M&S Series Atomic Absorption Systems Thermo M5 AA Performance Specifications
Thermo Electron Corporation's Atomic Absorption Range offers revolutionary stylish and fully automatic AA
Spectrometers, coupled with intelligent accessories capable of outstanding performance.
M Series Dual Atomizer Systems
The concept of the M Series is simplicity itself. Using a Stockdale double beam Echelle optical system means
that each beam passes through a separate atomizer. The user defines which is the sample beam and hence which
atomizer will be used. As a result flame to furnace changeover can be fully automatic and take place
unattended. Productivity is significantly enhanced and performance guaranteed.
S Series Single Atomizer Systems
The radical S Series instruments offer the highest levels of multi-element productivity and reduced cost of
ownership for no compromise, single atomizer analysis. Standard features include QuadLine background
correction, double beam optics, Universal inert burner, furnace and burner auto-optimization, plus full QC
protocols.
Mi and Si Spectrometers
Enhanced M and S Series AA Spectrometers with added Intelligence. Choosing the Mi or Sispectrometer gives
you a truly unique instrument that is capable of Intelligent Spectrometer Qualification (iSQ). The iSQ module
and dedicated software automatically tests the spectrometer hardware performance and can be set to perform the
tests even before you arrive in the lab. Arrive in the morning to find an iSQ report confirming that you're good
to go!
AA Series Atomic Absorption Spectrometer Specification: Spectrometer Models Si Spectrometer
S Series Spectrometer with added intelligence. Instrument capable of performing Intelligent Spectrometer
Qualification (iSQ).
S2 Spectrometer
AA Spectrometer with standard or wide range photomultiplier. 6 Lamp Automatic Carousel. Semi-Automatic
Gas System.
IET - Used Lab Equipment - Refurbished Analytical Laboratory Instruments
Page 2 of 23
S4 Spectrometer
AA Spectrometer with standard or wide range photomultiplier. 6 Lamp Automatic Carousel. Fully Automatic
Gas System.
Mi Spectrometer
M Series Spectrometer with added intelligence. Instrument capable of performing Intelligent Spectrometer
Qualification (iSQ).
M5 Spectrometer
Dual Atomiser AA Spectrometer with standard or wide range photomultiplier. Flame Atomization in LH
Sample Compartment. Furnace Atomization in RH Sample Compartment. Vapour Atomization in either Sample
Compartment.
M6 Spectrometer
Dual Atomizer Zeeman AA Spectrometer with wide range photomultiplier. Flame Atomization in LH Sample
Compartment. Zeeman Furnace Atomization in RH Sample Compartment. Vapour Atomization in either
Sample Compartment.
MQZ Spectrometer
Zeeman Furnace AA Spectrometer with wide range photomultiplier. Furnace or Vapour Atomization.
MQZe Spectrometer
Dual Atomizer Zeeman/Vapour AA Spectrometer with wide range photomultiplier. Vapour Atomization in LH
Sample Compartment. Zeeman Furnace Atomization in RH Sample Compartment
Main Spectrometer Descriptions Mi or Si Spectrometers
Choosing the Mi or Si will give you Intelligent Spectrometer Qualification (iSQ). The iSQ module consists of a
motorized filter wheel containing a set of calibrated filters (traceable to a primary set held by the National
Physical Laboratory, UK (NPL)), the iSQ module is mounted in the instrument (LH side for S Series, RH side
for M Series) and provides the following suite of automated system performance checks:
• Wavelength accuracy
• Photometric stability
• Monochromator resolution
• Photometric stability
• D2 Background correction accuracy
• Lamp Peaking Test
• Beam Selector Movement Test (Additionally for Zeeman instruments only)
• Polarizer Orientation Test
• Polarizer Repeatability Test
The iSQ software can be programmed to start automatically. Run the iSQ tests once a day, week, month or year
to check the performance of your spectrometer hardware and increase confidence in your data quality.
S Series
Automatic Atomic Absorption Spectrometer system, with full multi-element capability, controlled with a data
station running SOLAAR software under Windows®. Flame absorption /emission systems are extendible
IET - Used Lab Equipment - Refurbished Analytical Laboratory Instruments
Page 3 of 23
to Graphite Furnace, Vapour or Mercury Cold Vapour absorption modes by ordering the appropriate accessory
atomizer.
M Series
Automatic Dual Sample Compartment Atomic Absorption System, with full multielement capability, controlled
by a data station running SOLAAR software under Windows®. Flame absorption /emission systems can be
extended to Graphite Furnace, Vapour or Mercury Cold Vapour absorption modes by ordering the appropriate
accessory atomizer. The MQZ and MQZe spectrometers must be ordered with a Zeeman Graphite Furnace to
provide a working system. Optical system All systems (except the MQZ) are supplied with Stockdale double
beam optics giving maximum signal-to-noise ratio during signal measurement. All mirrors are silica coated,
and the optics are sealed against dust ingress. The self-calibrating Ebert (S Series) or Echelle (M Series)
monochromator provides automatic wavelength and bandpass set-up. The high energy QuadLine background
correction system corrects for up to 2A of background with less than 2 % error, and is fitted as standard to all
instruments. AA Series spectrometers are fitted with a 6 position data-coded, auto aligning lamp carousel, with
6 independent power supplies modulated at 200/240 Hz. With the carousel, automatic analyses of up to 16
elements in a single method are possible. The standard range photomultiplier covers the wavelength range from
185 nm to 760 nm, with a wide range option covering 180 nm to 900 nm. GFTV is an option on all systems.
Flame System
All AA flames are supported by the Universal 50 mm titanium burner with chamfered slot design giving high
solids handling, excellent flame stability and low carbon build-up. The burner height is automatically
adjustable, and controls are provided for the burner rotation and transverse position. An optional 100 mm burner
for air/acetylene flames is also available. Automatic flame ignition is provided by a high tension electric spark.
Flame systems are controlled by either a fully automatic gas control system (S4 and M Series) or a semi-
automatic gas control system (S2). The S4 automatic gas control system uses programmable array state logic
and binary flow switching technology for reliability. Oxidant gas changeover, and fuel and oxidant flow rates
are software controlled and interlocked to prevent operation with incorrect burners and gas flows. With the
S2 semi-automatic gas system, the fuel flow rate is controlled by a precision needle valve and rotameter display.
Oxidant changeover and fuel gas flow boost are software controlled, and the burner interlock ensures
burner/flame compatibility. On both systems, a flame sensor, responding to all flame types including hydrogen
fuelled flames, ensures safe gas handling if the flame goes out. All fuel gas lines are fitted with flame arrestors,
and with pressure sensors to provide automatic safe shutdown if the line pressure falls below the safe limits.
Both gas control systems shut down safely if the power fails. The totally enclosed, draught proof “kitchen” area
has a safety door containing a heat and UV absorbing window. The inert fluoroplastic spray chamber contains
an inert adjustable impact bead and flow spoiler, and low memory hydrophilic disc chamber front end. It has
built in over pressure relief and automatic empty drain protection. The standard spray chamber configuration is
suitable for all flame types. The inert nebulizer with a Pt/Ir capillary and PTFE venturi is standard and is factory
optimized.
Graphite Furnace Systems S Series
The GFS97 Integrated Graphite Furnace and Furnace Autosampler can be fitted without removing the spray
chamber or disconnecting the gas lines.
M Series
All Graphite Furnaces are fitted in the RH Sample Compartment of the M Series Spectrometers. The M5
Spectrometers will accept the GF95 Graphite Furnace, while the M6, MQZ and MQZe Spectrometers will
IET - Used Lab Equipment - Refurbished Analytical Laboratory Instruments
Page 4 of 23
accept the GF95Z Zeeman Graphite Furnace. The GF95Z Zeeman furnace is mounted on a prealigned Tilt
Mount, which can be tilted forward out of the sample compartment to allow access to the furnace for
maintenance purposes, and to allow other accessories to be fitted into the sample compartment. The
GF95 Furnace is fitted with a pre-aligned Fixed Mount as standard, but can be optionally fitted with the Tilt
Mount if required. The FS95 Graphite Furnace Autosampler is mounted in front of the RH Sample
Compartment with a simple self-locating 'push in' mount.Graphite Furnace TeleVision (GFTV) Optional on all
AA Series Spectrometers, the GFTV accessory provides high definition images of events inside the Graphite
Furnace cuvette, allowing monitoring of the sample injection, dry and ash phases of the furnace program. A
CCD camera mounted in the instrument produces live video for display on the Data Station's screen through the
GFTV Software and Interface Kit, or on an external monitor. The external monitor should be monochrome,
PAL compatible with a minimum screen resolution of 400 lines, and a coaxial video cable with BNC plug for
connection to the spectrometer.
Electronics
Spectrometer electronics are based on the Motorola 68340 microprocessor, with the system firmware residing in
Flash PROM. Firmware can be downloaded to the Flash PROM from an external PC without removing any
covers. Interfacing to the Data Station, autosamplers, furnace and associated accessories is via RS232C at 9600
baud. The vapour system is interfaced via a dedicated 8 pin socket.
Data Station Control
The spectrometer system and accessories are controlled via a Data Station running the SOLAAR AA Software
Package under the Microsoft Windows® 98SE, Me, NT v4, 2000 or XP operating systems. The Data Station is
a Personal Computer, with a typical configuration of: 800 MHz Pentium processor or equivalent. 128MB RAM.
CD ROM Drive. Fixed Disk Drive with at least 500 Mbyte free space. SVGA Colour Monitor. 9 way Serial
Port, with type 16550 buffered RS232 interface chip. Parallel Port. Windows® 98SE, Me, NTv4, 2000 or XP
Professional. Additional requirements for GFTV: Free USB port or free PCI expansion slot. Video adapter that
supports Direct Draw.
Detailed Specifications
Optics
S Series Configuration - Double Beam Stockdale optical system. M5 Configuration - Double Beam Stockdale
system M6, MQZe Configuration - Double Beam Stockdale system, with self-referencing Single Beam Zeeman
system MQZ Configuraion - Self-referencing Single Beam Zeeman system Monochromator S Series - Ebert
configuration with nominal 0.2, 0.5 and 1.0 nm spectral bandwidths. Half height slit settings available.
Reciprocal linear dispersion 1.5 to 2.0 nm/mm. Focal length 270 mm. Grating 1800 l/mm. M Series - Echelle
monochromator and prism post-monochromator. Nominal 1.0, 0.5 and 0.2 nm spectral bandpass with 0.1 nm
available below 400 nm. 0.5 nm/mm reciprocal linear dispersion at 200 nm
Absorbance Range
-0.150 A to 3.000 A (including background signal).
Background Correction
QuadLine (continuum source) is standard on all spectrometers. Background signals below 2 A are corrected
with less than 2 % error. Total signal up to 3 A.
IET - Used Lab Equipment - Refurbished Analytical Laboratory Instruments
Page 5 of 23
Gas Control
Fuel Flow Range
Air/Acetylene - 0.8 to 2.3 l/min. Nitrous Oxide/Acetylene - 3.6 to 5.1 l/min. S2 - Needle valve and rotameter
fuel flow control. S4 and M Series Automatic binary switching fuel flow control.
Safety
Automatic flame ignition. Software controlled, automatic oxidant changeover. Software controlled, automatic
fuel gas boost on oxidant changeover. Automatic flame shutdown. Fuel line flashback arrestor. Fuel line
pressure regulator. Fuel and oxidant line pressure sensors. Flame Present sensor. Burner Type sensor. Burner
Present sensor. Power failure protection. Empty drain protection. Spray chamber over-pressure protection.
Light Source
Single or multi-element hollow cathode lamps. Uncoded and Data Coded hollow cathode lamps. 6 independent
lamp power supplies, each providing 0 to 20 mA.
Wavelength Range
180 nm to 900 nm with wide range photomultiplier. 185 nm to 760 nm with standard photomultiplier.
Data/Control Systems Spectrometer
Motorola 68340 microprocessor. 512 kB Flash PROM. 256 kB RAM. 4 general purpose RS232C ports.
Dedicated vapour port.
Data Station
Typically Pentium PC (see below).
Local Control
Motorola 68340 microprocessor.
512 kB Flash PROM.
512 kB RAM.
Power
100 to 240 V (±10 %) at 50 or 60 Hz. Consumption 300 VA.
Dimensions (mm)
S Series 567W x 510H x 589D. M Series 780W x 520H x 595D.
Weight
S Series Approximately 33 kg. M Series Approximately 60 kg.
SOLAAR Software.
Data Station Software
The SOLAAR Data Station Software runs on the AA Series Data Station, and will control both S Series and M
Series spectrometers. The software is aware of the type of spectrometer that it is controlling, and will only
provide relevant parameters and options for each instrument type and configuration. The Data Station Software
supports all the facilities provided by all the AA Series Accessories.
The software features a range of Wizard based procedures and a tabbed Methods dialogue to make
setting up even complex analyses fast and easy. All raw data, results and parameters are stored in a single
IET - Used Lab Equipment - Refurbished Analytical Laboratory Instruments
Page 6 of 23
database for easy retrieval. True Windows® multi-tasking is available, so that SOLAAR can be iconized while
running an analysis, allowing concurrent use of other applications. Methods can be created, edited and saved,
and earlier results can be reviewed, edited and printed, while an analysis is underway. SOLAAR software
enables full 16 element Methods to be performed automatically. The flexible and comprehensive tabbed
Methods dialogue box guides the user through setting up complex, multielement analyses. Data coded HCLs are
fully supported, and facilities are provided to record lamp usage. Up to 10 calibration standards and a blank can
be used, with segmented curve, linear or quadratic least squares fit calibration algorithms. Normal, standard
addition and standard curve calibration methods are supported, with full graphical display and print out of the
calibration curve. Full alphanumeric sample details including sample mass and dilution correction can be
entered for each sample, or imported via text files. Comprehensive furnace signal graphics are provided, and
automated ash-atomize plots produce graphical data for furnace program optimization, and will even suggest
optimum values. For flame systems, fully automatic fuel gas flow and burner height optimization with graphical
presentation are provided.
Extensive on line, fully context sensitive, Help facilities are available. Carefully
structured “books” allow easy browsing and searching as well as direct access to any Help topic. Trouble-
shooting and diagnostics sections describe simple experiments to first locate the problem then to rectify it. A
full “Cookbook” with default parameters for all elements, common interferences, and the means of overcoming
them is included.
SOLAAR software provides comprehensive printer support through the Windows® operating system.
Report formatting enables customization of the sample details, results, summaries, methods, system parameters
and graphical displays that are to be included in the Report, and all Reports, as well as individual results, are
time and date stamped. Flexible and comprehensive Results database filters select and display the required data,
which can be exported to other applications for further processing. Fully undoable results editing is available
and is automatically audit trailed. Integrated Auto QC software provides comprehensive Quality Control
protocols automation with checking, testing and reanalysis options, SPC graphical results display, report
generation and data filing. QC Blanks, QC Checks, QC Duplicates and two forms of QC Spikes are all included,
each with user definable test criteria and failure actions. Full QC reporting, including pass/fail results and time
and date stamps is provided on the Results display, and hard copy reporting of all QC protocols, parameters,
results and actions is available. QC data is linked to the results to which it applies, and can be exported with the
results. SOLAAR software is available in English, USA English, French, German, Russian, Spanish, Japanese,
Chinese and Polish languages. SOLAAR security Software The SOLAAR securitysoftware package adds
additional facilities to the SOLAAR Data Station software. This provides the tools, facilities and functions
required to allow an organization to comply with the requirements of the US Food and Drugs Administration
set out in Part 11 "Electronic Records; Electronic Signatures" of Title 21 of the Code of Federal Regulations
(the 21 CFR part 11 Rule). Network compatible, this package provides User Authentication, Access
Controls, Audit Trails, Event Logs and Digital Signatures for the Data Station client software, with central
administration and storage of the user security settings. SOLAAR security is compatible with Windows NTv4,
2000 and XP Professional operating systems.
Accessory Specifications
Mi and Si Enhancement
Mi enhancement for M Series instruments and Si enhancement for S Series instruments. The Intelligent
spectrometer Qualification (iSQ) module is mounted on the LH side of the Spectrometer for S Series
instruments and on the RH side for M Series instruments.
Standard Items
Intelligent Spectrometer Qualification module. Coded Ca/Mg Hollow cathode lamp LH Accessory mount for Si
option RH Accessory mount for Mi option.
IET - Used Lab Equipment - Refurbished Analytical Laboratory Instruments
Page 7 of 23
Graphite Furnaces
GFS97 Integrated Graphite Furnace and Autosampler GFS97 is available for S Series spectrometers only.
GFS97 Furnace specifications are as the GF95 Furnace Specifications, and GFS97 Furnace Autosampler
specifications are as the FS95 Furnace Autosampler Specifications.
Furnace Head
Furnace Power supply, Furnace Autosampler, GFS97 Integrated Furnace and Autosampler, Mount. Normal
Electro graphite cuvettes (10). Necessary Additional Items: Additional cuvettes as required.
GF95 and GF95Z Graphite Furnaces
GF95 is available for M Series M5 model only. GF95Z is available for M Series M6, MQZ and MQZe models
only. GFS97 Furnace Specification is as GF95.
Standard Items
Standard (GF95) or Zeeman (GF95Z) Furnace Head, Power supply, Tilt Mount (GF95Z), Fixed Mount (GF95)
or Integrated Mount (GFS97).
Normal Electrographite cuvettes (10).
Necessary Additional Items
Additional cuvettes as required; Optional Tilt Mount for GF95.
Furnace Head
All cuvettes mount directly in an all-graphite containment with end loaded contacts. Cuvettes are self aligning,
and can be rapidly exchanged with a single lever movement. The binary flow controlled internal gas system,
with gas stop, offers a choice of the inert gas or an alternate gas, and the fixed external inert gas flow protects
the cuvette and purges the optical temperature sensor.
Temperature Control System
Voltage Feedback Control System. Optical Temperature Feedback Control System with a stabilized sensor and
fiber optic transmission system.
Gas System
Pressure 1.03+/-0.07 bar (15+/- 1psi) fixed external flow. Adjustable internal flow with choice of inert or
alternate gas.
Cooling Water
Mains or Recirculator. Minimum flow rate 0.7 l/min. Pressure 1.4 - 6.9 bar (20 - 100 psi).
Furnace Control
Control of the furnace is from the SOLAAR Data Station or Local Control software via the spectrometer. A full
handshaking RS232C link is provided to the furnace power supply. Furnace cycles are set up within a rapid fill
table allowing up to 20 phases to be programmed for temperature, time, ramp rate, gas type and flow and
special functions such as optical temperature control, read phase and non linear ramp. (Limited facilities
with Local Control). Individual temperature look-up tables for all cuvette types ensure accurate temperature
calibration. A lifetime display enables the cuvette to be changed before failure becomes likely. A reprogrammed
cuvette clean cycle is available, culminating in maximum temperature for 5 seconds to ensure complete
IET - Used Lab Equipment - Refurbished Analytical Laboratory Instruments
Page 8 of 23
contamination removal. Comprehensive storage of the furnace program, together with spectrometer and
autosampler parameters, is provided. Full furnace and autosampler status can be displayed, together with the
results during the analysis (Data Station control only). Complete monitoring of all furnace interlocks
is provided.
FS95 Furnace Autosampler
The attractively styled Autosampler will accommodate up to 60 samples in the large carousel. Carousel
changeover allows up to 500 samples per run. 6 large reagent cups enable up to 6 matrix modifiers to be
employed. Using the cup reducing ring, smaller sample cups can be used in the reagent cup positions to reduce
modifier consumption. The sample is collected and dispensed via an inert PTFE capillary with easily
interchangeable tips. Viscous and Normal modes of injection ensure that samples are handled appropriately and
an enhanced wash program eliminates contamination. Fast furnace operation provides concurrent operation of
the furnace, spectrometer, and autosampler sample uptake procedures, minimizing overall furnace cycle time.
The syringe is visible and easily accessible for routine maintenance. A large 1 litre capacity wash vessel for
long unattended analysis is complemented by an on-board waste vessel, removing the need for inconvenient
waste collection on the laboratory floor.
All facilities are programmed from the SOLAAR software and sampling facilities such as matrix
modification (wet or dry mixing), standards addition, dilution, reconcentration and automatic standards
preparations can be rapidly set-up. Automatic rescaling and re-calibration functions are provided. If a sample
falls outside the calibration range the system automatically and intelligently calculates the most appropriate
dilution factor required to bring the sample back within range. Autosampler loading guides indicate the
optimum carousel layout.
Standard Items
Furnace Autosampler, Polypropylene Sample Cups (1000), Polypropylene Reagent Cups (50), Spare Capillary
tips (10).
Carousel Capacity
60 sample cups, 6 reagent cups.
Cup Types
Polypropylene sample cups, Fluoroplastic sample cups, Reduced volume sample cups, Polypropylene reagent
cups.
Cup Volume (nominal)
Sample cups 2 ml., Reduced volume cups 1.5 ml., Reagent cups 25 ml.
Inert Gas Pressure
Nominal - 0.34 bar (5psi), Minimum - 0.20 bar (3psi), Maximum - 0.68 bar (10psi).
Injection
Sample Volume 1-70 uL, Reproducibility - better than 1% (by mass), at volumes equal to and greater than 10
uL.
221XL Flame Autosampler
Random access flame and vapour autosampler with 60 (Rack 21), 44 (Rack 22) or 14 (Rack 24) sample
capacities. Smaller sample capacities imply larger available volume per sample.
IET - Used Lab Equipment - Refurbished Analytical Laboratory Instruments
Page 9 of 23
Standard Items
Random access x-y autosampler mechanism, Rack 21 (60 position), PTFE probe, RS232C adaptor.
Necessary Additional Items
AA Wash Accessory, Polypropylene or glass sample/standard tubes for Rack 21.
Power
110 V, 120 V, 220 V, 240 V at 50 or 60 Hz, Consumption 50 VA.
222XL Flame Autosampler
Random access, large capacity flame and vapour autosampler with 70 to 254 sample capacity, dependent on
rack configuration. Smaller sample capacities imply larger available volume per sample. First rack (14 position)
is used for blank and standards.
Standard Items
Random access x-y autosampler mechanism. Rack 24 (14 position). PTFE probe. RS232C adaptor.
Necessary Additional Items
AA Wash Accessory. Polypropylene or glass sample/standard tubes for Rack 24. 4 of Rack 24 (14 positions),
Rack 22 (44 positions) or Rack 21 (60 positions) with appropriate polypropylene or glass sample/standard
tubes.
AA Wash Accessory
The AA Wash Accessory provides wash facilities for the 221XL and 222XL autosamplers.
Standard Items
Constant Head wash reservoir. Wash well for 221XL/ 222XL autosamplers. ASX-520 and ASX-260 Flame
Autosampler Random access, large capacity (ASX-520) or compact (ASX-260) flame/vapour autosampler
with 84 to 360 max.(ASX-520) and 42 to 180 max (ASX-260). sample capacity, dependent on rack
configuration. Optional racks for 21, 24, 40, 60 or 90 samples. Smaller sample capacities imply larger available
volume per sample.
Standard Items
Random access x-y autosampler mechanism. Standards rack (10 position). ASX-520 4 sample racks (60
position each). ASX-260 2 sample racks (60 position each). 10 standard and 240 sample polypropylene
sample tubes. 0.5 mm ID and 0.8 mm ID PTFE sample probes. Pumped wash facility. RS232C adaptor.
ID100 Autodilutor
The ID100 Autodilutor provides rapid, in-line preparation of calibration standards and intelligent or fixed ratio
dilution of samplesfor flame measurements. It is based on a high precision multi-piston pump that is calibrated
for life, and does not require any consumable items. It can be used with manual sampling or with any of the
flame autosamplers.
VP100 Continuous Flow Vapour Generator
The VP100 Continuous Flow Vapour System performs hydride and mercury vapour measurements with
significant sensitivity improvement over normal flame techniques. An air/acetylene flame or an electrically
heated cell is used for atomization, avoiding the need to use hydrogen as the fuel gas. The VP100 unit
incorporates the reagent reservoirs, a variable speed 4 channel peristaltic pump, control electronics and
IET - Used Lab Equipment - Refurbished Analytical Laboratory Instruments
Page 10 of 23
Gas Liquid separator. A integrated Mass Flow Controller allows the carrier gas flow to be controlled and
monitored through the system software, and the continuous flow principle eliminates the need to clean the
reaction vessel after each sample. The VP100 operates automatically under Data Station control and can provide
full autosampling operation with a suitable autosampler. When used with the EC90 Electrically Heated
Atomization Cell, unattended operation is possible.
Standard Items
T-Cell (2) - Open ended, T-shaped silica cell (120 mm long), Mercury Cell - Long path (150 mm), Spare pump
tubing, Spare glass Beads, Spare semi-permeable membrane.
Necessary Additional Items
An appropriate accessory mount is required to fit the long path length cell to the spectrometer.
Cell Heating
Air/acetylene flame on 50 mm Universal burner. Electrical heating using EC90.
Carrier/Purge Gas: Argon or nitrogen, Gas Inlet Pressure: 0.35 - 1 bar (5 - 14 psi).
Gas Consumption: Up to 500 ml/min.
Reducing Agents: Sodium borohydride, Tin (II) chloride may be used for mercury analysis.
Solution Transport
Variable speed, 4 channel peristaltic pump, using continuous flow principle.
Solution Flow Rates
Reductant - 3.2 ml/min, Sample - 7.8 ml/min.
Carrier Gas Flow Rates: Variable 50 - 500 mL/min, Electronic control via Mass Flow Controller.
Reagent Containers: Two 1 L plastic bottles.
Power: 220/240 V, 50 Hz or 110/120 V, 60 Hz. Consumption 30 VA.
Dimensions (mm): 475W x 300H x 200D.
EC90 Electrically Heated Atomization Cell
The EC90 is an electrically heated atomization cell for flameless vapour generation AAS. It provides improved
analytical performance, unattended operation and reduced operating costs in comparison with a flame heated
atomization cell. The EC90 must be used with the VP90 or VP100 and is suitable for all hydride forming
elements.
Standard Items
EC90 Furnace Head., EC90 Furnace Power Supply, 2 Silica T Cells.
Necessary Additional Items
VP90 or VP100 Continuous Flow Vapour System. Appropriate Accessory Mount.
IET - Used Lab Equipment - Refurbished Analytical Laboratory Instruments
Page 11 of 23
Control System: From external power supply.
AA Validator
Available in three separate kits designed for Flame, Furnace or Flame and Furnace installations, the unique AA
Validator provides full system validation for S Series and M Series AA systems, including comprehensive
qualification plans. Full SOP's and check sheets, together with complete supplier assessment questionnaires, are
supplied in the Validator Logbook for recording and demonstrating compliance. All necessary consumables are
provided and even method validation guidelines are included. The integrated PQ Test software contains a
complete set of SOPs and a Wizard to lead the user through the tests needed to validate the instrument
analytical performance against the manufacturer's specification.
Major Contents (depends on kit type):
Validator Logbook, Ca/Mg Hollow Cathode Lamp, Ni/Cr/Mn Hollow Cathode Lamp, Pyrolytically Coated
Cuvettes, Manganese Validation Standard, Nickel Validation Standard, Chromium Validation Standard.
Water Blank Validation Standard.
AA Validatorplus
The Validatorplus accessory contains a set of traceable and certified filters. It is fitted in the spectrometer and
controlled by the OQ Test software.
Standard Items
Calibrated Validation Unit (CVU), Power Supply, RS232 Data Lead
Necessary Additional Items
Either the Validator Logbook or the appropriate Validator Kit. RH Universal Accessory Mount for use with
M Series spectrometers. LH Universal Accessory Mount for use with S Series spectrometers.
Aliquot Micro-sampling Accessory
This is a simple accessory that provides manual Aliquot Micro-sampling facilities. Aliquot Micro-sampling is
particularly useful when the available volume of sample is limited, or the sample is too viscous, or contains too
high a concentration of dissolved solids, to allow conventional aspiration.
Standard Items: Teflon block. Uptake tube.
Necessary Additional Items: Adjustable Micropipette
Slotted Tube Atom Trap
The STAT accessory provides a useful increase in flame analytical sensitivity for certain more volatile
elements, such as Cd, Zn, Pb, and Cu.
Standard Items
STAT Holder, 5 Quartz STAT tubes.
Necessary Additional Items
50 mm Universal Titanium Burner, Variable Flow Auxiliary Oxidant Kit. The Variable Flow Auxiliary Oxidant
Kit allows the flow rate of the auxiliary oxidant gas that can be supplied to an air/acetylene flame to be varied
over a wide range, in order to allow the flame chemistry to be optimized when using less common organic
solvents. Note that all flame instruments fitted with automatic gas control modules include a fixed flow rate
IET - Used Lab Equipment - Refurbished Analytical Laboratory Instruments
Page 12 of 23
auxiliary oxidant facility for air/acetylene flames that is suitable for the organic solvents most commonly used.
The Variable Flow Auxiliary Oxidant Kit is not compatible with the S2 spectrometer.
Standard Items
Variable Flow Auxiliary Oxidant Kit, Solvent Resistant Flame Kit. This kit contains the necessary replacement
flame system parts recommended when the spectrometer is to be used extensively for the analysis of organic
solvents. It includes solvent resistant 'O' rings, a modified spray chamber, and solvent resistant drain tubing.
LH Universal Accessory Mount
Accessory Mount for use in the S Series and left hand M Series sample compartments. This is necessary for
mounting the EC90 Furnace Head and the Mercury Absorption Cell in these sample compartments. It is also
used for mounting the Validatorplus Calibration Validation Unit in the S Series sample compartment.
RH Universal Accessory Mount
Accessory Mount for use in the RH M Series sample compartment. This is necessary for mounting the EC90
Furnace Head, the Validatorplus Calibration Validation Unit, and the Mercury Absorption Cell in this sample
compartment.
Retrospective Accessory Mount Kit
This kit is required for mounting certain accessories in the LH sample compartment of older (part number 9423
400 20###), M Series Spectrometers, and earlier 969/989 Spectrometers.
Graphite Furnace Cuvettes
Normal (Electro graphite) Cuvettes
Description: The normal cuvette is made from a form of graphite called electrographite. This cuvette is suitable
for the determination of volatile elements such as Pb, and Cd in simple matrices, such as clean waters.
(Pyrolytically) Coated Electro graphite Cuvettes
Description: This coated cuvette is a modified electrographite cuvette, coated with a thin layer of pyrolytic
graphite. Coated cuvettes are required for the carbide forming, mediumvolatile and refractory elements and for
samples with complex high salt matrices.
Extended Lifetime Cuvettes (ELC)
Description: ELC's are unique to Thermo and have a pyrolytic coating that is up to 10 times thicker than the
standard coating. This gives them more stable performance and much longer useful lifetimes than either of the
electrographite-based cuvettes. These cuvettes are especially recommended for high throughout analyses with
complex samples and when measuring the most refractory elements.
Omega Platform Extended Lifetime
Description: Omega platform cuvettes have an integrated L'Vov platform built into the cuvettes, which
makes them particularly suitable for determining the more volatile elements in heavy matrices where vapour
phase interferences cause severe problems. The integral platform is designed to accommodate up to 50 µL
volumes and so can provide excellent concentration limits. The Omega cuvettes use the ELC coating
technology and have similar long, stable lifetime characteristics.
IET - Used Lab Equipment - Refurbished Analytical Laboratory Instruments
Page 13 of 23
Hollow Cathode Lamps Data Coded Multi-element Lamps
High quality hollow cathode lamps optimized for use with the M Series and S Series spectrometers. The
spectrometers read the list of elements contained in the lamp and the maximum permissible lamp current from
the Data Coding, providing maximum convenience and removing the possibility of damaging the lamp with
excessive currents. Multi-element lamps provide analytical performance comparable with single element lamps,
while reducing your lamp inventory and increasing the ease of use. All Thermo Electron Corporation Data
Coded Multi-element Lamps carry a 5000 mA/hr lifetime guarantee.
Data Coded Single Element Lamps
High quality hollow cathode lamps optimized for use with the M Series and S Series spectrometers. The
spectrometers read the element contained in the lamp and the maximum permissible lamp current from the
Data Coding, providing maximum convenience and removing the possibility of damaging the lamp with
excessive currents. Single element lamps provide the best possible analytical performance. All Thermo Electron
Corporation Data Coded Single Element Lamps carry a 5000 mA/hr lifetime guarantee.
Uncoded Multi-element Lamps
High quality hollow cathode lamps optimized for use with the M Series and S Series spectrometers. Multi-
element lamps provide analytical performance comparable with single element lamps, while reducing your
lamp inventory and increasing the ease of use. All Thermo Electron Corporation Uncoded Multielement Lamps
carry a 5000 mA/hr lifetime guarantee. Uncoded Single Element Lamps High quality hollow cathode lamps
optimized for use with the M Series and S Series spectrometers. Single element lamps provide the best possible
analytical performance. All Thermo Electron Corporation Uncoded Single Element Lamps carry a 5000 mA/hr
lifetime guarantee.
Value Range Lamps
A limited range of cost effective hollow cathode lamps for use with the M Series and S Series spectrometers.
Value Range Single Element lamps provide analytical performance comparable with normal lamps, at a
significantly reduced cost. All Thermo Electron Corporation Value Range Uncoded Single Element Lamps
carry a 5000 mA/hr lifetime guarantee.
How to Order Guide
Description Part Number
Si Spectrometers
to upgrade your S Series spectrometer to an Si instrument add the following Part number to
your instrument order 9423 420 30001
S2 AA Spectrometer (standard photomultiplier) 9423 400 30021
S2 AA Spectrometer (wide range photomultiplier) 9423 400 30022
S4 AA Spectrometer (standard photomultiplier) 9423 400 30041
S4 AA Spectrometer (wide range photomultiplier) 9423 400 30042
Mi Spectrometers
IET - Used Lab Equipment - Refurbished Analytical Laboratory Instruments
Page 14 of 23
to upgrade your M Series spectrometer to an Mi instrument add the following Part number
to your instrument order 9423 420 20001
M5 Dual Atomizer AA Spectrometer (wide range photomultiplier) 9423 400 21052
M6 Dual Atomizer Zeeman AA Spectrometer 9423 400 21062
MQZ Zeeman AA Spectrometer 9423 400 21072
MQZe Dual Atomizer Zeeman/Vapour AA Spectrometer 9423 400 21572
Major Options and Accessories Part Number
Graphite Furnace and Furnace Autosampler
For S Series Spectrometers
GFS97 Integrated Graphite Furnace and Autosampler
9423 490 30001 or 9423 490
30002 (N. America)
For M Series Spectrometers
GF95 Graphite Furnace (for M5 spectrometers only)
9423 490 20001 or 9423 490
20002 (N. America)
GF95Z Zeeman Graphite Furnace (for M6, MQZ and MQZe
spectrometers)
9423 490 20051 or 9423 490
20052 (N. America)
GF95Z - with Rhodium Plated centre block
9423 490 20061 or 9423 490
20062 (N. America)
Rhodium Plated centre block for GF95Z - retrospective fitting 9423 490 20069
FS95 Furnace Autosampler 9423 480 20001
Graphite Furnace TeleVision
GFTV (for M Series Spectrometers, Factory Fitted, 50 Hz) 9423 410 20011
GFTV (for M Series Spectrometers, Factory Fitted, 60 Hz) 9423 410 20021
GFTV (for S Series Spectrometers, Factory Fitted, 50 Hz) 9423 410 30011
GFTV (for S Series Spectrometers, Factory Fitted 60 Hz) 9423 410 30021
GFTV Software and Interface Kit (USB version) 9423 393 96051
GFTV Software and Interface Kit (PCI version) 9423 410 20005
GFTV Software and Interface Kit is necessary if an external monitor is
not used. It cannot be used if the spectrometer is controlled with a
Local Control module.
An appropriate TV license may be required.
GFTV Monitor 9423 420 20003
IET - Used Lab Equipment - Refurbished Analytical Laboratory Instruments
Page 15 of 23
A GFTV Monitor is necessary if the GFTV Software and Interface Kit
is not used, or if the spectrometer is controlled by a Local Control
module.
Control Options
All spectrometers include SOLAAR software as standard.
Local Control module (Factory Fitted) 9423 410 20002
Data Station (110 V/120 V) 9423 393 10003
Data Station (220 V/240 V) 9423 393 10004
SOLAAR Upgrade Software (includes software for earlier systems) 9423 430 20001
SOLAAR security Software (ordered with spectrometer) 9423 430 30011
SOLAAR security Software (upgrade for existing systems) 9423 430 30001
Note that this includes the SOLAAR Upgrade software.
Description Part Number
Flame and Vapour Autosamplers
Gilson 221XL Autosampler 110 - 120 V/200 - 240 V 9423 470 03001
Gilson 222XL Autosampler 110 - 120 V/200 - 240 V 9423 470 03002
AA Wash Kit for Gilson autosamplers 9423 470 03501
CETAC ASX-520 Autosampler 110 - 120 V/200 - 240 V 9423 470 04001
CETAC ASX-260 Autosampler 110 - 120 V/200 - 240 V 9423 470 04002
Autodilutor
ID100 Autodilutor 9423 470 03201
Continuous Flow Vapour Generation System
VP100 Continuous Flow Vapour System 9423 460 10001
Electrically Heated Atomization Cell
EC90 Electrically Heated Atomization Cell 9423 393 60061
Must be used with a VP90 or VP100 Vapour System and a Universal Accessory Mount
to provide a working system.
It is essential for vapour analysis with the MQZ and MQZe spectrometers.
Validator
Validator Log Book 9499 VAL 90006
FAAS Validator Package 9423 VAL 00001
GFAAS Validator Package 9423 VAL 00002
IET - Used Lab Equipment - Refurbished Analytical Laboratory Instruments
Page 16 of 23
FAAS/GFAAS Validator Package 9423 VAL 00003
Validatorplus Accessory 9423 VAL 00004
Aliquot Sampling
Manual Aliquot Micro-sampling Accessory 9423 390 05201
Slotted Tube Atom Trap
STAT Accessory (requires use of 50 mm Universal Burner) 9423 390 35011
Variable Flow Auxiliary Oxidant
Variable Flow Auxiliary Oxidant Kit (Factory Fitted) 9423 393 46111
Solvent Resistant Flame Kit 9423 420 31051
Universal Accessory Mounts
LH Sample Compartment Universal Accessory Mount 9423 420 21005
This is required for mounting certain accessories in the LH Sample Compartment of
current (part number 9423 400 21###) M Series Spectrometers, and in the Flame
Compartment of S Series Spectrometers.
RH Sample Compartment Universal Accessory Mount 9423 420 20005
This is required for mounting certain accessories in the RH Sample Compartment of all
M Series Spectrometers.
Retrospective Accessory Mount Kit 9423 420 22005
This is required for mounting certain accessories in the LH Sample Compartment of
older (part number 9423 400 20###) M Series Spectrometers, and earlier 969/989
Spectrometers.
Air Supply
Air Compressor (220/240 V, 50 Hz) (includes filter/regulator unit) 9423 393 34225
Air Compressor (220/240 V, 60 Hz) (includes filter/regulator unit) 9423 393 34226
Air Compressor (110/120 V, 50 Hz) (includes filter/regulator unit) 9423 393 34115
Air Compressor (100/120 V, 60 Hz) (includes filter/regulator unit) 9423 393 34116
Printers
Colour inkjet printer (NOT suitable for use with Local Control module) 9423 440 20001
Epson compatible Dot Matrix Printer (220/240 V) 9423 393 50011
Epson compatible Dot Matrix Printer (110/120 V) 9423 393 50021
Parallel Printer Cable 4013 172 59554
IET - Used Lab Equipment - Refurbished Analytical Laboratory Instruments
Page 17 of 23
General Accessories and Consumables
Description Part Number
Spectrometer
Wide Range Photomultiplier Field Upgrade 9423 390 08341
S Series Spectrometers GFTV Field Upgrade (50 Hz) 9423 410 30011
A GFTV Software and Interface Kit, or a GFTV Monitor will be
required to make a working system.
S Series Spectrometers GFTV Field Upgrade (60 Hz) 9423 410 30021
A GFTV Software and Interface Kit, or a GFTV Monitor will be
required to make a working system.
M Series Spectrometers GFTV Field Upgrade (50 Hz) 9423 410 20011
A GFTV Software and Interface Kit, or a GFTV Monitor will be
required to make a working system.
M Series Spectrometers GFTV Field Upgrade (60 Hz) 9423 410 20021
A GFTV Software and Interface Kit, or a GFTV Monitor will be
required to make a working system.
Accessory Trolley 9423 393 46131
User Spares Kit for older M Series Spectrometers 9423 450 20001
Older M Series Spectrometers have part numbers 9423 400 20###.
User Spares Kit for current M and S Series Spectrometers 9423 450 30001
Current M Series Spectrometers have part numbers 9423 400 21###.
Background Correction
QuadLine deuterium source for current M and S Series Spectrometers 9423 420 30004
Current M Series Spectrometers have part numbers 9423 400 21###.
QuadLine deuterium source for older M Series Spectrometers 9423 420 20004
Older M Series Spectrometers have part numbers 9423 400 20###.
Gas Handling
These items are suitable for British Oxygen Company equipment only.
Acetylene Gas Pressure Regulator 9423 352 43881
Acetylene Flashback Arrestor 9423 393 46061
Hydrogen Flashback Arrestor 9423 393 46071
Nitrogen/Argon/Air Gas Pressure Regulator 9435 179 21201
Nitrous Oxide Gas Pressure Regulator 9423 354 03111
IET - Used Lab Equipment - Refurbished Analytical Laboratory Instruments
Page 18 of 23
Flame Compartment
Universal (50 mm) Titanium Burner 9423 420 31011
Air/acetylene (100 mm) Titanium Burner 9423 420 31021
Retrofit Universal (50 mm) Titanium Burner 9423 420 35011
Retrofit Air/acetylene (100 mm) Titanium Burner 9423 420 35021
The Retrofit Burners are required for M Series Spectrometers, and
earlier 969/989 Spectrometers.
older (part number
9423 400 20###)
Spare Inert Nebulizer 9423 390 05481
Nebulizer Uptake Tube - 0.4 mm ID (for reduced uptake rate) 9423 390 05411
Nebulizer Uptake Tube - 0.5 mm ID 9423 390 05421
Nebulizer Cleaning Probe 9423 390 05441
Variable Flow Auxiliary Oxidant Kit Field Upgrade 9423 393 46111
Spray Chamber Renovation Kit 9423 390 05461
Micro-adjustable Impact Bead Kit 9423 390 05401
5 litre Drain Vessel 9423 390 05471
Solvent Resistant Drain Tubing 9423 390 05491
LH Universal Accessory Mount (S Series and current M Series) 9423 420 21005
RH Universal Accessory Mount (M Series only) 9423 420 20005
Retrofit Accessory Mount Kit (for older M Series, part number 9423 400
20###, and earlier instruments) Gilson Flame and Vapour Autosampler. 9423 420 22005
Polypropylene Tubes for Rack 21, 8 ml, 1000 per pack 9423 470 03311
Glass Tubes for Rack 21, 9 ml, 100 per pack 9423 470 03321
Rack 21 (60 position) 9423 470 03301
Polypropylene Tubes for Rack 22, 17 ml, 10 per pack 9423 470 03331
Glass Tubes for Rack 22, 20 ml, 100 per pack 9423 470 03341
Rack 22 (44 position) 9423 470 03391
Polypropylene Tubes for Rack 24, 50 ml, 10 per pack 9423 470 03351
Glass Tubes for Rack 24, 50 ml, 30 per pack 9423 470 03361
Rack 24 (14 position) 9423 470 03381
Spare Teflon Sample Probe 9423 470 03211
222XL Autosampler Rack Tray 9423 470 03471
IET - Used Lab Equipment - Refurbished Analytical Laboratory Instruments
Page 19 of 23
Gilson Autosampler to Nebulizer capillary tube 9423 393 68121
Rack 81 Kit (supports 96 well microplates on 222XL only) 9423 470 03481
Description Part Number
CETAC Flame and Vapour Autosampler
21 Position Sample Rack 9423 470 03901
24 Position Sample Rack 9423 470 03911
40 Position Sample Rack 9423 470 03921
60 Position Sample Rack 9423 470 03931
90 Position Sample Rack 9423 470 03941
Polypropylene Tubes for 21 position rack and Standards
rack, 50 ml, 500 per pack 9423 470 04151
Polypropylene Tubes for 24 position rack, 30 ml, 500 per
pack 9423 470 04111
Polypropylene Tubes for 40 position rack, 20 ml, 500 per
pack 9423 470 04161
Polypropylene Tubes for 60 position rack, 14 ml, 1000 per
pack 9423 470 04131
Polypropylene Tubes for 90 position rack, 8 ml, 1000 per
pack 9423 470 04141
Sample Probe Assembly 9423 470 03981
Sample Probe, 0.5 mm ID 9423 470 03991
Sample Probe, 0.8 mm ID 9423 470 04171
Tray Sub-assembly with a Standards Rack 9423 470 03951
Rinse Station 9423 470 03961
Drain Pump Tubing and Connector Kit 9423 470 03971
Autodilutor
ID100 Autodilutor Consumables Kit 9423 450 03201
Graphite Furnace
Water Recirculator (50 Hz) 9423 393 97005
Water Recirculator (60 Hz) 9423 393 97003
GF95 Tilt Mount 9423 490 20005
Graphite Contact Cones for GF95/GFS97 9423 393 95011
Contact Cone Replacement Tool for GF95/GFS97 9423 393 95101
IET - Used Lab Equipment - Refurbished Analytical Laboratory Instruments
Page 20 of 23
Graphite Contact Cones for GF95Z 9423 393 95161
Contact Cone Replacement Tool for GF95Z 9423 393 95181
Adjustable Micropipette, 0 - 20 µl 9423 357 05421
Pipette Tips, 20 µl, 1000 per pack 9423 357 05431
GF95 User Spares Kit 9423 450 20002
GF95Z User Spares Kit 9423 450 20003
GFS97 User Spares Kit 9423 450 30002
Furnace Autosampler
Polypropylene Sample Cups, 2 ml, 1000 per pack 9423 393 80031
Polypropylene Reagents Cups, 25 ml, 50 per pack 9423 393 80021
PTFE Sample Cups, 2ml, 20 per pack 9423 393 80051
Reduced Volume Sample Cups, 1.5 ml, 20 per pack 9423 393 80061
Cup Reducing Ring 9423 393 80071
Spare Capillary Tips 9423 393 81261
Replacement PTFE Sampling Capillary Tube 9423 393 82261
Carousel and Tray 9423 393 83261
Furnace Autosampler User Spares Kit 9423 450 20004
Slotted Tube Atom Trap
Spare Quartz Slotted Tubes, 5 per pack 9423 393 35021
Dot Matrix Printer
Spare Printer Paper 9423 299 40501
Spare Printer Ink Ribbon 9423 299 40801
Validation
Nickel Validation Standard 9423 VAL 00011
Chromium Validation Standard 9423 VAL 00012
Manganese Validation Standard 9423 VAL 00013
Water Blank Validation Standard 9423 VAL 00014
Re-calibration of Validatorplus and/or iSQ module
Accessory 9423 VAL R0004
IET - Used Lab Equipment - Refurbished Analytical Laboratory Instruments
Page 21 of 23
Graphite Furnace Cuvettes
Description Part Number
Graphite Furnace Cuvettes
Normal Graphite Cuvettes, Ridged and Uncoated, 10 per pack 9423 393 95031
Pyrolytically Coated Graphite Cuvettes, Ridged, 10 per pack 9423 393 95071
Extended Lifetime Cuvettes (ELC), 10 per pack 9423 393 95041
Extended Lifetime Cuvettes (ELC), 20 per pack 9423 393 95051
Omega Platform Cuvettes, 10 per pack 9423 490 20101
Hollow Cathode Lamps Multi-Element Lamps
Description Uncoded P/N Coded P/N Description Uncoded P/N Coded P/N
Ag/Cr/Cu 9423 393 21501 9423 393 31501 Co/Ni 9423 393 21211 9423 393 31211
Ag/Cu 9423 393 21491 9423 393 31491 Cr/Cu 9423 393 21131 9423 393 31131
Al/Mg 9423 393 21051
9423 393
31051 Cr/Mn
9423 393
21141 9423 393 31141
Al/Si 9423 393 21041 9423 393 31041 Cr/Ni 9423 393 21121 9423 393 31121
Ba/Sr 9423 393 21591 9423 393 31591 Cr/Ni/Cu/Mn 9423 393 21171 9423 393 31171
Cu/Fe/Zn 9423 393 21531 9423 393 31531 Cr/Ni/Mn
9423 393
21161 9423 393 31161
Ca/Ba 9423 393 21571 9423 393 31571 Cu/Fe
9423 393
21221 9423 393 31221
Ca/Ba/S 9423 393 21601 9423 393 31601 Cu/Fe/Mn/Zn 9423 393 21561 9423 393 31561
Ca/Mg 9423 393 21011 9423 393 31011 Cr/Ni/Cu 9423 393 21151 9423 393 31151
Ca/Mg/Si 9423 393 21061 9423 393 31061 Cu/Mn
9423 393
21231 9423 393 31231
Ca/Si 9423 393 21081 9423 393 31081 Cu/Mn/Zn
9423 393
21541 9423 393 31541
Ca/Sr 9423 393 21581
9423 393
31581 Cu/Ni 9423 393 21241 9423 393 31241
Cd/Zn 9423 393 21611 9423 393 31611 Fe/Mn
9423 393
21251 9423 393 31251
Co/Cu 9423 393 21181 9423 393 31181 Fe/Ni
9423 393
21261 9423 393 31261
Co/Cu/Fe 9423 393 21281 9423 393 31281 Fe/Zn
9423 393
21511 9423 393 31511
Co/Fe 9423 393 21191 9423 393 31191 Mg/Si 9423 393 21091 9423 393 31091
Co/Mn 9423 393 21201 9423 393 31201 Mn/Zn
9423 393
21521 9423 393 31521
IET - Used Lab Equipment - Refurbished Analytical Laboratory Instruments
Page 22 of 23
Co/Mn/Ni 9423 393 21291 9423 393 31291 Na/K
9423 393
21021 9423 393 31021
Co/Mo 9423 393 21111
9423 393
31111 Si/Ti 9423 393 21101 9423 393 31101
Hollow Cathode Lamps Single Element Lamps
Description
Uncoded
Data
Coded
Value
Lamp Description
Uncoded
Data
Coded Value Lamp
Ag 20471 30471 2047 V Nb 20411 30411
Al 20131 30131 2013 V Nd 20601 30601
As 20331 30331 2033 V Ni 20281 30281 2028 V
Au 20791 30791 2079 V Os 20761 30761
B 20051 30051 P 20151 30151
Ba 20561 30561 Pb 20821 30821 2082 V
Be 20041 30041 Pd 20461 30461
Bi 20831 30831 2083 V Pr 20591 30591
Ca 20201 30201 2020 V Pt 20781 30781
Cd 20481 30481 2048 V Rb 20371 30371
Co 20271 30271 2027 V Re 20751 30751
Cr 20241 30241 2024 V Rh 20451 30451
Cs 20551 30551 Ru 20441 30441
Cu 20291 30291 2029 V Sb 20511 30511 2051 V
Dy 20661 30661 Sc 20211 30211
Er 20681 30681 Se 20341 30341 2034 V
Eu 20631 30631 Si 20141 30141
Fe 20261 30261 2026 V Sm 20621 30621
Ga 20311 30311 Sn 20501 30501 2050 V
Gd 20641 30641 Sr 20381 30381
Ge 20321 30321 Ta 20731 30731
Hf 20721 30721 Tb 20651 30651
Hg 20801 30801 2080 V Te 20521 30521
Ho 20671 30671 Th 20901 30901
IET - Used Lab Equipment - Refurbished Analytical Laboratory Instruments
Page 23 of 23
In 20491 30491 Ti 20221 30221
Ir 20771 30771 Tl 20811 30811
K 20191 30191 2019 V Tm 20691 30691
La 20571 30571 U 20921 30921
Li 20031 30031 V 20231 30231
Lu 20711 30711 W 20741 30741
Mg 20121 30121 2012 V Y 20391 30391
Mn 20251 30251 2025 V Yb 20701 30701
Mo 20421 30421 2042 V Zn 20301 30301 2030 V
Na 20111 30111 2011 V Zr 20401 30401