OPERATING MANUAL MORCELLATOR - bowa-medical.com · MN009-938-S0 ERGO 300 3 1 General information...

46
OPERATING MANUAL MORCELLATOR

Transcript of OPERATING MANUAL MORCELLATOR - bowa-medical.com · MN009-938-S0 ERGO 300 3 1 General information...

Page 1: OPERATING MANUAL MORCELLATOR - bowa-medical.com · MN009-938-S0 ERGO 300 3 1 General information 1.1 Indications Morcellator set for morcellation and extraction of tissue and for

OPERATING MANUAL MORCELLATOR

Page 2: OPERATING MANUAL MORCELLATOR - bowa-medical.com · MN009-938-S0 ERGO 300 3 1 General information 1.1 Indications Morcellator set for morcellation and extraction of tissue and for

2 ERGO 300 MN009-938-S0

Table of contents 1 General information 3

1.1 Indications 3 1.2 Principle 3 1.3 Contraindications 3 1.4 Medical application 3 1.5 On receipt 4 1.6 Scope of delivery, accessories and spare parts 5 1.7 Technical Data «ERGO 300 - Morcellator» 6 1.8 Ambient conditions 6 1.9 Warranty coverage 6 1.10 Right to make technical changes reserved 6 2 Explanation of symbols 7

3 Safety information 7

3.1 EMC Manufacturer’s Declaration of Conformity 7 3.2 Modification and misuse 7 3.3 Essential requirements 8 3.4 During use 8 4 Device overview 9

4.1 ERGO 300 Morcellator system chart 11 5 Before use 12

5.1 Connection to the power supply 12 5.2 Device preparation 12 5.2.1 Assembly of the Morcellator set with protective tube (option 1) 13 5.2.2 Assembly of the Morcellator set with Trocar Cannula (option 2) 16 5.3 Mains connection 18 6 Use 19

6.1 Using the control unit 19 6.2 Operation 21 6.2.1 Option 1: morcellation with protective tube 23 6.2.2 Option 2: morcellation with Trocar Cannula 24 6.2.3 Option 3: Myomectomy 26 6.3 After use 27 6.4 Disassembling Trocar Cannula 28 6.5 Disassembly of motor 29 6.6 Disassembly of gear unit and Morcellator with protective tube 29 7 Instrument preparation 30

7.1 Control unit and pedal 30 7.2 Preparation of the Trocar Cannula 31 7.3 Preparation of motor 32 7.4 Preparation of gear unit 34 7.5 Preparation of cutting tubes, protective tubes, filling rods, seals, seal unit and

myoma drill 35 8 Maintenance and repair 37

8.1 General maintenance 37 8.2 Replacing the control unit fuse 37 8.3 Safety inspections 38 8.4 Malfunction and troubleshooting 38 8.5 Spare parts list with order numbers 39 8.6 Information on disposal 39 9 Electromagnetic compatibility (EMC) 40

Page 3: OPERATING MANUAL MORCELLATOR - bowa-medical.com · MN009-938-S0 ERGO 300 3 1 General information 1.1 Indications Morcellator set for morcellation and extraction of tissue and for

MN009-938-S0 ERGO 300 3

1 General information

1.1 Indications

Morcellator set for morcellation and extraction of tissue and for removing myomas or the uterus dur-ing laparoscopic procedures in surgery and gynaecology.

1.2 Principle

The ERGO 300 - Morcellator is a control unit for the BOWA Morcellator for use in morcellation and extraction of tissue. A rotating cylindrical tube with a blade at the distal end is inserted through an abdominal cavity that has been enlarged by an Insufflator. The tissue that is to be removed is guided to the cutting tube by a gripper, cut off and removed through the tube. This manual describes three different options for removing tissue by a laparoscopic procedure. Option 1: Tissue removal by morcellation with protective tube. Instruments (gripper) are inserted through a protective tube sealed by a valve system. The valve can be covered by a thumb for removing the gripper from the sealed protective tube to prevent gas loss. Option 2: Tissue removal by morcellation with Trocar Cannula. Instruments (cutter tube, gripper) are inserted through the Trocar Cannula. The Trocar Cannula con-tains a valve that prevents gas loss when withdrawing the instruments. Option 3: Tissue removal (myectomy) with myoma drill through the Trocar Cannula. The myoma drill is inserted through the Morcellator (protective tube or Trocar Cannula). The Trocar Cannula and the protective tube can be used with or without a cutting tube. No gas loss after re-moval of instruments with the Trocar Cannula, minimum gas loss during removal from the protective tube, sealing unit can be sealed with the thumb.

1.3 Contraindications

Do not use the Morcellator for the following:

• Treatment of malignant tumors

• Treatment of vascularized tissue

• Preparation of tissue

1.4 Medical application

This product is designed solely for use in medical facilities by medically qualified persons.

Page 4: OPERATING MANUAL MORCELLATOR - bowa-medical.com · MN009-938-S0 ERGO 300 3 1 General information 1.1 Indications Morcellator set for morcellation and extraction of tissue and for

4 ERGO 300 MN009-938-S0

1.5 On receipt

• The shipping package must contain all parts listed in 1.6 Scope of delivery.

• If any parts are damaged or missing, contact the BOWA agent from which the device was or-dered or BOWA.

Immediate inspection of the ERGO 300 - Morcellator motor system (art. no. 905-000):

1. Morcellator control unit (905-001) including 2. power cord 3 m 3. Morcellator Gear unit complete including

4. Seals (905-004) 5. Cross seal (905-009) (10 pcs.) 6. Seals for Ø 12-22 mm instruments

(905-512) (10 pcs.) 7. Micromotor with cable 3 m (905-002)

8. Fuses (2 pcs.) 9. Vario Foot switch with cable 3 m (905-003)

10. Morcellator Spray nozzle gear unit for Nou Clean Spray (905-006)

11. Morcellator Spray nozzle motor for Nou Clean Spray (905-007)

12. Handle (905-005) 13. Manual on CD-ROM

Page 5: OPERATING MANUAL MORCELLATOR - bowa-medical.com · MN009-938-S0 ERGO 300 3 1 General information 1.1 Indications Morcellator set for morcellation and extraction of tissue and for

MN009-938-S0 ERGO 300 5

1.6 Scope of delivery, accessories and spare parts

Art.-no. Description Quantity

Basic Set 905-000 Morcellator ERGO 300 basic set comprising the following products ---------------------------- 1 unit 905-001 Morcellator Control unit ERGO 300 including power cord 3 m ----------------------------------- 1 unit 905-002 Morcellator Micromotor ERGO 300 with cable 3 m ------------------------------------------------- 1 unit 905-003 Morcellator Vario Foot switch with cable 3 m --------------------------------------------------------- 1 unit 905-004 Morcellator Gear unit complete including seals ------------------------------------------------------ 1 unit 905-005 Morcellator Handle ------------------------------------------------------------------------------------------ 1 unit 905-006 Morcellator Spray nozzle gear unit ---------------------------------------------------------------------- 1 unit 905-007 Morcellator Spray nozzle motor -------------------------------------------------------------------------- 1 unit MN031-578 Manual on CD-ROM ---------------------------------------------------------------------------------------- 1 unit Application-Sets 905-612 Morcellator application set Ø 12 mm ------------------------------------------------------------------- 1 unit

905-112 Morcellator Cutting tube Ø 12 mm ---------------------------------------------------------------------- 1 unit 905-212 Morcellator Protective sleeve Ø 12 mm ---------------------------------------------------------------- 1 unit 905-012 Obturator Ø 12 mm ----------------------------------------------------------------------------------------- 1 unit 905-615 Morcellator application set Ø 15 mm ------------------------------------------------------------------- 1 unit

905-115 Morcellator Cutting tube Ø 15 mm ---------------------------------------------------------------------- 1 unit 905-215 Morcellator Protective sleeve Ø 15 mm ---------------------------------------------------------------- 1 unit 905-015 Obturator Ø 15 mm ----------------------------------------------------------------------------------------- 1 unit 905-620 Morcellator application set Ø 20 mm ------------------------------------------------------------------- 1 unit

905-120 Morcellator Cutting tube Ø 20 mm ---------------------------------------------------------------------- 1 unit 905-220 Morcellator Protective sleeve Ø 20 mm ---------------------------------------------------------------- 1 unit 905-020 Obturator Ø 20 mm ----------------------------------------------------------------------------------------- 1 unit Accessories and spare parts 905-312 Morcellator Trocar sleeve d12 mm ---------------------------------------------------------------------- 1 unit 905-315 Morcellator Trocar sleeve d15 mm ---------------------------------------------------------------------- 1 unit 905-008 Morcellator Carrier plate ERGO 300 -------------------------------------------------------------------- 1 unit 905-010 Morcellator Myoma drill d10 mm ------------------------------------------------------------------------- 1 unit 905-011 Morcellator Gasket holder ---------------------------------------------------------------------------- per 1 unit 905-009 Morcellator Cross seal ------------------------------------------------------------------------------------ 10 unit 905-512 Morcellator Seal d12-20 mm 12-20 mm--------------------------------------------------------------- 10 unit 905-013 Morcellator Motor cable 3 m ------------------------------------------------------------------------------ 1 unit

In line with regulations pertaining to hazardous materials, the following items are not deliv-ered with the control unit but can be ordered separately from any official BOWA service cen-tre. www.bowa.de

905-014 Morcellator Cleaning spray "Nou Clean" -------------------------------------------------------------------- 1 unit

Page 6: OPERATING MANUAL MORCELLATOR - bowa-medical.com · MN009-938-S0 ERGO 300 3 1 General information 1.1 Indications Morcellator set for morcellation and extraction of tissue and for

6 ERGO 300 MN009-938-S0

1.7 Technical Data «ERGO 300 - Morcellator»

Voltage -------------------------------------------------------------- variable: 100 V∼/ 115 V∼/ 230 V∼, 50–60 Hz Fuse, power supply ------------------------------------------------------------------------- 2 fuses T 1 AL 250 V AC Power consumption ------------------------------------------------------------------------------------------------- 60 VA Maximum motor speed -------------------------------------------------------------------------------------- 40,000 rpm Maximum torque, motor ------------------------------------------------------------------------------------------- 6 Ncm Torque at gear output -------------------------------------------------------------------------------------------- 80 Ncm Motor coupling ------------------------------------------------------------------------------ INTRA coupling ISO3964 Motor cable length ------------------------------------------------------------------------------------------------------ 3 m Pedal cable length ------------------------------------------------------------------------------------------------------ 3 m Morcellator speed ----------------------------------------------------------------------------------------- 50 – 1000 rpm Protection class ------------------------------------------------------------------------------------------------------class II Applied part type ------------------------------------------------------------------------------------------------- Typ BF * Foot switch -------------------------------------------------------------------------------------------------------------- IPX8 Dimensions (W x D x H) ------------------------------------------------------------------------ 120 x 180 x 107 mm Net weight control unit ---------------------------------------------------------------------------------------------- 1,8 kg

* Applied part is the Morcellator

1.8 Ambient conditions

Transport and storage: Operation:

Relative humidity: Max. 90 % Max. 80 %

Temperature: 10°C to 60°C 10°C to 30°C

Atmospheric pressure: 700 hPa to 1060 hPa 800 hPa to 1060 hPa

1.9 Warranty coverage

Purchasing the ERGO 300 - Morcellator entitles you to a 1-year warranty. If you return the warranty card for registration within four weeks of the date of purchase, warranty coverage will be extended for a further 6 months. Improper use or repair, or failure to observe these instructions, will relieve us from any obligation arising from warranty provisions or other claims.

1.10 Right to make technical changes reserved

Illustrations, dimensions and weights and technical data may vary slightly as a result of continuing development. We reserve the right to make changes and we are not responsible for printing errors.

Page 7: OPERATING MANUAL MORCELLATOR - bowa-medical.com · MN009-938-S0 ERGO 300 3 1 General information 1.1 Indications Morcellator set for morcellation and extraction of tissue and for

MN009-938-S0 ERGO 300 7

2 Explanation of symbols

3 Safety information

Your safety, the safety of your team and of course that of your patients is very important to us. It is therefore essential to bear the following information in mind:

3.1 EMC Manufacturer’s Declaration of Conformity

Please observe the information on electromagnetic compatibility provided with these operating instructions.

3.2 Modification and misuse

Important information Autoclavable at 135°C

Warning

Suitable for thermal disinfection

Manufacturer Protection against continuous submersion

in

water.

Applied part of type BF is the Morcella-

tor

Protection class II

Certified by the Canadian Standards

Association (CSA)

Electrical and electronic devices that have

reached the end of their service life com-

prise hazardous waste and may not be

disposed of together with household waste.

Local disposal regulations apply.

CE mark with Notified Body

Symbol indicating the order number

Intermittent operation, 1 min on, 1 min

off

for 8 cycles

Symbol indicating the serial number with

the

date of manufacture (year/month)

Observe the instructions for use

Motor

Pedal Biohazard

• Modification or manipulation of the ERGO 300 - Morcellator and its accessories is prohibited.

The manufacturer is not liable for any damages resulting from unauthorized modifications or manipulations. The warranty will be cancelled.

• Use of the ERGO 300 - Morcellator outside the indications described in Section 1.1 is prohibited.

The user or operator is solely responsible for any such use.

Page 8: OPERATING MANUAL MORCELLATOR - bowa-medical.com · MN009-938-S0 ERGO 300 3 1 General information 1.1 Indications Morcellator set for morcellation and extraction of tissue and for

8 ERGO 300 MN009-938-S0

3.3 Essential requirements

Before use read this manual and the

manuals for all other equipment used

in the procedure thoroughly.

The products described in this manual must be

used by trained and qualified medical personnel

only in accordance with the requirements and

with reference to the manual.

For protection of the patient, the user

and third parties the products used

must be prepared as specified in the

manual.

Improper use or repair of the device, or failure to

observe these instructions, will relieve us from

any obligation arising from warranty provisions or

other claims.

The use of third-party products is the

responsibility of the operator.

Functionality and patient safety cannot

be guaranteed with third-party acces-

sories.

Prior to using the device, before startup, and

before operation, the user must always ensure

that the device and accessories are in good

working order and are clean, sterile and opera-

tional.

Repairs may only be performed by

authorized BOWA service technicians.

Use Nou-Clean spray to care for the motor and

gear unit. Using other care products can result in

malfunctions and/or cause the warranty to be

revoked.

3.4 During use

The device is not sterile on delivery. All

sterilizable parts must be sterilized

before use (see Chapter 7 Instrument

preparation).

Never operate the clamping mechanisms of the

cutting tubes while the system is running. This

could result in instrument damage.

Operate the device only outside the

danger zone of explosive, flammable

mixtures or gases.

Keep hands clear of the cutting tubes when op-

erating the device. Danger of serious injury.

The motor ventilation slots must be

kept clear in order to prevent the motor

from overheating.

The ERGO 300 - Morcellator must be operated

under the continuous supervision of medically

qualified persons only.

Disregarding the guidelines on intermit-

tent operation (1 min on, 1 min off for 8

cycles) may cause burns at contact

with the gear unit.

Make sure that the operating voltage to the

mains voltage.

Page 9: OPERATING MANUAL MORCELLATOR - bowa-medical.com · MN009-938-S0 ERGO 300 3 1 General information 1.1 Indications Morcellator set for morcellation and extraction of tissue and for

MN009-938-S0 ERGO 300 9

4 Device overview Front view Control panel Rear view

Rear view Handle Motor

1. Display 2. Motor LED 3. Pedal connection 4. Motor connection

1. Mode button (variable speed range) 2. Low button (low speed range) 3. Medium button (medium speed range) 4. High button (high speed range) 5. Change rotational direction button 6. Up button (increase value) 7. Down button (reduce value)

1. Flange for protective tube connect 2. Thread for seal holder 3. Cutting tube holder 4. Seal holder 5. Thread for membrane screw 6. Membrane screw 7. Seal holder thread 8. Motor connection coupling

1. Release lever 2. Slots for motor ventilation

1. Mains connection 2. Power switch 3. Fuse module 4. Window with national voltage display 5. Type plate with type designation, reference

number, serial number, information on power supply and device fuse

1. Device connection plug 2. Motor cable 3. Motor cap 4. Motor 5. Handpiece holder 6. Motor connection plug

Page 10: OPERATING MANUAL MORCELLATOR - bowa-medical.com · MN009-938-S0 ERGO 300 3 1 General information 1.1 Indications Morcellator set for morcellation and extraction of tissue and for

10 ERGO 300 MN009-938-S0

Protective tube with filling rod

Cutting tube

Trocar Cannula

1. Protective tube 2. Lock nut 3. Tissue positioner 4. Filling rod

1. Protective tube 2. Seal cap holder 3. Valve actuator 6. Unterer Schenkel 7. Feder 8. Flexibler Arm der Raste

4. Valve body 5. Main body 6. Nose

1. Cutting edge 2. Locking pins 5. Fingerstütze 6. Unterer Schenkel 7. Feder 8. Flexibler Arm der Raste

6

Page 11: OPERATING MANUAL MORCELLATOR - bowa-medical.com · MN009-938-S0 ERGO 300 3 1 General information 1.1 Indications Morcellator set for morcellation and extraction of tissue and for

MN009-938-S0 ERGO 300 11

4.1 ERGO 300 Morcellator system chart

Page 12: OPERATING MANUAL MORCELLATOR - bowa-medical.com · MN009-938-S0 ERGO 300 3 1 General information 1.1 Indications Morcellator set for morcellation and extraction of tissue and for

12 ERGO 300 MN009-938-S0

5 Before use

The ERGO 300 Morcellator,and all required accessories and instruments must be placed on a level nonslip surface. Do not allow the operating range of the device with cable to be restricted by limiting factors. The display, keypad and indicator lights must be fully visible at all times. It is very important to ensure that no objects can fall on the pedal. The power plug at the rear of the device must be accessible at all times.

5.1 Connection to the power supply

• Before plugging the power cable into the power socket for the first time, you must check the sup-ply voltage setting next to the power switch.

• In order to prevent the risk of electric shock, the device may only be connected to a power network with a

PE protective earth conductor. If the voltage shown does not correspond to the local mains voltage, the grey fuse holder must be set to the correct voltage: 1. Swich the device off.

2. Unplug the power cable.

3. Use a screwdriver to open the fuse slot.

4. Remove the fuse holder.

5. Remove the grey fuse holder and reinsert it so that the local mains voltage setting is shown in the small window.

6. Slide the grey fuse holder back in and close the fuse slot.

7. Check the mains voltage shown on the fuse slot.

8. Plug the power cable back into the device.

5.2 Device preparation

The device is not sterile on delivery. All sterilizable parts must be sterilized before use (see Chapter 7 Instrument preparation).

If the accessories have already been sterilized: when removing them from the sterile packaging, ensure that the sterile packaging is not damaged and that the sterility indicator confirms sterility and shelf life of the sterile item has not expired.

The Morcellator (as per option 1 to option 3) must be assembled under sterile conditions (wear gloves and a mask, place Morcellator and accessories on a sterile surface).

Page 13: OPERATING MANUAL MORCELLATOR - bowa-medical.com · MN009-938-S0 ERGO 300 3 1 General information 1.1 Indications Morcellator set for morcellation and extraction of tissue and for

MN009-938-S0 ERGO 300 13

5.2.1 Assembly of the Morcellator set with protective tube (option 1)

There is a risk of injury to tissue by the sharp edge at the distal end of the cutting tube. • Handle the cutting tube carefully. • If the cutting tube is not used, rotate the protective tube to the NO CUT position. • If the cutting tube is not used, it must be placed on a flat, sterile surface so it cannot fall and cause injury

• Check the cutting tube. • Make sure that the cutting edge of the blade is sharp and undamaged (e.g. no cracks, defor-

mation or burrs). • Use only cutting tubes in perfect condition.

1. Inserting the cutting tube through the gear unit

Remove the seal unit from the gear unit.

Insert the cutting tube through the gear unit. The two guide pins on the cutting tube must fit into the slot on the flange of the gear unit.

2. Fixing the cutting tube to the gear unit

Hold the cutting tube with one hand.

Connect the sealing unit to the gear unit with the

bayone closure with the other hand.

3. Mounting protective tube

Select a suitable protective tube.

Push the protective tube over the distal end of the cutting tube. The guide pin of the protective tube must be in the distal slot of the gear unit.

Lock the protective tube in position with a click.

4. Protecting the cutting tube

The distal end of the cutting tube is very sharp. To prevent injury, the cutting tube must be covered at all times when it is not in use. To cover the cutting tube, rotate the protective tube until the words “NO CUT” are visible in the distal slot of the gear unit. The cutting tube is completely covered by the protective tube.

Page 14: OPERATING MANUAL MORCELLATOR - bowa-medical.com · MN009-938-S0 ERGO 300 3 1 General information 1.1 Indications Morcellator set for morcellation and extraction of tissue and for

14 ERGO 300 MN009-938-S0

5. Replacing seals

Remove the seal unit from the gear unit by rotating it slightly. Unscrew the membrane screw from the seal unit.

Check seals before use. Do not use dam-aged seals (frayed, discolored, yellowed, porous, non-elastic).

Position the edge of the cross seal (1) to point to the inside stop (2) of the seal holder (3) and press the cross seal in.

Position the edge of the membrane seal (4) to point to the slot (5) of the membrane screw (6) and press the membrane seal into the slot.

Screw the membrane screw (6) into the thread of the seal holder (3) to the stop and connect the seal unit to the gear unit.

6. Inserting filling rod

Insert the filling rod through the seal unit, gear unit and protective tube to the stop.

7. Attaching handle

Lift and hold the release lever (1) of the handle (2).

Push the handle (2) over the motor connection of the gear unit.

Release the handle release lever. The handle release lever must lock in position.

Do not open the handle release lever to remove the handle during the procedure..

8. Connecting motor

Insert the motor coupling into the bottom end of the handle.

Continue to insert the motor coupling until it locks into the gear unit motor connection.

Page 15: OPERATING MANUAL MORCELLATOR - bowa-medical.com · MN009-938-S0 ERGO 300 3 1 General information 1.1 Indications Morcellator set for morcellation and extraction of tissue and for

MN009-938-S0 ERGO 300 15

Align the red dot on the device connection plug of the motor with the red dot on the motor connection of the control unit.

Plug the motor device connection plug into the motor connection of the control unit. The plug must lock in position.

9. Connecting pedal

Align the red dot on the device connection plug of the pedal with the red dot on the pedal connection of the control unit.

Plug the pedal device connection plug into the pedal connection of the control unit.

The plug must lock in position.

10. Functional check

• Check Morcellator for loose parts

• Check that Morcellator is correctly assembled

• Are all labels easily readable?

• Check cutting tube: cutting tube evenly ground, no broken edges or bends?

• Switch on control unit

• Switch on motor (press foot switch).

• Run motor with Morcellator for about 20 seconds; it must reach the set speed.

• If it does not run evenly, switch off motor with Morcellator..

• Switch off control unit.

The control unit will only operate if the foot switch has been connected to the control unit.

• Check that Morcellator is correctly assembled.

• Check that the gear unit runs smoothly; if necessary treat it with Nou-Clean care spray.

• Check that the motor is clean and runs smoothly; if necessary clean it and treat with Nou-Clean care spray.

• Repeat the function check.

Risk of injury by motor failure Problem: the motor becomes very hot. Cause: the motor was damaged during preparation. Remedy: do not continue to use the motor. Always have a spare motor available.

Page 16: OPERATING MANUAL MORCELLATOR - bowa-medical.com · MN009-938-S0 ERGO 300 3 1 General information 1.1 Indications Morcellator set for morcellation and extraction of tissue and for

16 ERGO 300 MN009-938-S0

5.2.2 Assembly of the Morcellator set with Trocar Cannula (option 2)

Using the Trocar Cannula The Trocar Cannula is used instead of the protective tube. The integrated valve flap prevents the escape of gas when instruments are moved in and out. The nose of the Trocar Cannula partially extends beyond the cutter tube to protect organs in the abdominal cavity from the sharp blade. This requires the Morcellator to be rotated under visual supervision (endoscopy) so the nose of the Trocar Cannula covers the cutter tube blade to protect the organ.

1. Visual check

• Check the product for sealing, corrosion and sterility.

• Ensure that it is free from rust, dents and scratches.

• The distal end of the filling rod must not be damaged.

• The sealing flap and the seal must not be damaged.

Risk of injury Select blade. Make sure that the cutting edge of the blade is sharp and undamaged (e.g. no cracks or burrs).

2. Insert the cutting tube through the gear unit

Remove the seal unit from the gear unit.

Insert the cutting tube through the gear unit. The two guide pins on the cutting tube must fit into the slot on the flange of the gear unit.

3. Fixing the cutting tube to the gear unit

Hold the cutting tube with one hand.

Connect the seal unit to the gear unit with the bayonet closure with the other hand.

4. Mounting Trocar Cannula

Select a suitable Trocar Cannula. Open the Trocar Cannula valve by actuating the lever to prevent damage to the cutting tube by pushing the Trocar too far.

Push the Trocar Cannula over the distal end of the cutting tube. The guide pin of the Trocar Cannula must be in the distal slot of the gear unit.

Lock the Trocar Cannula in position with a click.

Page 17: OPERATING MANUAL MORCELLATOR - bowa-medical.com · MN009-938-S0 ERGO 300 3 1 General information 1.1 Indications Morcellator set for morcellation and extraction of tissue and for

MN009-938-S0 ERGO 300 17

5. Replacing seals

Remove the seal unit from the gear unit by rotat-ing it slightly.

Unscrew the membrane screw from the seal unit.

Position the edge of the cross seal (1) to point to the inside contact (2) of the seal holder (3) and press the cross seal in. Position the edge of the membrane seal (4) to point to the slot (5) of the membrane screw (6) and press the membrane seal into the slot.

Screw the membrane screw (6) into the thread of the seal holder (3) to the stop and connect the seal unit to the gear unit.

6. Inserting filling rod

Insert the filling rod through the seal unit, gear unit and Trocar Cannula.

7. Attaching handle

Lift and hold the release lever (1) of the handle (2). Push the handle (2) over the motor connection of the gear unit.

Release the handle release lever. The handle release lever must lock in position.

8. Connecting motor

Insert the motor coupling into the bottom end of the handle.

Continue to insert the motor coupling until it locks in-to the gear unit motor connection.

Align the red dot on the device connection plug of the motor with the red dot on the motor connection of the control unit.

Plug the motor device connection plug into the motor connection of the control unit. The plug must lock in position.

9. Connecting pedal

Align the red dot on the device connection plug of the pedal with the red dot on the pedal connection of the control unit.

Plug the pedal device connection plug into the pedal connection of the control unit. The plug must lock in position.

Page 18: OPERATING MANUAL MORCELLATOR - bowa-medical.com · MN009-938-S0 ERGO 300 3 1 General information 1.1 Indications Morcellator set for morcellation and extraction of tissue and for

18 ERGO 300 MN009-938-S0

Functional check

• Check Morcellator for loose parts

• Check that Morcellator is correctly assembled

• Are all labels easily readable?

• Check cutting tube: cutting tube evenly ground, no broken edges or bends?

• Switch on control unit

• Switch on motor (press foot switch).

• Run motor with Morcellator for about 20 seconds; it must reach the set speed.

• If it does not run evenly, switch off motor with Morcellator.

• Switch off control unit.

The control unit will only operate if the foot switch has been connected to the control unit.

• Check that Morcellator is correctly assembled.

• Check that the gear unit runs smoothly; if necessary treat it with Nou-Clean care spray.

• Check that the motor is clean and runs smoothly; if necessary clean it and treat with Nou-Clean care spray.

• Repeat the function check.

5.3 Mains connection

Risk of electric shock Connect the device to an earthed AC mains system that conforms to IEC guidelines only. Connect the device to a power supply that corresponds to the connection values on the type plate on the back of the device.

Risk of damage Do not switch on the control unit and the other equipment until all cables have been connected. Otherwise the equipment may be damaged.

The device is not water-proof. Keep the device away from liquids. Keep the mains connection socket dry. Do not use the device if liquid penetrates it.

Connecting the power cable

Switch off the power switch on the back of the device (“O” position).

Plug the included power cable into the socket on the back of the device.

Plug the other end of the power cable into the power socket.

Problem: the motor becomes very hot. Cause: the motor was damaged during preparation. Remedy: do not continue to use the motor. Always have a spare motor available.

Page 19: OPERATING MANUAL MORCELLATOR - bowa-medical.com · MN009-938-S0 ERGO 300 3 1 General information 1.1 Indications Morcellator set for morcellation and extraction of tissue and for

MN009-938-S0 ERGO 300 19

6 Use

Before use

Make sure that the product has been correctly prepared and inspected.

Risk of injury The ERGO 300 - Morcellator control unit has been designed for operation with the morcellation blades as a set. The user is solely responsible if different products are used with the set. Incom-patibility may risk injury to patients.

Switching on and off

Switch on the control unit. The display lights up. A short signal tone sounds. The control unit dis-plays the value “200 rpm”.

Operating mode (motor)

The motor is designed for intermittent operation (1 minute on, 1 minute off for 8 cycles).

6.1 Using the control unit

Setting the speed

The speed can be set manually in stages (within a

range of 50 to 100 rpm).

Press « » to increase the speed.

Press « » to reduce the speed.

The set speed is shown on the display.

Starting motor

Press foot switch.

The blade rotates at the set speed.

The motor LED lights up.

To stop the motor remove foot from foot switch.

Risk of injury and damage due to heat! The Morcellator may generate heat if operated for a long period. Remedy: Use Morcellator only in intermittent mode of “1 minute on, 1 minute off for 8 cycles”. Stop the Morcellator immediately if one or more components of the Morcellator set become very warm.

Precaution when changing parameters. Unusual behaviour of instruments while operating may provoke false reactions and jeopardize the patient. Every setting and the new behaviour of the instrument has to be verified and tested.

Page 20: OPERATING MANUAL MORCELLATOR - bowa-medical.com · MN009-938-S0 ERGO 300 3 1 General information 1.1 Indications Morcellator set for morcellation and extraction of tissue and for

20 ERGO 300 MN009-938-S0

Available rotary modes

« » (100 – 400 rpm)

« » (300 – 700 rpm)

« » (500 – 1000 rpm)

Activating mode setting

Press « » to activate mode setting.

The Mode LED lights up.

The Low LED lights up. (Default setting)

The display shows “VARIO LOW, 100 – 400 rpm”.

Selecting mode

Press « », « » or « » to select the corresponding mode.

The display shows the mode and the corresponding speed range that can be selected with the pedal.

The corresponding LED lights up.

Starting motor

Press foot switch. The motor LED lights. The blade rotates at a speed within the range set by how far down the foot switch is pressed.

To stop the motor remove foot from pedal.

Exit mode setting

Press « »

Changing direction of rotation

To change the direction of rotation, press« »

The display shows the mode and the corresponding speed range that can be selected with the pedal. The corresponding LED lights up.

Page 21: OPERATING MANUAL MORCELLATOR - bowa-medical.com · MN009-938-S0 ERGO 300 3 1 General information 1.1 Indications Morcellator set for morcellation and extraction of tissue and for

MN009-938-S0 ERGO 300 21

6.2 Operation

Risk of death or serious injury!

Unsupervised cutting may injure the abdominal wall and organs and cause the death of the pa-tient. Remedy: • Before every procedure make sure that an endoscopic procedure is preferable to a conven-

tional procedure. • Use only the products described in this manual. • During the procedure use the protective tube or Trocar Cannula as a precautionary measure.

Follow “Option 1: Morcellation with protective tube” and “Option 2: Morcellation with Trocar Cannula” in Chapter 5.

• Use Morcellator under visual control (endoscope) only. • Use Morcellator only for morcellation of tissue that has been fully prepared and is clearly visi-

ble. BOWA accepts no liability for any other applications. • BOWA also recommends the use of a second gripper or a similar retaining device to prevent

uncontrolled movement of large Morcellator pieces of tissue. This will require an additional percutaneous access.

• Fix the Morcellator in a ventrolateral position to prevent movements inside the uterus. • Prevent movement of the lateral blood vessels, kidneys and retroperitoneum. • Pull the tissue towards the rotating blade for morcellation with the gripper.

Risk of injury due to gas loss!

Gas may be lost during the procedure. This may result in the abdominal wall dropping.

• Select an Insufflator with adequate flow to prevent the abdominal wall from dropping. • Use the Morcellator with protective tube as in option 1: when removing or replacing instruments,

immediately close the proximal opening of the seal unit with the thumbs. • Use with Trocar Cannula as in option 2: pull the Morcellator with cutting tube out of the Trocar

Cannula. The Trocar Cannula valve closes to prevent gas loss..

• After removal of the morcellated tissue, insert the Morcellator into the Trocar Cannula and repeat the process.

Risk of injury! The Morcellator system is designed for removing tissue with the greatest possible safety. For this purpose the protective tube and the Trocar Cannula are both fitted with a valve flap.

Morcellation without protective tube or Trocar Cannula is prohibited.

Tissue escapes the gripper. Tissue may escape the gripper during the procedure and fall into the abdominal cavity. Causes: • The tissue is not gripped tightly enough. Remedy: • Grip the tissue at a different position • Press the jaws of the gripper together as tightly as possible.

Page 22: OPERATING MANUAL MORCELLATOR - bowa-medical.com · MN009-938-S0 ERGO 300 3 1 General information 1.1 Indications Morcellator set for morcellation and extraction of tissue and for

22 ERGO 300 MN009-938-S0

Risk of injury due to heat generated by friction! The user must ensure that as little friction heat as possible is generated during a procedure on a patient.

High speeds and high application pressure may cause thermal necrosis of tissue.

Risk of injury!

The distal end of the cutting tube is extremely sharp.

Precautions: • Handle the cutting tube particularly cautiously and carefully. • Where possible, have the Morcellator with protective tube in protected position (NO CUT), par-

ticularly if the cutting tube is not used for any period. • Insert and remove the protective tube and cutting tube under visual control (endoscopy).

Risk of injury! When inserting the filling rod through the percutaneous access, there is risk of injury to blood vessels, intestinal loops or the bladder. This can result in significant complications, which will require a Laparotomy.

Risk of damage to the blade of the cutting tube! If the jaws of the gripper are not completely closed when it is pulled from the protective tube, the blade will be damaged (see pictures).

Risk of injury! Damaged blades and cutting tubes must not be used.

Cutting tubes must not be sharpened.

Page 23: OPERATING MANUAL MORCELLATOR - bowa-medical.com · MN009-938-S0 ERGO 300 3 1 General information 1.1 Indications Morcellator set for morcellation and extraction of tissue and for

MN009-938-S0 ERGO 300 23

6.2.1 Option 1: morcellation with protective tube

1. Hysterektomie

Select the procedure for the hysterectomy in accordance with current medical standards.

2. Inserting cutting tube

Make sure that the filling rod is inside the cutting tube.

Make sure that the protective tube covers the blade on the cutting tube (set position to NO CUT; “PROTECTED” is also shown on the black ring between the gear unit and protective tube).

Insert Morcellator (with filling rod) into an existing percutaneous access under visual control (endos-copy).

3. Using cutting tube

To use the cutting tube, rotate the protective tube to uncover the blade of the cutting tube (unprotected CUT position is displayed).

To cover the blade again, rotate the protective tube (protected NO CUT position is displayed; “PROTECTED” is also shown on the black ring between the gear unit and protective tube).

4. Inserting gripper

Pull the filling rod out of the protective tube.

Close the proximal opening of the gear unit with the thumbs immediately.

Insert gripper carefully through the proximal opening of the gear unit.

5. Removing tissue

Grip the tissue with the gripper under good visual control (endoscopy).

The handle lock is set

Start the motor of the Morcellator set.

Pull the tissue towards the rotating blade for morcellation with the gripper. As the cut progresses pull the gripper with the tissue further out of the Morcel-lator.

Carefully pull the complete gripper with the separated tissue out of the Morcel-lator and the patient. The retracted mouth must not touch the blade of the cutting tube.

Repeat the process until the tissue has been completely removed.

Make sure that the proximal opening of the seal unit is closed immediately with the thumb when there is no instrument in the Morcellator to prevent large quantities of gas from escaping.

Endoscope

Page 24: OPERATING MANUAL MORCELLATOR - bowa-medical.com · MN009-938-S0 ERGO 300 3 1 General information 1.1 Indications Morcellator set for morcellation and extraction of tissue and for

24 ERGO 300 MN009-938-S0

6.2.2 Option 2: morcellation with Trocar Cannula

1. Hysterectomy

Select the procedure for the hysterectomy in accordance with current medical standards.

2. Inserting Trocar Cannula

Insert Trocar Cannula with filling rod into an existing access under visual control (endos-copy).

Remove Trocar mandrel.

3. Insert the cutting tube through the Trocar Cannula

Actuate the valve flap lever of the Trocar Cannula and hold in this position..

Carefully insert prepared cutting tube with gear unit and seal unit into the proximal opening of the Trocar Cannula.

Cutting tube must be fully inserted.

Release the valve flap lever of the Trocar Cannula.

The guide pin of the Trocar Cannula must be in the distal slot of the gear unit.

Lock the Trocar Cannula in position with a click.

Make sure that the cutting tube remains in the NO CUT position.

4. Inserting gripper

Insert gripper carefully through the proximal opening of the gear unit and Trocar Cannula.

5. Preparing cutting tube

Rotate the Trocar Cannula so the CUT posi-tion is visible from above.

If the cutting tube is completely inserted, the cutting edge is now only partially covered by the Trocar Cannula.

CUT

Page 25: OPERATING MANUAL MORCELLATOR - bowa-medical.com · MN009-938-S0 ERGO 300 3 1 General information 1.1 Indications Morcellator set for morcellation and extraction of tissue and for

MN009-938-S0 ERGO 300 25

6. Protecting tissue

The nose on the Trocar Cannula, which par-tially covers the sharp edge of the blade, must protect tissue that should not be injured.

Rotate the Morcellator to the desired position for this.

7. Removing tissue

Grip the required tissue with the gripper.

The handle lock is set

Start the Morecellator set motor

Pull the tissue towards the rotating blade with the gripper.

Pull the gripper out of the patient through the gear unit. The retracted mouth of the gripper must not touch the blade.

Repeat the process until the Morcellator tissue has been completely removed.

Protected area

Unprotected cutting area

Page 26: OPERATING MANUAL MORCELLATOR - bowa-medical.com · MN009-938-S0 ERGO 300 3 1 General information 1.1 Indications Morcellator set for morcellation and extraction of tissue and for

26 ERGO 300 MN009-938-S0

6.2.3 Option 3: Myomectomy

The myoma drill has been designed exclusively for removing tissue from a myoma.

1. Inserting Morcellator

Insert Morcellator with protective tube through the percutaneous access.

Keep gripper ready to remove the tissue from the myoma drill.

2. Inserting myoma drill

Insert Trocar Cannula through an additional percutaneous access.

Open valve flap of Trocar Cannula and insert myoma drill.

3. Upcoming myoma

Remove tissue from the myoma with the myoma drill

4. Inserting gripper

Insert gripper through the second access and remove separated tissue from the myoma drill.

Remove myoma drill.

5. Removing separated tissue

If the piece of tissue is larger than the diameter of the cutting tube, start the Morcellator set motor.

Pull the tissue towards the rotating blade with the gripper.

(Smaller pieces of tissue can be removed without morcellation)

Carefully pull the complete gripper out of the Morcellator and the patient. The retracted mouth must not touch the blade of the cutting tube.

6. Removing additional tissue

If necessary, re-insert the myoma drill and repeat the above process.

Page 27: OPERATING MANUAL MORCELLATOR - bowa-medical.com · MN009-938-S0 ERGO 300 3 1 General information 1.1 Indications Morcellator set for morcellation and extraction of tissue and for

MN009-938-S0 ERGO 300 27

6.3 After use

1. Switching off

Switch off the power switch on the back of the device (“O” position).

2. Removing protective tube (when using Morcellator with protective tube)

Rotate protective tube to NO CUT position.

Close the proximal opening of the gear unit with the thumb to prevent gas loss.

Hold the protective tube and carefully disconnect gear unit with cutting tube and remove from pa-tient.

The protective tube can also be removed from the percutaneous access.

If applicable, remove any other instruments.

3. Removing the cutting tube (when using the Morcellator with Trocar Cannula)

Close the proximal opening of the gear unit with the thumb to prevent gas loss.

Hold the Trocar Cannula and carefully disconnect gear unit with cutting tube from the Trocar Cannu-la and remove from patient. The valve flap closes automatically as soon as the cutting tube is pulled out.

The Trocar Cannula can remain as access throughout the remaining procedure.

4. Disconnecting gear unit connections

Pull the motor coupling from the gear unit mo-tor connection.

Pull the motor device connection plug from the motor connection of the control unit.

Pull the motor out of the handle.

Lift and hold the release lever of the handle.

Pull the handle from the gear unit motor connection.

5. Disconnecting pedal connections

Pull the pedal device connection plug from the pedal connection of the control unit.

6. Preparing for preparation at the location of use

Take the product to the preparation area imme-

diately.

Prepare the product as specified in this manual.

Page 28: OPERATING MANUAL MORCELLATOR - bowa-medical.com · MN009-938-S0 ERGO 300 3 1 General information 1.1 Indications Morcellator set for morcellation and extraction of tissue and for

28 ERGO 300 MN009-938-S0

6.4 Disassembling Trocar Cannula

Disassembling Trocar Cannula (four steps)

Disassemble Trocar Cannula before decontamination.

Disassemble Trocar Cannula and rinse in deionized water during disassembly.

1. Unscrew sealing cap holder from the

proximal end of the Trocar Cannula. 2. Remove the internal seal from the seal-

ing cap holder. 3. Unscrew the valve body from the main

body. 4. Remove the red O-ring from the valve

body.

Page 29: OPERATING MANUAL MORCELLATOR - bowa-medical.com · MN009-938-S0 ERGO 300 3 1 General information 1.1 Indications Morcellator set for morcellation and extraction of tissue and for

MN009-938-S0 ERGO 300 29

6.5 Disassembly of motor

Disassembling motor (three steps)

Disassemble motor before decontamination.

1. Unscrew handpiece holder from motor 2. Unscrew motor cap from motor 3. Pull motor plug from motor.

6.6 Disassembly of gear unit and Morcellator with protective tube

Disassembling seal unit

Remove membrane screw from seal unit and remove membrane seal.

Remove seal retainer from gear unit and remove cross seal.

Risk of injury for the user!

The distal end of the cutting tube is very sharp, handle the cutting tube very carefully..

Removing protective tube and cutting tube

Make sure that the lock nut of the protective tube covers the distal end of the cutting tube (protected NO CUT position is visible).

Hold the gear unit with one hand so the distal end of the cutting tube does not point to the user or any other person.

With the other hand carefully disconnect the protective tube from the gear unit and pull it off.

Hold the cutting tube with the same hand and press it lightly against the gear unit until it re-leases itself from the holder and slides out smoothly. It can now be pulled out of the proximal opening of the gear unit.

1.Remove protective tube Schutzrohr abziehen

2. Pull out cutting tube Schneidrohr herausziehen

Page 30: OPERATING MANUAL MORCELLATOR - bowa-medical.com · MN009-938-S0 ERGO 300 3 1 General information 1.1 Indications Morcellator set for morcellation and extraction of tissue and for

30 ERGO 300 MN009-938-S0

7 Instrument preparation

• Do not clean Morcellator with compressed air. • Use only cleaning and care spray recommended by BOWA – other care sprays may destroy

the Morcellator. • At unpacked steam sterilization of single items it must be regarded that they have a stable

seat in the tray to prevent them from rolling around, causing damage. • Allow the Morcellator to cool down after steam sterilization!

Immersing instruments in cleaning solution

• Never exceed the maximum concentration and exposure time specified by the manufacturer of the cleaning and disinfection solution.

• Air bubbles must not adhere to the instrument. • All components of the instrument must be completely immersed in the cleaning solution. • All lumina of the instrument must be completely filled with cleaning solution and free from

bubbles.

• Do not use cleaning agents that contain solvents. • If sterilized accessories are not used immediately, a sterilization indicator must be placed on the

packaging and it must be labelled with the sterilization date.

• Perform cleaning, disinfection and sterilization after every treatment. • To autoclave single items, such as cutting tubes, BOWA recommends the use of sterilization

packages to prevent of damage. • Make sure that sterilization packaging is no more than 80 % full. • Autoclave material for at least five minutes at 135 °C. • If sterilized material is not used immediately, the sterilization packaging must be labelled with

the sterilization date. • BOWA recommends including a sterility indicator.

7.1 Control unit and pedal

The control unit and pedal do not come into contact with the patient. Wipe the outside using microbiologically tested surface disinfectant or a 70 % -isopropyl alcohol solution. The front plate of the control unit is sealed accordingly for this purpose and can be wiped clean.

Risk of damage! Do not use incompatible preparation methods. The product may be damaged.

Risk of damage to the ERGO 300 - Morcellator control unit! • Wipe away dirt or dust with a soft cloth. • Use a moist cloth for stubborn dirt.

Risk of infection! Prepare the product before its first use and before every subsequent use as specified in this manual. Inadequate and/or incomplete preparation of the product may cause an infection of the patient. Observe the following specific instructions.

Page 31: OPERATING MANUAL MORCELLATOR - bowa-medical.com · MN009-938-S0 ERGO 300 3 1 General information 1.1 Indications Morcellator set for morcellation and extraction of tissue and for

MN009-938-S0 ERGO 300 31

7.2 Preparation of the Trocar Cannula

Essential requirements

Avoid long wait times until preparation due to risk of drying and corrosion. Maximum time between use and preparation: 2 hours. The Trocar Cannula is designed for 500 sterilization cycles.

Disassembly

Before decontamination disassemble the Trocar Cannula as described in section 6.4 “Disassem-bling Trocar Cannula”. Remove seal cap, internal seal and O-ring. Manual cleaning

• Rinse and clean all surfaces and lumina thoroughly with cleaning solution and suitable brushes.

• Place Trocar Cannula in an ultrasonic bath with cleaning solution, if necessary.

• Rinse all components of the Trocar Cannula thoroughly with cleaning gun and deionized water.

Machine cleaning and disinfection

• Connect all components of the Trocar to the cleaning connections of the machine, if necessary.

• Use a fine-mesh basket for small parts, such as seals and O-rings.

• Run a standard machine cleaning and thermal disinfection cycle.

Autoclaving

Autoclave in vacuum autoclave (class B or S as per EN 13060) at 135 °C for at least five minutes*. Do not exceed the maximum load of the sterilizer when sterilizing multiple instruments in one sterili-zation cycle. Autoclaving without vacuum must be followed by a drying phase. Allow the Trocar Cannula to dry in a bag for at least one hour at room temperature with the paper side up..

* The temperature holding times conform to the national guidelines and standards with the maximum holding time 25 minutes. Assembly

• Make sure that all parts of the Trocar Cannula are completely dry after preparation.

• The flap in the valve body must be open when screwing the spacer with internal seal to the valve body.

• Assembly the Trocar Cannula in reverse order of disassembly (see chapter 6.4 “Disassembling Trocar Cannula”).

Page 32: OPERATING MANUAL MORCELLATOR - bowa-medical.com · MN009-938-S0 ERGO 300 3 1 General information 1.1 Indications Morcellator set for morcellation and extraction of tissue and for

32 ERGO 300 MN009-938-S0

7.3 Preparation of motor

Frequent preparation has little effect on the electric motor. The end of the product service life nor-mally results from wear and damage caused by use. The electric motor is designed for 250 steriliza-tion cycles.

Risk of damage • Motor must not be cleaned with ultrasonic or compressed air. • Do not kink the motor cable, otherwise it may be broken. • Use BOWA cleaning and care spray exclusively, other care sprays may destroy the motor. • Steam sterilize the micromotor in a sterilization bag and then allow to cool.

At the site of use

Remove surface dirt with a disposable cloth or paper towel.

Storage and transport

Avoid long wait times before preparation because of the risk of drying and corrosion; maximum peri-od between use and preparation is eight hours.

Preparation for cleaning

Remove dirt on the electric motor with a disposable cloth or paper towel. Unscrew motor cap and handpiece holder and remove with cable. Do not place electric motor in ultrasonic bath.

Disassembly

Before decontamination, disassemble motor as described in “Disassembling motor”. Manual cleaning

Use a neutral cleaning agent, cleaning brushes and deionized water (< 38 °C).

• Rinse and brush off dirt on the electric motor, motor cap with cable and handpiece holder.

• Apply cleaning agent to all surfaces and intermediate spaces with brush and brush thoroughly.

• Rinse electric motor, motor cap with cable and handpiece carrier thoroughly under running water.

Automatic cleaning and disinfection

Equipment: cleaning and disinfecting machine with special racks that connect motors to the cleaning and disinfecting machine and rinse out all channels. Wash the motor from the front. Use only neutral cleaning agents.

• Position electric motor in rack (ensure that channels are washed by connecting them to the cleaning and disinfecting machine.

• Place the motor cap and cable, and handpiece holder, in the basket.

• Set a cleaning cycle that offers sufficient cleaning and rinsing. Perform the final rinse with deion-ized water.

• Perform a 10-minute rinse cycle (disinfection) at 93°C to facilitate thermal disinfection.

• When removing the electric motor, check the motor cap and cable and handpiece holder to verify whether soiling is still visible in the gaps and the grooves. If necessary, repeat the cycle or clean manually.

Page 33: OPERATING MANUAL MORCELLATOR - bowa-medical.com · MN009-938-S0 ERGO 300 3 1 General information 1.1 Indications Morcellator set for morcellation and extraction of tissue and for

MN009-938-S0 ERGO 300 33

Manual disinfection

For manual disinfection, spray the electric motor, handpiece holder, motor cap with the plug and cable underneath it, and the handpiece holder with isopropyl-based disinfectant spray.

Drying

If a drying programme is not provided by the washer-disinfector, the electric motor must be dried in a drying cabinet at 60 °C.

Inspection, maintenance and assembly

• Perform a visual inspection to check for damage, corrosion and wear and tear.

• Screw the motor to spray attachment (art. no. 905-007) and spray with Nou-Clean spray for about three seconds and briefly wipe with a moist cloth (see instructions on spray can).

• After spraying, attach the handpiece holder and screw the cable with motor cap onto the electric motor..

Packaging

• Storage of single unit: Pack the electric motor in individual packaging for sterile items. The bag must be large enough to ensure that the seal is not subject to strain. When using sterilization container, observe the instructions of the manufacturer.

• Storage of set: Sort electric motors on trays intended for this purpose or place them on all-purpose sterilization trays.

Sterilization

• Autoclave in the vacuum autoclave (Class B or S as per EN 13060) at 135 °C for at least 5 minutes*.

• When sterilizing several instruments during one sterilization cycle, do not exceed the maximum sterilizer load.

• A drying cycle must be added in the case of autoclaves without a post-vacuum function. Allow the electric motor to dry in the bag for at least 1 hour at room temperature with the paper side facing upwards.

* Temperature exposure times are based on country-specific guidelines and standards. Maxi mum exposure time: 25 minutes.

Storage

If the sterilized electric motor is not used immediately after sterilization, the material packaging must be labeled with the sterilization date.

Page 34: OPERATING MANUAL MORCELLATOR - bowa-medical.com · MN009-938-S0 ERGO 300 3 1 General information 1.1 Indications Morcellator set for morcellation and extraction of tissue and for

34 ERGO 300 MN009-938-S0

7.4 Preparation of gear unit

The gear unit is designed for 250 sterilization cycles.

Oil may leak from the pressure compensation opening of the gear unit under pressure. Wipe off with a moist cloth.

Do not immerse gear unit in liquids.

Disassembly

Disassemble gear unit as described in section 6.6 “Disassembly o gear unit and Morcellator”.

Manual cleaning

Use neutral and alkaline cleaning agents, cleaning brushes and deionized water.

• Rinse and brush surfaces and intermediate spaces thoroughly with neutral and alkaline cleaning agent.

• Brush and rinse lumen thoroughly with neutral and alkaline cleaning agent.

• Rinse all surfaces and lumina with deionized water for at least 10 minutes. Make sure that all in-termediate spaces are cleaned.

Machine cleaning and disinfection

• Equipment: Cleaner and disinfector with special racks that ensure that channels are cleaned. Use only neutral cleaning agents.

• Place the gear unit in the load carrier (ensure that the channels can be rinsed).

• Place seal holder and membrane screw in basket

• Set a cleaning cycle that offers sufficient cleaning and rinsing. Perform the final rinse with deionized water.

• Perform a 10-minute rinse cycle at 93 °C to facilitate thermal disinfection.

• When removing the gear unit, the seal holder and the membrane screw, check whether visible contamination still remains in the gaps and the grooves. If necessary, repeat the cycle or clean manually.

Maintenance, lubricating gear unit

• Perform a visual inspection to check for damage, corrosion and wear and tear.

• Use Nou-Clean spray exclusively.

• Attach the spray attachment for handpiec-es (art. no. 905-006) to Nou-Clean spray and sprayngear unit for about three sec-onds and briefly wipe with a moist cloth (see instructions on spray can).

Page 35: OPERATING MANUAL MORCELLATOR - bowa-medical.com · MN009-938-S0 ERGO 300 3 1 General information 1.1 Indications Morcellator set for morcellation and extraction of tissue and for

MN009-938-S0 ERGO 300 35

Sterilizing gear unit

Autoclave in the vacuum autoclave (Class B or S as per EN 13060) at 135 °C for at least 5 minutes*. When sterilizing several instruments during one sterilization cycle, do not exceed the max-imum sterilizer load. Adrying cycle must be added in the case of autoclaves without a post-vacuum function. Allow the gear unit to dry in the bag for at least 1 hour at room temperature with the paper side on top. * Temperature exposure times are based on country-specific guidelines and standards. Maximum exposure time: 25 minutes.

Assembly

• Make sure that all parts of the gear unit are completely dry after preparation.

• Assemble gear unit in reverse order of disassembly (see Chapter 6 Use).

7.5 Preparation of cutting tubes, protective tubes, filling rods, seals, seal unit and myoma drill

Clean all products with suitable cleaning brushes. Reprocessing restrictions

The end of the product service life is normally determined by wear and tear and damage through use. The cutting tubes are designed for a service life of 20 cycles. At the location of use

Remove surface soiling with a disposable cloth/paper towel. Storage and transport

No special requirements – avoid long periods before preparation to prevent the danger of drying-out and corrosion, maximum time between use and preparation is two hours. Preparation for cleaning

Disassemble seal unit into component parts under running water (see 6.6 Disassembly of gear unit and morcellator) Disassemble gripper and remove seal caps. Wash visible soiling from all components. Manual cleaning

• Use disinfectant solution, neutral or alkaline cleaning agents, suitable cleaning brushes and de-ionized water. Wipe all surfaces thoroughly with disinfectant (exposure time as specified by the disinfectant manufacturer).

• Immerse all parts in alkaline cleaning agent (exposure time as specified by the cleaning agent manufacturer). Use suitable brushes that can reach all parts of the lumen for cavities.

• Rinse all part thoroughly under running water. Automatic cleaning and disinfection

Equipment: Cleaning/disinfection device with a special load carrier which facilitates the connection of parts with cavities to the cleaning/disinfection device and the rinsing of channels. Use a fine-mesh basket for small parts. Use neutral or alkaline cleaning agents at the recommended concen-tration.

• Place the cutting tubes, protective tubes, myoma drills, grippers and ergonomical handle in the load carrier (ensure that the channels can be rinsed).

• Place seal holder, membrane screw, membrane seal, cross seal and seal caps of gripper in the fine-mesh basket and close the basket.

• Set a cleaning cycle that offers sufficient cleaning and rinsing. Perform the final rinse with deion-ized water.

• Perform a 10-minute rinse cycle at 93°C to facilitate thermal disinfection.

• When removing the products, check whether visible contamination still remains in the gaps and the grooves. If necessary, repeat the cycle or clean manually.

Page 36: OPERATING MANUAL MORCELLATOR - bowa-medical.com · MN009-938-S0 ERGO 300 3 1 General information 1.1 Indications Morcellator set for morcellation and extraction of tissue and for

36 ERGO 300 MN009-938-S0

Drying

If a drying programme is not provided by the washer-disinfector, all parts must be dried in a drying cabinet at 60 °C. Inspection, maintenance and assembly

• Perform a visual inspection to check for damage, corrosion and wear and tear.

• Check cutting edge for damage and rotary components for bending.

• Check that rotary components move smoothly.

• Check seals and seal caps for elasticity, cracking, deformation and porosity. Packaging

• Storage of single unit: Pack parts in individual packaging for sterile items. The bag must be large enough to ensure that the seal is not subject to strain. When using sterilization container, ob-serve the instructions of the manufacturer.

• Storage of set: Sort parts on trays intended for this purpose or place them on all-purpose sterili-zation trays.

Risk of damage The blade of the cutting tube may damage the sterile packaging.

Sterilization of handle, cutting tubes, protective tubes, trocar dome, myoma drills, seal hold-er and membrane screw

Autoclave in the vacuum autoclave (Class B or S as per EN 13060) at 135 °C for at least 5minutes*. When sterilizing several instruments during one sterilization cycle, do not exceed the maximum sterilizer load. A drying cycle must be added in the case of autoclaves without a post-vacuum func-tion. Allow products to dry in the bag for at least 1 hour at room temperature with the paper side on top. * Temperature exposure times are based on country-specific guidelines and standards. Maximum exposure time: 25 minutes.

The instructions throughout chapter 7 have been validated as being suitable for preparing a medical device to be reused. The preparer is responsible for ensuring that the actually performed preparation involving the equipment, materials and personnel used in the prepa-ration facility achieves the desired results. This normally requires validation and routine monitoring of the procedure. Likewise, each deviation from the provided instructions should be carefully analyzed for its effect and potential detrimental consequences.

Page 37: OPERATING MANUAL MORCELLATOR - bowa-medical.com · MN009-938-S0 ERGO 300 3 1 General information 1.1 Indications Morcellator set for morcellation and extraction of tissue and for

MN009-938-S0 ERGO 300 37

8 Maintenance and repair

8.1 General maintenance

Maintenance, motor cable

• Replace a damaged motor cable with a new one (art. no. 905-013) Maintenance, protective tube and filling rod

• Replace a damaged protective tube or filling rod with a new one. Protective tube Ø 12 mm (art. no. 905-212) Obturator Ø 12 mm (art. no. 905-012) Protective tube Ø 15 mm (art. no. 905-215) Obturator Ø 15 mm (art. no. 905-015) Protective tube Ø 20 mm (art. no. 905-220) Obturator Ø 20 mm (art. no. 905-020)

Maintenance, seals for Trocar

• Replace damaged seals with new ones. Set gasket holder with membrane screw (art. no. 905-011), package for 1 piece

Maintenance, seals in seal unit

• Replace damaged seals with new ones Membrane seal (art. no. 905-512) for use with protective tube Ø 12-20 mm Cross seal (art. no. 905-009) suitable for all cutting tube diameters

Wartung, Adapter zu NouClean-Spray

• Replace missing adapters with new ones Adapter for Nou-Clean spray for care of electric motor (art. no. 905-007) Adapter for Nou-Clean spray for care of gear unit (art. no. 905-006)

8.2 Replacing the control unit fuse

Users can replace faulty control unit fuses themselves. These are located at the rear of the device in the fuse slot beside the power switch:

1. Switch the device off. 2. Unplug the power plug. 3. Open the fuse slut using a screw driver 4. Replace the faulty fuse T 1 AL 205VAC 5. Slide the fuse holder back in and close the fuse slot. 6. Check the mains volatage shown on the fuse slot 7. Plug in the power plug again 1. Fuse slot locking mechanism 2. Display window for voltage setting 3. Fuse slot

4. Fuse 1 5. Fuse 2

1 2 3 4 5

Page 38: OPERATING MANUAL MORCELLATOR - bowa-medical.com · MN009-938-S0 ERGO 300 3 1 General information 1.1 Indications Morcellator set for morcellation and extraction of tissue and for

38 ERGO 300 MN009-938-S0

8.3 Safety inspections

The performance of safety inspections on medical devices is required by the legislation of several countries. The safety inspection is a regular safety check that is compulsory for those operating medical devices. The objective of this measure is to ensure that device defects and risks to patients, users or third parties are identified in good time.

An inspection interval of 2 years applies for the ERGO 300 - Morcellator.

BOWA offers a safety inspection service for its customers. See the appendix of the manual under “Service Centers”. Please contact our technical service department for more information.

Additional international service centers are listed on the BOWA web site:

www.nouvag.com > Service > Servicestellen

8.4 Malfunction and troubleshooting

Malfunction Cause Solution Refer to the operating

instructions

Device is not op-

erational

(display is not on)

Control unit not switched

on

Set the power switch

“I/O” to “I”

5.2.1 Assembly of the Morcellator set

with protective tube (option 1)

5.2.2 Assembly of the Morcellator set

with Trocar Cannula (option 2)

Power connection not

established

connect 5.2.1 Assembly of the Morcellator set

with protective tube (option 1)

5.2.2 Assembly of the Morcellator set

with Trocar Cannula (option 2)

Incorrect operating voltage Check the mains voltage 5.1. Connection to the power supply

Faulty fuse Replace the fuse 8.2 Replacing the control unit fuse

Motor does not

run

Motor not switched on Switch on the motor using

the treadplate

6.1 Using the control unit

Motor not connected Connect the motor cable to

the control unit

5.2.1 Assembly of the Morcellator set

with protective tube (option 1)

5.2.2 Assembly of the Morcellator set

with Trocar Cannula (option 2)

Gear unit not correctly in-

stalled

Press the electric motor

firmly onto the gear unit

coupling until it clicks into

position and check that it is

secure by moving it slightly

in the opposite direction.

5.2.1 Assembly of the Morcellator set

with protective tube (option 1)

5.2.2 Assembly of the Morcellator set

with Trocar Cannula (option 2)

Foot switch does

not operate

(display is on)

Foot switch not connected Plug foot switch cable into

the foot switch socket.

5.2.1 Assembly of the Morcellator set

with protective tube (option 1)

5.2.2 Assembly of the Morcellator set

with Trocar Cannula (option 2)

If a fault cannot be rectified, please contact your supplier or an authorized service centre. The addresses are provided on the last page of the operating instructions.

Page 39: OPERATING MANUAL MORCELLATOR - bowa-medical.com · MN009-938-S0 ERGO 300 3 1 General information 1.1 Indications Morcellator set for morcellation and extraction of tissue and for

MN009-938-S0 ERGO 300 39

8.5 Spare parts list with order numbers

Accessories Art. no.

Morcellator Gear unit complete including seals ---------------------------------------------------------------------------- 905-004

Trocar Cannula Ø 12 mm -------------------------------------------------------------------------------------------------- 905-312

Trocar Cannula Ø 15 mm -------------------------------------------------------------------------------------------------- 905-315

Protective tube Ø 12 mm -------------------------------------------------------------------------------------------------- 905-212

Protective tube Ø 15 mm -------------------------------------------------------------------------------------------------- 905-215

Protective tube Ø 20 mm -------------------------------------------------------------------------------------------------- 905-220

Obturator Ø 12 mm --------------------------------------------------------------------------------------------------------- 905-012

Obturator Ø 15 mm --------------------------------------------------------------------------------------------------------- 905-015

Obturator Ø 20 mm --------------------------------------------------------------------------------------------------------- 905-020

Myoma drill Ø 10 mm ------------------------------------------------------------------------------------------------------- 905-010

Cutting tube Ø 12 mm ------------------------------------------------------------------------------------------------------ 905-112

Cutting tube Ø 15 mm ------------------------------------------------------------------------------------------------------ 905-115

Cutting tube Ø 20 mm ------------------------------------------------------------------------------------------------------ 905-120

Handle for gear unit --------------------------------------------------------------------------------------------------------- 905-005

Electric motor with 3 m motor cable ------------------------------------------------------------------------------------- 905-002

Motor cable 3 m, spare part for electric motor ------------------------------------------------------------------------ 905-013

Membrane seals for seal unit for Ø 12-20 mm ----------------------------------------------------------------------- 905-512

Cross seal for all cutting tube diameters Ø 12-20 mm ------------------------------------------------------------- 905-009

Gasket holder ----------------------------------------------------------------------------------------------------------------- 905-011

To order additional parts, please contact our customer service department. The user manual is not available in printed form with this product. The user manual as PDF is on the CD-ROM that was included in the delivery. A CD-ROM drive and the Adobe Acrobat software is needed to read or print this user manual. To order the user manual in printed form please check on the type plate for the address or the website. If you send us the article number MN009-938 with the language code and the note “Printed copy”, a printed manual will be sent to you free of charge.

8.6 Information on disposal

Electrical and electronic devices that have reached the end of their service life comprise hazardous waste and may not be disposed of together with household waste. Prevailing national and local dis-posal regulations apply. When disposing of the device, device parts and accessories, the requirements specified in legislation must be followed. To ensure environmental protection, old devices can be returned to the dealer or manufacturer. Motors that have reached the end of their service life must not be disposed of with household waste. Motors must be sterilized before disposal. The national and local regulations for disposal of electron-ic waste must be observed. If they are not sterilized before disposal, please observe the national and local regulations for dis-posal of infectious waste.

Page 40: OPERATING MANUAL MORCELLATOR - bowa-medical.com · MN009-938-S0 ERGO 300 3 1 General information 1.1 Indications Morcellator set for morcellation and extraction of tissue and for

40 ERGO 300 MN009-938-S0

9 Electromagnetic compatibility (EMC) Bemerkung: The Product subsequently referred to herein always denotes the ERGO 300 Morcellator. Changes or modifications to this product not expressly approved by the manufacturer may result in in-creased emissions or decreased immunity performance of the product and could cause EMC issues with this or other equipment. This product is designed and tested to comply with applicable regulations regarding EMC and shall be installed and put into service according to the EMC information stated as follows. WARNING Use of portable phones or other radio frequency (RF) emitting equipment near the product may cause unex-pected or adverse operation. WARNING The product shall not be used adjacent to, or stacked with, other equipment. If adjacent or stacked use is necessary, the product shall be tested to verify normal operation in the configuration in which it is being used. Compliant Cables and Accessories WARNING The use of accessories, transducers and cables other than those specified may result in increased emis-sions or decreased immunity performance of the product. The table below lists cables, transducers, and other applicable accessories for which the manufacturer claims EMC compliance. NOTE: Any supplied accessories that do not affect EMC compliance are not listed.

Description Length max.

Electronicmotor 31ESS 2,9 m

Vario Footpedal IP68 2,9 m

Guidance and manufacturer’s declaration – electromagnetic emissions

The Product is intended for use in the electromagnetic environment specified below. The customer or the

user of the Product should assure that it is used in such an environment.

Emissions test Compliance Electromagnetic environment -

guidance

RF emissions

CISPR 11

Group 1 The Product uses RF energy only

for its internal function. Therefore,

its RF emissions are very low and

are not likely to cause any interfer-

ence in nearby electronic equip-

ment.

RF emissions

CISPR 11

Class B The Product is suitable for use in

all establishments, including do-

mestic establishments and those

directly connected to the public

low-voltage power supply network

that supplies buildings used for

domestic purposes.

Aussendung von Oberschwingungen

nach IEC 61000-3-2

Klasse A

Voltage fluctuations/flicker emissions

IEC 61000-3-3

Complies

Page 41: OPERATING MANUAL MORCELLATOR - bowa-medical.com · MN009-938-S0 ERGO 300 3 1 General information 1.1 Indications Morcellator set for morcellation and extraction of tissue and for

MN009-938-S0 ERGO 300 41

Guidance and manufacturer’s declaration – electromagnetic immunity

The Product is intended for use in the electromagnetic environment specified below. The customer or the

user of the Product should assure that it is used in such an environment.

Immunity tests IEC 60601 Test level Compliance level Electromagnetic envi-

ronment - guidance

Electrostatic discharge

(ESD)

IEC 61000-4-2

+/- 6 kV contact +/- 6 kV contact Floors should be wood,

concrete or ceramic tile. If

floors are covered with

synthetic material, the

relative humidity should

be at least 30 %.

+/- 8 kV air +/- 8 kV air

Electrical fast

transient/burst

IEC 61000-4-4

± 2 kV for power supply ± 2 kV for power supply Mains power quality

should be that of a typical

commercial or hospital

environment.

±1 kV for input/output

lines

±1 kV for input/output

lines

Stoßspannungen (Surg-

es) nach IEC 61000-4-5

+/- 1 kV differential mode

+/- 1 kV differential mode

Mains power quality

should be that of a typical

commercial or hospital

environment.

+/- 2 kV common mode +/- 2 kV common mode

Voltage dips, short

interruptions and voltage

variations on power

supply input lines

IEC 61000-4-11

< 5 % UT

(> 95 % dip in UT)

for 0,5 cycle

40 % UT

(60 % dip in UT)

for 5 cycles

70 % UT

(30 % dip in UT)

for 25 cycles

< 5 % UT

(> 95 % dip in UT)

for 5 sec

< 5 % UT

(> 95 % dip in UT)

for 0,5 cycle

40 % UT

(60 % dip in UT)

for 5 cycles

70 % UT

(30 % dip in UT)

for 25 cycles

< 5 % UT

(> 95 % dip in UT)

for 5 sec

Mains power quality

should be that of a typical

commercial or hospital

environment.

If the user of the Product

requires continued

operation during power

mains interruptions, it

is recommended that the

Product be powered

from an uninterruptible

power supply or a

battery.

Power frequency

(50/60Hz) magnetic field

IEC 61000-4-8

3 A/m 30 A/m Power frequency magnet-

ic fields should be at lev-

els characteristic of a

typical location in a typical

commercial or hospital

environment.

Note: UT is the a.c. mains voltage prior to application of the test level.

Page 42: OPERATING MANUAL MORCELLATOR - bowa-medical.com · MN009-938-S0 ERGO 300 3 1 General information 1.1 Indications Morcellator set for morcellation and extraction of tissue and for

42 ERGO 300 MN009-938-S0

Guidance and manufacturer’s declaration – electromagnetic immunity for not life support equipment

The Product is intended for use in the electromagnetic environment specified below. The customer or the

user of the Product should assure that it is used in such an environment.

Immunity tests IEC 60601

Test level

Compliance

level

Electromagnetic environment - guid-

ance

Portable and mobile RF communications

equipment should be used no closer to

any part of the Product, including cables,

than the recommended separation dis-

tance calculated from the equation appli-

cable to the frequency of the transmitter.

Recommended separation distance:

Conducted RF

IEC 61000-4-6

3 V rms

150 kHz to 80

MHz

outside ISM

bands

3 V rms

150 kHz to 80

MHz

outside ISM

bands

d = 1,2 × √P

Radiated RF

IEC 61000-4-3

3 V/m

80 MHz to 2.5

GHz

3 V/m

80 MHz to 2.5

GHz

d = 1,2 × √P 80 MHz to 800 MHz

d = 2,3 × √P 800 MHz to 2,5 GHz

where P is the maximum output power

rating in the transmitter in watts (W) ac-

cording to the transmitter manufacturer

and d is the recommended

separation distance in meters (m).

Field strengths from fixed RF transmitters,

as determined by an electromagnetic site

surveya, should be less than the compli-

ance level in each frequency rangeb.

Interference may occur in the vicinity of

equipment marked with the following

symbol:

Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.

Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by

absorption and reflection from structures, objects and people. a Fixed strengths from fixed transmitters, such as base stations for radio (cellular/cordless)

telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV

broadcast cannot be predicted theoretically with accuracy. To access the electromagnetic

environment due to fixed RF transmitters, and electromagnetic site survey should be con-

sidered. If the measured field strength in the location in which the Product is used exceeds

the applicable RF compliance level above, the Product should be observed to verify normal

operation. If abnormal performance is observed, additional measures may be necessary,

such as reorienting or relocating the Product b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

Page 43: OPERATING MANUAL MORCELLATOR - bowa-medical.com · MN009-938-S0 ERGO 300 3 1 General information 1.1 Indications Morcellator set for morcellation and extraction of tissue and for

MN009-938-S0 ERGO 300 43

Recommended separation distances between

portable and mobile RF communications equipment and the not life support equipment

The Product is intended for use in an electromagnetic environment in which radiated RF disturbances are

controlled. The customer or the user of the Product can help prevent electromagnet interference by main-

taining a minimum distance between portable and mobile RF communications equipment (transmitters) and

the Product as recommended below, according to the maximum output power of the communications

equipment.

Rated maximum output

power of transmitter

(W)

Separation distance according to frequency of transmitter (m)

150 kHz to 80 MHz

d = 1,2 × √P

80 MHz to 800 MHz

d = 1,2 × √P

800 MHz to 2,5 GHz

d = 2,3 × √P

0,01 0,12 0,12 0,23

0,1 0,38 0,38 0,73

1 1,2 1,2 2,3

10 3,8 3,8 7,3

100 12 12 23

For transmitters rated at a maximum output power not listed above, the recommended separation distance d

in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is

the maximum output power rating of the transmitter in watts (W) according to the higher frequency range

applies.

Note 1: At 80 MHz and 800 MHz, the separation distance fort the higher frequency range

applies.

Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is

affected by absorption and reflection from structures, objects and people.

Page 44: OPERATING MANUAL MORCELLATOR - bowa-medical.com · MN009-938-S0 ERGO 300 3 1 General information 1.1 Indications Morcellator set for morcellation and extraction of tissue and for

44 ERGO 300 MN009-938-S0

Servicecenters Switzerland Nouvag AG St. Gallerstr. 23-25 CH-9403 Goldach Phone: +41 (0) 71 846 66 00 Fax: +41 (0) 71 845 35 36 [email protected] │ www.nouvag.com Germany Nouvag GmbH Schulthaißstrasse 15 D-78462 Konstanz Phone: +49 (0) 7531 1290-0 Fax: +49 (0) 7531 1290-12 [email protected] │ www.nouvag.com USA Nouvag USA Inc. 18058 Albyn Court Lake Hughes, CA 93532 USA Phone: +1 (661) 724 0217 Fax: +1 (661) 724 1590 Free of charge: (800) 673 7427 [email protected] │ www.nouvag.com

Page 45: OPERATING MANUAL MORCELLATOR - bowa-medical.com · MN009-938-S0 ERGO 300 3 1 General information 1.1 Indications Morcellator set for morcellation and extraction of tissue and for
Page 46: OPERATING MANUAL MORCELLATOR - bowa-medical.com · MN009-938-S0 ERGO 300 3 1 General information 1.1 Indications Morcellator set for morcellation and extraction of tissue and for

Nouvag AG St. Gallerstr. 23-25 CH-9403 Goldach │ Switzerland Phone: +41 (0) 71 846 66 00 Fax: +41 (0) 71 846 66 70 [email protected] │ www.nouvag.com

Distributed by:

1275

MN

009-9

38-S

0-E

N 0

8/1

4

Prin

ted in G

erm

any

Subje

ct to

technic

al and d

esig

n c

hanges

Copyrig

ht

by B

OW

A-e

lectr

onic

, G

om

arin

gen │

Germ

any

BOWA-electronic GmbH & Co. KG Heinrich-Hertz Strasse 4–10 D-72810 Gomaringen │ Germany Phone: +49 (0) 7072-6002-0 Fax: +49 (0) 7072-6002-33 [email protected] │ www.bowa-medical.com

CE marked according to Medical Device 93/42/EWG