Technical Information TI 144C/07/en No. 50077402 Areas of application The analytical transmitter Mycom CLM 152 is intended for highly reliable and accurate conductivity or resistance measurement in all areas of process control and engineering. Modular instrument design and compatibility with a variety of measuring cells permit optimal adaptation to any measuring and control task. Chemical industry Foodstuffs industry Power plants Pharmaceutical industry Water conditioning Also suitable for Ex areas Benefits at a glance High reliability is assured by: – comprehensive self-monitoring functions – polarisation compensation – log function – calibration history Adaptability due to modular design: – conductive conductivity measurement, resistance measurement or inductive conductivity measurement – one-circuit or difference measurement – limit functions and output contacts according to NAMUR Special features for ultrapure water: – ultrapure water temperature compensation – measurement and monitoring according to USP 23 Concentration measurement Remote measuring range switching EEx em [ia/ib] IIC T4 approval PROFIBUS-PA interface Hauser + Endress Nothing beats know-how Conductivity Measurement mycom CLM 152 Conductivity transmitter for conductive or inductive systems with limit contactor functions

Transcript of mycom CLM 152 · 2010-05-03 · Technical Information TI 144C/07/en No. 50077402 Areas of...

Page 1: mycom CLM 152 · 2010-05-03 · Technical Information TI 144C/07/en No. 50077402 Areas of application The analytical transmitter Mycom CLM 152 is intended for highly reliable and

TechnicalInformationTI 144C/07/enNo. 50077402

Areas of applicationThe analytical transmitter MycomCLM 152 is intended for highly reliableand accurate conductivity or resistancemeasurement in all areas of processcontrol and engineering.Modular instrument design andcompatibility with a variety of measuringcells permit optimal adaptation to anymeasuring and control task.• Chemical industry• Foodstuffs industry• Power plants• Pharmaceutical industry• Water conditioning• Also suitable for Ex areas

Benefits at a glance• High reliability is assured by:

– comprehensive self-monitoringfunctions

– polarisation compensation– log function– calibration history

• Adaptability due to modular design:– conductive conductivity

measurement, resistancemeasurement or inductiveconductivity measurement

– one-circuit or differencemeasurement

– limit functions and output contactsaccording to NAMUR

• Special features for ultrapure water:– ultrapure water temperature

compensation– measurement and monitoring

according to USP 23• Concentration measurement• Remote measuring range switching• EEx em [ia/ib] IIC T4 approval• PROFIBUS-PA interface

Hauser+EndressNothing beats know-how

Conductivity Measurementmycom CLM 152

Conductivity transmitterfor conductive or inductive systemswith limit contactor functions

Page 2: mycom CLM 152 · 2010-05-03 · Technical Information TI 144C/07/en No. 50077402 Areas of application The analytical transmitter Mycom CLM 152 is intended for highly reliable and

Measuring and controlsystem

A measuring system generallycomprises:• a conductive two-electrode measuring

cell or an inductive measuring cellwith integrated Pt 100 temperaturesensor

• a welded fitting or an assembly forinstallation in a pipeline or tank

• the corresponding measuring cableand

• the Mycom CLM 152 transmitter.

The correct measuring cell for every measuring rangeConductivitymeasuring cells

The hatched areas designate themeasuring range expansion achievedwith polarisation compensation.

For further information on the measuringcells refer to the correspondingTechnical Information brochures.

PARAM

MEAS CAL

DIAG PARAM

MEAS CAL

DIAG

CLS 15

CLS 21

CLS 13

Conductivemeasuring cell

Measuring cablee.g. CYK 71

MycomCLM 152-xxxxxx0

conductive

MycomCLM 152-xxxxxx5

inductive

CLS 50

Available as:• one-circuit or difference instrument• Ex version• PROFIBUS version

Available as:• one-circuit or difference instrument• Ex version• PROFIBUS version

Inductivemeasuring cell with

fixed cable

LM152SYS.CDRExamples of possiblemeasuring systems

0.01 0.1

µS/cm

1.0 10 100 1 10

mS/cm

100 1000 2000

k = 1.98 cm–1

k = 5.9 cm–1

CLS 21

k = 10 cm–1

k = 1 cm–1

k = 0.1 cm–1

k = 0.1 cm–1

k = 0.01 cm–1

k = 0.01 cm–1

CLS 30

CLS 15

CLS 15

CLS 50

CLS 12/13/19

CLS 12/13/19

CLS 52

10 1.0 100 10 1 0.1 0.01 0.001100

k cmΩ⋅M cmΩ⋅

k = 0.01 cm–1CLS 12/13/15/19

k = 0.1 cm–1CLS 12/13/15/19

k = 1 cm–1CLS 21

LF-03-E.CDR

Conductivity measurement:

Two-electrodemeasuring cells

Inductivemeasuring cells

Resistance measurement:

2

Page 3: mycom CLM 152 · 2010-05-03 · Technical Information TI 144C/07/en No. 50077402 Areas of application The analytical transmitter Mycom CLM 152 is intended for highly reliable and

General information Precise calibration and measurementThe instrument supports two types ofcalibration:• Wet calibration

Cell adaptation by entry ofconductivity values of knowncalibration solutions.

• Data entryEntry of exact cell constant.

Temperature compensationThe electrolytic conductivity stronglydepends on temperature. Theconductivity display value is related toa reference temperature using atemperature coefficient specific to eachsolution.The instrument supports:• Linear compensation 0 ... 10 %/K with

user-definable reference temperature• Compensation acc. to IEC 746, part 3

for NaCl• Temperature compensation for

ultrapure water with trace impurities• User-programmable Tc curves.The temperature can either bemeasured continuously or programmedas a constant.

Polarisation compensationPolarisation effects in the boundarylayer between the electrode andmeasuring solution limit the measuringrange of conductive conductivitymeasuring cells.The Mycom CLM 152 transmitter candetect and compensate for polarisationeffects using an innovative, intelligentsignal evaluation procedure.This results in a very significantexpansion of the measuring range ofa measuring cell (cf. overview ofconductivity measuring cells).

Difference measurementThe “2-circuit” variant permits theconnection of two identical measuringcells for measurement and monitoringof a difference in conductivity, e.g. for:• Media separation• Monitoring of heat exchangers• Monitoring of ion exchangers.

Concentration measurementThe Mycom CLM 152 has the ability todirectly display and output substanceconcentration values.The concentration values of foursubstances are stored in the instrument:NaOH (sod. hydrox. sol.) 0 ... 15 %HNO3 (nitric acid) 0 ... 20 %H2SO4 (sulphuric acid) 0 ... 20 %H3PO4 (phosphoric acid) 0 ... 12 %Four additional concentration curves foruser-specific media can be programmed.

Contacts according to NAMURSwitching contacts for output of functioncheck, warning and failure messagesas recommended by the measuring andcontrol engineering standardscommittee (NAMUR).

Simple controlIn the basic version, the instrument isequipped with two potential-freeoutput contacts (non-Ex: relays, Ex:optocouplers). One of them serves asa failure contact, the other can beprogrammed by the user to function asa limit contactor or fault signalling contact.By adding a plug-in module, theinstrument can be expanded to a totalof five output contacts.The table below shows the configurationof these five output contacts accordingto NAMUR.

Output contact

1 2 3 4 5

Failu

re

Mai

nten

.re

qui

red

Func

tion

chec

k

Lim

it1

Lim

it2

Cleaning contacts(conductive measuring systems only)The contact outputs can be switched tocontrol spray cleaning systems, e.g. theinjector CYR 10. Cleaning can beinitiated manually or automatically byprogrammed cleaning cycles.

Clever service and start-up functionsThe instrument provides a number ofuseful service functions, such as log,current output simulation and relayswitching simulation, plus someadditional selective checking functions,e.g. keyboard, display or RAM test. Theuser is guided through the setupprocedure, assuring quick selection ofappropriate basic settings.

USP 23The USP (United States Pharmacopeia)is instrumental in defining therequirements for ultrapure water in thepharmaceutical industry.The Mycom CLM 152 fulfills all USPrequirements for conductivity measuringsystems:• Exact temperature measurement at

the same point where the conductivityis measured

• Simultaneous display ofuncompensated conductivity valueand temperature

• Display resolution of 0.001 µS/cm• Exact factory adjustment of transmitter

with traceable laboratory resistances• Precise factory calibration of

measuring cells according toASTM D 1125-91

• Temperature-dependent monitoringaccording to USP 23 stage 1.

3

Page 4: mycom CLM 152 · 2010-05-03 · Technical Information TI 144C/07/en No. 50077402 Areas of application The analytical transmitter Mycom CLM 152 is intended for highly reliable and

Features/options Modular designThe design of the CLM 152 transmitterwith plug-in modules permits optimaladaptation of the instrument to user-specific requirements.

Internal / external remotemeasuring range switchingWhen the analog/contact inputexpansion module is added, theinstrument has up to four externallyswitchable measuring ranges withparameters to be chosen by the user.Inductive measuring systems inparticular are used with measuringrange switching due to the wide rangespan encountered. For theseapplications, the inductive MycomCLM 152 transmitter offers the benefit

of automatic internal measuring rangeswitching: the instrument automaticallyswitches the instrument parameters,e.g. the current output, to the next rangedepending on user-definable switchingthresholds and signals the currentlyvalid measuring range via relaycontacts.Switching affects the measuring range,the temperature coefficient and thelimits.

Basic equipment Expansion module

Conductivityinput

Outputcontacts

Analog/contact input

Inputs Conductivity input

Temperature input

Conduct. 2

Temp. 2

2nd cond./temperatureinput fordifference/redundantmeasurement

Analoginput

(active)+ 2 contact

inputs

OutputsConductivity output

Temperature output

ContactsFailure and

maintenancerequired or

failure and limit

Cleaning/NAMURcontacts

Holdcontacts,remote

measuringrange

switching

Cd. Cd.

mA

ϑ

Cd.

mA

mA

ϑ

Measuring range (parameter settings)

1 2 3 4

Medium Beer Water Cleaning 1 Cleaning 2

Conductivity range 800 ... 1200µS/cm

300 ... 700µS/cm

100 ... 200mS/cm

10 ... 100mS/cm

Temperaturecoefficient

2.0 %/K 2.1 %/K 1.6 %/K 2.8 %/K

Limits1000 µS/cm

MAX500 µS/cm

MAX120 mS/cm

MIN80 mS/cm

MIN

Contactstate

1 open closed open closed

2 open open closed closed

Example for remotemeasuring rangeswitching: control ofa CIP system

ϑ

4

Page 5: mycom CLM 152 · 2010-05-03 · Technical Information TI 144C/07/en No. 50077402 Areas of application The analytical transmitter Mycom CLM 152 is intended for highly reliable and

Operation Menu-guided operationThe functions of the Mycom CLM 152transmitter are arranged in four maingroups:• Parameter settings• Measurement• Calibration• Diagnosis.Pressing the PARAM, MEAS, CAL andDIAG keys accesses the associatedselection menus where the availableoptions are shown in plain text. Thecurrently selected element is displayedin reverse video.Functions are selected with the arrowkeys which are also used to editnumerical values.

Everything at a glanceThe eight-line display has plenty ofspace for auxiliary information:• Secondary measured values• Temperature compensation setting• Options available for selection• Range of the currently edited value

No unauthorised accessConfiguration and calibration areeffectively protected againstundesirable changes by threeaccess code levels.

• No code enteredThe following can be accessed:– Measurement displays– Error list– Info list– Log– Calibration data history.

• The operator codeprovides access to:– Shortcut menu– Calibration type settings– Calibration functions– Temperature compensation– Cleaning functions.

• The advanced codeprovides access to all menus andfunctions.

PARAM

MEAS CAL

DIAG

Display

The illuminated dot matrix displayhas room for eight lines of text.The character set includes allletters and numbers in single-lineand multiple-line sizes.Reverse video is used to highlightthe currently selected element.

MEAS (measurement)

Displays the measuredconductivity value in largenumbers with user-selectableauxiliary information.

This key can be pressed to returnto the measuring mode from anysubmenu.

DIAG (diagnosis)

The red LED signals an errorcondition. The error messages canbe displayed in the diagnosismenu. The operating and errorhistory is stored in a logbook.

The access code-protected servicefunctions are also accessed withthis key.

CAL (calibration)

This key accesses the calibrationmenu in which the operator isguided through the calibrationsequence by means of plain-textmessages.

PARAM (parameter settings)

Following entry of the appropriatecode, the instrument configurationand controller settings can bechanged at the advanced level.Not all fields are accessible at theoperator level.

Operation

The arrow keys are used to selectmenu items and edit numericalvalues. Press the E key (Enter) toaccept a value and go on to thenext field.

Double arrow

Visible during active communicationvia PROFIBUS.

LM152BOE.CDR

Display andoperating elementsof Mycom CLM 152

5

Page 6: mycom CLM 152 · 2010-05-03 · Technical Information TI 144C/07/en No. 50077402 Areas of application The analytical transmitter Mycom CLM 152 is intended for highly reliable and

Electrical connection ofCLM 152-A innon-Ex area

Connections are established in aseparate connection compartmentwhich holds the power connection,the signal and contact outputs of themotherboard (basic assembly) anda connection module for conductivitymeasurement.The plug-in terminal blocks of up tothree additional plug-in modules

(depending on expansion; slots 2–4)are also located in this compartment:• Second conductivity input for

two-circuit measurement• Three additional output contacts for

controller or Chemoclean functions• Two binary input contacts (hold) and

one active 0 / 4 ... 20 mA current input(transmitter power supply).

88 87 85 86

Powersupply

31 32 33 34

I I1 2

K1

Analog out

0/4...20mAmax. 600

Relaycontact

250 Vac / 2A( cos =1)125 Vdc / 3A

PEmA mA

SL

OT

1

SL

OT

2

SL

OT

3

SL

OT

4

Cond.Temp.

F101

M3,

15A

/250

V

L L NL+ L+ N –

LM152AR.CDR

Connectioncompartment ofMycom CLM 152-Afor non-Ex area

Terminal block ofoptional modules

+ +- -11 12 13 14 15 16 17 31 32 33 34 85 86 87 88 L L N PE

Temperature Conductivity Conductivity Temp. Failure Contact 1 Mains

Pt1

00

mA mA

Conductivity input module Motherboard

Slot 1

(Jumpers fortwo-el. meas. cells)

Cd. Cd.

DC: L+ L+ L–

LM152AN.CDR

Connection diagram forconductive systems,e.g. CLM 152-A1Axx10

11 12 13 14 15 16 17 89 90 91 92 93 94 21 22 81 82 83 84

Temperature Conductivity Contact 2 Contact 3 Contact 4 Analog inp. Binary inputs

Pt1

00

Conductivity input module Output contact module Analog/contact input module

mA

(Jumpers fortwo-el. meas. cells)

Cd.

LM152MA.CDRConnection diagrams ofadd-in modules

+ +- -

Pt1

00

mA mA

Conductivity input module Motherboard

Slot 1

Cd. Cd.

Temperature Conductivity Conductivity Temp. Failure Contact 1 Mains

11 12 13 14 15 16 17 31 32 33 34 85 86 87 88 L L N PE

Tran

smit

Rec

eive

DC: L+ L+ L–

LM152CON.CDR

Connection diagram forinductive systems,e.g. CLM 152-A1Axx15

6

Page 7: mycom CLM 152 · 2010-05-03 · Technical Information TI 144C/07/en No. 50077402 Areas of application The analytical transmitter Mycom CLM 152 is intended for highly reliable and

Electrical connection ofCLM 152-Z inEx area

The Mycom CLM 152-Z is certified foroperation in explosive atmospheres(EEx em [ia/ib] IIC T4, BVS 95.D.2098and Swiss Electrotechnical Association(SEA) 96.1 10489 test certificates).This transmitter may be operated inzone 1 or 2. Endress+Hauserconductivity measuring cells can beused in zone 1 without requiring aspecial approval.

The Endress+Hauser basic informationpublication GI 003/11/de “Explosions-schutz von elektrischen Betriebsmittelnund Anlagen” (“Explosion protection ofelectric equipment and systems”)contains helpful information on theinstallation and operation of electricequipment in explosive atmospheres.This brochure may be obtained from allEndress+Hauser sales offices.

SL

OT

1

SL

OT

2

SL

OT

3

SL

OT

4

Cond. AnalogoutTemp.

PE

88 87 85 86

K1

2x Optocoupler contactmax. 30 Vdc / 100mA / 750mW

Power-supply

Sicherung / fuse:50076930

100V..230V50076931 24Vac50087807 24Vdc

+ +

L

L+

N

L-

F

F

LM152ARX.CDR

Connectioncompartment of MycomCLM 152-Z (Ex)

PARAM

MEAS CAL

DIAG PARAM

MEAS CAL

DIAG

Ex zone 1

Measuring currentcircuit ia

non-Ex range

Signal outputcircuit ia

Supply circuitLM152_EX.CDR

Inductive measuring celland transmitter inEx area

+ +- - L+ L-

mA mA

Temperature Conductivity Conductivity Temp. Failure Contact 1 Mains

11 12 13 14 15 16 17 31 32 33 34 85 86 87 88 L N PE

Conductivity input module Current output module (Ex) Motherboard

Pt1

00

Slot 1 Slot 2

(Jumpers fortwo-el. meas. cells)

Cd. Cd.

LM152ANX.CDR

Connection diagramof basic Ex versionCLM 152-Z (Ex)

11 12 13 14 15 16 17 89 90 91 92 93 94 21 22 81 82 83 84

Temperature Conductivity Contact 2 Contact 3 Contact 4 Analog inp. Binary inputs

Pt 100

module module module

mA

2nd conductivity input Output contact (Ex) Analog/contact input

(Jumpers fortwo-el. meas. cells)

C E C E C E

Cd.

LM152MAX.CDR

Connection diagrams ofplug-in modules (Ex)

An Exinstrumentversion maycontain up to twoplug-in modulesin addition to thestandardequipment.

7

Page 8: mycom CLM 152 · 2010-05-03 · Technical Information TI 144C/07/en No. 50077402 Areas of application The analytical transmitter Mycom CLM 152 is intended for highly reliable and

Electrical connection ofCLM 152-Z inEx area

IC_UCE_E.CDR

Characteristic ofswitching transistors onmotherboard and outputcontact module

U0

U0

U0

IK

IK

IK

UZ

Tl. 87

Tl. 88

+

UC

EC = 12 nFi,max

Tl. 85

Tl. 86

+

UC

EC = 12 nFi,max

Motherboard

Electrical characteristics of contact circuit:max. input voltage Ui,max = 30 Vmax. input current Ii,max = 100 mAmax. input power Pi,max = 750 mW

Consumer 1 Consumer 2

ASXEFCGA.CDR

External wiring ofoutput contacts onmotherboard:Only one intrinsicallysafe contact circuit(ia or ib) with➀ linear

Pa = ¼(U0⋅IK)➁ trapezoidal

1) UZ > 0.5 U0:Pa = ¼(U0⋅IK)

2) UZ < 0.5 U0:Pa = (U0 – UZ)⋅Ui/Ri

➂ or squarePa = U0⋅IK

output characteristic

Tl. 81

Tl. 82

Tl. 83

Tl. 84

Tl. 21

Tl. 22

+

+

+

Analog/contact input module

Parallel connection of the two input circuits (term. 81/82 and 83/84)in a single intrinsically safe circuit is also permissible.

Intrinsically safe ia/ib circuit,Max. load 500 Ω

Switch 1

Switch 2

Ci,max = 15 nFLi,max = 53 µH

Ci,max = 15 nFLi,max = 53 µH

C0,max = 40 nFL0,max = 440 µH

Intrinsically safe ia/ib circuit 1:Voltage supply Ui = 4 ... 30 VCurrent consumption max. 3 mA @ 30 V

Intrinsically safe ia/ib circuit 2:Voltage supply Ui = 4 ... 30 VCurrent consumption max. 3 mA @ 30 V

U0,max = 22.0 VI0,max = 50 mAP0,max = 640 mW

113 Ω

Ui,max = 30 VIi,max = 100 mAPi,max = 750 mW

ASXEFCXI.CDR

External wiring ofinput circuits on FCXImodule (analog/contact input)

Tl. 89

Tl. 90

+

UC

EC = 12 nFi,max

Tl. 91

Tl. 92

+

UC

EC = 12 nFi,max

Tl. 93

Tl. 94

+

UC

EC = 12 nFi,max

Output contact module (Ex)

Intrinsically safe ia/ib circuit 1:max. output voltage Ui,max = 30 Vmax. output current Ii,max = 100 mAmax. output power Pi,max = 750 mW

Parallel connection of the three contact circuits (term. 89/90, 91/92, 93/94)in a single intrinsically safe circuit is also permissible.

Consumer 1

Consumer 2

Consumer 3

Intrinsically safe ia/ib circuit 2:max. output voltage Ui,max = 30 Vmax. output current Ii,max = 100 mAmax. output power Pi,max = 750 mW

Intrinsically safe ia/ib circuit 3:max. output voltage Ui,max = 30 Vmax. output current Ii,max = 100 mAmax. output power Pi,max = 750 mW

ASXEFCKY.CDR

External wiring of outputcontacts on outputcontact module

8

Page 9: mycom CLM 152 · 2010-05-03 · Technical Information TI 144C/07/en No. 50077402 Areas of application The analytical transmitter Mycom CLM 152 is intended for highly reliable and

Dimensions

Mounting

Mounting version B is supplied withmounting elements to be used for panelor post mounting.

Mounting cutout: 161+0.5 × 241+0.5 mmInstallation depth: approx. 134 mmMaximum pipe diameter: 70 mm

PARAM

MEAS CAL

DIAG

167 111

247

82

MY152MAS.CDRDimensions ofMycom CLM 152

111167

142

247

154

∅8

M6 screw

MY152WAN.CDRWall mounting ofMycom CLM 152

207

MY152MTG.CDR

Panel mounting (left)and post mounting(right).

Weather protectioncover CYY 101,available as anaccessory, isrecommended foroutdoor installation

9

Page 10: mycom CLM 152 · 2010-05-03 · Technical Information TI 144C/07/en No. 50077402 Areas of application The analytical transmitter Mycom CLM 152 is intended for highly reliable and

Technical data,inductive

Data independent of measuring cell

Data of measuring cell CLS 50

Measuring range uncompensated 0 ... 2000 mS/cm

Measuring range compensated 0 ... 1000 mS/cm

Measurement deviation1 ±0.5 % of measured value ± 3 digits

Reproducibility1 ±0.2 % of measured value ± 3 digits

Max. cable length 55 m (with junction box VBM)

Current output transfer characteristic linear, bilinear

Subject to modifications.1acc. to IEC 746-1, for nominal operating conditions

General data Lower measuring range limit 5 µS/cm ... 2000 mS/cm

Cell constant approx. 1.98 cm–1

Storage temperature –20 ... +80 °C

Ingress protection (DIN 40050) IP 67 (measuring cell in mounted state)

Measurement deviation (–20 ... +100 °C) ±(5 µS/cm + 0.5 % of measuring range)

Measurement deviation (> 100 °C) ±(10 µS/cm + 0,5 % of measuring range)

Temperature measurement Temperature sensor Pt 100, class A acc. to IEC 751

Temperature response time t90 90 % of final temperature display: 10 ... 15 min(acc. to IEC 746-1)

Installation Required pipe cross-section > DN 80 (for pipe diameters < DN 110, noteadaptation factor)

Installation in reduced outflow ≥ DN 50

Supplementary documentation Technical Information CLS 50 order no. 50090385

Subject to modifications.

Order variantswithout flange

Order variantswith flange

Pre

ssur

e[b

ar]

Pre

ssur

e[b

ar]

20

16

10

20

16

10

PEEK

PFA

Variant up to125 °C

Variant up to180 °C

Variant up to180 °C

Variant up to125 °C

PVDF

Temperature [°C]

Temperature [°C]

DN 50 / ANSI 2";stainless steel

JIS; stainless steel

PT-DIAGR.CDR

–20 125 180

–20 125 18020

Pressure-temperaturecurves as a functionof material and flangevariant

10

Page 11: mycom CLM 152 · 2010-05-03 · Technical Information TI 144C/07/en No. 50077402 Areas of application The analytical transmitter Mycom CLM 152 is intended for highly reliable and

Technical data,inductive (cont.)

Data of measuring cell CLS 52

General data Lower measuring range limit (uncompensated) 10 µS/cm ... 2000 mS/cm

Reaction time T90 < 3 s over total measuring range (one-circuitinstrument)

Temperature measurement Temperature sensor Pt 100, class A acc. to IEC 751

Temperature response time t90 < 15 s

Heat transfer bushing with Pt 100

Material V4A (1.4571)

Seal O-ring, EPDM

Cell constant k (measurement deviation ±0.5 %) 5.9 cm–1

Storage temperature –25 ... +80 °C

Ingress protection (DIN 40050) IP 67

Measurement deviation (–5 ... +100 °C) ±(10 µS/cm + 0.5 % of measuring range)

Measurement deviation (+100 ... +140 °C) ±(30 µS/cm + 0.5 % of measuring range)

Medium temperature –5 ... +125 °C

Ambient temperature –10 ... +70 °C

for sterilisation +140 °C (max. 30 min)

Pressure max. 16 bar (20 °C)

Measuring cell material PEEK

Surface roughness Ra ≤ 0.5 µm

Installation Required pipe cross-section

Dairy pipe fitting, clamp nozzle G 1½ ≥ DN 65

APV, Varivent connection ≥ DN 40

Supplementary documentation Technical Information CLS 52 order no. 50086110

Subject to modifications.

11

Page 12: mycom CLM 152 · 2010-05-03 · Technical Information TI 144C/07/en No. 50077402 Areas of application The analytical transmitter Mycom CLM 152 is intended for highly reliable and

Technical data,conductive

Technical data ofPROFIBUS-PA

Output parameters Output signal digital communication signal, PROFIBUS-PA

PA function slave

Response time Slave approx. 20 ms

PLC approx. 600 ms at approx. 30 instruments

Alarm signal PROFIBUS-PA: signal status bit is set,last valid measured value is retained

Integration time 0 ... 99 s, default 0 s

Communication resistor none, separate PROFIBUS-PA terminating resistor

Physical layer IEC 1158-2

Integrated overvoltage protection 25 V AC / 250 A

Display and user interface Remote control via PC using operating program Commuwin II

Communication point PROFIBUS-PA

Power supply Supply voltage 9 ... 32 V DC

Current consumption 10 ± 1 mA

Inrush current corresponds to Table 4, IEC 1158-2

Subject to modifications.

Conductivity / resistance / concentrationmeasurement

Measuring and display ranges for conductivity

Measuring and display ranges for resistance

Measurement deviation2 in display ±0.5 % ± 2 digits of measured value

Reproducibility2 ±0.2 % ± 2 digits of measured value

Reference temperature adjustable –35 ... +250 °C, typ. +25 °C

Measuring frequency 128 ... 1024 Hz

Measuring voltage ≤ 300 mV

Max. cable length for resistance approx. 20 m

Max. cable length for conductivity/concentration approx. 100 m

Supplementary documentation Technical Information CLS 12 order no. 50059349

Technical Information CLS 13 order no. 50059350

Technical Information CLS 15 order no. 50065950

Technical Information CLS 19 order no. 50065951

Technical Information CLS 21 order no. 50059352

Technical Information CLS 30 order no. 50059353

Subject to modifications.

1The instrument has four internal switching stages in each measuring range.This achieves maximum precision and resolution automatically.

2acc. to IEC 746-1, for nominal operating conditions

Cell constant k Measuring range (MR) 1 Display range (DR)

0.01 cm–1 0.0 nS/cm ... 600.0 µS/cm 0.0 µS/cm ... 200.0 µS/cm

0.1 cm–1 0.000 µS/cm ... 6000 µS/cm 0.000 µS/cm ... 2000 µS/cm

1 cm–1 0.00 µS/cm ... 60.00 mS/cm 0.00 µS/cm ... 20.00 mS/cm

10 cm–1 0.0 µS/cm ... 600.0 mS/cm 0.0 µS/cm ... 200.0 mS/cm

Cell constant k Measuring range (MR) Display range (DR)

0.01 cm–1 20.0 kΩ⋅cm ... 20.00 MΩ⋅cm 20.0 kΩ⋅cm ... 37.99 MΩ⋅cm

0.1 cm–1 2.00 kΩ⋅cm ... 2000 kΩ⋅cm 2.00 kΩ⋅cm ... 3799 kΩ⋅cm

1 cm–1 0.200 kΩ⋅cm ... 200.0 kΩ⋅cm 0.200 kΩ⋅cm ... 379.9 kΩ⋅cm

12

Page 13: mycom CLM 152 · 2010-05-03 · Technical Information TI 144C/07/en No. 50077402 Areas of application The analytical transmitter Mycom CLM 152 is intended for highly reliable and

General technical data

General data Manufacturer Endress+Hauser

Product designation Mycom CLM 152

Limit and alarm functions Function Limit contactor

Function type MIN or MAX

Set point settings (in absolute values) 0 ... 100 % of display range

Hysteresis for switching contacts (in absolute values) 1 ... 10 % of display range

Pickup / dropout delay 0 ... 7200 s

Alarm threshold 0.5 ... 100 % of display range

Alarm delay 0 ... 6000 s

Electrical connection data Power supply AC 24 / 100 / 115 / 200 / 230 V +10 / –15 %

Frequency 47 ... 64 Hz

Power supply DC 24 V +20 / –15 %

Power consumption max. 10 VA

Contact outputs (optional) potential-free change-over contacts (Ex version:optocouplers), switchable as NO or NC contacts

Switching current max. 3 A

Switching voltage max. 250 V AC / 125 V DC

Switching power max. 750 VA

Signal outputs 2 × 0 / 4 ... 20 mA, electrically isolated from othercircuits, but not between outputs

Separation voltage 276 Vrms

Current output

Current range 0 / 4 ... 20 mA

Measurement deviation ≤ 0.2 % of current end value

Load max. 600 Ω

Terminals, max. cable cross-section 2.5 mm²

Temperature measurement Temperature sensor Pt 100 (three-wire circuit)

Measuring range (also displayable in °F and K) –35 ... +250 °C

Measured value resolution 0.1 °C

Measurement deviation1 at temperature signal output 0 ... 100 °C: ±0.5 K; remaining range: ±1 K

Reproducibility1 max. 0.1 % of measuring range

Temperature signal output transfer range adjustable ∆ 28.5 ... ∆ 285 °C

Concentration measurement

Temperature compensation Range for linear and freely programmableTc values –35 ... 250 °C

Range for NaOH 0 ... 85.0 °C

HNO3 0 ... 75.0 °C

H3PO4 0 ... 75.0 °C

H2SO4 0 ... 80.0 °C

Selection Conductivity range Concentration

NaOH 0.0 mS/cm ... 410 mS/cm 0 ... 15 %

HNO3 0.0 mS/cm ... 781 mS/cm 0 ... 20 %

H2SO4 0.0 mS/cm ... 723 mS/cm 0 ... 20 %

H3PO4 0.0 mS/cm ... 73 mS/cm 0 ... 12 %

USER 1 ... 4 0.0 µS/cm ... 2000 mS/cm 0 ... 99.99 %

13

Page 14: mycom CLM 152 · 2010-05-03 · Technical Information TI 144C/07/en No. 50077402 Areas of application The analytical transmitter Mycom CLM 152 is intended for highly reliable and

General technical data(cont.)

Ambient conditions Electromagnetic compatibility (EMC)

Interference emission acc. to EN 50081-1:1992

Interference immunity acc. to EN 50082-2:1995

Nominal operating conditions

Ambient temperature –10 ... +55 °C

Relative humidity 10 ... 95 %, non-condensing

Limit operating conditions

Ambient temperature –20 ... +60 °C

Storage and transport temperature –25 ... +85 °C

Ex permit EEx em [ia/ib] IIC T4

Ex conformity certificates BVS 95.D.2098; SEV 96.1 10489

Mechanical data Dimensions (H × W × D) 247 × 167 × 111 mm

Weight max. 6 kg

Ingress protection IP 65

Materials

Housing GD-AlSi 12 (Mg content < 0.05 %), plastic-coated

Front panel polyester, UV-resistant

Subject to modifications.1acc. to IEC 746-1, for nominal operating conditions

14

Page 15: mycom CLM 152 · 2010-05-03 · Technical Information TI 144C/07/en No. 50077402 Areas of application The analytical transmitter Mycom CLM 152 is intended for highly reliable and

Accessories Weather protection cover CYY 101Weather protection cover CYY 101 is tobe used for outdoor installation.Order no. CYY 101-A

Round post mounting for CYY 101For attachment of weather protectioncover to vertical or horizontal pipes withdiameters of up to 70 mm.Order no. 50062121

Calibration solutionsPrecision calibration solutions forconductivity, accuracy 0.5 % at 25 °C,related to SRM by NIST.Bottle containing 500 ml.

Type Conductivityat 25 °C1

Ordernumber

CLY 11-A 74.0 µS/cm 50081902

CLY 11-B 149.6 µS/cm 50081903

CLY 11-C 1.406 mS/cm 50081904

CLY 11-D 12.64 mS/cm 50081905

CLY 11-E 107.00 mS/cm 50081906

1Values may vary due to manufacturingtolerances. The accuracy is given forthe value specified on the bottle.

Pure water calibration system ConCalLow-conductivity calibration solutionsare unstable. For this reason, they areonly permissible above 74 µS/cm acc.to DIN/IEC. The alternative is to use thefactory-calibrated ConCal comparisonmeasuring system (factory calibrationtraceable to ASTM by NIST).

270

300

320

YH-101-6.CDR

66 55

30

7090

M6

∅8

YH-101-7.CDR

15

Page 16: mycom CLM 152 · 2010-05-03 · Technical Information TI 144C/07/en No. 50077402 Areas of application The analytical transmitter Mycom CLM 152 is intended for highly reliable and

Product structure

Endress+Hauser GmbH+Co.- Instruments International -

P.O. Box 2222D-79574 Weil am Rhein

Tel. (07621) 975 - 02Fax (07621) 975345

Hauser+EndressNothing beats know-how

TI 144C/07/en/07.99Printed in Germany / DT / CV5 (50077402)

Conductivity and resistance measuring transmitter Mycom CLM 152

Field housing, ingress protection IP 65, for wall mountingSignal outputs 0 / 4 ... 20 mA for conductivity / resistance and temperature, 2 output contacts

Equipment and certificateA1A 1-circuit; basic versionA1B 1-circuit; 3 relays / ChemocleanA1C 1-circuit; feedback/hold contact / remote MR switchingA1D 1-circuit; 3 relays / Chemoclean / feedback/hold contact / remote MR switchingA1E 1-circuit; PROFIBUSA1F 1-circuit; 3 relays / Chemoclean, PROFIBUSA1G 1-circuit; feedback/hold contact / remote MR switching, PROFIBUSA1H 1-circuit; 3 relays / Chemoclean / feedback/hold contact / remote MR switching, PROFIBUS

A2A 2-circuit; basic versionA2B 2-circuit; 3 relays / ChemocleanA2C 2-circuit; feedback/hold contact / remote MR switchingA2D 2-circuit; 3 relays / Chemoclean / feedback/hold contact / remote MR switchingA2E 2-circuit; PROFIBUSA2F 2-circuit; 3 relays / Chemoclean, PROFIBUSA2G 2-circuit; feedback/hold contact / remote MR switching, PROFIBUS

Z1A 1-circuit; EEx em [ia/ib] IIC T4Z1B 1-circuit; 3 optocouplers / Chemoclean, EEx em [ia/ib] IIC T4Z1C 1-circuit; feedback/hold contact / remote MR switching, EEx em [ia/ib] IIC T4Z1D 1-circuit; 3 optocouplers / Chemoclean, feedback/hold contact /

remote MR switching, EEx em [ia/ib] IIC T4Z1E 1-circuit; PROFIBUS, EEx em [ia/ib] IIC T4Z1F 1-circuit; 3 optocouplers / Chemoclean, PROFIBUS, EEx em [ia/ib] IIC T4Z1G 1-circuit; feedback/hold contact / remote MR switching, PROFIBUS, EEx em [ia/ib] IIC T4

Z2A 2-circuit; EEx em [ia/ib] IIC T4Z2B 2-circuit; 3 optocouplers / Chemoclean, EEx em [ia/ib] IIC T4Z2C 2-circuit; feedback/hold contact / remote MR switching, EEx em [ia/ib] IIC T4Z2E 2-circuit; PROFIBUS, EEx em [ia/ib] IIC T4

Power supply0 230 V, 50 / 60 Hz1 115 V, 50 / 60 Hz2 200 V, 50 / 60 Hz3 24 V, 50 / 60 Hz5 100 V, 50 / 60 Hz8 24 V DC

LanguageA D, E, F, I switchableC D, E, F, NL, J switchable

Measuring method / equipment10 conductive15 inductive20 conductive with moisture protection varnish25 inductive with moisture protection varnish

AttachmentA without additional attachmentB post mounting kit

complete order codeCLM 152-