Philips Technologies
![Page 2: Philips Technologies by Olu Albert](https://reader030.fdocuments.us/reader030/viewer/2022032700/55d2ff69bb61ebf0518b45e4/html5/thumbnails/2.jpg)
Philips Technologies
• Digital AutoTrak• Digital AutoTrak sensitive• AVAPS• AVAPS rate• AVAPS-AE
![Page 3: Philips Technologies by Olu Albert](https://reader030.fdocuments.us/reader030/viewer/2022032700/55d2ff69bb61ebf0518b45e4/html5/thumbnails/3.jpg)
Digital AutoTrak (DAT)
![Page 4: Philips Technologies by Olu Albert](https://reader030.fdocuments.us/reader030/viewer/2022032700/55d2ff69bb61ebf0518b45e4/html5/thumbnails/4.jpg)
Components of DAT
A. Recognizes and compensates for system leaks 1. Leak ToleranceB. Provides 2 methods of inspiratory Trigger 2. Volume Trigger 3. Shape TriggerC. Provides 4 methods of expiratory cycling 4. Spontaneous Expiratory Threshold (SET) 5. Shape Trigger/Cycle 6. Flow Reversal 7. Maximum IPAP time
![Page 5: Philips Technologies by Olu Albert](https://reader030.fdocuments.us/reader030/viewer/2022032700/55d2ff69bb61ebf0518b45e4/html5/thumbnails/5.jpg)
1. Leak Tolerance
21
1 = Increase in flow & stabilizes most part of the inspiratory phase
2 = Decrease in flow & stabilizes most part of the expiratory phase
= Original Baseline/Beginning of inspiration/End of expiratory cycle
= Change in leakage flow/New Baseline Flow
![Page 6: Philips Technologies by Olu Albert](https://reader030.fdocuments.us/reader030/viewer/2022032700/55d2ff69bb61ebf0518b45e4/html5/thumbnails/6.jpg)
Inspiratory Triggering
![Page 7: Philips Technologies by Olu Albert](https://reader030.fdocuments.us/reader030/viewer/2022032700/55d2ff69bb61ebf0518b45e4/html5/thumbnails/7.jpg)
2. Volume Trigger
![Page 8: Philips Technologies by Olu Albert](https://reader030.fdocuments.us/reader030/viewer/2022032700/55d2ff69bb61ebf0518b45e4/html5/thumbnails/8.jpg)
3. Shape Trigger
Patient Flow = Total Flow- Circuit Leak
Shape Signal (Appears as a shadow image of the patient actual flow)
![Page 9: Philips Technologies by Olu Albert](https://reader030.fdocuments.us/reader030/viewer/2022032700/55d2ff69bb61ebf0518b45e4/html5/thumbnails/9.jpg)
![Page 10: Philips Technologies by Olu Albert](https://reader030.fdocuments.us/reader030/viewer/2022032700/55d2ff69bb61ebf0518b45e4/html5/thumbnails/10.jpg)
The shape trigger functions as a “sensitivity threshold” for inspiration and expiration
![Page 11: Philips Technologies by Olu Albert](https://reader030.fdocuments.us/reader030/viewer/2022032700/55d2ff69bb61ebf0518b45e4/html5/thumbnails/11.jpg)
Expiratory Cycling
![Page 12: Philips Technologies by Olu Albert](https://reader030.fdocuments.us/reader030/viewer/2022032700/55d2ff69bb61ebf0518b45e4/html5/thumbnails/12.jpg)
4. Spontaneous Expiratory Threshold
Actual FlowThe slope of the SET is calculated at the beginning of the breath and signal rises in proportion to the inspiratory flow rate of each breath
![Page 13: Philips Technologies by Olu Albert](https://reader030.fdocuments.us/reader030/viewer/2022032700/55d2ff69bb61ebf0518b45e4/html5/thumbnails/13.jpg)
When the SET signal crosses the flow signal, EPAP is triggered
![Page 14: Philips Technologies by Olu Albert](https://reader030.fdocuments.us/reader030/viewer/2022032700/55d2ff69bb61ebf0518b45e4/html5/thumbnails/14.jpg)
5. Shape Trigger/Cycle
The shape signal also functions as a “sensitivity threshold” for Expiration, when patient flow crosses the shape signal , it cycles into EPAP
![Page 15: Philips Technologies by Olu Albert](https://reader030.fdocuments.us/reader030/viewer/2022032700/55d2ff69bb61ebf0518b45e4/html5/thumbnails/15.jpg)
6. Flow Reversal
During flow reversal, the unit in the process Of cycling into EPAP when the leak occurs
![Page 16: Philips Technologies by Olu Albert](https://reader030.fdocuments.us/reader030/viewer/2022032700/55d2ff69bb61ebf0518b45e4/html5/thumbnails/16.jpg)
Flow reversal does not allow trigger to IPAP without returning to EPAP. This keeps “breath stacking” from occurring
![Page 17: Philips Technologies by Olu Albert](https://reader030.fdocuments.us/reader030/viewer/2022032700/55d2ff69bb61ebf0518b45e4/html5/thumbnails/17.jpg)
7. Maximum IPAP Time
![Page 18: Philips Technologies by Olu Albert](https://reader030.fdocuments.us/reader030/viewer/2022032700/55d2ff69bb61ebf0518b45e4/html5/thumbnails/18.jpg)
![Page 19: Philips Technologies by Olu Albert](https://reader030.fdocuments.us/reader030/viewer/2022032700/55d2ff69bb61ebf0518b45e4/html5/thumbnails/19.jpg)
DAT Components Modes
1. Leak Tolerance All Modes
2. Volume Trigger All Modes (except T & CV)
3. Shape Trigger All Modes (except T and CV
4. SET All Modes (except T, CV, AC & SIMV)
5. Shape Cycle All Modes (except T and CV)
6. Flow Reversal All Modes
7. IPAP Time S, S/T, T, PC
![Page 20: Philips Technologies by Olu Albert](https://reader030.fdocuments.us/reader030/viewer/2022032700/55d2ff69bb61ebf0518b45e4/html5/thumbnails/20.jpg)
Bottom line!!!The DAT detects and compensates for unintentional leaks in the system and automatically adjusts trigger and cycle algorithms to maintain optimum performance in the presence of leaks
![Page 21: Philips Technologies by Olu Albert](https://reader030.fdocuments.us/reader030/viewer/2022032700/55d2ff69bb61ebf0518b45e4/html5/thumbnails/21.jpg)
DAT Sensitive
![Page 22: Philips Technologies by Olu Albert](https://reader030.fdocuments.us/reader030/viewer/2022032700/55d2ff69bb61ebf0518b45e4/html5/thumbnails/22.jpg)
Sensitive Auto-Trak is an enhancement to the DAT algorithm that improves patient and machine synchrony for patients with minimal respiratory effort. DAT refines the baseline trigger and cycle sensitivity thresholds.
![Page 23: Philips Technologies by Olu Albert](https://reader030.fdocuments.us/reader030/viewer/2022032700/55d2ff69bb61ebf0518b45e4/html5/thumbnails/23.jpg)
Hybrid Modes
1. AVAPS
2. AVAPS-AE
![Page 24: Philips Technologies by Olu Albert](https://reader030.fdocuments.us/reader030/viewer/2022032700/55d2ff69bb61ebf0518b45e4/html5/thumbnails/24.jpg)
What are hybrid modes?
Pressure Support Ventilation modes with targeted volume delivery
![Page 25: Philips Technologies by Olu Albert](https://reader030.fdocuments.us/reader030/viewer/2022032700/55d2ff69bb61ebf0518b45e4/html5/thumbnails/25.jpg)
What is AVAPS?
• Average Volume Assured Pressure Support
• Available in S, S/T, PC, and T modes (Not a standalone mode!!!)
• It maintains tidal volume (VT) equal to or greater than the target volume setting by automatically controlling the PS
![Page 26: Philips Technologies by Olu Albert](https://reader030.fdocuments.us/reader030/viewer/2022032700/55d2ff69bb61ebf0518b45e4/html5/thumbnails/26.jpg)
AVAPS rate
• Allows the adjustment of the maximum rate at which PS automatically changes to achieve the target volume
![Page 27: Philips Technologies by Olu Albert](https://reader030.fdocuments.us/reader030/viewer/2022032700/55d2ff69bb61ebf0518b45e4/html5/thumbnails/27.jpg)
AVAPS Speed per Device
Fixed 2.5 cmH2O / min
0.5 to 5 cmH2O / min
1 to 5cmH2O / min
![Page 28: Philips Technologies by Olu Albert](https://reader030.fdocuments.us/reader030/viewer/2022032700/55d2ff69bb61ebf0518b45e4/html5/thumbnails/28.jpg)
How does AVAPS Work? AVAPS automatically adapts pressure support to patient needs to guarantee an average tidal volume.
Target Vt
EPAP
IPAP
± 1 to 3 cmH2O over 1 min
IPAP min
IPAP max
![Page 29: Philips Technologies by Olu Albert](https://reader030.fdocuments.us/reader030/viewer/2022032700/55d2ff69bb61ebf0518b45e4/html5/thumbnails/29.jpg)
Main Settings:
• Can be accessed in S, S/T, PC and T modes• AVAPS (set based on IBW 6-8ml/kg)• IPAP max• IPAP min• EPAP• RR• Ti
![Page 30: Philips Technologies by Olu Albert](https://reader030.fdocuments.us/reader030/viewer/2022032700/55d2ff69bb61ebf0518b45e4/html5/thumbnails/30.jpg)
AVAPS-AE
![Page 31: Philips Technologies by Olu Albert](https://reader030.fdocuments.us/reader030/viewer/2022032700/55d2ff69bb61ebf0518b45e4/html5/thumbnails/31.jpg)
What is AVAPS-AE?
AVAPS-AE is a auto-titration mode of noninvasive ventilation designed to better treat respiratory insufficiency patients (OHS, COPD and NMD) in the hospital and homecare environments
![Page 32: Philips Technologies by Olu Albert](https://reader030.fdocuments.us/reader030/viewer/2022032700/55d2ff69bb61ebf0518b45e4/html5/thumbnails/32.jpg)
Components of AVAPS-AE
• AVAPS
• Automatic EPAP (AE)
• Auto RR
![Page 33: Philips Technologies by Olu Albert](https://reader030.fdocuments.us/reader030/viewer/2022032700/55d2ff69bb61ebf0518b45e4/html5/thumbnails/33.jpg)
33
AVAPS AE settings
![Page 34: Philips Technologies by Olu Albert](https://reader030.fdocuments.us/reader030/viewer/2022032700/55d2ff69bb61ebf0518b45e4/html5/thumbnails/34.jpg)
EPAP
Target Vt
IPAP PS min
PS max
AVAPS-AEMaintaining Tidal volume and Airway patency
EPAP min
EPAP max
Resistance
34
![Page 35: Philips Technologies by Olu Albert](https://reader030.fdocuments.us/reader030/viewer/2022032700/55d2ff69bb61ebf0518b45e4/html5/thumbnails/35.jpg)
Theory of Operation
Upper Airway Resistance
Device Pressure
P search
crit
Ptherapy
P searchopt
Illustration courtesy of Krames Medical Illustration.
AVAPS-AE: Auto EPAP proactive analysis
35
![Page 36: Philips Technologies by Olu Albert](https://reader030.fdocuments.us/reader030/viewer/2022032700/55d2ff69bb61ebf0518b45e4/html5/thumbnails/36.jpg)
Popt – Optimal Pressure Search
(High Pressure Search)
EPAP
P opt
P opt
P therapy
P therapy
Critical Pressure Searches
(Low Pressure Search)
P crit
P crit
P therapy
EPAP
P crit
P therapy
Popt
Pcrit
Auto EPAP Proactive Analysis
Pther
Popt
Pres
sure
Resistance
Pcrit
36
P opt
![Page 37: Philips Technologies by Olu Albert](https://reader030.fdocuments.us/reader030/viewer/2022032700/55d2ff69bb61ebf0518b45e4/html5/thumbnails/37.jpg)
SET BRUsed for the rest of the night
AVAPS-AEAuto Backup rate
121
2
3
4
567
8
9
10
11
1st hour
50 breaths Sliding window
BR continually Updates
1h
BR set
15
30
45
560
BR=Actual Rate minus 2
37
![Page 38: Philips Technologies by Olu Albert](https://reader030.fdocuments.us/reader030/viewer/2022032700/55d2ff69bb61ebf0518b45e4/html5/thumbnails/38.jpg)
Auto backup rateM
ovin
g br
eath
rate
Minimum autobackup rate
Spontaneous breaths
Time 50 breaths
10
Calculated spontaneous rateTarget auto
backup rate
Targeted auto backup rate is 2 bpm below avg. spontaneous
rate
38
Bpm
![Page 39: Philips Technologies by Olu Albert](https://reader030.fdocuments.us/reader030/viewer/2022032700/55d2ff69bb61ebf0518b45e4/html5/thumbnails/39.jpg)
Auto backup rate
Spontaneous BreathsTimed Breaths
Target auto backup rate
10 Auto backup rate buffer zone
The auto backup rate is buffered by spontaneous triggers to promote synchrony with
the device
• The buffer is reduced or eliminated by timed triggers.
• The patient will be guided back to the targeted backup rate
39
Bpm
M
ovin
g br
eath
rate
Time (Night time session)
![Page 40: Philips Technologies by Olu Albert](https://reader030.fdocuments.us/reader030/viewer/2022032700/55d2ff69bb61ebf0518b45e4/html5/thumbnails/40.jpg)
Bottom line: AVAPS-AE is an “auto-titrating” BIPAP mode that will respond to:1) Hypoventilation (AVAPS)2) Airway Closure (AE)3) Inconsistent Respiration and /or Apnea (Auto RR)