PCR 5 Lab Setup

download PCR 5 Lab Setup

of 22

Transcript of PCR 5 Lab Setup

  • 8/6/2019 PCR 5 Lab Setup

    1/22

    PCR Laboratory Set-up.

    The arrangement or design of a PCR laboratory is of greatimportance in maintaining a high standard of performance.

    As a minimum you need three dedicated areas (minimumrequirement).

    The reagent preparation area (1).The specimen

    preparation area (2). The PCR and detection area (3).

    If more space is available, it may be used to good

    advantage.

  • 8/6/2019 PCR 5 Lab Setup

    2/22

    AmplificationlaboratoryArea

    AnalysislaboratoryArea

    LAB 3

    LAB 2

    Extraction

    Area

    LAB 1

    Master mix

    Area

    OUT

    IN

    Minimum layout requirements for a basic PCR laboratory

  • 8/6/2019 PCR 5 Lab Setup

    3/22

    Ideal physical arrangement for a PCR laboratory

    LAB 3RT-PCRLaboratoryArea

    LAB 4PCR/AmpliLaboratoryArea

    LAB 5

    AnalysisLaboratoryArea

    LAB 2ExtractionLaboratory

    Area

    LAB 1Master-mixLaboratory

    Area

    O

    ffices

    Offices

    Gene

    ralArea

    LAB 6Nested-PCRLaboratoryArea

  • 8/6/2019 PCR 5 Lab Setup

    4/22

    Ideal physical arrangement for a real-time PCR laboratory

    LAB 3Real-time PCRRobotic extraction

    LaboratoryArea

    LAB 4

    LaboratoryArea

    LAB 5Analysis

    LaboratoryArea

    LAB 2ExtractionLaboratory

    Area

    LAB 1Master-mixLaboratory

    Area

    Offices

    Office

    s

    Gene

    ralArea

    Store

    Real-time PCR

  • 8/6/2019 PCR 5 Lab Setup

    5/22

    Reagent Preparation - Area 1

    The first steps in the PCR procedure starts in Area 1.

    Here, master mix (amplification reagents) reagents and primers are

    prepared along with controls.

    Prepared master mix and controls are then added to the PCR reactiontubes.

    If you can divide this area into two spaces so as to keep the controlDNA away from the master mixes.

  • 8/6/2019 PCR 5 Lab Setup

    6/22

    Equipment required in Area 1

    If using one room, a dead-air box with an optional ultraviolet(UV) light or, if using two rooms a clean, dedicated

    area.

    Freezer/Fridge and ice-boxes. Dedicated micropipettes with plugged (aerosol-barrier) tips

    (25 L, 50 L, and 100 L) or

    Dedicated positive-displacement pipette and tips. A repeat

    pipettor (optional).

    Perkin-Elmer GeneAmpTM PCR System reagents (Taq Pol,

    buffer etc), MicroAmpTM consumables (tubes, caps,

    base, tray, and retainer).

    Microcentrifuge.

    Lab coat

    Gloves

  • 8/6/2019 PCR 5 Lab Setup

    7/22

    Important considerations applicable to Area 1

    With PCR diagnostics, always work in a one-way direction from Area

    1 to Area 2 to Area 3 in order to avoid carryover contamination from

    amplified products.

    Area 1 should be located as far away from Area 3 as possible.

    Area 1 should not be in close proximity to the centrifuge area (Area 2)

    to avoid possible aerosol contamination. Clean, dedicated area working surface must be decontaminated prior

    to use with 10% bleach solution (freshly made daily) followed by

    a 70% ethanol rinse or equivalent. If a UV box is used, for your own safety, you must turn offthe

    ultraviolet light in the box beforeplacing your arms and hands

    into the working area. Master mix and controls are to be prepared in this area.

  • 8/6/2019 PCR 5 Lab Setup

    8/22

    Important considerations applicable to Area 1 (cont.)

    Prepared master mix and controls will be added to PCR reaction

    tubes and capped in this area.

    Dedicated pipettes with plugged (aerosol-barrier) tips or

    positive-displacement tips, or a repeat pipettor are required in

    Area 1 and should not be transferred to either Area 2 orArea 3.

    Lab coats designated for wear in Areas 1 and 2 should never be

    worn into Area 3 to avoid carryover contamination ofamplified product from Area 3 back to Area 1 or Area 2:-

    two coats may be used, disposable coats may be worn over a

    cloth coat

    Gloves must be worn at all times for your safety as well as for

    control of contamination from one area to another.

    Gloves are to be changed at each of the three work areas.

    Gloves worn in Area 3 must never be worn in Area 1 or Area 2.

  • 8/6/2019 PCR 5 Lab Setup

    9/22

    DNA/RNA Extraction - Area 2Specimens are prepared in this area and are added to PCR tubes with

    master mix. Specimens are now considered amplification-ready.

    Equipment required in Area 2

    Biological cabinet, flow cabinet, dead area space.

    Standard clinical tabletop centrifuge and/or microcentrifuge (max RCF 16,000g).

    Dedicated micropipettes with plugged (aerosol-barrier) tips (25l, 50l, & 100 l) or

    Dedicated positive-displacement pipette and tips.

    Repeat pipettor (optional).

    Balance/Scale.

    Spectrophotometer.

    Vortex mixer

    Dry-heat temperature block. Sterile calibrated transfer pipettes

    All reagents and chemicals.

    All consumables.

    Lab coat (different coat to Area 1).Gloves (new pair).

  • 8/6/2019 PCR 5 Lab Setup

    10/22

    Important considerations applicable to Area 2

    Working surface in cabinet must be decontaminated prior to specimen preparation

    with 10% bleach solution (freshly made daily) followed by a 70% ethanol

    rinse or equivalent.

    Cabinet must be on 30 minutes before specimen preparation procedure.

    All pipettors, pipettes, bulbs, and other equipment used in specimen preparation

    should be kept in this cabinet at all times.

    Specimens must be stored separately from reagents-two separate refrigerators are

    required.

    Dry baths are preferred over water baths water baths can contaminate specimens

    through seepage into poorly stoppered tubes.

    A set of pipettes should be dedicated exclusively to be used for specimen preparation

    To avoid specimen-to-specimen contamination, plugged (aerosol-barrier) pipette tips

    or positive-displacement tips, must be used in Area 2. Lab coats designated for wear in Areas 2 should never be worn into Area 3 or 1 to

    avoid carryover contamination.

    Gloves must beworn at all times for your safety as well as for control of

    contamination from one area to another gloves are to be changed at each of thethree work areas.

  • 8/6/2019 PCR 5 Lab Setup

    11/22

    Amplification (PCR) and Detection - Area 3

    The final steps in the procedure takes place in Area 3.

    Here, amplification-ready specimens undergo PCR and the PCR

    products (amplicons) of this PCR are then detected.

    This area should also be divided into two areas if possible.

    Amplification area

    Analyses area.

  • 8/6/2019 PCR 5 Lab Setup

    12/22

    Equipment required in Area 3

    Perkin-Elmer GeneAmpTM PCR Systems 9600/9700/2400/2700 Thermal Cyclers or

    equivalent.

    Dedicated micropipettes with plugged (aerosol-barrier) tips (25 L and 100 L) or

    dedicated positive-displacement pipette and tips Multichannel pipettor, Repeat pipettor, Nonplugged pipette tips (all optional)

    Electrophoreses apparatus and power supplies.

    Microcentrifuge

    Disposable reagent reservoirs Incubator (37C 2C) with or without CO

    2

    Microwell plate reader, Microwell plate washer. Key for microwell strip removal.

    UV trans-illuminator

    Gel documentation system (computer hardware/software & printer) Lab coat (different from the one worn in Areas 1 and 2)

    Gloves (new pair)

    Electrophoresis apparatus and documentation system.

    Balance/Scale and microwave oven

  • 8/6/2019 PCR 5 Lab Setup

    13/22

    Important considerations applicable to Area 3

    Area 3 should be kept as far away as possible from Areas 1 and 2 to avoid aerosol

    contamination.

    Incubator temperature should be kept stable.

    Traffic in and out of incubator should be reduced to a minimum.

    Area 3 pipettes should never be used in Areas 1 or 2.

    Pipettes with plugged (aerosol-barrier) tips are used for pipetting denaturing solution

    into PCR tubes and denatured amplified product into microwell plates.

    Non-plugged tips may be used for all other reagent additions a pipette contaminated

    with this highly concentrated product could cause false-positive results

    plugged tips therefore are required to prevent this potential carryover

    contamination.

    As this is a one-way workflow, the lab coat worn in Area 3 must never be worn inAreas 1 and 2.

    Gloves must beworn at all times for your safety as well as for control of

    contamination from one area to another. Gloves are to be changed at each of

    the three work areas. Gloves worn in Area 3 must never be worn in Area 1 or Area 2.

  • 8/6/2019 PCR 5 Lab Setup

    14/22

    Establishment of a PCR assay

    The selection of a PCR assay must be based on factors including scientific and

    international acceptance, cost, available resources, nature of the intended use,

    sensitivity and specificity, number of tests to be done and availability of standard

    reagents.

    Development and optimization of the assay must then be performed and should

    include a series of experimental procedures and the evaluation of the data

    generated. Analysis should determine a fixed protocol for use, the nature and

    number of controls required and specifications required of reagents and equipment.

    The optimisation of the PCR should be followed by evaluation of the test. This willinclude a period of testing samples with known histories.

  • 8/6/2019 PCR 5 Lab Setup

    15/22

    Validation of the assay

    Validation establishes the performance standards of the assay (e.g.,

    sensitivity, specificity, accuracy, precision, positive/negative cut-off criteria etc.), using appropriate statistical methods.

    Validation can include a comparison with other methods, with

    reference standards, collaborative studies with other laboratories

    using the same test procedure, experimental challenge studies or

    reproduction of data from accepted standard methods or a

    reputable publication

  • 8/6/2019 PCR 5 Lab Setup

    16/22

    Specificity is the ability of the assay to exclusively detect the agent of

    interest.

    Sensitivity is described as the ability to precisely detect small quantities

    of the test agent

    Stage 1 validation

    Involves the development of the assay performing a feasibility study to

    determine whether the assay can detect a range of agents (e.g., virus

    concentrations, virus serotypes/genotypes), without background activity.

    This stage include all aspects of development and optimisation of the

    test, including the identification of the test, the determination of thetarget template(s), the sequence determination of the primers and the test

    conditions and criteria.

    Validation

  • 8/6/2019 PCR 5 Lab Setup

    17/22

    Stage 2 validation

    Involves the evaluation of the test against a panel of known positive

    and negative template samples to determine the analytical sensitivity

    and specificity.

    Stage 3 validation

    This involves the determination of diagnostic sensitivity and

    specificity, I.e. involving field samples in 2 parallel tests (classical vs

    PCR). .

    Diagnostic sensitivity (DS) and specificity (DP)

    DS determines the proportion of known infected reference animals testing positive[TP/(TP+FN)], while DP determines the proportion of uninfected reference animals

    that test negative [TN/(TN+FP)]. (TP = true positive; FN = false negative; TN =

    true negative; FP = false positive).

    DS - how many false negatives you get.DP - how many false positives you get

  • 8/6/2019 PCR 5 Lab Setup

    18/22

    Stage 4 validation

    Entails the continued monitoring of the validity of assay

    performance in the field by calculating the predictive value of

    positive or negative results based on estimates of pathogen

    prevalence in the target animal population. This can only be done

    satisfactorily if a diagnostic sensitivity and specificity data (stage

    3) are available.

    Stage 5 validation

    Involves the maintenance of validation criteria using internal

    quality controls. Frequent monitoring for repeatability and

    accuracy are needed. The OIE also recommends biannual ring-testing to determine reproducibility between laboratories, although

    annual testing is also described.

  • 8/6/2019 PCR 5 Lab Setup

    19/22

    Quality assurance programme or accreditation

    The intention of accreditation or the formal recognition of a quality assurance

    programme by an independent third party, is to ensure that GLP, QA (QC -

    IQC/EQC) is in place within the diagnostic laboratory. Specific standards are used

    (e.g. ISO 17025) in accordance with the requirements of the inspecting body (e.g.,

    the South African National Laboratory Accreditation Service).

    Requirements will include a quality policy, a quality manual, procedures for

    evaluating and reviewing new and exciting test methods and protocols, the use of

    blind quality assurance within and between laboratories, monitoring of equipment,

    ongoing training, an audit programme, regular review of the quality system and

    documentation and records of all quality activities and evaluations. Standard

    operating procedures (SOPs) will be required for not only the test methods used,

    but also for the calibration and maintenance of instrumentation and equipment

    used.

  • 8/6/2019 PCR 5 Lab Setup

    20/22

    Proficiency testing

    Proficiency testing is the means used to determine the capability

    of a laboratory to perform the assay and effectively detect theagent (internal proficiency testing).

    Such testing will also contribute to ensuring that within or

    between laboratories performing routine diagnostic services, aspecific assay is performed according to established international

    standards (external proficiency testing

  • 8/6/2019 PCR 5 Lab Setup

    21/22

    PCR Controls

    Positive control samples

    These should include:

    Specimen material with a high pathogen concentration.

    Specimen material with a low concentration (preferably the lowest limit

    routinely detected, or preferably the lowest amount of detectable target

    agent present in sample material by conventional means and by reported

    PCR data. Specimen material with expected pathogen concentration in an infected case (not

    essential)

    Extracted positive DNA/RNA can be used only if the above are not available.

    In the case of a proficiency testing process, specimen material spiked with a

    dilution series of virus (e.g., 10-1 to 10-10) should also be included. Virus titre

    would have been established by means of a golden standard assay.

  • 8/6/2019 PCR 5 Lab Setup

    22/22

    Negative control samples

    A negative control consisting of all reagents except extracted sample DNAor RNA - this should confirm absence of amplicon or agent itself or

    other cross reactive entities in reagents.

    A negative specimen control sample (allantoic fluid, buffer, tissue), if

    available, which will undergo nucleic acid extraction to confirmabsence of contamination in the extraction procedure.

    Extracted nucleic acid derived from pathogen-free sample material obtained

    previously (e.g., tissue or other medium normally submitted,

    including tissue, blood, buffer) - should confirm that no non-

    specific binding to virus-free specimen DNA/RNA has taken place

    (eukaryotic and prokaryotic genomic DNA or RNA present in

    sample). Not essential.

    In the case of RT-PCR, an additional negative control consisting of all

    reagents as well as the extracted sample nucleic acid, but without

    reverse transcriptase included, this should confirm absence of

    DNA or amplicon contaminants in both reagents andtest material.