NORMAL CORNEA The highest diopter of human eye,about 43D at corneal apex The highest diopter of...

download NORMAL CORNEA The highest diopter of human eye,about 43D at corneal apex The highest diopter of human eye,about 43D at corneal apex Average radius curvature.

If you can't read please download the document

Transcript of NORMAL CORNEA The highest diopter of human eye,about 43D at corneal apex The highest diopter of...

  • Slide 1
  • NORMAL CORNEA The highest diopter of human eye,about 43D at corneal apex The highest diopter of human eye,about 43D at corneal apex Average radius curvature of 7.8 mm Average radius curvature of 7.8 mm Is not absolutely transparent,scatters about 10%of the incident light Is not absolutely transparent,scatters about 10%of the incident light
  • Slide 2
  • CORNEAL GEOGRAPHY The central zone (4mm diameter)almost spherical and called apex, responsible for the high definition vision the paracentral zone where the cornea begins to flatten The peripheral zone the limbal zone
  • Slide 3
  • Normal cornea A normal prolate cornea is steep in center and flat in periphery A normal prolate cornea is steep in center and flat in periphery the center of the normal cornea is steeper than the best fit sphere & midperiphery is flatter than reference sphere the center of the normal cornea is steeper than the best fit sphere & midperiphery is flatter than reference sphere Central cornea is red on a normal elevation map and midperiphery appears blue Central cornea is red on a normal elevation map and midperiphery appears blue
  • Slide 4
  • SUPER VISION Developing new tools and extremely promising laser surgical techniques that have proven to increase the human beings VA by reducing corneal aberrations Topographic & aberrometer linked LASIK are on the way to achieve this goal of better than normal vision Regularizing the corneal shape by means of reduction of halos,glare & other optical aberrations The influence of other dioptric surfaces (vitreous,lens,..)and interfaces still has to be ascertained
  • Slide 5
  • Posterior corneal power Is negative and much smaller than anterior corneal power. Is negative and much smaller than anterior corneal power. The posterior surface reduces corneal power The posterior surface reduces corneal power
  • Slide 6
  • 6
  • Slide 7
  • Instruments to measure the corneal surface Keratometry Keratometry Keratoscopy Keratoscopy Computerized videokeratoscopy (topography) Computerized videokeratoscopy (topography)
  • Slide 8
  • CORNEAL TOPOGRAPHY Wide acceptance as a clinical examination procedure with the advent of modern laser refractive surgery Measure a greater area of the cornea with a much higher number of points and produce permanent records
  • Slide 9
  • Topographic techniques Reflection perform indirect measurement of the corneal surface Reflection perform indirect measurement of the corneal surface Projection visualize the corneal surface directly Projection visualize the corneal surface directly
  • Slide 10
  • PLACIDODISC Illuminates the cornea by sending a mire of concentric rings Illuminates the cornea by sending a mire of concentric rings A videocamera captures the corneal reflex from the tear layer A videocamera captures the corneal reflex from the tear layer A computer & software perform the analysis of the data through different algorithms A computer & software perform the analysis of the data through different algorithms
  • Slide 11
  • Keratometry and corneal topography with placidodisc systems were originally invented to measure anterior corneal curvature Keratometry and corneal topography with placidodisc systems were originally invented to measure anterior corneal curvature The problem in the placidodisc system is that cannot perform a slit scan topography The problem in the placidodisc system is that cannot perform a slit scan topography
  • Slide 12
  • TOPOGRAPHY corneal topography plays a critical role in refractive surgery decisions corneal topography plays a critical role in refractive surgery decisions conventional axial & tangential topography are not enough to demonstrate a healthy cornea conventional axial & tangential topography are not enough to demonstrate a healthy cornea can not decide any type of laser refractive surgery based only on surface topographic evaluation can not decide any type of laser refractive surgery based only on surface topographic evaluation 12
  • Slide 13
  • ORBSCAN Combines both slit scan and placido images to give a very good composite picture from topographic analysis
  • Slide 14
  • 14
  • Slide 15
  • 15
  • Slide 16
  • Artifacts of topography Artifacts of topography Shadows on the cornea from large eyelashes or trichiasis Shadows on the cornea from large eyelashes or trichiasis Ptosis or non-sufficient eye opening Ptosis or non-sufficient eye opening Irregularities of the tear film layer (dry eye) Irregularities of the tear film layer (dry eye) Too short working distance of the small placidodisc cone Too short working distance of the small placidodisc cone Incomplete or distorted image (pathology) Incomplete or distorted image (pathology)
  • Slide 17
  • ORBSCAN II Is a fully integrated multidimensional diagnostic system Is a fully integrated multidimensional diagnostic system Acquires over 9000 data points in 1.5 seconds to meticulously map the entire corneal surface Acquires over 9000 data points in 1.5 seconds to meticulously map the entire corneal surface Analyze elevation & curvature measurement on both the anterior & posterior surfaces of the cornea Analyze elevation & curvature measurement on both the anterior & posterior surfaces of the cornea 17
  • Slide 18
  • ORBSCAN SYSTEM Use the principle of projection Use the principle of projection Forty scanning slit beams (20 from the left and 20 from the right with up to 240 data points per slit ) to scan the cornea and measure independently the X,Y & Z locations Forty scanning slit beams (20 from the left and 20 from the right with up to 240 data points per slit ) to scan the cornea and measure independently the X,Y & Z locations 18
  • Slide 19
  • Orbscan imaging Forty slit images are acquired in two 0.7 second periods Forty slit images are acquired in two 0.7 second periods Each of the 40 slit images triangulates one slice of ocular surface Each of the 40 slit images triangulates one slice of ocular surface Distance between data slices average 250 microns Distance between data slices average 250 microns
  • Slide 20
  • ORBSCAN A three dimensional slitscan topographic A three dimensional slitscan topographic Orbscan detect the abnormalities on the posterior surface of the cornea Orbscan detect the abnormalities on the posterior surface of the cornea
  • Slide 21
  • 21
  • Slide 22
  • ORBSCAN Orbscan I only slit scan topography Orbscan I only slit scan topography Orbscan II the placidodisc added in orbscan I Orbscan II the placidodisc added in orbscan I
  • Slide 23
  • 23
  • Slide 24
  • ORBSCAN II Is based on slit scanning technology in addition to traditional placido-based technique Is based on slit scanning technology in addition to traditional placido-based technique Placidodisc improve the accuracy of the curvature measurements and give information on axial keratometric readings Placidodisc improve the accuracy of the curvature measurements and give information on axial keratometric readings 24
  • Slide 25
  • ORBSCAN The images used to construct the anterior corneal surface, posterior corneal surface, anterior iris and anterior lens surfaces The images used to construct the anterior corneal surface, posterior corneal surface, anterior iris and anterior lens surfaces Data regarding the corneal pachymetry and anterior chamber depth Data regarding the corneal pachymetry and anterior chamber depth 25
  • Slide 26
  • Elevation Orbscan measure elevation Orbscan measure elevation Elevation is important the only complete scaller measure of surface shape Elevation is important the only complete scaller measure of surface shape Both slope & curvature can be mathematically derived from a single elevation map Both slope & curvature can be mathematically derived from a single elevation map
  • Slide 27
  • ORBSCAN Slit beam scanners and triangulation are used to derive the actual spatial location of thousands of points on the surface Slit beam scanners and triangulation are used to derive the actual spatial location of thousands of points on the surface Each beam sweep across the cornea gives information on corneal elevation or height from the anterior corneal surface, posterior surface & iris Each beam sweep across the cornea gives information on corneal elevation or height from the anterior corneal surface, posterior surface & iris 27
  • Slide 28
  • BEST FIT SPHERE (BFS) The computer calculates a hypothetical sphere that matches as close as possible to the actual corneal shape being measured The computer calculates a hypothetical sphere that matches as close as possible to the actual corneal shape being measured Compares the real surface to the hypothetical sphere showing areas above the surface of the sphere in warm colours and areas below the surface in cool colours Compares the real surface to the hypothetical sphere showing areas above the surface of the sphere in warm colours and areas below the surface in cool colours 28
  • Slide 29
  • Topography quad map The upper left : anterior float The upper left : anterior float The upper right : posterior float The upper right : posterior float The lower left : keratometric pattern The lower left : keratometric pattern The lower right : pachymetry map The lower right : pachymetry map
  • Slide 30
  • 30
  • Slide 31
  • 31
  • Slide 32
  • 32
  • Slide 33
  • 33
  • Slide 34
  • NORMAL BAND SCALE Highlights the abnormal areas in the cornea in orange to red colors Highlights the abnormal areas in the cornea in orange to red colors The normal areas are all shown in green The normal areas are all shown in green Helpful in generalized screening in preoperative examination Helpful in generalized screening in preoperative examination
  • Slide 35
  • 35
  • Slide 36
  • POWER MAP = AXIAL MAP A familiar sagital map from placido system A familiar sagital map from placido system The mean power map determines the location of a surface abnormality The mean power map determines the location of a surface abnormality Normal astigmatism appears a classic bowtie Normal astigmatism appears a classic bowtie
  • Slide 37
  • AXIAL MAP Provides detailed keratometric information across the diameter of the cornea Provides detailed keratometric information across the diameter of the cornea K readings are between certain values the cornea must be neither too steep nor too flat K readings are between certain values the cornea must be neither too steep nor too flat 37
  • Slide 38
  • AXIAL MAP To create a good quality corneal flap in LASIK if either extremes (too steep or too flat) is the case, this can lead to surgical flap complications To create a good quality corneal flap in LASIK if either extremes (too steep or too flat) is the case, this can lead to surgical flap complications K readings of more than 48 D are an indication of potential keratoconus K readings of more than 48 D are an indication of potential keratoconus 38
  • Slide 39
  • 39
  • Slide 40
  • 40
  • Slide 41
  • 41
  • Slide 42
  • Middle box Keratometric readings Keratometric readings White to white distance in mm White to white distance in mm Angle kappa readings Angle kappa readings The thinnest point of cornea The thinnest point of cornea irregularity within the central 3 mm & 5 mm irregularity within the central 3 mm & 5 mm
  • Slide 43
  • 43
  • Slide 44
  • Pachymetry Ultrasound provides a reading of corneal thickness from bowmans membrane to descemets membrane Ultrasound provides a reading of corneal thickness from bowmans membrane to descemets membrane Orbscan provides a reading from the precorneal tearfilm to the endothelium,slightly thicker readings can be expected Orbscan provides a reading from the precorneal tearfilm to the endothelium,slightly thicker readings can be expected 44
  • Slide 45
  • PACHYMETRY MAP The orbscan measures thickness from the tear film layer to descemets membrane and is thicker than that obtained with ultrasound The orbscan measures thickness from the tear film layer to descemets membrane and is thicker than that obtained with ultrasound Adjustment factor (acoustic factor),the default setting is 92% Adjustment factor (acoustic factor),the default setting is 92% Provides a reading showing the thinnest point of the cornea that may not necessarily be the central reading Provides a reading showing the thinnest point of the cornea that may not necessarily be the central reading
  • Slide 46
  • 46
  • Slide 47
  • PACHYMETRY Provides thickness information the cornea from limbus to limbus Provides thickness information the cornea from limbus to limbus The relationship between pachymetry readings can be looked,100 micron should be a cut-off criteria The relationship between pachymetry readings can be looked,100 micron should be a cut-off criteria 47
  • Slide 48
  • pachymetry Central pachymetry compare to peripheral readings considered abnormal if : -the peripheral readings are not at least 20 micron thicker The thinnest reading less than 30 micron thinner
  • Slide 49
  • PACHYMETRY MAP Thinnest point 100 microns from the thinnest point to the values at 7mm optical zone 49
  • Slide 50
  • ELEVATION MAPS The anterior elevation map ;the top left hand map The anterior elevation map ;the top left hand map The posterior elevation map ;the top right hand map The posterior elevation map ;the top right hand map Slit scanning provides elevation data, and also can create a 3D interpretation of the cornea Slit scanning provides elevation data, and also can create a 3D interpretation of the cornea 50
  • Slide 51
  • ELEVATION MAP The green colour is referred as reference sphere (at sea level ) The green colour is referred as reference sphere (at sea level ) The warmer colours are above this level and the cooler colours are below The warmer colours are above this level and the cooler colours are below 51
  • Slide 52
  • 52
  • Slide 53
  • 53
  • Slide 54
  • ANTERIOR ELEVATION MAP Looking at a proper scale in the cornea, can see height differences Looking at a proper scale in the cornea, can see height differences Compare the height of the actual cornea to a best fit sphere Compare the height of the actual cornea to a best fit sphere
  • Slide 55
  • Posterior map The highest elevation value as a keratoconus indicator or at least as a screen for patients may be at risk of developing keratectasia The highest elevation value as a keratoconus indicator or at least as a screen for patients may be at risk of developing keratectasia 55 D elevation as an absolute cut off 55 D elevation as an absolute cut off 55
  • Slide 56
  • ELEVATION DATA The difference between the highest and lowest points is a potential keratoconus indicator if over 100 microns (Rousch criteria) The difference between the highest and lowest points is a potential keratoconus indicator if over 100 microns (Rousch criteria) 56
  • Slide 57
  • DIAGNOSTIC CRITERIA Power map changes Power map changes Posterior elevation maps Posterior elevation maps Pachymetry Pachymetry Composite/integrated topography information Composite/integrated topography information 57
  • Slide 58
  • POWER MAP Mean corneal power >45D Mean corneal power >45D In addition to steep corneal curvatures-the bowtie or broken bowtie appearance indicative of early keratoconus In addition to steep corneal curvatures-the bowtie or broken bowtie appearance indicative of early keratoconus Central corneal asymetry a change within central 3mm optical zone of the cornea of more than 3D Central corneal asymetry a change within central 3mm optical zone of the cornea of more than 3D 58
  • Slide 59
  • Irregularity in central cornea Greater than 1.5 D in 3 mm zone and greater than 2.0 D in 5 mm zone is considered abnormal and cause for concern Greater than 1.5 D in 3 mm zone and greater than 2.0 D in 5 mm zone is considered abnormal and cause for concern
  • Slide 60
  • ELEVATION MAPS A ratio can be calculated between the posterior and anterior surfaces,which gives an indication of the relative difference in curvature between the two maps A ratio can be calculated between the posterior and anterior surfaces,which gives an indication of the relative difference in curvature between the two maps 60
  • Slide 61
  • POSTERIOR ELEVATION MAP Many surgeons think the first sign of keratoconus appears on the posterior surface of the cornea Many surgeons think the first sign of keratoconus appears on the posterior surface of the cornea 3.13% of population screened for laser surgery had posterior ectasia criteria by orbscan, despite having axial topography classified as normal 3.13% of population screened for laser surgery had posterior ectasia criteria by orbscan, despite having axial topography classified as normal 61
  • Slide 62
  • 62
  • Slide 63
  • POSTERIOR ELEVATION MAP The most common reference surface for viewing elevation maps is the best fit sphere The most common reference surface for viewing elevation maps is the best fit sphere A best fit sphere (BFS) >55D on the posterior profile, indicative of posterior ectasia A best fit sphere (BFS) >55D on the posterior profile, indicative of posterior ectasia 63
  • Slide 64
  • POSTERIOR ELEVATION MAP Increased forward shift of the posterior corneal surface is common after myopic LASIK and correlates with the residual corneal thickness Increased forward shift of the posterior corneal surface is common after myopic LASIK and correlates with the residual corneal thickness 64
  • Slide 65
  • Posterior float difference Greater than 50 micron generally accepted as abnormal Greater than 50 micron generally accepted as abnormal In corneas thinner than normal over 40 as abnormal In corneas thinner than normal over 40 as abnormal
  • Slide 66
  • Posterior elevation map The location of the steepest part of the posterior float should be relatively central, but is a more concern it be located away from the center and in an area of corneal thinning The location of the steepest part of the posterior float should be relatively central, but is a more concern it be located away from the center and in an area of corneal thinning Posterior float difference;40 to 50 microns seems to be the maximum difference Posterior float difference;40 to 50 microns seems to be the maximum difference
  • Slide 67
  • 67
  • Slide 68
  • Correlation of signs of the highest point Highest point on the posterior elevation coincides with the highest point of anterior elevation, the thinnest on pachymetry and the point of steepest curvature on the power map Highest point on the posterior elevation coincides with the highest point of anterior elevation, the thinnest on pachymetry and the point of steepest curvature on the power map 68
  • Slide 69
  • 69
  • Slide 70
  • Although high posterior elevation and ratio between two elevation maps rarely used as exclusion criteria alone, but by considering these together, more conclusive information can be obtained Although high posterior elevation and ratio between two elevation maps rarely used as exclusion criteria alone, but by considering these together, more conclusive information can be obtained 70
  • Slide 71
  • Risks of ectasia indices Number of abnormal maps Posterior float difference >0.050 3mm & 5mm irregularity Peripheral thickness changes Astigmatism variance between eyes Steep ks mean power map
  • Slide 72
  • 72
  • Slide 73
  • Three step rule One abnormal map ; perform with caution One abnormal map ; perform with caution Two abnormal map ; with concern Two abnormal map ; with concern Three abnormal map ;contraindicated Three abnormal map ;contraindicated
  • Slide 74
  • 74
  • Slide 75
  • 75
  • Slide 76
  • 76
  • Slide 77
  • Composite/integrated information Similarly between anterior & posterior profiles a forward bending of areas shown above the BFS and association with the thinnest point on the cornea Similarly between anterior & posterior profiles a forward bending of areas shown above the BFS and association with the thinnest point on the cornea Inferotemporal displacement of the highest point Inferotemporal displacement of the highest point 77
  • Slide 78
  • 78
  • Slide 79
  • 79
  • Slide 80
  • Abnormal tear film Can significantly distort the readings Can significantly distort the readings The significant change in surface quality and validity of the dry eye The significant change in surface quality and validity of the dry eye
  • Slide 81
  • 81