Particle Size Analyzers
C060-E005A
Particle Size Analyzer Applications
Particle Size Distribution Is One of the Largest Factors Determing the Characteristics of Powders and Particles.Powders and particles are used in a wide variety of fields for a wide range of objectives and applications. In some cases, they
are used directly as pharmaceuticals, catalysts, additives, or binders while in other situations they are used as raw ingredients.
In either case, the particle size distribution can have a major effect on the characteristics desired for a given application or
objective, or on the performance and quality of a final product. Consequently, measuring the particle size distribution is
essential for stabilizing or improving the characteristics, performance, or quality of powders or particles.
0.20.0005
3000.01
30000.05
25000.017
3500.1
3500.25
nm μm mmnanometers
1 5 10 50 100 500 1 5 1 510 50 100 500
micrometers millimeters
0.001 0.005 0.01 0.05 0.1 0.5 1 5 1000 500010 50 100 500
Particle Size: μm
nm μm mmnanometers
1 5 10 50 100 500 1 5 1 510 50 100 500
micrometers millimeters
0.001 0.005 0.01 0.05 0.1 0.5 1 5 1000 500010 50 100 500
Particle Size: μm
Zinc Sulfide Phosphor
IG-1000 Plus
SALD-7101
SALD-3101
SALD-2300
SALD-301V
SALD-201V
Fullerenes
Silica Nanoparticles
Nano-Inks
Liposomes
Titanium Oxide
Talc
Pearl Pigments
Acetaminophens
Cornstarch
Wheat Flour (cake flour)
Wheat Flour (bread flour)
Kaolinite
Alumina
PVC Powder
Acrylic Polymer Beads
Vanadium Oxide
Clay Silt Sand
Cement
Carbon Black
Nanoparticles
Pharmaceuticals
Cosmetics
Food Products
Ceramics
Macromolecules
Catalysts
Electronic Materials
Soil and Civil Engineering Materials
Developed for single-digit nanoparticles
Developed for nanoparticles
Standard model of SALD series
Developed for coarse or dense particles
Low-cost model for fine particles
Low-cost model
Shimadzu Particle Size Analyzers Are Used in a Wide Variety of Fields, for a Wide Range of Purposes and Applications.
Guide to Selecting Particle Size Analyzers
Want to measure single-digit nanoparticles (with diameters in single-digit nanometer range).
However, not sure if samples contain aggregates or contaminants.
Screen with an SALD-7101,
then measure with an IG-1000 Plus
Want to screen nanoparticle samples for aggregates or contaminants. Want to measure
nanoparticles in a viscous medium.SALD-7101
Nano Particle Size Analyzer
Want to mainly measure particles in the micrometer to millimeter range. Want to measure particles with high specific gravity, such as metals. * Want to measure dried powders in the dry state.
SALD-3101Laser Diffraction Particle Size Analyzer
* Add a dry measurement unit.
Cost is important. Want to mainly measure particles in the micrometer range. Wet measurement of low specific-gravity particles is sufficient.
SALD-201V or SALD-301VLaser Diffraction Particle Size Analyzer
Want to mainly measure particles in the submicron to micron range.
* Want to measure dried powders in the dry state.
SALD-2300Laser Diffraction Particle Size Analyzer
* Add a dry measurement unit.
Want to measure single-digit nanoparticles (with diameters in single-digit nanometer range).
Samples are already confirmed not to contain aggregates and contaminants.IG-1000 Plus
Single Nano Particle Size Analyzer
1. Nanoparticles
2. Pharmaceuticals
3. Cosmetics
4. Food Products
5. Ceramics
6. Macromolecules
7. Catalysts
8. Electronic Materials
9. Soil and Civil Engineering Materials
Nanoparticles are expected to provide benefits as a result of the special properties related to their size. Particle size analyzers provide an extremely important tool for developing methods to maintain good dispersion characteristics by ensuring contaminants and aggregates are identified and screened out.
The smaller the particles, the larger their specific surface area and the more quickly they dissolve. In the case of particles in medical injection, the particle size determines how they pass through or penetrate capillaries and blood vessel walls and which parts of the body they reach. This has a major influence on the efficacy and side effects of pharmaceuticals.
For lipstick, mascara, and eye shadow, subtle differences in color and shine are controlled by differences in the particle size distribution. The smoothness or UV light-blocking properties of creams also vary depending on the particle size distribution.
Many food products include powdered ingredients. The mouth, tooth, and tongue feel and other characteristics of bread, cakes, pasta, etc. depend on the particle size distribution. Also, controlling the particle size distribution in beverages is important to ensure consistent quality. For example, smaller particle sizes are used in milk and lactic acid beverages to prevent differences in concentration and taste between the upper and lower portions of the container.
The strength, density, hardness, heat resistance, water and air permeability, and other characteristics of ceramics depend not only on the type of ingredient particles, but also significantly on the particle size distribution.
When particles are used as ingredients in pipes, films, and sheets, the particle size distribution can affect the strength and light permeability of the final product.
Though chemical reactivity is affected by the specific surface area and pore structure, given the same material, the chemical reactivity can be controlled by varying the particle size distribution.
The manner and degree to which particle size affects electronic materials differ depending on the application and material. However, the quality control of particle size distribution is increasingly being required to ensure higher and more consistent quality of the final product.
The particle size distribution of soil and cement has a large effect on the stability and strength of supporting ground, the strength of buildings and other structures, and how much these change over time. Also, measuring the particle size distribution is an important factor in understanding the scale of environmental pollution in soil.
Single Nano Particle Size Analyzer
IG-1000 Plus
The IG method allows measuring particle sizes in the 0.5 to 200
nm range, with high sensitivity and high reproducibility.
With the IG method, the signal level does not depend on
particle size, so measurement is not easily affected by
contaminants or aggregates. Therefore, analyzers can be used
in regular laboratory environments.
The IG method does not use scattered light, so measurement is
not easily affected by color. Therefore, this model can even
measure black nanoparticles, which tend to absorb light.
The laser diffraction method can measure particle distributions
in the 10 nm to 300 μm (300,000 nm) range.
The single light source and broad measurement range allow
monitoring dispersion and aggregation processes in real time.
Thanks to technology that enables measuring highly
concentrated samples, nanoparticles can even be measured in
highly viscous liquid media, which was not possible with
previous models.
Voyaging into the Nano Region
Shimadzu Nano Particle Size Analyzers
Nano Particle Size Analyzer
SALD-7101
4
Dispersion and aggregation processes of unstable and easily changing nanoparticles can be evaluated with high sensitivity and high
reproducibility over a wide range.
Particle Size Range Evaluated: 0.5 nm to 300 μm (300,000 nm)
Samples can be reliably screened for contaminants and aggregates using the SALD-7101, then single-digit nanoparticles can be
accurately evaluated using the IG-1000 Plus.
Particles in the 10 to 200 nm range can be measured by both the IG-1000 Plus and SALD-7101.
Comparing the data from different measurement principles provides a multifaceted evaluation of nanoparticle dispersion and
aggregation.
Powerful Combination of IG-1000 Plus and SALD-7101 Analyzers
Accurately Assesses the Actual Status of Unstable Nanoparticles
1nm 10nm 100nm 1000nm
1μm 10μm 100μm
No
rmal
ized
Part
icle
Am
ou
nt
Primary Particles(smaller than 100 nm)
Aggregates(100 nm to several μm)
Contaminants(larger than several μm)
IG MethodSingle Nano Particle Size Analyzer IG-1000 Plus
UV Laser Diffraction Method
Nano Particle Size Analyzer SALD-7101
Primary Particle Size
Primary Particle Size
Primary Particle Size
Particle Size
Particle SizeParticle Size
With the capability of measuring the diameter of both aggregated and dispersed particles, it is a powerful tool for establishing dispersion methods.
Powerful evaluation technique achieves single-digit nanoparticles by dispersing aggregates down to primary particles.
5
Single Nano Particle Size Analyzer
IG-1000 Plus
Measures single-digit nanoparticle sizes with high sensitivity
and high reproducibility.
Accurately measures samples with broad size distributions.
Measures samples in normal environments, even without any
strict pretreatment (filtering) processes.
Effects from particle refractive index and color are negligible.
No need to even specify (input) refractive indices.
Allows verifying the validity of measurement results using raw
data (variations in intensity of diffracted light).
The IG method forms a diffraction grating made up of groups of particles in liquid, then determines the particle sizes based on how fast
the particle diffuses and breaks down. Large particles diffuse slowly and small particles, particularly single-digit nanoparticles, diffuse
quickly. This diffusion process is detected as a change in the intensity of diffracted light.
The sample (nanoparticles diffused in solution) is measured by
placing it in a batch cell and inserting an electrode substrate.
When an alternating voltage is applied to the electrodes,
dielectrophoresis acts to increase the particle concentration in the
area where the left and right electrodes face each other. This
generates cyclic variations in particle concentration, forming a
particle concentration diffraction grating. When the alternating
voltage is shut off, the nanoparticles that formed the particle
concentration diffraction grating begin to diffuse and the size of
the particles can be determined from the speed of that diffusion.
Single Nano Particle Size Analyzer
Utilizes Revolutionary IG Nanoparticle Size Measurement Technology
Diffusion of Particle Concentration Diffraction Grating External Force (dielectrophoresis)
Laser Light
Primary Diffracted Light
OFFONOFF
Large Particles
Inte
nsi
ty o
f D
iffr
acte
d L
igh
t
Time
DiffusionFormationParticle ConcentrationDiffraction Grating
Small Particles
Measurement Method:
Measurement Range:
IG method (induced grating method)
0.5 to 200 nm
* PITTCON 2009 Annual Editors’ Choice Awards Bronze Award
* Nikkan Kogyo Shimbun’s 51st Best 10 New Products Awards New Product Award
Measurement Principle
Diffraction Grating Formedfrom ParticlesComb-Shaped
Electrodes
Glass Slides
Alternating Voltage
Batch Cell
Sample Suspension
a
2a
a: Pitch of Comb- Shaped Electrodes
6
No
rmal
ized
Par
ticl
e A
mo
un
t
Particle Diameter (nm)N
orm
aliz
ed P
arti
cle
Am
ou
nt
Particle Diameter (nm)
Even the particle size of fullerene hydroxide, which typifies samples in the
single-digit nanometer region, can be measured with high sensitivity and
good reproducibility.
(measurement concentration: 0.1 wt%)
Even mixed samples can be measured. The IG method utilizes the diffusion of a diffraction grating formed from particles to measure samples, so the signal
level is not dependent on particle size. Therefore, this allows evaluating mixed samples. It is difficult to measure mixed samples using scattered light methods
because the signal level is proportional to the cube of the particle diameter, even for the same volume.
Even for nanoparticle samples with broad distributions where the largest
particle is 10 times the size of the smallest one, the presence of smaller
particles is accurately detected to ensure the distribution is not skewed
toward the large end.
Measurement Examples
Colloidal Silica ○ : A, ● : B, △ : A + B Polystyrene Latex ○ : 50nm, ● : 100nm, △ : 50nm + 100nm
No
rmal
ized
Par
ticl
e A
mo
un
t
Particle Diameter (nm)
No
rmal
ized
Par
ticl
e A
mo
un
t
Particle Diameter (nm)
Inte
nsit
y of
Diff
ract
ed L
ight
Diffusion Time (ms)
ItemMeasurement PrincipleMeasurement RangeMeasurement TimeMeasurement Method Compatible OSTemperature Adjustment Range Size and WeightConsumables
SpecificationsIG method (induced grating method)0.5 to 200 nm30 sec (from start of measurement until results are displayed)Batch cell (0.3 mL volume)Windows 7 / XP / Vista7 to 40 °C (when using optional thermostatic circulator) W600 × D400 × H200 mm and 15 kgElectrode tips
Specifications
(sample provided by professor Kokubo of Osaka University)
Measurement Example (1):
Fullerene Hydroxide
The intensity of diffracted light decreases at a steep slope for smaller
particles, but the slope is less steep for larger particles. Therefore, the
relation between small and large particles (average particle diameters) can be
verified at a glance.
Allows Verifying the Validity of Measurement Results Using Raw Data
(Variations in Intensity of Diffracted Light)
Measurement Example (2):
Samples with Broad Distributions
Measurement Example (3): Measuring Mixed Samples
* Particle concentration: 0.01 to 1 %
* Max. conductivity of liquid media: About 400 μS/cm
(microsiemens per centimeter)
* Nonpolar solvents cannot be used
(such as toluene or hexane)
7
Nano Particle Size Analyzer
SALD-7101 Measurement Range: 10 nm to 300 μm (300,000 nm)
Allows observing the status of unstable nanoparticles, from a few tens of nanometers to several hundred micrometers, in real time.
Allows reliably evaluating the dispersion and aggregation processes of particles, as well as the status of contaminants, including how
these change over time.
Allows monitoring changes in the status of unstable nanoparticles in real time over a wide range, from 10 nm to 300 μm (300,000 nm).
Offers the following advantages over DLS (dynamic light scattering).
Allows real-time measurements at 1-second intervals.
Even measures nanoparticles dispersed in highly viscous media or solids, such as films.
Measures particles in a wide range of particle concentrations, from a few ppm to 20 %.
The high measurement limit of 300 μm (300,000 nm) allows confirming particle aggregates and contaminants, which means it can
even be used for screening nanoparticle samples.
Designed specifically for wet measurement.
Allows inline real-time size measurements of micro and nano air bubbles.
Uses a UV laser light source (375 nm wavelength) and contiguous wide angle scattered light detection.
The laser diffraction method irradiates samples with laser light and detects the changes in intensities, which vary depending on the
angle of the diffracted and scattered light. It then uses calculations to determine the particle size distribution. The SALD-7101 uses a
UV laser light source and a contiguous wide angle scattered light detection system that allow observing not only primary particles but
also aggregates and contaminants, ranging from the nano region to several hundred micrometers, simultaneously and in real time.
SALD Series Particle Size Analyzers
Ultra VioletLaser
Lens
Sensors for Side / BackwardScattered Light
Cell
60 degrees
Focal Length
Virtual Shape of Sensor forForward Scattered Light
Wing parts
Sensor for ForwardScattered Light(Wing Sensor)
Sensor for ForwardScattered Light(Wing Sensor)
Optical System with Contiguous Wide Angle Scattered Light Detection
Function of the Wing sensor is equivalent to that of the concentric shape sensor by compensation of light intensity detected from respective sensor element. In the compensation, area of respective sensor element and effect of polarization must be considered.
8
Light Source: Red semiconductor laser (680 nm wavelength)
The SALD-2300 is the standard model in the SALD series. It is
capable of measuring a wide range of particle size distributions,
such as pharmaceuticals, foods, and electronic materials.
Operability and data analysis functions have been improved
from its preceding models, the SALD-2001/2101/2201, while
ensuring data remains inter-compatible.
Supports injection type dry measurement as well.
Laser Diffraction Particle Size Analyzer
SALD-2300 Measurement Range: 17 nm to 2500 μm
Light Source: Red semiconductor laser (690 nm wavelength)
Measures even coarse particles, such as soil or metal particles,
or particles with high specific gravity. It is used for civil
engineering and construction, research of environmental
problems with rivers and coastlines, disaster prevention
measures, and other purposes. It is also used to evaluate coal
ash in thermal power plants and other applications.
Supports both injection and free-fall type dry measurement.
SALD-3101 Measurement Range: 50 nm to 3000 μm
Light Source: Red semiconductor laser (670 nm wavelength)
Low cost particle size analyzer that offers high performance
and a small footprint.
Ideal for particles with relatively low specific gravity (less than 2),
such as foods, beverages, pharmaceuticals, cosmetics, and
emulsions.
Designed specifically for wet measurement.
SALD-201V Measurement Range: 250 nm to 350 μm
Light Source: Violet semiconductor laser (405 nm wavelength)
Based on the SALD-201V, the SALD-301V offers an improved
light source and optical system to provide better performance
in the microparticle region. It also enables measuring blue and
black type particles, such as in pigments and coatings.
Designed specifically for wet measurement.
SALD-301V Measurement Range: 100 nm to 350 μm
9
Scattered light from coarse particles is concentrated at low angles near the optical axis
and fluctuates vigorously within a tiny angle, but scattered light from microparticles
fluctuates slowly up to large angles away from center. Whereas the intensity of scattered
light from coarse particles is extremely high, the intensity of scattered light from
microparticles is very low. The SALD series achieves high resolution over a wide range of
particle sizes by utilizing the relationship between particle size and scattered light and
increasing the detection surface area of each of the 76 concentric detector elements in
the Wing sensor at a logarithmic rate from the center outward. In addition to the Wing
sensor, one sensor is used for side scattered light and four sensors are used for backward
scattered light.
SALD Series Particle Size Analyzers
Particle Size Distribution Data with Five Peaks
Reliably reproduces even particle size distributions with complicated
distribution shapes. This is an example of measuring a mixture of five types
of particles, with diameters of 0.6, 2.5, 10, 30, and 300 μm.
Rather large aggregates present in nanoparticle samples can be measured at
the same time. For example, this can also be used for screening when
reusing CMP.
Q 3 (%) q 3 (%)
0.01 0.05 0.1 0.5 1 5 10 50 100 5000
10
20
30
40
50
60
70
80
90
100
The above is an overlay of data from measuring the same sample 10 times,
showing high reproducibility.
NIST (National Institute of Standards and Technology) traceable particles can
be measured with extremely high reproducibility. The standard deviation
from multiple mean particle diameter measurements is practically zero.
Wing Sensor
Note: SALD-201V and SALD-301V models use a sensor with 54 concentric elements instead of a Wing sensor.
Particle Diameter (μm)
No
rmal
ized
Par
ticl
e A
mo
un
t
Particle Diameter (μm)
No
rmal
ized
Par
ticl
e A
mo
un
t(c
um
ula
tive
)
No
rmal
ized
Par
ticl
e A
mo
un
t(f
req
uen
cy)
Particle Diameter (μm)
No
rmal
ized
Par
ticl
e A
mo
un
t(c
um
ula
tive
)
No
rmal
ized
Par
ticl
e A
mo
un
t(f
req
uen
cy)
Particle Diameter (μm)
No
rmal
ized
Par
ticl
e A
mo
un
t(c
um
ula
tive
)
No
rmal
ized
Par
ticl
e A
mo
un
t(f
req
uen
cy)
SALD series analyzers accurately capture the true image of particle size distributions. They detect the presence of aggregates and
contaminants in nanoparticle samples.
Detects True Image of Particle Size Distributions with
Accuracy and High Resolution
Utilizing a single light source and highly reliable optical system provides extremely high reproducibility in terms of consistent sample
measurement results. This allows measuring and evaluating even subtle changes in samples with confidence.
Achieves High Reproducibility to
Accurately Measure Subtle Changes in Samples
10
Laser diffraction method ISO 13320 and JIS Z 8825-1 compliant
SALD systems comply with ISO 13320 and JIS Z 8825-1 laser diffraction and scattering standards.
Validation possible with JIS standard particles
System performance can be confirmed using MBP1-10 standard particles specified in JIS Z 8900-1 (for SALD-7101 and SALD-2300
models). These samples have a broad particle size distribution, which is specified by the JIS standard. Using these samples allows
verifying that the instrument is always accurate.
Models SALD-3101, SALD-301V, and SALD-201V use JIS class 11 test dust, specified in JIS Z 8901, for validation.
MBP1-10 standard particles (347-61001-01) are included standard with SALD-7101 analyzers.
MBP1-10 standard particles (347-61004) are included standard with SALD-2300 analyzers.
JIS class 11 test dust is included standard with SALD-3101, SALD-301V, and SALD-201V analyzers.
Respective part numbers are SALD-3101 (346-63722), SALD-301V (346-62474), and SALD-201V (346-62340-02).
Easy maintenance
Systems feature a powerful self-diagnostic function.
This allows checking the output signals from each sensor and detection element and the
system functional status. The operation log function stores detailed information with all
measurement data, such as the instrument operating status and cell contamination
status. This allows retroactively verifying the validity of measurement data and
confirming the cell contamination status.
Improved stability of the optical system
SALD systems utilize an OSAF (omnidirectional shock absorption frame). This completely isolates all elements of the optical system from
impacts, vibration, and other external disturbances. Therefore, optical axes rarely need adjustment.
Allows verifying the validity of measurement results (particle size distribution
data) by referencing light intensity distribution data (raw data)
Since light intensity distribution data (raw data) and measurement results (particle size
distribution data) can be displayed on the same screen, measurement results can be
verified while viewing both sets of data. This allows not only verifying whether the
detection signal level (particle concentration) is appropriate or not, but also confirming
the validity of measurement results from multiple aspects, such as in terms of the
distribution width and the presence of aggregates and contaminants.
High Reliability Established by ISO-/JIS-Based Validation
and Refined Hardware/Software
11
(1) Specify a range forthe refractive index
(3) Display candidatesfrom refractive index and particle size
calculation results
(2) Evaluate
SALD Series Particle Size Analyzers
Selecting a refractive index was an unavoidable part of using the laser diffraction method, where generally a published value was
entered, but such values were not necessarily appropriate, considering the effects of particle composition and shape. Therefore, tedious
trial and error processes were used to select refractive indices.
WingSALD II solves such problems by being the world's first software to include a function that automatically calculates an appropriate
refractive index based on the LDR method (light intensity distribution reproduction method).
Note: The LDR method automatically calculates an appropriate refractive index based on consistency between the actual measured light intensity distribution and one reproduced (recalculated) from particle size distribution data. This method was developed by Shimadzu and published in two technical papers. It is sometimes called the "Kinoshita Method", in academic communities, after the name of Shimadzu’s engineer.
Automatic Refractive Index Calculation Function Eliminates
the Mistake or Trouble of Selecting Refractive Indices
12
Creating, saving, and sharing measurement conditions and procedures, including pretreatment methods and conditions, ensures
measurements are performed using the same conditions and procedures, even if performed by a different operator or at a different
location or plant, and allows safely comparing data.
Furthermore, when the measurement assistant function is used, measurement instructions for the operator are displayed on the screen.
This enables even inexperienced operators to perform measurements correctly.
In addition, administrators and operators can be assigned different operating privileges to ensure security.
Note: SOP is an acronym for Standard Operating Procedure.
Measurement Assistant Function Allows Preparing an SOP to Ensure
Measurements Are Always Performed Using the Same Conditions and Procedures
Create and save measurementconditions and procedures (SOP)
Procedures, remarks,and other information are displayedinteractively during measurements.
This standardizes measurementprocedures and prevents mistakes.
Includes security features that limit the functionality
available to operators
13
Measured as-is
Particle Diameter (μm)
No
rmal
ized
Par
ticl
e A
mo
un
t Measured diluted
SALD Series Particle Size Analyzers
Concentrated samples that were not previously measurable with laser diffraction and scattering can now be measured without dilution.
If high-concentration samples are measured using a standard flow cell or batch cell, the optical path length becomes very long, causing
multiple scattering, which prevents accurate measurement. Therefore, samples are now sandwiched in a special sample cell that
shortens the optical path length, avoiding the negative effects of multiple scattering. This makes it possible to measure such samples
accurately.
Using optional WingSALD-CM software allows measuring changes to particle size distributions at intervals as short as 1 second and saving
the results.
This is an example of light intensity distribution data and particle size distribution data for the dissolution process of calcium carbonate. It shows how dissolution
progresses from smaller diameter particles and how the normalized amount of large particles increases.
This example compares measuring hand cream as-is (highly concentrated)
and diluted. It shows that diluting the sample made the distribution
narrower and prevented an accurate measurement.
Measuring the undiluted solution as-is allows obtaining accurate
measurement results.
q 3 (%)
0.1 0.5 1 5 10 500
2
4
6
8
10
12
14
16
18
20(%)
0
20
40
60
80
100
5 10 15 20 25 30 35 40 45 50 55 60 65 5
100% = 2000
Updates particle sizedistributions and lightintensity in real time
Particle Diameter (μm)
No
rmal
ized
Par
ticl
e A
mo
un
t
Sensor Element Number
1. Initial 2. After 20 sec3. After 40 sec4. After 60 sec
5. After 80 sec6. After 100 sec7. After 120 sec8. After 140 sec
9. After 160 sec10. After 180 sec
Measures Without Diluting Concentrated Samples, Such as Creams
Particle size distribution data and light intensity distribution data
can be displayed in real time. This means that sample changes
over time or shifts in the dispersion status can be monitored in
real time. Since both the light intensity distribution data, which is
the raw data, and particle size data can be monitored
simultaneously, they can be compared to keep track of any
changes in the status of samples.
Monitors Changes in Sample Status in Real Time
14
The following data analysis applications are included standard.
Evaluation of Scattering Angle
Graphs the components of scattered light intensity at each angle. This takes advantage of the features of the highly-integrated photodiode
array to allow evaluating the low-angle scattered light with high resolution.
Application Fields: Evaluating the scattering characteristics of films and sheets
Data Emulation Function
Based on SALD series measurement results, this function allows emulating measurement results obtained using other models and
measurement principles. This allows maintaining data compatibility with previous measurement methods.
Mixture Data Simulation Function
Allows simulating particle size distributions using any mixture ratio of multiple particle size distributions. This makes it possible to
determine the optimal mixture ratio for obtaining the desired particle size distribution, without the trouble of actually measuring the
particle size distribution of sample mixtures over and over.
Data Connecting Function
Allows combining the measurement results for two different measurement ranges at any particle size point to create a single particle
size distribution. For example, sieve data for particles above 2000 μm can be combined with SALD series data for particles below 2000 μm
to create a wide-ranging particle size distribution, which is required for civil engineering, disaster prevention, or environmental fields.
Evaluates Measurement Data from Multiple Facets
–Extensive Assortment of Data Analysis Applications Included Standard
Selectionof data Grouping
Save grouped data
DrawingAnalysis
Re-drawingReanalysis
Multiple sets of data can be stored as a group.
This makes not only organizing the data but also redisplaying and reanalyzing it easier.
Data can be loaded as a group and displayed or analyzed at the same time, rather than having to load each set of data separately.
Processes Multiple Sets of Data More Efficiently
15
SALD Series Particle Size Analyzers
Summary Chart of SALD Series Measurement Functions
680 690
*3
*7
*11
*15
*16
*4
*8
*12
*1
*5
*9
*4
*8
*13
*2
*6
*10
*14
16
*1 Use an SALD-MS71 sampler for SALD-7101 analyzers (P/N 347-60711-20 (115 V), 347-60711-30 (230 V)). Sold separately from the main analyzer (measurement unit).
*2 Use an SALD-MS23 sampler for SALD-2300 analyzers (P/N 347-61701-42 (115 V), 347-61701-44 (230 V)).Sold separately from the main analyzer (measurement unit).
*3 Use an SALD-MS30 sampler for SALD-3101 analyzers (P/N 346-63701-20 (115 V), 346-63701-30 (230 V)).Sold separately from the main analyzer (measurement unit).
*4 Samplers for models SALD-201V and SALD-301V are included standard with the analyzer.
*5 Use a batch cell for SALD-7101 analyzers (P/N 347-60712). Sold separately from the main analyzer (measurement unit).
*6 Use a batch cell for SALD-2300 analyzers (P/N 347-61702-42). Sold separately from the main analyzer (measurement unit).
*7 Use a batch cell for SALD-3101 analyzers (P/N 321-48338-02). Sold separately from the main analyzer (measurement unit).
*8 Batch cells for models SALD-201V and SALD-301V are included standard with the analyzer.
*9 Use a high-concentration sample measurement system for SALD-7101 analyzers (P/N 347-60713-21).Sold separately from the main analyzer (measurement unit).
*10 Use a high-concentration sample measurement system for SALD-2300 analyzers (P/N 347-61703-42).Sold separately from the main analyzer (measurement unit).
*11 Use a high-concentration sample measurement system for SALD-3101 analyzers (P/N 347-60001-21). old separately from the main analyzer (measurement unit).
*12 Use a high-concentration sample measurement system for SALD-301V analyzers (P/N 346-63420-21).Sold separately from the main analyzer (measurement unit).
*13 Use a high-concentration sample measurement system for SALD-201V analyzers (P/N 346-63141-21).Sold separately from the main analyzer (measurement unit).
*14 Use an SALD-DS5 Injection Type Dry Measurement Unit (P/N 347-61706-42 (115 V), 347-61706-44 (230 V)).Sold separately from the main analyzer (measurement unit).Requires separate, compressor, and dust collector.
*15 Use an SALD-DS5 Injection Type Dry Measurement Unit (P/N 347-61707-20 (115 V), 347-61707-30 (230 V)).Sold separately from the main analyzer (measurement unit).Requires separate detector case for SALD-3101 (P/N 347-60010-21), compressor, and dust collector.
*16 Use an SALD-DS3 Free-Fall Type Dry Measurement Unit (P/N 346-61206-21 (115 V), 346-61206-31 (230 V)).Sold separately from the main analyzer (measurement unit).Requires separate detector case for SALD-3101 (P/N 347-60010-21) and dust collector.
Dispersion Bath
Side ScatteredLight Sensor
Backward Scattered Light Sensor
Forward ScatteredLight Sensor
Light Source
Flow Cell
Pump
Stirrer
UltrasonicSonicator
Batch Cells for SALD-3101/SALD-2300/SALD-7101
SALD-BC30
SALD-BC23
SALD-BC71
12cm3
12cm3
7cm3
SALD-3101
SALD-2300
SALD-7101
SALD-MS71
SALD-MS23
SALD-MS30
280cm3
100~300cm3 *1
400cm3
1600cm3/min
1600cm3/min
5000cm3/min
Yes
Yes
No
Yes
Yes
No
SALD-7101
SALD-2300
SALD-3101
*1 The sample quantity is changable. 100mL, 200mL or 300mL can be selected.
17
Samplers for SALD-3101/SALD-2300/SALD-7101
SALD Series Particle Size Analyzers
Free-Fall Type Dry Measurement Unit for SALD-3101
SALD-DS5 for 3101
SALD-DS5 for 2300
SALD-3101
SALD-2300
*1 Requires separate compressor and dust collector
*2 SALD-3101 requires separate detector case (P/N 347-60010-21)
CondenserLens
Detection Surface
Laser LightSource
CollimatorLens
Diffraction/ScatteringImage
Diffracted/ScatteredLight Sensor
Suction
Vibrating Feeder
Free-Fall
SALD-DS3 free-fall type dry measurement
unit installed on SALD-3101
SALD-DS5 cyclone injection type dry measurement
unit installed on SALD-2300
Ejector
Hopper
Air Inlet to PreventBackflow
Ejector
Air Valve
(Compressed AirSource)
Sample can be injectedonly under the operationof a vacuum cleaner
3 Types
Injection Nozzle
Injection
Sensor for Forward Scattered Light
Scattered Light
Condensing Lens
Sample Particles
Pressure Sensor toConfirm the Operationof Dust Collector
Suction
Suction Tube
Suction Nozzle
Hand Switch
Pressure Regulator(PC Control)
PC
18
Powder Samples Can Be Measured Without Preparation
Continuous Measurement Software for SALD-7101
WingSALD-7101 CM SALD-7101
Measuring high-concentration
samples using regular methods
Using high-concentration
sample measurement system
SALD-HC71
SALD-HC23
SALD-HC30
SALD-301V-HC
SALD-201V-HC
SALD-7101
SALD-2300
SALD-3101
SALD-301V
SALD-201V
Glass sample plate (0.1mm)
Glass sample plate (0.2mm)
Glass sample plate (0.3mm)
Glass sample plate (0.4mm)
Glass sample plate (0.5mm)
Glass sample plate (0.05mm)
0.1mm (100μm)
0.2mm (200μm)
0.3mm (300μm)
0.4mm (400μm)
0.5mm (500μm)
0.05mm (50μm)
10
10
10
10
10
10
0.03cm3
0.06cm3
0.09cm3
0.12cm3
0.15cm3
0.015cm3
Few hundred ppm
to several percent
CrossSection
Indentation Depth
Glass Sample Plate(glass slide with indentation)
RoundIndentation
Glass SamplePlate
Glass Slide
Sample
Glass Sample Plates (glass slides with indentation sold separately)
Optional Software
19
High-Concentration Sample Measurement System for SALD Series
Printed in Japan 3655-12214-10A-IK
© Shimadzu Corporation, 2013
Measurement and Data Display Functions
Measurement of Particle Size Distribution
Refractive Index Setting
Real-Time Display of Particle Size Distribution
Real-Time Display of Light Intensity Distribution
Display of Particle Size Distribution Data
Display of Light Intensity Distribution
Recalculation of Particle Size Distribution
Diagnostics/Adjustments
Statistical Data Processing
Time-Series Processing
3-Dimensional Graphing
Data Transfer via Clipboard
Data Sorting
Report Function
Allows measurements using measurement assistant function (interactive process based on SOP)
Automatic refractive index calculation function makes setting the refractive index easy.
Min. 1-second intervals
Min. 1-second intervals
Displays overlay of max. 200 distributions
Displays overlay of max. 200 distributions
Batch recalculation of max. 200 distributions
Self-diagnostic function and cell check function
Max. 200 sets of data (also allows overlaying max. 200 data sets)
Max. 200 sets of data
Max. 200 sets of data
[Image Output]: Outputs entire data sheet or graph only.
[Text Output]: Outputs summary data, particle size distribution data, or light intensity distribution data.
Sorts by �le name, sample ID, sample number, or refractive index
Single data sets (6 templates), overlaid data (5 templates), statistical data, time-series data,
or 3D data can be selected and output using batch processing
Output Conditions
Particle Size (µm) Divisions
Particle Amount (%) Divisions
Distribution Basis
Expression of Cumulative Distribution
Expression of Frequency Distribution
Smoothing Levels
Distribution Function Fitting
Data Shifting
Fixed 51 or 101 divisions
Fixed 51 divisions
Count, length, area, or volume
Oversized or undersized
q, q / ∆×, q / ∆log ×
10 levels
Rosin-Rammler distribution, logarithmic Gaussian distribution
±10 levels
Data Analysis Functions
Scattering Angle Evaluation Function
Data Emulation Functions
Mixture Data Simulation Function
Data Connection Function
Evaluates scattering characteristics within micro angle regions for samples such as optical �lms and sheets.
Emulates measurement results from other instruments and measurement principles, using SALD series measurement results.
Simulates particle size distributions using any mixture ratio of multiple particle size distributions.
Combines two particle size distributions with different measurement ranges at any particle size point
to create a single particle size distribution.
OS
CPU
MEMORY
HDD
CD-ROM Drive
RS-232C Port
Display
Printer
Windows 7 * WingSALD CM series is compatible with Windows XP only.
Pentium Dual-Core 2.5 GHz min.
2GB min.
Min. 1 GB of free space required.
Required for software installation
One port required (However, two ports are required to control the sampler from the PC.)
SXGA (1280 × 1024 pixels) min.
Must be compatible with operating system.
User settable 51 divisions
User settable 51 divisions
PC Requirements
The software is included standard on a CD-R with the SALD system (optical system). Install the software on a PC that meets the following speci�cations.
*1 Windows is a registered trademark of Microsoft Corporation in the United States and other countries.
*2 Other company names and product names mentioned in this document are registered trademarks or trademarks of their respective companies.
*3 The TM and ® symbols are omitted in this document.
SALD Series Measurement and Data Processing Software
WingSALD II Speci�cations
Particle Size Analyzers