NASA NHB 5300.2 Apollo Metrology Requirements Manual
Transcript of NASA NHB 5300.2 Apollo Metrology Requirements Manual
-
7/25/2019 NASA NHB 5300.2 Apollo Metrology Requirements Manual
1/34
NHB 5300.2
APOLLO PROGRAM
DECEMBER1965
APOLLO
3
_09 W dOd A, ,L II I_ *4 d
METROLOGY
REQUIREMENTS
MANUAL
NATIONAL AERONAUTICS AND
Washington, D.C.
SPACE ADMINISTRATION
20546
-
7/25/2019 NASA NHB 5300.2 Apollo Metrology Requirements Manual
2/34
_w
V
r_
-
7/25/2019 NASA NHB 5300.2 Apollo Metrology Requirements Manual
3/34
NASA Issuance No. NHB 5300.2
Apollo Documentation No. RA-006-008-1
PREFACE
Prevention
of
interface anomalies during Apollo
system hardware integration will, to a large extent,
depend upon the compatibility of measurement pro-
cedures presently being utilized throughout the NASA
and Contractor network.
Implementation of the Apollo Metrology Requirements
Manual will insure that uniform measurements of
high accuracy are achieved during the Apollo Program.
The criteria established within this document are to
be considered a clarification of the requirements
contained in NPC 200-2, Quality Program Provisions
for Space System Contractors.
Major General, USAF
Director, Apollo Program
Distribution:
Special
-
7/25/2019 NASA NHB 5300.2 Apollo Metrology Requirements Manual
4/34
V
T
; j
1 I
i
il
-
7/25/2019 NASA NHB 5300.2 Apollo Metrology Requirements Manual
5/34
k_J
TABLE OF CONTENTS
Subject
Chapter 1 Introduction
i.I
I.Z
1.3
Purpose ...................................... 1- 1
Scope 1- 1
Applicability 1- 1
Z.I
Chapter 2: Definitions
Applicability 2- 1 . .
3.1
3.2
3.3
3.4
3.5
3.6
3.7
3.8
3.9
3.10
3.11
3.12
3.13
3.14
Chapter 3: Standards and Equipment Control Requirements
Major Activities ................................ 3- 1
Operating Procedures ............................ 3- 1
Selection ..................................... 3- 1
Evaluation and Acceptance ......................... 3-2
Traceability ................................... 3-3
Calibration Cycles ............................... 3-4
Calibration Procedures ........................... 3-4
Records ...................................... 3-7
Recall ....................................... 3-8
Corrective Action ............................... 3-9
Facilities ..................................... 3- 10
Personnel .................................... 3- 1 1
Interlaboratory Comparison ........................ 3- 12
Evaluation .................................... 3- 12
4.1
4.2
4.3
4.4
4.5
Chapter 4: Evaluation Procedures
Purpose and Scope ............................... 4- 1
Surveyor's Preparation ........................... 4-1
Applicable Documents ............................ 4- 1
Performance and Evaluation ........................ 4-2
Metrology Requirements Evaluation Checklists ............ 4-2
-
7/25/2019 NASA NHB 5300.2 Apollo Metrology Requirements Manual
6/34
i
-
7/25/2019 NASA NHB 5300.2 Apollo Metrology Requirements Manual
7/34
I.I PURPOSE
CHAPTER 1
INTRODUCTION
1.2
The Apollo Metrology Requirements Manual establishes specific require-
ments and associated criteria for a coordinated Metrology System,
including the establishment of an Evaluation Procedure (see Chapter 4).
This Manual provides for developing and controlling calibration practices
for standards and measuring equipment used in the development, manu-
facture, inspection and test of hardware.
SCOPE
This Manual includes control practices for all inspection and test
standards, gages, tooling, measuring, and test equipment used for
calibrating, data collecting, testing, and inspecting to determine the
conformance of hardware to design specifications.
1.3 APPLICABILITY
This Manual is applicable to all Centers, sites and their prime contrac-
tors who are responsible for the selection, evaluation, approval, cali-
bration, maintenance, and control of standards and measuring equipment.
1-1
-
7/25/2019 NASA NHB 5300.2 Apollo Metrology Requirements Manual
8/34
-
7/25/2019 NASA NHB 5300.2 Apollo Metrology Requirements Manual
9/34
M.J
CHAPTER 2
DEFINITIONS
2.1 APPLICABILITY
To promote mutual understanding and to assist in effective communication
of requirements, the following definitions are applicable to this Manual.
a.
Calibration--Comparison between two instruments or devices (one of
which is a standard of known accuracy) to detect, correlate, report,
and/or adjust any aviation in the accuracy of the instrument being
compared.
b. Calibration Cycle--An
established interval between calibrations of
standards or inspection, measuring and test equipment.
c.
CalibrationProcedures--The
detailed instructions for performing
calibration.
do
Clean Room--An
enclosed area employing control over the particu-
late matter in air with the additional control of temperature, humidity,
and pressure, as required.
eu
Commercial
Items--Catalog
standards and inspection, measuring and
test equipment considered to be off-the-shelf and/or available
comme r cally.
f.
Historical Record--Record of identification, and accumulation of main-
tenance and calibration data for each standard and inspection, measur-
ing and test equipment to provide the data required for determining
stability and performance trends.
go
Inspection, Measuring and Test Equipment--Equipment
(including
tooling) used to inspect, measure or test hardware to determine
conformance to applicable design specifications. Tooling includes
tools, gages, jigs and fixtures which measure dimensions, contours,
or locations affecting quality characteristics.
(Inspection, measuring and test equipment are hereafter referred to
in this Manual as equipment. )
h. Laboratory--Any segregated area specifically equipped for calibration
of standards and inspection, measuring, and test equipment.
io
Metrology--The
science and technology of measurement of any pa-
rameter of weight, mass, length, chemical and electrical unit, and
physical constant. This technology includes the selection, utilization,
and control of equipment used to provide measurement, and also,
includes assurance of a valid relationship between measured values
2-1
-
7/25/2019 NASA NHB 5300.2 Apollo Metrology Requirements Manual
10/34
Z.i Definitions
2-2
a
j.
I.
m.
n.
o.
P.
q_
ro
and values of measurement established by national reference standards
and international agreements.
Mobile Capability--Ability
to transport facilities, standards and
inspection, measuring and test equipment, and the ability to
perform
calibration at remote locations.
NASA (or Contractor) Reference Standard--Measure
m e n t standards
representing the highest level of accuracy available within the NASA
(or Contractor) facilities. These are standard references against
which lower level standards are compared, and are calibrated to
national reference standards or are defined and sanctioned by inter-
national agreement.
National Bureau of Standards (NBS)--A
central Government agency
that has the responsibility for the custody, maintenance, and develop-
ment of the national standards of measurement together with the
provisions of calibration services related to these standards. This
agency has standards laboratories in Washington, D.C., and Boulder,
Colorado. (In addition to NBS, reference standards are maintained
throughout the world, such as at the National Physical Laboratory in
England and the National Research Council in Canada.)
National Refernce Standards
--The highest level national references
which are: standards maintained by the National Bureau of Standards;
acceptable standards maintained by other recognized agencies where
an NBS reference is not available; or values of natural physical
constants, such as, the acceleration due to gravity or standard
reference materials.
Recall System--A documented system to insure that all standards
and inspection, measuring, and test equipment are recalled from the
users at predetermined intervals for the purpose of calibration.
Repeatability--The
ability of a standard or measurement equipment
to repeat a measurement within some known value when used by one
person under specified conditions.
Reproducibility--The
difference in the average of groups of repeated
measurements made by different persons using the same standard or
measurement equipment under specified conditions.
Shielded Room--
A screened or metal enclosed area to prevent out-
side electrical interference or radiation from affecting equipment
performance or calibration measurements, and is provided with a
filtered power supply.
Site--NASA installations and component field activities geographically
located apart from the parent installation (i.e., MSC White Sands
Test Facility, Mississippi Test Facility, Michoud Assembly Facility,
etc.)
-
7/25/2019 NASA NHB 5300.2 Apollo Metrology Requirements Manual
11/34
Definitions 2. I
Se
te
We
De
Special Equipment--Equipment
of special design utilized for specific
measurement(s) and not normally available commercially.
Standard-
-An item which is established as an authoriz'ed or recognized
measurement reference, and is used to calibrate other standards or
to calibrate inspection, measuring and test equipment.
Systern--A documented method for the consistent performance of a
task or associated group of tasks.
Transfer Standavd--Any standard employed to transmit accuracies
from a NASA (or Contractor) Reference Standard toworking standard.
Workir_ Standard--Any
standard used in the calibration of inspection,
measuring, and test equipment.
2-3
-
7/25/2019 NASA NHB 5300.2 Apollo Metrology Requirements Manual
12/34
_J
i
Ii
-
7/25/2019 NASA NHB 5300.2 Apollo Metrology Requirements Manual
13/34
%J
CHAPTER 3
STANDARDS AND EQUIPMENT CONTROL REQUIREMENTS
3.1 MA:_OR ACTIVITIES
The coordinated Apollo Metrology Program will include the major
activities as delineated in pars. 3.3 through 3.14. The activities provide
for the control of inspection, measuring and test equipment (hereafter
referred to as equipment ) and for the control of standards. Criteria
necessary for meeting these activity requirements are outlined and will
be included in operating systems at each Center, site or prime contrac-
tor.
3.2 OPERATING PROCEDURES
Written operating procedures will be developed and maintained for all
systems based on criteria specified in this document. These written
procedures
will be the basis for continuous review, evaluation and con-
sistent implementation regardless of personnel changes. Therefore,
the procedures will include all practices for the complete Metrology
Program as specified in the following paragraphs.
3.3 SELECTION
The selection of standards and equipment willbebased on the following
criteria:
a. The accuracy, stability, and range
of
standards and equipment will
be in conformance with end-use requirements. Within state-of-the-
art limitations, the standards used for calibration of inspection,
measuring and test equipment will have a tolerance no greater than
10 per cent of the allowable tolerance of the equipment being
calibrated, i.e., if the item to be calibrated has a tolerance of fl.0
per cent, the maximum allowable tolerance of the standard should
be -+0.1 per cent. If the state-of-the-art prevents the attainment of
this 10 to 1 ratio, authorization must be secured prior to use.
b. The selection will not result in unnecessary duplication of a capa-
bility that is available and canfeasiblybe utilized within the organiza-
tion, through an associated laboratory (NASA, DOD, etc.) or at another
Center, site or prime contractor.
c. Commercial or off-the-shelf standards or equipment will be
selected in lieu of special items, where feasible.
3-1
-
7/25/2019 NASA NHB 5300.2 Apollo Metrology Requirements Manual
14/34
3.4 Standards and Equipment Control Requirements
d. When several different types of standards or equipment are adequate
to perform a specific function(s), selection will be based on:
(1) Availability.
(2) Histories of like standards or equipment.
(3) Analysis of interlaboratory comparisons.
(4) Analysis of defect reports.
(5) Initial cost.
(8) Maintenance expenditures.
(7) Manufacturer's performance and service.
e. Verification of the laboratory's (or alternate facility's) capability to
perform calibration on the selected standard or equipment.
3.4 EVALUATION AND ACCEPTANCE
Standards and equipment that are received for evaluation and acceptance
will be processed in a consistent manner to include, but not be limited to,
the following actions:
a.
bo
Standards and equipment will be tested and verified for intended use
to the parameters given in the applicable standard and equipment
specification.
t
Evaluations will be performed and documented to determine that the
standard or equipment has the required accuracy for meeting end-
use specifications and that the standard or equipment can be
satisfactorily calibrated. The extent of these evaluations will be
dependent on the complexity and end-use requirements of the item as
follow s :
(i)
Commercial items (with sufficient information relative to accuracy,
stability and repeatability), when used in a manner consistent with
design, need not necessarily be subjected to evaluation tests.
(2)
Special standards or equipment will receive evaluation tests to in-
sure compatibility with the product and supporting or related equip-
ment. In addition, these special items, due to complexity and
specific requirements, will receive evaluation under intended
method of operation.
(3)
Complex and/or special standards or equipment will receive
comprehensive evaluations for assurance that the end item being
measured will be controlled within specified tolerances. These
evaluations will include repeatability and reproducibility testing
where necessary to provide assurance of required accuracies.
3-2
%J
-
7/25/2019 NASA NHB 5300.2 Apollo Metrology Requirements Manual
15/34
Standards and Equipment Control Requirements 3.5.1
c. An applicable defect report will be generated and maintained for
standards and equipment that do not meet specifications.
d. All standards and equipment that are accepted will be:
(i) Entered into an inventory system (see par. 3.8.1).
(2) Calibrated to a specific procedure (see par. 3.7).
(3) Assigned a calibration cycle (see par. 3.6).
(4) Sealed to prevent tampering by unauthorized personnel prior to
next calibration.
(5)
Labeled or marked when possible, and entered into appropriate
records to indicate date of calibration and due date for next cali-
bration.
(6) Entered into recall system (see par. 3.9).
(7)
Tagged for safety instructions, special usage {i.e., Indication
Only, Reference Only, Limited Use) or special handling, special
storage, and environmental requirements, as applicable.
e.
f.
Data generated from the initial verification of standards and equip-
ment will be maintained.
Special standards or equipment specifications will be properly
documented by drawings and/or specifications and maintained current.
go
All associated manufacturer's instructions or procedures, including
records of any modification made to the item, will be maintained and
kept current.
3.5 TRACEABILITY
.
The standards used to calibrate lower level standards or equipment
will have valid relationships to national reference standards. When this
is not feasible due to state-of-the-art measurement, standards will be
acceptable that are developed by ratio-type of self-calibrationtechniques
using standards whose history and drift rates are known to be satis-
factory.
5.1 Evidence. Traceability to national reference standards will be
evidenced by valid reports of calibration for each standard processed.
These reports will indicate the date of calibration. If calibration of
standards is performed by an alternate laboratory, their report
will be maintained and will cross-reference an NBS report date and
number or other national reference standard report to indicate the
traceability and calibration status of standards used. Any standards
that by their nature are not sent to NBS, or an alternate laboratory,
for example, frequency standards, will be regularly monitored by a
3-3
-
7/25/2019 NASA NHB 5300.2 Apollo Metrology Requirements Manual
16/34
3.5.2
Standards and EcLuipment Control Requirements
method that is traceable to national reference standards and evi-
denced by a permanent log ofthe date, time, and observed deviations.
3.5.2
Flow Charts. Appropriate flow charts will be developed to provide
evidence of traceability of all equipment capabilities. These flow
charts will show this traceability through a series of standards with
records to substantiate relationships to national reference standards.
3.6 CALIBRATION CYCLES
To insure the maintenance of required accuracies for a particular standard
or equipment, it is required that recall for calibration be performed at
predetermined intervals. These intervals may be specified in terms of
time or usage. Time intervals will be established in terms of date
(month/day/year) so that the calibration label indicates the specific day
of expiration. The cycles established will be implemented by the recall
system (see par. 3.9).
3.6.1 Initial Cycle. The establishment of the initial calibration cycle for
incoming items will be based on the following, as applicable:
a. Experience with like standards or equipment.
b. Purpose.
c.
Amount
of use.
d. Accuracy requirements.
e. Standard practices.
f. Stability.
3.7
3.6.2
Cycle Adjustment. Calibration cycles will be shortened as required
to insure continued accuracy as evidence by the results of
preceding
calibrations, and may be lengthened only whenthe results of analysis
of data provide definite indications that such actions will not
adversely affect the accuracy.
CALIBRATION PROCEDURES
Written procedues will be used for calibrating standards or equipment.
These written procedures will be readily available to laboratory personnel
performing the calibration. There will be a means of insuring that an up-
to-date procedure is used that includes a method of indicating and con-
trolling necessary revisions.
3.7.1
Procedures Index and Format. The Index and Format for Calibra-
tion Procedures,
provides
for the listing of all available procedures
developed or being developed by each MSF
Center
and outlines a
procedures format. The Index will be provided to all MSF
Centers
for their use in establishing the availability of existing and planned
3-4
-
7/25/2019 NASA NHB 5300.2 Apollo Metrology Requirements Manual
17/34
Standards and Equipment Control Requirements 3.7.3
procedures for calibrating standards or equipment. The use of
this Index by the Centers will:
a. Provide knowledge of availability of Center-proven procedures.
b. Reduce duplication of efforts in producing new procedures.
c. Provide for attaining standardization of new procedures as they
are developed.
3.7.2
d. Improve communication of like functions between Centers.
Procedure Development. Procedures that are developed will be
based on the Format in the Index and Format for Calibration
Procedures and will contain the following information:
a. Safety warnings and cautions.
b. Environmental control details
to
insure measurements of re-
quired accuracy including temperature, humidity, vibration,
cleanliness, etc.
C,
do
ea
List of calibration equipment required and a diagram of complex
setups (consider accuracy ratios as discussed in par. 3.3a).
The actual operations listed in a series of sequential steps with
applicable tolerances.
A record keeping form or data sheet to include identification of
personnel performing calibration, condition of items previous
to calibration, final results, calibration time/cost, and other
information pertinent to a specific item.
f. Instructions for special handling of a standard or equipment
and required environmental controls after calibration.
g. Provisions for sealing standard or equipment to prevent
tampering between calibrations.
h. Provisions for documenting for the user any adjustments neces-
sary due to changes in environments.
i. For special equipment which is composed of several individual
items, the calibration of any one item may change the interrelated
results of the configuration. Therefore, procedures should
provide for calibration of the entire special equipment, if this
situation is encountered.
3.7.3 NASA (or Contractor) Reference Standards
a. The highest level standards that are available at a particular
location will be calibrated by the National Bureau of Standards
3-5
-
7/25/2019 NASA NHB 5300.2 Apollo Metrology Requirements Manual
18/34
3.7.4 Standards and Equipment Control Requirernents
3-6
(NBS) by another approved laboratory either directly or by use
of a special transfer standard if the Reference Standard cannot
be transported or by ratio-type of self-calibration techniques
(see par. 3.5). If the NBS performs this calibration, a report
will be issued for information and file which will show its
traceability. If an alternate laboratory performs the calibration,
it will be insured that this laboratory is performing the calibra-
tion to an approved procedure which includes proper environ-
ments, and the use of standards having the required accuracies;
these standards will have current reports that indicate direct
traceability to a national reference standard.
b.
A calibration procedure may not be needed for certain Reference
Standards, i.e., frequency standards, and they might not be sent
to NBS or an alternate laboratory for calibration. In these cases,
there will be a monitoring method consisting of a permanent log
showing dates, times and observed deviations that are directly
traceable to national reference standards.
3.7.4 Working and Transfer Standards
a.
If the standards at the working and transfer levels require cali-
bration at a different laboratory due to lack of facilities or lack
of adequate standards available locally, it will be insured that the
alternate laboratory performing the calibration is using approved
procedures (including proper environments and standards having
the required accuracies that are traceable to national reference
standards).
b*
When calibration is performed locally, it is essentialthat formal,
approved procedures be used. Since most of these standards
would fall into the category of a commercial item, a common
procedure should be available as provided for in the Index and
Format for Calibration Procedures. When standards are special
or a common procedure is not available, a procedure will be
developed containing the information outlined in par. 3.7.2.
C.
Standards that are to be used only over a portion of the range or
at certain points in the range may require calibration at these
points or in the ranges used. If this is performed, the calibration
points or ranges will be clearly indicated on the standard.
3.7.5 Equipment Calibration
a.
The calibration of equipment will be performed with working
standards in the required environment and at the intervals
established to provide the required confidence level of the
equipment. Each distinct type of equipment requires that a
calibration procedure be available for consistent operating
practices, and that the standards that are used are supported by
a train of successive reports that trace back to a national
reference standard.
v
%J
i
-
7/25/2019 NASA NHB 5300.2 Apollo Metrology Requirements Manual
19/34
Standards and Equipment Control Requirements 3.8.2
b.
Equipment that is to be used only over a portion of the range
or at certain points in the range may require calibration at these
points or in the ranges used. If this is performed, the calibration
points or ranges will be clearly indicated on the equipment.
3.8 RECORDS
Each standard and equipment will be supported by individual records
detailing all inventory and historical data. These records are necessary
to implement each of the control systems that are developed.
3.8.1
Inventory Records and Control. Inventory records of standards and
equipment are required to provide availability, status and capa-
bility data essential to ensure proper inventory control. Inventory
control will include:
a.
A positive nonduplicating inventory identification code to be
assigned to all standards and equipmert and to be affixed to
the item wherever it does not interfere with usage.
b. A record of each standard and equipment by nomenclature,
model number, serial number, manufacturer, date acquired,
physical location (user), range, accuracy and identification code.
c. Provisions for removing standards or equipment from inventory
when the items become obsolete or beyond economical repair.
d.
All personally owned equipment which will be recorded with
owner's name or payroll number, complete description of
'equipment and physical location. {Personally owned equipment
may be used only when properly authorized.}
3.8.2
Historical.
These records will provide objective evidence that
calibration schedules are complied with and that the accuracy of
the standard or equipment is being maintained. The records also
provide a history of stability, which may be evaluated and utilized
as a basis for the adjustment of calibration cycles or utilized to
indicate trends of wear and deterioration and may become a basis
for future selection of standards and equipment. Historical
records will include, but not be limited to, the following:
a. Data generated during calibration as required by the applicable
procedures (see par. 3.7.2e).
b. Provisions for providing data that will accompany standards or
equipment when returned to the user.
Co
do
e.
Data generated during the initial verification and evaluation and
data from any subsequent evaluations of a standard or equipment.
Calibration cycle.
Range and accuracy.
3-7
-
7/25/2019 NASA NHB 5300.2 Apollo Metrology Requirements Manual
20/34
3.9 Standards and Equipment Control Requirements
f. Special usage information.
g, Calibration time/cost.
h. Maintenance data (repairs required and list of parts replaced).
i. Hours of usage for critical items where the usage factor is
important.
j. The individual record of a standard or equipment (whose
accuracy is reported via a formal report) will cross-reference
the report number. This is especially important for NBS reports
that are received with specific standards.
3.9 RECALL
3.9.1
Documented System. A documented recall system will be established
to insure thatal__ standards and al__jl equipment (incIuding personally
owned) are recalled for calibration at established intervals (cali-
bration cycles).
3.9.2
Maintenance of Records.
The required records for standards and
equipment will be maintained and keyed to a specific calibration
cycle and a specific calibration procedure. These records will
include identification, range, accuracy requirements, and special
usage infromation. The following subparagraphs state the basic
criteria that must be included in a recall system.
3.9.3
Notification.
Notification will be given to each user prior to the
established calibration due date for the assurance that standards and
equipment are:
a. Removed or made ready for calibration when required.
b. Not used after calibration due date.
The established calibration due dates do not relieve the user of the
responsibility of requesting supplementary calibration when a
standard or equipment is damaged, misused or suspected of being
out- of- tole r ance.
3.9.4
Mandatory
Removal.
A mandatory removal system will provide
delegation of authority for insuring the removal from use of stand-
ards or equipment that are beyond the calibration due date. These
items should be removed from service immediately. If physical
removal is impractical due to weight, service connections or other
characteristics of the item, it will be impounded by attachment of
signs, tags, or other means. However, consideration will be given
by the authority to the feasibility of temporarily extending the cali-
bration beyond the due date without degrading accuracies, when the
removal of the items would require teardown of costly setups or
interruption of tests. A tagging or other identification system will
3-8
-
7/25/2019 NASA NHB 5300.2 Apollo Metrology Requirements Manual
21/34
Standards and Equipment Control Requirements 3.10
3.9.5
3.9.6
3.9.7
3.10
be used for identifying items that have not been maintained, or
calibrated in accordance with an established schedule.
Delinquent Notices. Notification of delinquency will be given when
standards or equipment are not calibrated by the responsible
laboratory as scheduled. Responsibility will be established for
follow-up and positive corrective action involving delinquent items.
Labeling. Each standard or equipment will be labeled or marked
to indicate the date when next calibration is due. This label {or
appropriate record) will include date of calibration and identification
of the person performing the calibration. When item identification
does not appear on the label, a method will be established to insure
that labels
cannot
be interchanged from one item to another. If
the calibration cycle is based on usage or another system not as-
sociated with a time interval, the item and/or records will readily
indicate calibration status. To preclude the misuse of a standard or
equipment that is an integral part of anassembled item, the label-
ing or marking will identify the next higher assembly in which that
standard or equipment must be used.
Pickup and Delivery. A pickup and delivery service for all standards
and equipment will be established. This service is the responsi-
bility of the applicable laboratory, which will provide for handling
and transportation dependent upon the complexity and delicacy of
the standards and equipment. A receipt method will be used to
insure that the individual standard or equipment has been received
by the laboratory and as subsequent evidence that the item has been
returned to the user.
CORRECTIVE ACTION
When during normal use, calibration, or as a result of periodic inspec-
tions or crosschecks (see par. 3.14.2), it is determined that a standard
or equipment is out-of-specification or that there is indication of
malfunction, immediate corrective actions will be taken as follows:
a. Standards or equipment will be properly tagged to prevent Usage.
b. Standards or equipment in use will be returned to the calibration
laboratory.
Co
The user will be immediately notified and will determine if any
standard or equipment that was previously calibrated, or if any
material tested, by the out-of-specification item will require re-
calibration or retest and will take appropriate action.
d. Responsible organizational elements will take timely and positive
corrective actions to prevent recurrence.
M.J
3-9
-
7/25/2019 NASA NHB 5300.2 Apollo Metrology Requirements Manual
22/34
3.11 Standards and Equipment Control Requirements
3.11
FACILITIES
The individual facility requirements that are necessary for laboratories
performing calibration will depend upon the complexity of the standards
and equipment utilized. All standards and equipment will be cali-
brated and used in an environment that is controlled to the extent
necessary to insure continued measurements of the required accuracy.
3.11.1
Requisites.
The following facility requisites will be established and
maintained compatible with the performance requirements of the
standards and equipment.
a. Files for calibration procedures and record keeping facilities
for laboratory use.
b. Adequate space and good housekeeping for effective use of items
assigned to each area.
c. Secured cabinets or areas for custody of standards or com-
parison items to prevent damage and unauthorized usage.
d. Storage and control of repair parts for repair capability.
e. A
secured
area for storage of obsolete and defective standards
and
equipment.
f.
A cleaning area to be located other than in the laboratory to
remove dust and foreign particles from items before they
enter the laboratory.
g. Controlled temperature, humidity, vibration, and a cleanatmos-
phere, as required, including personnel cleanliness and dress.
h.
Freedom from vibration in critical areas.
i. Lighting to provide adequate shadowless lighting at working
surfaces.
j.
k.
Shielded room equipped with filtered power in which calibration
will be performed on articles sensitive to electrical, electro-
magnetic, or radio frequency interference (RFI).
Separate environmental controlled (where necessary to maintain
accuracy} and segregated areas for incoming, in-process, and
outgoing items.
I. Exclusion of all but authorized personnel inlaboratory facilities.
m, Permanent setup of items that are required for special or
frequently performed calibration to reduce possibility of human
error and to reduce labor costs in making repetitive setups.
V
%.t
3-10
-
7/25/2019 NASA NHB 5300.2 Apollo Metrology Requirements Manual
23/34
Standards and Equipment Control Requirements 3.12.2
. i
V
3.12
n.
Standby power for items that will lose accuracy due to primary
power failure.
OB
Environmental controlled mobile capability (if required)for
pickup or delivery of sensitive standards or equipment and
for transporting standards or equipment to make local cali-
bration of items not readily movable.
PERSONNEL
The effectiveness of the Apollo Metrology Program islargely dependent
upon the knowledge, skill and devotion to duty of the personnel assigned
to the laboratories. To attain these attributes, proper selection and
training are required.
3.12.1 Selection. Selection of personnel will be based on their knowledge
and skill and by the evaluation of the following characteristics:
a. Demonstrated understanding of the meaning of precision meas-
urements.
b. Sense of responsibility.
c. Manipulative skill.
d. Educational background of the basic physical sciences and
mathematics.
e. Motivation with the appreciation of the importance of the work
and its impact on the end item,
3.12.2
Training. Each facility will recognize its needs for training its
personnel in common, as well as, unique or specialized skills in
keeping with their responsibilities and types of measurements. These
training needs will be met by a training program including:
a. Indoctrination and familiarization.
b. Training in operations of a specialized nature prior to perform-
ance of duties.
c. On-the-job training.
d. Certification and recertification for personnelperforming spe-
cialized tasks.
e. Assessment to determine requirements for additional training
to revise activities compatible with the state-of-the art.
3-11
-
7/25/2019 NASA NHB 5300.2 Apollo Metrology Requirements Manual
24/34
3.13 Standards and Equipment Control Requirements
3.13 INTERLABORATORY COMPARISON
Enter Center, site and prime contractor will participate and contribute
to an interlaboratory comparison program for the purpose of verifying
the measuring capabilities among laboratories. A procedure for inter-
laboratory comparison will be implemented by a centrally controlled
authority as outlined in the Apollo Interlaboratory Comparison Pro-
cedure. This procedure will be based on the comparison of data taken
at the various laboratories with data from national reference standards,
through the use of an established comparison
package
of standards
and/or e quipment.
3.14 EVALUATION
3.14.1 Procedure. Chapter 4 establishes the procedure for surveying and
evaluating the conformance to the requirements of this Manual.
3.14.2 Responsibility
a. The Apollo Quality Assurance Office (Code MAR-Q), NASA
Headquarters, is responsible for evahating periodically the
conformance of the MSF Centers and their responsible organiza-
tions to the
provisions
of this Manual.
b. The MSF Centers are responsible for evaluating the compliance
of their sites and prime contractors to the provisions of this
Manual.
c. Each MSF Center, site and prime contractor is responsible
for evaluating its own performance against the requirements
of this Manual. All evaluations should be performed by an
impartial team composed of personnel not having specific line
responsibilities in the areas to be surveyed. Applicable written
instructions covering policy and procedures should be supplied
to each team. In addition, periodic inspections or cross-checks
will be made to detect damaged or inoperative calibration
standards or equipment, erratic readings, improper use of
standards and equipment, and other
performance
degrading
factors.
3-12
-
7/25/2019 NASA NHB 5300.2 Apollo Metrology Requirements Manual
25/34
k_/
4.1
4.2
4.3
EVALUATION
PURPOSE AND SCOPE
This Chapter sets forth:
a.
bo
CHAPTER 4
PROCEDURE
Methods and techniques for conducting surveys and evaluating degree
of conformance of Centers, sites and prime contractors.
Evaluation checklists which include major activity work elements to
provide for evaluating the degree of effective coverage of each
activity described in Chapter 3.
SURVE YOR' S PRE PARATION
The surveyor will:
a. Determine those rnetrology system requirements that are made
applicable by contractual documents.
b.
Ce
Determine, dependent on the requirements of the end-item, the degree
of compliance necessary to provide for the accuracy, stability, and
range that is required in the control and use of standards and
inspection, measuring and test equipment.
Be generally knowledgeable of measurement systems, requirements
and facilities.
APPLICABLE DOCUMENTS
NPC 200- 1A
NPC 200-2
NPC 200-3
NASA SP-6003
To Be Issued Later
To Be Issued Later
Quality Assurance Provisions for Govern-
ment Agencies
Quality Program Provisions for Space
System Contractors
Inspection System Provisions for Sup-
pliers of Space Materials, Parts, Com-
ponents, and Services
Quality Program Evaluation Procedures
Index and Format for Calibration Proce-
dures
Apollo Interlaboratory Comparison Pro-
cedure
4-I
-
7/25/2019 NASA NHB 5300.2 Apollo Metrology Requirements Manual
26/34
4.4 Evaluation Procedure
4.4 PERFORMANCE AND EVALUATION
4.4.1 Performance.
The performance of the survey should be conducted
by evaluating the work elements of the major activities established
in par. 4.5. For the adequate evaluation of each of these work
elements, it will be necessary to determine answers within the
following two areas:
a. Procedures-- Determine that documents are available which
describe the method of performance of each major activity as
detailed in Chapter 3.
b. Performance--Determine
that the implementation of the proce-
dures is accomplished to achieve the desired results. The deter-
mination of performance may be accomplished by actual observa-
tions of operations or by the review of applicable records that
would indicate degree of performance.
4.4.2 Evaluation. It is suggested that the initial approach to be used
during a survey should be to determine the existence and adequacy
of procedures that require the implementation of each work element
within each major activity [i.e., Selection, Evaluation and Acceptance,
Traceability, etc.). The next step would be to survey and evaluate
the performance against these established procedures. Each of these
steps require the use of the evaluation checklists in par. 4.5.
4.4.3 Reporting.
After completing the survey, a review meeting will be
held with the personnel responsible for the metrology program at
the
Center,
site or contractor being evaluated. This review should
point out major points of deviation from the requirements of this
Manual so that
corrective
action
may
be started. The discussion
with the Center, site or contractor should also emphasize the good
points in their metrology system. A formal report should be
developed and sent to the applicable Center, site, or contractor
for their use in correcting any discrepancies. This report will be
utilized by the evaluating function as a reference for corrective
action follow-up.
4.5 METROLOGY REQUIREMENTS EVALUATION CHECKLISTS
Each of the major activities set forth in Chapter 3 have been included
in the following checklists. These checklists contain work elements
including columns which will indicate the adequacy of coverage by nota-
tion of Satisfactory ( S ), Unsatisfactory ( U ), or Not Applicable ( N/A ).
When
an element is considered 'rN/A or U , it should be explained in
the Remarks column. If a more formal scoring is desired, the surveyor
may use a numerical grading system as outlined in the Quality Program
Evaluation Procedures (NASA SP-6003).
v
4-2
-
7/25/2019 NASA NHB 5300.2 Apollo Metrology Requirements Manual
27/34
Evaluation Procedure 4.5
I--
w
U
-1-
U
Z
0
..I
>-
uJ
I--
Z
IM
_E
gJ
e
i
CY
ev
0
-J
0
I--
IJ,J
0
u.
ul
.J
IE
_3
or..l
ul
v
0 1
l,o
.1
v
E_
__ _ __ o
_._:
_
d
::::::::: :
eiioll
eeeeee
oleeOeJJ
oeoole
eoeeaeoeo
: : : : : : : : :
d
::1::::':
._
1:::::::
d
........
_ : 1: ..
_.:.::::
i _ _
d
-
7/25/2019 NASA NHB 5300.2 Apollo Metrology Requirements Manual
31/34
Evaluation Procedure 4.5
o
P_
E_
v
b-
v
.a _ E
o._
U.I
4-?
-
7/25/2019 NASA NHB 5300.2 Apollo Metrology Requirements Manual
32/34
4.5 Evaluation
Procedure
t-
O
(J
I-
ra
_1
v
U
W
.1-
z
0
I-
_.1
z
Cy
Iii
Iy
>..
0
..I
0
I--
iii
0
II.
4-8
r_
8
11
OJ
1
0
0
-1
Z