Operator’s Manual
AGH6200/AGH6200F
Version:3.3.1901
1
AIYI Technologies
Dear customer:
Glad to have your trust and support on AIYI Technologies, we will
provide you best product and service in return.
As an ISO certified manufacturer. AIYI Technologies has been focus on
safety and environment many years, we provide you gas & dust detec-
tor and systems. The design and manufacture of product is strictly
follow the international standard and company regulations, and each
product get a normative QC control to ensure the best quality for you.
Please read and understand this operator’s manual before operating
instrument. Improper use of the gas monitor could result in bodily
harm or death. Please don't hesitate to contact us if you have any
questions or suggestions. Thanks!
We are honored to have the opportunity to serve you.
Yours sincerely.
2
AIYI Technologies
Copyright Statement
Copyright© 2017 AIYI Technologies .All Rights Reserved.
This document is the property of AIYI Technologies. And can be nei-
ther reproduced nor disclosed to a third party without a written authorization.
is the registered trademark of AIYI Technologies. The others inside of the
document belongs to other companies.
Disclaimer
This manual, as well as the software described in it, is furnished under license and may be used or
copied only in accordance with the terms of such license. The content of this manual is furnished
for informational use only, is subject to change without notice, and should not be construed as a
commitment by AIYI Technologies. (AIYI Technologies ) AIYI Technologies assumes no responsibility
or liability for any errors or inaccuracies that may appear in the manual.
Except as permitted by such license, no part of this publication may be reproduced, stored in a
retrieval system, or transmitted, in any form or by any means, electronic, mechanical, recording, or
otherwise, without the prior written permission of AIYI Technologies.
AIYI TECHNOLOGIES MAKES NO WARRANTIES, EXPRESS OR IMPLIED, INCLUDING WITHOUT LIMITA-
TION THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE,
REGARDING THE AIYI TECHNOLOGIES INSTRUMENT. AIYI TECHNOLOGIES DOES NOT WARRANT,
GUARANTEE, OR MAKE ANY REPRESENTATIONS REGARDING THE USE OR THE RESULTS OF THE USE
OF THE INSTRUMENT IN TERMS OF ITS CORRECTNESS, ACCURACY, RELIABILITY, CURRENTNESS, OR
OTHERWISE. THE ENTIRE RISK AS TO THE RESULTS AND PERFORMANCE OF THE INSRTUMENT IS
ASSUMED BY YOU. THE EXCLUSION OF IMPLIED WARRANTIES IS NOT PERMITTED BY SOME STATES.
THE ABOVE EXCLUSION MAY NOT APPLY TO YOU.
IN NO EVENT WILL AIYI TECHNOLOGIES, ITS DIRECTORS, OFFICERS, EMPLOYEES, OR AGENTS BE
LIABLE TO YOU FOR ANY CONSEQUENTIAL, INCIDENTAL, OR INDIRECT DAMAGES (INCLUDING
DAMAGES FOR LOSS OF BUSINESS PROFITS, BUSINESS INTERRUPTION, LOSS OF BUSINESS INFOR-
MATION, AND THE LIKE) ARISING OUT OF THE USE OR INABILITY TO USE THE AIYI TECHNOLOGIES
INSTRUMENT EVEN IF AIYI TECHNOLOGIES HAS BEEN ADVISED OF THE POSSIBILITY OF SUCH DAM-
AGES. BECAUSE SOME STATES DO NOT ALLOW THE EXCLUSION OR LIMITATION OF LIABILITY FOR
CONSEQUENTIAL OR INCIDENTAL DAMAGES, THE ABOVE LIMITATIONS MAY NOT APPLY TO YOU.
3
AIYI Technologies
Statement
The manual is applicable to the use and maintenance of AGH6200 & AGH6200F series gas detector.
This manual is suitable for the following personnel: instrument maintenance engineers, field users.
Attention: Before connecting and operating your device,
please be advised of the manual.
4
AIYI Technologies
CONTENT
Chapter 1: Product Introduction
1.1 Brief introduction …………………………………………………………05
1.2 Instrument description……………………………………………………07
1.3 Interface description………………………………………………………08
1.4 Specifications………………………………………………………………09
Chapter 2: Operation
2.1 Packing list ………………………………………………………………10
2.2 Power on/off………………………………………………………………10
2.3 Daily use …………………………………………………………………11
2.4 Operation …………………………………………………………………12
2.5 Charging …………………………………………………………………17
2.6 Data upload ………………………………………………………………17
Chapter 3: Maintenance
3.1 Cautions …………………………………………………………………17
3.2 Warranty …………………………………………………………………18
5
AIYI Technologies
Chapter 1 Product Introduction
1.1 Brief introduction
The AGH6200 and AGH6200F portable gas detector are independently developed and produced by AIYI Technologies, which can be used in industrial sites to detect the risks of combustible gases, carbon monoxide, hydrogen sulfide and oxygen deficiency. The overall design of the product is compact, which greatly improves the comfort of the customer. The high-precision sensor is used for accurate detection and easy mainte-nance. The IP66 protection grade can be applied to all kinds of bad occasions.
Note: The AGH6200F is a portable combustible gas detector with a measuring range of 0 to 100% LEL. The AGH6200 is a portable 4 gas detector, the following are referred to as detectors.
The main functions and features of the product:
▎Detect combustible gas, oxygen, carbon monoxide and hydrogen sulfide.
▎Imported sensors with high sensitivity.
▎Various alarm modes of sound and light and vibration, effectively alert.
▎Double case design, IP66 protection, dust and water proof .
▎Intrinsically safe design for use in hazardous locations.
▎Standard Micro USB interface for easy charging and data export.
▎Compact and lightweight, easy to carry.
The product design, manufacture, testing follow the below standards:
The AGH6200F is only equipped with a combustible gas sensor; the AGH6200 is equipped with four gas sensors: combustible gas, oxygen, carbon monoxide and hydrogen sulfide. Except for different gases detected, the two operations are exactly the same. The following is an example of AGH6200. Unless otherwise specified, AGH6200F is not separately described.
6
AIYI Technologies
▎GB 15322.1-2003 Combustible gas detectors--Part 3:Portable type detectors
for 0-100% LEL combustible gas
▎GB 3836.1-2010 Explosive atmospheres—Part 1:Equipment—General require-
ments
▎GB 3836.2-2010 Explosive atmospheres—Part 2:Equipment protection by
flameproof enclosures“d”
▎GB 3836.4-2010 Explosive atmospheres—Part 4:Equipment protection by
intrinsic safety i
▎GB 50493-2009 Specification for design of combustible gas and toxic gas de-
tection and alarm for petrochemical industry
▎GB 12358-2006 Gas monitors and alarms for workplace General technical
requirements
▎GB 16838-2005 Envionmental test methods and severities for fire electronic products
▎GB/T 4208-2017 Degrees of protection provided by enclosure(IP code)
▎GB/T 18664-2002 Selection, use and maintenance of respiratory protective
equipment
▎GBZ 2.1-2007 Occupational exposure limits for hazardous agents in the work-place Part1:Chemical hazardous agents
▎Q/3201 AECN 009-2018 AGH6200 portable gas detector
▎Q/3201 AECN 010-2018 AGH6200F portable gas detector
1.2 Instrument description
7
AIYI Technologies
1.3 Interface de-scription
1 Indicator light 2 LCD screen 3 Button
4 Calibration kit adapter 5 Sensors 6 Buzzer
7 Alarm LED light 8 Alligator clip 9 Nameplate
Function Choose, Exit
Power ON/OFF, Menu, Confirm
Fault Yellow light keeps on
Alarm Red light flashes
Power Green light flashes
Function Power
8
AIYI Technologies
1 Task bar
Zero calibration
Span calibration
Self-test
USB connection
Power
2 Status bar
Gas detected, eg: Ex O2 CO H2S
STEL Short term exposure limit
TWA Time weighted average
Low alarm
High alarm
3 Readings bar
Display the current gas concentration
5 Time bar
Display current time
4 Unit bar
%LEL lower explosive limit
ppm Parts per million
% Percentage
μmol/mol One millionth mol gas per
mol air
note:1ppm = 1μmol/mol
9
AIYI Technologies
1.4 Specifications
Model AGH6200F AGH6200
Gas detected Ex Ex O2 CO H2S
Measure range Ex: catalytic combustion CO: electrochemical O2: electrochemical H2S: electrochemical
Detection principle Ex: 0~100%LEL CO: 0~1000µmol/mol
O2: 0~30%VOL H2S: 0~100µmol/mol
Drift Ex: ±5%F.S CO: Absolute error ±5%µmol/mol or relative error ±10% O2: ±3%F.S H2S: ±5µmol/mol
Response time ≤30s (T90)
Sampling method Diffusion sampling standard
Alarm method Sound, light (3 LEDs), vibration
Operating temperature -40~70℃ -10~55℃
Operating humidity ≤95%RH
Explosion-proof grade ExdibIICT4 Gb ExdibIICT4 Gb
Ingress protection IP66
Battery Rechargeable 3.7v lithium battery in 2200mAh
Charge Voltage DC5V
Continuous operation ≥18h
Case material Antistatic TPU + Antistatic PC
Dimension 132mm*63mm*48mm(including alligator clip)
Weight About 270g About 285g
10
AIYI Technologies
Chapter 2 Operation
2.1 Packing list
Please confirm if the package is complete before use. If there is any loss, please con-tact us as soon as possible. Normally, the gas detector contains the following products and accessories:
▎Power on: press and hold the power button, the detector enters the boot pro-cess, and the full screen flashes 3 times during the boot process, and then displays the STEL alarm value, TWA alarm value, alarm-1 value, alarm-2 value and calibration value of each channel in turn.
Item name Quantity
Item name Quantity
Item name Quantity
Detector 1 Calibration certificate &
QC certificate 1 Calibration adapter 1
Manual 1 Charger(including USB
Cable) 1
Full screen interface STEL alarm value TWA alarm value
Alarm-1 value Alarm-2 value Calibration value
2.2 Power on/off
11
AIYI Technologies
▎Power off: Press and hold the power button to enter the shutdown interface. All the indicator lights are turned off, the detector enters the shutdown countdown, and the display is turned off from OFF 004 to OFF 000. After the countdown ends, the display is turned off, and the Green indicator keep flashing for about 10 seconds, then the instrument is turned off.
2.3 Daily use
▎Factory reset: After the boot process is displayed, the detector enters the facto-ry reset interface. If you need to restore the factory settings, press the function button twice in rapid succession, otherwise the detector will automatically enter the main interface.
▎Main interface: The main interface is mainly used to display: gas name, gas concentration, alarm type, time and battery status. The green indicator flashes at this time.
▎Fault indication: When the gas sensor is shorted or broken, the corresponding detection channel has a fault indication, as shown in the figure below. At this point, the green indicator is off and the yellow fault indicator is keeps on.
Shutdown interface
Factory reset interface Main interface
12
AIYI Technologies
▎Alarm indication: When the gas concentration in the environment is more than alarm value, the detector will give an alarm. At this point, the alarm indicator flashes, the buzzer sounds, and the detector vibrates. The STEL, TWA, alarm-1 and alarm-2 are shown below.
2.4 Operation
▎Menu:
1.Press and hold the power button to enter the shutdown interface,as shown in
the figure below.
2.During the shutdown countdown process, press the power and function button
at the same time to enter the menu. Menu is shown in the figure below.
3.Short press the function button to select the zero calibration, span calibration
or self-test. The selected icon will flash.
4.Press and hold the function button to exit the menu interface and return to the
main interface. Main interface is shown in the figure below.
Fault indication (LEL channel fault) Main interface
STEL alarm TWA alarm Alarm-1 Alarm-2
13
AIYI Technologies
▎Zero calibration:
1.After entering the menu, short press the function button to select the zero
calibration, the zero calibration indicator flashes, then short press the power button to enter the zero calibration interface. At this time, the zero calibration indicator keeps display, and the span calibration indicator and self-test indicator are not displayed. Zero calibration interface is shown in the figure below.
2.Short press the function button to select the specific channel, and the gas
name keep flashes. 3. Short press the power button to do zero calibration. If the calibration is com-pleted, the “End” prompt will be displayed. If the calibration fails, the “Err” prompt will be displayed, as shown in the figure. 4. Press and hold the function button to exit the zero calibration operation and return to the main interface.
Shutdown interface Menu interface Main interface
Menu interface Zero calibration Calibration completed Calibration failed
14
AIYI Technologies
▎Span calibration:
1.After entering the menu, short press the function button to select the span
calibration, the span calibration indicator flashes, the zero calibration indicator and the self-test indicator keep displayed, then short press the power button to enter the span calibration interface. At this time, the span calibration indicator keeps displayed, and the zero calibration and self-test indicator are not displayed. Span calibration interface is shown in the figure below.
2.Short press the function button to select the specific channel, and the gas
name in the selected channel flashes.
3.Short press the power button to do span calibration. If the calibration is com-
pleted, the “End” prompt will be displayed. If the calibration fails, the “Err” prompt will be displayed, as shown in the figure.
4.Press and hold the function button to exit the span calibration and return to
the main interface.
▎Self-test:
1. After entering the menu, short press the function button to select the self-test, the self-test indicator flashes, the zero calibration indicator and span calibration indicator keeps displayed, then short press the power button, the detector enters self-test.
2. The detector will display all the contents in turn, then alarm with sound, light,
vibration once. as shown in the figure.
3.After the self-test is completed, back to the menu interface.
Menu interface Span calibration Calibration completed Calibration failed
15
AIYI Technologies
▎Bump test:
1.Refer to indication diagram of calibration adapter and the ventilation, con-
necting the calibration adapter to the detector tightly, and passing the gas of the specified concentration from the vent of the calibration adapter.
Self-test interface Self-test interface
Menu interface Self-test interface Self-test interface
16
AIYI Technologies
2.In the main interface, press and hold the function button to enter the bump
test preparation interface. At this time, the self-test indicator flashes, short press the power button to enter the bump test.
3.Short press the function button to select the quick detection zone and the gas
name indicator flashes.
4.After a stable readings, press the power button to enter bump test, and the gas
name flashes. About 15 seconds later, if the test is normal, the concentration value will flash for 3 seconds, and then the "End" prompt will be displayed. If the test is not normal, the “Err” prompt will flash for 3 seconds, and then the gas concentration will be calibrated for processing. If the calibration is successful, the correct concentration value will be displayed for 3 seconds, and then the “End” prompt will be displayed. If the calibration fails, the “Err” prompt will flash for 3 seconds, then the “End” prompt will be displayed, followed by the sensor fault as “---”.
5.The bump test is end and the detector automatically back to the main interface.
Main interface Bump test interface
Bump test completed Bump test failed Fault indication
17
AIYI Technologies
2.5 Charging
There is a power indicator in the upper right corner of the display. Please keep charge in time with original charger and cable when low power. The charging port is a standard USB charging port. When charging, the charging indicator of the detector keeps red, and there is also a charging prompt in the upper right corner of the display. The charge indicator turns green when fully charged.
2.6 Data upload
In the power on state, connect the detector to the computer with the Micro USB cable, and the computer will automatically recognize and display the correspond-ing drive letter. After double-clicking on it, you can implement data export and other operations.
Chapter 3 Maintenance
3.1 Cautions
▎Take care to prevent the detector from falling or severe vibration.
▎It is strictly forbidden to expose the detector to a gas with high concentration,
corrosive gas, etc. for a long time, otherwise it will cause a serious damage to the
sensor.
▎In order to ensure the performance of the detector, calibration should be carried
out regularly. The calibration frequency depends on the specific working conditions.
It is recommended to calibrate the detector every 3 months. The calibration record
sheet should be filled in each calibration. Considering if factors such as high concen-
tration impact and corrosive gases happen, it is necessary to shorten the calibration
period. In either case, the calibration period must not exceed one year.
▎Keep the detector clean, and if the sensor vent is blocked, it may affects the
detection sensitivity and even damage it.
Do not charge in danger areas. Please recharge the battery before it is exhausted, or it will reduce the battery life.
18
AIYI Technologies
▎Replace sensor in time once the sensor life is over. The replacement of the sen-
sor should be carried out under the guidance of the manufacturer or returned to
the factory for replacement.
▎Keep use the origin parts provided by manufacturer while the maintenance and
the replacement of the parts. Otherwise it may affect the performance and safety
of the instrument itself.
▎The operation of the detector, calibration and other maintenance work should
be carried out by qualified professionals.
▎For the faults encountered in daily maintenance, please contact the manufactur-
er in time.
3.2 Warranty
AIYI Technologies promises that the product will be guaranteed for one year from
sale and will be serviced for life. If the product fails or any problem during the use,
please feel free to contact us, we will be happy to provide you with timely and
quality service!
3.3. Trouble-Shooting
For the faults encountered during use, you can refer to the following ways to solve
the problem, or you can contact the manufacturer directly for consultation.
Common Problems Recommended Solutions
Abnormal charging Replace the charger or cable
Readings drift Calibration
Cannot connect to PC Replace the cable or replace the PC
Unable to boot Try reboot after charging, or please contact the
manufacturer
Battery low alarm Stop operation, turn off and charging
The fault light keeps on Contact the manufacturer
19
AIYI Technologies
SCIENTIFIC INNOVATION