NSLS-II Beam Loss Monitor System

35
1 BROOKHAVEN SCIENCE ASSOCIATES NSLS-II Beam Loss Monitor System S.L. Kramer ERL 2011 October 16-21, 2011

Transcript of NSLS-II Beam Loss Monitor System

Page 1: NSLS-II Beam Loss Monitor System

1 BROOKHAVEN SCIENCE ASSOCIATES

NSLS-II Beam Loss Monitor System

S.L. Kramer ERL 2011

October 16-21, 2011

Page 2: NSLS-II Beam Loss Monitor System

2 BROOKHAVEN SCIENCE ASSOCIATES

Outline of Talk

• Beam Loss Control in Storage Rings or ERLs • Beam Loss Monitoring in Storage Rings or ERLs • Comparison of Cerenkov Beam Loss Monitors (CBLM) • Testing of prototype CBLM for NSLS-II • Design of CBLM for NSLS-II LCM System

Page 3: NSLS-II Beam Loss Monitor System

3 BROOKHAVEN SCIENCE ASSOCIATES

LCM Requirements • Measure beam charge losses (energy) between accelerators

to minimize total beam power lost: dipole current (Beam energy) and DCCT/ICT (Beam current / charge)

• SR shielding design limits were: <13nC/min at any point in injection region, (septum <3nC/sec) < 1.1nC/min any other point of the SR, • Need to control beam loss locations in SR and • Need to verify beam losses in injection region, insuring losses

below limit in rest of the ring (unaccounted beam)

Page 4: NSLS-II Beam Loss Monitor System

4 BROOKHAVEN SCIENCE ASSOCIATES

NSLS-II SR Loss Control Scrapers

• Blade (thin <1 Xrad) type scrapers inserted into vacuum chamber in injection region degrade beam energy

• Subsequent dipole separates and dumps intercepted beam • Minimize radiation off scrapers toward experimental floor • Blades must be cooled to take full synchrotron radiation power • Beam impedance of scraper design needs to be minimized • Track penetration electrons and radiation from shower • Model beam loss mechanisms, calculate fraction intercepted

versus aperture (operational studies will be done to determine aperture that captures largest fraction without significantly reducing lifetime), blades locked for ops. at that value

Page 5: NSLS-II Beam Loss Monitor System

5 BROOKHAVEN SCIENCE ASSOCIATES

3GeV Electron Penetration of 10mm Cu

10mm Cu (0.7 Xrad) to reduce Ionization loss peak ~1.2 MeV/gm/cm2

Thin enough to yield electron beam loss signal in magnets

δ < -20% lost upstream of exit from dipole chamber

Page 6: NSLS-II Beam Loss Monitor System

6 BROOKHAVEN SCIENCE ASSOCIATES

Scrapers for Beam Loss Control

Critical 2- δ<0 inner blades, 2- δ>0 outer blades: Hscraper1 &2 2-Vertical pair and 1-Horizontal pair scrapers for X &Y geometric apertures

Injection Septum

Page 7: NSLS-II Beam Loss Monitor System

7 BROOKHAVEN SCIENCE ASSOCIATES

Survival Distribution for 2 & 0 HScrapers

Control of loss for δ <-5% (dX =-23mm) but only if one ID ahead of scraper, i.e. minimum loss control for scatters in rest of the ring

Survival for dp/p = -5.2%

0.001

0.010

0.100

1.000

10.000

100.000

0 50 100

s [m]

Surv

ival [%

]

2-ScrapersHscraper1B30M2Hscraper2B1M2EndSP1Noscrapers

s >SP1 0.04%

Page 8: NSLS-II Beam Loss Monitor System

8 BROOKHAVEN SCIENCE ASSOCIATES

Demonstrated Loss Control in Xray Ring Xray Ring has 5mm Cu scraper near dispersion maximum in SP=3

0.91m CBLM in dipole end

Page 9: NSLS-II Beam Loss Monitor System

9 BROOKHAVEN SCIENCE ASSOCIATES

Demonstrated Loss Control in Xray Ring

0

1000

2000

3000

4000

5000

6000

-40 -35 -30 -25 -20 -15

Life

time {

hr*1

00],C

BLM[

mv*

20],B

LMs

X15 Inner Scraper Position [mm]

X15 Scan proCBLM Hi Gain

Lifetimex100CBLM*20 (mV)b2bm2 (raw)b1bm2 (raw)b1bm1 (raw)

Reduction of losses at b1bm1.. b2bm2 before Lifetime decrease is observed due to the reduction of momentum aperture by the scraper

Beam loss at CBLM after scraper, charge loss rate used to calibrate CBLM

] NaI | BLMs ]Ahead dipoles

Page 10: NSLS-II Beam Loss Monitor System

10 BROOKHAVEN SCIENCE ASSOCIATES

Signal Generation for CBLM

Number of photon for 200 - 800nm in 975/cm

θc(λ) = cos-1[1/(β n(λ))] ~ 46° for quartz n=1.458

dP/dx ~ α*2π*0.5*Eγ *f (dλ/λ2)

2 2 2 -1NE x sin [ (E )] 370 sin [ (E )] ev cm= ≈ c cd

d dZcγ γ γ

α θ θ

Convolving with diode sensitivity from 190-

1000nm

Assuming 50% photons propagate to photodiode (forward cone only TIR)

Gives 0.02nA per electron per turn per cm path

length with dark current 0.1 nA S/N ~0.2

Page 11: NSLS-II Beam Loss Monitor System

11 BROOKHAVEN SCIENCE ASSOCIATES

Lost Electrons Shower in Vacuum Chamber

CBLM

Model e- shower in VC with Shower (EGS4)

Lost e-

Page 12: NSLS-II Beam Loss Monitor System

12 BROOKHAVEN SCIENCE ASSOCIATES

Beam e- @ 3° e- & e+ Generated in VC Walls

1.6 +/- 0.12

e- per lost e-

All e- E > 10MeV

e- in Rod |Y| < 5mm

e+ in Rod |Y| < 5mm

| 10mm |

Page 13: NSLS-II Beam Loss Monitor System

13 BROOKHAVEN SCIENCE ASSOCIATES

Electrons shower in 25mm Al VC Wall

Shower yields more charged particles and longer path length for radiation

5-10 times signal for a single e- loss per turn approaches a S/N ~1 for photodiode

(dark current ~0.1nA)

Page 14: NSLS-II Beam Loss Monitor System

14 BROOKHAVEN SCIENCE ASSOCIATES

Generation of Photons in Radiator Rod

The incident electron direction (θe , Фe , x, y, z ) gives path length p

θe polar angle to axis of rod (z-axis), Фe azimuthal angle in x-y plane

Nγ photons = p* 975 /cm uniform in energy for (200-800nm)

With θc angle to electron direction

Page 15: NSLS-II Beam Loss Monitor System

15 BROOKHAVEN SCIENCE ASSOCIATES

Propagation of Photons to Detector

Fraction of energy transmitted

Out rod to detector (z= 200cm)

Peaks θe = 0: 85%, 90°: 18%

> 110° < 0.1%

Total energy transmitted

Out rod to detector (z= 200cm)

Page 16: NSLS-II Beam Loss Monitor System

16 BROOKHAVEN SCIENCE ASSOCIATES

Output Photon are at Large Angles

Output photon divergence angle

Since NA > 1

Fractional energy reflected symmetric except for electron path is closer to upstream

Page 17: NSLS-II Beam Loss Monitor System

17 BROOKHAVEN SCIENCE ASSOCIATES

Fiber Optics Cable has smaller NA

Fraction of energy transmitted

Out fiber detector (z= 200cm)

Peaks θe = 45: 12%, 5°: 2%

90° : <0.1%

Total energy transmitted

Peak at Cerenkov angle

Out fiber detector (z= 200cm)

Page 18: NSLS-II Beam Loss Monitor System

18 BROOKHAVEN SCIENCE ASSOCIATES

Output Photon at Smaller Angles

Output photon divergence angle

Since NA =0.22 = sin(θ) ~13°

Fractional energy reflected symmetric except for electron path is closer to upstream

Fraction of lost photons out sides of fiber

i.e. not TIR

Page 19: NSLS-II Beam Loss Monitor System

19 BROOKHAVEN SCIENCE ASSOCIATES

Rods Have Lower Radiation Induced Atten.

Suprasil 2B 99.9999% SiO2

Measured Radiation Induced Attenuation (RIA) to 26 Mrad

(One year NSLS-II Ops)

Similar peak RIA for UV fiber at 0.5Mrad exposure and larger in

visible fibers

Page 20: NSLS-II Beam Loss Monitor System

20 BROOKHAVEN SCIENCE ASSOCIATES

Testing Prototype CBLM in Xray Ring Prototype CBLM: Suprasil fused silica rods and Hamamatsu Photodiode Mod.

0

500

1000

1500

2000

2500

3000

3500

4000

0.0 10.0 20.0 30.0 40.0

Dose

Rat

e [ur

em/h

r]

Loss Rate [pC/sec]

Scan X15 Inner Blade Scan

X14 2ndX14 1st

10mm x 0.91m CBLM in dipole end

X14 Rad.Mon.

Beam Loss Rate Q’ = Io * To/τ

Page 21: NSLS-II Beam Loss Monitor System

21 BROOKHAVEN SCIENCE ASSOCIATES

Scraper Controls Local Loss Rate seen by CBLM

0

1000

2000

3000

4000

5000

6000

7000

8000

9000

10000

0 10 20 30 40 50 60 70

CBLM

Out

[mV]

Loss Rate [pC/sec]

CBLM vs Loss Rate 25April2011

InOutFit 213mV/pC/sec

0

50

100

150

200

250

0 10 20 30 40

CBLM

Out

[mV]

Loss Rate [pC/sec]

CBLM vs X13 Scan 25April 2011

X13inFit 5.3mV/pC/sec

CBLM signal scales linearly with local loss rate Io/τ over >40dB

Zero current level is 0.5 -0.7mV or Noise level ~2-3 fC/sec

Lab noise was 0.035mV or 0.2 fC/sec noise level

Scraper out CBLM ~ lifetime loss or I2 for Touschek lifetime

X13 scraper not local to CBLM signal ~40X less

0

1000

2000

3000

4000

5000

6000

7000

8000

0 10 20 30 40 50

CBLM

Out

[mV]

Loss Rate [pC/sec]

CBLM vs Loss Rate 25April2011 80-60mA

InOutFit 207mV/pC/sec

Page 22: NSLS-II Beam Loss Monitor System

22 BROOKHAVEN SCIENCE ASSOCIATES

Lower Gain (~100X) PD module

0

20

40

60

80

100

120

140

160

0 10 20 30 40 50 60

Vcbl

m [m

V]

Beam Loss Rate [pC/sec]

proCBLM Low Gain

dQ/dtFit 2.27mV/pC/s

0

0.5

1

1.5

2

2.5

3

3.5

4

4.5

5

0 10 20 30 40 50 60 70 80 90 100 110

Vcbl

m [m

V]

Loss Rate [pC/sec]

X13 Scan vs proCBLM signal

X13top

Fit0.065mV/pC/s

As for NSLS-II the upstream scraper will provide calibration signal once local loss rate ( Io /tau ) dominates total loss

Page 23: NSLS-II Beam Loss Monitor System

23 BROOKHAVEN SCIENCE ASSOCIATES

Gain too High for Injection Loss

High gain saturates with normal injection losses for ~32 msec

Loss in SP=3 dispersion region

BW= 10Hz so falltime 16msec or 320 samples

The lowest gain ~10-6 V/A of PDM has BW=64KHz or 2.4µsec

Testing in Linac with 120MeV e- beam at normal incidence with 108 (~16 pC) e-/pulse (2nsec long)

Integral or DC term 55nV/pC/sec

Split lowest gain (hi BW) output to multiple channels of gain or

Use log amplifier and count rate for low loss

Page 24: NSLS-II Beam Loss Monitor System

24 BROOKHAVEN SCIENCE ASSOCIATES

Inject. Beam Loss for Low Gain, BW~1KHz

Injection pulse in proCBLM saturates at 9.3V but integral still gives total charge lost, width changes with charge

At high current a 2nd loss occurs ~22ms but represents only a small fraction of the 600 to 800µA/pulse injection rate

Loss at 50ms after injection was seen with high gain as well.

Page 25: NSLS-II Beam Loss Monitor System

25 BROOKHAVEN SCIENCE ASSOCIATES

NSLS-II CBLM System Components

Modular design- Radiator Rods(RR)- 3 dipole, 2 quadrupole Light Output Couplers (LOC)- 5 and Photo Detector Module(PDM)- 5 Dipole RR- 29.5 mm diameter, 1.2m long and 25m bend radius Quad RR – 29.5mm diameter , 1.2m long straight LOC – 90° prisms with Light pipe to PDM for reduced Rad Exposure PDM’s – 9 decade DR detector module, R&D effort ongoing

LOC’s

PDM’s

RR’s

Page 26: NSLS-II Beam Loss Monitor System

26 BROOKHAVEN SCIENCE ASSOCIATES

Cerenkov Radiator Rods and Couplers

Hereaus Quartz: provides two ultra-pure Fused silica materials and rods Suprasil 2B: tested in Xray ring (120Krad), in Co60 (>26Mrad) Spectrosil 2000: (no Chlorine) (existing stock 29.5mm OD 1.2m) Free draw of rods: flame polished, excellent clarity, low scatter, TIR 10mm OD length up to 1meter , 25mm OD up to 2meters Corning Fused silica 7980 O-A similar quality fused silica but need fabricator to draw rods Prisms and tapered collectors will be made with 7980 raw material Rods and prisms supported with lexan spacers with minimum width for TIR

Page 27: NSLS-II Beam Loss Monitor System

27 BROOKHAVEN SCIENCE ASSOCIATES

Output Coupler and RR Components

Heraeus Quartz USA has existing Spectrosil 2000 rods 29.5mm OD and will bend to 25m. Use air gap between rod to prism and prism to down rod loses 8-10% light. Could be chemically activated bonded with only 4% loss.

Page 28: NSLS-II Beam Loss Monitor System

28 BROOKHAVEN SCIENCE ASSOCIATES

Goals for beam loss monitor sensitivity

3hr lifetime loss rate ~ 2000 e-/ turn (2.6 μsec) for 10 e-/ turn or 0.6 pC/sec in CBLM S/N > 10 or 1nA into 1KΩ ~ 10μV x 10K gain 100mV pulse for count rate measurements from 1K to 1M cps ( 1 to 1000 bunches/turn) Losses of 0.001μA/turn 16K e-/ turn 75KHz(5 turns) time for betatron losses 0.98 nC/sec 3.2μA into 1KΩ give 3.2mV x 100 gain Losses of 0.1μA/turn 1.6M e-/ turn 3.3 KHz (114 turns) time for synch. losses 0.1μC/sec 0.32mA into 1KΩ give 320mV x10 gain Losses of 3mA/turn 4G e-/turn over 100 turn pulse: dump, injection, steering 3 mC/sec 800mA peak pulse into 10 Ω 8V peak x 1 gain Large Dynamic Range ~>109 for detector output current with BW from 50MHz – 100 Hz and DC

Page 29: NSLS-II Beam Loss Monitor System

29 BROOKHAVEN SCIENCE ASSOCIATES

Progress in HDR Photodiode Modules

0.001

0.01

0.1

1

10

100

1000

10000

1.E-02 1.E-01 1.E+00 1.E+01 1.E+02 1.E+03 1.E+04 1.E+05

Vout

[mV]

Power [nW]

S1337_LT1222 Response at 405nm DC

Dvm-OffFit_0.22mV/nW

Low Noise Amplifier 7-Decades but boxed in by noise floor

and amplifier bias

Log-amplifier shows 8 to 10 decades but have a lower BW at low levels, not serious issue

Page 30: NSLS-II Beam Loss Monitor System

30 BROOKHAVEN SCIENCE ASSOCIATES

Design of High Dynamic Range Detector Photodiode Detectors-

QE > 25% down to 200nm

DC coupling, high peak current, no HV,

Compact, simpler, external gain, limited by noise floor and amplifier rail

low BW for high gain and large area

PMT Detectors-

High Gain, High BW, pulse count rate

QE < 25% peak , DC coupling HV issue, HV required, high peak current vs gain response complicated,

Page 31: NSLS-II Beam Loss Monitor System

31 BROOKHAVEN SCIENCE ASSOCIATES

PMT Output for Count Rate to Low levels

PMT signals at ~1Kcps clean single e- signal count rate C Q’ = e * C

At ~1Mcps pulse pile-up makes count rate miss charge signals but integrated signal still yields accurate Q’, need FPGA

Page 32: NSLS-II Beam Loss Monitor System

32 BROOKHAVEN SCIENCE ASSOCIATES

Summary of CBLM System for NSLS-II

• Use of thin scrapers gives clear beam loss signal, with physics of shower in VC walls modifiying collection of charge signal (+ & -)

• Scraper controlled beam loss rate for calibration of detector signal

• Calibration without scraper needs modeled capture efficiency

• CBLM can be used without scraper but need model for fraction of scraper beam intercepted by RR

• Fused silica rods have greater Cerenkov light collection and radiation resistance and less directional variation than Fiber Optics

• Can be used for continuous or pulsed beam losses, will need higher BW (smaller PDs or PMTs), Location determined by RR length.

• Dual detector and log amplifier could yield new generation of HDR PDM especially if FPGA can deal with data streaming and transition from pulse counting to analog levels in smooth way (overlapping ranges)

Page 33: NSLS-II Beam Loss Monitor System

33 BROOKHAVEN SCIENCE ASSOCIATES

Dipole CBLM/Coil Support New Design

Light Output Coupler (LOC)

Photo Detector Module (PDM)

^ Dipole Support for CBLM/Coils with Radiator Rod inside

Page 34: NSLS-II Beam Loss Monitor System

34 BROOKHAVEN SCIENCE ASSOCIATES

Replacement of Radiator Rods

Spectrosil 2000 is best glass for low RIA >10years ops

If attenuation is too low for calibration of high end loss rates then replace

1.Cut damaged rod too removable pieces as pulled out 2.Replace with new glass cut and polished to shorter lengths

with air splice joints (7-10% loss per joint)

Page 35: NSLS-II Beam Loss Monitor System

35 BROOKHAVEN SCIENCE ASSOCIATES

Need for HDR Detector

Assuming 500mA operations at 3hr lifetime for 5500 hrs per year Top-off charge injection rate into SR 7.3 nC/min Total = 2.42 mC/ yr CBLM rate 50 to 100 e-/turn 6pC/sec 0.36 pC/min 55 dumps per year = 73 µC /yr or 3% of total charge lost for operation year 1.32 µC/ turns 500 mC/sec dump rate into CBLM 20% inject losses /turn 1.46nC/turn 0.1mC/sec into CBLM