    OPERATING INSTRUCTIONS AND SPECIFICATIONS

    NI 9214 with NI TB-921416-Channel Thermocouple Input Module withTerminal Block

    ni.com/manuals

    DeutschFranais

  • 8/13/2019 National Instruments 9214 manual

    2/44

    NI 9214 Operating Instructions and Specifications 2 ni.com

    This doc u ment describes how to u se the National Instr u ments 9214 inconj u nction with an NI TB-9214 terminal block and incl u desspecifications for the NI 9214 and terminal assignments for theNI TB-9214. Visit ni.com/info and enter rd s oftwarever s ion to determine which software yo u need for the mod u les yo u are u sing.For information abo u t installing, config u ring, and programming thesystem, refer to the system doc u mentation. Visit ni.com/info andenter c s erie s doc for information abo u t C Series doc u mentation.

    Visit ni.com/info and enter compatibility for informationabo u t chassis and carrier compatibility for the mod u les yo u areu sing.

    Note The safety g u idelines and specifications in this

    docu

    ment are specific to the NI 9214u

    sed in conju

    nctionwith an NI TB-9214. The other components in the systemmight not meet the same safety ratings and specifications.Refer to the doc u mentation for each component in thesystem to determine the safety ratings and specificationsfor the entire system. Visit ni.com/info and enterc s erie s doc for information abo u t C Seriesdoc u mentation.

  • 8/13/2019 National Instruments 9214 manual

    3/44

    National Instruments Corp. 3 NI 9214 Operating Instructions and Specifications

    Safety GuidelinesOperate the NI 9214 only as described in these operating

    instr u ctions.Hot Surface This icon denotes that the component may behot. To u ching this component may res u lt in bodily inj u ry.

    Safety Guidelines for Hazardous VoltagesIf hazardo u s voltages are connected to the mod u le, take thefollowing preca u tions. A hazardo u s voltage is a voltage greaterthan 42.4 V pk or 60 VDC to earth gro u nd.

    Caution Ens u re that hazardo u s voltage wiring isperformed only by q u alified personnel adhering to localelectrical standards.

    Caution Do not mix hazardo u s voltage circ u its andhu man-accessible circ u its on the same mod u le.

    Caution Make s u re that devices and circ u its connected tothe mod u le are properly ins u lated from h u man contact.

  • 8/13/2019 National Instruments 9214 manual

    4/44

    NI 9214 Operating Instructions and Specifications 4 ni.com

    Caution When mod u le terminals are hazardo u s voltageLIVE (>42.4 V pk /60 VDC), yo u m u st ens u re that devicesand circ u its connected to the mod u le are properlyins u lated from h u man contact. Yo u must u se theNI TB-9214 terminal block incl u ded with the NI 9214to ens u re that the terminals are not accessible.

    Fig u re 1 shows the NI TB-9214 terminal block.

    Figure 1. NI TB-9214 Terminal Block

    Note The NI TB-9214 contains a plastic insert to preventaccidental wire contact with the metal enclos u re.

  • 8/13/2019 National Instruments 9214 manual

    5/44

    National Instruments Corp. 5 NI 9214 Operating Instructions and Specifications

    Safety Guidelines for Hazardous LocationsThe NI 9214 is s u itable for u se in Class I, Division 2, Gro u ps A, B,C, D, T4 hazardo u s locations; Class I, Zone 2, AEx nA IIC T4, andEx nA IIC T4 hazardo u s locations; and nonhazardo u s locationsonly. Follow these g u idelines if yo u are installing the NI 9214 in apotentially explosive environment. Not following these g u idelinesmay res u lt in serio u s inj u ry or death.

    Caution Do not disconnect I/O-side wires or connectorsu nless power has been switched off or the area is knownto be nonhazardo u s.

    Caution Do not remove mod u les u nless power has beenswitched off or the area is known to be nonhazardo u s.

    Caution Su bstit u tion of components may impairsu itability for Class I, Division 2 and Zone 2.

    Caution For Division 2 and Zone 2 applications, install

    the system in an enclosu

    re rated to at least IP 54 asdefined by IEC 60529 and EN 60529.

    Caution For Division 2 and Zone 2 applications,connected signals m u st be within the following limit:

    Capacitance..........................0.2 F max

  • 8/13/2019 National Instruments 9214 manual

    6/44

    NI 9214 Operating Instructions and Specifications 6 ni.com

    Special Conditions for Hazardous Locations Use in EuropeThis eq u ipment has been eval u ated as Ex nA IIC T4 eq u ipmentu nder DEMKO Certificate No. 07 ATEX 0626664X. Each mod u leis marked II 3G and is s u itable for u se in Zone 2 hazardo u slocations, in ambient temperat u res of 40 C Ta 70 C. If yo u are u sing the NI 9214 in Gas Gro u p IIC hazardo u s locations, yo u mu st u se the device in an NI chassis that has been eval u ated asEx nC IIC T4, EEx nC IIC T4, Ex nA IIC T4, or Ex nL IIC T4eq u ipment.

    Electromagnetic Compatibility GuidelinesThis prod u ct was tested and complies with the reg u latory

    req u irements and limits for electromagnetic compatibility (EMC)as stated in the prod u ct specifications. These req u irements andlimits are designed to provide reasonable protection againstharmf u l interference when the prod u ct is operated in its intendedoperational electromagnetic environment.

    This prod u ct is intended for u se in ind u strial locations. There is nogu arantee that harmf u l interference will not occ u r in a partic u larinstallation, when the prod u ct is connected to a test object, or if theprod u ct is u sed in residential areas. To minimize the potential forthe prod u ct to ca u se interference to radio and television reception

  • 8/13/2019 National Instruments 9214 manual

    7/44

    National Instruments Corp. 7 NI 9214 Operating Instructions and Specifications

    or to experience u nacceptable performance degradation, install andu se this prod u ct in strict accordance with the instr u ctions in theprod u ct doc u mentation.

    Fu rthermore, any changes or modifications to the prod u ct notexpressly approved by National Instr u ments co u ld void yo u rau thority to operate it u nder yo u r local reg u latory r u les.

    Caution When operating this prod u ct, u se shieldedcables and accessories.

    Special Guidelines for Marine ApplicationsSome prod u cts are Lloyds Register (LR) Type Approved formarine (shipboard) applications. To verify Lloyds Registercertification for a prod u ct, visit ni.com/certification andsearch for the LR certificate, or look for the Lloyds Register markon the prod u ct label.

    Caution In order to meet the EMC req u irements for

    marine applications, install the produ

    ct in a shieldedenclos u re with shielded and/or filtered power andinp u t/o u tp u t ports. In addition, take preca u tions whendesigning, selecting, and installing meas u rement probesand cables to ens u re that the desired EMC performance is

    attained.

  • 8/13/2019 National Instruments 9214 manual

    8/44

    NI 9214 Operating Instructions and Specifications 8 ni.com

    Connecting the NI 9214The NI TB-9214 provides connections for 16 thermoco u ple

    channels on the NI 9214.

    Figure 2. NI TB-9214 Terminal Reference

    TC0+

    TC0

    TC1+

    TC1

    TC5+

    TC5

    TC6+

    TC6

    TC2+

    TC4

    TC4+

    TC3+

    TC3

    TC2

    TC7+

    TC9

    TC9+

    TC8+

    TC8

    TC7

    TC10+

    TC12

    TC12+

    TC11+

    TC11

    TC10TC13+

    TC15

    TC15+

    TC14+

    TC14

    TC13

    COM

    COM

  • 8/13/2019 National Instruments 9214 manual

    9/44

    National Instruments Corp. 9 NI 9214 Operating Instructions and Specifications

    Yo u can connect thermoco u ple inp u t signals to the NI 9214.Connect the positive lead of the thermoco u ple to the TC+ terminaland the negative lead of the thermoco u ple to the TC terminal. If

    yo u are u ns u re which of the thermoco u ple leads is positive andwhich is negative, check the thermoco u ple doc u mentation or thethermoco u ple wire spool.

    The NI 9214 also has two common terminals, COM, that areinternally connected to the isolated gro u nd reference of themod u le. A connection to COM is not necessary for mostapplications. Specifically, leave COM u nconnected forconfig u rations in which all thermoco u ples are floating, allthermoco u ples are referenced to the same common-mode voltage,or one thermoco u ple is referenced to a common-mode voltage andall others are floating.

    However, for config u rations in which two or more thermoco u plesare referenced to different common-mode voltages, connectingCOM according to the following g u idelines improves

    common-mode rejection performance. Connect COM to chassisgro u nd if the common-mode to chassis voltages are smaller than1.2 V. Otherwise, connect COM to a valid common-mode voltagereference that is within 1.2 V of the common-mode voltages of allthe thermoco u ple inp u ts.

  • 8/13/2019 National Instruments 9214 manual

    10/44

    NI 9214 Operating Instructions and Specifications 10 ni.com

    Connecting the SignalsTo connect thermoco u ple wires to the NI TB-9214, complete thefollowing steps while referring to Fig u re 3.

    1. Loosen the captive screws and remove the top cover.

    2. Remove the foam pad from the NI TB-9214 opening.

    3. Prepare the screw-terminal wire by stripping the o u terins u lation and inner ins u lation. Refer to the Specifications section for more information on screw-terminal wiring.

    4. Insert the stripped end of the wire f u lly into the appropriateterminal. Make s u re no exposed wire extends past the screwterminal.

    5. Sec u re the connection by tightening the screw for the terminal.Refer to the Specifications section for information onscrew-terminal torq u e.

    6. Ro u te the wire thro u gh the NI TB-9214 opening.

    7. Remove slack in the NI TB-9214 wiring and sec u re the wiresby threading a zip tie thro u gh the two strain relief holes on theNI TB-9214 opening.

  • 8/13/2019 National Instruments 9214 manual

    11/44

    National Instruments Corp. 11 NI 9214 Operating Instructions and Specifications

    8. Replace the foam pad in the NI TB-9214 opening to fill anyexcess space in the opening.

    9. Reinstall the top cover on the NI TB-9214 and tighten the fo u rcaptive screws to sec u re the cover and compress the foam pad.

    Figure 3. Connecting Wires to the NI TB-9214

    1 Top Cover2 Foam Pad3 Terminals

    4 Strain Relief Holes5 Ground Lug

    1

    2 4

    5

    3

  • 8/13/2019 National Instruments 9214 manual

    12/44

    NI 9214 Operating Instructions and Specifications 12 ni.com

    Yo u can connect a shielded thermoco u ple to the NI 9214. Connectthe shield to the gro u nd l u g on the NI TB-9214 and COM to acommon-mode voltage reference on the thermoco u ple. A valid

    common-mode voltage reference is a voltage that is within 1.2 Vof the common-mode voltage of the thermoco u ple. If COM is leftfloating, the internal common-mode voltage of the inp u t circ u itryis the average of all the inp u ts. Refer to the Specifications sectionfor more information abo u t the common-mode voltage. The shield

    gro u nding methodology can vary depending on the application.Refer to Fig u re 4 for an ill u stration of a typical shieldingconfig u ration.

    Figure 4. Connecting a Shielded Thermocouple Input Signal to the NI 9214

    Thermoco u pleTC

    COM

    TC+

    S hield N

    I T B - 9

    2 1 4

    NI 9214

    Ch ass is Gro u nd

    Gro u ndLu g

  • 8/13/2019 National Instruments 9214 manual

    13/44

    National Instruments Corp. 13 NI 9214 Operating Instructions and Specifications

    Installing the NI TB-9214Connect the NI TB-9214 to the NI 9214 front connector as shown

    in Fig u re 5. Tighten the two jackscrews on the NI TB-9214 to holdit sec u rely in place. Do not overtighten the jackscrews.

    Figure 5. Installing the NI TB-9214

    NI 9214 CircuitryThe NI 9214 passes 16 thermoco u ple inp u t channels and oneau tozero channel thro u gh a differential filter. Each channel is thenmu ltiplexed and sampled by a 24-bit analog-to-digital converter(ADC). Refer to Fig u re 6 for an ill u stration of the inp u t circ u itry

    for one channel of the NI 9214.

  • 8/13/2019 National Instruments 9214 manual

    14/44

    NI 9214 Operating Instructions and Specifications 14 ni.com

    Figure 6. Input Circuitry for One Channel of the NI 9214

    Each thermoco u ple channel has a programmable openthermoco u ple detection (OTD) circ u it, which consists of a c u rrentso u rce between the TC+ and TC terminals. If an open

    thermoco u ple is connected to the channel, the c u rrent so u rce forcesa f u ll-scale voltage across the terminals. Yo u can config u re theOTD circ u it in software. Refer to the software help for informationabo u t config u ring the OTD circ u it. Visit ni.com/info and enterc s erie s doc for information abo u t C Series doc u mentation.

    TC+

    TC

    COM

    NI 9214

    Inp u tImped a nce

    10 M

    10 M

    OpenThermoco u ple

    DetectionCu rrent

    FilteredDifferenti a lAmplifier

    a ndMu ltiplexer

    Is ola tedADC

  • 8/13/2019 National Instruments 9214 manual

    15/44

    National Instruments Corp. 15 NI 9214 Operating Instructions and Specifications

    Each thermoco u ple channel also has a resistor between theTC+ and COM terminals and between the TC and COMterminals. Beca u se long thermoco u ple wires can act like resistors,

    the bias c u rrent from the OTD circ u it can ca u se small offset errors.The gain and offset errors res u lting from the so u rce impedance ofconnected thermoco u ples are negligible for many applications. Thermoco u ples with a higher lead resistance can introd u ce moresignificant errors. Refer to the Compensating for OTD Current

    section for information abo u t compensating for offset errorsca u sed by bias c u rrent and lead resistance. Refer to theSpecifications section for more information abo u t errors res u ltingfrom so u rce impedance.

    The channels share a common gro u nd, COM, that is isolated fromother mod u les in the system. The NI 9214 common-mode range isthe maxim u m voltage allowed between any channel and COM. TheNI 9214 meas u res the common-mode voltage level of each channeland ret u rns a warning in the software if the signal is o u tside thecommon-mode voltage range. Refer to the Specifications sectionfor more information abo u t the common-mode voltage range.

    The NI 9214 s u pports high-resol u tion and high-speed timingmodes. High-resol u tion timing mode optimizes acc u racy and noiseand rejects power line freq u encies. High-speed timing mode

  • 8/13/2019 National Instruments 9214 manual

    16/44

    NI 9214 Operating Instructions and Specifications 16 ni.com

    optimizes sample rate and signal bandwidth. Refer to theSpecifications section for more information abo u t the high-speedand high-resol u tion timing modes. Refer to the software help for

    information abo u t setting the timing mode, or conversion time,in software. Visit ni.com/info and enter c s erie s doc forinformation abo u t C Series doc u mentation.

    Temperature Measurement AccuracyConsiderationsTemperat u re meas u rement errors depend partly on thethermoco u ple type, the acc u racy of the thermoco u ple, thetemperat u re being meas u red, the resistance of the thermoco u ple

    wires, and the cold-j u nction temperat u re. Refer to the Temperature Measurement Accuracy section in the Specifications section for themod u le acc u racy for each thermoco u ple type when connected tothe NI 9214.

    For the best acc u racy res u lts, keep temperat u re gradients acrossNI 9214 terminals to a minim u m, n u ll the lead wire resistance, andenable the a u tozero channel. Refer to the Minimizing ThermalGradients section, the Compensating for OTD Current section,and the Using the Autozero Channel section for more information.

  • 8/13/2019 National Instruments 9214 manual

    17/44

    National Instruments Corp. 17 NI 9214 Operating Instructions and Specifications

    Cold-Junction Temperature Measurement AccuracyThe NI TB-9214 has three cold-j u nction compensation channels.Heat dissipated by adjacent mod u les or other nearby heat so u rcescan ca u se errors in thermoco u ple meas u rements by heating u p theNI 9214 terminals to a different temperat u re than the cold-j u nctioncompensation sensor. The thermal gradient across the terminalscan ca u se the terminals of different channels to be at differenttemperat u res, in which case the res u lting meas u rement createserrors not only in absol u te acc u racy b u t also in the relativeacc u racy between channels. Refer to the Specifications section forthe cold-j u nction compensation acc u racy specifications. Thethermoco u ple acc u racy specifications incl u de the errors ca u sedby the thermal gradient across the NI 9214 terminals forconfig u rations with the NI 9214 terminals facing forward oru pward. Refer to the Temperature Measurement Accuracy sectionin the Specifications section for the thermoco u ple acc u racyspecifications.

    Minimizing Thermal GradientsThermal gradients can be ca u sed by changes in the ambient airtemperat u re near the front connector or by the thermoco u ple wireif it cond u cts heat or cold directly to the terminal j u nctions. For the

  • 8/13/2019 National Instruments 9214 manual

    18/44

    NI 9214 Operating Instructions and Specifications 18 ni.com

    best acc u racy res u lts, follow these g u idelines for minimizingthermal gradients: Use small-ga u ge thermoco u ple wire. Smaller wire transfers

    less heat to or from the terminal j u nction. If yo u are u sing a low channel co u nt on the mod u le, insert the

    foam pad in the NI TB-9214 opening. Refer to the Connectingthe Signals section for more information.

    R u n thermoco u ple wiring together near the NI TB-9214 tokeep the wires at the same temperat u re.

    Avoid r u nning thermoco u ple wires near hot or cold objects. Minimize adjacent heat so u rces and air flow across the terminals. If possible, u se the foam pad in the NI TB-9214 opening to

    restrict airflow aro u nd the terminals. Keep the ambient temperat u re as stable as possible. Make s u re the mod u le terminals are facing forward or u pward.

    Keep the mod u le in a stable and consistent orientation. Allow the thermal gradients to settle after a change in system

    power or in ambient temperatu

    re. A change in system powercan happen when the system powers on, the system comes o u tof sleep mode, or yo u insert/remove mod u les. Refer to thewarm- u p time in the Specifications section for moreinformation.

  • 8/13/2019 National Instruments 9214 manual

    19/44

    National Instruments Corp. 19 NI 9214 Operating Instructions and Specifications

    Compensating for OTD CurrentThe OTD c u rrent ca u ses a voltage error on the inp u t when p u shingagainst a so u rce resistance. This error is minimal for manyapplications, b u t can be significant for some applications thatreq u ire high acc u racy and u se long, narrow ga u ge thermoco u plewires that ca u se large so u rce resistances. Refer to theSpecifications section for more information abo u t the OTD biascu rrent.

    Yo u can also eliminate the OTD bias c u rrent by disabling the OTD.Refer to the software help for information abo u t config u ring theOTD circ u it. Visit ni.com/info and enter c s erie s doc forinformation abo u t C Series doc u mentation.

    Yo u can compensate for the OTD bias c u rrent if yo u r applicationreq u ires high acc u racy, high resistance thermoco u ple wires, andthe ability to detect open thermoco u ples. Take a meas u rement withthe OTD enabled and a second meas u rement with the OTDdisabled. The voltage difference between the meas u rements isthe error voltage d u e to the OTD bias c u rrent. S u btract the errorvoltage from s u bseq u ent meas u rements.

    Visit ni.com/info and enter 9214OTD for more informationabo u t compensating for the OTD errors.

  • 8/13/2019 National Instruments 9214 manual

    20/44

    NI 9214 Operating Instructions and Specifications 20 ni.com

    Using the Autozero ChannelThe NI 9214 has an internal a u tozero channel, which can besu btracted from each thermoco u ple reading to compensate foroffset errors. Use of the a u tozero channel is optional, however theNI 9214 specifications ass u me that a u tozero is applied to everysample. Refer to the software help for information abo u t u sing theau tozero channel. Visit ni.com/info and enter c s erie s doc forinformation abo u t C Series doc u mentation.

    Sleep ModeThis mod u le s u pports a low-power sleep mode. S u pport for sleepmode at the system level depends on the chassis that the mod u le is

    plu gged into. Refer to the chassis man u al for information abo u tsu pport for sleep mode. If the chassis s u pports sleep mode, referto the software help for information abo u t enabling sleep mode.Visit ni.com/info and enter c s erie s doc for informationabo u t C Series doc u mentation.

    Typically, when a system is in sleep mode, yo u cannotcomm u nicate with the mod u les. In sleep mode, the systemcons u mes minimal power and may dissipate less heat than it doesin normal mode. Refer to the Specifications section for moreinformation abo u t power cons u mption and thermal dissipation.

  • 8/13/2019 National Instruments 9214 manual

    21/44

    National Instruments Corp. 21 NI 9214 Operating Instructions and Specifications

    SpecificationsThe following specifications are typical for the range 40 to 70 Cu nless otherwise noted. The specifications are for the NI 9214 u sedin conj u nction with an NI TB-9214.

    Warm- u p time 1 .................................. 15 min u tes

    Input CharacteristicsNu mber of channels

    NI 9214....................................... 16 thermoco u ple channels,1 internal a u tozero channel

    NI TB-9214 ................................3 internal cold-j u nction

    compensation channelsADC resol u tion................................. 24 bitsType of ADC..................................... Delta-SigmaSampling mode.................................Scanned

    Voltage meas u rement range ..............78.125 mV

    1 The warm- u p time ass u mes the mod u le is not in sleep mode, is facing forwardor u pward, and is in a constant ambient temperat u re. National Instr u mentsrecommends allowing the f u ll warm- u p time.

  • 8/13/2019 National Instruments 9214 manual

    22/44

  • 8/13/2019 National Instruments 9214 manual

    23/44

  • 8/13/2019 National Instruments 9214 manual

    24/44

    NI 9214 Operating Instructions and Specifications 24 ni.com

    Inp u t noiseHigh-resol u tion mode

    RMS..................................... 220 nV rmsCrest factor........................... 6

    High-speed modeRMS..................................... 2.8 VrmsCrest factor........................... 10

    Gain errorHigh-resol u tion mode................. 0.03% typ at 25 C,

    0.15% max at 40 to 70 CHigh-speed mode........................ 0.04% typ at 25 C,

    0.16% max at 40 to 70 COffset error

    High-resol u tion mode.................2 V typ, 8 V maxHigh-speed mode........................15 V typ, 23 V max

    Offset error from sou

    rceimpedance with OTD enabled .......... Add 0.2 V per

  • 8/13/2019 National Instruments 9214 manual

    25/44

    National Instruments Corp. 25 NI 9214 Operating Instructions and Specifications

    Inp u t c u rrentOTD enabled ..............................200 nAOTD disabled .............................400 pA

    OTD bias c u rrent drift....................... 200 pA/C maxCold-j u nction compensation acc u racy

    23 5 C ....................................0.25 C typ20 to 70 C ...............................0.6 C max40 to 70 C ...............................0.9 C max

    MTBF ...............................................Contact NI for BellcoreMTBF or MIL-HDBK-217Fspecifications.

  • 8/13/2019 National Instruments 9214 manual

    26/44

    NI 9214 Operating Instructions and Specifications 26 ni.com

    Temperature Measurement AccuracyMeas u rement sensitivity 1

    High-resol u tion modeTypes J, K, T, E, N ............... 0.01 CTypes R, S ............................ 0.03 CType B.................................. 0.04 C

    High-speed modeTypes J, K, T, E .................... 0.10 CType N.................................. 0.11 CTypes R, S ............................ 0.36 CType B.................................. 0.48 C

    1 Meas u rement sensitivity represents the smallest change in temperat u re that a sensorcan detect. It is a f u nction of noise. The val u es ass u me the median of the f u llmeas u rement range of the standard thermoco u ple sensor according to NISTMonograph 175.

  • 8/13/2019 National Instruments 9214 manual

    27/44

    National Instruments Corp. 27 NI 9214 Operating Instructions and Specifications

    The following thermoco u ple meas u rement tables and graphs showthe mod u le acc u racy for each thermoco u ple type u nder thefollowing conditions:

    A u tozero is enabled. Open thermoco u ple detection is disabled.

    0 V common-mode voltage.

    The tables incl u de all meas u rement errors of the mod u le andterminal block incl u ding RMS noise. The tables do not incl u de theacc u racy of the thermoco u ple itself.

  • 8/13/2019 National Instruments 9214 manual

    28/44

    NI 9214 Operating Instructions and Specifications 28 ni.com

    Table 1. Thermocouple Type J/N Measurement Accuracy (C)

    1 0 0 C

    0

    C

    1 0 0 C

    3 0 0 C

    5 0 0 C

    7 0 0 C

    9 0 0 C

    1 1 0 0 C

    H i g h

    - R e s o l u t i o n

    Typical,23 5 C

    0.53 0.40 0.37 0.39 0.44 0.45 0.50 0.59

    Max,20 to 70 C

    1.70 1.24 1.00 1.16 1.44 1.58 1.89 2.33

    Max,40 to 70 C

    1.70 1.26 1.24 1.41 1.69 1.80 2.10 2.57

    H i g h

    - S p e e d

    Typical,23 5 C 1.49 1.17 1.05 0.96 0.97 1.03 1.12 1.24

    Max,20 to 70 C

    2.79 2.12 1.76 1.78 1.96 2.24 2.59 2.99

    Max,40 to 70 C

    2.79 2.12 2.00 1.98 2.17 2.42 2.77 3.18

  • 8/13/2019 National Instruments 9214 manual

    29/44

    National Instruments Corp. 29 NI 9214 Operating Instructions and Specifications

    Table 2. Thermocouple Type K Measurement Accuracy (C)

    1 0 0 C

    0

    C

    1 0 0 C

    3 0 0 C

    7 0 0 C

    9 0 0 C

    1 1 0 0 C

    1 4 0 0 C

    H i g h

    - R e s o l u t i o n

    Typical,23 5 C

    0.50 0.36 0.37 0.42 0.52 0.60 0.69 0.85

    Max,20 to 70 C

    1.56 1.06 0.95 1.23 1.82 2.21 2.64 3.40

    Max,40 to 70 C

    1.56 1.10 1.20 1.49 2.08 2.48 2.93 3.71

    H i g h

    - S p e e d

    Typical,23 5 C

    1.17 0.86 0.87 0.95 1.11 1.25 1.41 1.70

    Max,20 to 70 C

    2.33 1.64 1.50 1.81 2.46 2.91 3.42 4.32

    Max,40 to 70 C

    2.33 1.66 1.76 2.08 2.72 3.19 3.71 4.64

    bl h l ( )

  • 8/13/2019 National Instruments 9214 manual

    30/44

    NI 9214 Operating Instructions and Specifications 30 ni.com

    Table 3. Thermocouple Type T/E Measurement Accuracy (C)

    1 0 0 C

    0

    C

    1 0 0 C

    3 0 0 C

    5 0 0 C

    7 0 0 C

    9 0 0 C

    H i g h

    - R e s o l u t i o n

    Typical,23 5 C

    0.54 0.37 0.33 0.33 0.37 0.43 0.49

    Max,20 to 70 C

    1.76 1.17 0.89 1.00 1.25 1.58 1.94

    Max,40 to 70 C

    1.76 1.17 1.04 1.17 1.42 1.74 2.11

    H i g h

    - S p e e d

    Typical,23 5 C

    1.25 0.88 0.77 0.69 0.69 0.78 0.90

    Max,20 to 70 C

    2.59 1.77 1.38 1.41 1.60 1.96 2.37

    Max,40 to 70 C

    2.59 1.77 1.53 1.53 1.77 2.13 2.55

    T bl 4 Th l T R/S M A (C)

  • 8/13/2019 National Instruments 9214 manual

    31/44

    National Instruments Corp. 31 NI 9214 Operating Instructions and Specifications

    Table 4. Thermocouple Type R/S Measurement Accuracy (C)

    0

    C

    1 0 0 C

    3 0 0 C

    5 0 0 C

    7 0 0 C

    9 0 0 C

    1 1 0 0 C

    1 4 0 0 C

    H i g h

    - R e s o l u t i o n

    Typical,23 5 C

    0.81 0.61 0.54 0.55 0.57 0.59 0.60 0.67

    Max,20 to 70 C

    2.80 1.94 1.84 1.98 2.15 2.31 2.48 2.86

    Max,40 to 70 C

    2.80 1.94 1.84 1.98 2.15 2.31 2.48 2.86

    H i g h

    - S p e e d

    Typical,23 5 C

    4.50 3.30 2.74 2.61 2.54 2.47 2.42 2.49

    Max,20 to 70 C

    6.85 4.91 4.26 4.25 4.32 4.38 4.47 4.85

    Max,40 to 70 C

    6.85 4.91 4.27 4.25 4.32 4.38 4.47 4.85

    T bl 5 Th l T B M t A (C)

  • 8/13/2019 National Instruments 9214 manual

    32/44

    NI 9214 Operating Instructions and Specifications 32 ni.com

    Table 5. Thermocouple Type B Measurement Accuracy (C)

    0

    C

    1 0 0 C

    3 0 0 C

    5 0 0 C

    7 0 0 C

    9 0 0 C

    1 1 0 0 C

    1 4 0 0 C

    H i g h

    - R e s o l u t i o n

    Typical,23 5 C

    0.94 0.61 0.51 0.46 0.43 0.45

    Max,20 to 70 C

    3.40 2.30 1.97 1.86 1.89 2.04

    Max,40 to 70 C

    3.45 2.34 2.00 1.88 1.89 2.05

    H i g h

    - S p e e d

    Typical,23 5 C

    7.36 4.56 3.46 2.88 2.55 2.33

    Max,20 to 70 C

    10.40 6.63 5.21 4.52 4.19 4.10

    Max,40 to 70 C

    10.45 6.66 5.23 4.54 4.21 4.11

  • 8/13/2019 National Instruments 9214 manual

    33/44

  • 8/13/2019 National Instruments 9214 manual

    34/44

    NI 9214 Operating Instructions and Specifications 34 ni.com

    Figure 8. Thermocouple Error Max (High-Res), 20 to 70 C

    M e

    a s u r e m e n

    t E r r o r (

    C )

    Me asu red Temper a tu re ( C )

    0

    2

    1

    4

    3

    200 50 300 550 800 1050 1300 1550 1800

    Type K

    Type T/E Type R/ SType J/NType B

    Power Requirements

  • 8/13/2019 National Instruments 9214 manual

    35/44

    National Instruments Corp. 35 NI 9214 Operating Instructions and Specifications

    Power RequirementsPower cons u mption from chassis

    Active mode ...............................300 mW maxSleep mode .................................30 W max

    Thermal dissipation (at 70 C)Active mode ...............................630 mW maxSleep mode ................................. 450 mW max

    Physical CharacteristicsIf yo u need to clean the mod u le, wipe it with a dry towel.

    Note For two-dimensional drawings and

    three-dimensional models of the C Series mod u le andconnectors, visit ni.com/dimen s ion s and search bymod u le n u mber.

    NI TB-9214 screw-terminalwiring................................................ 20 to 30 AWG thermoco u ple

    wire with 51 mm (2 in.) ofou ter ins u lation and 5.1 mm(0.2 in.) of inner ins u lationstripped from the end

  • 8/13/2019 National Instruments 9214 manual

    36/44

  • 8/13/2019 National Instruments 9214 manual

    37/44

    Safety Standards

  • 8/13/2019 National Instruments 9214 manual

    38/44

    NI 9214 Operating Instructions and Specifications 38 ni.com

    yThis prod u ct meets the req u irements of the following standards ofsafety for electrical eq u ipment for meas u rement, control, and

    laboratory u se: IEC 61010-1, EN 61010-1

    UL 61010-1, CSA 61010-1

    NoteFor UL and other safety certifications, refer to theprod u ct label or the Online Product Certification section.

    Electromagnetic CompatibilityThis prod u ct meets the req u irements of the following EMC

    standards for electrical equ

    ipment for measu

    rement, control,and laboratory u se:

    EN 61326-1 (IEC 61326-1): Class A emissions; Ind u strialimm u nity

    EN 55011 (CISPR 11): Gro u p 1, Class A emissions

    AS/NZS CISPR 11: Gro u p 1, Class A emissions

    FCC 47 CFR Part 15B: Class A emissions

    ICES-001: Class A emissions

    Note For EMC declarations and certifications, refer to

  • 8/13/2019 National Instruments 9214 manual

    39/44

    National Instruments Corp. 39 NI 9214 Operating Instructions and Specifications

    the Online Product Certification section.

    CE ComplianceThis prod u ct meets the essential req u irements of applicableEu ropean Directives as follows:

    2006/95/EC; Low-Voltage Directive (safety)

    2004/108/EC; Electromagnetic Compatibility Directive(EMC)

    Online Product Certification

    To obtain prod u ct certifications and the Declaration of Conformity(DoC) for this prod u ct, visit ni.com/certification , search bymod u le n u mber or prod u ct line, and click the appropriate link inthe Certification col u mn.

    Shock and Vibration

  • 8/13/2019 National Instruments 9214 manual

    40/44

    NI 9214 Operating Instructions and Specifications 40 ni.com

    To meet these specifications, yo u m u st panel mo u nt the system andu se the NI TB-9214 terminal block to protect the connections.

    Operating vibrationRandom (IEC 60068-2-64)......... 5 g rms , 10 to 500 HzSin u soidal (IEC 60068-2-6) .......5 g, 10 to 500 Hz

    Operating shock

    (IEC 60068-2-27).............................. 30 g, 11 ms half sine,50 g, 3 ms half sine,18 shocks at 6 orientations

    EnvironmentalNational Instr u ments C Series mod u les are intended for indoor u seonly b u t may be u sed o u tdoors if installed in a s u itable enclos u re.Refer to the man u al for the chassis yo u are u sing for moreinformation abo u t meeting these specifications.Operating temperat u re(IEC 60068-2-1, IEC 60068-2-2) ..... 40 to 70 CStorage temperat u re(IEC 60068-2-1, IEC 60068-2-2) .....40 to 85 C

    Ingress protection

  • 8/13/2019 National Instruments 9214 manual

    41/44

    National Instruments Corp. 41 NI 9214 Operating Instructions and Specifications

    NI 9214....................................... IP 40Operating h u midity

    (IEC 60068-2-56).............................. 10 to 90% RH,noncondensing

    Storage h u midity(IEC 60068-2-56).............................. 5 to 95% RH,

    noncondensingMaxim u m altit u de............................. 2,000 mPoll u tion Degree ............................... 2

    Environmental Management

    NI is committed to designing and man u fact u ring prod u cts inan environmentally responsible manner. NI recognizes thateliminating certain hazardo u s s u bstances from o u r prod u cts isbeneficial to the environment and to NI c u stomers.

    For additional environmental information, refer to the NI and the Environment Web page at ni.com/environment . This pagecontains the environmental reg u lations and directives with whichNI complies, as well as other environmental information notincl u ded in this doc u ment.

    Waste Electrical and Electronic Equipment (WEEE)

  • 8/13/2019 National Instruments 9214 manual

    42/44

    NI 9214 Operating Instructions and Specifications 42 ni.com

    EU Customers At the end of the prod u ct life cycle,all prod u cts must be sent to a WEEE recycling center.

    For more information abo u t WEEE recycling centers,National Instr u ments WEEE initiatives, and compliancewith WEEE Directive 2002/96/EC on Waste andElectronic Eq u ipment, visit ni.com/environment/ weee .

    CalibrationYo u can obtain the calibration certificate and information abo u tcalibration services for the NI 9214 at ni.com/calibration .Calibration interval ........................... 1 year

    RoHS

    National Instruments (RoHS)

    National Instruments RoHSni.com/environmen t /rohs_china (For informationabout China RoHS compliance, go to ni.com/environmen t /rohs_china .)

  • 8/13/2019 National Instruments 9214 manual

    43/44

    Korea 82 02 3451 3400, Lebanon 961 (0) 1 33 28 28,M l i 1800 887710 M i 01 800 010 0793

  • 8/13/2019 National Instruments 9214 manual

    44/44

    2010 National Instruments Corp. All rights reserved.

    375138B-01 Dec10

    LabVIEW, National Instruments, NI, ni.com, the National Instruments corporate logo, and the Eagle logo aretrademarks of National Instruments Corporation. Refer to the Trademark Information at ni.com/trademark s for other National Instruments trademarks. Other product and company names mentioned herein are trademarks ortrade names of their respective companies. For patents covering National Instruments products/technology, refer tothe appropriate location: HelpPatents in your software, the patent s .txt file on your media, or the NationalInstruments Patent Notice at ni.com/patent s .

    Malaysia 1800 887710, Mexico 01 800 010 0793,Netherlands 31 (0) 348 433 466, New Zealand 0800 553 322,

    Norway 47 (0) 66 90 76 60, Poland 48 22 328 90 10,Port u gal 351 210 311 210, R u ssia 7 495 783 6851,Singapore 1800 226 5886, Slovenia 386 3 425 42 00,So u th Africa 27 0 11 805 8197, Spain 34 91 640 0085,Sweden 46 (0) 8 587 895 00, Switzerland 41 56 2005151,Taiwan 886 02 2377 2222, Thailand 662 278 6777,Tu rkey 90 212 279 3031, United Kingdom 44 (0) 1635 523545