Monitoring CO Poisoning with the Rad-57 TM

31
Monitoring CO Poisoning with the Rad-57 TM Includes a review on Carbon Monoxide Poisoning For Emergency Responders V.10 14 Dec 11

description

Monitoring CO Poisoning with the Rad-57 TM. Includes a review on Carbon Monoxide Poisoning For Emergency Responders. V.10 14 Dec 11. Important Notice. - PowerPoint PPT Presentation

Transcript of Monitoring CO Poisoning with the Rad-57 TM

Page 1: Monitoring CO Poisoning with the Rad-57 TM

Monitoring CO Poisoningwith the Rad-57TM

Includes a review on Carbon Monoxide Poisoning For Emergency Responders

V.10 14 Dec 11

Page 2: Monitoring CO Poisoning with the Rad-57 TM

Important Notice

• This program is not meant as a substitute for a program or course of study in carbon monoxide recognition and treatment, or emergency medicine. This is for review only. Please refer to your physician or local Medical Director for approved content and medical protocols.

• This program is not meant as a substitute for the manufacturer’s operator manual for the Masimo Rad-57. Please refer to the manufacturer for complete operating instructions.

Page 3: Monitoring CO Poisoning with the Rad-57 TM

Section 1

Carbon Monoxide Poisoning For Emergency Responders

Page 4: Monitoring CO Poisoning with the Rad-57 TM

Introduction• Leading cause of poisoning deaths in

industrialized countries:▪ 50,000 emergency room visits in the US annually 1

▪ At least 3,800 deaths in the US annually 2

▪ 1,400-3,000 accidental deaths in the US annually 3,4

• More frequently involved in fatal work-related inhalations than any other substance

▪ 35% of fatal inhalations 5,6

1Hampson NB, Weaver LK. Carbon Monoxide poisoning: A new incidence for an old disease. Undersea and Hyperbaric Medicine 2007;34(3):163-168.2 Mott JA, Wolfe MI, Alverson CJ, MacDonald SC, Bailey CR, Ball LB, Moorman JE, Somers JH, Mannino DM, Redd SC. National Vehicle Emissions policies and practices and declining US carbon monoxide-related mortality. JAMA 2002;288:988-9953 Hampson NB, Stock AL. Storm-Related Carbon Monoxide Poisoning: Lessons Learned from Recent Epidemics. Undersea Hyperb Med 2006;33(4):257-2634 Cobb N, Etzel RA, Unintentional Carbon monoxide-related deaths in the United States, 1979 through 1988. JAMA 1991;266(5):659.5Valent, F et al. Fatal Work-Related Inhalation of Harmful Substances in the United States. Chest. 121:969-975, 2002. 6Lofgren DJ. Occupational Carbon Monoxide Poisoning in the State of Washington, 1994-1999. Applied Occup and Envi Hyg. 17(4):286-295, 2002.

Page 5: Monitoring CO Poisoning with the Rad-57 TM

Chemistry• Colorless, Odorless, Tasteless

• Chemical Asphyxiant

• Vapor Density- 0.97▪ Atmosphere- 1.0

• Combines with hemoglobin to form carboxyhemoglobin.

• Carbon Monoxide = CO Carbon Dioxide = CO2

• CO = Pulse CO-Oximetry CO2 = Capnography

Page 6: Monitoring CO Poisoning with the Rad-57 TM

• Incomplete combustion of any carbon-based material will produce carbon monoxide.

• Common sources:

• Remember, CO is a gas. If it’s found in one area of a building, hotel, apartment, house- assume its everywhere until proven otherwise!

Common Sources

Forklifts Generators Methylene Chloride

Smoke Concrete Saws Explosives

Floor Buffers Water Heaters Welders and Torches

Furnaces Compressors Cigarettes

Motor Vehicle Exhaust Foundry Processes Tractors

Ice Resurfacer (Zamboni) Refinery Processes Pressure Washers

BBQ’s Water Pumps Space Heaters

Lofgren DJ. Occupational Carbon Monoxide Poisoning in the State of Washington, 1994-1999. Applied Occup and Envi Hyg. 17(4):286-295, 2002.Erdogan, MS et al. Occupational Carbon Monoxide Poisoning Among West Virginia Workers’ Compensation Claims: Diagnosis, Treatment, Duration, and Utilization. J Occup Envi Med. 46(6):577-583, 2004.

Page 7: Monitoring CO Poisoning with the Rad-57 TM

Pathophysiology

• Hypoxia- hemoglobin binds CO with a greater affinity than oxygen.

• Vasodilation- secondary to increased formation of Nitric Oxide• Headache, Hypotension, Syncope

• Intracellular toxicity-• Cardiac and skeletal muscle dysfunction

• Results in short term (acute) and long term (chronic) symptoms.

• The organs most oxygen dependent are the most affected: the heart and the brain.

Hampson NB. UHM 2007. 34(3):163-7.2 Hampson NB et al. Crit Care Med. 2009; 37(6): 1941-47. Bledsoe BE. Journal of Emergency Medical Service. 32:54-59, 2007. Weaver LK, et al. N Engl J Med, 2002;347(14):1057-067. Henry CR, et al. JAMA. 2006;295(4):398-402.

Page 8: Monitoring CO Poisoning with the Rad-57 TM

Neurologic Effects• Acute Symptoms:• CONFUSION• Headache• Dizziness• Seizures

• Chronic Symptoms:• Cognitive/Personality changes• Permanent seizure disorders• Learning disabilities• Early onset Alzheimer’s Disease,

Parkinson’s Disease

• Delayed Neurologic Syndrome• Can take up to 8 months before symptoms manifest

Hampson NB. UHM 2007. 34(3):163-7.2 Hampson NB et al. Crit Care Med. 2009; 37(6): 1941-47. Bledsoe BE. Journal of Emergency Medical Service. 32:54-59, 2007. Weaver LK, et al. N Engl J Med, 2002;347(14):1057-067. Henry CR, et al. JAMA. 2006;295(4):398-402.

Page 9: Monitoring CO Poisoning with the Rad-57 TM

Cardiac Effects• Acute Symptoms: • Decreased oxygen supply interrupts normal

function• Decreased oxygen to body• Irregular heart beat

• Chronic Symptoms: • Increased risk of early heart disease, heart

attack or death.

• A single exposure to carbon monoxide can result in permanent heart and/or brain damage.

• Chronic exposure to low levels of carbon monoxide can be worse than a single, large exposure!!

Hampson NB. UHM 2007. 34(3):163-7.2 Hampson NB et al. Crit Care Med. 2009; 37(6): 1941-47. Bledsoe BE. Journal of Emergency Medical Service. 32:54-59, 2007. Weaver LK, et al. N Engl J Med, 2002;347(14):1057-067. Henry CR, et al. JAMA. 2006;295(4):398-402. 5 Weaver LK, et al. N Engl J Med, 2002;347(14):1057-067.6 Henry CR, et al. JAMA. 2006;295(4):398-402.

Page 10: Monitoring CO Poisoning with the Rad-57 TM

CO Poisoning

30-50 % of CO-exposed patients presenting to Emergency Departments were misdiagnosed.

Barker MD, et al. J Pediatr. 1988;1:233-43Barret L, et al. Clin Toxicol. 1985;23:309-13Grace TW, et al. JAMA. 1981;246:1698-700

Page 11: Monitoring CO Poisoning with the Rad-57 TM

Signs and Symptoms • Carbon monoxide poisoning CANNOT be reliably diagnosed from the patient’s signs and

symptoms.

• Patients can be sick and not have any complaints.

• Symptoms are often vague and mimic other conditions.

• Common symptoms include: Flu-like symptoms Fecal/urinary incontinence

Headache Behavioral changes

Nausea/vomiting/diarrhea Syncope

Dizziness Chest pain

Fatigue Gait disturbances

Impaired judgment Seizures

Confusion Coma

Palpitations Death

Page 12: Monitoring CO Poisoning with the Rad-57 TM

Treatment Considerations • Protect yourself and your crew:

• Consider SCBA

• Remove patient(s) from poisonous environment

• Treatment for elevated CO is 100% oxygen.• With a mask or intubation, per patient

condition.

• Monitor carboxyhemoglobin level with Rad-57. • Pregnancy- fetal Hb has a stronger bond to CO

than adult Hb• A fetus does NOT tolerate hypoxia well• Consider transport, even if mother

asymptomatic

Page 13: Monitoring CO Poisoning with the Rad-57 TM

Half Life• Half-life: The amount of time required to

reduce the carboxyhemoglobin level by 50%.

• Age and health will impact the half-life of CO.

• Breathing Room Air: ~4-6 hours

High Flow Oxygen: ~40-60 minutes

• Placing patients on high flow O2 as soon as possible can significantly reduce the damage from carbon monoxide.

Page 14: Monitoring CO Poisoning with the Rad-57 TM

Atmospheric CO vs. SpCO • Gas monitors measure how much CO is in the atmosphere

at that moment in time. • It does not tell how much CO is actually inside

the patient(s). • Doors and windows may be opened prior to

arrival, ventilating the house and masking the actual level of CO patients were exposed to.

• How much is in the air and how much is in the patient are two different things.• For example- if the CO level in a patient’s house

is 30ppm but has been like that for 4 months, they may be sick.

• Treatment decisions SHOULD NEVER be based on atmospheric readings alone!• Need to factor in: atmospheric readings, SpCO level, patient’s

presentation, medical history and the duration of exposure.

Page 15: Monitoring CO Poisoning with the Rad-57 TM

Firefighter Risk

• CO poisoning is often unrecognized due to the presence of exhaustion, heat stress, illness or injuries.

• CO levels typically highest post-fire, during overhaul.• Often SCBA’s not worn during this time.

• On the fire ground, elevated CO

levels can be immediate threats to health and safety: • Impaired judgment, confusion• Fatigue• Exacerbation of pre-existing conditions

Page 16: Monitoring CO Poisoning with the Rad-57 TM

Firefighter Rehab

• NFPA 1584 suggests monitoring CO levels in FF• CO induces death secondary to V Fib • VF initial rhythm in 90% interior FF deaths• Should not leave rehab if > 5% COHb

Page 17: Monitoring CO Poisoning with the Rad-57 TM

Section 2

The RAD 57 Pulse CO-Oximeter™

Page 18: Monitoring CO Poisoning with the Rad-57 TM

Rad-57 Carry Case

• Water resistant, protective case • Extra sensor pocket • Quick Reference Guide

pocket• Pen holder• Spare battery pocket

• For maximum protection, the device MUST remain in the protective case during field use.

Page 19: Monitoring CO Poisoning with the Rad-57 TM

Sensor Placement

• Step 1: Select the appropriately sized sensor for the patient’s finger

• Step 2: Carefully insert the patient’s finger into the sensor until it reaches the digit stop.

• Step 3: Shield the sensor in environments with excessive ambient light

Optimal LED path

3 STEPS FOR PROPER SENSOR PLACEMENT:

Page 20: Monitoring CO Poisoning with the Rad-57 TM

Sensor Placement Tips• Pedi Sensor- Small/slender fingers Adult Sensor- Average to large fingers

• Sensor should be used on fingers only. Do NOT use toes or ears.

• Using the non-dominant hand, the following order is preferred: Ring, middle, index finger.

• Ensure sensor is right side up- the cable should run over the top of the hand.

• To adjust the sensor or take additional readings, remove the sensor from the finger completely for at least 5 seconds, then replace.

• INCORRECT SENSOR PLACEMENT MAY CAUSE INACCURATE SpCO READINGS!

Adult Sensor

Pedi/Slender Digit Sensor

Page 21: Monitoring CO Poisoning with the Rad-57 TM

Ambient Light Shield• Shield the sensor from intense ambient light!

• Especially strobe lights or bright sunlight. • High-intensity light can interfere with the ability to

calculate an SpCO reading.

• Boot shield• Insert the sensor all the way into the light shield;

the cable should come out of the top of the opening.

• Once sensor is in light shield, insert finger into sensor.

• Bag shield • Place sensor on finger first• Place bag over sensor and secure with Velcro

strap

• If neither type of shield is available, anything that completely covers the sensor will do- i.e. jacket, pillow, blanket. • This does NOT include your hand.

Apply pressureto open sensor

Page 22: Monitoring CO Poisoning with the Rad-57 TM

Power Up

• Power Button• Press to turn ON • Press and HOLD to turn OFF• Do NOT have to turn the device on/off in

between readings.

• Battery Life Indicator• 4 Green LED’s• Each represents ~25% battery life• Do Not use rechargeable batteries

• Battery Compartment• Located on the back of device• Holds 4 AA alkaline batteries• Provides 8-10 hours of battery life

Page 23: Monitoring CO Poisoning with the Rad-57 TM

Start Up Sequence

• Upon power up, the Rad 57 performs a self test • Display screen will cycle through alarm setting• No external calibration needed. • Take ~10 seconds

• When display shows SEN OFF, device is ready• If device fails self test, machine will alarm very loudly.

Take out the batteries and call (800) 326-4890.

• To begin monitoring, apply sensor

• Calculation/Pulse search:• Display will show scrolling zeroes followed by

dashed lines. • Can take ~20 seconds for SpO2; up to 120

seconds for SpCO.

Page 24: Monitoring CO Poisoning with the Rad-57 TM

Measuring SpO2 and Pulse Rate • Oxygen saturation (SpO2) reading on top

in RED; pulse rate on bottom in GREEN.

• SIQ- Signal Identification and Quality. Identifies the quality of the SpO2 signal.

• Press YELLOW button to silence alarms for 120 seconds.

• Press button to scroll through display screens.

• Next/Enter/Up & Down Arrows- for higher menu functions (i.e. change alarm settings or screen brightness).

Alarm Silence Button

Page 25: Monitoring CO Poisoning with the Rad-57 TM

Measuring SpCO • Press button to switch to the CO

screen.• Carboxyhemoglobin (SpCO) on top,

in %; letters ‘CO’ on bottom

• May take additional time for the SpCO reading, even if the device already shows the SpO2 and pulse rate.• Will see dashed lines on the display while

the device calculates the SpCO value.

• Confirm all elevated SpCO readings by checking sensor placement and repeating the measurement.

• If SpO2 < 90% or SpMet > 2%, SpCO may NOT display. • If SpCO reading is disrupted,

display will show dashed lines.

Page 26: Monitoring CO Poisoning with the Rad-57 TM

Sample SpCO Protocol

Page 27: Monitoring CO Poisoning with the Rad-57 TM

Measuring Perfusion Index (PI)

• Press the button to cycle to the PI parameter.

• PI = Measurement of blood flow to the extremity the sensor is on. • Similar principal as

capillary refill

• Values range from 0.02-20. The lower the number, the lower the perfusion.

• Displayed both numerically and graphically.

Perfusion Index Bar Graph

Page 28: Monitoring CO Poisoning with the Rad-57 TM

Optional Features

• Auto ON/OFF• Device can automatically power

down after 5, 10 or 15 minutes of inactivity.

• Factory Default: Device stays on until the power button is pressed.

• Perfusion Index (PI)• Digital display can be turned on or

off• PI will always display on bar graph• Factory Default: PI display is on.

• Start up screen• Upon powering on, can select

SpCO screen to display first. • Factory Default: SpO2/pulse rate

parameters will display first.

Page 29: Monitoring CO Poisoning with the Rad-57 TM

SpCO Accuracy• The accuracy of the SpCO parameter is +/- 3% for 1 standard

deviation within a range of 1 to 40%.

• Said another way, the RAD 57 will be within 3% of a blood gas 68% of the time and within 6% of a blood gas 95% of the time.

• Example: If a patient has a SpCO of 5%, their carboxyhemoglobin could be as low as 2% or as high as 8%.

• Measurements may vary from reading to reading, even on the same patient, within the accuracy specification.

• If 2 readings vary greatly (i.e.- one finger reads 3%, another finger reads 10%):• Confirm the proper sensor is being used• Confirm proper sensor placement technique• Shield the sensor from light• Repeat readings

Page 30: Monitoring CO Poisoning with the Rad-57 TM

Obtaining Accurate SpCO Measurements

• Inaccurate SpCO measurements may be caused by:• Incorrect sensor placement• Excessive ambient light • Low arterial perfusion• Patient motion• Altitude-induced hypoxemia• Low arterial oxygen saturation• Elevated levels of methemoglobin • Intravascular dyes or externally applied coloring

(such as nail polish)