Transcript of Monte Carlo model and spectral measurements of a computed tomography x-ray tube Magdalena Bazalova...
![Page 1: Monte Carlo model and spectral measurements of a computed tomography x-ray tube Magdalena Bazalova Frank Verhaegen.](https://reader036.fdocuments.us/reader036/viewer/2022062421/56649d705503460f94a526ac/html5/thumbnails/1.jpg)
Monte Carlo model and Monte Carlo model and spectral measurements of a spectral measurements of a computed tomography x-ray computed tomography x-ray
tube tube
Magdalena BazalovaMagdalena BazalovaFrank VerhaegenFrank Verhaegen
![Page 2: Monte Carlo model and spectral measurements of a computed tomography x-ray tube Magdalena Bazalova Frank Verhaegen.](https://reader036.fdocuments.us/reader036/viewer/2022062421/56649d705503460f94a526ac/html5/thumbnails/2.jpg)
OverviewOverview• MotivationMotivation• Diagnostic x-ray beam productionDiagnostic x-ray beam production• Monte Carlo (MC) simulationsMonte Carlo (MC) simulations• Spectral measurementSpectral measurement
– in computed tomography (CT) geometryin computed tomography (CT) geometry– experimental setup and Compton scatterexperimental setup and Compton scatter– detectordetector
• Corrections of measured spectraCorrections of measured spectra• Results: Comparison of MC simulated spectra Results: Comparison of MC simulated spectra
with measured spectrawith measured spectra• ConclusionsConclusions
![Page 3: Monte Carlo model and spectral measurements of a computed tomography x-ray tube Magdalena Bazalova Frank Verhaegen.](https://reader036.fdocuments.us/reader036/viewer/2022062421/56649d705503460f94a526ac/html5/thumbnails/3.jpg)
MotivationMotivation
• It is very difficult to assess patient dose It is very difficult to assess patient dose from CT imaging and mostly approximative from CT imaging and mostly approximative methods are used.methods are used.
• A Monte Carlo model of the x-ray tube of a A Monte Carlo model of the x-ray tube of a CT scanner in combination with patient CT CT scanner in combination with patient CT images will allow for an accurate estimate images will allow for an accurate estimate of CT dose.of CT dose.
• However, the MC model has to be accurate However, the MC model has to be accurate in order to calculate the dose accurately.in order to calculate the dose accurately.
• To developed and validate (using spectral To developed and validate (using spectral measurements) a MC model of a Philips measurements) a MC model of a Philips PQ5000 CT simulator x-ray tube.PQ5000 CT simulator x-ray tube.
![Page 4: Monte Carlo model and spectral measurements of a computed tomography x-ray tube Magdalena Bazalova Frank Verhaegen.](https://reader036.fdocuments.us/reader036/viewer/2022062421/56649d705503460f94a526ac/html5/thumbnails/4.jpg)
X-ray beam productionX-ray beam production
tungsten target
![Page 5: Monte Carlo model and spectral measurements of a computed tomography x-ray tube Magdalena Bazalova Frank Verhaegen.](https://reader036.fdocuments.us/reader036/viewer/2022062421/56649d705503460f94a526ac/html5/thumbnails/5.jpg)
Monte Carlo model of the Monte Carlo model of the PQ5000 x-ray tubePQ5000 x-ray tube
• using the BEAM code and using the BEAM code and manufacturer’s specifications of the manufacturer’s specifications of the x-ray tubex-ray tube
• anode material: Wanode material: W• anode angle: 7anode angle: 7°°• tube voltage: 100-140 kVptube voltage: 100-140 kVp• additional filter: 9 mm Aladditional filter: 9 mm Al
![Page 6: Monte Carlo model and spectral measurements of a computed tomography x-ray tube Magdalena Bazalova Frank Verhaegen.](https://reader036.fdocuments.us/reader036/viewer/2022062421/56649d705503460f94a526ac/html5/thumbnails/6.jpg)
Spectral measurements of Spectral measurements of diagnostic x-raysdiagnostic x-rays
• Not trivial due to the high photon flux Not trivial due to the high photon flux causing pile-up in the detector of causing pile-up in the detector of choice.choice.
• Can be solved by using very small Can be solved by using very small collimators and/or increasing the x-collimators and/or increasing the x-ray source to detector distance; the ray source to detector distance; the alignment is then detrimental.alignment is then detrimental.
• Not applicable for CT geometry and Not applicable for CT geometry and our detector, therefore, the x-ray flux our detector, therefore, the x-ray flux has to be reduced in another way.has to be reduced in another way.
![Page 7: Monte Carlo model and spectral measurements of a computed tomography x-ray tube Magdalena Bazalova Frank Verhaegen.](https://reader036.fdocuments.us/reader036/viewer/2022062421/56649d705503460f94a526ac/html5/thumbnails/7.jpg)
Problems of spectral Problems of spectral measurements in CT geometrymeasurements in CT geometry
a few meters
80 cm
• Attenuation?• Scatter!
![Page 8: Monte Carlo model and spectral measurements of a computed tomography x-ray tube Magdalena Bazalova Frank Verhaegen.](https://reader036.fdocuments.us/reader036/viewer/2022062421/56649d705503460f94a526ac/html5/thumbnails/8.jpg)
Experimental setup using Experimental setup using Compton scatterCompton scatter
![Page 9: Monte Carlo model and spectral measurements of a computed tomography x-ray tube Magdalena Bazalova Frank Verhaegen.](https://reader036.fdocuments.us/reader036/viewer/2022062421/56649d705503460f94a526ac/html5/thumbnails/9.jpg)
Compton scatter- scatterer Compton scatter- scatterer materialmaterial
• interaction of a photon of energy hinteraction of a photon of energy hνν with a loosely bound orbital electron with a loosely bound orbital electron of an absorberof an absorber
![Page 10: Monte Carlo model and spectral measurements of a computed tomography x-ray tube Magdalena Bazalova Frank Verhaegen.](https://reader036.fdocuments.us/reader036/viewer/2022062421/56649d705503460f94a526ac/html5/thumbnails/10.jpg)
Scatterer materialScatterer material
Solid line: Z=6
Dashed line: Z=50
![Page 11: Monte Carlo model and spectral measurements of a computed tomography x-ray tube Magdalena Bazalova Frank Verhaegen.](https://reader036.fdocuments.us/reader036/viewer/2022062421/56649d705503460f94a526ac/html5/thumbnails/11.jpg)
Measurements of x-ray Measurements of x-ray spectra: CdTe detectorspectra: CdTe detector
• 33×3×1 mm×3×1 mm33 CdTe crystal (Z CdTe crystal (ZCdCd=48, Z=48, ZTeTe=52)=52)
• Pt and In contactsPt and In contacts
![Page 12: Monte Carlo model and spectral measurements of a computed tomography x-ray tube Magdalena Bazalova Frank Verhaegen.](https://reader036.fdocuments.us/reader036/viewer/2022062421/56649d705503460f94a526ac/html5/thumbnails/12.jpg)
Spectrum corrections-detector Spectrum corrections-detector responseresponse
2 mm
3 mm3 mm
1 mm
![Page 13: Monte Carlo model and spectral measurements of a computed tomography x-ray tube Magdalena Bazalova Frank Verhaegen.](https://reader036.fdocuments.us/reader036/viewer/2022062421/56649d705503460f94a526ac/html5/thumbnails/13.jpg)
Spectrum corrections-Spectrum corrections-Compton effectCompton effect
![Page 14: Monte Carlo model and spectral measurements of a computed tomography x-ray tube Magdalena Bazalova Frank Verhaegen.](https://reader036.fdocuments.us/reader036/viewer/2022062421/56649d705503460f94a526ac/html5/thumbnails/14.jpg)
Spectrum corrections-peak Spectrum corrections-peak broadeningbroadening
![Page 15: Monte Carlo model and spectral measurements of a computed tomography x-ray tube Magdalena Bazalova Frank Verhaegen.](https://reader036.fdocuments.us/reader036/viewer/2022062421/56649d705503460f94a526ac/html5/thumbnails/15.jpg)
Spectrum corrections - finalSpectrum corrections - final
![Page 16: Monte Carlo model and spectral measurements of a computed tomography x-ray tube Magdalena Bazalova Frank Verhaegen.](https://reader036.fdocuments.us/reader036/viewer/2022062421/56649d705503460f94a526ac/html5/thumbnails/16.jpg)
Comparison of MC simulations Comparison of MC simulations and spectral measurementsand spectral measurements
filte
ring
kVp
![Page 17: Monte Carlo model and spectral measurements of a computed tomography x-ray tube Magdalena Bazalova Frank Verhaegen.](https://reader036.fdocuments.us/reader036/viewer/2022062421/56649d705503460f94a526ac/html5/thumbnails/17.jpg)
Half value layer comparisonHalf value layer comparison
![Page 18: Monte Carlo model and spectral measurements of a computed tomography x-ray tube Magdalena Bazalova Frank Verhaegen.](https://reader036.fdocuments.us/reader036/viewer/2022062421/56649d705503460f94a526ac/html5/thumbnails/18.jpg)
ConclusionsConclusions
• An accurate Monte Carlo model of the An accurate Monte Carlo model of the PQ5000 CT simulator x-ray was developed.PQ5000 CT simulator x-ray was developed.
• The model was validated with spectral The model was validated with spectral measurements using a CdTe detector in a measurements using a CdTe detector in a Compton scattering setup.Compton scattering setup.
• A very good agreement between the MC A very good agreement between the MC simulated and measured spectra for the simulated and measured spectra for the extreme tube voltages with and without an extreme tube voltages with and without an additional 9 mm thick aluminum filter was additional 9 mm thick aluminum filter was found.found.
• The MC model can be used for CT dose The MC model can be used for CT dose calculation.calculation.