Phase Optimization: A New Approach to HPLC Method Development · HPLC column. A real benefit of...

4
Introduction By some estimates, there are close to 800 different HPLC stationary phases to choose from when beginning an HPLC method development project. Choosing a column usually comes down to an educated guess, and whether it will successfully separate your analytes is usually not obvious until many hours have been spent evaluating different mobile phase conditions. If the column fails, another is chosen and the process is repeated until the right combination of stationary phase and mobile phase is discovered. The reason there are so many stationary phases to choose from is quite simple; selectivity. It is widely accepted that stationary phase selectivity is the most important factor in determining success or failure of a separation. The problem with this classical method development approach is that of the 800 phases available, many offer similar selectivi- ties. A base deactivated C18 from one manufacturer is likely to be quite similar to that from another manufacturer. If your sample separates on the first C18 that you try, your problem is solved. If not, then you begin the arduous task of screening other columns until you find one that works. In Phase Optimized Liquid Chromatography (POPLC ® ), a technique developed by Bischoff Chromatography, the stationary phase is first optimized for a given sample. Once you are assured of having the optimum stationary phase, it is a much simpler task to adjust the mobile phase for the desired speed and resolution. The POPLC Approach to Method Development In POPLC, rather than making a rough estimate of the appropriate column and then optimizing the mobile phase, one does the reverse. A rough choice of mobile phase is made, and then the sample is run isocratically on short columns of three or more different bonded phases. (Figure 1) The mobile phase chosen is based on previous experience with the sample, or determined by running a scouting gradient and choosing an organic composition near the mid-point of the chromatogram. Retention times for all components on each stationary phase are input to the POPLC Optimizer software. The software is based on Phase Optimization: A New Approach to HPLC Method Development the simple principle that retention times are additive. In other words, the retention time of an analyte on a column made up of segments of different stationary phases will be the sum of its retention times on the individual stationary phases. Armed with these data, the software can determine the optimum column, consisting of various length segments of the different stationary phases used. Based on the requirements specified by the investigator, the method can be optimized for speed, resolution or both. The user can specify maximum column length, desired analysis time and required resolution. Once the optimum column has been indentified, the various segments are assembled using the POPLink coupling system (Figure 2). Mobile phase optimization is then carried out either manually or computer aided by using other programs. The POPLink system is ideal for quickly combining various stationary phase segments. Up to ten segments can be linked with no detectable loss in efficiency. Once a column combination has been finalized, the segments can be linked inside a solid stainless steel tube similar in appearance to a conventional HPLC column. A real benefit of this design is that as the column wears, there is no need to replace the entire column, only the first segment, as the inlet end is where column failure normally occurs. FIGURE 1 The POPLC approach to method development does not “force fit” a separation to a randomly selected column. Instead, as shown above, the Optimizer Software creates a column customized to the sample. It does so by combining retention data from up to five different, but complementary, stationary phases. P O P LC ® Phase O P timized Liquid Chromatography

Transcript of Phase Optimization: A New Approach to HPLC Method Development · HPLC column. A real benefit of...

Page 1: Phase Optimization: A New Approach to HPLC Method Development · HPLC column. A real benefit of this design is that as the column wears, there is no need to replace the entire column,

IntroductionBy some estimates, there are close to800 different HPLC stationary phases tochoose from when beginning an HPLCmethod development project. Choosinga column usually comes down to aneducated guess, and whether it willsuccessfully separate your analytes isusually not obvious until many hourshave been spent evaluating differentmobile phase conditions. If the columnfails, another is chosen and the processis repeated until the right combinationof stationary phase and mobile phase is discovered.

The reason there are so many stationaryphases to choose from is quite simple;selectivity. It is widely accepted thatstationary phase selectivity is the mostimportant factor in determining successor failure of a separation. The problemwith this classical method developmentapproach is that of the 800 phasesavailable, many offer similar selectivi-ties. A base deactivated C18 from onemanufacturer is likely to be quite similar to that from another manufacturer. If your sample separates on the firstC18 that you try, your problem is solved. If not, then youbegin the arduous task of screening other columns until you find one that works.

In Phase Optimized Liquid Chromatography (POPLC®), a technique developed by Bischoff Chromatography, thestationary phase is first optimized for a given sample.Once you are assured of having the optimum stationaryphase, it is a much simpler task to adjust the mobile phasefor the desired speed and resolution.

The POPLC Approach to Method DevelopmentIn POPLC, rather than making a rough estimate of theappropriate column and then optimizing the mobile phase,one does the reverse. A rough choice of mobile phase ismade, and then the sample is run isocratically on shortcolumns of three or more different bonded phases. (Figure 1) The mobile phase chosen is based on previousexperience with the sample, or determined by running ascouting gradient and choosing an organic compositionnear the mid-point of the chromatogram. Retention timesfor all components on each stationary phase are input tothe POPLC Optimizer software. The software is based on

Phase Optimization:A New Approach to HPLC Method Development

the simple principle that retention times are additive. Inother words, the retention time of an analyte on a columnmade up of segments of different stationary phases will bethe sum of its retention times on the individual stationaryphases. Armed with these data, the software can determine the optimum column, consisting of various length segments of the different stationary phases used.Based on the requirements specified by the investigator, the method can be optimized for speed, resolution or both. The user can specify maximum column length, desiredanalysis time and required resolution. Once the optimumcolumn has been indentified, the various segments areassembled using the POPLink coupling system (Figure 2).Mobile phase optimization is then carried out either manually or computer aided by using other programs.

The POPLink system is ideal forquickly combining various stationary phase segments.Up to ten segments can be linked with no detectable loss in efficiency. Once a column combination has beenfinalized, the segments can be linked inside a solid stainless steel tube similar in appearance to a conventionalHPLC column. A real benefit of this design is that as thecolumn wears, there is no need to replace the entire column, only the first segment, as the inlet end is where column failure normally occurs.

FIGURE 1

The POPLC approach to method developmentdoes not “force fit” a separation to a randomlyselected column. Instead, as shown above, theOptimizer Software creates a column customizedto the sample. It does so by combining retentiondata from up to five different, but complementary,stationary phases.

POPLC ®

Phase OPtimized Liquid Chromatography

Page 2: Phase Optimization: A New Approach to HPLC Method Development · HPLC column. A real benefit of this design is that as the column wears, there is no need to replace the entire column,

NHCH2CH3

NH

O

N

NN

NH N

H

S

N

NN

NH

NH

Cl

N

NN

NH

NH

S

N

NN

NH

NH

S

N

NN

NH

NH

S

N

NN

NH N

NN

Cl

NHCH2CH3

CH3CHNH N

NN

Cl

NHCH2CH3

CH3

(1) Prometon (2) Simetryn (3) Ametryn (4) Prometryn

(5) Terbutryn (6) Simazin (7) Atrazine (8) Propazine

FIGURE 5Structures of Triazine Pesticides

The stationary phases included in the POPLC system are as follows:

ProntoSIL C18 EPS-2: This is a C18 phase with an embedded amide group. As such, it offers additional polarselectivity over a conventional C18 column. In general, itwill show greater retention for acids, and slightly less retention for bases. See Figure 3.

ProntoSIL C18 SH-2: This is a classic C18 stationaryphase with maximum carbon loading and effective endcapping. It separates purely by hydrophobic interaction. See Figure 3.

Complementary SelectivitiesRather than work with hundreds of random stationary phases, the POPLC approach uses up to five different stationary phase types which offer complementary, or evenorthogonal selectivity. The POPLC system includes columnsegments of varying lengths and stationary phases whichcan be coupled together to form a custom column optimized for your specific sample (Figure 2).

ProntoSIL Phenyl-2: This modern phenyl phase offers orthogonal selectivity to a C18 phase through π – π interactions.

ProntoSIL CN-2: This is an endcapped cyanopropylphase with very low hydrophobic interaction. Polar selectivity is excellent.

ProntoSIL C30: This non-endcapped C30 has very high shape selectivity in that it can separate planar andnon-planar analytes such as cis – trans isomers and fusedring aromatic compounds. See Figure 4.

Example – Separation of Triazine PesticidesThe structures of eight triazine pesticides are shown inFigure 5. POPLC was used to develop an isocraticmethod with baseline resolution of all pesticides. Figure 6shows the separation of a triazine pesticide mixture onfour different stationary phases. Notice that no individualstationary phase adequately separated all analytes.Retention data were then entered into the Optimizer Software along with criteria regarding the required separation. The user could specify a minimum resolution,a maximum run time, or the “best” separation possiblegiven these stationary phases. In this case, the computersimulated best separation is shown in Figure 7 along withthe actual chromatogram obtained.

BaP

PhP

h TB

N

BaP

PhP

h TB

N

0 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30 32 34 36

02468

101214161820

mV

02468

1012141618

mV

02468

101214

mV

BaP

PhP

h TB

N

Time (min)

0 1 2 3 4 5 6 7 8 9

Time (min)

0.02

0.04

0.06

0.08

0.10

0.12

0.14AU

1

2

3 45 6

0 1 2 3 4 5 6 7 8 9

Time (min)

0.020.040.060.080.100.120.140.16AU

1

2

3 45 6

10

10

This figure illustrates differences in selectivity for a classicallybonded C18 phase (top) and a polar embedded phase (bottom).Acidic compounds are retained longer while bases are retainedslightly less on the polar embedded phase.

FIGURE 3Comparison of SelectivityAcid, Basic and Neutral Solutes

Benzo(α)pyrene (BαP) is lower molecular weight than tetrabenzon-aphthalene (TBN) and, as predicted, elutes earlier on columns with low shape selectivity. The bottom chromatogram, however,illustrates the high degree of shape selectivity of ProntoSIL C30.Since BαP is a planar molecule, it has greater access to the stationary phase, and is retained longer than non-planar TBN.

FIGURE 4Influence of Shape Selectivity

FIGURE 2

Column segments of varying lengths can be coupled with thePOPLink system. Individual cartridges are color coded by stationary phase so that the column sequence can be viewedfrom the outside.

Classical Bonded C18 Phase

ProntoSIL SparselyBonded C18

ProntoSIL C18 SH

ProntoSIL 30

ProntoSIL C18-EPS

Supertrap
Page is trapped with Supertrap 3.0.32 Copyright 2002 Heidelberger Druckmaschinen AG To view traps, delete traps or to get detailed trapping information, please donwload free Supertrap Viewer Plug-In from: http://www.heidelberg.com Settings: Width: 0.088 mm = 0.250 pt Printorder: Cyan / Magenta / Yellow / Black / Step Limit: 5.0% Common Density Limit: 4.00 Centerline Trap Limit: 100% Trap Color Scaling: 100.0% Image to Object Trapping: yes Image to Image Trapping: no Black Width Scaling: 100.0% Black Color Limit: 95.0% Overprint Black Text: 8.0pt Overprint Black Strokes: no Overprint Black Graphics: no
Page 3: Phase Optimization: A New Approach to HPLC Method Development · HPLC column. A real benefit of this design is that as the column wears, there is no need to replace the entire column,

SummarySationary phase selectivity is highly important in HPLCmethod development yet conventional method developmentprocedures randomly select a column with the hope thatmethodical mobile phase changes will provide adequateresolution of the sample. In phase optimized liquid chromatography (POPLC), column selectivity is optimizedfor a given sample by using a combination of stationaryphases with complementary or orthogonal selectivity. Thisprocess requires a minimum number of experiments and is aided by a very powerful software package. Once theappropriate type and number of stationary phases is determined, these are combined through a unique connection system called POPLink. Methods thus createdare isocratic, with the user selecting conditions optimal for speed, resolution or both. By having a column with optimum selectivity for a given sample, the operator cannow, if desired, further refine the separation by optimizingthe mobile phase either manually or through a computer-aided method development program.

10 20 30 40Time (min)

12,3

4 5,6

7 8

0 50

10 20 30 40

Time (min)

0 50

12,3

45 7,86

7 8

1

2

4,5

3

6

10 20 30 40Time (min)

0 50

1 2

7 8

12

3,4,5

6

Time (min)

5 10 15 20 25 30 35 40 45 50

12

3 45

6 7

8

0

1

23

45 6 7

8

2 4 6 8 10 12 14 16 18 20 22 24 26 28 30 32 34 36 38 40 42 44 46 48 500

Time (min)

Time (min)

POPLC® Kits

Description Part Number

POPLC Basic Kit 150-3 MSCQ-1503-3 POPLC Basic Kit 250-3 MSCQ-2503-3POPLC Basic Kit 150-5 MSCQ-1503-5 POPLC Basic Kit 250-5 MSCQ-2503-5

Note: All Kits have the POPLC Optimizer Software

Differences in the KitsThe 250-3 and 150-3 Kits contain the following stationary phases:

• C18 SH-2• C18 EPS-2• Phenyl-2

The 250-5 and 150-5 Kits contain the following stationary phases:• C18 SH-2 • CN-2• C18 EPS-2 • C30• Phenyl-2

The 150 Kits (150-3 and 150-5) contain the following POPLink cartridges of each of the stationary phases in the Kit:

• 1 each 1 cm• 1 each 2 cm• 1 each 4 cm• 1 each 8 cm

The 250 Kits (250-3 and 250-5) contain the following POPLink cartridges of each of the stationary phases in the Kit:

• 1 each 1 cm • 1 each 6 cm• 1 each 2 cm • 1 each 8 cm• 2 each 4 cm

Note: All Kits have the necessary POPLink holder segments andinlet and outlet fittings for configuring columns. POPLink car-tridges are packed with 5 micron particles and have an internaldiameter of 3.0 mm.

FIGURE 6Separation of Triazine Pesticides

FIGURE 7Separation of Triazine Pesticides

The above chromatograms show the separation of a mixture ofeight triazine pesticides chromatographed under the same mobilephase conditions on four different stationary phases.

The lower chromatogram represents an actual pesticide sampleanalyzed on the column combination suggested by the optimizersoftware. The top trace is the predicted chromatogram.

ProntoSIL C18H

Software PredictedChromatogram

POPLC Predicted Stationary Phase:6cm C18-EPS + 18cm Phenyl + 6cm CN

Actual Chromatogram

ProntoSIL C18-EPS

ProntoSIL Phenyl

ProntoSIL CN

Page 4: Phase Optimization: A New Approach to HPLC Method Development · HPLC column. A real benefit of this design is that as the column wears, there is no need to replace the entire column,

POPLink Column Segment CartridgesDescription Part Number ProntoSIL C18 SH-2, 1cm, 3 each UB03E182PS050-3ProntoSIL C18 SH-2, 2cm, 3 each UC03E182PS050-3ProntoSIL C18 SH-2, 4cm, 2 each UD03E182PS050-2 ProntoSIL C18 SH-2, 6cm, 1 each UE03E182PS050-1ProntoSIL C18 SH-2, 8cm, 1 each UF03E182PS050-1ProntoSIL C18 EPS-2, 1cm, 3 each UB03E18BPS050-3ProntoSIL C18 EPS-2, 2cm, 3 each UC03E18BPS050-3 ProntoSIL C18 EPS-2, 4cm, 2 each UD03E18BPS050-2ProntoSIL C18 EPS-2, 6cm, 1 each UE03E18BPS050-1ProntoSIL C18 EPS-2, 8cm, 1 each UF03E18BPS050-1ProntoSIL Phenyl-2, 1cm, 3 each UB03E052PS050-3ProntoSIL Phenyl-2, 2cm, 3 each UC03E052PS050-3ProntoSIL Phenyl-2, 4cm, 2 each UD03E052PS050-2ProntoSIL Phenyl-2, 6cm, 1 each UE03E052PS050-1ProntoSIL Phenyl-2, 8cm, 1 each UF03E052PS050-1ProntoSIL CN-2, 1cm, 3 each UB03E202PS050-3ProntoSIL CN-2, 2cm, 3 each UC03E202PS050-3ProntoSIL, CN-2, 4cm, 2 each UD03E202PS050-2ProntoSIL CN-2, 6cm, 1 each UE03E202PS050-1ProntoSIL CN-2, 8cm, 1 each UF03E202PS050-1ProntoSIL C30, 1cm, 3 each UB03H300PS050-3ProntoSIL C30, 2cm, 3 each UC03H300PS050-3ProntoSIL C30, 4cm, 2 each UD03H300PS050-2ProntoSIL C30, 6cm, 1 each UE03H300PS050-1ProntoSIL C30, 8cm, 1 each UF03H300PS050-1

POPLink Holder Segments and Inlet and Outlet Port AssembliesDescription Part NumberPOPLink Holder Segments 1cm, 3 each MSCH10-08-03POPLink Holder Segments 2cm, 3 each MSCH20-08-03POPLink Holder Fitting Assy, Inlet Port MSCI-08POPLink Holder Fitting Assy, Outlet Port MSCO-08

® POPLC is a registered trademark of Bischoff Chromatography, Leonberg, Germany LC487EU

FIGURE 8Column Segments and POPLink Connecting Hardware

Cartridge Column Segment

Cartridge Holder Segment

Outlet Nut

Inlet Nut

Inlet AdapterOutletAdapter

To order or for more information:

BISCHOFF Analysentechnik u. geraete GmbHP.O. Box 1155Boeblinger Str. 23D-71229 LEONBERG, Germany

Phone: +49-(0)7152-6064-0FAX: +49-(0)7152-606435www.bischoff-chrom.de or [email protected]

Supertrap
Page is trapped with Supertrap 3.0.32 Copyright 2002 Heidelberger Druckmaschinen AG To view traps, delete traps or to get detailed trapping information, please donwload free Supertrap Viewer Plug-In from: http://www.heidelberg.com Settings: Width: 0.088 mm = 0.250 pt Printorder: Cyan / Magenta / Yellow / Black / Step Limit: 5.0% Common Density Limit: 4.00 Centerline Trap Limit: 100% Trap Color Scaling: 100.0% Image to Object Trapping: yes Image to Image Trapping: no Black Width Scaling: 100.0% Black Color Limit: 95.0% Overprint Black Text: 8.0pt Overprint Black Strokes: no Overprint Black Graphics: no