Orthopaedic Applications of Ferromagnetic Shape Memory Alloys

40
Orthopaedic Applications of Ferromagnetic Shape Memory Alloys MASSACHUSETTS iNSl T"UT~: OCT 2 2 2008 -. OTRARIES ----- ··-- rU1:. ~ I;.. by Weimin Guo B.Eng., Bioengineering Nanyang Technological University, 2007 Submitted to the Department of Materials Science and Engineering in partial fulfillment of the requirements for the degree of Master of Engineering in Materials Science and Engineering at the Massachusetts Institute of Technology September 2008 C 2008 Weimin Guo. All rights reserved. The author hereby grants to MIT permission to reproduce and to distribute publicly paper and electronic copies of this thesis document in whole or in part. /9A 4- Signature of author: Certified by: Accepted by: V I Department of Materials Science and Engineering June 30, 2008 Samuel M. Allen POSCO Professor of Physical Metallurgy Thesis Supervisor Samuel M. Allen POSCO Professor of Physical Metallurgy Chair, Departmental Committee on Graduate Students ARCHIVE-S

Transcript of Orthopaedic Applications of Ferromagnetic Shape Memory Alloys

Page 1: Orthopaedic Applications of Ferromagnetic Shape Memory Alloys

Orthopaedic Applications of Ferromagnetic ShapeMemory Alloys

MASSACHUSETTS iNSl T"UT~:

OCT 2 2 2008

-.OTRARIES-----··--rU1:. ~ I;..

by

Weimin Guo

B.Eng., BioengineeringNanyang Technological University, 2007

Submitted to the Department of Materials Science and Engineeringin partial fulfillment of the requirements for the degree of

Master of Engineering in Materials Science and Engineeringat the

Massachusetts Institute of Technology

September 2008

C 2008 Weimin Guo. All rights reserved.

The author hereby grants to MIT permission to reproduce and to distribute publicly paper andelectronic copies of this thesis document in whole or in part.

/9A 4-

Signature of author:

Certified by:

Accepted by:

V I

Department of Materials Science and EngineeringJune 30, 2008

Samuel M. AllenPOSCO Professor of Physical Metallurgy

Thesis Supervisor

Samuel M. AllenPOSCO Professor of Physical Metallurgy

Chair, Departmental Committee on Graduate Students

ARCHIVE-S

Page 2: Orthopaedic Applications of Ferromagnetic Shape Memory Alloys

Orthopaedic Applications of Ferromagnetic ShapeMemory Alloys

by

Weimin Guo

Submitted to the Department of Materials Science and Engineeringon June 30, 2008 in partial fulfillment of the requirements for the degree of

Master of Engineering in Materials Science and Engineering

Abstract

Ferromagnetic shape memory alloys (FSMAs) are a new class of magnetic field-actuated active

materials with no current commercial applications. By applying a magnetic field of around 0.4 T,

they can exert a stress of approximately 1.5 MPa, exhibiting a strain of up to 6%. This thesis

evaluates their technical and commercial feasibility in orthopaedic applications. Remote

actuation is a key advantage FSMAs have over current implant materials. Also, the human body

temperature is constant, providing a stable environment for FSMAs to operate.

A number of potential orthopaedic applications are proposed and evaluated. Out of these, the

most prominent application is the spinal traction device. It is a temporary implantable device,

intended to perform internal spinal traction. A design has been proposed, with suggestions of

suitable materials for its various components and appropriate device dimensions. Preliminary

market and cost analyses have been conducted. This orthopaedic technology is currently in its

infant stage. To commercialize this device, more trials are needed.

Thesis Supervisor: Samuel M. AllenTitle: POSCO Professor of Physical Metallurgy

Page 3: Orthopaedic Applications of Ferromagnetic Shape Memory Alloys

Acknowledgements

I would like to thank Prof Sam Allen for being there for me every week and devoting a huge

amount of his time and attention guiding this project. He is the type of supervisor most of my

fellow students would be envious of having. Next, I would like to thank Dr Bob O'Handley for

providing wonderful insights on the technical aspects of this project, despite retiring from MIT.

Also, I would like to express my appreciation for Prof James Goh, of the National University of

Singapore, who suggested an important potential application of FSMAs and guided me through

the physiological aspects of this project.

I am also grateful to the following professors and surgeons who spent time answering my queries

and contributed to my overall understanding of this project:

* Prof Wong Hee KitHead of Department of Orthopaedic Surgery, NUS School of Medicine

* Prof Casey ChanDivision of Bioengineering, NUS

* Assoc Prof Daniel John BlackwoodDepartment of Materials Science and Engineering, NUS

* Assoc Prof Fuss Franz KonstantinDivision of Bioengineering, NTU

* Dr Ang Kian ChuanConsultant Sports & Orthopaedic Surgeon, Gleneagles Hospital

* Dr Ngian Kite SengConsultant Orthopaedic Surgeon, Island Orthopaedic Group

* Dr Wong Yue ShuenConsultant Sports & Orthopaedic Surgeon, Consultant Foot & Ankle Surgeon, IslandOrthopaedic Group

Lastly, I would like to thank all my fellow students for making the past year such a wonderful

and unforgettable one.

Page 4: Orthopaedic Applications of Ferromagnetic Shape Memory Alloys

Table of Contents

Abstract 2

Acknowledgements 3

1. Overview of Ferromagnetic Shape Memory Alloys 6

1.1 Introduction 6

1.2 Background of Ferromagnetic Shape Memory Alloys 7

2. Potential Orthopaedic Applications 11

2.1 Search for Applications 11

2.2 Bone Fixation Devices 11

2.2.1 Cost 12

2.2.2 Potential Market of Enhanced Bone Fixation Devices 13

2.2.3 Supply Chain Analysis 13

2.2.4 Competing Technologies 14

2.3 Fracture Union Detector 15

2.4 Magnetic Field Exposure Concerns 16

3. Traction for Spinal Disorders 18

3.1 Spinal Traction Overview 18

3.2 Proposed Treatment: FSMA Spinal Traction Device 19

3.3 Competing Technologies 23

3.4 Market Analysis 24

3.5 Cost Modeling 24

3.6 Intellectual Property: Patents 27

4. Future Direction 30

4.1 Issues and Risks 30

4.2 In Vitro Compatibility Testing 31

4.2.1 Pitting Corrosion Test 33

4.2.2 Weight Loss Test and Trace Nickel Detection 34

Conclusions 35

References 36

Page 5: Orthopaedic Applications of Ferromagnetic Shape Memory Alloys

Table of Figures

Figure 1: Photo of Ni 49.4Mn 29.7Ga 20.9 at room temperature in zero external field 6

Figure 2: Application of magnetic field induces twin boundary movement 8

Figure 3: Magnetic field-induced transformation of austenite to martensite 8

Figure 4: General purpose plating for a variety of midfoot and rearfoot surgical procedures 12

Figure 5: Collaboration with existing manufacturing companies 14

Figure 6: Diagram of fractured bone fixated with FSMA bone plate 16

Figure 7: Encapsulated FSMA 20

Figure 8: Diagram of the FSMA spinal traction device 20

Figure 9: Top view of a titanium casing 21

Figure 10: Technological barriers of FSMA orthopaedic technology 31

Index of Tables

Table 1: Global sales summary of Synthes in 2007 13

Table 2: Compositions of the main alloys used in orthopaedic prostheses 32

Page 6: Orthopaedic Applications of Ferromagnetic Shape Memory Alloys

1. Overview of Ferromagnetic Shape Memory Alloys

1.1 Introduction

This thesis evaluates the technical and commercial feasibility of ferromagnetic shape memory

alloys (FSMAs) in orthopaedic applications. A photo of a Ni-Mn-Ga single crystal (a member of

the FSMA family) is shown in Figure 1. The background of the development of such alloys is

provided in the following section. Their properties and capabilities are briefly discussed. Some

potential orthopaedic applications of these alloys are proposed and analyzed in this thesis.

During the author's search for viable orthopaedic applications, a key application identified is the

incorporation of FSMAs into spinal implants. An overview of spinal problems is given in

Chapter 3, which also introduces the means by which FSMAs may be utilized to mitigate and

possibly cure these disorders. Besides these, competing technologies, supply chain, and

intellectual property issues are analyzed. Risks and the future direction are outlined at the end.

2.6 cm

Figure 1. Photo of single-variant sample of Ni49 .4 Mn 29 .7Ga20.9 at room temperature in zero external field

[1].

Page 7: Orthopaedic Applications of Ferromagnetic Shape Memory Alloys

1.2 Background of Ferromagnetic Shape Memory Alloys

Ferromagnetic Shape Memory Alloys (FSMAs) officially became recognized as a new class of

active materials when a 0.2% magnetic-field-induced strain in a single crystal of Ni2MnGa was

observed in 1996 [2]. This was only a fraction of the 6 to 7% strain ultimately attainable by field-

induced twin selection in such alloys [3], which can exert a stress of 1 to 2 MPa as one twin

variant is completely converted to another [4,5]. This strain is reversible upon the application of

a magnetic field in an orthogonal direction [5]. Comparatively, other active materials exhibit

much lower strains. Piezoelectric materials show strains on the order of 0.1%, while the leading

magnetostrictive material, Terfenol-D, exhibits a field-induced strain of about 0.24% [6].

FSMAs exhibit martensitic phase transformation, allowing control of large strains by magnetic

field application. The field-induced strain is the result of twin boundary motion in the martensitic

phase. It can be induced by either magnetic field or mechanical stress. These large strains make

FSMAs suitable for actuator and sensor applications. Although other members of this family of

materials, including Co2MnGa, FePt, CoNi and FeNiCoTi, have been explored during the past

few years [5], Ni-Mn-Ga alloys remain the best-performing room-temperature candidate.

The strain phenomenon of Ni-Mn-Ga results from twin-boundary movement in martensitic

single crystals. The magnetic easy axis is in the same direction as the c-axis of the tetragonal

martensitic unit cell. This axis direction is different for each of the twin variants. For martensitic

crystals with large uniaxial magnetic anisotropy, applying a magnetic field parallel to the easy

axis of one variant would destabilize the others present [7]. This process is shown in Figure 2.

The minimum required magnetic field to initiate twin boundary motion is typically 0.4 T [8].

Any field beyond this can induce the twin boundary to move such that the volume of the variant

which best aligns with the applied field is increased. The magnetic field-induced austenite to

martensite transformation is illustrated in Figure 3, which shows how an increasing field or stress

can induce the transformation.

Page 8: Orthopaedic Applications of Ferromagnetic Shape Memory Alloys

' IiH

+41

Figure 2. Application of magnetic field in the same direction as the easy axis of one twin variant induces

the twin boundary to move such that the volume of that variant is increased [7].

Magnetic field-inducedA->M transformation

r------I-

I Ir - - - -- - - - -- -

M I

I--- -- -- -- - ---

PA

stenite

C~Figure 3. Top: Magnetic field-induced transformation of an austenitic sample to a single-variant

martensitic structure. Bottom: The A-M phase boundary is shown below in field-stress-temperature space;

the arrow indicates how an increasing field can induce the transformation (horizontal e-H path). This

process may also be assisted by the application of a suitable stress (vertical e-o path) [1].

~c;i Ihrr

;5. .·:·

·'·

r~:l.~.a ::I·~~~

G1

Page 9: Orthopaedic Applications of Ferromagnetic Shape Memory Alloys

In order to exhibit magnetic field-induced strain, four essential material property criteria need to

coexist [4]. Firstly, a low-symmetry phase, martensite, must exist under actuation conditions for

twinning deformation to occur. For tetragonal martensites, the crystallographic ratio c/a

determines the maximum twinning strain. Next, the twin-boundary yield stress must be

sufficiently low to allow the magnetic moments of the material to maneuver the twin defect

structure. Also, the material has to be ferromagnetic with a large saturation magnetization. This

magnetization magnitude plays a role in generating sufficient magnetic forces within the material.

Finally, the magnetocrystalline anisotropy must be sufficiently large along a particular

crystallographic axis of easy magnetization, rather than in a plane.

For martensitic Ni-Mn-Ga, large magnetocrystalline anisotropy causes magnetic moment

alignment with the c-axis on either side of the twin-boundary. The magnetic energy stored as a

result of hard-axis magnetization exerts pressure on the twin-boundary. This occurs until the

twin-boundary yield stress is reached and twin-boundary motion occurs. If twin boundaries are

pinned on defects, which prevent the former from sweeping through the whole alloy, actuation

stops and further energy storage occurs.

The aforementioned twin-boundary-driven strain effect has only been observed in single crystals,

which are brittle and expensive to grow [9]. Under uniaxial loading, however, cyclic magnetic

field actuation has demonstrated hundreds of thousands of strain cycles without fracture [4].

These factors limit their commercial usefulness. Therefore, there is current research on

composites fabricated from Ni-Mn-Ga powder mixed with polymer [10]. Such composites are

easier to make and are more economical. Also, they are capable of withstanding higher stresses

in both compression and tension than the single-crystal form. The viability of such composites

may be pivotal to the commercial success of FSMAs.

Due to the large linear deformation-induced change in magnetization, FSMAs can be

implemented into stress sensors. They have been explored as possible magnetoelectric converters

[11], although the conversion efficiency is currently too low to be commercially feasible. There

has also been an interest in the thin film technology of Ni-Mn-Ga alloys for micro- and nano-

system applications [12, 13].

Page 10: Orthopaedic Applications of Ferromagnetic Shape Memory Alloys

In the field of active flow control technology, FSMAs are being researched as actuators in

synthetic jet applications [14]. To produce string synthetic jet flow at high frequency, a new

membrane actuator based on an FSMA composite and hybrid mechanism was designed and

constructed. Due to the large force and martensitic transformation on the FSMA composite

diagram, the membrane actuator designed can produce 190 m/s synthetic jets at 220 Hz.

Although FSMAs have been widely investigated in the laboratory, little research has been

conducted to apply the unique properties of FSMAs to biomedical devices. The following

chapter analyzes some potential orthopaedic applications of these alloys.

Page 11: Orthopaedic Applications of Ferromagnetic Shape Memory Alloys

2. Potential Orthopaedic Applications

2.1 Search for Applications

Since the first demonstrated magnetic-field-induced strain in 1996 at MIT [2], there has been

significant interest in FSMAs from the scientific community from a materials research point of

view. However, there is yet to be any large-scale commercial application. The United States

Navy has experimented with them as the active material for sonar transducers, but has achieved

limited success [15].

The author believes that FSMAs have great potential to be used as a medical implant material.

This is because of two reasons. Firstly, the human body is a constant temperature environment,

which is very stable for the operation of FSMAs. Secondly, FSMAs can be remotely actuated by

a magnetic field. Such remote actuation means that the conformation of implants may be

remotely altered after surgery. However, there is no current research on the orthopaedic

applications of FSMAs. Thus, it has been the author's task to identify and evaluate various

potential applications since the commencement of this project.

2.2 Bone Fixation Devices

One such possible application lies in the field of bone fixation. Bone fixation devices are tools

which hold fractured bones in place to enable them to heal. They include bone plates, rods and

screws. An initial compression is usually applied on the fracture site by these devices because

compressive stresses promote faster healing. This is referred to as compression healing, which is

only suitable for simple fractures, where the cortex of the bone is still intact, with good cortical

contact [16]. For convoluted fractures, where the bone has been broken into several pieces,

compression cannot be used because the bones will otherwise collapse.

Simple fractures usually heal well using current bone plates and screws [16, 17]. Difficult

fractures may occur for cases of high trauma, such as for high impact fractures. For such

fractures, bone loss may result at the fracture site. The osteocytes at the connecting ends of the

bones may die off, leaving a gap. Such cases are known as delayed unions or non unions, which

Page 12: Orthopaedic Applications of Ferromagnetic Shape Memory Alloys

may not heal even with the aid of current bone plates. In these situations, the usual course of

action taken by surgeons is to conduct another surgery and use bone grafting [16].

- *ehs: 1 4, 1 s4d4 synmm- 5 enohs., 12, 14, 2X, 24 and 3Omm

Figure 4. General purpose plating for a variety of midfoot and rearfoot surgical procedures [ 18]. Such

plating is useful for isolated tarsal fusions and for Evans lengthening procedures with interpositional bone

grafts.

This is a field where FSMAs can play a useful role. Figure 4 shows a conventional bone plate

being implanted during a foot surgical procedure. FSMAs may be incorporated into the inner

shaft of the bone plate, so that medical personnel can induce further compression externally to

stimulate healing for difficult fractures, negating the need for additional surgery. Whenever

surgeons identify high-impact fractures, which are likely to result in delayed or non unions, they

can implant bone plates with FSMA enhancement. These can be furthered compressed after

surgery for treatment of the fractures. Cost would play a large role in determining the popularity

of such enhanced bone plates in hospitals. Thus, a preliminary estimate of how much more

expensive such plates would be, compared to conventional ones, is important in determining

their viability.

2.2.1 Cost

Since there has not been any large-scale commercial production of FSMAs, it is challenging to

obtain an accurate prediction of the cost of such FSMA-enhanced bone plates. MIT obtains

FSMA crystals for research purposes from the Ames Laboratory, United States Department of

Page 13: Orthopaedic Applications of Ferromagnetic Shape Memory Alloys

Energy. A single Ni-Mn-Ga crystal, 7.5 cm in length with a diameter of 2.5 cm, costs

approximately $2,000 [19]. A preliminary estimate of the cost of FSMA incorporation would be

close to at least an additional $800 per device. It is reasonable to assume that this cost would

drop as FSMAs are mass produced. Also, the cost of the external electromagnet has to be

factored into the overall cost of treatment. However, the electromagnet can be re-used and shared

among several patients. Thus, its cost can also be distributed.

2.2.2 Potential Market of Enhanced Bone Fixation Devices

The global demand for bone fixation devices is rapidly growing. Synthes, a leading global

medical device company, currently holds a 50% share of the US orthopaedic implant market [20].

Specializing in fabricating bone plates, rods and screws, its global net sales was close to 2.8

billion dollars in 2007. Its growth of more than 15% from the preceding year shows that the

market is rapidly expanding. A summary of its global sales is provided in Table 1. Although this

market size appears promising, FSMA-enhanced bone fixation devices would take up only a

niche segment of it, probably in the domain of difficult fracture treatment.

In US$ millions Full Year 2007 (January - December)

Consolidated % Change % ChangeNet ales 2007 2006 (in US) (in local

currencies)'

North America 1,721,0 1,5251 12.8% 12.6%

Europe 637.3 519.1 22.8% 13.4%

Asia Pacific 248.4 220.1 12.8% 8.7%

Rest of the World 153.0 127.3 20 3% 13.9%

Total . 2,759.7 2,391.6 15.4% 12.5%

" Local currencies - 2007 results translated at 2006 forgn exchange rates.

Table 1. Global sales summary of Synthes in 2007. The net sales have grown by more than 15% from theprevious year [20].

2.2.3 Supply Chain Analysis

In order to bring FSMA-enhanced fixation devices into the market, collaboration with existing

device manufacturers is the most cost-effective route to take. AdaptaMat is a company in Finland,

13

Page 14: Orthopaedic Applications of Ferromagnetic Shape Memory Alloys

specializing in the growth of FSMA crystals and actuators. One strategy is to obtain appropriate

patents and license them out to companies like AdaptaMat. A better way would be to collaborate

with AdaptaMat to produce specific FSMA components suitable for incorporation into bone

fixation devices. Such sharing of expertise and facilities is helpful to producing top quality and

low-cost FSMA products. The components could then be sold to large medical implant

companies such as Synthes, which would incorporate them into some of their fixation devices. It

is not cost-effective to produce entire implants solely by a new start-up. Major implant firms like

Synthes run hundreds of product lines. Thus, from an economy-of-scale perspective, it is unwise

for a start-up to run entire production lines just to produce a few types of products. Therefore,

smooth virtual vertical integration with companies like AdaptaMat and Synthes is the best route

to commercialization of the products. Such collaboration is illustrated in Figure 5.

SSynthes Hospitals

OurResearch AdaptaMatGroup

Figure 5. Collaboration with existing manufacturing companies is a cost-effective route to bring the

products into the market. Mutual exchange of expertise is helpful in improving product quality and

lowering cost.

2.2.4 Competing Technologies

In the treatment of difficult fractures, the main competing technology is bone grafting. The cost

of a bone graft ranges from $250 to $900 [21], excluding other hospital charges such as surgeons'

fees and equipment costs. This is relatively inexpensive, compared to FSMA-enhanced bone

plates. In a study of over 1000 patients who received very large allografts after surgery for bone

cancer, 85% of them were able to return to work or normal physical activities without using

clutches [21]. Less severe bone defects, such as fractures, should heal completely without serious

complications. This shows that bone grafting is an effective treatment for difficult fractures.

Page 15: Orthopaedic Applications of Ferromagnetic Shape Memory Alloys

When treating difficult fractures, surgeons may prescribe the use of bone grafts directly. Thus,

FSMA-enhanced bone fixation devices do not achieve a truly unique treatment purpose.

Compared to its main competitor, bone grafting, FSMA bone fixation is not cheaper. Also, bone

grafting is effective in treating difficult fractures. The disadvantages of FSMA bone fixation

outweigh its merits. Therefore, it is not advisable to continue with the development of FSMA

bone fixation.

2.3 Fracture Union Detector

Knowing exactly when a fracture has healed optimizes treatment for the patient. It provides an

indication of when various rehabilitation programmes can be initiated [22]. Studies of fracture

treatment often rely on radiographically defined end-points to make treatment recommendations

[23]. Since union is considered the end-point of fracture healing, it must be accurately defined.

Its clinical definition must be precise enough to avoid both over- and under-treatment. Currently,

it is challenging to use radiographs for defining fracture union in conservatively treated fractures

[24, 25]. Clinical assessment of such union is also unreliable [26].

Internal fixation of fracture affects its mechanical and biological environment. Thus, the amount

of callus formation and the extent of cortical bridging may be affected. The definition of union

by radiographs depends on these two factors. In a recent study of fracture union, 47 radiographic

series of forearm, femoral and tibial fractures treated using internal fixation over a 3-year period

were reviewed [27]. Callus formation and fracture line filling with time were measured on each

radiograph. The conclusion of the study was that radiographs do not define union in internally

fixated fractures with sufficient accuracy to enable their use as end-points of fracture healing.

Thus, radiographic end-points are not sufficiently dependable for treatment recommendations.

Therefore, there is a need for a more accurate method of fracture union detection.

This is an area where FSMA-enhanced bone fixation devices may play a part. Figure 6 shows a

diagrammatic side view of a fracture fixated by a bone plate with an FSMA component as part of

its shaft. As the fracture heals, callus formation at the interface of the fracture shares more of the

load originally borne by the bone plate. To quantify the amount of callus formation, and thus the

extent of fracture healing, an approach is to detect the stiffness of the fracture region. With the

Page 16: Orthopaedic Applications of Ferromagnetic Shape Memory Alloys

FSMA-enhancedbone plate Screws

\ /I

Fractured bone Region of callusformation

Figure 6. Diagrammatic side view of a fractured bone fixated by a bone plate with an FSMA component

as part of its shaft. Note the parallel load-bearing arrangement between the fractured bone and the bone

plate.

patient at rest and hence, in the absence of external forces, apply a time-varying magnetic field

on the bone plate. The strain response of the bone plate would provide an indication of how stiff

the fracture region is. The stiffer the fracture region, the more the strain response of the bone

plate would be impeded, indicating more callus formation and better fracture healing. Thus, the

progress of fracture healing and its end-point can be determined with FSMA-enhanced bone

plates. Care has to be taken such that these assessments of the progress of fracture healing do not

have detrimental effects on the healing itself. The author feels that this potential application

deserves further exploration.

Another promising orthopaedic application of FSMAs is in the treatment of spinal problems.

This concept is attributed to Professor James Goh and his research group at the National

University of Singapore. It is discussed in detail in the following chapter.

2.4 Magnetic Field Exposure Concerns

Since the actuation of FSMA devices in the body would inevitably entail human exposure to

magnetic fields, it is necessary to check whether such exposure would involve any risks or cause

long term effects. Some of the established effects of exposure to static magnetic fields include

Page 17: Orthopaedic Applications of Ferromagnetic Shape Memory Alloys

vertigo, nausea and phosphenes, as a result of movement in the field [28]. The magnitude of the

magnetic field required to actuate the FSMA component is up to 0.4 T. To achieve this value at

the site of the FSMA component, higher field levels may be inflicted upon the regions of the

body nearer to the external magnet. Thus, a safety margin is needed.

According to the International Commission on Non-Ionizing Radiation Protection, the ceiling

value of static magnetic field exposure is 2 T [28]. The magnetic field level required for the

actuation of FSMA devices is significantly lower than this value. Also, the exposure limit

averaged for a working day is 0.2 T. This average value is not a cause for concern, since patient

exposure to magnetic field is only needed for actuation to occur and hence, the exposure time is

very short.

Page 18: Orthopaedic Applications of Ferromagnetic Shape Memory Alloys

3. Traction for Spinal Disorders

3.1 Spinal Traction Overview

Since at least 1800 BC, traction has been used to treat spinal disorders [29]. By the 19th century,

the traction bed was used to treat scoliosis, backache, rickets and spinal deformity [30]. During

the last century, traction became a common treatment technique for chronic low back pain,

although opinions differ regarding how traction should be applied [31]. Currently, it is still

popular as a treatment technique for lumbar spine conditions, being used in approximately 7% of

physiotherapy sessions in The Netherlands [32]. It is most commonly used for the relief of pain,

normalization of neurological deficits or painfully restricted neuromeningeal tension signs [33,

34], and for improving joint mobility [35]. It has also been promoted for lumbar disc lesions,

relying on the theory that it would produce negative pressure in the disc and thus reduce disc

herniations [36].

Traction could improve signs and symptoms of spinal disorders by biomechanical effects,

primarily the separation of the intervertebral motion segments [37]. This mechanical vertebral

separation may induce neurophysiological changes which result in pain reduction. It may provide

relief from radicular symptoms by removing direct pressure or contact forces from sensitized

neural tissue [38].

Axial lumbar traction has been reported to increase intervertebral disk height [39, 40]. In the area

of stenosis relief, it has also been shown to produce an increase in bony foraminal area, although

the central canal area remains unchanged [41]. It is believed that any force or movement that

causes a measurable anatomical increase in central or lateral canal size will result in

physiological neural decompression at least temporarily [42]. The apparent lack of a dose-

response relationship suggests that low doses may be sufficient to achieve benefit [38].

By its non-surgical nature, traction is an externally administered technique which is unable to

target specific problematic regions of the spinal vertebrae. Current traction techniques are unable

to limit their effects to targeted treatment regions. The applied traction forces are distributed over

a large portion of the spinal column and inadvertently affect other healthy spinal regions. This

Page 19: Orthopaedic Applications of Ferromagnetic Shape Memory Alloys

may reduce the forces acting on the targeted region, as well as the displacement experienced by

it.

After discussion with Prof James Goh of the National University of Singapore, the author

proposes the design of an FSMA spinal traction device which could cater to this issue. Such a

device is intended to be temporarily implanted, for up to a few months, to achieve regulated and

specific traction from within the body. Its purposes are to provide pain relief and promote tissue

recovery. Similar to current traction protocols, the FSMA device can be actuated to provide

traction for a fixed number of hours every day. The amount of force and displacement required

by the treatment can be regulated by adjusting the externally applied magnetic field. This device

is designed to be positioned in between two spinous processes and to exert a force to push the

processes apart when actuated, resulting in separation of intervertebral segments. The following

section describes this proposed device in detail.

3.2 Proposed Treatment: FSMA Spinal Traction Device

The author has undertaken the task of designing an FSMA device which can fit into the confines

of the space between two spinous processes, exert sufficient traction force and exhibit the

required displacement. Schematics of the proposed design are presented in Figures 7, 8 and 9.

To make an FSMA component, a piece of FSMA (4 cm x 1 cm x 1 cm) is cut into 10 equal slices

for better actuation. To minimize biocompatibility issues with the body, it is encapsulated with a

rigid polymer cap and a flexible polymer cover, as shown in Figure 7. The rigid cap has to

possess sufficient load-bearing capability, while the flexible cover must not hinder the extension

of the FSMA.

Page 20: Orthopaedic Applications of Ferromagnetic Shape Memory Alloys

1 cm

4cm

- Rigid cap: SIBS

-. Flexible coverPDMS or PET

FSMA (cut into10 slices)

Figure 7. Encapsulated FSMA. Each FSMA component is encapsulated by rigid caps on both ends and by

a flexible cover in the middle, as shown in this figure. The rigid cap can be made of poly(styrene-block-

isobutylene-block-styrene) (SIBS). The flexible cover can be made of either polydimethylsiloxane

(PDMS) or polyethylene terephthalate (PET). The dimensions of the FSMA component are 4 cm x 1 cm x

1 cm. It is cut into 10 equal slices for better actuation.

Spinous process

Titanium casing

EncapsulatedFSMA component

Figure 8. Diagram of the FSMA spinal traction device (side view). Two encapsulated FSMA components

would be slotted into two titanium casings and positioned in between two spinous processes of the body.

External magnetic actuation of the FSMA components would push the spinous processes further apart

from each other.

Page 21: Orthopaedic Applications of Ferromagnetic Shape Memory Alloys

Slots to insert encapsulatedFSMA components

/

Figure 9. Top view of a titanium casing. One FSMA component would be inserted into each of the slots

shown. When the FSMA components have been sandwiched in between two titanium casings, they would

be prevented from falling out of the spinal traction device by the side barriers of the casings.

The proposed candidate for the rigid cap is poly(styrene-block-isobutylene-block-styrene)

(SIBS). The development of SIBS for medical devices evolved from deficiencies encountered

with the long-term in vivo performance of polyurethanes, such as their degradation and resulting

inflammatory and fibrotic reactions [43]. These deficiencies limited the use of polyurethanes for

long-term load-bearing implant applications and for applications in contact with a metal, as metal

ions contribute to polyurethane degradation by oxidative pathways. SIBS has been developed to

overcome these deficiencies. The inertness of SIBS has enabled some novel medical devices,

including a synthetic trileaflet aortic valve to possibly replace tissue and mechanical valves.

Several other medical devices made from SIBS, including spinal implants, are in their early

stages of development and may be commercialized over the next few years. From the processing

perspective, SIBS can be injection and compression molded, as well as extruded and solvent cast

from non-polar solvents.

However, there are some drawbacks to the thermoplastic nature of SIBS [43]. Firstly, it is

susceptible to stress cracking in the presence of organic solvents. Also, it has poor creep

properties and therefore requires fiber reinforcement for certain load-bearing applications. The

non-polar nature of SIBS does not allow sites for hydrogen bonding, unlike polyurethanes, thus

limiting its ultimate tensile strength to less than that of dry polyurethane. SIBS has poor gas

Page 22: Orthopaedic Applications of Ferromagnetic Shape Memory Alloys

permeability, which renders it more cumbersome to sterilize with ethylene oxide. All surfaces

must be exposed to this gas. Furthermore, SIBS is not gamma-ray sterilizable. At the present

production rate, the cost of synthesis and purification of SIBS is relatively high. Finally, due to

intellectual property constraints, there is no source for implant-grade SIBS. To make use of SIBS,

an agreement would have to be reached with Boston Scientific Corporation and Innovia LLC.

This intellectual property issue is further discussed later in this chapter. Despite these drawbacks,

the author believes SIBS is still the best candidate as the material for the rigid cap.

For the flexible cover, the author proposes the use of either polydimethylsiloxane (PDMS) or

polyethylene terephthalate (PET). PDMS, commonly known as silicone rubber, has been proven

to be very stable in contact with body fluids over a period of up to 2 years [44, 45]. However,

PDMS heart valve failures have been reported and are believed to be caused by the absorption of

blood lipids into the implant over years [46]. Although PDMS may suffer from property changes

after implantation over time [47], the spinal traction device is intended to be a temporary implant

of at most a few months. Hence, the author believes that lipid absorption will not be an issue here

and that PDMS is still a viable candidate as the flexible cover material.

Besides PDMS, PET may also be a good material choice for the flexible cover. At present, it is

the preferred material for most large-caliber textile vascular grafts [48]. Being reasonably inert,

flexible, resilient, durable and resistant to biological degradation, it is one of the two materials

that withstood the test of time and are still used for vascular graft applications (the other being

polytetrafluoroethylene PTFE). Further research may be needed to determine the final choice of

material for the flexible cover.

After discussing the material selection, it is important to perform a few basic calculations to

determine whether the device dimensions are suitable for its required function. For a person

lying on his or her back, the amount of force needed to distract adjacent spinous processes is

approximately 15% of the body weight [22]. For a person weighing 800 N, that force would be

equivalent to 120 N.

For each FSMA component, the relevant cross-sectional area is I cm x 1 cm = 0.0001 m2

Page 23: Orthopaedic Applications of Ferromagnetic Shape Memory Alloys

Taking the maximum stress exerted by the FSMA component to be 1.5 MPa [4, 5], the maximum

force which can be exerted by each FSMA component would be 150 N. Since each spinal

traction device has two FSMA components, the maximum force which can be exerted by each

device would be 300 N. This upper limit is significantly higher than the 120 N required to

distract adjacent spinous processes of a patient lying down.

As shown in Figure 7, each FSMA component is 4 cm long. Taking its maximum strain to be 6%,

the device is capable of achieving a displacement of 2.4 mm. With data from future research, this

FSMA component length may be modified to give the required displacement values.

3.3 Competing Technologies

One obvious competitor with the FSMA spinal traction device is conventional traction treatment.

The current typical cost for a session of traction therapy in the United States is between $50 and

$100 [49]. A popular form of it is the vertebral axial decompression (VAX-D) system. By its

very nature, it is non-invasive and does not require hospitalization, giving it a strong advantage

over the FSMA spinal traction device. It claims to be successful in over 70% of patients,

providing relief of acute or chronic low back pain and associated leg pain generally within a

month [50]. A recent study was conducted to verify its effectiveness [51]. Its participants

consisted of 296 patients with low back pain and evidence of a degenerative or herniated

intervertebral disk at one or more levels of the lumbar spine. The intervention was an 8-week

course of prone lumbar traction, using the VAX-D system, consisting of five 30-minute sessions

a week for 4 weeks, followed by one 30-minute session a week for 4 additional weeks. The

system has shown to reduce pain intensity, although the results are not fully conclusive. Since

the FSMA spinal traction device is able to target specific areas of the spinal column, its

performance is expected to be better.

Current surgical treatment options, such as spinal fusion and artificial disc replacement, involve

passive and permanent implants. Also, spinal fusion permanently limits spinal motion. These

implants cannot be actuated externally to induce tissue repair and recovery, which is the premise

of the FSMA spinal traction device. The latter is intended as a temporary treatment device, and

would be removed from the body after recovery.

Page 24: Orthopaedic Applications of Ferromagnetic Shape Memory Alloys

3.4 Market Analysis

This section attempts to predict the level of market demand for the FSMA spinal traction device.

It would be erroneous to make such a prediction based on the number of patients undergoing

current conventional spinal traction procedures. This is because patients are likely to pursue non-

invasive options first and only consider surgery after such options failed to work. One approach

would be to evaluate the potential market based on the number of patients undergoing surgery

for related spinal disorders.

Lumbar spinal surgery is one of the most common types of surgeries performed in the United

States with over 500,000 surgeries performed for lumbar herniated disks and lumbar spinal

stenosis annually [52]. Assuming that the FSMA spinal traction device gains a 5% penetration of

this current market, it would be equivalent to a demand for 25,000 units per year. The actual

market share of the device will depend on how effective the treatment is proven to be and the

cost of the entire treatment compared to current and emerging alternatives, as well as the

surgeons' and patients' receptivity to this new form of treatment. One factor which is likely to

keep this market share low is the fact that such implants are temporary and non-bioresorbable,

necessitating an additional surgery for their removal.

3.5 Cost Modeling

This section attempts to do a preliminary analysis on how much each unit of the spinal traction

device would eventually cost when it is out in the market. This is important in determining its

competitiveness and market penetration.

The eventual unit cost depends heavily on the supply chain pathway taken to make the device.

The supply chain in this case is similar to the one discussed in Section 2.2.3, which is for the

supply chain of bone fixation devices. The strategy would be to set up a company to manufacture

the FSMA components used in the spinal traction device, with involvement from experienced

FSMA companies like AdaptaMat. The FSMA components would then be sold to large medical

implant firms, such as Synthes, which would manufacture the titanium casings and in turn

introduce the final product to hospitals and medical centers. This virtual vertical integration of

the supply chain is intended to keep the manufacturing costs at the minimum. The distinctive

Page 25: Orthopaedic Applications of Ferromagnetic Shape Memory Alloys

feature in this case is that involvement of Boston Scientific Corporation is needed, because it

holds the intellectual property rights to SIBS, which is the chosen material for the rigid

encapsulation caps. The following is a brief analysis of the costs of the device materials.

Poly(dimethylsiloxane)

Although both poly(dimethylsiloxane) (PDMS) and poly(ethylene terephthalate) (PET) are being

considered as candidates for the FSMA flexible cover material, PDMS will be used here for the

cost modeling analysis. PDMS can be obtained from Sigma-AldrichTM at $208.50 for 250 ml

[53], which should be sufficient for encapsulating 5 FSMA sticks. Since each device comprises 2

sticks, it requires 100 ml. Thus, the cost of PDMS for each device is $83.40.

Poly(styrene-block-isobutylene-block-styrene)

Leonard Pinchuk of Innovia LLC patented poly(styrene-block-isobutylene-block-styrene) (SIBS)

in 2003 [54]. The SIBS intellectual property is currently owned by Boston Scientific Corporation

and a license is required for its use [55]. It has been mentioned that the current cost of SIBS is

relatively high [43]. Therefore, for the purpose of this cost analysis exercise, the cost of SIBS is

taken as $166.80, twice that of PDMS for each component.

Ni-Mn-Ga Crystal

As mentioned in Section 2.2.1, there is no current large-scale production of FSMAs. Thus, it is

challenging to obtain an accurate prediction of the cost of such FSMA components. MIT obtains

FSMA crystals for research purposes from the Ames Laboratory, United States Department of

Energy. A single Ni-Mn-Ga crystal, 7.5 cm in length with a diameter of 2.5 cm, costs

approximately $2,000 [19]. The usable proportion of such crystals is 50 - 75%. Taking this into

consideration, each such crystal should be sufficient for making 4 FSMA components. Since

each spinal traction device requires 2 FSMA components, the cost of FSMA for each device

currently stands at $1,000. The author believes that with mass production and further research,economy of scale may bring the production cost of FSMA for each device down to

approximately $500.

Page 26: Orthopaedic Applications of Ferromagnetic Shape Memory Alloys

Titanium

Titanium can be obtained from Sigma-AldrichTM at $249 for 100 g [56]. Taking into

consideration the process yield, it is expected that approximately 100 g of titanium would be

needed for each device.

Electromagnet

The cost of the electromagnet will not be added here for the computation of the total production

cost of each device. One of the main reasons is that the electromagnet can be shared amongst

several patients, and re-used for new patients. The strategy to adopt would be to license out the

intellectual property to a third party company, which has the expertise of manufacturing general

electromagnets, for it to make electromagnets compatible with the FSMA device. This company

would then sell its electromagnets to hospitals, which would rent them out to patients. Thus,

from the perspectives of patients, the electromagnet cost comes in the form of rental fees of

hospital equipment. Even though the electromagnet cost is to be spread out via the

aforementioned system, it is imperative to keep the final overall cost low to maintain the

competitiveness of this treatment approach.

Cost Computation

Total cost of materials = $83.40 + $166.80 + $500 + $249

= $999.20

Assuming that the processing and sterilization costs add up to $500, the total cost for each device

would be $1499.20.

The effect of production yield has to be taken into consideration to give a more accurate figure

for the production cost. The overall yield Y of all the processes can be computed by:

Y= YI * Y2 * Y3 * ...... * YN, where N is the number of processing steps.

For simplicity of computation, the overall yield Y is taken as 80 %.

After accounting for yield, the actual production cost of each device is $1874. This amount can

be further reduced if the production costs of FSMA and SIBS become much lower as their

Page 27: Orthopaedic Applications of Ferromagnetic Shape Memory Alloys

manufacturing technology becomes more mature. Also, if the demand is high, economy of scale

can push the cost down further.

3.6 Intellectual Property: Patents

The number of patents related to FSMAs is limited. Currently, there are no patents for

orthopaedic applications of FSMAs. The most relevant patent to the application outlined in this

thesis is shown in the following:

Patent Title: High-strain, magnetic field-controlled actuator materials

Patent Number: 5, 958, 154

Date of Patent: Sep 28, 1999

Inventors: Robert C. O'Handley, Kari M. Ullakko

Assignee: Massachusetts Institute of Technology, Cambridge, Massachusetts

This patent is about FSMAs being used as actuator materials. The inventors are Dr O'Handley

(MIT) and Dr Ullakko (AdaptaMat, former Visiting Scientist at MIT), who were among the first

researchers to demonstrate magnetic field-induced strain in FSMAs [2]. The assignee is MIT.

This means that commercialization of FSMA enhanced products may be carried out by paying

licensing fees to the MIT Technology Licensing Office (TLO).

There are some existing patents which document remote actuation of medical implants through

the application of an external magnetic field. An example is given in the following:

Patent Title: Method and apparatus for stimulating the healing of medical implants

Patent Number: 6, 032, 677

Date of Patent: Mar 7, 2000

Inventors: Abraham M. Blechman, Jonathan J. Kaufman

This patent describes a technique to enhance the stability of dental and orthopaedic implants by

inducing implant micromotion via a permanent magnet coupled to the implant which is vibrated

using a cyclic magnetic force to enhance growth and apposition of the surrounding bone. The

Page 28: Orthopaedic Applications of Ferromagnetic Shape Memory Alloys

implant temporarily holds a rigidly fixed, permanent magnet that is oscillated by an externally

applied, time-varying magnetic field of controlled frequency and intensity. Movement of the

internal magnet transmits movement to the implant and effects stimulation of the implant-bone

interface.

Although patents such as these describe remote actuation of medical implants using an external

magnetic field, their claims involve an implanted permanent magnet. The device described in

this thesis involves FSMAs for actuation, and not permanent magnets. Thus, these existing

patents are not directly related to it. However, these patents are an issue to be aware of during

any possible future patent application.

One of the potential materials to be used to encapsulate Ni-Mn-Ga in the spinal traction device is

poly(styrene-block-isobutylene-block-styrene) (SIBS). The patent describing this material is

presented in the following:

Patent Title: Drug delivery compositions and medical devices containing block copolymer

Patent Number: 6, 545, 097 B2

Date of Patent: Apr 8, 2003

Inventors: Leonard Pinchuk, Sepideh Nott, Marlene Schwarz, Kalpana Kamath

Leonard Pinchuk of Innovia LLC patented poly(styrene-block-isobutylene-block-styrene) (SIBS)

in 2003. The SIBS intellectual property is currently owned by Boston Scientific Corporation and

a sublicense from the company is required for its use [55]. Therefore, use of SIBS in making the

spinal traction device would involve negotiation, and probably collaboration, with Boston

Scientific Corporation.

Another patent worthy of mention is one documenting a method for bonding silicone rubber and

polyurethane:

Page 29: Orthopaedic Applications of Ferromagnetic Shape Memory Alloys

Patent Title: Method for bonding silicone rubber and polyurethane materials and articles

manufactured thereby

Patent Number: 5, 147, 725

Date of Patent: Sep 15, 1992

Inventor: Leonard Pinchuk

Coincidentally, the inventor of this method, Leonard Pinchuk, is also the inventor in the

preceding patent. The method described in this patent may form the basis of a method used to

bond SIBS with PDMS or PET, which is required in the production of the spinal traction device.

On the whole, the author expects this field to be less litigious than other more mature fields, such

as the semiconductor industry.

Page 30: Orthopaedic Applications of Ferromagnetic Shape Memory Alloys

4. Future Direction

4.1 Issues and Risks

There are several issues with using FSMAs as part of medical implants. A leading concern is the

biocompatibility of FSMAs, which has not been tested before. Ni-Mn-Ga crystals have never

been implanted into the human body for extended periods of time. Direct contact with body

tissues may lead to cytotoxicity issues. The extent of these biocompatibility issues has to be

determined. Since the spinal traction device is expected to stay in the body for up to a few

months, the FSMA components would have to be properly isolated from body tissues. Suitable

encapsulation materials have been suggested in the preceding chapter. These candidates have to

undergo rigorous in vitro and in vivo tests to prove their suitability. Also, how the encapsulation

materials are to be bonded together and used to enclose the FSMA requires further investigation.

Concern needs to be placed on the cost and ease of mass production.

The effectiveness of the device in treating spinal disorders is another uncertainty. It is unknown

whether the device can significantly improve spinal conditions, even if it works according to its

specifications. Seeking FDA approval is another challenge, which is likely to take a long time.

At this stage of development, it is challenging to give accurate predictions of the manufacturing

yield and costs. Since there has not been any large-scale production of FSMAs before, there is no

reliable information about the yield and costs involved. The exact materials to be used for

making various components of the device are still unconfirmed. At this point, it is unknown

which material or combination of materials is the cheapest effective candidate for FSMA

encapsulation. Thus, the current prediction of production costs has a large margin of error.

There are also market risks to consider. Although this new device is slated to be the most precise

means to perform traction, how receptive surgeons and the general public are to this new

treatment technique remains to be seen.

Page 31: Orthopaedic Applications of Ferromagnetic Shape Memory Alloys

This FSMA orthopaedic technology is very early in the development stage. It is still quite some

time away from the state desired when starting a company. As seen in Figure 10, there are likely

to be several hurdles and uncertainties ahead on the road to commercialization. These include the

aforementioned issues and risks. Therefore, the current strategy would be to conduct relevant

research (prototype fabrication and testing) and secure appropriate patents. Constant monitoring

of the patent arena is needed. Sufficient coverage in the field of intellectual property is crucial to

protecting the research team's interests, in anticipation of future litigation.

~I-

- fe 1 u11 i)tU

Figure 10. This FSMA orthopaedic technology is in its embryonic stage, with much development needed

before concrete commercialization steps (such as company formation) should be considered.

The setup and strength of the electromagnet required to generate sufficient magnetic field at the

FSMA location for device actuation have to be determined. The characteristic magnetic field-

strain behavior of such a setup has to be ascertained. In vitro trials are expected to take

approximately 1.5 years [22]. These include bench-top tests and measurements in cadavers. A

couple of these in vitro compatibility tests are outlined in the following section. If these achieve

satisfactory results, animal trials should follow. These in vivo trials may be conducted on rabbits

[22]. Spinal disorders would be deliberately created in rabbits. The prototype device would then

be implanted and tested for its effectiveness in treating these disorders.

4.2 In Vitro Compatibility Testing

Any metal in direct contact with biological systems will corrode, resulting in the release of ions

which cause adverse physiological effects [57, 58]. These include toxicity, carcinogenicity,

Page 32: Orthopaedic Applications of Ferromagnetic Shape Memory Alloys

genotoxicity and metal allergy. Although metal ions are naturally present in the body,

concentrations beyond physiological levels can be toxic. Therefore, release of metal ions from

implants remains an important concern. However, the boundary line between acceptable and

toxic levels of metal ions is currently unclear [59]. The compositions of the main alloys used in

orthopaedic implants outlined by the American Society of Testing and Materials (ASTM) are

presented in the following table.

Group Co Cr Ni Mo W Ti Al V Cu Mn

F 75 cast 59-60w/ 27-3(0% <1% 5-7% N/A N/A N/A N/A N/A 1%Co/Cr/Mo alloy

F 90 wrought -45, 19-21% 9-11% N/A 14-16% N/A N/A N/A N/A N/ACo/Cr/Mo alloy

F 799 forged -59% 26-30•, <1% 5-7% N/A N/A N/A N/A N/A N/ACo/Cr alloy

F 562 wrought -29V% 19-21% 33-37% 9-1, 9 N/A N/A N/A N/A N/A N/ACo/Ni/Cr alloy

F 1537 wrought -59% 26-30% . 1% 5-7% N/A N/A N/A N/A N/A N/ACo/Cr/Mo alloy

SM 21 58.9-69% 26-30% <1% 5-7% N/A N/A N/A N/A N/A N/ACo/Cr alloy

ASTM F138 N/A 17-19% 13-15.5% 2-4% N/A N/A N/A N/A 0.5 N/AStainless Steel

ASTM F136 N/A N/A N/A N/ N/A N/A A 9-91% 5.5-6.5% 35-4.5% N/A N/ATitanium alloy

Table 2. Compositions of the main alloys used in orthopaedic prostheses [59].

Toxic concentrations of transition metals, especially chromium and nickel, disrupt the body's

oxidation-reduction balance, preventing normal cell signaling and gene expression, and resulting

in toxicity and carcinogenicity [60]. Nickel, which is present in Ni-Mn-Ga alloys, has been

identified as a human carcinogen [61]. With their participation in oxidation-reduction reactions,

metal ions can also shift the pH in the body [62]. In terms of cytotoxicity, nickel from stainless

steel is one of the most harmful components in alloys used as implant material [63]. It has been

shown that soluble forms of nickel cause significant increases in cell transformation in

fibroblasts of mice [64]. This transforming ability is directly related to the toxicity of the metal.

Compared to nickel, less is known about the effects of increased concentrations of manganese

and gallium in the body. In general, more research is required to have a better understanding of

the effects of metals used in implants and their resulting ion concentrations [59].

Page 33: Orthopaedic Applications of Ferromagnetic Shape Memory Alloys

Although the FSMA components of the implant would be encapsulated, certain compatibility

tests still have to be conducted to ascertain possible outcomes of improper encapsulation. The

following are some in vitro corrosion tests to be conducted before animal testing. Note that they

are in no way exhaustive.

4.2.1 Pitting Corrosion Test

Chloride ions in the body cause pitting corrosion of implants. Thus, a pitting corrosion test is

required to evaluate how Ni-Mn-Ga responds to pitting. The cyclic pitting scan is to be

conducted in saline solution of 0.9% NaCI, at 370 C with a sweep rate of 0.2 mV s 1'. From a plot

of the corrosion characteristics curve of current density against potential, the critical pitting

potential and the protection potential can be obtained. The critical pitting potential is the

potential above which new pits will initiate and existing pits will propagate. The protection

potential is the potential below which there is zero pitting. Between these two values, existing

pits will propagate with no new pit formation. Comparing these two potential values of Ni-Mn-

Ga with those of implant materials in literature would show how relatively resistant to pitting Ni-

Mn-Ga is. The more positive the potential values, the more resistant to pitting a material will be.

Being in the single-crystal form, Ni-Mn-Ga is expected to be relatively resistant to pitting [65],although this needs to be confirmed by experimental data. For this test, several sets of data are

needed because there are usually significant variations in the test results [65]. This may be

accomplished either with several samples of the same dimensions, or by polishing the sample

surface after each measurement.

For a detailed guide to conducting the test, one may refer to ASTM G61, which documents the

standard test method for conducting cyclic potentiodynamic polarization measurements for

localized corrosion susceptibility of iron-, nickel-, or cobalt-based alloys in a chloride

environment. This test method also describes an experimental procedure which can be used to

check one's experimental technique and instrumentation.

When conducting pitting corrosion tests, take note of crevice corrosion, which refers to localized

corrosion of a surface, usually a metal, at or immediately adjacent to an area which is shielded

from full exposure to the environment due to close proximity between the metal and the surface

Page 34: Orthopaedic Applications of Ferromagnetic Shape Memory Alloys

of another material. This corrosion is prominent because the solution stagnates in such crevices,

and can turn more acidic and become more effective in corroding the material. This form of

corrosion should be avoided during pitting corrosion tests, because it would affect the potential

values obtained. In the event that crevice corrosion is detected, the pitting corrosion test results

would become void.

4.2.2 Weight Loss Test and Trace Nickel Detection

This test is to be conducted on NiTi, 316L stainless steel and Ni-Mn-Ga, with one sample of

each material. This juxtaposition allows comparison of Ni-Mn-Ga with materials already used

for implants [66]. Obtain one sample of each material and weigh them. Completely immerse and

suspend each in a separate saline solution of 0.9% NaCl. Suspend them in such a way as to

minimize surface contact, reducing the chances of crevice corrosion. Leave them in solution for

1 month, before removing and weighing them again. Compute and compare the weight loss. The

less the weight loss, the better Ni-Mn-Ga would perform as an implant material.

Determine the amount of nickel released into the solutions of all three samples with inductively

coupled plasma (ICP), which is an analytical technique used for the detection of trace metals in

solution, commonly used for environmental samples. Compared to the solutions of NiTi and

316L stainless steel, a comparable or lower concentration of nickel in that of Ni-Mn-Ga would

negate the nickel toxicity issue, while a much higher concentration would raise a cause for

concern.

Page 35: Orthopaedic Applications of Ferromagnetic Shape Memory Alloys

Conclusions

The optimal strategy to take now is to perform the device fabrication and testing, and obtain

relevant patents to cover the IP arena in anticipation of future litigation. In vitro trials are

expected to take about 1.5 years, while in vivo trials are expected to take longer. Two of the in

vitro trials have been outlined in Section 4.2. If the trials prove successful, seeking FDA

approval would be the following step. After that, commercialize the product by collaborating

with companies such as AdaptaMat, Synthes and Boston Scientific Corporation, through virtual

vertical integration of manufacturing processes. Collaboration with AdaptaMat could mean

licensing the technology to them or to form a specialized start-up with the help of their expertise

in FSMA growth and fabrication. The decision between these two choices may lie in technical

considerations. This orthopaedic technology is currently in its infant stage. Its technical

feasibility and commercial viability remain to be seen.

Page 36: Orthopaedic Applications of Ferromagnetic Shape Memory Alloys

References

1) R.C. O'Handley, S.M. Allen (2002), Shape-memory alloys, magnetically activated

ferromagnetic shape-memory materials, Encyclopedia of Smart Materials (pp. 936-951), John

Wiley and Sons 2002.

2) K. Ullakko, J.K. Huang, C. Kantner, V.V. Kokorin, and R.C. O'Handley, Large magnetic-

field-induced strains in Ni2MnGa single crystals, Applied Physics Letters 69, 1966-1968 (1996).

3) S.J. Murray, M. Marioni, S.M. Allen, R.C. O'Handley, 6% magnetic-field-induced strain by

twin-boundary motion in ferromagnetic Ni-Mn-Ga, Applied Physics Letters 77, 886-888 (2000).

4) C.P. Henry, Dynamic actuation properties of Ni-Mn-Ga ferromagnetic shape memory alloys,

MIT Doctoral Thesis (2002).

5) A.A. Likhachev, K. Ullakko, Magnetic-field-controlled twin boundaries motion and giant

magneto-mechanical effects in Ni-Mn-Ga shape memory alloy, Physics Letters A 275, 142-151

(2000).

6) J.R. Cullen, A.E. Clark, K.B. Hathaway, Handbook of Materials Science, Chapter 16, edited

by K.H.J. Buschow, VCH, Amsterdam (1997).

7) R.C. O'Handley et al., Magnetic-field-induced structure and property changes in

ferromagnetic shape-memory alloys, presented at MRS Fall 2000, Boston, MA, 2000.

8) R.C. O'Handley, S.J. Murray, M. Marioni, H. Nembach, S.M. Allen, Phenomenology of giant

magnetic-field-induced strain in ferromagnetic shape-memory materials, Journal ofApplied

Physics 87, 4712-4717 (2000).

9) R.H. Ivester, Fabrication and characterization of Ni-Mn-Ga ferromagnetic shape-memory

alloy composites, MIT Bachelor Thesis (2002).

10) J. Feuchtwanger, M.L. Richard, Y.J. Tang, A.E. Berkowitz, R.C. O'Handley, S.M. Allen,

Large energy absorption in Ni-Mn-Ga/polymer composites, Journal ofApplied Physics 97,

10M319 (2005).

11) I. Suorsa, J. Tellinen, K. Ullakko, E. Pagounis, Voltage generation induced by mechanical

straining in magnetic shape memory materials, Journal ofApplied Physics 95, 8054-8058 (2004).

12) M. Kohl, D. Brugger,M. Ohtsuka, T. Takagi, A novel actuation mechanism on the basis of

ferromagnetic SMA thin films, Sensors and Actuators, A, Physical 114, 445-450 (2004).

13) V.A. Chernenko, M. Ohtsuka, M. Kohl, V.V. Khovailo, T. Takagi, Transformation behavior

of Ni-Mn-Ga thin films, Smart Materials and Structures 14, S245-S252 (2005).

Page 37: Orthopaedic Applications of Ferromagnetic Shape Memory Alloys

14) Y. Liang, Y. Kuga, M. Taya, Design of membrane actuator based on ferromagnetic shape

memory alloy composite for synthetic jet applications, Sensors andActuators, A, Physical 125,

512-518 (2006).

15) R.C. O'Handley, Personal communication, December 2007.

16) K.C. Ang, Personal communication, March 2008.

17) K.S. Ngian, Personal communication, March 2008.

18) Wright Medical Technology, Inc. Retrieved February 27, 2008, from

http://www.wmt.com/Physicians/Products/Extremities/

19) S.M. Allen, Personal communication, February 2008.

20) Synthes media release, West Chester (PA), USA, February 28, 2008.

21) L. Christenson, Gale Encyclopedia of Medicine, published December 2002, updated on

August 14, 2006. Retrieved April 6, 2008, from Health A to Z,

http://www.healthatoz.com/healthatoz/Atoz/common/standard/transform.jsp?requestURI=/h

ealthatoz/Atoz/ency/bone_grafting.jsp

22) C.H. Goh, Personal communication, April 2008.

23) T.J. Blokhuis, J.H.D. de Bruine, J.A.M. Bramer, et al., The reliability of plain radiography in

experimental fracture healing, Journal of Skeletal Radiology 30, 151-156 (2001).

24) R. Hammer, S. Hammerby, B. Lindholm, Accuracy of radiologic assessment of tibial shaft

fracture union in humans, Clinical Orthopaedics and Related Research 199, 233-238 (1985).

25) J.J. Dias, Definition of union after acute fracture and surgery for fracture non-union of the

scaphoid, Journal of Hand Surgery 26B(4), 321-325 (2001).

26) J. Webb, G. Herling, T. Gardner, et al., Manual assessment of fracture stiffness, Injury 27(5),

319-320 (1996).

27) B.J. Davis, P.J. Roberts, C.I. Moorcroft, M.F. Brown, P.B.M. Thomas, R.H. Wade,

Reliability of radiographs in defining union of internally fixed fractures, Injury 35, 557-561

(2004).

28) P. Vecchia, Scientific rational for occupational EMF exposure restrictions, International

Commission on Non-Ionizing Radiation Protection, Current trends in health and safety risk

assessment of work-related exposure to EMFs, Joint international workshop, 14-16 February

2007, Milan, Italy.

29) K. Kumar, Spinal deformity and axial traction, Spine 21, 653-655 (1996).

Page 38: Orthopaedic Applications of Ferromagnetic Shape Memory Alloys

30) M.V. Shterenshis, The history of modern spinal traction with particular reference to neural

disorders, Spinal Cord 35, 139-146 (1997).

31) C. Hinterbuchner (1985), Traction, In J.V. Basmajian (editor), Manipulation, traction and

massage, pp. 178, Baltimore, MD, Williams and Wilkins.

32) G.J.M.G. van der Heijden, A.J.H.M. Beurskens, B.W. Koes, W.J.J. Assendelft, H.C.W. de

Vet, L.M. Bouter, The efficacy of traction for back and neck pain: a systematic, blinded review

of randomized clinical trial methods, Physical Therapy 75, 93-104 (1995).

33) P. Gillstrom, A. Ehmberg, Long-term results ofautotraction in the treatment of lumbago and

sciatica, Archives of Orthopaedic and Traumatic Surgery 104, 294-298 (1985).

34) E. Knuttson, C. Skoglund, E. Natchev, Changes in voluntary muscle strength, somatosensory

transmission and skin temperature concomitant with pain relief during autotraction in patients

with lumbar and sacral root lesions, Pain 33, 173-179 (1988).

35) G. Grieve, Neck traction, Physiotherapy 68, 260-265 (1982).

36) J. Cyriax, Conservative treatment of lumbar disc lesions, Physiotherapy 50, 300-303 (1964).

37) L.T. Twomey, Sustained lumbar traction: an experimental study of long spine segments,

Spine 10, 146-149 (1985).

38) M. Krause, K.M. Refshauge, M. Dessen, R. Boland, Lumbar spine traction: evaluation of

effects-and recommended application for treatment, Manual Therapy 5(2), 72-81 (2000).

39) S.C. Colachis, B.R. Strohm, Effects of intermittent traction on separation of lumbar vertebrae,

Archives of Physical Medicine and Rehabilitation 50, 251-258 (1969).

40) R.C. Gupta, S.V. Ramarao, Epidurography in reduction of lumbar disc prolapse by traction,

Archives of Physical Medicine and Rehabilitation 59, 322-327 (1978).

41) J.D. Schlegel, J. Champine, M.S. Taylor, J.T. Watson, M. Champine, R.L. Schleusener, et al.,The role of distraction in improving the space available in the lumbar stenotic canal and foramen,Spine 19, 2041-2047 (1994).

42) E.G. Ralph, G. Bronfort, R.L. Evans, Distraction manipulation of the lumbar spine: A review

of the literature, Journal of Manipulative and Physiological Therapeutics 28, 266-273 (2005).

43) L. Pinchuk, G.J. Wilson, J.J. Barry, R.T. Schoephoerster, J.M. Parel, J.P. Kennedy, Medical

applications of poly(styrene-block-isobutylene-block-styrene) ("SIBS"), Biomaterials 29, 448-

460 (2008).

Page 39: Orthopaedic Applications of Ferromagnetic Shape Memory Alloys

44) J.W. Swanson, J.E. LeBeau, The effect of implantation on the physical properties of silicone

rubber, Journal of Biomedical Materials Research 8, 357-367 (1974).

45) E.J. Roggendorf, The biostability of silicone rubbers, a polyamide, and a polyester, Journal

of Biomedical Materials Research 10, 123-143 (1976).

46) R. Carmen, S.C. Mutha, Lipid absorption by silicone rubber heart-valve poppets - in vivo

and in vitro results, Journal of Biomedical Materials Research 6, 327-346 (1972).

47) P. Vondracek, B. Dolezel, Biostability of medical elastomers: a review, Biomaterials 5, 209-

214(1984).

48) S. Weinberg, M.W. King (2004), Medical fibers and biotextiles, In B.D. Ratner, A.S.

Hoffinan, F.J. Schoen, J.E. Lemons (editors), Biomaterials Science: An Introduction to Materials

in Medicine, 2nd Edition, pp. 87, Elsevier Academic Press.

49) E.G. Ralph, S.B. Jeffrey, Evidence-informed management of chronic low back pain with

traction therapy, Spine 8, 234-242 (2008).

50) The VAX-D Network. Retrieved June 5, 2008, from http://www.vaxd.net/index.html

51) P.F. Beattie, R.M. Nelson, L.A. Michener, J. Cammarata, J. Donley, Outcomes after a prone

lumbar traction protocol for patients with activity-limiting low back pain: a prospective case

series study, Archives of Physical Medicine and Rehabilitation 89(2), 269-274 (2008).

52) K.L. Saban, S.M. Penckofer, I. Androwich, F.B. Bryant, Health-related quality of life of

patients following selected types of lumbar spinal surgery: A pilot study, Health and Quality of

Life Outcomes 5, 71-81 (2007).

53) Sigma-AldrichTM. Retrieved June 3, 2008, from

http://www.sigmaaldrich.com/catalog/search/ProductDetail/ALDRICH/469300

54) US Patent # 6, 545, 097: "Drug delivery compositions and medical devices containing block

copolymer" Pinchuk et al. (Scimed Life Systems, Inc.), April 8, 2003.

55) Innovia LLC and Affiliates. Retrieved June 4, 2008, from http://www.innovia-

llc.com/Fall%202007/Technology%20-%20Biomaterials%20-%20SIBS%20.html

56) Sigma-AldrichTM. Retrieved June 3, 2008, from

http://www.sigmaaldrich.com/catalog/search/ProductDetail?ProdNo=305812&Brand=ALDRIC

H

57) S.G. Steinemann, Metal implants and surface reactions, Injury 27, SC 16-22 (1996).

Page 40: Orthopaedic Applications of Ferromagnetic Shape Memory Alloys

58) N. Hallab, K. Merritt, J.J. Jacobs, Metal sensitivity in patients with orthopaedic implants,

Journal ofBone and Joint Surgery (US Volume) 83A(3), 428-436 (2001).

59) A. Sargeant, T. Goswami, Hip implants - Paper VI - Ion concentrations, Materials and

Design 28, 155-171 (2007).

60) G. Buzard, K. Kasprzak, Possible roles of nitric oxide and redox cell signaling in metal-

induced toxicity and carcinogenesis: a review, Journal of Environmental Pathology Toxicology

and Oncology 19(3), 179-199 (2000).

61) M. Pulido, A. Parrish, Metal-induced apoptosis: mechanisms, Mutation Research 533, 227-

241 (2003).

62) D.E. Carter, Oxidation-reduction reactions of metal ions, Environmental Health Perspectives

103, 17-19 (1995).

63) T. Rae, The toxicity of metals used in orthopaedic prostheses: An experimental study using

cultured human synovial fibroblasts, Journal of Bone and Joint Surgery (British Volume) 63B(3),

435-440 (1981).

64) D.W. Howie, S.D. Rogers, M.A. McGee, D.R. Haynes, Biologic effects of cobalt chrome in

cell and animal models, Clinical Orthopaedics and Related Research, 329S, S217-S232 (1996).

65) D.J. Blackwood, Personal communication, June 2008.

66) J. Ryhlinen, M. Kallioinen, J. Tuukkanen, J. Junila, E. Niemeld, P. Sandvik, W. Serlo, In

vivo biocompatibility evaluation of nickel-titanium shape memory metal alloy: muscle and

perineural tissue responses and encapsule membrane thickness, Journal of Biomedical Materials

Research 41, 481-488 (1998).