Perils of Replacing Liquid in Glass Thermometers

60
Perils of Replacing Liquid-in-Glass Thermometers Dean Ripple, Greg Strouse, Dawn Cross, and Karen Garrity NIST Thermometry Group Overview of thermometer types Discussion of application issues Detailed discussion of different thermometer types

Transcript of Perils of Replacing Liquid in Glass Thermometers

Page 1: Perils of Replacing Liquid in Glass Thermometers

Perils of Replacing Liquid-in-Glass Thermometers

Dean Ripple, Greg Strouse, Dawn Cross, and Karen Garrity

NIST Thermometry Group

Overview of thermometer types

Discussion of application issues

Detailed discussion of different thermometer types

Page 2: Perils of Replacing Liquid in Glass Thermometers

Overview of Different Thermometer Types

Page 3: Perils of Replacing Liquid in Glass Thermometers

Thermometer TypesThermometer Types

Standard Platinum Resistance Thermometers (SPRTs) (very accurate, but susceptible to shock)13.8 K to 962 °C

Industrial Platinum Resistance Thermometers (IPRTs)–196 °C to 850 °C

Thermistors–50 °C to 100 °C

Thermocouples–196 ° C to 2100 °C

Liquid-in-Glass Thermometers–150 °C to 400 °C

Digital Thermometers (PRT, thermistor, or thermocouple in disguise)–196 °C to 850 °C

Page 4: Perils of Replacing Liquid in Glass Thermometers

NIST Thermometry Group Calibration Staff

Dean Ripple – Group Leader– [email protected], 301 975 4801

Karen Garrity – Thermocouples– [email protected], 301 975 4818

Dennis Minor – SPRTs for T > 83 K, Fixed-points, and Bridges– [email protected], 301 975 4821

Gregory Strouse – ITS-90 for T > 83 K, SPRTs, Fixed-points, and Bridges– [email protected], 301 975 4803

Wes Tew – ITS-90 for T < 84 K, SPRTs– [email protected], 301 975 4811

Dawn Cross – Industrial – thermistors, resistance, liquid-in glass– [email protected], 301 975 4822

www.nist.gov/thermometry

Page 5: Perils of Replacing Liquid in Glass Thermometers

Considerations in Selecting a Thermometer

Accuracy: Uncertainties range from < 0.1 m°C to >1 °C

Cost of Thermometer: Range from $6 to $6000

Cost of Calibration: from $50 to $12,000

Temperature Range– thermistors 10 °C to 100 °C vs. W-Re thermocouples 0 °C to 2100 °C

Stability and Durability during use– chemical contamination– resistance to high temperatures, moisture, vibrations, and shock

Compatibility with measurement equipment– resistance thermometers, thermocouples easy to integrate to

electronics– liquid-in-glass, digital thermometers much easier for quick visual

inspectionCompatibility with object being measured

– sheath diameter, length chosen for good thermal equilibrium

Page 6: Perils of Replacing Liquid in Glass Thermometers

Thermometer Types:Calibration Ranges and Uncertainties

0.01

0.1

1

10

100

1000

10000

-300 0 300 600 900 1200 1500 1800Temperature / °C

Exp

ande

d U

ncer

tain

ty /

m°C

Pt-Rh alloy thermocouples

Base metal thermocouples

Standard Platinum Resistance Thermometers(SPRTs)

Liquid-in-glass (LiG)

Industrial Platinum Resistance Thermometers (IPRTs), Thermistors

Au/Pt thermocouples

Page 7: Perils of Replacing Liquid in Glass Thermometers

Comparative Thermometer Types:Calibration Methods, Uncertainties, and Costs

60

6 to 50

600

50 to 1000

100 to 1000

3000

3000

Nominal Cost$

1000<0.5Comparison–20 to 400LiG, Hg

10001Comparison0 to 1100Thermocouple,Base Metal

10000.3Comparison0 to 1100Thermocouple,Noble Metal

1200<0.005Comparison–10 to 100Thermistor

1750<0.01Comparison–196 to 500IPRT

1750<0.01Comparison–196 to 500SPRT

7000<0.001Fixed-Point–196 to 660SPRT

Calibration Cost, $

Calibration U (k=2), °C

Calibration Method

Temperature Range, °C

Thermometer Type

Page 8: Perils of Replacing Liquid in Glass Thermometers

General Issues with Replacing Liquid-in-Glass (LiG) Thermometers

Page 9: Perils of Replacing Liquid in Glass Thermometers

Tolerances vs. Calibration Uncertainties

0.001

0.01

0.1

1

10

-200 0 200 400 600Temperature / °C

Tole

ranc

e or

unc

erta

inty

/ °C

Thermistor, accuracy class 1

LiG, total immersion

LiG, partial immersion

PRT, grade A

TC, type K, special

Colored lines: ASTM tolerances (ASTM E1, E1137, E230, and E879).

Dashed lines: NIST calibration uncertainties (k=2)

Page 10: Perils of Replacing Liquid in Glass Thermometers

Tolerances vs. Calibration Uncertainties

Tolerance band: manufacturer’s guarantee that the instrument response will conform to a standard response function to within an error equal to the tolerance.

Calibrated thermometer: may or may not have a response close to the nominal response function for that thermometer type.

Response of individual unit is reported, along with uncertainties of the calibration process.

Individually calibrated thermometers cannot be considered directly interchangeable, unless the readouts or software are adjusted toincorporate the individual response function.

Page 11: Perils of Replacing Liquid in Glass Thermometers

Temperature Non-Uniformity

Total-Immersion Liquid-in-Glass Thermometer: Immersion depth varies with temperature

Partial-Immersion Liquid-in-Glass Thermometer: Immersion depth specified on thermometer

Platinum Resistance Thermometer: Placing thermometer at a fixed depth may introduce a bias, due to temperature variations in apparatus

LiG

PRT

2 °C

Page 12: Perils of Replacing Liquid in Glass Thermometers

Bias of Liquid-in-Glass Thermometers

1. For a partial immersion thermometer, if the stem temperature during use differs significantly from the ASTM E 1 stem temperature specified in Table 4 of E 1 and a correction is not applied, there will be an error (see ASTM E 77).

2. Total-immersion thermometer is used at a fixed, partial immersion, with no correction applied. Extreme care must be taken in selecting an alternative thermometer for these applications, because use of a different thermometer type, while reducing the measurement error, may cause changes in the bias of the standard.

3. If the thermometer is not in good thermal contact with the body being measured, there may be significant errors due to thermal conduction along the thermometer sheath. Temperature reading biased even though the precision is acceptable.

Page 13: Perils of Replacing Liquid in Glass Thermometers

Typical Measurement Uncertainty Budget: Thermocouples

Component Method of evaluation Calibration uncertainty or tolerance

Manufacturer or calibration laboratory, or ASTM E 230 tolerance

Thermocouple drift Results from literature, or in situ comparisons

Reference junction uncertainty

Manufacturer or independent evaluation

Readout uncertainty Manufacturer or independent evaluation

Readout drift Manufacturer or independent evaluation

Items in italics—examples of components generally not addressed with liquid-in-glass thermometers

Page 14: Perils of Replacing Liquid in Glass Thermometers

Examples of Subtle Device/Readout Failures

Long-term drift of readouts is expected, and addressed by periodic recalibration, but there are other risks:

Device loses calibration values in memory & reverts to default coefficients

Resistance bridge balances correctly, but circuitry to compensate for cable resistance is faulty

Incorrect entry of calibration coefficients into readout

Probes switched without updating coefficients

Consequence: Greater need for measurement cross-checks / measurement assurance / check standards

Routine checks of performance

Checks at ice point

Comparison of readings of different thermometers

Page 15: Perils of Replacing Liquid in Glass Thermometers

Specific Types of Thermometers: Liquid-in-Glass Thermometers

Page 16: Perils of Replacing Liquid in Glass Thermometers

Types of Liquid-in-Glass (LiG) Thermometers

Types of ASTM LiG thermometers– Over 120 types– Total Immersion– Partial Immersion

Liquids used– Mercury (Hg) - ASTM– Organic– Mercury Thallium– Proprietary (non-toxic)

Page 17: Perils of Replacing Liquid in Glass Thermometers
Page 18: Perils of Replacing Liquid in Glass Thermometers

Total Immersion Partial ImmersionTop of the liquid column about 1 cm

above the top of the bath fluid.Immersed to an indicated line or

depth

Page 19: Perils of Replacing Liquid in Glass Thermometers

Emergent Stem-TemperatureCorrection

Partial Immersion Thermometers above 150 °C

Faden Thermometers used to make emergent stem-temperature (EST) correction

Uncertainty of EST correction

– Faden temperature measurement

where k is the differential expansion coefficient of Hg

and n is the number of scale degrees equivalent to the length of emergent stem

)( spobs ttknEST

Page 20: Perils of Replacing Liquid in Glass Thermometers

Tolerances & Calibration Uncertainties

Typical ASTM tolerances: 1 or 2 scale divisions.

Typical calibration uncertainties: 0.2 to 0.5 of a scale division

Additional uncertainties in use:• Emergent stem temperature different from ASTM specification (partials only)

• Drift due to changes in bulb

• Drift due to capillary distress at high temperatures (> 150 °C)

Page 21: Perils of Replacing Liquid in Glass Thermometers

Recalibration of LiG Thermometers

• NIST SP 819 “A Procedure for the Effective Recalibration of Liquid-in-Glass Thermometers” recommends recalibration by adjusting old calibration values by the ice-point shift.

• Presently, we believe that LiG thermometers used above 150 °Cshould additionally be periodically recalibrated at the highest temperature of use.

• New NIST study in progress to revisit SP819 results

Page 22: Perils of Replacing Liquid in Glass Thermometers

Temporary Changes in Bulb VolumeAfter use at high temperatures, there is a temporary enlargement of the bulb

This type of temporary bulb expansion will cause a depression in the reading at the ice-point.

It takes 72 hours at room temperature for a thermometer to recover from exposure to high temperatures.

Easiest solution is to use thermometer with increasing temperature.

Alternative solution is to apply a scale correction (contact Dawn Cross)

Permanent Changes in Bulb Volume

High temperature use sometimes causes a permanent increase in the ice-point reading due to annealing effects

Page 23: Perils of Replacing Liquid in Glass Thermometers

Non-Mercury Liquid-in-Glass Thermometers

• Organic liquids generally have inferior performance to mercury, but are a reasonable alternative if uncertainty requirements aremodest (ASTM standard just begun)

• Beware of drainage of organic liquid down capillary wall on cooling

• “Next-generation” organic liquids under development (Existing ASTM standard E2251); good accuracy, but check for separation ofliquid column

• For all non-mercury LiG thermometers, capillary and bulb dimensions will be different, with different time response and immersion characteristics!!!

Page 24: Perils of Replacing Liquid in Glass Thermometers

Choice of a LiG Thermometer

Advantages of LiG thermometers– Relatively inexpensive

– When used at moderate temperatures (<150 °C), recalibration at the Ice MP suffices

– Damage to thermometer is usually visually apparent (!!!)

Disadvantages of LiG thermometers– Very difficult to automate

– Total immersion require adjustment of immersion with changing temperature/Partial immersion not too accurate

– Mercury is banned in some circumstances; prohibitively expensive to clean up in other instances

Page 25: Perils of Replacing Liquid in Glass Thermometers

Thermocouples

Page 26: Perils of Replacing Liquid in Glass Thermometers

Thermocouples...any two dissimilar conductors, joined at one end

Although total signal depends on temperature of two ends (1 & 4), thermocouples generate signal primarily in regions of strong thermal gradients. (Region 2-3)

The junction itself does not generate a voltage!!

T0

T1

Material A

Material BVoltmeter

Page 27: Perils of Replacing Liquid in Glass Thermometers

Advantages of Thermocouples

CheapWide temperature range (270 C to 2100 C) Small (down to 0.25 mm diameter) Easy to integrate into automated data systems

Disadvantages of Thermocouples

Small signals, limited temperature resolution (1 mK to 1 K) Thermocouple wires must extend from the measurement point to the

readout. Signal generated wherever wires pass through a thermalgradient.

At higher temperatures, thermocouples may undergo chemical and physical changes, leading to loss of calibration.

Recalibration of certain types of thermocouples or in certain applications is very difficult.

Page 28: Perils of Replacing Liquid in Glass Thermometers

Emf-Temperature Relationshipsfor the 8 Letter-Designated Thermocouple Types

BT

E JK

NR

S

Temperature / °C

Em

f / m

V

Notation: E = emf = Electromotive Force = Thermoelectric VoltageS = dE/dT = Seebeck Coefficient = Sensitivity

Base-metal types

Page 29: Perils of Replacing Liquid in Glass Thermometers

ASTM E230 Tolerances for Type K and Type N Thermocouples

-10

-5

0

5

10

-200 0 200 400 600 800 1000 1200 1400

Temperature, °C

Tole

ranc

e, °

C

Standard toleranceSpecial tolerance

Page 30: Perils of Replacing Liquid in Glass Thermometers

Soft-Insulated Thermocouples

• Choose polymer insulation based on upper temperature limit • High-temperature: use single lengths of alumina, not woven ceramics

Inner polymer insulationOuter polymer insulation

Weld or solder junction by any method

Mineral-Insulated, Metal-Sheathed (MIMS) Thermocouples

• MIMS thermocouples are available in small diameters (0.25 mm)• Sheath protects thermoelements from contamination

Page 31: Perils of Replacing Liquid in Glass Thermometers

Selecting a Thermocouple Type

type E: High Seebeck coefficient, homogeneous materials. Good for low temperatures.

type J: Cheap!type K: Fairly cheap high temperature thermocouple.type N: Good base metal thermocouple for high temperatures.type T: Homogeneous materials. Direct connection of differential

pairs to voltmeters.

Use type K, E, or T at room temp., type K up to 200 C, type N in the range 300 C to 600C, type N or K above 600 C

type R, S: Noble metal thermocouple for range 0 C to1400 C. type B: Noble metal thermocouple used from 800 C to 1700 C.

Use type R or S below 1300 C, type B above 1300 C.

Page 32: Perils of Replacing Liquid in Glass Thermometers

Emf MeasurementsVoltmeter

• voltmeter: accurate, but no direct temperature display

• Analog-to-digital PC card: beware high noise, high voltage offsets

• dedicated readout: equivalent to voltmeter + reference junction compensation + software.

Reference Junction Compensation• Thermocouple tables assume 0 °C reference junction temperature• Reference junction compensators mimic “missing” section of

thermocouple when reference junction temperature is not at 0 °C• Emf addition by hardware or software

Copper wires

DVMIce Point

TC

TC connector & extension wire

Page 33: Perils of Replacing Liquid in Glass Thermometers

Care and Feeding of Thermocouples

• Monitor drifts in base metal thermocouples by in situ tests• Protect from mechanical strain and kinks

• Protect from contamination using alumina or silica tubes, or usemineral-insulated-metal-sheathed thermocouples.

• For each temperature environment to be measured, a new thermocouple should be made, and it should always be used at thesame immersion.

• Obey the ASTM upper temperature limits for bare wire thermocouples.

Page 34: Perils of Replacing Liquid in Glass Thermometers

Limitations on Thermocouple Performance

• Chemical contamination (potentially very large)• Physical strain (moderate)• Preferential oxidation, volatilization (potentially large)• Hysteresis in structural phase changes (small to moderate,

worst for type E and K)• Extension wires, thermocouple connectors, feedthroughs

(potentially large, but avoidable)

Rule of thumb for fractional uncertainty of t, in °C

Base-Metal Noble-MetalVery best possible 3x10-4 10-4

(t < 100 °C)

Very best possible 10-3 10-4

(t > 100 °C)

Typical 10-2 10-3

Page 35: Perils of Replacing Liquid in Glass Thermometers

Recalibration of Used Thermocouples is Problematic!

If the furnace is isothermal, there will be NO difference between the original TC calibration and the recalibration.

Page 36: Perils of Replacing Liquid in Glass Thermometers

Platinum Resistance Thermometers (PRTs)

Page 37: Perils of Replacing Liquid in Glass Thermometers

What is an IPRT?2, 3, or 4-wire resistance element –nominally 100 @ 0 °C

– Wire wound– Thick film– Thin film

Resistance changes as a function of temperaturePositive temperature coefficient Nominal temperature range of use:

– –200 °C to 850 °CNominal resistance at 0 °C

– 100

Pictures courtesy of SDI and Minco

Page 38: Perils of Replacing Liquid in Glass Thermometers

IPRT Wiring Configurations

2-Wire: Rmeas = V/i = i(RL + RL + RS)/i = 2RL + RS

possible error of (2 RL/RS)(1/4103)°C

• calibration will change if RL changes

V

R L

R L

R S

i

4-Wire: No current passes through voltage-sensing leads. Consequently, there is no extra voltage drop along these leads, and measured voltage is RS.

I+ I-

V+ V-to DMM

to DMM

to current source to current source

• compensates for changes in RL

• preferred method

2-wire for non-demanding applications (±5 °C)3-wire for ±1 °C measurements, or ±5 °C over long cables4-wire for all high-accuracy measurements

Page 39: Perils of Replacing Liquid in Glass Thermometers

Nominal Resistance vs. Temperature Curve for an IPRT

0

100

200

300

400

Temperature, °C

Res

ista

nce,

IEC 751ASTM E1137

Page 40: Perils of Replacing Liquid in Glass Thermometers

ASTM E1137 “Off the Shelf” Tolerance and Uncertainty

0.0

0.5

1.0

1.5

2.0

2.5

Temperature, °C

Tole

ranc

e, °C

Class AClass B

Calibrate individual units for better uncertainty

Page 41: Perils of Replacing Liquid in Glass Thermometers

Types and Construction of IPRT elements

Wire-wound element– Alumina insulator: –200 °C to 850 °C– Glass insulator: –200 °C to 400 °C– Often loose or compacted MgO powder

for support– Sometimes hermetic via glass

compound– Best accuracy with loose powder

Thick and Thin Film element– Alumina substrate: –200 °C to 850 °C– Polyimide: –200 °C to 200 °C– Small with fast time response

Trade off between shock resistance and high accuracy

2-wire IPRT element

Pictures courtesy of SDI and Minco

Thick Film IPRT element

Page 42: Perils of Replacing Liquid in Glass Thermometers

Construction of IPRT ProbeMIMS IPRT Probe

– –200 °C to 650 °C– 2, 3, or 4-wire device– Element is either

wire-wound or film– Specify length,

fittings, element type

Support of winding– Not strain free– Thin film is bendable

(strain sensor)

Sealing– Element is usually

hermetically sealed– MIMS cable may not

be sealed

MIMS

Pictures courtesy of Minco

Page 43: Perils of Replacing Liquid in Glass Thermometers

Which IPRT Should I Use?

Probes vs. Bare Element: probes recommended unless element is permanently mounted

Film IPRTs: good time response, small size, shock resistant

Wire-wound IPRTs with constrained coils: low accuracy, but shock resistant

Wired wound IPRTs with slightly constrained coils: best accuracy (approaching ±0.01 °C over 400 °C span), sensitive to shock. Performance is highly variable with model.

Resistor configuration– 2-wire for non-demanding applications (±5 °C)– 3-wire for ±1 °C measurements, or ±5 °C over long cables– 4-wire for all high-accuracy measurements

Page 44: Perils of Replacing Liquid in Glass Thermometers

Considerations in Selecting IPRTs

ADVANTAGES– Wide temperature range– R vs. T is well characterized– Rugged construction– Cost is less than an SPRT– Available in different shapes and sized to meet most application

requirements– Can be used with a digital temperature read-out device

DISADVANTAGES– Resistance element is not strain free - hysteresis– Deterioration at elevated temperatures– 2- and 3- wire devices need lead-wire compensation– Non-hermetically sealed IPRTs will deteriorate in environments

with excessive moisture– Not as accurate as an SPRT– Not a defining standard of the ITS-90

Page 45: Perils of Replacing Liquid in Glass Thermometers

Sources of Errors When Using IPRTsStability

– Repeatability at 0 °C or 0.01 °CImmersion

– Check immersion characteristics of IPRT on insertion in thermal environment

Insulation Resistance– May degrade at t > 500 °C and high-moisture environments

Self Heating– Calibration and measurement current must be the same (nominally 1 mA)– Thermal contact with temperature of interest is important– Air probe or fluid probe will influence calibration method

Mechanical Shock and Vibration– Vibration or dropping the IPRT will cause the IPRT to drift or fail

Hysteresis– Measurement of temperature with increasing temperature may be

different than measurement of temperature with decreasing temperature

Page 46: Perils of Replacing Liquid in Glass Thermometers

Immersion Test of an IPRT

Immersion Profile of IPRT on Insertion– Different designs will exhibit different immersion characteristics

5 10 15 20

Immersion depth in oil bath, cm

T, °

C

IPRT 1IPRT 2

Page 47: Perils of Replacing Liquid in Glass Thermometers

Hysteresis Test of an IPRT

Causes of Hysteresis– Strain in platinum – most IPRTs are not strain free– Moisture produces shunting of sensor

-0.4

-0.2

0.0

0.2

0.4

0 100 200 300 400 500 600

Temperature, °C

T, °

CIncreasingDecreasing

Page 48: Perils of Replacing Liquid in Glass Thermometers

Recalibration Interval for an IPRT

Widely varies by designWidely varying performance based on use

– Thermal history– Mechanical shock

Behavior is not as predictable as an SPRTDrift at 0 °C or 0.01 °C does not always correlate well at other

temperatures

Recommendation:– Measurement at 0 °C or 0.01 °C as a minimum– Measurement at highest temperature of use is better

Page 49: Perils of Replacing Liquid in Glass Thermometers

Thermistors

Page 50: Perils of Replacing Liquid in Glass Thermometers

Thermistors (Thermal Resistor)

Semiconductors of ceramic material made by sintering mixtures of metallic oxides such as manganese, nickel, cobalt, copper, iron and uranium.

Temperature Range: –50 °C to 100 °C

Standard Forms:bead 300 to 100 probe bead in glass roddisc 0.5 cm to 1.3 cm thick, 5 k to 10 kwasher 2 cm diameterrod moderate power capacity, 1 k to 150 k

NTC: Negative Temperature Coefficient - The vast majority of commercial thermistors used as thermometers are in the NTC category.

Page 51: Perils of Replacing Liquid in Glass Thermometers

Anatomy of a Thermistor

Pictures courtesy of Thermometrics

Page 52: Perils of Replacing Liquid in Glass Thermometers

Temperature, °C

R(t

) / R

(25

°C),

k

IPRTThermistor

Sensitivity Comparison between Thermistors and IPRTs

Thermistor Sensitivity– Resistance changes by

more than a factor of 300 from –50 °C to 90 °C

– Higher sensitivity than an IPRT

– Non-linear• –50 °C: 5000 / °C• 90 °C: 8 / °C

Page 53: Perils of Replacing Liquid in Glass Thermometers

Interchangeability of a Thermistor

Thermistors are interchangeable to within the manufacturer’s tolerance band

Practical limits from –80 °C to 105 °C– ±0.05 °C for a 50 °C span– ±0.10 °C for a 75 °C span– ±0.20 °C for a 100 °C span

For a greater accuracy than the tolerance band, then a calibration must be performed– 0.001 °C for a 15 °C span– 0.003 °C for a 70 °C span

Page 54: Perils of Replacing Liquid in Glass Thermometers

Advantages and Disadvantages of Thermistors

ADVANTAGESEasy to miniaturizeRuggedFast response timeEasy to useDigital thermometer readoutInexpensiveHigh sensitivitySmall-size beads may be used for point-sensingStability: 4000 h at 100 °C

bead-in-glass 0.003 °C to 0.02 °Cdisc 0.01 °C to 0.02 °C

DISADVANTAGESSmall temperature rangeNon-linear deviceNeeds frequent checks on calibration when exposed to t > 100 °CInterchangeability is limited unless the thermistors are matchedSelf-heating may be large

Page 55: Perils of Replacing Liquid in Glass Thermometers

Cautions for Using Thermistors

Self-heating error may be large– Use low current and match calibration and use current– Calibrate air probe in air-filled glass tube

Epoxy coated thermistors are susceptible to degradation from moisture

Glass sealed thermistors are susceptible to degradation if the lead wires are bent apart at the base of the glass– Micro cracks can form in the glass

Heat sink the leads when making electrical connections with solder

Do not expose thermistors above 100 °C– Elevated temperatures will excessively age the thermistor and

cause drift

Page 56: Perils of Replacing Liquid in Glass Thermometers

Digital Thermometers

Page 57: Perils of Replacing Liquid in Glass Thermometers

Digital ThermometersWhat is a digital thermometer?

– An electronic measurement box that converts either resistance or emf of a thermometer to temperature

Pictures courtesy of Agilent, ASL, Brookstone, Hart Scientific, and Omega Engineering

Page 58: Perils of Replacing Liquid in Glass Thermometers

Digital Thermometers

Easy to use in conjunction with a resistance thermometer

Device displays temperature directly by using the calibration coefficients of the thermometer

Different temperature scales available

Uncertainty: 0.001 °C to 1 °C

Resolution: 0.0001 °C to 1 °C

Device may allow two thermometers to connected directly to unit for differential thermometry

Some have software that allow “real time” calibration

SMART thermometers have electronics in the thermometer head for an RS232 (or similar) port connection

Page 59: Perils of Replacing Liquid in Glass Thermometers

Caveats

Careful use requires careful reading: look at manufacturer’s specifications & know the probe type and limitations

“Digital Thermometer” only means that you can watch the numbers on a display

Calibration can be as a system or as readout + separate probeSystem calibration is cheaper and simpler

Probe and readout calibration allows easier repair if one element fails, allows identification of source of drift

Page 60: Perils of Replacing Liquid in Glass Thermometers

Conclusions & Roadmap

1. Identify the level of uncertainty needed2. Identify the temperature range3. Identify unique aspects of the test apparatus or method (e.g.,

inherent temperature non-uniformity)

4. Identify adequacy of presently specified mercury thermometer (anywhere from overkill to just adequate)

Make judgments on how tightly to prescribe the thermometerwhether to require calibration, measurement assurancewhat tests/round robins are needed to validate the

revised standard