Automatic Blood Pressure Monitor !" Model SEM-1 INSTRUCTION MANUAL === SEM-1

Transcript of Model === SEM-1 - Omron Healthcare · PDF fileAutomatic Blood Pressure Monitor!" Model SEM-1...

Page 1: Model === SEM-1 - Omron Healthcare · PDF fileAutomatic Blood Pressure Monitor!" Model SEM-1 INSTRUCTION MANUAL === S E M-1 En_TC-Cover[SEM-1]-2 Page 1 8/4/05, 4:19 PM Adobe PageMaker

Automatic Blood Pressure Monitor�� !"

Model SEM-1

INSTRUCTION MANUAL

�===�

SEM-1

En_TC-Cover[SEM-1]-2 8/4/05, 4:19 PMPage 1 Adobe PageMaker 6.5C/PPC

Page 2: Model === SEM-1 - Omron Healthcare · PDF fileAutomatic Blood Pressure Monitor!" Model SEM-1 INSTRUCTION MANUAL === S E M-1 En_TC-Cover[SEM-1]-2 Page 1 8/4/05, 4:19 PM Adobe PageMaker

Before Using the Monitor

Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3Important Safety Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4

Operating Instructions

Know Your Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6Unit Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7How to Use Your Blood Pressure Monitor . . . . . . . . . . . . . . . . . .8How to Install/Replace the Batteries . . . . . . . . . . . . . . . . . . . . . .9How to Sit Correctly When Taking a Measurement . . . . . . . . . .11How to Apply the Arm Cuff . . . . . . . . . . . . . . . . . . . . . . . . . . .13How to Take a Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . .15How to Use the Memory Function . . . . . . . . . . . . . . . . . . . . . .19

Care and Maintenance

Maintenance and Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22Using the Optional AC Adapter . . . . . . . . . . . . . . . . . . . . . . . . .24Error Indicator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25Troubleshooting Tips . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .29

CONTENTS

2

EN-02-29[SEM-1]-4 8/23/05 11:10 AM Page 2

Page 3: Model === SEM-1 - Omron Healthcare · PDF fileAutomatic Blood Pressure Monitor!" Model SEM-1 INSTRUCTION MANUAL === S E M-1 En_TC-Cover[SEM-1]-2 Page 1 8/4/05, 4:19 PM Adobe PageMaker

Before Using the Monitor

Thank you for purchasing OMRON SEM-1 Automatic Blood PressureMonitor.

INTRODUCTION

3

Your new digital blood pressure monitor uses the oscillometric method ofblood pressure measurement. This means the monitor detects yourblood’s movement through your brachial artery and converts themovements into a digital reading. An oscillometric monitor does not needa stethoscope so the monitor is simple to use.

The SEM-1 Automatic Blood Pressure Monitor is intended for home use.

SAVE THESE INSTRUCTIONS

EN-02-29[SEM-1]-4 8/23/05 11:10 AM Page 3

Page 4: Model === SEM-1 - Omron Healthcare · PDF fileAutomatic Blood Pressure Monitor!" Model SEM-1 INSTRUCTION MANUAL === S E M-1 En_TC-Cover[SEM-1]-2 Page 1 8/4/05, 4:19 PM Adobe PageMaker

To assure the correct use of the product basic safety measures shouldalways be followed including the precautions listed below:

Self-diagnosis of measured results and treatment are dangerous.Please follow the instructions of your doctor.

Do not use the unit on infants or persons who cannot express their consent.

Do not inflate the arm cuff over 299 mmHg.

Do not use the unit for any purpose other than measuring blood pres-sure.

Do not use a cellular phone near the unit. This may result in incorrectoperation of the unit.

Keep batteries out of the reach of young children and infants. If a bat-tery swallowed, consult a doctor immediately.

If the battery fluid should get in your eyes, immediately rinse with plenty of clean water. Consult a doctor immediately.

If the battery fluid should get on your skin or clothing, immediatelyrinse with plenty of clean water.

IMPORTANT SAFETY NOTES

4

EN-02-29[SEM-1]-4 8/23/05 11:10 AM Page 4

Page 5: Model === SEM-1 - Omron Healthcare · PDF fileAutomatic Blood Pressure Monitor!" Model SEM-1 INSTRUCTION MANUAL === S E M-1 En_TC-Cover[SEM-1]-2 Page 1 8/4/05, 4:19 PM Adobe PageMaker

IMPORTANT SAFETY NOTES

5

When using batteries, the battery liquid may leak and damage the main unit. Please observe the following points.

• Do not insert the batteries with their polarities incorrectly aligned.

• Replace old batteries with new ones immediately.

• When you are not going to use the unit for a long period of time(approximately three months or more), take out the batteries.

Do not use new and used batteries together. Do not use batteries of adifferent type together.

Do not forcibly bend the arm cuff or bend the air tube excessively.

When removing the air tube, pull on the air plug at the connection with the main unit not the tube itself.

Do not expose the unit to strong shocks or vibrations. Do not drop theunit.

Do not inflate the arm cuff when it is not wrapped around your arm.

Avoid eating, alcohol, smoking, exercising and bathing at least 30

minutes before taking measurement.

EN-02-29[SEM-1]-5 9/12/05 1:36 PM Page 5

Page 6: Model === SEM-1 - Omron Healthcare · PDF fileAutomatic Blood Pressure Monitor!" Model SEM-1 INSTRUCTION MANUAL === S E M-1 En_TC-Cover[SEM-1]-2 Page 1 8/4/05, 4:19 PM Adobe PageMaker

Operating Instructions

SEM-1

6

KNOW YOUR UNIT

Air Jack

Main Unit:

Battery Compartment

START/STOPButtonMEMORY Button

Display

AC Adapter Jack

Arm Cuff:

Air Tube

Air Plug

The arm cuff should last for 18 months, when used four times a day.To ensure accurate measurement results, you should change to anew arm cuff after this period.

If air starts to leak from the arm cuff, please purchase a new one.

NOTE: An air plug is not included with thereplacement arm cuff. Do not disposeof the current air plug when youpurchase a new arm cuff.

EN-02-29[SEM-1]-4 8/23/05 11:10 AM Page 6

Page 7: Model === SEM-1 - Omron Healthcare · PDF fileAutomatic Blood Pressure Monitor!" Model SEM-1 INSTRUCTION MANUAL === S E M-1 En_TC-Cover[SEM-1]-2 Page 1 8/4/05, 4:19 PM Adobe PageMaker

7

KNOW YOUR UNIT

Diastolic Blood Pressure

Systolic Blood Pressure

Memory Symbol

Pulse Display

Memory Number DisplayBattery Low Symbol

Average Value Symbol

Heartbeat Symbol

Deflation Symbol

UNIT DISPLAY

Storage Case

Components:

Four “AA” Alkaline Batteries

EN-02-29[SEM-1]-4 8/23/05 11:10 AM Page 7

Page 8: Model === SEM-1 - Omron Healthcare · PDF fileAutomatic Blood Pressure Monitor!" Model SEM-1 INSTRUCTION MANUAL === S E M-1 En_TC-Cover[SEM-1]-2 Page 1 8/4/05, 4:19 PM Adobe PageMaker

HOW TO USE YOUR BLOOD PRESSURE MONITOR

8

Preparation

Measurement

Sit correctly

Apply the arm cuff

Start measurement

View Results

Check themeasurement result

View previousmeasurement results

page 9

page 11

page 13

page 15

page 17

page 19

Measurement results automatically stored in memory (Up to 21 measurements)

Insert the batteries

EN-02-29[SEM-1]-5 9/12/05 1:36 PM Page 8

Page 9: Model === SEM-1 - Omron Healthcare · PDF fileAutomatic Blood Pressure Monitor!" Model SEM-1 INSTRUCTION MANUAL === S E M-1 En_TC-Cover[SEM-1]-2 Page 1 8/4/05, 4:19 PM Adobe PageMaker

9

1. Turn the main unit over.

2. Slide the battery cover in thedirection of the arrow while pressingthe ribbed part of the cover.

3. Install or replace four “AA” sizebatteries so that the + (positive) and - (negative) polarities match thepolarities indicated on the batterycompartment.

4. Replace the battery cover.

Slide the battery cover as indicated, until it clicks into place.

NOTE: The monitor memorizes previous measurement results even whenthe batteries are replaced.

HOW TO INSTALL/REPLACE THE BATTERIES

EN-02-29[SEM-1]-4 8/23/05 11:10 AM Page 9

Page 10: Model === SEM-1 - Omron Healthcare · PDF fileAutomatic Blood Pressure Monitor!" Model SEM-1 INSTRUCTION MANUAL === S E M-1 En_TC-Cover[SEM-1]-2 Page 1 8/4/05, 4:19 PM Adobe PageMaker

10

Battery Life & Replacement

If the battery low symbol ( ) appears on the display, replace all fourbatteries at the same time.

Even if the battery low symbol ( ) starts to blink, you will still be ableto use the unit for a short while. However, you should replace thebatteries with new ones when the symbol remains lit.

• Turn the monitor off before replacing the batteries.

• Remove the batteries if the monitor will not be used for more than 3 months.

• Dispose of batteries according to applicable local regulations.

Four new “AA” alkaline batteries will last for approximately 1500measurements, when used to take two measurements a day.

Since the supplied batteries are for monitoring use only, they may have ashorter life and not last for 1500 measurements.

HOW TO INSTALL/REPLACE THE BATTERIES

EN-02-29[SEM-1]-6 9/13/05 10:22 AM Page 10

Page 11: Model === SEM-1 - Omron Healthcare · PDF fileAutomatic Blood Pressure Monitor!" Model SEM-1 INSTRUCTION MANUAL === S E M-1 En_TC-Cover[SEM-1]-2 Page 1 8/4/05, 4:19 PM Adobe PageMaker

You can take a measurement on either your left or right arm.

NOTE: Measurements should be taken in a quiet place and you shouldbe in a relaxed, seated position. Make sure that the room is nottoo hot or cold.

Do not take a measurement after bathing, alcohol or exercise.

HOW TO SIT CORRECTLY WHEN TAKING A MEASUREMENT

11

CorrectPosture

Remove tight fitting clothing fromyour upper arm, and any thickclothing such as a sweater.Do not place the cuff over thickclothes and do not roll up your sleeveif it is too tight.

Place your arm on a table sothat the cuff will be at the samelevel as your heart.

Sit uprightwith yourback straight.

The gap between thechair and the top of thetable should be between10" to 12".

EN-02-29[SEM-1]-4 8/23/05 11:10 AM Page 11

Page 12: Model === SEM-1 - Omron Healthcare · PDF fileAutomatic Blood Pressure Monitor!" Model SEM-1 INSTRUCTION MANUAL === S E M-1 En_TC-Cover[SEM-1]-2 Page 1 8/4/05, 4:19 PM Adobe PageMaker

NOTES:• Correct posture during measurement is necessary to get accurate

results. • You should also try to measure your blood pressure at the same time

each day. (Within 1 hour of waking is recommended.)

HOW TO SIT CORRECTLY WHEN TAKING A MEASUREMENT

12

Incorrect Posture

• Arched back (leaning forwards)• Sitting cross-legged• Sitting on a sofa or at a low table

so that you tend to lean forwards

These situations could lead to higher blood pressure reading due to strain or thearm cuff being lower than the heart.

If the arm cuff is at a lower position than your heart use cushions etc., to adjustthe height of your arm.

EN-02-29[SEM-1]-4 8/23/05 11:10 AM Page 12

Page 13: Model === SEM-1 - Omron Healthcare · PDF fileAutomatic Blood Pressure Monitor!" Model SEM-1 INSTRUCTION MANUAL === S E M-1 En_TC-Cover[SEM-1]-2 Page 1 8/4/05, 4:19 PM Adobe PageMaker

Be sure to wrap the arm cuff correctly so that you get accurate results.

Measurements can be taken in light clothing. However, please removethick clothes, such as sweaters, before taking a reading.

1. Insert the air plug into the air jack on theleft side of the main unit.

2. Put your left arm through the cuff loop.

The air tube should run down the inside ofyour forearm and be in line with yourmiddle finger.

NOTE: If the cuff is not assembled, pass theend of the cuff furthest from thetubing through the metal D-ring toform a loop. The smooth cloth shouldbe on the inside of the cuff loop.

HOW TO APPLY THE ARM CUFF

13

EN-02-29[SEM-1]-4 8/23/05 11:10 AM Page 13

Page 14: Model === SEM-1 - Omron Healthcare · PDF fileAutomatic Blood Pressure Monitor!" Model SEM-1 INSTRUCTION MANUAL === S E M-1 En_TC-Cover[SEM-1]-2 Page 1 8/4/05, 4:19 PM Adobe PageMaker

3. Position the arm correctly.

1) The air tube should run down theinside of your forearm and be inline with your middle finger.

2) The bottom of the cuff should beapproximately 1 to 2 cm aboveyour elbow.

3) Apply the cuff to your upper armso that the colored marker iscentered on the middle of yourinner arm and points down theinside of the arm.

TAKING MEASUREMENTS ON THE RIGHT ARM

Apply the cuff so that the air tube is at the side of your elbow.

• Be careful not to rest your arm on theair tube, or otherwise restrict the flowof air to the cuff.

• The cuff should be 1 to 2 cm abovethe elbow.

4. When the cuff is positioned correctly, close the fabric fastenerFIRMLY.

HOW TO APPLY THE ARM CUFF

14

1 to 2 cm

EN-02-29[SEM-1]-4 8/23/05 11:10 AM Page 14

Page 15: Model === SEM-1 - Omron Healthcare · PDF fileAutomatic Blood Pressure Monitor!" Model SEM-1 INSTRUCTION MANUAL === S E M-1 En_TC-Cover[SEM-1]-2 Page 1 8/4/05, 4:19 PM Adobe PageMaker

15

1. Press the START/STOP button to turn on the monitor.

Sit in a chair with your feet flat on the floor and place your arm on atable so that the cuff will be at the same level as your heart. Keep stilland do not talk during measurement.

NOTE: To cancel a measurement, press the START/STOP button toturn off the unit and to release the air in the arm cuff.

HOW TO TAKE A MEASUREMENT

CorrectPosture

Do not talk duringthe measurement.

Arm cuff atheart level.

Sit uprightwith yourback straight.

EN-02-29[SEM-1]-4 8/23/05 11:10 AM Page 15

Page 16: Model === SEM-1 - Omron Healthcare · PDF fileAutomatic Blood Pressure Monitor!" Model SEM-1 INSTRUCTION MANUAL === S E M-1 En_TC-Cover[SEM-1]-2 Page 1 8/4/05, 4:19 PM Adobe PageMaker

2. When the measurement is complete, the arm cuff completely deflatesand your blood pressure and pulse rate are displayed.

NOTE: The monitor automatically stores blood pressure and pulserate into its memory.

HOW TO TAKE A MEASUREMENT

16

Common Causes of Inaccurate Results

Arm cuff is too loose. Movement or talkingduring measurement.

Leaning forwards.

NOTE: If the arm cuff is at a lower positionthan your heart use cushions etc., toadjust the height of your arm.

EN-02-29[SEM-1]-4 8/23/05 11:10 AM Page 16

Page 17: Model === SEM-1 - Omron Healthcare · PDF fileAutomatic Blood Pressure Monitor!" Model SEM-1 INSTRUCTION MANUAL === S E M-1 En_TC-Cover[SEM-1]-2 Page 1 8/4/05, 4:19 PM Adobe PageMaker

17

3. Check the measurement results.

The monitor automatically storesblood pressure and pulse rate into itsmemory. Refer to “How to Use theMemory Function” on page 19.

NOTES:

• Self-diagnosis of measured results and treatment are dangerous.Please follow the instructions of your doctor.

• If an error symbol “E” or “EE” appears, measurement was notcompleted successfully. Refer to “Error Indicator” on page 25.

• Wait 2-3 minutes before taking another blood pressure measurement.Waiting between readings allows the arteries to return to thecondition prior to taking the blood pressure measurement.

4. Remove the arm cuff.

5. Press the START/STOP button to turn the monitor off.

NOTE: If you forget to turn the monitor off, it will automatically shutitself off after five minutes.

HOW TO TAKE A MEASUREMENT

Systolic BloodPressure

Pulse Display

Diastolic BloodPressure

Guide to Blood Pressure LevelsRecent research suggests that the following values can be used as aguide to high blood pressure for measurements taken at home.

Use these figures as a guide. * Consult your doctor if necessary.

Systolic Blood Pressure: Above 135 mmHgDiastolic Blood Pressure: Above 85 mmHg

EN-02-29[SEM-1]-4 8/23/05 11:10 AM Page 17

Page 18: Model === SEM-1 - Omron Healthcare · PDF fileAutomatic Blood Pressure Monitor!" Model SEM-1 INSTRUCTION MANUAL === S E M-1 En_TC-Cover[SEM-1]-2 Page 1 8/4/05, 4:19 PM Adobe PageMaker

INSTRUCTIONS FOR SPECIAL CONDITIONS

Instructions for special conditions If your systolic pressure is known tobe more than 220 mmHg, press and hold the START/STOP button untilthe arm cuff inflates 30 to 40 mmHg higher than your suspected systolicpressure.

1. Press the START/STOP button to turnthe unit on.

Measurement starts.

2. When the cuff starts to inflate, press the START/STOP button andkeep it pressed until the pressure is 30 to 40 mmHg higher than yoursuspected systolic pressure.

NOTE: You cannot inflate the cuff above 299 mmHg. (An error willbe displayed if you try to inflate the cuff above 300mmHg.)

3. Release the START/STOP button when the cuff has been inflated tothe desired pressure. The cuff starts to deflate and measurement starts.

4. The rest of the procedure is the same as for normal measurement.Refer to steps 2 to 4 on page 16 to 17.

NOTE: Do not apply more pressure than necessary.

HOW TO TAKE A MEASUREMENT

18

EN-02-29[SEM-1]-4 8/23/05 11:10 AM Page 18

Page 19: Model === SEM-1 - Omron Healthcare · PDF fileAutomatic Blood Pressure Monitor!" Model SEM-1 INSTRUCTION MANUAL === S E M-1 En_TC-Cover[SEM-1]-2 Page 1 8/4/05, 4:19 PM Adobe PageMaker

This monitor has a memory capable of storing 21 sets of readings. Everytime you complete the measurement, the monitor automatically stores theblood pressure and pulse rate. The monitor also calculates an averagereading based on the measurements from the last three readings.

NOTE: When 21 sets of readings are stored in memory, memory number21 will be deleted to store a new set. (The most recent reading isset to number 1.)

1. Press the MEMORY button.

The average values based on the three mostrecent measurements is displayed.

NOTE: If there are no measurementsresults stored in memory, thescreen to the right is displayed.

HOW TO USE THE MEMORY FUNCTION

19

EN-02-29[SEM-1]-4 8/23/05 11:10 AM Page 19

Page 20: Model === SEM-1 - Omron Healthcare · PDF fileAutomatic Blood Pressure Monitor!" Model SEM-1 INSTRUCTION MANUAL === S E M-1 En_TC-Cover[SEM-1]-2 Page 1 8/4/05, 4:19 PM Adobe PageMaker

2. Press the MEMORY button again, whilethe average reading screen is displayed.

The result from the most recentmeasurement is displayed.

The memory number of the stored readingis also displayed. (The most recentmeasurement is set to number 1, with theoldest being number 21.)

3. Press the MEMORY button repeatedly tocycle through the previous measurementresults.

Keep the button pressed down to cyclerapidly through the previous results.

4. Press the START/STOP button to turn off the power.

NOTE: If you forget to turn the monitor off, it will automatically shutitself off after five minutes.

HOW TO USE THE MEMORY FUNCTION

20

EN-02-29[SEM-1]-4 8/23/05 11:10 AM Page 20

Page 21: Model === SEM-1 - Omron Healthcare · PDF fileAutomatic Blood Pressure Monitor!" Model SEM-1 INSTRUCTION MANUAL === S E M-1 En_TC-Cover[SEM-1]-2 Page 1 8/4/05, 4:19 PM Adobe PageMaker

TO DELETE ALL THE READING

You cannot delete individual stored readings, all the readings will bedeleted.

1. To delete stored measurement results,while the memory values aredisplayed, first press the MEMORYbutton. Then while holding it down,press the START/STOP buttonsimultaneously for about 2-3 seconds.All results will then be deleted.

NOTE: Be careful not to press theSTART/STOP button first.If the START/STOP button ispressed first, the monitor will beturned off.

2. Press the START/STOP button to turn the monitor off.

If you forget to turn the monitor off, it will automatically shut itself offafter five minutes.

HOW TO USE THE MEMORY FUNCTION

21

1

2

EN-02-29[SEM-1]-4 8/23/05 11:10 AM Page 21

Page 22: Model === SEM-1 - Omron Healthcare · PDF fileAutomatic Blood Pressure Monitor!" Model SEM-1 INSTRUCTION MANUAL === S E M-1 En_TC-Cover[SEM-1]-2 Page 1 8/4/05, 4:19 PM Adobe PageMaker

Care and Maintenance

MAINTENANCE

To keep your digital blood pressure monitor in the best condition andprotect the unit from damage follow the directions listed below:

• Do not subject your monitor and cuff to extreme temperatures,humidity, moisture or direct sunlight.

• Do not fold the cuff or tubing tightly.

• Do not inflate the monitor over 299 mmHg.

• Do not disassemble the monitor.

• Do not subject the monitor to strong shocks or vibrations (for example,dropping the monitor on the floor).

• Do not use volatile liquids to clean the monitor. The monitor should becleaned with a soft, dry, cloth.

• Do not wash the arm cuff or immerse it in water.

MAINTENANCE AND STORAGE

22

EN-02-29[SEM-1]-4 8/23/05 11:10 AM Page 22

Page 23: Model === SEM-1 - Omron Healthcare · PDF fileAutomatic Blood Pressure Monitor!" Model SEM-1 INSTRUCTION MANUAL === S E M-1 En_TC-Cover[SEM-1]-2 Page 1 8/4/05, 4:19 PM Adobe PageMaker

STORAGE

Keep the unit in its storage case when not in use.

1. Unplug the air tube from the air jack.

2. Gently fold the air tube into the armcuff.

NOTE: Do not bend the air tubeexcessively.

3. Place the arm cuff and the main unit inthe storage case.

Do not store the monitor in the following situations:

• If the unit is wet.

• Locations exposed to extreme temperatures, humidity, direct sunlight,dust or corrosive vapours.

• Locations exposed to vibrations, shocks or where it will be at an angle.

• Locations exposed to chemicals or corrosive vapours.

MAINTENANCE AND STORAGE

23

EN-02-29[SEM-1]-4 8/23/05 11:10 AM Page 23

Page 24: Model === SEM-1 - Omron Healthcare · PDF fileAutomatic Blood Pressure Monitor!" Model SEM-1 INSTRUCTION MANUAL === S E M-1 En_TC-Cover[SEM-1]-2 Page 1 8/4/05, 4:19 PM Adobe PageMaker

Insert batteries into the battery compartment even when using the ACadapter.

NOTES:

• Never unplug the power cord with wet hands.

• Use only exclusive AC adaptor specified by Omron dealers. Use ofunsupported adapters may damage the unit.

• When storing the optional AC adapter in the storage case with themain unit, be careful not to damage the main unit or cuff.

1. Insert the AC adapter plug into theAC adapter jack on the rear side ofthe main unit.

2. Plug the AC adapter into anelectrical outlet.

To disconnect the AC adapter, unplug the AC adapter from the electricaloutlet first and then immediately remove the AC adapter plug from themain unit.

USING THE OPTIONAL AC ADAPTER

24

EN-02-29[SEM-1]-4 8/23/05 11:10 AM Page 24

Page 25: Model === SEM-1 - Omron Healthcare · PDF fileAutomatic Blood Pressure Monitor!" Model SEM-1 INSTRUCTION MANUAL === S E M-1 En_TC-Cover[SEM-1]-2 Page 1 8/4/05, 4:19 PM Adobe PageMaker

ERROR INDICATOR

25

Release your finger from theSTART/STOP button before thepressure reaches 299 mmHg,when inflating the cuff manually.Refer to page 18.

The arm cuff wasinflated above 299mmHg when inflatingthe cuff manually.

Carefully read and repeat thesteps listed under “How to Applythe Arm Cuff” on page 13.

Cuff is under inflated.

Replace the arm cuff with a newone.

Air is leaking fromthe arm cuff.

Remove any clothing interferingwith the arm cuff.

Clothing is interferingwith the arm cuff.

SYMBOL CAUSE CORRECTION

Movement duringmeasurement.

Repeat measurement. Remainstill and do not talk duringmeasurement. Refer to page 15.

Air plugdisconnected.

Insert the air plug securely.Refer to page 13.

Arm cuff not appliedcorrectly.

Apply the arm cuff correctly.Refer to page 13.

EN-02-29[SEM-1]-4 8/23/05 11:10 AM Page 25

Page 26: Model === SEM-1 - Omron Healthcare · PDF fileAutomatic Blood Pressure Monitor!" Model SEM-1 INSTRUCTION MANUAL === S E M-1 En_TC-Cover[SEM-1]-2 Page 1 8/4/05, 4:19 PM Adobe PageMaker

ERROR INDICATOR

26

Device error.Contact your OMRON servicerepresentative.

Batteries are worn.Replace all four “AA” batterieswith new ones.Refer to page 9.

SYMBOL CAUSE CORRECTION

Blinks or appearscontinuously

EN-02-29[SEM-1]-4 8/23/05 11:10 AM Page 26

Page 27: Model === SEM-1 - Omron Healthcare · PDF fileAutomatic Blood Pressure Monitor!" Model SEM-1 INSTRUCTION MANUAL === S E M-1 En_TC-Cover[SEM-1]-2 Page 1 8/4/05, 4:19 PM Adobe PageMaker

Cannot measure orreadings are too lowor too high.

Has the arm cuffbeen inflatedsufficiently?

Inflate the cuff so that itis 30 to 40 mmHg aboveyour previousmeasurement result.Refer to page 19.

Replace the arm cuffwith a new one.

Air is leaking fromthe arm cuff.

The reading isextremely low (orhigh).

Arm cuff not appliedcorrectly.

Apply the arm cuffcorrectly. Refer to page 13.

Movement or talkingduring measurement.

Remain still and do nottalk duringmeasurement. Refer to page 15.

Clothing isinterfering with thearm cuff.

Remove any clothinginterfering with the armcuff. Refer to page 11.

Arm cuff pressuredoes not rise.

Is the air plugsecurely inserted intothe main unit?

Make sure that the airplug is connectedsecurely. Refer to page 13.

The arm cuff isloose.

Apply the cuff correctlyso that it is firmlywrapped around thearm. Refer to page 13.

Arm cuff deflates toosoon.

REMEDY

27

TROUBLESHOOTING TIPS

PROBLEM CAUSES

EN-02-29[SEM-1]-4 8/23/05 11:10 AM Page 27

Page 28: Model === SEM-1 - Omron Healthcare · PDF fileAutomatic Blood Pressure Monitor!" Model SEM-1 INSTRUCTION MANUAL === S E M-1 En_TC-Cover[SEM-1]-2 Page 1 8/4/05, 4:19 PM Adobe PageMaker

Nothing happenswhen you press thebuttons.

TROUBLESHOOTING TIPS

28

Replace all four “AA”batteries with new ones.Refer to page 9.

The batteries areempty.

The unit loses powerduring measurementor the date and timeare reset.

The batteries areworn or have beenremoved for a longtime.

Replace all four “AA”batteries with new ones.Refer to page 9.

The batteries havebeen inserted withtheir polaritiesincorrectly aligned.

Check the batteryinstallation for properplacement of the batterypolarities.

Press the START/STOP button and repeatmeasurement.If the problem continues, try replacing thebatteries with new ones.

If this still does not solve the problem, contactyour OMRON service representative.

Other problems.

REMEDYPROBLEM CAUSES

EN-02-29[SEM-1]-4 8/23/05 11:10 AM Page 28

Page 29: Model === SEM-1 - Omron Healthcare · PDF fileAutomatic Blood Pressure Monitor!" Model SEM-1 INSTRUCTION MANUAL === S E M-1 En_TC-Cover[SEM-1]-2 Page 1 8/4/05, 4:19 PM Adobe PageMaker

29

SPECIFICATIONS

Name: OMRON Automatic Blood Pressure MonitorModel: SEM-1Display: LCD Digital DisplayMeasurement Range: Pressure: 0 to 299 mmHg, Pulse: 40-180 beats/min.Accuracy/Calibration: Pressure: ±3mmHg or 2% of reading

Pulse: ±5% of readingInflation: Automatic by electric pumpDeflation: Automatic pressure release valveRapid Air Release: Automatic exhaust valvePressure Detection: Capacitive pressure sensorMeasurement Method: Oscillometric methodPulse Wave Detection: Capacitance type pressure sensorPower Source: 4 “AA” batteries (or AC Adapter)Battery Life: Approx. 1500 uses when used twice a day

with new “AA” alkaline batteriesOperating Temperature 10ºC to 40ºC /30 to 85% RH/Humidity:Storage Temperature -20ºC to 60ºC /10 to 95% RH/Humidity:Main Unit Weight: Approximately 340 g not including batteriesMain Unit Dimensions: Approx. 121 mm (w) × 86 mm (h) × 141 mm (l)Cuff Size: Approximately 146 mm (w) × 446 mm (l)

Cuff tube 600 mm (l)Cuff Circumference: Fits arm circumferences 220mm to 320mmContents: Main Unit, Arm Cuff, Storage Case,

Four “AA” Alkaline Batteries, Instruction Manual

NOTE: These specifications are subject to change without notice.

EN-02-29[SEM-1]-5 9/12/05 1:36 PM Page 29