Optical, Confocal, Fluorescence, and Two- Photon...
Transcript of Optical, Confocal, Fluorescence, and Two- Photon...
1
Optical, Optical, ConfocalConfocal, , Fluorescence, and TwoFluorescence, and Two--
Photon MicroscopyPhoton Microscopy
Dr. McShaneMSE 505
April 28, 2004
LOUISIANA TECH UNIVERSITYINSTITUTE FOR MICROMANUFACTURING / BIOMEDICAL ENGINEERING Outline
• Optical Microscopy• Conventional Microscopes• Fluorescence Microscopy• Scanning Microscopes• Confocal Systems• Two-Photon Microscopy• Near-field Scanning Optical Microscopy• Comparisons
NOTE
• This presentation is not mathematically rigorous…
• Discussion and comparison of optical microscopy can be done by considering the instrumentation (optical/ mechanical/ electrical system configuration used to form images) and impact on resolution
Magnification
The aim of using a microscope is not to magnify an imagebut to see finer details in the image.
Image: scales on the wing of a mosquito – identical mag
2
Resolving Ability
The performance of the microscope is expressed as its “resolving power”, the ability to separate (“resolve”) fine details.
Components of Microscope
1. Foot2. Limb3. Tube4. Objectives5. Eyepiece6. Coarse focus7. Fine focus8. Stage9. Turret10.Mirror11.Condenser12.Condenser iris
diaphragm13. Swing-out filter ring14. Control for oblique illumination
Modern Light Microscope Magnification in MicroscopesMagnification of the image takes place in 2 stages: in the objectives and in the eyepieces. Both have their proper magnification engraved.
Magnification of objectives ranges from 3X to 100X, and that ofeyepieces from about 5X to about 15X. The total magnificationthus ranges from about 15X to 1500X.
Exercise: at maximum magnification, how large will a10nm structure appear?
If you have two 10nm structures separated by 10nm,will you be able to see them?
3
Optical Aberrations - Chromatic Corrections for Chromatic Aberrations
Spherical Aberrations Longitudinal / Transverse Aberrations
4
Astigmatism Astigmatism Aberration
Coma - Geometrical Aberration Note on Aberrations
THE SMALLER THE OBJECT, THE GREATER THE IMPACT OF ABERRATIONS ON IMAGE QUALITY
5
Wave Nature of Light Constructive Interference
Young’s Double Slit Experiment Diffraction of Light - Effects of Apertures
The wave pattern that passes the slit can be constructed by representing the wave front in the slit as a collection of point sources all emitting inphase. Diffraction can be considered as “interference” between this collection of point sources.
6
Fringe Separation
Fringe separation is expressed in terms of angular separation.Sin θ = λ / W (as W decreases, θ increases)
θ = λ / W, when W is not too small
For circular apertures, the diffraction pattern is also circular.The angular separation between the central maximum and thefirst dark ring is given by,
Sin θ = 1.22 λ / D (D is the diameter of the aperture)
θ = 1.22 λ /D (for large D or small L/D)
Diffraction and Light Intensity Distribution
Numerical Aperture Numerical Apertures
7
Numerical Aperture - Airy Disc size The Point Spread Function
• Mathematically, airy disks can be described as a PSF– PSF – output of an imaging
system for an input point source
– General Form
• J1 = Bessel function
www-inst.eecs.berkeley.edu/~ee243/ sp03/lectures/ps_mc.pdf
Airy Disks and the Point Spread Function
• Pinhole at x1
– E-field at any x2 is dependent only on distance:
– Light intensity = (above function) x (complex-conjugate) = airy image intensity distribution function
Rayleigh Criterion
A point source imaged by a lenswill not be imaged to a point, butto a diffraction pattern (Airy disk)of that point source withthe first dark ring having aradius of 1.22 λf/D.
θR = 1.22 λ/D
Rayleigh criterion:Two point sources are just resolvable if the maximum of the diffraction pattern of one point source falls on the first dark ring of the pattern of the second point source.
8
Resolution
� The limit of resolution is the smallest separation at whichtwo points can be seen as distinct entities.
� Resolution in light microscope is limited primarily by thewave nature of the light.
The relationship between resolution and wavelength is given by Abbe’s equation: d = 0.612 λ / NA
� There are several equations that have been derived to express the relationship between Numerical Aperture, wavelength, and resolution:
R = λ / 2(NA)R = 0.61 λ / NAR = 1.22 λ / (NA(obj) + NA(cond))
Resolution Versus Wavelength
NA~0.95
Depth of Field/ Image Depth
One important aspect to resolutionis the axial resolving power of anobjective, which is measured parallel to the optical axis, and isoften referred to as DOF (d).
Thickness of object space in focus is inversely proportional to resolution
2
22
)()(
NA
NAnd
−= λ
Conventional Microscopy
• Illuminator– light source and condenser
• Collecting Lens (Objective)• Eye Piece• Detector (Eye, Camera)• No scanning req’d
– Simple– Inexpensive
9
Fluorescence Microscopy
• Sometimes we cannot resolve the tiny nanostructures that we make or use, do to size or optical contrast…
• …but if we can specifically “tag” them with another molecule that provides a unique signature, we can use that to determine presence and image distributions
Fluorescence SpectroscopyFluorescence Spectroscopy
• Emission of Photons from Excited States
• Sensitive, Specific• Large choice of dyes
– “Labeling,” “tagging”– Immunostaining, etc.
hhννexex
RelaxationRelaxation
hhννemem
EXEX EMEM
WavelengthWavelength
Ext
inct
ion
Ext
inct
ion
Fluor. IntensityFluor. Intensity
Quantum Yield (QY)Quantum Yield (QY)
# fluorescent photons emitted# fluorescent photons emitted
# photons absorbed# photons absorbed
Epi-illumination
• Illuminate from same side as view
• Less loss due to propagation through turbid samples
Fluorescence Microscopy
Epifluorescent Illumination of Upright Microscope
Epifluorescent Illumination of Inverted Microscope
10
“Hot” and “Cold” Mirrors Fluorescence Microscopy
http://micro.magnet.fsu.edu/primer/java/lightpaths/ix70fluorescence/index.html
Point Imaging Microscopy
• Illuminator– light source and condenser
• Collecting Lens (Objective)
• Single detector with pinhole, scanned across fully illuminated sample
• Slower, but only single detector required for digital image acquisition
Scanning Microscopy
• Illuminator– Point source and
objective (projector)– Spot scanned across
specimen
• Collecting Lens• Single Detector
• Slower image formation
• Reduced overall exposure
11
Confocal Scanning Microscopy
• Illuminator– Point source and
objective (projector)– Scanned across sample
• Collecting Lens (Objective)
• Point Detector• Pinhole matches source
pinhole (conjugate)
• Both lenses play key roles in image quality
• Slower, more complex instrumentation
• Higher resolution, optical sectioning
The Confocal Principle - Summary
• Provides…– Increased resolution– Optical sectioning capability
• Confocal Necessities– Laser, Light Source Pinhole, Dichromatic
Mirror, Objective Lens, Specimen, Detector Pinhole, Photodetector
Theory of a Confocal System
• Light is directed through objective to object points on specimen near the focal point of lens
• Light is reflected back through objective, or through a second lens, to a pinhole aperture covering a photodetector.
• The pinhole is CONjugate to the FOCAL point of the lens
– Therefore, only the part of the specimen located at the focal point of the lens is detected
• Scanning is achieved via galvanometer mirrors, or translating the stage of the specimen
Theory of a Confocal System
• In-focus plane reaches detector through conjugate pinhole• Out-of-focus plane not able to reach detector
Tilted IC� Blurred Image
Confocal� only in-focus
light accepted
12
Resolution in Confocal Microscopy
• Lateral resolution: small advantage– Pinhole � 0– R�
• Axial resolution: major advantage– Reflected light:
– Photoluminescence: 2
77.1
NA
λ
2sin
22.0
2 θλ
n
NA2
22.1
14
1 λ
Comparing with Standard Optical
Standard ConfocalLateral
Best (µm) Typical (µm)
0.013640.0151550.01705
0.0189440.0208390.0227330.0246270.026522
0.1252170.13913
0.1565220.1739130.1913040.2086960.2260870.243478
Axial
resolution (�m)wavelength (nm) Conventional Confocal
360 0.54442344 0.381096400 0.604914934 0.42344450 0.680529301 0.476371500 0.756143667 0.529301550 0.831758034 0.582231600 0.907372401 0.635161650 0.982986767 0.688091700 1.058601134 0.741021
Confocal provides ~30% better resolution, in addition to optical sectioning capability
Confocal Layouts - Minsky
• From Minsky (US Patent 03013467).
Confocal Layout
13
Typical CLSM Layout The Need for a Fluor-Confocal System
• Out-of-focus fluorescence overwhelms in-focus details.
• CLSM uses high intensity laser/pinhole.– Intensity decreases with 3rd potency above
and below the focal plane.– Pinhole further reduces out of-focus light.
Confocal Fluorescence Microscopy
• http://www.loci.wisc.edu/images/movies/lpath.mpg
14
Ultimate Visualization The Leica TCS SP
Confocal Detection Pinhole Spectrophotometer
Prism
Photomultiplier Tubes
The Effect of Slit on PMTs
The Photomultiplier
• Measures intensity• Slit which allows specific detection
increments, while reflecting other λ’s to other PMTs
• PMT-readings are depicted on monitor
• Slits can be controlled from software:
Slits control increment of detectable wavelengths
The Dichromatic Mirror
• Dichroic mirror can split different colors for simultaneous detection of different fluorophoresin same focal plane
+ =
15
Optical Sectioning in CM
• The capability of observing selected thin layers– Less fluorescence blur– Less out-of-focus light
• Only features located within thin plane are seen Conventional
Microscopy
Optical Sectioning in CM
• Optical sectioning depends on axial resolution.– Parts of specimen several axial resolutions away
from focal plane will not be detected– Parts of specimen at different z coordinates within
one axial resolution are different intensities– Extended focus image is the sum of all optical
sections• IEF(x,y) = ∫ I(x,y,z)dz This is the
key factor for optical
sectioning
Optical Sectioning in CM Optical Sectioning in CM
Focal Plane 1: Fig. BDetects S2
Focal Plane 2Nothing Detected
Focal Plane 3: Fig. CDetects S3 and S4Note Intensities
16
Optical Sectioning in CM 3D Reconstruction of Optical Sections
Optical Sectioning in CM Optical Sectioning Examples
17
Sectioning / Reconstruction IC imaging
• http://micro.magnet.fsu.edu/primer/java/confocal/nikon/index.html
Advantages of CM
• Clearer Images – only in-focus light is detected
• Optical Sectioning– Allows 3D reconstruction
• Increased Sensitivity – PMT for low intensities– Controlled excitation intensity to control
photobleaching (fluorescence)
• Reduced Photobleaching– reduced due to point-illumination
Advantages of CM
• Three-four Dimensional Measurements– through optical sectioning
• Scan through z by changing focal plane
– can use time as a variable
• More Accurate Quantification• Multiple Simultaneous Analyses
18
Photobleaching in CM
• CM Enemy #1• Confocal Pattern of Photobleaching
– Photobleaching is proportional to illumination at thatplane
• Illumination level = (photons/µm2) x time• Therefore, assuming absorption of photons is low (number of
photons is constant/plane), as µm2 increases, severity of bleaching decreases.
Photobleaching in CM
• However, as the illuminating cone scans, less severely bleached planes are naturally illuminated for longer periods of time– Volume of space that is equally photobleached is in the shape of
an octahedron, defined by the convergence angle of the beam (half-angle of objective), α
• Note, during reversal of scan, increased time causes increased photobleaching
Focal Plane
Resolution in CM
• Via Point of Illumination
• Via Pinhole
Influence of Pinhole Size in CM
• Airy Disk – disk seen around image of a point source due to diffraction of light– Governs size of pinhole
• Confocal pinhole allows elimination of out-of-focus light!
• Pinhole too big – interference by out-of-focus light
• Pinhole too small – loss of in-focus light � dimmer image
Image obtained at focal plane
Images obtained away from focal plane
19
Airy Disks and Pinhole Size in CM
• Intensity of Light as a Function of Radius
Overexposed picture of airy disk; note secondary ring
Optimization of light
Secondary ring
Airy Disks and Resolution
• 2 image points are resolvable if their distance is larger than the radius of the airy disk– Therefore, smaller airy radii allow smaller
distances to be resolved
• Confocal rAiry = 0.4λo/NAobj
– NAobj = numerical aperture of the objective lens– λo = wavelength of light in a vacuum
• If pinhole…– <0.5 rAiry, x-y resolution is improved by 40%, but
signal level reduced by 95%!• As pinhole increases, resolution is reduced
– = rAiry, total focused signal captured, but resolution about 10%
PSF – Key to confocal resolution A Virtual Confocal Instrument
http://micro.magnet.fsu.edu/primer/virtual/confocal/
20
Cellular Imaging Confocal Applications
• Latex or silica nanoparticlesserve as template
• Thin films assembled on surface• Inner core removed (acid)• Polyion “shell” acts like cell
membrane Confocal & AFM images5µm shell, 20nm wall (PSS/PAH)4
Shells Produced by Core Removal
Template Dissolution Fragment migration
5µm
5µmNanocapsule from SiO2 core
Decomposition Expts/Results• Capsule walls: {PVS/PAH}4 and
{PVS/PAH}2/{PSS/PAH}2
• Monitored over 15 min w/ 45 sec between images
{PVS/PAH}4
{PVS/PAH}2{PSS/PAH}2
Added stability through the addition of PSS/PAH layers
Confocal Transmission Imaging
• Fluorescent Recovery After Photobleaching(FRAP)– After bleaching area of interest, replacement of the
fluorochrome is a measure of diffusion.
Confocal Applications
21
FRAP analysis for diffusion measurements
• Shells loaded 24hr prior to CLSM
• Cover slip sealed to prevent loading solution evaporation
• Data collected over various time periods, depending on Mw.
• Area selected for analysis• 4 data sets collected for
each capsule/Mw combination
FRAP PVS/PAH 4400mw
0
50
100
150
200
250
0 100 200 300 400
Time (sec)
Flu
ore
sc
enc
e In
ten
sity
(AU
)
Capsule Interior
Capsule Exterior
Confocal Applications
• Fluorescence Resonance Energy Transfer (FRET)– Excitation does not release photon, but transfers
energy to nearby molecule– Detects changes in molecular proximity– Can detect changes in conformation of a protein
Two-Photon Microscopy
• Elimination of out of-focus fluorescence• Two photons’ energies combine to excite a fluorophore
1-Photon Excitation 2-Photon Excitation
λ1
λ2
λ3
λ3 = (1/ λ1 + 1/ λ2)-1
Multiphoton Physics
• http://microscopy.fsu.edu/primer/java/multiphoton/jablonski/index.html
22
Two-Photon Microscopy
• Minimization of Photobleaching
http://micro.magnet.fsu.edu/primer/java/multiphoton/excitationbleaching/index.html
Two-Photon Microscopy
• 2-photon absorption leads to more localized excitation• Note:1-P intensity constant above and below focal plane• Note: 2-P intensity maximized more locally (A)
A
http://microscopy.fsu.edu/primer/java/multiphoton/excitationregion/index.html
Multi-photon Excitation
• Superposition of energy of photons arriving simultaneously• 2-P Absorption cross-section ~10-50 cm4s/photon-molecule
• C = concentration (molecules/cm3)
• δ = 2-P cross section• P = average laser power (photons/sec)• l = absorption pathlength (cm)• A = cross-sectional area of beam (cm2)
• ΦF = Fluorescence quantum yield (QY/2)
econd)(photons/s 2
22 A
lPKCFl F
h
Φδ=ν
Multi-photon Excitation
23
Pawley. Handbook of Biological Confocal Microscopy. 1995
Two-Photon Microscopy
• Pulsing increases two-photon absorption probability• Probability of absorption:
na ≈ δ⟨P⟩2Fp-1(πANA
2/hcλ)2ξ
δ = 2-photon cross section
⟨P⟩ = average power
Fp = repetition frequency
ξ = “advantage” factor
ξ = ⟨p2⟩/⟨p⟩2 = (t1 – t2) ∫ p2(t)dt / (∫ p(t)dt)2
(Integrals over t1 to t2)
Advantages of TPE
Decreased photobleaching: bleaching only occurs on the focal plane
True optical section: fluorescence excitation/emission only from focal plane
Deeper penetration: elimination of using lower wavelength light (higher scattering)
Versatility: tunable excitation wavelength (720-930nm)..one laser for many fluorophores
TPE Manufacturers/Vendors
• Conneaut Lake Scientific, Hartstown, PA 16131 Phone: 814-382-1604, Fax: 814-382-8349 http://www.conneautlakescientific.com
• JASCO, Inc., 649 Commerce Dr., Easton, MD 21601 Phone: 410-822-1220, Fax: 410-822-7526 http://www.jascoinc.com
• Micro Photonics, P.O. Box 3129, Allentown, PA 18106-0129 Phone: 610-366-7103, Fax: 610-366-7105 http://www.microphotonics.com
• VayTek, Inc, 305 West Lowe Ave. P.O. Box 732, Fairfield, IA 52556 Phone: 641-472-2227, Fax: 641-472-8131 http://www.vaytek.com
• WITec Instruments Corp, 101 Tomaras Ave., Savoy, IL 61874 , Fax: 217-352-6655 http://www.WITec-Instruments.com
Near-Field Scanning Optical Microscopy
(NSOM, SNOM)
24
Introduction
Near-field scanning optical microscopy (NSOM) is a technique that can achieve spatial resolution performance beyond the
classical diffraction limit by employing a sub-wavelength light source or detector positioned in close proximity to a specimen.
NSOM is also one of the scanning probe techniques…It is a combination of optical microscopy and SPM.
NSOM probes – used for imaging and metrology
Optical Diffraction Limit
Optical systems of any kind that use lenses or mirrors to form an image are limited in their spatial resolution, even with the best designs. Diffraction limit or Abbé's limit:
To increase resolutionDecrease wavelength- X-ray, electron microscopyIncrease NA- immerse the lens into oil
The typical resolution of conventional optical microscopy is about
half the wavelength; for argon laser light (blue/green line, λ= 488 nm)the resolution is ~250nm.
NA/61.0 λδ =
Near-field Optical Microscopy
Light in the visible range is diffraction limited on lengthscales of about 1 micron.
To circumvent this diffraction limit and obtain true nm-scalespatial resolution, a near-field system is used to scan a small 100 nm aperture positioned very close to the surface of interest.
This aperture couples to the high spatial-frequency (evanescent)modes of light that decay exponentially from the surface and that are thus never seen with traditional optical methods.
Principles of NSOM
Fourier optics analysis shows that the diffraction limit on resolution in
optical microscopy is the fundamental constraint for far-field light…
but has little effect on the near-field light.
So, if we can “see” the near-field part of light, we can get images with
resolution dependent on the probe size and the probe-to-sample separation.
That is the key aspect of NSOM.
Both probe size and probe-to-sample separation can, in principle, be made
much smaller than the wavelength of light by micromachining technology.
So we can get resolution beyond the diffraction limit.
25
Principle of NSOM
Light from an object can consist of two components: the far-field and the near-field light.
Conventional optical microscopy can only “see” the far-field light. So we can also call it “far-field microscopy.”
NSOM (or SNOM?)
Uses the interaction of a sharp (sub-λ) probe tip with near-fieldlight on a sample or sub-λ aperture near-field light source and/or detector, in order to image the surface at sub-λ opticalresolution.
The spatial resolution is determined by the size and shape of the probe tip.
SNOM employs nanometer precision piezoelectric raster-scanning together with nm-sized sharp probes to obtain lightoptical images at higher resolution
Far-field vs. near-field
Since SNOM can provide very high spatial resolution, one canapply this method to chemical and structural characterizationby the addition of spectroscopic analysis.
The light field at the surface of an object actually contains moreinformation – higher spatial frequencies- than we can imageby using a far-field lens system. Only the spatial frequenciesthat reach the imaging lens (pass through the NA) are “seen”.These are the propagating, low frequencies.Higher spatial frequencies exist at the sample surface, but decay exponentially within a distance less than the wavelength,so never reach the detector.
General NSOM System
26
Internal Configuration NSOM Probe – Sample Interaction
http://www.chem.ukans.edu/rdunngroup/nsom.html
Fabrication of NSOM Probes Modes of Operation
27
Sample-Probe-Detector Configurations Topographic and Near-field Images
NSOM – Modified Configuration SNOM vs. STM and AFM
28
NSOM Manufacturers/Vendors
• Conneaut Lake Scientific, Hartstown, PA 16131 Phone: 814-382-1604, Fax: 814-382-8349 http://www.conneautlakescientific.com
• JASCO, Inc., 649 Commerce Dr., Easton, MD 21601 Phone: 410-822-1220, Fax: 410-822-7526 http://www.jascoinc.com
• Micro Photonics, P.O. Box 3129, Allentown, PA 18106-0129 Phone: 610-366-7103, Fax: 610-366-7105 http://www.microphotonics.com
• VayTek, Inc, 305 West Lowe Ave. P.O. Box 732, Fairfield, IA 52556 Phone: 641-472-2227, Fax: 641-472-8131 http://www.vaytek.com
• WITec Instruments Corp, 101 Tomaras Ave., Savoy, IL 61874 , Fax: 217-352-6655 http://www.WITec-Instruments.com
Other Resources
• http://micro.magnet.fsu.edu/primer/resources/confocal.html
• http://microscopy.fsu.edu/primer/resources/multiphotonweb.html