Ocular Ultrasound for raised ICP

18
Optic Nerve Sheath Diameter (ONSD)Measurement Dr Shweta Gidwani MBBS FCEM

description

How to perform bedside ocular ultrasound for raised ICP

Transcript of Ocular Ultrasound for raised ICP

Page 1: Ocular Ultrasound for raised ICP

Optic Nerve Sheath Diameter

(ONSD)MeasurementDr Shweta Gidwani MBBS FCEM

Page 2: Ocular Ultrasound for raised ICP

•The purpose of ultrasound measurement of ONSD in the ED is to provide a safe, effective, accurate and reproducible bedside test for measuring intracranial pressure (ICP).

•This is a demonstration of the technique used to measure ONSD in the ED.

Page 3: Ocular Ultrasound for raised ICP
Page 4: Ocular Ultrasound for raised ICP

Anatomy

The optic nerve is part of the central nervous system (CNS). It is surrounded by a dural sheath within which is contained subarachnoid space with cerebrospinal fluid.

Page 5: Ocular Ultrasound for raised ICP

The anterior part of the optic nerve, is loosely attached to the dural sheath and is only surrounded by orbital fat, thus making it distensible.

• Due to the direct communication with the brain, the intraorbital subarachnoid space around the optic nerve is subject to the same pressure changes as the intracranial compartment.

• This means it can inflate in response to raised ICP.

Page 6: Ocular Ultrasound for raised ICP

Patient Position

Page 7: Ocular Ultrasound for raised ICP

Probe •To do this procedure you need a

High Frequency Linear probe.

• Most EDs commonly have this probe as it is used for vascular and musculoskeletal ultrasound in the ED)

Page 8: Ocular Ultrasound for raised ICP
Page 9: Ocular Ultrasound for raised ICP

•A thick layer of ultrasound gel is placed on the closed upper eyelid of the patient.

•Some studies demonstrate the use of a barrier such as a clear dressing (eg Tegaderm) or an eyepad, to make the exam more comfortable for the patient.

Page 10: Ocular Ultrasound for raised ICP
Page 11: Ocular Ultrasound for raised ICP

The probe should then be gently placed over the eyelid paying careful attention not to exert too much pressure.

Page 12: Ocular Ultrasound for raised ICP

•The probe is placed in two planes:

•The horizonal or CROSS-SECTIONAL axis and then rotated 90 degrees to lie in along the VISUAL axis

Page 13: Ocular Ultrasound for raised ICP

• In each plane, the probe is adjusted from the temporal to the nasal end to find a suitable angle for displaying the entry of the optic nerve into the globe.

Page 14: Ocular Ultrasound for raised ICP

The optic nerve must then be placed in the centre of the image before making the measurements.

The ONSD is measured 3mm behind the globe by convention.

Page 15: Ocular Ultrasound for raised ICP

Ref: Tayal VS, Neulander M, Norton HJ, et al. Emergency depart- ment sonographic measurement of optic nerve sheath diameter to detect findings of increased intracranial pressure in adult head injury patients. Ann Emerg Med 2007;49: 508—14.

Page 16: Ocular Ultrasound for raised ICP

MeasurementThe probe is placed in two planes: The horizonal

or CROSS-SECTIONAL axis and then rotated 90 degrees to lie in along the VISUAL axis

2 measurements in 2 planes (axial and visual are made for each eye and the average is recorded as the final measurement.

Page 17: Ocular Ultrasound for raised ICP
Page 18: Ocular Ultrasound for raised ICP