NASA NHB 5300.2 Apollo Metrology Requirements Manual

download NASA NHB 5300.2 Apollo Metrology Requirements Manual

of 34

Transcript of NASA NHB 5300.2 Apollo Metrology Requirements Manual

  • 7/25/2019 NASA NHB 5300.2 Apollo Metrology Requirements Manual

    1/34

    NHB 5300.2

    APOLLO PROGRAM

    DECEMBER1965

    APOLLO

    3

    _09 W dOd A, ,L II I_ *4 d

    METROLOGY

    REQUIREMENTS

    MANUAL

    NATIONAL AERONAUTICS AND

    Washington, D.C.

    SPACE ADMINISTRATION

    20546

  • 7/25/2019 NASA NHB 5300.2 Apollo Metrology Requirements Manual

    2/34

    _w

    V

    r_

  • 7/25/2019 NASA NHB 5300.2 Apollo Metrology Requirements Manual

    3/34

    NASA Issuance No. NHB 5300.2

    Apollo Documentation No. RA-006-008-1

    PREFACE

    Prevention

    of

    interface anomalies during Apollo

    system hardware integration will, to a large extent,

    depend upon the compatibility of measurement pro-

    cedures presently being utilized throughout the NASA

    and Contractor network.

    Implementation of the Apollo Metrology Requirements

    Manual will insure that uniform measurements of

    high accuracy are achieved during the Apollo Program.

    The criteria established within this document are to

    be considered a clarification of the requirements

    contained in NPC 200-2, Quality Program Provisions

    for Space System Contractors.

    Major General, USAF

    Director, Apollo Program

    Distribution:

    Special

  • 7/25/2019 NASA NHB 5300.2 Apollo Metrology Requirements Manual

    4/34

    V

    T

    ; j

    1 I

    i

    il

  • 7/25/2019 NASA NHB 5300.2 Apollo Metrology Requirements Manual

    5/34

    k_J

    TABLE OF CONTENTS

    Subject

    Chapter 1 Introduction

    i.I

    I.Z

    1.3

    Purpose ...................................... 1- 1

    Scope 1- 1

    Applicability 1- 1

    Z.I

    Chapter 2: Definitions

    Applicability 2- 1 . .

    3.1

    3.2

    3.3

    3.4

    3.5

    3.6

    3.7

    3.8

    3.9

    3.10

    3.11

    3.12

    3.13

    3.14

    Chapter 3: Standards and Equipment Control Requirements

    Major Activities ................................ 3- 1

    Operating Procedures ............................ 3- 1

    Selection ..................................... 3- 1

    Evaluation and Acceptance ......................... 3-2

    Traceability ................................... 3-3

    Calibration Cycles ............................... 3-4

    Calibration Procedures ........................... 3-4

    Records ...................................... 3-7

    Recall ....................................... 3-8

    Corrective Action ............................... 3-9

    Facilities ..................................... 3- 10

    Personnel .................................... 3- 1 1

    Interlaboratory Comparison ........................ 3- 12

    Evaluation .................................... 3- 12

    4.1

    4.2

    4.3

    4.4

    4.5

    Chapter 4: Evaluation Procedures

    Purpose and Scope ............................... 4- 1

    Surveyor's Preparation ........................... 4-1

    Applicable Documents ............................ 4- 1

    Performance and Evaluation ........................ 4-2

    Metrology Requirements Evaluation Checklists ............ 4-2

  • 7/25/2019 NASA NHB 5300.2 Apollo Metrology Requirements Manual

    6/34

    i

  • 7/25/2019 NASA NHB 5300.2 Apollo Metrology Requirements Manual

    7/34

    I.I PURPOSE

    CHAPTER 1

    INTRODUCTION

    1.2

    The Apollo Metrology Requirements Manual establishes specific require-

    ments and associated criteria for a coordinated Metrology System,

    including the establishment of an Evaluation Procedure (see Chapter 4).

    This Manual provides for developing and controlling calibration practices

    for standards and measuring equipment used in the development, manu-

    facture, inspection and test of hardware.

    SCOPE

    This Manual includes control practices for all inspection and test

    standards, gages, tooling, measuring, and test equipment used for

    calibrating, data collecting, testing, and inspecting to determine the

    conformance of hardware to design specifications.

    1.3 APPLICABILITY

    This Manual is applicable to all Centers, sites and their prime contrac-

    tors who are responsible for the selection, evaluation, approval, cali-

    bration, maintenance, and control of standards and measuring equipment.

    1-1

  • 7/25/2019 NASA NHB 5300.2 Apollo Metrology Requirements Manual

    8/34

  • 7/25/2019 NASA NHB 5300.2 Apollo Metrology Requirements Manual

    9/34

    M.J

    CHAPTER 2

    DEFINITIONS

    2.1 APPLICABILITY

    To promote mutual understanding and to assist in effective communication

    of requirements, the following definitions are applicable to this Manual.

    a.

    Calibration--Comparison between two instruments or devices (one of

    which is a standard of known accuracy) to detect, correlate, report,

    and/or adjust any aviation in the accuracy of the instrument being

    compared.

    b. Calibration Cycle--An

    established interval between calibrations of

    standards or inspection, measuring and test equipment.

    c.

    CalibrationProcedures--The

    detailed instructions for performing

    calibration.

    do

    Clean Room--An

    enclosed area employing control over the particu-

    late matter in air with the additional control of temperature, humidity,

    and pressure, as required.

    eu

    Commercial

    Items--Catalog

    standards and inspection, measuring and

    test equipment considered to be off-the-shelf and/or available

    comme r cally.

    f.

    Historical Record--Record of identification, and accumulation of main-

    tenance and calibration data for each standard and inspection, measur-

    ing and test equipment to provide the data required for determining

    stability and performance trends.

    go

    Inspection, Measuring and Test Equipment--Equipment

    (including

    tooling) used to inspect, measure or test hardware to determine

    conformance to applicable design specifications. Tooling includes

    tools, gages, jigs and fixtures which measure dimensions, contours,

    or locations affecting quality characteristics.

    (Inspection, measuring and test equipment are hereafter referred to

    in this Manual as equipment. )

    h. Laboratory--Any segregated area specifically equipped for calibration

    of standards and inspection, measuring, and test equipment.

    io

    Metrology--The

    science and technology of measurement of any pa-

    rameter of weight, mass, length, chemical and electrical unit, and

    physical constant. This technology includes the selection, utilization,

    and control of equipment used to provide measurement, and also,

    includes assurance of a valid relationship between measured values

    2-1

  • 7/25/2019 NASA NHB 5300.2 Apollo Metrology Requirements Manual

    10/34

    Z.i Definitions

    2-2

    a

    j.

    I.

    m.

    n.

    o.

    P.

    q_

    ro

    and values of measurement established by national reference standards

    and international agreements.

    Mobile Capability--Ability

    to transport facilities, standards and

    inspection, measuring and test equipment, and the ability to

    perform

    calibration at remote locations.

    NASA (or Contractor) Reference Standard--Measure

    m e n t standards

    representing the highest level of accuracy available within the NASA

    (or Contractor) facilities. These are standard references against

    which lower level standards are compared, and are calibrated to

    national reference standards or are defined and sanctioned by inter-

    national agreement.

    National Bureau of Standards (NBS)--A

    central Government agency

    that has the responsibility for the custody, maintenance, and develop-

    ment of the national standards of measurement together with the

    provisions of calibration services related to these standards. This

    agency has standards laboratories in Washington, D.C., and Boulder,

    Colorado. (In addition to NBS, reference standards are maintained

    throughout the world, such as at the National Physical Laboratory in

    England and the National Research Council in Canada.)

    National Refernce Standards

    --The highest level national references

    which are: standards maintained by the National Bureau of Standards;

    acceptable standards maintained by other recognized agencies where

    an NBS reference is not available; or values of natural physical

    constants, such as, the acceleration due to gravity or standard

    reference materials.

    Recall System--A documented system to insure that all standards

    and inspection, measuring, and test equipment are recalled from the

    users at predetermined intervals for the purpose of calibration.

    Repeatability--The

    ability of a standard or measurement equipment

    to repeat a measurement within some known value when used by one

    person under specified conditions.

    Reproducibility--The

    difference in the average of groups of repeated

    measurements made by different persons using the same standard or

    measurement equipment under specified conditions.

    Shielded Room--

    A screened or metal enclosed area to prevent out-

    side electrical interference or radiation from affecting equipment

    performance or calibration measurements, and is provided with a

    filtered power supply.

    Site--NASA installations and component field activities geographically

    located apart from the parent installation (i.e., MSC White Sands

    Test Facility, Mississippi Test Facility, Michoud Assembly Facility,

    etc.)

  • 7/25/2019 NASA NHB 5300.2 Apollo Metrology Requirements Manual

    11/34

    Definitions 2. I

    Se

    te

    We

    De

    Special Equipment--Equipment

    of special design utilized for specific

    measurement(s) and not normally available commercially.

    Standard-

    -An item which is established as an authoriz'ed or recognized

    measurement reference, and is used to calibrate other standards or

    to calibrate inspection, measuring and test equipment.

    Systern--A documented method for the consistent performance of a

    task or associated group of tasks.

    Transfer Standavd--Any standard employed to transmit accuracies

    from a NASA (or Contractor) Reference Standard toworking standard.

    Workir_ Standard--Any

    standard used in the calibration of inspection,

    measuring, and test equipment.

    2-3

  • 7/25/2019 NASA NHB 5300.2 Apollo Metrology Requirements Manual

    12/34

    _J

    i

    Ii

  • 7/25/2019 NASA NHB 5300.2 Apollo Metrology Requirements Manual

    13/34

    %J

    CHAPTER 3

    STANDARDS AND EQUIPMENT CONTROL REQUIREMENTS

    3.1 MA:_OR ACTIVITIES

    The coordinated Apollo Metrology Program will include the major

    activities as delineated in pars. 3.3 through 3.14. The activities provide

    for the control of inspection, measuring and test equipment (hereafter

    referred to as equipment ) and for the control of standards. Criteria

    necessary for meeting these activity requirements are outlined and will

    be included in operating systems at each Center, site or prime contrac-

    tor.

    3.2 OPERATING PROCEDURES

    Written operating procedures will be developed and maintained for all

    systems based on criteria specified in this document. These written

    procedures

    will be the basis for continuous review, evaluation and con-

    sistent implementation regardless of personnel changes. Therefore,

    the procedures will include all practices for the complete Metrology

    Program as specified in the following paragraphs.

    3.3 SELECTION

    The selection of standards and equipment willbebased on the following

    criteria:

    a. The accuracy, stability, and range

    of

    standards and equipment will

    be in conformance with end-use requirements. Within state-of-the-

    art limitations, the standards used for calibration of inspection,

    measuring and test equipment will have a tolerance no greater than

    10 per cent of the allowable tolerance of the equipment being

    calibrated, i.e., if the item to be calibrated has a tolerance of fl.0

    per cent, the maximum allowable tolerance of the standard should

    be -+0.1 per cent. If the state-of-the-art prevents the attainment of

    this 10 to 1 ratio, authorization must be secured prior to use.

    b. The selection will not result in unnecessary duplication of a capa-

    bility that is available and canfeasiblybe utilized within the organiza-

    tion, through an associated laboratory (NASA, DOD, etc.) or at another

    Center, site or prime contractor.

    c. Commercial or off-the-shelf standards or equipment will be

    selected in lieu of special items, where feasible.

    3-1

  • 7/25/2019 NASA NHB 5300.2 Apollo Metrology Requirements Manual

    14/34

    3.4 Standards and Equipment Control Requirements

    d. When several different types of standards or equipment are adequate

    to perform a specific function(s), selection will be based on:

    (1) Availability.

    (2) Histories of like standards or equipment.

    (3) Analysis of interlaboratory comparisons.

    (4) Analysis of defect reports.

    (5) Initial cost.

    (8) Maintenance expenditures.

    (7) Manufacturer's performance and service.

    e. Verification of the laboratory's (or alternate facility's) capability to

    perform calibration on the selected standard or equipment.

    3.4 EVALUATION AND ACCEPTANCE

    Standards and equipment that are received for evaluation and acceptance

    will be processed in a consistent manner to include, but not be limited to,

    the following actions:

    a.

    bo

    Standards and equipment will be tested and verified for intended use

    to the parameters given in the applicable standard and equipment

    specification.

    t

    Evaluations will be performed and documented to determine that the

    standard or equipment has the required accuracy for meeting end-

    use specifications and that the standard or equipment can be

    satisfactorily calibrated. The extent of these evaluations will be

    dependent on the complexity and end-use requirements of the item as

    follow s :

    (i)

    Commercial items (with sufficient information relative to accuracy,

    stability and repeatability), when used in a manner consistent with

    design, need not necessarily be subjected to evaluation tests.

    (2)

    Special standards or equipment will receive evaluation tests to in-

    sure compatibility with the product and supporting or related equip-

    ment. In addition, these special items, due to complexity and

    specific requirements, will receive evaluation under intended

    method of operation.

    (3)

    Complex and/or special standards or equipment will receive

    comprehensive evaluations for assurance that the end item being

    measured will be controlled within specified tolerances. These

    evaluations will include repeatability and reproducibility testing

    where necessary to provide assurance of required accuracies.

    3-2

    %J

  • 7/25/2019 NASA NHB 5300.2 Apollo Metrology Requirements Manual

    15/34

    Standards and Equipment Control Requirements 3.5.1

    c. An applicable defect report will be generated and maintained for

    standards and equipment that do not meet specifications.

    d. All standards and equipment that are accepted will be:

    (i) Entered into an inventory system (see par. 3.8.1).

    (2) Calibrated to a specific procedure (see par. 3.7).

    (3) Assigned a calibration cycle (see par. 3.6).

    (4) Sealed to prevent tampering by unauthorized personnel prior to

    next calibration.

    (5)

    Labeled or marked when possible, and entered into appropriate

    records to indicate date of calibration and due date for next cali-

    bration.

    (6) Entered into recall system (see par. 3.9).

    (7)

    Tagged for safety instructions, special usage {i.e., Indication

    Only, Reference Only, Limited Use) or special handling, special

    storage, and environmental requirements, as applicable.

    e.

    f.

    Data generated from the initial verification of standards and equip-

    ment will be maintained.

    Special standards or equipment specifications will be properly

    documented by drawings and/or specifications and maintained current.

    go

    All associated manufacturer's instructions or procedures, including

    records of any modification made to the item, will be maintained and

    kept current.

    3.5 TRACEABILITY

    .

    The standards used to calibrate lower level standards or equipment

    will have valid relationships to national reference standards. When this

    is not feasible due to state-of-the-art measurement, standards will be

    acceptable that are developed by ratio-type of self-calibrationtechniques

    using standards whose history and drift rates are known to be satis-

    factory.

    5.1 Evidence. Traceability to national reference standards will be

    evidenced by valid reports of calibration for each standard processed.

    These reports will indicate the date of calibration. If calibration of

    standards is performed by an alternate laboratory, their report

    will be maintained and will cross-reference an NBS report date and

    number or other national reference standard report to indicate the

    traceability and calibration status of standards used. Any standards

    that by their nature are not sent to NBS, or an alternate laboratory,

    for example, frequency standards, will be regularly monitored by a

    3-3

  • 7/25/2019 NASA NHB 5300.2 Apollo Metrology Requirements Manual

    16/34

    3.5.2

    Standards and EcLuipment Control Requirements

    method that is traceable to national reference standards and evi-

    denced by a permanent log ofthe date, time, and observed deviations.

    3.5.2

    Flow Charts. Appropriate flow charts will be developed to provide

    evidence of traceability of all equipment capabilities. These flow

    charts will show this traceability through a series of standards with

    records to substantiate relationships to national reference standards.

    3.6 CALIBRATION CYCLES

    To insure the maintenance of required accuracies for a particular standard

    or equipment, it is required that recall for calibration be performed at

    predetermined intervals. These intervals may be specified in terms of

    time or usage. Time intervals will be established in terms of date

    (month/day/year) so that the calibration label indicates the specific day

    of expiration. The cycles established will be implemented by the recall

    system (see par. 3.9).

    3.6.1 Initial Cycle. The establishment of the initial calibration cycle for

    incoming items will be based on the following, as applicable:

    a. Experience with like standards or equipment.

    b. Purpose.

    c.

    Amount

    of use.

    d. Accuracy requirements.

    e. Standard practices.

    f. Stability.

    3.7

    3.6.2

    Cycle Adjustment. Calibration cycles will be shortened as required

    to insure continued accuracy as evidence by the results of

    preceding

    calibrations, and may be lengthened only whenthe results of analysis

    of data provide definite indications that such actions will not

    adversely affect the accuracy.

    CALIBRATION PROCEDURES

    Written procedues will be used for calibrating standards or equipment.

    These written procedures will be readily available to laboratory personnel

    performing the calibration. There will be a means of insuring that an up-

    to-date procedure is used that includes a method of indicating and con-

    trolling necessary revisions.

    3.7.1

    Procedures Index and Format. The Index and Format for Calibra-

    tion Procedures,

    provides

    for the listing of all available procedures

    developed or being developed by each MSF

    Center

    and outlines a

    procedures format. The Index will be provided to all MSF

    Centers

    for their use in establishing the availability of existing and planned

    3-4

  • 7/25/2019 NASA NHB 5300.2 Apollo Metrology Requirements Manual

    17/34

    Standards and Equipment Control Requirements 3.7.3

    procedures for calibrating standards or equipment. The use of

    this Index by the Centers will:

    a. Provide knowledge of availability of Center-proven procedures.

    b. Reduce duplication of efforts in producing new procedures.

    c. Provide for attaining standardization of new procedures as they

    are developed.

    3.7.2

    d. Improve communication of like functions between Centers.

    Procedure Development. Procedures that are developed will be

    based on the Format in the Index and Format for Calibration

    Procedures and will contain the following information:

    a. Safety warnings and cautions.

    b. Environmental control details

    to

    insure measurements of re-

    quired accuracy including temperature, humidity, vibration,

    cleanliness, etc.

    C,

    do

    ea

    List of calibration equipment required and a diagram of complex

    setups (consider accuracy ratios as discussed in par. 3.3a).

    The actual operations listed in a series of sequential steps with

    applicable tolerances.

    A record keeping form or data sheet to include identification of

    personnel performing calibration, condition of items previous

    to calibration, final results, calibration time/cost, and other

    information pertinent to a specific item.

    f. Instructions for special handling of a standard or equipment

    and required environmental controls after calibration.

    g. Provisions for sealing standard or equipment to prevent

    tampering between calibrations.

    h. Provisions for documenting for the user any adjustments neces-

    sary due to changes in environments.

    i. For special equipment which is composed of several individual

    items, the calibration of any one item may change the interrelated

    results of the configuration. Therefore, procedures should

    provide for calibration of the entire special equipment, if this

    situation is encountered.

    3.7.3 NASA (or Contractor) Reference Standards

    a. The highest level standards that are available at a particular

    location will be calibrated by the National Bureau of Standards

    3-5

  • 7/25/2019 NASA NHB 5300.2 Apollo Metrology Requirements Manual

    18/34

    3.7.4 Standards and Equipment Control Requirernents

    3-6

    (NBS) by another approved laboratory either directly or by use

    of a special transfer standard if the Reference Standard cannot

    be transported or by ratio-type of self-calibration techniques

    (see par. 3.5). If the NBS performs this calibration, a report

    will be issued for information and file which will show its

    traceability. If an alternate laboratory performs the calibration,

    it will be insured that this laboratory is performing the calibra-

    tion to an approved procedure which includes proper environ-

    ments, and the use of standards having the required accuracies;

    these standards will have current reports that indicate direct

    traceability to a national reference standard.

    b.

    A calibration procedure may not be needed for certain Reference

    Standards, i.e., frequency standards, and they might not be sent

    to NBS or an alternate laboratory for calibration. In these cases,

    there will be a monitoring method consisting of a permanent log

    showing dates, times and observed deviations that are directly

    traceable to national reference standards.

    3.7.4 Working and Transfer Standards

    a.

    If the standards at the working and transfer levels require cali-

    bration at a different laboratory due to lack of facilities or lack

    of adequate standards available locally, it will be insured that the

    alternate laboratory performing the calibration is using approved

    procedures (including proper environments and standards having

    the required accuracies that are traceable to national reference

    standards).

    b*

    When calibration is performed locally, it is essentialthat formal,

    approved procedures be used. Since most of these standards

    would fall into the category of a commercial item, a common

    procedure should be available as provided for in the Index and

    Format for Calibration Procedures. When standards are special

    or a common procedure is not available, a procedure will be

    developed containing the information outlined in par. 3.7.2.

    C.

    Standards that are to be used only over a portion of the range or

    at certain points in the range may require calibration at these

    points or in the ranges used. If this is performed, the calibration

    points or ranges will be clearly indicated on the standard.

    3.7.5 Equipment Calibration

    a.

    The calibration of equipment will be performed with working

    standards in the required environment and at the intervals

    established to provide the required confidence level of the

    equipment. Each distinct type of equipment requires that a

    calibration procedure be available for consistent operating

    practices, and that the standards that are used are supported by

    a train of successive reports that trace back to a national

    reference standard.

    v

    %J

    i

  • 7/25/2019 NASA NHB 5300.2 Apollo Metrology Requirements Manual

    19/34

    Standards and Equipment Control Requirements 3.8.2

    b.

    Equipment that is to be used only over a portion of the range

    or at certain points in the range may require calibration at these

    points or in the ranges used. If this is performed, the calibration

    points or ranges will be clearly indicated on the equipment.

    3.8 RECORDS

    Each standard and equipment will be supported by individual records

    detailing all inventory and historical data. These records are necessary

    to implement each of the control systems that are developed.

    3.8.1

    Inventory Records and Control. Inventory records of standards and

    equipment are required to provide availability, status and capa-

    bility data essential to ensure proper inventory control. Inventory

    control will include:

    a.

    A positive nonduplicating inventory identification code to be

    assigned to all standards and equipmert and to be affixed to

    the item wherever it does not interfere with usage.

    b. A record of each standard and equipment by nomenclature,

    model number, serial number, manufacturer, date acquired,

    physical location (user), range, accuracy and identification code.

    c. Provisions for removing standards or equipment from inventory

    when the items become obsolete or beyond economical repair.

    d.

    All personally owned equipment which will be recorded with

    owner's name or payroll number, complete description of

    'equipment and physical location. {Personally owned equipment

    may be used only when properly authorized.}

    3.8.2

    Historical.

    These records will provide objective evidence that

    calibration schedules are complied with and that the accuracy of

    the standard or equipment is being maintained. The records also

    provide a history of stability, which may be evaluated and utilized

    as a basis for the adjustment of calibration cycles or utilized to

    indicate trends of wear and deterioration and may become a basis

    for future selection of standards and equipment. Historical

    records will include, but not be limited to, the following:

    a. Data generated during calibration as required by the applicable

    procedures (see par. 3.7.2e).

    b. Provisions for providing data that will accompany standards or

    equipment when returned to the user.

    Co

    do

    e.

    Data generated during the initial verification and evaluation and

    data from any subsequent evaluations of a standard or equipment.

    Calibration cycle.

    Range and accuracy.

    3-7

  • 7/25/2019 NASA NHB 5300.2 Apollo Metrology Requirements Manual

    20/34

    3.9 Standards and Equipment Control Requirements

    f. Special usage information.

    g, Calibration time/cost.

    h. Maintenance data (repairs required and list of parts replaced).

    i. Hours of usage for critical items where the usage factor is

    important.

    j. The individual record of a standard or equipment (whose

    accuracy is reported via a formal report) will cross-reference

    the report number. This is especially important for NBS reports

    that are received with specific standards.

    3.9 RECALL

    3.9.1

    Documented System. A documented recall system will be established

    to insure thatal__ standards and al__jl equipment (incIuding personally

    owned) are recalled for calibration at established intervals (cali-

    bration cycles).

    3.9.2

    Maintenance of Records.

    The required records for standards and

    equipment will be maintained and keyed to a specific calibration

    cycle and a specific calibration procedure. These records will

    include identification, range, accuracy requirements, and special

    usage infromation. The following subparagraphs state the basic

    criteria that must be included in a recall system.

    3.9.3

    Notification.

    Notification will be given to each user prior to the

    established calibration due date for the assurance that standards and

    equipment are:

    a. Removed or made ready for calibration when required.

    b. Not used after calibration due date.

    The established calibration due dates do not relieve the user of the

    responsibility of requesting supplementary calibration when a

    standard or equipment is damaged, misused or suspected of being

    out- of- tole r ance.

    3.9.4

    Mandatory

    Removal.

    A mandatory removal system will provide

    delegation of authority for insuring the removal from use of stand-

    ards or equipment that are beyond the calibration due date. These

    items should be removed from service immediately. If physical

    removal is impractical due to weight, service connections or other

    characteristics of the item, it will be impounded by attachment of

    signs, tags, or other means. However, consideration will be given

    by the authority to the feasibility of temporarily extending the cali-

    bration beyond the due date without degrading accuracies, when the

    removal of the items would require teardown of costly setups or

    interruption of tests. A tagging or other identification system will

    3-8

  • 7/25/2019 NASA NHB 5300.2 Apollo Metrology Requirements Manual

    21/34

    Standards and Equipment Control Requirements 3.10

    3.9.5

    3.9.6

    3.9.7

    3.10

    be used for identifying items that have not been maintained, or

    calibrated in accordance with an established schedule.

    Delinquent Notices. Notification of delinquency will be given when

    standards or equipment are not calibrated by the responsible

    laboratory as scheduled. Responsibility will be established for

    follow-up and positive corrective action involving delinquent items.

    Labeling. Each standard or equipment will be labeled or marked

    to indicate the date when next calibration is due. This label {or

    appropriate record) will include date of calibration and identification

    of the person performing the calibration. When item identification

    does not appear on the label, a method will be established to insure

    that labels

    cannot

    be interchanged from one item to another. If

    the calibration cycle is based on usage or another system not as-

    sociated with a time interval, the item and/or records will readily

    indicate calibration status. To preclude the misuse of a standard or

    equipment that is an integral part of anassembled item, the label-

    ing or marking will identify the next higher assembly in which that

    standard or equipment must be used.

    Pickup and Delivery. A pickup and delivery service for all standards

    and equipment will be established. This service is the responsi-

    bility of the applicable laboratory, which will provide for handling

    and transportation dependent upon the complexity and delicacy of

    the standards and equipment. A receipt method will be used to

    insure that the individual standard or equipment has been received

    by the laboratory and as subsequent evidence that the item has been

    returned to the user.

    CORRECTIVE ACTION

    When during normal use, calibration, or as a result of periodic inspec-

    tions or crosschecks (see par. 3.14.2), it is determined that a standard

    or equipment is out-of-specification or that there is indication of

    malfunction, immediate corrective actions will be taken as follows:

    a. Standards or equipment will be properly tagged to prevent Usage.

    b. Standards or equipment in use will be returned to the calibration

    laboratory.

    Co

    The user will be immediately notified and will determine if any

    standard or equipment that was previously calibrated, or if any

    material tested, by the out-of-specification item will require re-

    calibration or retest and will take appropriate action.

    d. Responsible organizational elements will take timely and positive

    corrective actions to prevent recurrence.

    M.J

    3-9

  • 7/25/2019 NASA NHB 5300.2 Apollo Metrology Requirements Manual

    22/34

    3.11 Standards and Equipment Control Requirements

    3.11

    FACILITIES

    The individual facility requirements that are necessary for laboratories

    performing calibration will depend upon the complexity of the standards

    and equipment utilized. All standards and equipment will be cali-

    brated and used in an environment that is controlled to the extent

    necessary to insure continued measurements of the required accuracy.

    3.11.1

    Requisites.

    The following facility requisites will be established and

    maintained compatible with the performance requirements of the

    standards and equipment.

    a. Files for calibration procedures and record keeping facilities

    for laboratory use.

    b. Adequate space and good housekeeping for effective use of items

    assigned to each area.

    c. Secured cabinets or areas for custody of standards or com-

    parison items to prevent damage and unauthorized usage.

    d. Storage and control of repair parts for repair capability.

    e. A

    secured

    area for storage of obsolete and defective standards

    and

    equipment.

    f.

    A cleaning area to be located other than in the laboratory to

    remove dust and foreign particles from items before they

    enter the laboratory.

    g. Controlled temperature, humidity, vibration, and a cleanatmos-

    phere, as required, including personnel cleanliness and dress.

    h.

    Freedom from vibration in critical areas.

    i. Lighting to provide adequate shadowless lighting at working

    surfaces.

    j.

    k.

    Shielded room equipped with filtered power in which calibration

    will be performed on articles sensitive to electrical, electro-

    magnetic, or radio frequency interference (RFI).

    Separate environmental controlled (where necessary to maintain

    accuracy} and segregated areas for incoming, in-process, and

    outgoing items.

    I. Exclusion of all but authorized personnel inlaboratory facilities.

    m, Permanent setup of items that are required for special or

    frequently performed calibration to reduce possibility of human

    error and to reduce labor costs in making repetitive setups.

    V

    %.t

    3-10

  • 7/25/2019 NASA NHB 5300.2 Apollo Metrology Requirements Manual

    23/34

    Standards and Equipment Control Requirements 3.12.2

    . i

    V

    3.12

    n.

    Standby power for items that will lose accuracy due to primary

    power failure.

    OB

    Environmental controlled mobile capability (if required)for

    pickup or delivery of sensitive standards or equipment and

    for transporting standards or equipment to make local cali-

    bration of items not readily movable.

    PERSONNEL

    The effectiveness of the Apollo Metrology Program islargely dependent

    upon the knowledge, skill and devotion to duty of the personnel assigned

    to the laboratories. To attain these attributes, proper selection and

    training are required.

    3.12.1 Selection. Selection of personnel will be based on their knowledge

    and skill and by the evaluation of the following characteristics:

    a. Demonstrated understanding of the meaning of precision meas-

    urements.

    b. Sense of responsibility.

    c. Manipulative skill.

    d. Educational background of the basic physical sciences and

    mathematics.

    e. Motivation with the appreciation of the importance of the work

    and its impact on the end item,

    3.12.2

    Training. Each facility will recognize its needs for training its

    personnel in common, as well as, unique or specialized skills in

    keeping with their responsibilities and types of measurements. These

    training needs will be met by a training program including:

    a. Indoctrination and familiarization.

    b. Training in operations of a specialized nature prior to perform-

    ance of duties.

    c. On-the-job training.

    d. Certification and recertification for personnelperforming spe-

    cialized tasks.

    e. Assessment to determine requirements for additional training

    to revise activities compatible with the state-of-the art.

    3-11

  • 7/25/2019 NASA NHB 5300.2 Apollo Metrology Requirements Manual

    24/34

    3.13 Standards and Equipment Control Requirements

    3.13 INTERLABORATORY COMPARISON

    Enter Center, site and prime contractor will participate and contribute

    to an interlaboratory comparison program for the purpose of verifying

    the measuring capabilities among laboratories. A procedure for inter-

    laboratory comparison will be implemented by a centrally controlled

    authority as outlined in the Apollo Interlaboratory Comparison Pro-

    cedure. This procedure will be based on the comparison of data taken

    at the various laboratories with data from national reference standards,

    through the use of an established comparison

    package

    of standards

    and/or e quipment.

    3.14 EVALUATION

    3.14.1 Procedure. Chapter 4 establishes the procedure for surveying and

    evaluating the conformance to the requirements of this Manual.

    3.14.2 Responsibility

    a. The Apollo Quality Assurance Office (Code MAR-Q), NASA

    Headquarters, is responsible for evahating periodically the

    conformance of the MSF Centers and their responsible organiza-

    tions to the

    provisions

    of this Manual.

    b. The MSF Centers are responsible for evaluating the compliance

    of their sites and prime contractors to the provisions of this

    Manual.

    c. Each MSF Center, site and prime contractor is responsible

    for evaluating its own performance against the requirements

    of this Manual. All evaluations should be performed by an

    impartial team composed of personnel not having specific line

    responsibilities in the areas to be surveyed. Applicable written

    instructions covering policy and procedures should be supplied

    to each team. In addition, periodic inspections or cross-checks

    will be made to detect damaged or inoperative calibration

    standards or equipment, erratic readings, improper use of

    standards and equipment, and other

    performance

    degrading

    factors.

    3-12

  • 7/25/2019 NASA NHB 5300.2 Apollo Metrology Requirements Manual

    25/34

    k_/

    4.1

    4.2

    4.3

    EVALUATION

    PURPOSE AND SCOPE

    This Chapter sets forth:

    a.

    bo

    CHAPTER 4

    PROCEDURE

    Methods and techniques for conducting surveys and evaluating degree

    of conformance of Centers, sites and prime contractors.

    Evaluation checklists which include major activity work elements to

    provide for evaluating the degree of effective coverage of each

    activity described in Chapter 3.

    SURVE YOR' S PRE PARATION

    The surveyor will:

    a. Determine those rnetrology system requirements that are made

    applicable by contractual documents.

    b.

    Ce

    Determine, dependent on the requirements of the end-item, the degree

    of compliance necessary to provide for the accuracy, stability, and

    range that is required in the control and use of standards and

    inspection, measuring and test equipment.

    Be generally knowledgeable of measurement systems, requirements

    and facilities.

    APPLICABLE DOCUMENTS

    NPC 200- 1A

    NPC 200-2

    NPC 200-3

    NASA SP-6003

    To Be Issued Later

    To Be Issued Later

    Quality Assurance Provisions for Govern-

    ment Agencies

    Quality Program Provisions for Space

    System Contractors

    Inspection System Provisions for Sup-

    pliers of Space Materials, Parts, Com-

    ponents, and Services

    Quality Program Evaluation Procedures

    Index and Format for Calibration Proce-

    dures

    Apollo Interlaboratory Comparison Pro-

    cedure

    4-I

  • 7/25/2019 NASA NHB 5300.2 Apollo Metrology Requirements Manual

    26/34

    4.4 Evaluation Procedure

    4.4 PERFORMANCE AND EVALUATION

    4.4.1 Performance.

    The performance of the survey should be conducted

    by evaluating the work elements of the major activities established

    in par. 4.5. For the adequate evaluation of each of these work

    elements, it will be necessary to determine answers within the

    following two areas:

    a. Procedures-- Determine that documents are available which

    describe the method of performance of each major activity as

    detailed in Chapter 3.

    b. Performance--Determine

    that the implementation of the proce-

    dures is accomplished to achieve the desired results. The deter-

    mination of performance may be accomplished by actual observa-

    tions of operations or by the review of applicable records that

    would indicate degree of performance.

    4.4.2 Evaluation. It is suggested that the initial approach to be used

    during a survey should be to determine the existence and adequacy

    of procedures that require the implementation of each work element

    within each major activity [i.e., Selection, Evaluation and Acceptance,

    Traceability, etc.). The next step would be to survey and evaluate

    the performance against these established procedures. Each of these

    steps require the use of the evaluation checklists in par. 4.5.

    4.4.3 Reporting.

    After completing the survey, a review meeting will be

    held with the personnel responsible for the metrology program at

    the

    Center,

    site or contractor being evaluated. This review should

    point out major points of deviation from the requirements of this

    Manual so that

    corrective

    action

    may

    be started. The discussion

    with the Center, site or contractor should also emphasize the good

    points in their metrology system. A formal report should be

    developed and sent to the applicable Center, site, or contractor

    for their use in correcting any discrepancies. This report will be

    utilized by the evaluating function as a reference for corrective

    action follow-up.

    4.5 METROLOGY REQUIREMENTS EVALUATION CHECKLISTS

    Each of the major activities set forth in Chapter 3 have been included

    in the following checklists. These checklists contain work elements

    including columns which will indicate the adequacy of coverage by nota-

    tion of Satisfactory ( S ), Unsatisfactory ( U ), or Not Applicable ( N/A ).

    When

    an element is considered 'rN/A or U , it should be explained in

    the Remarks column. If a more formal scoring is desired, the surveyor

    may use a numerical grading system as outlined in the Quality Program

    Evaluation Procedures (NASA SP-6003).

    v

    4-2

  • 7/25/2019 NASA NHB 5300.2 Apollo Metrology Requirements Manual

    27/34

    Evaluation Procedure 4.5

    I--

    w

    U

    -1-

    U

    Z

    0

    ..I

    >-

    uJ

    I--

    Z

    IM

    _E

    gJ

    e

    i

    CY

    ev

    0

    -J

    0

    I--

    IJ,J

    0

    u.

    ul

    .J

    IE

    _3

    or..l

    ul

    v

    0 1

    l,o

    .1

    v

    E_

    __ _ __ o

    _._:

    _

    d

    ::::::::: :

    eiioll

    eeeeee

    oleeOeJJ

    oeoole

    eoeeaeoeo

    : : : : : : : : :

    d

    ::1::::':

    ._

    1:::::::

    d

    ........

    _ : 1: ..

    _.:.::::

    i _ _

    d

  • 7/25/2019 NASA NHB 5300.2 Apollo Metrology Requirements Manual

    31/34

    Evaluation Procedure 4.5

    o

    P_

    E_

    v

    b-

    v

    .a _ E

    o._

    U.I

    4-?

  • 7/25/2019 NASA NHB 5300.2 Apollo Metrology Requirements Manual

    32/34

    4.5 Evaluation

    Procedure

    t-

    O

    (J

    I-

    ra

    _1

    v

    U

    W

    .1-

    z

    0

    I-

    _.1

    z

    Cy

    Iii

    Iy

    >..

    0

    ..I

    0

    I--

    iii

    0

    II.

    4-8

    r_

    8

    11

    OJ

    1

    0

    0

    -1

    Z