Neonatal Chest Compression Device
description
Transcript of Neonatal Chest Compression Device
![Page 1: Neonatal Chest Compression Device](https://reader035.fdocuments.us/reader035/viewer/2022062410/56816245550346895dd28248/html5/thumbnails/1.jpg)
Neonatal Chest Compression DeviceCourtney GallagherJillian ZeberAdvisor: Dr. Walsh, Vanderbilt NICU
![Page 2: Neonatal Chest Compression Device](https://reader035.fdocuments.us/reader035/viewer/2022062410/56816245550346895dd28248/html5/thumbnails/2.jpg)
Problem Statement• When performed in
conjunction with operations of the neck and lower abdomen, there is limited space for manual chest compressions.
• Manual chest compressions require both hands to be wrapped around the chest.
• 2 occurrences at the Vanderbilt NICU this past year.
![Page 3: Neonatal Chest Compression Device](https://reader035.fdocuments.us/reader035/viewer/2022062410/56816245550346895dd28248/html5/thumbnails/3.jpg)
Project Goal
• Small, easy to setup device that will perform automatic chest compressions on a 0-30 day old infant for use in the OR during a simultaneous procedure.
![Page 4: Neonatal Chest Compression Device](https://reader035.fdocuments.us/reader035/viewer/2022062410/56816245550346895dd28248/html5/thumbnails/4.jpg)
Performance Criteria• Small but adjustable to fit the size of the baby.
• Provide enough force for sufficient chest compressions to reduce the width to one third, 2 to 3 cm.
• Apply the 11-12 lbs of force directly below the nipples in the center of the chest. • Verified with bathroom scale and free weights.
• Maintain a rate of 80-100 compressions per minute for blood pumping.
• Simple and easy to use by Dr. Walsh in the NICU
![Page 5: Neonatal Chest Compression Device](https://reader035.fdocuments.us/reader035/viewer/2022062410/56816245550346895dd28248/html5/thumbnails/5.jpg)
Force Required for Chest Compressions
Applied Force (lbs) Adequate Compression
11.1 Yes
11.6 Yes
12.3 Yes
13.1 Yes
14.7 Yes
18 Yes
10.8 No
10.1 No
9.9 No
Desired Force Range: 11 – 12 lbs
![Page 6: Neonatal Chest Compression Device](https://reader035.fdocuments.us/reader035/viewer/2022062410/56816245550346895dd28248/html5/thumbnails/6.jpg)
Solution Proposal
Solenoid ValveElectronic Timer
Power Supply
15 cm
6-9 cm
2-3 cm
Adjustable depending on size of the baby
![Page 7: Neonatal Chest Compression Device](https://reader035.fdocuments.us/reader035/viewer/2022062410/56816245550346895dd28248/html5/thumbnails/7.jpg)
Solution Proposal
Solenoid ValveElectronic Timer
Power Supply
15 cm
5 cm
4-6 cm
Adjustable depending on size of the baby
![Page 8: Neonatal Chest Compression Device](https://reader035.fdocuments.us/reader035/viewer/2022062410/56816245550346895dd28248/html5/thumbnails/8.jpg)
Solution Proposal
3- WAY
Exhaust
Air Cylinder Air Compressor
![Page 9: Neonatal Chest Compression Device](https://reader035.fdocuments.us/reader035/viewer/2022062410/56816245550346895dd28248/html5/thumbnails/9.jpg)
Device Components• Single-acting spring return pneumatic cylinder
• Air compressor with coil hose
• Solenoid valve controlled by electronic timer• On/off time range: 0.1 s to 99 hr
• Power supply: wall outlet
• 1.1 inch diameter plunger to displace chest 2-3 cm
• Wheel-up cart supporting electronic components & air compressor
• Support structure:• Currently: stand with base and adjustable arm• Ideally: Swivel arm connected to side panel of GE OmniBed
![Page 10: Neonatal Chest Compression Device](https://reader035.fdocuments.us/reader035/viewer/2022062410/56816245550346895dd28248/html5/thumbnails/10.jpg)
Calculations
Air Cylinder Specs
• Bore size:F = p A = p π d2/4
whereF = force exerted (N)p = gauge pressure (N/m2, Pa)A = full bore area (m2)d = full bore piston diameter (m)
F = 11 lbs = 50 N p = 100 psi = 689.5 kPa
d = 10 mm ≈ 7/16”
• Stroke length:> 3 cm with extra space 5 cm ≈ 2”
Timer Settings
80/min 90/min 100/min
Cycle Frequency (s-1)
1.333 1.50 1.667
Cycle Length (s) 0.750 0.667 0.600On/Off time (s) 0.375 0.335 0.300
• On/Off time:Cycle Length = 1/Cycle FreqencyOn/Off time = Cycle Length/2
![Page 11: Neonatal Chest Compression Device](https://reader035.fdocuments.us/reader035/viewer/2022062410/56816245550346895dd28248/html5/thumbnails/11.jpg)
Factors
• Provide necessary but not excessive force (11-12 lbs = 5 kg)• Materials that can be sterilized• Comfortable• Quick setup• Potentially portable• Safety• Adjustable
![Page 12: Neonatal Chest Compression Device](https://reader035.fdocuments.us/reader035/viewer/2022062410/56816245550346895dd28248/html5/thumbnails/12.jpg)
Implementation of Factors into Design
• Sterilizing plunger and stopper with plastic drapes
• Dr. Walsh
• Vary force applied with the psi of compressed air applied to cylinder
• Variable rate with micro-timer
• Adjustable with flexible arm coming out of NICU bed
![Page 13: Neonatal Chest Compression Device](https://reader035.fdocuments.us/reader035/viewer/2022062410/56816245550346895dd28248/html5/thumbnails/13.jpg)
Evaluation• Isabel in the NICU Simulation Lab• Responds like an alive baby• Provides vitals• Can determine if chest compression are adequate
![Page 14: Neonatal Chest Compression Device](https://reader035.fdocuments.us/reader035/viewer/2022062410/56816245550346895dd28248/html5/thumbnails/14.jpg)
Previous Work• Researched existing adult compression devices
• AutoPulse, Lucas, Thumper
• Established goals and criteria that must be met• Discussed with advisors
• Measured force required to provide chest compressions
• Decided on a prototype design• Pneumatic cylinder
• Researched and purchased materials to build prototype• Pneumatic cylinder and Air compressor• Solenoid valve, timer• All necessary connections
• Tested simplified version of prototype• Verified applied force was correct• Ordered more parts
![Page 15: Neonatal Chest Compression Device](https://reader035.fdocuments.us/reader035/viewer/2022062410/56816245550346895dd28248/html5/thumbnails/15.jpg)
Current Work
• Finishing DesignSafe
• Getting in touch with GE contact• Kent Meeks
• Building complete version of prototype
• Testing full prototype
![Page 16: Neonatal Chest Compression Device](https://reader035.fdocuments.us/reader035/viewer/2022062410/56816245550346895dd28248/html5/thumbnails/16.jpg)
Future Work• After testing, make modifications and re-test
• Build flexible arm• Test
• Make Demo
• Evaluate effectiveness against traditional method
![Page 17: Neonatal Chest Compression Device](https://reader035.fdocuments.us/reader035/viewer/2022062410/56816245550346895dd28248/html5/thumbnails/17.jpg)
References• http://www.zoll.com/medical-products/cardiac-support-pump
/autopulse/
• http://content.onlinejacc.org/cgi/content/full/44/11/2214/FIG2
• http://www.omega.com/auto/pdf/SimpValvesguide.pdf
• http://www.gehealthcare.com/euen/maternal-infant-care/products/microenvironments/giraffe_omnibed/index.html
• http://www.engineeringtoolbox.com/pneumatic-cylinder-force-d_1273.html