Optically interrogated MEMS pressure sensor...
Transcript of Optically interrogated MEMS pressure sensor...
14th Int Symp on Applications of Laser Techniques to Fluid Mechanics Lisbon, Portugal, 07-10 July, 2008
- 1 -
Optically interrogated MEMS pressure sensor array
Lukas Prochazka1, Thomas Roesgen
2
1: Institute of Fluid Dynamics, ETH, Zurich, Switzerland, [email protected]
2: Institute of Fluid Dynamics, ETH, Zurich, Switzerland, [email protected]
Abstract A novel pressure measurement technique is developed with the objective to simplify wind tunnel investigations while providing full field measurement capability with highest measurement performance. The static pressure distributions are recorded by interferometric means using an array of silicon diaphragm microresonators as pressure sensing elements. They are remotely excited into free vibration using a capacitive ultrasound transducer (CUT) transmitting high power narrow band (50 – 80kHz) acoustic noise. Dependent on their quasi-static deflection caused by a pressure load the resonance frequency varies with an average pressure sensitivity of 3Hz/Pa at 70kHz. Further requirements for the system performance are a maximum pressure, temporal and spatial resolution of 0.2% of FS (10kPa), 1Hz and 5mm, respectively.
1. Introduction
During aerodynamic and mechanical load performance surveys in wind tunnel testing it is often
desirable to obtain the surface pressure on a specific component of the investigated model with high
spatial resolution, good measurement accuracy and low uncertainty. Despite the great demand for
static pressure measurements only the pressure tap method tends to be used in commercial and
industrial wind tunnel facilities. This technique, where the transducers are linked by a tube to the
measurement locations, provides high measurement performance but is also very costly due to the
required mechanical and hydraulic interfacing in the test model. The application of pressure
sensitive paints for low speed testing at atmospheric pressure conditions is often unsuitable due its
low sensitivity and large temperature cross-sensitivity (Liu et al. 2005 and Airaghi 2006).
The objective of the present work is to develop an alternative pressure field measurement technique
with measurement performance (i.e. high accuracy, low measurement uncertainty) comparable to
the pressure tap system but more simple handling and installation requirements. The requirements
on the measurement performance are summarized in Table 1.
Table 1 Required measurement performance
Temporal resolution
Spatial resolution
Measurement range
High sensitivity
Low measurement uncertainty (drift, cross sensitivity, …)
≈ 1Hz
> 5mm
< 10kPa
0.2% FS or 20Pa
In order to define a suitable design concept different techniques and devices for static pressure
measurements of different application fields were compared and evaluated regarding the above
mentioned criteria (Prochazka et al. 2005). Based on this study two basic conclusions were
identified.
1. The need for high measurement performance (sensitivity, uncertainty) in low speed wind
tunnel testing requires a pressure sensing device with a deflecting sensing element operating
in differential mode, e.g. membrane or diaphragm.
14th Int Symp on Applications of Laser Techniques to Fluid Mechanics Lisbon, Portugal, 07-10 July, 2008
- 2 -
2. To allow for a simple installation of the device, the sensor unit has to be mounted directly
on the model surface and its interrogation has to be wireless (no electrical or hydraulic
interfacing)
Wireless interrogation of a sensor array by electromagnetic waves in the microwave or radio
frequency (RF) range has the advantage of permitting an operation even without visual contact
between sensor and interrogation unit. On the other side the transmitter on the sensor side
(electronics and power supply) would again require complex installation within generally small
wind tunnel models. The delivery of the necessary power by RF over a distance of around 2 meters
would require high transmitting power with the potential for electromagnetic interference with other
highly sensitive electronic devices within the wind tunnel environment. An optical interrogation
causes no electromagnetic interference and the necessary visual contact between sensor and detector
is usually given due to several observation points in a test section. The challenging task hereby is to
detect the expected small signal carrying the pressure information within a bright, vibrating
background and over the mentioned interrogation distance.
The concept which is regarded as the best suitable measurement technique for the mentioned
application is based on tiny and passive MEMS pressure sensor dies which are mounted directly on
the model surface and are optically interrogated using a highly sensitive active pixel CMOS lock-in
detector.
2. Method
The pressure sensing system is based on the optical interrogation of an array of passive MEMS
pressure sensors (silicon diaphragms with a sealed cavity) which may be embedded in a highly
elastic filler material and are mounted directly on the model surface (Fig.2.1). Dependent on the
surface geometry, the model size and application, a thin (0.2 mm) sensor foil with a certain number
of sensing elements or individual tags equipped with a single sensor are conceivable. In case of
small-scale models, where even 0.2mm thick and 5mm large obstacles could alter the flow
characteristics, the tags may be recessed in tiny milled cavities in the model surface (Fig.2.2).
Fig. 2.1 Measurement setup including the remote imaging
interferometer and the ultrasound transducer
Fig. 2.2 Sensor design and tags mounted flush to the
surface of the model (one possible arrangement)
As a measure for the applied pressure the resonance frequency of the clamped silicon diaphragm is
used for the present setup which depends on its quasi-static deflection and thus on the applied
pressure. We believe that in combination with a wireless and remote readout capability of the
sensors of about 2 meters the measurement of frequency instead of static deflection is more accurate
14th Int Symp on Applications of Laser Techniques to Fluid Mechanics Lisbon, Portugal, 07-10 July, 2008
- 3 -
and robust. An intensity modulating signal with the particular frequency may arise due to the
interference between the object and reference beam, due to an alternating speckle pattern or even
due to the deflection of the reflected light on the vibrating diaphragm. The resonant sensor
characteristic of a ‘digital’ frequency output signal (i.e. basically independent of analog levels)
simplifies the installation and adjustment of the optical setup significantly.
The readout is achieved in a two-stage configuration. First, an ultrasound transducer (UT) (emitting
band-limited noise) excites the diaphragms into free oscillation. After that, the actual resonant
vibration responses of the individual sensing elements are measured with an imaging interferometer
(or a speckle setup). This device is based on a highly sensitive, “active-pixel” CMOS camera
(CSEM, Switzerland) that works as a full field lock-in detector with a spatial resolution of 144 x 90
pixels (> 104 measurement points) (Beer et al. 2005). It can resolve signal frequencies up to 100kHz
while providing very strong background (i.e. static image) rejection and a frame rate of 6kHz.
Fig. 2.3 CUT (SensComp, Inc.) used in the present setup
and its power supply (self-made). Transducer diameter ≈
40mm
Fig. 2.4 Transmitting characteristic of the CUT
(SensComp, Inc.) operated with 200 VoltDC and
350 VAC p-p (Sound pressure measured at 1m from
transducer)
Using ultrasound allows the excitation of several sensing devices in parallel without altering the
sensing behavior due to heating as would be a problem with laser excitation or due to specific
actuators as in the electrical case. The difficulty associated with this excitation principle is mainly
related to the high losses caused by strong attenuation of ultrasound in air and thus the need for a
high power UT to allow for sufficiently high excitation over an operating distance of about 2
meters. High power UTs for applications in air are mainly of capacitive type (CUT). Advantages of
CUTs are their high transfer coefficient (high transmitting efficiency in air) in combination with a
wide bandwidth and the simple design that is reflected in their low cost (Manthey et al. 1992 and
Rafiq et al. 1991). Figure 2.3 shows the CUT used in the present setup (SensComp, Inc.) with its
large noise generating diaphragm of about 40mm and its DC and AC power supply. At almost
maximum operating power (200 VDC and 350 VAC p-p) the specific CUT provides a sound pressure
around 120dB (at 1m from transducer) in a wide frequency range from 50 to 100 kHz (Fig. 2.4).
3. Experimental verification of the sensing principle
The pressure sensing behavior of a plate resonator can be described by simple semi-empirical
formulas (Fabula et al. 1997), eq. 3.1 – 3.
( ) ( )2
25.110
+==
t
ddfdf resres (3.1)
14th Int Symp on Applications of Laser Techniques to Fluid Mechanics Lisbon, Portugal, 07-10 July, 2008
- 4 -
)1(122
16.35)0(
22 νρπ −==
E
a
tdf res (3.2)
+
−
=
32
258.12.4
)1(16
t
d
t
d
a
tEp
ν (3.3)
Eq. 3.1 describes the variation of the resonance frequency as function of the static diaphragm
deflection d and its thickness t whereas eq. 3.2 gives a formula for the resonance frequency of a
square, non deflected, clamped diaphragm. The variable a represents the edge length of the
diaphragm and the material properties are defined by the Young’s modulus E, the Poisson’s ration ν
and the material density ρ. The static deflection due to loading of the diaphragm, i.e. pressure load
p, can be calculated iteratively using the implicit formula given by eq. 3.3.
Although experimental data associated with the plate resonator can be found in literature (Defaÿ et
al. 2002, Smits et al. 1985 and Uttamchandani et al. 1987) they are insufficient to establish a
reliable validation of the above mentioned approach. Therefore experimental tests with common
pressure sensor dies were carried out using a pressure and temperature controlled test cell. The
pressure outside the resonator and in the cavity behind the diaphragm could be adjusted
independently of each other between 0.8bar below and above ambient pressure with an accuracy
better than 1mbar (< 0.1% FS). The temperature was set and controlled with a Peltier heater and
two heat exchangers to values between 20 and 35ºC with an accuracy better than 0.02K (Fig.3.1).
Fig. 3.1 Experimental setup for the investigation of sensor behavior
Due to the hermetically sealed cell and limited internal space it was not possible to combine
acoustic excitation and pressure control inside the cell. Instead an external, pulsed laser diode
(Flexpoint 10mW, Laser Components GmbH) was used to heat the diaphragm locally and thus to
generate mechanical stresses which excite the diaphragm into vibration with a certain excitation
frequency. The dynamic amplitude was then measured with a Laser Doppler Vibrometer OFV-551
14th Int Symp on Applications of Laser Techniques to Fluid Mechanics Lisbon, Portugal, 07-10 July, 2008
- 5 -
(Polytec GmbH) from outside of the test cell as well (Fig.3.1). During a test run the excitation
frequency was swept across the range of interest with frequency steps up to 20Hz in the
neighborhood of the resonance peak. The resonance frequency was then found at the maximum of
the amplitude spectrum (Fig. 4.2).
Figure 3.2 shows a comparison between the theoretical approach and experiment (red crosses)
carried out with a 1.4mm (left) and 1.1mm (right) large square diaphragm both about 10µm thick.
The static diaphragm deflection was varied by an adjustment of the back pressure in the cavity
between 0 and -900mbar. The pressure and the temperature in the test cell were kept constant
during the measurement. The red dashed line represents the resonance frequency as function of the
pressure load acting on the diaphragm determined with the above mentioned approach and the
dotted black line gives its derivative in respect to the pressure load (i.e. pressure sensitivity). Due to
the uncertainty in the value of diaphragm thickness of the individual sensor dies the particular
thickness for calculation was adjusted to match the resonance frequency at zero static deflection
based on theory and experiment. As is obvious, eq. 3.1 strongly overestimates the pressure
sensitivity determined by the experiment and under the aforementioned assumption of thickness
setting. Therefore eq. 3.1 was adjusted by a linear term inside the square root and the coefficients
are determined using the best correlation with all conducted measurements, eq. 3.4. Furthermore the
new approach rather underestimates the pressure sensitivity compared with the old one and
therefore seems to be more suitable to find the right resonator design for the present application.
( ) ( )2
95.021.010
+−==
t
d
t
ddfdf resres (3.4)
Fig. 3.2 Resonance frequency as function of static diaphragm deflection (diff. pressure across diaphragm) for a 10µm
thick silicon diaphragm with an edge length of 1.4mm (left) and 1.1mm (right). Red crosses: experiment, red solid:
eq. 3.1, red dashed: eq. 3.4. black dotted: pressure sensitivity derived from eq. 3.1 and black dash-dot derived from
eq. 3.4
4 Design considerations
The criteria regarding the sensor design partially followed from observations made during the above
mentioned experimental tests. The selection of the appropriate sensor design is based on the
validated approach given by eq. 3.2, 3.3 and 3.4.
Operating frequency range
The sensing devices are designed to operate in a frequency range between 50 and 80 kHz. At higher
frequencies the attenuation of ultrasound which increases with the frequency squared becomes too
high and thus excitation over a longer distance difficult. Furthermore the pressure sensitivity of a
14th Int Symp on Applications of Laser Techniques to Fluid Mechanics Lisbon, Portugal, 07-10 July, 2008
- 6 -
plate resonator with chip dimensions suitable for the present application is maximized in this
frequency range. Working at frequencies considerably below 50 kHz is not suitable because
environmental vibrations could influence the measurement or interaction between sound and the
boundary layer (BL) or the resonator and BL could alter the flow characteristics. Scenarios like an
interaction with the Tollmien-Schlichting waves in a laminar BL and thus an influence on the
transition between laminar and turbulent BL, or an alteration of the kinetic energy (larger eddies) of
a turbulent BL are conceivable. The latter interaction scenario could have an influence on the
friction drag or separation behavior of the BL. The main carriers of kinetic energy in a turbulent BL
are the larger eddies with characteristic frequencies far below 10 kHz (at wind speeds around
60m/s) (Blackwelder et al. 1976 and Schlichting et al. 2000). Thus an interaction with ultrasound is
not plausible. All studies where active flow control using acoustic sound or oscillating wall
structures was investigated, were carried out in the lower audible frequency range (Yarusevych et
al. 2006, Nishri et al. 1998, Ahuja et al. 1984, Greenblatt et al. 2000 and Sinha 2001).
Pressure sensitivity
Despite the fact that pressure sensors with a large sensing area (i.e. diaphragm) are generally more
sensitive to pressure variations due to the larger static diaphragm deflection, the measurement
sensitivity of a plate resonator is also influenced by its damping characteristics. The main damping
source of a plate oscillator results from acoustic radiation in air (Stemme 2001) which is dependent
on the diaphragm area squared, the vibrating frequency and the air density (Kinsler et al. 1982). The
influence of the two former parameters on the acoustic damping is shown in Figure 4.1. Basically
the damping is reduced (i.e. the pressure sensitivity increased) with decreasing diaphragm area. This
gain in measurement performance is reduced if the device operates in a frequency range below 100
kHz where the influence of the frequency on the acoustic damping is high and where the damping is
reduced with decreasing frequency (i.e. increasing diaphragm area). Figure 4.2 shows the
experimentally determined damping characteristics of a 1.1mm (blue circles) and 1.4mm (red star)
large diaphragm, expressed as a quality factor which is inversely related to the damping of an
oscillator (Q = fres/∆f3dB) (Stemme 2001).
Fig. 4.1 Damping fraction of the acoustic radiation as
function of the diaphragm size (R radius of a circular
diaphragm) and operating frequency
Fig. 4.2 Quality factor as function of the static diaphragm
deflection and area. 1.1mm (blue circles, 1.4mm (red
stars).
It is obvious that despite the higher resonance frequency (see Fig. 3.2 (right)) the smaller diaphragm
leads to higher Q-factors, i.e. lower damping compared with the larger device. The weak decay of
the Q-factor with decreasing pressure in the cavity (increasing static deflection or resonance
frequency) may be interpreted as an influence of the frequency on the acoustic damping.
Further restrictions toward an increasing diaphragm size are the required spatial resolution of 5mm
and a device size which should not modify the model geometry (contour). Sufficiently high pressure
14th Int Symp on Applications of Laser Techniques to Fluid Mechanics Lisbon, Portugal, 07-10 July, 2008
- 7 -
sensitivity and the capability of being acoustically excited and optically interrogated give the
limiting minimum size.
In general the maximum pressure sensitivity becomes higher if the diaphragm thickness decreases.
This is not necessarily true if the overall pressure sensitivity averaged over the required
measurement range is considered. For a very thin diaphragm the pressure sensitivity decays rapidly
after passing its maximum and may fall below the sensitivity of a thicker diaphragm. Furthermore
with a thinner diaphragm the fragility increases and a patterning of the upper surface (for optical
reasons as we will see later) becomes more difficult or unfeasible.
To be able to perform signed static pressure measurements the diaphragm must have a certain initial
static deflection which depends on the particular, signed measurement range (e.g. -6 kPa to 4 kPa).
The initial deflection is adjusted by setting the right pressure in the hermetically sealed cavity below
the diaphragm during the fabrication process.
A pre-stressed operation of the resonator is also favorable if its measurement performance is
considered. Figure 4.3 shows amplitude spectra of a 1.4 and 1.1mm large plate resonator for an
increasing static deflection, i.e. decreasing pressure in the cavity from 0 to -600mbar (-900mbar for
the right Figure) (from left to right). The shape of the individual amplitude spectrum gives certain
information about the oscillation behavior of the resonator. One may see that the resonance peaks at
zero or low pre-stressing are high and narrow and lead therefore to quality factors higher 100. If one
considers the larger diaphragm a sudden degradation of the oscillation behavior occurs at relatively
small static deflection (Fig. 4.3 left). The high drop of the resonance peak disappears again with
further increasing the pre-stressing of the diaphragm. The oscillation behavior improves again and
remains almost constant. Only a slight decay and broadening of the resonance peaks with increasing
static deflection remains visible. It probably is caused by the stiffening of the diaphragm together
with a constant excitation power and the broadening from the increasing resonance frequency (i.e.
increasing acoustic damping). The sudden drop in oscillation behavior could be located in the
vicinity of the maximum pressure sensitivity. As can be seen in figure 3.2 (right) this point is not
entirely reached and probably therefore no sudden drop of the resonance peak occurs in the case of
the smaller diaphragm (Fig. 4.3 right). That could confirm the assumption of different stress
conditions in the diaphragm before and after the point of maximum pressure sensitivity.
Temperature cross-sensitivity
An important requirement for the design is to keep the temperature cross-sensitivity far below the
measurement resolution of the system to avoid the need for correction of the pressure data. A single
material sensor design (single crystal silicon) with no electronics or intermediate oxide and nitride
Fig. 4.3 Characteristics of amplitude spectra with increasing static diaphragm deflection (from left to right) for a
large sensor with 1.4mm, Pcav: 0- -600mbar (left fig.) and small sensor with 1.1mm large diaphragm,
Pcav: 0- -900mbar (right fig.)
14th Int Symp on Applications of Laser Techniques to Fluid Mechanics Lisbon, Portugal, 07-10 July, 2008
- 8 -
layers together with a proper choice of the embedding material should minimize stresses due to
thermal expansion coefficient mismatch.
Fig. 4.4 Temperature cross-sensitivity of the large
sensor (1.4mm diaphragm edge length) embedded in
soft epoxy for a cavity pressure between 0 and
-800mbar (increasing static diaphragm deflection). A
constant sensitivity in the two given temperature ranges
(21 and 25ºC, 25 and 31ºC) is assumed
Fig. 4.5 Change in resonance frequency of the larger
sensor due to varying pressure in the test cell (i.e. air
density) between -500 and 300mbar. The static
diaphragm deflection of the sensor was kept constant
during the measurement. Solid line represents the
analytical approach (Di Gioanni 1982)
The experimental investigation of the influence of different embedding materials (silicon, rigid and
soft epoxy) on the temperature sensitivity of the sensing device did not provide conclusive results,
probably because of an unsuitable sensor fitting within the sample holder for such kind of
investigation. Further work on this subject will follow using the actual devices with a closed cavity.
Despite that an interesting observation could be made, namely a decreasing temperature cross-
sensitivity with an increasing static deflection of the diaphragm (i.e. an increasing pre-stressing)
(Fig. 4.4). The values in figure 4.4 are based on the assumption of constant temperature sensitivity
in the range between 21 and 25ºC and 25 and 31ºC and also take into account the uncertainty in
resonance frequency determination (error bars). The data were recorded for a large sensor
(diaphragm edge length 1.4mm) embedded in soft epoxy and cured at 35ºC for several hours.
The influence of variations of the surrounding gas density due to a change in temperature on the
vibration behavior is analytically described in literature (Di Giovanni 1982) and was also verified
by experimental tests (Fig. 4.5). Figure 4.5 shows the change in resonance frequency with
increasing pressure (from -500 to 300 mbar) and thus density within the test cell and compares it
with the analytical approach (solid line). During all measurements a slight pre-stressing of the
diaphragm (diff. pressure 25mbar) was preserved by adjusting the pressure in the cavity. If one
assumes that the air density changes due to temperature variations then the resulting cross-
sensitivity is small and should not exceed 4Hz/K at fres of about 70 kHz (compare temperature
sensitivity due to packaging, Fig. 4.4).
Dimensioning of the diaphragm
To evaluate the optimal size and thickness of the diaphragm the pressure sensitivity characteristics
for different diaphragm dimensions (edge length 1 to 2mm and thickness 3 to 10µm) was computed
using eq. 3.2 to 3.4. Subsequently the data were analyzed so as to fulfill following requirements:
(1) Maximum operating frequency range 50 to 80 kHz
(2) Minimum measurement range 10 kPa
(3) Minimum diaphragm pre-stressing corresponds to the point where maximum pressure
sensitivity occurs
14th Int Symp on Applications of Laser Techniques to Fluid Mechanics Lisbon, Portugal, 07-10 July, 2008
- 9 -
Fig. 4.6 Mean, max. and min. pressure sensitivity and operating
frequency range for different diaphragm dimensions identified
with an evaluation procedure
Fig. 4.7 Retro-reflecting characteristic of square
grooves with perpendicular side walls and
different aspect ratios (AR), β minimum incident
angle for retro-reflection, x usable part of the
bottom surface for retro-reflection and at a certain
incident angle α
Figure 4.6 shows the pressure sensitivity of all designs which fulfill the above mentioned criteria.
The markers represent the mean (average over measurement range) and the bar the maximum and
minimum sensitivity. The maker type indicates the corresponding diaphragm thickness and the
different colors the operating frequency range.
If one considers the advantage of smaller but thicker diaphragms (sensor size, acoustic damping and
mechanical strength) a plate resonator with an edge length of 1.4mm and a thickness between 5µm
and 6µm working in a frequency range between 50kHz and 80kHz was identified as the most
suitable configuration. Based on the assumption that the minimum sensitivity limits the
measurement performance (i.e. ≈ 2Hz/Pa), the resonance frequency has to be measured with an
uncertainty of 40Hz to meet the required resolution of 0.2% FS.
Adaptation of the sensor design for an improved optical interrogation
The main advantage of an imaging measurement technique is the possibility to interrogate many
sensors at the ‘same’ time. That requires an interrogation capability without the need for a precise
alignment of sensor and detector which is mainly fulfilled by frequency instead of displacement
measurement in the present application. Additionally the top surface of the diaphragm which
modulates the incoming light must have the capability to direct as much as light as possible to the
detector.
One possibility to obtain such an optical characteristic is to artificially roughen the top surface to
create a certain degree a diffusive reflectance. Because of the thin sensor diaphragm (t = 5-6µm) the
roughness height is limited to about the wave length of the incoming light (λ ≈ 700nm). Thus, only
weak diffusive reflectance may be expected. Besides the drawback of a diffusive surface losing
light due to hemispheric scatter, a further difficulty could arise during the manufacture of such a
rough surface on a thin diaphragm with common micro fabrication processes.
The most suitable sensor top surface would feature retro-reflecting properties, which are usually
obtained with a surface structure being similar to an array of corner prisms. As described in
literature there are ways to fabricate corner prism arrays with an element size down to 20µm on
silicon substrates (Neudeck et al. 1996 and Reimer et al. 1999). The drawbacks are the difficulty to
14th Int Symp on Applications of Laser Techniques to Fluid Mechanics Lisbon, Portugal, 07-10 July, 2008
- 10 -
further scale down the structures to suitable dimensions for the present application (1 to 2µm) and
special and unusual fabrication processes like the selective epitaxial growth technique or grey tone
lithograph.
With common wet and dry etching processes triple prisms are not producible on silicon substrates
but structures with partially retro-reflecting properties can be fabricated. Such structures are tiny 45º
V-grooves (Fig. 2.2) or square grooves with perpendicular side walls with respect to the top surface
(Fig. 4.7) both with an edge length between 1 and 2µm and a depth of 1µm. The former structures
can be manufactured by wet anisotropic etching using a surfactant added TMAH solution and an
alignment of the groove edge with the ⟨100⟩ direction (45º to the wafer flat) on a {100}-silicon
wafer (Sekimura 1999). The main drawback is that only incoming light parallel to the groove edges
is reflected back to the light source. Square grooves are simply fabricated using dry etching
processes which allows for highly anisotropic etching and thus perpendicular walls to the top
surface. It is further conceivable to smooth the side walls and the bottom surface performing a short
oxidation and subsequent stripping of the oxide in a buffered HF etchant. Though square structures
have a triple prism in each of their corners the retro-reflecting property works only for a certain
incident angular range which is dependent on the aspect ratio of the structure (Fig. 4.7). Shallow
structures works better for flat and deep ones for steep irradiation and thus a mixed array with
structures of different aspect ratios should be preferable. Nevertheless a blind spot range remains.
Due to structure dimensions near the wavelength of the interrogation light diffraction on the
patterned surface should appear which would hinder the interrogation process. Therefore the
individual grooves are pseudo-randomly distributed over a regular grid with a pitch of 2µm. As can
be shown the power spectral density spectrum of such a grid is independent of spatial frequency, i.e.
white-noise-like (Yashchuk et al. 2007).
We intend to develop fabrication processes for a rough, diffusive surface and for the two above
mentioned (partially) retro-reflective groove patterns and compare them regarding their capability
to be detected by coherent illumination from different observation points.
5 Sensor fabrication
As mentioned before the targeted low temperature cross-sensitivity of a highly sensitive pressure
sensor requires a sensing device made out of one material to minimize stresses due to thermal
expansion coefficient mismatch. Thus the common fabrication technique for pressure sensors where
the sensing diaphragm is manufactured by wet anisotropic etching from the back side of the wafer
with a precise etch stop at the highly doped area represented by the diaphragm is not applicable for
the present application.
The plate resonator with a hermetically sealed cavity is fabricated using low temperature direct
bonding of a silicon-on-insulator (SOI) and a silicon 4 inch wafer. The thin silicon wafer (200 µm)
forms the support for the diaphragm and features on its front side the 30µm deep pressure sensor
cavities. They are manufactured by common photolithography processes and wet anisotropic
etching in KOH (Fig. 5.1a and b). The device layer of the SOI wafer with a total thickness variation
(TTV) of less than 0.5µm forms the 6µm thick sensing diaphragm. Its front side is structured with a
retro-reflecting pattern (45º V- or 90º square grooves) performing e-beam lithography and
dependent on the groove geometry wet (surfactant added TMAH solution) or dry anisotropic
etching processes (Fig. 5.1c and d). E-beam lithography is necessary to obtain square structures
with preferably small rounding of their corners. The patterned surface is then bonded (low
temperature fusion bonding) with a further silicon support wafer (200µm) (Fig. 5.1e). For its later
removal by KOH etching the pattern on the SOI wafer is masked with an oxide layer (0.5µm) from
being etched as well. The diaphragm is released from the remaining SOI wafer using KOH (silicon
handle layer) and BHF (oxide sacrificial layer) wet etching (Fig. 5.1f). Low temperature fusion
14th Int Symp on Applications of Laser Techniques to Fluid Mechanics Lisbon, Portugal, 07-10 July, 2008
- 11 -
bonding is also used for joining the diaphragm (supported by wafer III) and its bottom part
including the cavity. This time the boding has to be performed in a precisely controlled nitrogen
environment to adjust the low pressure of -200mbar in the cavity and therefore the necessary initial
static deflection of the diaphragm (Fig. 5.1g). During the final step the support wafer (wafer III) and
the oxide layer are removed in a KOH and BHF bath respectively (Fig. 5.1h). a)
d)
b)
e)
c)
f)
g)
h)
Fig. 5.1 Fabrication of the sensing device
6 Conclusion
A sensor concept for wall pressure measurements in low speed wind tunnel applications is
presented. The sensor is designed for full field acquisition of a pressure distribution with a discrete
spatial resolution in the millimeter range and a temporal resolution around 1Hz. An integration of
the sensor unit directly on the surface of the model and a wireless interrogation should reduce the
handling and installation costs significantly.
The sensor concept is based on an array of discrete vibrating plate micro-resonators using their shift
of the resonance frequency as a measure for the applied pressure. These resonant structures have to
be excited wirelessly by ultrasound and their resonance frequency is optically interrogated from a
distance of about 2 meters.
The sensor design was optimized with regard to highest pressure sensitivity and measurement
performance and was carried out using a semi-empirical approach which was validated by
experimental tests. The most suitable resonator design has a 5 to 6µm thick diaphragm with an edge
length of 1.4mm. It is expected that the sensor is working in a frequency range between 50 and
80kHz and covers thereby a pressure range of around 10kPa. A certain initial static diaphragm
deflection allows for signed pressure measurement and improves also the measurement
performance. A minimum pressure sensitivity of about 2Hz/Pa and hence an uncertainty in
14th Int Symp on Applications of Laser Techniques to Fluid Mechanics Lisbon, Portugal, 07-10 July, 2008
- 12 -
resonance frequency measurement of 40Hz is expected to meet the required resolution of 0.2% FS.
A type of retro-reflecting pattern on the top surface of the diaphragm should simplify the optical
interrogation of the sensor array.
In order to reduce the temperature cross-sensitivity the whole sensor is made out of one material,
namely single crystal silicon. The fabrication relies on common photolithography and anisotropic
etching processes and also involves two low temperature direct fusion bonding processes.
References Airaghi S (2006) Self-Illuminating Pressure Sensitive Paints Using Electroluminescent Foils. PhD Thesis ETH No.
16522, Swiss federal Institute of Technology (ETHZ), Zurich, Switzerland, 2006
Ahuja KK, Burrin RH (1984) Control of flow separation by sound. AIAA paper, 84–2298
Beer S, Seitz P (2005) Real-time tomographic imaging without x-rays: a smart pixel array with massively parallel signal
processing for real-time optical coherence tomography performing close to the physical limits. Research in
Microelectronics and Electronics, 2005 PhD, 2, pp. 135-138
Blackwelder RF, Kaplan RE (1976) On the wall structure of the turbulent boundary layer, J. Fluid Mech. (1976), vol.
76, part 1, pp. 89-112
Defaÿ E, Millon C, Malhaire C, Barbier D (2002) PZT thin films integration for the realization of a high sensitivity
pressure microsensor based on a vibrating membrane, Sensors and Actuators A 99 (2002), pp. 64-67
Di Giovanni M (1982) Flat and corrugated diaphragm design handbook. Dekker Verlag (1982), ISBN 0-8247-1281-1
Fabula T, Buettgenbach S (1997) Analytical and Finite- Element Modeling of Resonant Silicon Microsensors, Sensors
and Material, 9(8): pp. 501-519, 1997
Greenblatt D, Wygnanski IJ (2000) The control of flow separation by periodic excitation, Progress in Aerospace
Sciences 36 (2000), pp. 487-545
Kinsler LE, Frey AR, Coppens AB, Sanders JV (1982) Fundamentals of Acoustics, John Wiley & Sons, Third Edition
(1982)
Liu T, Sullivan JP (2005) Pressure and Temperature Sensitive Paints. Springer, 2005, ISBN 3-540022241-3
Manthey W, Kroemer N, Magori V (1992) Ultrasonic transducers and transducer arrays for applications in air. Meas.
Sci. Technol., (3): pp. 249-261, 1992
Neudeck GW, Spitz J, Chang JCH, Denton JP, Gallagher N (1996) Precision crystal corner cube arrays for optical
gratings fromed by (100) silicon planes with selective epitaxial growth, Applied Optics, Vol. 35, No. 19, July 1996,
pp. 3466-3470
Nishri B, Wygnanski I (1998) Effects of Periodic Excitation on Turbulent Flow Separation from a Flap, AIAA Journal,
Vol. 36, No. 4, April 1998, pp. 547-556
Prochazka L, Roesgen T (2005) Remote Sensing of Pressure Distributions on Aerodynamic Models. Technical
Proposal, Institute of Fluid Dynamics (IFD), Swiss federal Institute of Technology (ETHZ), Zurich, Switzerland
Rafiq M, Wykes C (1991) The performance of capacitive ultrasonic transducers using v-grooved backplates. Meas. Sci.
Technol., (2): pp. 168-174, 1991
Reimer K, Engelke R, Hofmann U, Merz P, Kohlmann-von Platen KT, Wagner B (1999) Progress in gray-tone
lithography and replication techniques for different materials, Proceedings of SPIE, Vol. 3879, August 1999, pp.98-
105
Schlichting H, Gersten K (2000) Boundary-Layer Theory, Springer-Verlag, 2000, ISBN: 3-540-66270-7
Sekimura M (1999) Anisotropic Etching of Surfactant-Added TMAH Solution, Micro Electro Mechanical Systems,
Twelfth IEEE International Conference in Orlando, Florida, 1999, pp. 650-655
Sinha SK (2001) Flow Separation Control with Microflexural Wall Vibrations, Journal of Aircraft, Vol. 38, No. 3, 2001
Smits JG, Tilmans HAC, Lammerink TSJ (1985) Pressure Dependence of Resonant Diaphragm Pressure Sensor, Third
International Conference on Solid-State Sensors and Actuators", 1985, pp. 93-96.
Stemme G (1991) Resonant silicon sensors, J. Micromech. Microeng, 1 (1991), pp. 113-125
Uttamchandani D, Thornton KEB, Nixon J, Culshaw B (1987) Optically excited resonant diaphragm pressure sensor,
Electronics Letters, 23(4), 1987
Yarusevych S, Kawall JG, Sullivan PE (2006) Airfoil Performance at Low Reynolds Numbers in the Presence of
Periodic Disturbances, Journal of Fluids Engineering, May 2006, Vol. 128, pp. 587-595
Yashchuk VV, McKinney WR, Takacs PZ (2007) Binary pseudo-random grating as a standard test surface for
measurement of modulation transfer function of interferometric microscopes, Proceedings of SPIE, Vol. 6704,
September 2007