Physics of Radiation Therapy Lecture III: X-ray Production/Clinical Radiation Generators.

30
Physics of Radiation Therapy Lecture III: X-ray Production/Clinical Radiation Generators

Transcript of Physics of Radiation Therapy Lecture III: X-ray Production/Clinical Radiation Generators.

Page 1: Physics of Radiation Therapy Lecture III: X-ray Production/Clinical Radiation Generators.

Physics of Radiation Therapy

Lecture III: X-ray Production/Clinical Radiation Generators

Page 2: Physics of Radiation Therapy Lecture III: X-ray Production/Clinical Radiation Generators.

Production of X Rays• The X-Ray Tube• Components

– Glass tube – maintains vacuum necessary to minimize electron interactions outside of the target area

– Cathode – contains filament and focusing cup

– Anode – contains x-ray target Figure 3.1 from Faiz Khan

Page 3: Physics of Radiation Therapy Lecture III: X-ray Production/Clinical Radiation Generators.

The X-Ray Tube

• The Cathode– Tungsten filament (high melting point – 3370 ° C)– Thermionic emission – electron production as a consequence

of heating– Focusing cup – “directs” electrons to anode– Dual filaments (diagnostic tubes) – necessary to balance

small focal spots and larger tube currents

Page 4: Physics of Radiation Therapy Lecture III: X-ray Production/Clinical Radiation Generators.

The X-Ray Tube

• The Anode– Tungsten target

• High melting point• High Z (74) – preferred since bremsstrahlung production Z2

– Heat dissipation• Copper anode – heat conducted outside glass into oil / water / air• Rotating anode (diagnostic tubes) – larger dissipation area

– Anode hood – copper and tungsten shields intercept stray electrons and x rays

Page 5: Physics of Radiation Therapy Lecture III: X-ray Production/Clinical Radiation Generators.

The Anode

• Focal-spot size• Line focus principle

– apparent focal spot (a) is smaller then actual (A) due to target angle :

a = A sin

Apparent focal spots– 0.1 – 2.0 mm in diagnostic

tubes, – 5 – 7 mm in therapy tubes

Figure 3.2 from Faiz Khan

Page 6: Physics of Radiation Therapy Lecture III: X-ray Production/Clinical Radiation Generators.

The Anode• Heel effect

– Since the x-rays are produced at various depths in the target, they suffer varying amounts of attenuation in the target

– There is greater attenuation for x-rays coming from greater depths than those from near the surface of the target.

– Consequently, the intensity of the x-ray beam decreases from the cathode to the anode direction of the beam.

– This variation across the x-ray beam is called the heel effect.

Page 7: Physics of Radiation Therapy Lecture III: X-ray Production/Clinical Radiation Generators.

Basic X-Ray Circuit

• Consists of two parts:– High-voltage circuit –

provides x-ray tube accelerating potential

– Filament circuit – provides filament current

Figure 3.3 from Faiz Khan

Page 8: Physics of Radiation Therapy Lecture III: X-ray Production/Clinical Radiation Generators.

Basic X-Ray Circuit

• Voltages can be increased or decreased using “step-up” or “step-down” transformers– Wire coils wound around iron rings– Coils create a magnetic field within the ring– Current and voltages (V) on opposite sides (primary and

secondary) of the ring are proportional to the number of “turns” (N):

VP / VS = NP / NS

– Stepwise and continuous voltage adjustments can be made with autotransformers and rheostats

Page 9: Physics of Radiation Therapy Lecture III: X-ray Production/Clinical Radiation Generators.

Basic X-Ray Circuit

• Voltage Rectification – Alternating current is used to energize x-ray tubes – X-ray tubes are designed to operate at a polarity

where the anode is positive– Rectifiers permit current flow in only one direction– Rectifiers placed in series in HV circuit provide half-

wave rectification

Page 10: Physics of Radiation Therapy Lecture III: X-ray Production/Clinical Radiation Generators.

Basic X-Ray Circuit

• Rectifiers can be arranged in a “diamond” arrangement to provide full-wave rectification – conduction during both halves of alternating (AC) voltage Figure 3.4 from Faiz Khan

Page 11: Physics of Radiation Therapy Lecture III: X-ray Production/Clinical Radiation Generators.

X-Ray Production

• Bremsstrahlung (“braking” radiation)

• Electromagnetic radiation emitted when an electron losses energy as a consequence of coulomb interaction with the nucleus of an atom Figure 3.6 from Faiz Khan

Page 12: Physics of Radiation Therapy Lecture III: X-ray Production/Clinical Radiation Generators.

X-Ray Energy Spectrum

• The bremsstrahlung photon’s energy is equal to the difference between electron’s incident and final energies– This leads to an

“energy spectrum” (the Kramer’s spectrum):

IE = K Z (Em- E)

Page 13: Physics of Radiation Therapy Lecture III: X-ray Production/Clinical Radiation Generators.

X-Ray Energy Spectrum• Characteristic X Rays

– K x rays – most “intense” transition to K-shell (most commonly by transitions from adjacent (L) shell), K x rays – second most intense, etc.

– For tungsten: K ≈ 60 keV, K ≈ 70 keV Figure 3.8 from Faiz Khan

Page 14: Physics of Radiation Therapy Lecture III: X-ray Production/Clinical Radiation Generators.

X-Ray Angular Distribution

• Angular distribution of x rays – Angular

distribution becomes more “forward peaked” as the electron energy increases

Figure 3.7 from Faiz Khan

Page 15: Physics of Radiation Therapy Lecture III: X-ray Production/Clinical Radiation Generators.

X-Ray Spectrum

• X-Ray Spectrum – Composite of Kramer’s

spectrum and characteristic x rays

– Filtration reduces lower-energy component

– Rule of thumb is average energy ≈ 1/3 maximum energy

– Half Value Layer (HVL) is a common descriptor

Figure 3.9 from Faiz Khan

Page 16: Physics of Radiation Therapy Lecture III: X-ray Production/Clinical Radiation Generators.

Clinical Radiation GeneratorsTreatment Unit Kilovoltage Range Half-Value Layer Clinical Use

Grenz Ray » 20 kVp ≈ 0.04-0.09 mm Al Dermatological Use

Contact Therapy 40-50 kVp ≈ 1-2 mm Al Few mm

Superficial Therapy

50-150 kVp ≈ 1-8 mm Al Skin Lesions

Orthovoltage 150-500 kVp ≈ 1-4 mm Cu Deep Therapy

Supervoltage 500-1000 kVp ≈ 10 mm Pb Replaced by Co-60

Page 17: Physics of Radiation Therapy Lecture III: X-ray Production/Clinical Radiation Generators.

Clinical Radiation Generators

Page 18: Physics of Radiation Therapy Lecture III: X-ray Production/Clinical Radiation Generators.

The Linear Accelerator

Page 19: Physics of Radiation Therapy Lecture III: X-ray Production/Clinical Radiation Generators.

The Linear Accelerator Components

DC Power Section

Microwave Power Section

Accelerator Guide Treatment Head

Page 20: Physics of Radiation Therapy Lecture III: X-ray Production/Clinical Radiation Generators.

The Linear Accelerator

• DC Power Section– Produces properly-shaped pulses of DC power;

these pulses are shaped in the pulse-forming network of the modulator and delivered to the electron gun and microwave power section at the proper frequency through a high-voltage switching device (the thyratron).

Page 21: Physics of Radiation Therapy Lecture III: X-ray Production/Clinical Radiation Generators.

The Linear Accelerator

• Microwave Power Section – Provides microwave power amplification (utilizing either a magnetron or klystron) and transmits the amplified microwaves to the accelerator guide.

• Accelerator Guide – A cylindrical tube in which electrons, injected by the electron gun, are accelerated by the amplified microwaves. The accelerated electrons exit the waveguide and enter the treatment head.

Page 22: Physics of Radiation Therapy Lecture III: X-ray Production/Clinical Radiation Generators.

The Linear Accelerator

• Treatment Head– Contains the beam shaping, steering, and control

components of the linear accelerator. These components are: the bending magnet, x-ray target, electron scattering foils (most accelerators), x-ray flattening filter, dose monitoring chambers, and beam collimation system.

Page 23: Physics of Radiation Therapy Lecture III: X-ray Production/Clinical Radiation Generators.

Microwave Power

• Electrons gain energy through continued exposure to an increasing electric field

• The process is analogous to a “surfer riding a wave” – Microwave cavities create

that environment – they are machined to dimensions that resonate at microwave frequencies (S-Band accelerators – 3000 MHz)

Page 24: Physics of Radiation Therapy Lecture III: X-ray Production/Clinical Radiation Generators.

Microwave Power• In a fashion similar to that

by which electrons gain energy from microwaves, microwave power can be amplified by the deposition of electron kinetic energy if electrons (grouped in “buncher” cavities) arrive in “catcher” cavities at the cavities’ resonant frequencies

Figure 4.7 from Faiz Khan

Page 25: Physics of Radiation Therapy Lecture III: X-ray Production/Clinical Radiation Generators.

Electron Acceleration• Electrons can be

accelerated in microwave cavities that are arranged in a linear configuration such that the microwaves’ “phase velocity” is matched to the velocity of the electrons traveling through the guide – (Note that energy gain will

be proportional to waveguide length)

Figure 1-11 from Hendee

Page 26: Physics of Radiation Therapy Lecture III: X-ray Production/Clinical Radiation Generators.

The Treatment Head• Shielded Housing – lead

shielding reduces unwanted radiation

• Bending Magnet – provides electron energy selection – the magnetic field intensity B

is set such that electrons possessing the appropriate energy (momentum mev) are bent through the radius r that allows passage through the magnet’s exit port:

• mev B r

Page 27: Physics of Radiation Therapy Lecture III: X-ray Production/Clinical Radiation Generators.

The Treatment Head

Page 28: Physics of Radiation Therapy Lecture III: X-ray Production/Clinical Radiation Generators.

The Treatment Head

• X-Ray Target – transmission-type tungsten target in which electron produce bremsstrahlung radiation; inserted only during x-ray beam production, removed during electron-beam production

• Flattening Filter – (photon beams) metal filter placed in the x-ray beam to compensate for the “forward peaked” photon distribution and produce a “flat” beam

Page 29: Physics of Radiation Therapy Lecture III: X-ray Production/Clinical Radiation Generators.

The Treatment Head

• Scattering Foils – (electron beams) thin metallic foils inserted in the electron beam to spread the beam and obtain a uniform electron fluence

• Monitoring Chambers – transmission ionization chambers used to monitor dose rate, total integrated dose and beam symmetry.

• Collimation System – fixed and movable beam limiting devices, normally made of lead, used to shape and size the beam

Page 30: Physics of Radiation Therapy Lecture III: X-ray Production/Clinical Radiation Generators.

Cobalt-60 (Megavoltage unit)• Co-60

– Co-60 is most commonly used because of its superior specific activity per gram, its greater beam intensity per curie, and higher average photon energy.

– A typical Co-60 source is 1 to 2 cm in diameter, and possesses an activity of about 3000 to 5000 Curies.

• Co-60– The source activity is limited by

the physical size of the source.– The large size of the Co-60

source produces a significant geometric penumbra (will be discussed later on)

– Does not give appropriate dose profile for deep seated tumors (in thoracic area)

– Replaced by linear accelerators