ModFit Tutorial –Cell Cycle Analysis Software · sure G2M position looks good. Let the S -phase...

19
1. Start by choosing all the files to progress in batch mode. ModFit Tutorial – Cell Cycle Analysis Software NYULMC DART Cytometry and Cell Sorting ModFit Tutorial, 12/27/2016

Transcript of ModFit Tutorial –Cell Cycle Analysis Software · sure G2M position looks good. Let the S -phase...

Page 1: ModFit Tutorial –Cell Cycle Analysis Software · sure G2M position looks good. Let the S -phase options remain the default and check the auto -aggregate and debris boxes. Click

1.Startbychoosingallthefilestoprogressinbatchmode.

ModFit Tutorial– CellCycleAnalysisSoftware

NYULMC DART Cytometry and Cell Sorting ModFit Tutorial, 12/27/2016

Page 2: ModFit Tutorial –Cell Cycle Analysis Software · sure G2M position looks good. Let the S -phase options remain the default and check the auto -aggregate and debris boxes. Click

2.Setthebatchcontrolstonotanalyzebutwithfileprompton.Thisallowsyoutoopen filesandalterthegatesforanalysisbutwillnot trytofitthedata.This

letsyoujustlookattheunfittedhistogramsfirst.

Page 3: ModFit Tutorial –Cell Cycle Analysis Software · sure G2M position looks good. Let the S -phase options remain the default and check the auto -aggregate and debris boxes. Click

3.Choosetheparameterforanalysis(PI-A).

4.Selectthefirsttwogatesandcreatethemwideopen (tocoverthewholeplot).Thisissoyoucanseethegeneralpositionofpopulationsandhowtheychangebetweenfiles.Wecanchangethesegateslaterif

needed.

Page 4: ModFit Tutorial –Cell Cycle Analysis Software · sure G2M position looks good. Let the S -phase options remain the default and check the auto -aggregate and debris boxes. Click

5.WecannowsetuptheSynchWizard.

Page 5: ModFit Tutorial –Cell Cycle Analysis Software · sure G2M position looks good. Let the S -phase options remain the default and check the auto -aggregate and debris boxes. Click

It’sgood tostartbylettingtheG0G1peakpositionandSDadjustautomatically.

Page 6: ModFit Tutorial –Cell Cycle Analysis Software · sure G2M position looks good. Let the S -phase options remain the default and check the auto -aggregate and debris boxes. Click

G2Mposition looksgoodsononeedtochangethis.

Page 7: ModFit Tutorial –Cell Cycle Analysis Software · sure G2M position looks good. Let the S -phase options remain the default and check the auto -aggregate and debris boxes. Click

It’salsogoodtomaintaintheS-phasedefaultoptions tostart,whicharethesimplermodels.

Underthe‘Other’tabyoucanstartwithhavingtheDebrisandAggregateschecked.

UsuallythereisnoreasontochangeModFit’s defaultssothisiswhywewanttostartwiththemfirst.

Page 8: ModFit Tutorial –Cell Cycle Analysis Software · sure G2M position looks good. Let the S -phase options remain the default and check the auto -aggregate and debris boxes. Click

6.Thisfirstfileproducesthefollowingerror,tellingusthattheG0G1positionhasshiftedbeyondtheexpectedamountandwantsyoutotellitwhattodonext.Ifthenewvaluewereclosetotheoriginalposition(around50orso),thenyoucanletitmovetothenewposition.However,a

positionof-1doesnotmakesense,sowewillchoosetolocktheG0G1position.Oncethemodelislocked,itwillstaylockeduntilyouchangeit.

Page 9: ModFit Tutorial –Cell Cycle Analysis Software · sure G2M position looks good. Let the S -phase options remain the default and check the auto -aggregate and debris boxes. Click

• Insomeinstancesyouwillseesubsequent filesthathaveaverydistinctG0G1peakthathasshifted totherightofleftfromthislockedpositionsoyoucanclickthemodeleditbuttontobringup theeditdialogagain.

• Ifitappearstohaveshiftedbacktotheoriginal,expectedlocationthenyoucanchoosetore-selecttheAdjustautomaticallymode.Otherwise,youcanjustdragthearrowoverthenewG0G1position.

Page 10: ModFit Tutorial –Cell Cycle Analysis Software · sure G2M position looks good. Let the S -phase options remain the default and check the auto -aggregate and debris boxes. Click

6.IfthenormalcontrolforsynchexperimentshasawideCV,thenModFit willbeunabletogothroughtheSynchWizardproperlyandyouwillgethighRCSvaluesback.

Inthiscaseyouwillhavetoproceedmanually.

Page 11: ModFit Tutorial –Cell Cycle Analysis Software · sure G2M position looks good. Let the S -phase options remain the default and check the auto -aggregate and debris boxes. Click

VerityalsomentionedthatiftherenoblacktrianglespresentunderneaththeG0G1andG2MpeaksthenthisisanotherindicationofCV’sthataretoohightouseSynchWizardwith.

Page 12: ModFit Tutorial –Cell Cycle Analysis Software · sure G2M position looks good. Let the S -phase options remain the default and check the auto -aggregate and debris boxes. Click

6.YouwillhavetoexitoutofSynchWizardandclickontheMod icon.FromhereyoucantellModFit tolook forG2Masavisiblepeak.

Page 13: ModFit Tutorial –Cell Cycle Analysis Software · sure G2M position looks good. Let the S -phase options remain the default and check the auto -aggregate and debris boxes. Click

KeepAutoaggregatesanddebrison.

Page 14: ModFit Tutorial –Cell Cycle Analysis Software · sure G2M position looks good. Let the S -phase options remain the default and check the auto -aggregate and debris boxes. Click

RCSvaluewentfrom24.914downto1.958!

Page 15: ModFit Tutorial –Cell Cycle Analysis Software · sure G2M position looks good. Let the S -phase options remain the default and check the auto -aggregate and debris boxes. Click

SynchWizard

1.Openallthefilesinabatch,makethegateswideopen,openSynchWizard.

2.SetthelocationforG0G1andletitadjustthepositionautomatically(thedefault).CheckandmakesureG2Mpositionlooksgood.LettheS-phaseoptionsremainthedefaultandchecktheauto-aggregateanddebrisboxes.ClickAnalyze.

Page 16: ModFit Tutorial –Cell Cycle Analysis Software · sure G2M position looks good. Let the S -phase options remain the default and check the auto -aggregate and debris boxes. Click

3.Sometimesthesynchwizarddoesn’tdoagood jobatcalculatingaggregatesautomatically.

.

Page 17: ModFit Tutorial –Cell Cycle Analysis Software · sure G2M position looks good. Let the S -phase options remain the default and check the auto -aggregate and debris boxes. Click

Youcanfixthisbyturning offtheauto-aggregatesandmakingyourdoublet-discriminator gatetightaround thesinglecells.

Page 18: ModFit Tutorial –Cell Cycle Analysis Software · sure G2M position looks good. Let the S -phase options remain the default and check the auto -aggregate and debris boxes. Click

Bydoing this,itbrings theRCSvaluedownto2.601.

Page 19: ModFit Tutorial –Cell Cycle Analysis Software · sure G2M position looks good. Let the S -phase options remain the default and check the auto -aggregate and debris boxes. Click

4.Whengoing through thesubsequent filesinyourbatchanalysisyoucaneditthebatchproperties to‘Fitwithcurrentmodel’ soyoudonothavetoclick

analyzeeverytime.YoucanalsounchecktheDatabaseandFilepromptsowhenyouswitchthrough thefilesyouarenotprompted tochoosetheparameterand

gateseverytime.