Neptune Service Manual
-
Upload
juan-moya-blas -
Category
Documents
-
view
3.112 -
download
473
Transcript of Neptune Service Manual
NEPTUNE ELECTRONIC VENTILATOR
TECHNICAL MANUAL
2nd edition January 2006
MEDEC BENELUX NV
All rights reserved. No part of this publication may be reproduced, stored in a retrieval system or transmitted, in any form or by any means, electronically, mechanically, by photocopying, recording or otherwise, without the prior written permission of Medec Benelux NV. Medec Benelux NV reserves the right to change specifications without prior notification. Careful attention has been paid to the compilation of this publication. Medec Benelux NV Lion D’Orweg 19 9300 Aalst Belgium Telephone : (32) 53 / 70.35.44 Fax : (32) 53 / 70.35.33 Website : www.medecbenelux.be E-mail : [email protected] 2nd edition January 2006
User responsibility
The equipment described in this manual has been built to confirm with the specifications and instructions stated in this manual. To ensure proper and safe operation of the equipment, it must be checked and serviced at least according to the minimum standards laid out in this manual. The equipment must be repaired and serviced only in accordance with written instructions issued by Medec Benelux N.V. and must not be altered or modified in any way without written approval of Medec Benelux N.V. The user of this equipment shall have the responsibility for any malfunction which results from improper use, maintenance, repair, damage or alteration by anyone other than Medec Benelux N.V. or its appointed agents. The Neptune anaesthesia combination has been specially developed for anaesthetic applications. The system is built around the so-called bag in bottle principle. This system has for many years proved to be a reliable and safe system. One important aspect in this connection is that the gas exchange of the patient is completely separated from the machine by use of the bag mentioned. Another important aspect is that the Neptune anaesthesia system, whatever catastrophe might occur (compressed air, mains voltage failure or technical problems), always goes into the spontaneous / manual respiration mode. This way, the (manual) respiration of the patient is always guaranteed as long as the fresh gas supply is assured. The Neptune anaesthesia system may only be used in anaesthetic rooms which are conducting and are provided with proper ventilation and electrical wiring. The Neptune anaesthesia system may only be used in anaesthetic rooms conform with EN60601-1-2 level. RS232 output (optional) may only be used with devices conform with EN60601-1-2 level.
Table of Contents
1. Electronic system 1.1 Electronic system overview 1 1.2 Power supply 1.2.1 Replacing the AC mains fuses 5 1.2.2 Block diagram 5 1.2.3 Schematic diagram 8 1.2.4 Power supply board layout 11 1.3 Backplane board 1.3.1 Block diagram 12 1.3.2 Schematic diagram 14 1.3.3 Backplane board layout 16 1.4 MMI board 1.4.1 Block diagram 17 1.4.2 Schematic diagram 19
1.4.3 MMI board layout 25 1.4.4 Build-in test software 27 1.4.5 Graphics display 53 1.5 O2 measurement 1.5.1 General 54
1.5.2 Operation 54 1.5.3 O2 measurement specifications 55 1.6 Master board 1.6.1 Block diagram 56 1.6.2 Schematic diagram 58 1.6.3 Build-in test software 61 1.6.4 Master board layout 69 1.7 Pneumatic board 1.7.1 Block diagram 70 1.7.2 Schematic diagram 72 1.7.3 Built-in test software 77 1.7.4 Pneumatic board layout 85 2. Pneumatic system 2.1 Pneumatic diagram 86 2.2 Manual/Spontaneous mode (MAN) 88 2.3 Controlled mandatory ventilation (CMV) 90 2.4 Pressure controlled ventilation (PCV) 92 2.5 Peep 94 2.6 Flush safety 94 2.7 Mechanical construction 2.7.1 Patient breathing unit 95 2.7.2 Bottle 99 2.7.3 Supply tank 101 2.7.4 Solenoid valves 102
3. Maintenance and calibration 3.1 Visual inspection 103 3.2 Battery backup 104 3.3 The calibration menu 105 3.4 The pressure transducers 108 3.5 The pressure regulators 113 3.6 The flow regulators 116 3.7 O2 flush 118 3.8 The input pressure switches 119 3.9 The bottle safety valve 124 3.10 Peep valve calibration 125 3.11 Valve test 127 3.12 Performing a leaktest 3.12.1 Entering the leaktest menu 130 3.12.2 Performing the leaktest 132 3.12.3 Leaving the leaktest menu 135 3.13 Maintenance instructions 136 3.14 Parts list 141 4. Classification and discard 4.1 Classification 142 4.2 Discard 143 5. Inspection 5.1 Visual inspection 144 5.2 Other inspections 145 6. Troubleshooting 148 A. Checklist Neptune 165 B. List replaced parts 166
Neptune ventilator Technical manual 1
1. ELECTRONIC SYSTEM
1.1 ELECTRONIC SYSTEM OVERVIEW The Neptune contains several electronic printed circuit boards, which are located on different locations in the machine.
Electronic system overview
Neptune ventilator Technical manual 2
• Mains AC input The AC inlet at the back of the ventilator contains a fuse holder and power switch. The AC mains voltage is connected to the AC/DC converter board.
• AC/DC converter board
The AC/DC converter board converts the AC mains voltage to +24V DC voltage. The +24V DC is fully isolated from the AC mains voltage to ensure safe working conditions.
• DC/DC power supply
The DC/DC power supply board generates different DC voltages and controls the battery charging current. A special battery controller device will charge the battery as safe as possible and under the best conditions to have a long battery lifetime. The DC/DC power supply output voltages are:
o +5V DC o +12V DC o +24V DC o +12V DC
• Extendable system set
The extendable system set consists of 4 different boards that are connected by means of 2 elevated multiple pins connectors:
o MMI board o Backplane board o Master board o Pneumatic board
The 4 different boards have their own specific tasks. The MMI board, master board and the pneumatic board have one or more microprocessors and other electronic devices on board. The backplane board only contains connectors and has no microprocessor. The extendable system set is built so that further expansion of the electronics is possible by means of the elevated multiple pins connectors.
• Backplane board
The backplane board contains buzzer, O2 sensor, power switch and power supply connectors. The power connector is connected to the DC/DC power supply. The board is distributing the different DC voltages through the extendable system set.
Neptune ventilator Technical manual 3
• MMI board
The MMI board contains only one microprocessor. It’s reprogrammable so it can be upgraded if necessary. The microprocessor controls all the graphical display functions and is communicating - by means of the master board - with the pneumatic board. The microprocessor is also controlling the keyboard board, the control knob and the speaker.
• Master board
The master board contains one microprocessor and is also reprogrammable if necessary. The master board looks after the communication between the MMI board and pneumatic board. The master board microprocessor checks the proper working of the MMI and pneumatic board microprocessors. Otherwise, the MMI board microprocessor checks the proper working of the master and pneumatic board microprocessor and the pneumatic board microprocessor checks the MMI and master board microprocessors. Each microprocessor is checking the other microprocessors for errors or malfunctions, which makes the ventilator a very safe system. Once an error is found, the electronics stop working and a continuous beep is audible.
Note: The ventilator automatically switches over to manual mode after an error or malfunction is detected.
• Pneumatic board
The pneumatic board contains one microprocessor. This microprocessor is reprogrammable for upgrades if necessary. The pneumatic board of course controls the pneumatic functions of the ventilator. There are six pressure transducers mounted on the board for measuring all kinds of parameters. The pneumatic board also drives the electrical valves on the manifold and bottle. The microprocessor receives settings from the MMI board and processes the values. The pneumatic board microprocessor sends information back to the MMI board like e.g.: actual airway pressure, alarms, patient status values, etc.
• Keyboard board
The keyboard board contains several switches and leds for interaction between the operator and the electronic system. The keyboard board is connected to the MMI board and processed by the microprocessor on the MMI board.
Neptune ventilator Technical manual 4
• Graphical display
The graphical display shows all kind of information. It consists of menus, settings, parameters, graphs, etc. The display is also responsible for interaction between the operator and the ventilator electronics. Along with the keyboard switches and leds, they form the operator’s interface.
Neptune ventilator Technical manual 5
1.2 POWER SUPPLY The AC mains inlet at the back of the ventilator is equipped with a power switch and a fuse holder block. The fuse holder contains two fuses type 2A slow. The AC mains voltage must be in the range from 100V AC to 240V AC. The AC mains frequency must be in the range from 50 Hz to 60 Hz. These are the absolute maximum ratings for the AC mains supply. 1.2.1 Replacing the AC mains fuses Replace the AC mains fuses by mean of the following instructions:
- Turn off the Neptune and disconnect the power cord from the mains inlet.
- Insert a screwdriver in the small AC receptacle and pull out the fuse
holder block.
- Replace the blown fuses and place the fuse holder block back into the AC receptacle.
- Connect the power cord back to the AC mains inlet.
Note: Always replace the blown fuses with the same type and ratings. 1.2.2 Block diagram The power supply block diagram is represented on the next page.
Neptune ventilator Technical manual 6
Block diagram power supply
Neptune ventilator Technical manual 7
The battery-charging unit charges the battery to maintain a maximum battery capacity and life. If the AC mains supply voltage is present, the battery is charged with a maximum current of approximately 0,9 Ampere. After a while, the charging current decreases and a minimum trickle current of 0.09 Ampere flows through the battery. To guarantee full recharge of an exhausted battery, connect the ventilator to the mains supply for at least a couple of hours. Note: The switch at the back of the ventilator must be turned ON to activate the battery charging. Note: The ventilator uses high voltages, capable of causing personal injury. Do not touch the AC mains voltage electronics during operation. If an AC mains power failure occurs, the power supply automatically switches over to battery supply. You can work approximately 1 hour on battery supply. The battery-charging unit monitors the battery voltage and checks for a battery voltage lower than 10,5 Volt. If the battery voltage is getting lower than 10,5 Volt, the power supply electronics is disconnected from the battery to prevent a totally exhausted battery. A totally exhausted battery will shrink the battery lifetime enormously. An audible alarm is activated during 20 seconds after the ventilator shuts off because of an empty battery. The power supply board is equipped with 5 fuses:
- battery fuse F1: 3,15A slow - +5V DC fuse F2: 3,15A slow - +12V DC fuse F3: 3,15A slow - +24V DC fuse F4: 3,15A slow - +12V DC fuse F5: 3,15A slow
The green LEDs indicate if a voltage is present:
- LED D19 indicates the +24V DC input voltage from AC/DC converter - LED D14 indicates the +5V DC output voltage - LED D15 indicates the +12V DC output voltage - LED D16 indicates the +12V DC output voltage - LED D17 indicates the +24V DC output voltage
The +24 Volt DC output voltage is not available when working on battery. This means that LED D19 and D17 aren’t lit in this situation. You can rapidly determine a blown fuse by taking a look at the LED indicators. There is no LED indicator provided on the battery input voltage to prevent a waste of battery power.
Neptune ventilator Technical manual 8
The power supply board is provided with four connectors:
• Connector P3:
Pin number Description 1 +24V DC input voltage 2 +24V DC input voltage 3 GND 4 GND 5 Battery + input voltage 6 Battery – input voltage
• Connector P1:
Pin number Description 1 +5V DC output voltage 2 GND 3 +12V DC output voltage 4 GND 5 +24V DC output voltage 6 GND
• Connector P2:
Pin number Description 1 +12V DC output voltage 2 +12V DC output voltage 3 GND 4 GND
• Connector P4:
Pin number Description 1 AC mains LED indicator voltage 2 Ventilator ON/OFF switch input 3 Ventilator ON/OFF switch input
1.2.3 Schematic diagram The schematic diagram of the power supply is represented on the following pages.
Neptune ventilator Technical manual 9
1 2 3 4
A
B
C
D
43 21
D
C
B
A Title
Number RevisioSizeA4
Date: 6-Jan-2003 Sheet of File: C:\DOCUMENT\..\psu042000v1 Drawn
IN+ 3
IN+ 14 IN-2
IN-15
IN+ OP15
IN+ OP212 IN- OP16
IN- OP211
OUT 1
OUT 16
OUT 7
OUT 10
FB8 CATHOD 9
V+4
V-13
U1 LM61
GND
R193K1
R213k
R31k2
D1 MBR36
D2 MBR36
D3 MBR36
Q1BS17
GND
R4 100K
R5 1M
C1 1µF/25
C2 1µF/25
GND
D11 1n414
GND
R618k
R7R243
2
13Q5
BS25
C310µF/25
GND
c/s- 2 c/l- 4 c/s+ 3 v.in 5 c/s 1 o.c.t. 8 gnd 6 p.i. 7 o.ch.ind 9 st.lev.con 10 trik.bia 11 char.en. 12 vol.sens 13 comp. 14 d.so. 15 d.si. 16 U2 UC390
Q8 BD242R8
R0.2R9107K
R10R787
R1123k
R12R470
D4
MBR36
R13 294K
C4100N
GND
GND
GND
R14 1k
D13
BZX79C20
GND GND
R16100K
R19
470K
Q2BS17
GND
Q6
BS25
R17100K
GND GND
B1
BUZZE
C62200µF/25
GND
R20R330
D121N414
+12V
+12V
+12V
+12V
+24v
+VBA
GND
2
13
Q9RFP30P0
Q7BS25
Q3BS17
1 2 3
14
7
U3A
CD409
5 6 4
14
7
U3B
CD409
8
910
14
7
U3C
CD409
12 13
11
14 7
U3D
CD409
GND
GND
GND
GND
+12V
+12V +12V
+12V
GND
+12V
GND
F1
3.15A/
GND
GND
GND
+12V
+12V
GND
+12V
+12V S 6 CLK 3 D 5 R 4
Q 1Q 2
VCC 14 GND 7U4A CD4013
S8CLK11
D9R10
Q 13
Q 12
VCC14 GND 7U4B CD4013
D5 MBR36
R2122K
C5 100N
C7100N/40
GND
GND
R18100
+12V
C910µF/25
GND
POWER SUPPLY
1 2JP - KDP
PSU042000V1
R22 680K
1 2 3
P4
GND
+24v
1 2 3 4 5 6
P3 +24v
GND
+24V
DC INPUT
+24V
+24V
GND GND BAT+ BAT-
KEYBOAR
D6 MBR36
D7 MBR36
+24V
+24V
+24V
SW5V
SW12
+12Vou
SW16
+16Vou
SW_ON_OFF
SW_ON_OFF
POWER regulators.sc
SW5V
SW12
+12Vou
SW16
+16Vou
SW_ON_OFF
SW_ON_OFF
R23
100R+24V
C8100N/40
GND
+24V
+16Vou
SW16
+12Vou
SW12
SW5V
SW_ON_OFF
SW_ON_OFF
GND
D19
GREER3610K
Neptune ventilator Technical manual 10
1 2 3 4
A
B
C
D
4321
D
C
B
A Title
Number RevisionSize
A4
Date: 6-Jan-2003 Sheet of File: C:\DOCUMENT\..\regulators.sch Drawn By:
FB4
ON/OFF5
GN
D3
IN1 OUT 2U5 MIC4576BT
FB4
ON/OFF5
GN
D3
IN1 OUT 2U6 MIC4576BT
+ C10470µF/63V
+ C11470µF/63V
GND
GND
L1
68µH
L2
68µH
D8
MBR360
D9
MBR360
GND
GND
C131000µF/50V
C14
1000µF/50V
GND
GND
R24
13K
R251K5
R27R4530
R261K5
GND
GND
F2
3.15A/T
F3
3.15A/T
D14
GR
EEN
R151K
GND
GND
D15
GR
EEN
R282K7
GND
GND
C16
3.3N/63V
C17
3.3N/63V
POWER SUPPLY - REGULATORS
2 2JP - KDP
PSU042000V1
SW5V
SW12V
+12Vout
SW16V
+16Vout
SW_ON_OFF1
SW_ON_OFF2
POWER SUPPLY
SW5V
SW12V
+12Vout
SW16V
+16Vout
SW_ON_OFF1
SW_ON_OFF2
1
2
3
4
5
6
P1
GND
GND
GND
GND
+24V
GND
+12V
GND
+5V
+ C12470µF/63V
GND
D10MBR745
L3
68µH
R30
9090
R311500
GND
C151000µF/50V
GND
C19
10N
DC OUTPUT CONNECTOR 1
F5
3.15A/T
GND
D16
GR
EEN
R292K7
+24v
GND
D17
GR
EEN
R3210K
1
2
3
4
P2
GND
DC OUTPUT CONNECTOR 2
+16V
GND
GND
+16V
F4
3.15A/T
+16Vout
SW16V
+12Vout
SW12V
SW5V
SW_ON_OFF1D1820V
GND
Vin5 Vsw 4
GN
D3
Vc
2
FB1
U7 LT1074CT
GNDGND
R33
2K7
GND
Q4
BS170
GND
2
13
Q10RFP30P05
R34100K
R35100K
SW_ON_OFF2
Neptune ventilator Technical manual 11
1.2.4 Power supply board layout
Note: Fan drive output voltage indicated by D16 = 12 Volt
Neptune ventilator Technical manual 12
1.3 BACKPLANE BOARD 1.3.1 Block diagram The block diagram of the backplane board is represented on the next page. The DC voltages coming from the power supply are distributed on the backplane board. Connector P1 is the connection to the power board. Connector P2 and P3 are supplying the extendable system set of electrical power. Connector P5 is the power on/off connection to the power supply (P4). Connector P4 connects to the MMI board through a flatcable. Connector P6 is used to connect the buzzer and the O2 sensor.
Neptune ventilator Technical manual 13
Block diagram backplane board
Neptune ventilator Technical manual 14
1.3.2 Schematic diagram The schematic diagram of the backplane board is represented on the next page.
Neptune ventilator Technical manual 15
Neptune ventilator Technical manual 16
1.3.3 Backplane board layout
Neptune ventilator Technical manual 17
1.4 MMI BOARD 1.4.1 Block diagram The block diagram of the MMI board is represented on the following page. There is only one microprocessor provided on the MMI board. This main microprocessor controls the control knob, the keyboard, the speaker sound, LCD and the communication. It is reprogrammable by means of the programming interface connector P1. All the graphical data is stored in the flash memory. This flash memory is also reprogrammable with the use of connector P1. The RAM memory is used as video memory. The video memory is copied through the LCD data bus to the on-board LCD controller. The LED indicators are located on the top of the MMI board and consist of four LEDs: one green LED and three red LEDs. They give you helpful information when an error occurs. More about errors and malfunctions of the ventilator is described later in this manual. The main microprocessor is communicating with the master board by means of connector P2. The MMI board sends to and receives information from the master board. The master board can transmit to and receive data from the pneumatic board. In this way the MMI board is communication with the pneumatic board.
Neptune ventilator Technical manual 18
Block diagram MMI board
Neptune ventilator Technical manual 19
1.4.2 Schematic diagram The schematic diagram of the MMI board is represented on the following pages.
Neptune ventilator Technical manual 20
1 2 3 4
A
B
C
D
4321
D
C
B
A Title
Number RevisionSize
A4
Date: 24-Mar-2004 Sheet of File: C:\G\..\MMI.sch Drawn By:
Checked
OC1
C11
1D2
2D3
3D4
4D5
5D6
6D7
7D8
8D9
1Q 19
2Q 18
3Q 17
4Q 16
5Q 15
6Q 14
7Q 13
8Q 12
U1
74VHC573
AD0AD1AD2AD3AD4AD5AD6AD7
AD0AD1AD2AD3AD4AD5AD6AD7
ALEALE
GND
A0A1A2A3A4A5A6A7
MEMORY
MEMORY
MMI_MEMORY.sch
MEMORY
A8A9A10A11A12A13A14A15
7SEG - LED
7 SEG-LED BAR
MMI_LED.sch
7SEG - LED
RDWR
SPI
RS485 DRIVER
MMI_SPI.sch
SPI
LP
ADC_CSADC_SCK7SEG_CSLEDBAR_CSDATACLOCKXSCLCONTRAST_CS
PE0PE1
KEY_Y3
KEY_Y1KEY_Y2
PB4MISOMOSI
SS
BUZ_CSBUZ_ON_OFF
PB1
GND GND
VCC
PB7
MMI board
HDH
GNDPF7PF6PF5
VCC
PF4PB7
VCC
PB4PB5PB6
PB6PB5
C112P. C212P.
GND
R1100K
R2100K
R3100K
R4100K
12345
109876
SW1
DS-05 APEM
R51K5
R61K5
R71K5
R81K5
GND
VCC
C3
100N
C4
100N
C5
100N
GND
VCC
PF4PF5PF6PF7
X116MHz
1 5
MMI 092003v1 v1.0
ENC_BENC_A
ADC_DATA
VCC 1
Vref 2
Vin 3
GND 4/CS5
SDO6
SCK7
Fo8U2
LTC2400CS8
VCC1
Vout 2
GND3
U3
MAX6120EUR
R12
4K7
+ C22
2,2µ
/ 10
V T
ant.
C6100N C23
1N
AGND AGND AGNDAGND
AGND
VCC VCC
VCC
1,2V
ADC_DATA
ADC_SCK
ADC_CS
+
-
DIN2 SCLK1
CS3
VCC14
GND5
H 11
W 12
L 13
OUT 7
SHD
N4
IN-
8
IN+
9
VSS
6
VD
D10
U4
MA
X54
37EU
D
R13
18K
+12V
+12VGND
GND
VCC
CLOCKDATABUZ_CS
BUZZER
BUZ_ON_OFF
KB_INT
LCD
LCD
MMI_LCD.sch
LCD
PA0(AD0) 51PA1(AD1) 50PA2(AD2) 49PA3(AD3) 48PA4(AD4) 47PA5(AD5) 46PA6(AD6) 45PA7(AD7) 44
PC0(A8) 35PC1(A9) 36PC2(A10) 37PC3(A11) 38PC4(A12) 39PC5(A13) 40PC6(A14) 41PC7(A15) 42
PD0(INT0)25 PD1(INT1)26 PD2(INT2)27 PD3(INT3)28 PD4(IC1)29 PD530 PD6(T1)31 PD7(T2)32
PB0(ss) 10PB1(SCK) 11PB2(MOSI) 12PB3(MISO) 13PB4(OC0/PWM0) 14PB5(OC1A/PWM1A) 15PB6(OC1B/PWM1B) 16PB7(OC2/PWM2) 17
PE0(PDI/RXD)2 PE1(PDO/TXD)3 PE2(AC+)4 PE3(AC-)5 PE4(INTR4)6 PE5(INTR5)7 PE6(INTR6)8 PE7(INTR7)9
PF0(ADC0)61 PF1(ADC1)60 PF2(ADC2)59 PF3(ADC3)58 PF4(ADC4)57 PF5(ADC5)56 PF6(ADC6)55 PF7(ADC7)54
RD 34
WR 33
XTAL1 24XTAL2 23
ALE 43
TOSC119 TOSC218
RESET20
AREF62
AGND63
AVCC64
PEN 1
VCC52
GND 22GND53
VCC 21
ATmega103L
U6
ATMEGA128-16AI
D1 D2 D3 D4
OC1
CLK11
1D2
2D3
3D4
4D5
5D6
6D7
7D8
8D9
1Q 19
2Q 18
3Q 17
4Q 16
5Q 15
6Q 14
7Q 13
8Q 12
U7
74VHC574
GND
AD0AD1AD2AD3AD4AD5AD6AD7
1 2
U8A
74VHC14
1
23U9A
74VHC32
4
56U9B
74VHC32
3 4
U8B
74VHC14
89
10U9C
74VHC32
1112
13U9D
74VHC32
1
23U10A
74VHC32
5 6
U8C
74VHC14
A15
A14CS_LCD
A15
A14
WR
F_A15F_A16F_A17F_A18R_A15R_A16
4
56U10B
74VHC32
89
10U10C
74VHC32
1112
13U10D
74VHC32
RST
KEY_Y1
KEY_Y2
KEY_Y3 ENC_SWKB_INT
CS_F
CS_R
GND
1
2
Pspare
HEADER 2X1
O2
Neptune ventilator Technical manual 21
1 2 3 4 5 6
A
B
C
D
654321
D
C
B
A
Title
Number RevisionSize
B
Date: 24-Mar-2004 Sheet of File: C:\G\..\MMI_MEMORY.sch Drawn By:
Checked
MEMORY
MMI MEMORYC7100N
C8100N
VCC
HDH2 5
MMI 092003v1 v1.0
A012
A111
A210
A39
A48
A57
A66
A75
A827
A926
A1023
A1125
A124
A1328
A1429
A153
A162
A1730
A181
WR31
OE24
CE22
I/O0 13
I/O1 14
I/O2 15
I/O3 17
I/O4 18
I/O5 19
I/O6 20
I/O7 21
U11
AM29F040B_90EC
4 Mbit FLASH MEMORY 1 Mbit SRAM
A0A1A2A3A4A5A6A7A8A9A10A11A12A13A14F_A15F_A16F_A17F_A18
WRRD
AD0AD1AD2AD3AD4AD5AD6AD7
A0A1A2A3A4A5A6A7A8A9A10A11A12A13A14R_A15R_A16
WRRD
AD0AD1AD2AD3AD4AD5AD6AD7
GND
CS_FCS_R
VCC
A01
A12
A23
A34
A413
A514
A615
A716
A817
A918
A1019
A1120
A1221
A1329
A1430
A1531
A1632
WR12
OE28
CE5
I/O1 6
I/O2 7
I/O3 10
I/O4 11
I/O5 22
I/O6 23
I/O7 26
I/O8 27
U12
HY628100B LLT1-70
Neptune ventilator Technical manual 22
1 2 3 4 5 6
A
B
C
D
654321
D
C
B
A
Title
Number RevisionSize
B
Date: 24-Mar-2004 Sheet of File: C:\G\..\MMI_LED.sch Drawn By:
Checked7 SEGMENT DISPLAY - LED BAR
HDH3 5
MMI 092003v1 v1.0
a16
b15
c3
d2
e1
f18
g17
DP
4
DIG
114
DIG
213
a11
b10
c8
d6
e5
f12
g7
DP
9
D5
a16
b15
c3
d2
e1
f18
g17
DP
4
DIG
114
DIG
213
a11
b10
c8
d6
e5
f12
g7
DP
9
D6
C-5
62G
PA
RA
LIG
HT
a16
b15
c3
d2
e1
f18
g17
DP
4
DIG
114
DIG
213
a11
b10
c8
d6
e5
f12
g7
DP
9
D7
C-5
62G
PA
RA
LIG
HT
a16
b15
c3
d2
e1
f18
g17
DP
4
DIG
114
DIG
213
a11
b10
c8
d6
e5
f12
g7
DP
9
D8
C-5
62G
PA
RA
LIG
HT
a16
b15
c3
d2
e1
f18
g17
DP
4
DIG
114
DIG
213
a11
b10
c8
d6
e5
f12
g7
DP
9
D9
C-5
62G
PA
RA
LIG
HT
a16
b15
c3
d2
e1
f18
g17
DP
4
DIG
114
DIG
213
a11
b10
c8
d6
e5
f12
g7
DP
9
D10
C-5
62G
PA
RA
LIG
HT
a16
b15
c3
d2
e1
f18
g17
DP
4
DIG
114
DIG
213
a11
b10
c8
d6
e5
f12
g7
DP
9
D11
C-5
62G
PA
RA
LIG
HT
Aa
1
Ab
2120 111213141516171819
Ba
2
Bb
22
Ca
3
Cb
23
Da
4
Db
24
Ea5
Eb25
Fa6
Fb26
Ga
7
Gb
27
Ha
8
Hb
28
Ia9
Ib29
Ja10
Jb30
D14
MV
5916
4
Aa
1
Ab
2120 111213141516171819
Ba
2
Bb
22
Ca
3
Cb
23
Da
4
Db
24
Ea5
Eb25
Fa6
Fb26
Ga
7
Gb
27
Ha
8
Hb
28
Ia9
Ib29
Ja10
Jb30
D15
MV
5916
4
Aa
1
Ab
2120 111213141516171819
Ba
2
Bb
22
Ca
3
Cb
23
Da
4
Db
24
Ea5
Eb25
Fa6
Fb26
Ga
7
Gb
27
Ha
8
Hb
28
Ia9
Ib29
Ja10
Jb30
D16
MV
5916
4
Aa
1
Ab
2120 111213141516171819
Ba
2
Bb
22
Ca
3
Cb
23
Da
4
Db
24
Ea5
Eb25
Fa6
Fb26
Ga
7
Gb
27
Ha
8
Hb
28
Ia9
Ib29
Ja10
Jb30
D17
MV
5916
4
Aa
1
Ab
2120 111213141516171819
Ba
2
Bb
22
Ca
3
Cb
23
Da
4
Db
24
Ea5
Eb25
Fa6
Fb26
Ga
7
Gb
27
Ha
8
Hb
28
Ia9
Ib29
Ja10
Jb30
D13
MV
5916
4
Aa
1
Ab
2120 111213141516171819
Ba
2
Bb
22
Ca
3
Cb
23
Da
4
Db
24
Ea5
Eb25
Fa6
Fb26
Ga
7
Gb
27
Ha
8
Hb
28
Ia9
Ib29
Ja10
Jb30
D12
MV
5916
4
SEG
A14
SEG
B16
SEG
C20
SEG
D23
SEG
E21
SEG
F15
SEG
G17
SEG
DP
22
DIG
02
DIG
111
DIG
26
DIG
37
DIG
43
DIG
510
DIG
65
DIG
78
DIN1
CLK13
LOAD(CS)12
GN
D4
GN
D9
V+
19
Iset18
Dout 24
U13
MAX7221CWG
SEG
A14
SEG
B16
SEG
C20
SEG
D23
SEG
E21
SEG
F15
SEG
G17
SEG
DP
22
DIG
02
DIG
111
DIG
26
DIG
37
DIG
43
DIG
510
DIG
65
DIG
78
DIN1
CLK13
LOAD(CS)12
GN
D4
GN
D9
V+
19
Iset18
Dout 24
U14
MAX7221CWG
GNDGND GNDGND GNDGND GNDGND
DA
TA
CLO
CK
7SEG
_CS
R1412K
R1512K
R1612K
R1712K
VCC VCC VCC VCC
DA
TA
1_SE
GA
1_SE
GB
1_SE
GC
1_SE
GD
1_SE
GE
1_SE
GF
1_SE
GG
1_SE
GD
P
1_D
IG0
1_D
IG1
1_D
IG2
1_D
IG3
1_D
IG4
1_D
IG5
1_D
IG6
1_D
IG7
2_SE
GA
2_SE
GB
2_SE
GC
2_SE
GD
2_SE
GE
2_SE
GF
2_SE
GG
2_SE
GD
P
2_D
IG0
2_D
IG1
2_D
IG2
2_D
IG3
2_D
IG4
2_D
IG5
2_D
IG6
2_D
IG7
3_SE
GA
3_SE
GB
3_SE
GC
3_SE
GD
3_SE
GE
3_SE
GF
3_SE
GG
3_SE
GD
P
3_D
IG0
3_D
IG1
3_D
IG2
3_D
IG3
3_D
IG4
3_D
IG5
3_D
IG6
3_D
IG7
4_SE
GA
4_SE
GB
4_SE
GC
4_SE
GD
4_SE
GE
4_SE
GF
4_SE
GG
4_SE
GD
P
4_D
IG0
4_D
IG1
4_D
IG2
4_D
IG3
4_D
IG4
4_D
IG5
4_D
IG6
4_D
IG7
LEDB
AR
_CS
1_SE
GA
1_SE
GB
1_SE
GC
1_SE
GD
1_SE
GE
1_SE
GF
1_SE
GG
1_SE
GD
P
1_SE
GA
1_SE
GB
1_SE
GC
1_SE
GD
1_SE
GE
1_SE
GF
1_SE
GG
1_SE
GD
P
1_SE
GA
1_SE
GB
1_SE
GC
1_SE
GD
1_SE
GE
1_SE
GF
1_SE
GG
1_SE
GD
P
1_SE
GA
1_SE
GB
1_SE
GC
1_SE
GD
1_SE
GE
1_SE
GF
1_SE
GG
1_SE
GD
P
2_SE
GA
2_SE
GB
2_SE
GC
2_SE
GD
2_SE
GE
2_SE
GF
2_SE
GG
2_SE
GD
P
2_SE
GA
2_SE
GB
2_SE
GC
2_SE
GD
2_SE
GE
2_SE
GF
2_SE
GG
2_SE
GD
P
2_SE
GA
2_SE
GB
2_SE
GC
2_SE
GD
2_SE
GE
2_SE
GF
2_SE
GG
2_SE
GD
P
2_SE
GA
2_SE
GB
2_SE
GC
2_SE
GD
2_SE
GE
2_SE
GF
2_SE
GG
2_SE
GD
P
2_SE
GA
2_SE
GB
2_SE
GC
2_SE
GD
2_SE
GE
2_SE
GF
2_SE
GG
2_SE
GD
P
2_SE
GA
2_SE
GB
2_SE
GC
2_SE
GD
2_SE
GE
2_SE
GF
2_SE
GG
2_SE
GD
P
2_SE
GA
2_SE
GB
2_SE
GC
2_SE
GD
2_SE
GE
2_SE
GF
2_SE
GG
2_SE
GD
P
2_SE
GA
2_SE
GB
2_SE
GC
2_SE
GD
2_SE
GE
2_SE
GF
2_SE
GG
2_SE
GD
P
1_SE
GA
1_SE
GB
1_SE
GC
1_SE
GD
1_SE
GE
1_SE
GF
1_SE
GG
1_SE
GD
P
1_SE
GA
1_SE
GB
1_SE
GC
1_SE
GD
1_SE
GE
1_SE
GF
1_SE
GG
1_SE
GD
P
1_D
IG0
1_D
IG1
1_D
IG2
1_D
IG3
2_D
IG0
2_D
IG1
2_D
IG2
2_D
IG3
2_D
IG4
2_D
IG5
2_D
IG6
2_D
IG7
1_D
IG4
1_D
IG5
TIDAL VOLUMEMINUTE VOLUME
PEAK PLATEAU MEAN PEEP O2
3_D
IG0
3_D
IG1
3_D
IG2
3_D
IG3
3_D
IG4
3_D
IG5
3_D
IG6
3_D
IG7
4_D
IG0
4_D
IG1
4_D
IG2
4_D
IG3
4_D
IG4
4_D
IG5
4_D
IG6
3_SE
GA
3_SE
GB
3_SE
GC
3_SE
GD
3_SE
GE
3_SE
GF
3_SE
GG
3_SE
GD
P3_
SEG
A3_
SEG
B3_
SEG
C3_
SEG
D3_
SEG
E3_
SEG
F3_
SEG
G3_
SEG
DP
3_SE
GA
3_SE
GB
3_SE
GC
3_SE
GD
3_SE
GE
3_SE
GF
3_SE
GG
3_SE
GD
P3_
SEG
A3_
SEG
B3_
SEG
C3_
SEG
D3_
SEG
E3_
SEG
F3_
SEG
G3_
SEG
DP
3_SE
GA
3_SE
GB
3_SE
GC
3_SE
GD
3_SE
GE
3_SE
GF
3_SE
GG
3_SE
GD
P
3_SE
GA
3_SE
GB
3_SE
GC
3_SE
GD
3_SE
GE
3_SE
GF
3_SE
GG
3_SE
GD
P3_
SEG
A3_
SEG
B3_
SEG
C3_
SEG
D3_
SEG
E3_
SEG
F3_
SEG
G3_
SEG
DP
3_SE
GA
3_SE
GB
3_SE
GC
3_SE
GD
3_SE
GE
3_SE
GF
3_SE
GG
3_SE
GD
P4_
SEG
A4_
SEG
B4_
SEG
C4_
SEG
D4_
SEG
E4_
SEG
F4_
SEG
G4_
SEG
DP
4_SE
GA
4_SE
GB
4_SE
GC
4_SE
GD
4_SE
GE
4_SE
GF
4_SE
GG
4_SE
GD
P
4_SE
GA
4_SE
GB
4_SE
GC
4_SE
GD
4_SE
GE
4_SE
GF
4_SE
GG
4_SE
GD
P4_
SEG
A4_
SEG
B4_
SEG
C4_
SEG
D4_
SEG
E4_
SEG
F4_
SEG
G4_
SEG
DP
4_SE
GA
4_SE
GB
4_SE
GC
4_SE
GD
4_SE
GE
4_SE
GF
4_SE
GG
4_SE
GD
P4_
SEG
A4_
SEG
B4_
SEG
C4_
SEG
D4_
SEG
E4_
SEG
F4_
SEG
G4_
SEG
DP
4_SE
GA
4_SE
GB
4_SE
GC
4_SE
GD
4_SE
GE
4_SE
GF
4_SE
GG
4_SE
GD
P
7SEG - LED
C9100N
C10100N
C11100N
GND
VCC
C12100N
SEG
A14
SEG
B16
SEG
C20
SEG
D23
SEG
E21
SEG
F15
SEG
G17
SEG
DP
22
DIG
02
DIG
111
DIG
26
DIG
37
DIG
43
DIG
510
DIG
65
DIG
78
DIN1
CLK13
LOAD(CS)12
GN
D4
GN
D9
V+
19
Iset18
Dout 24
U15
MAX7221CWG
SEG
A14
SEG
B16
SEG
C20
SEG
D23
SEG
E21
SEG
F15
SEG
G17
SEG
DP
22
DIG
02
DIG
111
DIG
26
DIG
37
DIG
43
DIG
510
DIG
65
DIG
78
DIN1
CLK13
LOAD(CS)12
GN
D4
GN
D9
V+
19
Iset18
Dout 24
U16
MAX7221CWG
+ C100
10µ / 10V Tant.
+ C101
10µ / 10V Tant.
+ C102
10µ / 10V Tant.
+ C103
10µ / 10V Tant.
Neptune ventilator Technical manual 23
1 2 3 4 5 6
A
B
C
D
654321
D
C
B
A
Title
Number RevisionSize
B
Date: 24-Mar-2004 Sheet of File: C:\G\..\MMI_SPI.sch Drawn By:
Checked
DE1/RE11
DE2/RE24
DE3/RE35
DR12
DR23
DR36
RO47
DI48
DE49
RE410 GND 11OB4 12OA4 13OB3 14OA3 15
OB2 17OA2 18OB1 19OA1 20U17
DS36954
SS
SS
MISOMISOMOSI
SCK
MISOAMISOBMOSIAMOSIB
SCKASCKBSSSLV1ASSSLV1B
SPI
GND
RST
PB1
PE1
PE0
PSCK
PMISO
PMOSI
SCK
KEY_X2
KEY_X1
GND
VCC 4
MR 3RST2
GND1U18
MAX811L
GND
VCC
1 23 45 67 89 10
P1
HEADER 5X2 HAAKS
VCC
GNDPSCKPMISO
PMOSI
GND
RST
MMI SPI
C13100N
C14100N
C15100N
GND
VCC
MOSIAMISOASCKASSSLV1A
MOSIBMISOBSCKBSSSLV1B
AGND
GND
R18
5K6R19
5K6R20
5K6R21
5K6
MOSIA
MISOA
SCKA
SSSLV1A
MOSIB
MISOB
SCKB
SSSLV1B
+12V VCC
GND
C16
100N
+ C244µ7
/ 16V
Tan
t.
+ C25
4µ7
/ 16V
Tan
t.
C17
100N
GND
AX12
AY13
BX2
BY1
CX5
CY3
A11
B10
C9
INH6
O/IA 14
O/IB 15
O/IC 4
U19
74HC4053
HDH4 5
MMI 092003v1 v1.0
1 23 45 67 89 1011 1213 1415 1617 1819 20
P2
HEADER 10X2
BUZZER +24V
O2
GND
PS_ON_OFF
VCC+12V123456789
10111213141516171819
P3
R22
1K8
GND
R243K3
D18
ZEN
ER 2
0V 0
,4W
+24V
PS_ON_OFF
1_SEGA1_SEGB1_SEGC1_SEGD1_SEGE1_SEGF1_SEGG1_SEGDP1_DIG61_DIG7
R9
1K5
R101K5
GNDGND
KEY_X1KEY_X2KEY_Y1KEY_Y2KEY_Y3
GND
CO
NN
ECT
TO K
EYB
OA
RD
R253K3
R263K3
VCC VCC
ENC
_SW
ENC
_A
ENC
_B
Q1
MM
BF1
70
R231K8
9 8
U8D
74AHC14
1110
U8E
74VHC14
13 12
U8F
74VHC14
R11
1K5
GND
R273K3
VCC
GNDGND
CY1CY2CY3
GND GNDGND
VCC
1 23 45 67 89 10
P4
HEADER 5X2
Neptune ventilator Technical manual 24
1 2 3 4 5 6
A
B
C
D
654321
D
C
B
A
Title
Number RevisionSize
B
Date: 24-Mar-2004 Sheet of File: C:\G\..\MMI_LCD.sch Drawn By:
CheckedLCD INTERFACE
HDH5 5
MMI 092003v1 v1.0
D059
D160
D21
D32
D43
D54
D6
5
D7
6
OSC
154
OSC
255
VA
043
VA
142
VA
241
VA
340
VA
439
VA
538
VA
637
VA
736
VA
835
VA
934
VA
1033
VA
1132
VA
1231
VA
1330
VA
1428
VA
1527
VD0 26
VD1 25
VD2 24
VD3 23
VD4 22
VD5 21
VD6 20
VD7 19
YSCL 18
YD 17
YDIS 16
WF 15
LP 14
XSCL 12
XECL 11
VRW44
VCE45
RES47
RD50
WR51
SEL252
SEL153
CS56
A057
XD
010
XD
19
XD
28
XD
37
NC
29
NC
46
NC
48
NC
49
U20
S1D13305F00A1
A010
A19
A28
A37
A46
A55
A64
A73
A825
A924
A1021
A1123
A122
A1326
WR27
CE20 OE22
A141
D0 11
D1 12
D2 13
D3 15
D4 16
D5 17
D6 18
D7 19
256
U21
BS62LV256SC-70
VA0VA1VA2VA3VA4VA5VA6VA7VA8VA9VA10VA11VA12VA13VA14
VD0VD1VD2VD3VD4VD5VD6VD7
VA
0V
A1
VA
2V
A3
VA
4V
A5
VA
6V
A7
VA
8V
A9
VA
10V
A11
VA
12V
A13
VA
14
VD0VD1VD2VD3VD4VD5VD6VD7
AD0AD1AD2AD3AD4AD5
AD
6A
D7
A0
GND
GND
RST
WRRD
CS_LCD
X2
10 M
Hz
C26
10P.
C27
10P.
GND GND
XD
0X
D1
XD
2X
D3
YD
LPXSCL
C18
100N
C19
100N
GND GND
VCC VCC
VA15
VA
15
C20
100N
GND
VCC
VWR
VW
R
123456789
1011121314
P5XD0XD1XD2XD3
VCC
GND
YD
LPXSCL
CO
NN
ECT
TO L
CD
DIS
PLA
Y
POL2
ISET5
SHDN1
VC
C8
LX 7
FB 4
REF 3
GN
D6
U22
MAX629ESA
VCC
C21
100N+C28
10µ
/ 35V
Tan
t. Lo
w E
SR
GND GND
L147µH
C30
2.2µF / 40VD19
MB
R05
40T1
D20
MBR0540T1
+
C2910µ / 35V Tant. Low ESRR28
120K
R306k8
C31
100p
F / 4
0 V
C322.2µF / 10V
GND GND
GNDGND
-22V
+
-
DIN2 SCLK1
CS3
VCC14
GND5
H 11
W 12
L 13
OUT 7
SHD
N4
IN-
8
IN+
9
VSS
6
VD
D10
U5
MA
X54
37EU
D
VCC
GND
R31
270K
VCC
R29
68K
D0D1
D3D2
/DOFFFLMN.C.CL1CL2VDD
VSSVo
VSSVEEVo
1
2
P6
CONNECT TO LED BACKLIGHTVout 1
GND 2
GND 3
NC 4NC5
GND6
GND7
Vin8U23
LM78L05ACM
+12V
R32
82E
GND
R33
2E
DATACONTRAST_CS
CLOCK
1 2STR1
VCC
12
STR2
Neptune ventilator Technical manual 25
1.4.3 MMI board layout
Neptune ventilator Technical manual 26
Neptune ventilator Technical manual 27
1.4.4 Built-in test software
• Start-up test After turning the ventilator on, the MMI board will test the different functions.
During start-up the following tests are executed:
o Microprocessor internal functions tests: test internal program memory test internal SRAM memory test the timers test internal EEPROM memory test Watchdog
o Microprocessor external functions tests:
test external SRAM memory test external flash memory test LCD driver with Medec logo test keyboard for shorts buzzer test
Neptune ventilator Technical manual 28
Visual test of the leds and the displays on the keyboard panel. All segments and leds are on.
Visual test of the barograph
The green leds light up from left to right and the red leds from right to left.
o Microprocessor communication test: Tests the communication between MMI board and master
board.
When all these tests are executed successfully, the MMI board is ready to start working. If an error occurred, the type of error is displayed on the LED indicators.
G R R R
Neptune ventilator Technical manual 29
After start-up, the LED indicators can show the following errors:
LED indicators Type error
G R R R
Device is working properly
G R R R
Microprocessor internal function error
G R R R
Microprocessor external function error
G R R R
Communication error
Note: During error free operation of the ventilator, the LED indicators might be turned on and off by the software. This is quite normal and has nothing to do with error codes. To determine the exact error on the MMI board, you have to perform a specific test by using the build-in test software. Note: When executing a specific test on the MMI board, there will be no communication between the master board and the MMI board. The master board will interprete this as a communication error and a continuous beep is audible. This can be very annoying. To prevent this error, put the master board in internal test mode by setting dipswitch 1 high. For detailed instructions, see the master board section later in this manual.
Neptune ventilator Technical manual 30
• Microprocessor internal function test If a microprocessor internal error is detected, execute an internal function test as follows:
o Turn the Neptune ventilator off o Remove the upper cover plate at the back of the Neptune o Change the dipswitch setting of the MMI board with a small
screwdriver to
1 2 3 4 5 o Turn the ventilator on o The graphics display continuously shows the Medec logo
o Display changes from normal to inverted to flashing o If no errors where found, all the red LEDs are off and the green LED is
blinking o If an error is detected, the error is indicated on the red error LEDs and
the green LED is off. The possible errors are illustrated in the table on the next page:
Neptune ventilator Technical manual 31
INTERNAL FUNCTION TEST ERROR CODES Error LEDs Error code Description
G R R R
0
NO ERRORS detected, test loop is running
G R R R
1
Internal PROGRAM MEMORY error
G R R R
2
Internal SRAM error
G R R R
3
Internal TIMER error
G R R R
4
Internal EEPROM error
G R R R
5
Internal WATCHDOG error
G R R R
6
Reserved
G R R R
7
Reserved
Neptune ventilator Technical manual 32
Resolve internal function test error code 4 - Internal EEPROM error To fix error code 4 or internal EEPROM error, follow the next steps:
o Turn the Neptune ventilator off o Change the dipswitch setting of the MMI board with a small
screwdriver to
1 2 3 4 5 o Turn the ventilator on o You will notice that all the LED indicators are blinking fast. This
means that the MMI board EEPROM settings are restored to the factory settings. You need to check the values in the setup menu afterwards
o Turn the ventilator off o Repeat the microprocessor internal functions test described on the
previous page. If the EEPROM error still remains, contact an authorized service engineer
Note: Restoring the MMI EEPROM settings will not affect the pneumatic board EEPROM settings. Calibration of the ventilator is not needed in this case. Just check the power-on parameters and the speaker sound volume in the setup menu.
Warning: Try to resolve internal function error code 4 (EEPROM error) with the instructions described above. For all other internal function test errors, contact an authorized service engineer.
Neptune ventilator Technical manual 33
• Microprocessor external function test If a microprocessor external error is detected, execute an external function test as follows:
o Turn the Neptune ventilator off o Remove the upper cover plate at the back of the Neptune o Change the dipswitch setting of the MMI board with a small
screwdriver to
1 2 3 4 5 o Turn the ventilator on o The graphics display continuously shows the Medec logo o The Medec logo changes from time to time from normal to inverted to
flashing
o If no errors where found, all the red LEDs are off and the green LED is blinking. Some devices need to be visually checked for failures.
o Visual test for the barograph: check that all red leds light up from right to left and the green leds from left to right.
Neptune ventilator Technical manual 34
o 7 segment displays: all segments light up once. When a digit is done, the Dp segment is lit.
After this test, a counter from 0 to F is displayed on all digits.
Neptune ventilator Technical manual 35
o Leds on the keyboard panel: the test starts with all leds off. Each led is lit up once. Battery => Tidal volume => Minute volume => O2 High => O2 Low => CMV => PCV => Manual => Standby => High pressure Alarm => Low pressure alarm => Apnea alarm => Trigger => LCD led.
After this test, the leds light up in the following sequence: 1° Standby (1) 2° Standby (1), High pressure (2) 3° Standby (1), High pressure (2), Low pressure (3) 4° Standby (1), High pressure (2), Low pressure (3), Apnea alarm (4) 5° Standby (1), High pressure (2), Low pressure (3), Apnea alarm (4), Trigger (5) 6° Standby (1), High pressure (2), Low pressure (3), Apnea alarm (4), Trigger (5) and LCD led (6)
o LCD contrast
In the FiO2 display appears a counter from 1 to 5. The contrast of the LCD changes with each value.
1
234 5
6
Neptune ventilator Technical manual 36
o Alarm buzzer volume In the FiO2 display appears a counter from 1 to 5. The audio volume of the buzzer changes with each value.
o O2 sensor
In the FiO2 display appears O2.
When no sensor attached, the value should be around 7FFFF When shorted, the value should be around 000000 A good sensor, should have a value around 3F8000
Neptune ventilator Technical manual 37
o If an error is detected, the error is indicated on the red error LEDs and the green LED is off. The possible errors are illustrated in the table below.
EXTERNAL FUNCTION TEST ERROR CODES Error LEDs Error code Description
G R R R
0
NO ERRORS detected, test running
G R R R
1
External FLASH/ROM error
G R R R
2
External SRAM error
G R R R
3
LCD DISPLAY DRIVER error
G R R R
4
Reserved
G R R R
5
Reserved
G R R R
6
Reserved
G R R R
7
Reserved
Note: The external function test is running continuously. When an error is detected, the error is indicated on the LED indicators, and the external function test is repeated. It’s not necessary to turn the ventilator on and off to repeat the external function test.
Neptune ventilator Technical manual 38
Resolve external function test error code 1 - External flash error There is a problem with the external flash memory. The microprocessor is only reading data from the flash memory for display on the graphics display. Contact an authorized service engineer if the problem cannot be resolved.
Resolve external function test error code 2 - External SRAM error There is a problem with the external SRAM memory. The microprocessor is reading data from and writing data to the external SRAM memory. Contact an authorized service engineer if the problem cannot be resolved.
Resolve external function test error code 3 - External display driver error There is a problem with the external display driver. It’s not necessary to connect the graphics display with the MMI board to perform an external display driver test. The microprocessor is writing data to and reading data from the display driver. Contact an authorized service engineer if the problem cannot be resolved.
Resolve external function test error code 4 - Keyboard error There is a problem with the keyboard. To define which error is detected on the keyboard controller, you can perform a specific keyboard controller test:
o Turn the Neptune ventilator off o Remove the upper cover plate at the back of the Neptune o Change the dipswitch setting of the MMI board with a small
screwdriver to
1 2 3 4 5 o Turn the ventilator on o The graphics display shows the Medec logo and all leds are off
Neptune ventilator Technical manual 39
o Press the volume mode key (1).
The leds tidal volume (2) and minute volume (3) should come on. Release the key. The leds should be off.
o Press the CMV key (5).
The led CMV (4) should be on. Release the key. The leds should be off.
o Press the PCV key (7).
The led PCV (6) should be on. Release the key. The leds should be off.
o Press the Manual key (9).
The led Manual (8) should be on. Release the key. The leds should be off.
o Press the Standby key (11).
The led Standby (10) should be on. Release the key. The leds should be off.
o Press the Silent key (13).
The led LCD (12) should be on. Release the key. The leds should be off.
o Press the rotary knob (14).
The led battery (15) should be on. Release the key. The leds should be off.
12
3
4
10
11
12
13
5 6 7 8 9
14
15
Neptune ventilator Technical manual 40
• Microprocessor communication test To perform a microprocessor communication test, perform the following test:
o Turn the Neptune ventilator off o Remove the upper cover plate at the back of the Neptune o Change the dipswitch setting of the MMI board with a small
screwdriver to
o Change the dipswitch setting of the master board with a small
screwdriver to
o Turn the ventilator on. o If no errors found, the green LED is blinking. If an error is detected,
the error is indicated on the red error LED’s and the green LED is off. The red LED’s indicating the numbers of error’s detected (binary number between 1 and 7).
• LED indicator test To test the error LED indicators, perform the following test:
o Turn the Neptune ventilator off o Remove the upper cover plate at the back of the Neptune o Change the dipswitch setting of the MMI board with a small
screwdriver to
1 2 3 4 5 o Turn the ventilator on o The graphics display continuously shows the Medec logo o The error LEDs are blinking one by one
Neptune ventilator Technical manual 41
• Showing a test pattern on the graphics display To test the display driver and/or graphics display, you can show a test pattern on the graphics display. Perform the following instructions:
o Turn the Neptune ventilator off o Remove the upper cover plate at the back of the Neptune o Change the dipswitch setting of the MMI board with a small
screwdriver to
1 2 3 4 5 o Turn the ventilator on o The graphics display shows a test pattern
Neptune ventilator Technical manual 42
• Alarm buzzer test To test the buzzer, perform the following test:
o Turn the Neptune ventilator off o Remove the upper cover plate at the back of the Neptune o Change the dipswitch setting of the MMI board with a small
screwdriver to
1 2 3 4 5 o Turn the ventilator on o The graphics display continuously shows the Medec logo o In the FiO2 display (1) appears a counter from 1 to 5
The audio volume of the buzzer changes with each value
Neptune ventilator Technical manual 43
• LCD contrast test To test the buzzer, perform the following test:
o Turn the Neptune ventilator off o Remove the upper cover plate at the back of the Neptune o Change the dipswitch setting of the MMI board with a small
screwdriver to
1 2 3 4 5 o Turn the ventilator on o The graphics display continuously shows the Medec logo o In the FiO2 display (1) appears a counter from 1 to 5
The LCD contrast changes with each value
Neptune ventilator Technical manual 44
• Barograph test To test the barograph, perform the following test:
o Turn the Neptune ventilator off o Remove the upper cover plate at the back of the Neptune o Change the dipswitch setting of the MMI board with a small
screwdriver to
1 2 3 4 5 o Turn the ventilator on o The graphics display continuously shows the Medec logo o First test: the green leds light up from left to right in the airway
pressure window.
o Second test: the red leds light up from left to right in the airway
pressure window.
o Third test: both green and red leds light up from left to right in the
airway pressure window.
Neptune ventilator Technical manual 45
• Keyboard test To test the barograph, perform the following test:
o Turn the Neptune ventilator off o Remove the upper cover plate at the back of the Neptune o Change the dipswitch setting of the MMI board with a small
screwdriver to
1 2 3 4 5 o Turn the ventilator on o The graphics display continuously shows the Medec logo
o Press the volume mode key (1).
The leds tidal volume (2) and minute volume (3) should come on. Release the key. The leds should be off.
o Press the CMV key (5).
The led CMV (4) should be on. Release the key. The leds should be off.
o Press the PCV key (7).
The led PCV (6) should be on. Release the key. The leds should be off.
12
3
4
10
11
12
13
5 6 7 8 9
14
15
Neptune ventilator Technical manual 46
o Press the Manual key (9). The led Manual (8) should be on. Release the key. The leds should be off.
o Press the Standby key (11).
The led Standby (10) should be on. Release the key. The leds should be off.
o Press the Silent key (13).
The led LCD (12) should be on. Release the key. The leds should be off.
o Press the rotary knob (14).
The led battery (15) should be on. Release the key. The leds should be off.
Neptune ventilator Technical manual 47
• Keyboard leds test To test the keyboard leds, perform the following test:
o Turn the Neptune ventilator off o Remove the upper cover plate at the back of the Neptune o Change the dipswitch setting of the MMI board with a small
screwdriver to
1 2 3 4 5 o Turn the ventilator on o The graphics display continuously shows the Medec logo o First test: the leds light up one by one
Neptune ventilator Technical manual 48
o Second test: the leds light up in the following way: 1° Standby (1) 2° Standby (1), High pressure (2) 3° Standby (1), High pressure (2), Low pressure (3) 4° Standby (1), High pressure (2), Low pressure (3), Apnea alarm (4) 5° Standby (1), High pressure (2), Low pressure (3), Apnea alarm (4), Trigger (5) 6° Standby (1), High pressure (2), Low pressure (3), Apnea alarm (4), Trigger (5) and LCD led (6)
1
234 5
6
Neptune ventilator Technical manual 49
• 7 segment display test To test the 7 segment displays, perform the following test:
o Turn the Neptune ventilator off o Remove the upper cover plate at the back of the Neptune o Change the dipswitch setting of the MMI board with a small
screwdriver to
1 2 3 4 5 o Turn the ventilator on o The graphics display continuously shows the Medec logo o First test: each segment lights up once. When a digit is done, the
decimal point segment is on.
Neptune ventilator Technical manual 50
o Second test: on each display a counter from 0 to F is displayed
Neptune ventilator Technical manual 51
• O2 sensor test
To test the 7 segment displays, perform the following test:
o Turn the Neptune ventilator off o Remove the upper cover plate at the back of the Neptune o Change the dipswitch setting of the MMI board with a small
screwdriver to
1 2 3 4 5 o Turn the ventilator on o The graphics display continuously shows the Medec logo
In the FiO2 display appears O2.
When no sensor attached, the value should be around 7FFFF When shorted, the value should be around 000000 A good sensor, should have a value around 3F8000
Neptune ventilator Technical manual 52
• Encoder test To test the encoder, perform the following test:
o Turn the Neptune ventilator off o Remove the upper cover plate at the back of the Neptune o Change the dipswitch setting of the MMI board with a small
screwdriver to
1 2 3 4 5 o Turn the ventilator on o The graphics display continuously shows the Medec logo o The displays are blank
Turn the knob to the right. In the peep display appears a number. This number increments with each click of the encoder. Turn the knob to the right. This number decrements with each click of the encoder
Note: Make sure that in normal operating mode of the Neptune ventilator the dipswitches of the MMI board, master board and pneumatic board are put in the following position:
1 2 3 4 5
Neptune ventilator Technical manual 53
1.4.5 Graphics display The LCD display is quarter-VGA display. It has a contrast adjustment. The graphics display excels in a wide range of ambient lighting environments while effectively eliminating the blooming. The display consists of a cristal panel and control electronics. Note: The display generates voltages capable of causing personal injury (high voltage up to -24 VDC). Do not touch the display electronics during operation. The display is connected to connector P5 and P6 of the MMI board. After switching the ventilator on, the display will respond after a few seconds.
Neptune ventilator Technical manual 54
1.5 O2 MEASUREMENT 1.5.1 General The O2 measurement is integrated into the MMI board. 1.5.2 Operation The oxygen measurement is done with an O2 fuel cell. The oxygen sensor function is similar to a typical battery. The O2 fuel cell generates a small voltage that is linear with the oxygen concentration. The output voltage of a new O2 fuel cell is about 15,0 ± 2 mV at dry ambient air. The voltage is converted to a digital value through the 24-bit ADC converter. U2 generates a reference voltage equal to 1,2V used by the ADC converter U1. LED D1 is lit when the ADC converter is working properly.
• O2 sensor test To test the O2 sensor, perform the following test:
o Turn the Neptune ventilator off o Remove the upper cover plate at the back of the Neptune o Change the dipswitch setting of the MMI board with a small
screwdriver to
1 2 3 4 5 o Turn the ventilator on o The graphics display continuously shows the Medec logo o The display fields plateau, mean and peep are showing a value
When no sensor attached, the value should be around 7FFFF When shorted, the value should be around 000000 A good sensor should have a value around 3F8000
Neptune ventilator Technical manual 55
1.5.3 O2 Measurement specifications
• Ambient air calibration
Minimum sensor input voltage: ± 4,3 mV
Maximum sensor input voltage: ± 230 mV
• 100% O2 calibration
Minimum sensor input voltage: Ambient air cal. voltage x 4
Maximum sensor input voltage: 1200 mV
• Recommended calibration
Calibrate every 24 hours with ambient air and/or 100% O2.
Neptune ventilator Technical manual 56
1.6 MASTER BOARD 1.6.1 Block diagram The block diagram of the master board is represented on the following page. The master board looks after the communication between the MMI board and the pneumatic board. The master board contains a reprogrammable microprocessor (reprogrammable by means of connector P3). The real time clock keeps up the actual time and date. It contains also timer information like the total working time of the ventilator, service time, etc. The audible safety circuit contains a buzzer. The buzzer is driven by the microprocessor and the reset signal. If an error is detected, the buzzer generates an audible sound. The master board is also provided with a RS-232 interface.
Neptune ventilator Technical manual 57
microprocessor
Rammemory
Programminginterface
(connector P3)
SPI communication with other boards(connector P2)
LED indicators
Master board
Real time clock
RS-232interface
RS-232 interface(connector P4)
audible safety circuit
Block diagram master board
Neptune ventilator Technical manual 58
1.6.2 Schematic diagram The schematic diagram of the master board is represented on the following pages.
Neptune ventilator Technical manual 59
1 2 3 4
A
B
C
D
4321
D
C
B
A Title
Number RevisionSize
A4
Date: 15-Jan-2003 Sheet of File: C:\DOCUMENT\..\master.sch Drawn By:
PA0(AD0) 51PA1(AD1) 50PA2(AD2) 49PA3(AD3) 48PA4(AD4) 47PA5(AD5) 46PA6(AD6) 45PA7(AD7) 44
PC0(A8) 35PC1(A9) 36PC2(A10) 37PC3(A11) 38PC4(A12) 39PC5(A13) 40PC6(A14) 41PC7(A15) 42
PD0(INT0)25 PD1(INT1)26 PD2(INT2)27 PD3(INT3)28 PD4(IC1)29 PD530 PD6(T1)31 PD7(T2)32
PB0(ss) 10PB1(SCK) 11PB2(MOSI) 12PB3(MISO) 13PB4(OC0/PWM0) 14PB5(OC1A/PWM1A) 15PB6(OC1B/PWM1B) 16PB7(OC2/PWM2) 17
PE0(PDI/RXD)2 PE1(PDO/TXD)3 PE2(AC+)4 PE3(AC-)5 PE4(INTR4)6 PE5(INTR5)7 PE6(INTR6)8 PE7(INTR7)9
PF0(ADC0)61 PF1(ADC1)60 PF2(ADC2)59 PF3(ADC3)58 PF4(ADC4)57 PF5(ADC5)56 PF6(ADC6)55 PF7(ADC7)54
RD 34
WR 33
XTAL1 24XTAL2 23
ALE 43
TOSC119 TOSC218
RESET20
AREF62
AGND63
AVCC64
PEN 1
VCC52
GND 22GND53
VCC 21ATmega103L
U1
103L
OC1
C11
1D2
2D3
3D4
4D5
5D6
6D7
7D8
8D9
1Q 19
2Q 18
3Q 17
4Q 16
5Q 15
6Q 14
7Q 13
8Q 12
U2
74HC573
AD0AD1AD2AD3AD4AD5AD6AD7
AD0AD1AD2AD3AD4AD5AD6AD7
ALE
ALE
GND
A0A1A2A3A4A5A6A7
A8A9A10A11A12A13A14A15
SPI BUS
SPIBUS.sch
RDWR
A1
B2
C3
E14
E25
E36
Y0 15
Y1 14
Y2 13
Y3 12
Y4 11
Y5 10
Y6 9
Y7 7
U3
74HC138
A15
CS0CS1CS2CS3CS4CS5CS6CS7
A1
B2
C3
E14
E25
E36
Y0 15
Y1 14
Y2 13
Y3 12
Y4 11
Y5 10
Y6 9
Y7 7
U4
74HC138
SS0SS1SS2SS3SS4SS5SS6SS7
SS
SS
MOSIMISO
PB1
PE0PE1PE2PE3PE4PE5PE6PE7
GND
GND
PF0PF1PF2PF3PF4PF5PF6PF7
A010
A19
A28
A37
A46
A55
A64
A73
A825
A924
A1021
A1123
A122
A1326
WR27
CE20 OE22
A141
D0 11
D1 12
D2 13
D3 15
D4 16
D5 17
D6 18
D7 19
256
U6
RAM/ROM/FLASH
AD0AD1AD2AD3AD4AD5AD6AD7
A0A1A2A3A4A5A6A7A8A9A10A11A12A13
A010
A19
A28
A37
A46
A55
A64
A73
A825
A924
A1021
A1123
A122
A1326
WR27
CE20 OE22
A141
D0 11
D1 12
D2 13
D3 15
D4 16
D5 17
D6 18
D7 19
256
U7
RAM/ROM/FLASH
AD0AD1AD2AD3AD4AD5AD6AD7
A0A1A2A3A4A5A6A7A8A9A10A11A12A13
01 S1
01 S2
01 S3
01 S4
A14
WR A14
WR
CS0RD RD
CS1
01 S5
01 S6
X1 4MHZ
GND GND
VCC
PWR1
X12
X23
AD04
AD15
AD26
AD37
AD48
AD59
AD610
AD711
GND12 CS 13ALE 14WR 15GND 16RD 17KS 18IRQ 19Vbat 20RCLR 21Vbaux 22SQW 23VCC 24U8
DS1685
AD0AD1AD2AD3AD4AD5AD6AD7
GND
BAT1
LITHIUM 3V
D1
HLM
P-17
90
R11K5
PB7
VCC
MASTER
JP-KDP
R5
100K
R6
100K
R7
100K
R8
100K
VCC
PF7PF6PF5
R9100K
R143K9
GND
R10100K
R1510K
GND
+12V
+
C3
10µF
GND
+24V
+
C410µFGND
PF0
PF1
GND
+C510µF
C6100nF
GND
VCC
GND
PB7VCC
D2
HLM
P-17
00
R21K5
VCC
D3
HLM
P-17
00
R31K5
VCC
D4
HLM
P-17
00
R41K5
VCC
PB4PB5PB6
PB4PB5PB6
PF4 12345
109876
SW1
DS-05 APEM
A1
B2
C3
E14
E25
E36
Y0 15
Y1 14
Y2 13
Y3 12
Y4 11
Y5 10
Y6 9
Y7 7
U5
74HC138
A15A14
A12
A8A9A10
CS8CS9CS10CS11CS12CS13CS14CS15
CS8
RD
WRALE
GND
R11
100K
VCC
VCC
R12
100K
VCC
PD0PD1PD2PD3PD4PD5PD6PD7
RST
PD0PD1PD2
PD3PD4PD5
PE4
X2
32.768KHz
GND
C133pF
C233pF
MASTER012000v1
1 2
v1.0
VCC1
Vout 2
GND3
U14
MAX6045AEUR-T
VCC
C2010nF
Neptune ventilator Technical manual 60
1 2 3 4
A
B
C
D
4321
D
C
B
A Title
Number RevisionSize
A4
Date: 9-Jan-2003 Sheet of File: C:\DOCUMENT\..\SPIBUS.sch Drawn By:
DE1/RE11
DE2/RE24
DE3/RE35
DR12
DR23
DR36
RO47
DI48
DE49
RE410 GND 11OB4 12OA4 13OB3 14OA3 15
OB2 17OA2 18OB1 19OA1 20U9
SP495
SS0
MISOMOSI
SCK
MISOAMISOBMOSIAMOSIB
SCKASCKBSSSLV0ASSSLV0B
AX12
AY13
BX2
BY1
CX5
CY3
A11
B10
C9
INH6
O/IA 14
O/IB 15
O/IC 4
U11
74HC4053
RST
PB1
PE1
PE0
PSCK
PMISO
PMOSI
SCK
TPE1
RPE0
GND
VCC 4
MR 3RST2
GND1U12
max811L
GND
VCC
1 23 45 67 89 10
P3
HEADER 5X2
VCC
GNDPSCKPMISO
PMOSI
C7
100nF
DE1/RE11
DE2/RE24
DE3/RE35
DR12
DR23
DR36
RO47
DI48
DE49
RE410 GND 11OB4 12OA4 13OB3 14OA3 15
OB2 17OA2 18OB1 19OA1 20U10
SP495
SSSLV1ASSSLV1B
SSSLV2B
SSSLV3B
SSSLV4BSSSLV4A
SSSLV3A
SSSLV2ASS1SS2
SS3
SS4
VCC
GND
GND
GND
SPIBUS
JP-KDP
R13100KVCC
+ C8
0.1µF 16V
+C9
0.1µF 16V
R1 IN13
R2 IN8
T1 IN11
T2 IN10
V+2 V- 6
R1 OUT 12
R2 OUT 9
T1 OUT 14
T2 OUT 7
C1+1
C1 -3 C2+ 4
C2 - 5
U13
MAX202ECWE
+C10
0.1µF 16VVCC
+
C11
0.1µF 16V
GND
TPE1
RPE0
GND GND
MOSIAMISOASCKA
SSSLV0A
SSSLV2ASSSLV3ASSSLV4A
MOSIBMISOBSCKBSSSLV0BSSSLV1B
SSSLV3BSSSLV2B
SSSLV4B
GND GND
1 23 45 67 89 1011 1213 1415 1617 1819 2021 2223 2425 2627 2829 3031 3233 3435 36
P2
HEADER 18X2
R17
5k6R18
5k6R19
5K6
R20
5k6
MOSIA
MISOA
SCKA
MOSIB
MISOB
SCKB
R21
5k6R22
5k6R23
5k6
SSSLV1A
SSSLV2A
SSSLV3A
SSSLV4A
SSSLV2B
SSSLV3B
SSSLV4B
SSSLV1B
C12100N
C13100N
C14100N
C15100N
VCC
1 23 45 67 89 10
P4
HEADER 5X2
GND
VCC
GND
GND
RST
PE2
PE3
C16100N
C17100N
VCC
GND
+12V
GND
GND
VCC
+24V
1 23 45 67 89 1011 1213 1415 1617 1819 2021 2223 2425 2627 2829 3031 3233 3435 36
P1
HEADER 18X2
GND
C18100N
+ C2122µF
+ C2222µF C19
100N
GND GND
VCC
GND+12V
GND
C24100N
C25100N
VCC
GND
C26100N
C27100N
VCC
GND
SSSLV1A
MAS012000v1 v1.02 2
R24
5k6SSSLV0A SSSLV0B
BUZ1
GND
Q1
BS170M
Q2
BS170M
Q3
BS170M
+12V
R2510K
C28100N
GND GND
RST
R27
1K
GND GND GND
C29100N
R26100K
PD7
D6
1N4148
VCCD71N4148
Neptune ventilator - 61 - Technical manual
1.6.3 Build-in test software After turning the ventilator on, the master board will test all his different functions. During start-up the following tests are executed:
o Microprocessor internal functions tests:
test internal program memory test internal SRAM memory test the timers test EEPROM memory test Watchdog
o Microprocessor external functions test
test external SRAM memory test external RS-232 test external real time clock
o Microprocessor communication test
test the communication between master board and display board
test the communication between master board and pneumatic board
When all these tests are executed successfully, the master board is ready to start working. If an error occurred, the type of error is displayed on the LED indicators.
G R R R After start-up the LED indicators can show the following errors:
LED indicators Type error
G R R R
Device is working properly
G R R R
Microprocessor internal function error
G R R R
Microprocessor external function error
G R R R
Communication error
Neptune ventilator - 62 - Technical manual
To determine the exact error on the master board, you have to perform a specific test by using the build-in test software. Note: During executing a specific test on the master board there will be no
communication between the master board and the display board. The MMI board will interpret this as a communication error and a communication error code is displayed and also an alarm sound is audible. This can be very annoying. To prevent this error put the MMI board in internal test mode by setting dipswitch 1 high. See the MMI board section in this manual for detailed instructions.
Note: The pneumatic board will also detect a communication error during a
executing a specific test on the master board. Don’t pay any attention to it in this situation.
• Microprocessor internal function test If a microprocessor internal error is detected, execute an internal function test as follow:
o Turn the Neptune ventilator off o Remove the upper cover plate at the back of the Neptune. o Change the dipswitch setting of the master board with a small
screwdriver to
1 2 3 4 5 o Turn the ventilator on o If no errors where found, all the red LED’s are off and the green LED
is blinking. o If an error is detected, the error is indicated on the red error LED’s and
the green LED is off. The possible errors are illustrated in the table on the next page:
Neptune ventilator - 63 - Technical manual
INTERNAL FUNCTION TEST ERROR CODES
Error LED’s Error code Description
G R R R
0
NO ERRORS detected, test loop is running
G R R R
1
internal PROGRAM MEMORY error
G R R R
2
Internal SRAM error
G R R R
3
Internal TIMER error
G R R R
4
Internal EEPROM error
G R R R
5
Internal WATCHDOG error
G R R R
6
Reserved
G R R R
7
Reserved
Neptune ventilator - 64 - Technical manual
Resolve internal function test error code 4 – Internal EEPROM error
To fix error code 4 or internal EEPROM error follow the next steps:
o Turn the Neptune ventilator off. o Change the dipswitch setting of the master board with a small
screwdriver to
1 2 3 4 5 o Turn the ventilator on o You will notice that all the LED indicators are blinking fast. This
means that the master board EEPROM settings are restored to the factory settings.
o Turn the ventilator off o Repeat the microprocessor internal functions test described on the
previous page. If the EEPROM error still remains contact an authorized service engineer.
Note: Restoring the master EEPROM settings will not affect the pneumatic
board EEPROM settings. Calibration of the ventilator is not needed in this case.
Warning: Try to resolve internal function error code 4 (EEPROM error) with the instructions described above. For all other internal function test errors contact an authorized service engineer.
Neptune ventilator - 65 - Technical manual
• Microprocessor external function test If a microprocessor external error is detected, execute an external function test as follow:
o Turn the Neptune ventilator off o Remove the upper cover plate at the back of the Neptune. o Change the dipswitch setting of the master board with a small
screwdriver to
1 2 3 4 5 o Turn the ventilator on o If no errors where found, all the red LED’s are off and the green LED
is blinking. o If an error is detected, the error is indicated on the red error LED’s and
the green LED is off. The possible errors are illustrated in the table on the next page:
Neptune ventilator - 66 - Technical manual
EXTERNAL FUNCTION TEST ERROR CODES
Error LED’s Error code Description
G R R R
0
NO ERRORS detected, test running
G R R R
1
reserved
G R R R
2
reserved
G R R R
3
reserved
G R R R
4
External REAL TIME CLOCK error
G R R R
5
reserved
G R R R
6
reserved
G R R R
7
reserved
Note: The external function test is running continuously. When an error is
detected, the error is indicated on the LED indicators, and the external function test is repeated. It’s not necessary to turn the ventilator on and off to repeat the external function test.
Neptune ventilator - 67 - Technical manual
Resolve external function test error code 4 – External real time clock error There is a problem with the real time clock. Device U8 contains the actual time and date and timer information. The microprocessor can send data or read data from device U8. Device U8 is equipped with a very accurate crystal and a backup battery. Contact an authorized service engineer if the problem cannot be resolved. If the test is running correctly the buzzer is set on and off every second. Communication test between master board and MMI board
To test the communication between the master board and the MMI board perform the following test:
o Turn the Neptune ventilator off. o Remove the upper cover plate at the back of the Neptune. o Change the dipswitch setting of the master board with a small
screwdriver to
1 2 3 4 5 o Change the dipswitch setting of the MMI board with a small
screwdriver to
1 2 3 4 5
o Turn the ventilator on o If no errors found, the green LED is blinking. If an error is detected,
the error is indicated on the red error LED’s and the green LED is off. The red LED’s indicating the numbers of error’s detected (binary number between 1 and 7).
Neptune ventilator - 68 - Technical manual
Communication test between master board and pneumatic board
To test the communication between the master board and the pneumatic board perform the following test:
o Turn the Neptune ventilator off o Remove the upper cover plate at the back of the Neptune. o Change the dipswitch setting of the master board with a small
screwdriver to
1 2 3 4 5 o Change the dipswitch setting of the pneumatic board with a small
screwdriver to
1 2 3 4 5
o Turn the ventilator on. o If no errors found, the green LED is blinking. If an error is detected,
the error is indicated on the red error LED’s and the green LED is off. The red LED’s indicating the numbers of error’s detected (binary number between 1 and 7).
• LED indicator test
To test the error LED indicators perform the following test:
o Turn the Neptune ventilator off o Remove the upper cover plate at the back of the Neptune. o Change the dipswitch setting of the MMI board with a small
screwdriver to
1 2 3 4 5 o Turn the ventilator on o The error LED’s are blinking one by one.
Neptune ventilator - 69 - Technical manual
1.6.4 Master board layout
Neptune ventilator - 70 - Technical manual
1.7 PNEUMATIC BOARD 1.7.1 Block diagram The block diagram of the pneumatic board is represented on the following page. The pneumatic board is provided with one microprocessor. The microprocessor is re-programmable by means of connector P3. The A/D converter converts the signal of the pressure transducer, the +12V valves voltage and the +12V voltage to a digital value. There are two valve driver IC’s that can drive each 8 valves. Each output is provided with a green LED. You can see the valve state directly on the LED’s. Connectors P4 and P5 connecting the valve driver IC’s with the valves. The pneumatic microprocessor receives settings from the MMI board. On the basis of these settings and the signals from the transducers the microprocessor is calculating all the parameters needed to drive the pneumatic valves. The microprocessor must trigger the 12V safety circuit in a certain time. During a microprocessor failure the +12V safety circuit will not be triggered and the +12V voltage on the valves shuts off. The ventilator switches automatically over to MAN. mode in this situation (because no valves are driven).
Neptune ventilator - 71 - Technical manual
microprocessor
RAMmemory
A/D converterPTR1
Patient pressure
Programminginterface
(connector P3)
SPI communication with master board(connector P2)
LED indicators
Pneumatic board
PTR2Peep valve
exp.pressure
PTR3Patient flow
PTR4freshgas
PTR5Tank pressure
PTR6Peep valve
insp. pressure
Valve driver 2Valve driver 1
12VSAFTY
CIRCUIT
Connector P4 Connector P5
LED IndicatorsLED Indicators
12v valve 12v
Connector P6
Input switches
Block diagram pneumatic board
Neptune ventilator - 72 - Technical manual
1.7.2 Schematic diagram
The schematic diagram of the pneumatic board is represented on the following pages.
Neptune ventilator - 73 - Technical manual
1 2 3 4
A
B
C
D
4321
D
C
B
A Title
Number RevisionSize
A4
Date: 9-Jan-2003 Sheet of File: C:\DOCUMENT\..\PNEUMATIC.sch Drawn By:
PA0(AD0) 51PA1(AD1) 50PA2(AD2) 49PA3(AD3) 48PA4(AD4) 47PA5(AD5) 46PA6(AD6) 45PA7(AD7) 44
PC0(A8) 35PC1(A9) 36PC2(A10) 37PC3(A11) 38PC4(A12) 39PC5(A13) 40PC6(A14) 41PC7(A15) 42
PD0(INT0)25 PD1(INT1)26 PD2(INT2)27 PD3(INT3)28 PD4(IC1)29 PD530 PD6(T1)31 PD7(T2)32
PB0(ss) 10PB1(SCK) 11PB2(MOSI) 12PB3(MISO) 13PB4(OC0/PWM0) 14PB5(OC1A/PWM1A) 15PB6(OC1B/PWM1B) 16PB7(OC2/PWM2) 17
PE0(PDI/RXD)2 PE1(PDO/TXD)3 PE2(AC+)4 PE3(AC-)5 PE4(INTR4)6 PE5(INTR5)7 PE6(INTR6)8 PE7(INTR7)9
PF0(ADC0)61 PF1(ADC1)60 PF2(ADC2)59 PF3(ADC3)58 PF4(ADC4)57 PF5(ADC5)56 PF6(ADC6)55 PF7(ADC7)54
RD 34
WR 33
XTAL1 24XTAL2 23
ALE 43
TOSC119 TOSC218
RESET20
AREF62
AGND63
AVCC64
PEN 1
VCC52
GND 22GND53
VCC 21
ATm
ega103L
U1
ATMEGA103
AD0AD1AD2AD3AD4AD5AD6AD7
ALE
VALVES
VALVES.sch
A8A9A10A11A12A13A14A15
SENSORS
SENSORS.sch
RDWR
RS485 DRIVER
PNEUSPI.sch
OC1
C11
1D2
2D3
3D4
4D5
5D6
6D7
7D8
8D9
1Q 19
2Q 18
3Q 17
4Q 16
5Q 15
6Q 14
7Q 13
8Q 12
U2
74HC573
AD0AD1AD2AD3AD4AD5AD6AD7
ALE
GND
A0A1A2A3A4A5A6A7
A8A9A10
A11
A7
A13A14
A1
B2
C3
E14
E25
E36
Y0 15
Y1 14
Y2 13
Y3 12
Y4 11
Y5 10
Y6 9
Y7 7
U3
74HC138
A4$100*
$107*
A5A6
A15
A12
12
89
U4A
74HC4075
34
56
U4B
74HC4075
1112
1310
U4C
74HC4075
CS0CS1CS2CS3CS4CS5CS6CS7
A010
A19
A28
A37
A46
A55
A64
A73
A825
A924
A1021
A1123
A122
A1326
WR27
CE20 OE22
A141
D0 11
D1 12
D2 13
D3 15
D4 16
D5 17
D6 18
D7 19
256
U5
RAM/ROM/FLASH
AD0AD1AD2AD3AD4AD5AD6AD7
A0A1A2A3A4A5A6A7A8A9A10A11A12A13
01 S1
01 S2
A14
WR
RD
01 S3
01 S4
A15
A14
WR
12
3
U6A
74HC02
45
6
U6B
74HC02
8
910
U6C
74HC02
WR
WR
CS0
CS1
VALSEL1
VALSEL2
PF0PF1PF2PF3PF4PF5PF6PF7
RST
PE1PE0
MISOMOSIPB1SS
VAL.CL12V ON/OFFPD3PD4PD5PD6PD7
VAL.EN.
GND GND
PNEUMATIC
JP-KDP
R1
100K
R2
100K
R3
100K
R4
100K
PF7PF6PF5
VCC
D4
HLM
P-17
00
R5
1K5
PB4
D3
HLM
P-17
00
R6
1K5
D2
HLM
P-17
00
R7
1K5
D1
HLM
P-17
90
R8
1K5
PB7PB6PB5
GND
PF4
VCC
GNDGND
VCC
PB4PB5PB6PB7VCC
R16100R
GND
ADINT
X1 6MHz
12345
109876
SW1
DS-05 APEM
C3100N
C133pF
C233pF
PNEU012000v1 v1.0
1 4
Neptune ventilator - 74 - Technical manual
1 2 3 4
A
B
C
D
4321
D
C
B
A Title
Number RevisionSize
A4
Date: 9-Jan-2003 Sheet of File: C:\DOCUMENT\..\VALVES.sch Drawn By:
AD0AD1AD2AD3AD4AD5AD6AD7
12345678910
11121314151617181920
P4
HEADER 10X2
V1V2V3V4V5V6V7V8
D5 D6 D7 D8 D9 D10 D11 D12
12345678
161514131211109
R17
8*4K7
VCCGND
VAL.CLVALSEL1
VAL.EN.
AD0AD1AD2AD3AD4AD5AD6AD7
12345678910
11121314151617181920
P5
HEADER 10X2
V9V10V11V12V13V14V15V16
D13 D14 D15 D16 D17 D18 D19 D20
12345678
161514131211109
R18
8*4K7
VCCGND
VAL.CLVALSEL2
VAL.EN.
D1 2
D2 3
D3 4
D4 5
D5 6
D6 7
D7 8
D8 9
Q118
Q217
Q316
Q415
Q514
Q613
Q712
Q811
E1 1
E2 19
U7
74HC541
1 2 3 4 5 6 7 8
16 15 14 13 12 11 10 9
R198*100K
VCC
GND
RD
Q1
RFP30P05
+12v
Q2
BS170M
GND
12V
ON
/OFF
AD0AD1AD2AD3AD4AD5AD6AD7
VCC
GND
CS2
VALVES
JP-KDP
A 1B 2CLR 3Q4
Q13
Cext 14
RCext 15
U8A
74HC123
RST
GND
VCC
12v val.
C21
100nF
C22
100nF
C23
100nF
C24
100nF
R20
100K
A 9B 10CLR 11Q12
Q5
Cext 6
RCext 7
U8B
74HC123
GND
COM 12OUT8 13OUT7 14OUT6 15OUT5 16OUT4 17OUT3 18OUT2 19OUT1 20
VDD 21
OE 22CLEAR1
IN13 STROBE2
INT24
INT35
INT46
INT57
INT68
INT79
INT810
GND11
MIC5801BN
U9
MIC5801BN
COM 12OUT8 13OUT7 14OUT6 15OUT5 16OUT4 17OUT3 18OUT2 19OUT1 20
VDD 21
OE 22CLEAR1
IN13 STROBE2
INT24
INT35
INT46
INT57
INT68
INT79
INT810
GND11
MIC5801BN
U10
MIC5801BN
D21 R29
4K7GND
1 23 45 67 89 10
P6
HEADER 5X2
PNEU012000v1 v1.0
R30
22K
R9
10K
C20
330N
2 4
Neptune ventilator - 75 - Technical manual
1 2 3 4
A
B
C
D
4321
D
C
B
A Title
Number RevisionSize
A4
Date: 9-Jan-2003 Sheet of File: C:\DOCUMENT\..\SENSORS.sch Drawn By:
AGND 15CH0 16CH1 17CH2 18CH3 19CH4 20CH5 21CH6 22CH7 23INT 24REFADJ 25REF 26VDD 27DGND 28CLK1
CS2
WR3
RD4
HBEN5
SHDN6
D77
D68
D59
D410
D3/D1111
D2/D1012
D1/D913
D0/D814
MA
X19
7
U11
MAX197
AD0AD1AD2AD3AD4AD5AD6AD7
C26100pF
AGND
WRRDA0
CS3
VCC
ADINT
AGND
C270.01µF
AGNDAGND
SENSORS
JP-KDP
Vref COM12
Vref BG13
Vref 2.514
Vref 515
Vref 1016
Vref Out4
V+in6 RG9 RG8
V-in7
v- 3v+
1
slee
p2
Vo 10
SENSE 11
IAref 5
U12 INA125
VO-1
-Vexc 2
VO+3
+Vexc4 PTR1
SM5652-015-G-3-L
3
26
74
U18
OPA237
R10
RES
1
AGNDAGND Vref COM12
Vref BG13
Vref 2.514
Vref 515
Vref 1016
Vref Out4
V+in6 RG9 RG8
V-in7
v- 3
v+1
slee
p2
Vo 10
SENSE 11
IAref 5
U13 INA125
VO-1
-Vexc 2
VO+3
+Vexc4 PTR2
SM5652-015-G-3-L
3
26
74
U19
OPA237
R11
RES
1
AGNDAGND Vref COM12
Vref BG13
Vref 2.514
Vref 515
Vref 1016
Vref Out4
V+in6 RG9 RG8
V-in7
v- 3
v+1
slee
p2
Vo 10
SENSE 11
IAref 5
U14 INA125
VO-1
-Vexc 2
VO+3
+Vexc4 PTR3
SM5652-003-D-3-L
3
26
74
U20
OPA237
R12
RES
1
AGNDAGND Vref COM12
Vref BG13
Vref 2.514
Vref 515
Vref 1016
Vref Out4
V+in6 RG9 RG8
V-in7
v- 3
v+1
slee
p2
Vo 10
SENSE 11
IAref 5
U15 INA125
VO-1
-Vexc 2
VO+3
+Vexc4 PTR4
SM5652-003-D-3-L
3
26
74
U21
OPA237
R13
RES
1
AGNDAGND Vref COM12
Vref BG13
Vref 2.514
Vref 515
Vref 1016
Vref Out4
V+in6 RG9 RG8
V-in7
v- 3
v+1
slee
p2
Vo 10
SENSE 11
IAref 5
U16 INA125
VO-1
-Vexc 2
VO+3
+Vexc4 PTR5
SM5612-030-G-3-L
3
26
74
U22
OPA237
R14
RES
1
AGNDAGND
+12v SENSORS
CH
0
CH
1
CH
2
CH
3
CH
4
PATIENT PRESSURE PATIENT FLOWPEEP VALVE EXP. PRESSURE FRESH GAS FLOW TANK PRESSURE
Vref COM12
Vref BG13
Vref 2.514
Vref 515
Vref 1016
Vref Out4
V+in6 RG9 RG8
V-in7
v- 3
v+1
slee
p2
Vo 10
SENSE 11
IAref 5
U17 INA125
VO-1
-Vexc 2
VO+3
+Vexc4 PTR6
SM5652-015-G-3-L
3
26
74
U23
OPA237
R15
RES
1
AGNDAGND
+12v SENSORS
CH
5
PEEP VALVE INSP. PRESSURE
R21100K
R2310K
AGND
+12v
CH6
AGND
AGND
R22100K
R2410K
AGND
CH7
12v
val.
+C28
22µF
/35v
+
C29
47µF +
C30
47µF
C4
100N
C5
100N
C6
100N
C7
100N
C8
100N
CH0CH1CH2CH3CH4CH5CH6CH7
VCC
PNEU012000v13 4
v1.0
C34100N
C35
100N
C36
100N
C37
100N
C40
100N
C43
100N
C46
100N
C38
100N
C39
100N
C41
100N
C42
100N
C44
100N
C45
100N
C47
100N
C48
100N
C49100N
C11
100N
C33
100N
L1
22µH
C19
100N
AGND
+12v SENSORS +12v
R31 10
C18100N
AGND
Neptune ventilator - 76 - Technical manual
1 2 3 4
A
B
C
D
4321
D
C
B
A Title
Number RevisionSize
A4
Date: 9-Jan-2003 Sheet of File: C:\DOCUMENT\..\PNEUSPI.sch Drawn By:
DE1/RE11
DE2/RE24
DE3/RE35
DR12
DR23
DR36
RO47
DI48
DE49
RE410 GND 11OB4 12OA4 13OB3 14OA3 15
OB2 17OA2 18OB1 19OA1 20U24
SP495
SS
SS
MISOMISOMOSI
SCK
MISOAMISOBMOSIAMOSIB
SCKASCKBSSSLV0ASSSLV0B
GND
AX12
AY13
BX2
BY1
CX5
CY3
A11
B10
C9
INH6
O/IA 14
O/IB 15
O/IC 4
U25
74HC4053
RST
PB1
PE1
PE0
PSCK
PMOSI
SCK
PMISO
GND
VCC 4
MR 3RST2
GND1U26
max811L
GND
VCC
1 23 45 67 89 10
P3
HEADER 5X2
VCC
GNDPSCKPMISO
PMOSI
GND
RST
PNEUMATIC SPI
JP-KDP
1
23
U27A
74HC00
GND
C25100nF
+12V
GND
GND
VCC
1 23 45 67 89 1011 1213 1415 1617 1819 2021 2223 2425 2627 2829 3031 3233 3435 36
P1
HEADER 18X2
GND
C9100N
+ C31
22µF
+ C32
22µF
/35v
C10100N
GND GND
VCC
GND+12V
GND
MOSIAMISOASCKA
SSSLV0A
MOSIBMISOBSCKBSSSLV0B
GND GND
1 23 45 67 89 1011 1213 1415 1617 1819 2021 2223 2425 2627 2829 3031 3233 3435 36
P2
HEADER 18X2
R25
5K6R26
5K6R27
5K6R28
5K6
MOSIA
MISOA
SCKA
MOSIB
MISOB
SCKB
SSSLV0A SSSLV0B
4
56
U27B
74HC00
9
108
U27C
74HC00
12
1311
U27D
74HC00
GND GND GND
C12100N
C13100N
C14100N
C15100N
C16100N
C17100N
VCC
GND
PNEU012000v14 4
v1.0
Neptune ventilator Technical manual 77
1.7.3 Build-in test software
• Start-up test After turning the ventilator on, the pneumatic board will test all his different functions. During start-up the following tests are executed:
o Microprocessor internal functions tests:
test internal program memory test internal SRAM memory test the timers test EEPROM memory test Watchdog
o Microprocessor external functions test
test external SRAM memory test external 12V safety valve test external A/D converter
o Microprocessor communication test
test the communication between pneumatic board and master board
When all these tests are executed successfully, the MMI board is ready to start working. If an error occurred, the type of error is displayed on the LED indicators.
G R R R After start-up the LED indicators can show the following errors:
LED indicators Type error
G R R R
Device is working properly
G R R R
Microprocessor internal function error
G R R R
Microprocessor external function error
G R R R
Communication error
Neptune ventilator Technical manual 78
To determine the exact error on the pneumatic board, you have to perform a specific test by using the build-in test software. Note: During executing a specific test on the pneumatic board there will be
no communication between the master board and the pneumatic board. The master board will interpret this as a communication error and a continuous beep is audible. This can be very annoying. To prevent this error put the master board in internal test mode by setting dipswitch 1 high. See the master board section later in this manual for detailed instructions.
Note: The MMI board will also detect a communication error during a
executing a specific test on the pneumatic board. To prevent this error put the MMI board in internal test mode by setting dipswitch 1 high. See the MMI board section in this manual for detailed instructions.
• Microprocessor internal function test If a microprocessor internal error is detected, execute an internal function test as follow:
o Turn the Neptune ventilator off o Remove the upper cover plate at the back of the Neptune. o Change the dipswitch setting of the pneumatic board with a small
screwdriver to
1 2 3 4 5 o Turn the ventilator on o If no errors where found, all the red LED’s are off and the green LED
is blinking. o If an error is detected, the error is indicated on the red error LED’s and
the green LED is off. The possible errors are illustrated in the table on the next page:
Neptune ventilator Technical manual 79
INTERNAL FUNCTION TEST ERROR CODES
Error LED’s Error code Description
G R R R
0
NO ERRORS detected, test loop is running
G R R R
1
internal PROGRAM MEMORY error
G R R R
2
Internal SRAM error
G R R R
3
Internal TIMER error
G R R R
4
Internal EEPROM error
G R R R
5
Internal WATCHDOG error
G R R R
6
Reserved
G R R R
7
Reserved
Neptune ventilator Technical manual 80
Resolve internal function test error code 4 – Internal EEPROM error
To fix error code 4 or internal EEPROM error follow the next steps:
o Turn the Neptune ventilator off. o Change the dipswitch setting of the pneumatic board with a small
screwdriver to
1 2 3 4 5 o Turn the ventilator on o You will notice that all the LED indicators are blinking fast. This
means that the pneumatic board EEPROM settings are reset. Turn the ventilator off
o Repeat the microprocessor internal functions test described on the previous page. If the EEPROM error still remains contact an authorized service engineer.
Warning: Reset from the pneumatic board EEPROM settings will AFFECT
the calibration parameters. Recalibration of the ventilator is necessary in this case. Refer to the maintenance and calibration procedure described later in this manual!
Neptune ventilator Technical manual 81
• Microprocessor external function test If a microprocessor external error is detected, execute an external function test as follow:
o Turn the Neptune ventilator off o Remove the upper cover plate at the back of the Neptune. o Change the dipswitch setting of the pneumatic board with a small
screwdriver to
1 2 3 4 5 o Turn the ventilator on o If no errors where found, all the red LED’s are off and the green LED
is blinking. o If an error is detected, the error is indicated on the red error LED’s and
the green LED is off. The possible errors are illustrated in the table on the next page:
Neptune ventilator Technical manual 82
EXTERNAL FUNCTION TEST ERROR CODES
Error LED’s Error code Description
G R R R
0
NO ERRORS detected, test running
G R R R
1
reserved
G R R R
2
External SRAM error
G R R R
3
reserved
G R R R
4
reserved
G R R R
5
reserved
G R R R
6
reserved
G R R R
7
reserved
Note: The external function test is running continuously. When an error is
detected, the error is indicated on the LED indicators, and the external function test is repeated. It’s not necessary to turn the ventilator on and off to repeat the external function test.
Neptune ventilator Technical manual 83
Resolve external function test error code 2 – External SRAM error There is a problem with the external SRAM memory. The external SRAM memory is located on socket U5 on the pneumatic board. The microprocessor is reading data from and writing data to the external SRAM memory. Contact an authorized service engineer if the problem cannot be resolved.
• Pneumatic function test You can test all the pneumatic functions of the pneumatic board as follow:
o Turn the Neptune ventilator off o Remove the upper cover plate at the back of the Neptune. o Change the dipswitch setting of the pneumatic board with a small
screwdriver to
1 2 3 4 5 o Turn the ventilator on o If no errors where found, all the red LED’s are off and the green LED
is blinking. Each valve of each valve driver is driven one by one. o If an error is detected, the error is indicated on the red error LED’s and
the green LED is off. The possible errors are illustrated in the table on the next page:
Neptune ventilator Technical manual 84
PNEUMATIC FUNCTION TEST ERROR CODES
Error LED’s Error code Description
G R R R
0
NO ERRORS detected, test running
G R R R
1
Not able to start A/D converter
G R R R
2
A/D converter busy
G R R R
3
All A/D converter readings = $FFF
G R R R
4
All A/D converter readings = $000
G R R R
5
Not able to switch 12V safety circuit
OFF within 20 ms
G R R R
6
Not able to switch 12V safety circuit
ON within 20 ms
G R R R
7
No +12V input or +12V input too low
Neptune ventilator Technical manual 85
1.7.4 Pneumatic board layout
Neptune ventilator Technical manual 86
2. PNEUMATIC SYSTEM
2.1 PNEUMATIC DIAGRAM The pneumatic diagram of the Neptune ventilator is represented on the next page.
Neptune ventilator Technical manual 87
Neptune ventilator Technical manual 88
2.2 MANUAL / SPONTANEOUS MODE (MAN) If the Neptune is set to Man/Spont mode, the hand balloon is connected through the patient breathing circuit and absorber to the patient. Squeezing the manual bag will initiate an inspiration phase. The CO2 is removed by the soda lime in the canister. Mixed gas from hand balloon and fresh gas flow goes directly to the patient. During this phase, with the ventilator switched to Man/Spont mode, the patient pressure is monitored and shown on the display. If the patient pressure exceeds the upper limit pressure setting, an audible and visual alarm is set. In this mode, the safety valve on the manual bag will guarantee the maximum pressure level that can be set.
Man/Spont INSPIRATION
Neptune ventilator Technical manual 89
Releasing the hand balloon will initiate an expiration phase. The patient can breath out. At the end of the expiration, parameters like PEEP, frequency, PIP and tidal volume are calculated and shown on the display.
Man/Spont EXPIRATION
Neptune ventilator Technical manual 90
2.3 CONTROLLED MANDATORY VENTILATION (CMV) The system is built around the so-called "bag in bottle principle". The figure on the next page represents the inspiration and expiration phase in CMV mode. Tank T is filled with an initial pressure calculated by the pneumatic microprocessor. The initial pressure is calculated on the basis of an algorithm comprising the following parameters:
o Fresh gas flow o Total circuit resistance o Patient compliance o Hose system compliance o Frequency o I/E ratio o Tidal volume o PEEP o Leakage
The contents of tank T flows into the bottle during inspiration, so that a pressure rise will occur in it, and this will compress the bag. The gas flows via the MAN/CMV valve, flow sensor and absorber to the patient. During inspiration, the gas is routed through the soda lime. The patient pressure is shown on the barograph or on the graphic. During inspiration, the PEEP/upper limit pressure valve will contain the set upper limit pressure. As soon as the patient pressure exceeds this limit, gas will be evacuated from the patient circuit.
CMV INSPIRATION
Neptune ventilator Technical manual 91
During expiration, the patient breathes out into the bag via the soda lime in the absorber, the flow sensor and the MAN/CMV valve. The bottle venting valve is open, causing the pressure in the bottle to become equal to atmospheric pressure. If the fresh gas flow is higher than the consumption by the patient or leakage, as appropriate, the system would become overfilled. This is prevented by the evacuation of the surplus gas via the PEEP/upper limit valve (X6).
CMV EXPIRATION
Neptune ventilator Technical manual 92
2.4 PRESSURE CONTROLLED VENTILATION (PCV) In PCV mode the ventilator has to deliver the set peak pressure. At the start of inspiration the tank who is filled during last expiration is connected with bottle. This gives a decelerating flow, high at the start that comes to zero when the peak level is reached. V7 and V8 are used to give extra flow to keep the pressure at same level. For a smooth regulation activation from this valves is done in ‘PWM’ pulse wide modulation . V1,V2,V14,V15&V16 are used to open the bottle and keep the pressure at peak level. During the inspiration time ,the electronics will measure the patient pressure. If the pressure is not within limits a calculation is done to adjust the tank pressure setting for the next inspiration.. If the patient pressure becomes higher than the set peak pressure +5hPa/+5mbar, valve V13 will be set in expiration position, the amount of pressure that is too much can escape via the evacuation. When the patient pressure is within limits, valve V13 will be set in inspiration position. Each time valve V13 switches during inspiration the pressure in reservoir upper limit becomes lower.” The start pressure is 100hPa/100mbar”. During the expiration the exhaled patient flow is measured and tidal volume is calculated. If the PEEP (positive end expiration pressure) is higher than the set PEEP + 20hPa/+20mbar, the ventilator goes in Man/Spont mode and an error message is shown on display.
Neptune ventilator Technical manual 93
PCV EXPIRATION
PCV INSPIRATION
Neptune ventilator Technical manual 94
2.5 PEEP During expiration and a PEEP is set, a certain pressure is set behind the PEEP valve X6. This pressure is derived from the X6 back pressure during inspiration. At the start op expiration valve V13 is set to expiration mode. With a PEEP set, valve V9 is always on. Valve 10 is off for a certain time, during this time the back pressure from valve X6 can escape rapidly, to create the best expiration situation. Valve V10 is switched on after a certain time to load the PEEP reservoir with pressure. The time that valve V10 is switched on is calculated at the end of expiration, this time will be shorter for higher PEEP level and longer for lower PEEP level. PEEP regulation is the same for CMV and PCV mode. 2.6 Flush safety There is a detection when an oxygen flush is generated to prevent high pressure if the upper limit was set too high. The Max pressure allowed during an O2 flush is last peak pressure +20hPa / +20mbar, If the volume was stable and when this level (last pip +20 is below upper limit setting). When the upper limit alarm is lower ,the upper limit alarm level is used. The min level is 30hPa / 30mbar if upper limit is higher. When the upper limit alarm level is lower then the calculated safety, the upper limit alarm level has priority.
Neptune ventilator Technical manual 95
2.7 MECHANICAL CONSTRUCTION 2.7.1 Patient breathing unit There are four main parts in the patient unit that need to be distinguished:
o PEEP and upper limit pressure valve o MAN/CMV switching valve o Flow sensor o Latex free autoclavable bag
PEEP and upper limit pressure valve
MAN/CMV switching valve
Flow sensor
Neptune ventilator Technical manual 96
● PEEP and upper limit pressure valve Regulation of the PEEP and upper limit pressure is provided by the mushroom valve (C) which operates a plastic valve (D).
During inspiration, the value set for the upper limit pressure is applied to the mushroom valve causing this to expand and thus exert the same pressure on the plastic valve. When a pressure higher than the upper limit pressure is built up on the underside, the valve will be lifted and gas is able to escape via the evacuation. The plastic valve ensures that the pressure is uniformly distributed so that the noise made by the escaping air is attenuated. During expiration, the value set for the PEEP is applied to the mushroom valve. The exhalation curve thus approaches the PEEP gradually. There are no sudden transitions, which results in a “smoothed shape”. The upper limit pressure thus ensures that the patient is not exposed to too high pressures during inspiration, while the PEEP ensures - when expiration occurs - that sufficient pressure remains in the lung. This is only applicable in CMV and PCV mode.
Latex free autoclavable bag
C
D
Neptune ventilator Technical manual 97
● MAN/CMV switching valve The changeover valve is always in the unpowered state in the Man/Spont mode. The valve (G) is in fact pressed up by a spring so that the absorber is connected directly to the hand balloon. If the ventilator mode is changed to CMV or PCV, pressure reaches the switching diaphragm via connection E. The valve is thereby pressed down and the hand balloon shut off. When this occurs, the connection to the bag (F) is opened so that the absorber is now connected to the bag.
G
E
F
Neptune ventilator Technical manual 98
● Flow sensor The flow sensor is a derivative, developed by Medec Benelux NV, of the well- known Fleisch principle. To ensure laminar flow, special perforated rings are fitted. The flow sensor can measure flow up to 96 l/min. Differential pressure (∆P) is measured across the restriction by means of connections A and B, which are connected to transducer TRX3.
● Latex free autoclavable bag The bag is connected to the connection F intended for this purpose. It should be clearly understood that in the unpowered state (i.e. in the Man/Spont mode), the bag is not used. The bag is a special balloon and therefore cannot be replaced with a different type of balloon. Warning: Medec Benelux NV can only guarantee correct ventilator operation with the same type of balloon.
Neptune ventilator Technical manual 99
2.7.2 Bottle The feed-through plate of the bottle is provided with five connections. This plate ensures a gas-tight connection between the bottle and the patient breathing unit. This makes the patient breathing unit easy to remove and to clean.
A. MAN/CMV switching valve. B. PEEP and upper limit pressure valve. C. Outlet for gas evacuation from the mushroom valve. D. Measurement point, flow sensor positive. E. Measurement point, flow sensor negative and patient pressure.
The bag is inserted in the bottle after which the patient breathing unit is attached to the bottle by means of the whale clip. The bottle is manufactured from aluminum. On the back of the bottle, there are two valves:
o The bottle safety valve located on the top is adjusted to 120hPa/120mbar. So the pressure in the bottle (and thus also in the patients lung) never exceeds this value.
o The bottle venting valve located on the bottom ensures that during
inspiration, the bottle is shut off from the outside air so that the bag can be compressed. During expiration, this valve ensures that the bottle is vented, so that the bag can be filled again with the expired gas from the patient.
A B C
D
E
Neptune ventilator Technical manual 100
bottle safety valve
bottle venting valve
V12
V13
V1V2
V14V15V16
V9
V10V11V3
X3V7V9V6V5V8
X2
Neptune ventilator Technical manual 101
2.7.3 Supply tank The supply tank T (contents = 1,2 liter) provides the inspiration stroke volume. The tank is located around the bottle. The tank can be filled with the use of two low flow valves or one high flow valve. There’s another valve provided to connect the tank with the bottle.
The supply tank T
Bottle
Neptune ventilator Technical manual 102
2.7.4 Solenoid valves
The solenoid valves are special maintenance-free low-power valves. Because of the special design, in which there is no sliding core but a corrosion-resistant metal diaphragm, the valve processes a number of unique properties. The ring which connects the coil and the valve housing may be unscrewed one turn in order to attend to the electrical connections. Never unscrew the ring entirely, to prevent vital parts being lost or damaged. Note: Never change the order of diaphragm and spacer ring, since these are specially calibrated.
Neptune ventilator Technical manual 103
3. MAINTENANCE AND CALIBRATION
The aim of the maintenance and calibration procedure is to ensure proper and safe working of the Neptune ventilator. It must be performed by a qualified service engineer every 6 months. Always recalibrate the ventilator and execute a leaktest after replacing a component. 3.1 VISUAL INSPECTION Remove the cover plate of the Neptune ventilator.
• Electronic system
Ensure that all the boards of the extendable system set are mounted correctly.
Check that all electrical connectors are latched. Check all keyboard board switches for proper working Check control knob switch and rotation for proper working
• Pneumatic
Make sure that all tubes are properly connected and none are kinked or pinched.
Check the clear tubing for signs of water. If any is detected, replace the tubing or blow out the water by means of compressed air.
Check the air input water trap. Any presence of water must be removed by pressing the bottom pin of the water trap. Make sure that no water is flowing away into the ventilator.
Check the patient breathing system, the absorber and all external tubing.
Neptune ventilator Technical manual 104
3.2 BATTERY BACKUP The ventilator can work for at least 1 hour on battery when fully charged. Connect the ventilator for a couple of hours to the AC mains supply to assure a fully charged battery. Check the ventilator battery backup by means of the following instructions:
o Connect the power cord to the AC mains inlet o Turn the AC mains switch at the back of the ventilator ON o Turn the ventilator ON o There is a mains plug drawn at the top of the display which means that
the ventilator is working on AC mains supply o Set the ventilator to CMV mode o Check the battery backup by pulling the mains plug o There is a battery drawn at the top of the display which means that the
ventilator is working on battery If the ventilator shuts off in less than 1 hour, replace the battery by a new one. We recommend replacing the battery every 3 years. Always replace the battery with the same type and ratings. Note: The battery does not require maintenance under normal circumstances. Always replace with same type every three years.
Neptune ventilator Technical manual 105
3.3 THE CALIBRATION MENU Enter the calibration menu as follows:
o Turn the ventilator ON o Go to the Info menu o Press the buttons in the exact order as represented:
1
23
4
5
Neptune ventilator Technical manual 106
The calibration menu looks as follows:
1. Control buttons: with these buttons you can change the calibration settings, drive the +12 voltage of the valves, set valves on and off, etc.
2. Exit button: you can leave the calibration menu by pressing the exit button. The
changes made in the calibration menu are stored to the EEPROM memory of the pneumatic board.
3. Input indicators: the input pressure switches, connected to connector P6 on the
pneumatic board, are indicated over here. INPUT1 is connected to the air input pressure switch, INPUT2 is connected to the O2 input pressure switch and INPUT6 is connected to the N2O input pressure switch. INPUT3, INPUT4, INPUT5, INPUT7 and INPUT8 are not used.
4. Valve ON/OFF: you can set each valve individually ON and OFF.
5. +12V VALVES: you can turn the +12V voltage of the valves ON and OFF.
Make sure that you set the +12V VALVES voltage on when you are testing each valve individually. The ADC reading value is the value that the analog / digital converter (ADC), located on the pneumatic board, indicates during converting the +12V VALVES voltage to a digital value.
6. +12V SUPPLY: over here you can see a read-out of the digital value of the
+12V supply generated by the ADC converter. The +12V supply must always be present after turning the ventilator on.
7. Selection beam: you can scroll the selection beam through the calibration menu
by rotating the control knob. The function of the control buttons at the left is depending on the position of the selection beam.
8. Pressure transducers: each pressure transducer (TRX) can be calibrated
individually. TRX1 measures the airway pressure, TRX2 measures the PEEP pressure, TRX3 measures the patient flow located in the patient breathing unit,
1
2 3
4
5
7
8
9
6
Neptune ventilator Technical manual 107
TRX4 measures the fresh gas flow, TRX5 measures the tank pressure and TRX6 measures the upper limit pressure.
9. Service time: the service time is used to check the time between the last
maintenance and now. The service engineer has the possibility to zero this service time after the maintenance and calibration procedure is executed. The service timer is always running when the ventilator is on, regardless of standby mode, manual mode, PCV or CMV mode.
Neptune ventilator Technical manual 108
3.4 THE PRESSURE TRANSDUCERS
Zeroing the pressure transducer TRX1, TRX2, TRX5 and TRX6
o Disconnect patient breathing unit. o Close fresh gas. o Activate valve V5 to discharge the tank. o Move the selection beam to transducer TRX1. o Press the zero button. o Transducer TRX1 is now zeroed. o Repeat the 3 previous instructions for zeroing transducer TRX2, TRX5
and TRX6. Zeroing the pressure transducer TRX3 and TRX4 Warning: A zero and/or max. calibration of transducer TRX3 and/or TRX4
will erase the previous GAIN setting of the transducer. Make a note of these GAIN setting (they can be used later) before starting any zero or Max calibration transducer TRX3 and TRX4.
The pressure transducer TRX3 and TRX4 are used to measure flow. They measure the differential pressure across the flow sensor. The pneumatic microprocessor uses a look-up table of the sensor to convert differential pressure to flow (liter/minute). There can be a very small deviation between the look-up table and the flow sensor. You can eliminate this deviation by adjusting the GAIN of the selected pressure transducer (only applicable
TRX1 TRX2 TRX3
TRX4 TRX5 TRX6
1
2
34 5
6++
- -
Neptune ventilator Technical manual 109
on TRX3 and TRX4). You need a calibrated flow meter to compare the adjusted flow in the calibration menu with the reference flow measured. Warning: A zero and/or max. calibration of transducer TRX3 and/or TRX4
will erase the previous GAIN setting of the transducer. Adjustment of the GAIN is necessary to eliminate the deviation again.
o Be sure that the fresh gas is closed.
NOTE: It is not necessary to open the ventilator to do the zeroing TRX3. Be sure that patient breathing unit is disconnected.
o Move the selection beam to transducer TRX3. o Press the zero button. o Transducer TRX3 is now zeroed.
NOTE: Set back the factory setting from TRX3 where you have made a note from when only a zeroing from TRX3 is necessary.
When also a Max calibration will be done, restore of factory gain you have to do after the Max calibration.
NOTE: It is not necessary to open the ventilator to do the zeroing TRX4. Be sure that patient breathing unit is disconnected.
o Move the selection beam to transducer TRX4. o Set a fresh flow of 0,3 liters/min. o Press the zero button. o Close the fresh gas flow. o Transducer TRX4 is now zeroed.
NOTE: Set back the factory setting from TRX4 where you have made a note
from when only a zeroing from TRX4 is necessary. When also a Max calibration will be done, restore of factory gain you
have to do after the Max calibration.
Neptune ventilator Technical manual 110
Maximum calibration of pressure transducer TRX1, TRX2 and TRX6 To perform the maximum calibration of the pressure transducers, you need a calibrated pressure gauge, a T-piece, silicon tubes, syringe and a pneumatic test connector. Connect in accordance with the next drawing:
Calibration set
o Connect calibration set with transducer, see picture page 108 section 3.4 THE PRESSURE TRANSDUCERS
o Move the selection beam to transducer TRX1. o Connect the calibration set to pressure transducer TRX1 by way of the
pneumatic test connector. o Set a pressure of exactly 100hPa/100mbar by way of the syringe. o Check the OUT2 value of TRX1. If equal to 100hPa/100mbar, go to the
next step. If not, perform a max. calibration by pressing the max. button. The TRX1 OUT2 value is now equal to 100hPa/100mbar.
o Repeat the 4 previous instructions for the max. calibration of transducer
TRX2, and TRX6 in the same way. Maximum calibration of pressure transducer TRX3
o Move the selection beam to transducer TRX3. o Connect the calibration set to the positive port pressure transducer TRX3
see picture page 108. o Set a pressure of exact 20hPa/20mbar by way of the syringe. o Perform a maximum calibration of TRX3 by pressing the max. button. ▪ It’s possible to eliminate the deviation between the microprocessor look-
up table and the flow sensor to create an extremely accurate measurement of the flow. These steps are optional and can only performed when a calibrated digital flowmeter is available:
Connect to pressure transducer TRX by means of pneumatic test connector.
Neptune ventilator Technical manual 111
NOTE: When you don’t have a calibrated digital flowmeter, set back the factory GAIN by pressing the gain+ or gain- button. Gain+ and GAIN- are in steps of 5, set back the value that is nearest the factory value where you have made a note from.
Connect the digital flowmeter with one side to the left input connector of the patient breathing unit.
Disconnect handset from right side patient breathing unit. Connect other side from flowmeter with fresh gas front outlet.
Select fresh gas outlet front and push O2 flush. Make sure that no valves are active. Adjust the TRX3 OUT2 reading by manipulating the TRX3
GAIN. Change the GAIN by pressing the gain+ or gain- button. The digital flowmeter value must be equal to the TRX3 OUT2 reading.
Maximum calibration of pressure transducer TRX4
o Move the selection beam to transducer TRX4. o Connect the calibration set to pressure transducer TRX4 + port. o Set a pressure of exact 20hpa/20mbar by way of the syringe. o Perform a maximum calibration of TRX4 by pressing the max. button. ▪ It’s possible to eliminate the deviation between the microprocessor look-
up table and the flow sensor to create an extremely accurate measurement of the flow. These steps are optional and can only performed when an calibrated digital flowmeter is available:
NOTE: When you don’t have a calibrated digital flowmeter, set back the factory GAIN by pressing the gain+ or gain- button. Gain+ and GAIN- are in steps of 5, set back the value that is nearest the factory value where you have made a note from.
Set a fresh gas flow of 10 liters/min air. Make sure that no valves are active. Connect one side of the digital flowmeter to the fresh gas outlet at
the front of the Neptune ventilator. Make sure the fresh gas front output is selected. Leave the other side of the digital flowmeter open.
Adjust the TRX4 OUT2 reading by manipulating the TRX4 GAIN. Change the GAIN by pressing the gain+ or gain- button. The digital flowmeter value must be equal to the TRX4 OUT2 reading.
Neptune ventilator Technical manual 112
Maximum calibration of pressure transducer TRX5
o Place the pressure gauge between the tube that connects the pressure transducer TRX5 with the tank.
o Make use of VALVE7 or VALVE8 to build a pressure of 2000hPa/2bar in the tank.
o Regulate the pressure so the pressure gauge is indicating 2000hPa/2bar. Check the OUT2 value of TRX5. If equal to approximately 2000hPa/2bar, go to the next step. If not, perform a max. calibration by pressing the max. button. The TRX5 OUT2 value is now equal to 2000hPa/2bar
Checking the flow sensors readings
o Reconnect the absorber and the patient breathing system in a correct way. o Set a fresh gas flow for example of 4 liters/min. The TRX3 OUT2 and
TRX4 OUT2 value must be a positive value. If one of these readings is negative, you must check all tubes inside and outside the ventilator for proper connection.
Warning: Leave the calibration menu by pressing the exit button. Now you’re sure that all the changes made in the calibration menu are stored to the EEPROM memory!
Neptune ventilator Technical manual 113
3.5 THE PRESSURE REGULATORS
1. Air pressure regulator 3000hPa/3bar.
2. Air pressure regulator 500hPa/0.5bar.
12
X9
X10X6X7
Neptune ventilator Technical manual 114
Check / calibrate AIR pressure regulator 3000hPa/3bar (1).
o For the adjustment of the AIR pressure regulators, you need a test patient breathing unit.
Test patient breathing unit
o Make sure that the AIR input pressure is connected. Check the AIR input pressure on the manometer located on the front. The pressure must be higher than 3000hPa /3 bar.
o Remove the patient breathing unit. o Connect the test patient breathing unit with the bottle. o Connect a pressure gauge with the connector output G. o Enter the calibration menu. o Set VALVE6 ON. Don’t forget to set the +12V VALVES voltage
ON. o Check the pressure gauge readings. If equal to approximately
3000hPa /3bar, go to the next step. Otherwise, unfasten the locking ring of the AIR pressure regulator 1 and adjust the pressure so it’s equal to 3000hPa/3 bar (+/- 100hPa /0.1bar) by means of rotating the spindle. Fasten the locking ring again so that the adjustment spindle is locked.
o Set VALVE6 OFF.
Neptune ventilator Technical manual 115
Check / calibrate AIR pressure regulator 500hPa/0,5 bar (2).
o Make sure that the AIR input pressure is connected. Check the AIR input pressure on the manometer located on the front. The pressure must be higher than 3000hPa /3bar.
o Remove the patient breathing unit. o Connect the test patient breathing unit with the bottle. o Connect a pressure gauge with the connector output A. o Enter the calibration menu. o Set valve 13 and 11 ON. Don’t forget to set the +12V VALVES
voltage ON. o Check the pressure gauge readings. If equal to approximately 500hPa
/0,5 bar go to the next step. Otherwise, unfasten the locking ring of the AIR pressure regulator 2 and adjust the pressure so it’s equal to 500hPa /0,5bar (+/- 50hPa/0.5bar) by means of rotating the spindle. Fasten the locking ring again so that the adjustment spindle is locked.
o Set valve 13 &11 OFF.
Neptune ventilator Technical manual 116
3.6 THE FLOW REGULATORS For the adjustment of the flow regulators you need the test patient breathing unit and a flowmeter. Follow the next steps:
o Remove the patient breathing unit. o Connect the test patient breathing unit with the bottle. o Connect a flowmeter with the connector output G. o Enter the calibration menu. o Set VALVE6 ON. Don’t forget to set the +12V VALVES voltage ON. o Check the flowmeter readings. If equal to 4 liters/min, go to the next step.
Otherwise, unfasten the locking ring of the flow regulator and adjust the flow so it’s equal to 4 liters/min (+/-0.1lpm) by means of rotating the spindle. Fasten the locking ring again so that the adjustment spindle is locked.
Neptune ventilator Technical manual 117
o Set VALVE6 OFF. o Connect the flowmeter with the connector output A. o Set VALVE11 and VALVE13 ON. o Check the flowmeter readings. If equal to 1.6 liters/min, go to the next
step. Otherwise, unfasten the locking ring of the flow regulator and adjust the flow so it’s equal to 1.6 liters/min (+/-0.1lpm) by means of rotating the spindle. Fasten the locking ring again so that the adjustment spindle is locked.
o Set all the valves OFF.
Neptune ventilator Technical manual 118
3.7 O2 FLUSH
o Connect flowmeter to front outlet. o Select the front fresh gas output. o Push O2 flush button.
o Check the flowmeter readings. If equal to 35 liters/min, go to the next
step. Otherwise, unfasten the locking ring of the flow regulator and adjust the flow so it’s equal to 35 liters/min (+/- 0.5lpm) by means of rotating the spindle. Fasten the locking ring again so that the adjustment spindle is locked.
Neptune ventilator Technical manual 119
3.8 THE INPUT PRESSURE SWITCHES To adjust the input pressure switches, follow the next steps:
o Regulate the pressure at the O2 NIST input connector until it’s equal to 2500hPa/2.5 bar (+/-100hPa /0.1bar).
o The input indicator I2 / O2 must be lit. If not, adjust the screw in the middle of the pressure switch until the I2 / O2 indicator is lit.
o Let the O2 NIST input pressure decrease slowly.
The I2 / O2 indicator will go out with a pressure equal to 2200hPa 2.2bar (+/-200hPa/0.2bar). Otherwise, adjust the screw a little bit more.
O2
Neptune ventilator Technical manual 120
o Repeat the previous steps to adjust the input pressure switch of N2O.
N2O
Neptune ventilator Technical manual 121
o Regulate the pressure at the AIR NIST input connector until it’s equal to 2 bar and set switch drive gas to AIR.
o The input indicator DRV must be lit. If not, adjust the screw in the
middle of the drive gas pressure switch until the I1 / DRV indicator is lit.
Set switch drive gas to AIR.
DRV
Neptune ventilator Technical manual 122
o Let the AIR NIST input pressure decrease slowly. o The I1 / DRV indicator will go out with a pressure equal to
1700hPa/1.7bar (+/-170hPa/0.17bar). Otherwise, adjust the screw a little bit more.
o Regulate the pressure at the AIR NIST input connector until it’s equal to
2000hPa /2 bar (+/-200hPa/0.2bar). o The input indicator AIR must be lit. If not, adjust the screw in the
middle of the drive gas pressure switch until the I3 / AIR indicator is lit.
AIR
Neptune ventilator Technical manual 123
o Let the AIR NIST input pressure decrease slowly. The I3 / AIR indicator will go out with a pressure equal to 2200hPa 2bar (+/-200hPa/0.2bar). Otherwise, adjust the screw a little bit more.
Neptune ventilator Technical manual 124
3.9 THE BOTTLE SAFETY VALVE This section of the manual describes the procedure to check and adjust the bottle safety valve:
o Remove the patient breathing unit. o Connect the test patient breathing unit with the bottle. o Connect a calibrated pressure gauge to connector output C. o Connect connector output G to connector output D. o Set VALVE3 ON. o Set VALVE6 ON. o The pressure will increase until the bottle safety valves opens. o Check the pressure gauge readings. The valve relief pressure must be
equal to 120hPa/120mbar. If not, adjust the screw at the top of the bottle safety valve so that the pressure is equal to 120hPa/120mbar.
o Leave the calibration menu by pressing the exit button. o Reconnect the absorber and the patient breathing system in a correct way.
Neptune ventilator Technical manual 125
3.10 Peep valve calibration
• Connect tubes on patient circuit. • Cork patient Y • Go to calibration menu • Activate 12V valves. • Activate V3 & V6. • Set fresh gas flow 8Lpm
• Turn housing peep valve clockwise or counter clockwise and regulate until the reading on OUT2 for TRX1 = 01hPa /1mbar
• Set lock screw free before turning the peep housing.
Tighten lock screw ,don’t use high force when turning it in.
+ -
Neptune ventilator Technical manual 126
3.11 VALVES TEST This section of the manual describes how to check the different valves on there functionality and for possible leaks:
1. Leakage check V7,V8 and X2. o Enter the calibration menu. o Activate valve 5 and 12V valves to empty the TANK. o When Tank is empty, Set 12 v valves OFF and valve 5 OFF. o Verify or there is a pressure increase on sensor TRX5. If so, you have to
determine which valve is leaking. Check valves V7, V8, and X2/V4.
2. Leakage check Tank
o Enter the calibration menu. o Activate 12V valves. o Activate valve V7 and let the pressure in Tank increase to
1000hPa/1bar. o Close V7. o Verify or there is a pressure drop on sensor TRX5. o Pressure in tank may not drop more than 5hPa/5mbar in 1 minute. o In case of leakage, check connections with tank, check valve
X3/V5,check tank (O-rings).
3. Test V7, V8 , X2/V4 , X3/V5. o Enter the calibration menu. o Activate 12V valves. o Activate valve V7 and let the pressure in Tank and bottle increase to
1000hPa/1bar. o Close V7. o Check/replace V7 if loading from tank is not possible o Activate V5 to discharge tank. o Activate valve V8 and let the pressure in Tank and bottle increase to
1000hPa/1bar. o Close V8.
Neptune ventilator Technical manual 127
o Check/replace V8 if loading from tank is not possible Activate V5 to discharge tank.
o Activate valve V4 and let the pressure in Tank and bottle increase to 1000hPa/1bar.
o Close V4. o Check/replace X2/V4 if loading from tank is not possible o Activate V5 to discharge tank.
4. Leakage check Bottle.
o Enter the calibration menu. o Remove the patient breathing unit. o Connect the test patient breathing unit with the bottle. o Use a tube to connect output connection C with output connection D. o Activate valve 5 and valve 3. o Activate valve V7 and let the pressure in Tank and bottle increase to
50hPa/50mbar o Switch valve 7 OFF. o Verify or there is a pressure drop on sensor TRX5. If so, you have to
determine which valve is leaking. Check valves X5, X1, V1,V2,V14,V15 and V16
o Leave the calibration menu by pressing the exit button. o Reconnect the absorber and the patient breathing system in a correct way.
5. Test V1, V2 , V14, V15and V16.
o Enter the calibration menu. o Remove the patient breathing unit. o Connect the test patient breathing unit with the bottle. o Use a tube to connect output connection C with output connection D. o Activate valve 5 and valve 3. o Activate valve V7 and let the pressure in Tank and bottle increase to
50hPa/50mbar o Switch valve 7 OFF. o ActivateV1 o Pressure on TRX5 has be zero within 15 sec if not check/replace valve
V1 o Set valve V1 OFF o Activate valve V7 and let the pressure in Tank and bottle increase to
50hPa/50mbar o Switch valve 7 OFF. o ActivateV2 o Pressure on TRX5 has be zero within 15 sec if not check/replace valve
V2 o Set valve V2 OFF o Activate valve V7 and let the pressure in Tank and bottle increase to
50hPa/50mbar o Switch valve 7 OFF. o ActivateV14
Neptune ventilator Technical manual 128
o Pressure on TRX5 has be zero within 15 sec if not check/replace valve V14
o Set valve V14 OFF o Activate valve V7 and let the pressure in Tank and bottle increase to
50hPa/50mbar o Switch valve 7 OFF. o ActivateV15 o Pressure on TRX5 has be zero within 15 sec if not check/replace valve
V15 o Set valve V15 OFF o Activate valve V7 and let the pressure in Tank and bottle increase to
50hPa/50mbar o Switch valve 7 OFF. o ActivateV16 o Pressure on TRX5 has be zero within 15 sec if not check/replace valve
V16 o Set valve V16OFF
o Leave the calibration menu by pressing the exit button. o Reconnect the absorber and the patient breathing system in a correct way.
Neptune ventilator Technical manual 129
3.12 PERFORMING A LEAKTEST 3.12.1 Entering the leaktest menu Set the ventilator to standby mode and select leak test. Always disconnect the patient before entering the leak test menu. Note: It is not possible to enter the leak test menu when the airway pressure is greater than 2hPa/2mbar. Disconnect patient, so that pressure can decrease to 0hPa/0mbar. The leak test menu is drawn in accordance with the next drawing.
Neptune ventilator Technical manual 130
1. Patient circuit + internal bag button: press this button to select the patient breathing circuit and the internal balloon leaktest.
2. Patient circuit + handset button: press this button to select the patient
breathing circuit and the external handset.
3. EXIT: select and press encoder knob to leave the leaktest menu.
4. PEEP valve leakage detected: this message appears when the PEEP valve leakage test fails.
5. Information window: follow these instructions to perform a leaktest.
6. Barograph: indicates airway pressure.
6
1
4
2
5 3
Neptune ventilator Technical manual 131
3.12.2 Performing the leaktest You can perform a leaktest on:
- the patient breathing circuit and the external handset. - the patient breathing circuit and the internal balloon.
To perform a leaktest on the patient breathing circuit and the external handset, proceed as follows:
- Select . - Make all the connections in the patient breathing unit; see the operating
instructions of the CO2 absorber. - Close safety valve on handset. - Close / disconnect sample tube of gas monitor (optional). - Close the output of the Y-piece. - Set a fresh gas flow. - Let the airway pressure increase to 30 hPa. - Close the fresh gas flow. - The time in which the pressure decreases to 20 hPa must be
15 seconds or longer. If the leak test is not within the specifications, check all connections in the patient breathing circuit. If no leaks can be found, contact your service department. After successfully performing the previously described leaktest, you can perform a leaktest on the patient breathing circuit and the internal balloon. Proceed as follows:
- Select . - Close the output of the Y-piece. - Set a fresh gas flow. - Let the airway pressure increase to 30 hPa. - Close the fresh gas flow. - The time in which the pressure decreases to 20 hPa must be
15 seconds or longer. If the leak test is not within the specifications, check all connections in the patient breathing circuit. Also check the internal balloon for leaks. If no leaks can be found, contact your service department.
Neptune ventilator Technical manual 132
When is selected, at the same time the PEEP valve located in the patient breathing unit is filled once with a pressure of ±100hPa/±100mbar. The pressure in the PEEP valve will continuously be monitored. If a pressure drop is detected within 15 sec, the message "PEEP valve leakage detected" will be displayed:
To repeat the PEEP valve test, select
followed by To determine either the PEEP leakage is intern in the ventilator or extern in the patient breathing unit, remove the patient circuit from bottle and close the PEEP output with finger in accordance with next drawing.
Keep your finger on the PEEP output and repeat the PEEP valve test. If the PEEP valve test succeeds, the problem is located in the patient breathing unit, otherwise the problem is probably intern in the ventilator.
PEEP output
Neptune ventilator Technical manual 133
. Leakage in the patient breathing unit
o Check/replace mushroom valve and O-ring in the patient breathing unit. o Check/replace blue O-ring in feed-through plate.
. Leakage located in the ventilator
o Remove the patient breathing unit. o Connect the test patient breathing unit with the bottle. o Connect a calibrated pressure gauge to connector output A. o Set valve 13 and 12 ON. o Set valve 11 ON until the pressure is between 80-100hPa/80 – 100mbar. o Set valve 11 OFF. o Check the pressure gauge readings and check for pressure drop.
Neptune ventilator Technical manual 134
o Set valve 10 and valve 9 ON. o Set valve 13 OFF. o The PEEP reservoir will be filled with a certain amount off pressure
delivered by the UPPER LIMIT reservoir. o Check the pressure gauge readings and check for pressure drop. o Leave the calibration menu by pressing the exit button. o Reconnect the absorber and the patient breathing system in a correct way.
3.12.3 Leaving the leaktest menu
Leave the leaktest menu by selecting the exit . The ventilator will return to standby mode.
Neptune ventilator Technical manual 135
3.13 MAINTENANCE INSTRUCTIONS Warning: It is recommended that the Neptune should be checked at least twice a year by the Medec Benelux NV service department or by specially trained personnel. There are a number of parts which require either periodic maintenance or replacement.
Patient breathing unit
• The mushroom valve and the switching diaphragm are to be replaced each service.
• The MAN/CMV changeover valve is to be inspected at each service and if
necessary cleaned. It is replaced annually.
• The expiration valve (under the mushroom valve) is replaced annually. When fitting, it’s important to guide the valve gently onto its seating.
• Check the flow sensor at each service and clean with alcohol if necessary.
• Replace the O-rings (see detailed drawing patient breathing unit).
• Replace the bag.
Warning: Always replace with same type of bag to ensure satisfactory performance. In fact, the specifications of the Neptune are greatly dependent on the elasticity of the balloon. Warning: Never immerse the patient breathing unit in a liquid and only fit a dry unit!
Neptune ventilator Technical manual 136
8
16
16
2 7
9
10
10
4
1
10
10
6
Neptune ventilator Technical manual 137
Bottle
• The blue O-ring connections between the patient breathing unit and the bottle are checked and tested for gas tightness at each service. Replace the O-rings every two years.
3 11
12
13
13
14
Neptune ventilator Technical manual 138
Solenoid valves
• The solenoid valves must be checked for proper working each service time.
Bottle venting valve
• Replace the bottle venting valve on the back of the bottle at each service.
Bottle safety valve
• Check the bottle safety valve annually and, if necessary, re-adjust to ±120hPa/120mbar.
5
Neptune ventilator Technical manual 139
Fresh gas flow sensor
• Inspect the sensor annually for pollution. The sensor can be checked easily by removing the plug on the side.
Neptune ventilator Technical manual 140
3.14 PARTS LIST
• Service Parts – Patient Circuit (half-yearly check)
Item Quantity Description Part number 1 1 Switching diaphragm 121213000 2 1 Mushroom valve 121227000 3 1 Gas bag 250035000 4 1 O-ring 19X1 501190100
• Service Parts – Patient Circuit (annual check)
Item Quantity Description Part number 1 1 Switching diaphragm 121213000 2 1 Mushroom valve 121227000 3 1 Gas bag 250035000 4 1 O-ring 19X1 501190100 5 1 Venting valve 121250000 6 1 Valve 121236000 7 1 Expiration valve 124228000 8 1 O-ring 14 x 1,78 silic. 70° shore 505140100 9 1 O-ring 18 x 2 501180200 10 4 O-ring 32 x 2 501320200 11 1 O-ring 58 x 3,5 501580300 12 1 O-ring 25 x 2,5 501250100 16 2 O-ring 23.52 x 1,78 501235100
• Service Parts (two-yearly check)
Item Quantity Description Part number 13 4 O-ring 3 x 3 silic. 70° blue 505030300 14 1 O-ring 10 x 3 silic. 70° blue 505100500
• Service Parts (tree yearly check)
Item Quantity Description Part number 15 1 Battery 12v 7.2 Ah 668202000
Neptune ventilator Technical manual 141
4. CLASSIFICATION AND DISCARD
4.1 CLASSIFICATION
• Class I/ Type B/ IPX1. • Do not use in presence of flammable anaesthetics.
Neptune ventilator Technical manual 142
4.2 DISCARD There is no risk for explosion when discarding. The electronic boards and display contain a small amount of lead. The battery is a sealed lead acid type. Please adhere to governmental related laws or regulations when discarding the Neptune.
Neptune ventilator Technical manual 143
5. Inspection
5.1 VISUAL INSPECTION o Wheels and brakes.
Check or the can rotate freely and verify or the brakes can be activated. o Drawer unit.
The drawer must go in and out drawer case smoothly. o Absorber.
Empty canister, check glass on visual damage. o Clean valve domes ,the dome glass must be free of contamination, valves must be visual without
removing the valve domes. o Check suction jar on visual damage. o Check ventilator case, front layers, pressure gauges , keyboard , display and knobs on visual
damage. o Check flowmeter glass and spindles on visual damage. o Check cover plate on visual damage. o Check manual balloon and arm, the arm must move freely with no visual damage. o Check patient circuit tubing o Check / clean AIR input pressure watertrap reservoir . The watertrap glass must be free of
contamination so that, in case there is water inside it is visual from the outside. Check filter inside watertrap and replace when contaminated. Check wheels and brakes.
Neptune ventilator Technical manual 144
5.2 OTHER INSPECTIONS 1.Ventilator keyboard & encoder: The functionality from the keys and encoder can be tested in the mode setup, the silent button can be tested in MAN mode. 2.Speaker test: The functionality speaker can be tested in setup A beep is generated when the volume level is changed. 3.Mains/ battery indicator : Set mains switch on and check mains indicator on Keyboard. Set ventilator ON. Check mains indicator on display. Set mains switch OFF and check batttery indicator, Led on keyboard must go OFF. 4.Fan and filter: The Fan has to rotate without making high noise level. Check filter at the bottom, replace/ clean if necessary.
5.Battery: Set mains switch OFF and verify or the ventilator can work for 1 hour on battery. Before starting this test be sure that the ventilator was connected with the mains power for at least 4 hours.
6.Internal connections: Check internal tubing and electrical connections. All tubing must be clear ,replace tubing who have indications from contamination or humidity. 7.Pressure regulators: See technical manual section 3.5 THE PRESSURE REGULATORS.
Neptune ventilator Technical manual 145
8.Flow regulators: See technical manual section 3.6 THE FLOW REGULATORS 9.O2 Flush: See technical manual section 3.7 O2 FLUSH 10.Pressure switches: See technical manual section 3.8 THE INPUT PRESSURE SWITCHES 11.ZERO and MAX calibration transducers : See technical manual section 3.4 THE PRESSURE TRANSDUCERS 12.Bottle safety valve: See technical manual section 3.9 THE BOTTLE SAFETY VALVE 13.Valve test: See technical manual section 3.10 VALVES TEST 14.Front absorber selector: Set selector knob in front position and verify or fresh gas comes out on front. Try to go in MAN mode. Check message on screen “Low Drive pressure” or “Fresh gas front output selected”. Set selector knob in absorber position and verify or fresh gas comes out on absorber. 15.Suction:
Be sure that vacuum is connected on ventilator.
Set Suction ON , block input suction jar and check for negative pressure on pressure gauge. Set suction OFF, No negative pressure may be seen on pressure gauge when blocking the input from the suction jar with the suction selector in OFF position. 16. Handset safety valve: Set the valve in close position and perform a leakage test. Close patient Y and set flow of 10Lpm Open fresh gas flow and verify max pressure, when valve is completely closed The max pressure that can build up is 60 hPa +/_ 5hPa 17.O2 Measurement: Go to setup menu section O2 calibration and perform a 21% and 100% calibration.
Neptune ventilator Technical manual 146
Check also wire and connection with O2 cell. See users manual section
18.Connections: Check all external connection’s. 19.Absorber:
See technical manual Absorber
20.Flowmeter:
See technical manual Flowmeter
21.Gasovac: See technical manual Gasovac 22.Leaktest: See technical manual section 3.11 PERFORM A LEAKTEST 24.AERTH connection and mains output:
Disconnect the mains power cable and measure resistance between earth connection on mains plug and earth connection on mains outlet on top cover plate. The resistance must be less then 0.2 Ohm. Measure resistance between earth connection on mains plug and cover plate The resistance must be less then 0.2 Ohm. 25.Working: Connect ventilator with test lung. Let the ventilator work in CMV mode with following parameters. TV: 500 PEEP 0hPa Freq: 16 Insp.Pause 0 I:E 1/2 Freshgas 1LPM Check volume on display with volume on testlung.
Neptune ventilator Technical manual 147
6. TROUBLESHOOTING
Error code: 01 7 segment display code
Tidal volume : Err Peak : 01 Plateau : 01
LCD display Error detected / error 01 / MMI board / Internal program memory error
Problem: MMI board - Internal program memory error
Action: See section 1.4 MMI BOARD- Microprocessor internal function test.
Error code: 02 7 segment display code
Tidal volume : Err Peak : 01 Plateau : 02
LCD display Error detected / error 02 / MMI board / Internal SRAM error
Problem: MMI board - Internal SRAM error
Action: See section 1.4 MMI BOARD- Microprocessor internal function test.
Error code: 03 7 segment display code
Tidal volume : Err Peak : 01 Plateau : 03
LCD display Error detected / error 03 / MMI board / Internal timer error
Problem: MMI board - Internal timer error
Action: See section 1.4 MMI BOARD- Microprocessor internal function test.
Neptune ventilator Technical manual 148
Error code: 04 7 segment display code
Tidal volume : Err Peak : 01 Plateau : 04
LCD display Error detected / error 04 / MMI board / Internal EEprom error
Problem: MMI board - Internal EEprom error Action: Reset display parameters. See section 1.4 MMI BOARD- Microprocessor internal function test Resolve internal function test error code 4 – Internal EEPROM error
Error code: 05 7 segment display code
Tidal volume : Err Peak : 01 Plateau : 05
LCD display Error detected / error 05 / MMI board / Internal watchdog error
Problem: MMI board - Internal watchdog error
Action: See section 1.4 MMI BOARD- Microprocessor internal function test.
Error code: 11 7 segment display code
Tidal volume : Err Peak : 02 Plateau : 01
LCD display Error detected / error 11 / MMI board / External flash error
Problem: MMI board - External flash error
Action: See section 1.4 MMI BOARD- Microprocessor external function test Resolve external function test error code 1 – External flash error
Neptune ventilator Technical manual 149
Error code: 12 7 segment display code
Tidal volume : Err Peak : 02 Plateau : 02
LCD display Error detected / error 12 / MMI board / SRAM error
Problem: MMI board - External SRAM error
Action: See section 1.4 MMI BOARD- Microprocessor external function test Resolve external function test error code 2 – External SRAM error
Error code: 13 7 segment display code
Tidal volume : Err Peak : 02 Plateau : 03
LCD display Error detected / error 13 / MMI board / external display driver error
Problem: MMI board - External display driver error
Action: See section 1.4 MMI BOARD- Microprocessor external function test Resolve external function test error code 3 – External display driver error
Error code: 15 7 segment code
Tidal volume : Err Peak : 02 Plateau : 05
LCD display Error detected / error 15 / MMI board / 7 Segment lockup
Problem: MMI board – 7 segement lockup
Action: See section 1.4 MMI BOARD- Microprocessor internal function test Perform the internal function test See section 1.4 MMI BOARD- 7 segment display test
Perform the 7 segment display test
Neptune ventilator Technical manual 150
Error code: 16 7 segment code
Tidal volume : Err Peak : 02 Plateau : 06
LCD display Error detected / error 16 / MMI board / Barograph lockup
Problem: MMI board – barograph lockup
Action: See section 1.4 MMI BOARD- Microprocessor internal function test Perform the internal function test See section 1.4 MMI BOARD- barograph test
Perform the barograph test
Error code: 17 7 segment code
Tidal volume : Err Peak : 02 Plateau : 07
LCD display Error detected / error 17 / MMI board / Frontleds lockup
Problem: MMI board – Frontleds lockup
Action: See section 1.4 MMI BOARD- Microprocessor internal function test Perform the internal function test See section 1.4 MMI BOARD- led indicator test
Perform the led indicator test
Error code: 18 7 segment code
Tidal volume : Err Peak : 02 Plateau : 08
LCD display Error detected / error 18 / MMI board / Buzzer lockup
Problem: MMI board – buzzer lockup
Action: See section 1.4 MMI BOARD- Microprocessor internal function test Perform the internal function test See section 1.4 MMI BOARD- buzzer test
Perform the buzzer test
Neptune ventilator Technical manual 151
Error code: 19 7 segment code
Tidal volume : Err Peak : 02 Plateau : 09
LCD display Error detected / error 19 / MMI board / LCD lockup
Problem: MMI Board - LCD lockup
Action: See section 1.4 MMI BOARD- Microprocessor internal function test Perform the internal function test See section 1.4 MMI BOARD- showing a test pattern on the graphic display
Perform the lcd pattern test See section 1.4 MMI BOARD- Microprocessor external function test Resolve external function test error code 3 – External display driver
Error code: 20 7 segment code
Tidal volume : Err Peak : 02 Plateau : 04
LCD display Error detected / error 20 / MMI board / keyboard / encoder failure
Problem: MMI board - Keyboard controller – Other failure Problems with keyboard layer/circuit or the encoder knob.
Action: Verify that there is no key pressed. Verify the encoder knob is not locked. See section 1.4 MMI BOARD- Keyboard test
Perform the keyboard test See section 1.4 MMI BOARD- Encoder test Perform the encoder test
Error code: 28 7 segment code
Tidal volume : Err Peak : 03 Plateau : 01
LCD display Error detected / error 28 / communication 25 ms timeout
Problem: Board communication error - 25 ms timeout Communication with other boards is not possible.
Action: A communication problem can be the result from problems on the display board but also from problems on other boards.
Neptune ventilator Technical manual 152
In case of communication error: Perform an internal test for all boards.
And verify or the tests on all boards pass. See section 1.4 MMI BOARD - Microprocessor internal function test. See section 1.7 MASTER BOARD - Microprocessor internal function test. See section 1.8 PNEUMATIC BOARD - Microprocessor internal function test.
Perform an external test for all boards. See section 1.4 MMI BOARD – Microprocessor external function test. See section 1.7 MASTER BOARD - Microprocessor external function test. See section 1.8 PNEUMATIC BOARD - Microprocessor external function test.
Perform a communication test between MASTER and DISPLAY board See section 1.7 MASTER BOARD - Communication test between master board and display board
Perform a communication test between MASTER and PNEUMATIC board
See section 1.7 MASTER BOARD - Communication test between master board and pneumatic board
Error code: 29
7 segment code Tidal volume : Err Peak : 03 Plateau : 02
LCD display Error detected / error 29 / Parity error start-up
Problem: Board communication error - Parity error start-up Communication with other boards is not possible.
Action: Same action as for Error code: 28
Error code: 30 7 segment code
Tidal volume : Err Peak : 03 Plateau : 03
LCD display Error detected / error 30 / parity error start-receive byte
Problem: Board communication error - Parity error start-receive byte Communication with other boards is not possible.
Action: Same action as for Error code: 28
Neptune ventilator Technical manual 153
Error code: 31 7 segment code
Tidal volume : Err Peak : 03 Plateau : 04
LCD display Error detected / error 31 / Receive string length error
Problem: Board communication error - Receive string length error Communication with other boards is not possible.
Action: Same action as for Error code: 28
Error code: 32 7 segment code
Tidal volume : Err Peak : 03 Plateau : 05
LCD display Error detected / error 32 / Eot counter out of range
Problem: Board communication error - Eot counter out of range Communication with other boards is not possible.
Action: Same action as for Error code: 28
Error code: 33 7 segment code
Tidal volume : Err Peak : 03 Plateau : 06
LCD display Error detected / error 33 / Target slave not ready
Problem: Board communication error - Target slave not ready Communication with other boards is not possible.
Action: Same action as for Error code: 28
Neptune ventilator Technical manual 154
Error code: 34 7 segment code
Tidal volume : Err Peak : 03 Plateau : 07
LCD display Error detected / error 34 / Transmit string wrong
Problem: Board communication error - Transmit string wrong Communication with other boards is not possible.
Action: Same action as for Error code: 28
Error code: 41 7 segment code
Tidal volume : Err Peak : 00 Plateau : 00
LCD display Error detected / error 41 / Pneumatic board / Global error
Problem: Pneumatic board global error.
Action: Perform an internal test on pneumatic board.
See section 1.8 PNEUMATIC BOARD - Microprocessor internal function test.
Perform an external test on pneumatic board. See section 1.8 PNEUMATIC BOARD - Microprocessor external
function test.
Error code: 44 7 segment code
Tidal volume : Err Peak : 00 Plateau : 00
LCD display Error detected / error 44 / Pneumatic board / Not able to start A/D converter
Problem: Pneumatic board – Not able to start A/D converter Problem with analog digital converter on pneumatic board.
Action: Perform an external test on pneumatic board. See section 1.8 PNEUMATIC BOARD - Microprocessor external function test.
Neptune ventilator Technical manual 155
Error code: 45 Tidal volume : Err Peak : 00 Plateau : 00
LCD display Error detected / error 45 / Pneumatic board / A/D converter busy time-out
Problem: Pneumatic board – A/D converter busy time-out Problem with analog digital converter on pneumatic board.
Action: Perform an external test on pneumatic board. See section 1.8 PNEUMATIC BOARD - Microprocessor external function test.
Error code: 46 Tidal volume : Err Peak : 00 Plateau : 00
LCD display Error detected / error 46 / Pneumatic board / A/D readings equal to $FFF
Problem: Pneumatic board – A/D readings equal to $FFF Problem with analog digital converter on pneumatic board.
Action: Perform an external test on pneumatic board. See section 1.8 PNEUMATIC BOARD - Microprocessor external function test.
Error code: 47 Tidal volume : Err Peak : 00 Plateau : 00
LCD display Error detected / error 47 / Pneumatic board / A/D readings equal to $000
Problem: Pneumatic board – A/D readings equal to $000 Problem with analog digital converter on pneumatic board.
Action: Perform an external test on pneumatic board. See section 1.8 PNEUMATIC BOARD - Microprocessor external function test.
Neptune ventilator Technical manual 156
Error code: 48 Tidal volume : Err Peak : 00 Plateau : 00
LCD display Error detected / error 48 / Pneumatic board / Internal EEPROM memory error
Problem: Pneumatic board – Internal EEPROM memory error
Action: Perform an internal test on pneumatic board. See section 1.8 PNEUMATIC BOARD - Microprocessor internal function test Resolve internal function test error code 4 – Internal EEPROM error Reset from the pneumatic board EEPROM settings will AFFECT the calibration parameters. Recalibration of the ventilator is necessary in this case. Refer to the maintenance and calibration procedure described this manual!
Error code: 49 Tidal volume : Err Peak : 00 Plateau : 00
LCD display Error detected / error 49 / Pneumatic board / +12V input voltage too low
Problem: Pneumatic board – 12V input voltage too low
Action: Check 12V voltage on DC/DC converter.
Connector P1:
Pin number Description 3 +12V DC output voltage 4 GND
Perform a Pneumatic function test.
See section 1.8 PNEUMATIC BOARD - Pneumatic function test.
Neptune ventilator Technical manual 157
Error code: 50 Tidal volume : Err Peak : 00 Plateau : 00
LCD display Error detected / error 50 / Pneumatic board / +12V valves enable circuit error
Problem: Pneumatic board – 12V valves enable circuit error Input voltage is OK but circuit can’t switch ON the 12V valve voltage
Action: Disconnect valve connectors Perform. Perform a Pneumatic function test.
See section 1.8 PNEUMATIC BOARD - Pneumatic function test. Check for bad valve or short circuit if test is OK when valve
connectors are disconnected. Check circuit Q1 / Q2 when test is not OK and valve connectors
are disconnected. Q1
RFP30P05
+12v
Q2
BS170M
GND
A 1B 2CLR 3Q4
Q13
Cext 14
RCext 15
U8A
74HC123
GND
VCCR20
100K
R30
22K
R9
10K
C20
330N
Neptune ventilator Technical manual 158
Error code: 51 Tidal volume : Err Peak : 00 Plateau : 00
LCD display Error detected / error 51 / Pneumatic board / +12V valves disable circuit error
Problem: Pneumatic board – 12V valves disable circuit error Input voltage is OK but circuit can’t switch the 12V valve voltage OFF.
Action: Disconnect valve connectors Perform. Perform a Pneumatic function test.
See section 1.8 PNEUMATIC BOARD - Pneumatic function test. Check circuit Q1 / Q2
Q1
RFP30P05
+12v
Q2
BS170M
GND
A 1B 2CLR 3Q4
Q13
Cext 14
RCext 15
U8A
74HC123
GND
VCCR20
100K
R30
22K
R9
10K
C20
330N
Neptune ventilator Technical manual 159
Error code: 51
Tidal volume : Err Peak : 00 Plateau : 00
LCD display Error detected / error 52 / Pneumatic board / Not able to fill tank
Problem: Pneumatic board – Not able to fill tank
A. When first starting in CMV or PCV mode the tank is filled with pressure for the first inspiration. There is a time limit on this filling and when this time is exceeds ERROR message 52 “Not able to fill tank ”is set on screen. B. During CMV or PCV mode, the tank will be filled with a calculated pressure during expiration, this filling has to be done before new inspiration starts. When this filling is still busy when new inspiration starts, ERROR message 52 “Not able to fill tank ”is set on screen.
Action: Situation A:
Low input pressure or selector knob fresh gas in front position and input pressure switch disabled.
Check selector knob fresh gas front absorber. Check input pressure ,more than 3000hPa/3bar. Check or input pressure switches are in ON position
Go to next step when previous checks are OK When the switches are in OFF position, the software will not check for low level input pressure and will try to fill tank by activating valve V4 or V7 and V8.When this filling is not possible within time, ERROR message 52 “Not able to fill tank ”is set on screen.
Go to calibration menu. Activate 12v valves. Activate valve V5 to discharge Tank. Verify tank pressure must be 0hPa/0mbar. Perform a zero calibration tank if not zero. Set valve V5 OFF. Activate valve 4 and let pressure in tank increase up to 2000hPa
2bar and set valve V4 OFF. If not possible, check valve 4 and X2.and connection V4.
Activate valve V5 to discharge Tank. .Set valve V5 OFF. Activate valve 7,let pressure in tank increase up to 2000hPa / 2bar
and set valve V7 OFF. If not possible, check valve 7 and connection V7.
Activate valve V5 to discharge Tank. .Set valve V5 OFF. Activate valve 8,let pressure in tank increase up to 2000hPa / 2bar
Neptune ventilator Technical manual 160
set valve V8 OFF. If not possible, check valve 8 & connection V8.
Situation B: Low input pressure or selector knob fresh gas to front position
switched during CMV or PCV mode and input pressure switch disabled.
Check selector knob fresh gas front absorber. Check input pressure ,more than 3000hPa / 3bar. Check or input pressure switches are in ON position
Go to next step when previous checks are OK When the switches are in OFF position, the software will not check for low level input pressure and will try to fill tank by activating valve V4 or V7 and V8.When this filling is not possible before next inspiration starts, ERROR message 52 “Not able to fill tank ”is set on screen.
Go to calibration menu. Activate 12v valves. Activate valve V5 to discharge Tank. Verify tank pressure must be 0hPa / 0mbar. Perform a zero calibration tank if not zero. Set valve V5 OFF. Activate valve 4 and let pressure in tank increase up to 2000hPa
2bar and set valve V4 OFF. If not possible, check valve 4 and X2. and connection V4.
Activate valve V5 to discharge Tank. .Set valve V5 OFF. Activate valve 7,let pressure in tank increase up to 2000hPa / 2bar
and set valve V7 OFF. If not possible, check valve 7 and connection V7.
Activate valve V5 to discharge Tank. .Set valve V5 OFF. Activate valve 8,let pressure in tank increase up to 2000hPa / 2bar
and set valve V8 OFF. If not possible, check valve 8 and connection V8.
Error code: 53
Tidal volume : Err Peak : 00 Plateau : 00
LCD display Error detected / error 53 / Pneumatic board Not able to stop tank filling
Problem: Pneumatic board –Not able to stop tank filling When the tank is filled and the valves to fill the tank are switched OFF a check is done or the pressure is within limits. If the pressure is higher than the calculated value ERROR message 53 “Not able to stop tank filling ”is set on screen.
Neptune ventilator Technical manual 161
A. When first starting in CMV or PCV mode, the tank will be discharged before filling. This has to be done within time .if not ERROR message 53 “Not able to stop tank filling ”is set on screen, because of high pressure in tank. B. During CMV or PCV mode, the tank will be filled with a calculated pressure during expiration, at the end of expiration the pressure in the tank is checked with the calculated pressure. When this pressure is higher then the calculated pressure value ERROR message 53 “Not able to stop tank filling ”is set on screen.
Action: Situation A:
Low input pressure or selector knob fresh gas in front position and input pressure switch disabled.
Check selector knob fresh gas front absorber. Check input pressure ,more than 3000hPa/3bar. Check or input pressure switches are in ON position
Go to next step when previous checks are OK When the switches are in OFF position, the software will not check for low level input pressure and will try to discharge tank pressure by activating valve V5.Activation from valve 5 is done electrical but X3 is pneumatically driven and activation will be not possible when there is no pressure. When discharging is not possible within time, ERROR message 53 “Not able to stop tank filling ”is set on screen.
Go to calibration menu. Activate 12v valves. Activate valve V5 to discharge Tank. If not possible check X3,V5 and connection V5. When X3/V5 is OK verify tank pressure, must be 0hPa / 0mbar. Perform a zero calibration tank if not zero. Perform a leakage check V7,V8 and X2
See section 3.10 VALVES TEST / Leakage check V7,V8 and X2.
Situation B: Checked Tank pressure at end of expiration is higher than
calculated value.
Go to calibration menu. Activate 12v valves. Activate valve V5 to discharge Tank. Verify tank pressure must be 0hPa / 0mbar. Perform a zero calibration tank if not zero. Set valve V5 OFF. Perform a leakage check V7,V8 and X2 See section 3.10 VALVES TEST / Leakage check V7,V8 and X2.
Neptune ventilator Technical manual 162
Error code: 54
Tidal volume : Err Peak : 00 Plateau : 00
LCD display Error detected / error 54 / Pneumatic board / Tank low flow valve close time
Problem: Pneumatic board – Tank low flow valve close time When the tank is filled and the valves to fill the tank are switched OFF a check is done or the pressure is within limits. If the pressure is higher than the calculated value and valves V7 and V8 are used to fill tank, ERROR message 54 “Tank low flow valve close time” is set on screen.
Action: Go to calibration menu. Activate 12v valves. Activate valve V5 to discharge Tank. Verify tank pressure must be 0hPa / 0mbar. Perform a zero calibration tank if not zero. Set valve V5 OFF. Perform a leakage check V7,V8
See section 3.10 VALVES TEST / Leakage check V7,V8 and X2.
Error code: 55
Tidal volume : Err Peak : 00 Plateau : 00
LCD display Error detected / error 55 / Pneumatic board / Tank high flow valve close time
Problem: Pneumatic board – Tank high flow valve close time When the tank is filled and the valves to fill the tank are switched OFF a check is done or the pressure is within limits. If the pressure is higher than the calculated value and valve V4 / X2 is used to fill tank, ERROR message 55 “Tank high flow valve close time” is set on screen.
Action: Go to calibration menu. Activate 12v valves. Activate valve V5 to discharge Tank. Verify tank pressure must be 0hPa / 0mbar. Perform a zero calibration tank if not zero. Set valve V5 OFF. Perform a leakage check valve X2.
See section 3.10 VALVES TEST / Leakage check V7,V8 and X2.
Neptune ventilator Technical manual 163
Error code: 56 Tidal volume : Err Peak : 00 Plateau : 00
LCD display Error detected / error 56 / Pneumatic board / End exp. pressure to high
Problem: Pneumatic board – End exp. pressure to high In CMV and PCV, the pressure is checked at the end of expiration. When this pressure is higher than the PEEP setting + 20 the ventilator Stops the working mode, switch over to manual and error message is set on screen.
Action: Check evacuation port at back side ventilator for obstruction.
Error code: 57
Tidal volume : Err Peak : 00 Plateau : 00
LCD display Error detected / error 57 / Pneumatic board / PIP press. higher than upp.lim
Problem: Pneumatic board – PIP press. higher than upp.lim. At end of inspiration CMV or PCV, the pip pressure is checked. If higher than the upper limit +20 hPa / +20mbar (MAX 105hPa / 105mbar) the ventilator stops the working mode, switch over to manual and error message is set on screen.
Action: Check PEEP valve in patient breathing unit. Check valve V13 and V10
Error code: 58 Tidal volume : Err Peak : 00 Plateau : 00
LCD display Error detected / error 57 / Pneumatic board / Not able to lower tank pressure
Problem: Pneumatic board – Not able to lower tank pressure. At start of expiration the tank is filled with pressure. When there is already a pressure in the tank that is higher than the calculated pressure, the tank pressure will be discharged. If this is not possible before next inspiration starts, the software will see that the discharging was still busy and an error message 58 is set on screen.
Action: Check valve V5 and X3.
Neptune ventilator Technical manual 164
CHECKLIST NEPTUNE. Done by:…………………….. Serial number:……………… Date:………/………/………… Total working hours:………
conform
Y/N
Remarks
1
Visual inspection
2
Keyboard & encoder
3
Speaker
4
Mains / battery indicator
5
Fan & filter
6
Battery
7
Internal connections
8
Pressure regulators
9
Flow regulators
10 O2 Flush
11
Pressure switches
12
Calibration transducer
13
Bottle safety valve
14
Valve test
15
Front/ absorber selector
16
Handset safety valve
17
Suction
18
O2 Measurement
19
Connections
20
Absorber
21
Flowmeter
22
Gasovac
23
Leak test
24
Earth connection
25
Working
Neptune ventilator Technical manual 165
CHECKLIST NEPTUNE. Replaced parts:
Part Nummer Quantity
Remark
121213000 Switching diaphragm
121227000 Mushroom valve
250035000 Gas bag
501190100 O-ring 19X1
121250000 Venting valve
121236000 Valve
124228000 Expiration valve
505140100 O-ring 14 x 1,78 silic. 70° shore
501180200 O-ring 18 x 2
501320200 O-ring 32 x 2
501580300 O-ring 58 x 3,5
501250100 O-ring 25 x 2,5
501235100 O-ring 23.52 x 1,78
505030300 O-ring 3 x 3 silic. 70° blue
505100500 O-ring 10 x 3 silic. 70° blue
668202000 Battery 12v 7.2 Ah