MORPHOLOGICAL ANALYSIS AND MICROTENSILE BOND …
Transcript of MORPHOLOGICAL ANALYSIS AND MICROTENSILE BOND …
MORPHOLOGICAL ANALYSIS AND MICROTENSILE BOND
STRENGTH EVALUATION OF A SELF-ETCH ADHESIVE TO
CARIES EXCAVATED HUMAN DENTIN PREPARED BY
MECHANICAL, CHEMOMECHANICAL AND Er, Cr: YSGG
LASER SYSTEMS: AN IN VITRO STUDY.
Dissertation submitted to
THE TAMILNADU Dr. M. G. R. MEDICAL UNIVERSITY
In partial fulfillment for degree of
MASTER OF DENTAL SURGERY
BRANCH IV – CONSERVATIVE DENTISTRY &
ENDODONTICS
APRIL 2012
RR AA JJ AA SS DD EE NN TT AA LL CC OO LL LL EE GG EE
RAJA NAGAR, KAVALKINARU - 627 105, TIRUNELVELI DISTRICT.
DCI Recognition No.DE-3 (44) - 93/2246, dated 09/11/1993
Affiliated to The Tamil Nadu Dr. M.G.R. Medical University, Chennai.
DEPARTMENT OF CONSERVATIVE DENTISTRY & ENDODONTICS
CERTIFICATE
This is to certify that this dissertation entitled “Morphological Analysis and Microtensile Bond
Strength Evaluation of a Self-Etch Adhesive to Caries Excavated Human Dentin Prepared
by Mechanical, Chemomechanical and Er, Cr: YSGG Laser Systems: An in vitro study” is
a genuine work done by Dr. S. K. AJILAL under my guidance during his postgraduate study
period between 2009-2012.
This dissertation is submitted to THE TAMILNADU Dr. M. G. R. MEDICAL UNIVERSITY, in
partial fulfillment for the degree of MASTER OF DENTAL SURGERY in
CONSERVATIVE DENTISTRY & ENDODONTICS – BRANCH IV. It has not been
submitted (partial or full) for the award of any other degree or diploma.
HOD, Professor & Guide
Dr. R. Jonathan Emil Sam. MDS.
Professor & HOD
Department of Conservative Dentistry & Endodontics
Rajas Dental College & Hospital, Raja Nagar, Kavalkinaru.
ACKNOWLEDGEMENT
I take this opportunity to express my sincere and heartfelt gratitude to my postgraduate
teacher and guide Prof. Dr. R. Jonathan Emil Sam, Professor and Head of the department of
Conservative Dentistry & Endodontics, Rajas Dental College, for his inspiring guidance,
constant encouragement, advice, unlimited help and kind support throughout the course of this
study as well as during my entire course.
I express my profound gratitude to co-guide Dr. Renjith Babu, Senior Lecturer,
Department of Conservative Dentistry & Endodontics, Rajas Dental College, for having
reviewed my work and for his support, valuable guidance and supervision during the preparation
of this dissertation.
My sincere gratitude to Dr. Suresh Mohan Kumar professor, Dr. Arvind.V, assistant
professor, Department of Conservative Dentistry & Endodontics, Rajas Dental College, for their
invaluable guidance, constant encouragement and constructive suggestions throughout the course
of my study.
It had been a privilege and honor to have worked under esteemed teachers, Dr. Kashi
Vishwanath. P, Dr. Bejoy John Thomas, Dr. T S P Ram Balaji, Dr. Anoop Samuel, Dr.
Saravanan. Their in-depth knowledge and authority on subject has helped me to learn scientific
look approach to the subject.
I owe my sincere thanks to Dr. Premila Suganthan BDS, Diploma in Laser Dentistry
(IALD, AALZ Aachen University), Director, KP Tooth Care Clinic, Chennai, for her valuable
suggestions, sound advices and for granting me permission to make use of the facilities needed
for this work.
I wish to thank Dr. Neeraja Raju, Associate Professor, Deparment of Pedodontics,
Oxford Dental College Banagalore, for her valuable help during the study.
I sincerely thank Dr. Prabhakar Rao. Senior Principal Scientist & Mr. M. R.
Chandran Senior Technical Officer, of Electron Microscopy & Materials Characterization
Division of National Institute of Interdisciplinary Science & Technology (NIIST)
Thiruvananthapuram, for granting me permission to use the facilities needed for SEM study.
I am extremely thankful to Dr. Narayana Swamy, Scientist-F & Mr. Eldhose. K,
Technical Officer, Geo Science Division, Centre for Earth Science Studies (CESS), Akkulam,
Thiruvananthapuram, for their esteemed co-operation in utilizing the facilities for the
microtensile sample preparation.
My sincerely grateful to Mr. Muraleedharan C V, Engineer-G, Mr. Renjith G,
Engineer-C & Ms. Sreedevi V, Project Assistant, Organ Testing Department, BMT wing, Sree
Chitra Tirunal Institute of Medical Science & Technology (SCTIMST), Poojappura,
Thiruvananthapuram, for their valuable guidance and help in carrying out the microtensile bond
strength determination.
I am thankful to Mr. Muraleedharan Nair, Retired Associate Professor & Consultant
Bio Statistician, Government Medical College, Thiruvananthapuram for the statistical analysis.
It is my extreme pleasure to extend my gratitude to my beloved chairman Dr. Jacob
Raja and founder Chairman Dr. S. A. Raja for their valuable support and constant
encouragement throughout the period of my study.
It gives me immense pleasure to convey my deep indebtness to our respected Principal,
Dr. Suresh Sathiasekhar, Academic Director and Professor Dr. Marykutty Joseph and
Administrative Director Dr. I. Pakiaraj for the permission, help and guidance throughout my
course.
I owe my sincere thanks to Dr. S. Ignatius Rex, former professor and HOD, for his
constant encouragement and unlimited help.
I express my sincere gratitude and is indebted to my colleagues Dr. Arun Mathew
Thomas, Dr. Pramod S Prasad, Dr. Arun Balakrishnan, Dr. Gnanaseelan, Dr. Joan
Mathew, for sharing their views and ideas all through my post graduation course.
I am obliged to Dr. Ambu Suresh and Dr. Segin Chandran for their invaluable
suggestions, help and support during the study.
I am greatly thankful to my Father (Late) S. Krishna pillai, Mother (Late) P.
Saraswathy Amma, Brother Mr. S. K. Anilal, in laws Commander (Rtd) K P Rajendran
Pillai and Mrs. Radhamany Amma, who helped me to tide through all my problems with their
understanding and timely advices.
Above all I bow in gratitude to God Almighty for showering his grace and blessings all
through my life in achieving unexpected goals and proceed towards new height of destination.
Dedicated to-
My beloved wife SINDHU R PILLAI
& my dearest son RIDHIMAN
CONTENTS
1.
INTRODUCTION
1 - 4
2.
AIMS & OBJECTIVES
5
3.
REVIEW OF LITERATURE
6 – 27
4.
MATERIALS & METHODS
28 - 32
5.
RESULTS
33 - 45
6.
DISCUSSION
46 - 58
7.
SUMMARY & CONCLUSION
59 - 61
8.
BIBLIOGRAPHY
62 - 70
LIST OF FIGURES
Fig 1 Extracted human carious tooth with grounding of occlusal surface
Fig 2 a, b, c Chemomechanical preparation with CarisolvTM
and special instruments
Fig 3 a, b, c Laser preparation with Er, Cr: YSGG laser (BIOLASETM
WATERLASE).
Fig 4 Scanning electron microscope (JEOL JSM – 5600 LV)
Fig 5 a, b, c Photomicrographs of tooth prepared with TC bur (X800, X2000 and X5000)
Fig 6 a, b, c Photomicrographs of tooth prepared with CarisolvTM
(X800, X2000 and
X5000)
Fig 7 a, b, c Photomicrographs of tooth prepared with Er: Cr: YSGG laser (BIOLASETM
)
(X800, X2000 and X5000)
Fig 8 Clearfil SE Bond (Kuraray Medical Inc, Tokyo, Japan)
Fig 9 QTH curing unit (Confident Dental – India)
Fig 10 Prepared samples for microtensile bond strength test
Fig 11 Buehler fine sectioning machine (Buehler – USA)
Fig 12 Prepared 1mm² samples
Fig 13 Universal Testing Machine (BIOPULS – INSTRON, USA)
Fig 14 Loading at Universal Testing Machine
Fig 15 Stereomicroscope (Leica, Switzerland)
Fig 16a, b, c, d Photomicrographs of the fractured specimens
LIST OF TABLES
Table I The Micro Tensile Bond Strength (µTBS) values in MPa for the ‘residual caries-
excavated dentin’ according to the caries excavation technique tested (Mechanical (G1).
Table II The Micro Tensile Bond Strength (µTBS) values in MPa (SD) for the ‘residual caries-
excavated dentin’ according to the caries excavation technique tested (Carisolv (G2).
Table III The Micro Tensile Bond Strength (µTBS) values in MPa (SD) for the ‘residual caries-
excavated dentin’ according to the caries excavation technique tested (Laser (G3).
Table IV Mean Micro Tensile Bond Strength (µTBS) values comparison between Mechanical
(G1), Carisolv (G2) and the Laser (G3) groups using Kruskal-Wallis test.
Table V Comparison of Mean Micro Tensile Bond Strength (µTBS) values of two groups at a
time. Inter group comparison was carried out using Man-Whitney ‘U’ test.
Table VI Distribution of fracture types using different caries removing techniques
LIST OF BAR-DIAGRAMS
Chart I
Bar diagram of Mean Micro Tensile Bond Strength (µTBS) values of different caries
removing techniques.
Chart II
Bar diagram of Distribution of tensile strength in different groups.
Chart III
Bar diagram of Mean Micro Tensile Bond Strength (µTBS) values of different caries
removing techniques in MPa.
Chart IV
Bar diagram of Distribution of type of fractures as seen using the stereomicroscope
Introduction
Introduction
1
Dental caries is ubiquitous in all populations throughout the world and is the
key factor responsible for dental pain and tooth loss in populations throughout the
world43
. It is a process that may take place on any tooth surface in the oral cavity
where dental plaque is allowed to develop over a period of time31
.
The advent of “Adhesive Dentistry” has simplified the guidelines for
cavity preparation enormously. The design and extent of the current preparations are
basically defined by the extent and shape of the caries lesion, potentially slightly
extended by beveling the cavity margins in order to meet the modern concept of
minimally invasive dentistry41
. The superficial necrotic zone of caries-infected dentin
that harbors the core bacterial biomass should be excavated, leaving only ‘residual’
caries-affected dentin lining the cavity with sound enamel margins and dentin
adjacent to the enamel dentin junction, which enables the best peripheral seal to be
achieved with the current adhesive dentin bonding system3.
Caries dentin removal by mechanical means is non-specific nature of
excavation that may result in excessive loss of tissue thus affecting the prognosis of
the treated tooth33
. Other inherent fundamental drawbacks of these approaches are
unpleasantness to patient, need for local anesthesia, and potential effect to the pulp
due to heat and pressure30
.
Numerous caries excavation techniques have been introduced such as plastic
and ceramic burs, caries-disclosing dyes, enzymatic caries dissolving agents, sono-
Introduction
2
abrasion, air abrasion, laser ablation, Photo active disinfection (PAD), Ozone therapy,
Chemical Vapor Deposition (CVD)13
and caries infiltration of low viscosity resin.
They all aim to remove caries-infected tissue as selectively as possible, while being
minimally invasive through maximum preservation of caries-affected tissue. Minimal
invasive preparation techniques claim to achieve controlled removal of infected and
softened dentin while preserving healthy and hard dental tissues and causes minimal
discomfort for the patient63
.
Chemomechanical caries removal system removes only the softened carious
dentin whilst preserving the healthy tissue. The CarisolvTM
system, based on non-
specific proteolytic effect of sodium hypochlorite into which three naturally occurring
amino acids, namely glutamic acid, leucin and lysine are incorporated to form a gel.
The amino acids involve in the chlorination of partially degraded collagen in carious
dentin and initiates disruption of the altered collagen fibers in carious dentin21
.
Recently fluorescence controlled laser was introduced for the selective
removal of carious dentin20.
Laser light on absorption, is converted to heat, which
overheats the water to vaporize and cause micro explosions that can carry away
surrounding tooth fragment. Carbon dioxide, Neodymium and Erbium lasers were
used for the removal of carious dentin. Among these, erbium lasers are more
commonly used due to their ability to ablate enamel and dentin more effectively
because of their highly efficient absorption in both water and hydroxyapatite. Er: Cr:
YSGG laser would be favorable for caries removal because it does not damage dental
pulp tissue62
. More over dentin irradiated using erbium lasers provide microscopically
irregular surfaces and open dentinal tubules without smear layer5.
Introduction
3
During cavity preparation, the surface to bond will be covered by a smear
layer, which is not attached firmly to the tooth surface. The basic mechanism of
bonding to enamel and dentin is essentially an exchange process involving
replacement of minerals removed from the hard dental tissues by resin monomers that
upon setting becomes micro mechanically interlocked in the created porosities37
.
Bonding effectiveness may be impaired by thick smear layers. The recognition of the
role of smear layer in dentin bond strength highlights the importance of the cavity-
preparation method. Current adhesives should be able to dissolve the smear layer
without demineralizing the tooth surface too profoundly, thereby removing
hydroxyapatite at the interface38
.
Mild self-etch adhesives demineralize the dentin surface sufficiently to
provide micro-mechanical retention, while preserving hydroxyapatite within the
hybrid layer to enable additional chemical interaction53
. Phosphoric-acid etching of
dentin could nowadays be considered too aggressive for dentin, given all the
consequences related to exposure of the vulnerable collagen46
. Selective phosphoric-
acid etching of the enamel cavity margins is therefore highly recommended, followed
by applying a mild self-etch procedure to both the beforehand etched enamel and
(unetched) dentin26
.
Micro Tensile Bond Strength (µTBS) test method is one of the most
commonly used methodology to mechanically assess the strength of the resin-dentin
interface complex, since it has several advantages over traditional Tensile and shear-
bond test methodologies, as µTBS test is more versatile as multiple specimens can be
obtained from a single tooth53
. Micro Tensile Bond Strength (µTBS) test produces
Introduction
4
more adhesive failure than cohesive failure, permits testing of bond to irregular
surfaces as well very small areas68
.
This invitro investigation was undertaken to evaluate the morphological
changes, microtensile bond strength and interfacial characteristics of three different
caries excavation techniques namely mechanical (rotary), chemomechanical
(CarisolvTM
) and Er, Cr: YSGG Laser ((BIOLASETM
WATERLASE-MD) using a
two-step self-etch adhesive (Clearfil SE Bond).
Aim & Objectives
Aims & Objectives
5
The aim of this in vitro study was to evaluate the surface characteristics of caries-excavated
dentin and to compare the microtensile bond strength values of an adhesive, bonded to the
residual dentin after three different contemporary caries-excavation methods namely Mechanical
(handpiece and bur), Chemomechanical (CarisolvTM
) and Laser Ablation ( Er, Cr: YSGG laser -
BiolaseTM
) and to assess the failure mode after microtensile testing.
The objective of this in vitro study was to compare the efficacy of three caries-excavation
methods as Mechanical (handpiece and bur), Chemomechanical (CarisolvTM
) and Laser Ablation
( Er, Cr: YSGG laser - BiolaseTM
).
Review of Literature
Review of Literature
6
1. Tao L, Pashley DH and Boyd L (1988)67
in an in vitro study compared the effect of
different types of smear layers created on dentin, using burs or sandpaper. Clinically
dentin is shaped with burs prior to dentin bonding, and in laboratory studies bonding is
done on dentin prepared with sandpaper. Human third molars were selected and the
crowns were cemented to plastic cylinders. Teeth were sectioned along the long axis;
smear layer on the dentin and the enamel were created using either bur or sand paper. The
prepared surfaces were restored using composite resin and the shear bond test was
undertaken. There were no statistically significant differences among bond made to
enamel versus dentin smear layer. Dentin smear layer created with slow-speed burs and
silicon carbide sandpaper gave similar bond strengths. Microphotographs of the sides of
sheared surfaces indicated a cohesive failure of the smear layer. The authors suggested
that tooth surfaces prepared with rotary burs should be finished with silicon carbide
sandpaper.
2. Yip HK & Samaranayake LP (1998)77
reviewed the development of various caries
removal techniques and the evolutionary philosophies of cavity preparation which
promulgated over the last century. Caries excavation techniques such as hand
instruments, air-abrasion, air-polishing, atraumatic restorative technique, laser, enzymes,
ultrasonics and chemomechanical agents were discussed elaborately. The article
concluded that, the development of caries removal techniques in restorative dentistry is
progressing towards a more biological and conservative direction.
Review of Literature
7
3. Ericson D et al (1999)21
conducted a multicentre clinical study to test the efficacy and
safety of Carisolv for chemomechanical removal of dentin caries. The mean caries
removal time of 10.62min with Carisolv and 4.42min with rotary instruments were
recorded. The volume of Carisolv used was 0.45ml per person. 64% experienced no taste,
77% no smell, 29% experienced taste and 20% experienced smell. 52% of the patients
perceived that Carisolv method was faster than drilling, 25% estimated it to have lasted
the same and 11% estimated it to have lasted longer or much longer. In the study, the
complete caries removal was obtained in 106 of 107 cases and no short-term adverse
effects were observed.
4. Banerjee A, Kidd E A M & Watson T F (2000)11
in an invitro study evaluated five
alternate methods of carious dentin excavation using rotary instruments (no.3 round
carbon-steel bur in a slow speed handpiece), air-abrasion, sono-abrasion and carisolv
with hand excavation. Though bur excavation was found to be quickest and carisolv the
slowest, it was found that bur excavation tended to over prepare, sono-abrasion under-
prepare and carisolv removed adequate amount of caries
5. Banerjee A, Watson T F & Kidd E A M (2000)9 reviewed and discussed some of the
techniques available to excavate demineralized dentin clinically. These methods can be
classified as mechanical rotary, mechanical non-rotary, chemomechanical and laser
photo-ablation that include: dental handpieces/burs, manual excavators, air-abrasion, air-
polishing, ultrasonication, sono-abrasion, chemo-mechanical methods, lasers and
enzymes. The advantages and disadvantages of each technique were discussed.
Review of Literature
8
6. Banerjee A, Kidd EAM and Watson TF (2000)10
evaluated the time taken and quantity
of dentin removed using four caries-dentin excavation techniques namely bur, air-
abrasion, sono-abrasion and carisolv gel, and were compared to conventional hand
excavation. Eighty freshly extracted molars were assigned to four experimental groups of
20 teeth each, which were sectioned longitudinally through the occlusal lesion and pre-
excavation color microphotographs were obtained. Using natural autofluorescence of
carious dentin as an objective, carious dentin removal was assessed in each half of the
split-tooth sample. A confocal laser scanning microscope was used for this study. The
end-point of excavation was checked using a dental probe and the final color
microphotographs were taken. The study concluded that bur excavation was quickest but
resulted in over preparation. Carisolv excavation was the slowest, but removed adequate
amount of tissue. Sono-preparation tends to under prepare where as air- abrasion was
more comparable to hand excavation in both the time and the amount of dentin removed.
7. J. A. Beeley,1 H. K. Yip,2 and A. G. Stevenson (2000)8 reviewed Chemomechanical
caries removal techniques which involve the chemical softening of carious dentin
followed by removal with gentle excavation. The reagent involved is generated by
mixing amino acids with sodium hypochlorite; N-monochloroamino acids were formed
which selectively degrade demineralized collagen in carious dentin. The procedure
requires 5–15 minutes but avoids the painful removal of sound dentin thereby reducing
the need for local anaesthesia. It is well suited for the treatment of deciduous teeth, dental
phobics and medically compromised patients. The dentin surface formed is highly
irregular and well suited to bonding with composite resin or glass ionomer. When
Review of Literature
9
complete caries removal is achieved, the remaining dentin is sound and properly
mineralized. The system was originally marketed in the USA in the 1980’s as Caridex.
Large volumes of solution and a special applicator system were required. A modification
of this system known as Carisolv was introduced in 1998, that comes as a gel, which
requires volumes of 0.2–1.0 ml and is accompanied by specially designed instruments.
8. Splieth et al (2001)60
assessed the efficacy of carisolv and rotary instruments in
removing dentin caries. Caries removal with Carisolv or rotating round bur was
monitored by checking the hardness of the dentin with a dental explorer and was stopped
when a leather hard texture was reached or a sharp scratching sound was heard. After
embedding and sectioning, caries activity of the residual dentin was assessed using
methyl red dye. Carisolv treatment resulted in higher mean depth of caries-active dentin
(71 to 78µm) than conventional caries removal using round bur (19-51 µm). The authors
concluded that carisolv treatment leaves about 50 µm more carious dentin than caries
removal with round burs.
9. Bart Van Meerbeek et al (2003)38
stated that bonding to tooth tissue can be achieved
through etch & rinse, self-etch or glass ionomer approach. The basic bonding mechanism
to enamel and dentin of these three approaches were demonstrated by means of ultra-
morphological and chemical characterization of tooth-biomaterial interfacial interactions.
Bond-strength testing and measurement of marginal-sealing effectiveness were evaluated
upon their value and relevance in predicting clinical performance. Benefits and
drawbacks of etch and rinse versus self-etch was thoroughly evaluated. It was concluded
Review of Literature
10
that the conventional three-step etch & rinse adhesives performed more favorably and are
more reliable in long-term. Mild self-etch adhesives that bond through a combined
micromechanical and chemical interaction with tooth tissue were comparable with
conventional three-step systems in bonding performance.
10. Laura Ceballos (2003)14
evaluated the microtensile bond strength (µTBS) of total-etch
and self-etch adhesives to caries affected versus normal dentin. The bond strength results
were correlated with DIAGNOdent laser fluorescence and Knoop microhardness (KH)
measurements of the dentin substrates. Extracted carious human molars were ground to
expose flat surfaces with caries surrounded by normal dentin. Surfaces were bonded
respectively with Prime & Bond NT, Scotchbond 1, Clearfil SE Bond and Prompt L-
prop, and core build up was done using resin composite. 1 mm² samples were prepared
and the microtensile bond strength (µTBS) test was carried out. The results showed high
bond strength values for total-etch than self-etch samples. This study concluded that
total-etch adhesives produced higher bond strengths to normal and caries-affected dentin
than self-etching systems.
11. Sonada H, Banerjee A, Watson TF et al (2005)59
conducted an invitro study to
compare the micro-tensile bond strengths of two different adhesive systems, ABF
(Clearfil Protect Bond) and PBNT Prime & Bond NT, bonded to caries-affected dentin
retained after chemo-mechanical caries removal using Carisolv gel, and to the residual
dentin after conventional hand instrumentation. Matchstick-shaped samples were
prepared and the microtensile bond strength values were recorded. Scanning electron
Review of Literature
11
microscope (SEM) was used to ascertain the mode of failure at the restoration–dentin
interface. ABF group showed no difference in bond strengths between the controls and
Carisolv group but these values were significantly higher than those for the hand-
excavated samples. PBNT samples showed no significant differences in any of the three
test groups. They concluded that microtensile bond strengths of PBNT/composite
restorations to caries-affected dentin in clinical cavities were statistically comparable to
those to sound dentin. In the ABF/composite restored group (self-etched), the use of
conventional hand excavation appeared to weaken the bond strength to the remaining
caries-affected dentin. However, the use of Carisolv gel excavation did not compromise
bond strengths to caries-affected dentin.
12. William J Dunn et al (2005)18
evaluated dentin and enamel bond strength to resin
composite following high-speed rotary or Er: YAG laser preparation with total-etch
adhesive system. Extracted human molar teeth without visible caries or surface defects
were selected. Specimens having enamel and dentin were prepared with high-speed
rotary or Er: YAG laser and etched either with 37% phosphoric acid or not etched.
Composite was bonded to specimens and after thermocycling, samples were tested for
shear bond strength. There was significant difference between the etched and the not-
etched samples. Acid etched specimens had higher mean bond strengths, with rotary
preparation specimens having significantly higher mean strength versus laser prepared
specimens. Within the group comparison, acid etching was better than laser etch, and
laser etch was better than nonetch samples. Scanning electron microscopy of laser ablated
specimens demonstrated significant surface scaling and subsurface fissuring beyond
Review of Literature
12
normal resin penetration depth. This study concluded that adhesion to laser-ablated or
laser-etched dentin and enamel was inferior to that of conventional rotary preparation and
acid etching.
13. Inoue G et al (2006)28
conducted a study to examine the ultrastructure of both intact and
caries affected dentin-adhesive interface after artificial secondary caries formation, using
scanning electron microscope (SEM) and nanoindentation testing. Half of the prepared
specimens were bonded with Clearfil SE Bond and a resin composite for the
nanoindentation test. The other specimens were stored in a buffered demineralizing
solution for ninety minutes, and then observed under SEM. An acid-base resistant zone
(ABRZ) was observed beneath the hybrid layer. The ABRZ of caries-affected dentin was
thicker than that of normal dentin. It is suggested that, the monomer of Clearfil SE Bond
penetrated deeper into the residual dentin, which created a thicker hybrid layer.
14. Lennon AM et al (2006)35
conducted an in vitro study to compare the efficiency of four
caries excavation methods namely fluorescence aided caries excavation (FACE), caries
detector dye (CD), chemomechanical excavation (CS) and conventional excavation (CE);
and to compare the time taken to excavate and successfully remove bacterially infected
dentin. In the FACE group, the operating field was illuminated with violet light and the
operator observed the teeth through a 530nm yellow glass filter and removed the infected
dentin which was seen as orange-red fluorescing areas using a slow-speed bur. In the CS
group, Carisolv was applied to the cavity and allowed to act for 30 sec and the caries
excavation was performed using special hand instruments. In the CD group, caries was
Review of Literature
13
removed using caries detector and in the CE group, conventional excavation was carried
out using visual tactile criteria. The excavation time was recorded. Results showed that
excavation time was significantly shorter for FACE (3min 3sec), compared to CS (5min
8sec), CD (5min 26sec) and CE (4min 2sec). Microscopic examination revealed; bacteria
were significantly less in FACE samples when compared to CS and CD groups, but the
bacterial count were comparable with CE group. The study concluded that the excavation
with fluorescence aided caries excavation (FACE) is equal to conventional excavation
(CE) and superior to caries detector dye (CD) and chemomechanical excavation (CS).
15. Nelson R F A Silva et al (2006)56
conducted a study to test the hypothesis that there is a
reduction in bond strength when a micro tensile load is applied to adhesive junctions
prepared at 10, 20 and 30 degrees to the usual perpendicular interface. Extracted human
third molars were selected, occlusal enamel was removed, self-etch adhesives were
applied, and composite resin restorations were done. The teeth were sectioned at 10, 20
and 30 degrees to the bond interface and the control group sectioned at 0 degree bond
angle. The bond strength resulted in lower values as the angle on the interface increases.
They proved that there is a reduction in bond strength when a micro tensile load is
applied to different angles.
16. Bor-Shiunn Lee et al (2007)34
investigated the tensile bond strength of composite resin
to Er:Cr:YSGG laser irradiated dentin. The study analyzed the resin-dentin interface
among bur-cut/acid etched, Er:Cr:YSGG laser ablated / acid etched and Er:Cr:YSGG
laser ablated dentin. Crown dentin disks were prepared from extracted human third
Review of Literature
14
molars and SEM analysis was undertaken. SEM analysis showed all the three groups
were free of dentin debris and smear layer. The peritubular dentin was found protruded
from the surrounding intertubular dentin after laser irradiation. No statistical difference
was found in the tensile bond strength of the first and second groups, whereas the third
group values were lower. It was proved that Er:Cr:YSGG laser ablation adversely affects
the adhesion of resin to dentin and the acid etching followed by laser ablation method
was able to increase the tensile bond strength.
17. Omar H et al (2007)44
compared the ability of two self-etch adhesives and a
conventional three-step adhesive to bond composite to both intact and caries-affected
dentin with and without thermocycling. Thirty extracted human teeth with occlusal caries
were randomly assigned to three groups according to the adhesives used: Scotchbond
multi-purpose, Clearfil SE bond and Xeno IV. The occlusal surfaces were ground;
adhesives were applied, and restored using a composite material. 1mm² thickness sections
were obtained using a micro-slicer. Half the specimens were subjected to thermocycling
prior to testing. All the specimens were subjected to microtensile bond strength (µTBS)
testing. Mean µTBS values were: Scotchbond multi-purpose – 22.19 and 15.7MPa;
Clearfil SE bond – 24.25 and 22.3MPa and for Xeno IV – 21.43 and 18.3 MPa
respectively of initial and after thermocycling procedure. This study concluded that two-
step self-etch adhesives produced comparable µTBS values to both sound and residual
dentin and thermocycling significantly reduces µTBS values of specimens.
Review of Literature
15
18. Zuhal Kirzioglu et al (2007)30
studied the clinical efficacy of Carisolv™ system and
hand excavation methods in the removal of occlusal dentin caries of primary molar teeth.
Both Carisolv system and hand excavation methods were applied for the removal of
caries on different teeth and were restored. The clinical follow-up was made every 3
months within a year. At the end of 1 year, outcome of the Carisolv and the hand
excavation groups in terms of marginal adaptation were found to be comparable.
19. Arlene Tachibana et al (2008)62
evaluated the influence of the dental surfaces obtained
after the use of different caries removal techniques on bonding of a self-etching system.
Extracted, carious, human molar teeth were selected and the caries lesions were removed
by mechanical, Carisolv and Er: Cr: YSGG laser. The prepared teeth were submitted to a
bonding system followed by construction of resin based composite crown. Hour glass
shaped samples were obtained and submitted to the micro-tensile bond strengths (µTBS).
The highest bond strengths were observed with the sound dentin treated with burs and
Carisolv. The samples of the sound dentin presented higher bond strength than samples of
the caries affected dentin. Among the caries removal methods tested, the Er: Cr: YSGG
laser ablation was the poorest in providing a substrate for bonding with the self-etch
system used.
20. J Eberhard et al (2008)20
evaluated the extension of cavities prepared conventionally by
bur or by Er:YAG laser. Extracted human teeth with dentin caries were bisected through
the caries lesion and was treated by Er:YAG laser and by mechanical caries removal. The
specimens were subjected to histological staining and a quantitative evaluation of the
Review of Literature
16
cavity area (mm²) by computer-assisted alignment. 23 out of 29 cavities were smaller
after caries removal with laser compared to the bur. This invitro study indicates that
caries removal by Er:YAG laser resulted in less dentin than that of caries removal by a
rotary instruments.
21. Marcio V Cardoso, BV Meerbeek et al (2008)13
studied the influence of dentin cavity
surface finishing on microtensile bond strength (µTBS) of adhesives. Tooth preparation
techniques such as chemical vapor deposition (CVD), laser irradiation and conventional
mechanical methods were followed. The morphological characteristics of dentin prepared
with these techniques were also evaluated. Scanning electron microscopic examination
revealed different morphological features on the dentin surface depending on the caries
excavation method. Microtensile bond strength (µTBS) evaluation resulted in lower
values for chemical vapor deposition (CVD) and Er, Cr: YSGG laser irradiation
techniques than to conventional rotary prepared dentin. This study concluded that
alternate cavity preparation techniques negatively influenced the microtensile bond
strength (µTBS) of the adhesives.
22. Michal Staninec et al (2009)57
conducted a study to prove the hypothesis that, laser
initiated cracks resulted in lower bending strength of dentin tested. Dentin beam
specimens were prepared from human molar teeth and divided into three groups as
control, wet & dry and was irradiated with either Er:YAG or Er, Cr: YSGG laser. The
bending strength of each beam was tested in a universal testing machine. They concluded
Review of Literature
17
that, Er:YSGG laser without water caused cracks in the surface that significantly
decreases the bending strength of dentin.
23. Mine A, De Munk, B Van Meerbeek et al (2009)39
in an invitro study, evaluated the
bonding effectiveness of two new self-etch adhesives (Adper Easy Bond and Adper
ScotchBond SE, 3M ESPE) to enamel and dentin and to characterize the interfacial ultra-
structure at enamel and dentin using transmission electron microscope (TEM). The
adhesives were applied to coronal human enamel and dentin surfaces and core build up
was done with micro-hybrid resin composite Z100 (3M ESPE). The ‘gold-standard’ two-
step self-etch adhesive Clearfil SE Bond (Kuraray) served as control. Results showed that
the µTBS of the two self-etch adhesives to enamel was significantly lower than that of
the control. To dentin, the µTBS of Adper Easy Bond was significantly lower than that of
Adper ScotchBond SE and the control. TEM showed a tight interface to enamel for all
the three self-etch adhesives but observed spot and cluster like nano-leakage in dentin.
This study concluded that the new two self-etch adhesives revealed a tight interaction at
both enamel and dentin, their bond strength to both tooth tissues was generally lower than
that of the adhesive Clearfil SE Bond (control).
24. Maria Paula Gandolfi Paranhos et al (2009)45
evaluated the Microtensile Bond
Strength (µTBS) of two adhesive systems to carious or normal dentin. Adhesives used
were Adper single bond plus and Clearfil SE bond. The prepared teeth were divided into
twelve groups: group one to six were submitted to pH cycling for artificial caries and
group seven to twelve with normal dentin. Dentin surfaces were treated with either, laser
Nd: YAG irradiation for one minute, laser Nd: YAG irradiation associated with fluoride
Review of Literature
18
gel, or no treatment. Adhesive systems were applied and a composite resin block was
built up. Teeth were sectioned serially to get a 0.8 mm² samples and were submitted for
Microtensile Bond Strength (µTBS) testing. The study results showed the highest mean
bond strength was obtained in groups of normal dentin treated with Clearfil SE Bond
(40.65MPa). Laser irradiation with fluoride decreased the bond strength value of both the
adhesives. Presence of carious dentin significantly reduces the bond strength value of
both adhesive systems. This study concluded that after excavating the caries affected
dentin, the use of erbium laser followed by a self-etching adhesive system is the best
clinical choice when considering the bond strength; compared with the total-etching
based system.
25. Nathalie Brulat et al (2009)5 conducted a study to compare the shear bond strength of
composite resin bonded to Er:YAG laser or bur-prepared dentin surfaces using three self-
etching adhesive systems. Occlusal surface of the extracted human third molars were
grounded to expose dentin and the dentin was prepared using either a carbide bur or Er:
YAG laser. Three different self-etching adhesive systems were applied: iBondTM
, Xeno
IIITM
and Clearfil SE BondTM
. Samples were restored with composite resin and shear
bond tests were carried out. It was found that Xeno III showed no difference in shear
bond strength to bur and laser. Shear bond strength values of iBond and Clearfil SE Bond
to laser prepared dentin were lower than preparation with burs.
26. Avijit Banerjee et al (2010)3 evaluated the amount of residual dentin retained after using
three excavation techniques; namely hand, Carisolv and an enzymatic gel (SFC-V). The
Review of Literature
19
microtensile bond strengths (µTBS) to residual dentin were assessed using an offset
micro-tensile testing device, comparing etch-rinse versus self-etching adhesives. They
found that surface roughness and smear layer formation is more with hand excavation
when compared with chemomechanical system. µTBS value from etch-rinse sample
showed statistical difference between hand and Carisolv groups. Self-etch samples
showed significant difference between hand and Carisolv as well hand and Biosolv. This
study concluded that different caries excavation techniques have an effect on the
microtensile bond strength (µTBS) of adhesives to dentin.
27. Snejana et al (2010)63
conducted an in vitro investigation aimed to study the
morphological changes in the hard dental tissues after using different caries removal
methods by means of scanning electron microscope. Caries excavations in freshly
extracted human teeth were done using Er: YAG laser, Carisolv gel and mechanical
preparation. Results showed considerable differences in surface characteristics of dental
tissues. This study concluded that, dental tissue prepared using steel and diamond burs
showed surfaces covered with a thick smear layer that may be relevant to the subsequent
bonding of adhesive restorative material. Carisolv samples showed a rough retentive
surface, where as Er: YAG laser prepared teeth surface remained without smear layer and
clearly exposed dentinal tubule orifices.
28. Sarr M et al (2010)53
evaluated mechanically and ultra-morphologically eleven different
adhesive systems bonded to dentin. The resultant interfacial ultra-structure in dentin was
characterized by Transmission electron microscope (TEM). The Micro Tensile Bond
Review of Literature
20
Strength (µTBS) of eleven adhesives, including two three-step etch & rinse, three two-
step etch & rinse, two two-step self-etch and four one-step self-etch adhesives to dentin,
were measured. The microtensile bond strength (µTBS) varied from 11.1 to 63.6 MPa;
the highest bond strengths were obtained with the three-step etch & rinse adhesives and
the lowest with one-step self-etch adhesives. Mild self-etch adhesives demineralized the
dentin surface sufficiently to provide micro-mechanical retention, while preserving
hydroxyapatite within the hybrid layer to enable additional chemical interaction. This
study concluded that adhesives with simplified application procedure still underperform
when compared to conventional three-step adhesives. “Mild” two-step self-etch adhesives
provide additional chemical bonding, which appears to be the most acceptable bonding
effectiveness with a simplified application protocol.
29. J D Scholtanus et al (2010)55
conducted a study to determine the microtensile bond
strength of three different adhesive systems to caries-affected dentin. Extracted human
molars with primary carious lesions were ground flat to expose dentin and the caries-
infected dentin was excavated with the help of caries detector dye. Three adhesives
namely Adper Scotchbond(2 step etch & rinse adhesive), Clearfil S Bond(one step self-
etch) and Clearfil SE Bond(two step self-etch) were used and a resin composite buildup
was done. Specimens were sectioned into 1.0 x 1.0mm bars and µTBS was measured in a
UTM at a crosshead speed of 1 mm/min. All the three adhesives showed comparable
result to sound dentin whereas Scotchbond and Clearfil S bond showed significantly
lower bond strength value to caries-affected dentin. They concluded that two step self-
Review of Literature
21
etch adhesive Clearfil SE Bond showed better result than etch & rinse and one step self-
etch adhesive.
30. Ahmed A EL Zohairy (2010)78
in an in vitro study evaluated the efficacy of the
microtensile bond strength (µTBS) and micro shear bond (µSBS) in ranking dental
adhesives according to bond strength tests. Forty four caries-free extracted human molars
were divided into two groups. After occlusal grinding, different types of adhesive
systems were applied and the resin core build was done. Samples were stored in water for
24h and the microtensile bond strength (µTBS) tests were carried out. The resulted test
ranks of adhesives were analyzed. The authors concluded that, microtensile bond strength
(µTBS) tests appeared to be more accurate in differentiating among adhesives.
31. Steve Armstrong et al (2010)1 reviewed the mechanics, geometry, load application and
other testing parameters of ‘micro’ tensile and shear tests, and to outline their advantages
and limitations. The testing of multiple specimens from a single tooth conserves
specimens and allows research designs not possible using conventional ‘macro’ methods.
Specimen fabrication, gripping and load application methods, in addition to material
properties of various components comprising of the resin-tooth adhesive bond, will
influence the stress distribution and consequently the nominal bond strength and failure
mode. For the foreseeable future, both ‘micro’ and ‘macro’ bond strength tests will be an
important tool for improving resin-tooth adhesion to increase the service life of dental
resin-based composite restorations.
Review of Literature
22
32. Ulrich Salz & Thorsten Bock (2010)68
reviewed testing adhesion of direct restoratives
to enamel and dentin. They aimed to survey available methods of adhesion testing and
influential parameters affecting experimental outcome. The longevity and success of
modern dental restorations very often relies on potent dental adhesives to provide durable
bonds between the dental hard substance and the restorative composite. This review
provided a current overview of bond strength testing methods and their applicability to
the characterization of dental adhesives. Preparatory parameters possess tremendous
influence on the validity and scope of obtained data. A subtle variation in sample
preparation may severely impact test results. This article discusses the most influential
parameters, such as substrate nature, age, health status, storage, pre-treatment and sample
preparation.
33. A clinical study conducted by Yazici AR et al (2010)75
evaluated the two year clinical
performance of two minimally invasive cavity preparation techniques using bur and laser;
in class I occlusal resin composite restorations. Twenty seven patients, each having at
least one pair of occlusal caries, were selected for the study. For each patient, one cavity
was prepared with a diamond bur and the other was prepared with Er, Cr: YSGG laser
and the cavities were restored with a nano filled flowable resin composite using an etch
& rinse adhesive. The restorations were evaluated at baseline, 6, 12, 18 and 24 months
according to modified Cvar/Ryge criteria (alpha, Bravo & Charlie). Patient recall rate
was 100%. The retention rates of restorations at 24 months were 98.1% for bur and 100%
for the laser-prepared group. After 24 months, 5.6% of bur prepared and 7.4% of laser-
prepared were rated bravo in marginal discoloration. Marginal adaptation for bur
Review of Literature
23
prepared was 9.3% and laser-prepared was 13% in 24 months and was rated bravo. This
study concluded that there were no significant difference between the two cavity
preparation techniques and both the techniques performed equally, with excellent
outcomes after a 24-month period.
34. Aline de A. Neves et al (2011)41
reviewed current concepts and techniques for caries
excavation and adhesion to residual dentin. New caries excavation techniques have been
introduced, such as the use of plastic and ceramic burs, improved caries disclosing dyes,
enzymatic caries-dissolving agents, caries-selective sono/air abrasion and laser ablation.
They all aim to remove or help remove caries-infected tissues as selectively as possible,
while being minimally invasive through maximum preservation of caries affected tissue.
Each technique entails a specific caries-removal endpoint and produces residual dentin
substrates of different natures and thus different receptiveness for adhesive procedures.
They summarized that bond strength of caries-affected dentin is lower than that at sound
dentin interfaces and also no difference in bond strength between etch and rinse as well
self-etch approach. They concluded that the design and extent of the current preparations
are basically defined by the extent and shape of the caries lesion, potentially slightly
extended by beveling the cavity margins in order to meet the modern concept of
minimally invasive dentistry.
35. Aline de A. Neves et al (2011)42
determined the caries-removal effectiveness (CRE) and
minimal-invasiveness potential (MIP) of contemporary caries-removal techniques.
Carious molars were scanned using micro-CT, after which dentin caries was removed by
Review of Literature
24
9 contemporary caries-removal techniques. The micro-CT was repeated and CRE was
determined on basis of the relative volume of residual caries and the mineral density
(MD) at the cavity floor. MIP was determined by measuring the cavity size relative to the
initial size of the caries lesion. The study concluded that, Er:YAG-laser aided by Laser
Induced Fluorescence resulted in non-selective caries removal. Rotary/ oscillating caries
removal may lead to over-excavation. The risk for over-excavation was reduced with
carisolv group and the chemo-mechanical methods were most selective in removing
caries, whilst preserving sound tissue.
36. Aline de A. Neves et al (2011)40
evaluated the micro-tensile bond strength (µTBS) and
interfacial characteristics of adhesive–dentin bonds produced after caries-removal with
contemporary techniques. Carious molars were sectioned at the base of the fissure, and
after caries-excavation, a composite was bonded using a 2-step self-etch adhesive.
Results showed µTBS to residual caries-excavated dentin was lower than to sound dentin.
They found that various caries-removing techniques had different effects on the µTBS.
Er:YAG-laser excavation guided by a LIF-feedback system resulted in a statistically
significantly lower µTBS as compared to sound dentin and produced a thick layer of
demineralized/ unprotected collagen at the interface with the adhesive. Carisolv resulted
in the highest µTBS to ‘residual caries-excavated’ dentin, followed by the use of a
tungsten-carbide-bur.
37. Bart Van Meerbeek et al (2011)37
reviewed the current self-etch adhesives. After
presenting the general characteristics of self-etch adhesives, the major shortcomings of
Review of Literature
25
the most simple-to-use one-step (self-etch) adhesives are addressed. Special attention was
devoted to the AD-concept and the benefit of chemical interfacial interaction with regard
to bond durability. Finally, issues like the potential interference of surface smear and the
more challenging bond to enamel for ‘mild’ self-etch adhesives are discussed. They
concluded that selective phosphoric-acid etching of the enamel cavity margins is highly
recommended, followed by applying a mild self-etch procedure to both the beforehand
etched enamel and (unetched) dentin. Phosphoric-acid etching of dentin is believed to be
too aggressive for dentin, given all the consequences related to exposure of the vulnerable
collagen.
38. David H Pashley et al (2011)46
conducted a study aimed to explore the therapeutic
opportunities of each step of three-step etch & rinse adhesives. Etch & rinse adhesive
systems are the oldest of the multi-generation evolution of resin bonding systems. The
three-step version involve acid-etching, priming and application of a separate adhesive.
Acid-etching, using 32-37% phosphoric acid (pH 0.1 to 0.4) etches enamel and dentin as
well destroys residual bacteria due to its low pH. Primers are water and HEMA-rich
solutions that ensure complete expansion of the collagen fibril meshwork and wet the
collagen with hydrophilic monomers. In the future, ethanol or water-free solvents may
serve as dehydrating primers which can increase the resin-dentin bond. Three-step Etch
& rinse adhesives produce higher resin-dentin bonds that are more durable than one and
two step adhesives. Incorporation of protease inhibitors in etchants and/or cross-linking
agents in primers may increase the durability of resin-dentin bonds.
Review of Literature
26
39. Arlene Tachibana et al (2011)61
conducted a study to detect the influence of blood
contamination on bond strength and the efficacy of cleaning agents on contaminated
enamel and dentin during the adhesion process. One hundred and four extracted human
molars were sectioned into halves along the long axis. Heparinized and fresh blood were
obtained from the same donor, applied and dried to maintain a layer of dry blood on the
top of samples. The cleansing agents used were hydrogen peroxide, anionic detergent and
antiseptic solution. Adhesive was applied and composite resin core build up was done.
Bond strength result showed higher values for groups without contamination than of
contaminated group. There were no statistically significant difference among cleansing
agents and they were as effective as water stream in counter acting the effect of blood
contamination. The study concluded that a water stream is sufficient enough to remove
blood contamination from dental tissues, before the application of adhesive agent.
40. Cecilia Goracci et al (2011)26
conducted an in vitro study to determine the influence of
mechanical properties of resin based composites on the microtensile bond strength to
dentin. Three dimensional models of the microtensile beam were created for finite
element analysis. The tensile strength, flexural strength, tensile elastic modulus, shear
elastic modulus, poissons ratio, Vickers hardness and contraction stress of the resin
composite were assessed. The study found that adhesives did not significantly influence
the microtensile bond strength; but the properties of the resin composite were a
significant factor. The authors concluded that when comparing the bonding potential of
several adhesives with the microtensile technique, the same resin composite should be
used in all the experimental groups for building up of the coronal portion.
Review of Literature
27
41. Anne-Maria Vuorinen et al (2011)70
in their study evaluated the effect of water storage
on the microtensile bond strength of composite resin to dentin. Experimental primers
were prepared and applied on to dentin and the resin build up was done. Teeth were
sectioned after 48h, 6 months and 12 months of water storage and tested for microtensile
bond strength. The authors found that long-term water storage decreased the microtensile
bond strength. The study concluded that one-year water storage lowered the dentin bond
strength.
42. Heintze et al (2011)27
evaluated whether the correlation between in vitro bond strength
data and estimated clinical retention rates of cervical restorations depends on pooled data
obtained from multicentre studies or single-test data. Pooled mean data for six dentin
adhesive systems ( Adper Prompt L-Pop, Clearfil SE, Optibond FL, Prime & Bond NT,
Single Bond and Scotchbond Multi-Purpose) and four laboratory methods (macroshear,
microshear, macrotensile and micro-tensile bond strength (µTBS) test) were correlated to
estimated pooled two-year retention rates of Class V restorations using the same adhesive
system. The results of the regression analysis of the pooled data revealed that only the
macrotensile and the micro-tensile bond strength (µTBS) tests correlated well with the
clinical findings. This study concluded that the tensile bond strength tests showed an
adequate correlation rate with the clinical data’s.
Materials & Methods
Materials and Methods
28
1. Sampling procedure.
Forty five extracted cavitated human permanent molars were used in this study. Teeth
with caries lesions limited to the occlusal surface and extending at least half the distance from
the enamel-dentin junction to the pulp chamber were included. Only central dentin portion that is
located directly above the pulp was used in order to minimize any regional variation between the
periphery and the central dentin substrate25
. These characters were determined by visual and
radiographical inspections.
2. Sample preparation.
The occlusal enamel was removed perpendicular to the long axis of the tooth to expose
the dentin surface completely, which contain a central zone of caries-infected dentin surrounded
by sound dentin, by using a low speed diamond saw ( IsoMetTM
Beuhler – USA) under running
water. The exposed flat surfaces were manually polished with a wet 600-grit silicon carbide
paper (Fig. 1).
3. Experimental groups.
The teeth were randomly divided into three groups (n=15), according to the caries
removal methods, as follows:
Materials and Methods
29
Group I : Mechanical rotary preparation by round tungsten carbide burs/air turbine (NSK
PANA-MAX, Japan).
Group II : Chemo-mechanical preparation with carisolvTM
gel (MediTeam Sweden).
Group III : Laser preparation by Er. Cr: YSGG Laser (BIOLASETM
WATERLASE-MD,
USA).
Preparations were made strictly according to the manufacturer’s instructions. For all the
caries-excavation methods, the infected dentin was removed till clinically detectable hardness of
the dentin was felt.
3.1. Mechanical rotary excavation (GI).
Round tungsten carbide burs No.2 and No.3 were used in an airotor handpiece ( NSK
PANA-MAX, Japan) to remove carious dentin. The full extent of carious dentin including a
periphery of sound dentin was excavated. A spoon shaped hand excavator was used for the
removal of softened caries. Infected-dentin removal was terminated when the soft dentin has
been removed from the cavity surface as clinically detected by using a probe to check the
firmness.
Materials and Methods
30
3. 2 Chemomechanical preparation with CarisolvTM
(G2).
Caries excavation using carisolvTM
(MediTeam Sweden) was done as per manufacturer’s
instructions. Pre-mixed gel was introduced into the carious dentin for 40sec. The mixture was
agitated against the dentin using a metal mace tip instrument. Once the gel has become cloudy
with a muddy consistency, it was rinsed away and a second fresh mix of gel was applied and
further agitated. Excavation was deemed complete, when the gel failed to become cloudy and the
cavity surface was checked with dental probe for hardness (Fig. 2 a, b, c).
3.3 Er, Cr: YSGG Laser preparation (G3).
Er: Cr: YSGG laser (BIOLASETM
WATERLASE-MD) with a power output of 3.5W,
pulse duration of 140µs, frequency of 20Hz, energy density of 175mj, and a Turbo handpiece
with MX700 micron focusing lens was used. The non-contact mode was used at a distance of 3-
5mm. Cooling was obtained by setting the equipment at the 80% air and 70% water level. (Fig: 3
a, b). caries-excavation was terminated once the firmness of the prepared surface was confirmed
with a probe.
4. Scanning Electron Microscopy (SEM).
After caries excavation, the samples were randomly selected from each group for
scanning electron microscopy (JEOL JSM – 5600 LV) (Fig. 4). These specimens were immersed
in 4% glutaraldehyde solution for 1hr at room temperature. Then they were rinsed in distilled
Materials and Methods
31
water. The samples were then placed in cold buffer solution of sodium cacodylate for 90 minutes
to fix the organic matter. Specimens were then dehydrated in ascending grades of ethanol (30,
50, 70, 80, 95 and 100% for one hour in each series) and then dried in a Critical Point Drier
based desiccator. The dried specimens were mounted on a metal stand and gold sputter coated
(200 – 250 nm) by cathode atomization under vacuum. SEM images of the caries excavated
surfaces were obtained. For each specimen, three microphotographs of different magnification
(X800, X2000 and X5000) were made. Each SEM photomicrographs were evaluated, described
and the morphological findings were compared.
5. Sample restoration.
After caries excavation, the samples were thoroughly washed with air-spray and gently
air-dried. Then the surfaces were bonded with a two-step self-etch adhesive; Clearfil SE Bond
(Kuraray Medical Inc, Tokyo, Japan) with the aid of a QTH curing unit (Confident Dental India)
according to the manufacturer’s instruction. The bonded surface was restored using a nanohybrid
resin composite (Filtek – Z 250, 3M ESPE. USA). Composite resin was constructed with 1.5mm
increments to reach a height of approximately 6mm. Each increment was light cured for 20
seconds. The samples were then immersed in water at 37°C for 24h.
6. Micro-tensile bond strength test (µTBS).
The specimens were serially sectioned into multiple slices using a low speed diamond
saw ( IsoMetTM
Beuhler – USA) under water cooling. Each slab was ground with an ultra-fine
Materials and Methods
32
diamond disc in a ‘high-speed precision saw’ (Buehler – USA) under water spray to create a
stick (rod) shaped configuration with a cross-sectional area of approximately 1mm2. The final
width and thickness of the samples prepared were 1mm X 1mm. The prepared rods comprised of
residual dentin, adhesive and composite resin. Silicon sleeves were attached to both the end of
the samples. Specimens were mounted on a pneumatic jig. The tests were taken out using a
universal testing machine (BIOPULS – INSTRON, USA), at a tension of 0.5mm/min until it
failed. Bond strengths were calculated by dividing the load at failure by the cross-sectional
bonding area. The pre-testing failures were considered as 0 MPa.
7. Failure modes analysis.
Dentin-resin samples after microtensile bond strength test were randomly selected and
the interfaces were analyzed using a digital stereoscopic microscope (Leica, Switzerland) at
X100 magnification to evaluate the fracture mode. To determine the fracture mode, the
classification used by Trajtenberg et al64
was adopted:
Type 1 - Complete adhesive failure between resin and dentin
Type 2 - Partial adhesive failure between resin and dentin and partial cohesive failure
within resin
Type 3 - Complete cohesive failure within resin
Type 4 - Partial cohesive failure within dentin.
Fig 1: Extracted carious tooth with Fig 2a: CarisolvTM
grounded occlusal surface
Fig 2b: Special instruments for CarisolvTM
Fig 2c: Chemomechanical preparation with
CarisolvTM
Fig 3a: Laser preparation with Fig 3b: BIOLASETM
WATERLASE
Er, Cr: YSGG laser
Fig 4: Scanning electron microscope (JEOL JSM – 5600 LV)
Fig: 5a
Fig. 5 a b c: SEM images of tooth prepared with TC bur (X800, X2000 and X5000)
Fig: 5b
Fig: 5c
Fig. 5 a b c: SEM images of tooth prepared with TC bur (X800, X2000 and X5000)
Fig. 5 a b c: SEM images of tooth prepared with TC bur (X800, X2000 and X5000)
Fig: 6a
Fig. 6 a b c: SEM images of tooth prepared with Carisolv
Fig: 6a Fig: 6b
Fig: 6c
Fig. 6 a b c: SEM images of tooth prepared with CarisolvTM
(X800, X2000 and X5000
(X800, X2000 and X5000
Fig: 7a
Fig. 7 a b c: SEM images of tooth prepared with Er: Cr: YSGG laser (BIOLASE
Fig: 7b
Fig: 7c
Fig. 7 a b c: SEM images of tooth prepared with Er: Cr: YSGG laser (BIOLASE
(X800, X2000 and X5000)
Fig. 7 a b c: SEM images of tooth prepared with Er: Cr: YSGG laser (BIOLASETM
)
Fig 8: Clearfil SE Bond Fig. 9. QTH curing unit
Fig 10a, b: prepared samples for microtesile bond strength test
Fig 11a, b: Buehler fine sectioning machine
Fig 12a, b: Prepared 1mm² samples
Fig 13: Universal Testing Machine Fig 14: Loading at UTM
Fig 15: Stereomicroscope
Fig 16a, b, c, d: Photomicrographs of the fractured specimens
Results
Results
33
Data from the bond strength testing were statistically compared using non-parametric
Kruskal-Wallis one-way ANOVA. Mann-Whitney ‘U’ test was used to identify
significant difference between the different caries removing techniques. A significant
level of 5% was employed for all analysis (p = ≤ 0.05)
1. Scanning Electron Microscopy.
Morphological analysis of the cavity floor prepared with tungsten carbide bur, air
turbine and water cooling appeared rougher with irregular particles scattered throughout
the surface area. A well defined smear layer was detected. Missing of smear layer was
observed in some areas (Fig 5a b c). In the areas of water turbulence; there were patent
dentinal tubule orifices, but without a clear outline of both tubule lumens and peri-
tubular as well inter- tubular dentin.
CarisolvTM
produced a reduced and more inconsistent smear layer with large areas of
open tubular orifices. The dentinal tubule orifices were visible and there were almost no
smear layer (Fig 6a b c). Preparing the organic matrix using chemo-mechanical
preparation with CarisolvTM
protected the mineralized dental tissue and at the same time
resulted in rough appearance of the treated surface.
Results
34
The dentin surfaces that received Er, Cr: YSGG laser irradiation showed a scaly,
irregular and rugged appearance (Fig 7a b c). There were rough and irregular surfaces
with no smear layer and dentinal tubule orifices were open without widening. The peri
tubular dentin protruded slightly from the surrounding intertubular dentin.
Results
35
Table I – The Micro Tensile Bond Strength (µTBS) values in MPa (SD) for the ‘residual
caries-excavated dentin’ according to the caries excavation technique tested (Mechanical
(G1).
Frequency Percent
Valid
Percent
Cumulative
Percent
Valid
0.93513 1 6.7 6.7 6.7
2.73848 1 6.7 6.7 13.3
2.92532 1 6.7 6.7 20
3.51224 1 6.7 6.7 26.7
3.95952 1 6.7 6.7 33.3
4.43319 1 6.7 6.7 40
6.279 1 6.7 6.7 46.7
6.33294 1 6.7 6.7 53.3
6.93334 1 6.7 6.7 60
7.32774 1 6.7 6.7 66.7
8.306 1 6.7 6.7 73.3
8.60986 1 6.7 6.7 80
9.77605 1 6.7 6.7 86.7
9.8309 1 6.7 6.7 93.3
12.3649 1 6.7 6.7 100
Total 15 100 100
Results
36
Table II – The Micro Tensile Bond Strength (µTBS) values in MPa (SD) for the
‘residual caries-excavated dentin’ according to the caries excavation technique tested
Carisolv(G2) group.
Frequency Percent Valid Percent Cumulative Percent
Valid
1.6044 1 6.7 6.7 6.7
3.113 1 6.7 6.7 13.3
3.28716 1 6.7 6.7 20
3.3461 1 6.7 6.7 26.7
4.07832 1 6.7 6.7 33.3
4.19492 1 6.7 6.7 40
4.3214 1 6.7 6.7 46.7
4.34274 1 6.7 6.7 53.3
5.80472 1 6.7 6.7 60
6.368 1 6.7 6.7 66.7
8.53832 1 6.7 6.7 73.3
8.6999 1 6.7 6.7 80
8.7961 1 6.7 6.7 86.7
9.64364 1 6.7 6.7 93.3
19.3857 1 6.7 6.7 100
Total 15 100 100
Results
37
Table III – The Micro Tensile Bond Strength (µTBS) values in MPa (SD) for the
‘residual caries-excavated dentin’ according to the caries excavation technique tested:
Laser (G3) group.
Frequency Percent
Valid
Percent
Cumulative Percent
Valid
1.01609 1 6.7 6.7 6.7
1.15049 1 6.7 6.7 13.3
1.41134 1 6.7 6.7 20
1.47562 1 6.7 6.7 26.7
1.91447 1 6.7 6.7 33.3
3.34997 1 6.7 6.7 40
3.61219 1 6.7 6.7 46.7
4.17611 1 6.7 6.7 53.3
4.4698 1 6.7 6.7 60
5.42358 1 6.7 6.7 66.7
5.60265 1 6.7 6.7 73.3
7.03815 1 6.7 6.7 80
7.80213 1 6.7 6.7 86.7
8.39711 1 6.7 6.7 93.3
10.2984 1 6.7 6.7 100
Total 15 100 100
Results
38
Table IV – Mean Micro Tensile Bond Strength (µTBS) values in MPa (SD) for the
‘residual caries-excavated dentin’ according to the caries excavation techniques tested
(Mechanical (G1), Carisolv(G2) and the Laser(G3) groups).
Caries
Removing
Agent
No of
Samples
Mean
Tensile
strength
Median
Tensile
strength
Standard
Deviation
Test of
Significance
Group – I
Mechanical
15 6.2843073 6.3329400 3.1818350 Kruskal-wallis
test
Chi-square
2.424
p-value 0.298
Group – II
Carisolv
15 6.3682973 4.4327400 4.3647770
Group – III
Laser
15 4.4758747 4.1761100 2.9174428
Chart I -
0
1
2
3
4
5
6
7
1 2 3
laser
carisolv
mechanical
Results
39
Table V – Comparison of Mean Micro Tensile Bond Strength (µTBS) values of two
groups at a time.
V A – Comparison between Mechanical (G1) and Carisolv (G2).
Caries
Removing
Agent
No of
Samples
Mean
Tensile
strength
Median
Tensile
strength
Standard
Deviation
Test of
Significance
Group – I
Mechanical
15 6.2843073 6.3329400 3.1818350
Man-
Whitney
‘U’=
106.000
p-value =
0.787
Group – II
Carisolv
15 6.3682973 4.4327400 4.3647770
V B - Comparison between Mechanical (G1) and Laser (G3).
Caries
Removing
Agent
No of
Samples
Mean
Tensile
strength
Median
Tensile
strength
Standard
Deviation
Test of
Significance
Group – I
Mechanical
15 6.2843073 6.3329400 3.1818350
Man-
Whitney
‘U’= 79.000
p-value =
0.165
Group – III
Laser
15 4.4758747 4.1761100 2.9174428
Results
40
V C - Comparison between Carisolv (G2) and Laser (G3).
Caries
Removing
Agent
No of
Samples
Mean
Tensile
strength
Median
Tensile
strength
Standard
Deviation
Test of
Significance
Group – II
Carisolv
15 6.3682973 4.4327400 4.3647770
Man-
Whitney
‘U’= 82.000
p-value =
0.206
Group – III
Laser
15 4.4758747 4.1761100 2.9174428
2. Micro tensile bond strength ( µTBS)
The highest bond strength values were observed in the Carisolv group (6.37MPa) and
the Mechanical group (6.25MPa), and the lowest value was found with the Laser group
(4.47MPa). The data was statistically analyzed using Kruskal-Wallis non-parametric one-
way ANOVA and Mann-Whitney ‘U’ test.
Results
41
Chart II
Chart III
Results
42
Table VI- Fracture Mode Analysis
Types Mechanical(G1) Carisolv(G2) Laser(G3)
1 – A 0 1 5
2 – B 9 5 8
3 – C 1 2 0
4 – D 5 7 2
Chart IV
3. Failure mode analysis
A substantial increase in cohesive dentin failure was observed in CarisolvTM
group. For
the mechanical caries excavation, fractured specimens typically showed a ‘mixed’
fracture pattern. The predominant fracture modes of laser-ablated dentin were Type II
(53%) followed by Type I (33%).
0
1
2
3
4
5
6
7
8
9
10
1 - A 2 – B 3 – C 4 – D
Mechanical(G1)
Carisolv(G2)
Laser(G3)
Results
43
Table I – The Micro Tensile Bond Strength (µTBS) values in MPa (SD) for the ‘residual
caries-excavated dentin’ according to the caries excavation technique tested (Mechanical
(G1).
Table II – The Micro Tensile Bond Strength (µTBS) values in MPa (SD) for the
‘residual caries-excavated dentin’ according to the caries excavation technique tested
Carisolv(G2) group.
Table III – The Micro Tensile Bond Strength (µTBS) values in MPa (SD) for the
‘residual caries-excavated dentin’ according to the caries excavation technique tested:
Laser (G3) group.
Table IV – Mean Micro Tensile Bond Strength (µTBS) values comparison between
Mechanical (G1), Carisolv (G2) and the Laser (G3) groups. The values for the ‘residual
caries-excavated dentin’ according to the caries excavation techniques tested were 6.28 ±
3.18, 6.36 ± 4.36 and 4.47 ± 2.91 respectively. Non-parametric Kruskal-Wallis test
analysis (0.298) indicated that they were not statistically significant (p� 0.05).
Results
44
Table V - Comparison of Mean Micro Tensile Bond Strength (µTBS) values of two
groups at a time. Inter group comparison was carried out using Man-Whitney ‘U’ test.
Table II A shows comparison between Mechanical (G1), Carisolv (G2), and the p-value
is 0.787, which is statistically not significant. Table II B is the Comparison
between Mechanical (G1), Laser (G3) and the p-value is 0.165, which is also statistically
not significant. Table II C represents the Comparison between Carisolv (G2), Laser (G3)
groups with the p-value of 0.206 and that showed no significant statistical difference
(p� 0.05).
Table VI – Distribution of fracture types for different caries removing techniques.
Fracture mode analysis exhibited an increase in the percentage of cohesive failure in
dentin with Carisolv (G2) group. The Mechanical (G1) group specimen showed a high
number of ‘mixed’ fracture pattern and the Laser (G3) group revealed a mixed failure
pattern.
Results
45
Chart I - Mean Micro Tensile Bond Strength (µTBS) values of different caries removing
techniques. Kruskal-Wallis test analysis indicated no significant difference (0.298)
between the caries-excavated groups.
Chart II – Distribution of tensile strength in different groups: Carisolv (G2), Laser (G3)
and Mechanical (G1). The dots represent the minimum and the maximum values.
Chart III – Mean Micro Tensile Bond Strength (µTBS) values of different caries
removing techniques in MPa. The boxes represent the spreading of the data between the
first and third quartile. The central line represents the median. The whiskers denote the
5th
and 95th
percentiles. Group I is Mechanical (6.28 ± 3.18), Group II Laser (4.47 ± 2.91)
and Group III is Carisolv (6.36 ± 4.36) in this chart.
Chart IV – Distribution of type of fractures as seen using the stereomicroscope after
Micro Tensile Bond Strength (µTBS) have been performed. An increased number of
cohesive dentin fractures were observed with carisolv group.The mechanical caries
excavation group, fractured specimens typically showed a ‘mixed’ fracture pattern. The
predominant fracture modes of laser-ablated dentin group were of mixed type.
Discussion
Discussion
46
Minimal Invasive Dentistry has gained popularity with the development of new adhesive
systems and technological improvements in tooth preparation75
. Minimal invasive techniques
claimed to achieve controlled removal of infected and softened dentin, while preserving healthy
and hard dental tissues and perform with a minimal discomfort to the patient63,66
. The superficial
necrotic zone of caries-infected dentin which harbors the core bacterial biomass should be
excavated leaving only “residual” caries-affected dentin lining the cavity with sound enamel
margins and dentin adjacent to the enamel-dentin junction3,23
.
Fusayama (1966) claimed that cariogenic bacteria were never found beyond the softening
front of dentin24
. Elimination of the heavily infected dentin and preservation of the residual
affected dentin were thus defined as pre-requisites for effectively arresting the carious process
without harming the long-term survival of the pulp and the restoration32
. This will enable the best
peripheral seal to be achieved with the current adhesive dentin bonding agents55
.
The transition zone between infected and affected dentin is difficult to assess. Change in
color is not a good indicator as the gradual pathologic changes are not consistently color-
dependent60
. Defining the actual end point of caries excavation is the start point of restoration
and is often clinically challenging40
. EAM Kidd (2004) suggested removing carious dentin to the
level where it is ‘firm’42
. The ability of a restorative material to achieve a strong and durable
Discussion
47
bond with the tooth substrate is of paramount importance for the clinical success56
. Complete
removal of infected-dentin is mandatory to attain this goal.
The conventional methods of carious dentin removal by using rotary instruments are
proved to be quick and efficient. However it may result in unnecessary removal of the healthy or
even the affected dentin that shows the ability for remineralization. Moreover this technique is
usually associated with pain and discomfort to the patient77
. As a result of the former drawbacks,
a growing interest has been noticed to develop alternative minimally invasive techniques. The
search for a more gentle, comfortable and conservative caries excavation has led to the
development of methods which aim at providing minimal thermal changes, less vibration and
removal of infected dentin only.
Techniques such as laser, chemomechanical excavation and air abrasion have been shown
to be more or less successful in overcoming these problems15
. These methods offer interesting
advantages in comparison to the conventional approach, but are still far from fulfilling all
requirements. In most cases they are more time consuming than bur preparation, require more
investment and space, and are still dependent on conventional burs to gain access to the lesion
and to finish the preparation margins. Moreover, many of these techniques tend to over or under
prepare or do not completely eliminate the smear layer9. Chemochemical caries removal involves
the selective removal of carious dentin and is proved to be clinically effective in the removal of
infected-dentin and also harmless to healthy tissue and bio-compatible to the pulp8. Carisolv
TM
system, provides efficient removal of dental caries with no harm expected either on the healthy
Discussion
48
dentin or the pulp tissues19
. The chemo mechanical caries removing agent ‘CarisolvTM
’ consists
of 0.5% w/v sodium hypochlorite, 0.1M of an amino acid mixture (glutamic acid, leucine and
lysine) and water11
. It is available in two tube system and the contents of the tubes were mixed
and introduced into the prepared cavity using specific instruments provided by the manufacturer.
The gel was applied in the cavity, after which the carious dentin is scrapped off using specially
designed non-cutting hand instrument30, 16
.
Lasers can be considered as a possible replacement for current caries excavation
methods. Erbium lasers have been pointed out as the most promising, due to their specificity in
ablating enamel and dentin without side effects to the pulp and surrounding tissue57
. The Er:
YAG (Erbium loaded Yttrium – Aluminum –Garnet) and the Er, Cr: YSGG (Erbium,
Chromium: Yttrium – Scandium – Gallium – Garnet) are the two types of erbium lasers currently
available. Both devices present very similar wavelength (2.78µm for Er, Cr: YSGG and 2.94µm
for Er: YAG), although the Er, Cr: YSGG laser is discretely more absorbed by hydroxyapatite
than Er: YAG41,71
. The wavelength emitted by Er, Cr: YSGG laser (2.78µm) coincides with the
absorption peak of water and is well absorbed by all biological tissues including enamel and
dentin34
.
The hard tissue erbium lasers have the capability to prepare enamel, dentin, caries,
cementum and bone in addition to cutting soft tissue. The ability of hard tissue lasers to reduce or
eliminate vibrations, the audible whine of drills, microfractures, and some of the discomfort that
many patients fear and commonly associate with high-speed handpieces is impressive. In
addition, these lasers can be used with a reduced amount of local anesthetic for many procedures,
Discussion
49
which is another feature that makes the hard tissue laser very exciting for needle-phobic
patients62
.
Lee et al (2007) claimed that the dentin irradiated with laser power output above 3.5W
would exhibit micro cracks under SEM examination. Hence in this study dentin surfaces were
irradiated with a laser power output of 3.5W to avoid the micro cracks. Frequency is fixed at
20Hz. Air pressure level of 80% could create the roughest surface and the water pressure level
was elevated to 70% to get a least charred or carbonated dentin surface34
.
Conventional excavation with burs and spoon excavators are the most popular current
caries excavation techniques. The selective removal of caries-infected dentin is not easily
achieved with this method due to relative tactile insensitivity and operator variability; which
leads to varying quantities of tissue removed35
. Most of the caries excavation techniques leave a
layer of smear layer in the residual dentin75
. A major factor to consider in terms of restoring a
cavity with current adhesives are the final dentin surface characteristics such as surface
roughness and the presence of smear layer, which can affect the final bond and the seal achieved
by adhesive systems3, 10, 72
.
The smear layer in caries-affected dentin may be more resistant to the action of self-
etching primers, as they include acid-resistant crystals and extrinsic proteins that might have
permeated into the mineral phase during demineralization cycles76
. If a residual smear layer is
Discussion
50
left on the surface, the adhesive resin will bond to crystals within it rather than to underlying
dentin49
. Moreover, the acidity of the primer could also be buffered by the mineral content of the
smear layer29
. D H Pashley (1995) found that, the presence of mineral casts within the tubules of
caries-affected dentin, would prevent the formation of resin tags in dentinal tubules, and it is
believed that tags are thought to contribute to bond strength67
.
Analysis of the morphological changes in the residual dentin is critical in assessing the
adherence capacity of the caries-excavated dentin. Arguably Scanning Electron Microscope
(SEM) is the best instrument for the topographical analysis as it is easy to operate, large depth of
field which allows more of a specimen to be in focus at one time, much higher resolution and the
data acquisition is rapid28
. Hence, this in vitro study aimed to evaluate morphological changes in
caries-excavated dentin using a scanning electron microscope52
.
Samples for scanning electron microscopic (SEM) evaluation were immersed in 4%
glutaraldehyde for one hour, rinsed in distilled water and placed for one hour in sodium
cacodylate for the fixation of the tooth substrate. Following fixation, the samples were placed in
ethanol in ascending series for the dehydration63
. For SEM analysis, the specimens should be
dried and gold sputter coated for electrical conduction.
Er, Cr: YSGG laser irradiation produced a scaly, irregular and rugged appearance of
dentin in SEM evaluation. Absence of smear layer and partially closed dentinal tubule orifices
Discussion
51
were observed. The peri-tubular dentin was found protruding from the intertubular dentin. It
could be due to the higher mineral content and the lower water content of peri-tubular dentin.
Signs of carbonation were not found. The marked surface irregularities and lack of smear layer
provided solid evidence for the physical mechanism of bonding with the composite material after
laser treatment34, 36
.
The dentin surface treated with CarisolvTM
observed under scanning electron microscope
in the present study showed uneven surface with many undermined areas leaving residual-caries
affected dentin which is partially demineralized and the intertubular dentin exhibiting a high
degree of porosity. There were partially patented dentinal tubules and residues of contaminant
smear layer covering the dentinal surfaces. The dentin topography after Carisolv treatment was
granular and rough comparable to laser prepared dentin. This surface roughness and structural
changes may play a crucial role in adhesion to the composite material, possibly without using
etching agents3, 12
. The ‘Mild’ two-step self-etch adhesives have a potential to penetrate deeper
and form a thick and homogenous hybrid layer.
The scanning electron microscopic examination of the dentin surfaces prepared using
tungsten carbide bur and spoon excavator produced smearing and smear plugs in the tubular
orifice. This method leaves a homogenous smear layer with more or less uniform roughness, and
dentinal tubules visibly obstructed with smear plugs. Though the residual dentin after mechanical
excavation showed similar hardness to that of sound dentin, there are chances of sclerotic dentin
formation within the dentinal tubules that can lead to complete obliteration14
. With regard to
Discussion
52
bonding receptiveness, the smear covered surface does not interfere with the etch-and-rinse
adhesive, but may reduce the bonding effectiveness of self-etching adhesives14, 59
.
The basic mechanism of bonding to enamel and dentin is essentially an exchange process
involving replacement of minerals removed from the hard dental tissues by resin monomers that
upon setting becomes micro mechanically interlocked in the created porosities37, 38
. The etch-
and-rinse and the self-etch methods are employed to produce an effective bond between the
direct restoration and dentin46
. The total-etch technique relies on the removal of the smear layer
and exposure of collagen matrix by acid-etching, followed by the application of a self-priming
agent that combines the primer and the adhesive resin into one solution51
, 37
. Self-etching
adhesive systems have been used to simplify adhesive procedure as it decreases technique
sensitivity45
. The self-etch approach is the use of self-etching primers, in which the acid and the
primer are combined in one solution to form a highly hydrophilic and acidic monomers, that
make hybrid layer more permeable and sensitive to water sorption from the underlying dentin65
.
A two-step mild self-etch adhesive (Clearfil SE Bond, Kuraray, Japan), which have
repeatedly shown to be excellent performer in clinical and laboratory studies and is considered to
be ‘gold-standard’ adhesives for their class, was used in this study39
. Two-step self-etch
adhesives were characterized by separate chemical formulations for priming and bonding,
utilizing a self-etching hydrophilic primer that is followed by the application of a comparatively
mild hydrophobic bonding agent22
. ‘Mild’ self-etch adhesives demineralize the dentin surface
sufficiently to provide micro-mechanical retention, while preserving hydroxyapatite within the
Discussion
53
hybrid layer to enable additional chemical interaction53, 69
. ‘Mild’ two-step self-etch adhesives
that provide additional chemical bonding and that appear to be the most optically combine
bonding effectiveness with a simplified application protocol 53, 74
. Omar et al (2007) proved that
the microtensile bond strength of residual dentin, when bonded with two-step self-etch adhesive
is comparable to that of sound dentin44
.
The performance of enamel and dentin adhesives can be investigated by shear or tensile
bond strength17
. Micro Tensile Bond Strength (µTBS) testing is one of the most commonly used
methodologies to mechanically assess the strength of resin dentin interface complex1. Adhesive
performance of the enamel and dentin can be tested by macro or micro bond strength approaches,
depending upon the size of the bonded area27
. The macro bond strength tests as tensile bond
strength and shear bond strength are performed in specimens with bond area of 3 mm² or more68,
4. The use of macro test methods has been eclipsed in the recent years because of the advantages
of micro test methods78
. Currently Micro Tensile Bond Strength (µTBS) test is the most popular
method47
.
The advantages of the Micro Tensile Bond Strength (µTBS) test can be enumerated as68, 48
:
1. More adhesive failure, fewer cohesive failure
2. Higher initial bond strength
3. Permits measurements of regional bond strength
4. Means and variances can be calculated for single teeth
Discussion
54
5. Permits testing bonds to irregular surface
6. Permits testing of very small areas
7. Facilitate SEM examination of the failed bonds since the surface area is
approximately 1 mm².
1 mm² samples were prepared for the Micro Tensile Bond Strength (µTBS) test. The
adhesive (Clearfil SE Bond, Kurraray, Japan) was applied over the prepared dentin and
composite resin was incrementally polymerized onto the adhesive26
. Four millimeter to six
millimeter thickness of composite resin was bonded over the prepared surface; which helped to
prepare samples containing residual dentin, adhesive junction and the composite resin, which
favors easy holding of samples during testing6. After a storage period of 24hours in water
70, the
specimen was sectioned to 1 mm² stick shaped samples using a high-speed precision saw
(Buehler – USA) under water spray50
.
Sample preparation for the Micro Tensile Bond Strength (µTBS) test can be of stick
shaped, dumbbell shaped or hour-glass shaped. Studies have shown that stick and dumbbell
shaped specimens with a rectangular bonded area shows similar bond strength and failure
modes54
. Sawing and trimming of the test samples appears to be the most technique-sensitive
part of the Micro Tensile Bond Strength (µTBS) test. The interfacial stress during the preparation
can leads to pre-test failure of samples. The hour-glass or dumbbell shaped trimming of the
samples places additional pre-test strain and stress concentration on the bonded area; hence a
stick shaped samples were preferred for this study.
Discussion
55
Following the sample preparations, the test specimens were mounted on a pneumatic jig
using suitable sized silicon-sleeves on both the end of the specimens. The conventional method
of using fast-setting adhesives for the fixation of samples to the jig which can lead to glue
contamination of the samples; as well it can influence the test results.
The Mean Micro Tensile Bond Strength (µTBS) value of the CarisolvTM
group (G2)
showed highest among the three groups: 6.36MPa (SD 4.36). This value was comparable with
the values of the Mechanical group G1 (6.28 MPa [SD 3.18]). The third group; Er, Cr: YSGG
laser showed lower values (4.47 MPa [SD 2.91]) than that of the other two groups. The higher
values shown by CarisolvTM
group could be because of the removal of smear layer by the action
of CarisolvTM
, which allowed the self-etch adhesive to penetrate deeper into the dentinal
surfaces; resulting in a thick and homogenous hybrid layer3, 59
.
Residual dentin which remained after chemomechanical (CarisolvTM
) excavation showed
a roughened surface, with minimum smear layer and many open tubular orifices. The rough and
granular dentin topography might have allowed the ‘Mild’ self-etch adhesive to penetrate deeper
and form a thicker and smoother hybrid layer; that resulted as the highest Mean Micro Tensile
Bond Strength (µTBS) value among all the experimental groups.
Discussion
56
Mechanical caries excavation exhibited over-preparation with more histologically sound
dentin at the cavity margins. SEM examination showed that hand excavation produced smearing
and smear plugs in the tubular orifice. Although the dentin that is left after caries removal with
hand excavation has similar hardness to that of normal sound dentin, there are ultra-structural
differences. It is well known that as caries progress; sclerotic dentin (translucent dentin) is
deposited within the tubules at the advancing front of the lesion, which can lead to their complete
obliteration58
. This factor, in conjunction with the formation of the smear layer that has a higher
organic content, due to the increased bacterial load, may interfere with the ability of the ‘Mild’
self-etch adhesive to penetrate deeper into smear layer and thus demineralizing the underlying
dentin, leading to a reduced Micro Tensile Bond Strength (µTBS) value.
Er, Cr: YSGG laser irradiation produced a scaly, irregular and rugged appearance with
the absence of smear layer with exposed dentinal tubule orifices. All these features could
increase the adhesive area and strength, but lack of good hybrid layer could counteract these
effects and cause a lower Micro Tensile Bond Strength (µTBS) value. The laser initially
vaporizes water and the other hydrated organic components of the dentin. During the process,
internal pressure increases in the dentin until the explosive destruction of inorganic substances
occurs. Since intertubular dentin contains more water and has a lower mineral content than the
peritubular dentin, it is selectively more ablated than the peritubular dentin, leaving protruding
dentinal tubules with a cuff-like appearance2. Thus it would be expected to have better bond
strength to irradiated dentin. However, not only the morphology of the dentin surface is
important for bonding, but the chemical composition of the intertubular dentin is also an
Discussion
57
important factor. During the bonding procedure, this dentin is demineralized and then permeated
by the hydrophilic monomers; the bonding resin can then hybridize with the network of collagen
fibers. Thus the bonding to irradiated dentin was less than the other groups; it can be assumed
that, the composition of the intertubular dentin has been modified by the laser irradiation. This
modification could have lead to a dentin surface more resistant to demineralization, impairing the
action of the mild pH hydrophilic primer used62
.
The fracture mode analysis of the resin-dentin interface revealed that the three caries
excavation techniques have no effect on the failure mode and loci of failure. It is apparent from
the data that samples in all groups suffered similar proportions of cohesive failure in the
composite aspect of the restorative interface. A substantial increase in cohesive dentin failure
was observed in the CarisolvTM
group. Sodium hypochlorite might have reduced the mechanical
properties of dentin, causing the specimens to fail more cohesively in dentin59, 7
. For the
mechanical caries excavation group, fractured specimens typically showed a ‘mixed’ fracture
pattern, including areas that fractured in dentin, composite and within the adhesive. The
predominant fracture modes of laser-ablated dentin were Type II (53%) followed by Type I
(33%). More number of adhesive failures could be because laser ablation did not create an
appropriate hybrid layer.
Within the limitations of this study it was concluded that; the surface remaining after
CarisolvTM
excavation and bonding with a mild self-etch adhesive seem to be very compatible,
allowing deeper penetration of the adhesive and resulted in a thick and homogenous hybrid layer.
Discussion
58
In this study, the inclusion criteria for samples were teeth with caries lesion limited to the
occlusal surface and extending at least half of the distance from the enamel-dentin junction to the
pulp chamber62
. This was to minimize any regional variation between the periphery and the
central dentin substrate and to increase the validity of the results53
. Generally micro specimens
for the microtensile bond strength studies were fixed to a jig with the aid of a cyanoacrylate-
based glue, whereas silicon sleeves were used in this study. This method might have resulted in
lower values when compared to the published data’s. Further studies are recommended with
larger number of samples for the microtensile bond strength evaluation.
Summary
Summary
59
The advent of ‘Adhesive Dentistry’ has simplified the guidelines for cavity preparation
enormously. In light of the minimally invasive dentistry concept, the macroretentive cavities
have been replaced by cavities limited to removal of carious dentin that are at most extended
with some additional rounding off sharp margin edges and/or beveling. Caries dentin removal by
mechanical means is non-specific nature of excavation that may result in excessive loss of tissue
thus affecting the prognosis of the treated tooth. This in vitro study was aimed to evaluate the
surface characteristics of caries-excavated dentin and to compare the microtensile bond strength
values of an adhesive bonded to the residual dentin remained after three different contemporary
caries-excavation methods namely Mechanical, Chemomechanical (CarisolvTM
) and Er, Cr:
YSGG laser (BiolaseTM
).
Forty five extracted cavitated human permanent molars with the lesion limited to the
central portion of the occlusal surface were selected. The teeth were randomly divided into three
groups (n=15), according to the caries removal methods, as follows: Group I: Mechanical rotary
preparation by diamond bur/air turbine. Group II: Chemo-mechanical preparation with
carisolvTM
gel. Group III: Laser preparation by Er. Cr: YSGG Laser(BiolaseTM
).
Morphological changes in the residual dentin surfaces were evaluated using scanning
electron microscopy (SEM). These specimens were immersed in 4% glutaraldehyde solution for
1hr at room temperature. Then they were rinsed in distilled water. The samples were then placed
in cold buffer solution of sodium cacodylate for 90 minutes to fix the organic matter. Specimens
were then dehydrated in ascending grades of ethanol and then dried in a Critical Point Drier
Summary
60
based desiccator. The dried specimens were mounted on a metal stand and gold sputter coated by
cathode atomization under vacuum. SEM images of the caries excavated surfaces were obtained.
After caries excavation the samples were bonded using a “Mild” two-step self-etch
adhesive, Clearfil SE Bond and then restored using a nanohybrid resin composite. The specimens
were serially sectioned into multiple slices with a diamond disc under water cooling to create a
1mm² stick shaped samples for the Micro-tensile bond strength test (µTBS). The prepared rods
comprised of residual dentine to which composite resin was bonded using suitable dentin
adhesive. Silicon sleeves were attached to both the end of the samples. Specimens were mounted
on a pneumatic jig. The test was taken out at a tension of 0.5mm/min until it failed. Bond
strengths were calculated by dividing the load at failure by the cross-sectional bonding area.
In this study, the morphological analysis of the residual dentin excavated with CarisolvTM
produced a reduced and more inconsistent smear layer with large areas of open tubular orifices.
The dentine surfaces that received Er, Cr: YSGG laser irradiation showed a scaly, irregular and
rugged appearance whereas the teeth prepared by mechanical method appeared rougher with
irregular particles scattered throughout the surface area. A well defined smear layer was
detected.
The results were subjected to statistical analysis by means of parametric Kruskal-Wallis
one-way ANOVA and Mann-Whitney ‘U’ test.
The Microtensile bond strength evaluation showed the highest bond strength values for
the CarisolvTM
group (6.37MPa) and the mechanical group (6.25MPa), and the lowest value was
found with the laser group (4.47MPa).
Conclusion
Conclusion
61
Within the limitations of this in vitro study it can be concluded that the caries-removal
effectiveness and minimal-invasiveness potential varied amongst the three caries-excavation
techniques tested. It was found that the surface remaining after CarisolvTM
excavation and
bonding with a mild self-etch adhesive seem to be very compatible, allowing deeper penetration
of the adhesive and resulted in a thick and homogenous hybrid layer.
The fracture mode analysis of the resin-dentin interface revealed that the three caries
excavation techniques have no effect on the failure mode and loci of failure. A substantial
increase in cohesive dentin failure was observed in the Carisolv group, whereas mechanical and
laser groups showed a mixed fracture pattern.
Bibliography
Bibliography
62
1. Armstrong S, Geraldeli S, Maia R, Raposo L H A, Soares C J, Yamagawa J. Adhesion to
tooth structure: A critical review of micro bond strength test methods. Dental Materials
(2010); 26: e50-e62.
2. Aoki A, Ishikawa I, Yamada T, Otsuki M, Watanabe H, Tagami J, Ando Y, Yamamoto H.
Comparison between Er: YAG laser and conventional technique for root caries treatment in
vitro. J Dent Res (1998); 77: 1404-1414.
3. Banerjee A, Kellow S, Mannocci F, Cook R J, Watson T F: An in vitro evaluation of
microtensile bond strengths of two adhesive bonding agents to residual dentin after caries
removal using three excavation techniques: Journal of Dentistry 38 (2010) 480-489.
4. Braga RR, Meira JBC, Boaro LCC, Xavier TA. Adhesion in Dentistry – Analyzing Bond
strength test methods, variables and outcomes. As: Ulrich Salz, Thorsten Bock: Testing
Adhesion of Direct Restoratives to Dental Hard Tissue: J Adhes Dent (2010); 12: 343-371.
5. Brulat N, Rocca JP, Leforestier E, Fiorucci, Nammour S, Bertrand MF: Shear bond strength
of self-etching adhesive systems to Er: YAG – laser- prepared dentin: Lasers Med Sci
(2009) 24: 53-57.
6. Betamar N, Cardew G, Van Root R. Influence of specimen designs on the microtensile bond
strength to dentin. J Adhes Dent (2007); 9: 159-168.
7. Burrow MF, Bokas J, Tanumiharja M, Tyas MJ. Microtensile bond strengths to caries-
affected dentine treated with Carisolv. Australian Dental Journal (2003); 48: 110–114.
8. Beeley JA, Yip HK, Stevenson AG; Chemochemical Caries removal: A Review of the
Techniques and Latest Developments; British Dental Journal (2000) 188, 427-430.
9. Banerjee A, Watson T F, and Kidd E. A. M: Dentine Caries Excavation; A Review of
Current Clinical Techniques: British Dental Journal (2000) 188, 476-482.
Bibliography
63
10. Banerjee A, Kidd EAM, Watson TF. Scanning Electron Microscopic Observations on
Human Dentin after Mechanical Caries Excavation. Journal of Dentistry (2000); 28: 179-
186.
11. Banerjee A, Kidd E. A. M. and Watson T F: In vitro Evaluation of Five Alternative
Methods of Carious Dentin Excavation; Caries Res (2000) 34: 144-150.
12. Cahreli ZC, Yazici AR, Akra T, Ozgunaltay GA. Morphological and microtensile bond
strength evaluation of a single-bottle adhesive to caries-affected dentin after four different
caries- removal techniques. Journal of Dentistry (2003); 31; 429-435.
13. Cardoso M V, Coutinho E, Ermis R B, Poitevin A, Landuyt K V, Munk J D, Carvalho C R,
Meerbeek B V. Influence of dentin cavity surface finishing on microtensile bond strength of
adhesives. Dental Materials (2008); 24: 492-501.
14. Ceballos L, Camejo DG, Victoria Fuentes M, Osorio R, Toledano M, Carvalho RM,
Pashley DH. Microtensile bond strength of Total-etch and Self-etching Adhesives. Journal
of Dentistry (2003); 31: 469-477.
15. Celiberti P, Francescut P, Lussi A; Performance of Four Dentin Excavation methods in
Deciduous Teeth: Caries Res (2006); 40: 117-123.
16. Dahlin S (2001) Chemical mode of action of Carisolv. In: Albrektsson T, Bratthall D,
Glantz PO, Lindhe J. (eds) Tissue preservation in caries treatment. Quintessence, London,
pp 167–175.
17. DeVries KL, Borgmeier PR. Testing of Adhesives. In: Handbook of Adhesive Technology;
Marcel Dekker Inc (2003); 223.
18. Dunn W J, Davis J T, Bush A C. Shear bond strength and SEM evaluation of composite
bonded to Er: YAG laser prepared dentin and enamel. Dental Materials (2005); 21: 616-624.
Bibliography
64
19. Elkhonay NR, Abdelaziz KM, Zaghloul NM, Aboulenine N; Chemomechanical Method: A
Valuable Alternative for Caries Removal; J Minim Interv Dent (2009) 2: 4; 248-260.
20. Eberhard J, Bode K, Hedderich J, Jepsen S: Cavity size difference after caries removal by a
fluorescence-controlled Er: YAG laser and by conventional bur treatment: Clin Oral Invest
(2008) 12: 311-318.
21. Ericson D, Zimmerman M, Raber H, Gotrick B, Bornstein R, Thorell J: Clinical evaluation
of efficacy and safety of a new method for chemomechanical removal of caries. A multi-
center study. Caries Res. 1999; 33: 171-177
22. Frankenberger R, Tay FR: Self-etch versus etch-and-rinse adhesive: effect of thermo
mechanical fatigue loading on marginal quantity of bonded resin composite restoration:
Dental Materials (2005); 21: 397-412.
23. Fusayama T, Two layers of Carious Dentin: diagnosis and treatment. Operative Dentistry
1979; 4: 63-70.
24. Fusayama, T.; Okuse, K.; and Hosoda, H.: Relationship between Hardness, Discoloration,
and Microbial Invasion in Carious Dentin, JDent Res 45:1033-1046, 1966.
25. Giannini M, Carvalho RM, Martins LR, Dias CT & Pashley DH (2001): The influence of
tubular density and area of solid dentin on bond strength of two adhesive systems to dentin:
Journal of Adhesive Dentistry: 3(4); 315-376.
26. Goracci C, Margvelashvili M, Apicella D, Sedda M, Magni E, Ferrari M. Influence of resin
composite mechanical properties on adhesive microtensile bond strength to dentin. J Adhes
Dent (2011); 13: 323-331.
27. Heintze S D, Rousson V. Pooling of microtensile bond strength data improves clinical
correlation. J Adhes Dent (2011); 13: 107-110.
Bibliography
65
28. Inoue G, Tsuchiya S, Nikaido T, Foxton R M, Tagami J. Morphological and mechanical
characterization of the acid-base resistant zone at the adhesive-dentin interface of intact and
caries-affected dentin. Operative Dentistry (2006); 31-4: 466-472.
29. Itou K, Torii Y, Suziki K, Makai H, Inoue K. Adhesion of Restorative Resin to Tooth-
adhesion Promoted by Linear Bond. J Adhes Dent (1994); 12: 174-181.
30. Kirzioglu Z, Gurbuz T, Yilmaz Y: Clinical evaluation of chemomechanical and mechanical
caries removal: status of the restoration at 3, 6, 9 and 12 months. Clin Oral Invest (2007)
11:69-76
31. Kidd E A M. Essentials of Dental Caries. 3rd
edn. Oxford University Press (2005). Page 1-4.
32. Kidd E A M. How ‘Clean’ must a cavity be before restoration? Caries Res 2004; 38: 305-
313.
33. Kidd E A M, Ricketts D, Beighton D. Criteria for caries removal at the enamel dentin
junction. Bri Dent J. (1996); 287:9: in Fontana M, Young DA, Wolff MS, Pitts NB,
Longbottom C. Editors. Current Concepts in Cariology. Dent Cli Nor America (2010) page
541 –5 49.
34. Lee B S, Lin P Y, Chen M H, Hsieh T T, Lin C P, Lai J Y, Lan W H: Tensile bond strength
of Er, Cr: YSGG laser-irradiated human dentin and analysis of dentin-resin interface: Dental
Materials 23 (2007) 570-578.
35. Lennon AM, Buchalla W, Rassner B, Becker K, Attin T: Efficiency of Four Caries
Excavation Methods Compared: Operative Dentistry (2006), 31-5; 551-555.
36. Lee BS, Lin CP, Lin FH, Lan WH. Ultrastructural Changes of Human Dentin after
irradiation by Nd: YAG Laser. Lasers Surg Med (2002) 30; 3: 246-252.
Bibliography
66
37. Meerbeek V B, Yoshihara K, Yoshida Y, Mine A, De Munck J: State of the art of self-etch
adhesives: Dental Materials 27 (2011) 17-28.
38. Meerbeek V B, De Munck J, Yoshida Y, Inoue S,Vargas M, Vijay P, Landuyt, K Van,
Lambrechts P, Vanherle G:Bunocore memorial lecture; Adhesion to Enamel and Dentin:
Current Status and Future Challenges: Operative Dentistry. 2003, 28-3, 215-235.
39. Mine A, Munk J D, Cardoso M V, Landuyt K L V, Poitevin A, Kuboki T, Yoshida Y,
Suzuki K, Lambrechts P, Meerbeek B V. Bonding effectiveness of two contemporary self-
etch adhesives to enamel and dentin. Journal of Dentistry (2009); 37: 872-883.
40. Neves A de A, Coutinho E, Cardoso MV, De Munk J, Meerbeek V B: Micro-tensile bond
strength and interfacial n charecterizaton of an adhesive bonded to dentin prepared by
contemporary caries-excavation techniques: Dental Materials 27 (2011) 552-562.
41. Neves A de A, Coutinho E, Cardoso M, Lambrechts P, Meerbeek BA: Current concepts
and techniques for Caries Excavation and Adhesion to Residual Dentin. J Adhes Dent 2011;
13: 7-22.
42. Neves A de A, Coutinho E, De Munk J, Meerbeek V B. Caries-removal effectiveness and
minimal-invasiveness potential of caries-excavation techniques: A micro-CT investigation.
Journal of Dentistry (2011); 39: 154-162.
43. Ole Fejerskov, EAM Kidd. Editors. Dental Caries: The Disease and its Clinical
Management.2nd
edition. Oxford (UK): Blackwell Munksgaard (2008). Page 2- 5.
44. Omar H, El-Badrawy W, El-Mowafi O, Atta O, Saleem B. Microtensile bond strength of
resin composite bonded to caries-affected dentin with three adhesives. Operative Dentistry
(2007); 32-1: 24-30.
Bibliography
67
45. Paranhos M P G, Spohr A M, Marcondes M, Oshima H M S, Mota E G, Burnett L H.
Influence of Nd: YAG laser irradiation on microtensile bond strength of adhesive system to
sound or carious dentin. Quintessence Int (2009); 40: 145-153.
46. Pashley D H, Tay FR, Breschi L, Tjaderhane L, Carvalho RM, Carrilho M, Mutluay A T:
State of the art etch-and-rinse adhesives: Dental Materials 27 (2011) 1-16.
47. Pashley DH, Carvalho RM, Sano H, Nakajima M, Yoshiyama M, Shono Y, Fernandes CA,
Tay F. The Microtensile Bond Strength Test – A Review. J Adhes Dent (1999); 1: 299-309.
48. Pashley DH, Sano H, Ciucchi B, Yoshiyama M, Carvalho RM. Adhesion Testing of Dentin
Bonding Agents – A Review. Dent Mater, (1995); 11: 117-125.
49. Pashley EL, Tao L, Pashley DH. Effects of Oxalates on Dentin Bonding. Am J Dent
(1993); 6: 116-118.
50. Poitevin A, De Munk J, van Landuyt K, Coutinho E, Peumans M, Lambrechts P, Van
Meerbeek B. Critical analysis of the influence of different parameters on the microtensile
bond strength of adhesives to dentin. J Adhes Dent (2008); 10: 7-16.
51. Proenca JP, Franklin Garcia-Godoy et al: Dentin regional bond strength of self-etch and
total-etch adhesive systems: Dental Materials 23 (2007) 1542-1548.
52. Russell SD, Daghlian CP. "Scanning electron microscopic observations on deembedded
biological tissue sections: Comparison of different fixatives and embedding materials". J of
Electron Microscopy Technique (1985) 2; 5: 489–495.
53. Sarr M, Kane AW, Vreven J, Mine A, Landuyt KL Van, Peumans M, Lambrechts P,
Meerbeek V B, De Munck J: Microtensile Bond Strength and Interfacial Characterization
of 11 Contemporary Adhesives Bonded to Bur-cut Dentin: Operative Dentistry. 2010, 35-1,
94-104.
Bibliography
68
54. Scherrer SS, Cesar PF, Swain MV. Direct comparison of bond strength results on the
different test methods: A critical literature review. Dent Mater (2010); 26: e78-93.
55. Scholtanus J D, Purwanta K, Dogan N, Kleverlaan C J, Feilzer A J. Microtensile bond
strength of three simplified adhesive systems to caries-affected dentin. J Adhes Dent (2010);
12: 273-278.
56. Silva N R F A, Calamia C S, Harsono M, Carvalho R M. Bond angle effects on microtensile
bonds: Laboratory and FEA comparison. Dental Materials (2006); 22: 314-324.
57. Staninec M, Meshkin N, Manesh S K, Ritchie R O, Fried D: Weakening of dentin from
cracks resulting from laser irradiation: Dental Materials 25 (2009) 520-525.
58. Sattabanasuk V, Burrow MF, Shimada Y, Tagami J. Resin adhesion to caries-affected
dentine after different removal methods. Australian Dental Journal (2006); 51: 162–169.
59. Sonoda H, Banerjee A, Sheriff M, Tagami J, Watson TF. An Invitro Investigation of
Microtensile bond strengths of Two Dentin Adhesives to Caries-affected Dentin. Journal
of Dentistry (2005); 33: 335-342.
60. Splieth C, Rosin M, Gellissen B: Determination of residual dentin caries after conventional
mechanical and chemomechanical caries removal with carisolv: Clin Oral Invest (2001) 5:
250-253.
61. Tachibana A, Castanho G M, Viera S N, Matos AB. Influence of blood contamination on
bond strength of self-etching adhesives to dental tissue. J Adhes Dent (2011); 13: 349-358.
62. Tachibana A, Marques MM, Soler JMP, Matos AB: Er, Cr: YSGG laser for caries
removal: Influence on bonding of a self-etching adhesive system: Laser Med Sci (2008) 23:
435-441
Bibliography
69
63. Tsanova S Ts, Tomov G T. Morphological changes in Hard Dental Tissues Prepared by Er:
YAG Laser, Carisolv and Rotary Instruments. A Scanning Electron Microscopy Evaluation.
Folia Medica 2010; 52(3): 46-55.
64. Trajtenberg CP, Pereira PNR, Powers JM. Resin bond strength and micromorphology of
human teeth prepared with an Er: YAG laser. Am J Dent (2004); 17: 331-336; as in Lee B
S, Lin P Y, Chen M H, Hsieh T T, Lin C P, Lai J Y, Lan W H: Tensile bond strength of Er,
Cr: YSGG laser-irradiated human dentin and analysis of dentin-resin interface: Dental
Materials 23 (2007) 570-578.
65. Tay FR, King NM, Suh BI, Pashley DH. Effect of delayed activation of light-cured resin
composites on bonding of all-in-one adhesives. J Adhes Dent (2001); 3: 207–225.
66. Tyas MJ, Anusavice KJ, Frencken JE, Mount GJ: Minimal Intervention dentistry : A
Review – FDI commissioned Project. International Dental Journal 2000; 50: 1-12
67. Tao L, Pashley DH, Boyd L: Effect of Different Types of Smear layers on Dentin and
Enamel Shear Bond Strengths: Dental Materials; (1998), 15; 4:4; 208-216.
68. Ulrich Salz, Thorsten Bock: Testing Adhesion of Direct Restoratives to Dental Hard
Tissue: J Adhes Dent (2010); 12: 343-371.
69. Van Landuyt KL, Snauwaert J, De Munk J, Peumans M, Yoshida Y, Poitevin A, Coutinho
E, Suzuki K, Lambrechts P & Van Meerbeek B. Systematic Review of the Chemical
Composition of Contemporary Dental Adhesives. J Biomaterials (2007), 28; 26: 3757-3785
70. Vuorinen A M, Dyer S R, Vallittu P K, Lassila L V J. Effect of water storage on the
microtensile bond strength of composite resin to dentin using experimental rigid rod
polymer modified primers. J Adhes Dent (2011); 13: 333-340.
Bibliography
70
71. Walsh LJ. The Current Status of Laser Applications in Dentistry; Aus Dent J (2003); 48:
146-155.
72. White RR, Goldman M, Lin P. The Influence of the Smeared layer upon Dentinal Tubule
Penetration by Plastic Filling Materials. Part II. Journal of Endodontics 1987; 13: 369-374.
73. Yazici AR, Baseren M, Gorucu J: Clinical Comparison of Bur-and Laser-prepared
Minimally Invasive Occlusal Resin Composite Restorations: Two-yea Follow-up: Operative
Dentistry, 2010, 35-5, 500-507.
74. Yoshida Y, Nagakane K, Fukuda R, Nakayama Y, Okazaki M, Shintani H, Inoue S, Tagawa
Y, Suzuki K, De Munk J & Van Meerbeek B. Comparative study on Adhesive Performance
of Functional Monomers. Journal of Dental Research (2004), 83; 6: 454-458.
75. Yazici AR, Ozgunaltay G, Dayangac B. A Scanning Electron Microscopic study of
Different Caries Removal Techniques on Human Dentin; Oper Dent (2002); 27: 360-366.
76. Yoshiyama M, Urayama A, Kimochi T, Matsuo T, Pashley DH. Comparison of
conventional versus self-etching Adhesive bonds to caries-affected Dentin. Oper Dent;
(2000); 25: 163-169.
77. Yip HK, Samaranayake LP; Caries Removal Techniques and Instrumentation: A Review;
Clin Oral Invest (1992) 2: 148-154.
78. Zohairy A A E, Saber M H, Abdalla A I, Feilzer A J. Efficacy of microtensile versus
microshear bond testing for evaluation of bond strength of dental adhesive systems to
enamel. Dental Materials (2010); 26: 848-854.