Page 1: Operating Manual - Hello from Web Server micro operating... · DOC. No. OM-28373 REV.5 06/05 AUTION MICRO PU-4210 OPERATING MANUAL e. Introduction This manual contains outlines, instructions

Compact Urine Analyzer

Operating Manual

Page 2: Operating Manual - Hello from Web Server micro operating... · DOC. No. OM-28373 REV.5 06/05 AUTION MICRO PU-4210 OPERATING MANUAL e. Introduction This manual contains outlines, instructions

1 AUTION MICRO

This manual contains important information on thefunctions of the AUTION MICRO PU-4210.

Supplied by : ARKRAY, Inc.distributed by : A. Menarini I.F.R. srl

Diagnostic Division

This manual has been issued by: ARKRAY, Inc.Read carefully prior to starting up the instrument.It is recommended to retain this manual for futureuse.

This product conforms to European Directive98/79/EC.

PremiseAUTION MICRO PU-4210 OPERATING MANUALDOC. No. OM-28373 REV.5 06/05

e

Page 3: Operating Manual - Hello from Web Server micro operating... · DOC. No. OM-28373 REV.5 06/05 AUTION MICRO PU-4210 OPERATING MANUAL e. Introduction This manual contains outlines, instructions

IntroductionThis manual contains outlines, instructions for the

operation, maintenance and troubleshooting for the

AUTION MICRO PU-4210. Read this manual carefully

before use the Analyzer in order to operate it correctly.

After you read the manual, keep it near the Analyzer to

refer to it when necessary.

● This instrument uses urine as a sample. Urine may becontaminated by pathogenic microbes that causeinfectious diseases. Always be careful when handlingurine samples. If operated incorrectly, there is a dangerof infection to you and those around you by pathogenicmicrobes.

● Never touch places where samples may be contactedwith bare hands. When you perform maintenance,always wear protective gloves to prevent exposure topathogenic microbes.

● Separate used Test Strips, etc. from general wasteand dispose of them according to local regulations onbiohazardous waste.

● This instrument may becomeinfectious in the course of use.Discard the product in accordancewith local regulations for biohazardouswaste.

AUTION MICRO 2

AUTION MICRO PU-4210 OPERATING MANUAL DOC. No. OM-28373 REV.5 06/05

e

Page 4: Operating Manual - Hello from Web Server micro operating... · DOC. No. OM-28373 REV.5 06/05 AUTION MICRO PU-4210 OPERATING MANUAL e. Introduction This manual contains outlines, instructions

3 AUTION MICRO

Pay careful attention to these caution marks in this Operating Manual.

Caution Marks

To prevent infection of operator or others from pathogenic

microbes, follow the instructions described.

To prevent injuries to operator or others, or material

damages, follow the instructions described.

Failure to follow the instructions may lead to incorrect

results.

To prevent damage to the unit and assist you in

making best use of the capabilities of the Analyzer,

additional explanations and notes are provided herein.

Regarding accidents resulting injury.

Regarding damage and performance of Analyzer.

Reference information on operation, additional

explanations and related functions are provided herein.

Important

Note

Reference

AUTION MICRO PU-4210 OPERATING MANUALDOC. No. OM-28373 REV.5 06/05

e

Page 5: Operating Manual - Hello from Web Server micro operating... · DOC. No. OM-28373 REV.5 06/05 AUTION MICRO PU-4210 OPERATING MANUAL e. Introduction This manual contains outlines, instructions

AUTION MICRO 4

■For installation, read thefollowing precautions carefully

and always follow safety

measures.

●Place the Analyzer on a stable and levelsurface, free of vibration.

●Select a room in which to set up theAnalyzer where the temperature can becontrolled between 10℃ and 30℃, andhumidity in a range of 20 to 80 %.

●Do NOT place the Analyzer where it maybe affected by magnetic materials includingmagnets or speakers.

●Do NOT place the Analyzer where it maybe affected by chemicals, corrosive gasesand electronic noise.

●Do NOT place the Analyzer where it maybe affected by direct sunlight or wind.

Precautions for Installation AUTION MICRO PU-4210 OPERATING MANUAL DOC. No. OM-28373 REV.0 04/01

Page 6: Operating Manual - Hello from Web Server micro operating... · DOC. No. OM-28373 REV.5 06/05 AUTION MICRO PU-4210 OPERATING MANUAL e. Introduction This manual contains outlines, instructions

AUTION MICRO PU-4210 OPERATING MANUALDOC. No. OM-28373 REV.0 04/01

5 AUTION MICRO

●When the instrument is not in use, unplugthe AC adapter.

●If an abnormal sound, odor or smoke isfound, turn the power off immediately andconsult your distributor after making surethat the abnormal situation is not continuing.

●Do NOT drop or hit the Analyzer.

●Should the Analyzer stop functioning,contact your distributor for repairs. Do NOTattempt to repair it your self. Doing so mayresult in damaging the Analyzer or injury.

●Do NOT place the Analyzer where it maybe affected by water. Samples may causedamage to the Analyzer.

●Plug in an exclusive AC adapter.

Precautions for Installation

Page 7: Operating Manual - Hello from Web Server micro operating... · DOC. No. OM-28373 REV.5 06/05 AUTION MICRO PU-4210 OPERATING MANUAL e. Introduction This manual contains outlines, instructions

AUTION MICRO PU-4210 OPERATING MANUAL DOC. No. OM-28373 REV.0 04/01

AUTION MICRO 6

AUTION MICRO(PU-4210)Operating Manual

Premise

Introduction

Caution Marks

Precautions for Installation

Table of Contents

Outline of AUTION MICRO

Shipping Carton

Parts Description and Functions

Display Screen and Key Arrangement

Flow of Measurement

Cautions

Start Up

Setting Up Parameters

Normal Measurement

Check Measurement

Auxiliary Function

Special Parameters

Printer

Printing Item

Maintenance

Troubleshooting

…………………………………………………………1

……………………………………………………2

…………………………………………………3

…………………………………4

……………………………………………6

…………………………………7

………………………………………………8

……………………………9

……………………10

………………………………………11

………………………………………………………12

………………………………………………………14

……………………………………15

………………………………………16

………………………………………19

……………………………………………21

…………………………………………26

…………………………………………………………27

…………………………………………………28

…………………………………………………29

………………………………………………30

Introduction

Measurement

Printing

Maintenance and

Troubleshooting

Table of Contents

Page 8: Operating Manual - Hello from Web Server micro operating... · DOC. No. OM-28373 REV.5 06/05 AUTION MICRO PU-4210 OPERATING MANUAL e. Introduction This manual contains outlines, instructions

AUTION MICRO PU-4210 OPERATING MANUALDOC. No. OM-28373 REV.0 04/01

7 AUTION MICRO

The AUTION MICRO (PU-4210) is a small portable urine-analyzing device

which anyone can use anytime, anywhere. Since the Printer is removed

from the Analyzer, you can operate it by a minimal key operation.

The device is equipped with transmission capability to adapt to

systematization.

Characteristics

Minimum MaintenanceThe only necessary maintenance item is cleaning the Strip Holder.

Long Life Light SourceUse of LED and short-time lighting enables a longer light source life.

Temperature Correction FunctionAn automatic correction function, which is provided by the internal thermalsensor, solves any errors caused by temperature changes in the environment.

Power SourceUse a dry battery or a AC adapter when using the Analyzer alone, and use theAC adapter when using it with a Printer.

Regent SelectionAUTION Sticks 10EA and Uriflet 9UB are available to use.

100 memoriesMaximum 100 measurement results can be stored.When more than 100 samples have been measured, older data will be automaticallydeleted in order.

Outline of PocketChem UA Introduction

Page 9: Operating Manual - Hello from Web Server micro operating... · DOC. No. OM-28373 REV.5 06/05 AUTION MICRO PU-4210 OPERATING MANUAL e. Introduction This manual contains outlines, instructions

AUTION MICRO PU-4210 OPERATING MANUAL DOC. No. OM-28373 REV.0 04/01

AUTION MICRO 8

Analyzer

OptionalAccessories

OptionalConsumables

RS-232C Cable Thermal Printer Paper    Check Strip Set

Strip Holder

Printer

Shipping carton of AUTION MICRO (PU-4210) contains the following.

Make sure that all of the following are included:

Shipping Carton Introduction

AC Adapter     Check Strip Set  Thermal Printer Paper  Operating Manual(this booklet)

Compact Urine Analyzer

Operating Manual

Page 10: Operating Manual - Hello from Web Server micro operating... · DOC. No. OM-28373 REV.5 06/05 AUTION MICRO PU-4210 OPERATING MANUAL e. Introduction This manual contains outlines, instructions

AUTION MICRO PU-4210 OPERATING MANUALDOC. No. OM-28373 REV.0 04/01

9 AUTION MICRO

Front

Bottom (Analyzer) Rear (Analyzer)

8

21

3

74

5

9

6

    Name

① Display Screen

② Operator Panel

③ Optical System

④ Power Input Terminal

⑤ Connection Terminal

⑥ DATA OUT Terminal

⑦ Battery Box

⑧ FEED key

⑨ Lock Release Button

         Function

Displays measurement results and error codes

For setting conditions and starting measurement

Measures samples

For connection of exclusive AC adapter

For electrical connection to the Printer

Terminal for connection to a PC or other devices

Place batteries here

Feeds printer paper

Unlocks connection between the Analyzer and Printer

Parts Description and Functions Introduction

Parts description in this Operating Manual is as follows:

Page 11: Operating Manual - Hello from Web Server micro operating... · DOC. No. OM-28373 REV.5 06/05 AUTION MICRO PU-4210 OPERATING MANUAL e. Introduction This manual contains outlines, instructions

AUTION MICRO PU-4210 OPERATING MANUAL DOC. No. OM-28373 REV.0 04/01

AUTION MICRO 10

Display Screen and Key Arrangement Introduction

4. Delete1. Memory2. ID

3. No.

11. Test Strip Icon

12. Waiting Icon 13. Transmission Icon

14. Printer Icon

15. Battery Icon

10. figure display13 digits

9. Item Name

7. Test Strip

6. Time

LCD

8. Semiquantitative Symbol

Memory

ID

No.

Delete

Date

Time

Test Strip

Semiquantitative Symbol

Item Name

Figure display 13 digits

Test Strip Icon

Waiting Icon

Transmission Icon

Printer Icon

Battery Icon

1.

2.

3.

4.

5.

6.

7.

8.

9.

10.

11.

12.

13.

14.

15.

Displayed when the data is redisplayed.

Displayed (Blinks during ID setting) when ID is displayed.

Displayed (Blinks during No. setting) when No. is displayed.

Blinks when the memory is deleted. For deletion, enter [ENT].

Light is on (Blinks during Date setting) when the Date is displayed.

Light is on (Blinks during Time setting) when the Time is displayed.

Displayed when the Test Strip type is displayed. Blinks during Test Strip type setting.

Displayed according to measurement results when measurement results are displayed or redisplayed.

Displayed according to test items when measurement results are displayed or redisplayed.

Measurement No., ID, Date, Time, Test Strip name, etc. are displayed. (13 digits)

Blinks when the Test Strip is dipped or the Test Strip Holder is moved.

Displayed from the measurement start to finish.

Displayed for the external output.

Displayed when printing is performed.

Displayed in three different signs depending on the degree of exhaustion.

5. Date

1. Function Key or Numeric Key

2. Start Key, Enter Key or Power Switch

3. Stop Key or Hyphen Key

Key Arrangement Specification

Page 12: Operating Manual - Hello from Web Server micro operating... · DOC. No. OM-28373 REV.5 06/05 AUTION MICRO PU-4210 OPERATING MANUAL e. Introduction This manual contains outlines, instructions

AUTION MICRO PU-4210 OPERATING MANUALDOC. No. OM-28373 REV.5 06/05

11 AUTION MICRO

Flow chart of measurementThe following flow chart explains a measurement procedure.

Before measurement, understand the whole procedure.

Turn on the main power.

Start Measurement

Dispose of Test Strip.

Dip the Test Strip in the sample for 2 seconds in accordance with the alarm.

Remove the excess sample.

Set the Test Strip onto the Strip Holder.

The Strip Holder moves inside of the Analyzer and measurement starts.

Measurement Results are displayed and the Strip Holder comes out from the Analyzer.

Enter Measurement No. and Patient ID.Select Test Strip Type

Continue measurement

Finish measurement

Clean the Strip Holder, and turn off the main power.

Flow of Measurement Measurement

Wear protective gloves to prevent exposure to pathogenic

microbes.

Separate used test papers, etc. from general waste anddispose of them according to local regulations onbiohazardous waste.

e

Page 13: Operating Manual - Hello from Web Server micro operating... · DOC. No. OM-28373 REV.5 06/05 AUTION MICRO PU-4210 OPERATING MANUAL e. Introduction This manual contains outlines, instructions

AUTION MICRO PU-4210 OPERATING MANUAL DOC. No. OM-28373 REV.0 04/01

AUTION MICRO 12

Wear protective gloves to prevent exposure to pathogenic microbes.

Use a fresh urine sample. If the sample urine is not measured within anhour after collection, cap it tightly and keep it refrigerated. Be sure toallow it to reach to room temperature before measurement.

Mix the sample thoroughly before measurement. Do NOT centrifuge them.

Prepare an adequate amount of samples so that the whole pad portion ofthe Test Strip can be immersed.

Do NOT add any preservative, disinfectant, or detergent to the sample.

Keep the sample out of direct sunlight.

If samples containing ascorbic acid are measured, the measured valuesof glucose and occult blood are occasionally lower than the real values.

If drug-administered urine or visually observed hematuria are measured,correct measurement results sometimes cannot be obtained.

Handling Samples

Cautions Measurement

Page 14: Operating Manual - Hello from Web Server micro operating... · DOC. No. OM-28373 REV.5 06/05 AUTION MICRO PU-4210 OPERATING MANUAL e. Introduction This manual contains outlines, instructions

AUTION MICRO PU-4210 OPERATING MANUALDOC. No. OM-28373 REV.0 04/01

Do NOT use Test Strips of other manufacturers.

Use only our Test Strips,“AUTION Sticks, Uriflet”.Read the attached manuals of the Test Strips.

Do NOT use expired Test Strips.

Do NOT use expired Test Strips and/or Test Strips if the reagent padsshow any signs of discoloration, deformation or deterioration.

Prepare Test Strips just before the measurement.

Take out only the necessary number of Test Strips to use and close thecap. If Test Strips are left out for a long time, they may absorb humidityor collect dust, causing incorrect measurement results.

Do NOT touch the reagent pad of the Test Strip.

Do NOT touch the reagent pad of the Test Strip, otherwise correctmeasurement results may not be obtained.

Set Test Strip type.

Set the Test Strip type before measurement. If different Test Strips areused, correct measurement results may not be obtained.

Keep the desiccant.

Keep the desiccant contained in the Test Strip container until all TestStrips are used.

13 AUTION MICRO

Handling Test Strips

Cautions Measurement

Page 15: Operating Manual - Hello from Web Server micro operating... · DOC. No. OM-28373 REV.5 06/05 AUTION MICRO PU-4210 OPERATING MANUAL e. Introduction This manual contains outlines, instructions

AUTION MICRO PU-4210 OPERATING MANUAL DOC. No. OM-28373 REV.5 06/05

AUTION MICRO 14

1. Preparation.When using the Printer, make sure that the

Printer is connected to the Analyzer correctly

and paper is loaded. If there is a red strip on

the both sides of the paper, replace the roll.

2. Turn on the power.Connect the exclusive AC adapter with the

power input terminal on the rear of the device

(while using the AC adapter, battery is not

consumed). When only the Analyzer is

operated with batteries, insert batteries into

the battery box at the bottom of the Analyzer.

Press [START-ENT] on the operator panel to

turn on the power. After initialization, the

Ready Screen appears.

There are two screens for the measurement-standby stage.(Ready Screen).

In either case, press [START] to start measurement.

Make sure that the Strip Holder is set up correctly.

Start Up Measurement

Do NOT touch the optical system.

Doing so may result in injury.

Ready Screen 1The latest measurement

result is displayed on

the LCD.

Ready Screen 2Parameters for the next

measurement are

displayed.

(Measurement No., ID,

Test Strips, Measurement

Item for Test Strip)

e

Page 16: Operating Manual - Hello from Web Server micro operating... · DOC. No. OM-28373 REV.5 06/05 AUTION MICRO PU-4210 OPERATING MANUAL e. Introduction This manual contains outlines, instructions

AUTION MICRO PU-4210 OPERATING MANUALDOC. No. OM-28373 REV.0 04/01

1. Setting up the measurement number.Press [No.] from the Ready Screen.

Measurement No. setting mode appears and“No.”and

the cursor blinks.

Enter any numbers using the numeric keys.(Max. 4 digits)

Press [ENT] to confirm the entry. Press [STOP] to cancel,

and the Ready Screen is restored.

Measurement No. can be set arbitrarily.Once the Measurement No. is set, the number increaseswith every measurement.

2. Setting up of Patient ID.Enter [ID] from the Ready Screen. The patient ID setting

mode appears and“ID”and the cursor blinks.

Enter any numbers using the numeric keys.

(Max. 13 digits including hyphens)

Press [ENT] to confirm the entry. The Ready Screen is

restored.

1: [STOP] functions as a hyphen key.To cancel the entry, press [ENT] once before re-entry.

2: When the patient ID is entered, the Ready or Measurementscreen displays Measurement No. and Patient IDalternately.

3: Patient ID is included in the printing results and externaltransmission. (When not entered, -----------is displayed.)

3. Setting of Test Strip Type.Press [STRIP] from the Ready Screen, Test Strip type

selection mode appears and“STRIP”blinks, and theTest Strip type and Item are displayed.

To change the Test Strip type and Item, press [STRIP].

(10EA→9Ub(9UB)→C(CHECK)→10EA)Press [ENT] to confirm the entry.

Press [STOP] to cancel, and the Ready Screen is restored.

15 AUTION MICRO

Setting Up Parameters Measurement

Reference

Reference

Page 17: Operating Manual - Hello from Web Server micro operating... · DOC. No. OM-28373 REV.5 06/05 AUTION MICRO PU-4210 OPERATING MANUAL e. Introduction This manual contains outlines, instructions

AUTION MICRO PU-4210 OPERATING MANUAL DOC. No. OM-28373 REV.5 06/05

AUTION MICRO 16

1. Prepare sample.Prepare sample in a sample cup.

Make sure that sample isenough to cover the whole reagent pads.

Stir sample well.

Do NOT use a centrifuge for mixing.

Prepare tissue paper to wipe off excess urine.

2. Prepare the Test Strip.Take out only the necessary number of strips

from the Test Strip container.

Close the cap.

Read the attached manualsfor the Test Strip.

Reagent Pads

Close the lid tightly

Normal Measurement Measurement

Wear protective gloves to prevent exposure to pathogenic

microbes.

Important

Important

Important

e

Page 18: Operating Manual - Hello from Web Server micro operating... · DOC. No. OM-28373 REV.5 06/05 AUTION MICRO PU-4210 OPERATING MANUAL e. Introduction This manual contains outlines, instructions

AUTION MICRO PU-4210 OPERATING MANUALDOC. No. OM-28373 REV.0 04/01

3. Press [START] to start Normal Measurement.Test Strip Icon blinks and the timer starts to count

down the timing of dipping of the Test Strip from [3

sec]. (A timing buzzer also sounds in accordance with

the count down.)

To stop the measurement,keep pressing [STOP] for 2-3 seconds during measurement. Do NOT move the Analyzer duringmeasurement. If impact or shock is given to the Analyzer,measurement cannot be done correctly. The Test Stripmay become jammed in the Analyzer.

4. Dip a Test Strip in the sample.Following the indicator timing (--sec), a timing buzzer

sounds for 2 seconds.

Keep dipping the Test Strip while the buzzer is

sounding.

Dip the Test Strip properly for 2seconds following the sound of thetiming buzzer. Shorter dipping timemay result in insufficient coloringand longer dipping time may causedischarging of the reagent.

5. Remove the excess sample.When the timing buzzer stops, pull out the Test

Strip and remove the excess sample with tissue

paper.

17 AUTION MICRO

Do NOT dip the black marker

Normal Measurement Measurement

Reference

Important

Important

Page 19: Operating Manual - Hello from Web Server micro operating... · DOC. No. OM-28373 REV.5 06/05 AUTION MICRO PU-4210 OPERATING MANUAL e. Introduction This manual contains outlines, instructions

AUTION MICRO PU-4210 OPERATING MANUAL DOC. No. OM-28373 REV.0 04/01

AUTION MICRO 18

6. Set the Test Strip.While the Test Strip Icon is blinking, place the Test

Strip onto the Strip Holder.

A few seconds before the Strip Holder moves inside of the Analyzer, the Test Strip Iconstops blinking. When the StripHolder is set inside the Analyzer,the Test Strip Icon disappears.If the Test Strip is not correctlyset on the Strip Holder, correct measurement results cannot beobtained.

7. Measurement.The Strip Holder is pulled into the Analyzer

and the measurement starts.

8. Finish Measurement.When measurement is completed,the

measurement result is displayed and the

Strip Holder comes out from the Analyzer.

Discard the used Test Strip, and when

finishing the whole measurement process,

remove the Strip Holder from the Analyzer

and clean it.

9. Turn off the power.Keep pressing [START] for 2-3 seconds to turn off the power (the Strip Holder is

pulled into the Analyzer automatically).

When no key entry is made for more than 3 minutes, the power automatically turns off.

Normal Measurement Measurement

Reference

Important

Make sure that the tip of Test Strip

reaches to the left end of the holder.

Reference

Page 20: Operating Manual - Hello from Web Server micro operating... · DOC. No. OM-28373 REV.5 06/05 AUTION MICRO PU-4210 OPERATING MANUAL e. Introduction This manual contains outlines, instructions

AUTION MICRO PU-4210 OPERATING MANUALDOC. No. OM-28373 REV.5 06/05

For installation, read the following precautions carefully and always

follow safety measures.

1. Clean the Strip Holder.Wash the Strip Holder and dry it completely with

the soft cloth. See“Maintenance”.

Make sure the holder is clean,otherwise dirt may adhere tothe Check Strip and damageit.

2. Set the Test Strip Type.Set the Test Strip Type to C (Check).

See“Setting Up Parameters”.

3. Prepare a Check Strip.Remove 1 Check Strip from the Check Strip

container.

Do NOT touch the gray-colored part of Check Strip.

19 AUTION MICRO

Check Strip(Gray)

Semiquantitative Value Label

Check Measurement Measurement

Wear protective gloves to prevent exposure to pathogenic

microbes.

Note

Important

e

Page 21: Operating Manual - Hello from Web Server micro operating... · DOC. No. OM-28373 REV.5 06/05 AUTION MICRO PU-4210 OPERATING MANUAL e. Introduction This manual contains outlines, instructions

AUTION MICRO PU-4210 OPERATING MANUAL DOC. No. OM-28373 REV.0 04/01

AUTION MICRO 20

4. Press [START] to start Check

Measurement.While the Test Strip Icon is blinking, place the

Check Strip onto the Strip Holder.

A few seconds before the StripHolder moves inside of the Analyzer, the Test Strip Icon stops blinking.When the Strip Holder is setinside the Analyzer, the Test Strip Icon disappears.

5. Confirm the results.Compare the Check Measurement results with the ranges of semiquantitative

values attached to the Check Strips container.

●When the Check Measurement results are within the range:The Analyzer is performing correctly. End Check Measurement.

●When the Check Measurement results are out of the range:Repeat the Check Measurement with the other Check Strip.

6. Repeat the Check Measurement with

the other Check Strip.

●If the second Check Measurement results are within the range:The Check Strip used for the first measurement was defective.

Do NOT use it again.

●If the second Check Measurement results are still out of the range:Contact your distributor.

Check Measurement Measurement

Reference

Page 22: Operating Manual - Hello from Web Server micro operating... · DOC. No. OM-28373 REV.5 06/05 AUTION MICRO PU-4210 OPERATING MANUAL e. Introduction This manual contains outlines, instructions

AUTION MICRO PU-4210 OPERATING MANUALDOC. No. OM-28373 REV.0 04/01

1. Calendar Display and Setting.Press [DATE] from the Ready Screen,the current

date is displayed. (The Ready Screen is restored

in 2-3 seconds automatically.)

Press [DATE] again while the date is displayed,

the Date Setting Screen appears and“DATE”and cursor blink.

Enter the current date.

Press [ENT] to confirm the entry. Press [STOP]

to cancel, and the Ready Screen is restored.

2. Time Display and Setting.Press [TIME] from the Ready Screen, current

time is displayed.

(The Ready Screen is restored in 2-3 seconds

automatically).

Press [TIME] again while the time is displayed,

the Time Setting Screen appears and [TIME]

and cursor blink.

Enter the current time.

Press [ENT] to confirm the entry.

Press [STOP] to cancel, and the Ready Screen

is restored.

21 AUTION MICRO

Auxiliary Function Measurement

Page 23: Operating Manual - Hello from Web Server micro operating... · DOC. No. OM-28373 REV.5 06/05 AUTION MICRO PU-4210 OPERATING MANUAL e. Introduction This manual contains outlines, instructions

AUTION MICRO PU-4210 OPERATING MANUAL DOC. No. OM-28373 REV.0 04/01

AUTION MICRO 22

3. Redisplay of Measurement

Results.

Press [MEMORY▲] or [MEMORY▼] from theReady screen, the latest data is redisplayed

([MEMORY] is displayed at the top left of the

screen).

Press [MEMORY▼] again, one previous data isdisplayed.

Repeat this procedure, if necessary.

Press [MEMORY▲], the next data is displayed.

Press [STOP], the Ready Screen is restored.

If No., ID, DATE and TIME are pressed during Redisplay ofdata, Measurement No.,Patient ID, Measurement Dateand Measurement Time on theRedisplayed data are displayed.(Measurement No. is displayedas default. This setting isreflected in all data during aseries of redisplay operation.)

Auxiliary Function Measurement

Reference

Page 24: Operating Manual - Hello from Web Server micro operating... · DOC. No. OM-28373 REV.5 06/05 AUTION MICRO PU-4210 OPERATING MANUAL e. Introduction This manual contains outlines, instructions

AUTION MICRO PU-4210 OPERATING MANUALDOC. No. OM-28373 REV.0 04/01

4. Delete Data.A: Delete Single Data.

Press [DELETE] on the Redisplay screen of

measurement results,“DELETE”blinks.(The measurement results are also displayed.)

Press [ENT] to delete the displayed measurement

data and the next data is displayed.

To cancel it, press [STOP] and the Redisplay

screen is restored.

B: Delete All Data.Keep pressing [DELETE] on the Redisplay screen

of the measurement results for 2-3 seconds,

“DELETE”blinks and“ALL”is displayed.(Other displayed items disappear.)

Press [ENT] to delete all measurement data in the

memory.

To cancel it, press [STOP] and the Redisplay

screen is restored.

5. Retransmit.A: Retransmit Single Data.

Press [SEND] on the Redisplay screen of the

measurement results, the Transmission Icon blinks.

(The measurement Results are also displayed.)

Press [ENT] to transmit the displayed measurement

data. To cancel it, press [STOP] and the Redisplay

screen is restored.

23 AUTION MICRO

Auxiliary Function Measurement

Page 25: Operating Manual - Hello from Web Server micro operating... · DOC. No. OM-28373 REV.5 06/05 AUTION MICRO PU-4210 OPERATING MANUAL e. Introduction This manual contains outlines, instructions

AUTION MICRO PU-4210 OPERATING MANUAL DOC. No. OM-28373 REV.0 04/01

AUTION MICRO 24

B: Retransmit All Data.Keep pressing [SEND] on the Redisplay screen of

the Measurement Results for 2-3 seconds, the

Transmission Icon blinks and“ALL”is displayed.

(Other displayed items disappear.)

Press [ENT] to transmit all measurement data in

the memory from the latest data.

To cancel it, press [STOP] and the Redisplay screen

is restored.

To stop transmission, press [STOP] during transmission.

6. Reprint.

(when the Printer is connected.)A: Reprint Single Data.

Press [PRINT] on the Redisplay screen of the

measurement results, the Printer Icon blinks.

(The measurement results are also displayed.)

Press [ENT] to print the displayed measurement

data. To cancel it, press [STOP] and the Redisplay

screen is restored.

B: Reprint All Data.Keep pressing [PRINT] on the Redisplay screen of

the measurement results for 2-3 seconds, the

Printer Icon blinks and [All] is displayed.

(Other displayed items disappear.)

Auxiliary Function Measurement

Reference

Page 26: Operating Manual - Hello from Web Server micro operating... · DOC. No. OM-28373 REV.5 06/05 AUTION MICRO PU-4210 OPERATING MANUAL e. Introduction This manual contains outlines, instructions

AUTION MICRO PU-4210 OPERATING MANUALDOC. No. OM-28373 REV.01 01/01

25 AUTION MICRO

Press [ENT] to print all measurement data in

memory from the latest data. To cancel it, press

[STOP] and the Redisplay screen is restored.

To stop printing, press [STOP] during printing.To print the latest data,press[PRINT] on the Ready Screen.

7. Battery Icon.A battery icon at the lower right of the analyzer

display tells you the timing of battery exchange.

Follow the instructions below, and replace the

battery to a new one if necessary.

The supplied AC adapter can be used as well.

The battery is full with power.

Although measurement can be continued

with this battery, the power is getting

short. Prepare a new battery.

This icon sometimeschanges to the next icon(blinking) even if enoughpower remains in.

The battery runs out.

To continue measurement, replace the

battery to a new one.

Auxiliary Function Measurement

Reference

(Blinking)

② ③

②③

Note

Page 27: Operating Manual - Hello from Web Server micro operating... · DOC. No. OM-28373 REV.5 06/05 AUTION MICRO PU-4210 OPERATING MANUAL e. Introduction This manual contains outlines, instructions

AUTION MICRO PU-4210 OPERATING MANUAL DOC. No. OM-28373 REV.0 04/01

The following special parameters are available to set with the PU-4210.

(The parameters are stored until the settings are changed.)

●Orientation of the Test Strip set.Setting of Test Strip from the right side or from the left side of the Analyzer

(default is set to the right side).

To change the Test Strip orientation, remove the StripHolder from the Analyzer beforehand.

●Date Order.Choose Date order among Y-M-D, M-D-Y, and D-M-Y.

(The default setting is Y-M-D).

Setting is reflected on the screen display and printing. (Fixed for external output.)

●Setting of measurement units and symbols for semi-quantitative ranks.Set the measurement units for the measurement results (SI or usual) and format

of semi-quantitative ranks ( +++ or +n ) ( default is set to usual, +++).

Setting is reflected on the printing and external output. (Fixed for the screen display.)

1. Operation. Turn on the power while pressing [STOP].

(Keep pressing [STOP] during initialization.)

After initialization, the entry screen for 3 digits

numbers is displayed and the parameter setting

mode appears.

The numeral of each digit corresponds to the

following. Input the numerals for the desired setting.

In this case, if numerals other than the designated

ones are input, they are not accepted.

Hundreds 1:Right setting 2:Left settingTens 1:Y-M-D 2:M-D-Y 3:D-M-Y

Ones 1:mg/dl,+++ 2:mg/dl,+n 3:mmol/l,+++ 4:mmol/l,+n

Press [ENT] after the entry to change the settings.

The Ready Screen is restored.

e.g.) Change from the default (Right setting, Y-M-D, mg/dl,+++)

●When“232”is entered,(Left setting, D-M-Y, mg/dl, +n) is set. Then the Ready Screen is restored.

●When“063”is entered,(Right setting, Y-M-D, mmol/l, +++) is set. Then the Ready Screen is restored.

AUTION MICRO 26

previous setting

Special Parameters Measurement

Note

Page 28: Operating Manual - Hello from Web Server micro operating... · DOC. No. OM-28373 REV.5 06/05 AUTION MICRO PU-4210 OPERATING MANUAL e. Introduction This manual contains outlines, instructions

AUTION MICRO PU-4210 OPERATING MANUALDOC. No. OM-28373 REV.0 04/01

27 AUTION MICRO

1. Attachment or Removal of

Printer.When connecting the Printer to the Analyzer,

insert the hook of the Printer in the hole at the

bottom of the Analyzer, and then slide it firmly

enough to achieve a secure fit.

When removing the Printer, slide the Analyzer

to the front while pressing the lock release button of the Printer.

Before you attach or remove the Printer, make sure to turnoff the power and pull out the AC adapter of the Printer.

2. Paper Jam.When paper is jammed, press [FEED] of the

Printer to remove the paper.

3. Loading of Paper.After confirming the face and the back of the

paper, put the thermal printer paper into the

insertion position, and then the paper will be

fed automatically.

When inserting a new thermal Printer paper,cut the edge with scissorsto make it even, and set it inthe position.

4. Error Comments.When an error occurs, error codes and error comments are printed.

(Only error codes are displayed on the screen.)

※See“Troubleshooting”.

lock release button

FEED

Printer Printing

Note

Note

Page 29: Operating Manual - Hello from Web Server micro operating... · DOC. No. OM-28373 REV.5 06/05 AUTION MICRO PU-4210 OPERATING MANUAL e. Introduction This manual contains outlines, instructions

AUTION MICRO PU-4210 OPERATING MANUAL DOC. No. OM-28373 REV.0 04/01

AUTION MICRO 28

Abnormal Mark

Marks for abnormal values(Normal:Blank)

Measurement Item

Measurement No.

Patient IDDate・Time・ Type of Test Strip・Temperature in Instrument

Measurement Results

Ranks (Semi-quantitative symbols)

Concentration values

Measurement Units

Abnormal Mark

Measurement No.

Patient ID

Date, Time

Type of Strip

Temperature in

Instrument

Measurement

Results

When the measurement results is abnormal (Rank is more than“+1”.),

sample abnormal mark‘●’is displayed.

4 digits

ID is entered using keys. When there is no ID, [------------------]is printed.

There are 3 printing patterns of the date

YMD, MDY, DMY

Example 2001-07-05 09:53

      07-05-2001 09:53

      05-07-2001 09:53

The Test Strip Type is displayed.

Example 9UB

Temperature in the instrument is printed in ℃

Example 29℃

Abnormal Mark / When an abnormal measurement result (rank is over “+1”) is recognized,

‘*’ is printed before the name of item.

Measurement Item / GLU, URO, PRO, KET, BLD, BIL, NIT, LEU, pH, S.G.

Ranks (Semi-quantitative symbols), Concentration,

Measurement Units.

Printing Item Printing

Page 30: Operating Manual - Hello from Web Server micro operating... · DOC. No. OM-28373 REV.5 06/05 AUTION MICRO PU-4210 OPERATING MANUAL e. Introduction This manual contains outlines, instructions

AUTION MICRO PU-4210 OPERATING MANUALDOC. No. OM-28373 REV.5 06/05

Clean the Strip Holder after finishing measurement at the end of the day.

29 AUTION MICRO

1. Slide the Strip Holder to remove. 2. Wash the Strip Holder with

detergent or alcohol and then rinse

it with tap water. Especially, clean

the groove with special care.

3. Dry the Strip Holder completely

with a soft cloth.

4. Insert the Strip Holder completely

by using grooves of the backside of

the Strip Holder.

Maintenance Troubleshooting and Maintenance

Wear protective gloves to prevent exposure to pathogenic

microbes.

Discard used cotton swabs and used gloves according to

local regulations for biohazardous waste.

e

Page 31: Operating Manual - Hello from Web Server micro operating... · DOC. No. OM-28373 REV.5 06/05 AUTION MICRO PU-4210 OPERATING MANUAL e. Introduction This manual contains outlines, instructions

AUTION MICRO PU-4210 OPERATING MANUAL DOC. No. OM-28373 REV.0 04/01

AUTION MICRO 30

When measurement and/or printing trouble occurs, the Analyzer displays

error codes. Perform the appropriate remedy described in the following

chart. If measurement is still impossible, contact your distributor.

Optical System does not function correctly.

E001Optical Movement

Make sure that the Strip Holder and Test Strip are set properly and turn on the power.

Lamp intensity of LED is too strong.

E002Lamp Intensity High

Clean the White Plate of the Strip Holder with a soft cloth moistened with water or alcohol and turn on the power.

Lamp intensity of LED is too weak.

E003Lamp Intensity Low

Clean the White Plate of the Strip Holder with a soft cloth moistened with water or alcohol and turn on the power.

No Printer PaperE004 Load printer paper.(Press [STOP] to restore the Ready Screen.)

Printer does not function correctly.

E005 Make sure that the Analyzer and the printer are connected correctly. (Press [STOP] to restore the Ready Screen.)

The Test Strip is not inserted correctly.

E006Strip Error

Insert the Test Strip and Strip Holder correctly and start measurement again.(Press [STOP] to restore the Ready Screen.)

Error CodeError Comment Cause Remedy

Test Strip type is not judged.

Important

Troubleshooting Troubleshooting and Maintenance

Page 32: Operating Manual - Hello from Web Server micro operating... · DOC. No. OM-28373 REV.5 06/05 AUTION MICRO PU-4210 OPERATING MANUAL e. Introduction This manual contains outlines, instructions

84-00259E

SampleTest StripMeasurement ItemMeasurement Method

Measurement WavelengthSample Supply Reaction TimeCalibration FunctionDisplayOperator Panel Memory CapacityThermal Correction FunctionSG Correction FunctionExternal OutputMeasurement Environment Condition DimensionsWeight Power Supply

 1234

56789

101112131415161718

UrineAUTION Sticks 10EA, Uriflet 9UBGLU, URO, PRO, KET, BLD, BIL, NIT, LEU, pH, S.G. Test stripDual or Single (BLD only) Wavelength Reflectance MethodTriple Wavelength, Individual LED (565, 632, 755nm)Manual DippingApprox. 45 seconds/testAutomatic Calibration by CAL padLCD12 keys100 testsAutomatic Correction Function (10-30℃) Automatic Correction according to pH measurement valueConformed to RS-232CTemperature:10~30℃, Humidity:20~80%RH 125(W)×87(D)×31(H) mmApproximately 300gSize AA alkali cells (×2) (Or AC adapter)

No. Items Contents

PrinterPrinter PaperDimensionsWeightPower Supply

 12345

36 digits of half-sized characters, Line Thermal PrinterThermal Printer Paper 58mm (Width)×φ26mm 125(W)×163(D)×38(H) mmApproximately 300g AC adapter (Approximately 20W)

No. Items Contents

Printer

AnalyzerSpecifications