Applying the innovations offered by additive Layer Manufacturing to solve traditional limitations Limitations in Implant Design Joint replacements are becoming a more common procedure as our ageing population’s expectations of being active into later life exceed the life expectancy of natural joints. These procedures are most frequently performed on the hip and knee, although the number of other joints now being treated, such as the shoulder, elbow and ankle, is increasing. No matter the location, the process typically involves the removal of the patient’s diseased or damaged joint and replacement with metal, plastic or ceramic bearing surfaces attached to a metal support structure which must be fixed into the surrounding bone. Bone has evolved to be a complex composite material, a stiff dense shell encapsulating a honeycomb structure with anisotropic behaviour thanks to an intricate substructure. The introduction of a ‘lump’ of metal can have a dramatic impact on the load path from the joint, as the relatively stiff metal transfers the load and shields the surrounding bone from the stress it would otherwise see. The biological response of this unloaded bone is for it to be resorbed, so this area reduces in density and weakens. This is linked to the most commonly observed failure mechanism, with the potential of reduced implant fixation, pain and even bone fracture. A Solution A suggested solution to this problem has been to reduce the stiffness of the implant, thus lowering the differential between the metal and bone. This could be done by changing the component material; however there are extremely stringent rules and limitations on the materials which can be implanted into the human body. A lot of work has been done to optimise the external profile of implants; therefore an approach to reducing the stiffness is to make the component semi-solid or PARTNER SHOWCASE MSC Software: Partner Showcase - Simpleware Image Based Modeling for Biomedical Implant Design Partner Showcase: Simpleware | 1 Key Highlights: Industry Biomedical Challenge To reduce the stiffness of the hip replacement implant MSC Software Solutions Marc to analyze identical boundry conditions & loads Benefits Sucessful testing of theoretical ideas Improve implant design Simpleware, an MSC Software Community Partner, offers ScanIP+FE software tools which generate volume and/or surface meshes, contact surfaces and material properties from segmented scan data. ese high quality meshes can be directly imported into Patran and MSC Nastran, as well as a range of other commercial FE and CFD Packages.

Transcript of MSC Software: Partner Showcase - Simpleware Image Based … · MSC Software is the worldwide leader...

Page 1: MSC Software: Partner Showcase - Simpleware Image Based … · MSC Software is the worldwide leader of multidiscipline simulation solutions that help companies improve quality, save

Applying the innovations offered by additive Layer Manufacturing to solve traditional limitations

Limitations in Implant DesignJoint replacements are becoming a more common procedure as our ageing population’s expectations of being active into later life exceed the life expectancy of natural joints. These procedures are most frequently performed on the hip and knee, although the number of other joints now being treated, such as the shoulder, elbow and ankle, is increasing. No matter the location, the process typically involves the removal of the patient’s diseased or damaged joint and replacement with metal, plastic or ceramic bearing surfaces attached to a metal support structure which must be fixed into the surrounding bone.

Bone has evolved to be a complex composite material, a stiff dense shell encapsulating a honeycomb structure with anisotropic behaviour

thanks to an intricate substructure. The introduction of a ‘lump’ of metal can have a dramatic impact on the load path from the joint, as the relatively stiff metal transfers the load and shields the surrounding bone from the stress it would otherwise see. The biological response of this unloaded bone is for it to be resorbed, so this area reduces in density and weakens. This is linked to the most commonly observed failure mechanism, with the potential of reduced implant fixation, pain and even bone fracture.

A SolutionA suggested solution to this problem has been to reduce the stiffness of the implant, thus lowering the differential between the metal and bone. This could be done by changing the component material; however there are extremely stringent rules and limitations on the materials which can be implanted into the human body. A lot of work has been done to optimise the external profile of implants; therefore an approach to reducing the stiffness is to make the component semi-solid or

PARTNER SHOWCASE

MSC Software: Partner Showcase - Simpleware

Image Based Modeling for Biomedical Implant Design

Partner Showcase: Simpleware | 1

Key Highlights:

IndustryBiomedical

ChallengeTo reduce the stiffness of the hip replacement implant

MSC Software SolutionsMarc to analyze identical boundry conditions & loads

Benefits•Sucessful testing of theoretical ideas

•Improve implant design

Simpleware, an MSC Software Community Partner, offers ScanIP+FE software tools which generate volume and/or surface meshes, contact surfaces and material properties from segmented scan data. These high quality meshes can be directly imported into Patran and MSC Nastran, as well as a range of other commercial FE and CFD Packages.

Page 2: MSC Software: Partner Showcase - Simpleware Image Based … · MSC Software is the worldwide leader of multidiscipline simulation solutions that help companies improve quality, save

PARTNER SHOWCASE

2 | MSC Software

MSC Software: Partner Showcase - Simpleware

hollow. Traditional manufacturing techniques have not been appropriate for creating such a structure, either hollow or with an internal micro-architecture, but recent developments in additive layer manufacturing or 3D printing using metals have opened up many new possibilities.

Using Simpleware and Marc to Generate and Analyze Novel Implant ConstructionsProfessor Mark Taylor, from the University of Southampton in the UK, and Simpleware have been collaborating to investigate how lowering femoral implant stiffness through design can influence load transfer through the femur following hip replacement surgery.

Simpleware provides a world leading solution for the conversion of 3D image data into high quality surface and volume meshes for 3D printing, CAD, FEA and CFD. The software was used to generate several implanted models of a femur, with externally identical implant designs but internally either being solid, hollow or containing a microstructure.

The ScanIPTM module of Simpleware was used to generate the femur model from clinical CT scans of a healthy male. Once a 3D representation of the femur was produced it was modified using 3D editing tools to simulate surgical procedure, i.e. removal of the femoral head. Simpleware’s +CAD module was then used to import and position a basic CAD implant into the image (Figure 1).

Once positioned, the implanted CAD model was used to create three instances for this initial investigation; model A with a solid stem, model B with a hollow stem and model C with an internal structure. The internal

details of models B and C were made using Simpleware’s Internal Structures Wizard. This tool allows the user to interactively select a unit cell shape from a library and use this to fill any volume, defining the cell unit size, its volume fraction and an encasing shell thickness.

Simpleware’s robust and flexible meshing algorithms were used to mesh the model geometries. Simpleware has been developed to segment, reconstruct and mesh the complex and arbitrarily complex structures that can be captured from volumetric imaging modalities such as CT, µCT and MRI. Therefore meshing the intricate structure of the internal micro-architecture was possible (Figure 2). Each mesh produced had matched nodes and elements at the interface between bone and implant, with high mesh quality suitable for direct use in MSC Software’s Marc nonlinear simulation software.

Simpleware exports an input file, not just a mesh; this allowed the material properties of the femur to be directly assigned from the image. There is a linear relationship between the grey level in the image data and apparent bone density, and researched empirical relationships between density and modulus. These factors were simply typed into Simpleware allowing the effect of the inhomogeneous material properties of the femur to be incorporated into the later analysis through the assignment of element-wise modulus values (Figure 3).

The Findings...The models were all imported directly into Marc for analysis, where identical boundary conditions and loads were applied; simulating the peak forces acting through the femur

“The combination of Simpleware and Marc has been able to test the potential for new manufacturing techniques to address long standing problems to improve implant design.” Rebecca Bryan, Simpleware, MSC Technology Partner

Figure 1: Simpleware +CAD module, showing positioning of the femoral implant into the segmented and edited femur model.

Page 3: MSC Software: Partner Showcase - Simpleware Image Based … · MSC Software is the worldwide leader of multidiscipline simulation solutions that help companies improve quality, save

PARTNER SHOWCASE

Partner Showcase: Simpleware | 3

MSC Software: Partner Showcase - Simpleware

during normal walking. The results metric chosen was equivalent von Mises stress and plots of this are shown through the implant and bone for the three configurations.

The results show that the conventional, solid stem represented by Model A transfers the majority of the load through the stem and to the bone surrounding its lower end. The stress in the bone above decreases away to very low levels, particularly on the medial side (left side in the image). It is possible that this shielding could result in bone loss around this area, potentially destabilising the implant.

The hollow stem, Model B, and structured stem, Model C, show an improved stress distribution in that there is load transfer more evenly down the length of the implant. This

would hopefully make bone resorbtion less likely and reduce the related complications.

However, the hollow stem is an extreme example and may not be practical in service due to the possibility of buckling. The introduction of an internal microstructure was simple to design within Simpleware and the model used in the analysis can be sent directly to a 3D printer to be made. The structure reduces the weight of the component as well as its stiffness compared to the solid version. Traditional methods would not be able manufacture this design, however direct metal laser sintering machines can build a design like this layer by layer.

As a preliminary study this shows the potential for new manufacturing techniques to address

long standing problems. The combination of Simpleware and Marc has been able to test the theoretical ideas suggested for improving implant design. The workflow established will allow future studies to interrogate this concept much further, as the flow from image to CAD model integration to internal structure addition to solution was straightforward and robust. Further work will be able to extend the scope of the study, for example comparing different internal structure designs and densities and pushing the analysis to include further measures which can predict the ‘health’ of an implant such as micro-motion between it and the surrounding bone.

For more information please visit www.simpleware.com

Dr. Rebecca Bryan, Simpleware Ltd.

Professor Mark Taylor, University of Southampton, UK

Figure 2: Example image of meshed implanted femur with introduced internal structure. Highlighted zone shows matched nodes and elements across part interfaces.

Figure 3: Plot of the modulus distribution through the femur, automatically assigned from the image grey level.

Figure 4: Plots of equivalent von Mises stress for the three models analysed; Model A – solid stem, Model B – hollow stem and Model C – stem with internal structure.

Page 4: MSC Software: Partner Showcase - Simpleware Image Based … · MSC Software is the worldwide leader of multidiscipline simulation solutions that help companies improve quality, save

PARTNER SHOWCASE

Europe, Middle East, AfricaMSC Software GmbHAm Moosfeld 1381829 Munich, GermanyTelephone 49.89.431.98.70

Asia-PacificMSC Software Japan LTD. Shinjuku First West 8F23-7 Nishi Shinjuku1-Chome, Shinjuku-KuTokyo, Japan 160-0023Telephone 81.3.6911.1200

Asia-PacificMSC Software (S) Pte. Ltd. 100 Beach Road#16-05 Shaw Tower Singapore 189702Telephone 65.6272.0082

CorporateMSC Software Corporation2 MacArthur PlaceSanta Ana, California 92707Telephone 714.540.8900www.mscsoftware.com

The MSC Software corporate logo, MSC, and the names of the MSC Software products and services referenced herein are trademarks or registered trademarks of the MSC Software Corporation in the United States and/or other countries. All other trademarks belong to their respective owners. © 2012 MSC Software Corporation. All rights reserved.

SIMPLEWARE*2012SEPT*PS

About MarcAdvanced Nonlinear & Multiphysics

Marc is a powerful, general-purpose, nonlinear finite element analysis solution to accurately simulate the response of your products under static, dynamic and multi-physics loading scenarios. Marc’s versatility in modeling nonlinear material behaviors and transient environmental conditions makes it ideal to solve your complex design problems. With its innovative technologies and modeling methodologies, Marc enables you to simulate complex real world behavior of mechanical systems making it best suited to address your manufacturing and design problems in a single environment.

With the solution schemes that are smarter and designed to provide the performance that you need by taking full advantage of your hardware combined with an easy to use modeling solution, you can truly discover and explore nature’s inherent nonlinearities. Whether your problems involve large deformation and strains, nonlinear materials, complex contact or interaction between multiple physics, you have reached the end of your search and with Marc, you can now focus on your improving your designs.

Please visitwww.mscsoftware.com

for more partner showcases

About MSC SoftwareMSC Software is the worldwide leader of multidiscipline simulation solutions that help companies improve quality, save time and reduce costs associated with designing and testing manufactured products. MSC Software works with thousands of companies worldwide to develop better products faster with simulation technology, software, and services. MSC Software is a global company with offices in 20 countries. For additional information about MSC Software’s products and services, please visit www.mscsoftware.com.