  • Multipurpose Readout Unit

    CRD-400Rev.0

  • 2Multipurpose Readout Unit CRD-400 - Rev.0

    INTERACTIVE INDEX

    Pag.

    Notes on the use of the product 3

    Disposal 4

    Symbols 4

    Identification labels 4

    CRD-400 overview 5

    Introduction 6

    Accessories and spare parts 8

    Travel bag 8

    Batteries 9

    Battery charger 9

    How to use the charger 10

    Charging cycle and led displays 11

    Taking care of the batteries 12

    Temperature effect 13

    User interface 14

    Turning on and off the CRD-400 16

    Startup pages 17

    Main display 18

    Reading the instruments 20

    Reading a bi-channel instrument 22

    Configuration 31

    Example of Sisgeo calibration report 34

    Troubleshooting 35

    Potentially explosive atmosphere 35

    Keep the CRD-400 within acceptable temperatures 35

    Care of the CRD-400 35

    Repair or modification of the CRD-400 35

    Service 36

    Appendix 37

    Technical specifications 40

  • 3BACK INDEXMultipurpose Readout Unit CRD-400 - Rev.0

    NOTES ON THE USE OF THE PRODUCT

    For a safe and efficient operation of the product, please read the following instructions carefully before use.

    Any other use beyond that described in this manual is considered be under the sole responsibility of the user.The same applies for any unauthorised modifications.In addition to the standards listed here, the user must comply with the provisions of current legislation on the safety and health of persons in the workplace.SISGEO is not liable for any failure, breakdown, accident, etc., due to lack of knowledge of (or non-compliance with) the requirements contained in this manual.Check that the product was not damaged during transport.Verify that the package includes all standard accessories as well as any requested optional accessories. If necessary contact the device retailer.The equipment should be used only after connecting it properly, according to the configuration needed, and the user must therefore perform carefully all the operations described in this manual.Use and any service or repair operation are permitted only for authorized operators. These operators must be physically and intellectually capable.

    To obtain information or order spare parts, always specify the data on the identification plate. When replacing parts, use only ORIGINAL SPARE PARTS.

    The manufacturer reserves the right to make changes without notice for whatever technical or commercial reason. We will attempt to ensure that the manuals are updated to reflect functional revisions of the products.

  • 4BACK INDEXMultipurpose Readout Unit CRD-400 - Rev.0

    DISPOSAL

    Under the European Directive 2002/96/EC, the appliance must be disposed of in a proper manner. The recyclable materials used in the device are recovered in order to prevent environmental pollution. For more information, please contact your dealer or local waste disposal service.

    The CRD-400 readout contains batteries and electronic components; materials must be disposed of separately from household waste.When this product reaches the end of its useful life, it must be brought to a collection point for the recycling of electronic components. Incorrect disposal by the user could be subject to fines.

    SYMBOLSBelow are the symbols used in the manual to draw the reader's attention:

    IDENTIFICATION LABELS

    Each CRD-400 readout, hereafter named "CRD-400" or "readout", has two types of identification labels.

    It is important to communicate these data when requesting information about your product or accessories.

    The first label, placed on top of theCRD-400, shows the registered Serial Number of the product.

    Warning! These operations must be performed by qualified personnel.

    Pay particular attention to the following information.

  • 5BACK INDEXMultipurpose Readout Unit CRD-400 - Rev.0

    The second label on the back of the CRD-400 provides information about the electrical characteristics of the product.

    CRD-400 OVERVIEW

    Respect the requirements on the label.Failure to observe these requirements could damage the product and endanger the operator.

    DisplayLabel

    Battery housing

    Function keys

    Key pad

    On/off key

    Connectors

    “Charger” connector

    “Sensor” connector

  • 6BACK INDEXMultipurpose Readout Unit CRD-400 - Rev.0

    INTRODUCTION

    The CRD-400 is a mono-channel hand-held readout capable of reading different types of SISGEO instruments. The CRD-400 can not perform measurements with removable inclinometer probes.

    For more information on the family of SISGEO instruments that can be read with theCRD-400, see Appendix.

    The CRD-400 and its accessories come in a package upon which are indicated the contents and main technical characteristics of the unit. This package can also be used for the storage of the CRD-400 when it is not in use.

  • 7BACK INDEXMultipurpose Readout Unit CRD-400 - Rev.0

    The CRD-400 is composed principally by:• Travel bag• Charger for NiMH batteries (0ECABCRD400)• 6-alligator clip cable + ground (0ECAV8P6A00)• CD-ROM (containing this manual)

  • 8BACK INDEXMultipurpose Readout Unit CRD-400 - Rev.0

    ACCESSORIES AND SPARE PARTS

    Flying cables:• Flying cable for CRD-400 - switch panel connection (0ECAV08V2S0)• Flying cable for CRD-400 - connectorized instruments connection (0ECAV08V2J0)

    CRD-400 Switch panel connection CRD-400 Connectorized instruments connection

    If the connectorized instrument is equipped with a bi-channel sensor, the CRD-400 readout, being mono-channel, will read only the first channel.

    TRAVEL BAG

    The travel bag can be used in two ways:• with the shoulder strap;• attached to the belt.

    Shoulder bag Bag - Belt

  • 9BACK INDEXMultipurpose Readout Unit CRD-400 - Rev.0

    BATTERIES

    Each CRD-400 is powered by internal rechargeable NiMH batteries (4 x AA, min. 2700 mAh). If properly used, NiMH batteries have a rated life of up to 1000 charging cycles.However an effect known as "memory effect" could diminish the battery life. This effect may occur if the batteries are left on charge for a few days or if they are repeatedly partially discharged. To prevent this problem, we recommend using the charger provided with the CRD-400 and performing a complete battery cycle (fully charge and discharge) at least once a month.

    To replace the battery, remove the lid of the battery compartment on the back of the readout by unscrewing the two screws.

    BATTERY CHARGER

    The charger supplied is controlled by a microprocessor that runs a quick charge in less than 3 hours. The range of the input voltage is universal (100-240 V~).The charger uses a method called "-dv detection"; this method relies on the fact that the voltage of the NiMH batteries decreases when they are fully charged.This voltage drop is detected when the voltage fall to a certain percentage of the maximum: in this case the charge will be automatically interrupted.When the reduction of voltage does not occur, the charger has a safety timer that will stop charging in any case, after a given period of time, to avoid battery overcharging.If the batteries have not been used for a long time, the decrease in voltage may occur in the first part of the charge cycle; for this reason there is a timer inside the charger to prevent "dv-detection" in the first few minutes of the charge cycle.

    The batteries should be recharged only when they are fully discharged and not before. Charge them before they are completely discharged only when it is necessary to use the CRD-400 immediately.

    A fully charged battery allows for continuous use of the CRD-400 for at least 6 hours: take this into account before going to the work site and if greater autonomy is needed, bring a spare battery pack.

    DO NOT perform measurements while the CRD-400 is being charged.Never use any charger other than the one provided.

  • 10BACK INDEXMultipurpose Readout Unit CRD-400 - Rev.0

    HOW TO USE THE CHARGER

    To recharge the batteries, the charger must be connected to the "Charger" connector on the readout, and only then should be connected to the wall outlet.

    Battery charger Charger connector

    The LED on the charger will be yellow before rapid charging and orange when charging is underway.

    When charging is completed the LED will be green. A full charge takes approximately 3-4 hours.

    The charger should not be connected to the readout for more than 12 hours.If the battery voltage is far below normal, the LED will indicate "error", flashing green and orange. In this case you must replace the batteries because they could be damaged.

    The charger may become warm during the charging process, so it is always necessary to leave adequate ventilation space around the charger.

    Do not use the charger if:• the charger has been in contact with liquid;• the charger is damaged;• the cable or the charger plug is damaged or frayed;

    After charging, disconnect the charger from the wall outlet only then disconnect the "Charger" connector from the CRD-400.

  • 11BACK INDEXMultipurpose Readout Unit CRD-400 - Rev.0

    CHARGING CYCLE AND LED DISPLAYS

    After connecting the charger to the wall outlet, the LED will be orange for the first 5-7 seconds, becoming yellow when initialization and analysis begin. If the battery is connected, the actual charging will begin a few seconds later, when the LED turns orange.When charging is complete, the LED will turn green after about 4 seconds: once the -dv has been detected, the start of the final stage of charging is indicated by a flashing green and yellow LED. The LED will be green during the trickle charge.

    LED MEANING

    Yellow Disconnected battery

    Yellow Initialisation and battery analysis

    Orange Quick charge

    Green with intermittent Charge completeflashing Yellow

    Green Trickle charger

    Orange – Green alternating Error

  • 12BACK INDEXMultipurpose Readout Unit CRD-400 - Rev.0

    TAKING CARE OF THE BATTERIES

    What to doFully charge the batteries before use, using only the supplied charger; it can provide enough power to safely charge the unit completely.When charging the batteries for the first time, or when the batteries have not been used for a long time, it is best to recharge at least three times consecutively to achieve maximum efficiency over time.If the batteries will not be used for a long time, it is best to remove them from the CRD-400.When storing batteries for long periods, it is best if they are charged at least 30-40% of their capacity and are stored in a cool and dry place.For greater efficiency, it is recommended to keep the contacts of both the batteries and the compartment where they are lodged clean and dry.When battery performance decreases substantially, it is recommended to replace the batteries with identical models or ones with equivalent features.

    What not to doUse a charger other than the one provided.Recharge batteries of different capacities.Recharge batteries of a different or non-homogeneous state of residual discharge.Overcharge the batteries: this may determine a rapid decrease in their efficiency, with consequent reduction of their useful life.Short-circuit the batteries.Drop or expose the batteries to unnecessary shocks.Try to open or disassemble the batteries.

  • 13BACK INDEXMultipurpose Readout Unit CRD-400 - Rev.0

    TEMPERATURE EFFECT

    NiMH batteries are extremely sensitive to temperature. Below is a chart which shows an example of the variation of battery discharge in relation to temperature:

    Keep this effect in mind before going to the work site: if greater autonomy is needed, bring along an extra battery pack.

    80

    100

    120

    Variation of Discharge Capacity with Temperature

    Avai

    labl

    e(%

    Of R

    ated

    Cap

    acity

    )

    0

    20

    40

    60

    -20 -10 0 10 20 30 40 50

    Discharge Temperature (Degrees c)

    Capa

    city

    Avai

    labl

    e(%

    Of R

    ated

    Cap

    acity

    )

  • 14BACK INDEXMultipurpose Readout Unit CRD-400 - Rev.0

    USER INTERFACE

    Keyboard and keysThe CRD-400 has 9 keys.

    “SWITCH” Key

    It has two modes:• Change of measurement units mode:In the measurement phase, it allows you to select the units of measurement (for example, mV, Ω, °C) of the active phase; for example, in the measurement "V.W and Thermistor", pressing this key repeatedly you will visualize periodically the reading in Digit, °C, microseconds, Hz, μ , etc.• Cancel Mode:It is used in the sub-menu or in the various configuration pages to cancel the phase and return to the previous window.

    “RUN” Key

    It is active only during measurement and is used to activate or deactivate the measurement.

    “kN” Key

    It is used in the "Ratiometric" measurement phase to choose one of the various available full scales.

  • 15BACK INDEXMultipurpose Readout Unit CRD-400 - Rev.0

    “ARROWS” Key

    They are used to navigate through the various menu options, in the configuration pages or during data entry.

    “ENTER/ESC” Key

    Lets you:• Select a menu item, an option, or when typing data, a letter or number.• Exit a measurement.

    “LIGHT+” Key

    This key is used to increase the LCD backlight.

    “LIGHT-” Key

    This key is used to decrease the LCD backlight.

    “ON/OFF” Key

    Turns the unit on and off.

  • 16BACK INDEXMultipurpose Readout Unit CRD-400 - Rev.0

    TURNING ON AND OFF THE CRD-400

    Turning onTo turn on the CRD-400, press the ON/OFF key, and hold it down until the LCD turns on.

    Turning offTo turn off the CRD-400, press the ON/OFF key, and hold it down until the LCD turns off. After the seconds visualized on the LCD display have counted down, the CRD-400 will turn off automatically.To cancel shutdown, press any key during the countdown.

  • 17BACK INDEXMultipurpose Readout Unit CRD-400 - Rev.0

    STARTUP PAGES

    Hold the readout upright and turn it on.At startup, the readout shows two pages: • Logo and readout identification;• Mode selection page.

    Logo and readout identification• the SISGEO logo;• the readout nameplate: “CRD-400”.

    Mode selectionIn this second page, the system provides information on:• System state. The first line shows any errors in calibration or internal components;• Readout calibration date;• Firmware version;• Serial number of the unit;• Lets you select “right-handed” or “left-handed” use of the readout; • Position the readout vertically and use the arrow keys to select the desired mode.

    From now on you can hold the readout in a horizontal position, depending on which mode (right-handed or left-handed) is selected.

  • 18BACK INDEXMultipurpose Readout Unit CRD-400 - Rev.0

    MAIN DISPLAY

    Title bar

    Main menu

    Status bar

    Title barIn all screens, except in those where a virtual keyboard is shown for data entry, the "Title Bar" is shown at the top of the display, and contains information about the currently selected phase.

    The main menu and menu treeBelow are the main menu options and sub-menu, displayed in the central part of the display:

    • Current Loop• 3 Wire Current• Voltage• Wheatstone Bridge• Servo MonoAx• RTD Pt100• NTC Thermistor• V.W. and Thermistor• Setup• System Parameters• Polynomials

    Use the arrow key to select the desired item, then press the ENTER/ESC key to activate the stage or change the menu. Press the SWITCH key to go back the previous menu.The cursor highlighting the option becomes: • Red when you reach the first and last item in a menu;• Blue when you are scrolling within the menu items and the highlighted option is neither the first nor the last of the menu.

  • 19BACK INDEXMultipurpose Readout Unit CRD-400 - Rev.0

    Status bar

    • Battery status• Flag of active scanning• Measurement unit of the active reading if a size has been selected, or date and system time• Temperature of the readout in Celsius degrees.

    Battery statusThis indicator will show the following colours based on the percentage of battery power or if the battery is being charged.

    • Green: over 70% battery charge• Yellow: percentage between 50% and 70% charge• Red: less than 50% charge. When the battery reaches this threshold for the first time, the buzzer is activated for about 5 seconds.• Blue: if the battery is charging.

    Flag of active scanningThis flag indicates that the CRD-400 is taking a reading (alternating red and blue flashes).

    Battery status

    Active read flag Measure unit/date and hour Internal temperature

  • 20BACK INDEXMultipurpose Readout Unit CRD-400 - Rev.0

    READING THE INSTRUMENTS

    • Connect the flying cable supplied to the "Sensor" connector on the readout.

    Never force a connector into a port. Verify that there are no possible obstructions in the port. If the connector and port are not easily connected, they are probably not compatible. Make sure the connector is compatible with the port and you have positioned the connector correctly in relation to the port.

    • Connect the wires of the instrument to be measured to the alligator clips of the flying cable.

    To avoid damage to the CRD-400 or the instrument, make sure you do not ever cause a short circuit with the alligator clips of the flying cable.Make sure the cable is properly connected and that the type of instrument is consistent with the type of measurement which has been selected.

    • To read the instruments, just select the desired type of measurement in the main menu. The CRD-400 shows the wiring diagram for SISGEO mono-channel instruments:

    The wiring diagram shows the functions (+ Vcc, + Out, etc.) of each connector pin: refer to the calibration report of the instrument to be read for the proper correspondence of functions.The colour scheme is purely indicative and in any case, refers to the colour of the mono-channel instrument cables or to channel A in the case of two-channel instruments.

  • 21BACK INDEXMultipurpose Readout Unit CRD-400 - Rev.0

    • Press the RUN key to begin measurement of the instrument connected to the readout.

    Read value

    It is always advisable to wait at least 10 seconds before noting the displayed value to allow the stabilisation of the electronics of the instrument measured.If the input signal exceeds the full scale, the readout will interrupt the measurement to avoid damage to the readout and/or the instrument.

    Press the RUN key to interrupt the reading of the instrument.Press the ENTER/ESC key to interrupt the reading and return to the previous screen.

    During the reading, the measurement unit shown on the display can be changed by pressing the SWITCH key repeatedly; the engineering unit must be configured previously in the “Polynomials” menu.

    When a Ratiometric reading is performed, pressing the kN key allows you to choose between the following full scales of the standard SISGEO L2XXVXXX cells:• 2.0 mV/V• 300 kN• 500 kN• 750 kN• 1200 kN• 1500 kN• 1800 kN• 2500 kN• 3000 kN• 5000 kN• 10000 kN

    Conversion to kN is done by calculating a sensitivity of 1.5mV / V F.S. (for F.S. from 300kN to 1500kN) and a sensitivity of 2mV / V F.S. (for F.S. from 1800kN 10000kN); the zero reading of the instrument is not considered. The measurement in kN is therefore considered merely indicative.

  • 22BACK INDEXMultipurpose Readout Unit CRD-400 - Rev.0

    READING A BI-CHANNEL INSTRUMENT

    To read a Bi-channel instrument, proceed as follows:• Consult the Calibration Report of the instrument issued by SISGEO to verify the wirings.

    • Connect channel A (in the example of the certificate, RED wire= + loop, BLACK wire= -loop), respectively to alligator clips A and C of the CRD-400 as described in the Current Loop wiring diagram of the CRD-400.

    • Then, after reading the A channel, repeat the procedure by connecting the B channel (in the example of the certificate WHITE wire= + loop, GREEN wires= -loop), respectively

    to alligator clips A and C of the CRD-400 as described in the Current Loop wiring diagram of the CRD-400.

  • 23BACK INDEXMultipurpose Readout Unit CRD-400 - Rev.0

    Current loop measurment

    Power: 24V Maximum current drawn by the instrument: 25mA

    Selecting “Current Loop”, the wiring diagram will be shown. Press the ENTER/ ESC key to enter the measurement phase.

    On this page are shown measurements in a range from 0,000 to 21,000 mA.

    The following keys are active: • The RUN key activates or deactivates the measurement.• The ENTER/ESC key cancels the measurement.• The SWITCH key changes the unit of measurement if the corresponding polynomial has been previously enabled: - mA - user-defined units.

    A

    A

    B

    C

    DEF

    G

    A+ Supply

    Return Signal

  • 24BACK INDEXMultipurpose Readout Unit CRD-400 - Rev.0

    3 wire current measurement

    Power: 24V Maximum current drawn by the instrument: 100mA

    Selecting “3 Wire Current”, the wiring diagram will be shown. Press the ENTER/ ESC key to enter the measurement phase.

    On this page are shown measurements in a range from 0,000 to 21,000 mA.

    The following keys are active: • The RUN key activates or deactivates the measurement.• The ENTER/ESC key cancels the measurement• The SWITCH key changes the unit of measurement if the corresponding polynomial has been previously enabled: - mA - user-defined units.

    A

    AB

    C

    DEF

    G

    A+ Supply

    - Signal

    Signal

  • 25BACK INDEXMultipurpose Readout Unit CRD-400 - Rev.0

    Voltage measurement

    Power: 24V Maximum current drawn by the instrument: 100mA

    Selecting “Voltage”, the wiring diagram will be shown. Press the ENTER/ ESC key to enter the measurement phase.

    On this page are shown measurements in a range from -10000 mV to +10000 mV

    The measurement will be always shown in mV, only the number of decimal places shown on the screen will change: • 3 decimal places for values ranging from ± 10.000 mV to ± 99.999 mV• 2 decimal places for values ranging from ± 10000 mV to ± 99999 mV• 1 decimal place for values ranging from ± 1000.0 mV to ± 9999.9 mV

    The following keys are active: • The RUN key activates or deactivate the measurement.• The ENTER/ESC key cancels the measurement.• The SWITCH key changes the unit of measurement if the corresponding polynomial has been previously enabled: - mV or V - user-defined units.

    A

    A1B

    C

    DEF

    G

    +CH

    -CH

  • 26BACK INDEXMultipurpose Readout Unit CRD-400 - Rev.0

    Wheatstone bridge measurement

    Power: 5V

    Maximum current drawn by the instrument: 100mA

    Selecting “Wheatstone Bridge”, the wiring diagram will be shown. Press the ENTER/ ESC key to enter the measurement phase.

    On this page are shown measurements in a range from -10 mV/V to +10 mV/V

    The following keys are active: • The RUN key activates or deactivates the measurement.• The ENTER/ESC key cancels the measurement.• The SWITCH key changes the unit of measurement if the corresponding polynomial has been previously enabled: - mV/V. - user-defined units.

    • The kN key, when measurement is not active, lets you select the full scale of the measurement. The selection is made using the menu "Select Full Scale" and navigating with the arrow keys. Press the ENTER/ESC key to select the new full scale. Press the SWITCH key to return to the measurement phase without changing the full scale.

    For more details refer to the paragraph "Reading the instruments".

    A

    B

    C

    DEF

    G

    +SENSING

    +SIGNAL

    -SENSING-SIGNAL

  • 27BACK INDEXMultipurpose Readout Unit CRD-400 - Rev.0

    Servoinclinometer monoaxial measurement

    Power: ±12V

    Maximum current drawn by the instrument: 100mA

    Selecting “Servo MonoAx”, the wiring diagram will be shown. Press the ENTER/ ESC key to enter the measurement phase.

    On this page are shown measurements in a range from -10000 mV to +10000 mV

    The measurement will be always shown in mV, only the number of decimal places shown on the screen will change: • 3 decimal places for values ranging from ± 10.000 mV to ± 99.999 mV• 2 decimal places for values ranging from ± 10000 mV to ± 99999 mV• 1 decimal place for values ranging from ±1000.0 mV to ±9999.9 mV

    The following keys are active: • The RUN key activates or deactivate the measurement.• The ENTER/ESC key cancels the phase.• The SWITCH key changes the unit of measurement if the corresponding polynomial has been previously enabled: - mV - user-defined units.

    A

    B

    C

    DEF

    G

    GND PWR

    +SGN

    GND SGN

    V1

    V1+ Supply

    - Signal

    +Signal -Signal

    GNQ

  • 28BACK INDEXMultipurpose Readout Unit CRD-400 - Rev.0

    RTD PT100 measurement

    Maximum current drawn by the instrument: 1mA

    Selecting “RTD PT100”, the wiring diagram will be shown. Press the ENTER/ ESC key to enter the measurement phase.

    On this page are shown measurements in a range from -150 °C to +150 °C.

    The following keys are active: • The RUN key activates or deactivates the measurement.• The ENTER/ESC key cancels the measurement.• The SWITCH key selects the visualization of the measurement in: - °C - °F - Ω - mV.

    A

    B

    C

    DEF

    G

    +I (1mA)

    +SIGNAL

    -SIGNAL

    -I (1mA)

    100Ω at 0°C0.385Ω per °C

  • 29BACK INDEXMultipurpose Readout Unit CRD-400 - Rev.0

    NTC thermistor measurement

    Maximum current drawn by the instrument: 0,1 mA

    Selecting “NTC Thermistor”, the wiring diagram will be shown. Press the ENTER/ ESC key to enter the measurement phase.

    On this page are shown measurements in a range from -30 °C to +150 °C.

    The following keys are active: • The RUN key activates or deactivates the measurement.• The ENTER/ESC key cancels the measurement.• The SWITCH key selects the visualization of the measurement in: -°C -°F -Ω -mV.

    A

    B

    C

    DEF

    G

    +I (0.1mA) and +signal

    -I (0.1mA) and -signal

    3000Ω at 25°C

  • 30BACK INDEXMultipurpose Readout Unit CRD-400 - Rev.0

    Digit = Hz2

    1000

    Vibrating Wire and Thermistor measurement

    Selecting “V.W. and Thermistor”, the wiring diagram will be shown. Press the ENTER/ ESC key to enter the measurement phase.

    The following keys are active: • the RUN key activates or deactivates the measurement.• the ENTER/ESC key cancels the measurement• the SWITCH key selects the visualization of the measurement in:• Digit

    - °C - μs (micro seconds) the period of oscillation - Hz (Hertz) the measurement of frequency oscillation - μ VK4000 (micro Epsilon) calculated as: - μ VK4100 - μ VK4200

    - user-defined units.

    The measurement of vibrating wire instruments is the measurement of resonant frequency and is performed by transforming the signal, acquired by an A/D converter, into a spectral signal using a Fast Fourier Transform (FFT) algorithm.The quality parameter shown on the display compares the amplitude of the resonant frequency with the background noise: values ≥ 7 ensure an acceptable quality of the measurement.

    µ = gauge · Digit = gauge · Hz2

    1000

    A

    B

    C

    DEF

    G

    VW excitation and return

    VW excitation and return

    +I (0.1mA) and +signal

    -I (0.1mA) and -signal

    3000Ω at 25°C

  • 31BACK INDEXMultipurpose Readout Unit CRD-400 - Rev.0

    CONFIGURATION

    Selecting this item in the main menu, you enter the configuration menu of the readout. The options in this menu are:

    • System Parameters, to set the general parameters of the readout.

    • Polynomials, to activate or set the parameters of polynomial equations of I, II and III degrees.

    General ParametersSelecting this phase you can modify the following parameters:• Languages: language of the readout's messages. Selecting the item and pressing

    the ENTER/ESC key will bring up the language selection screen. Select the desired language and press the ENTER/ESC key to change language again.

    • Keypads Tones: if “ON”, pressing a key will trigger acoustic confirmation.• Backlight Time. [s]: this is the time limit after which, when a key is pressed, the readout

    will decrease the backlighting of the LCD to save battery power.

    Set the seconds to zero to disable the function.

    • Backlight [%]: This is the percentage of backlighting currently selected, and which will be set the next time the unit is turned on. It can vary from 49 to 100%.

    Note that battery consumption is proportional to backlighting

    • Shut Down [s]: This is the time limit after which, when a key is pressed, the readout turns itself off, informing the operator with an acoustic signal that shutdown is starting. It can be set from 30 to 3600 seconds.

    Set the seconds to zero to disable the function. In this case the readout does not ever shut itself down.

    • Date: Allows you to set the date of the readout, which is visible in the status bar. Pressing the ENTER/ESC key, the virtual keyboard appears. Set the date in dd / mm / yyyy format.

    • Time: Allows you to set the time of the readout, which is visible in the status bar. Pressing the ENTER/ESC key, the virtual keyboard appears. Set the time in hh: mm: ss format.

  • 32BACK INDEXMultipurpose Readout Unit CRD-400 - Rev.0

    NOTE: when the batteries are removed, the date and time will be lost. If this operation is performed in less than 10 seconds, the time and date will not be lost.

    PolynomialsThe readout has the ability to enable a polynomial for each measurement, from first to third degree, to set how the measurements will be transformed.The measurements:

    • RTD PT100• NTC Thermistors

    have these polynomials enabled by default to allow the display of temperature in °C.

    Select an item in the Polynomials menu to enter the parameter settings.Press the ENTER/ESC key when Polynomial is selected to choose these options:

    • disable• 1st degree• 2st degree• 3rd degree

  • 33BACK INDEXMultipurpose Readout Unit CRD-400 - Rev.0

    Enabling the first degree polynomial (Ax + B) it is possible to set:

    • Measuring Unit: Allows you to enter a description of the unit of measure, which will appear in the status bar.

    • Parameters A and B of the polynomial.

    Enabling the second degree polynomial (Ax2 + Bx + C) it is possible to set:

    • Measuring Unit: Allows you to enter a description of the unit of measure, which will appear in the status bar.

    • Parameters A, B and C of the polynomial.

    Enabling the third degree polynomial (Ax3 + Bx2 + Cx + D) it is possible to set:

    • Measuring Unit: Allows you to enter a description of the unit of measure, which will appear in the status bar.

    • Parameters A, B, C and D of the polynomial.

    The polynomials are taken from the instrument calibration report.

  • 34BACK INDEXMultipurpose Readout Unit CRD-400 - Rev.0

    EXAMPLE OF SISGEO CALIBRATION REPORT

  • 35BACK INDEXMultipurpose Readout Unit CRD-400 - Rev.0

    TROUBLESHOOTING

    Readout freezes upIn the event that the CRD-400 no longer responds to the commands of the operator, remove and replace the batteries to reset the readout.

    POTENTIALLY EXPLOSIVE ATMOSPHERE

    In hazardous areas the CRD-400 must be turned off. Do not charge the CRD-400.Sparks could cause an explosion or fire, causing serious injury and even death.Areas with a potentially explosive atmosphere are usually indicated.

    KEEP THE CRD-400 WITHIN ACCEPTABLE TEMPERATURES

    The CRD-400 is designed to operate at temperatures between -20 ° C and +60 ° C and to be stored at temperatures between -30 ° C and +70 ° C. Extreme temperatures may temporarily shorten battery life or cause temporary malfunction of the CRD-400.

    CARE OF THE CRD-400

    The CRD-400 contains delicate parts and components.Do not drop, disassemble, open, crush or deform the readout. Do not use the CRD-400 if it is damaged. Failure to observe this warning could damage installed instruments.

    REPAIR OR MODIFICATION OF THE CRD-400

    SISGEO does not allow the end user to repair or modify the CRD-400. Failure to comply with this provision could cause damage not covered by warranty.In case of malfunction, contact SISGEO customer service.

  • 36BACK INDEXMultipurpose Readout Unit CRD-400 - Rev.0

    SERVICE

    SISGEO is aware of the importance of after sale customer service. Therefore SISGEO Customer Service is structured to offer the following services:

    Repairs and maintenanceDuring the period covered by the guarantee we ensure that the equipment supplied is free from defects in materials and appropriate for standard use.

    Any repairs and/or recalibration should always be carried out at SISGEO or by persons technically trained and authorized by SISGEO. If repair and/or recalibration should be necessary, send the CRD-400 and all its accessories in the provided box, in order to prevent possible damages during transport.

    In order to provide appropriate and efficient Customer Service, SISGEO ensures a "Return Manufacturer Authorization" (RMA) for the shipping of material needing repair/maintenance. The procedure involves filling out the form on the website www.sisgeo.com, clicking on the link service - repairs; once the request has been received, a confirmation will be sent along with order code.

    The material must be sent with appropriate packaging, freight prepaid (shipping charges prepaid) to:

    SISGEO S.r.l.Via F. Serpero, 4/F1 - 20060 MASATE (MI)To the attention of: Customer ServiceTel. +39-02-95764130 - Fax: +39-02-95762011 - e-mail: [email protected]

    Phone supportTelephone support regarding the installation and configuration of instruments, measurement systems and software.

    Assistance on the installation of the equipment suppliedTechnical assistance on the installation of monitoring systems with its commissioning and start up, with the help of teams of experts.

    TrainingTraining courses are offered at our facility or on-site regarding installation procedures, system configurations and the use of measurement systems.

  • 37BACK INDEXMultipurpose Readout Unit CRD-400 - Rev.0

    APPENDIX

    Instrument Code Instrument Name Measure Type Supply Signal Unit A B C D E F G

    420MAGEN2W Generic 4-20mA 2 wires

    Current Loop 24V 20mA mA shield

    D232TXXX Embankment extensiometer

    Current Loop 24V 20mA mA red black shield

    D241A200 Wire Crackmeter-Deformometer

    Current Loop 24V 20mA mA red black shield

    D313FAXX Electrical Crackmeter-Jointmeters

    Current Loop 24V 20mA mA red black shield

    D313MAXX D313MAXXX

    Electrical Deformeter Current Loop 24V 20mA mA red black shield

    D313SAXX Electrical Crackmeter Current Loop 24V 20mA mA red black shield

    D314CAXX Embedment joinmeter Current Loop 24V 20mA mA red black shield

    D422XXXX DSM gauge Current Loop 24V 20mA mA red black shield

    DEX35SXXX 3D borehole sensor (inclinometer, CH A)

    Current Loop 24V 20mA mA red black shield

    DEX35SXXX 3D borehole sensor (inclinometer, CH B)

    Current Loop 24V 20mA mA white green shield

    DTE1AXXX Displacement transdu-cer

    Current Loop 24V 20mA mA red black shield

    DTM0AEXX DTM displacement transducer 1000mm

    Current Loop 24V 20mA mA red black shield

    L141XXXX Earth pressure cell 100x200mm

    Current Loop 24V 20mA mA red black shield

    L2CE0XXX Stell lining pressure cell

    Current Loop 24V 20mA mA red black shield

    L3400AS0 Surface strain-gauge 400mm

    Current Loop 24V 20mA mA red black shield

    P235I0XX Drive-in piezometer Current Loop 24V 20mA mA red black shield

    P235SXXX Electrical piezometer Current Loop 24V 20mA mA red black shield

    P251QXXX V-notch level gauge 1.0m

    Current Loop 24V 20mA mA red black shield

    P252AXXX Pressure transducer 100KPa

    Current Loop 24V 20mA mA red black shield

    P252CXXX Removable piezometer Current Loop 24V 20mA mA red black shield

    P252RXXX Vented pressure transducer

    Current Loop 24V 20mA mA red black shield

    QVLIV100 V-notch float/level transducer

    Current Loop 24V 20mA mA red black shield

    S411HAXX Uniaxial in-place inclinometer

    Current Loop 24V 20mA mA red black shield

    S411P1XX Vertical in-place incli-nometer 1.0m mono

    Current Loop 24V 20mA mA red black shield

    S421HAXX Uniaxial (SUB) horizon-tal IPI

    Current Loop 24V 20mA mA red black shield

    S521HAXX Waterproof clinometer Current Loop 24V 20mA mA red black shield

  • 38BACK INDEXMultipurpose Readout Unit CRD-400 - Rev.0

    Instrument Code Instrument Name Measure Type Supply Signal Unit A B C D E F G

    S531PAXX Clinometer 4-20mA mono

    Current Loop 24V 20mA mA red black shield

    S700TLXX Tilt-beam sensor +/-10°

    Current Loop 24V 20mA mA red black shield

    S7RDSXXX RDS Railway Deformation System

    Current Loop 24V 20mA mA red black shield

    Generic 4-20mA 3 wires

    3 Wire Current 24V 20mA mA red black white shield

    D313FVXX Electrical Deformeter Voltage 24V 2000mV mV red black white green shield

    D313MVXXX Electrical Deformeter Voltage 24V 2000mV mV red black white green shield

    D313SVXX Electrical Crackmeter Voltage 24V 2000mV mV red black white green shield

    DEX35SXXX 3D borehole sensor (extensometer)

    Voltage 24V 10000mV mV brown grey yellow pink shield

    DEX35XXX 3D borehole sensor (extensometer)

    Voltage 24V 10000mV mV red black white green shield

    DTE1VXXX Displacement transdu-cer

    Voltage 24V 2000mV mV red black yellow white shield

    L3400VSC Surface strain-gauge 400mm

    Voltage 24V ±10mV mV red black white green shield

    S411HV10 Uniaxial in-place inclinometer

    Voltage 24V ±700mV mV red black white green shield

    S411HV20 Uniaxial in-place inclinometer

    Voltage 24V ±1400mV mV red black white green shield

    S411HV90 Uniaxial in-place inclinometer

    Voltage 24V ±2100mV mV red black white green shield

    S421PVXX In-place inclinometer mono

    Voltage 24V ±400mV mV red black yellow white shield

    S521PVXX Waterproof clinometer mono

    Voltage 24V ±400mV mV red black yellow white shield

    L2CT0XXX Electrical load cell Wheatstone Bridge

    ±2.5V ±2.0mV/V mV/V red black blue green yellow white shield

    L2XXVXXX Electrical load cell Wheatstone Bridge

    ±2.5V ±2.0mV/V mV/V red black blue green yellow white shield

    L3400VS0 Surface strain-gauge 400mm

    Wheatstone Bridge

    ±2.5V ±2.0mV/V mV/V red black blue green yellow white shield

    P235SXXV Electrical piezometer Wheatstone Bridge

    ±2.5V ±2.0mV/V mV/V red black blue green yellow white shield

    P252AXXV Pressure transducer Wheatstone Bridge

    ±2.5V ±3.0mV/V mV/V red black blue green yellow white shield

    P252CXXV Removable piezometer Wheatstone Bridge

    ±2.5V ±2.0mV/V mV/V red black blue green yellow white shield

    P252RXXV Vented pressure transducer

    Wheatstone Bridge

    ±2.5V ±3.0mV/V mV/V red black blue green yellow white shield

    S411SVXX Servo in-place inclino-meter

    Servo MonoAx ±12V ±5000mV mV red white black yellow green shield

    S521SVXX Waterproof servo clinometer mono

    Servo MonoAx ±12V ±5000mV mV red white black yellow green shield

    S531PVXX Clinometer mono Servo MonoAx 24V ±400mV mV red black yellow white shield

  • 39BACK INDEXMultipurpose Readout Unit CRD-400 - Rev.0

    Instrument Code Instrument Name Measure Type Supply Signal Unit A B C D E F G

    S531SVXX Servo clinometer mono Servo MonoAx ±12V ±5000mV mV red white black yellow green shield

    SCLIN45X Tilli portable tiltmeter Servo MonoAx ±12V ±5000mV mV red white black yellow green shield

    T111PT10 Thermometer Pt100 RTD Pt100 1mA 1000_ °C, mV, _

    red white black green shield

    T3800GKA Built-in thermistor NTC NTC Thermistor 0.1mA 10000_ °C, mV, _

    white green shield

    D232TXVW Vibrating Wire Embankment extensometer

    V.W.and Ther-mistor

    400-6000 Hz

    Hz red black white green shield

    D313F0XX Vibrating Wire deformometer

    V.W.and Ther-mistor

    400-6000 Hz

    Hz red black white green shield

    D313SXXX Vibrating Wire crackmeter

    V.W.and Ther-mistor

    400-6000 Hz

    Hz red black white green shield

    D314CXXX V.W. embedmaent jointmeter

    V.W.and Ther-mistor

    400-6000 Hz

    Hz red black white green shield

    D422RXXVW V.W. DSM embank. V.W.and Ther-mistor

    400-6000 Hz

    Hz red black white green shield

    DTEXXXVW V.W. Linear transducer V.W.and Ther-mistor

    400-6000 Hz

    Hz red black white green shield

    L143DXXX Vibrating wire pressure cell 230mm

    V.W.and Ther-mistor

    400-6000 Hz

    Hz red black white green shield

    PK45AXXX Vibrating wire piezo-meter

    V.W.and Ther-mistor

    400-6000 Hz

    Hz red black white green shield

    PK45CXXX Removible vibrating wire piezometer

    V.W.and Ther-mistor

    400-6000 Hz

    Hz red black white green shield

    PK45HXXX Vibrating wire pressure gauge

    V.W.and Ther-mistor

    400-6000 Hz

    Hz red black white green shield

    PK45IXXX Drive-in vibrating wire piezometer 350KPa

    V.W.and Ther-mistor

    400-6000 Hz

    Hz red black white green shield

    PK45SXXX Vibrating wire piezo-meter

    V.W.and Ther-mistor

    400-6000 Hz

    Hz red black white green shield

    VK4000SM V.W. Shotcreate strain-gauge

    V.W.and Ther-mistor

    400-6000 Hz

    Hz red black white green shield

    VK4000VS Vibrating wire surface strain-gauge

    V.W.and Ther-mistor

    4,043E+9 400-6000 Hz

    Hz red black white green shield

    VK4100VK Spot weldable Geokon 4100

    V.W.and Ther-mistor

    3,910E+8 400-6000 Hz

    Hz red black white green shield

    VK4200VC Embedment vibrating wire strain-gauge

    V.W.and Ther-mistor

    3,814E+9 400-6000 Hz

    Hz red black white green shield

    VK4200VN No-stress vibrating wire embedment strain-gauge

    V.W.and Ther-mistor

    3,814E+9 400-6000 Hz

    Hz red black white green shield

    VK4210VC Strain-gauge Geokon 4210

    V.W.and Ther-mistor

    3,568E+8 400-6000 Hz

    Hz red black white green shield

    VKBAR0XX Vibrating wire Rebar V.W.and Ther-mistor

    400-6000 Hz

    Hz red black white green shield

  • 40BACK INDEXMultipurpose Readout Unit CRD-400 - Rev.0

    Type of measurement

    mA - mV - V - mV/V - °C - Hz - μS - digit - μ

    A/D converter 24-bit Sigma-Delta ADC (22 true bit)

    Measuring range 0-20mA - ±20mV - ±1V - ±10V - ±10mV/V - Pt100 (from -150 to +150°C) - NTC (from -30 to +150°C) - from 400 to 6000Hz

    Measuring resolution

    1μA with F.S. 20mA1μV with F.S. ±20mV10μV with F.S. ±1V100μV with F.S. ±10V0.001mV/V with F.S. 10mV/V0.1°C with F.S. 1000Ω (Pt100)0.1°C with F.S. 10000Ω (NTC)0.1 Hz with F.S. from 400 to 6000Hz

    Accuracy 0.01% F.S.

    Temperature drift 0.001 % F.S. / °C

    Imput impedance >1 0MΩ for voltage < 2.5V, > 1MΩ for voltage > 2.5V

    TECHNICAL SPECIFICATIONS

    Additional information

    Rechargeable battery

    4 x AA, NiMH, 2700 mAh

    Battery average working time

    6h (always in measurement, maximum backlight, 2700mAh batteries, at 25°C)

    Battery charger Programmable charger, IP41, input voltage: 100 ÷ 240 V~, 50-60 Hz, 1.3A

    Display Amorphous silicon TFT LCD panel with LED backlight unit,320 x 240, 3.5", sunlight reliability

  • 41BACK INDEXMultipurpose Readout Unit CRD-400 - Rev.0

    Operating temperature from -20°C to +60°C

    Storage temperature from -30°C to +70°C

    Humidity max. 80%

    Max. height of use 2000 m

    Enviromental Conditions

    Physical Characteristics

    Weight 507 gr

    Dimensions (L x W x H) 100 x 35 x 220 mm

    Protection Degree IP67

    Material ABS

    Connectors 1 x instrument, 1 x battery charger

    Certifications

    Eletromagnetic compatibility EN 61326-1 (2006)

    Safety requirements EN 61010-1 (2001)