MRI Physics

★ MRI physics recap★ Pulse sequences

✦ Gradient vs. spin echo★ Diffusion MRI

✦ Isotropic and anisotropic diffusion✦ Useful quantities derived from the

tensor

14

Page 2: MRI Physics - momilab · 2016-01-06 · MRI Physics ★ MRI physics recap ... pendence of MRI resonance frequency on the orientation of brain microstructure relative to the main magnetic

What is diffusion?

15

Page 3: MRI Physics - momilab · 2016-01-06 · MRI Physics ★ MRI physics recap ... pendence of MRI resonance frequency on the orientation of brain microstructure relative to the main magnetic

What is diffusion?

Random motion of particles due to thermal energy

Water molecules collide and experience net displacement

Displacement described by diffusion coefficient (D)

Normally, diffusion is isotropic (equal in all directions)

√2DT

16

Page 4: MRI Physics - momilab · 2016-01-06 · MRI Physics ★ MRI physics recap ... pendence of MRI resonance frequency on the orientation of brain microstructure relative to the main magnetic

Why is diffusion interesting?

Diffusion is restricted by tissue boundaries, membranes, etcMarker for tissue microstructure (healthy and pathology)

Unrestricted Tissue boundaries

22

Page 5: MRI Physics - momilab · 2016-01-06 · MRI Physics ★ MRI physics recap ... pendence of MRI resonance frequency on the orientation of brain microstructure relative to the main magnetic

Diffusion anisotropy in white matter

Water can diffuse more freely along white matter fibers than across them

23

Page 6: MRI Physics - momilab · 2016-01-06 · MRI Physics ★ MRI physics recap ... pendence of MRI resonance frequency on the orientation of brain microstructure relative to the main magnetic

Δ

+A+A

Diffusion contrast

b ∝ A2∆

S = S0e−bD

Sig

nal

Tissue 1(low D)

Tissue 2(high D)

24

Page 7: MRI Physics - momilab · 2016-01-06 · MRI Physics ★ MRI physics recap ... pendence of MRI resonance frequency on the orientation of brain microstructure relative to the main magnetic

Δ

+A+A

Diffusion contrast

S = S0e−bD

b ∝ A2∆

Sig

nal

Direction 1(low D)

Direction 2(high D)

Lee, et al. PNAS, 2010;107(11):5130-35

25

Page 8: MRI Physics - momilab · 2016-01-06 · MRI Physics ★ MRI physics recap ... pendence of MRI resonance frequency on the orientation of brain microstructure relative to the main magnetic

Δ

+A+A

Diffusion contrast

S = S0e−bD

b ∝ A2∆

Sig

nal

Direction 1(low D)

Direction 2(high D)

showing parallel orientation for condition A and alternatingparallel and perpendicular orientations for condition B (Fig. 3 Aand B). Macroscopic tissue orientation was parallel to B0, and theoverall shape was similar for both conditions.Images representing signal amplitude and resonance fre-

quency from the two different microstructural orientations areshown in Figs. 3 C–F. The average resonance frequency of thetissues in condition A was −3.89 ± 0.76 Hz relative to the salinesolution. Under condition B, the resonance frequency of therotated segments C2 and C4 showed a positive shift of 0.56 ±0.67 Hz (mean difference ± pooled SD over voxels) (Fig. 3F)relative to condition A (Fig. 3E). This effect is shown moreclearly in Fig. 4, in which the average cross-sectional profile ofthe difference image is plotted. The large SD is attributed to thenonuniform phase shift within a tissue segment and the truesusceptibility variation within the tissue.Apart from the frequency shift observed inside the rotated

tissue segments, subtle effects also were seen outside the rotatedtissue segments. For example, Fig. 3F shows a slightly positivefrequency shift in the fluid immediately lateral to the rotatedtissue segments. The frequency profile in Fig. 4 suggests thepresence of a shift in the unrotated segments (C1, C3, and C5) as

well, with the strongest effect seen in C3. Inside of C1 and C5,the frequency gradually decreases close to the C1 side of theC1–C2 boundary and the C5 side of the C4–C5 boundary. Thiscan be interpreted as the effect of C2 and C4 extending outsideof those tissue sections, with an amplitude that decreases withdistance. In summary, the results suggest that the orientation ofthe microscopic structure affects MRI resonance frequency bothlocally and remotely.

DiscussionOur findings represent direct experimental evidence for a de-pendence of MRI resonance frequency on the orientation ofbrain microstructure relative to the main magnetic field. Ourexperimental design allowed the separation of microscopic andmacroscopic orientation effects that often coexist under generalin-vivo imaging conditions. Such effects have complicated the

Fig. 1. High-resolution (250 × 250-μm in-plane,2 mm slice thickness, axial slices) GRE frequencyimages at 7 T (f0 = 298.095 MHz) in vivo. Strongcontrast is seen in association with the major whitematter fiber bundles: (1) corpus callosum body; (2)corona radiata; (3) corpus callosum splenium; (4)optic radiation. Frequency was zeroed in areas withlow or absent NMR signal (e.g. outside the brain).

Fig. 2. Selection of white matter sections for postmortem MRI. (A) Fivesquare cylindrical pieces of fixed tissue were sectioned along the primaryfiber orientation in the corpus callosum (coronal view). (B) The tissues wereplaced in a groove at the bottom of a disk-shaped container. (C and D)Representations of condition A (C) and condition B (D) showing fiber ori-entations. The zoomed images show the detailed tissue structure.

Fig. 3. Results of MRI frequency measurement at the original (condition A: A,C, and E) and rotated positions of fiber bundle segments C2 and C4 (conditionB: B, D, and F). (A and B) DTI confirmation of microstructure orientation ofsegments, with red representing parallel orientations and blue representingperpendicular (through-plane) orientations relative to B0. (C and D) Magnitudeimages. (E and F) Frequency images. In F, rotation leads to a positive frequencyshift within and near the lateral edges of the rotated segments. A slightlynegative frequency shift is observed in unrotated segments. A few small dipolefield patterns, likely originating from air bubbles, are also seen in E. Theimages are from coronal scans in which the slice selection was along thephysical y-axis, the readout gradient was along the physical z-axis, and thephase encoding was along the physical x-axis. B0 was along the z-axis.

Lee et al. PNAS | March 16, 2010 | vol. 107 | no. 11 | 5131

MED

ICALSC

IENCE

S

B0

Lee, et al. PNAS, 2010;107(11):5130-35

25

Page 9: MRI Physics - momilab · 2016-01-06 · MRI Physics ★ MRI physics recap ... pendence of MRI resonance frequency on the orientation of brain microstructure relative to the main magnetic

If diffusion is present, gradients cause a drop in signal.

Faster Diffusion = Less Signal

Diffusion contrast

Posterior region of the Corona RadiataCorpus CallosumForceps MajorPosterior limb of internal capsuleAnterior limb of internal capsuleExterior capsule

26

Page 10: MRI Physics - momilab · 2016-01-06 · MRI Physics ★ MRI physics recap ... pendence of MRI resonance frequency on the orientation of brain microstructure relative to the main magnetic

Diffusion-weighted imaging

Directional encodingb=0

Fittedparameter

maps

b ∝ A2∆

Sig

nal

Direction 1

Direction 2

27

Page 11: MRI Physics - momilab · 2016-01-06 · MRI Physics ★ MRI physics recap ... pendence of MRI resonance frequency on the orientation of brain microstructure relative to the main magnetic

Diffusion anisotropy in white matter

Directionality of diffusion tells us about fiber integrity/structure and orientation

Diffusion in white matter fibers is “anisotropic”

28

Page 12: MRI Physics - momilab · 2016-01-06 · MRI Physics ★ MRI physics recap ... pendence of MRI resonance frequency on the orientation of brain microstructure relative to the main magnetic

The diffusion tensor

Displacement due to diffusion is approximately ellipsoidal

29

Page 13: MRI Physics - momilab · 2016-01-06 · MRI Physics ★ MRI physics recap ... pendence of MRI resonance frequency on the orientation of brain microstructure relative to the main magnetic

The diffusion tensor

Displacement due to diffusion is approximately ellipsoidal

29

Page 14: MRI Physics - momilab · 2016-01-06 · MRI Physics ★ MRI physics recap ... pendence of MRI resonance frequency on the orientation of brain microstructure relative to the main magnetic

The diffusion tensor

Displacement due to diffusion is approximately ellipsoidal

Eigenvectors = axes of ellipsoid (direction of fibers)Eigenvalues = size of axes (strength of diffusion)

29

Page 15: MRI Physics - momilab · 2016-01-06 · MRI Physics ★ MRI physics recap ... pendence of MRI resonance frequency on the orientation of brain microstructure relative to the main magnetic

TENSOR

λ2

λ1

λ3

Page 16: MRI Physics - momilab · 2016-01-06 · MRI Physics ★ MRI physics recap ... pendence of MRI resonance frequency on the orientation of brain microstructure relative to the main magnetic

The tensor shape follows the water diffusion

λ2

λ1

λ3

λ2 λ1

λ3

Page 17: MRI Physics - momilab · 2016-01-06 · MRI Physics ★ MRI physics recap ... pendence of MRI resonance frequency on the orientation of brain microstructure relative to the main magnetic

The Tensor

λ2 λ1 λ2

λ1

Isotropic Anisotropic

ɉ1 = longitudinal (axial) diffusivity(AD) (ɉ2 + ɉ3)/2 = radial diffusivity (RD)(ɉ1 + ɉ2 + ɉ3)/3 = mean diffusivity (MD)

λ3

λ3

Page 18: MRI Physics - momilab · 2016-01-06 · MRI Physics ★ MRI physics recap ... pendence of MRI resonance frequency on the orientation of brain microstructure relative to the main magnetic

Fractional Anisotropy (FA)

= The relative degree of anisotropy

as measured by the weighted average – dominated by the largest component

λ2 λ1 λ2

λ1

IsotropicFA = 0

AnisotropicFA = 0.8

ɉ1 = longitudinal (axial) diffusivity(AD) (ɉ2 + ɉ3)/2 = radial diffusivity (RD)(ɉ1 + ɉ2 + ɉ3)/3 = mean diffusivity (MD)

λ3

λ3 λ2

λ1 λ2

λ1

IsotropicFA = 0

AnisotropicFA = 0.8

ɉ1 = longitudinal (axial) diffusivity(AD) (ɉ2 + ɉ3)/2 = radial diffusivity (RD)(ɉ1 + ɉ2 + ɉ3)/3 = mean diffusivity (MD)

λ3

λ3

Page 19: MRI Physics - momilab · 2016-01-06 · MRI Physics ★ MRI physics recap ... pendence of MRI resonance frequency on the orientation of brain microstructure relative to the main magnetic
Page 20: MRI Physics - momilab · 2016-01-06 · MRI Physics ★ MRI physics recap ... pendence of MRI resonance frequency on the orientation of brain microstructure relative to the main magnetic

The diffusion tensor: Useful quantities

30

Page 21: MRI Physics - momilab · 2016-01-06 · MRI Physics ★ MRI physics recap ... pendence of MRI resonance frequency on the orientation of brain microstructure relative to the main magnetic

The diffusion tensor: Useful quantities

Mean diffusivity (MD): Info about tissue integrity

30

Page 22: MRI Physics - momilab · 2016-01-06 · MRI Physics ★ MRI physics recap ... pendence of MRI resonance frequency on the orientation of brain microstructure relative to the main magnetic

The diffusion tensor: Useful quantities

Fractional anisotropy (FA): how elongated is the ellipsoid? Info about fiber integrity

Mean diffusivity (MD): Info about tissue integrity

30

Page 23: MRI Physics - momilab · 2016-01-06 · MRI Physics ★ MRI physics recap ... pendence of MRI resonance frequency on the orientation of brain microstructure relative to the main magnetic

The diffusion tensor: Useful quantities

Principal diffusion direction (PDD): what direction is greatest diffusion along? Info about fiber orientation

Fractional anisotropy (FA): how elongated is the ellipsoid? Info about fiber integrity

Mean diffusivity (MD): Info about tissue integrity

30

Page 24: MRI Physics - momilab · 2016-01-06 · MRI Physics ★ MRI physics recap ... pendence of MRI resonance frequency on the orientation of brain microstructure relative to the main magnetic

Diffusion tensor imaging

Principal diffusion

direction (PDD)

Fractionalanisotropy (FA)

Meandiffusivity (MD)

At each voxel, fit the diffusion tensor modelCan then calculate MD, FA, PDD from fitted parameters

31

Page 25: MRI Physics - momilab · 2016-01-06 · MRI Physics ★ MRI physics recap ... pendence of MRI resonance frequency on the orientation of brain microstructure relative to the main magnetic

Mean diffusivity (MD)

Mean diffusion coefficient across all directionsCorrelate of tissue integrity (white and gray matter)Example: MD is altered in acute and chronic stroke

Control MD Acute Stroke

32

Page 26: MRI Physics - momilab · 2016-01-06 · MRI Physics ★ MRI physics recap ... pendence of MRI resonance frequency on the orientation of brain microstructure relative to the main magnetic

Fractional Anisotropy (FA)

Variance of diffusion coefficient across different directionsHigh in regions where diffusion is most directionalRelates to integrity of white matter fiber bundles

FA=0.8

FA=0

33

Page 27: MRI Physics - momilab · 2016-01-06 · MRI Physics ★ MRI physics recap ... pendence of MRI resonance frequency on the orientation of brain microstructure relative to the main magnetic

Principal diffusion direction (PDD)

Direction along which greatest diffusion occursRelates to direction of fiber orientations

34

Page 28: MRI Physics - momilab · 2016-01-06 · MRI Physics ★ MRI physics recap ... pendence of MRI resonance frequency on the orientation of brain microstructure relative to the main magnetic

Diffusion tractography

Follow PDD to trace white matter fibers (“tractography”)

35

Page 29: MRI Physics - momilab · 2016-01-06 · MRI Physics ★ MRI physics recap ... pendence of MRI resonance frequency on the orientation of brain microstructure relative to the main magnetic

Diffusion tractography

Follow PDD to trace white matter fibers (“tractography”)

35

Page 30: MRI Physics - momilab · 2016-01-06 · MRI Physics ★ MRI physics recap ... pendence of MRI resonance frequency on the orientation of brain microstructure relative to the main magnetic

Diffusion tractography

Jones et al

Follow PDD to trace white matter fibers (“tractography”)

35

Page 31: MRI Physics - momilab · 2016-01-06 · MRI Physics ★ MRI physics recap ... pendence of MRI resonance frequency on the orientation of brain microstructure relative to the main magnetic

Typical* Diffusion Imaging Parameters

Parameter Value Relevant points

TE(echo time) 100 ms Limited by b-value

Matrix size / Resolution

128x128 / 2 mm

Limited by distortion, SNR

Number of directions 6-60 Lower limit: tensor model

Upper limit: scan time

b-value 1000 s/mm2 Larger b = more contrast Smaller b = more signal

* Typical, not fixed!!

36

Page 32: MRI Physics - momilab · 2016-01-06 · MRI Physics ★ MRI physics recap ... pendence of MRI resonance frequency on the orientation of brain microstructure relative to the main magnetic

Motion in DWI

Diffusion gradients encode tiny displacementSubject motion is also accidentally encodedImage artefacts if we try to combine data from multiple excitations (different motion)

Linescan diffusion image

43

Page 33: MRI Physics - momilab · 2016-01-06 · MRI Physics ★ MRI physics recap ... pendence of MRI resonance frequency on the orientation of brain microstructure relative to the main magnetic

Eddy Currents

Diffusion gradients create large eddy currents, which persist into acquisition windowDistort the k-space trajectory, casing shears/scaling of images

eddy currents

48

Page 34: MRI Physics - momilab · 2016-01-06 · MRI Physics ★ MRI physics recap ... pendence of MRI resonance frequency on the orientation of brain microstructure relative to the main magnetic

Eddy Currents

Fractional Anisotropy(“variance”)

Diffusion-weighteddirections

49

Page 35: MRI Physics - momilab · 2016-01-06 · MRI Physics ★ MRI physics recap ... pendence of MRI resonance frequency on the orientation of brain microstructure relative to the main magnetic

myelin

micro-tubules

glia

Changes in FA could reflect:myelin sheathmembrane integrityaxon densityglial densityfiber coherenceetc etc

Biological interpretation:what is “integrity”?

50

Page 36: MRI Physics - momilab · 2016-01-06 · MRI Physics ★ MRI physics recap ... pendence of MRI resonance frequency on the orientation of brain microstructure relative to the main magnetic

DTI and Microstructure

Page 37: MRI Physics - momilab · 2016-01-06 · MRI Physics ★ MRI physics recap ... pendence of MRI resonance frequency on the orientation of brain microstructure relative to the main magnetic

The diffusion tensor

Diffusion tensor:3D ellipsoid

The diffusion tensor is just one possible signal model

51

Page 38: MRI Physics - momilab · 2016-01-06 · MRI Physics ★ MRI physics recap ... pendence of MRI resonance frequency on the orientation of brain microstructure relative to the main magnetic

Voxels containing multiple fibres

52

Page 39: MRI Physics - momilab · 2016-01-06 · MRI Physics ★ MRI physics recap ... pendence of MRI resonance frequency on the orientation of brain microstructure relative to the main magnetic

Voxels containing multiple fibres

52

Page 40: MRI Physics - momilab · 2016-01-06 · MRI Physics ★ MRI physics recap ... pendence of MRI resonance frequency on the orientation of brain microstructure relative to the main magnetic

How to encapsulate multiple fibres?

How to model more complex architecture• Multiple sticks?• Multiple tensors?• More sophisticated models?

53