Palltronic ® Flow Check II Unit Simple verification of flow measurement instruments Features u Actual flow rate is automatically calculated and shown on the display u A precise pressure sensor determines the atmospheric pressure u Robust, sanitary stainless steel design makes it ideal for use in the pharmaceutical industry u Easy to use Stäubli u connections u Internal Pall Gaskleen ® filter protects the capillary in order to ensure maximum service life Benefits u No need to calculate flow rate using conversion table u Even small changes in height or weather conditions are detected u The unit can be used with all integrity test instruments in laboratories or cleanrooms u The unit can be easily and quickly connected to and disconnected from filter integrity test instruments u Internal filter protects against particles and liquids destroying the capillary USD 2643

Transcript of Palltronic® Flow Check II Unit,Palltronic® Flow Check II Unit · Palltronic ® Flow Check II Unit...

Page 1: Palltronic® Flow Check II Unit,Palltronic® Flow Check II Unit · Palltronic ® Flow Check II Unit Simple verification of flow measurement instruments Features u Actual flow rate

Palltronic® Flow Check II Unit

Simple verification of flow measurement instrumentsFeaturesu Actual flow rate is automaticallycalculated and shown on the display

u A precise pressure sensor determinesthe atmospheric pressure

u Robust, sanitary stainless steel designmakes it ideal for use in thepharmaceutical industry

u Easy to use Stäubliu connections

u Internal Pall Gaskleen® filter protects thecapillary in order to ensure maximumservice life

Benefitsu No need to calculate flow rate usingconversion table

u Even small changes in height or weatherconditions are detected

u The unit can be used with all integritytest instruments in laboratories or cleanrooms

u The unit can be easily and quicklyconnected to and disconnected fromfilter integrity test instruments

u Internal filter protects against particlesand liquids destroying the capillary

USD 2643

Page 2: Palltronic® Flow Check II Unit,Palltronic® Flow Check II Unit · Palltronic ® Flow Check II Unit Simple verification of flow measurement instruments Features u Actual flow rate

DescriptionThe Palltronic Flow Check II unit enables the user to perform afast, simple check on a filter integrity test instrument to confirmthat the instrument is measuring the air or nitrogen gas flowcorrectly. A ‘Flow Check’ test can be performed as often as theuser requires, in order to ensure complete confidence in theintegrity test instrument. The user simply connects the FlowCheck II unit to the integrity test instrument and performs eithera Flow Check test (Palltronic Flowstar integrity test instrumentsonly), Forward Flow or Diffusion test. The flow measured by theintegrity test instrument is then compared with the calibratedflow of the Flow Check II unit.

The gas flow rate through a capillary is dependent on thedifferential pressure across the capillary. The inlet pressure iscontrolled by the connected integrity test instrument, as pressureabove the actual atmospheric pressure. The atmosphericpressure is dependent on the weather situation and local heightabove sea level. The Flow Check II unit has a built-in pressuresensor. When the button on the side of the unit is pressed, theatmospheric pressure is measured and the unit displays theactual flow rate. There is no need to use a table to calculate acorrection factor.

SpecificationsDimensions180 x 150 x 90 mm (6 x 5 x 3 in.)

Weight2 kg (4.4 lbs)

Battery:The unit include a standard battery (1.5 V battery type AA(LR06/AM-3). The battery is replaced during annual service and calibration.

Scope of SupplyEach Palltronic Flow Check II unit is supplied with a calibrationcertificate, Instructions for Use, pneumatic tubing and Stäubli* fittings.

The Palltronic Flow Check II unit is available in three configurations- which version should be used depends on the flow range of theintegrity test instrument being checked. For the FC02M andFC02H models (see table below), a volume must be installedbetween the unit and the instrument.

Ordering InformationUpstream Volume

Part Number Flow Rate Range RequirementsFC02 8 ± 1 mL/min 0.5 to 2 meter hose

FC02M 90 ± 10 mL/min 0.5 to 2 liter vessel

FC02H 900 ± 15 mL/min 2.5 to 6 liter vessel

Europe+44 (0)23 9230 3303 phone+44 (0)23 9230 2506 [email protected] e-mail

United States800.717.7255 toll free (USA)516.484.5400 phone516.801.9548 [email protected] e-mail

International OfficesPall Corporation has offices and plants throughout the world in locations such as: Argentina, Australia, Austria, Belgium, Brazil, Canada, China, France, Germany, Hong Kong, India, Indonesia, Ireland, Italy, Japan, Korea,Malaysia, Mexico, the Netherlands, New Zealand, Norway, Poland, Puerto Rico, Russia, Singapore, South Africa,Spain, Sweden, Switzerland, Taiwan, Thailand, the United Kingdom, the United States, and Venezuela. Distributorsin all major industrial areas of the world.

The information provided in this literature was reviewed for accuracy at the time of publication. Product data may besubject to change without notice. For current information consult your local Pall distributor or contact Pall directly.

© 2009, Pall Corporation. Pall, , Palltronic and Gaskleen are trademarks of Pall Corporation. ® indicates atrademark registered in the USA. Filtration. Separation. Solution.SM is a service mark of Pall Corporation. uStäubli is atrademark of Stäubli A.G.10/2009, PDF, UK GN09.3737 USD 2643

Visit us on the Web at www.pall.com/biopharme-mail us at [email protected]