Noxious Gases Ref Peaks:An Alternative to Dual Inlet Bellows

Paul Eby

Page 2: Noxious Gases Ref Peaks: An Alternative to Dual Inlet Bellows Paul Eby.

Normal Conflo Ref Gas

GC / EA

Ref Gas

He

IRMS

- Ref Gas is always flowing (0.2mL/min)

- Not okay for noxious gases

- CO, CH3Cl, SO2

- Small quantities / expensive gases

Page 3: Noxious Gases Ref Peaks: An Alternative to Dual Inlet Bellows Paul Eby.

Using the BellowsGC / EA

- Consumption is very low (0.02 to 0.002mL/min)

- Continuously flows

- Noxious gas exits with the pump exhaust

IRMS

REF SAMPLE

- Requires dual inlet

Page 4: Noxious Gases Ref Peaks: An Alternative to Dual Inlet Bellows Paul Eby.

37Cl Measurements

1. Precipitate chloride out as AgCl by adding AgNO3

2. Convert AgCl to CH3Cl gas by reacting with CH3I

3. Run by GC-IRMS

4. Measured as CH3Cl

ProblemsRequires an extra cup to measure masses 50 and 52

Bellows needed for toxic CH3Cl (also expensive)

$ $ $ $

Page 5: Noxious Gases Ref Peaks: An Alternative to Dual Inlet Bellows Paul Eby.

Thermo Conflo IV

Allows for 5 different reference gases, 2 at a time

Page 6: Noxious Gases Ref Peaks: An Alternative to Dual Inlet Bellows Paul Eby.

Thermo Conflo IV

Ref A

Ref B

Ref C

IRMS

Can we leave the ref gas off until a peak is needed?

Does the gas need to reach a steady state by constantly flowing?

How long does it take to reach a steady state?

Can I test this before upgrading the IRMS?

Page 7: Noxious Gases Ref Peaks: An Alternative to Dual Inlet Bellows Paul Eby.

Thermo Conflo IV in Isodat

Extra columns added to the method using Configurator

Page 8: Noxious Gases Ref Peaks: An Alternative to Dual Inlet Bellows Paul Eby.

Carbon Dioxide On/Off Test

ave = -0.03 stdev = 0.02 ave = -0.01 stdev = 0.04

Page 9: Noxious Gases Ref Peaks: An Alternative to Dual Inlet Bellows Paul Eby.

Carbon Dioxide On/Off Test

-0.2

-0.15

-0.1

-0.05

0

0.05

0.1

0.15

0.2

1 2 3 4 5 6 7 8 9 10 11 12

d13C

/ p

erm

il

Page 10: Noxious Gases Ref Peaks: An Alternative to Dual Inlet Bellows Paul Eby.

Methyl Chloride Trials

1. Upgrade the detectors

2. Add ventilation to the Conflo IV

3. Pump exhaust vented

4. Small bomb of CH3Cl

Page 11: Noxious Gases Ref Peaks: An Alternative to Dual Inlet Bellows Paul Eby.

Methyl Chloride On/Off Test

ave = 0.01 stdev = 0.01 ave = -0.02 stdev = 0.02

Page 12: Noxious Gases Ref Peaks: An Alternative to Dual Inlet Bellows Paul Eby.

-0.2

-0.15

-0.1

-0.05

0

0.05

0.1

0.15

0.2

1 2 3 4 5 6 7 8 9 10 11

d37C

l / p

erm

il

Methyl Chloride On/Off Test

Page 13: Noxious Gases Ref Peaks: An Alternative to Dual Inlet Bellows Paul Eby.

Methyl Chloride On/Off Test

Page 14: Noxious Gases Ref Peaks: An Alternative to Dual Inlet Bellows Paul Eby.

Conclusions

1. ConFlo IV can be used to limit consumption of ref gas

2. Comparing to Bellows:

Flow rate in Conflo IV 0.2mL/min

Flow rate in bellows 0.02 - 0.002mL/min?