Sept 6, 2007 L. Torbeck 1
OOSGuidance
SOCMASeptember 6, 2007
![Page 2: Oos Guidance](https://reader033.fdocuments.us/reader033/viewer/2022061406/577cc6951a28aba7119e9d4c/html5/thumbnails/2.jpg)
Sept 6, 2007 L. Torbeck 2
Overview
• CGMP’s 21 211.192 • U.S.A. vs. Barr Laboratories, Inc.• Able Laboratories• Reportable Results• Specifications• OOS flowchart• Specific OOS topics
![Page 3: Oos Guidance](https://reader033.fdocuments.us/reader033/viewer/2022061406/577cc6951a28aba7119e9d4c/html5/thumbnails/3.jpg)
Sept 6, 2007 L. Torbeck 3
Specific Topics
• OOS definition• Initial and full investigations• Testing and retesting• Sampling and resampling• Averaging and Outliers
![Page 4: Oos Guidance](https://reader033.fdocuments.us/reader033/viewer/2022061406/577cc6951a28aba7119e9d4c/html5/thumbnails/4.jpg)
Sept 6, 2007 L. Torbeck 4
CGMP’s 21 CFR 211.192
• “Any unexplained discrepancy of the failure of a batch or any of its contents to meet any of its specifications shall be thoroughly investigated, whether or not the batch has already been distributed.”
![Page 5: Oos Guidance](https://reader033.fdocuments.us/reader033/viewer/2022061406/577cc6951a28aba7119e9d4c/html5/thumbnails/5.jpg)
Sept 6, 2007 L. Torbeck 5
CGMP’s 21 CFR 211.192
• “The investigation shall extend to other batches of the same drug product and other drug products that may have been associated with the specific failure or discrepancy.
• A written record of the investigation shall be made and shall include the conclusions and follow-up.”
![Page 6: Oos Guidance](https://reader033.fdocuments.us/reader033/viewer/2022061406/577cc6951a28aba7119e9d4c/html5/thumbnails/6.jpg)
Sept 6, 2007 L. Torbeck 6
Barr Court Case
• Civil action by FDA, June 12, 1992.• Judge Alfred M. Wolin• Can the FDA expand the CGMP’s
interpretation into the statistical areas of outliers, retesting, resampling, averaging and sample size and other areas of failure investigations, …”
![Page 7: Oos Guidance](https://reader033.fdocuments.us/reader033/viewer/2022061406/577cc6951a28aba7119e9d4c/html5/thumbnails/7.jpg)
Sept 6, 2007 L. Torbeck 7
Barr Background
• “The current conflict surrounding these rules is best characterized as a confrontation between a humorless warden and his uncooperative prisoner. … These witnesses revealed an industry mired in uncertainty and conflict, guided by vague regulations which produce tugs-of-war of varying intensity.”
![Page 8: Oos Guidance](https://reader033.fdocuments.us/reader033/viewer/2022061406/577cc6951a28aba7119e9d4c/html5/thumbnails/8.jpg)
Sept 6, 2007 L. Torbeck 8
Able Laboratories
• On May 23, 2005, Able labs issued a Class II recall of all of its 46 drug products
• The company immediately suspended operations and laid off 200 employees.
• The stock price in the last two weeks of May dropped by more than 90%.
• The company went out of business.
![Page 9: Oos Guidance](https://reader033.fdocuments.us/reader033/viewer/2022061406/577cc6951a28aba7119e9d4c/html5/thumbnails/9.jpg)
Sept 6, 2007 L. Torbeck 9
OOS Philosophy
• “Testing lies at the heart of a drug manufacturer’s successful operation. Through testing companies validate their processes and ensure the quality of batches for release.” Judge Wolin
• If we can’t trust our measurements, we don’t have anything.
![Page 10: Oos Guidance](https://reader033.fdocuments.us/reader033/viewer/2022061406/577cc6951a28aba7119e9d4c/html5/thumbnails/10.jpg)
Sept 6, 2007 L. Torbeck 10
Draft Version
• 30 September 1998• “Guidance for Industry – Investigating Out
of Specification (OOS) Test Results for Pharmaceutical Production”
• Submitted comments by 30 November 1998• These comments can be inspected by going
to the FDA offices.
![Page 11: Oos Guidance](https://reader033.fdocuments.us/reader033/viewer/2022061406/577cc6951a28aba7119e9d4c/html5/thumbnails/11.jpg)
Sept 6, 2007 L. Torbeck 11
Final Guidance
• 12 October 2006 in the Federal Register• FDA authors include:
– Richard Friedman – Director of the Division of Manufacturing & Product Quality
– Paul Haynie is responsible for the Guidance– (301) 827-9020– [email protected]
![Page 12: Oos Guidance](https://reader033.fdocuments.us/reader033/viewer/2022061406/577cc6951a28aba7119e9d4c/html5/thumbnails/12.jpg)
Sept 6, 2007 L. Torbeck 12
Controversial Topics
• Definition of reportable values?• Use of averaging?• Number of retests?• Second analyst?• Use of outlier testing?• What specification limits?• Defining testing into compliance?
![Page 13: Oos Guidance](https://reader033.fdocuments.us/reader033/viewer/2022061406/577cc6951a28aba7119e9d4c/html5/thumbnails/13.jpg)
Sept 6, 2007 L. Torbeck 13
OOS Definition
• “… the term OOS results includes all test results that fall outside the specification or acceptance criteria established in drug applications, drug master files, official compendia, or by the manufacturer.
• Two Major Issues:– What test results?– What specifications?
![Page 14: Oos Guidance](https://reader033.fdocuments.us/reader033/viewer/2022061406/577cc6951a28aba7119e9d4c/html5/thumbnails/14.jpg)
Sept 6, 2007 L. Torbeck 14
Scope
• “These laboratory tests are performed on active pharmaceutical ingredients, excipients and other components, in-process materials and finished drug products.
• “Laboratory testing … is necessary to confirm that components, container and closures, in-process materials and finished products conform to specifications, including stability specifications.
![Page 15: Oos Guidance](https://reader033.fdocuments.us/reader033/viewer/2022061406/577cc6951a28aba7119e9d4c/html5/thumbnails/15.jpg)
Sept 6, 2007 L. Torbeck 15
Scope
• This guidance applies to chemistry-based laboratory testing of drugs regulated by CDER.
• “The term also applies to all in-process laboratory tests ….”
• “… applies to in-house testing of drug product components that are purchased …”
![Page 16: Oos Guidance](https://reader033.fdocuments.us/reader033/viewer/2022061406/577cc6951a28aba7119e9d4c/html5/thumbnails/16.jpg)
Sept 6, 2007 L. Torbeck 16
Reportable Valuesfor Out-of-Specification
Test resultsLynn Torbeck
Pharmaceutical TechnologyVol. 23, No. 2, February 1999
Special Supplement
![Page 17: Oos Guidance](https://reader033.fdocuments.us/reader033/viewer/2022061406/577cc6951a28aba7119e9d4c/html5/thumbnails/17.jpg)
Sept 6, 2007 L. Torbeck 17
FDA Definition
• “It should be noted that a test might consist of replicates to arrive at a result. For instance, an HPLC assay result may be determined by averaging the peak responses from a number of consecutive, replicate injections from the same preparation.The assay result would be calculated using the peak response average.”
![Page 18: Oos Guidance](https://reader033.fdocuments.us/reader033/viewer/2022061406/577cc6951a28aba7119e9d4c/html5/thumbnails/18.jpg)
Sept 6, 2007 L. Torbeck 18
FDA Definition
•“This determination is considered one test and one result.”
![Page 19: Oos Guidance](https://reader033.fdocuments.us/reader033/viewer/2022061406/577cc6951a28aba7119e9d4c/html5/thumbnails/19.jpg)
Sept 6, 2007 L. Torbeck 19
Implied Definition
• A reportable value is the end result of the complete measurement method as documented.
• It is the value compared to the specifications.
• It is the value used for official reports.• It is the value used for statistical analysis.
![Page 20: Oos Guidance](https://reader033.fdocuments.us/reader033/viewer/2022061406/577cc6951a28aba7119e9d4c/html5/thumbnails/20.jpg)
Sept 6, 2007 L. Torbeck 20
Figure 1Batch
Sample
Preparation
Figure 1
ReportableValue, RV
Inj
![Page 21: Oos Guidance](https://reader033.fdocuments.us/reader033/viewer/2022061406/577cc6951a28aba7119e9d4c/html5/thumbnails/21.jpg)
Sept 6, 2007 L. Torbeck 21
Figure 2Batch
Sample
Inj1
Preparation 3Preparation 2Preparation 1
Inj 2 Inj 3
Figure 2
Inj 7 Inj 8 Inj 9
ReportableValue, RV
Inj 4 Inj 5 Inj 6
![Page 22: Oos Guidance](https://reader033.fdocuments.us/reader033/viewer/2022061406/577cc6951a28aba7119e9d4c/html5/thumbnails/22.jpg)
Sept 6, 2007 L. Torbeck 22
Figure 3Batch
Sample Resample
Reinjection
ReportableValue, RV
Inj1
Preparation 1C
Preparation 1B
Preparation 1A
Inj 2 Inj 3 Inj 4 Inj 5 Inj 6
Repreparation2C
Repreparation2B
Inj 7 Inj8 Inj 9
Retest
Remeasure Remeasure
RV RV
Figure 3
![Page 23: Oos Guidance](https://reader033.fdocuments.us/reader033/viewer/2022061406/577cc6951a28aba7119e9d4c/html5/thumbnails/23.jpg)
Sept 6, 2007 L. Torbeck 23
Interpretation
• The individual determinations do not have to meet the specification.
• Determinations are not reported out of the lab.
• The variability of determinations is like a system suitability issue.
• Set an upper limit on the standard deviation or %RSD.
![Page 24: Oos Guidance](https://reader033.fdocuments.us/reader033/viewer/2022061406/577cc6951a28aba7119e9d4c/html5/thumbnails/24.jpg)
Sept 6, 2007 L. Torbeck 24
Interpretation
• All reportable values must be documented• Do not average OOS with in spec to get an
in spec results to release with.• Do not average reportable values for QA to
make a decision. QA must see all R.V.• If after QA makes a decision, and a value is
needed for a COA, then average them.
![Page 25: Oos Guidance](https://reader033.fdocuments.us/reader033/viewer/2022061406/577cc6951a28aba7119e9d4c/html5/thumbnails/25.jpg)
Sept 6, 2007 L. Torbeck 25
Specifications
Types of Limits ?1.Regulatory specifications (External)2.Accept/reject limits (Includes stability)3.Action (Cpk=1.33 ?)4.Alert or Trend (Cpk=1.0 ?) 5.In-process adjustment limits ?
![Page 26: Oos Guidance](https://reader033.fdocuments.us/reader033/viewer/2022061406/577cc6951a28aba7119e9d4c/html5/thumbnails/26.jpg)
Sept 6, 2007 L. Torbeck 26
Phase I Investigations
• An investigation is required for an OOS.• The purpose is to find the cause of the OOS.• Is it measurement or manufacturing?• “Batch rejection does not negate the need to
perform the investigation.”• The first phase should include an assessment
of the accuracy of results.
![Page 27: Oos Guidance](https://reader033.fdocuments.us/reader033/viewer/2022061406/577cc6951a28aba7119e9d4c/html5/thumbnails/27.jpg)
Sept 6, 2007 L. Torbeck 27
Phase I Investigations
• “For contract laboratories, the laboratory should convey its data, findings, and supporting documentation to the manufacturing firm’s quality control unit who should then initiate the full scale investigation.?
![Page 28: Oos Guidance](https://reader033.fdocuments.us/reader033/viewer/2022061406/577cc6951a28aba7119e9d4c/html5/thumbnails/28.jpg)
Sept 6, 2007 L. Torbeck 28
Responsibility of the Analysts
• “The first responsibility for achieving accurate laboratory results lies with the analysts who is performing the test.”
• “The analysts should be aware of potential problems that could occur …”
![Page 29: Oos Guidance](https://reader033.fdocuments.us/reader033/viewer/2022061406/577cc6951a28aba7119e9d4c/html5/thumbnails/29.jpg)
Sept 6, 2007 L. Torbeck 29
Responsibility of the Analysts
• “Analysts should check the data for compliance with test specifications before discarding test preparations or standard preparations.”
• An assessment of the accuracy of the results should be started immediately.
![Page 30: Oos Guidance](https://reader033.fdocuments.us/reader033/viewer/2022061406/577cc6951a28aba7119e9d4c/html5/thumbnails/30.jpg)
Sept 6, 2007 L. Torbeck 30
Responsibility of the Supervisor
• Supervisors assessment should be:– Objective– Timely– No preconceived assumptions– Prompt assessment of data– Laboratory or manufacturing?
![Page 31: Oos Guidance](https://reader033.fdocuments.us/reader033/viewer/2022061406/577cc6951a28aba7119e9d4c/html5/thumbnails/31.jpg)
Sept 6, 2007 L. Torbeck 31
Supervisors Steps
1. Discuss the method with analysts2. Examine raw data3. Verify calculations4. Confirm performance of instruments.5. Confirm reference standards, reagents6. Evaluate performance of method7. Document, document, document
![Page 32: Oos Guidance](https://reader033.fdocuments.us/reader033/viewer/2022061406/577cc6951a28aba7119e9d4c/html5/thumbnails/32.jpg)
Sept 6, 2007 L. Torbeck 32
Phase II Investigations
• Unconfirmed OOS requires a full scale OOS investigation with predefined procedure.
• Find the root cause and do CAPA• Review production and sampling
procedures• Investigations given the highest priority
![Page 33: Oos Guidance](https://reader033.fdocuments.us/reader033/viewer/2022061406/577cc6951a28aba7119e9d4c/html5/thumbnails/33.jpg)
Sept 6, 2007 L. Torbeck 33
Review of Production
• Quality Control Unit conducts the investigation.
• Other departments participate:1. Manufacturing2. Process Development3. Maintenance4. Engineering
![Page 34: Oos Guidance](https://reader033.fdocuments.us/reader033/viewer/2022061406/577cc6951a28aba7119e9d4c/html5/thumbnails/34.jpg)
Sept 6, 2007 L. Torbeck 34
Review of Production
• “In cases where manufacturing occurs off-site (i.e., performed by a contract manufacturer or at multiple manufacturing sites), all sites potentially involved should be included in the investigation.”
• Review all documents and records of the manufacturing process.
![Page 35: Oos Guidance](https://reader033.fdocuments.us/reader033/viewer/2022061406/577cc6951a28aba7119e9d4c/html5/thumbnails/35.jpg)
Sept 6, 2007 L. Torbeck 35
Full Scale Investigation
• A written record of the review includes:1. A clear statement of the reason2. Summary of manufacturing process aspects
that could cause the problem3. Results of documentation review with
probable cause.4. Results of previous reviews5. Description of corrective actions to be taken.
![Page 36: Oos Guidance](https://reader033.fdocuments.us/reader033/viewer/2022061406/577cc6951a28aba7119e9d4c/html5/thumbnails/36.jpg)
Sept 6, 2007 L. Torbeck 36
Retesting
• A retest is another test of a portion of the original sample brought into the lab.
• FDA prefers a second analysts do the retest.• Don’t “test into compliance.”• Specify the number of retests.• Prepare a protocol before retesting.• If the error is found the retest substitutes.
![Page 37: Oos Guidance](https://reader033.fdocuments.us/reader033/viewer/2022061406/577cc6951a28aba7119e9d4c/html5/thumbnails/37.jpg)
Sept 6, 2007 L. Torbeck 37
How Big the Retest Sample?
• This is still and unresolved issue and the statisticians are still publishing journal articles and discussing it.
• Bar case n=7 not fully supported by statistics
• Could be too much or not enough• Currently n= 3 to n=9
![Page 38: Oos Guidance](https://reader033.fdocuments.us/reader033/viewer/2022061406/577cc6951a28aba7119e9d4c/html5/thumbnails/38.jpg)
Sept 6, 2007 L. Torbeck 38
Testing Into Compliance
• Testing into compliance is the practice of ignoring valid information that should be used to make decisions.
• Such a practice is at best not scientific and at worst is fraudulent, illegal, and immoral.
• Such practices must be found and stopped.• See Torbeck, Pharm Tech, Oct 2002
![Page 39: Oos Guidance](https://reader033.fdocuments.us/reader033/viewer/2022061406/577cc6951a28aba7119e9d4c/html5/thumbnails/39.jpg)
Sept 6, 2007 L. Torbeck 39
Testing Into Compliance
• Averaging OOS with in specification results to get an in specification result.
• Physically averaging powers, granulations and liquids to get in specifications results.
• Not recording data until is known to be in specification.
• Missing samples
![Page 40: Oos Guidance](https://reader033.fdocuments.us/reader033/viewer/2022061406/577cc6951a28aba7119e9d4c/html5/thumbnails/40.jpg)
Sept 6, 2007 L. Torbeck 40
Not Testing Into Compliance
• Large initial sample sizes are acceptable if all data generated is reported.
• Large number of retests are acceptable if all data generated is reported.
• Failing system suitability is not an OOS• Out of limits for an in-process adjustment is
not an OOS
![Page 41: Oos Guidance](https://reader033.fdocuments.us/reader033/viewer/2022061406/577cc6951a28aba7119e9d4c/html5/thumbnails/41.jpg)
Sept 6, 2007 L. Torbeck 41
Resampling
• “Resampling involves analyzing a specimen from any additional units collected as part of the original sampling procedure or from a new sample collected from the batch.
• Original should be large enough for retests• Original sample must representative• Resampling should be a rare event.
![Page 42: Oos Guidance](https://reader033.fdocuments.us/reader033/viewer/2022061406/577cc6951a28aba7119e9d4c/html5/thumbnails/42.jpg)
Sept 6, 2007 L. Torbeck 42
Averaging
• We can average determinations to get the reportable value.
• We do not average reportable values.• QA must see all of the reportable values.• Don’t average OOS with in spec results.
![Page 43: Oos Guidance](https://reader033.fdocuments.us/reader033/viewer/2022061406/577cc6951a28aba7119e9d4c/html5/thumbnails/43.jpg)
Sept 6, 2007 L. Torbeck 43
Outliers
• "In a sample of n observations it is possible for a limited number to be so far separated in value from the remainder that they give rise to the question whether they are not from a different population, or that the sampling technique is at fault. Such values are called outliers. Tests are available to ascertain whether they can be accepted as homogeneous with the rest of the sample." Marriott, 1990.
![Page 44: Oos Guidance](https://reader033.fdocuments.us/reader033/viewer/2022061406/577cc6951a28aba7119e9d4c/html5/thumbnails/44.jpg)
Sept 6, 2007 L. Torbeck 44
Outliers
• "The USP expressly allows firms to apply this test to biological and antibiotic assays, ..., but is silent on its use with chemical tests."
• "In the Court's view the silence of the USP with respect to chemical testing and outliers is prohibitory."
![Page 45: Oos Guidance](https://reader033.fdocuments.us/reader033/viewer/2022061406/577cc6951a28aba7119e9d4c/html5/thumbnails/45.jpg)
Sept 6, 2007 L. Torbeck 45
Outlier Tests
• Use of outlier tests is determined in advance• Specify the test and the sample size.• Can reject biological data but not chemical.
– The outlier can be omitted.• Outlier tests to be used sparingly.• No application variability or homogeneity is
being assessed.
![Page 46: Oos Guidance](https://reader033.fdocuments.us/reader033/viewer/2022061406/577cc6951a28aba7119e9d4c/html5/thumbnails/46.jpg)
Sept 6, 2007 L. Torbeck 46
Concluding the Investigation
• If a cause is found, invalidate the initial result and use the retest value(s) in its place.
• If the OOS is confirmed the batch is rejected.
• If the OOS is inconclusive and the retests are within specification, then QA may be able to justify releasing the batch.
![Page 47: Oos Guidance](https://reader033.fdocuments.us/reader033/viewer/2022061406/577cc6951a28aba7119e9d4c/html5/thumbnails/47.jpg)
Sept 6, 2007 L. Torbeck 47
Field Alerts
• “For those product that are the subject of approved full and abbreviated new drug applications, regulation require submitting within 3 working days a field alert report of information concerning any failure of a distributed batch to meet any of the specifications ….”
![Page 48: Oos Guidance](https://reader033.fdocuments.us/reader033/viewer/2022061406/577cc6951a28aba7119e9d4c/html5/thumbnails/48.jpg)
Sept 6, 2007 L. Torbeck 48
Conclusion