Operating Instructions Excellence Micro- Ultra … of Contents 1 Introduction 5 1.1 Symbols and...
Transcript of Operating Instructions Excellence Micro- Ultra … of Contents 1 Introduction 5 1.1 Symbols and...
XP / XS Models – Part 1O
pera
ting
Inst
ruct
ions Excellence Micro- Ultra-Micro Balances
Table of Contents
Introduction1 5
Symbols and presentations used1.1 6
Safety Information2 7
Definition of signal warnings and symbols2.1 7Product specific safety notes2.2 7
Overview of XP/XS Micro and Ultra Micro Balances3 9
Installation and Putting into Operation4 11
Unpacking and checking the delivered items4.1 11Unpacking the balance4.1.1 11Scope of delivery4.2 12Selecting a location4.3 13Assembling the balance4.4 13Power supply4.5 13Operating of the glass draft shield4.6 14Setting the reading angle and positioning the terminal4.7 15Optimise the readability of the terminal4.7.1 15Remove terminal and place close to the balance4.7.2 15Transporting the balance4.8 16Transporting over short distances4.8.1 16Transporting over long distances4.8.2 16Below-the-balance weighing4.9 17
First Steps5 18
Switching on / off5.1 18Leveling the balance5.2 18Leveling aid of the XP models5.3 18
Maintenance6 20
Cleaning6.1 20Disposal6.2 22
Technical Data7 23
General data7.1 23Explanatory notes for the METTLER TOLEDO AC adapter7.2 23Model-specific data7.3 24Dimensions7.4 27Dimensions of the terminal and control unit of XP models7.4.1 27Dimensions of the terminal and control unit of XS models7.4.2 28Dimensions of the weighing cell (XS and XP models)7.4.3 29Interfaces7.5 30Specifications of RS232C7.5.1 30Specifications of "Aux" connection7.5.2 30
Accessories and Spare Parts8 31
Accessories8.1 31Spare parts8.2 38
Table of Contents 3
Appendix9 41
MT-SICS interface commands and functions9.1 41Procedure for certified balances9.2 41
Index10 43
Table of Contents4
1 IntroductionThank you for choosing a METTLER TOLEDO balance.
The balances of the XP/XS line combine a large number of weighing and adjustment possibilities with exceptionally convenient operation.
In this chapter you will be given basic information about your balance. Please read right through this chaptercarefully even if you already have experience with METTLER TOLEDO balances. Please pay special attention tothe safety warnings!
The different models have different characteristics regarding equipment and performance. Special notes in thetext indicate where this makes a difference to operation.
The XP/XS line comprises a range of balances which differ from each other in relation to their weighing rangeand resolution.
The following features are common to all models of the XP/XS lines:
Glass draft shield for precise weighing even in unstable environments (XP models with motorized drive).
Fully automatic adjustment using internal weights ("ProFACT" on XP models, "FACT" on XS models).
Built-in level sensor, illuminated level indicator and Leveling Assistant for fast and easy leveling (XP modelsonly).
Built-in applications for normal weighing, statistics, formulation, piece counting, percent weighing, density,differential weighing (XP models only) and LabX Client.
Integral RS232C interface.
Slot for second interface (optional).
Touch-sensitive graphics terminal ("Touch screen") with color (XP models) or monochrome display (XSmodels).
Two programmable sensors for hands-off operation ("SmartSens") to speed up frequently recurring tasks(XP models).
A brief word about standards, guidelines, and methods of quality assurance: The balances comply with usualstandards and guidelines. They support standard procedures, specifications, working methods, and reportsaccording to GLP (Good Laboratory Practice). In this connection, records of working procedures and adjustments become very important; for this purpose we recommend you to use a printer from the METTLER TOLEDOrange, since these are optimally adapted to your balance. The balances conform to the applicable standardsand guidelines and possess a EC declaration of conformity. METTLER TOLEDO is certified as manufactureraccording to ISO 9001 and ISO 14001.
The operating instructions for the XP/XS balances consist of 3 separate documents, whose contents arelisted below.
Part 1, this documentContents
Introduction Safety information Installation and putting into operation Leveling the balance Cleaning and service Technical data Accessories Spare parts Interface commands and MT-SICS functions
5Introduction
6 Introduction
Part 2, separate documentContents: Terminal, System, and Applications
Basic principles for using the terminal and the firmware System settings User-specific settings (XP models only) Applications Firmware (Software) updates Error and status messages Conversion table for weight units Recommended printer settings
Part 3, separate documentContents: Adjustments, and Tests
Adjustments Tests
Finding more informationInternet http://www.mt.com/excellence
1.1 Symbols and presentations usedThe following conventions apply to the operating instructions: part 1, part 2 and part 3.
Key and button designations are indicated by a picture or text in square brackets (e.g. [ ] or [On/Off] XPmodels, [ ] or [On/Off] XS models).
This symbol indicates press key briefly (less than 1.5 s).
This symbol indicates press and hold key down (longer than 1.5 s).
These symbols indicate an instruction:prerequisites
1
2 …
steps
results
2 Safety Information
2.1 Definition of signal warnings and symbolsSafety notes are indicated by signal words and warning symbols and contain warnings and information aboutsafety issues. Ignoring safety notes can lead to personal injury, damage to the instrument, malfunctions anderroneous results.
Signal words
WARNING for a hazardous situation with medium risk, possibly resulting in severeinjuries or death if not avoided.
CAUTION for a hazardous situation with low risk, resulting in damage to the device orthe property or in loss of data or minor or medium injuries if not avoided.
Attention (no symbol)for important information about the product.
Note (no symbol)for useful information about the product.
Warning symbols
General hazard Electrical shock
2.2 Product specific safety notes
Intended useYour balance is used for weighing. Use the balance exclusively for this purpose. Any other type of use andoperation beyond the limits of technical specifications without written consent from Mettler-Toledo AG, is considered as not intended.
It is not permitted to use the instrument in explosive atmosphere of gases, steam, fog, dustand flammable dust (hazardous environments).
General safety informationYour instrument meets the state of the art technology and complies with all recognized safety rules, however,certain hazards may arise in extraneous cirmumstances. Do not open the housing of the instrument: It does notcontain any parts which can be maintained, repaired or replaced by the user. If you ever have problems withyour instrument, contact your authorized METTLER TOLEDO dealer or service representative.
Always operate and use your balance only in accordance with the operating instructions part 1, part 2 andpart 3.
The instructions for setting up your new balance must be strictly observed.
If the instrument is not used according to the manufacturer’s operating instructions (part 1, part 2 andpart 3), protection of the instrument may be impaired.
Staff safetyIn order to use the instrument, you must have read and understood the operating instructions. Keep the operating instructions for further reference.
Never make any modifications or constructional alterations to the instrument and use only original spare partsand optional equipment from METTLER TOLEDO.
7Safety Information
8 Safety Information
Safety notes CAUTION
Damage of device
– For use only in dry interior rooms.
– Do not use sharply pointed objects to operate the keyboard! Although your balance is veryruggedly constructed, it is nevertheless a precision instrument. Treat it with correspondingcare.
– Do not open the balance: It does not contain any parts which can be maintained, repaired,or replaced by the user. If you ever have problems with your balance, contact your METTLERTOLEDO dealer.
– Use only balance accessories and peripheral devices from METTLER TOLEDO; they areoptimally adapted to your balance.
CAUTION
Risk of electric shock
Use only the original universal AC adapter delivered with your balance, and check that thevoltage printed on it is the same as your local power supply voltage. Only plug the adapterinto a socket which is grounded.
3 Overview of XP/XS Micro and Ultra Micro Balances
1 Terminal (XS models: Type "S", monochrome / XPmodels: Type "P", color), (details see OperatingInstructions – Part 2)
2 Display (Touch-sensitive "Touch screen")3 Operating keys4 SmartSens sensors (terminal type "P" only)5 Type name6 Control unit7 Drawer with weighing tweezers, cleaning brush,
and cleaning tweezers
8 Door handle9 Weighing chamber plate10 Weighing pan11 Glass draft shield12 Level indicator13 Weighing cell14 Foot screws
15 Socket for control unit
9Overview of XP/XS Micro and Ultra Micro Balances
10 Overview of XP/XS Micro and Ultra Micro Balances
16 Slot for second interface (optional)17 Socket for AC adapter18 Socket for terminal19 RS232C serial interface20 Socket for weighing cell21 Aux sockets for hand- or foot-switch (XS models) or
"ErgoSens" (XP models)
4 Installation and Putting into OperationThis chapter explains how to unpack your new balance, and how to set it up and prepare it for operation. Whenyou have carried out the steps described in this chapter, your balance is ready for operation.
4.1 Unpacking and checking the delivered items
4.1.1 Unpacking the balance
1 Open the outer packaging box.
2 Grasp the strap on the inner box and pull this outof the outer box along with the padding.
1 Remove the padding.
2 Take the inner box out of the plastic bag and placeit on a level surface with the opening flap facingup.
3 Open the inner box (open the flap and remove thecardboard sleeve).
– Remove the following parts from the upper part ofthe packaging:
Documents (1), already removed here. Connecting cable (2) for weighing cell – control
unit. Glass cover (3) of the draft shield. Power cable (4) country-specific for the AC
adapter. AC adapter (5).
11Installation and Putting into Operation
12 Installation and Putting into Operation
1 Lift off the upper part of the inner packaging.
You will find the following parts in the lowerpart:
2 Remove the following parts from the packaging:
Weighing cell (6) with draft shield. Plastic box (7), contains the parts for the draft
disk. Control unit (8) with mounted terminal (9) and
protective cover for the terminal.
1 Remove the parts from the packaging.
2 Remove the shipping lock (10) (plastic protection) from the draft shield.
NotePlease keep all parts of the packaging. This packaging guarantees best possible protection of your balance fortransportation.
See Transporting the balance (page 16).
4.2 Scope of deliveryThe standard scope of delivery contains the following items:
Weighing cell and control unit with terminal installed, protective cover for the terminal• RS232C interface• Slot for second interface (optional)• Feedthroughs for below-the-balance weighing and for antitheft device
Weighing pan is installed, draft disk and hook weighing pan (XP6U) are delivered separately and must beinstalled by the user
AC adapter with country-specific power cable
Connection cable for connecting the weighing cell to the control unit
Cleaning brush
Cleaning tweezers
Weighing tweezers
Production certificate
CE declaration of conformity
Operating instructions part 1 (this document), part 2 and part 3
4.3 Selecting a location
Choose a position which is stable, free from vibration, and as nearlyhorizontal as possible. The supporting surface must be able to bearthe weight of the fully loaded balance safely.
Avoid the following:
Direct sunlight Draft (e.g. from fans or air conditioning) Excessive fluctuations in temperature.
Further information can be found in Weighing the Right Way.
Observe ambient conditions.
See General data (page 23).
NoteIf the balance is not horizontal from the beginning, it will have to beleveled during initial operation.
See Leveling the balance (page 18).
4.4 Assembling the balance
1 Remove the parts for the draft disk from the black plastic box.
2 Assemble the parts according to the instructions in the cover of theplastic box.
3 Connect the terminal cable (1) to the control unit.
4 Use the cable delivered (2) to connect the control unit to the weighing cell.
XP6U only
A hook weighing pan is also delivered with this model.
1 If you wish to use this, remove the standard round weighing pan.
2 Install the hook weighing pan according to the instructions in thecover of the black plastic box.
4.5 Power supply
WARNING
Risk of electric shock
– Ensure that the AC power pack for the balance is only used in accordance with the specifications listed in chapter General Data.
– Your instrument is supplied with a 3-pin power cable with an equipment grounding conductor. Only extension cables which meet this relevant standards and also have an equipment grounding conductor may be used. Intentional disconnection of the equipment grounding conductor is prohibited.
13Installation and Putting into Operation
14 Installation and Putting into Operation
Your balance is delivered complete with an AC adapter and a country-specific power supply cable. The ACadapter is suitable for all power supply voltages in the range of: 100 – 240 VAC, 50/60 Hz, for exact specifications.
See Technical Data (page 23).
First, check whether the voltage of the power supply matches your local line voltage. If this is not the case,on no account connect the balance to the power, but contact your METTLER TOLEDO sales representative ordealer.
Guide the cables so that they cannot become damaged and will not be in your way during your daily work!Take care that the AC adapter cannot come into contact with liquids!
The power plug must be always accessible. Before operating, check all cables for damage.
– Plug the AC adapter (1) into the socket (2) in the back of your balance and into the power supply.
After the balance has been connected to the power supply, it carriesout a self test and is then ready for operation.
NoteIf the display field remains dark, even though the power supply connection functions.
1 First disconnect the balance from the power supply.
2 Open the terminal.
Terminal of XP models
– Press both buttons (1) on the back of the terminal and open theupper part of the terminal.
1 1
Terminal of XS models
– Press both flaps (1) on the side of the terminal and open the upperpart of the terminal.
1 1
1 Check that the plug for the terminal cable (1) is connected correctlyinside the terminal.
2 Ensure that the ferrite core (2) makes full contact on the plug.
4.6 Operating of the glass draft shieldThe glass draft shield on your balance can be opened and closed by turning the door handle.
On XP models, the draft shield can also be operated via the [ ] buttonor the "SmartSens" sensors, see Operating Instructions – Part 2.
AttentionDuring weighing, always ensure that the draft shield is closed!
4.7 Setting the reading angle and positioning the terminal
4.7.1 Optimise the readability of the terminal
Changing the reading angle
1 Press both buttons (1) or flaps (1) used to open the terminal
The top of the terminal can then be pulled up or pushed downuntil it engages in the desired position. A total of 3 setting positions are available.
2 Move it in an appropriate position.
XP Terminal
1 1
XS Terminal
1 1
4.7.2 Remove terminal and place close to the balance
The terminal is attached to the control unit, but can be removed and positioned individually if necessary.
15Installation and Putting into Operation
16 Installation and Putting into Operation
1 Switch off the balance and disconnect it from the power supply.
2 Open the terminal by pressing both buttons or flaps.See Power supply (page 13).
3 Unplug the cable from the connection socket in the terminal.
4 Detach the terminal from the terminal support.On the XP models, unscrew both knurled screws (1) on the insideof the terminal.On XS models, the terminal is attached with just one knurled screw(2).
5 Pull the terminal cable out of the terminal.
6 Release the cable guide (3) on the underside of the control unit (2Torx T-10 screws).
The cable is now free and you can position the terminal separately.The terminal support (4) is attached to the underside of thecontrol unit with 2 screws (Torx T-20). You can leave the terminal support on the control unit or remove it.
7 Reconnect the balance to the power supply.
4.8 Transporting the balance1 Switch off the balance.
2 The balance must be disconnected from the power supply.
3 Remove any interface cable from the balance.It is not necessary to disconnect the control unit from the weighing cell.
4.8.1 Transporting over short distances
If you wish to move your balance over a short distance to a new location, proceed as follows.
CAUTION
Damage of device
Never lift the balance by the glass draft shield, as this can cause damage!
– Grasp the control unit and weighing cell by the sides of the housing and carry them to their new location.See Selecting a location (page 13).
4.8.2 Transporting over long distances
If you want to transport or ship your balance over long distances, or if it is not certain that the balance will betransported upright, use the complete original packaging.
– Place the shipping lock (plastic protection) in the glass draft shield!
4.9 Below-the-balance weighingSo that weighings can be carried out below the working surface (below-the-balance weighing), your balance isprovided with a special hanger.
1 Switch off the balance.
2 Unplug the connection cable for the control unit from the back panel of the weighing cell.
3 Remove the glass cover, the weighing pan, and the draft disk.
4 Remove the draft shield from the unit.
5 Carefully tip the weighing cell toward the back.
6 Turn the cover plate of the hanger for below-the-balance weighinguntil the hole for the feedthrough is exposed.
The weighing cell is now ready for installing the feedthrough forbelow-the balance weighing.
17Installation and Putting into Operation
18 Installation and Putting into Operation
5 First Steps
5.1 Switching on / off
Switching on
– Press [On/Off].
The display appears.
OnOff
Switching off
– Press [On/Off] until "Off" appears in the display.
NoteDo not disconnect the balance from the power supply except if you willnot be using the balance for an extended period.
OnOff
5.2 Leveling the balanceCheck the position of the air bubble in the level indicator on the top of the weighing cell. If the air bubble is notin the inner circle, the weighing cell needs to be leveled.
Turn the two leveling screws on the back of the weighing cell until theair bubble is in the inner circle of the level indicator (left figure = leveled correctly, right figure = leveled incorrectly).
5.3 Leveling aid of the XP modelsThe XP models have a built-in leveling aid.
On the terminal, tap the large empty space below the weighing result.
1 To start the Leveling Assistent, tap [Show].
The Leveling Assistant will guide you step by step through theleveling process.
2 Watch the levelcontrol located of your balance and press theappropriate button of the current position.
The Leveling Assistant will show you with red arrows in whichdirection you need to turn the two footscrews on the back of theweighing cell.
3 Screw the footscrew until the air bubble is in the inner circle.
4 Tap [Exit].
A message appears that advises you to adjust the balance.
5 Confirm with [OK].
Status icon will no longer appear and balance returns to normaloperation.
19First Steps
20 First Steps
6 Maintenance
6.1 CleaningPeriodically clean the weighing chamber, the housing, and the terminal of your balance using the brush supplied with it. The maintenance interval depends on your standard operating procedure (SOP).
Please observe the following notes
WARNING
Risk of electric shock
- The balance must be disconnected from the power supply.
- Ensure that no liquid comes into contact with the balance, the terminal or the AC adapter.
- Never open the balance, terminal or AC adapter – they contain no components, which canbe cleaned, repaired or replaced by the user.
CAUTION
Damage of balance
On no account use cleaning agents which contain solvents or abrasive ingredients, as thiscan result in damage to the terminal overlay.
Cleaning
Your balance is made from high quality, resistant materials and can therefore be cleaned with a commerciallyavailable, mild cleaning agent.
1 To clean the weighing chamber thoroughly, pull thedraft disk (which on the XP2U and XP6U balances ismade up of several parts) and the weighing pan vertically up and off.It may be necessary to turn the weighing pan slightly toremove it.
2 When you replace these parts, make sure they are in thecorrect position.
open
close
NotePlease contact your METTLER TOLEDO dealer for details of the available service options. Regular servicing byan authorized service engineer ensures constant accuracy for years to come and prolongs the service life ofyour balance.
21Maintenance
22 Maintenance
6.2 Disposal
In conformance with the European Directive 2002/96/EC on Waste Electrical and ElectronicEquipment (WEEE) this device may not be disposed of in domestic waste. This also appliesto countries outside the EU, per their specific requirements.
Please dispose of this product in accordance with local regulations at the collecting pointspecified for electrical and electronic equipment. If you have any questions, please contactthe responsible authority or the distributor from which you purchased this device. Should thisdevice be passed on to other parties (for private or professional use), the content of this regulation must also be related.
Thank you for your contribution to environmental protection.
7 Technical Data
7.1 General data
CAUTION
Use only with a tested AC Adapter with SELV output current.Ensure correct polarity
Power supplyPower supply connector with AC/DCadapter:
11107909Primary: 100-240 VAC, -15%/+10%, 50/60 HzSecondary: 12 VDC ±3%, 2.0 A (with electronic overload protection)
Cable to AC adapter: Design: 3-core, with country-specific plugNoteMake sure the power supply plug is freely accessible
Power supply to the balance: 12 VDC ±3%, 2.0 A, maximum ripple: 80 mVpp
Protection and standardsOvervoltage category: Class IIDegree of pollution: 2Standards for safety and EMC: See Declaration of ConformityRange of application: For use only in closed interior rooms
Environmental conditionsHeight above mean sea level: Up to 4000 mAmbient temperature: 5-40 °CRelative air humidity: Max. 80% at 31 °C, linearly decreasing to 50% at 40 °C, non
condensingWarm-up time: 24 hours after connecting the balance to the power supply; when
switched on from standby-mode, the balance is ready for operation immediately.
MaterialsHousing: Die-cast aluminum, plastic, chrome steel and glassTerminal: Die-cast zinc, chromed and plasticsWeighing pan: Aluminum, chromed (AlMgSi1 coated chem Ni 15 µm,
Cr 0.3 – 0.5 µm)
7.2 Explanatory notes for the METTLER TOLEDO AC adapterThe certified external power supply which conforms to the requirements for Class II double insulated equipmentis not provided with a protective earth connection but with a functional earth connection for EMC purposes. Thisearth connection IS NOT a safety feature. Further information about conformance of our products can be foundin the brochure "Declaration of Conformity" which is coming with each product.
In case of testing with regard to the European Directive 2001/95/EC the power supply and the balance have tobe handled as Class II double insulated equipment.
Consequently an earth bonding test is not required. Similarly it is not necessary to carry out an earth bondingtest between the supply earth conductor and any exposed metalwork on the balance.
23Technical Data
24 Technical Data
Because the balance are sensitive to static charges a leakage resistor, typically 10 kΩ, is connected betweenthe earth connector and the power supply output terminals. The arrangement is shown in the equivalent circuitdiagram. This resistor is not part of the electrical safety arrangement and does not require testing at regularintervals.
10 kΩ coupling resistor for
electrostatic discharge
Input 100…240 VAC Output 12 VDC
Double Insulation
Plastic Housing
P
N
E
AC
DC
Equivalent circuit diagram
7.3 Model-specific dataXP2U XP6U
Limit valuesMaximum capacity 2.1 g 6.1 gReadability 0.0001 mg 0.0001 mg Tare range (from…to) 0 … 2.1 g 0 … 6.1 g Repeatability (at nominal load) 1) sd 0.00025 mg (2 g) 0.0004 mg (5 g) Repeatability (at low load) 1) sd 0.0002 mg (0.2 g) 0.00025 mg (0.2 g) Linearity deviation 0.0015 mg 0.004 mg Eccentricity deviation (test load) 2) 0.0025 mg (1 g) 0.002 mg (2 g) Sensitivity offset (test weight) 3) 0.03 mg (2 g) 0.048 mg (6 g) Sensitivity temperature drift 4) 0.0001%/°C 0.0001%/°C Sensitivity stability 5) 0.0001%/a 0.0001%/a Typical valuesRepeatability 1) sd 0.00015 mg –Linearity deviation 0.0008 mg 0.0019 mg Eccentric deviation (test load) 0.0016 mg (1 g) 0.0012 mg (2 g) Sensitivity offset (test load) 0.012 mg (2 g) 0.018 mg (6 g) Minimum sample weight (according to USP) 1) 0.45 mg –Minimum sample weight (U=1%, k=2) 1) 0.03 mg –Settling time 10 s 15 sDimensionsBalance dimensions (WxDxH) 128x287x113 mm 128x287x113 mm Weighing pan dimensions ø 16 mm ø 16 mmTypical uncertainties and supplementary dataRepeatability 1) sd 0.00015mg+0.0000025%·Rgr –Differential linearity deviation sd √(0.08pg·Rnt) √(0.15pg·Rnt) Differential eccentric load deviation sd 0.00008%·Rnt 0.00003%·Rnt Sensitivity offset 3) sd 0.0003%·Rnt 0.00015%·Rnt Minimum sample weight (according to USP) 1) 0.45mg+0.0075%·Rgr –Minimum sample weight (U=1%, k=2) 1) 0.03mg+0.0005%·Rgr –Interface update rate 23 /s 23 /s Usable height of draft shield 55 mm 55 mm Weight of balance 7.5 kg 7.5 kgNumber of built-in reference weights 2 2Weights for routine testing
XP2U XP6UOIML CarePac
Weights
2 g E2, 0.1 g E2
#11123004
5 g E2, 0.2 g E2
#11123005
ASTM CarePac
Weights
2 g 1, 0.1 g 1
#11123104
5 g 1, 0.2 g 1
#11123105
sd = Standard deviation Rnt = Net weight (sample weight)
Rgr = Gross weight a = Year (annum)
1) Valid for compact objects 2) According to OIML R763) After adjustment with built-in reference weight 4) In the temperature range 10...30 °C5) After putting into operation for the first time, with the self-
adjustment function activated (ProFACT or FACT)
XP6 XS3DULimit valuesMaximum capacity 6.1 g 3.1 g Readability 0.001 mg 0.01 mg Tare range (from…to) 0 … 6.1 g 0 … 3.1 g Maximum capacity, fine range – 0.8 g Readability, fine range – 0.001 mg Repeatability (at nominal load) 1) sd 0.0008 mg (5 g) 0.006 mg (3 g) Repeatability (at low load) 1) sd 0.0006 mg (0.2 g) 0.005 mg (0.2 g)Repeatability, fine range (at low load) 1) sd – 0.0008 mg (0.2 g)Linearity deviation 0.004 mg 0.01 mg Eccentricity deviation (test load) 2) 0.003 mg (2 g) 0.004 mg (1 g) Sensitivity offset (test weight) 3) 0.048 mg (6 g) 0.045 mg (3 g) Sensitivity temperature drift 4) 0.0001%/°C 0.0001%/°CSensitivity stability 5) 0.0001%/a 0.0001%/a Typical valuesRepeatability 1) sd 0.0004 mg 0.003 mgRepeatability, fine range 1) sd – 0.0005 mgLinearity deviation 0.0019 mg 0.0038 mgEccentric deviation (test load) 0.002 mg (2 g) 0.0024 mg (1 g)Sensitivity offset (test load) 0.018 mg (6 g) 0.018 mg (3 g)Minimum sample weight (according to USP) 1) 1.2 mg 9 mg Minimum sample weight (according to USP),fine range 1)
– 1.5 mg
Minimum sample weight (U=1%, k=2) 1) 0.08 mg 0.6 mg Minimum sample weight (U=1%, k=2), finerange 1)
– 0.1 mg
Settling time 7 s 6 sSettling time, fine range – 10 s DimensionsBalance dimensions (WxDxH) 128x287x113 mm 128x287x113 mmWeighing pan dimensions ø 27 mm ø 27 mmTypical uncertainties and supplementary dataRepeatability 1) sd 0.0004mg+0.000003%·Rgr 0.003mg+0.00006%·RgrRepeatability, fine range 1) sd – 0.0005mg+0.000012%·RgrDifferential linearity deviation sd √(0.15pg·Rnt) √(1.2pg·Rnt)Differential eccentric load deviation sd 0.00005%·Rnt 0.00012%·RntSensitivity offset 3) sd 0.00015%·Rnt 0.0003%·RntMinimum sample weight (according to USP) 1) 1.2mg+0.009%·Rgr 9mg+0.18%·RgrMinimum sample weight (according to USP),fine range 1)
– 1.5mg+0.036%·Rgr
Minimum sample weight (U=1%, k=2) 1) 0.08mg+0.0006%·Rgr 0.6mg+0.012%·RgrMinimum sample weight (U=1%, k=2), finerange 1)
– 0.1mg+0.0024%·Rgr
Interface update rate 23 /s 23 /s Usable height of draft shield 55 mm 55 mm
25Technical Data
26 Technical Data
XP6 XS3DUWeight of balance 7.5 kg 7 kgNumber of built-in reference weights 2 2Weights for routine testingOIML CarePac
Weights
5 g E2, 0.2 g E2
#11123005
2 g E2, 0.1 g E2
#11123004
ASTM CarePac
Weights
5 g 1, 0.2 g 1
#11123105
2 g 1, 0.1 g 1
#11123104
sd = Standard deviation Rnt = Net weight (sample weight)
Rgr = Gross weight a = Year (annum)
1) Valid for compact objects 2) According to OIML R763) After adjustment with built-in reference weight 4) In the temperature range 10...30 °C5) After putting into operation for the first time, with the self-
adjustment function activated (ProFACT or FACT)
7.4 Dimensions
7.4.1 Dimensions of the terminal and control unit of XP models
Dimensions in mm.
74
80
143129 10
195
176
195
58
6
163146
309 4
Terminal and Control Unit of XP Models
27Technical Data
28 Technical Data
7.4.2 Dimensions of the terminal and control unit of XS models
Dimensions in millimeters.
74
80
143104.5 10
195
176
178
56
6
163121.5
284.5 4
Terminal and Control Unit of XS Models
7.4.3 Dimensions of the weighing cell (XS and XP models)
Dimensions in millimeters.
260
174 26
287.5
53
42
113.5
35.5
80
114
104°
22067.5
100
116
128
Weighing Cell XP and XS Models
29Technical Data
30 Technical Data
7.5 Interfaces
7.5.1 Specifications of RS232CInterface type: Voltage interface according to EIA RS-232C/DIN 66020 (CCITT V24/V.28)Max. cable length: 15 mSignal level: Outputs:
+5 V ... +15 V (RL = 3 – 7 kΩ)
–5 V ... –15 V (RL = 3 – 7 kΩ)
Inputs:
+3 V ... 25 V
–3 V ... 25 V
Connector: Sub-D, 9-pole, femaleOperating mode: Full duplexTransmission mode: Bit-serial, asynchronousTransmission code: ASCIIBaud rates: 600, 1200, 2400, 4800, 9600, 19200, 384001) (firmware selectable)Bits/parity: 7-bit/even, 7-bit/odd, 7-bit/none, 8-bit/none (firmware selectable)Stop bits: 1 stop bitHandshake: None, XON/XOFF, RTS/CTS (firmware selectable)End-of-line: <CR><LF>, <CR>, <LF> (firmware selectable)
6
15
9
DataGND
Handshake
Pin 2: Balance transmit line (TxD)
Pin 3: Balance receive line (RxD)
Pin 5: Ground signal (GND)
Pin 7: Clear to send (hardware handshake) (CTS)
Pin 8: Request to send (hardware handshake) (RTS)
1) 38400 baud is only possible in special cases, such as:• Weighing platform without terminal, or• Weighing platform with terminal, only via the optional RS232C interface.
7.5.2 Specifications of "Aux" connection
You can connect the METTLER TOLEDO "ErgoSens" or an external switch to sockets "Aux 1" and "Aux 2". Thisallows you to start functions such as taring, zeroing, printing and others.
External connection Connector: 3.5 mm stereo jack connectorElectrical data: Max. voltage 12 V
Max. current 150 mA
∅ 3
.5 m
m
GND
Do not connect!
Connection contact
8 Accessories and Spare Parts
8.1 AccessoriesYou can increase the functionality of your balance with accessories from the METTLER TOLEDO range. The following options are available:
Description Part No.
Printers
BT-P42 printer with Bluetooth connection to instrument 11132540Paper roll, set of 5 pcs 00072456Paper roll, self-adhesive, set of 3 pcs 11600388
Printer
17.375 g
19.319 g
8.003 g
7.773 g
6.554 g
10.506 g
8.097 g
5.876 g
3.205 g
1.098 g
METTLER TOLEDO
Ribbon cartridge, black, set of 2 pcs 00065975
RS-P42 printer with RS232C connection to instrument 00229265Paper roll, set of 5 pcs 00072456Paper roll, self-adhesive, set of 3 pcs 11600388
Printer
17.375 g
19.319 g
8.003 g
7.773 g
6.554 g
10.506 g
8.097 g
5.876 g
3.205 g
1.098 g
METTLER TOLEDO
Ribbon cartridge, black, set of 2 pcs 00065975
RS-P25 printer with RS232C connection to instrument 11124300Paper roll, set of 5 pcs 00072456Paper roll, self-adhesive, set of 3 pcs 11600388Ribbon cartridge, black, set of 2 pcs 00065975
RS-P26 printer with RS232C connection to instrument (withdate and time)
11124303
Paper roll, set of 5 pcs 00072456Paper roll, self-adhesive, set of 3 pcs 11600388Ribbon cartridge, black, set of 2 pcs 00065975
LC-P45 application printer with additional functions 00229119Paper roll, set of 5 pcs 00072456Paper roll, self-adhesive, set of 3 pcs 11600388Menu
Def
Menu
Clear
ID
Code
Result
F
3
2
1
4
5
6
7
8
9
0
LC-P45 Printer
31
1 2 3
METTLER TOLEDO
Ribbon cartridge, black, set of 2 pcs 00065975
Optional interfaces
Second RS232C Interface 11132500
31Accessories and Spare Parts
32 Accessories and Spare Parts
Ethernet Interface for connection to an Ethernet network 11132515
BT option: Bluetooth Interface for multipoint connection for upto 6 Bluetooth devices
11132530
BTS option: Bluetooth Interface, single-point connection 11132535
PS/2 option: Interface for connecting commercial keyboardsand barcode readers
11132520
LocalCAN option: Interface for connection of up to 5 LC(LocalCAN) instruments
11132505
MiniMettler option: Interface MiniMettler, for connection to older(legacy) METTLER TOLEDO systems
11132510
RS232 - USB converter cable – Cable with converter to connecta balance (RS232) to a USB port
64088427
Cables for RS232C interface
RS9 – RS9 (m/f): connection cable for PC, length = 1 m 11101051
RS9 – RS25 (m/f): connection cable for PC, length = 2 m 11101052
Cables for LocalCAN interface
LC – RS9: Cable for connecting a PC with RS232C, 9-pin (f),lenght = 2 m
00229065
LC – RS25: Cable for connecting a printer or PC with RS232C,25-pin (m/f), lenght = 2 m
00229050
LC – CL: Cable for connecting a device with METTLER TOLEDOCL interface (5-pin), length = 2 m
00229130
LC – LC2: Extension cable for LocalCAN, length = 2 m 00229115
LC – LC5: Extension cable for LocalCAN, length = 5 m 00229116
LC – LCT: Cable branch (T-connector) for LocalCAN 00229118
Cables for MiniMettler interface
MM – RS9f: RS232C connection cable to MiniMettler interface,length = 1.5 m
00229029
33Accessories and Spare Parts
34 Accessories and Spare Parts
Cables for terminal
Terminal extension cable, length = 4.5 m 11600517
Auxiliary displays
LC/RS-BLD auxiliary display on bench stand, backlit (incl. RScable and separate AC adapter)
00224200
Sensors
ErgoSens, optical sensor for hands-free operation 11132601
LC-Switchbox
For connection of up to 3 balances with LocalCAN interface toa printer
00229220
Footswitches
Footswitch with selectable function for balances (Aux 1, Aux 2) 11106741
LC-FS foot switch with selectable function for balances withLocalCAN interface
00229060
Universal AntiStatic kit
Universal AntiStatic Kit complete (U-shaped), including electrode and power supplyOptional: Second U-electrode* for universal AntiStatic Kit* Power supply for optional, second U-elektrode (11107764)
111077671110776411107766
Filter kit
Filter kit for XP/XS/MX/UMX balances ø 110 mm 00211227
Filter kit for XP/XS/MX/UMX balances ø 47 mm and ø 70 mm 11122136
Funnel kit
Funnel kit for XP/XS/UMX/MX balances 00211220
Barcode reader
RS232C Barcode Reader 21901297The following accessories are needed for operation (not included):
RS232 F cable 21901305Null modem adapter 21900924
Plus one of the following: AC adapter 5 V for EU 21901370AC adapter 5 V for US 21901372AC adapter 5 V for GB 21901371AC adapter 5 V for AU 21901370
+ 71209966
RS232C Barcode Reader – Cordless 21901299The following accessories are needed for operation (not included):
Cradle 21901300RS232 F cable 21901305Null modem adapter 21900924
Plus one of the following: AC adapter 12 V for EU 21901373AC adapter 12 V for US 21901375AC adapter 12 V for GB 21901374AC adapter 12 V for AU 21901373
+ 71209966
PS/2 Barcode Reader, without cable 21901297PS/2 wedge single cable 21901307
35Accessories and Spare Parts
36 Accessories and Spare Parts
PS/2Y Barcode Reader, without cable 21901297PS/2 wedge twin (Y) cable 21901308
Transport cases
Transport case for micro balances 11122760
Protective covers
Protective cover for XS terminal 11106870
Protective cover for XP terminal 11132570
Dust covers
Dust cover 30038799
Software
LabX software for one click™ weighing solutionsEnables you to perform One Click™ Standard Preparation, OneClick™ Loss on Drying, One Click™ Sieve Analysis and manyother applications.Simply start the method with the One Click™ shortcut on thebalance touchscreen. LabX guides you step-by-step throughthe SOP on the balance, performs your calculations automatically, and takes care of saving all your data. The completesolution can be tailored to match your process requirements.Visit www.mt.com/one-click-weighing for more information
on request
Freeweigh.
Net
Freeweigh.Net 21900895
Various
Wall fixture for terminal 11132665
AC/DC adapter (without power cable) 100–240 VAC,50/60 Hz, 0.3 A, 12 VDC 2.25 A
11107909
Country-specific 3-Pin power cable with grounding conductor.Power cable AU 00088751Power cable BR 30015268Power cable CH 00087920Power cable CN 30047293Power cable DK 00087452Power cable EU 00087925Power cable GB 00089405Power cable IL 00225297Power cable IN 11600569Power cable IT 00087457Power cable JP 11107881Power cable TH, PE 11107880Power cable US 00088668Power cable ZA 00089728
IP54 protective housing for AC adapter 11132550
Weighing pan ø 15.7 mm, chrome-nickel steel X5CrNi 18-10 11100437
Weighing table 11138044
37Accessories and Spare Parts
38 Accessories and Spare Parts
8.2 Spare partsWeighing chamber
Pos Description Part No.Glass cover for draft shield1 Glass cover for XP6 and XS3DU 002110821 Glass cover for XP2U and XP6U 00211177Weighing pan2 Weighing pan for XP6 and XS3DU 002110553 Weighing pan for XP2U and XP6U 002111974 Hook weighing pan for XP2U and
XP6U00211295
5 Draft disk complete 111000756 Ring nut 111003417 Weighing chamber plate 00211155Weighing chamber complete8 Weighing chamber for XP6 and
XS3DU11100861
8 Weighing chamber for XP2U andXP6U
11100862
1
2
5
6
7
8
34
9
9 Sealing cover 00211122
Weighing cellPos Description Part No.1 Cover 11122623
2
2
1
1
2 Footscrew 11122612
Control unitPos Description Part No.
1
1 Drawer 00211163
Terminal Type "P" (color, for XP balances)Pos Description Part No.1 Terminal Type "P" 111306922 Protective cover for Terminal Type "P" 111325703 Terminal support for Terminal Type
"P"11122950
1
3
24
4 Terminal cable 11122830
Terminal Type "S" (monochrome, for XS balances)Pos Description Part No.1 Terminal Type "S" 111078992 Protective cover for Terminal Type "S" 111068703 Terminal support for Terminal Type
"S"11122951
1
3
24 4 Terminal cable 11122830
39Accessories and Spare Parts
40 Accessories and Spare Parts
Small partsPos Description Part No.1 Cleaning brush 000701142 Cleaning tweezers 00211124
1 2 3
3 Weighing tweezers 00070661
TransportPos Description Part No.
Packaging complete 30019748
Export box 11122751
9 Appendix
9.1 MT-SICS interface commands and functionsMany of the instruments and balances used have to be capable of integration in a complex computer or dataacquisition system.
To enable you to integrate balances in your system in a simple manner and utilize their capabilities to the full,most balance functions are also available as appropriate commands via the data interface.
All new METTLER TOLEDO balances launched on the market support the standardized command set "METTLERTOLEDO Standard Interface Command Set" (MT-SICS). The commands available depending on the functionalityof the balance.
For further information please refer to the Reference Manual MT-SICS downloadable from the Internet under
u http://www.mt.com/micro
9.2 Procedure for certified balances
PrefaceCertified balances are subject to the national, legal requirements of "non-automatic balances".
Switching on the balance Switching on
• Immediately after being switched on, the balance displays 0.000.. g.• The balance is always started up with the "Factory setting" unit.
Switch-on range• At maximum 20% of the type load, otherwise overload is displayed (OIML R76 4.5.1).
Stored value as switch-on zero point• It is not permissible to use a stored value as a switch-on zero point; the MT-SICS M35 command is not
available (OIML R76 T.5.2).
Display Display of the weight value
• The "e" certification value is always shown in the display and is specified at the model designation plate(OIML R76 T.3.2.3 and 7.1.4).
• If the display increment is lower than the "e" certification value, this is variably displayed for the net, grossand weighed tare. (Graying of the digits or certification brackets) (OIML R76 T.2.5.4 and 3.4.1).
In accordance with guidelines, the tested display increment (certification value) is never lower than 1 mg(OIML R76 T.3.4.2).
At balances with d = 0.1 mg, the digits below 1 mg are displayed in gray. These digits in brackets areprinted. In accordance with legal metrology requirements, this illustration does not affect the accuracy of theweighing results.
Units of measurement• The display and info unit are firmly set to g or mg (depending on the model).• The following applies for the "Custom unit":
– No certification brackets.– The following names are blocked, this applies to upper and lower case letters.
– All official units (g, kg, ct etc.).– c, ca, car, cm, crt, cart, kt, gr, gra, gram, grm, k, kilo, to, ton.– All names with "o" which can be replaced by a zero (0z, 0zt etc.).
41Appendix
42 Appendix
Identification of the weight display• Gross, net, tare and other weight values are accordingly marked (OIML R76 4.6.5).
– Net for net when a tare value has been used.– B or G for gross.– T for the weighed tare.– PT for the specified tare.– * or diff for the difference between the net or gross.
Info field• The info weight value is handled metrologically in the same way as the weight value in the main display.
Printout (OIML R76 4.6.11) If a tare value is entered manually (PreTare), the PreTare value is always printed along with the net value
(PT 123.45 g).
The printed weight values are identified in the same way as the weight value on the display. I.e. N, B or G, T, PT, diff or *, with differentiation.Example:Single-range balance.
N 123.4[5] gPT 10.00 g for PreTareG 133.4[5] g
DR balance with 100.00 g fine range.
N 80.4[0] gT 22.5[6] g for weighed tareG 102.9[ ] g
Balance functions Reset to zero
• The zero range is limited to a maximum of ± 2% of the full load (OIML R76 4.5.1).
Tare• No negative tare values are permitted.• Tare immediate (TI) is not permitted, the MT-SICS TI command is not available (OIML R76 4.6.4).
1/xd• e = d
The 1/xd switchover is not permitted (OIML R76 3.1.2).• e = 10d
This is only permitted in the case of the 1/10d switchover.• e = 100d
Only the 1/10d and 1/100d switchover are permitted.
10 Index
AAC adapter 23, 23Accessories 31Assembling the balance 13Aux connections 30
BBalance functions 42Below-the-balance weighing 17
CCertified balances 41Changing the reading angle 15Cleaning 20Control unit 13, 39Conventions 6
DDimensions 27Display 41Display field remains dark 14Display of the weight value 41Disposal 22
EEnvironmental conditions 23ErgoSens 30
FFeatures 5
GGeneral safety information 7Glass draft shield 14GLP 5Good Laboratory Practice 5
HHook weighing pan 13
IIdentification of the weight display
42
Info field 42Interface
MT-SICS 41ISO 14001 5ISO 9001 5
LLeveling 18Leveling aid 18Leveling Assistent 19
MMaterials 23MT-SICS 41
PPower supply 14, 23Power supply voltages 14Printout 42Protection and standards 23
RRemove terminal 15Reset to zero 42RS232C interface 30
Ssafety information 7Safety information
General 7Staff safety 7
Scope of delivery 12Self test 14Setting up 11Shipping lock 12, 17Spare parts 38Staff safety 7Switching off 18Switching on 18
TTare 42Technical data 23Terminal 39Transport 40Transporting over long distances 16Transporting over short distances
16
Transporting the balance 16
UUnits of measurement 41
WWeighing cell 13, 38Weighing chamber 38
Index 43
Mettler-Toledo AG, Laboratory & Weighing TechnologiesCH-8606 Greifensee, SwitzerlandTel. +41 (0)44 944 22 11Fax +41 (0)44 944 30 60www.mt.com
Subject to technical changes.© Mettler-Toledo AG 03/201311781193C en
www.mt.com/excellenceFor more information
*11781193*