PLV®·102bPORTABLE VOLUME VENTILATOR
,
£RESPIRONICS
PLV-102b
When you need a portable volume ventilator capable of ventilating a wide range of patients, you need thePLV-102b. The PLV-102b uses a proven microprocessor-controlled, piston-driven ventilator platform to delivervolume ventilation in a range of settings. Its features address the needs of many patients who require volumeventilation and portability or transport in the sub-acute care, nursing home or home setting.
FEATURES
Portable· The PLV-102b is well suited for transport and portable applications due to its small size and lightweight. The ventilator is capable of operating from any of three power sources: 120 volts AC, an internal 12 voltDC battery and an external 12 volt DC battery, all of which are automatically switched by the PLV-102b.
Controls • The PLV-102b provides three modes: Control and Control-Assist, each of which can be used wit~optional Sigh, and the SIMV mode. The sensitivity of the PLV-102b is adjustable from -6 to +18 cm H20, which allowsfor use with PEEP up to 20 cm H20. The sensitivity is automatically and continuously corrected for barometriJpressure changes. IAlarms and Displays· Alarms are provided for Low Pressure, High Pressure, Apnea, Inverse I:E ratio, IncreaseIn11l1rlllflJ rum IlIlU runur OUUI UU iOUUOOI TIIOI 0 UIU UIOU Oill UlrrUi Dill UIUlIlI UUUUU IU UUUIUI III lruUUIOUllUUlIllYLarge, backlit LeDs display Tidal Volume, Respiratory Rate, I:E Ratio and Inspiratory Flow Rate.
FRONT PANEL CONTROLS
Mode Selection:
Tidal Volume:
Sigh Volume:
Sigh Frequency:
Breath Rate:
Inspiratory Flow Rate:
Battery Test Switch:
Main Power Switch:
DC Circuit Breaker:
AC Circuit Breaker:
Airway Pressure Limit:
Sensitivity:
Low Pressure Alarm:
Manual Sigh:
30 Second Alarm Silence:
POWER SOURCES
CONTROLCONTROL + SIGHASSIST/CONTROLASSIST/CONTROL + SIGHSIMV
0.05 to 0.20 +/- 0.02 liters0.20 to 3.00 liters +/- 10%
150% of Set Tidal Volume
1 Sigh Every 100 Tidal Breaths
2 to 35 BPM +/- 0.536 to 40 BPM +/- 2
10 to 120 LPM +/- 10%
120 VAC 50/60 Hz
12 VDC Internal Battery
12 VDC External Battery
PHYSICAL DIMENSIONS
OverallDimensions: 9" H x 12.25" W x 12.' 5" D(22.9 cm x 31.1 cm x 1.1 cm)
28.2 lb. (12.8 kg)Weight:
ALARMS
Test Internal/External Battery
On, Off/Recharge
Push/Pull Actuation
Push/Pull Actuation
5 to 100 cm H20 +/- 5
+18 cm H20 +/- 1to at least -6 cm H20
2 to 40 cm H20 +/- 241 to 50 cm H20 +/- 5
Push to Activate One Breath
Push to Activate
Low Pressure:
High Pressure:
Apnea:
Inverse I:E Ratio:
Visual and Continuo s Audible
Intermittent Audible
Visual and Continuo s Audible
Visual Warning
Visual Warning
Visual and Continuo s Audible
Visual and Continuo s Audible
Continuous Audible
Increase Inspiratory Flow:
Low Internal Battery:
Low External Battery:
Reverse Battery Connection:
Power Failure: Continuous Audible
Unit Malfunction: Intermittent Audible
Continuous AudiblePower Source Change:
ORDERING INFORMATIONFRONT PANEL VISUAL INDICATORS
Manual Sigh:
Tidal Volume:
Rate: Patient Breaths Per Minute:
I:E Ratio:
Inspiratory Flow Rate:
Increase Inspiratory Flow Rate:
120V AC Power:
Internal Battery Source:
External Battery Source:
Patient Airway Pressure:
Assist/Spontaneous:
15-Second Delay (Low Pressure Alarm):
30-Second Alarm Silence:
Alarm Code Display (in Front Window):
Yellow Indicator Light
Digital Display
Digital Display
Digital Display
Digital Display
Red Indicator Light
Green Indicator Light
Amber Indicator Light
Yellow Indicator Light
-10 to +100 cm H20
Green Indicator Light
Green Indicator Light
Yellow Indicator/Push Button
Digital Display
Description
PLV-l02b Ventilator
Part Number
1003111
!&1RESPIRONICS@
Global Headquarters:
Pittsburgh, PennsylvaniaUSA· 412-131-2100Customer Service: 1-800-345-6443Respironics Europe: +33-(0) 1-55-60- 9-80
Respironics Asia Pacific: +852-234-3 2-18www.resplronlcs.com
PLV-102b ventilators are the subiscts of U.S. patents #4,493,614, #4,617,637. #4.58 967,