Optimize plasma freezing - Helmer Scientific | Medical Refrigerator & Freezer … · 2019-04-12 ·...
Transcript of Optimize plasma freezing - Helmer Scientific | Medical Refrigerator & Freezer … · 2019-04-12 ·...
ThermoLine™ Ultra-Rapid Freezing Systems
Optimize plasma freezingfor higher protein yields.
• Higher FACTOR VIII yields than air blast systems
• Vertical bag freezing - Prevents plasma migration to ports - Permits flat storage and transport of frozen bags - Reduces bag breaks from air pocket entrapment
• Increased production capacity - Core temperature of -30° C is reached in 22 minute freeze cycles
- Independent freezing chambers permit efficient processing
• Environmentally-friendly, non-toxic InstaCoolant IV is ideally suited for heat transfer in low temperature systems
• Easy-to-use with - Top-loading design - No overwraps required - No heavy trays to lift - One-touch operation
• Reduce risk of contamination with Snap-Seal membrane pockets
RAPID PLASMA FREEZERS RAPID PLASMA FREEZERS
For ordering information, contact Helmer Scientific or visit us at: www.helmerinc.com
Helmer Scientific14400 Bergen BlvdNoblesville, IN 46060Toll-free: 800.743.5637 (US/Canada)Direct: +1.317.773.9073Fax: +1.317.773.9082
From stem cells to solutions.
©2009 ThermoGenesis. All rights reserved.
SPECIFICATIONS MP1100
Height (min/max) 111 cm (44”)/185 cm (73”)
Width 201 cm (80”)
Depth 82 cm (33”) Model 8-1100-27 only 95 cm (38”)
Condenser Type Integral-cascade***
Electric Defrost Heater 1000W
Freezer Weight (Empty) 475 kg (1048 lbs)
Freezer Weight (with Coolant) 715 kg (1577 lbs)
Coolant Volume 284 L (75 US gallons)
Product Final Core Temperature -40° C (-40° F)
Freezer Line 230 VAC/1 Ph/60 Hz/36 FLA Voltage/Amperage
Minimum Circuit Breaker 50A
Ambient 15° C to 26° C (59° F to 79° F) Operating Temperature
Relative Humidity 10%-70% RH
Compressor Refrigerant High Side: R-507†
Low Side: R-508b††
Compressor -62° C (-80° F) Operating Temperature Saturated Suction Temperature
†Allied Signal AZ-50 ††DuPont SUVA 95
***Requires adequate space from wall to proper ventillation/airflow.
The ThermoLine™ MP1100 meets the demands of low to high-volume labs with rapid
freezing times and multiple features to ensure cellular integrity throughout the
freezing process.
The MP1100 employs an integrated cascade condenser for rapid freezing performance.
FREEZER DECK CONFIGURATIONS AND ORDERING INFORMATION
MP1100
ORDER # POCKETS CAPACITY
8-1100-21 16 32 x 300 ml FFP Bags 16 x 1 L Terumo Bags
8-1100-21—16 pockets
Other bottle and FFP bag deck configurations are available. Contact your sales representative or Helmer Scientific for further information.