NCAR Airborne Oxygen Instrument (AO2)

19
NCAR Airborne Oxygen Instrument (AO2) Vacuum ultraviolet absorption technique Xe lamp (147 nm) and CsI detector Adapted from shipboard design (Stephens et al., 2003) Active pressure and flow control to 10 -6 Switches every 2.5 seconds between sample and WT gas 5-second 1-sigma precision of ± 2 per meg Factor of 2-5 motion degradation is correctable O 2 sensor Flow control:

description

NCAR Airborne Oxygen Instrument (AO2). Flow control:. Vacuum ultraviolet absorption technique Xe lamp (147 nm) and CsI detector Adapted from shipboard design (Stephens et al., 2003) Active pressure and flow control to 10 -6 Switches every 2.5 seconds between sample and WT gas - PowerPoint PPT Presentation

Transcript of NCAR Airborne Oxygen Instrument (AO2)

Page 1: NCAR Airborne Oxygen Instrument (AO2)

NCAR Airborne Oxygen Instrument (AO2)

• Vacuum ultraviolet absorption technique

• Xe lamp (147 nm) and CsI detector

• Adapted from shipboard design (Stephens et al., 2003)

• Active pressure and flow control to 10-6

• Switches every 2.5 seconds between sample and WT gas

• 5-second 1-sigma precision of ± 2 per meg

• Factor of 2-5 motion degradation is correctable

O2 sensor

Flow control:

Page 2: NCAR Airborne Oxygen Instrument (AO2)

NCAR Airborne Oxygen Instrument (AO2)

System components:

(fridge trap)

Page 3: NCAR Airborne Oxygen Instrument (AO2)

January 20, 2009

HIPPO1 Profile at 65 S

Southern Ocean O2 outgassing

Page 4: NCAR Airborne Oxygen Instrument (AO2)
Page 5: NCAR Airborne Oxygen Instrument (AO2)

(shorter preflight flow period)

Page 6: NCAR Airborne Oxygen Instrument (AO2)

(fridge trap + pre- and post- flight dry air purge)

Page 7: NCAR Airborne Oxygen Instrument (AO2)

(after linear time-in-flight correction)

Page 8: NCAR Airborne Oxygen Instrument (AO2)

Atmospheric Potential Oxygen:APO = O2 + 1.1*CO2

per meg

Page 9: NCAR Airborne Oxygen Instrument (AO2)
Page 10: NCAR Airborne Oxygen Instrument (AO2)
Page 11: NCAR Airborne Oxygen Instrument (AO2)
Page 12: NCAR Airborne Oxygen Instrument (AO2)
Page 13: NCAR Airborne Oxygen Instrument (AO2)

Summary

• AO2 has operated on every flight• HIPPO1 RF02 CO2_AO2 suffered from a cabin pressure effect and is not

reported• HIPPO3 RF06 O2_AO2 has several hours of missing data due to a noise

problem• HIPPO2 and HIPPO3 CO2_AO2 quality is degraded by inlet humidity effect

and not useful for rigorous comparison to other sensors• O2_AO2ADJ and APO_AO2ADJ are recommend variables for O2 analyses

Between now and HIPPO4

• Laboratory experiments to characterize inlet humidity effects on O2 and CO2

• Software improvements to reduce remaining motion sensitivity

Page 14: NCAR Airborne Oxygen Instrument (AO2)

HIPPO1 descent into ANC

Page 15: NCAR Airborne Oxygen Instrument (AO2)

NCAR Airborne Oxygen Instrument (AO2)8-hour flight

3-min cal

Page 16: NCAR Airborne Oxygen Instrument (AO2)

NCAR Airborne Oxygen Instrument (AO2)

• Inlet delay ~ 40 seconds with only a few second altitude dependency• Checked against OMS and VCSEL for 10 low and 10 high-alt

transitions per flight

Page 17: NCAR Airborne Oxygen Instrument (AO2)
Page 18: NCAR Airborne Oxygen Instrument (AO2)
Page 19: NCAR Airborne Oxygen Instrument (AO2)

January 12, 2009

HIPPO1 Profile at 80 N