Multivendor Algorithm Analysis Nuclear Medicine Applications

14
THE DICOM 2013 INTERNATIONAL CONFERENCE & SEMINAR March 14-16 Bangalore, India Multivendor Algorithm Analysis Nuclear Medicine Applications Author : Y.Kiran Kumar Co-Authors: A.K.Narayan, Kishan Harwalkar Philips Healthcare, Philips Innovation Campus Bangalore India

description

Multivendor Algorithm Analysis Nuclear Medicine Applications. Author : Y.Kiran Kumar Co-Authors: A.K.Narayan, Kishan Harwalkar Philips Healthcare, Philips Innovation Campus Bangalore India. Multivendor Algorithm Analysis for Nuclear Medicine Applications. Outline Introduction - PowerPoint PPT Presentation

Transcript of Multivendor Algorithm Analysis Nuclear Medicine Applications

Page 1: Multivendor Algorithm Analysis Nuclear Medicine Applications

THE DICOM 2013 INTERNATIONALCONFERENCE & SEMINARMarch 14-16 Bangalore, India

Multivendor Algorithm Analysis Nuclear Medicine Applications

Author : Y.Kiran KumarCo-Authors: A.K.Narayan, Kishan HarwalkarPhilips Healthcare,

Philips Innovation Campus

Bangalore

India

Page 2: Multivendor Algorithm Analysis Nuclear Medicine Applications

Multivendor Algorithm Analysis for Nuclear Medicine Applications

•Outline•Introduction•Clinical Workflow Background•Interoperability Issues & Clinical Relevance•Algorithm analysis •Conclusions

March 2013 DICOM International Conference & Seminar Multivendor Algorithm Analysis for Nuclear Medicine Applications 2

Page 3: Multivendor Algorithm Analysis Nuclear Medicine Applications

Outline

Customers are concerned to see variations in measurements of the same patient from various vendors that worries them about patient safety .

Indian Scenarios – Multivendor Workflow

March 2013 DICOM International Conference & Seminar Multivendor Algorithm Analysis for Nuclear Medicine Applications 3

Page 4: Multivendor Algorithm Analysis Nuclear Medicine Applications

Introduction

March 2013 DICOM International Conference & Seminar Title of Presentation (may abbreviate but identify the presentation.) 4

.

Page 5: Multivendor Algorithm Analysis Nuclear Medicine Applications

Vendor 1: Modality

Reporting Room - WorkStations

March 2013 DICOM International Conference & Seminar Multivendor Algorithm Analysis for Nuclear Medicine Applications 5

Vendor 2: Modality

Clinical Workflow - Nuclear Medicine

Page 6: Multivendor Algorithm Analysis Nuclear Medicine Applications

Clinical Workflow - Nuclear Medicine

•Technologist:

Scan Data Preparation

Physician:

Review the studyPerform post processing and interpretation.

March 2013 DICOM International Conference & Seminar Multivendor Algorithm Analysis for Nuclear Medicine Applications 6

Page 7: Multivendor Algorithm Analysis Nuclear Medicine Applications

Interoperability Issues

Issues found in Pragmatic level Interoperability :

• Analysis shows – significant variations of results

• Launching of Multivendor Datasets

March 2013 DICOM International Conference & Seminar Multivendor Algorithm Analysis for Nuclear Medicine Applications 7

Page 8: Multivendor Algorithm Analysis Nuclear Medicine Applications

Algorithm Analysis Variation -RENAL

RENAL Datasets

RENAL GFR – Vendor 1 (ml/min)

RENAL GFR – Vendor 2 (ml/min)

Difference of GFR (ml/min)

Patient -1114 47.3

66.7

Patient-2 85.42 40.5 44.92

Patient-3 90 45.8 44.2

Patient-4 144 96.4 47.6

Patient-5 70.69 35.4 35.29

March 2013 DICOM International Conference & Seminar Multivendor Algorithm Analysis for Nuclear Medicine Applications 8

Normal results range - 90 – 120 mL/min/1.73 m2

Page 9: Multivendor Algorithm Analysis Nuclear Medicine Applications

Algorithm Analysis & Clinical Relevance - MUGA

MUGA:

•MUGA (Multi-Gated Acquisition) allows you to automatically segment and quantify gated blood pool datasets, and create statistical Information about the cardiac cycle.

• A Left Ventricle ROI is required for MUGA. The widely used Edge detection algorithm in the MUGA scan is MUGA – GBP. The same algorithm is used between various vendors.

Clinical Significance of the statistical Results:

•In the MUGA Application the results like Ejection Fraction values are varied between the various vendors.

March 2013 DICOM International Conference & Seminar Title of Presentation (may abbreviate but identify the presentation.) 9

Page 10: Multivendor Algorithm Analysis Nuclear Medicine Applications

Algorithm Analysis & Clinical Relevance-Cardiac SPECT

• Result analysis for the reorientation algorithms outputs.

•The inputs for the algorithms are the projection data from Vendor-1,

and Vendor -2.

•The findings of the variation is the wrong assumption of the recon limits.

March 2013 DICOM International Conference & Seminar Title of Presentation (may abbreviate but identify the presentation.) 10

Page 11: Multivendor Algorithm Analysis Nuclear Medicine Applications

Algorithm Analysis Variation- Cardiac SPECT

March 2013 DICOM International Conference & Seminar Title of Presentation (may abbreviate but identify the presentation.) 11

Page 12: Multivendor Algorithm Analysis Nuclear Medicine Applications

Conclusions

•Variability in the Renal results, MUGA, Cardiac SPECT Results, has significant impact in the terms of the treatment, diagnosis, and therapy planning.

•When the variation is more than the expected normal range, the physician has to decide not only by the quantification results but also by the Visual Interpretation.

• For Diagnosis, Treatment and Therapy planning in NM - pragmatic interoperability level is required.

March 2013 DICOM International Conference & Seminar Multivendor Algorithm Analysis for Nuclear Medicine Applications 12

Page 13: Multivendor Algorithm Analysis Nuclear Medicine Applications

References

 www.ihe.net

 

ttp://www.accp8.org/web/data/Plenary%20Session/Session%203/Alan%20Lau%20PS.pdf

http://www.jacmp.org/index.php/jacmp/article/viewArticle/2977/1655

 

Functional assessment of the right ventricle with gated myocardial perfusion SPECT. Surjit Singh Wadhwa, Martin Caralon, Donna Abbati , Clinical Nuclear Medicine 2002.

 

Evaluation of radiocardiographic methods for determination of cardiac output and right and left LVEF using a cardiac phantom . Henning Kel February 1987, Nuclear medicine Communications. 

March 2013 DICOM International Conference & Seminar Multivendor Algorithm Analysis for Nuclear Medicine Applications 13

Page 14: Multivendor Algorithm Analysis Nuclear Medicine Applications

Contact info of Presenter

Y.Kiran Kumar

Philips Healthcare,

Philips Innovation Campus

Bangalore-560045

[email protected]

March 2013 DICOM International Conference & Seminar Multivendor Algorithm Analysis for Nuclear Medicine Applications 14