PA-100 RP-10 RAM R-200 … · Tuesday, December 27, 2016 PA-100 RP-10 RAM R-200 Report on the...
Transcript of PA-100 RP-10 RAM R-200 … · Tuesday, December 27, 2016 PA-100 RP-10 RAM R-200 Report on the...
Tuesday, December 27, 2016
PA-100
RP-10
RAM R-200
Report on the inspection for compliance with
United States Military Standards
MIL-STD-810E
MIL-STD-462
MIL-STD-1512
Name Date Signature
Conducted by: Valentin Ladizhensky 30/11/2016
Approved by: Dimitry Ginzburg 04/12/2016
Page 2 of 74
Contents
1. Scope………….….. ............................................................................................................. 4
2. Standard test conditions ........................................................................................................ 6
3. Executive Summary ............................................................................................................. 8
4. Detailed requirements and test procedures ........................................................................... 9
5. Environmental factors…………………………………………………………………….10
6. EMC……………………………………………………………………………….……...40
ANNEX A – Qualitech-Environment Test Report ......................................................
ANNEX B – Qualitech-EMC Test Report ...................................................................
Page 3 of 74
This Page Is Left Blank
Page 4 of 74
1. Scope
1.1. Purpose.
- The test report describes all the requirements, procedures and results in accordance
with the United States Military Standards. This report includes all the test activities
that were performed to demonstrate compliance with the standard.
- Basis for inspections is a document 720-ERAV2016001/1 that lists the requirements
for inspection and special conditions
- Report Arrangement :
Section 4 of this test report is an executive summary with test results.
The numbering of the tests is according to the MIL-STD- numbers.
1.2. Equipment under test:
- PA-100M – Alpha Detector contains atmospheric cells made of aluminium containing
a thin anode wire. The surface of the cell is a membrane of aluminized Mylar, protected
by a fine stainless steel wire mesh.
- RP-10 – Gamma Detector scintillation probe. Contains a highly efficient, hermetically
sealed, NaI(Tl) scintillator, 2″ diameter and 0.04″ thick, coupled to a 2″ photomultiplier
tube. This is encased in a rugged, window splash proof housing for protection against
shock, vibration and humidity, with 1mm Aluminium.
- RAM R-200 - This Monitor connect with the two above-mentioned detectors (PA-100
and RP-10). RAM-R-200 is portable multifunctional, rugged survey meter designed for
measuring wide range gamma radiation fields and contamination. The RAM R-200
along with its internal detectors is waterproof to a depth of 1 meter.
- RAM R-200 Detector Cable: This is a special cable for connection between the
Monitor and Detectors.
- Case – (BAK-1425) Case for storage and transportation of the above equipment.
Watertight, crushproof, and dustproof with Foam. (By Polycart Technologies Ltd.)
Table 1. Detectors under Test
No Hardware Component Manufacturer PN Serial N Quantity
1 RAM R-200 13220 10106290 1
2 RAM R-200 13220 10106302 1
3 RAM R-200 4-0037 8104-027 1
4 RAM R-200 4-0037 8104-029 1
5 PA-100M BAK-1340 9615-003 1
6 PA-100M BAK-1340 9615-007 1
7 RP-10 BAK-2175 3307-004 1
8 RP-10 BAK-2175 3307-012 1
Page 5 of 74
1.3. References
Testing is carried out according the recommendations and requirements of the following
sources:
- 720-ERAV2016001/1 – Document Basis for inspections
- DoD Test Readiness Review Checklist (TRR)
- TEST AND EVALUATION MANAGEMENT GUIDE, PUBLISHED BY THE
DEFENSE ACQUISITION UNIVERSITY PRESS FORT BELVOIR, VA 22060-
55654
- IEC 1000-4-2 (ESD Immunity)
- MIL-STD-810 MILITARY STANDARD: ENVIRONMENTAL TEST METHODS AND
ENGINEERING GUIDELINES
- MIL-STD-462 MILITARY STANDARD: MEASUREMENT OF ELECTROMAGNETIC
INTERFERENCE CHARACTERISTICS
- MIL-STD-1512 MILITARY STANDARD: ELECTROEXPLOSIVE SUBSYSTEMS,
ELECTRICALLY INITIATED, DESIGN REQUIREMENTS AND TEST METHODS. MILITARY
STANDARD ELECTROEXPLOSIVE SUBSYSTEMS, ELECTRICALLY INITIATED, DESIGN
REQUIREMENTS AND TEST METHODS
Page 6 of 74
2. Standard test conditions
Table 2. Compliance test standards
## Test Name Standard Notes
1 Temperature Storage MIL-STD-810E, Meth 501.3/502.3, Proc I Environment
2 Operation Temperature MIL-STD-810E, Meth 501.3/502.3, Proc II Environment
3 Low Pressure MIL-STD-810E, Meth 500.3, Proc II. Environment
4 Humidity MIL-STD-810E, Meth 507.3, Proc I, Cycle 3 Environment
5 Sand and Dust MIL-STD-810E, Meth 510.3, Proc I. Environment
6 Vibration MIL-STD-810E, Meth 514.4 Environment
7 Rain MIL-STD-810E, Meth 506.3 Section-II Environment
8 Radiated Emissions MIL-STD-462, RE02 EMC
9 Radiated Susceptibility MIL-STD-462, RS03 EMC
10 ESD MIL-STD-1512, Proc.205 EMC
11 RFI/EMI IEC 61503, Ed.1 EMC
12 Mechanical Shock MIL-STD-810E, Meth 516.4, Proc IV Environment
Page 7 of 74
2.1. List of tests and responsibilities
Table 3. List of tests and responsibilities
## Test Name Performed by Date
Reference detailed Report No
1 Temperature (Operation/Storage) Qualitech-Environment 18-20/09/2016
20160918-1605
2 Low Pressure Qualitech-Environment 20/09/2016 20160918-1605
3 Humidity Qualitech-Environment 20-22/09/2016 20160918-1605
4 Sand and Dust Carmel Laboratories 3-6/11/2016 20160918-1605
5 Vibration Qualitech-Environment 25/10/2016 20160918-1605
6 Rain Qualitech-Environment 25/10/2016 20160918-1605
7 Radiated Emissions Qualitech-EMC 6/10/2016 RMI 031116
8 Radiated Susceptibility Qualitech-EMC 9/10/2016 RMI 031116
9 ESD Qualitech-EMC 13/10/2016 RMI 031116
10 RFI/EMI Qualitech- EMC 5/10/2016 20160918-1605
11 Mechanical Shock Qualitech-Environment 7/11/2016 20160918-1605
- Test Source: Thorium-232
Page 8 of 74
3. Executive Summary Tested instruments demonstrated compliance with the referenced standards. Few deficiencies
were identified and recommendations were made for their correction. One failure was
observed during one the Operation Temperature test that requires reviewing the instrument
specifications for this characteristic. All the results are detailed in Table 4.
Table 4. Test summary and detailed results
## Test Name Standard Test Result
1 Temperature Storage MIL-STD-810E, Meth 501.3/502.3, Proc I Passed
2 Operation Temperature MIL-STD-810E, Meth 501.3/502.3, Proc II Passed – for RP-10
Failed – for PA-100
3 Low Pressure MIL-STD-810E, Meth 500.3, Proc II. Passed
4 Humidity MIL-STD-810E, Meth 507.3, Proc I, Cycle 3 Passed
5 Sand and Dust MIL-STD-810E, Meth 510.3, Proc I. Passed
6 Vibration MIL-STD-810E, Meth 514.4 Passed
7 Rain MIL-STD-810E, Meth 506.3 Section-II Passed
8 Radiated Emissions MIL-STD-462, RE02 Passed
9 Radiated Susceptibility MIL-STD-462, RS03 Passed
10 ESD MIL-STD-1512, Proc.205 Passed
11 RFI/EMI IEC 61503, Ed.1 Passed
12 Mechanical Shock MIL-STD-810E, Meth 516.4, Proc IV Passed
Page 9 of 74
4. Detailed requirements and test procedures
Detailed requirements and test methods are described in the attached documents laboratory
reports:
“QualiTech Compliance Engineering Environmental & Mechanical Lab” Annex A -
Report No: 20160918-1605
“QualiTech EMC Laboratory” Annex B - Report No: RMI 031116
Page 10 of 74
5. Environmental factors
The Environmental tests were conducted at the accredited laboratory
“QualiTech Compliance Engineering Environmental & Mechanical Lab”.
See Annex A - Report No: 20160918-1605
5.1. Storage Temperature 5.1.1. Scope
This document describes the testing process of the EUT response in the Storage
temperature according to the MIL-STD-810E Meth 501.3/502.3 Procedure I. Use high and
low temperature tests to obtain data to help evaluate effects of high/low temperature
conditions on material safety, integrity, and performance.
An instrument shall not exhibit a response greater than that stated by the manufacturer to
radiation other than the type of radiation for which it is designed.
5.1.2. Experiment
Use Procedure-I to investigate how high/low temperatures during storage affect the
material (integrity of materials, and safety/performance of the material). This test procedure
includes exposing the test item to high/low temperatures (and low humidity where
applicable) that may be encountered in the material's storage situation, followed by an
operational test at controlled or high temperature ambient conditions.
Storage -20°C/+70°C.
Total test duration of 50 hours, 24 hours in each extreme temperature.
Page 11 of 74
Fig.5.1.2. Temperature profile on Storage mode
All devices in the test are packed in Case (Pic.5.1.2.)
Pic.5.1.2
Page 12 of 74
Table 5.1.2. UUT configuration
5.1.3. Results:
a) Instrument functionality
Table 5.1.3. Storage test result.
Instruments show solid performance after the test, the measurement is correct and
does not differ from measurement before the storage. The loss of product functionality is not
detected.
Page 13 of 74
b) Mechanical status
- Observed slight compression and deformation of foam inside the case. (Pic.5.1.3.1)
- In one place foam damage found (Pic.5.1.3.2)
Despite the damage, the functionality of the Case is according to requirements. The loss of
product functionality is not detected.
(Pic.5.1.3.1) (Pic.5.1.3.2)
Page 14 of 74
5.1.4. Conclusion
The results of this test confirm that the EUT withstands the Storage temperature test
according to the MIL-STD-810E Meth 501.3/502.3 Procedure I standard.
Foam in the case needs improvement.
5.1.5. Solution
Representative of “Polycart Technologies Ltd.” proposed changes of material and
technology, which should solve the problem and eliminate the disadvantages.
The requirement is met
Page 15 of 74
5.2. Operation Temperature 5.2.1. Scope
This document describes the testing process of the EUT response in the Operation
temperature according to the MIL-STD-810E Meth 501.3/502.3 Procedure II.
Use Procedure II to investigate how high/low ambient temperatures may affect material
performance while it is operating.
Specification of UUT:
a) Detector PA-100:
- Temperature Range Operation: -10°C to +50°C
- Accuracy ± 15%
b) Detector RP-10:
- Temperature Range Operation: -30°C to +60°C
- Instrument accuracy is not determined to be a function of temperature and does
not affect the functionality of the device.
5.2.2. Experiment
Operation -10°C/+55°C, 3 cycle of 4 hours in each extreme temperature. Total test duration
of 40 hours.
Fig.5.2.2. Temperature profile on Operation temperature test
Page 16 of 74
5.2.3. Results:
5.2.3.1. Checking prior to the test.
All instruments have been tested in the chamber in a stable geometry before starting
Temperature testing at room temperature +24°C
Table. 5.2.3.1
Page 17 of 74
5.2.3.2. Results of Temperature Operation Test
Table. 5.2.3.2
- RP-10 S/N 3307-012 (Det-8): Demonstrated stable operation over the entire
temperature range. Test Passed.
- RP-10 S/N 3307-004 (Det-7): Demonstrated stable operation over the entire
temperature range. Registered unintentional entering into the calibration mode. The
UUT then continued normal operation. Test Passed.
- PA-100 S/N 9615-003 (Det-5): It does not operate at temperatures -10°С, to absence
of measurement. At other temperatures operation is normal. Test Failed.
- PA-100 S/N 9615-007 (Det-6): does not operate at temperatures -10°С, unacceptable
measurement error. At other temperatures operation is normal. Test Failed.
Page 18 of 74
5.2.4. Conclusion:
a) The results of this test confirm that the EUT RP-10 withstands the Temperature Operation
requirement according to the MIL-STD-810E.
The requirement is met
b) The results of this test confirm that EUT PA-100 withstands, except the operation at low
temperatures preventing Temperature Operation requirement according
to the MIL-STD-810E.
The requirement is met partially
5.2.5. Solution:
Adjust instrument specification of the device PA-100 with revised Operational Temperature
conditions.
Page 19 of 74
5.3. Low pressure (altitude) test
5.3.1. Scope
This document describes the testing process of the EUT response in the Low pressure
(altitude) test according to the Mil-Std-810E, Method 500.3 Procedure I & II standard.
Use low pressure (altitude) tests to determine if material can withstand and/or operate in a
low pressure environment and/or withstand rapid pressure changes. This method includes
four low pressure tests. In this test, only the first two procedures: Procedure I (Storage);
Procedure II (Operation).
a) Procedure I - Storage/Air Transport. Procedure I is appropriate if the material is to be
transported or stored at high ground elevations or transported by air in its shipping/storage
configuration.
b) Procedure II - Operation/Air Carriage. Use Procedure II to determine the performance of
the material under low pressure conditions.
5.3.2. Experiment
a) Storage at altitude of 15,000 ft. for 1 hour
b) Operation at altitude of 5,000 ft. for 1 hour
Pic.5.3.2.1. Equipment located in Low Pressure (Altitude) chamber.
Page 20 of 74
5.3.3. Results:
Table. 5.3.3. Results of measuring during Low Pressure Test.
5.3.4. Conclusion
The results of this test confirm that the EUT withstands the Low pressure (altitude) test
requirement according to the MIL-STD-810E.
The requirement is met
Page 21 of 74
5.4. Humidity Test
5.4.1. Scope
This document describes the testing process of the EUT response in the Humidity
Test according to the MIL-STD-810E, Meth 507.3, Proc I, Cycle 3.
The purpose of this method is to determine the resistance of material to the effects of
a warm, humid atmosphere.
5.4.2. Experiment
Test condition of 30°C/94% RH for duration of 48 hours.
Pic.5.4.2.1. Equipment located in Temperature/Humidity chamber.
- RP-10 Instruments are exposed the test while it is Operating mode.
- Instruments PA-100 during the test storage into Cases.
Page 22 of 74
5.4.3. Results:
Table. 5.4.3. Randomly monitoring of measurements during Humidity test progress
The loss of product functionality is not detected. Test Passed.
5.4.4. Conclusion
The results of this test confirm that the EUT withstands the Humidity Test
requirement according to the MIL-STD-810E.
The requirement is met
Page 23 of 74
5.5. Sand and Dust
5.5.1. Scope
This document describes the testing process of the EUT response in the Dust according to the
MIL-STD-810E Meth 510.3 Procedure I.
Dust (< 150μm) procedure. This test is performed to help evaluate the ability of
material to resist the effects of dust that may obstruct openings, penetrate into cracks,
crevices, bearings, and joints, and to evaluate the effectiveness of filters.
Procedure I - Blowing Dust. Use Procedure I to investigate the susceptibility of
material to concentrations of blowing dust.
5.5.2. Experiment
Total test duration of 12 hours, 6 hours in each extreme temperature (+70°C)
Table 5.5.2. UUT configuration
Page 24 of 74
EUT is packed in Cases and placed in the Chamber:
Before Test:
Pic. 5.5.2.1
After Test:
Pic. 5.5.2.2
Page 25 of 74
Pic. 5.5.2.3
5.5.3. Results:
5.5.3.1. Leak tightness of Case: The penetration of dust inside of the case is not detected.
Test Passed.
5.5.3.2. Operating after test:
Check measurement function after the test
Page 26 of 74
Deviations in dimensions of less than 10 percent.
The loss of product functionality is not detected. Test Passed.
5.5.4. Conclusion
The results of this test confirm that the EUT withstands the Dust requirement according to
the MIL-STD-810E.
The requirement is met
Page 27 of 74
5.6. RANDOM VIBRATION
5.6.1. Scope
This document describes the testing process of the EUT response in the Vibration according
to the MIL-STD-810E Meth 514.4 Procedure I standard.
5.6.2. Experiment
During the test instruments were attached to vibration exciter. Vibration is carried out on
three axes - Vertical, Longitudinal and Transverse, of 15 minutes in each axis.
Table 5.6.2.1 Vibration exposures
Page 28 of 74
Pic. 5.6.2. UUT mounted on Electrodynamic Shaker
5.6.3. Results:
Table 5.6.3.1 Pre-Test (Before the start of stage of tests Vibration resistance)
Page 29 of 74
Table 5.6.3.2. Test after vibration on Vertical axe.
Fig. 5.6.3.2. Test after vibration on Vertical axe.
-100.0-90.0-80.0-70.0-60.0-50.0-40.0-30.0-20.0-10.0
0.010.020.030.040.050.060.070.080.090.0
100.0
1 2 3 4 5 6 7 8 9 10
ERR
OR
[%
]
Time [min]
Vertical
UUT-1 UUT-2 UUT-3 UUT-4
Page 30 of 74
Table 5.6.3.3. Test after vibration on Transverse axe.
Fig. 5.6.3.3. Test after vibration on Transverse axe.
-100.0-90.0-80.0-70.0-60.0-50.0-40.0-30.0-20.0-10.0
0.010.020.030.040.050.060.070.080.090.0
100.0
1 2 3 4 5 6 7 8 9 10
ERR
OR
[%
]
Time [min]
Transverse
UUT-1 UUT-2 UUT-3 UUT-4
Page 31 of 74
Table 5.6.3.4. Test after vibration on Longitudinal axe.
Fig.5.6.3.4. Test after vibration on Longitudinal axe.
Note: AVERAGE* - after stabilization of Detector PA-100M 5 minutes after stopping
vibration.
The loss of product functionality is not detected. Test Passed.
5.6.4. Conclusion
The results of this test confirm that the EUT withstands the Vibration requirement according
to the MIL-STD-810E.
The requirement is met
-100.0-90.0-80.0-70.0-60.0-50.0-40.0-30.0-20.0-10.0
0.010.020.030.040.050.060.070.080.090.0
100.0
1 2 3 4 5 6 7 8 9 10
ERR
OR
[%
]
Time [min]
Longitudinal
UUT-1 UUT-2 UUT-3 UUT-4
Page 32 of 74
5.7. Rain
5.7.1. Scope
This document describes the testing process of the EUT response in the Rain according to the
MIL-STD-810E, Meth 506.3 Section-II standard. This method to evaluate material likely to
be exposed to rain, water spray, or dripping water during storage, transit, or operation.
5.7.2. Experiment
- UUT-2 (RAM R-200 + PA-100M) - tested on Storage into Case (BAK-1425)
(See Annex A: Sec.8 - DRIP BOX WATER TEST - 16/7024)
- UUT-1 (RAM R-200 + RP-10) - tested in Operation Mode
(See Annex A: Sec.8 - DRIP BOX WATER TEST - 16/7061)
Pic.5.7.2. Instruments under Drip water test.
Page 33 of 74
5.7.3. Results:
a) Instrument functionality
Table.5.7.3.1. Measurements before the Rain test.
Table.5.7.3.2. Measurements after the Rain test.
The loss of product functionality is not detected. Test Passed.
Page 34 of 74
b) Impermeability
- UUT-2 (RAM R-200 + PA-100M) - tested on Storage into Case (BAK-1425)
Pic.5.7.3.1. The Case (BAK-1425) after the Rain test.
Page 35 of 74
Pic.5.7.3.2. Instruments RAM R-200 + PA-100M into The Case (BAK-1425) after the Rain
test.
During visual inspection, at completion of the test, no water penetration was observed. Test
Passed.
5.7.4. Conclusion
The results of this test confirm that the EUT withstands the Rain requirement according to
the MIL-STD-810E.
The requirement is met
Page 36 of 74
5.8. Mechanical Shock
5.8.1. Scope
This document describes the testing process of the EUT response in Mechanical Shock
(Drop) the according to the MIL-STD-816E, Meth 516.4 Paragraph IV standard.
Shock tests are performed to:
a) Provide a degree of confidence that material can physically and functionally withstand the
relatively infrequent, non-repetitive shocks encountered in handling, transportation, and
service environments. This may include an assessment of the overall material system
integrity for safety purposes in any one or all of the handling, transportation, and service
environments;
b) Determine the material's fragility level, in order that packaging may be designed to protect
the material's physical and functional integrity; and
c) Test the strength of devices that attach material to platforms that can crash.
5.8.2. Experiment
5.8.2.1. Free Fall (Drop) Test – All instruments packed into Cases (BAK-1425).
Pic.5.8.2.1.1. Packaging instruments in the cases before the Free Fall (Drop) Test.
Page 37 of 74
Pic.5.8.2.1.2. UUT stranded on LANSMONT - PDT56ED Drop Tester
Free Fall (Drop) Test.
Test description: Drop Height: 1 m.
Number of Drops of Test Item: One drop was performed on each of the 4 corners of test item.
Total Number of Drops: 4 drops.
5.8.2.2. Bench Handling Drop Test One edge was raised to a height of 10 cm, or to a position that the chassis forms an angle of
45° with the horizontal bench top; 4 drops at each unit.
Pic.5.8.2.2. Bench Handling Drop Test procedure
Page 38 of 74
5.8.3. Results:
a) Functional test:
Table.5.8.3.1. Measurements before the Drop test.
Table.5.8.3.2. Measurements After Free Fall (Drop) Test (Packed into Case)
Page 39 of 74
Table.5.8.3.3. Measurements after Bench Handling Drop Test
The loss of product functionality is not detected. Test Passed.
b) Visual inspection: Damage is not detected. Test Passed.
5.8.4. Conclusion
The results of this test confirm that the EUT withstands the Mechanical Shock requirement
according to the MIL-STD-810E, Meth 516.4 Paragraph IV.
The requirement is met
Page 40 of 74
6. Electromagnetic Compatibility Test (EMC)
The Electromagnetic Compatibility Test tests were conducted at the accredited
laboratory “QualiTech EMC Laboratory” , See Annex B - Report No: RMI 031116.
6.1. Immunity to Electrostatic Discharge
6.1.1. Scope
This document describes the testing process of the EUT response in Electrostatic
Discharge the according to IEC 61000-4-2 standard.
IEC 61000-4-2 outlines the international immunity standard for electronic equipment
ability to withstand ESD generated from a human body or metal objects with a built up static
charge. The standard assumes that the source is an electrified human body discharge, and
testing simulates the current waveform generated in those conditions.
6.1.2. Experiment
Equipment: Noiseken, ESS-2000 Electrostatic Discharge Simulator.
Contact Discharge kV
Air Discharge kV
Pic. 6.1.2. Immunity to Electrostatic Discharge Test Station
6.1.3. Results:
a) Functional test:
Page 41 of 74
Table.6.1.3.1. Measurements before ESD-Test
Table.6.1.3.2. Measurements after ESD-Test
The loss of product functionality is not detected. Test Passed.
Page 42 of 74
b) Visual Inspection: During the test, break segments appeared on the display and after the
test the unit returned to normal operation. . Test Passed.
6.1.4. Conclusion
The results of this test confirm that the EUT withstands the Electrostatic Discharge
requirement according to the to IEC 61000-4-2 standard.
The requirement is met
6.2. Immunity to Radiated Electromagnetic Field
6.2.1. Scope
This document describes the testing process of the EUT response in Immunity to
Radiated Electromagnetic Field the according to IEC 61000-4-3.
This standard is part of a set of basic EMC standards that outline test procedures
that must be successfully accomplished before products can be sold. The object of
this particular standard is to establish a common reference of a product or device to
radiated RF immunity cause by any emitting source. Products must be designed and
tested to insure that they are immune to both intentional transmitters, such as walkie
talkies and cell phones, and unintentional RF emitting devices like electric motors and
welders.
Page 43 of 74
6.2.2. Experiment
The EUT was placed in a full-anechoic chamber on a non-conductive table 0.8m above the
absorbing ground plane, and was configured, arranged and operated in a manner consistent with
typical application and load conditions. Normal performance of the EUT was verified.
The EUT was positioned parallel to the pre-calibrated “uniform area” at a distance of 2.7m for
frequency range below 1GHz and for the frequency range above 1GHz at 2.5m from the transmitting
antenna.
The frequency range was swept in steps of 1% with 1kHz sine-wave 80% Amplitude
modulated with a dwell time of 1 sec. A list of clock and sensitive frequencies were analyzed
separately if made available.
Special exercising S/W was used in order to precisely generate the required electromagnetic
field based on the calibrated field uniformity data.
All tests were performed four times, with the antenna facing each of the four sides of the EUT. Each
test was performed for both, horizontal and vertical antenna polarization
During the tests, the EUT and external equipment were monitored to verify normal functional
performance.
Pic.6.2.2. The EUT placed in a full-anechoic chamber.
Page 44 of 74
6.2.3. Results:
6.2.3.1. Radiated Immunity 80MHz - 1GHz
Table 6.2.3.1. Pre-Test (Before the start of stage of tests Radiated Immunity on range 80MHz -
1GHz )
Page 45 of 74
a)
Table 6.2.3.1.1.
Fig. 6.2.3.1.1.
-50.0
-40.0
-30.0
-20.0
-10.0
0.0
10.0
20.0
30.0
40.0
50.0
0 200 400 600 800 1000 1200Erro
r [%
]
Frequency [MHz]
Deviation measurements while changing the frequency
UUT-1 UUT-2 UUT-3 UUT-4
Page 46 of 74
Table 6.2.3.1.2.
Fig. 6.2.3.1.2.
-50.0
-40.0
-30.0
-20.0
-10.0
0.0
10.0
20.0
30.0
40.0
50.0
0 200 400 600 800 1000 1200Erro
r [%
]
Frequency [MHz]
Deviation measurements while changing the frequency
UUT-1 UUT-2 UUT-3 UUT-4
Page 47 of 74
Table 6.2.3.1.3.
Fig. 6.2.3.1.3.
-50.0
-40.0
-30.0
-20.0
-10.0
0.0
10.0
20.0
30.0
40.0
50.0
0 200 400 600 800 1000 1200Erro
r [%
]
Frequency [MHz]
Deviation measurements while changing the frequency
UUT-1 UUT-2 UUT-3 UUT-4
Page 48 of 74
Table 6.2.3.1.4.
Fig. 6.2.3.1.4.
-50.0
-40.0
-30.0
-20.0
-10.0
0.0
10.0
20.0
30.0
40.0
50.0
0 200 400 600 800 1000 1200Erro
r [%
]
Frequency [MHz]
Deviation measurements while changing the frequency
UUT-1 UUT-2 UUT-3 UUT-4
Page 49 of 74
Table 6.2.3.1.5.
Fig. 6.2.3.1.5.
-50.0
-40.0
-30.0
-20.0
-10.0
0.0
10.0
20.0
30.0
40.0
50.0
0 200 400 600 800 1000 1200Erro
r [%
]
Frequency [MHz]
Deviation measurements while changing the frequency
UUT-1 UUT-2 UUT-3 UUT-4
Page 50 of 74
Table 6.2.3.1.6.
Fig. 6.2.3.1.6.
-50.0
-40.0
-30.0
-20.0
-10.0
0.0
10.0
20.0
30.0
40.0
50.0
0 200 400 600 800 1000 1200Erro
r [%
]
Frequency [MHz]
Deviation measurements while changing the frequency
UUT-1 UUT-2 UUT-3 UUT-4
Page 51 of 74
6.2.3.2. Radiated Immunity 1.4GHz – 2.7GHz
(AM, 80%; 1kHz; 10V/M)
Table 6.2.3.2. Pre-Test (Before the start of stage of tests Radiated Immunity on range 80MHz -
1GHz )
Page 52 of 74
Table 6.2.3.2.1.
Fig. 6.2.3.2.1.
-50.0
-40.0
-30.0
-20.0
-10.0
0.0
10.0
20.0
30.0
40.0
50.0
1300 1500 1700 1900 2100 2300 2500 2700Erro
r [%
]
Frequency [MHz]
Deviation measurements while changing the frequency
UUT-1 UUT-2 UUT-3 UUT-4
Page 53 of 74
Table 6.2.3.2.2.
Fig. 6.2.3.2.2.
-50.0
-40.0
-30.0
-20.0
-10.0
0.0
10.0
20.0
30.0
40.0
50.0
1300 1500 1700 1900 2100 2300 2500 2700Erro
r [%
]
Frequency [MHz]
Deviation measurements while changing the frequency
UUT-1 UUT-2 UUT-3 UUT-4
Page 54 of 74
Table 6.2.3.2.3.
Fig. 6.2.3.2.3.
-50.0
-40.0
-30.0
-20.0
-10.0
0.0
10.0
20.0
30.0
40.0
50.0
1300 1500 1700 1900 2100 2300 2500 2700Erro
r [%
]
Frequency [MHz]
Deviation measurements while changing the frequency
UUT-1 UUT-2 UUT-3 UUT-4
Page 55 of 74
Table 6.2.3.2.4.
Fig. 6.2.3.2.4.
-50.0
-40.0
-30.0
-20.0
-10.0
0.0
10.0
20.0
30.0
40.0
50.0
1300 1500 1700 1900 2100 2300 2500 2700Erro
r [%
]
Frequency [MHz]
Deviation measurements while changing the frequency
UUT-1 UUT-2 UUT-3 UUT-4
Page 56 of 74
Table 6.2.3.2.5.
Fig. 6.2.3.2.5.
-50.0
-40.0
-30.0
-20.0
-10.0
0.0
10.0
20.0
30.0
40.0
50.0
1300 1500 1700 1900 2100 2300 2500 2700Erro
r [%
]
Frequency [MHz]
Deviation measurements while changing the frequency
UUT-1 UUT-2 UUT-3 UUT-4
Page 57 of 74
Table 6.2.3.2.6.
Fig. 6.2.3.2.6.
-50.0
-40.0
-30.0
-20.0
-10.0
0.0
10.0
20.0
30.0
40.0
50.0
1300 1500 1700 1900 2100 2300 2500 2700Erro
r [%
]
Frequency [MHz]
Deviation measurements while changing the frequency
UUT-1 UUT-2 UUT-3 UUT-4
Page 58 of 74
The loss of product functionality is not detected.
Test Passed.
6.2.4. Conclusion
The results of this test confirm that the EUT withstands the Immunity to Radiated
Electromagnetic Field the according to IEC 61000-4-3 standard.
The requirement is met
Page 59 of 74
6.3. RE02 Radiated Emission, Electric Fields 14 kHz to 10000 MHz
6.3.1. Scope
This document describes the testing process of the EUT response in Radiated
Emission the according to the MIL-STD-462 C.
This standard establishes general techniques for use in the measurement and
determination of the electromagnetic emission and susceptibility characteristics of electronic,
electrical, and electromechanical equipment and subsystems designed or procured for use by
activities and agencies of the Department of Defense.
6.3.2. Experiment
UUT are set in the Operating mode and checked by different types of antennas depending on
the frequency range under test: 10 kHz to 30 MHz, 104 cm rod with impedance matching network EMCO 3301B
30 MHz to 200 MHz, Biconical, 137 cm tip to tip, Schwarzbeck VHBB9124 w/BBA9106
200 MHz to 1GHz Log periodic Schwarzbeck VUSLP9111
1GHz – 18 GHz DRG-118/A
Pic.6.3.2. UUT on Operating mode into Camber whit different types of antennas
Page 60 of 74
6.3.3. Results
UUT-1 (Monitor RAM-R200 + Detector Rp-10): in the frequency range of 30MHz to
90MHZ spikes found (38MHz, 40MHz, 44MHz) with peaks 40 dBuV/m.
The cause of the fall was UUT-1 damage to the RAM R-200 Detector Cable (BAK-0400)
shield. This problem is fixed by adding to Cable the Ferrite Filter (ferrite number -
0444164281)
Pic. 6.3.3. RAM R-200 Detector Cable (BAK-0400) with Ferrite Filter (0444164281)
After adding to Cable the Ferrite Filter UUT-1 passed the test successfully.
All other UUT passed the test successfully.
Test Passed.
6.3.4. Conclusion
The results of this test confirm that the EUT withstands the Radiated Emission the
according to the MIL-STD-462 C.
Possible damage to the cable shield of RAM R-200 Detector Cable (BAK-0400) ,
which do not affect the work of the tested Instruments, but the cause of Radiated Emission.
For fix this problem by adding Ferrite Filter.
The requirement is met
6.3.5. Solution:
a) Develop a methodology for testing cables RAM R-200 Detector Cable (BAK-0400.)
serviceability
b) Develop a version of the cable RAM R-200 Detector Cable (BAK-0400) with ferrite filter.
Page 61 of 74
6.4. RS03 Radiated Susceptibility, Electric Field, 14 kHz to 18 GHz
6.4.1. Scope
This document describes the testing process of the EUT response in Radiated
Susceptibility the according to the MIL-STD-462 C.
The radiated susceptibility test is performed to determine a device's ability to operate
in the presence of an external interference signal propagated via free space.
6.4.2. Experiment
A transmit antenna is placed in front of the EUT's most susceptible RF pickup area at
the separation distance prescribed by the test specification.
The EF(Electric Field) is established at the specified frequencies using a signal
generator and an RF power amplifier to drive the transmit antenna. Used the suite of signal
generators, power amplifiers, and antennas to cover the entire frequency range.
Pic. 6.4.2. UUT on Operating mode into Camber whit different types of antennas
Page 62 of 74
6.4.3. Results
Table 6.4.3. Pre-Test (Before the start of stage of tests Radiated Susceptibility)
Page 63 of 74
Table 6.4.3.1. Horizontal Polarization -- from 30 MHz to 1GHz
* Problem identified RAM R-200 Detector Cable (BAK-0400). The problem fixed by adding to Cable the Ferrite Filter. (ferrite number -0444164281). Retest with Ferrite Filter: PASSED
Page 64 of 74
Graph 6.4.3.1. Horizontal Polarization -- from 30 MHz to 1GHz
-50.0
-40.0
-30.0
-20.0
-10.0
0.0
10.0
20.0
30.0
40.0
50.0
0 200 400 600 800 1000 1200
Erro
r [%
]
Frequency [MHz]
Deviation measurements while changing the frequency
UUT-1 UUT-2 UUT-3 UUT-4
Page 65 of 74
Table 6.4.3.2. Vertical Polarization -- from 30 MHz to 1GHz
Page 66 of 74
Fig. 6.4.3.2. Vertical Polarization -- from 30 MHz to 1GHz
-50.0
-40.0
-30.0
-20.0
-10.0
0.0
10.0
20.0
30.0
40.0
50.0
0 200 400 600 800 1000 1200
Erro
r [%
]
Frequency [MHz]
Deviation measurements while changing the frequency
UUT-1 UUT-2 UUT-3 UUT-4
Page 67 of 74
Table 6.4.3.3. Horizontal Polarization -- from 1 GHz to 10 GHz
Page 68 of 74
Fig. 6.4.3.3. Horizontal Polarization -- from 1 GHz to 10 GHz
-50.0
-40.0
-30.0
-20.0
-10.0
0.0
10.0
20.0
30.0
40.0
50.0
500 1500 2500 3500 4500 5500 6500 7500 8500 9500 10500
Erro
r [%
]
Frequency [MHz]
Deviation measurements while changing the frequency
UUT-1 UUT-2 UUT-3 UUT-4
Page 69 of 74
Table 6.4.3.4. Vertical Polarization-- from 1 GHz to 10 GHz
Page 70 of 74
Fig. 6.4.3.4. Vertical Polarization-- from 1 GHz to 10 GHz
-50.0
-40.0
-30.0
-20.0
-10.0
0.0
10.0
20.0
30.0
40.0
50.0
500 2500 4500 6500 8500 10500
Erro
r [%
]
Frequency [MHz]
Deviation measurements while changing the frequency
UUT-1 UUT-2 UUT-3 UUT-4
Page 71 of 74
Table 6.4.3.5. From 40 kHz to 30 MHz
Page 72 of 74
Fig. 6.4.3.5. From 40 kHz to 30 MHz
-50.0
-40.0
-30.0
-20.0
-10.0
0.0
10.0
20.0
30.0
40.0
50.0
0 5 10 15 20 25 30
Erro
r [%
]
Frequency [MHz]
Deviation measurements while changing the frequency
UUT-1 UUT-2 UUT-3 UUT-4
Page 73 of 74
6.4.3. Results:
On Frequency range 30-1000 MHz, Horizontal antenna polarization the UUT-1
(Monitor-1(RAM R-200) + Detector-8 (RP-100)) switched to Calibration mode. After Reset
UUT return to normal operation. Problem identified RAM R-200 Detector Cable (BAK-
0400). The problem fixed by adding to Cable the Ferrite Filter. (See Sec.6.3.3 and
Pic.6.3.3)
After adding to Cable the Ferrite Filter UUT-1 passed the test successfully.
All other UUT passed the test successfully.
Test Passed.
6.4.4. Conclusion
The results of this test confirm that the EUT withstands the Radiated Susceptibility
the according to the MIL-STD-462 C.
The requirement is met
6.4.5. Solution: See Sec.6.3.5.
Page 74 of 74
END OF REPORT