions
Weight
Sample Gas Conditions
Filter element
Signal cable
Power cord
Tube
Tube
Tube
Tube
Joint
Cover
SD memory card
• Zirconia O2 analyzer is available. Please contact Horiba for the details.
For reference line*
For analog output (2 m) with connector
2.5 m
φ6/φ4PTFE tube 0.12 m (for mist catcher short)
φ6/φ4PTFE tube 5 m (for sample)
φ9/φ5 Imron tube 5 m (for exhaust)
φ9/φ5 Imron tube 1 m (for drain discharge)
φ6 straight (for sample tube)
Dust cover (for storage)
512 MB
24
1
1
1
1
1
1
1
1
1
Specifications QuantityPart Name
Mist catcher
Scrubber
Air filter element
MC-025
For reference line*
For reference line*
NotesName
Pump
NOx converter catalyst
Zero gas purifier unit catalyst
Ozone generator
Deozonizer
CR2032 battery
Galvanic O2 cell
1 year
1 year
1 year
1 year
1 year
5 years
1 year
3 months
3 months
2 weeks
Replace when broken
For NOx analyzer*
*
For NOx analyzer*
For NOx analyzer*
For clock backup
Replace when broken*
Replace Every(general guideline)
Replace Every(general guideline)
Name
Replacement part intervals assume 8 hours of operation per day. Replacement interval may be more frequent depending on measurement gas conditions and use conditions.
* Differs depending on model
* Differs depending on model
Notes
[Consumable Items]
[Replacement Parts]
• Separate tubing and joint are required if a pretreatment unit is added. * Differs depending on model.
●SD memory card slot ●Color LCD touch screen ●Screen capture function ●On screen guidance ●Color trend graph
2 componentAnalyzer
3 componentAnalyzer
4 componentAnalyzer
5 componentAnalyzer
2 componentAnalyzer
4 componentAnalyzer
Type of Analyzers
PG-320 PG-325
NOx/O2
PG-330
CO/CO2/O2
PG-335
NOx/CO/O2
PG-337
NOx/SO2/O2
PG-340
NOx/CO/CO2/O2
PG-344
CH4/CO/CO2/O2
PG-350
NOx/SO2/CO/CO2/O2
PG-324
CH4/CO2CO/CO2
NOx: Cross-Flow Modulation Chemiluminescence Detection MethodSO2,CO,CH4 : Cross-Flow Modulation Non-Dispersive Infrared Absorption Method
CO2 : Non-Dispersive Infrared Absorption MethodO2 : Galvanic Method, Paramagnetic Method(only in EU area)
Approx. 0.5 L/min.
Measurement (3 or 4 digit display), range, flow rate, etc.
DC 4-20 mA (non-insulated) / Ethernet
With 30 min. warm-up, ±2.0% of Full scale / 2 hours
SD memory card
0-40°C
85% R.H. or less
AC 100 V - 120 V, 220 V - 240 V
Approx. 220 VA in a steady state
260 (W) x 510 (D) x 265 (H) mm (Without side guards) 300 (W) x 520 (D) x 265 (H) mm (With side guards)
Approx. 14 kg (Without side guards) Approx. 15 kg (With side guards)
Temperature : Less than 40°C, H2O Content : Standard or less at ambient temperature, Dust : 0.1 g/m3 or less, Pressure : ±0.98 kPa
NOx : 0-25/50/100/250/500/1000/2500 ppmSO2 : 0-200/500/1000/3000 ppm CO : 0-200/500/1000/2000/5000 ppmCO2 : 0-10/20/30 vol%O2 : 0-5/10/25 vol%
±0.5% of Full scale (NOx : ≧ 100 ppm range / CO : ≧ 1000 ppm range)±1.0% of Full scale (Except above)
±2.0% of Full scale
■Equipped with an SD memory card slot to enable data to be saved immediately
SD memory card slot accessed from the front of the instrument enables necessary data to be saved on the spot in the universal CSV format.
■Screen capture function enables data to be saved immediately as a bitmap image onto the SD memory card.
No paper or pen required - simply touch the SCREEN CAPTURE icon and a screen shot is stored in memory.
■On screen guidance function allows you to confirm review operating procedures instantly
The simple guidance function provides assistance when you forget how to perform an operation. You can review regular operational procedures or important points right on the screen.
Simple, intuitive operation makes on-site measurement easy. The PG-300 has a highly visible and easy to operate LCD color touch screen. Data is readily saved on an SD memory card for easy transfer to a PC. The unit is equipped with a screen capture function as a standard feature, enabling necessary data to be saved on the spot. There is also an intuitive on screen guidance function, when the operator’s manual is not at hand.
Operation is simple and intuitive, making it easy to perform measurements in the laboratory or the field.
■LCD touch screen improves ease of operation
All operations, including calibration, measurement and saving on-screen data, can be performed on the touch screen. The high visibility color display makes it easy to check the status.
■Easy real time analysis using the color trend graph
There is a convenient color trend graph function, enabling gas component trends as a function of time to be confirmed at a glance.
The SD card slot is located on the front of the unit for easy access.
Current date and time
Measurement gas flow rate
Current measurement range
Measurements for each gas component
Name of measuredcomponent
Status and messagedisplay area
POWER icon
MENU icon
SCREEN CAPTURE icon
SD CARD icon
WARNING icon (red)INFORMATION icon (yellow)
GUIDE icon
PAGE FORWARD / BACK icons
Note: Calibration requires separately purchased calibration gas and pressure regulator.When you press the GUIDE button... ... guidance appropriate for the currently displayed screen appears.
[Sample display screens] [Color trend graph] [Calibration screen]
●Unit status is clearly displayed on the LEDs on the front of the unit.
●Easy-to-operate unit yields precision analysis results. ●The touch screen on the front makes operation easy.
3 4
asy peration Analyzer Specifications
Standard Accessories Replacement parts
Analyzers except SO2 analyzer : 45 sec. or less(From sample inlet, response time setting of electrical system : 10 sec.)
SO2 analyzer : 180 sec. or less(From sample inlet, response time setting of electrical system : 10 sec.)
Moving average selectable (10 or 30 sec.)
±1.0% of Full scale / day (For SO2 analyzer only : ±2.0% of Full scale / day
CH4 : 0-2000/5000 ppm CO : 0-2000/5000 ppm CO2 : 0-5/10/20 vol% O2 : 0-5/10/25 vol%
±1.0% of Full scale
±1.0% of Full scale / day
Model
Components Measured
Analysis Principle
Ranges
Repeatability
Linearity
Drift
Response Time (T90)
Sample Gas Flow Rate
Display
Output
Warm-up Time
Data Saving
Ambient Temperature
Ambient Humidity
Power
Power Consumption
Dimensions
Weight
Sample Gas Conditions
Filter element
Signal cable
Power cord
Tube
Tube
Tube
Tube
Joint
Cover
SD memory card
• Zirconia O2 analyzer is available. Please contact Horiba for the details.
For reference line*
For analog output (2 m) with connector
2.5 m
φ6/φ4PTFE tube 0.12 m (for mist catcher short)
φ6/φ4PTFE tube 5 m (for sample)
φ9/φ5 Imron tube 5 m (for exhaust)
φ9/φ5 Imron tube 1 m (for drain discharge)
φ6 straight (for sample tube)
Dust cover (for storage)
512 MB
24
1
1
1
1
1
1
1
1
1
Specifications QuantityPart Name
Mist catcher
Scrubber
Air filter element
MC-025
For reference line*
For reference line*
NotesName
Pump
NOx converter catalyst
Zero gas purifier unit catalyst
Ozone generator
Deozonizer
CR2032 battery
Galvanic O2 cell
1 year
1 year
1 year
1 year
1 year
5 years
1 year
3 months
3 months
2 weeks
Replace when broken
For NOx analyzer*
*
For NOx analyzer*
For NOx analyzer*
For clock backup
Replace when broken*
Replace Every(general guideline)
Replace Every(general guideline)
Name
Replacement part intervals assume 8 hours of operation per day. Replacement interval may be more frequent depending on measurement gas conditions and use conditions.
* Differs depending on model
* Differs depending on model
Notes
[Consumable Items]
[Replacement Parts]
• Separate tubing and joint are required if a pretreatment unit is added. * Differs depending on model.
●SD memory card slot ●Color LCD touch screen ●Screen capture function ●On screen guidance ●Color trend graph
2 componentAnalyzer
3 componentAnalyzer
4 componentAnalyzer
5 componentAnalyzer
2 componentAnalyzer
4 componentAnalyzer
Type of Analyzers
PG-320 PG-325
NOx/O2
PG-330
CO/CO2/O2
PG-335
NOx/CO/O2
PG-337
NOx/SO2/O2
PG-340
NOx/CO/CO2/O2
PG-344
CH4/CO/CO2/O2
PG-350
NOx/SO2/CO/CO2/O2
PG-324
CH4/CO2CO/CO2
NOx: Cross-Flow Modulation Chemiluminescence Detection MethodSO2,CO,CH4 : Cross-Flow Modulation Non-Dispersive Infrared Absorption Method
CO2 : Non-Dispersive Infrared Absorption MethodO2 : Galvanic Method, Paramagnetic Method(only in EU area)
Approx. 0.5 L/min.
Measurement (3 or 4 digit display), range, flow rate, etc.
DC 4-20 mA (non-insulated) / Ethernet
With 30 min. warm-up, ±2.0% of Full scale / 2 hours
SD memory card
0-40°C
85% R.H. or less
AC 100 V - 120 V, 220 V - 240 V
Approx. 220 VA in a steady state
260 (W) x 510 (D) x 265 (H) mm (Without side guards) 300 (W) x 520 (D) x 265 (H) mm (With side guards)
Approx. 14 kg (Without side guards) Approx. 15 kg (With side guards)
Temperature : Less than 40°C, H2O Content : Standard or less at ambient temperature, Dust : 0.1 g/m3 or less, Pressure : ±0.98 kPa
NOx : 0-25/50/100/250/500/1000/2500 ppmSO2 : 0-200/500/1000/3000 ppm CO : 0-200/500/1000/2000/5000 ppmCO2 : 0-10/20/30 vol%O2 : 0-5/10/25 vol%
±0.5% of Full scale (NOx : ≧ 100 ppm range / CO : ≧ 1000 ppm range)±1.0% of Full scale (Except above)
±2.0% of Full scale
Additional optional components expand the PG-300 portable analyzer capabilities.ptions
5 6
Drain separator unit / Drain pot unit specifications
Primary side filter probe specifications
When the gas sample includes moisture ranging from ambient temperature saturation to 40 °C saturation, a Drain Separator and Drain Pot are installed at the stage before the analyzer unit.
When the gas sample includes moisture exceeding 40 °C saturation, or when conducting continuous measurement (for five days or less), a thermoelectric cooler is installed at the stage before the analyzer unit. The electronic cooler unit can also accommodate low-temperature SO2 measurements.
Drain separator unit Drain pot unit
Electronic cooler unit
Primary side filter probe
The Cl2 scrubber can be built into the electronic cooling unit as an option. It is used to prevent corrosion of the cells, tubes and other internal components when the gas analyzer is operated at waste incineration facilities or in other situations where the gas sample includes Cl2.
[Cl2 scrubber] (optional)
Either of two types may be selected depending on use.
40 vol% or less
0.1 g/m3 or less
±2.94 kPa
0 - 50 °C* 0 - 120 °C*
Sample conditions (at feed port)
Sample conditions (at feed port)
Flue probe Simple probe
Drain separator Drain pot
Probe length (standard) 10 cm 1 m
*At flange inletNote: Please contact Horiba if the analyzer will be used in environments in which the temperature exceeds 120 °C. Please contact Horiba in case of use under the environmental that the pressure condition is other than ±2.94 kPa.
Temperature
Moisture
Dust
Pressure
Model Simple probe SE3 (flue probe)
±0.98 kPa ±4.9 kPa
0 - 40 °C
Ambient temperature saturation ~ 40 °C saturation
0.1 g/m3 or less
Model DS-300 (drain separator) DP-200 (drain pot)
Sample Preprocessing Unit (optional) UsesPlease select the ideal combination according to your needs.
Temperature
Moisture
Dust
Pressure
SteelmakingBoiler gas turbines
Simple probe
PG-300 Electronic cooler Electronic cooler
PG-300 PG-300
PG-300 PG-300
Simple probe Simple probe
Flue probe Flue probe Flue probe
Simple probe Simple probe
Stack gas Waste incineration Combustion gas appliances
Electronic cooler Electronic cooler
PG-300 PG-300
Drain separator Drain pot
Drain separator Drain pot Drain separator Drain pot Drain pot
Drain separator Drain pot Drain separator Drain pot Drain separator Drain potDrain pot
Electronic cooler
PG-300Note: Consult Horiba regarding probe material and other matters relating to applications.* For measurements exceeding 5 days, please contact Horiba.
Ap
plications
Intermittent M
easurement (8 hours or less)
Continuous M
easurement (5 d
ays or less)*
Electronic Cooler Unit Specifications
Material of sections contacting gas
Sampling flow rate
Dehumidifying capacity
Ambient temperature
Ambient humidity
Power
External dimensions
Mass
Sample conditions
SUS, PVC, PTFE, FKM, PVDF
Approx. 2 ~ 3 L/min
15 °C
0 ~ 40 °C
85% or less
AC100 V~240 V
260(W) x 375(D) x 223 (H) mm
Approx. 10 kg
Temperature: Ambient temperature, Dust: 0.1 g/m3 or less, Moisture: 20 vol% or less, Pressure: ±4.9 kPa
PS-300Model
Additional optional components expand the PG-300 portable analyzer capabilities.ptions
5 6
Drain separator unit / Drain pot unit specifications
Primary side filter probe specifications
When the gas sample includes moisture ranging from ambient temperature saturation to 40 °C saturation, a Drain Separator and Drain Pot are installed at the stage before the analyzer unit.
When the gas sample includes moisture exceeding 40 °C saturation, or when conducting continuous measurement (for five days or less), a thermoelectric cooler is installed at the stage before the analyzer unit. The electronic cooler unit can also accommodate low-temperature SO2 measurements.
Drain separator unit Drain pot unit
Electronic cooler unit
Primary side filter probe
The Cl2 scrubber can be built into the electronic cooling unit as an option. It is used to prevent corrosion of the cells, tubes and other internal components when the gas analyzer is operated at waste incineration facilities or in other situations where the gas sample includes Cl2.
[Cl2 scrubber] (optional)
Either of two types may be selected depending on use.
40 vol% or less
0.1 g/m3 or less
±2.94 kPa
0 - 50 °C* 0 - 120 °C*
Sample conditions (at feed port)
Sample conditions (at feed port)
Flue probe Simple probe
Drain separator Drain pot
Probe length (standard) 10 cm 1 m
*At flange inletNote: Please contact Horiba if the analyzer will be used in environments in which the temperature exceeds 120 °C. Please contact Horiba in case of use under the environmental that the pressure condition is other than ±2.94 kPa.
Temperature
Moisture
Dust
Pressure
Model Simple probe SE3 (flue probe)
±0.98 kPa ±4.9 kPa
0 - 40 °C
Ambient temperature saturation ~ 40 °C saturation
0.1 g/m3 or less
Model DS-300 (drain separator) DP-200 (drain pot)
Sample Preprocessing Unit (optional) UsesPlease select the ideal combination according to your needs.
Temperature
Moisture
Dust
Pressure
SteelmakingBoiler gas turbines
Simple probe
PG-300 Electronic cooler Electronic cooler
PG-300 PG-300
PG-300 PG-300
Simple probe Simple probe
Flue probe Flue probe Flue probe
Simple probe Simple probe
Stack gas Waste incineration Combustion gas appliances
Electronic cooler Electronic cooler
PG-300 PG-300
Drain separator Drain pot
Drain separator Drain pot Drain separator Drain pot Drain pot
Drain separator Drain pot Drain separator Drain pot Drain separator Drain potDrain pot
Electronic cooler
PG-300Note: Consult Horiba regarding probe material and other matters relating to applications.* For measurements exceeding 5 days, please contact Horiba.
Ap
plications
Intermittent M
easurement (8 hours or less)
Continuous M
easurement (5 d
ays or less)*
Electronic Cooler Unit Specifications
Material of sections contacting gas
Sampling flow rate
Dehumidifying capacity
Ambient temperature
Ambient humidity
Power
External dimensions
Mass
Sample conditions
SUS, PVC, PTFE, FKM, PVDF
Approx. 2 ~ 3 L/min
15 °C
0 ~ 40 °C
85% or less
AC100 V~240 V
260(W) x 375(D) x 223 (H) mm
Approx. 10 kg
Temperature: Ambient temperature, Dust: 0.1 g/m3 or less, Moisture: 20 vol% or less, Pressure: ±4.9 kPa
PS-300Model
PORTABLE GAS ANALYZER
NOx SO2 CO CO2 O2 CH4
Portable Gas Analyzer
PG-300 Series
Laboratory-precise analyses, anywhere.
External Dimensions (mm)
PG-300 Series Analyzer Unit (Side guards excluded)
265 510
265
Bulletin:HRE-2879A Printed in Japan TH-G(SK)53
300 520
265
PG-300 Series Analyzer Unit (Side guards included)