NANOpure Analytical Deionization System bioresearch2.pdfAuxiliary Valve - Type I Reagent Grade Water...

48
1 NANOpure Analytical Deionization System OPERATION MANUAL AND PARTS LIST SERIES 851 ANALYTICAL GRADE SYSTEMS LT851X1 • 6/3/98 Serial Number BARNSTEAD|THERMOLYNE CORPORATION Model # Voltage Mount D4741 120 Wall D4742 240 Wall D4743 100 Wall D4744 120 Bench D4745 240 Bench D4746 100 Bench

Transcript of NANOpure Analytical Deionization System bioresearch2.pdfAuxiliary Valve - Type I Reagent Grade Water...

Page 1: NANOpure Analytical Deionization System bioresearch2.pdfAuxiliary Valve - Type I Reagent Grade Water (RGW) per ASTM-D1193, NCCLS ASC-3, and CAP. Flow rate (Maximum)2 Type 1 RGW3 Type

1

NANOpure AnalyticalDeionization System

OPERATION MANUALAND PARTS LIST

SERIES 851

ANALYTICAL GRADE SYSTEMS

LT851X1 • 6/3/98 Serial Number

BARNSTEAD|THERMOLYNE CORPORATION

Model # Voltage MountD4741 120 WallD4742 240 WallD4743 100 WallD4744 120 BenchD4745 240 BenchD4746 100 Bench

Page 2: NANOpure Analytical Deionization System bioresearch2.pdfAuxiliary Valve - Type I Reagent Grade Water (RGW) per ASTM-D1193, NCCLS ASC-3, and CAP. Flow rate (Maximum)2 Type 1 RGW3 Type

2

This manual contains important operating and safety information. You must carefullyread and understand the contents of this manual prior to the use of this equipment.

Control Panel

RemoteDispenser

Canisters

ultrapure water systemNANOpure

Auxiliary Valve

Page 3: NANOpure Analytical Deionization System bioresearch2.pdfAuxiliary Valve - Type I Reagent Grade Water (RGW) per ASTM-D1193, NCCLS ASC-3, and CAP. Flow rate (Maximum)2 Type 1 RGW3 Type

3

Table of Contents

Safety Information .............................................................................................................................................. 4Alert Boxes .................................................................................................................................................. 4Warnings ..................................................................................................................................................... 4

Technical Specifications..................................................................................................................................... 6Installation .......................................................................................................................................................... 7

Unpacking .................................................................................................................................................... 9Choosing a Site ........................................................................................................................................... 9Mounting and Utility Connections .............................................................................................................. 10Tubing Connector Installation .................................................................................................................... 12Initial Operation ......................................................................................................................................... 13

Normal Operation ............................................................................................................................................. 18Smart Purity Meter ........................................................................................................................................... 19

Modes of Operation ................................................................................................................................... 19Turning Unit On or Off ............................................................................................................................... 19Selecting the Set Point .............................................................................................................................. 20Displaying the Temperature ...................................................................................................................... 21Checking the Calibration ........................................................................................................................... 21Inlet Cell Installation .................................................................................................................................. 22Inlet Cell Operation .................................................................................................................................... 23

Remote Display Mounting ................................................................................................................................ 24Remote Dispenser ........................................................................................................................................... 26

Operation ................................................................................................................................................... 26Installing the 0.2 Micron Final Filter ........................................................................................................... 26

Float Switch/Pressure Switch Installation ........................................................................................................ 28Maintenance and Servicing .............................................................................................................................. 29

Replacing Cartridges ................................................................................................................................. 29Replacing the Remote Dispenser Filters ................................................................................................... 29Replacing the Auxiliary Valve 0.2 Micron Final Filter ................................................................................ 30Replacing Fuses ........................................................................................................................................ 31System Sanitization ................................................................................................................................... 31Cleaning the Resistivity Cell ...................................................................................................................... 34Shutdown ................................................................................................................................................... 35

Problem Solving Guide .................................................................................................................................... 36Replacement Parts Listing ............................................................................................................................... 39

Recommended Spares .............................................................................................................................. 39Consumables..................................................................................................................................... 39General Maintenance Parts ............................................................................................................... 40Safety Stock ...................................................................................................................................... 40

Ordering Procedures ........................................................................................................................................ 41Exploded View: 100 & 120 Volt Models ........................................................................................................... 42Exploded View: 240 Volt Models...................................................................................................................... 43Wiring Diagrams............................................................................................................................................... 44One Year Limited Warranty.............................................................................................................................. 47

Page 4: NANOpure Analytical Deionization System bioresearch2.pdfAuxiliary Valve - Type I Reagent Grade Water (RGW) per ASTM-D1193, NCCLS ASC-3, and CAP. Flow rate (Maximum)2 Type 1 RGW3 Type

4

Your Barnstead NANOpure has been designedwith function, reliability and safety in mind. It is yourresponsibility to install it in conformance with localelectrical codes. For safe operation, please payattention to the alert signals throughout the manual.

Warnings

To avoid electrical shock, always:1. Use with a properly grounded electrical

outlet of correct voltage and currenthandling capacity.

2. Do not mount the NANOpure directly overequipment that requires electrical service.Routine maintenance of this unit mayinvolve water spillage and subsequentelectrical shock hazard if improperlylocated.

3. Replace fuses with those of the same typeand rating.

4. Disconnect from the power supply prior tomaintenance and servicing.

To avoid personal injury:

1. Do not use in the presence of flammableor combustible materials; fire or explosionmay result. This device containscomponents which may ignite suchmaterials.

2. This device is to be used with water feedsonly. Sanitizing/cleaning agents must beused in compliance with instructions in thismanual. Failure to comply with the abovecould result in explosion and personalinjury.

Safety Information

WarningWarnings alert you to apossibility of personal injury.

CautionCautions alert you to apossibility of damage to theequipment.

NoteNotes alert you to pertinentfacts and conditions.

Alert Signals

Page 5: NANOpure Analytical Deionization System bioresearch2.pdfAuxiliary Valve - Type I Reagent Grade Water (RGW) per ASTM-D1193, NCCLS ASC-3, and CAP. Flow rate (Maximum)2 Type 1 RGW3 Type

5

3. Do not mount the NANOpure directly overequipment that requires electrical service.Routine maintenance of this unit mayinvolve water spillage and subsequentelectrical shock hazard if improperlylocated.

4. Replace fuses only with the same typeand rating for continued protection againstpossible fire hazard.

5. Wall composition and construction, as wellas fastener type, must be consideredwhen mounting this unit. The mountingsurface and fasteners selected must becapable of supporting a minimum of 275lbs. Inadequate support and/or fastenersmay result in damage to mounting surfaceand/or equipment. If you are unsure ofmounting surface composition, conditionand construction, or correct fasteners,consult your building maintenance groupor contractor.

6. Avoid splashing disinfecting solutions onclothing or skin. Ensure all pipingconnections are tight to avoid chemicalleakage. Always depressurize chemicallines before disassembly. Ensureadequate ventilation. Carefully followmanufacturer’s safety instructions onlabels of chemical containers and materialsafety data sheets.

7. Depressurize system prior to attempting toremove canisters.

8. Refer servicing to qualified personnel.

SAFETY INFORMATION

Page 6: NANOpure Analytical Deionization System bioresearch2.pdfAuxiliary Valve - Type I Reagent Grade Water (RGW) per ASTM-D1193, NCCLS ASC-3, and CAP. Flow rate (Maximum)2 Type 1 RGW3 Type

6

Technical Specifications

Analytical GradeFeed Water Requirements

Types1

Tap (Potable), RO, DI, Distilled.Pressure Range Gravity feed to 100 psig (7 kg/cm

2) maximum.

Temperature Range 4-49°C (40-120°F)Product Water

Water Quality1

Remote Dispenser - Type I Reagent Grade Water (RGW) per ASTM-D1193, NCCLS ASC-3, and CAP.Auxiliary Valve - Type I Reagent Grade Water (RGW) per ASTM-D1193, NCCLS ASC-3, and CAP.Flow rate (Maximum)

2

Type 1 RGW3

Type 1 RGW3

From Dispenser From Auxiliary ValvePressured Feed(40 psig (3 kg/cm

2) Inlet Min.) 50 HZ 1.5 LPM 1.3 LPM

60 Hz 1.6 LPM 1.5 LPMGravity Feed(12" (305 mm) Minimum) 50 Hz 1.0 LPM 0.8 LPM

60 Hz 1.1 LPM 1.0 LPM

DimensionsWall mounted models

Width 22" (559 mm)Depth 14 1/4" (362 mm)Height 29 1/2" (749 mm)

Bench mounted modelsWidth 22" (559 mm)Depth 16 1/4" (413 mm)Height 32 1/2" (826 mm)

Plumbing ConnectionsFeed water Inlet 3/8" OD tubing or 1/4" NPTF

Electrical Requirements (depending on model supplied)Voltage and Frequency (Nominal)

100 VAC, 50/60 Hz 85-110 VAC, 47-63 Hz, 1 phase115 VAC, 50/60 Hz 98-127 VAC, 47-63 Hz, 1 phase230 VAC, 50/60 Hz 196-253 VAC, 47-63 Hz, 1 phase

Protection100 VAC service 3 ampere slow blow fuse115 VAC service 3 ampere slow blow fuse230 VAC service 2 ampere slow blow fuse

Resistivity MeasurementRange 0.1-18.3 megohm-cm [temperature compensated to 25°C (77°F)]Accuracy ± 3 % FSCell 0.1 constantDisplay Digital

1NANOpure will produce Type I water usingpretreated water (RO, DI, Distilled) or highquality tap water, provided feedwater suitabilityis qualified by laboratory analysis andrecommended feed flowrate is maintained.

2Flowrates are dependent on operatingconditions and filter usage. Flowrates will alsodepend on filter compaction.

3Typical flowrate with 40 psig inlet and 30 psigoutlet pressure at faucet block.

Notes

Page 7: NANOpure Analytical Deionization System bioresearch2.pdfAuxiliary Valve - Type I Reagent Grade Water (RGW) per ASTM-D1193, NCCLS ASC-3, and CAP. Flow rate (Maximum)2 Type 1 RGW3 Type

7

It is your responsibility to read and understandthe contents of this manual prior to installation anduse of this equipment.

This manual contains the information you willneed to install, operate, and maintain theNANOpure, Series 851, ultrapure water systemmanufactured by Barnstead|ThermolyneCorporation.

The NANOpure is designed to produce Type IReagent Grade Water equal to or exceedingstandards established by ASTM, CAP, and NCCLS.

Careful attention to the following instructions willassure that the NANOpure runs properly andproduces water to specification.

Illustrated parts lists are found on pages 42-43.Take a few minutes to familiarize yourself with thehardware before installation.

The NANOpure deionization system can be usedon pretreated or high quality tap water. Somemunicipal tap water supplies contain a very highconcentration of suspended particulates, colloids,dissolved organic and inorganic materials thatshould be removed by pretreatment before thewater is processed by the NANOpure. If you plan touse tap water feed for your NANOpure, Barnsteadencourages the use of our water analysis service to

Installation

Figure 2 NANOpure water pathway diagram

WarningTo avoid personal injury, thisdevice is to be used withwater feeds only. Sanitizing/cleaning agents must beused in compliance withinstructions in this manual.Failure to comply with theabove could result inexplosion and personalinjury.

Do not use in the presenceof flammable or combustiblematerials; fire or explosionmay result. This devicecontains components whichmay ignite such materials.

Page 8: NANOpure Analytical Deionization System bioresearch2.pdfAuxiliary Valve - Type I Reagent Grade Water (RGW) per ASTM-D1193, NCCLS ASC-3, and CAP. Flow rate (Maximum)2 Type 1 RGW3 Type

8

WarningTo avoid electrical shock,use with a properly groundedelectrical outlet of correctvoltage and current handlingcapacity.

verify feedwater suitability. A sample collection kitmay be obtained by contacting our offices or yourpreferred laboratory supply dealer.

The NANOpure requires expendablepretreatment and deionization cartridges and FinalFilters which are not supplied with the unit (exceptfor one Auxiliary Valve 0.2 micron Final Filter) andmust be purchased separately. These expendablesare available as individual components or inexpendables kits (see Table 1 below).

Your NANOpure is supplied with a pre-wiredjumper in the "pump interlock" connector.Installation of options D0603 or D0606 (FloatSwitch) or D2706 (Pressure Switch) require removalof this jumper plug. DO NOT discard this plug; it willbe needed for certain maintenance operations.

Your NANOpure Analytical is also supplied withone FL703X2 Auxiliary Valve Final Filter. Thesefilters are not included with the cartridge listsdescribed below. They must be ordered separately.

All models are provided with a power cord andplug to be connected to a standard groundedelectrical outlet. Refer to TECHNICALSPECIFICATIONS and Figure 14, 15 & 16 in thisinstruction manual for the electrical requirements.

D4801 NANOpure 4-Cartridge Expendables Kit

1 each D0835 Pretreatment Cartridge

1 each D0803 High Capacity Cartridge

2 each D5027 Ultrapure SG Cartridge

2 each D3751 Remote Dispenser Filter

D4802 NANOpure 4-Cartridge Expendables Kit with

ORGANICfree

1 each D0836 Pretreament MACROpure

Cartridge

1 each D0803 High Capacity Cartridge

1 each D5027 Ultrapure SG Cartrdge

1 each D5021 ORGANICfree Cartridge

2 each D3751 Remote Dispenser Filter

D5025 NANOpure Pretreat Feed 4-Cartridge

Expendables Kit

1 each D0835 Pretreatment Cartridge

1 each D0809 Ultrapure SG Upflow Cartridge

2 each D5027 Ultrapure SG Cartridge

2 each D3751 Remote Dispenser Filter

D5026 NANOpure Pretreat Feed 4-Cartridge

Expendables Kit with ORGANICfree

1 each D0836 Pretreatment MACROpure

Cartridge

1 each D0809 Ultrapure SG Upflow Cartridge

1 each D5027 Ultrapure SG Cartridge

1 each D5021 ORGANICfree Cartridge

2 each D3751 Remote Dispenser Filter

Table 1 Expendable Kits

INSTALLATION

Page 9: NANOpure Analytical Deionization System bioresearch2.pdfAuxiliary Valve - Type I Reagent Grade Water (RGW) per ASTM-D1193, NCCLS ASC-3, and CAP. Flow rate (Maximum)2 Type 1 RGW3 Type

9

UnpackingUnpack the NANOpure carefully. Remove all

contents carefully. Inspect the packaging foradditional materials before discarding. Lift theNANOpure carefully from the box, holding ontoeither the mounting bracket or bench stand. Lay theunit down on its back. Remove the contents of theaccessory box. Locate the canisters and handlerings and set them aside. The remaining contentsof the accessory box are the power cord, theRemote Dispenser, tubing with fitting, o-rings, fillerpanel with bracket, 10 ft. interconnect cable and anAuxiliary Valve 0.2 micron Final Filter.Do notinstall any of these components until told to doso.

Choosing a SiteThe NANOpure system features remote controls

and a remote dispenser, allowing system to bemounted almost anywhere within the laboratory.Use the mounting bracket for wall mounted systems

INSTALLATION

Figure 3 NANOpure Inlet and Outlets

CautionLifting the NANOpure bycartridge canisters orallowing the NANOpure to siton the canisters may causedamage.

WarningTo avoid electrical shock andpersonal injury, do not mountthe NANOpure directly overequipment that requireselectrical service. Routinemaintenance of this unit mayinvolve water spillage andsubsequent electrical shockhazard if improperly located.

Page 10: NANOpure Analytical Deionization System bioresearch2.pdfAuxiliary Valve - Type I Reagent Grade Water (RGW) per ASTM-D1193, NCCLS ASC-3, and CAP. Flow rate (Maximum)2 Type 1 RGW3 Type

10

as a template to drill mounting holes. (TheNANOpure does not include screws and fastenersfor mounting.) Allow a minimum of 6 inches (150mm) clearance on all sides of the unit for servicingand 16 inches (410 mm) in front for top coverremoval.

Mounting and Utility Connections1. Remove the two screws securing the top

cover to the NANOpure. Remove thepump interlock plug (see Figure 3 forlocation). Slide the top cover off away fromthe NANOpure. Familiarize yourself withvarious components and install anyoptional equipment, i.e. inlet cell, pressureor float switch. See appropriate section forinstructions.

2. Remove the hold down bolt securing theNANOpure to the wall bracket. It is locatedon the center of the back wall inside theNANOpure cabinet. (Figure 3)

3. Disengage the NANOpure from the wallbracket and lay the NANOpure on its back.

4. Remove the packing material from the wallbracket.

5. Use the wall bracket as a template andlocate and drill the mounting holes in themounting surface. A minimum of fourfasteners will be required — two on the topand two on the bottom.

INSTALLATION

CautionDo not connect to electricalservice until instructed to doso.

WarningTo avoid personal injury, wallcomposition, condition andconstruction, as well asfastener type, must beconsidered when mountingthis unit. The mountingsurface and fastenersselected must be capable ofsupporting a minimum of 275lbs. Inadequate support and/or fasteners may result indamage to mounting surfaceand/or equipment. If you areunsure of mounting surfacecomposition, condition andconstruction or correctfasteners, consult yourbuilding maintenance groupor contractor.

NoteFor bench mounted models,steps 2 through 6 are notrequired – begin with step 7.

Page 11: NANOpure Analytical Deionization System bioresearch2.pdfAuxiliary Valve - Type I Reagent Grade Water (RGW) per ASTM-D1193, NCCLS ASC-3, and CAP. Flow rate (Maximum)2 Type 1 RGW3 Type

11

6. Hang the NANOpure unit on the bracketby sliding the mounting pins into thebracket grooves. Reinstall the hold downbolt.

7. Install the front cover by sliding it forward.Ensure the pins on the cover align with theholes on the forward cabinet section.Install the cover screws and the pumpinterlock plug removed in step 1. (Figure3).

8. Remove the Remote Dispenser frompackaging.

9. Remove the tubing connector from theends of the Remote Dispenser tubing.Retain for use as replacement parts. SeeFigure 4 for identification of the connectorcomponents.

10. Remove the tubing nut, grab ring, backupring and O-ring from the RemoteDispenser connection on the NANOpure.Save as replacement parts.

11. Connect the Remote Dispenser tubing toNANOpure at elbow connections on theNANOpure (Figure 3). It is not importantwhich nut on the dispenser is attached towhich connector on the NANOpure.

12. Connect 3/8" OD tubing (supplied withunit) to water service.

13. Locate the quick disconnect valve fittingon the inlet to the NANOpure.

14. Remove the inlet quick disconnect valveinsert by pressing the stainless steelthumb pad and removing the valve insert.

INSTALLATION

NoteThe Tubing ConnectorInstallation section (on nextpage) will aid you incompleting steps 9-12.

NoteTo disrupt the flow of waterto the NANOpure, press thestainless steel thumb pad.The insert with the valve iseasily removed. To resupplythe NANOpure water, pushthe insert into the couplingbody and listen for the clickthat indicates full closure.See Figure 13 foridentification.

Page 12: NANOpure Analytical Deionization System bioresearch2.pdfAuxiliary Valve - Type I Reagent Grade Water (RGW) per ASTM-D1193, NCCLS ASC-3, and CAP. Flow rate (Maximum)2 Type 1 RGW3 Type

12

15 Remove the tube nut from the valve insertand slide the nut over the tubing.

16. Wet the tubing with water and slip it overthe hose barb. Tighten the nut. Do notattach to the NANOpure until told to doso.

Tubing Connector Installation

1. Completely disassemble the fitting. Referto Figure 4 to familiarize yourself with thenames of the component parts.

2. Make sure the tubing is cut off reasonablysquare and that no plastic burrs or ridgesare present.

3. Place the grab ring and back-up ring in thehex nut in the order and orientation shownin Figure 4. Thread the nut onto theconnector. Do not use the O-ring at thistime.

4. Push the tubing through the nut until itbottoms out in the connector.

CautionDo not tighten tube fitting hexnut with a wrench. Tightconnections can be easilymade by hand.

Figure 4 Typical Polypropylene Tubing Connector Installation

INSTALLATION

Page 13: NANOpure Analytical Deionization System bioresearch2.pdfAuxiliary Valve - Type I Reagent Grade Water (RGW) per ASTM-D1193, NCCLS ASC-3, and CAP. Flow rate (Maximum)2 Type 1 RGW3 Type

13

5. Remove the hex nut and tubing. Place theO-ring over the tubing. Be careful not topush the back-up ring or grab ring furtherback on the tubing when installing the O-ring.

6. Install the hex nut on the connector andhand tighten.

Initial OperationInstall and rinse cartridges and filters as follows:

1. Clean the sealing surfaces of theheads and the top of the canisterswith a clean, wet cloth. Install the O-rings in the O-ring grooves of the 4canisters. Wet the O-rings with waterand install a D0835 or D0836Pretreament Cartridge, with the smallopening up, into a canister and installthis canister at position 1 (Figure 6).This is done by depressing the thumblever and rotating the canister handlering 1/4 turn to the right. Install theremaining canisters empty.

INSTALLATION

Secure locking pin beforeoperating. Pin must be fullyengaged before operating.

Caution

Ensure O-ring is in place inthe groove and wet O-ringprior to securing on thehead.

Caution

An extra set of head-to-canister O-rings aresupplied. These can be usedto replace any O-rings thatmay have been damaged ordeformed in shipment. Besure O-ring is in place beforereplacing canister beyondpoint where locking pin andpin hole line up.

Note

Figure 5 Canister Locking Pin Positioning

Page 14: NANOpure Analytical Deionization System bioresearch2.pdfAuxiliary Valve - Type I Reagent Grade Water (RGW) per ASTM-D1193, NCCLS ASC-3, and CAP. Flow rate (Maximum)2 Type 1 RGW3 Type

14

The locking pin on the canister handle ringmust be fully engaged into hole in the head beforesystem is operated (Figure 5 for proper positioning).

2. Install the electrical cord to the powerentry module on the right hand portionof the NANOpure (Figure 13 #31).Plug the power cord into a properlygrounded outlet of correct voltage andcurrent handling capacity.

3. Ensure that the system flush housingis in the Remote Dispenser. Opencustomr supplied inlet valve. Push theinsert of the inlet quick disconnectincluding the inlet tubing into thecoupling body to supply water to theNANOpure. Turn on power to unit andopen the Remote Dispenser valve.Allow water to run to drain for 10 to 15minutes through the RemoteDispenser.

4 Turn of the system power anddisconnect the inlet tubing bydepressing the stainless steel thumbpad on the inlet quick disconnectvalve and pulling gently.

5. With the Remote Dispenser open,also open the Auxiliary Valve todepressurize system prior toattempting to remove canisters.

6. Remove the canisters in positionstwo, three, and four and pour out thewater. Clean as required.

7. Install the remaining cartridges asshown in Figure 6 and Table 2.

INSTALLATION

CautionDo not operate the pump dry– dry running will damagethe pump.

WarningTo avoid personal injurydespressurize system priorto attempting to removecanisters.

Leave system flush housingin Remote Dispenser untilinstructed to do otherwise.

Note

Figure 6 Cartridge Sequence

Page 15: NANOpure Analytical Deionization System bioresearch2.pdfAuxiliary Valve - Type I Reagent Grade Water (RGW) per ASTM-D1193, NCCLS ASC-3, and CAP. Flow rate (Maximum)2 Type 1 RGW3 Type

15

INSTALLATION

The correct sequence ofcartridges is important inproducing the desired qualityof water.

Note

Standard Cartridge KitCatalog Number D4801

Poaition Type Catalog no.1. Pretreatment D08352. High capacity D08033. Ultrapure SG D50274. Ultrapure SG D5027

Pretreated Feed Standard Cartridge KitCatalog Number D5025

Position Type Catalog no.1. Pretreatment D08352. Ultrapure SG Upflow D08093. Ultrapure SG D50274. Ultrapure SG D5027

Cartridge Kit with ORGANICfreeCatalog Number D4802

Position Type Catalog no.1. MACROpure D08362. High capacity D08033. Ultrapure SG D50274. ORGANICfree D5021

Pretreated Feed Cartridge Kit withORGANICfree

Catalog Number D5026Position Type Catalog no.

1. MACROpure D08362. Ultrapure SG Upflow D08093. Ultrapure SG D50274. ORGANICfree D5021

Table 2 Correct Cartridge Sequence

Page 16: NANOpure Analytical Deionization System bioresearch2.pdfAuxiliary Valve - Type I Reagent Grade Water (RGW) per ASTM-D1193, NCCLS ASC-3, and CAP. Flow rate (Maximum)2 Type 1 RGW3 Type

16

8. Reconnect the inlet tubing valveinsert. Turn on power to unit and runthe system to drain through theRemote Dispenser for 10 to 15minutes.

9. Turn off the system power anddisconnect the inlet tubing. Allow thesystem to depressurize through theRemote Dispenser and by openingthe Auxiliary Valve.

10. Remove the system flush housingfrom the Remote Dispenser by gentlypushing the housing in the direction ofthe grooves in the Remote Dispenser(Figure 7). Ensure that the O-ringsremain in place.

11. Install a D3751, 0.2 Micron Final Filterinto the Remote Dispenser by slidingthe filter into the grooves on side ofthe Remote Dispenser (Figure 7).

12. Install a new 0.2 Micron Final Filter(FL703X2) by removing the filter fromthe package and placing it onto the

INSTALLATION

Always wet O-rings before0.2 Micron Final Filterinstallation.

Note

Figure 7 Remote Dispenser Final Filter Installation

0.2 Micron Final Filter/Flush Housing

O-rings

Page 17: NANOpure Analytical Deionization System bioresearch2.pdfAuxiliary Valve - Type I Reagent Grade Water (RGW) per ASTM-D1193, NCCLS ASC-3, and CAP. Flow rate (Maximum)2 Type 1 RGW3 Type

17

Luer connection on the AuxiliaryValve. Gently turn the 0.2 MicronFinal Filter clockwise until it is fullyseated on the Luer fitting.

13. Resupply the NANOpure with waterand turn the power on. Allowapproximately 8 liters of water to runto drain through both the RemoteDispenser and the Auxiliary Vavle torinse the D3751 0.2 Micron FinalFilter and the FL703X2 Final Filter.

14. Close the Remote Dispenser and theAuxiliary Valve and allow the systemto recirculate until the desired purity isachieved. The NANOpure system isnow ready for use.

INSTALLATION

Page 18: NANOpure Analytical Deionization System bioresearch2.pdfAuxiliary Valve - Type I Reagent Grade Water (RGW) per ASTM-D1193, NCCLS ASC-3, and CAP. Flow rate (Maximum)2 Type 1 RGW3 Type

18

For best results and optimum of cartridge life,leave the NANOpure in the standby modeduring periods of non-use and over night. Thestandby mode is designed to automaticallyrecirculate water through the entire NANOpuresystem (even the Remote Dispenser) for 10minutes of each hour of inactivity. When you will notbe using the NANOpure for an interval during theday, and at the end of the day, put your NANOpureinto Standby mode by pushing “On/Standby/Off”membrane switch on controls until display reads“SbY.” Ensure that adequate water supply isavailable to the NANOpure when the unit is instandby. When you are ready to use your NANOpureagain, press the "On/Standby/Off" membraneswitch twice. The display will show the currentresistivity of the water. Allow the system torecirculate until the desired purity is achieved, ifnecessary.

During normal operation, the NANOpure'selectronics are set in the resistivity mode. Themeter automatically compensates readings to 25°C.If the resistivity falls below the programmed setpoint (see Selecting the Set Point , page 20), thedisplay will flash numerals.

When the NANOpure display reads “Err,” it is anindication that there is air in the system or aproblem in the resistivity monitoring system. Pleaserefer to the troubleshooting section at the back ofthis manual for problem identification and solution.On initial startup, the unit will read “ERR” until waterhas passed through the system and out the RemoteDispenser.

Normal Operation

Page 19: NANOpure Analytical Deionization System bioresearch2.pdfAuxiliary Valve - Type I Reagent Grade Water (RGW) per ASTM-D1193, NCCLS ASC-3, and CAP. Flow rate (Maximum)2 Type 1 RGW3 Type

19

Modes of Operation

The NANOpure has three membrane switcheswhich control its functions.

• ModeAllows you to choose from resistivity, setpoint, or temperature.

• ControlAllows you to change set point values,check resistivity meter calibration, ormonitor the optional inlet cell.

• On/Standby/OffAllows you to turn unit on, off or operateunit in the standby mode.

Turning Unit On or Off• To turn unit on, press the “On/Standby/Off”

membrane switch once. The display willlight.

Figure 8 Display Indicators

Smart Purity Meter

Indicator Lights

Display

Membrane Switches

Page 20: NANOpure Analytical Deionization System bioresearch2.pdfAuxiliary Valve - Type I Reagent Grade Water (RGW) per ASTM-D1193, NCCLS ASC-3, and CAP. Flow rate (Maximum)2 Type 1 RGW3 Type

20

• To turn unit off, press “On/Standby/Off”membrane switch until the displaybecomes blank.

Selecting the Set PointThe NANOpure electronics include a userprogrammable set point which alerts you when thewater quality falls below the programmed set point.The set point is user selectable from 0.1 megohm-cm to 18.3 megohm-cm. The display will flashnumerals when actual resistivity measurement is ator below programmed set point.

1. Push the “Mode” membrane switch untilthe orange light (Figure 8) is lit alongsidethe set point indicator. The preset set pointin the electronics will appear on thedisplay.

2. To change the set point, press and holdthe “Control” membrane switch and thenpress the “Mode” membrane switch. Theset point display will automatically begin toscroll backward. To reverse the scrolling,release the “Control” membrane switchuntil the desired set point is attained.

3. When the desired set point is reached,release the “Mode” switch. The NANOpureelectronics will automatically retain the setpoint until you re-enter the set point modeand repeat the procedure.

SMART PURITY METER

Page 21: NANOpure Analytical Deionization System bioresearch2.pdfAuxiliary Valve - Type I Reagent Grade Water (RGW) per ASTM-D1193, NCCLS ASC-3, and CAP. Flow rate (Maximum)2 Type 1 RGW3 Type

21

Displaying the Temperature

To display the water temperature, simply pushthe “Mode” switch until the orange light on the lefthand side of the display is aligned with the “Temp°C” indicator. The display will now read the watertemperature in degrees centigrade.

Checking the Calibration

The NANOpure system is equipped with amanual electronics calibration check which allowsyou to verify meter accuracy.

1. Ensure that the reference/inlet cell toggleswitch inside the top housing (Figure 9) isin the reference position.

2. Ensure that the display is in the resistivitymode.

3. Depress and hold the control switch onuntil a reading is given.

If the electronics are within calibration, thereading will be between 9.0 and 11.0 megohm-cmfor resistivity. If the reading is not between thesevalues, call Barnstead|Thermolyne CustomerService for further assistance.

SMART PURITY METER

Page 22: NANOpure Analytical Deionization System bioresearch2.pdfAuxiliary Valve - Type I Reagent Grade Water (RGW) per ASTM-D1193, NCCLS ASC-3, and CAP. Flow rate (Maximum)2 Type 1 RGW3 Type

22

Inlet Cell Installation

The NANOpure system offers as an optionalfeature the ability to monitor the resistivity andtemperature of the NANOpure inlet water. To installthe inlet cell:

1. Disconnect the inlet tubing and disconnectelectrical service to unit. Open the RemoteDispenser to depressurize the unit.

2. Remove the top cover.

3. Remove the plug from the top of the inletcell well (Figure 9). Install the cell (part no.E550X1A) into the NANOpure inlet cellwell.

4. Thread all lead wires through theNANOpure system as shown in Figure 9.

5. Connect leads to the terminal (Figure 9)according to wiring scheme shown inTable 3.

6. Ensure that the reference/inlet cell toggleswitch is in the inlet cell position.

7. Reconnect the inlet tubing and check forleaks.

8. Reinstall the top cover and the powercord.

SMART PURITY METER

Figure 9 Inlet Cell Installation and Connections

Inlet cell well

Routing ofinlet cell leadwires

Inlet CellPC board

Terminalblock

Referenceinlet celltoggleswitch

E550X1A includes 1/2" NPSthreaded bushing and ano-ring.

Note

Cell lead Terminal block position

White Position 1

Red Position 2

Black Position 3

Table 3 Inlet Cell Wire Connections

Page 23: NANOpure Analytical Deionization System bioresearch2.pdfAuxiliary Valve - Type I Reagent Grade Water (RGW) per ASTM-D1193, NCCLS ASC-3, and CAP. Flow rate (Maximum)2 Type 1 RGW3 Type

23

Inlet Cell Operation

To monitor inlet cell readings, ensure reference/inlet cell toggle switch is in the inlet cell position,select the mode you desire (resistivity ortemperature) and depress and hold the “control”membrane switch. The reading shown on thedisplay will be for the inlet cell and will revert to theprimary cell when the switch is released.

SMART PURITY METER

Page 24: NANOpure Analytical Deionization System bioresearch2.pdfAuxiliary Valve - Type I Reagent Grade Water (RGW) per ASTM-D1193, NCCLS ASC-3, and CAP. Flow rate (Maximum)2 Type 1 RGW3 Type

24

The NANOpure includes as a standard featurethe ability to remotely mount the electronic controldisplay up to10 feet away by means of an umbilicalcord included with each unit.

1. Disconnect the inlet water line and theelectrical service, and remove the powercord from the NANOpure.

2. Remove the top cover of the unit byremoving the cover screws and the pumpinterlock cable or jumper plug and pull thetop cover straight out.

3. Locate the 10 ft. display interconnectcable. Remove the 6" cord from itsconnection points on the main circuitboard and the display unit. On the D4742and D4745 230V NANOpure, suppressormust first be removed from the cable.

4. Remove the display unit from themounting bracket by pulling the displaystraight up until the pins in back line upwith mounting holes, then pulling straightforward.

5. Attach the 10 foot cord to the connector onthe main circuit board and route theinterconnect cable through the NANOpureas shown on Figure 10. Attach suppressorremoved in Step 3 to 10 foot cable (on230V systems only).

6. A bracket is supplied to permanentlymount the display in a remote location.Attach the bracket at the desired locationusing user supplied screws. Attach thecable to the display. Attach the display tothe bracket by lining up the mounting pinson the display with the bracket grooves.

Remote Display Mounting

Page 25: NANOpure Analytical Deionization System bioresearch2.pdfAuxiliary Valve - Type I Reagent Grade Water (RGW) per ASTM-D1193, NCCLS ASC-3, and CAP. Flow rate (Maximum)2 Type 1 RGW3 Type

25

7. A filler panel is supplied to fill the void leftwhen the display is removed. Prior toattaching the filler panel, move themounting bracket forward by looseningand reattaching the mounting screws.

8. Remove the screws from the filler paneland reassemble through the slots on themounting bracket.

9. Reinstall the top cover and reattach thecover screws and the pump interlock cableor jumper plug.

10. Reconnect the inlet water line and thepower cord, and connect to electricalservice.

REMOTE DISPLAY MOUNTING

Do not mount display wherewater may spill or drip onelectronics.

Caution

Figure 10 Remote Display Mounting

Mounting bracket

Display

10 ft. displayinterconnectcable Filler panel

Page 26: NANOpure Analytical Deionization System bioresearch2.pdfAuxiliary Valve - Type I Reagent Grade Water (RGW) per ASTM-D1193, NCCLS ASC-3, and CAP. Flow rate (Maximum)2 Type 1 RGW3 Type

26

Operation

Each NANOpure system includes a RemoteDispenser unit. The Remote Dispenser unitincludes a replaceable 0.2 Micron Final Filter(packaged separately). The Remote Dispenser isshipped with an empty system flush housinginstalled. This housing is to be used duringcartridge rinsing procedures.

The dispenser incorporates an "easy to use"thumb wheel dispensing mechanism which isdesigned to deliver a steady stream when thethumb wheel is completely forward; drop by dropwhen it is slightly forward of the center; and spraywhen it is back. When the thumb wheel is in thecenter, no water will flow out of the dispenser.

When you are not using the Remote Dispenser,water automatically recirculates through the entirelength of the Remote Dispenser's feed tubing andthrough the 0.2 micron filter, ensuring that you havehigh purity water immediately available at yourfingertips. For applications requiring the highestpurity water, allow 500 ml of water to flow to drainfrom the Remote Dispenser before using theproduct water.

Installing the 0.2 Micron FinalFilter

1. Turn off the system power anddisconnect the inlet tubing. Allow thesystem to depressurize through theRemote Dispenser and by openingthe Auxiliary Valve.

Remote Dispenser

NoteDo not discard system flushhousing after initial cartridgerinsing. Keep for futurecartridge changes.

Page 27: NANOpure Analytical Deionization System bioresearch2.pdfAuxiliary Valve - Type I Reagent Grade Water (RGW) per ASTM-D1193, NCCLS ASC-3, and CAP. Flow rate (Maximum)2 Type 1 RGW3 Type

27

2. Remove the system flush housingfrom the Remote Dispenser by gentlypushing the housing in the direction ofthe grooves in the Remote Dispenser(Figure 7). Ensure that the O-ringsremain in place.

3. Install a D3751, 0.2 Micron Final Filterinto the Remote Dispenser by slidingthe filter into the grooves on side ofthe Remote Dispenser (Figure 7).

4. Resupply the NANOpure with waterand turn the power on. Allowapproximately 8 liters of water to runto drain through the RemoteDispenser to rinse the D3751, 0.2Micron Final Filter.

5. Close the Remote Dispenser andallow the system to recirculate untilthe desired purity is achieved. TheNANOpure system is now ready foruse.

6. Periodically replace the RemoteDispenser filter.

REMOTE DISPENSER

NoteTo prevent leakage at filterconnection, do not allowO-rings to become deformedduring installation. If O-ringsare damaged, replace.

Page 28: NANOpure Analytical Deionization System bioresearch2.pdfAuxiliary Valve - Type I Reagent Grade Water (RGW) per ASTM-D1193, NCCLS ASC-3, and CAP. Flow rate (Maximum)2 Type 1 RGW3 Type

28

Accessories D0603, D0606 (float switches) andD2706 (pressure switch) are designed to protectthe NANOpure pump by alerting the NANOpure ofan inadequate feedwater condition so that thepump can be shut down. Use the followinginstructions for installation.

1. Disconnect electrical service.

2. If using the D0603 or D0606 float switch,follow installation instructions included withunit for installation to tank.

3. If using the D2706 low pressure switch,install the PVC tee (supplied with D2706)into the incoming water line (see Figure11). Screw the switch into the top of thetee, then connect the inlet tubing toNANOpure with the remaining opening.

4. Route the cable from float or low pressureswitch either above or below NANOpurecover as shown in Figure 11.

5. Remove the jumper plug and save forfuture use.

6. Plug the cable into jumper plug outlet.

7. Reconnect the electrical service.

Float Switch/Pressure SwitchInstallation

Figure 11 Pump Protector/Pressure Switch Installation

Page 29: NANOpure Analytical Deionization System bioresearch2.pdfAuxiliary Valve - Type I Reagent Grade Water (RGW) per ASTM-D1193, NCCLS ASC-3, and CAP. Flow rate (Maximum)2 Type 1 RGW3 Type

29

Replacing Cartridges

When the resistivity of the water drops below thedesired level, change all of the cartridges together.

1. Turn off system power and disconnectinlet tubing by depressing the stainlesssteel thumb pad on the inlet quickdisconnect valve and pulling gently.

2. Open the Remote Dispenser and AuxiliaryValve to depressurize system prior toattempting to remove canisters.

3. Place a container under the cartridgecanisters to collect any spillage.

4. Carefully remove the canisters from theheads by depressing thumb lever androtating the canisters from right to left 1/4turn. Pull the cartridges from heads anddiscard, drain and clean the canisters.

5. Install the new cartridges as explained inINITIAL OPERATION section of thismanual.

Replacing Remote DispenserFilter

It is recommended that the Remote Dispenser0.2 micron Final Filter be replaced every 45 days,when there is an unacceptable bacteria passage orwhen flow decreases to less than one liter perminute.

To replace the filter follow the instructions in the0.2 MICRON FINAL FILTER INSTALLATIONsection of this manual.

Maintenance and Servicing

WarningTo avoid electrical shock,always disconnect frompower supply beforemaintenance and servicing.Refer servicing to qualifiedpersonnel.

WarningTo avoid personal injury,depressurize system prior toattempting to removecanisters.

NoteDo not tighten canisterbeyond point where lockingpin and pin hole line up.

Page 30: NANOpure Analytical Deionization System bioresearch2.pdfAuxiliary Valve - Type I Reagent Grade Water (RGW) per ASTM-D1193, NCCLS ASC-3, and CAP. Flow rate (Maximum)2 Type 1 RGW3 Type

30

Figure 12 Auxiliary Valve 0.2 Micron Filter Installation

Always run at least 6-10 liters of deionized waterthrough a new 0.2 Micron Final Filter before usingwater.

Replacing the Auxiliary Valve 0.2Micron Final Filter

Replace the Auxiliary Valve 0.2 Micron FinalFilter whenever any of the following conditionsoccur: every 30 days, the product water flow isreduced or bacteria and/or pyrogen break through.The Auxiliary Valve 0.2 Micron Final Filter isshipped assembled with a bell. To replace theAuxiliary Valve 0.2 micron filter assembly:

1. Remove the old Auxiliary Valve 0.2 micronfilter assembly by turning itcounterclockwise until it is free from theLuer fitting.

2. Remove the new Auxiliary Valve 0.2Micron Final Filter assembly from its bagand insert it into the Luer fitting. Gentlyturn it clockwise until it is fully seated inthe Luer fitting.

3. Rinse at least 5 liters of water through theAuxiliary Valve 0.2 Micron Final Filter todrain prior to using the product water.

NoteUse care when handling thenew Auxiliary Valve 0.2Micron Final Filter assemblyto prevent environmentalcontamination of filter.

MAINTENANCE AND SERVICING

NoteIf a newly installed 0.2Micron Filter clogs rapidlyafter installation, theNANOpure Analytical mayneed to be sanitized toremove bacterialcontaminants. See SystemSanitization.

CautionDo not overtighten theAuxiliary Valve 0.2 MicronFinal Filter assembly ontothe Luer fitting or useexcessive force in seating it.The filter and/or Luer fittingcan be damaged byovertightening or excessiveforce.

Page 31: NANOpure Analytical Deionization System bioresearch2.pdfAuxiliary Valve - Type I Reagent Grade Water (RGW) per ASTM-D1193, NCCLS ASC-3, and CAP. Flow rate (Maximum)2 Type 1 RGW3 Type

31

Replacing Fuses

The “main fuse” is located in a fuse holder at theupper right side of the top housing. To access,remove the top cover, squeeze the clip on the fuseholder and remove. Replace with a slow blow 2ampere fuse for 230 VAC units and a slow blow 3ampere fuse for 100 or 115 VAC units. Toreassemble, simply insert the fuse in the fuseholder and push the fuse holder until it snaps intoplace.

System Sanitization

Whenever cartridges are replaced, the systemshould be sanitized. The system should also besanitized if the 0.2 micron filters clog rapidly afterinstallation. Sanitize the system by using the new,easy-to-use sanitization cartridge (#D50223).

Sanitize your NANOpure Analytical as follows:

1. Turn the system off and disconnect thepower.

2. Remove inlet tubing. Open the RemoteDispenser and Auxiliary Valve todepressurize the system prior toattempting to remove canisters.

3. Remove the canisters by depressing thethumb lever and rotating 1/4 turn to theleft. Discard used cartridges. With thecartridges removed from the canisters,wash the inside of the canisters and theinside heads with soap or detergent, usinga sponge or clean cloth. Rinse out thecanisters and the heads with clean waterseveral times to remove the detergentresidues.

MAINTENANCE AND SERVICING

WarningReplace fuses only with thesame type and rating forcontinued protection againstpossible fire hazard andprotection of the equipment.

WarningDepressurize system prior toattempting to removecanisters.

Page 32: NANOpure Analytical Deionization System bioresearch2.pdfAuxiliary Valve - Type I Reagent Grade Water (RGW) per ASTM-D1193, NCCLS ASC-3, and CAP. Flow rate (Maximum)2 Type 1 RGW3 Type

32

Warning• Avoid splashing disinfecting

solutions on clothing or skin.

• Ensure all piping connectionsare tight to avoid chemicalleakage.

• Always depressurize chemicallines before disassembly.

• Ensure adequate ventiliation.

• Carefully follow manufacturer'ssafety instructions on labels ofchemical containers andmaterial safety data sheets.

MAINTENANCE AND SERVICING

4. Remove the 0.2 Micron Final Filter fromthe Remote Dispenser and replace withthe system flush housing. Remove theAuxiliary Valve 0.2 Final Filter. Do notattempt to sanitize the 0.2 Micron FinalFilters with chemical solutions.

5. Place the sanitizating cartridge containingthe chlorine pellet onto the head ofposition number one.

6. Install the four canisters onto the heads bydepressing the thumb lever and rotatingthe handle ring 1/4 turn to the right.

7. Place a suitable container under theAuxiliary Valve to catch the sanitizatingsolution. Reconnect the inlet tubing andopen the Auxiliary Valve and RemoteDispenser. Turn on the water supply andpower to the unit.

8. When the sanitizating solution begins toexit the Remote Dispenser and AuxiliaryValve, close the Auxiliary Valve.

9. Let the solution exit the Remote Dispenserfor thirty seconds in the stream pattern.After thirty seconds, change to the spraypattern and allow the unit to spray thesolution for an additional thirty seconds.This allows for the proper sanitization ofthe inside and outside tip and nozzle of theRemote Dispenser.

10. Close the Remote Dispenser.

11. Allow the unit to remain in recirculation for30-45 minutes.

Page 33: NANOpure Analytical Deionization System bioresearch2.pdfAuxiliary Valve - Type I Reagent Grade Water (RGW) per ASTM-D1193, NCCLS ASC-3, and CAP. Flow rate (Maximum)2 Type 1 RGW3 Type

33

12. After 30-45 minutes, open the AuxiliaryValve and the Remote Dispenser. Allowthe sanitizing solution to exit the unit.Leave these valves open forapproximately 5 minutes.

13. Turn the system off and disconnect thepower. Remove the inlet tubing. Open theRemote Dispenser and Auxiliary Valve todepressurize the system prior toattempting to remove canisters.

14. Carefully remove all canisters from thesystem and discard the remaining solution.DO NOT RINSE THE CANISTERS.

15. Install fresh cartridges in the system asindicated in the REPLACINGCARTRIDGES section of this manual. Donot reinstall used cartridges (they maycontain large amounts of bacteria).

16. Reconnect the feedwater line to thefeedwater source, and reconnect thepump protector or pressure switch to thereceptacle in the top housing. Save thejumper plug for future use.

17. Connect the power to the unit, and pressthe control panel “on/off” button to start thepump and fill the system. Run waterthrough the system to drain any remainingdisinfecting solution. A flush of 10 liters issufficient.

18. Close the Remote Dispenser valve, andallow the resistivity of the water to riseabove the “set point” setting on theresistivity meter. Install and a new RemoteDispenser 0.2 Micron Final Filter asindicated under REPLACING THEREMOTE DISPENSER FILTER" section ofthis manual.

MAINTENANCE AND SERVICING

NoteThe pellet in the sanitizationcartridge may not becompetely dissolved. Thisdoes not constitute anincomplete sanitization. Donot reuse a spentsanitization cartridge.

WarningDepressurize system prior toattempting to removecanisters.

CautionSecure locking pin beforeoperating. Pin must be fullyengaged before operating.

Page 34: NANOpure Analytical Deionization System bioresearch2.pdfAuxiliary Valve - Type I Reagent Grade Water (RGW) per ASTM-D1193, NCCLS ASC-3, and CAP. Flow rate (Maximum)2 Type 1 RGW3 Type

34

19. Install a new Auxiliary Valve 0.2 micronFinal Filter as indicated in theREPLACING THE AUXILIARY VALVE 0.2MICRON FINAL FILTER section of thismanual.

Cleaning the Resistivity Cell1. Disconnect power to the unit. Disconnect

the inlet tubing and depressurize thesystem.

2. Remove the top cover screws and slidetop cover toward you to remove it.

3. Remove the front housing by pushing upon the clip located at the bottom of thehousing and pulling the housing straightforward.

4. For Models D4741, D4743, D4744 andD4746, carefully remove the ResistivityMonitor circuit board. The D4742 andD4745 do not require the Monitor to beremoved.

5. Disconnect three cell leads at the printedcircuit board connector and gently pull thecable down through the grommet towardthe cell.

6. Unscrew and remove the cell. Carefullyremove O-ring to clean the cell.

7. Wash the cell in a mild detergent solutionor a 10% inorganic acid solution (followacid manufacturers recommendedhandling procedure). This may be done inan ultrasonic cleaner or with a soft brush.

MAINTENANCE AND SERVICING

CautionDo not immerse the entirecell assembly in cleaningsolution, only the electrodeportion.

The cell electrodes areetched to improve wettingcharacteristics. Do notmechanically abrade ordamage this surface.

Caution

Page 35: NANOpure Analytical Deionization System bioresearch2.pdfAuxiliary Valve - Type I Reagent Grade Water (RGW) per ASTM-D1193, NCCLS ASC-3, and CAP. Flow rate (Maximum)2 Type 1 RGW3 Type

35

8. Thoroughly rinse the cell in deionized ordistilled water following the detergent oracid cleaning.

9. After cleaning, check the o-ring; replace ifnecessary.

10. Reinstall the cell into the cell well andhand tighten.

11. Reroute the cable up through the housingand reconnect the leads. Refer to thewiring diagram for proper lead terminalposition.

12. Replace the Resistivity Monitor circuitboard and front housing. Replace the topcover.

Shutdown

If NANOpure is to be shut down for an extendedperiod of time, the system should be completelydrained and the cartridges removed to prevent thegrowth of bacteria.

If the system has remained inactive and full ofwater, then the system should be drained, sanitizedand new cartridges and final filters installed prior touse.

MAINTENANCE AND SERVICING

Page 36: NANOpure Analytical Deionization System bioresearch2.pdfAuxiliary Valve - Type I Reagent Grade Water (RGW) per ASTM-D1193, NCCLS ASC-3, and CAP. Flow rate (Maximum)2 Type 1 RGW3 Type

36

Problem Solving Guide

Problem Possible Cause SolutionsNANOpure completely No electrical power to Ensure that the NANOpure powerinactive (pump not operating, NANOpure. cord is connected to a live powercontrol panel not lit, etc.) source and completely plugged into

electrical outlet.

Main fuse blown. Replace the main fuse as indicated inthe REPLACING FUSES section.

Printed circuit fuse blown. Replace the printed circuit fuse asindicated in the REPLACING FUSESsection.

Pump runs, but no display Printed circuit board inter- Reconnect cable.(no digital display). connect cable disconnected.

Reduced or no product flow 0.2 micron final filter clogged. Replace the Final Filter asfrom the Remote Dispenser. indicated in the REPLACING THE

REMOTE DISPENSER0.2 MICRON FINAL FILTER section.

Leaking canisters. Large o-ring in canister is Replace or position correctly.missing, damaged or notsealed properly.

Loose Head/Canister fit. Tighten handle ring.

Pump does not run. Display Pump protector (in reservoir), Connect the pump protector orlight. feedwater line pressure switch pressure switch cord to the receptacle

or jumper plug not connected located inside top cover. If a Barnsteadto pump interlock. pressure switch is installed in the

feedwater line, the pump will not startuntil the line pressure risesto 0.35 kg/cm

2 (5 psi).

Open the feedwater line shut off valveor fill the feedwater reservoir.

Make sure the jumper plug isinstalled.

Page 37: NANOpure Analytical Deionization System bioresearch2.pdfAuxiliary Valve - Type I Reagent Grade Water (RGW) per ASTM-D1193, NCCLS ASC-3, and CAP. Flow rate (Maximum)2 Type 1 RGW3 Type

37

Problem Possible Cause SolutionsRecirculated water will not Exhausted cartridge. Replace all the cartridges asrinse up to desired purity level. indicated in the REPLACING

CARTRIDGES section (page 21).

Cartridges out of order. Install the cartridges in the properorder as indicated in the INITIALOPERATION section.

Cartridges upside down. Install the cartridges right side upas indicated in the INITIALOPERATION section.

Feed water bypassing Be sure that small O-ring insidecartridge(s). head is not damaged

and is properly installed.

Check valve malfunctioning. Remove tubing from checkvalve; turn unit on. If water leaksfrom end of check valve, washto remove any particulates.

Display reads "Err" when Resistivity cell disconnected Check resistivity cell wiring.checking resistivity. or wired improperly.

Air in system. Purge air from system byopening Remote Dispenserand/or Ultrafiltered WaterDrawoff.

System electronics or cell out Check resistivity of reference cell.of calibration. If resistivity displayed is not

between 9.7 and 10.3 electronicsneed recalibration. If resistivity readingis proper, clean cell and reinstall. Ifproblem persists, replace cell.

Display reads E_A. System electronics failure. Replace main PC board.

Display locks up in Standby Inlet power of poor quality. Disconnect power and thenmode. reconnect.

PROBLEM SOLVING GUIDE

Page 38: NANOpure Analytical Deionization System bioresearch2.pdfAuxiliary Valve - Type I Reagent Grade Water (RGW) per ASTM-D1193, NCCLS ASC-3, and CAP. Flow rate (Maximum)2 Type 1 RGW3 Type

38

Problem Possible Cause SolutionsShort cartridge life. Cartridges being used are Check the expiration date.

beyond expiration date. Cartridges begin to lose capacityafter being stored two years from the

date of manufacture. Replace thecartridges with unexpired ones.

Change in feedwater If a Barnstead ROpure is the feedwatercharacteristics. source, check that the membrane is

functioning properly.

If a Barnstead Still is the feedwatersource, ensure that the distillatetemperature to the NANOpure doesnot exceed 49°C (120°F).

If tap water is the feedwater source,check the quality of the water. In somecases the quality of the water willchange with the seasons. Changing thesource (city water to well water or wellwater to city water) will result in a waterquality change.

If feedwater is from a central waterpurification system, verify waterquality and proper functioning of thesystem.

PROBLEM SOLVING GUIDE

Page 39: NANOpure Analytical Deionization System bioresearch2.pdfAuxiliary Valve - Type I Reagent Grade Water (RGW) per ASTM-D1193, NCCLS ASC-3, and CAP. Flow rate (Maximum)2 Type 1 RGW3 Type

39

Recommended Spares

Consumables

Consumable parts are those required to support the day-to-day operation of thisequipment.

Barnstead |Thermolyne establishes two types of consumables; those items that mustperiodically be replaced to maintain performance (filters, resin cartridges, etc.) and otheritems of limited life (indicator lights, fuses, etc.) that you can expect to replace on a more orless random basis. Where practical, Barnstead |Thermolyne recommends the frequency ofreplacement, or provides information on life expectancy from which you may calculate areplacement interval compatible with your usage pattern.

The replacement of consumable parts is discussed in the Maintenance and Servicingsection of this manual to assist you in accomplishing your own service.

Consumables may be ordered separately and in some cases, as an expendables kit.Check with your Barnstead/Thermolyne representative for additional information on theexpendables kit.

Description Catalog No. Recommended QuantityRemote Dispenser Final Filter D3751 (package of 2) 2Auxiliary Valve 0.2 micron Final Filter FL703X2 (1/package) 2Pretreatment Cartridge D0835 or D0836 1High Capacity DI Cartridge D0803 1Ultrapure SG Cartridge D5027 2Ultrapure SG Upflow Cartridge D0809 1OrganicFree Cartridge D5021 13.0 Ampere SlowBlow Fuse, 120 Volt Models 04455 12.0 Ampere Slow Blow Fuse, 230 Volt Models 04420 1Teflon® Tape, Roll 06078 1Main Printed Circuit Board Fuse, 100 & 120 Volt Models FZX43 1Main Printed Circuit Board Fuse, 240 Volt Models FZX48 1O-ring (between heads) 06440 2O-ring (head-to-canister) GSX28 4O-ring (head-to-canister) GSX27 4Sanitization Cartridge D50223 1

Replacement Parts Listing

Page 40: NANOpure Analytical Deionization System bioresearch2.pdfAuxiliary Valve - Type I Reagent Grade Water (RGW) per ASTM-D1193, NCCLS ASC-3, and CAP. Flow rate (Maximum)2 Type 1 RGW3 Type

40

General Maintenance PartsGeneral maintenance parts are defined as laboratory level repair parts which do not

require great expertise or special tools for installation. Barnstead |Thermolyne recommendsthat you stock the general maintenance parts as an aid to ensuring the continued operationof this equipment.

Description Catalog No. Recommended QuantityFastener Pin FP550X1 2Connector (head to head) BR550X4 1Adapter (head end) BR550X2 1Quick Disconnect Body 1/4 NPT CUX8 1Quick Disconnect Insert 3/8 Tube CUX9 1Check Valve 02214 1

1/4" O.D. x 1/4" NPT Connector 05931 1

Safety StockFor critical applications where performance with minimum downtime is required,

Barnstead |Thermolyne recommends that you maintain a local stock of those parts listed inthe General Maintenance parts and Safety Stock sections.

Description Catalog No. Recommended QuantityResistivity Display SW550X1A 1Main PC Board 100 & 120 VAC Models PC550X1A 1Main PC Board 240 VAC Models PC747X2A 1Recirculation Pump & Motor 100 & 120 VAC Models PU687X1A 1Recirculation Pump & Motor 240 VAC Models PU733X2A 1Cartridge Canister Head BK550X2 1Cartridge Canister Handle Ring HN550X1A 1Cartridge Canister CS550X1 1Pressure Regulator 02280 1Resistivity Cell E550X1A 1

REPLACEMENT PARTS LISTING

Page 41: NANOpure Analytical Deionization System bioresearch2.pdfAuxiliary Valve - Type I Reagent Grade Water (RGW) per ASTM-D1193, NCCLS ASC-3, and CAP. Flow rate (Maximum)2 Type 1 RGW3 Type

41

Ordering Procedures

Please refer to the Specification Plate for thecomplete model number, serial number, and seriesnumber when requesting service, replacementparts or in any correspondence concerning thisunit.

All parts listed herein may be ordered from theBarnstead|Thermolyne dealer from whom youpurchased this unit or can be obtained promptlyfrom the factory. When service or replacementparts are needed we ask that you check first withyour dealer. If the dealer cannot handle yourrequest, then contact our Customer ServiceDepartment at 319-556-2241 or 800-553-0039.

Prior to returning any materials toBarnstead|Thermolyne Corp., please contact ourCustomer Service Department for a “Return GoodsAuthorization” number (RGA). Material returnedwithout a RGA number will be refused.

Page 42: NANOpure Analytical Deionization System bioresearch2.pdfAuxiliary Valve - Type I Reagent Grade Water (RGW) per ASTM-D1193, NCCLS ASC-3, and CAP. Flow rate (Maximum)2 Type 1 RGW3 Type

42

Exploded View100 & 120 Volt Models

Figure 13 Exploded View

Key Part # Description1 DL687X1A Top cover2 05766 1/4" OD x 1/4 NPT Elbow3 PU687X1A Pump assembly, 100/120V Models4 CRX72 Cord set 100/120V Models5 04247 Pump interlock plug6 CUX8 Quick Disconnect7 CUX9 Quick Disconnect Insert8 PC550X3A PC board, reference call9 SW550X1A Display10 WH550X2 Interconnect cable 10' (WH550X1 6")11 PC550X1A Main PC board, 100/120V Models12 02280 Pressure reducing valve13 02214 Stainless steel check valve14 BR550X2 Head end fitting15 06440 O-ring between heads16 FP550X1 Head connector pins17 BR550X4 Head connector18 BK550X2 Head19 GSX27 Head o-ring20 GSX28 Head to canister o-ring21 HN550X1A Canister handle ring22 CS550X1 Canister23 CV851X1A Auxiliary Valve24 DL630X18 Front Housing25 E550X1A Resistivity Cell26 CS895X1 Remote Dispenser28 FL563X2 System Flush Housing29 GSX29 O-Ring30 D4740 Floor Stand31 CEX172 Power Entry Module32 CAX44 Line Filter33 PM703X3 Luer Connector34 FS550X1 1/4" NPT Plug

Not ShownTN687X1A Transformer 100 VoltFL703X2 Filter

Page 43: NANOpure Analytical Deionization System bioresearch2.pdfAuxiliary Valve - Type I Reagent Grade Water (RGW) per ASTM-D1193, NCCLS ASC-3, and CAP. Flow rate (Maximum)2 Type 1 RGW3 Type

43

Exploded View240 Volt Models

Key Part # Description1 DL687X1A Top cover2 05766 1/4" OD x 1/4 NPT Elbow3 PU733X2A Pump assembly, 240V Models4 CRX70 Cord set 240V Models5 04247 Pump interlock plug6 CUX8 Quick Disconnect7 CUX9 Quick Disconnect Insert8 PC550X3A PC board, reference call9 SW550X1A Display10 WH550X2 Interconnect cable 10' (WH550X1 6")11 PC747X2A Main PC board, 240V Models12 02280 Pressure reducing valve13 02214 Stainless steel check valve14 BR550X2 Head end fitting15 06440 O-ring between heads16 FP550X1 Head connector pins17 BR550X4 Head connector18 BK550X2 Head19 GSX27 Head o-ring20 GSX28 Head to canister o-ring21 HN550X1A Canister handle ring22 CS550X1 Canister23 CV851X1A Auxiliary Valve24 DL630X18 Front housing25 E550X1A Resistivity cell26 CS895X3 Remote Dispenser28 FL563X2 System flush housing29 GSX29 O-ring30 D4740 Floor stand31 CEX172 Power Entry Module32 CAX44 Line Filter33 PM703X3 Luer Connector34 FS550X1 1/4" NPT Plug

Not ShownFL703X2 Filter

Figure 13 Exploded View (continued)

Page 44: NANOpure Analytical Deionization System bioresearch2.pdfAuxiliary Valve - Type I Reagent Grade Water (RGW) per ASTM-D1193, NCCLS ASC-3, and CAP. Flow rate (Maximum)2 Type 1 RGW3 Type

44

Wiring Diagrams

Figure 14 Wiring Diagram for 100V Model

Page 45: NANOpure Analytical Deionization System bioresearch2.pdfAuxiliary Valve - Type I Reagent Grade Water (RGW) per ASTM-D1193, NCCLS ASC-3, and CAP. Flow rate (Maximum)2 Type 1 RGW3 Type

45Figure 15 Wiring Diagram for 120V Model

WIRING DIAGRAMS

Page 46: NANOpure Analytical Deionization System bioresearch2.pdfAuxiliary Valve - Type I Reagent Grade Water (RGW) per ASTM-D1193, NCCLS ASC-3, and CAP. Flow rate (Maximum)2 Type 1 RGW3 Type

46

WIRING DIAGRAMS

Figure 16 Wiring Diagram for 230V Model

Page 47: NANOpure Analytical Deionization System bioresearch2.pdfAuxiliary Valve - Type I Reagent Grade Water (RGW) per ASTM-D1193, NCCLS ASC-3, and CAP. Flow rate (Maximum)2 Type 1 RGW3 Type

47

Barnstead|Thermolyne Corporation warrants that if a product manufactured byBarnstead|Thermolyne and sold by it within the continental United States or Canadaproves to be defective in material or construction, it will provide you, without charge, for aperiod of ninety (90) days, the labor, and a period of one (1) year, the parts, necessary toremedy any such defect. Outside the continental United States and Canada, the warrantyprovides, for one (1) year, the parts necessary to remedy any such defect. The warrantyperiod shall commence either six (6) months following the date the product is sold byBarnstead|Thermolyne or on the date it is purchased by the original retail consumer,whichever date occurs first.

All warranty inspections and repairs must be performed by and parts obtained froman authorized Barnstead|Thermolyne dealer or Barnstead|Thermolyne (at its owndiscretion). Heating elements, however, because of their susceptibility to overheating andcontamination, must be returned to our factory, and if, upon inspection, it is concluded thatfailure is not due to excessive high temperature or contamination, warranty replacement willbe provided by Barnstead|Thermolyne . The name of the authorizedBarnstead|Thermolyne dealer nearest you may be obtained by calling 1-800-446-6060 orwriting to:

Barnstead|ThermolyneP.O. Box 797

2555 Kerper BoulevardDubuque, IA 52004-0797

USAFAX: (319) 589-0516

E-MAIL ADDRESS: [email protected]|Thermolyne’s sole obligation with respect to its product shall be to repair

or (at its own discretion) replace the product. Under no circumstances shall it be liable forincidental or consequential damage.

THE WARRANTY STATED HEREIN IS THE SOLE WARRANTY APPLICABLE TOBarnstead|Thermolyne PRODUCTS. Barnstead|Thermolyne EXPRESSLY DISCLAIMSANY AND ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDINGWARRANTIES OF MERCHANTABILITY OR FITNESS FOR USE.

One Year Limited Warranty

Page 48: NANOpure Analytical Deionization System bioresearch2.pdfAuxiliary Valve - Type I Reagent Grade Water (RGW) per ASTM-D1193, NCCLS ASC-3, and CAP. Flow rate (Maximum)2 Type 1 RGW3 Type

48

Barnstead Thermolyne2555 Kerper Blvd.P.O. Box 797Dubuque, IA 52004-0797 USAPHONE: 319-556-2241 • 800-553-0039FAX: 319-589-0516E-Mail: [email protected]

a subsidiary of