945-2 Thalamic Infarct. Neuroimaging Figure 1. Right medial thalamic infarct.
MRS Magnetic Resonance SpectroscopyApplications Summary Thalamic proton MRS measures acquired soon...
Transcript of MRS Magnetic Resonance SpectroscopyApplications Summary Thalamic proton MRS measures acquired soon...
![Page 1: MRS Magnetic Resonance SpectroscopyApplications Summary Thalamic proton MRS measures acquired soon after birth in neonatal encephalopathy had the highest accuracy to predict neurodevelopment](https://reader030.fdocuments.us/reader030/viewer/2022040812/5e5721608deb6716420fb852/html5/thumbnails/1.jpg)
MRSMagnetic Resonance
SpectroscopyIn vivo biochemistry
![Page 2: MRS Magnetic Resonance SpectroscopyApplications Summary Thalamic proton MRS measures acquired soon after birth in neonatal encephalopathy had the highest accuracy to predict neurodevelopment](https://reader030.fdocuments.us/reader030/viewer/2022040812/5e5721608deb6716420fb852/html5/thumbnails/2.jpg)
Chemical shift
Ø www.imaios.com/en/e-Courses/e-MRI/Image-quality-and-artifacts/chemical-shift
![Page 3: MRS Magnetic Resonance SpectroscopyApplications Summary Thalamic proton MRS measures acquired soon after birth in neonatal encephalopathy had the highest accuracy to predict neurodevelopment](https://reader030.fdocuments.us/reader030/viewer/2022040812/5e5721608deb6716420fb852/html5/thumbnails/3.jpg)
Chemical shift and MR spectrum
Ø www.imaios.com/en/e-Courses/e-MRI/Image-quality-and-artifacts/chemical-shift
FourierTransform
Received Signal Image
![Page 4: MRS Magnetic Resonance SpectroscopyApplications Summary Thalamic proton MRS measures acquired soon after birth in neonatal encephalopathy had the highest accuracy to predict neurodevelopment](https://reader030.fdocuments.us/reader030/viewer/2022040812/5e5721608deb6716420fb852/html5/thumbnails/4.jpg)
1H NMR spectra
ü Peaks pattern is footprint of molecules
ü Peak area is proportional to the number of 1H nuclei
![Page 5: MRS Magnetic Resonance SpectroscopyApplications Summary Thalamic proton MRS measures acquired soon after birth in neonatal encephalopathy had the highest accuracy to predict neurodevelopment](https://reader030.fdocuments.us/reader030/viewer/2022040812/5e5721608deb6716420fb852/html5/thumbnails/5.jpg)
Proton spectroscopy
Ø The water signal has to be suppressed
H2O
lipids
![Page 6: MRS Magnetic Resonance SpectroscopyApplications Summary Thalamic proton MRS measures acquired soon after birth in neonatal encephalopathy had the highest accuracy to predict neurodevelopment](https://reader030.fdocuments.us/reader030/viewer/2022040812/5e5721608deb6716420fb852/html5/thumbnails/6.jpg)
Proton spectroscopy
Ø The water signal has to be suppressed
H2O
lipids
Lipids extract
![Page 7: MRS Magnetic Resonance SpectroscopyApplications Summary Thalamic proton MRS measures acquired soon after birth in neonatal encephalopathy had the highest accuracy to predict neurodevelopment](https://reader030.fdocuments.us/reader030/viewer/2022040812/5e5721608deb6716420fb852/html5/thumbnails/7.jpg)
Water suppression
Spectrum without and with water suppressionDifferent scaling
![Page 8: MRS Magnetic Resonance SpectroscopyApplications Summary Thalamic proton MRS measures acquired soon after birth in neonatal encephalopathy had the highest accuracy to predict neurodevelopment](https://reader030.fdocuments.us/reader030/viewer/2022040812/5e5721608deb6716420fb852/html5/thumbnails/8.jpg)
1H MRS metabolites
![Page 9: MRS Magnetic Resonance SpectroscopyApplications Summary Thalamic proton MRS measures acquired soon after birth in neonatal encephalopathy had the highest accuracy to predict neurodevelopment](https://reader030.fdocuments.us/reader030/viewer/2022040812/5e5721608deb6716420fb852/html5/thumbnails/9.jpg)
Brain spectroscopy
Ø Occipital cortex
In vivo 1H spectrum
ATP
PCr
Pi
PDEPME
NAACrCho
Ø The 1H (or 31P) nuclei in different molecules have slightly different resonance frequencies
Ø Each peak is related to a molecules (metabolite)In vivo 31P spectrum
![Page 10: MRS Magnetic Resonance SpectroscopyApplications Summary Thalamic proton MRS measures acquired soon after birth in neonatal encephalopathy had the highest accuracy to predict neurodevelopment](https://reader030.fdocuments.us/reader030/viewer/2022040812/5e5721608deb6716420fb852/html5/thumbnails/10.jpg)
In vivo biochemistry
![Page 11: MRS Magnetic Resonance SpectroscopyApplications Summary Thalamic proton MRS measures acquired soon after birth in neonatal encephalopathy had the highest accuracy to predict neurodevelopment](https://reader030.fdocuments.us/reader030/viewer/2022040812/5e5721608deb6716420fb852/html5/thumbnails/11.jpg)
1H spectroscopy
![Page 12: MRS Magnetic Resonance SpectroscopyApplications Summary Thalamic proton MRS measures acquired soon after birth in neonatal encephalopathy had the highest accuracy to predict neurodevelopment](https://reader030.fdocuments.us/reader030/viewer/2022040812/5e5721608deb6716420fb852/html5/thumbnails/12.jpg)
1H spectroscopy
![Page 13: MRS Magnetic Resonance SpectroscopyApplications Summary Thalamic proton MRS measures acquired soon after birth in neonatal encephalopathy had the highest accuracy to predict neurodevelopment](https://reader030.fdocuments.us/reader030/viewer/2022040812/5e5721608deb6716420fb852/html5/thumbnails/13.jpg)
31P spectroscopy
![Page 14: MRS Magnetic Resonance SpectroscopyApplications Summary Thalamic proton MRS measures acquired soon after birth in neonatal encephalopathy had the highest accuracy to predict neurodevelopment](https://reader030.fdocuments.us/reader030/viewer/2022040812/5e5721608deb6716420fb852/html5/thumbnails/14.jpg)
31P MRSIn vivo 31P spectra acquired from
the human occipital lobe at (B) 4 T and (C) 7 T:
Ø PE phosphoethanolamineØ PC phosphocholineØ Pi inorganic phosphate Ø GPE
glycerophosphoethanolamineØ GPC, glycerophosphocholineØ PCr phosphocreatine Ø ATP adenosine triphosphate Ø NADP nicotinamide adenine
dinucleotide phosphate
Qiao T. et al Magn Reson Imaging. 2006 24:1281-6
![Page 15: MRS Magnetic Resonance SpectroscopyApplications Summary Thalamic proton MRS measures acquired soon after birth in neonatal encephalopathy had the highest accuracy to predict neurodevelopment](https://reader030.fdocuments.us/reader030/viewer/2022040812/5e5721608deb6716420fb852/html5/thumbnails/15.jpg)
muscle 31PMRSØ Muscoli gastrocnemi normal subject
rest exercise
Pi
PCr
ATP
![Page 16: MRS Magnetic Resonance SpectroscopyApplications Summary Thalamic proton MRS measures acquired soon after birth in neonatal encephalopathy had the highest accuracy to predict neurodevelopment](https://reader030.fdocuments.us/reader030/viewer/2022040812/5e5721608deb6716420fb852/html5/thumbnails/16.jpg)
Signal localizationIn in vivo MRS the signal localization is mandatory
1H MRS
![Page 17: MRS Magnetic Resonance SpectroscopyApplications Summary Thalamic proton MRS measures acquired soon after birth in neonatal encephalopathy had the highest accuracy to predict neurodevelopment](https://reader030.fdocuments.us/reader030/viewer/2022040812/5e5721608deb6716420fb852/html5/thumbnails/17.jpg)
Signal localizationØ In in vivo
MRS the signal localization is mandatory
1H MRS
![Page 18: MRS Magnetic Resonance SpectroscopyApplications Summary Thalamic proton MRS measures acquired soon after birth in neonatal encephalopathy had the highest accuracy to predict neurodevelopment](https://reader030.fdocuments.us/reader030/viewer/2022040812/5e5721608deb6716420fb852/html5/thumbnails/18.jpg)
Volume selection
www.imaios.com/en/e-Courses/e-MRI/Magnetic-Resonance-Spectroscopy-MRS/single-voxel-spectroscopy
üThe localization of the MRS signal is essential for in vivo applicationüGradients are used localized spectroscopyüThe simplest localization technique is the use of the surface coil
![Page 19: MRS Magnetic Resonance SpectroscopyApplications Summary Thalamic proton MRS measures acquired soon after birth in neonatal encephalopathy had the highest accuracy to predict neurodevelopment](https://reader030.fdocuments.us/reader030/viewer/2022040812/5e5721608deb6716420fb852/html5/thumbnails/19.jpg)
Spectroscopic imaging(Chemical shift imaging CSI)
üCSI is an acquisition sequence that allows the acquisition of a spectrum per each voxelüThe acquisition time is large (> 10 minutes)
![Page 20: MRS Magnetic Resonance SpectroscopyApplications Summary Thalamic proton MRS measures acquired soon after birth in neonatal encephalopathy had the highest accuracy to predict neurodevelopment](https://reader030.fdocuments.us/reader030/viewer/2022040812/5e5721608deb6716420fb852/html5/thumbnails/20.jpg)
Spectroscopic imaging
üCSI is an acquisition sequence that allows the acquisition of a spectrum per each voxelüThe acquisition time is large
• > 10 minutes
![Page 21: MRS Magnetic Resonance SpectroscopyApplications Summary Thalamic proton MRS measures acquired soon after birth in neonatal encephalopathy had the highest accuracy to predict neurodevelopment](https://reader030.fdocuments.us/reader030/viewer/2022040812/5e5721608deb6716420fb852/html5/thumbnails/21.jpg)
Spectroscopic imaging(Chemical shift imaging CSI)üCSI is an acquisition sequence that allows the acquisition of a spectrum per each voxelüThe acquisition time is large (> 10 minutes)
AJNR 22, 4 (2007)
![Page 22: MRS Magnetic Resonance SpectroscopyApplications Summary Thalamic proton MRS measures acquired soon after birth in neonatal encephalopathy had the highest accuracy to predict neurodevelopment](https://reader030.fdocuments.us/reader030/viewer/2022040812/5e5721608deb6716420fb852/html5/thumbnails/22.jpg)
Spectroscopic imaging(Chemical shift imaging CSI)
![Page 23: MRS Magnetic Resonance SpectroscopyApplications Summary Thalamic proton MRS measures acquired soon after birth in neonatal encephalopathy had the highest accuracy to predict neurodevelopment](https://reader030.fdocuments.us/reader030/viewer/2022040812/5e5721608deb6716420fb852/html5/thumbnails/23.jpg)
Spectroscopic imaging(Chemical shift imaging CSI)
![Page 24: MRS Magnetic Resonance SpectroscopyApplications Summary Thalamic proton MRS measures acquired soon after birth in neonatal encephalopathy had the highest accuracy to predict neurodevelopment](https://reader030.fdocuments.us/reader030/viewer/2022040812/5e5721608deb6716420fb852/html5/thumbnails/24.jpg)
Quantitative data analysis
![Page 25: MRS Magnetic Resonance SpectroscopyApplications Summary Thalamic proton MRS measures acquired soon after birth in neonatal encephalopathy had the highest accuracy to predict neurodevelopment](https://reader030.fdocuments.us/reader030/viewer/2022040812/5e5721608deb6716420fb852/html5/thumbnails/25.jpg)
Applications
Preliminary evidence of the ability of 1H fMRS to detect changes in glutamate during various perceptual, motor, and cognitive tasks.
![Page 26: MRS Magnetic Resonance SpectroscopyApplications Summary Thalamic proton MRS measures acquired soon after birth in neonatal encephalopathy had the highest accuracy to predict neurodevelopment](https://reader030.fdocuments.us/reader030/viewer/2022040812/5e5721608deb6716420fb852/html5/thumbnails/26.jpg)
Applications
SummaryThalamic proton MRS measures acquired soon after birth in neonatal encephalopathy had the highest accuracy to predict neurodevelopment 2 years later.
Magnetic resonance spectroscopy assessment of brain injury after moderate hypothermia in neonatal encephalopathy:a prospective multicentre cohort study Peter J Lally et al for the MARBLE consortium
Lancet Neurol 2019; 18: 35–45
![Page 27: MRS Magnetic Resonance SpectroscopyApplications Summary Thalamic proton MRS measures acquired soon after birth in neonatal encephalopathy had the highest accuracy to predict neurodevelopment](https://reader030.fdocuments.us/reader030/viewer/2022040812/5e5721608deb6716420fb852/html5/thumbnails/27.jpg)
Applications
Methodsü 3.0 Tesla scanner ü single 15×15×15 mm3 voxel centred on the
left thalamus ü 1H MRS metabolite peak area ratios (7 min)ü 1H MRS metabolite absolute concentrations
(25 min) üdiffusion weighted MRI (DW MRI; 7 min)
Magnetic resonance spectroscopy assessment of brain injury after moderate hypothermia in neonatal encephalopathy:a prospective multicentre cohort study