91xx pH/Electrolyte Analyzers Service Manual

Transcript of mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf ·...

Page 1: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

91xx pH/Electrolyte Analyzers Service Manual

Page 2: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

COBASis a trademark of Roche.©2007 Roche Diagnostics

Roche Diagnostics GmbHD-68298 MannheimGermanywww.roche-diagnostics.com

Page 3: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service Manual · Version 2.0 3

91xx pH/Electrolyte Analyzers

Version History

e For more information, see:

Software on page G-5

Service Manual on page G-6

Edition notice

91xx pH/Electrolyte AnalyzersService Manual

This Service Manual is for the maintenance and repair of 91xx pH/Electrolyte Analyzers (9110 pH Analyzers and 9120, 9130, 9140, 9180 and 9181 Electrolyte Analyzers).

91xx pH/Electrolyte Analyzers Every effort has been made to ensure that all the information contained in this manual is correct at the time of printing. However, Roche Diagnostics GmbH reserves the right to make any changes necessary without notice as part of ongoing product development.

Any customer modification to the instrument will render the warranty or service agreement null and void.

Software updates are done by Roche Service representatives.

9180 Electrolyte Analyzers SN > 12001 are IVD compliant.

Intended use

This manual contains all of the information required for maintenance and repair of 91xx pH/Electrolyte Analyzers (9110 pH Analyzers and 9120, 9130, 9140, 9180 and 9181 Electrolyte Analyzers).

The user must be familiar with the function and operation of the instrument to fully understand the processes described here.

e For more information about 91xx pH/Electrolyte Analyzers, refer to corresponding

Instructions for Use.

Observe the service and repair procedures described in this manual and use only genuine Roche replacement parts and Roche-approved materials to guarantee the full functionality of the 91xx pH/Electrolyte Analyzers.

e For the order numbers of replacement parts, refer to:

91xx pH/Electrolyte Analyzers Spare Part List

Manual Version Software Version Version date Changes

1.0 n/a July 2004 First edition

2.0 n/a April 2007 New template

Language Order Number

English 03077063001

Page 4: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

4 Service Manual · Version 2.0

91xx pH/Electrolyte Analyzers

e For an overview of possible revisions and available software versions, see:

Software on page G-5

Service Manual on page G-6

Copyrights

© 2007 Roche Diagnostics GmbH, All rights reserved.

The contents of this document may not be reproduced in any form or communicated to any third party without the prior written consent of Roche Diagnostics. Every effort is made to ensure its correctness. Subject to change without notice.

Trademarks

COBAS, LIFE NEEDS ANSWERS, ROCHE MICROSAMPLER, SNAPPAK, ISETROL are trademarks of Roche.

All other trademarks are the property of their respective owners.

Contact addresses

Manufacturer Roche Diagnostics GmbH

D-68298 Mannheim / Germany

www.roche.com

Page 5: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 5

91xx pH/Electrolyte Analyzers

Contents

Version History 3Edition notice 3Intended use 3Copyrights 4Trademarks 4Contact addresses 4Contents 5Preface 9How to use this manual 9Where to Find Information 9Conventions used in this manual 10

System Description Part A

1 Safety informationImportant information A-5Operating safety information A-6Important notes and warnings A-6Disinfectants A-7Deproteinizer A-7Other disinfectants A-7ESD protection measures A-8Explanation of the phenomenon A-8Influence of electrostatic charges on components A-8Why is ESD protection so important today? A-9How can ESD protection be guaranteed? A-9Conclusion A-9

2 SpecificationsSpecifications A-13Specifications A-13Operating Parameters A-13Classifications A-15

3 Interface specificationsInterface specifications A-19Interface information A-19Example data string information A-20Data Link Information (9180/9181 only) A-20

4 Description of modulesDescription of modules A-23Mechanical assemblies A-23Fluidic Module A-27Electronics A-33

5 Electronic diagramsElectronic diagrams A-399110 analyzer block diagram A-399120, 9130, 9140, 9180 analyzer block diagram A-409181 analyzer block diagram A-41

System interconnect - 9110, 9120, 9130, 9140 A-42SBC PCB - power supply - 9110, 9120, 9130, 9140 A-43SBC PCB - valve drivers - 9110, 9120, 9130, 9140 A-44SBC PCB - pump motor driver / door detect - 9110, 9120, 9130, 9140 A-45SBC PCB - input amplifiers - 9110, 9120, 9130, 9140 A-46SBC PCB - temperature circuit - 9110, 9120, 9130, 9140 A-47SBC PCB - analog selector and A/D converter - 9110, 9120, 9130, 9140 A-48SBC PCB - microprocessor - 9110, 9120, 9130, 9140 A-49SBC PCB - printer display drivers - 9110, 9120, 9130, 9140 A-50Printer daughter PCB - 9110, 9120, 9130, 9140, 9180, 9181 A-51Lamp PCB - 9110, 9120, 9130, 9140 A-52Display daughter PCB - 9110 A-53Preheater PCB - 9110 A-54Display daughter PCB - 9120, 9130, 9140 A-55System interconnect - 9180, 9181 A-56SBC PCB - power supply - 9180, 9181 A-57SBC PCB - valve drivers - 9180, 9181 A-58SBC PCB - pump motor driver/door detect - 9180, 9181 A-59SBC PCB - input amplifier - 9180, 9181 A-60SBC PCB - temperature circuit - 9180, 9181 A-61SBC PCB - analog selector and A/D converter - 9180, 9181 A-62SBC PCB - microprocessor - 9180, 9181 A-63SBC PCB - microprocessor - 9180, 9181 A-64SBC PCB - Fluid pack detector - 9180, 9181 A-65Lamp PCB - 9180, 9181 A-66Display daughter PCB - 9180 A-67

Software operation Part B

6 Software operation 9110Software operation 9110 B-5Overview B-5Overall program operation flow B-6Main menu B-7Measurement sequence B-8Print functions menu B-9QC sample menu B-10Daily maintenance menu B-11Operator functions menu B-12Program instrument menu B-13

Page 6: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

6 Service manual · Version 2.0

91xx pH/Electrolyte Analyzers

Service functions menu B-14Calibration sequence B-15

7 Software operation 9120, 9130, 9140Software operation 9120, 9130, 9140 B-19Overview B-19Overall program operating flow B-20Main menu B-21Measurement sequence B-22Print functions menu B-23QC/standard/urine sample menu B-24Daily maintenance menu B-25Operator functions menu B-26Program instrument menu B-27Service functions menu B-28Calibration sequence B-29

8 Software operation 9180Software operation 9180 B-33Overview B-33Overall program operation flow B-34Main menu B-35Measurement sequence B-36Print functions menu B-37QC/standard/dialysate/urine sample menu B-38Daily maintenance menu B-39Operator functions menu B-40Program instrument menu B-41Service functions menu B-42Calibration sequence (1) (Example Na/K/Li calibration) B-43Calibration sequence (2) (Example Na/K/Li calibration) B-44Power-up B-45

9 Software operation 9181Software operation 9181 B-49Overview B-49Overall program operting flow B-50Main menu B-51Manual measurement sequence B-52Automated measurement sequence B-53Print functions menu B-54QC/std/dialysate/urine sample menu B-55Daily maintenance menu B-56Operator functions menu B-57Program instrument menu B-58Service functions menu B-59Calibration sequence (1) (example Na/K/Li calibration) B-60Calibration sequence (2) (example Na/K/Li calibration) B-61Power-up sequence B-62

10 Service functionsService functions B-65TEST ELECTRODES B-65TEST SAMPLE SENSOR B-65TEST SnapPak SENSOR (9180 and 9181 only) B-65TEST LANGUAGE SWITCH (9180 and 9181 only) B-65TEST SAMPLE DOOR (except 9181) B-66TEST PROBE (9181 only) B-66TEST AUTOSAMPLER (9181 only) B-66TEST PUMP B-66TEST PINCH VALVES B-67TEST INTERFACE B-67TEST AMPLIFIER B-67ENTER SERVICE CODE B-67Miscellaneous functions B-69

Maintenance Part C

11 AdjustmentsAdjustments C-5Sample contacts adjustment (9110 only) C-5Sample sensor adjustment (9120/9130/9140/9180/9181 only) C-5Temperature Adjustment (9110/9140/9180/9181 only) C-6

12 MaintenanceMaintenance C-9Daily maintenance C-9Weekly maintenance C-9Monthly maintenance C-9Six month maintenance C-9Annual preventive maintenance C-9

Troubleshooting Part D

13 TroubleshootingError messages and troubleshooting D-5Status-LEDs D-5STATUS: NOT CALIBRATED D-5STANDARD A NOT DETECTED D-5STANDARD B NOT DETECTED, STANDARD C NOT DETECTED (9180/9181 only) D-6CHECK SAMPLE CONTACTS(9110), CHECK SAMPLE SENSOR (9120/30/40/80/81) D-6CHECK REFERENCE HOUSING D-7PLEASE CLOSE SAMPLE DOOR (except 9181) D-7NO SAMPLE D-8CLEANING FLUID NOT DETECTED D-8CONDITIONING FLUID NOT DETECTED (except 9110) D-8INTERFACE ERROR D-9

Page 7: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 7

91xx pH/Electrolyte Analyzers

PAPER JAM OR PRINTER DEFECT D-9CHECK ELECTRODES D-9## NOT CAL’D D-9CLOG IN SAMPLE PATH (9110 only), CLOG CHECK FLUID PATH (except 9110) D-11↑↑↑↑ ↓↓↓↓ (Out of range) D-11↑ ↓ (Temparature out of range) D-11ERR D-12PERFORM DAILY MAINTENANCE (except 9110) D-12REPLACE FLUID PACK D-12CHECK TEMPERATURE (9140/9180/9181 only) D-12VALVE OVERTEMP! CHECK VALVES D-13ERROR: UPPER NEEDLE SENSOR D-13ERROR: LOWER NEEDLE SENSOR D-13WHEEL MISSING OR SAMPLER DEFECT D-13SAMPLER JAMMED OR DEFECTIVE D-14

Glossary Part E

Glossary E-3

Index Part F

15 Index

Versions Part G

16 VersionsSoftware G-5Service Manual G-6Version 1.0 G-6Version 2.0 G-6

Page 8: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

8 Service manual · Version 2.0

91xx pH/Electrolyte Analyzers

Page 9: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 9

91xx pH/Electrolyte Analyzers

Preface

This manual contains all of the information required for the maintenance and repair of the 91xx pH/Electrolyte Analyzers.

The user must be familiar with the function and operation of the instrument to fully understand the processes described here.

e For more information about instructions, refer to:

9180 Electrolyte Analyzers Instructions for Use

Observe the service and repair procedures described in this manual and use only genuine Roche replacement parts and Roche-approved materials to guarantee the full functionality of the 91xx pH/Electrolyte Analyzers.

e For the order numbers of replacement parts, refer to:

91xx pH/Electrolyte Analyzers Spare Part List

e For an overview of possible revisions and available software versions, see:

Software on page G-5

Service Manual on page G-6

How to use this manual

To help finding information quickly, there is a table of contents at the beginning of the book and each chapter. In addition, a complete index can be found at the end.

Where to Find Information

In addition to the Service manual, the following documents are also provided to assist in finding desired information quickly:

o 91xx pH/Electrolyte Analyzers Instructions for Use

o 9180 Electrolyte Analyzer Short Instruction

o 91xx pH/Electrolyte Analyzers Service manual (PDF or iSDoc version in GRIPS)

o 91xx pH/Electrolyte Analyzers Spare Part List (PDF or iSDoc version in GRIPS)

o Keep this Service manual in a safe place to ensure that it is not damaged and remains available

for use.

o This Service manual should be easily accessible at all times.

Page 10: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

10 Service manual · Version 2.0

91xx pH/Electrolyte Analyzers

Conventions used in this manual

Visual cues are used to help locate and interpret information in this manual quickly. This section explains formatting conventions used in this manual.

Symbols The following symbols are used:

For more information about ESD protection measures, see ESD protection measures on page A-8.

Symbol Used for

a Start of procedure

o List item

e Cross-reference

h Call-up (software reference)

Note

All sections or text locations marked with "NOTE" describe safe

procedures that are intended to provide the user with additional help.

Caution

All sections or text passages that are marked with this symbol describe

procedures or indicate conditions or dangers that could damage or

lead to malfunctions in the instrument, and which therefore should

never be attempted.

Warning

Sections marked with this symbol contain information that must be

observed for the prevention of personal injury (to patients, users or

third parties).

Risk of infection!

(according to the standard DIN EN 61010-2-101:2002) (Instrument)

Risk of infection!

(according to the standard DIN ISO 15223-1:2005) (Consumables)

High Voltage

Passages that are marked with this symbol warn of an immediate

danger in connection with electrical wiring or components.

ESD protection measures

All sections or passages that are marked with this symbol warn of

specific dangers in connection with static discharge. Packages that are

marked with this symbol must only be opened by trained technical

staff.

Page 11: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 11

91xx pH/Electrolyte Analyzers

IVD symbols Instruments: applicable for 9180 Electrolyte Analyzers SN > 12001.

The IVD symbols are used in accordance with DIN EN 980:2003, ISO 15223:2000 (as per the In Vitro Diagnostics Directive 98/79/EC), and DIN EN ISO 780:1997.

Symbol Description

This product complies with the requirements in the directive

98/79/EC on in vitro diagnostic medical devices.

Lot designation

Consumables: use by... (expiry date)

The consumables must be completely consumed by the indicated

date.

If a day is not indicated, apply the last day of the respective month.

Store at ...

The conditions necessary to preserve the product's shelf life before

opening.

For in vitro diagnostic use

"Grüner Punkt" (in Germany)

Manufacturer — according to directive 98/79/EC on in vitro

diagnostic medical devices

Store upright

Risk of infection!

(according to the standard DIN ISO 15223-1:2005) (Consumables)

Catalogue number

Caution (refer to accompanying documents). Please refer to safety-

related notes in the manual, accompanying this instrument.

Please consult instructions for use

Serial number (model plate)

Page 12: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

12 Service manual · Version 2.0

91xx pH/Electrolyte Analyzers

Other symbols The following symbols are listed as additional information:

Symbol Description

Electrodes:

This date indicates the limit of the maximum storage time of an

electrode. The electrode must be installed in the instrument no later

than the imprinted date.

If the installation takes place on the imprinted date, it still falls within

the specifications. The calculation of the “Install before” date is based

on the production date of the elctrode.

T: R23/33/50-53

Danger symbol: T: Toxic (on the packaging of the reference electrode)

Rating: Inhalation, swallowing or skin contact with even small

quantities can lead to serious health risks, including fatal risks.

Characteristics of this type of item are severe, possible irreversible

damage to health through repeated or prolonged contact, particular

with carcinogenic, genetic or reproductive (danger to reproductive

capabilities) effects.

Caution: Avoid any contact with the human body. If you feel unwell,

contact a doctor immediately. Any substances with carcinogenic,

genetic or reproductive dangers are indicated appropriately. Always

observe the regulations when handling such substances.

N: S45/60/61/7

Danger symbol: N: Dangerous to the environment (on the packaging

of the reference electrode)

Rating: If released into aquatic and non aquatic environments, can

cause immediate or delayed damage to ecosystems through a change

in environmental conditions.

These substances or their by-products can cause simultaneous

damage to sensitive environmental areas.

Caution: Depending on the potential for damage, do not allow the

substance to enter sewers, soils or the environment. Observe the

specific disposal regulations.

XI: R36/37/38, S 26-37

Danger symbol: Xi: Irritant

Rating: Although not corrosive, momentary, longer-lasting, or

repeated contact with skin or mucous membrane may result in

inflammation. Danger of sensitization during contact with skin

(when classified with R 43).

Caution: Avoid contact with eyes and skin, do not inhale vapors.

Do not use content if the packaging is damaged

Protective gloves, protective goggles and suitable protective clothing

must be worn

Page 13: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 13

91xx pH/Electrolyte Analyzers

Abbreviations The following abbreviations are used:

Abbreviation Definition

A

ADC Analogue to digital converter

ANSI American National Standards Institute

C

CMOS Complementary Metal-Oxide Semiconductor

E

EEPROM Electrically erasable programmable read-only memory

e.g. exempli gratia – for example

EC European community

EN European standard

ESD Electrostatic Discharge

H

HIV Human Immunodeficiency Virus

HW Hardware

I

IEC International Electrical Commission

ISE Ion selective electrode

IVD In vitro Diagnostics

L

LCD Liquid Cristal Display

LED Light Emitting Diode

M

MC Measuring chamber

MOS Metal-Oxide Semiconductor

MSDS Material safety data sheet

P

PCB Printed Circuit Board

Q

QC Quality control

R

REF Reference solution

S

SBC board Mainboard of the instrument

SW Software

T

TTL Transisitor-Transistor Logic

Page 14: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

14 Service manual · Version 2.0

91xx pH/Electrolyte Analyzers

Page 15: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

April 2007

1 Safety information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-3

2 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-11

3 Interface specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-17

4 Description of modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-21

5 Electronic diagrams . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-37

System Description A

Page 16: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte
Page 17: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 A-3

91xx pH/Electrolyte Analyzers 1 Safety informationTable of contents

Safety information

Important information ...............................................................................................A-5

Operating safety information ..................................................................................... A-6

Important notes and warnings ................................................................................... A-6

Disinfectants ................................................................................................................ A-7

Deproteinizer .........................................................................................................A-7

Other disinfectants ................................................................................................A-7

ESD protection measures ............................................................................................A-8

Explanation of the phenomenon .......................................................................... A-8

Influence of electrostatic charges on components ...............................................A-8

Why is ESD protection so important today? ........................................................A-8

How can ESD protection be guaranteed? ............................................................. A-9

Conclusion ............................................................................................................. A-9

In this chapter Chapter 1

Page 18: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

A-4 Service manual · Version 2.0

1 Safety information 91xx pH/Electrolyte AnalyzersTable of contents

Page 19: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 A-5

91xx pH/Electrolyte Analyzers 1 Safety informationImportant information

Important information

This Service Manual contains vital warning and safety information.

This instruments are intended to be used only for the specialized purpose described in the instructions. The most important prerequisites for use, operation, and safety are explained to ensure smooth operation. No warranty or liability claims will be covered if the instrument is used in ways other than those described or if the necessary prerequisites and safety measures are not observed.

The instrument may be operated only by persons whose qualifications enable them to comply with the safety measures that are necessary during operation of the instrument.

Suitable protective equipment, like laboratory clothing, protective gloves, protective goggles and if necessary mouth protectors, must be worn to prevent direct contact with biological working materials. In addition, a face mask is required if there is a risk.

Adjustments and maintenance performed with removed covers and connected power may be attempted only by a qualified technician who is aware of the associated dangers.

Instrument repairs are to be performed only by the manufacturer or qualified service personnel.

Only accessories and supplies either delivered by or approved by Roche are to be used with the instrument. These items are manufactured especially for use with this instrument and meet the highest quality requirements.

Operation of the instrument with solutions whose composition is not consistent with that of the original solutions can negatively affect, above all, the long term measurement accuracy. Deviations in the composition of the solutions can also decrease the service life of the electrodes.

The quality control requirements must be completed at least once daily for safety reasons. Because accurate measurement results depend not only on the proper functioning of the instrument, but also on a number of other factors (such as preanalytics), the results produced by the instrument should be examined by a trained expert before subsequent decisions are reached that are based on the measurement values.

Explanation:

"Caution, refer to accompanying documents“

Page 20: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

A-6 Service manual · Version 2.0

1 Safety information 91xx pH/Electrolyte AnalyzersOperating safety information

Operating safety information

The instrument has been constructed and tested according to the following European Standards:

o IEC/EN 61010-1:2001

o IEC/EN 61010-2-101:2002 (valid for 9180 Electrolyte Analyzers SN > 12001 only)

o IEC/EN 61010-2-081:2002 + A1:2003

It was delivered from the factory in flawless condition with regards to safety features.

In order to preserve this condition and ensure safe operation, the user must respect the notices and warnings that are contained in this Service Manual

o This instrument is classified under the protection class I according to IEC/EN 61010-1.

o The instrument meets the conditions for overvoltage category II.

o The instrument meets the conditions for contamination level 2.

o Do not operate the instrument in an explosive environment or in the vicinity of explosive anesthetic mixtures containing oxygen or nitrous oxide.

o If an object or liquid enters the internal areas of the instrument, remove the instrument from its power supply and allow an expert to check it thoroughly before using it again.

o The instrument is suitable for long-term operation indoors.

Important notes and warnings

o The power cord may be plugged into a grounded socket only. When using an extension cord,

make sure it is properly grounded.

o Any rupture of the ground lead inside or outside the instrument or a loose ground connection

may result in hazardous operating conditions. Intentional disconnection of the grounding is not

permitted.

o The instrument is not suitable for operation with a direct current power supply. Use only the

original mains plug delivered with the instrument.

o Never operate the analyzer in the vicinity of volatile or explosive gases

(e.g. anesthetic gases, etc.)!

o The instrument must be connected to a 2-pin, grounded socket.

o The power cable and the plug must be undamaged. Damaged power cables and plugs must be

replaced immediately.

o Before opening the back panel switch off the instrument and disconnect the power cable.

o Replace damaged fuses with the specified type of fuse.

Page 21: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 A-7

91xx pH/Electrolyte Analyzers 1 Safety informationDisinfectants

Disinfectants

Deproteinizer

Composition Aqueous NaOCl solution with active chlorine (≤ 2%)

Hazards identification Due to the basic and oxidizing character of the reagent ("Deproteinizer") local irritations after contact with eyes, skin or mucous membranes cannot be excluded.

First aid measures

Other disinfectants

Use standard alcohol-based (70%) disinfectants to clean the surface of the instrument. Read the product information first.

o Service and repair work must be done only as specified in this manual. Unqualified service or

repair work may result in warranty claims not being granted.

o Use only suitable tools and test instruments for service and repair work.

o Do not allow fluids to enter the interior of the instrument, because this may damage the

electronics.

o Use only slightly moistened cloths or cotton sticks to clean the instrument.

Use only liquid disinfectant such as protein remover (Roche deproteinizer) or an alcohol-based

(about 70%) surface disinfectant.

Do not spray disinfectant directly onto the instrument because this could cause malfunctions in the

electronics.

Do not use any type of bleaching agent. Exception: Roche Deproteinizer.

Do not attempt to clean/decontaminate any part of the instrument before shutting it down and

unplugging it from the power source.

Before plugging in the instrument again and switching it on always wait for 15 minutes to allow the

disinfectant to evaporate.

After inhalation: breath fresh air, drink large amounts of water

After skin contact: wash with generous amounts of water, remove contaminated clothing

After eye contact: rinse eyes with generous amounts of water, contact an eye specialist

After drinking: drink large amounts of water, avoid vomiting, contact a doctor

Never use standard disinfectant to clean the tubes and tubing paths under any circumstances!

Do not use any type of bleaching agent. Exception: Roche Deproteinizer.

Page 22: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

A-8 Service manual · Version 2.0

1 Safety information 91xx pH/Electrolyte AnalyzersESD protection measures

ESD protection measures

Explanation of the phenomenon

The most frequent cause of electrostatic discharge is friction of various materials such as plastic, synthetic fiber, hard rubber or paper.

It can also be caused by bending or applying pressure to a material.

Discharges that are dangerous to components occur when "charged" bodies or components do not have a discharge connection (ground) to discharge the electrostatic charge.

The discharges normally do not have a high capacity, but there are often voltage differences in the range of several thousand volts. The discharges are perceived as small shocks or visible sparks.

Examples: o Shoes with rubber soles:

Friction is caused by walking. A person develops an electrical charge different from the ground. A discharge occurs when an object (e.g. door handle) is contacted.

o Synthetic fiber clothing:

Discharge is audible and is visible in darkness.

Influence of electrostatic charges on components

If a person with an electrostatic charge touches a component, a discharge over a connection of an IC or semiconductor element may occur. The resulting voltages may damage the component.

Critical situations can occur while repairing or testing components if they are lying on a more or less conductive surface (e.g. table top) and a person with an electrostatic charge touches them. A discharge through a critical component connection can also occur in this case.

ESD = Electrostatic Sensitive Device

The dryer the air the greater the risk of electrical discharge caused by friction.

Eectrostatic charges are less likely to build up in humid air, particularly when saturated with water

vapor.

The probability of ESD phenomena is therefore particularly great in a northern hemisphere winter

in centrally heated rooms where the humidity is low.

Page 23: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 A-9

91xx pH/Electrolyte Analyzers 1 Safety informationESD protection measures

Why is ESD protection so important today?

Formerly, current controlling semiconductors were mostly used (TTL, normal transistors, etc.).

Today the principle of current logic is virtually the only technology in use in MOS and CMOS components.

Currents generated by an electrostatic discharge (as much as several kV!) destroy the sensitive component inputs or cause hidden damage. This damage is generally not immediately detectable.

Another effect is that the clearances inside the ICs continuously become smaller. The internal wires become thinner and thinner, the allowable maximum input voltages become smaller and the effects of any discharges become more and more critical.

How can ESD protection be guaranteed?

A continuous discharge is required when working on an electronic assembly. Do this as follows:

o Use ESD wrist bands (special wrist bands connected to a protective ground).

o Repairs and tests on assemblies must only be conducted on tables with ESD mats connected to grounds or ESD wrist bands.

o Always pick up components at the edge (e.g. like a photo).

o Components must always be transported in ESD packages or appropriate storage or shipping containers (use original packaging!).

o Shoes with rubber soles or clothing of synthetic fibers must not be worn in workshops where electronic components are repaired.

o Use a humidifier if necessary to maintain optimum humidity in the work area.

o Do not touch assemblies or components with the hand after testing.

o Always transport and ship assemblies or components for repair in ESD protective packaging only. This will prevent further damage that may result in misinterpretation of the original cause of the fault.

Conclusion

Of course, not all circuit boards and electronic assemblies require such careful handling. An electrical board that only carries simple plug connectors does not require ESD packaging.

In cases of doubt always use ESD packaging.

The ESD mat consists of materials that have a very low, defined conductivity (1012 ohms).

The following materials prevent charges caused by friction from building up and protect the

component from damage.

o ESD mats

o ESD packages

o Shipping containers

Page 24: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

A-10 Service manual · Version 2.0

1 Safety information 91xx pH/Electrolyte AnalyzersESD protection measures

Page 25: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 A-11

91xx pH/Electrolyte Analyzers 2 SpecificationsTable of contents

Specifications

Specifications ............................................................................................................. A-13

Specifications .......................................................................................................A-13

9110 ................................................................................................................ A-13

9120/9130/9140/9180/9181 ........................................................................... A-13

Operating Parameters .........................................................................................A-13

9110 ................................................................................................................ A-13

9120/9130/9140/9180/9181 ........................................................................... A-14

Common operating parameter ..................................................................... A-14

Classifications ......................................................................................................A-15

Product data ...................................................................................................A-15

Approvals .......................................................................................................A-15

In this chapter Chapter 2

Page 26: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

A-12 Service manual · Version 2.0

2 Specifications 91xx pH/Electrolyte AnalyzersTable of contents

Page 27: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 A-13

91xx pH/Electrolyte Analyzers 2 SpecificationsSpecifications

Specifications

Specifications

9110

Sample types Whole blood, Serum, Plasma, QC solutions

9120/9130/9140/9180/9181

Sample types Whole blood, Serum, Plasma

Sample type Urine

Operating Parameters

9110

Sample types Whole blood, serum, plasma, aqueous and QC solutions

Sample device Syringe, sample cup, capillary, Roche MICROSAMPLER

Reported parameters Measuring range Resolution

pH 6.00 - 8.00 0.001 pH units

Table A-1 9110 reported parameters

Reported parameters Measuring range Resolution

Sodium (Na+) 40 - 205 mmol/L 0.1 mmol/L

Potassium (K+) 1.5 - 15 mmol/L

(0.8 - 15 mmol/L dialysate)

0.01 mmol/L

Chloride (Cl–) 50 - 200 mmol/L 0.1 mmol/L

Calcium (Ca2+) 0.2 - 5.0 mmol/L 0.01 mmol/L

Lithium (Li+) 0.1 - 6.0 mmol/L 0.001 mmol/L

Table A-2 9120, 9130, 9140, 9180, 9181 reported parameters (1)

Reported parameters Measuring range Resolution

Sodium (Na+) 1 - 300 mmol/L 1.0 mmol/L

Potassium (K+) 4.5 - 120 mmol/L

(60 - 120 mmol/L

with add. dilution)

0.1 mmol/L

Chloride (Cl–) 1 - 300 mmol/L 1.0 mmol/L

Table A-3 9120, 9130, 9140, 9180, 9181 reported parameters (2)

Calcium and Lithium are not measured in urine samples.

Lithium is not measured in dialysate samples.

Page 28: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

A-14 Service manual · Version 2.0

2 Specifications 91xx pH/Electrolyte AnalyzersSpecifications

Sample size 25 μL

Analysis time 70 seconds

Sample rate o 40 per hour with printout

o 50 per hour without printout

pH Sensor pH sensitive flow-through glass capillary electrode in transparent acrylic plastic

Reference System Open liquid junction, flow-through electrode

Calibration o Fully automatic

o 1 point with each sample

o 2 point every 3 hours

Warm-up time 10 minutes

9120/9130/9140/9180/9181

Sample types Whole blood, serum, plasma, urine, dialysate (acetate or bicarbonate)

Sample device Syringe, sample cup, capillary, Roche MICROSAMPLER

Sample size 95 μL

Analysis time 50 seconds

Sample rate o 45 per hour with printout

o 60 per hour without printout

Sodium (Na+) sensor Ion-selective, flow-through, glass capillary electrode

Potassium (K+) sensor Ion-selective, flow-through, liquide membrane electrode

Chloride (Cl–) sensor Ion-selective, flow-through, liquide membrane electrode

Calcium (Ca2+) sensor Ion-selective, flow-through, liquide membrane electrode

Lithium (Li+) sensor Ion-selective, flow-through, liquide membrane electrode

Reference system Open liquid junction, flow-through electrode

Calibration o Fully automatic

o 1 point with each sample

o 2 point every 3 hours (9180/9181 every 4 hours)

Warm-up time 10 minutes

Common operating parameter

Standby mode Suspends calibrations

Temperature Room temperature, 15 - 32°C, 60 - 90 °F

Page 29: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 A-15

91xx pH/Electrolyte Analyzers 2 SpecificationsSpecifications

Humidity maximum 85 % relative humidity, non-condensing

Data management Quality control memory storage, 3 levels, 35 values; calculation of mean, standard deviation and coefficien of variation (CV)

Diganostic programs user-controlled diagnostics, YES/NO operation via the display

Electronics Microprocessor-controlled

Display LCD dot-matrix, 2 lines, 16 characters per line

Printer Integral thermal printer, 16 characters width

Autosampler (9181 only) Integral turntable, 18 positions, 2 mL or 0.25 mL sample cups

Interface RS232C serial port

DataLink Data link to COMAPCT 2/3 (9180/9181)

Electrical requirements 100 - 240 VAC, 50/60 Hz (90 - 250 VAC for 9110), self adjusting

Nominal power consumption 50 W

Dimensions

Classifications

Product data

Safety category I

Device type B (according to ÖVE-MG/EN 60601-1,

IEC 601-1)

Mode of operation continuous operation

Protection classification IP 20

Explosion protection the device is not designed for operation in explosive environments.

Approvals

CSA, CE, FCC Class B

CLIA Complexity Category: Moderate

Test System Code: 04739

Height 335 mm (12.2")

Width 315 mm (12.4")

Depth 295 mm (12.0")

Weight approx. 6 kg (13 lbs)

Table A-4 Instrument dimensions

Page 30: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

A-16 Service manual · Version 2.0

2 Specifications 91xx pH/Electrolyte AnalyzersSpecifications

FDA 510(k) numbers: K932642 (9110)

K961458 (9180)

K972673 (9181)

Analyte Codes: Na+: 5805

K+: 4910

Cl–: 1018

Ca2+: 1004

Li+: 3712

Special requirements for IVDproducts

EN 61010-2: 2001-12

EN 61010-2: 2002-01

(9180 SN > 12001 only)

Data subject to change without notice. Technical information is supplied for general informational

purposes only!

Page 31: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 A-17

91xx pH/Electrolyte Analyzers 3 Interface specificationsTable of contents

Interface specifications

Interface specifications ..............................................................................................A-19

Interface information ..........................................................................................A-19

Software .........................................................................................................A-19

Example data string information ....................................................................... A-20

Automatic Calibration Report ...................................................................... A-20

Serum sample report .....................................................................................A-20

Data Link Information (9180/9181 only) .......................................................... A-20

In this chapter Chapter 3

Page 32: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

A-18 Service manual · Version 2.0

3 Interface specifications 91xx pH/Electrolyte AnalyzersTable of contents

Page 33: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 A-19

91xx pH/Electrolyte Analyzers 3 Interface specificationsInterface specifications

Interface specifications

Interface information

The 9110, 9120, 9130, 9140, 9180 and 9181 analyzers are equipped with a standard serial interface output. This interface output is intended to be used with standard commercially available computer systems.

The data transmitted through the serial interface port employs the ASCII code.

The serial interface is terminated on the rear cover with a 9-pin male DB-9 connector.

The signal levels are as follows:

o Binary 1 = -12 V to -3 V

o Binary 0 = +3 V to +12 V

Two stop bits follow the eight data bits to complete the 10 bit word.

The baud rate is set at 9600 Baud fixed.

The maximal recommended cable length is 40 feet.

The pin assignment is as follows:

o pin 1...sample ground...GND

o pin 2...receive data....RxD

o pin 3...send data.......TxD

o pin 4...NC

o pin 5...signal ground...GND

o pin 6...NC

o pin 7...RTS (not used)

o pin 8...CTS(not used)

o pin 9...NC

(NC=Not Connected)

Software

The patient sample data is sent at the end of each measurement, the calibration report is sent at the end of each calibration.

The interface is always on, independent of the printer settings; the data is always sent.

Higher discharge current

A higher discharge current can be expected when using the serial interface. This must be checked by

suitably qualified personnel, depending on the local regulations.

Note

The arrow up (e.g. out of normal range) is sent as HEX 18 (↑), the arrow down as HEX19 (↓) and

the ° (degree) is sent as HEX1A (→).

Page 34: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

A-20 Service manual · Version 2.0

3 Interface specifications 91xx pH/Electrolyte AnalyzersInterface specifications

Example data string information

Automatic Calibration Report

<sx>* AVL 9130*<cr><crlf>ELECTROLYTE ISE<cr><crlf> 03JAN92

10:51crcrlf*CALIBR REPORT*<cr><crlf><cr><crlf>

DailyMaintenance<cr><crlf>Performed Last: <cr><crlf> 02JAN92 10:35

<cr><crlf><cr><crlf> Standard A<cr><crlf>Na = -112mV (3)<cr><crlf>

K = -1392mV (3)<cr><crlf>Cl = -106mV (3)<cr><crlf><cr><crlf>

Difference A-B<cr><crlf>Na = 1402mV ( )<cr><crlf>K = 1032mV ( )

<cr><crlf>Cl = -1006mV ( ) <cr><crlf><cr><crlf>Fluid Pack:<cr><crlf>

68% Remaining<cr><crlf><cr><crlf><ex>

Serum sample report

<sx>* AVL 9130*<cr><crlf>ELECTROLYTE ISE<cr><crlf> 03JAN92

10:59<cr><crlf><cr><crlf> Name: ..........<cr><crlf> ..........

<cr><crlf>Sample: SERUM<cr><crlf><cr><crlf> Sample No.13

<cr><crlf><cr><crlf>Na= 159soh mmol/L<cr><crlf>K = 5.4 mmol/L

<cr><crlf>Cl= 122soh mmol/L<cr><crlf><cr><crlf>*PERFORM DAILY *

<cr><crlf>*MAINTENANCE ! *<cr><crlf><cr><crlf><ex>

Data Link Information (9180/9181 only)

The data link with the COMPACT 2/3 blood gas analyzer allows to combine ISE results with pH/blood gas results on one printout. If Ca2+ is activated on the 9180/9181, a pH-corrected Ca2+ value will be calculated and printed on the combined sample report.

For connection of the 9180/9181 analyzer to the COMPACT 2/3 analyzer, the optional Interface Kit is required. To install the kit, first turn both instruments off. Connect the interface filter provided in the kit to the RS232 port on the 9180/9181. Then connect one end of the cable to the interface filter, the other end to the COM 2 port on the COMPACT 2/3. On the COMPACT 2/3, select 9180 under the COM 2 interface options.

e See COMPACT 2/3 Operator's Manual for details.

Page 35: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 A-21

91xx pH/Electrolyte Analyzers 4 Description of modulesTable of contents

Description of modules

Insert introducing text here.

Description of modules ............................................................................................A-23

Mechanical assemblies ........................................................................................A-23

Front door assembly ......................................................................................A-23

Needle unit assembly .....................................................................................A-23

Electrode holder assembly ............................................................................ A-24

Peristaltic pump assembly ............................................................................. A-24

Solenoid assembly .........................................................................................A-25

Printer assembly ............................................................................................A-25

Rear panel assembly ......................................................................................A-25

SBC board ......................................................................................................A-26

Lamp board ....................................................................................................A-27

Display board .................................................................................................A-27

Fluidic Module ....................................................................................................A-27

Standard A ....................................................................................................A-27

Standard B .....................................................................................................A-27

Standard C (not on 9110) ............................................................................. A-27

Cleaning solution (9110 only) ...................................................................... A-28

Reference solution .........................................................................................A-28

Waste liquids ..................................................................................................A-28

Main tube set (all models)/ Intake tube set (9110 only) ............................. A-28

Sample Contacts (9110 only) ........................................................................ A-29

Electronics ............................................................................................................ A-33

ISE SBC card ..................................................................................................A-33

Lamp board ....................................................................................................A-34

Display board .................................................................................................A-35

In this chapter Chapter 4

Page 36: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

A-22 Service manual · Version 2.0

4 Description of modules 91xx pH/Electrolyte AnalyzersTable of contents

Page 37: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 A-23

91xx pH/Electrolyte Analyzers 4 Description of modulesDescription of modules

Description of modules

Mechanical assemblies

Front door assembly

For the 9110, 20, 30, 40 and 80 analyzers, the front door can be removed by moving the analyzer to the edge of a work surface so the analyzer door, when opened, will extend past the edge of the work surface. With one hand, hold the analyzer door near the right side hinge pin and, with the other hand, gently apply pressure to the middle rear area of the door. This will allow the right hinge pin to clear the retaining hole in the main chassis. The door can then be removed from the analyzer.

The 9181 front panel can be removed by tilting it slightly away from the analyzer and lifting it straight up. The plastic window can also be replaced in the field by gently pressing from the rear of the window and snapping the plastic window out toward the front of the analyzer door. The door magnets cannot be replaced in the field as they are glued in place with conductive adhesive.

e See the 91xx Spare Parts List for a detailed reference of the parts of this assembly.

Needle unit assembly

To remove the 9110 fill port for replacement or cleaning, grasp the white tab on the fill port and rotate the bayonet-style connector counter-clockwise. Gently pull the fill port forward to remove. Replacement can be accomplished in the reverse order. During removal, be careful not to bend the needle.

The 9110 fill port assembly is desinged for easy removal without the use of tools, for cleaning, changing tubing or replacement. First, disconnect the sample intake tube on the left side of the fill port assembly from the sample preheater. Grasp the fill port assembly and pull it forward carefully. Disconnect the wire from the needle arm. To reassemble, reconnect the wire, press the fill port assembly into place and reconnect the tubing.

The 9110 wash port located on the underside of the flap can be removed by appying pressure to one side to overcome the snap fit. To reinsert, align wash port and apply pressure to snap into place.

The 9120, 30, 40, 80 and 81 sample needle and fill port are both designed for easy replacement by the user. To remove the fill port assembly for replacement, pull the fill

After changing any part of the analyzer always perform a calibration of the analyzer as well as a full

QC measurement. QC measurements must be performed in their entirety (i.e., all three QC levels

must be measured). Omitting QC measurements or ignoring QC measurement results may lead to

incorrect patient measurements, which may result in incorrect clinical decisions, possibly

endangering the patient's health.

Due to the update of the 9180 Electrolyte Analyzer according to IVDD compliance a new shielded

front door "FRONT DOOR 9180" is available and compatible with any 9180 Electrolyte Analyzer

(old type SN < 11999 and new type SN > 12001) and with 9120, 30, 40 Electrolyte Analyzers.

Further information: see the 91xx Spare Parts List.

Page 38: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

A-24 Service manual · Version 2.0

4 Description of modules 91xx pH/Electrolyte AnalyzersDescription of modules

port holder towards the front for easy access. Then, press the two plastic tabs on the fill port to allow removal of the fill port from the fill port holder. The sample needle can also be removed and is supplied as a one-piece assembly including the intake tubing.

To remove the 9120, 30, 40, 80 and 81 sample needle, grasp the needle near the white holder block and pull up to unsnap the sample needle from its position. To replace the assembly, align a replacement sample needle and press the replacement assembly into place.

Two (9110) or three screws (9120, 30, 40, 80, 81) located on the left side of the assembly hold the needle unit assembly in place. To access the two rear screws, turn power off and remove the rear panel. The rear panel can be moved away from the unit for easier access by unplugging the wire harness from the power supply. The SBC board can now be removed by removing the screws which secure the SBC board and disconnecting the cables on the SBC board.

Disconnect the connector to the lamp board which supplies the Door Open signal. The two rear screws which secure the needle unit can now be removed using a Philips screwdriver (9110) or an Allen wrench (9120, 30, 40, 80, 81). Then, remove the front securing screw (n./a. on the 9110), the sample intake tubing and the reagent supply tubing (red tag) from the fill port. The needle unit can now be removed from the front of the analyzer for replacement.

e See the 91xx Spare Parts List for a detailed reference of the parts of the needle unit.

Electrode holder assembly

To remove the electrode holder assembly from the analyzer, first unplug the sample intake tubing on the right side of the module, then the preheater cable (9110 only) or the sample sensor cable (9120, 30, 40, 80 and 81), then remove the reference solution tubing connector and last, take out the waste tube (green tag) at the left side of the electrode holder. The electrode holder can now be removed by pressing the two plastic tabs (right and left side of the module) and sliding the module forward. Press the plastic tabs again to release the second detent and completely remove the module from the analyzer. The sample preheater or the sample sensor and/or the left side electrode holder can be removed and replaced by removing the screws on the under side of this module.

e See the 91xx Spare Parts List for a detailed reference of the parts of this assembly.

Peristaltic pump assembly

Replacement of the roller assembly of the peristaltic pump can be accomplished by removing the pump tubing and firmly pulling the roller assembly toward the front. To replace the roller assembly, align the roller and the flat on the motor shaft and press the replacement roller into position.

To remove the motor, first remove the rear panel assembly and SBC Board.

e For details, see Rear panel assembly on page A-25, SBC board on page A-26.

Now, the electrical connector on the lamp board can be accessed to unplug the motor connector. The motor can now be removed by removing four screws located on the front of the housing near the peristaltic pump.

Page 39: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 A-25

91xx pH/Electrolyte Analyzers 4 Description of modulesDescription of modules

Solenoid assembly

Removal of each solenoid is identical. Each solenoid valve has a removable pressure piece which is held in place by a solenoid cap. To remove the pressure piece, locate the arrow on the solenoid cap and remove the solenoid cap by sliding it off the solenoid shaft in the direction of the arrow. This exposes the pressure piece which can now be removed from the solenoid shaft. To replace pressure piece and cap, energize the solenoid as described in TEST PINCH VALVES on page B-67. This extends the solenoid shaft to the outermost position for ease in replacing pressure piece and cap. Removal of the pressure piece and cap is required prior to removing the solenoid and allows for easier replacement of the tubing under each solenoid.

To remove the solenoid assembly, the rear panel and the SBC Board (described in Rear panel assembly on page A-25 and SBC board on page A-26) must be removed first.

e For details, see Rear panel assembly on page A-25, SBC board on page A-26.

Each solenoid has an electrical connection to the lamp board which must be unplugged prior to removal of the solenoid. At this point, remove the two screws on the front panel to remove the solenoid assembly.

Printer assembly

The printer assembly is designed to allow for easy removal by the user which can be accomplished without removal of the electrical power to the analyzer. Slide the paper tray forward to allow access to the printer, tear the paper roll and completely remove it together with the paper tray. Slide fingers under the printer assembly and pull the printer forward. This will disengage the printer from the interconnector and enable removal of the assembly from the front of the analyzer. Removal of the printer should be performed for replacement and for removal of a paper jam.

To replace the printer, locate the printer slide and insert the printer assembly. Press firmly into place to ensure electrical connection of the printer.

Rear panel assembly

To remove the rear panel assembly, ensure that the power cord has been disconnected from the analyzer. Remove the four corner screws to expose the rear panel. The power supply module located on the rear panel assembly can now be removed by removing the three screws securing the circuit board and by disconnecting the wiring

Note

The 9110 R-solenoid incorporates two heat sinks and is not interchangeable with the other

solenoids.

Note

Never attempt to dislodge paper from the printer with a paper clip or similar object to avoid

damage to the print head or printer platen.

Due to the update of the 9180 Electrolyte Analyzer according to IVDD compliance a new printer

assembly "PRINTER MODULE, ISE" is available and compatible with 9110 pH Analyzers and

with any 9120, 30, 40, 80 and 81 Electrolyte Analyzer.

e Further information: see the 91xx Spare Parts List.

Page 40: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

A-26 Service manual · Version 2.0

4 Description of modules 91xx pH/Electrolyte AnalyzersDescription of modules

interconnections. The main power receptacle can also be replaced by removing the two screws holding this assembly.

SBC board

After the rear panel has been removed, the SBC Board is accessible for removal. This circuit board can be removed by removing the seven screws securing the board to the housing and by disconnecting electrical connectors from the power supply module, display board and lamp board.

e For details, see Rear panel assembly on page A-25.

The electrode push pins can now be replaced by simply pulling the connecting pin from its socket. When installing the board, make sure to place the plastic washers under the screw heads.

Due to the update of the 9180 Electrolyte Analyzer according to IVDD compliance a new power

supply module "POWER SUPPLY 9180 >SN 12001" is available and compatible with 9180

Electrolyte Analyzers serial number > 12001 only.

The former power supply "MODULE POWER SUPPLY, ISE" is still available and has to be used

for 9180 Electrolyte Analyzers SN < 11999, for 9110 pH Analyzers and for 9120, 30, 40 and 81

Electrolyte Analyzers.

e Further information: see the 91xx Spare Parts List.

After changing the main board or the power supply always perform a measurement of protective

ground resistance to ensure a safe connection of the system ground. Always measure with the power

cord attached to the analyzer. The resistance measurement has to be done between the ground pin

of the power plug and several ground connectors placed on each of the analyzer's electronic

components such as power supply or main board and shielding. The total resistance should not

exceed 0.5 ohms. Be sure to consider the resistance of the ohmmeter's connecting cables.

Due to the update of the 9180 Electrolyte Analyzer according to IVDD compliance a new SBC

board "MAINBOARD 9180 >SN 12001" is available and compatible with 9180 Electrolyte

Analyzers serial number > 12001 only. The former SBC board "MAINBOARD 9180 <SN 11999"

is still available and has to be used for 9180 Electrolyte Analyzers SN < 11999.

Further information: see the 91xx Spare Parts List.

After changing the main board or the power supply always perform a measurement of protective

ground resistance to ensure a safe connection of the system ground. Always measure with the power

cord attached to the analyzer. The resistance measurement has to be done between the ground pin

of the power plug and several ground connectors placed on each of the analyzer's electronic

components such as power supply or main board and shielding. The total resistance should not

exceed 0.5 ohms. Be sure to consider the resistance of the ohmmeter's connecting cables.

Note

Due to the update of the 9180 Electrolyte Analyzer according to IVDD compliance the printouts,

display messages and data links sequences of the 9180 Electrolyte Analyzers serial number > 12001

will show "Roche" instead of "AVL". This also applies to 9180 Electrolyte Analyzers serial number

< 11999 in case of replacing the main board "MAINBOARD 9180 >SN 12001" or the program set

"PROGRAM SET 9180".

Page 41: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 A-27

91xx pH/Electrolyte Analyzers 4 Description of modulesDescription of modules

Lamp board

The rear panel and SBC board must be removed to gain access to the lamp board. Remove one screw and washer which hold the board in place and remove the electrical connectors to remove this board.

Display board

The rear panel and SBC board must be removed to gain access to the display board. Remove three screws and 50 degree spacers and unplug the preheater or sample sensor cable from the front and the ribbon cable to the SBC board to remove the display board.

Fluidic Module

Standard A

Standard A is drawn to the electrode module by vacuum provided by the peristaltic pump.

9110 pH Analyzers: When Standard A is to be aspirated into the electrode module, solenoid valve A and R are opened. Standard A is then drawn from the ISE SnapPakTM until Sample Contact 1 detects Standard A bridging the T-piece and Sample Contact 1. At this point, solenoid valve V is opened and solenoid valve A is closed. The standard is pumped up the tubing, through the wash portand into the sample needle until Sample Contact 2 detects the standard. The peristaltic pump slows down and the standard moves into the electrode module until detected by Sample Contact 3 (bridge between sample preheater tube and reference electrode). Then, the pump stops.

9120, 30, 40, 80, 81 ElectrolyteAnalyzers:

When Standard A is to be aspirated into the electrode module, solenoid valve A is opened and solenoid valves B, V and R are closed. Standard A is then drawn from the ISE SnapPakTM, to the fill port, through the sample needle and is sensed by the sample sensor. At this point, solenoid valve V is opened and solenoid valve A is closed as the peristaltic pump continues to pump Standard A into the electrode module. As the trailing edge of the Standard A sample is sensed by the sample sensor, the peristaltic pump stops.

During the time Standard A is aspirated into the electrode module, the reference housing is pressurized due to the peristaltic pump providing pressure to the reference solution line and solenoid valve R being closed. An amount of reference solution is forced out through the reference junction to provide contact to the sample.

Standard B

The process for Standard B is identical to that of Standard A with the exception of the operation of solenoid valve B. This solenoid valve is operated in reverse order of Standard A aspiration.

Standard C (not on 9110)

The process for Standard C is identical to that of Standard A with the exception of the operation of solenoid valve C. This solenoid valve is operated in reverse order of Standard A aspiration.

Page 42: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

A-28 Service manual · Version 2.0

4 Description of modules 91xx pH/Electrolyte AnalyzersDescription of modules

Cleaning solution (9110 only)

The cleaning routine within the 9110 pH Analyzer is automatic with cleaning solution being aspirated through the system every 40th sample or every 24 hours. Automatic operator-initiated cleaning can also be performed, by selecting ELECTRODE CLEANING from the main menu. The cleaning routine is very similar to the aspiration of Standard A or B with solenoid valve C being opened to allow cleaning solution to be aspirated through the fluidic path. The timing of the cleaning solution routine allows for a larger volume of cleaning solution to be aspirated than the Standard A or B cycles to ensure that cleaning solution covers the electrodes, wash port, fill port, sample needle and preheater during the cleaning cycle.

Reference solution

The reference housing is filled automatically using the second winding of the peristaltic pump and solenoid valve R. The reference solution connector allows for the reference housing tubing to be plugged into the reference solution circuit. As the peristaltic pump aspirates reference solution, solenoid valve R is opened to allow excess solution to be pumped into the reference return line of the ISE SnapPakTM.

A check valve is placed in the return line to ensure that the KCl output is constant (9110 only).

Waste liquids

Calibration standard and sample waste are pumped out of the left side of the electrode module (green-banded tubing) through the peristaltic pump to the waste line of the ISE SnapPakTM. An internal one-way valve is incorporated into the waste line of each ISE SnapPakTM to prevent any waste products from leaking out of the ISE SnapPakTM.

Main tube set (all models)/ Intake tube set (9110 only)

The main tube set is supplied as a pre-cut tubing harness with the fluidic interconnections preassembled. Replacement of the tube set should be performed annually. For removal of the tubing, first remove the ISE SnapPakTM from the reagent compartment. This will allow access for removal of the white TPR block which is the main interconnection between the ISE SnapPakTM and the tubing. This block can be removed by grasping the front surface and pulling forward. Next, remove tubing from each solenoid valve. Follow the procedure outlined in Section 4.1.5 for disassembly of the solenoid valves to enable fluidic tubing to be easily removed at each solenoid.

e For details, see Rear panel assembly on page A-25, SBC board on page A-26.

Pull the two pump windings at the peristaltic pump tube set off the pump roller and remove the white reference solution tube connector. Remove the white TPR reference solution block by grasping the front of the block and pulling forward. The three tube connections are labeled with green, blue and red tags. All remaining tubes can now be removed and the entire tubing harness can be discarded.

Fit the replacement tubing harness and assemble in reverse order as disassembly. Reference the Tube Diagram on the next page for correct tube locations.

Intake Tube Set (9110 only) The intake tube set comprises three pieces of tubing. A short piece (red tag) fits between the white T-piece and the standard contact. The long piece can be positioned between the standard contact and the wash port. The remaining tube can be fitted between the sample needle and the preheater tube.

Page 43: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 A-29

91xx pH/Electrolyte Analyzers 4 Description of modulesDescription of modules

Sample Contacts (9110 only)

The 9110 fluid detection is based on the principle of measuring the impedance in a sample contact path. There are three sample contacts in the 9110 pH Analyzer. Sample Contact 1 senses standard solution draws.

Page 44: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

A-30 Service manual · Version 2.0

4 Description of modules 91xx pH/Electrolyte AnalyzersDescription of modules

Figure A-1 9110 fluidic diagram

AB

FLU

ID P

ACK

REF.

RET

URN RE

F

WA

STE

CLEA

NIN

G

FLU

ID

CON

NEC

TOR

PERI

PU

MP

FILL

PORT

SAM

PLE

PRO

BE

SAM

PLE

PRO

BE

ASS

EMBL

Y

AIR

VEN

T

ELEC

TRO

DE

ASS

Y

REFE

REN

CECO

NN

ECTO

R

SOLE

NO

ID R

GRE

EN

REF

pHPR

E

HEA

TER

BLU

E

SOLE

NO

ID V

SOLE

NO

ID A

SOLE

NO

ID B

RED

YELL

OW

DO

TYE

LLO

W D

OT 9

110

FLU

IDIC

DIA

GRA

M

SOLE

NO

ID C

CHEC

K VA

LVE

Page 45: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 A-31

91xx pH/Electrolyte Analyzers 4 Description of modulesDescription of modules

Figure A-2 9120, 9130, 9140 fluidic diagram

AB

FLU

ID P

ACK

REF.

RET

URN RE

F

WA

STE

NO

T U

SED

C

FLU

ID

CON

NEC

TOR

PERI

PU

MP

FILL

PORT

SAM

PLE

PRO

BE

SAM

PLE

PRO

BE

ASS

EMBL

Y

AIR

VEN

T

ELEC

TRO

DE

ASS

Y

REFE

REN

CECO

NN

ECTO

R

SOLE

NO

ID R

GRE

EN

REF

Na

KCl Ca

BLU

E

SOLE

NO

ID V

SOLE

NO

ID A

SOLE

NO

ID B

RED

YELL

OW

DO

TYE

LLO

W D

OT 9

120,

913

0, 9

140

FLU

IDIC

DIA

GRA

M

Page 46: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

A-32 Service manual · Version 2.0

4 Description of modules 91xx pH/Electrolyte AnalyzersDescription of modules

Figure A-3 9180, 9181 fluidic diagram

B

FLU

ID P

ACK

REF.

RET

URN RE

F

WA

STE

FLU

ID

CON

NEC

TOR

PERI

PU

MP

FILL

PORT

SAM

PLE

PRO

BE

SAM

PLE

PRO

BE

ASS

EMBL

Y

AIR

VEN

T

ELEC

TRO

DE

ASS

Y

REFE

REN

CECO

NN

ECTO

R

SOLE

NO

ID R

GRE

EN

REF

Na

K

BLU

E

SOLE

NO

ID V

SOLE

NO

ID A

SOLE

NO

ID B

RED

YELL

OW

DO

TYE

LLO

W D

OT

SOLE

NO

ID C

C A

Li Cl Ca

918

0/91

81 F

LUID

IC D

IAG

RAM

Page 47: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 A-33

91xx pH/Electrolyte Analyzers 4 Description of modulesDescription of modules

Electronics

ISE SBC card

Power circuits Power is supplied to this circuit board through connector JP3 from the power supply assembly. Test points are provided to measure supply voltages, and light emitting diodes D1 through D5 are turned on to indicate each supply voltage present on the card. Since each LED is connected in series after the fuse, a blown fuse results in the respective LED turned off.

Four additional voltage regulator circuits are mounted on the ISE SBC II Card. IC U2 and IC U3 provide -8 VDC and +8 VDC (9180/9181: -5 VDC and +5 VDC) and are used to provide the supply voltages for the analog circuitry. IC U1 develops the supply voltage used to operate the peristaltic pump and provide either +10 VDC or +22 VDC. The +22 VDC is supplied only for the FAST speed and +10 VDC is used for all other pump speeds. Signal FAST/SLOW determines which supply voltage is selected.

U5 switches the voltage between 22 VDC and 12.5 VDC (16 VDC on the 9110) which supplies the solenoids. When a solenoid is turned on, the voltage switches to 22 VDC for 0.5 sec, then it returns to 12.5 VDC (16 VDC on the 9110).

The circuit which includes Q7 and R5 provides necessary switching for the measuring chamber illuminator LEDs.

The following list identifies the test points, fuses and LEDs for each respective DC supply voltage:

Test point Voltage Fuse Rating LED

TP1 +24 VDC F3 1.25 AT D3

TP2 +5 VDC F1/F2 2.0 AT/1.6 AT D1/D2

TP3 -12 VDC F4 0.3 AT D4

TP4 +12 VDC F5 0.3 AT D5

TP5 Ground

Table A-5 Voltage test points, LEDs

Test point Voltage

TP6 -8 VDC (9180/9181: -5 VDC)

TP7 +22/12 VDC

TP8 +8 VDC (9180/9181: +5 VDC)

TP9 Na+ channel (pH channel on the 9110)

TP10 K+ channel (not used on the 9110)

TP11 Ca2+/Cl– channel (Li+ 9180/9181 only; not used on the 9110)

TP12 Temperature

TP13 +10.000 V ref. voltage

TP14 -2.500 V ref.voltage (-2.500 V on the 9110)

TP15 Sample sensor (sample contact voltage on the 9110)

TP16 +22/12.5 VDC (+22/+16 VDC on the 9110)

Table A-6 Voltage test points

Page 48: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

A-34 Service manual · Version 2.0

4 Description of modules 91xx pH/Electrolyte AnalyzersDescription of modules

Valve drivers The solenoids are activated by transistors Q9 - Q13. To open the solenoid, a 0.5-second pulse of 22 V is applied, then the voltage drops down to 12.5 V (16 V on the 9110) to hold the solenoid open.

Pump motor driver/cover sensor The pump driver circuit IC U9 and IC U10 provide the drive and control for the peristaltic pump. Signals PUMPEN, Pump Enable and PUMPSTEP, Pump Step Pulses enable IC U10 to provide the necessary control to driver IC U9 which supplies pump drive signals PHASE A through PHASE D to control peristaltic pump operation.

IC U11 operates with signal DCLOSED from the sample door light gate to provide signal DOOR which detects the status of the sample door (open or closed).

Input amplifiers IC U13, U14 and U15 (9110/9180/9181: U13 only) provide high to low input impedance matching for each Electrode input. Offset generator IC U12D and amplifiers IC U12A, U12B and U12C provide the necessary gain and offset for each electrode channel. TP9, TP10 and TP11 provide test points for the Na+ (TP9: pH on the 9110), K+ and Cl–/Ca2+/Li+ channels respectively.

9110 only: IC U300 is used as the input amplifier for the reference electrode, IC U12A is used for the subtraction of the reference electrode signal from the pH electrode signal. For sample detection, the reference electrode is switched to ground (relay K300) during aspiration.

Temperature sensor This circuit is used in the 9110 pH Analyzer to amplify the signal of the temperature sensor in the preheater. This is an optional circuit which is used in the 9140, 9180 and 9181.

The circuit creates a temperature-dependent voltage which is converted by software to degrees Celsius. R54 can be adjusted for correct temperature display.

Analog channel selector and A/Dconverter

Analog inputs from the input amplifiers are selected via signals ADSEL0 through ADSEL2 and are multiplexed by IC U19. This multiplexed signal is then fed to the A/D Converter IC U20. IC U21 supplies the necessary reference voltage (2500mV) for the A/D Converter and can be measured at TP14.

Microprocessor Microprocessor IC U31, EPROM IC U25 (9180/9181: U40) and MOSTEK Real Time Clock/RAM IC U22 form the components of the central processing system. IC U23 Address Demultiplexer, IC U24 Programmable Array Logic (PAL) and IC U28 provide serial data buses (9180/9181: U23, U24 and U28 not present). IC U34 through IC U38 convert serial data to parallel data and provide printer data, display data, solenoid valve control and peristaltic pump control (9110: U34 not present). This fashion of serial to parallel data conversion is used to prevent analyzer lock-up by peripheral devices. The RS232 serial interface is a direct input/output from the microprocessor utilizing interface driver and receiver U29 and U30.

SnapPak Sensor (9180/9181only)

Optical detector ISO1 provides a signal indicating if the ISE SnapPak™ is in place. This signal is fed to U41 which converts the light gate input to a digital logic output indicating SnapPak position (IN or OUT).

Lamp board

The lamp board contains the LED array used to illuminate the measuring chamber and the interconnectors used for all solenoids, the peristaltic pump, door sensor and sample turntable (9181 only). A ribbon cable from the SBC Board supplies the electrical connection for each signal through connector JP6.

Page 49: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 A-35

91xx pH/Electrolyte Analyzers 4 Description of modulesDescription of modules

Display board

The display board contains the LCD display, sample contact circuit (9110 only) or sample sensor connector, the preheater (9110 only) and the YES and NO switches used for operation of the analyzer. Power and signal interconnection is accomplished via a ribbon cable from the SBC board through JP2 on the display board.

9110 only The sample contact circuit consists of a square wave generator U1B which generates a 500 Hz signal. U1A amplifies the signal, D2 and C3 rectify the square wave. The gain can be adjusted with R10. The sample contact output provides a DC voltage, which is proportional to the square wave amplitude. During sample aspiration, the preheater tube and the sample contact 1 tube are connected to the square wave via relay K1.

The heater control circuit controls two transistors which are placed in the preheater block. A thermistor provides temperature feedback to the setting of potentiometer R20. The signal difference drives the base of the heating transistors via buffer U2C. U2D shuts the system off in case the thermistor opens. U2B limits the maximum heating current during warm-up. U1C and attached circuit is used as an AC/DC converter to create negative voltage -VSS, using square wave generator U1B.

9181 only Display boards utilized in the 9181 analyzer also contain circuitry for both the needle mechanism motor control and position sensor. The autosampler "Table Present" signal is also buffered through U1 on this board.

Page 50: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

A-36 Service manual · Version 2.0

4 Description of modules 91xx pH/Electrolyte AnalyzersDescription of modules

Page 51: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 A-37

91xx pH/Electrolyte Analyzers 5 Electronic diagramsTable of contents

Electronic diagrams

Electronic diagrams ...................................................................................................A-39

9110 analyzer block diagram .............................................................................. A-39

9120, 9130, 9140, 9180 analyzer block diagram ................................................A-40

9181 analyzer block diagram .............................................................................. A-41

System interconnect - 9110, 9120, 9130, 9140 ...................................................A-42

SBC PCB - power supply - 9110, 9120, 9130, 9140 ...........................................A-43

SBC PCB - valve drivers - 9110, 9120, 9130, 9140 .............................................A-44

SBC PCB - pump motor driver / door detect - 9110, 9120, 9130, 9140 ........... A-45

SBC PCB - input amplifiers - 9110, 9120, 9130, 9140 .......................................A-46

SBC PCB - temperature circuit - 9110, 9120, 9130, 9140 ................................. A-47

SBC PCB - analog selector and A/D converter - 9110, 9120, 9130, 9140 ......... A-48

SBC PCB - microprocessor - 9110, 9120, 9130, 9140 ........................................A-49

SBC PCB - printer display drivers - 9110, 9120, 9130, 9140 ............................. A-50

Printer daughter PCB - 9110, 9120, 9130, 9140, 9180, 9181 ............................. A-51

Lamp PCB - 9110, 9120, 9130, 9140 .................................................................. A-52

Display daughter PCB - 9110 .............................................................................. A-53

Preheater PCB - 9110 ..........................................................................................A-54

Display daughter PCB - 9120, 9130, 9140 .......................................................... A-55

System interconnect - 9180, 9181 ....................................................................... A-56

SBC PCB - power supply - 9180, 9181 ............................................................... A-57

SBC PCB - valve drivers - 9180, 9181 ................................................................. A-58

SBC PCB - pump motor driver/door detect - 9180, 9181 ................................. A-59

SBC PCB - input amplifier - 9180, 9181 ............................................................ A-60

SBC PCB - temperature circuit - 9180, 9181 .....................................................A-61

SBC PCB - analog selector and A/D converter - 9180, 9181 ............................. A-62

SBC PCB - microprocessor - 9180, 9181 ............................................................ A-63

SBC PCB - microprocessor - 9180, 9181 ............................................................ A-64

In this chapter Chapter 5

Page 52: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

A-38 Service manual · Version 2.0

5 Electronic diagrams 91xx pH/Electrolyte AnalyzersTable of contents

SBC PCB - Fluid pack detector - 9180, 9181 .....................................................A-65

Lamp PCB - 9180, 9181 ......................................................................................A-66

Display daughter PCB - 9180 .............................................................................. A-67

Page 53: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 A-39

91xx pH/Electrolyte Analyzers 5 Electronic diagramsElectronic diagrams

Electronic diagrams

9110 analyzer block diagram

Figure A-4 9110 analyzer block diagram

Preamplifier

Ph Amp

not used not used

Analog Selector

andA/D Converter

Power Supply

+22 VDC / +10 VDC+8 VDC / -8 VDC

+22 VDC / +16 VDC

Interface Driver

RS232

90 - 250 V~

MAIN

50 - 60 Hz

Temperature Control

Microprocessor

Solenoid Valve Drivers

Power Receptacle

1.6 AT

1.6 AT

Display Module YES/NO Switches

Temp Sensor Sample Contacts

PrinterAssembly

Lamp Board

Power Supply Module

Fuse F1 2.5 A/250 V +5 VDC/+24 VDC+12 VDC/-12 VDC

Ph Electrode

Reference

ElectrodeGND

SBC BoardDisplay Board

Sample Contact1 / Preheater

P. Pump Motor

JP6

Sample Door

JP7

Valve A JP1

Valve B JP2

Valve C JP3

Valve V JP4

Valve R JP5

JP8

JP1

JP7JP3

JP1

Display DriverJP4

Printer DriverJP1

Peristaltic Pump DriverJP5

JP2

Page 54: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

A-40 Service manual · Version 2.0

5 Electronic diagrams 91xx pH/Electrolyte AnalyzersElectronic diagrams

9120, 9130, 9140, 9180 analyzer block diagram

Figure A-5 9120, 9130, 9140, 9180 analyzer block diagram

Preamplifier

Na Amp

K Amp Cl/Ca/Li Amp

Analog Selector

andA/D Converter

Power Supply

+22VDC / +10VDC+5VDC / -5VDC (9180)

+8VDC / -8VDC

+22VDC / +12.5VDC

Interface Driver

RS232

100 - 240 V~

MAIN

50 - 60 Hz

Fluid Pack Detector

Sensor

(9180 only)

Display Driver

Temperature Control(9140/9180 only)

Printer Driver

Microprocessor

Peristaltic Pump Driver

Solenoid Valve Drivers

Power Receptacle

1.6 AT

1.6 AT

Display Module YES/NO Switches

Temp Sensor

(9140/9180 only)Sample Sensor

PrinterAssembly

Lamp Board

Power Supply Module

Fuse F1 2.5A/250V +5VDC/+24VDC+12VDC/-12VDC

Na Electrode

KCl

Reference

ElectrodeGND

SBC Board

Display Board

Sample Sensor/Temp Sensor

P. Pump Motor

JP6

Sample Door

JP7

Valve A JP1

Valve B JP2

Valve C JP3

Valve V JP4

Valve R JP5

JP8

JP5

JP1

JP1

JP4

JP2

JP7JP3

JP1

Page 55: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 A-41

91xx pH/Electrolyte Analyzers 5 Electronic diagramsElectronic diagrams

9181 analyzer block diagram

Figure A-6 9181 anayzer block diagram

Preamplifier

Na Amp

K Amp Cl/Ca/Li Amp

Analog Selector

andA/D Converter

Power Supply

+22 VDC / +10 VDC+5 VDC / -5 VDC

+8 VDC / -8 VDC

+22 VDC / +12.5 VDC

Interface Driver

RS232

100 - 240 V~

MAIN

50 - 60 Hz

Fluid Pack Detector

Sensor

Display Driver

Temperature Control

Printer Driver

Microprocessor

Peristaltic Pump Driver

Solenoid Valve Drivers

Power Receptacle

1.6 AT

1.6 AT

Display Module YES/NO Switches

Temp Sensor Sample Sensor

PrinterAssembly

Lamp Board

Power Supply Module

Fuse F1 2.5 A/250 V +5 VDC/+24 VDC+12 VDC/-12 VDC

Na Electrode

KCl

Reference

ElectrodeGND

SBC BoardDisplay Board

Sample Sensor/Temp Sensor

P. Pump Motor

JP6

Sample Door

JP7

Valve A JP1

Valve B JP2

Valve C JP3

Valve V JP4

Valve R JP5

JP8

JP5

JP1

JP1

JP4

JP2

JP7JP3

JP1

Needle SensorPosition Buffer

Needle Motor Driver Probe Motor

Position Sensor

NeedleMechanismAssembly

JP1 JP1

JP

6JP

5

Turntable Assembly

Driver BDHome Sensor

Position Sensor

Bulkhead

BK5013BK5014

JP9

Ctrl LinesTurntable

Turntable Ctrl Lines

Needle Ctrl Lines

Page 56: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

A-42 Service manual · Version 2.0

5 Electronic diagrams 91xx pH/Electrolyte AnalyzersElectronic diagrams

System interconnect - 9110, 9120, 9130, 9140

Figure A-7 System interconnect - 9110, 9120, 9130, 9140

P8

AG

ND

DA

TA/I

NS

T-

PH

AS

EB

DD

[0..

7]

JP1

EZ

C10

DR

XH

A1

B1

A2

B2

A3

B3

A4

B4

A5

B5

A6

B6

A7

B7

A8

B8

A9

B9

A10

B10

MO

TOR

-

R90

30

12

DIS

PE

N

CH

EC

K

LAMP BOARD

PR

INT

VC

C

TO LAMP BOARD J

VA

LVE

B-

DO

T1-

CO

NTR

AS

T

DD

7

DG

ND

DISPLAY BOARD

THEY ARE TO ALLOW GROUNDS TO BE SEPARATED

TAC

HO

VA

LVE

C-

DO

T5-

RS

232

OU

T

10Amp TRACE - NO VIA's ALLOWED

P-1

2V

PRINTER

|MB6.SCH

ZD

50

05

F

SC

HE

MA

TIC

, MA

IN B

OA

RD

91

XX

19

Th

urs

day

, Ju

ne

27,

200

2

Title

Siz

eD

ocu

men

t N

um

ber

Rev

Dat

e:

She

et

of

DIS

PR

/W-

DD

4

R900-R903 DO NOT PHYSICALLY EXIST

JP3

26-6

0-40

70

1 2 3 4 5 6 7

R90

20

1

DC

LOS

ED

JP7

DB

9P

5 9 4 8 3 7 2 6 1

PV

CC

SYSTEM GROUND

|MB5.SCH

R90

42

70K

_1/4

W

1 2

mb1.sch

VC

C

PH

AS

EC

RS-232

|MB4.SCH

YE

S-

AMP 745061-4

PG

ND

VC

C

DD

1

A

DO

T2-

SA

MP

LE

|MB3.SCH

SAMPLEGND IS RUN INDIVIDUALLY

PV

CC

DD

6

VA

LVE

V-

DO

T6-

PG

ND

CUT TRACE AND INSERT R904

|MB2.SCH

RS

232

IN

AREA.

+24V

PG

ND

DD

3

VA

LVE

+

PG

ND

|MB1.SCH

AND CONNECTED AT THE SHIELD

PH

AS

ED

P+1

2V

R90

10

12

NOTE: FOR AVL ONLY

NO

-

DLE

D+

SG

ND

R90

00

12

LAM

P+

DO

T3-

|MB9.SCH

|LINK

DD

0

+5V

A

NOTE:

VA

LVE

R-

DO

T7-

SG

ND

DD

[0..

7]

|MB8.SCH

GN

D

DD

5

PH

AS

EA

TO DISPLAY BOARD JP2

SA

MP

LE

LED

-

TEM

P

MOLEX

JP2

626

50-1

1

|MB7.SCH

OV

ER

TE

MP

-

P+2

4V

DD

2

SS

EN

POWER

SULLINS

+10V

FE

ED

CTS

DO

T4-

JP5

CH

R26

G-N

D

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26

VA

LVE

A-

AMP

RTS

JP4

CH

R2

6G-N

D

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26

MATES WITH MOLEX 09-50-8073

Page 57: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 A-43

91xx pH/Electrolyte Analyzers 5 Electronic diagramsElectronic diagrams

SBC PCB - power supply - 9110, 9120, 9130, 9140

Figure A-8 SBC PCB - power supply - 9110, 9120, 9130, 9140

+C

6410

00uF

/16V

1 2

ZC

LIP

6

FU

SE

CLI

P

D4

LED

21

R7

560

12

C2

0.1u

F

12

ZC

LIP

2

FU

SE

CLI

P

TP5

UT3

5-01

7

1

D2

LED

21

C5

0.1u

F

12

ZC

LIP

3

FU

SE

CLI

P

"GND"

C3

0.1u

F

1 2

C23

0.1u

F

12

C20

0.1u

F

12

C4

0.1u

F

1 2

C1

0.1u

F

1 2

"+12V"

P-1

2V

C9

0.1u

F

1 2ZC

LIP

8

FU

SE

CLI

P

F1

2AT

12

C21

0.1u

F

12

PR

INTV

CC

C6

0.1u

F

1 2

D3

LED

21

TP2

UT3

5-01

7

1

D1

LED

21

R10

220

12

"-12V"

P+2

4V

TP1

UT3

5-01

71

C22

0.1u

F

12

D6

ICTE

.5

21

"+24V"

V-

-8V

VC

C

C8

0.1u

F

1 2

ZC

LIP

4

FU

SE

CLI

P

"+5V"

"0.300AT"

F5

0.30

0AT

12

-8V

P+1

2V

C14

0.1u

F

1 2

"+5V"

"0.300AT"

ZC

LIP

9

FU

SE

CLI

P

C16

0.1u

F

1 2

U2

LM79

082

3

1

VIN

VO

UT

C

C19

0.1u

F

12

R4

180

1 2

C15

0.1u

F

1 2

"-8V"

"1.25AT"

C17

0.1u

F

1 2

R1

1K

12

FA

ST/

SLO

W-

+5V

D5

LED

21

+C

2710

uF/1

6V

1 2

Q7

MJE

800

"1.6AT"

VC

C

V-

TP3

UT3

5-01

7

1

ZC

LIP

5

FU

SE

CLI

P

"2AT"

R2

1.8K

1 2

Q1

2N39

04

R3

1.2K

1 2

ZH

S1

HE

ATS

INK

mb2.sch

"+12V/F5"

F4

0.30

0AT

12

U3 LM

7805

13

2

VIN

VO

UT

C

+VS

LAM

P+

C69

0.1u

F

1 2

PR

INTV

CC

+

C28

10uF

/16V

12

"+22V/12V"

F3

1.25

AT

12

OPTIONAL

R9

220

12

"-12V/F4"

PV

CC

TP8

UT3

5-01

7

1

C70

0.1u

F

1 2

A

C12

0.1u

F

1 2

R6

560

12

+24V

ZC

LIP

10

FU

SE

CLI

P

C11

0.1u

F

1 2

C67

0.1u

F

1 2

C71

0.1u

F

1 2

TP4

UT3

5-01

7

1

C10

0.1u

F

1 2

C13

0.1u

F

1 2

+C

3022

00uF

/16V

1 2

PV

CC

F2

1.6A

T

12

+

C25

470u

F/2

5V

12

C24

0.1u

F

12C

680.

1uF

1 2

TP6

UT3

5-01

7

1

+C

2647

0uF

/25V

1 2

ZC

LIP

1

FU

SE

CLI

P

"+24V/F3"

ZC

LIP

7

FU

SE

CLI

P

+C

2922

00uF

/50V

1 2C

650.

1uF

1 2

V+

R8

1.2K

12

+5V

V+

TP7

UT3

5-01

7

1

"+5V/F2"

LAM

PE

N

U1

LM31

73

2

1

VIN

VO

UT

ADJ

C66

0.1u

F

1 2

ZD50

05F

SC

HEM

ATI

C, M

AIN

BO

AR

D 9

1XX

29

Thur

sday

, Jun

e 27

, 200

2

Title

Siz

eD

ocum

ent

Num

ber

Rev

Dat

e:S

heet

of

C18

0.1u

F

1 2

R10

31K

12

+24V

R5

3.9/

10W

12

C7

0.1u

F

1 2

"+5V/F1"

Page 58: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

A-44 Service manual · Version 2.0

5 Electronic diagrams 91xx pH/Electrolyte AnalyzersElectronic diagrams

SBC PCB - valve drivers - 9110, 9120, 9130, 9140

Figure A-9 SBC PCB - valve drivers - 9110, 9120, 9130, 9140

VA

LV

EC

-

Q13

MJE

800

+

C39

2.2

uF

/50V

12

OP

EN

3

Q12

MJE

800

SG

ND

VA

LV

EB

-

C32

0.1

uF

1 2

+

C40

2.2

uF

/50V

12

R14

1K

12

OP

EN

5

Q11

MJE

800

R20

1.5

K

1 2

R21

4.7

K

12

C31

0.1

uF

1 2R

15

1K

12

OP

EN

1

+24V

TP

16

UT35-0

17

1

Q10

MJE

800

"+22V/12.5V"

Q8

2N

3904

mb3.sch

R13

1K

12

A

OP

EN

2

U5

LM

317

32

1

VIN

VO

UT

ADJ Q9

MJE

800

+

C38

2.2

uF

/50V

12

VA

LV

E-

R12

1K

12

SG

ND

R18

200 1

%

1 2

OP

EN

4

TH

ER

M-

R17

180K

1 2

R19

1.8

2K

1%

1 2

VALVE DRIVERS

OV

ER

TE

MP

-+

C37

2.2

uF

/50V

12

R16

360K

12

VA

LV

EA

-

VA

LV

EV

-

R11

1K

12

ZD

50

05F

SC

HE

MA

TIC

, MA

IN B

OA

RD

91

XX

39

Th

urs

day

, Ju

ne

27,

200

2

Title

Siz

eD

ocum

ent

Num

ber

Rev

Date

:S

heet

of

ZH

S2

HE

ATS

INK

+

C36

2.2

uF

/50V1

2

VA

LV

E+

VA

LV

ER

-

Page 59: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 A-45

91xx pH/Electrolyte Analyzers 5 Electronic diagramsElectronic diagrams

SBC PCB - pump motor driver / door detect - 9110, 9120, 9130, 9140

Figure A-10 SBC PCB - pump motor driver / door detect - 9110, 9120, 9130, 9140

VC

C

PH

AS

EC

U10

L297

4 5 6 7 8 9 14

13

15

17

18

19

20

11

10

212

161

3

AIN

H1 B C

INH

2 D

SE

NS

E1

SE

NS

E2

VR

EF

CW

/CC

WS

TE

PH

ALF

/FU

LL

RE

SE

T

CO

NTR

OL

EN

AB

LE

GN

DV

S

OS

C

SY

NC

HO

ME

D12

1N

4004

21

D13

1N

4004

21

D9

1N

4004

21

DC

LO

SE

D

D11

1N

4004

21

SG

ND

SG

ND

C44

0.1

uF

1 2

R29

10K

1 2

SG

ND

PH

AS

EA

DLE

D+

VC

C

SG

ND

SG

ND

R27

100

1 2

A

+V

S

+ -V-

V+BBS

GND

U11

LM

311N

2 37

85

6

4

1

VC

C

ZS

2

DIP

20

ZD

50

05

F

SC

HE

MA

TIC

, MA

IN B

OA

RD

91

XX

49

Th

urs

da

y, J

un

e 2

7, 2

002

Title

Siz

eD

ocum

ent

Num

ber

Rev

Date

:S

heet

of

VC

C

SG

ND

PH

AS

ED

D7

1N

4004

21

SG

NDPUMP DRIVER

SG

ND

PH

AS

EB

SG

ND

DO

OR

R28

10K

1 2

SG

ND

C42

0.1

uF

1 2

R26

1

1 2

+V

S

R33

2.2

K

1 2

SAMPLE DOOR SWITCH DETECT

C45

0.1

uF

1 2

+C

46

100uF

/50V

1 2

VC

C

SG

ND

SG

ND

SAMPLE DOOR

VC

CV

CC

PU

MP

STE

P

SG

ND

R32

562 1

%

1 2

D10

1N

4004

21

R25

1

1 2

+22V/12V

+V

S

VC

C

C41

0.1

uF

1 2

R24

10K

1%

1 2

+C

47

10uF

/16V

1 2

SG

ND

mb4.sch

SG

ND

R23

22K

1 2

+V

S

SG

ND

C43

3.3

nF

1 2

R31

100K

12

U9

L293E

2 1 9

12

11

19

3 8 13

18

5 6 15

20

10

16

4 7 14

17

IN1

EN

AIN

2

IN3

EN

BIN

4

OU

T1

OU

T2

OU

T3

OU

T4

GN

DG

ND

GN

D

VS

S

VS

GN

D

SE

NS

E1

SE

NS

E2

SE

NS

E3

SE

NS

E4

+V

S

PU

MP

EN

D8

1N

4004

21

VC

C

SG

ND

SG

ND

ZS

1

DIP

20

R30

33K

1 2

D14

1N

4004

21

Page 60: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

A-46 Service manual · Version 2.0

5 Electronic diagrams 91xx pH/Electrolyte AnalyzersElectronic diagrams

SBC PCB - input amplifiers - 9110, 9120, 9130, 9140

Figure A-11 SBC PCB - input amplifiers - 9110, 9120, 9130, 9140

"Na/pH"

A

+8V

JP

202

HE

AD

ER

3

123

R47

165K

1%

IS

E102K

1%

pH

12

Cl/C

a/L

iOU

TR

65

1K

12

NOT STUFFED

+5V

+-

U12A

TLC

279C

N

321

11 4

ISE

CLOSED ISE

+8V

C51

0.1

uF

12

NOT STUFFED

ZP

RO

BE

5

R4C

R

OPEN pH

Na/pH

-8V

C300

0.2

2uF

1 2

pH 2-3

-8V

-8V

D15

1N

827

1 2

R300

100K

12

mb5.sch

- +

U13

LM

C6061

382

7

6

4

NOTE: PROBE THRU IN+ SHIELDED

Li

ISE 1-2

"Cl/Ca/Li"

ZP

RO

BE

4

R4W

W

FOR 9110

R46

68.1

K 1

%

12

or

ZP

RO

BE

3

R4W

W

OFFSET = -700mV

+-

U12D

TLC

279C

N

12

13

14

"K"

Na/p

HO

UT

C49 NOT STUFFED FOR 9110

R64

1K

12

C48

0.2

2uF

12

K

P5

PR

OB

E

+5V

Q300

2N

3904

R40

1K

1%

12

MASK.

R301

1K

1%

12

- +

U300 L

MC

6061

382

7

6

4

C50 NOT STUFFED

P3

PR

OB

E

P4

PR

OB

E

R39

1K

1%

12

+5V

Cl

-8V

R42

2.2

K

1 2

R302

18.2

K 1

%

1 2

+-

U12C

TLC

279C

N

109

8

R63

1K

12

SHOULD BE CLEAR OF SOLDER

OFFSET = -2200mV Na

R35

100K

12

D300

1N

4148

21

TP

9U

T35-0

17

1

Ref

NOTE: PROBE THRU IN+ SHIELDED

GUARD RING AND PROBE AREAS

OFFSET = -1100 mv pH T

P10

UT35-0

17

1

FOR 9110 ONLY

+-

U12B

TLC

279C

N

567

ZP

RO

BE

2

R4W

W

-6.2V

R36

100K

12

JP

201

HE

AD

ER

2

1 2

-8V

- +

U14

LM

C6061

382

7

6

4

NOTE:

NOT STUFFED FOR 9110

GP

O

R41

1K

1%

12

R111

100K

1 2

R114

1K

1 2

NOTE: NO SOLDER MASK IN PROBE AREA

*

- +

U15

LM

C6061

382

7

6

4

INPUT AMPLIFIERS

NOTE: 300 SERIES COMPONENTS

+5V

OFFSET GENERATOR

*

R48

604K

1%

12

VC

C

C50

0.2

2uF

12

P2

PR

OB

E

NOTE: PROBE THRU IN+ SHIELDED

KO

UT

R112

100K

1 2

*

C49

0.2

2uF

12

-8V

R113

1K

1 2

-8V

K300

SIP

RE

LA

Y S

PS

T

71

53

ZD

50

05

F

SC

HE

MA

TIC

, MA

IN B

OA

RD

91

XX

59

Th

urs

day

, Ju

ne

27,

200

2

Title

Siz

eD

ocum

ent

Num

ber

Rev

Date

:S

heet

of

R45

43.2

K 1

%

12

NOTE VALUES OF R44 AND R47

R34

100K

12

OFFSET = 0

R115

1K

1 2

TP

11

UT35-0

17

1

C52

0.1

uF

1 2

ARE DIFFERENT FOR ISE AND pH

R44

59K

1%

IS

E18.2

K 1

% p

H

12

R303

10K

12

PH

Page 61: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 A-47

91xx pH/Electrolyte Analyzers 5 Electronic diagramsElectronic diagrams

SBC PCB - temperature circuit - 9110, 9120, 9130, 9140

Figure A-12 SBC PCB - temperature circuit - 9110, 9120, 9130, 9140

AD

DA

T4

R69

1K

12

AD

HB

EN-

ADD

AT[0

..7]

AD

DA

T[0.

.7]

X1 3.58

MH

z

12

ADS

EL1

+5V

ADS

EL0

AD

DA

T2

R72

1K 1

%

12

VCC

R68

1K

12

A

Cl/C

a/Li

OU

T

AD

DA

T3

R71

200K

1%

1 2

VCC

ADS

EL2

AD

DA

T6

AD

CE

-

Na/

pHO

UT

8 TO 1 MULTIPLEXER

AD

DA

T4

U19

4051

13 14 15 12 1 5 2 4 6 11 10 9

3X0 X1 X2 X3 X4 X5 X6 X7 IN

HA B C

X

ZD50

05F

SCHE

MAT

IC, M

AIN

BOAR

D 9

1XX

79

Thur

sday

, Jun

e 27

, 200

2

Title

Size

Doc

umen

t Num

ber

Rev

Dat

e:S

heet

of

mb7.sch

AD

DA

T5

AD

DA

T0

R66

1K

1 2

AD

DA

T3

TP14

UT3

5-01

7

1

1 LSB APPROXIMATELY 1mV

ADST

ATU

S

KOU

T

AD

DA

T2

NOTE:

R60

100

1 2

C56

150n

F

1 2

2.05V

AD

RU

N

AD

DA

T1

U21

MC

1403

1 32

VIN

GN

DVO

UT

C57

1uF 1 2

C55

330n

F

1 2

AD

DA

T1

R70

1K

1 2

"2.500V"

AD

LBEN

-

+2.500V

-8V

C54

1nF

12

R61

100K

12

TEM

POU

T

+5V

AD

DA

T7

U20

ICL7

109

16 15 14 13 12 11 10 9 8 7 6 5 4 3 2

35 34 33 36 39 37 38 29 30 31 32 17 21 18 19 20 26 27 24 25 22 2328140

B1 B2 B3 B4 B5 B6 B7 B8 B9 B10

B11

B12

OR

PO

L

STA

TUS

INH

IIN

LOC

OM

MO

N

REF

IN+

REF

IN-

REF

CAP

+R

EFC

AP-

REF

OU

T

BU

FFE

RA

ZIN

T

TEST

MO

DE

LBEN

HBE

NC

E/LO

ADR

UN

/HO

LDS

EN

D

OSC

SEL

BU

FFO

SC

OU

TO

SCIN

OSC

OU

TV-

GN

DV+

AD

DA

T0

AD

DA

T5

A/D CONVERTER

R73

4.53

K 1%

1 2

-8V

R67

1K

12

Page 62: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

A-48 Service manual · Version 2.0

5 Electronic diagrams 91xx pH/Electrolyte AnalyzersElectronic diagrams

SBC PCB - analog selector and A/D converter - 9110, 9120, 9130, 9140

Figure A-13 SBC PCB - analog selector and A/D converter - 9110, 9120, 9130, 9140

AD

DA

T4

R69

1K

12

AD

HB

EN-

ADD

AT[0

..7]

AD

DA

T[0.

.7]

X1 3.58

MH

z

12

ADS

EL1

+5V

ADS

EL0

AD

DA

T2

R72

1K 1

%

12

VCC

R68

1K

12

A

Cl/C

a/Li

OU

T

AD

DA

T3

R71

200K

1%

1 2

VCC

ADS

EL2

AD

DA

T6

AD

CE

-

Na/

pHO

UT

8 TO 1 MULTIPLEXER

AD

DA

T4

U19

4051

13 14 15 12 1 5 2 4 6 11 10 9

3X0 X1 X2 X3 X4 X5 X6 X7 IN

HA B C

X

ZD50

05F

SCHE

MAT

IC, M

AIN

BOAR

D 9

1XX

79

Thur

sday

, Jun

e 27

, 200

2

Title

Size

Doc

umen

t Num

ber

Rev

Dat

e:S

heet

of

mb7.sch

AD

DA

T5

AD

DA

T0

R66

1K

1 2

AD

DA

T3

TP14

UT3

5-01

7

1

1 LSB APPROXIMATELY 1mV

ADST

ATU

S

KOU

T

AD

DA

T2

NOTE:

R60

100

1 2

C56

150n

F

1 2

2.05V

AD

RU

N

AD

DA

T1

U21

MC

1403

1 32

VIN

GN

DVO

UT

C57

1uF 1 2

C55

330n

F

1 2

AD

DA

T1

R70

1K

1 2

"2.500V"

AD

LBEN

-

+2.500V

-8V

C54

1nF

12

R61

100K

12

TEM

POU

T

+5V

AD

DA

T7

U20

ICL7

109

16 15 14 13 12 11 10 9 8 7 6 5 4 3 2

35 34 33 36 39 37 38 29 30 31 32 17 21 18 19 20 26 27 24 25 22 2328140

B1 B2 B3 B4 B5 B6 B7 B8 B9 B10

B11

B12

OR

PO

L

STA

TUS

INH

IIN

LOC

OM

MO

N

REF

IN+

REF

IN-

REF

CAP

+R

EFC

AP-

REF

OU

T

BU

FFE

RA

ZIN

T

TEST

MO

DE

LBEN

HBE

NC

E/LO

ADR

UN

/HO

LDS

EN

D

OSC

SEL

BU

FFO

SC

OU

TO

SCIN

OSC

OU

TV-

GN

DV+

AD

DA

T0

AD

DA

T5

A/D CONVERTER

R73

4.53

K 1%

1 2

-8V

R67

1K

12

Page 63: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 A-49

91xx pH/Electrolyte Analyzers 5 Electronic diagramsElectronic diagrams

SBC PCB - microprocessor - 9110, 9120, 9130, 9140

Figure A-14 SBC PCB - microprocessor - 9110, 9120, 9130, 9140

A15

A5

A6

RA

ME

N0-

RS

232I

N

A12

C60

22pF1

2

AD

6

V+

A8

NON-VOLATILE

FE

ED

A3

R11

74.

7K

12

D18

1N41

48

21

AD

2

AD

0

AD

7

R10

1

10K

12

ZS

5

DIP

28

ZS

6

DIP

24/3

AD

4 A1

A[0

..15

]

VC

C

JUMPER U27 PIN1 TO PIN3

A14

AD

4

R77

1K

1 2P

RO

ME

N-

PR

OM

EN

-

GP

O

A11

HANDSHAKE

VC

C

RA

ME

N0-

PR

STR

A0

R75

10K

12

U22

MK

48T0

8

10 9 8 7 6 5 4 3 25 24 21 23 2 20 26 27 22

11 12 13 15 16 17 18 19 14281

A0

A1

A2

A3

A4

A5

A6

A7

A8

A9

A10

A11

A12

CS

1C

S2

WE

OE

D0

D1

D2

D3

D4

D5

D6

D7

GN

D

VC

C

INT

RD

-

PU

MP

STE

P

A2

R11

610

K

1 2

RAM

AD

4

A4

A1

VC

C

RE

SE

T-

A12

DE-MUX

A15

A4

THE

RM

-

X2 8MH

z

12

D17

1N41

48

21

AD

5

U24

22V

10

1 2 3 4 5 6 7 8 9 10 11 13

23 22 21 20 19 18 17 16 15 14

I1/C

LKI2 I3 I4 I5 I6 I7 I8 I9 I1

0I1

1I1

2

O1

O2

O3

O4

O5

O6

O7

O8

O9

O10

A9

C58

0.1u

F

1 2

VC

C

AD

6

A6

A8

R83

10K

12

AD

3

A[0

..15

]V

CC

A15

RAM & CLOCK

DIS

PB

F

A7

R82

10K

12

NOT STUFFED

MO

SI

A10

A12

DECODE

+5V

V+

AD

STA

TUS

AD

7

CH

EC

K

AD

3

A10

A12

ZP

WB

1

PW

B

AD

2

A2

RTS

AD

1

RD

-

AD

[0..

7]

AD

5A

D4

A14

R74

10M

1 2

PROGRAM

A9

1-2

RS

232O

UT

A14

AD

0

VC

C

AD

1

AD

7

ZS

3

PLC

C52

YE

S-

A7

A3

AD

6

SPI STROBES

WR

-

MIS

O

2-3

VA

LVE

-

A[0

..15

]

A8

AD

0

AD

7

RA

ME

N1-

VC

CA

D1

A5

A10

ZD50

05F

SC

HEM

ATI

C, M

AIN

BO

AR

D 9

1XX

89

Th

urs

day

, Ju

ne

27,

200

2

Title

Siz

eD

ocum

ent

Num

ber

Rev

Dat

e:S

heet

of

AD

2

VC

C +5V

A8

A6

U25

2751

2

10 9 8 7 6 5 4 3 25 24 21 23 2 26 27 1 20 22

11 12 13 15 16 17 18 19

A0

A1

A2

A3

A4

A5

A6

A7

A8

A9

A10

A11

A12

A13

A14

A15

CE

OE

O0

O1

O2

O3

O4

O5

O6

O7

VC

C

SC

K

LAM

PE

N

A0

U23

74H

CT3

73

3 4 7 8 13 14 17 18 1 11

2 5 6 9 12 15 16 19

D0

D1

D2

D3

D4

D5

D6

D7

OC

G

Q0

Q1

Q2

Q3

Q4

Q5

Q6

Q7

N/C

A3

A10

A5

A4

U30

A14

89

13

2

VC

C

V+

A0

A3

AD

3

A7

A9

U28

74H

CT1

38

1 2 3 6 4 5

15 14 13 12 11 10 9 7

A B C G1

G2A

G2B

Y0

Y1

Y2

Y3

Y4

Y5

Y6

Y7

VC

CV

AS

TR3-4

A5

A6

TAC

HO

A0

AD

5

A0

ADDRESS

A7

A9

A

A1

CPU

NO

-

AD

2

A10

SA

MP

LE

BU

ZZ

ER

AD

5

A2

OEM

VC

C

AD

[0..

7]

A11

A11

U30

B14

89

46

5

JP6

BE

RG

2

1 2

U30

C14

89

108

9

CTS

AD

1

R81

10K

12

A3

AVL ISE

VC

C

AD

4

DO

OR

IOS

TRO

BE

-

A1

AD

7

A2

R86

10K

1 2

VC

C

A13

R80

10K

12

A6

R76

10K

12

U26

6264

10 9 8 7 6 5 4 3 25 24 21 23 2 20 26 27 22

11 12 13 15 16 17 18 19

A0

A1

A2

A3

A4

A5

A6

A7

A8

A9

A10

A11

A12

CS

1C

S2

WE

OE

D0

D1

D2

D3

D4

D5

D6

D7

A13

AD

5A

D6

A13

AD

[0..

7]

ZS

4

DIP

28

AD

0

FA

ST/

SLO

W-

AD

3

AD

0

R78

10K

12

A8

U29

A

1488

23

AD

OS

TR

AD

3

VC

C

A2

pH JUMPER 3 TO VCC

A5

A4

VC

C

DIS

TR

AD

ISTR

A12

EPROM

VC

C

A9

A11

A11

R85

10K

12

A1

A4

U31

MC

68H

C11

A0F

N

34 33 32 31 30 29 28 27 9 10 11 12 13 14 15 16 42 41 40 39 38 37 36 35 4 6 5 20 21 22 23 24 25

7 8 17 3 2 19 18 52 51 43 45 47 49 44 46 48 50

IC3/

PA

0IC

2/P

A1

IC1/

PA

2O

C5/

OC

1/P

A3

OC

4/O

C1/

PA

4O

C3/

OC

1/P

A5

OC

2/O

C1/

PA

6P

AI/

OC

1/P

A7

AD

0/P

C0

AD

1/P

C1

AD

2/P

C2

AD

3/P

C3

AD

4/P

C4

AD

5/P

C5

AD

6/P

C6

AD

7/P

C7

A8/

PB

0A

9/P

B1

A10

/PB

2A

11/P

B3

A12

/PB

4A

13/P

B5

A14

/PB

6A

15/P

B7

AS

/STR

AR

/W/S

TRB E

RXD

/PD

0TX

D/P

D1

MIS

O/P

D2

MO

SI/

PD

3S

CK

/PD

4S

S/P

D5

EXT

AL

XTA

L

RE

SE

T

MO

DA

/LIR

MO

DB

/VS

TBY

IRQ

XIR

Q

VR

HV

RL

AN

0/P

E0

AN

1/P

E1

AN

2/P

E2

AN

3/P

E3

AN

4/P

E4

AN

5/P

E5

AN

6/P

E6

AN

7/P

E7

C59

22pF1

2

PU

MP

ENW

R-

RA

ME

N1-

AD

[0..

7]

RTS

O

AD

1

U29

B

1488

4 56

U29

C

1488

9 108

U29

D

1488

12 1311

JP40

0

1234

AD

6

AD

2

A7

R84

10K

12

A[0

..15

]

mb8.sch

Page 64: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

A-50 Service manual · Version 2.0

5 Electronic diagrams 91xx pH/Electrolyte AnalyzersElectronic diagrams

SBC PCB - printer display drivers - 9110, 9120, 9130, 9140

Figure A-15 SBC PCB - printer display drivers - 9110, 9120, 9130, 9140

OP

EN

2

NOT STUFFED FOR 9110

V+

+5V

A

DD

[0..

7]

DIS

PB

F

BU

ZZ

ER

Q5

2N39

06

SS

EN

DD

1

+5V

A

DD

7

BZ

1

BU

ZZ

ER

1 2

R98

10K

12

"SS ADJ."

DO

T6-

SA

MP

LELE

D-

R95

100

1 2

-8V

ZD50

05F

SC

HEM

ATI

C, M

AIN

BO

AR

D 9

1XX

99

Th

urs

day

, Ju

ne

27,

200

2

Title

Siz

eD

ocum

ent

Num

ber

Rev

Dat

e:S

heet

of

DSPDATA

DO

T2-

R89

5KP

OT

1 3

2

ADDATA

DO

T5-

R91

10K

12

DIS

TR

R10

210

K1

2

AD

SE

L1

VC

C

OP

EN

3

DD

4

DIS

PR

/W-

AD

DA

T7

VACTRL

AD

DA

T3

"SS"

AD

DA

T1

AD

LBE

N-

R96

1K

1 2

U30

D14

89

1311

12

OP

EN

5

AD

OS

TR

AD

DA

T[0.

.7]

DO

T7-

DD

0

PG

ND

MIS

O

SA

MP

LE

R99

5.1K

1 2

DO

T3-

U35

74H

C59

5

14 11 10 12 13

15 1 2 3 4 5 6 7 9

SE

R

SR

CLK

SR

CLR

RC

LKG

QA

QB

QC

QD

QE

QF

QG

QH

QH

'

VA

STR

DD

3

AD

HB

EN

-

Q3

2N39

04

AD

DA

T0

AD

DA

T6

AD

SE

L2

Q4

2N39

04

AD

ISTR

MO

SI

"CONTRAST"

OP

EN

4

AD

CE

-

PG

ND

U33

LB12

56

10 18 17 16 15 14 13 12 11 9

1 2 3 4 5 6 7 8

VC

CO

UT1

OU

T2O

UT3

OU

T4O

UT5

OU

T6O

UT7

OU

TMG

ND

IN1

IN2

IN3

IN4

IN5

IN6

IN7

INM

U36

74H

C59

4

14 11 10 12 13

15 1 2 3 4 5 6 7 9

SE

R

SR

CLK

SR

CLR

RC

LKR

CLR

QA

QB

QC

QD

QE

QF

QG

QH

QH

'

VC

C

R88

4.7K

1 2

DA

TA/I

NS

T-

MO

TOR

-

VC

C

AD

RU

N

AD

DA

T5

VC

C

TP15

UT3

5-01

7

1

R94

1K

12

mb9.sch

SS

EN

OP

EN

1

PR

STR

AD

DA

T2

VC

C

CO

NTR

AS

T

AD

DA

T[0.

.7]

DD

2

STUFFED

R87

150

1 2

SC

K

+5V

A

+ -

U39

A

LF35

3

3 21

8 4

PRCTRL

AD

SE

L0

NOT

U34

74H

C59

4

14 11 10 12 13

15 1 2 3 4 5 6 7 9

SE

R

SR

CLK

SR

CLR

RC

LKR

CLR

QA

QB

QC

QD

QE

QF

QG

QH

QH

'

U37

74H

C59

4

14 11 10 12 13

15 1 2 3 4 5 6 7 9

SE

R

SR

CLK

SR

CLR

RC

LKR

CLR

QA

QB

QC

QD

QE

QF

QG

QH

QH

'

DO

T1-

AD

DA

T4

PR

INTV

CC

+5V

DO

T4-

DD

6

DD

[0..

7]

SA

MP

LE

DIS

PE

N

U38

74H

C16

5

10 11 12 13 14 3 4 5 6 2 15 1

9 7

SE

RA B C D E F G H C

LKIN

HS

H/L

D

QH

QH

R92

1K

1 2

RE

SE

T-

R97

10K

1 2

A

ADCTRL

R93

5KP

OT

1 3

2

DD

5

R90

1K

1 2

Q6

2N39

04

Page 65: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 A-51

91xx pH/Electrolyte Analyzers 5 Electronic diagramsElectronic diagrams

Printer daughter PCB - 9110, 9120, 9130, 9140, 9180, 9181

Figure A-16 Printer daughter PCB - 9110, 9120, 9130, 9140, 9180, 9181

JP1

ED

GE2

0

A1 B1 A2 B2 A3 B3 A4 B4 A5 B5 A6 B6 A7 B7 A8 B8 A9 B9 A10

B10

PAPER

SWITCH

J2 5203

15-6

1 2 3 4 5 6

TAC

HO

J1 5203

15-8

1 2 3 4 5 6 7 8

FEED

DO

T4-

SPST

ZM

5003

PW

B

D1

1N40

04

21

MO

TOR

-

S1

DO

T1-

2

NOTE: Only stuffed for "A" and "B" type printers

DO

T7-

ON SOLDER SIDE

DO

T3-

1

ZD50

03E

SCHE

MAT

IC, P

RINT

ER D

AUGH

TER

BOAR

D

11

Mon

day,

May

10,

199

9

Title

Siz

eD

ocum

ent

Num

ber

Rev

Dat

e:Sh

eet

of

B1

SCHURTER

MATES WITH SULLINS EZC10DRXH

DO

T2-

B2

CH

ECK

VCC

B10

JP2

JUM

PE

R

12

A

COMPONENT SIDE

AMP

A1

DO

T6-

AMP

.100 CENTERS

FEED

DO

T5-

Page 66: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

A-52 Service manual · Version 2.0

5 Electronic diagrams 91xx pH/Electrolyte AnalyzersElectronic diagrams

Lamp PCB - 9110, 9120, 9130, 9140

Figure A-17 Lamp PCB - 9110, 9120, 9130, 9140

JP5

1036

70-2

1 2 3

VA

LVE

A-

ZP

WB

1

PW

B

VALVER

LAM

P+

VA

LVE

R-

U3

HLM

P28

852 3 6 7 10 11 14 15

1 4 5 8 9 12 13 16

VALVEV

PH

AS

ED

OV

ER

TEM

P-

VALVEC

JP1

1036

70-2

1 2 3

PH

AS

EA

U2

HLM

P28

852 3 6 7 10 11 14 15

1 4 5 8 9 12 13 16

VALVEB

U1

HLM

P28

852 3 6 7 10 11 14 15

1 4 5 8 9 12 13 16

AMP

VALVEA

VA

LVE

V-

AMP

A

VA

LVE

+

VA

LVE

+

AMP

VA

LVE

B-

JP2

1036

70-2

1 2 3 AMP

JP3

1036

70-2

1 2 3

C

DC

LOS

ED

AMP

ZD5007

SC

HE

MA

TIC

, LA

MP

BO

AR

D

11

Frid

ay, N

ove

mb

er

21, 1

997

Title

Siz

eD

ocum

ent

Num

ber

Rev

Dat

e:S

heet

of

lamp.sch

LED ARRAY

DLE

D+

JP8

C3A

PT-

2606

G-N

D

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26

PH

AS

EB

AMP

LED ARRAY

VA

LVE

C-

JP4

1036

70-2

1 2 3

AMP

LED ARRAY

NOTE: I = 480 ma.

DOOR SENSOR

PH

AS

EC

JP6

1036

70-5

1 2 3 4 5 6

JP7

1036

70-3

1 2 3 4

PUMP

Page 67: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 A-53

91xx pH/Electrolyte Analyzers 5 Electronic diagramsElectronic diagrams

Display daughter PCB - 9110

Figure A-18 Display daughter PCB - 9110

+24V

R13

90.9

K 1

%

1 2

+24V

DD

2

R34

5.1

2W

12

R11

90.9

K 1

%

1 2

C6

0.1uF

1 2

VB

KL

+ -

U1C

TLC

279

10 9

8

R30

220

1 2

+5V

A

R22

9.0

9K

1%

1 2

DD

0

+-

U2A

LM32

4

321

11 4

+ -

U2C

LM

324

10

98

+5V

A

JP

4

AM

P 1

03670-2

1 2 3

BR

EA

DB

OA

RD

VE

RS

ION

FO

R F

EA

SIB

ILIT

Y

R26

10M

1 2

R28

TB

D

1 2

CO

NTR

AS

T

R1

10K

12

R5

100K

1 2

R16 1K

12

+

C9

10uF

12

D1

1N

4148

K1 D

IP R

ELA

Y S

PD

T

71

1426

BR

EA

DB

OA

RD

VE

RS

ION

FO

R F

EA

SIB

ILIT

Y

DD

2

YE

S

D4

1N

4148

+5V

A

R35

470K

12

R8

1M

1 2

VC

C

R23

562

1%

1 2

K SO

LDE

R P

AD

1

MO

LE

X

R17

1K

1 2

JP

6

HE

AD

ER

14

1 2 3 4 5 6 7 8 9 10

11

12

13

14

THE

RM

ISTO

R2

R4

100

K

12

SA

MP

2

D2

1N

4148

JP

1

RJ1

21 2 3 4 5 6

R7

100K

12

R9

470

12

R21

10K

12

R20

500

R25

7.5

K 1

%

1 2

DD

1

SS

EN

R18

10K

12

R10

50K

VC

C

+5V

AV

CC

+ -

U1D

TLC

279

12

13

14

VC

C

DD

6

YE

S-

Q2

2N

3904

1

2

3

+

C8

10uF

12

+24V

VC

C

DD

1

R14

1M

12

C1

1nF

1 2

VC

C

DD

6

+10V

S1

IP1241.1

082.3

DD

5

ZD

5010

E

SC

HE

MA

TIC

, P

CB

, pH

DIS

PLA

Y,

9110

B

22

Monday

, M

arch 0

3, 2

003

Title

Siz

eD

ocu

ment

Num

ber

Rev

Date

:S

hee

tof

DD

4

R24

1K

12

R6

150K

12

Q1

2N

3904

1

2

3

C4

1uF

1 2

C5

1.0

uF

12

TP

1

LC

D D

ISP

LAY

R3

1M

12

S2

IP1241.1

082.3

R29

68K

12

+ -

U2D

LM

324

12

13

14

PH

TE

MP

BR

EA

DB

OA

RD

+-

U2B

LM

324

567

R33

11

2

DIS

PE

N

DD

7

R15

10K

12

DD

0

D6

1N

4148

-Vss

D7

LE

D

12

MO

LE

X 15-4

3-8

566

+10V

+5V

A

R12

3.0

9K

1%

1 2

ZP

WB

1

PW

B

DD

3

PH

TEM

P B

RE

AD

BO

AR

D

TEM

P

DD

7

C7

0.1

uF

1 2

DD

3

JP

5

HE

AD

ER

26

1 2 3 4 5 6 7 8 910

11

12

13

14

15

16

17

18

19

20

21

22

23

24

25

26

A SO

LDE

R P

AD

1

TE

MP

BLO

CK

THE

RM

ISTO

R1

R19

470

12

DIS

PR

/W-

R32

1.5

K

12

D3

1N

4148

R31

1.5

K

12

SC

HU

RTE

R

BLO

CK

R27

TB

D

1 2

+ -

U1A

TLC

279

3 21

114

R J

UN

CTIO

N

+10V

SC

HU

RTE

R

SA

MP

1

D5

1N

4148

NO

-

"TE

MP

SE

T"

SA

MP

CO

NT

NO

+-

U1B

TLC

279

567

DD

4D

D5

C2

0.1uF

12

-Vss

+24V

DA

TA/I

NS

T-

R2

5.1

2W

1 2

+C

32.2

uF

1 2

CLL 2

4 F

EB

1992

BA

SE

SA

MP

CO

NT

Page 68: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

A-54 Service manual · Version 2.0

5 Electronic diagrams 91xx pH/Electrolyte AnalyzersElectronic diagrams

Preheater PCB - 9110

Figure A-19 Preheater PCB - 9110

R(T)

10K @ 25C1 2

RED

GRN

JP1

MODULAR PLUG

123456 YEL

BLU

FOR Q1 AND Q2

R2

1.0

12

USE INSULATOR PADS

R11.0

12

Q2MJE800

12

3

CONNECT TO BLOCK

WHT

Q1

MJE800

12

3

BLOCKTP1

1

BLK

PHb.sch

ZD5009 A

SCHEMATIC, SAMPLE HEATER BLOCK

A

1 1Thursday , Nov ember 14, 1996

Title

Size Document Number Rev

Date: Sheet of

Page 69: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 A-55

91xx pH/Electrolyte Analyzers 5 Electronic diagramsElectronic diagrams

Display daughter PCB - 9120, 9130, 9140

Figure A-20 Display daughter PCB - 9120, 9130, 9140

TP1

UT35-017

DD5

SAMPLEGND

S2

IP1241.1082.3

DD3

DD4

DISPR/W-

LCD DISPLAY

SAMPLELED-

display .sch

NOT STUFFED

DD7

DD5

ALTERNATE

+5VA

JP1

RJ12

123456

DD2

YES-

JP2

HEADER 26

123456789

1011121314151617181920212223242526

DD1

MOLEX

VCC

DD2

DATA/INST-

DD6

DD0

SCHURTER

DD6

DD4

SCHURTER

DD[0..7]

NOT STUFFED

S1

IP1241.1082.3

MOLEX

TEMP

JP3

HEADER 14

1234567891011121314

DD0

DD2

MOLEX 15-43-8566

CONTRAST

SAMPLEPT

LCISE C

SCHEMATIC, DISPLAY DAUGHTER BOARD

B

1 1Thursday , February 27, 2003

Title

Size Document Number Rev

Date: Sheet of

+5VA

SAMPLE SENSOR

MAINDOOR

+5VA

DD4

VCC

DD0

DISPEN

DD1

LCD DISPLAY

DD5

ZPWB1

PWB

DD6

DD3

NO

VCC

DD3

DD7

VCC

DD1

YES

NO-

DD7

Page 70: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

A-56 Service manual · Version 2.0

5 Electronic diagrams 91xx pH/Electrolyte AnalyzersElectronic diagrams

System interconnect - 9180, 9181

Figure A-21 System interconnect - 9180, 9181

R92

627

0K

1 2

TAC

HO

PH

AS

EA

R90

710

K

1 2 DD

[0..

7]

PG

ND

DD

4

PV

CC

DD

1

SS

EN

LGS

EL

3

R900-R903 DO NOT PHYSICALLY EXIST

ZS

CR

W-J

P7-

2

SC

RE

W

ZD

5015

G

SC

HE

MA

TIC

, M

AIN

BO

AR

D 9

180

/ 91

81

211

Mo

nday

, M

ay

03

, 1

999

Title

Siz

eD

ocu

men

t N

umbe

rR

ev

Dat

e:S

heet

of

R90

10

12

CO

NTR

AS

T

TBM

OTE

N-

ZF

ST-

JP1-

2

FA

STE

NE

R2

LAMP BOARD

JP1

EZ

C10

DR

XH

A1

B1

A2

B2

A3

B3

A4

B4

A5

B5

A6

B6

A7

B7

A8

B8

A9

B9

A10

B10

VC

C

AMP 745061-4

CH

EC

K

R90

00

12

DO

T5-

CW

/CC

W

JP4

CW

N-3

00-3

4-00

00

1 2 3 4 5 6 7 8 9 10

11

12

13

14

15

16

17

18

19

20

21

22

23

24

25

26

27

28

29

30

31

32

33

34

VA

LVE

C-

YE

S-

CONTROL LINES

RS

232O

UT

PG

ND

VC

C

R11

8

330

MOLEX

P+2

4V

DIS

PR

/W-

TA

BLE

PR

ES

R93

2

0

DC

LOS

ED

R90

30

12

R93

30

SULLINS

PG

ND

+24V

PH

AS

EB

R90

510

K

1 2

PG

ND

DD

6

AMP

TURNTABLE

UP

ND

PO

S

DD

3

+24

VD

D[0

..7]

JP3

26-6

0-40

70

1 2 3 4 5 6 7

R93

510

K

1 2S

AM

PLE

SG

ND

MATES WITH MOLEX 09-50-8073

AREA.

VC

C

DO

T1-

ZS

CR

W-J

P7-

1

SC

RE

W

R93

410

K

1 2

PH

AS

ED

SG

ND

DO

T6-

R9

0810

K

1 2

NO

-

DD

0

DISPLAY BOARD

AND CONNECTED AT THE SHIELD

LGS

EL

0PHYSICALLY INSTALLED

D2

4

1N41

48

21

+5V

AS

AM

PLE

LED

-

DO

T3-

VA

LVE

+

PRINTER

SAMPLEGND IS RUN INDIVIDUALLY

R11

9

330

PH

AS

ECJP

7

DB

9P

5 9 4 8 3 7 2 6 1

R933 IS NOT

VC

C

ZE

J-J

P5

-2

EJ

EC

TOR

P-1

2V

SYSTEM GROUND

NOTE:

VA

LVE

V-

LWN

DP

OS

DG

ND

ZE

J-JP

4-2

EJE

CTO

R

RT

S

UP

TBP

OS

DD

2

RS-232

CONTROL LINES

PV

CC

DO

T2-

RS

232I

N

DLE

D+DO

T7-

CT

S

VA

LVE

A-

NEEDLE MECHANISM

VC

C

DD

5

+10

V

TEM

P

LGS

EL

1

VC

C

LGS

EL

2

D23

1N4

148

21

A

OV

ER

TEM

P-

DO

T4-

DG

ND

ND

MO

TEN

-

AG

ND

SW

1

78B

04

1 2 3 4

8 7 6 5

MO

TOR

-

JP5

CW

N-3

00-

34-0

000

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34

R90

20

12

LANGUAGE SELECT

VC

C

ZE

J-JP

4-1

EJE

CTO

R

VA

LVE

R-

PR

INTV

CC

LAM

P+

VA

LVE

B-

DD

7

R9

0610

K

1 2

ZE

J-J

P5

-1

EJ

EC

TOR

DIS

PE

N

THEY ARE TO ALLOW GROUNDS TO BE SEPARATED

LWTB

PO

S

R90

4

10K

12

ZF

ST-

JP1-

1

FA

STE

NE

R2

JP2

626

50-1

1

POWER

DA

TA/I

NS

T-

FE

ED

C82

3.3u

F/5

0V

VC

C

P+1

2V

RO

TCLK

Page 71: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 A-57

91xx pH/Electrolyte Analyzers 5 Electronic diagramsElectronic diagrams

SBC PCB - power supply - 9180, 9181

Figure A-22 SBC PCB - power supply - 9180, 9181

+24V

V-

PR

INTV

CC

TP5

1

C20

0.1u

F

12

C24

0.1u

F

12

C1

0.1u

F

1 2

VC

C

FA

ST/

SLO

W-

C21

0.1u

F

12

C65

0.1u

F

1 2

TP2

1

D2

LED

21

TP4

1TP1

1

C17

0.1u

F

1 2

C18

0.1u

F

1 2

C22

0.1u

F

12

C23

0.1u

F

12

C72

0.1u

F

1 2

G

V-

F5

0.30

0AT

12

Q7

MJE

800

C19

0.1u

F

12

R8

1.2K

12

C3

0.1u

F

1 2

+5V

A

R2

1.8K

1 2

TP6

TP 1

A

ZH

LDR

5

FU

SE

HO

LDE

R

P+2

4V

U1

LM31

73

2

1

VIN

VO

UT

ADJ

R10

220

12

D1

LED

21

F2

1.6A

T

12

"+5VA"

+C

2710

uF/1

6V

1 2

PV

CC

C8

0.1u

F

1 2

TP7

TP 1

C12

0.1u

F1 2

"-12V/F4"

"-5VA"

+VS

R4

180

1 2

ZH

S1

HE

ATS

INK

C7

0.1u

F

1 2

R10

31K

12

C73

0.1u

F

1 2

C13

0.1u

F

1 2

C67

0.1u

F

1 2

+

C25

470u

F/2

5V

12

"+24V/F3"

"0.300AT"

PR

INTV

CC

+C

6410

00uF

/16V

1 2C

90.

1uF

1 2

C68

0.1u

F

1 2

R7

560

12

C74

0.1u

F

1 2

PV

CC

P+1

2V

C69

0.1u

F

1 2

C70

0.1u

F

1 2

"+5V/F2"

"0.300AT"

C75

0.1u

F

1 2

V+

D5

LED

21

F3

1.25

AT

12

TP8

TP 1

+C

2922

00uF

/50V

1 2C

710.

1uF

1 2

LAM

P+

"+5V/F1"

"1.25AT"

U2

LM79

052

3

1

VIN

VO

UT

C

F1

2AT

12

C11

0.1u

F

1 2

R6

560

12

R9

220

12

ZD50

15

SC

HEM

ATI

C, M

AIN

BO

AR

D 9

180

/ 918

1 411

Mon

day

, M

ay 0

3, 1

999

Title

Siz

eD

ocum

ent

Num

ber

Rev

Dat

e:S

heet

of

C14

0.1u

F

1 2

"GND"

"1.6AT"

-5V

A

ZH

LDR

1

FU

SE

HO

LDE

R

ZF

ST-

U2

FA

STE

NE

R1

C5

0.1u

F

12

TP3

1

C10

0.1u

F

1 2

+C

2647

0uF

/25V

1 2

Q1

2N39

04

"+12V"

"2AT"

+C

3022

00uF

/16V

1 2

R5

3.9/

10W

12

C4

0.1u

F

1 2

C6

0.1u

F

1 2

+24V

R1

1K

12

+

C28

10uF

/16V

12

ZH

LDR

3

FU

SE

HO

LDE

R

VC

C

"-12V"

D4

LED

21

F4

0.30

0AT

12

P-1

2V

U3

LM78

051

3

2

VIN

VO

UT

C

-5V

A

"+24V"

D3

LED

21

C15

0.1u

F

1 2

"+12V/F5"

D6

ICTE

.5

21

LAM

PE

N

ZH

LDR

2

FU

SE

HO

LDE

R

"+5V"

ZH

LDR

4

FU

SE

HO

LDE

R

C16

0.1u

F

1 2

C66

0.1u

F

1 2

+5V

A

V+

R3

1.2K

1 2

C2

0.1u

F

12

ZF

ST-

U3

FA

STE

NE

R1

"+22V/12V"

Page 72: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

A-58 Service manual · Version 2.0

5 Electronic diagrams 91xx pH/Electrolyte AnalyzersElectronic diagrams

SBC PCB - valve drivers - 9180, 9181

Figure A-23 SBC PCB - valve drivers - 9180, 9181

C32

0.1u

F

1 2

VA

LVE

V-

U5

LM31

73

2

1

VIN

VO

UT

ADJ

D19

1N75

0

12

C31

0.1u

F

1 2

SG

ND

THE

RM

-

+

C36

2.2u

F/5

0V12

OV

ER

TEM

P-

+

C39

2.2u

F/5

0V12

R19

1.82

K 1

%

1 2

VA

LVE

-

R15

1K

12

TP161

OP

EN

2

R18

200

1%

1 2

OP

EN

1

R21

4.7K

12

OP

EN

5

Q11

MJE

800

R14

1K

12

A

VA

LVE

A-

Q12

MJE

800

Q8

2N39

04

"+22V/12.5V"

OP

EN

3

Q10

MJE

800

+

C40

2.2u

F/5

0V12

R13

1K

12

+24V

+

C38

2.2u

F/5

0V12

VALVE DRIVERS

SG

ND

ZH

S2

HE

ATS

INK

Q9

MJE

800

R16

10K

12

VA

LVE

B-

VA

LVE

C-

R11

1K

12

OP

EN

4

VA

LVE

R-

R12

1K

12

VA

LVE

+

R20

1.5K

1 2ZD

5015

G

SC

HEM

ATI

C, M

AIN

BO

AR

D 9

180

/ 918

1 511

Mon

day

, M

ay 0

3, 1

999

Ti

tle

Siz

eD

ocum

ent

Num

ber

Rev

Dat

e:S

heet

of

+

C37

2.2u

F/5

0V12

Q13

MJE

800

Page 73: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 A-59

91xx pH/Electrolyte Analyzers 5 Electronic diagramsElectronic diagrams

SBC PCB - pump motor driver/door detect - 9180, 9181

Figure A-24 SBC PCB - pump motor driver/door detect - 9180, 9181

VC

C

SG

ND

C45

0.1u

F

1 2

ZS

1

DIP

20

+VS

SG

ND

VC

C

C41

0.1u

F

1 2

R29

10K

1 2

VC

C

SG

ND

DO

OR

SG

ND

D9

1N40

04

21

SG

ND

DC

LOS

ED

D10

1N40

04

21

+VS

D14

1N40

04

21

D8

1N40

04

21

PH

AS

ED

PH

AS

EB

D12

1N40

04

21

R30

33K

1 2

R25

1

1 2

SG

ND

SG

ND

ZD50

15G

SC

HEM

ATI

C, M

AIN

BO

AR

D 9

180

/ 918

1 611

Mon

day

, M

ay 0

3, 1

999

Title

Siz

eD

ocum

ent

Num

ber

Rev

Dat

e:S

heet

of

SG

ND

D13

1N40

04

21

SG

ND

D7

1N40

04

21

SG

ND

PH

AS

EC

+VS

DG

ND

A

SG

ND

PU

MP

EN

VC

C

+ -V-

V+

BBS

GND

U11

LM31

1N

2 37

85

6

4

1

R28

10K

1 2

SG

ND

VC

C

VC

C

PH

AS

EA

C44

0.1u

F

1 2DOOR SWITCH DETECT

+VS

C43

3.3n

F

1 2

R31

100K

12

VC

C

R32

562

1%

1 2

PUMP DRIVER

R23

22K

1 2

SG

ND

GN

D

VC

C

R26

1

1 2

R24

10K

1%

1 2

SG

ND

SG

ND

U9

L293

E

2 1 9 12 11 19

3 8 13 18 5 6 15

20 10164 7 14 17

IN1

EN

AIN

2

IN3

EN

BIN

4

OU

T1

OU

T2

OU

T3

OU

T4

GN

DG

ND

GN

D

VS

S

VS

GN

D

SE

NS

E1

SE

NS

E2

SE

NS

E3

SE

NS

E4

+VS

DLE

D+

U10

L297

4 5 6 7 8 9 14 13 15

17 18 19 20 11 10

212161

3

AIN

H1 B C

INH

2 D

SE

NS

E1

SE

NS

E2

VR

EF

CW

/CC

WS

TEP

HA

LF/F

ULL

RE

SE

T

CO

NTR

OL

EN

AB

LE

GN

DV

S

OS

C

SY

NC

HO

ME

R33

2.2K

1 2

C42

0.1u

F

1 2

+C

4710

uF/1

6V

1 2

PU

MP

STE

P

R27

100

1 2

D11

1N40

04

21

ZS

2

DIP

20

+C

4610

0uF

/50V

1 2

VC

C

SG

ND

Page 74: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

A-60 Service manual · Version 2.0

5 Electronic diagrams 91xx pH/Electrolyte AnalyzersElectronic diagrams

SBC PCB - input amplifier - 9180, 9181

Figure A-25 SBC PCB - input amplifier - 9180, 9181

+5V

A

ZP

RO

BE

2

R4W

W

-

U12D

TLC

279C

N

13

12

14

+5V

A

Na

C50

3.3

uF

/50V

P3

PR

OB

E

A

ZF

ST

-P2P

5-1

FA

ST

EN

ER

3

NOTE: PROBE THRU IN+ SHIELDED

"Na"

R45

43.2

K 1

%

12

Cl,Li,or Ca

KO

UT

R115

1K

1 2

R48

243K

1%

12

ZP

RO

BE

3

R4W

W

"Cl/Ca/Li"

+5V

A

C51

0.1

uF

12

R41

1K

1%

12

TP

11

1

Cl/C

a/L

iOU

T

P2

PR

OB

E

"K"

V-

C49

3.3

uF

/50V

-

U12B TLC

279C

N

6 57

R47

66.5

K 1

%

12

OFFSET = 0

-

U12A

TLC

279C

N

2 31

114

U42

LM

385B

XZ

-2.5

1 2

3

C52

0.1

uF

1 2TP

9

1

OFFSET = 700mV

NOTE:

R35

100K

12

U13D

LM

C6064

13

12

14

ZD

501

5G

SC

HE

MA

TIC

, MA

IN B

OA

RD

91

80 /

91

81

711

Monday

, M

ay

03,

1999

Title

Siz

eD

ocum

ent

Num

ber

Rev

Date

:S

heet

of

-5V

ANOTE: PROBE THRU IN+ SHIELDED

SHOULD BE CLEAR OF SOLDER MASK

NOTE: NO SOLDER MASK IN

PROBE AREA

R64

1K

12

R114

1K

1 2

GUARD RING AND PROBE AREAS

R44

59K

1%

12

ZP

RO

BE

5

R4C

R

OFFSET GENERATOR

R40

1K

1%

12

R65

1K

12

-5V

A

P4

PR

OB

E

OFFSET = 2200mV

NOTE: PROBE THRU IN+ SHIELDED

R63

1K

12

R36

100K

12

R113

1K

1 2

R34

100K

12

ZP

RO

BE

4

R4W

W

K

NaO

UT

R46

68.1

K 1

%

12

C48

3.3

uF

/50V

Ref

R42

82K

1 2Z

SP

HLD

R1

PR

OB

EH

OLD

ER

P5

PR

OB

E

TP

10

1

U13C

LM

C6064

9

10

8

-2.5V

-5V

A

R39

1K

1%

12

INPUT AMPLIFIERS

ZF

ST

-P2P

5-2

FA

ST

EN

ER

3

-

U12C TLC

279C

N

9

10

8

U13B

LM

C6064

6 57 U

13A

LM

C6064

2 31

114

Page 75: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 A-61

91xx pH/Electrolyte Analyzers 5 Electronic diagramsElectronic diagrams

SBC PCB - temperature circuit - 9180, 9181

Figure A-26 SBC PCB - temperature circuit - 9180, 9181

TEM

P

R52

10K

1%

12

R62

1K

12

R11

010

0K

1 2

R51

10K

1%

1 2

A

U18

MA

X674

2 456

VIN

GN

DTR

IMVO

R53

274K

1%

1 2

TP13

1

R49

7.5K

1%

1 2

R59

118K

1%

12

R55

68.1

K 1

%

12

+5V

A

TEM

PO

UT

R56

90.9

K 1

%

1 2

ZD50

15G

SCHE

MAT

IC, M

AIN

BOAR

D 91

80 /

9181

811

Mon

day,

May

03,

199

9

Title

Siz

eD

ocum

ent

Num

ber

Rev

Dat

e:S

heet

of

D16

1N41

48

21

R54

10K

POT

1 3

2

-5V

A

"TEMP ADJ."

+-

U17

B

LF41

2

567

"10.00V"

+10V

+24V

TP12 1

+-

U17

A

LF41

2

321

84

"TEMP"

Page 76: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

A-62 Service manual · Version 2.0

5 Electronic diagrams 91xx pH/Electrolyte AnalyzersElectronic diagrams

SBC PCB - analog selector and A/D converter - 9180, 9181

Figure A-27 SBC PCB - analog selector and A/D converter - 9180, 9181

ADR

UN

-5VA

AD

DA

T5

R71

200K

1%

1 2

VC

C

AD

DA

T0

AD

HB

EN

-

U20

ICL7

109

16 15 14 13 12 11 10 9 8 7 6 5 4 3 2

35 34 33 36 39 37 38 29 30 31 32 17 21 18 19 20 26 27 24 25 22 2328140

B1

B2

B3

B4

B5

B6

B7

B8

B9

B10

B11

B12 OR

PO

L

STA

TUS

INH

IIN

LOC

OM

MO

N

REF

IN+

REF

IN-

RE

FC

AP

+R

EF

CA

P-

REF

OU

T

BU

FF

ER

AZ

INT

TES

TM

OD

ELB

EN

HB

EN

CE

/LO

AD

RU

N/H

OLD

SE

ND

OS

CS

EL

BU

FF

OS

CO

UT

OS

CIN

OS

CO

UT

V-

GN

DV+

AD

STA

TUS

C57

1uF 1 2

R69

1K

12

2.05V

R60

100

1 2

AD

DA

T1

"2.500V"

R66

1K1 2

AD

DA

T2

R68

1K

12

+2.500V

AD

SE

L2

U21

MC

1403

1 32

VIN

GN

DV

OU

T

R73

4.53

K 1

%

1 2

R61

100K

12

A/D CONVERTER

AD

DA

T4

VC

C

+5V

A

Cl/C

a/Li

OU

T

-5V

A

ANALOG CHANNEL SELECTOR

AD

CE

-

ADD

AT[0

..7]

AD

DA

T0U

19

4051

13 14 15 12 1 5 2 4 6 11 10 9

3 8716

X0 X1 X2 X3 X4 X5 X6 X7 INH

A B C

X

VS

S

VE

E

VD

D

AD

SE

L0

DG

ND

TEM

PO

UT

AD

DA

T7

R70

1K

1 2

A

GN

D

C54

1nF

12

AD

DA

T1

AD

DA

T3

TP14

TP 1

1 LSB APPROXIMATELY 1mV

NaO

UT

+5V

AA

DD

AT6

AD

DA

T5

AD

DA

T2

AD

DA

T3

NOTE:

R67

1K

12

AD

LBE

N-

KO

UT

ZD50

15G

SCHE

MAT

IC, M

AIN

BOAR

D 91

80 /

9181

911

Mon

day

, M

ay 0

3, 1

999

Ti

tle

Siz

eD

ocum

ent

Num

ber

Rev

Dat

e:S

heet

of

AD

DA

T[0.

.7]

C55

330n

F

1 2

X1 3.58

MH

z

12

AD

DA

T4

C56

150n

F

1 2

AD

SE

L1

R72

1K 1

%

12

Page 77: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 A-63

91xx pH/Electrolyte Analyzers 5 Electronic diagramsElectronic diagrams

SBC PCB - microprocessor - 9180, 9181

Figure A-28 SBC PCB - microprocessor - 9180, 9181

TB

MO

TE

N-

D8

A10

A[0

..1

8]

R84

10K

12

U30

A1

48

9

13

2

A17

D8

D1

4

+5V

A

VC

C

GN

D

A13

D1

1

MO

SI

VA

LV

E-

RO

TC

LK

BE

RR

FE

ED

RE

SE

T-

A6

A9

A10

DIS

PB

F

A[0

..18

]

D1

2

SA

MP

LE

A3

A8

VC

C

VC

C

LG

SE

L1

ZS

4

DIP

32

D[8

..1

5]

VC

C

D9

ZS

5

DIP

28

A6

A9

R92

71

K

1 2

A2

U2

9A

14

88

23

A0

V+

AD

OS

TR

A0

CT

SIN

D1

71N

41

48

2 1

D1

0

A18

C58

0.1

uF

1 2

MIS

O

D1

3

A11

U2

6

62

64

10 9 8 7 6 5 4 3

25

24

21

23 2

20

26

27

22

11

12

13

15

16

17

18

19

A0

A1

A2

A3

A4

A5

A6

A7

A8

A9

A1

0A

11

A1

2

CS

1C

S2

WE

OE

D0

D1

D2

D3

D4

D5

D6

D7

D1

5

C59

22

pF

1 2

VC

C

CW

/CC

W

D1

0

VC

C

BE

RR

D1

1

D20

1N

41

48

2 1

PR

ST

R

A4

R91

03

30

K

12

VC

C

A7

ZP

WB

1

PW

B

DO

OR SC

K

R9

12

10K

12

ZD

50

15

G

SC

HE

MA

TIC

, MA

IN B

OA

RD

918

0 /

918

1

10

11

Mo

nd

ay

, M

ay

03

, 19

99

T

itle

Siz

eD

ocu

me

nt

Num

be

rR

ev

Date

:S

he

et

of

TA

CH

O

A1

2

D1

4

LG

SE

L0

A2

A15

FP

DE

T

A0

D[8

..15

]

VC

C

A9

A18

VC

C

D9

C7

60.1

uF

1 2

VC

C

D1

4

A7

A3

R9

13

10K

12

A1

D1

2

A12

D8

A8

RAM

LW

ND

PO

S

VC

C

A4

AD

STA

TU

S

R7

8

10K

12

R91

1

10

K

12

VC

C

VC

C

A2

D1

2

CPU

A11

R8

010

K

1 2

CT

S

PU

MP

EN

VC

C

RS

232

OU

T

NO

-

A12

D8

A6

U2

9B

14

88

4 56

R7

6

10

K

12

DIS

TR

X2

32

.768

KH

z

12

D1

1

U3

0C

14

89

10

8

9

CONN

V+

LA

MP

EN

LG

SE

L3

A7

C6

022

pF

1 2

D1

5

A9

D1

3

+C

77

1u

F2

0V

1 2

A15

NON-VOLATILE

VC

C

RS

23

2IN

D1

5

A16

V+

FA

ST

/SLO

W-

R9

09

10

K

12

BDM

A

D9

A7

VC

C

R8

510

K

12

R11

6

10K

12

R83

10K

12

A3

A5

A4

A14

U2

9D

14

88

12

13

11

TH

ER

M-

A1

D1

2

D1

4

R81

10K

12

PU

MP

STE

P

R86

10K

1 2

EPROM

OPTIONAL

YE

S-

D22

1N

41

48

21

U40

27C

04

0

12

11

10 9 8 7 6 5

27

26

23

25 4

28

29 3

22

24

13

14

15

17

18

19

20

21

230

31

1

A0

A1

A2

A3

A4

A5

A6

A7

A8

A9

A1

0A

11

A1

2A

13

A1

4A

15

CE

OE

O0

O1

O2

O3

O4

O5

O6

O7

A1

6A

17

A1

8

VP

P

A3

A12

R9

25

10

K

12

VC

C

CH

EC

K

UP

TB

PO

S

A8

VC

C

U30

B1

48

9

46

5

TA

BL

EP

RE

S

DG

ND

A1

CT

SIN

PROGRAM

VC

C

D1

0

A1

1

UP

ND

PO

S

VC

C

A16 D

11

D8

A1

0

A1

RAM

A0

A4

LG

SE

L2

VC

C

D1

3

A13

D1

3

VA

STR

A6

JP

9

HE

AD

ER

2X

5

1 2 3 4 5 6 7 8 9 10

A5

A10

U2

9C

14

88

9

10

8

U2

2

MK

48T

08

10 9 8 7 6 5 4 3

25

24

21

23 2

20

26

27

22

11

12

13

15

16

17

18

19

14

28

1

A0

A1

A2

A3

A4

A5

A6

A7

A8

A9

A10

A11

A12

CS

1C

S2

WE

OE

D0

D1

D2

D3

D4

D5

D6

D7

GN

D

VC

C

INT

D1

5

U3

1

MC

68

HC

16Z

1

1 82

14

05

96

18

49

811

312

4

71

70

69

42

78

77

76

75

74

73

72

44

45

46

47

48

49

52

53

50

51

12

812

712

612

9 313

213

113

0

17

185 6 7

12

10

11

13

14

15

16

54

43 2 9

22

29

41

58

62

83

97

10

511

212

5

90

19

20

23

24

25

26

27

28

30

31

32

33

34

35

36

37

38

39

11

912

012

112

212

3

10

910

810

710

610

410

310

210

110

099

96

95

94

93

92

91

85

86

80

89

88

82

81

87

66

67

68

64

65

11

011

611

711

811

111

411

563

56

60

57

55

79

4

VD

DI

VD

DE

VD

DE

VD

DE

VD

DI

VD

DE

VD

DE

VD

DE

VD

DE

VD

DE

BE

RR

HA

LT

RE

SE

T

VD

DA

I1/P

F1

I2/P

F2

I3/P

F3

I4/P

F4

I5/P

F5

I6/P

F6

I7/P

F7

AD

A0

/PA

D0

AD

A1

/PA

D1

AD

A2

/PA

D2

AD

A3

/PA

D3

AD

A4

/PA

D4

AD

A5

/PA

D5

AD

A6

/PA

D6

AD

A7

/PA

D7

VR

HV

RL

PW

MA

PW

MB

PC

LK

PA

I

OC

2/O

C1

/PG

P4

OC

3/O

C1

/PG

P5

OC

4/O

C1

/PG

P6

IC4/O

C5

/OC

1/P

GP

RX

DTX

D

IC3/P

GP

2IC

2/P

GP

1IC

1/P

GP

0

SC

KM

ISO

MO

SI

PC

S0

/SS

PC

S1

PC

S2

PC

S3

VS

TB

YV

SS

AV

SS

IV

SS

EV

SS

EV

SS

IV

SS

EV

SS

IV

SS

EV

SS

EV

SS

EV

SS

IV

SS

EV

SS

E

A0

A1

A2

A3

A4

A5

A6

A7

A8

A9

A10

A11

A12

A13

A14

A15

A16

A17

A18

A1

9/C

S6

/PC

3A

20/C

S7

/PC

4A

21/C

S8

/PC

5A

22/C

S9

/PC

6A

23/C

S10

D0

D1

D2

D3

D4

D5

D6

D7

D8

D9

D10

D11

D12

D13

D14

D15

AS

/PE

5D

S/P

E4

R/W

DS

AC

K0/P

E0

DS

AC

K1/P

E1

SIZ

0/P

E6

SIZ

1/P

E7

AV

EC

/PE

2

BK

PT/D

SC

LK

IPIP

E0

/DS

0IP

IPE

1/D

S1

FR

EE

ZE

/QU

OT

TS

TM

E/T

SC

CS

BO

OT

BR

/CS

0B

G/C

S1

BG

AC

K/C

S2

FC

0/C

S3

/PC

0F

C1/C

S4

/PC

1F

C2/C

S5

/PC

2C

LC

KO

UT

VD

DS

YN

XF

CE

XT

AL

XT

AL

MO

DC

LK

/PF

0

OC

1/P

GP

3

RTS

LW

TB

PO

S

A8

R9

24

1K

12

VC

C

A2

A5

D21

1N

41

48

21

L1

10u

H1

2

VC

C

A14

A17

R1

17

4.7

K

12R8

210

K

12

AD

IST

R

A11

D9

A5

FP

CL

K

BU

ZZ

ER

ND

MO

TE

N-

D1

0

R7

4

10

M

12

Page 78: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

A-64 Service manual · Version 2.0

5 Electronic diagrams 91xx pH/Electrolyte AnalyzersElectronic diagrams

SBC PCB - microprocessor - 9180, 9181

Figure A-29 SBC PCB - microprocessor - 9180, 9181

ZD50

15G

SCHE

MAT

IC, M

AIN

BOAR

D 91

80 /

9181 11

11M

onda

y, M

ay 0

3, 1

999

Ti

tle

Siz

eD

ocum

ent

Num

ber

Rev

Dat

e:S

heet

of

Q3

2N39

04

SAM

PLE

LED

-

DO

T5-

MIS

O

R99

5.1K

1 2

+5VA

ADD

AT0

DIS

TR

DO

T7-

SAM

PLE

ADD

AT3

R97

10K

1 2

R91

10K

12

DIS

PBF

DD

[0..

7]

R92

9

4.7K

12

U33

LB12

56

10 18 17 16 15 14 13 12 11 9

1 2 3 4 5 6 7 8

VC

CO

UT1

OU

T2O

UT3

OU

T4O

UT5

OU

T6O

UT7

OU

TMG

ND

IN1

IN2

IN3

IN4

IN5

IN6

IN7

INM

OP

EN2

BZ1

PKM

22E

PP-

4001

1 2

R93

5KP

OT

1 3

2

SSE

N

R92

1K

1 2

R96

1K

1 2

VC

C

R92

8

4.7K

12

ADDATA

DD

6

Q4

2N39

04

ADD

AT2

DSPDATA

+5VA

RE

SET-

DO

T2-

ADH

BE

N-

DG

ND

AD

CE-

TP15 1

R95

100

1 2

DO

T4-

PRCTRL

SS

EN

Q6

2N39

04

AD

DA

T[0.

.7]

Q5

2N39

06

VAS

TR

OP

EN3

ADD

AT7

VCC

DIS

PEN

DD

5

VC

C

ADS

EL0

R87

150

1 2

VCC

U36

74H

C59

4

14 11 10 12 13

15 1 2 3 4 5 6 7 9

SER

SR

CLK

SR

CLR

RC

LKR

CLR

QA

QB

QC

QD QE

QF

QG

QH

QH

'

PGN

D

DD

4

R88

4.7K

1 2

DD

1

DD

7

R94

1K

12

SCK

U30

D14

89

1311

12

DO

T1-

CO

NTR

AST

DD

3

ADCTRL

ADD

AT6

+5V

A

VC

C

U37

74H

C59

4

14 11 10 12 13

15 1 2 3 4 5 6 7 9

SER

SR

CLK

SR

CLR

RC

LKR

CLR

QA

QB

QC

QD QE

QF

QG

QH

QH

'

VACTRL

DD

2 + -

U39

A

LF35

3

3 21

8 4

U34

74H

C59

4

14 11 10 12 13

15 1 2 3 4 5 6 7 9

SER

SR

CLK

SR

CLR

RC

LKR

CLR

QA

QB

QC

QD QE

QF

QG

QH

QH

'

DD

[0..

7]

DIS

PR

/W-

DD

0

VC

C

OP

EN1

OP

EN4

ADLB

EN

-

R93

1

1.5K

12

DO

T3-

A

ADD

AT1

R93

0

4.7K

12

+5VA

PG

ND

R98

10K

12

"SS"

ADD

AT5

R90

1K

1 2

R10

210

K1

2

DO

T6-

-5V

A

"SS ADJ."

VCC

BU

ZZE

R

ADS

EL1

U38

74H

C16

5

10 11 12 13 14 3 4 5 6 2 15 1

9 7

SER

A B C D E F G H CLK

INH

SH

/LD

QH

QH

PR

STR

ADS

EL2

PR

INTV

CC

MO

SI

AD

DA

T[0.

.7]

DAT

A/I

NST

-

V+

VC

C

U35

74H

C59

5

14 11 10 12 13

15 1 2 3 4 5 6 7 9

SER

SR

CLK

SR

CLR

RC

LKG

QA

QB

QC

QD QE

QF

QG

QH

QH

'

AD

OST

R

MO

TOR

-

ADIS

TR

AD

RU

N

OP

EN5

GN

D

ADD

AT4

Page 79: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 A-65

91xx pH/Electrolyte Analyzers 5 Electronic diagramsElectronic diagrams

SBC PCB - Fluid pack detector - 9180, 9181

Figure A-30 SBC PCB - Fluid pack detector - 9180, 9181

ZD50

15G

SCHE

MAT

IC, M

AIN

BOAR

D 91

80 /

9181

311

Mon

day,

May

03,

199

9

Ti

tle

Siz

eD

ocum

ent

Num

ber

Rev

Dat

e:S

heet

of

D18

1N41

48

21

CP

U41

LM29

07N

-8

18

2

3

45

6

7

VC

C

C81

0.01

uF

1 2

R92

03.

3K

1 2

R91

83.

9K

1 2

R91

962

0

1 2

R91

4

1K

12

C79

1.0u

F

1 2

R91

616

0K

1 2

R92

13.

0K

1 2

R91

516

01 2

ISO

1

OP

B71

2

213 4

R91

720

0

1 2

Q14

2N39

04C

7810

00pF

1 2

VC

C

V+

C80

0.01

uF

12

R92

34.

3K

1 2

FP

DET

V+

VC

C

FP

CLK

A

R92

2

4.3K

12

Page 80: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

A-66 Service manual · Version 2.0

5 Electronic diagrams 91xx pH/Electrolyte AnalyzersElectronic diagrams

Lamp PCB - 9180, 9181

Figure A-31 Lamp PCB - 9180, 9181

JP7

AM

P 1

0367

0-3

1 2 3 4JP1

AM

P 1

0367

0-2

1 2 3

PUMP

JP2

AM

P 1

0367

0-2

1 2 3

VALVE R

DC

LOS

ED

PH

AS

EA

VA

LVE

B-

VALVE V

VALVE C

VA

LVE

V-

U1

HLM

P28

852 3 6 7 10 11 14 15

1 4 5 8 9 12 13 16

JP3

AM

P 1

0367

0-2

1 2 3

ZP

WB

1

PW

B

U2

HLM

P28

852 3 6 7 10 11 14 15

1 4 5 8 9 12 13 16

U3

HLM

P28

852 3 6 7 10 11 14 15

1 4 5 8 9 12 13 16

VALVE B

VA

LVE

+

NEEDLE MECHANISM

DLE

D+

JP6

AM

P 1

0367

0-5

1 2 3 4 5 6

JP4

AM

P 1

0367

0-2

1 2 3

LED ARRAY

ZD50

14B

SCHE

MAT

IC, L

AMP

BOAR

D, 9

180/

9181

.

B1

1F

riday

, Ja

nuar

y 3

1, 1

997

Title

Siz

eD

ocum

ent

Num

ber

Rev

Dat

e:S

heet

of

VA

LVE

A-

VALVE A

PH

AS

ED

VA

LVE

R-

JP9

AM

P 1

0367

0-9

1 2 3 4 5 6 7 8 9 10

LED ARRAY

OV

ER

TEM

P-

PH

AS

EC

JP8

HE

AD

ER

34

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34

JP5

AM

P 1

0367

0-2

1 2 3

LED ARRAY

lamp.sch

VA

LVE

C-

DOOR SENSOR

Page 81: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 A-67

91xx pH/Electrolyte Analyzers 5 Electronic diagramsElectronic diagrams

Display daughter PCB - 9180

Figure A-32 Display daughter PCB - 9180

SAMPLEGND

S1

IP1241.1082.3

1 2

SCHURTER

DD4 DD4

DD2

JP2

HEADER 34

123456789

10111213141516171819202122232425262728293031323334

DD5

DATA/INST-

DD1

SAMPLEPT

CONTRAST

+5VA

DISPR/W-

DD3

DD1

+5VAJP4

22-03-2141

1234567891011121314

MOLEX 15-43-8566

VCC

DD3

DD0

DD5DD6

SAMPLESENSOR

NO-

ZPWB1

PWB

SAMPLELED-

YES-

VCC

DD[0..7]

DD7

YES+24V

NOS2

IP1241.1082.3

1 2

DD0

TEMP

LCDDISPLAY

VCC

ZD5016

SCHEMATIC, DISPLAY BOARD, 9180

1 1Monday , July 22, 2002

Title

Size Document Number Rev

Date: Sheet of

A C

DD2

JP1

RJ12

123456

MOLEX

DD7DD6

SCHURTER

+5VA

Page 82: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

A-68 Service manual · Version 2.0

5 Electronic diagrams 91xx pH/Electrolyte AnalyzersElectronic diagrams

Page 83: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

April 2007

6 Software operation 9110 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-3

7 Software operation 9120, 9130, 9140 . . . . . . . . . . . . . . . . . . . . B-17

8 Software operation 9180 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-31

9 Software operation 9181 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-47

10 Service functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-63

Software operation B

Page 84: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte
Page 85: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 B-3

91xx pH/Electrolyte Analyzers 6 Software operation 9110Table of contents

Software operation 9110

Software operation 9110 ............................................................................................. B-5

Overview ................................................................................................................ B-5

Overall program operation flow ........................................................................... B-6

Main menu ............................................................................................................ B-7

Measurement sequence ......................................................................................... B-8

Print functions menu ............................................................................................ B-9

QC sample menu ................................................................................................. B-10

Daily maintenance menu .................................................................................... B-11

Operator functions menu ................................................................................... B-12

Program instrument menu ................................................................................. B-13

Service functions menu ....................................................................................... B-14

Calibration sequence ........................................................................................... B-15

In this chapter Chapter 6

Page 86: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

B-4 Service manual · Version 2.0

6 Software operation 9110 91xx pH/Electrolyte AnalyzersTable of contents

Page 87: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 B-5

91xx pH/Electrolyte Analyzers 6 Software operation 9110Software operation 9110

Software operation 9110

Overview

Software operation of the 9110 pH Analyzers can be accessed by YES/NO selection on the analyzer front panel. The following flow chart diagrams are provided to identify operating sequences of the 9110 pH Analyzers:

1 Overall Program Operation Flow

2 Main Menu

3 Measurement Sequence

4 Print Functions Menu

5 QC/Standard/Urine Sample Menu

6 Electrode Cleaning Menu

7 Operator Functions Menu

8 Program Instrument Menu

9 Service Functions Menu

10 Calibration Sequence

Page 88: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

B-6 Service manual · Version 2.0

6 Software operation 9110 91xx pH/Electrolyte AnalyzersSoftware operation 9110

Overall program operation flow

Figure B-1 9110 program flow

Page 89: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 B-7

91xx pH/Electrolyte Analyzers 6 Software operation 9110Software operation 9110

Main menu

Figure B-2 9110 main menu

Page 90: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

B-8 Service manual · Version 2.0

6 Software operation 9110 91xx pH/Electrolyte AnalyzersSoftware operation 9110

Measurement sequence

Figure B-3 9110 measurement sequence

Page 91: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 B-9

91xx pH/Electrolyte Analyzers 6 Software operation 9110Software operation 9110

Print functions menu

Figure B-4 9110 print functions menu

Page 92: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

B-10 Service manual · Version 2.0

6 Software operation 9110 91xx pH/Electrolyte AnalyzersSoftware operation 9110

QC sample menu

Figure B-5 9110 QC sample menu

Page 93: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 B-11

91xx pH/Electrolyte Analyzers 6 Software operation 9110Software operation 9110

Daily maintenance menu

Figure B-6 9110 daily maintenance menu

Page 94: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

B-12 Service manual · Version 2.0

6 Software operation 9110 91xx pH/Electrolyte AnalyzersSoftware operation 9110

Operator functions menu

Figure B-7 9110 operator functions menu

Page 95: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 B-13

91xx pH/Electrolyte Analyzers 6 Software operation 9110Software operation 9110

Program instrument menu

Figure B-8 9110 program instrument menu

Page 96: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

B-14 Service manual · Version 2.0

6 Software operation 9110 91xx pH/Electrolyte AnalyzersSoftware operation 9110

Service functions menu

Figure B-9 9110 service functions menu

Page 97: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 B-15

91xx pH/Electrolyte Analyzers 6 Software operation 9110Software operation 9110

Calibration sequence

Figure B-10 9110 calibration sequence

Page 98: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

B-16 Service manual · Version 2.0

6 Software operation 9110 91xx pH/Electrolyte AnalyzersSoftware operation 9110

Page 99: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 B-17

91xx pH/Electrolyte Analyzers 7 Software operation 9120, 9130, 9140Table of contents

Software operation 9120, 9130, 9140

Software operation 9120, 9130, 9140 ....................................................................... B-19

Overview .............................................................................................................. B-19

Overall program operating flow ......................................................................... B-20

Main menu .......................................................................................................... B-21

Measurement sequence ....................................................................................... B-22

Print functions menu .......................................................................................... B-23

QC/Standard/urine sample menu ...................................................................... B-24

Daily maintenance menu .................................................................................... B-25

Operator functions menu ................................................................................... B-26

Program instrument menu ................................................................................. B-27

Service functions menu ....................................................................................... B-28

Calibration sequence ........................................................................................... B-29

In this chapter Chapter 7

Page 100: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

B-18 Service manual · Version 2.0

7 Software operation 9120, 9130, 9140 91xx pH/Electrolyte AnalyzersTable of contents

Page 101: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 B-19

91xx pH/Electrolyte Analyzers 7 Software operation 9120, 9130, 9140Software operation 9120, 9130, 9140

Software operation 9120, 9130, 9140

Overview

1 Overall Program Operating Flow

2 Main Menu

3 Measurement Sequence

4 Print Functions Menu

5 QC/Standard/Urine Sample Menu

6 Daily Maintenance Menu

7 Operator Functions Menu

8 Program Instrument Menu

9 Service Functions Menu

10 Calibration Sequence!

Page 102: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

B-20 Service manual · Version 2.0

7 Software operation 9120, 9130, 9140 91xx pH/Electrolyte AnalyzersSoftware operation 9120, 9130, 9140

Overall program operating flow

Figure B-11 9120, 9130, 9140 program flow

Page 103: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 B-21

91xx pH/Electrolyte Analyzers 7 Software operation 9120, 9130, 9140Software operation 9120, 9130, 9140

Main menu

Figure B-12 9120, 9130, 9140 main menu

Page 104: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

B-22 Service manual · Version 2.0

7 Software operation 9120, 9130, 9140 91xx pH/Electrolyte AnalyzersSoftware operation 9120, 9130, 9140

Measurement sequence

Figure B-13 9120, 9130, 9140 measurement sequence

Page 105: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 B-23

91xx pH/Electrolyte Analyzers 7 Software operation 9120, 9130, 9140Software operation 9120, 9130, 9140

Print functions menu

Figure B-14 9120, 9130, 9140 print functions menu

Page 106: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

B-24 Service manual · Version 2.0

7 Software operation 9120, 9130, 9140 91xx pH/Electrolyte AnalyzersSoftware operation 9120, 9130, 9140

QC/standard/urine sample menu

Figure B-15 9120, 9130, 9140 QC/Standard/urine sample menu

Page 107: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 B-25

91xx pH/Electrolyte Analyzers 7 Software operation 9120, 9130, 9140Software operation 9120, 9130, 9140

Daily maintenance menu

Figure B-16 9120, 9130, 9140 daily maintenance menu

Page 108: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

B-26 Service manual · Version 2.0

7 Software operation 9120, 9130, 9140 91xx pH/Electrolyte AnalyzersSoftware operation 9120, 9130, 9140

Operator functions menu

Figure B-17 9120, 9130, 9140 operator functions menu

Page 109: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 B-27

91xx pH/Electrolyte Analyzers 7 Software operation 9120, 9130, 9140Software operation 9120, 9130, 9140

Program instrument menu

Figure B-18 9120, 9130, 9140 program instrument menu

Page 110: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

B-28 Service manual · Version 2.0

7 Software operation 9120, 9130, 9140 91xx pH/Electrolyte AnalyzersSoftware operation 9120, 9130, 9140

Service functions menu

Figure B-19 9120, 9130, 9140 service functions menu

Page 111: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 B-29

91xx pH/Electrolyte Analyzers 7 Software operation 9120, 9130, 9140Software operation 9120, 9130, 9140

Calibration sequence

Figure B-20 9120, 9130, 9140 calibration sequence

Page 112: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

B-30 Service manual · Version 2.0

7 Software operation 9120, 9130, 9140 91xx pH/Electrolyte AnalyzersSoftware operation 9120, 9130, 9140

Page 113: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 B-31

91xx pH/Electrolyte Analyzers 8 Software operation 9180Table of contents

Software operation 9180

Software operation 9180 ........................................................................................... B-33

Overview .............................................................................................................. B-33

Overall program operation flow ......................................................................... B-34

Main menu .......................................................................................................... B-35

Measurement sequence ....................................................................................... B-36

Print functions menu .......................................................................................... B-37

QC/standard/dialysate/urine sample menu ....................................................... B-38

Daily maintenance menu .................................................................................... B-39

Operator functions menu ................................................................................... B-40

Program instrument menu ................................................................................. B-41

Service functions menu ....................................................................................... B-42

Calibration sequence (1) (Example Na/K/Li calibration) ................................. B-43

Calibration sequence (2) (Example Na/K/Li calibration) ................................. B-44

Power-up .............................................................................................................. B-45

In this chapter Chapter 8

Page 114: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

B-32 Service manual · Version 2.0

8 Software operation 9180 91xx pH/Electrolyte AnalyzersTable of contents

Page 115: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 B-33

91xx pH/Electrolyte Analyzers 8 Software operation 9180Software operation 9180

Software operation 9180

Overview

1 Overall Program Operating Flow

2 Main Menu

3 Measurement Sequence

4 Print Functions Menu

5 QC/Std/Dialysate/Urine Sample Menu

6 Daily Maintenance Menu

7 Operator Functions Menu

8 Program Instrument Menu

9 Service Functions Menu

10 Calibration Sequence

11 Power-up Sequence

Page 116: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

B-34 Service manual · Version 2.0

8 Software operation 9180 91xx pH/Electrolyte AnalyzersSoftware operation 9180

Overall program operation flow

Figure B-21 9180 program flow

Page 117: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 B-35

91xx pH/Electrolyte Analyzers 8 Software operation 9180Software operation 9180

Main menu

Figure B-22 9180 main menu

Page 118: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

B-36 Service manual · Version 2.0

8 Software operation 9180 91xx pH/Electrolyte AnalyzersSoftware operation 9180

Measurement sequence

Figure B-23 9180 measurement sequence

Page 119: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 B-37

91xx pH/Electrolyte Analyzers 8 Software operation 9180Software operation 9180

Print functions menu

Figure B-24 9180 print functions menu

Page 120: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

B-38 Service manual · Version 2.0

8 Software operation 9180 91xx pH/Electrolyte AnalyzersSoftware operation 9180

QC/standard/dialysate/urine sample menu

Figure B-25 9180 QC/standard/dialysate/urine sample menu

Page 121: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 B-39

91xx pH/Electrolyte Analyzers 8 Software operation 9180Software operation 9180

Daily maintenance menu

Figure B-26 9180 daily maintenance menu

Page 122: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

B-40 Service manual · Version 2.0

8 Software operation 9180 91xx pH/Electrolyte AnalyzersSoftware operation 9180

Operator functions menu

Figure B-27 9180 operator functions menu

Page 123: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 B-41

91xx pH/Electrolyte Analyzers 8 Software operation 9180Software operation 9180

Program instrument menu

Figure B-28 9180 program instrument menu

Page 124: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

B-42 Service manual · Version 2.0

8 Software operation 9180 91xx pH/Electrolyte AnalyzersSoftware operation 9180

Service functions menu

Figure B-29 9180 service functions menu

Page 125: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 B-43

91xx pH/Electrolyte Analyzers 8 Software operation 9180Software operation 9180

Calibration sequence (1) (Example Na/K/Li calibration)

Figure B-30 9180 calibration sequence (1) (Example Na/K/Li calibration)

Page 126: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

B-44 Service manual · Version 2.0

8 Software operation 9180 91xx pH/Electrolyte AnalyzersSoftware operation 9180

Calibration sequence (2) (Example Na/K/Li calibration)

Figure B-31 9180 calibration sequence (2) (Example Na/K/Li calibration)

Page 127: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 B-45

91xx pH/Electrolyte Analyzers 8 Software operation 9180Software operation 9180

Power-up

Figure B-32 9180 power-up

Page 128: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

B-46 Service manual · Version 2.0

8 Software operation 9180 91xx pH/Electrolyte AnalyzersSoftware operation 9180

Page 129: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 B-47

91xx pH/Electrolyte Analyzers 9 Software operation 9181Table of contents

Software operation 9181

Software operation 9181 ........................................................................................... B-49

Overview .............................................................................................................. B-49

Overall program operting flow ........................................................................... B-50

Main menu .......................................................................................................... B-51

Manual measurement sequence ......................................................................... B-52

Automated measurement sequence .................................................................... B-53

Print functions menu .......................................................................................... B-54

QC/std/dialysate/urine sample menu ................................................................. B-55

Daily maintenance menu .................................................................................... B-56

Operator functions menu ................................................................................... B-57

Program instrument menu ................................................................................. B-58

Service functions menu ....................................................................................... B-59

Calibration sequence (1) (example Na/K/Li calibration) ................................. B-60

Calibration sequence (2) (example Na/K/Li calibration) ................................. B-61

Power-up sequence .............................................................................................. B-62

In this chapter Chapter 9

Page 130: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

B-48 Service manual · Version 2.0

9 Software operation 9181 91xx pH/Electrolyte AnalyzersTable of contents

Page 131: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 B-49

91xx pH/Electrolyte Analyzers 9 Software operation 9181Software operation 9181

Software operation 9181

Overview

1 Overall Program Operating Flow

2 Main Menu

3 Measurement Sequence

4 Print Functions Menu

5 QC/Std/Dialysate/Urine Sample Menu

6 Daily Maintenance Menu

7 Operator Functions Menu

8 Program Instrument Menu

9 ervice Functions Menu

10 Calibration Sequence

11 Power-up Sequence

Page 132: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

B-50 Service manual · Version 2.0

9 Software operation 9181 91xx pH/Electrolyte AnalyzersSoftware operation 9181

Overall program operting flow

Figure B-33 9181 program flow

Page 133: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 B-51

91xx pH/Electrolyte Analyzers 9 Software operation 9181Software operation 9181

Main menu

Figure B-34 9181 main menu

Page 134: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

B-52 Service manual · Version 2.0

9 Software operation 9181 91xx pH/Electrolyte AnalyzersSoftware operation 9181

Manual measurement sequence

Figure B-35 9181 manual measurement sequence

Page 135: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 B-53

91xx pH/Electrolyte Analyzers 9 Software operation 9181Software operation 9181

Automated measurement sequence

Figure B-36 9181 automated measurement sequence

Page 136: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

B-54 Service manual · Version 2.0

9 Software operation 9181 91xx pH/Electrolyte AnalyzersSoftware operation 9181

Print functions menu

Figure B-37 9181 print functions menu

Page 137: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 B-55

91xx pH/Electrolyte Analyzers 9 Software operation 9181Software operation 9181

QC/std/dialysate/urine sample menu

Figure B-38 9181 QC/std/dialysate/urine sample menu

Page 138: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

B-56 Service manual · Version 2.0

9 Software operation 9181 91xx pH/Electrolyte AnalyzersSoftware operation 9181

Daily maintenance menu

Figure B-39 9181 daily maintenance menu

Page 139: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 B-57

91xx pH/Electrolyte Analyzers 9 Software operation 9181Software operation 9181

Operator functions menu

Figure B-40 9181 operator functions menu

Page 140: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

B-58 Service manual · Version 2.0

9 Software operation 9181 91xx pH/Electrolyte AnalyzersSoftware operation 9181

Program instrument menu

Figure B-41 9181 program instrument menu

Page 141: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 B-59

91xx pH/Electrolyte Analyzers 9 Software operation 9181Software operation 9181

Service functions menu

Figure B-42 9181 service functions menu

Page 142: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

B-60 Service manual · Version 2.0

9 Software operation 9181 91xx pH/Electrolyte AnalyzersSoftware operation 9181

Calibration sequence (1) (example Na/K/Li calibration)

Figure B-43 9181 calibration sequence (1) (example Na/K/Li calibration)

Page 143: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 B-61

91xx pH/Electrolyte Analyzers 9 Software operation 9181Software operation 9181

Calibration sequence (2) (example Na/K/Li calibration)

Figure B-44 9181 calibration sequence (2) (example Na/K/Li calibration)

Page 144: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

B-62 Service manual · Version 2.0

9 Software operation 9181 91xx pH/Electrolyte AnalyzersSoftware operation 9181

Power-up sequence

Figure B-45 9181 power-up

Page 145: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 B-63

91xx pH/Electrolyte Analyzers 10 Service functionsTable of contents

Service functions

Insert introducing text here.

Service functions ....................................................................................................... B-65

TEST ELECTRODES ........................................................................................... B-65

TEST SAMPLE SENSOR .................................................................................... B-65

TEST SnapPak SENSOR (9180 and 9181 only) ................................................. B-65

TEST LANGUAGE SWITCH (9180 and 9181 only) ......................................... B-65

TEST SAMPLE DOOR (except 9181) ................................................................ B-66

TEST PROBE (9181 only) ................................................................................... B-66

TEST AUTOSAMPLER (9181 only) ................................................................... B-66

TEST PUMP ........................................................................................................ B-66

TEST PINCH VALVES ........................................................................................ B-67

TEST INTERFACE .............................................................................................. B-67

TEST AMPLIFIER ............................................................................................... B-67

ENTER SERVICE CODE .................................................................................... B-68

Miscellaneous functions ...................................................................................... B-70

Deleting data .................................................................................................. B-70

Warm-up bypass ............................................................................................ B-70

In this chapter Chapter 10

Page 146: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

B-64 Service manual · Version 2.0

10 Service functions 91xx pH/Electrolyte AnalyzersTable of contents

Page 147: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 B-65

91xx pH/Electrolyte Analyzers 10 Service functionsService functions

Service functions

The Service Functions mode provides a menu of service test functions to assist in failure diagnosis and identification. A logic flow chart is provided for reference of the sequences included in the Service Functions mode.

TEST ELECTRODES

This mode allows for the measurement and display of electrode voltages.

Standard A, Standard B or an external sample can be measured to determine electrode performance and provide measurement data useful in locating electrode calibration faults. Each electrode parameter will be displayed and the electrode voltage associated with the sample type selected can be observed.

Example displays

TEST SAMPLE SENSOR

In this mode, the sample sensor output is displayed which is useful for electrical adjustment or for troubleshooting sample sensor faults. A dry sample sensor should display a value of 100 with a range of 80 to 120. After verifying a dry sensor, aspiration of a clear fluid (i.e., water) should increase the displayed value by at least 40 units.

Example displays

TEST SnapPak SENSOR (9180 and 9181 only)

In this test, the SnapPak Sensor detects whether the ISE SnapPak™ is in place. "I" indicates that the ISE SnapPak™ is in place, "O" indicates that it is either not present or not completely seated.

Example display

TEST LANGUAGE SWITCH (9180 and 9181 only)

This mode displays the language selected for display and printout. To change the selection, set the language switch (located behind printer) to the desired language. The display will indicate the language choice according to the switch setting. To activate a new language, turn the analyzer off and back on with the switch set to the desired language.

St.A: Na=1582

K=654 Cl=-904

St.B:

pH=-808

Ext: Na=1321

K=-598 Cl=-127

(9110 pH Analyzer)

DRY 80-120: 102

YES=Pump/NO=Exit

External: 153

Pump: NO=OFF

DRY 135-145: 141

YES=Pump/NO=Exit

(9110 pH Analyzer)

SnapPak: I

I=IN O=OUT

Page 148: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

B-66 Service manual · Version 2.0

10 Service functions 91xx pH/Electrolyte AnalyzersService functions

Example display

TEST SAMPLE DOOR (except 9181)

This test verifies the operation of the sample door light gate and its associated circuitry. The display indicates an “O” or “C” depending on the position of the door.

Example display

TEST PROBE (9181 only)

This test verifies the operation of the sample probe motor, position sensor and associated circuitry. The "U" represents the output of the upper position sensor, the "L" represents the output of the lower position sensor.

Example display

TEST AUTOSAMPLER (9181 only)

This test verifies the operation of the Autosampler turntable assembly. The display indicates the output of both the position and home sensors.

Example display

TEST PUMP

This test turns on the peristaltic pump at each of the four pump speeds. The display indicates for two seconds each pump speed while each pump speed is being tested. Observe that the peristaltic pump changes speed as is indicated on the display.

LanguageSwitch

English

Figure B-46 Example 9180 language label

Polski

Português

Spare 3

Spare 4

English

Deutsch

Français

Español

Italiano

SampleDoor:C

C:Closed/O:Open

U=0 L=1

Sample Position?

Note

If the Autosampler is not plugged in or the TABLE PRESENT bit is not activated, this test will not

appear under the [SERVICE FUNCTIONS] menu.

Home=0

Pos=1

Page 149: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 B-67

91xx pH/Electrolyte Analyzers 10 Service functionsService functions

Example displays

TEST PINCH VALVES

This test turns on/off each solenoid valve as indicated and is useful to verify both electronic control and mechanical operation of each solenoid valve.

Example displays

TEST INTERFACE

The Interface Test can be used to verify correct operation of the RS232C serial port. To operate this test, jumper pin 2 and pin 3 of the rear interface connector prior to invoking the test. During the interface test, the analyzer sends out a test string and checks if it is received within a set period of time. If correct operation is observed after completion of the test, [INTERFACE OK] will be displayed.

TEST AMPLIFIER

The Amplifier Test displays ground, reference and output voltage of each input amplifier. The values displayed for each electrode channel will be dependent on the electrode inputs and may vary. Display of the ground should read between -10 mV and +10 mV with the reference voltage reading between -2490 mV and -2510 mV. With all inputs open, the display should read +/- 4095 for each channel. On the 9140/9180/9181, an additional display indicating the temperature voltage and temperature in degrees C is available to verify the temperature adjustment. On the 9110, the temperature reading should be between 36.8 and 37.2°C.

Example display

---Test Pump!---

Very Slow Speed

---Test Pump!---

Slow Speed

---Test Pump!---

Medium Speed

---Test Pump!---

Fast Speed

Test Valve A:

YES/NO = ON/OFF

Test Valve B:

YES/NO = ON/OFF

Test Valve C:

YES/NO = ON/OFF

(9180/9181 only)

Test Valve V:

YES/NO = ON/OFF

Test Valve R:

YES/NO = ON/OFF

Na=#### K=####

Cl/Ca/Li=####

GND =0

-2500mV=-2499

Temp(mV) =-####

Temp(°C) = ##.#

(depending on parameter configuration) (9180/9181 only)

pH=####

(9110 only)

Page 150: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

B-68 Service manual · Version 2.0

10 Service functions 91xx pH/Electrolyte AnalyzersService functions

ENTER SERVICE CODE

Service Codes are available and allow additional programming features. Following is a listing of the available Service Codes and functions:

Set Reset Service code description

ALL LLA This service code will enable the Ca2+ value during auto-(9181 only)

matic serum/blood measurement. Special precautions are necessary

to obtain the correct Ca2+ value due to sample exposure to room air.

CDC – Clears all Service Codes.

9110: Except SLC.

DEC CED Provides one additional digit of resolution for each measured

parameter for blood and serum samples. QC and Standard samples

are always displayed with high resolution. Urine samples are always

displayed in low resolution.

ECO OCE Allows the operator to run the instrument in the economy mode. The

instrument will enter the STANDBY MODE. This occurs at the time

of the next 2-point calibration when no samples have been analyzed

since the last 2-point calibration.

FIF IFI Allows printing of electrode voltages during printout of

measurements and calibrations. (not available for 9140)

LEM MEL 9180/9181 only!

This service code will print the last 20 error messages.

MGL LGM This service code is available only in the model 9140/9180/9181

analyzer and enables Ca2+ values to be displayed in mg/dL. The

normal mode displays values in mmol/L. QC and Standard results are

always displayed in mmol/L.

NOB BON 9180/9181 only!

This service code disables the beep.

QCC CCQ When this code is enabled, QC values will be reported in direct ISE

measurement instead of flame-equivalent values. In this mode,

correlation factors are NOT applied to the measured values.

SFC – 9180/9181 only!

This code allows the input of a percentage for the ISE SnapPakTM

counter.

SLC CLS 9110/9180/9181 only!

When this code is set, the current setting of the leasing counter, which

cannot be reset by the operator, is printed on the calibration report.

SSC CSS 9181 only!

This code toggles needle height to accommodate smaller (9181 only)

sample cups. In the default setting, the analyzer works with 2.0 mL

cups. To select 0.25 mL sample cups, SSC has to be activated. To

return to the default setting, CSS has to be activated.

CHP PHC This code allows for the display and printout of pH results in

concentration of H+ ions. The cH+ will be shown in addition to the

pH value. cH+ is not shown on QC values and statistics reports.

Table B-1 Service code descriptions

Page 151: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 B-69

91xx pH/Electrolyte Analyzers 10 Service functionsService functions

EXW WXE 9110 only!

This code extends the wash cycle following a serum or blood

measurement. Normally 10 Standard A segments are used for the

wash cycle; with EXW set, 20 Standard A segments will be used.

NTD DTN 9110 only!

When the code TMP is set, NTD prevents the temperature from

(9110 only) being reset after each sample and causes the last

temperature value to remain until changed.

TMP PMT 9110 only!

When this code is set, the [CHANGE PATIENT TEMPERATURE?]

menu will appear in the sampling process. The analyzer will perform

a temperature correction calculation to convert the measured results

for the desired patient temperature. The temperature will be reset

after each sample to 37°C unless NTD is entered. Temperature

corrections can only be performed in the range of 15 - 45°C.

Code Description 9110 9120 9130 9140 9180 9181

ALL Enables Ca2+ during automat. blood/

serum measuement

X

CHP Display results in concentratio of cH+ ions X

CDC Clears all Service Codes X X X X X X

DEC Additional display resolution X X X X X X

ECO Economy mode X X X X X X

EXW Extended wash following serum/blood

samples

X

FIF Cal. voltages on display X X X X X

LEM Last 20 error messges printed X X X

MGL Display mg/dL X X X

NOB No beep X X X

NTD Prevents temperature from being reset

following each sample

X

QCC QC values direct ISE measurement X X X X X X

SFC SnapPak counter input X X

SLC Set "Leasing counter" X X

SSC Small sample cups X

TMP Allows temperature correction for

measured results

X

Table B-2 Instrument / service code matrix

Set Reset Service code description

Table B-1 Service code descriptions

Page 152: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

B-70 Service manual · Version 2.0

10 Service functions 91xx pH/Electrolyte AnalyzersService functions

Miscellaneous functions

Deleting data

Clearing sample count

a To reset the sample count to 0

1 At the [READY] display and press NO until the display [OPERATOR FUNCTIONS?] appears. Confirm with YES.

2 Press NO until [Reset Sample Number?] appears. Press YES.

3 When the prompt [Are you sure?] appears, press YES again.

Clearing all data

a To reset the analyzer software to the default values

1 Press the YES and NO keys simultaneously and switch the analyzer off and on again. This function can be performed optionally upon completion of decommissioning.

All QC values and statistics will be deleted and the sample number reset to 0. QC and normal values as well as correlation factors, printer settings, date/time and liquid content of the SnapPak will be reset to default values. All service codes will be deactivated. It is recommended that all data be deleted if the analyzer is to be used by other persons at a later date.

Warm-up bypass

a To bypass the instrument warm-up

1 Press YES and NO keys simultaneously during instrument warm-up.

CAUTION!

All QC data and instrument programming will be lost.

Page 153: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

April 2007

11 Adjustments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-3

12 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-7

Maintenance C

Page 154: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte
Page 155: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 C-3

91xx pH/Electrolyte Analyzers 11 AdjustmentsTable of contents

Adjustments

Insert introducing text here.

Adjustments .................................................................................................................C-5

Sample contacts adjustment (9110 only) .............................................................C-5

Sample sensor adjustment (9120/9130/9140/9180/9181 only) ...........................C-5

Temperature Adjustment (9110/9140/9180/9181 only) ......................................C-6

In this chapter Chapter 11

Page 156: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

C-4 Service manual · Version 2.0

11 Adjustments 91xx pH/Electrolyte AnalyzersTable of contents

Page 157: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 C-5

91xx pH/Electrolyte Analyzers 11 AdjustmentsAdjustments

Adjustments

The 9110, 20, 30, 40, 80 and 81 Analyzers have been designed to require minimal electronic adjustment. The sample contact circuit may require adjustment following a change of the sample contact components or during a routine troubleshooting (9110 only) or the sample sensor circuit may require adjustment following a change of the sample sensor or during routine troubleshooting. The temperature adjustment is active only for models 9110/9140/9180/9181 and requires the temperature adjustment test plug for correct calibration of the temperature circuit.

Sample contacts adjustment (9110 only)

In preparation for correct electronic adjustment, several steps must be performed.

a To adjust the sample contacts

1 Slide the electrode holder forward and remove the electrode from the analyzer.

2 Disconnect the intake tube from the preheater.

3 Remove the trim panel (located below the needle mechanism) and disconnect the plug for sample contact 1.

4 Select [TEST SAMPLE CONTACTS] in the [SERVICE FUNCTIONS] menu and adjust R10 (Display Board) for a reading of 140.

5 Reconnect sample contact 1 plug, intake tubing and electrode.

6 Check to ensure the value (140) remains the same.

e For details, see Error messages and troubleshooting on page D-5.

Sample sensor adjustment (9120/9130/9140/9180/9181 only)

a To adjust the sample sensor

1 Select [TEST SAMPLE SENSOR] in the [SERVICE FUNCTIONS] menu which will display the current sample sensor setting.

2 Adjust R93 Sample Sensor Adjust located at the top of the ISE SBC card to a value of 100.

3 Press YES to turn on peristaltic pump and ensure displayed value remains constant.

Raise sample door and aspirate clear fluid (e.g. water). Displayed reading should now be 150 or greater.

Note

If the value drops after reconnecting each of the three sample contacts, follow the troubleshooting

guidelines for fault isolation.

Page 158: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

C-6 Service manual · Version 2.0

11 Adjustments 91xx pH/Electrolyte AnalyzersAdjustments

Temperature Adjustment (9110/9140/9180/9181 only)

a To adjust the temperature amplifier

1 Select [AMPLIFIER TEST] in the [SERVICE FUNCTIONS] menu which will display the current amplifier voltages.

2 Press NO twice to display the temperature voltage and actual temperature display.

3 Unplug the preheater (9110 only) or the sample sensor and fit the Temperature Adjustment Test Plug in place of the preheater or sample sensor plug.

4 Adjust R54 located on the SBC Board for a display of 3000 mV.

5 Unplug the Temperature Adjustment Test Plug and fit the preheater or the sample sensor connector to display the actual temperature.

9110 Wait 5 minutes for the preheater to stabilize. The display should now read 37.0±0.2°C. Adjust R20 on the display board (clockwise to increase) for 37.0±0.1°C.

9140/9180/9181 Due to the elevated temperature inside the analyzer, it is normal for the displayed temperature to be several degrees C higher than the actual room temperature.

If no Temperature Adjustment Test Plug is available, place a quality thermometer (preferably mercury) next to the right side electrode holder. With the front door open, wait approx. 10 min. Then adjust the displayed temperature to match the reading on the thermometer.

Note

R93 can be accessed through the rear cover vent slot and is located at the top center on the circuit

card.

Page 159: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 C-7

91xx pH/Electrolyte Analyzers 12 MaintenanceTable of contents

Maintenance

Insert introducing text here.

Maintenance ................................................................................................................C-9

Daily maintenance .................................................................................................C-9

Weekly maintenance ..............................................................................................C-9

Monthly maintenance ...........................................................................................C-9

Six month maintenance ........................................................................................C-9

Annual preventive maintenance ...........................................................................C-9

In this chapter Chapter 12

Page 160: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

C-8 Service manual · Version 2.0

12 Maintenance 91xx pH/Electrolyte AnalyzersTable of contents

Page 161: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 C-9

91xx pH/Electrolyte Analyzers 12 MaintenanceMaintenance

Maintenance

Maintenance procedures for the 9110, 20, 30, 40, 80, 81 Analyzers require minimal time by the operator to perform. The procedures outlined below should be performed by the schedule indicated. Detailed instructions for these procedures can be found in the Instructions for Use describing the correct method to perform routine analyzer maintenance.

Daily maintenance

m Perform cleaning cycle.

m Perform conditioning cycle.

Weekly maintenance

m Clean sample probe and fill port.

m Clean analyzer surfaces.

Monthly maintenance

m Clean reference electrode housing.

m Check level of pH Filling Solution (9110 only).

Six month maintenance

m Replace peristaltic pump tubing.

Annual preventive maintenance

m Replace complete tubing set.

m Replace peristaltic pump tubing.

m Replace intake tubing (9110 only).

m Replace fill port assembly.

m Replace sample probe.

m Replace sample sensor Quad-ring (9120/30/40/80/81 only).

m Replace preheater Quad-ring (9110 only).

m Check sample sensor voltage (9120/30/40/80/81 only).

m Check sample contact voltage (9110 only).

m Check sample temperature (9140/80/81 only).

Page 162: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

C-10 Service manual · Version 2.0

12 Maintenance 91xx pH/Electrolyte AnalyzersMaintenance

Page 163: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

April 2007

13 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-3

Troubleshooting D

Page 164: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte
Page 165: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 D-3

91xx pH/Electrolyte Analyzers 13 TroubleshootingTable of contents

Troubleshooting

Error messages and troubleshooting ......................................................................... D-5

Status-LEDs .......................................................................................................... D-5

STATUS: NOT CALIBRATED ............................................................................. D-5

STANDARD A NOT DETECTED ....................................................................... D-5

STANDARD B NOT DETECTED, STANDARD C NOT DETECTED (9180/9181 only) ...................................................................................................................... D-6

CHECK SAMPLE CONTACTS(9110), CHECK SAMPLE SENSOR (9120/30/40/80/81) .................................................................................................................... D-6

CHECK REFERENCE HOUSING ...................................................................... D-7

PLEASE CLOSE SAMPLE DOOR (except 9181) ............................................... D-7

NO SAMPLE ......................................................................................................... D-8

CLEANING FLUID NOT DETECTED ............................................................... D-8

CONDITIONING FLUID NOT DETECTED (except 9110) ............................. D-8

INTERFACE ERROR ........................................................................................... D-9

PAPER JAM OR PRINTER DEFECT .................................................................. D-9

CHECK ELECTRODES ....................................................................................... D-9

## NOT CAL’D ..................................................................................................... D-9

CLOG IN SAMPLE PATH (9110 only), CLOG CHECK FLUID PATH (except 9110) ................................................................................................................... D-11

↑↑↑↑ ↓↓↓↓ (Out of range) ............................................................................... D-11

↑ ↓ (Temparature out of range) ........................................................................ D-11

ERR ..................................................................................................................... D-12

PERFORM DAILY MAINTENANCE (except 9110) ........................................ D-12

REPLACE FLUID PACK .................................................................................... D-12

CHECK TEMPERATURE (9140/9180/9181 only) ........................................... D-12

VALVE OVERTEMP! CHECK VALVES ............................................................ D-13

ERROR: UPPER NEEDLE SENSOR ................................................................. D-13

ERROR: LOWER NEEDLE SENSOR ................................................................ D-13

In this chapter Chapter 13

Page 166: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

D-4 Service manual · Version 2.0

13 Troubleshooting 91xx pH/Electrolyte AnalyzersTable of contents

WHEEL MISSING OR SAMPLER DEFECT .................................................... D-13

SAMPLER JAMMED OR DEFECTIVE ............................................................ D-14

Page 167: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 D-5

91xx pH/Electrolyte Analyzers 13 TroubleshootingError messages and troubleshooting

Error messages and troubleshooting

Status-LEDs

The 5 status LEDs at the bottom of the chassis back panel are lit simultaneously during normal operation. See below if any LED is blank.

Possible causes and remedies

o One ore more supply voltages are out of range or missing. Check the voltage test points.

e For details, see Power circuits on page A-33.

o If the voltages are correct:

O Check subsequent cicuits and cabling for obvious damage and replace, if necessary.

O Check and replace the corresponding fuse.

o Replace the SBC board.

o Replace the power supply.

STATUS: NOT CALIBRATED

This message is only displayed when a calibration has been interrupted by pressing NO or opening the sample door. Perform a calibration to return to the [READY] Mode.

STANDARD A NOT DETECTED

This message is displayed when sample contact 1 (9110 only) or the sample sensor is unable to properly detect the Standard A solution in a programmed time period. As the analyzer draws a sample of Standard A, the leading edge of the solution is detected by sample contact 1 (by bridging solution between both parts of sample contact 1) or the sample sensor. The sample sensor must also detect a continuous flow of Standard A Solution into the sample chamber, lasting several seconds, without encountering air bubbles or air. After several seconds, the trailing edge of the

Figure D-1 Status LEDs (back panel, bottom)

Page 168: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

D-6 Service manual · Version 2.0

13 Troubleshooting 91xx pH/Electrolyte AnalyzersError messages and troubleshooting

Standard A sample is sensed, the peristaltic pump stops and measurement occurs. If aspiration fails, the analyzer will wash and repeat up to 3 times.

Possible causes and remedies

o Check for air leaks preventing the Standard from being drawn into the chamber. Ensure that all O-rings are in place.

There must be a continuous draw of standard solution, free of air bubbles.

o Check solenoid V for proper sealing of the vent line.

o Check fill port to ensure the absence of air leaks.

o Verify that the sample needle is not bent.

o Check/replace fill port and wash port to ensure the absence of air leaks (9110 only).

o Check for clots or crystals formed in the Standard A tubing or electrode chamber.

o Replace peristaltic pump tube set to ensure proper sample aspiration.

o Ensure that the preheater cable (9110 only) or the sample sensor is plugged in securely and perform h [TEST SAMPLE CONTACTS] or h [TEST SAMPLE SENSOR] to ensure correct operation. Clean, adjust or replace sample sensor to correct sample sensor operation.

o Check the fluid remaining in the ISE SnapPakTM. If less than 5% remains, replace the ISE SnapPakTM.

STANDARD B NOT DETECTED, STANDARD C NOT DETECTED (9180/9181 only)

The system for drawing Standard B or C is like that of Standard A. If Standard A is drawn properly and Standard B or C is not detected, check Standard B tubing for crystallization. Perform checks for Standard A detection problems to correct. During the calibration sequence, Standard B or C is drawn into the measuring chamber prior to Standard A, therefore check if Standard A calibrant can be properly detected to isolate possible fluidic faults. If neither Standard A, B nor C are detected check for air leaks or blockages within the fluidic path.

CHECK SAMPLE CONTACTS(9110), CHECK SAMPLE SENSOR (9120/30/40/80/81)

The sample contacts must provide a reading of 135 - 145 when air is detected and the sample sensor must provide a reading of 80 - 120 when air is detected.

Perform h [TEST SAMPLE CONTACTS] or h [TEST SAMPLE SENSOR] to observe sample contact or sample sensor response. When clear fluid (e.g. water) is aspirated, the sample sensor display must indicate an increase of of at least 40 units. For blood samples (not transparent), the reading should decrease by at least 40 units. The sample contacts can be adjusted (see cross reference below), the sample sensor is calibrated with air during each calibration.

e For details, see Sample contacts adjustment (9110 only) on page C-5.

Perform a calibration to remove the error message after troubleshooting.

Page 169: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 D-7

91xx pH/Electrolyte Analyzers 13 TroubleshootingError messages and troubleshooting

Possible causes and remedies

o Check for correct drying of measuring chamber during the wash cycle.

o Replace peristaltic pump tube set.

9110 o Check for clots and leaks in the sample path, especially in the fill port area.

o Clean the sample contacts using Electrode Cleaning procedure.

o Test and adjust sample contacts to correct adjustment.

o Replace sample contacts as needed.

9120/30/40/80/81 o Clean sample sensor using Daily Cleaning procedure.

o Test and adjust sample sensor to correct adjustment.

o Replace sample sensor assembly.

CHECK REFERENCE HOUSING

At the beginning of a calibration, pressure is formed in the reference housing while the R valve is closed and the V valve is open. After Standard A is detected in the sample chamber, the back pressure formed in the reference housing causes the Standard A solution to move back toward the preheater (9110) or the sample sensor. When the standard A solution is detected by the preheater (9110) or the sample sensor, the R valve is opened to release the pressure. If the preheater (9110) or the sample sensor does not detect standard A as it moves back, [CHECK REFERENCE HOUSING] will be displayed.

Possible causes and remedies

o Check for clogged reference housing and clean per monthly maintenance procedures.

o Check and ensure that reference tubing is securely connected at the tubing connector.

o Check for proper filling of the reference housing and ensure that reference solution reagent level allows complete filling of the reference housing.

o Check for bubble-free aspiration of Standard A.

PLEASE CLOSE SAMPLE DOOR (except 9181)

This message occurs and prints [NO SAMPLE] on the print-out, when either of two conditions exists.

o Sample door is not closed within 20 seconds after samples are in place.

o Sample door has been opened and no sample is detected within 20 seconds.

Note

Sample contacts can be tested by invoking [TEST SAMPLE CONTACTS] within [SERVICE FUNCTIONS]. If reading is low, disconnect each sample contact and notice a change in the

displayed result to isolate a specific contact fault.

Page 170: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

D-8 Service manual · Version 2.0

13 Troubleshooting 91xx pH/Electrolyte AnalyzersError messages and troubleshooting

Possible causes and remedies

o Close sample door within the time allowed.

o Check needle mechanism light gate.

o Check whether preheater (9110) or sample sensor is plugged in.

NO SAMPLE

This message is displayed when any of the following conditions occur:

o During sampling, either a bubble in the sample is detected or the sample size is smaller than the minimum sample required.

o The sample door is opened and no sample is fed into the analyzer.

o Sample door is not closed within 20 seconds after sample is detected.

Possible causes and remedies

o Check whether the sample sensor is plugged in and perform h [TEST SAMPLE CONTACTS] or h [TEST SAMPLE SENSOR] to verify that it operates correctly.

o Check sample draw and look for clots or leaks in the sampling system.

CLEANING FLUID NOT DETECTED

When the cleaning cycle has been initiated, the analyzer checks for proper aspiration of the cleaning solution using the sample contacts (9110) or the sample sensor. If solution is not detected in the proper time, above error message will be displayed.

Possible causes and remedies

o Ensure that a sufficient volume of cleaning solution is presented for aspiration and no air bubbles are present as solution is aspirated.

o Check to ensure that the solution is being aspirated through the sampling mechanism correctly. The sample path must be free of leaks or clots.

o Verify that sample sensor is properly plugged in and perform h [TEST SAMPLE CONTACTS] or h [TEST SAMPLE SENSOR] to verify the correct operation of the sample sensor.

o Check pump tubing and replace as necessary to ensure a correct pump rate.

CONDITIONING FLUID NOT DETECTED (except 9110)

When the conditioning cycle has been initiated, the analyzer checks for proper aspiration of the conditioning solution using the sample sensor. If the solution is not detected in the proper time, this error message will be displayed.

Possible causes and remedies

o Ensure that a sufficient volume of conditioning solution is presented for aspiration and no air bubbles are present as solution is aspirated.

Page 171: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 D-9

91xx pH/Electrolyte Analyzers 13 TroubleshootingError messages and troubleshooting

o Check to ensure that the solution is being properly aspirated through the sampling mechanism. The sample path must be free of leaks or clots.

o Verify that sample sensor is correctly plugged in and perform h [TEST SAMPLE SENSOR] to verify its proper operation.

o Check pump tubing and replace as necessary to ensure correct pump rate.

INTERFACE ERROR

This message is displayed only during the h [INTERFACE TEST]. Pins 2 and 3 must be shorted together to perform the h [INTERFACE TEST]. Make sure that the pins are not shorted to chassis ground.

Possible cause and remedy

o Replace SBC Card.

PAPER JAM OR PRINTER DEFECT

This message is displayed when the printer head is jammed and tries to print. It is a temporary message that is displayed for 2 seconds and then sample results are displayed. To clear paper jam remove printer from the analyzer and remove jammed paper. Replace printer and retry. The printer can be removed by grasping the rear edge of the printer assembly and pulling it to the front. The printer can be removed with the Analyzer power on. To free up jammed paper, turn the small gears on the left side of the printer module.

CHECK ELECTRODES

This message is displayed when either of the following conditions are present.

1 More than six aspirations of Standard A are required for all electrode channels.

2 The difference between voltage A and B is out of specification for all channels.

Possible causes and remedies

o Check the level of the filling solution in the pH electrode (9110 only).

o Electrodes are not plugged in. Check the slide mechanism to ensure it is in the correct position and firmly pushed into electrode connectors.

o Perform Electrode Cleaning (9110 only) or Daily Maintenance.

o Check or replace reference electrode.

o Replace ISE SnapPakTM.

## NOT CAL’D

o NA NOT CAL’D (all except 9110)

o K NOT CAL’D

o CL NOT CAL’D

Page 172: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

D-10 Service manual · Version 2.0

13 Troubleshooting 91xx pH/Electrolyte AnalyzersError messages and troubleshooting

o CA NOT CAL’D

o LI NOT CAL’D

These messages will be displayed when a specific electrode does not calibrate properly during a calibration sequence. If more than six standard draws are required or the difference between Standard A and Standard B voltages is out of range, the [NOT CALIBRATED] message will be displayed for each channel which exhibits either of these conditions. A copy of the calibration report should be printed to provide the electrode output voltages measured during calibration.

The electrode voltage range table identifies correct voltage ranges for each electrode parameter.

If the difference between Standard A and Standard C (9180/9181 only) is out of range for Li+ or Na+, [LI NOT CAL’D] will be displayed.

Possible causes and remedies

o Check to ensure calibration reagents are transported correctly. Test for leaks, blockages, bubbles, or improper pump flow rate.

o Check or replace ISE SnapPak™.

o Check whether electrodes are plugged in properly. Check the slide mechanism to ensure it is in the correct position and firmly pushed into electrode connectors. Clean and/or replace electrode.

o If A-C is out of range for Na+, repeat the calibration or replace the ISE SnapPak™.

o Check or replace SBC Card.

ELECTRODE VOLTAGE RANGES 9110

Electrode Standard A Standard B Allowable Difference A-B

pH -2100 to -100 -1600 to +900 -430 to -650

Table D-1 9110 electrode voltage ranges

ELECTRODE VOLTAGE RANGES 9120, 9130, 9140

Electrode Standard A Standard B Allowable Difference A-B

9120/9130 9140

Na+ -600 to +2400 -1600 to +2200 +720 to +2000 +250 to +680

K+ -700 to +1000 -2500 to +500 +520 to +1800 +470 to +120

Cl– -3100 to -100 -1100 to +1800 -520 to -1200 not used

Ca2+ -2600 to +1400 -2100 to +2700 not used -480 to -900

Table D-2 9120, 9130, 9140 electrode voltage ranges

ELECTRODE VOLTAGE RANGES 9180, 9181

Electrode Standard A Standard B Standard C Allowable Difference

A-B A-C

Na+ -600 to +2400 -1600 to +2000 -600 to +2400 +250 to +680 -50 to +50

K+ -700 to +1000 -2500 to +500 -700 to +1000 +470 to +1200 -40 to +40

Cl– -3100 to -100 -1000 to +3000 -3100 to -100 -370 to -860 not used

Ca2+ -3100 to +1000 -2300 to +2500 -3100 to +1000 -350 to -660 -150 to +150

Li+ -3100 to +1900 -3600 to +1400 -2600 to +3400 +1 to +760 -1730 to -285

Table D-3 9180, 9181 electrode voltage ranges

Page 173: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 D-11

91xx pH/Electrolyte Analyzers 13 TroubleshootingError messages and troubleshooting

CLOG IN SAMPLE PATH (9110 only), CLOG CHECK FLUID PATH (except 9110)

After completion of sample measurement or calibration, the analyzer will monitor the sample sensor to determine if the sample path has been cleared. This message will be displayed if the sample path has not been cleared.

Possible causes and remedies

o Check for leaks in the fill port and electrode area.

o Check for blockages in the sample path, especially in the sample probe, the tubing to the preheater or the sample sensor and in the preheater or the sample sensor.

o Perform Electrode Cleaning (9110 only).

o Check sample contacts (9110 only) or sample sensor is securely plugged in and check Sample Contact Test (9110 only) or Sample Sensor Test to ensure correct operation.

o Check peristaltic pump and tubing are correct and all tubing is properly in place.

↑↑↑↑ ↓↓↓↓ (Out of range)

In case the analyzer displays arrows up or arrows down instead of sample results, the concentration of the sample is outside of the measurement range.

Possible causes and remedies

o If the sample is a urine sample, arrows up instead of the K+ result indicate that further dilution of the specimen is required (except 9110).

o Check for proper sample preparation.

o Ensure that the sample is correctly aspirated into the measuring chamber and ensure small air bubbles are not present.

o Check for proper aspiration of Standard A.

o Check the programmed conversion factors. The sample result has to be within the measurement range after conversion factors have been applied.

o Check the level of Filling Solution in the pH Electrode.

↑ ↓ (Temparature out of range)

On the 9140, 9180 and 9181 models, the calibration report will print an arrow up or down instead of the actual temperature, if the temperature measured is out of range (10.0°C - 40.0°C). The temperature sensor is located in the right side electrode holder.

Possible causes and remedies

o Ensure that sample sensor cable is securely plugged in.

o Check whether room temperature is within specified limits (15°C to 32°C / 60°F to 90°F).

o Perform temperature adjustment procedure.

o Replace SBC board or right side electrode holder.

Page 174: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

D-12 Service manual · Version 2.0

13 Troubleshooting 91xx pH/Electrolyte AnalyzersError messages and troubleshooting

ERR

The analyzer will display ERR in place of sample results when the analyzer is unable to obtain valid voltage readings from the electrode (A/D over- or underflow).

Possible causes and remedies:

o Ensure that the electrodes are securely in place and plugged into the analyzer.

o Check to ensure proper filling of the reference electrode has occurred.

o Ensure proper sampling and proper sample preparation. Check for air bubbles in the sample.

o Check level of filling solution in the pH electrode and refill if necessary.

PERFORM DAILY MAINTENANCE (except 9110)

This message will be printed at the end of a sample report when cleaning or conditioning have not been performed within the last 24 hours. This message is printed only. Perform Daily Maintenance to clear this message.

REPLACE FLUID PACK

When the monitored fluid level in the ISE SnapPakTM reaches 5% remaining, the analyzer will print REPLACE FLUID PACK at the end of each sample report. This message is printed only. Replace the ISE SnapPakTM following the instructions outlined in the Operator Manual.

To maximize the life of the ISE SnapPakTM, continue running the analyzer until a STANDARD A NOT DETECTED message is displayed. This will ensure that the ISE SnapPakTM is completely depleted before replacement.

CHECK TEMPERATURE (9140/9180/9181 only)

On the 9140/80/81 models, the analyzer will display CHECK TEMPERATURE at the end of each measurement, if the temperature measured is out of range (10.0°C to 40°C). The temperature sensor is located in the right side electrode holder.

On the 9110, this message will be displayed if the temperature of the preheater is out of range (36.0°C to 38.0°C).

Possible causes and remedies:

o Ensure that the preheater cable (9110 only) or sample sensor cable is securely plugged in.

o Check whether room temperature is within specified limits (15°C to 32°C / 60°F to 90°F).

o Check fuse F3 (LED D3).

o Perform temperature adjustment procedure.

o Replace SBC board or preheater (9110 only) or right side electrode holder.

Page 175: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 D-13

91xx pH/Electrolyte Analyzers 13 TroubleshootingError messages and troubleshooting

VALVE OVERTEMP! CHECK VALVES

In case one of the valve solenoids overheats, the message VALVE OVERTEMP! CHECK VALVES will be displayed. Each solenoid incorporates a thermo-fuse, which will reset automatically.

Possible causes and remedies

o Turn power off for at least one hour.

o Check fuse F3 (24V).

o Check whether room temperature is within specified limits (15°C to 32°C / 60°F to 90°F).

ERROR: UPPER NEEDLE SENSOR

If probe movement to the upper position is not detected within a certain time period, the above message is displayed.

Possible remedies

o Check for foreign material that may prevent the probe from moving freely.

o Contact Roche Diagnostics for technical assistance.

ERROR: LOWER NEEDLE SENSOR

If probe movement to the lower position is not detected within a certain time period, the above message is displayed.

Possible remedies

o Make sure the fill port is installed correctly.

o Check for foreign material that may prevent the probe from moving freely.

o Contact Roche Diagnostics for technical assistance.

WHEEL MISSING OR SAMPLER DEFECT

During the homing sequence, the sampler checks for the sample wheel to be present.

Possible remedies

o Make sure the wheel is present and seated correctly.

o Replace sample wheel.

o Replace sampler.

Page 176: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

D-14 Service manual · Version 2.0

13 Troubleshooting 91xx pH/Electrolyte AnalyzersError messages and troubleshooting

SAMPLER JAMMED OR DEFECTIVE

The analyzer checks for correct movement of the sample wheel. If the sample wheel does not move at the expected speed, the above error message is displayed.

Possible remedies

o Check for obstructions preventing the sample wheel from moving freely.

o Replace sampler.

Page 177: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

April 2007

14 Glossary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-3

Glossary E

Page 178: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte
Page 179: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 E-3

91xx pH/Electrolyte Analyzers

Glossary

Alkaline Basic.

Arterial blood Blood taken from aorta.

Autosampler Sample wheel for automatic measurements with 9181 Electrolyte Analyzers.

Clot Catcher Coagulate trap.

Dummy electrode The dummy electrode is simply a flow electrode and is only intended as a spacer and it has no additional measuring functions.

Electrodes The electrodes are flow-through electrodes with a visible sample channel.

Electrode simulator Test device for the Measuring chamber.

Heparin salts Are the only clot inhibitors that may be used for analyses in the 91xx pH/Electrolyte Analyzers.

Input unit The sample insertion as well as the aspiration of solutions is carried out via input unit which consists of the sample probe mechanism behind the sample door.

ISE abbr. for ion sensitive electrodes

ISE measuring chamber The ISE measuring chamber and its sensors are used to measure the hematocrit value and the electrolyte values Na+, K+, Ca2+, Cl– and Li+.

ISETROL QC material for checking ISE.

Locking lever This lever fixes the electrodes in the measuring chamber.

Measurement evaluation The validity of the test results from the instrument must be carefully examined by a clinical-medical specialist who will take the patient's clinical condition into consideration before any clinical decisions are reached based on the test results.

Measuring chamber Underneath the front door is the ISE measuring chamber with the electrodes.

MC abbr. for "measuring chamber"

NIST standards Are precise serums with accredited target values.

Peristaltic pump see Pump

Plasma Plasma samples are obtained by centrifuging heparinized whole blood, during which the cel-lular components of the blood are removed from plasma.

PP abbr. for peristaltic pump

Printer Low-noise thermoprinter.

A

C

D

E

H

I

L

M

N

P

Page 180: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

E-4 Service manual · Version 2.0

91xx pH/Electrolyte Analyzers

Pump One peristaltic pump transports the sample and the operating fluids inside the instrument.

QC abbr. for Quality control

QC material see ISETROL

Quality control A quality control program includes the analysis of sample materials with known ranges of expected values and the comparison of these values with analyzer results.

"Ready" mode Top level of the analyzer mode.

Reference electrode The reference electrode must always produce the same signal, regardless of the sample's composition. This is achieved by mixing the sample flow with a liquid of a higher KCl concentration (reference solution).

Rubber sealing sealing of SnapPak.

Sample contact Sample contacts support monitoring (conductivity) when filling the measuring chamber with fluid.

Sample sensor Sample sensors support monitoring (optically) when filling the measuring chamber with fluid.

Sample throughput Number of samples per hour.

SnapPak Calibration solutions for ISE including waste container.

Volume limitation Is the maximum volume of sample which is aspirated from the container.

Waste Container Waste water, part of the SnapPak.

Q

R

S

V

W

Page 181: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

April 2007

15 Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-3

Index F

Page 182: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte
Page 183: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 F-3

91xx pH/Electrolyte Analyzers

Index

Adjustment– Sample contacts (9110 only), C-5– Sample sensor (9120/9130/9140/9180/9181 only), C-5– Temperature (9110/9140/9180/9181 only), C-6Approvals, A-15

Block diagram– 9110 analyzer, A-39– 9120, 9130, 9140, 9180 analyzer, A-40– 9181 analyzer, A-41

Classifications– Approvals, A-15– Product data, A-15Cleaning solution (9110 only), A-28Clear– all data, B-69– Sample count, B-69Common operating parameter, A-14Contact addresses, 4Copyrights, 4

Deproteinizer, A-7Description of modules– Electronics

– Display board, A-35– ISE SBC card, A-33– Lamp board, A-34

– Fluidic Module, A-27– Cleaning solution (9110 only), A-28– Main tube set (all models)/ Intake tube set (9110

only), A-28– Reference solution, A-28– Sample Contacts (9110 only), A-29– Standard A, A-27– Standard B, A-27– Standard C (not on 9110), A-27– Waste liquids, A-28

– Mechanical assemblies– Display board, A-27

– Electrode holder– Electrode holder assembly, A-24

– Front door, A-23– Lamp board, A-27– Needle unit, A-23– Peristaltic pump, A-24– Printer, A-25– Rear panel, A-25– SBC board, A-26– Solenoid, A-24

Disinfectants– Deproteinizer, A-7– Other disinfectants, A-7Display board, A-27, A-35Document information, 3

Edition notice, 3Electronic diagrams, A-39– 9110 analyzer

– Display daughter PCB, A-53– Preheater PCB, A-54

– 9110, 9120, 9130, 9140 analyzer– Lamp PCB, A-52– Printer daughter PCB, A-51– SBC PCB

– Analog selector and A/D converter, A-48– Input amplifiers, A-46– Microprocessor, A-49– Power supply, A-43– Printer display drivers, A-50– Pump motor driver / door detect, A-45– Temperature circuit, A-47– Valve drivers, A-44

– 9120, 9130, 9140 analyzer– Display daughter PCB, A-55

– 9180 analyzer– Display daughter PCB, A-67

– 9180, 9181 analyzer– Lamp PCB, A-66– Printer daughter PCB, A-51– SBC PCB

– Analog selector and A/D converter, A-62– Fluid Pack detector, A-65– Input amplifier, A-60– Microprocessor, A-63, A-64– Power supply, A-57– Pump motor driver/door detect, A-59– Temperature circuit, A-61

A

B

C

D

E

Page 184: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

F-4 Service manual · Version 2.0

91xx pH/Electrolyte Analyzers

– Valve drivers, A-58– System interconnect, A-56

– Block diagram– 9110 analyzer, A-39– 9120, 9130, 9140, 9180 analyzer, A-40– 9181 analyzer, A-41

– System interconnect - 9110, 9120, 9130, 9140, A-42Electronics, A-33– Display board, A-35– ISE SBC card, A-33– Lamp board, A-34Error messages, D-5– ↑ ↓, D-11– ↑↑↑↑ ↓↓↓↓, D-11– CHECK ELECTRODES, D-9– CHECK REFERENCE HOUSING, D-7– CHECK SAMPLE CONTACTS(9110), D-6– CHECK TEMPERATURE (9140/9180/9181 only),

D-12– CLEANING FLUID NOT DETECTED, D-8– CLOG IN SAMPLE PATH (9110 only), D-11– CONDITIONING FLUID NOT DETECTED (except

9110), D-8– ERR, D-12– ERROR

– LOWER NEEDLE SENSOR, D-13– UPPER NEEDLE SENSOR, D-13

– INTERFACE ERROR, D-9– NO SAMPLE, D-8– NOT CAL’D, D-9– PAPER JAM OR PRINTER DEFECT, D-9– PERFORM DAILY MAINTENANCE (except 9110),

D-12– PLEASE CLOSE SAMPLE DOOR (except 9181), D-7– REPLACE FLUID PACK, D-12– SAMPLER JAMMED OR DEFECTIVE, D-14– STANDARD A NOT DETECTED, D-5– STANDARD B NOT DETECTED STANDARD C NOT

DETECTED (9180/9181 only), D-6– STATUS

– NOT CALIBRATED, D-5– VALVE OVERTEMP! CHECK VALVES, D-13– WHEEL MISSING OR SAMPLER DEFECT, D-13ESD protection, A-8– Conclusion, A-9– Explanation of the phenomenon, A-8– How can ESD protection be guaranteed?, A-9– Influence of electrostatic charges on components, A-8– Why is ESD protection so important today?, A-9

Fluidic Module, A-27– Cleaning solution (9110 only), A-28

– Main tube set (all models)/ Intake tube set (9110 only), A-28

– Reference solution, A-28– Sample Contacts (9110 only), A-29– Standard A, A-27– Standard B, A-27– Standard C (not on 9110), A-27– Waste liquids, A-28Front door assembly, A-23

How to Use This Manual, 9

Important information, A-5Important notes and warnings, A-6Intended use, 3Interface information, A-19Interface specifications– Data Link Information (9180/9181 only), A-20– Example data string information, A-20– Interface information, A-19ISE SBC card, A-33

Lamp board, A-27, A-34

Main tube set (all models)/ Intake tube set (9110 only), A-28Maintenance, C-9– Annual preventive, C-9– Daily, C-9– Monthly, C-9– Six month, C-9– Weekly, C-9Mechanical assemblies– Display board, A-27– Electrode holder, A-24– Front door, A-23– Lamp board, A-27– Needle unit, A-23– Peristaltic pump, A-24– Printer, A-25– Rear panel, A-25– SBC board, A-26– Solenoid, A-24

F

H

I

L

M

Page 185: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 F-5

91xx pH/Electrolyte Analyzers

Needle unit assembly, A-23

Operating Parameters– 9110, A-13Operating Parameters 9120/9130/9140/9180/9181, A-14Operating safety information, A-6Other disinfectants, A-7

Peristaltic pump assembly, A-24Printer assembly, A-25Product data, A-15

Rear panel assembly, A-25Reference solution, A-28

Safety information– Important information, A-5– Operating safety information, A-6Sample Contacts (9110 only), A-29Sample contacts adjustment (9110 only), C-5Sample sensor adjustment (9120/9130/9140/9180/9181 only), C-5SBC board, A-26Service functions, B-65– ENTER SERVICE CODE, B-67– TEST AMPLIFIER, B-67– TEST AUTOSAMPLER (9181 only), B-66– TEST ELECTRODES, B-65– TEST INTERFACE, B-67– TEST LANGUAGE SWITCH (9180 and 9181 only),

B-65– TEST PINCH VALVES, B-67– TEST PROBE (9181 only), B-66– TEST PUMP, B-66– TEST SAMPLE DOOR (except 9181), B-66– TEST SAMPLE SENSOR, B-65– TEST SnapPak SENSOR (9180 and 9181 only), B-65Service Manual– Abbreviations, 13– Conventions used, 10– Finding information, 9

– How to use, 9– IVD symbols, 11– Other symbols, 12– Symbols, 10– Version, 3, G-6Software– Version, 3, G-5Software operation 9110, B-5– Calibration sequence, B-15– Daily maintenance menu, B-11– Main menu, B-7– Measurement sequence, B-8– Operator functions menu, B-12– Overall program operation flow, B-6– Print functions menu, B-9– Program instrument menu, B-13– QC sample menu, B-10– Service functions menu, B-14Software operation 9120, 9130, 9140, B-19– Calibration sequence, B-29– Daily maintenance menu, B-25– Main menu, B-21– Measurement sequence, B-22– Operator functions menu, B-26– Overall program operating flow, B-20– Print functions menu, B-23– Program instrument menu, B-27– QC/standard/urine sample menu, B-24– Service functions menu, B-28Software operation 9180, B-33– Calibration sequence (Example Na/K/Li calibration),

B-43, B-44– Daily maintenance menu, B-39– Main menu, B-35– Measurement sequence, B-36– Operator functions menu, B-40– Overall program operation flow, B-34– Power-up, B-45– Print functions menu, B-37– Program instrument menu, B-41– QC/standard/dialysate/urine sample menu, B-38– Service functions menu, B-42Software operation 9181, B-49– Automated measurement sequence, B-53– Calibration sequence (example Na/K/Li calibration),

B-60, B-61– Daily maintenance menu, B-56– Main menu, B-51– Manual measurement sequence, B-52– Operator functions menu, B-57– Overall program operting flow, B-50– Power-up sequence, B-62– Print functions menu, B-54– Program instrument menu, B-58– QC/std/dialysate/urine sample menu, B-55

N

O

P

R

S

Page 186: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

F-6 Service manual · Version 2.0

91xx pH/Electrolyte Analyzers

– Service functions menu, B-59Solenoid assembly, A-24Specifications– 9110, A-13– 9120/9130/9140/9180/9181, A-13System interconnect - 9110, 9120, 9130, 9140, A-42

Temperature Adjustment (9110/9140/9180/9181 only), C-6Trademarks, 4Troubleshooting, D-5– Status-LEDs, D-5

Waste liquids, A-28

T

W

Page 187: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

April 2007

16 Versions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . G-3

Versions G

Page 188: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte
Page 189: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 G-3

91xx pH/Electrolyte Analyzers 16 VersionsTable of contents

Versions

This chapter provides information about the released software versions and the released service manuals.

Software .......................................................................................................................G-5

Service Manual ............................................................................................................G-6

Version 1.0 .............................................................................................................G-6

Version 2.0 .............................................................................................................G-6

In this chapter Chapter 16

Page 190: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

G-4 Service manual · Version 2.0

16 Versions 91xx pH/Electrolyte AnalyzersTable of contents

Page 191: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

Service manual · Version 2.0 G-5

91xx pH/Electrolyte Analyzers 16 VersionsSoftware

Software

For the 9110, 9120, 9130 and 9140 software, substitute X by one of the following language codes: English (E), French (F), German (D), Italian (I), Japanese (J) and Spanish (S).

The 9180 and 9181 software was published as multi-language version.

Version Nr. Release Date Remark

A10X1.18 – 9110 only

A20X3.18 – 9120 only

A30X2.18 – 9130 only

A30X2.24 – 9140 only

Table G-1 9110, 9120, 9130 and 9140 software releases

Version Nr. Release Date Remark

A80X1.00 – initial version for 9180 only

A80X2.00 – 9181 software included

A80X2.04 – minor bug fixes and text changes

A80X2.05 17 May 2004 9180/81 SN > 12001 only

"AVL" replaced by "Roche", minor text changes

Table G-2 9180, 9181 software releases

Page 192: mx1.frankshospitalworkshop.commx1.frankshospitalworkshop.com/.../Roche_91xx_-_Service_manual.pdf · Roche Diagnostics April 2007 Service Manual · Version 2.0 3 91xx pH/Electrolyte

Roche Diagnostics April 2007

G-6 Service manual · Version 2.0

16 Versions 91xx pH/Electrolyte AnalyzersService Manual

Service Manual

Version 1.0

Version 2.0

Version Nr. Date of Release Modified Chapters Applicable from Software Version

1.0 July 2004 First edition A80X2.05

Table G-3 Version 1.0 release information

Version Nr. Date of Release Modified Chapters Applicable from Software Version

2.0 April 2007 all n/a

Table G-4 Version 2.0 release information

Chapter–Page Modification / Addition

all FrameMaker Templates applied for usage with iSDoc

(Preface) 10 symbols: updated

A-5 important information: protective equipment: updated

A-15 nominal power consumption: modified

A-20 data link information (9180/81 only): new

B-69 deleting data: new

D-5 Status-LEDs: new

Glossary, Versions new

Table G-5 Version 2.0 modifications