User’s Manual Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: S2] IM 11M10B01-01E IM 11M10B01-01E 9th Edition

Transcript of Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s...

Page 1: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

User’sManual Model OX400

Low Concentration Zirconia Oxygen Analyzer[Style: S2]

IM 11M10B01-01E

IM 11M10B01-01E9th Edition

Page 2: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

i

IM 11M10B01-01E

<INTRODUCTION>

INTRODUCTIONThank you for purchasing the OX400 Low Concentration Zirconia Oxygen Analyzer.Please read the following respective documents before installing and using the OX400.The related documents are as follows. GeneralSpecifications:GS11M10B01-01E User’sManual: IM11M10B01-01E(thismanual)

* the “E” in the document number is the language code.

AnexclusiveUser’sManualmightbeattachedtotheproductswhosesuffixcodesoroptioncodescontainthecode“Z”(madetocustomers’specifications).Pleasereaditalongwiththismanual.

n Safety,Protection,andModificationoftheProduct• In order to protect the system controlled by the product and the product itself and ensure

safe operation, observe the safety precautions described in this user’s manual. We assume no liability for safety if users fail to observe these instructions when operating the product.

• Ifthisinstrumentisusedinamannernotspecifiedinthisuser’smanual,theprotectionprovided by this instrument may be impaired.

• If any protection or safety circuit is required for the system controlled by the product or for the product itself, prepare it separately.

• BesuretousethesparepartsapprovedbyYokogawaElectricCorporation(hereaftersimplyreferredtoasYOKOGAWA)whenreplacingpartsorconsumables.

• Modificationoftheproductisstrictlyprohibited.• The following symbols are used in the product and user’s manual to indicate that there are

precautionsforsafety:

n NotesonHandlingUser’sManuals• Please hand over the user’s manuals to your end users so that they can keep the user’s

manuals on hand for convenient reference.• Please read the information thoroughly before using the product.• The purpose of these user’s manuals is not to warrant that the product is well suited to any

particular purpose but rather to describe the functional details of the product.• No part of the user’s manuals may be transferred or reproduced without prior written

consent from YOKOGAWA.• YOKOGAWA reserves the right to make improvements in the user’s manuals and product at

any time, without notice or obligation.• Ifyouhaveanyquestions,oryoufindmistakesoromissionsintheuser’smanuals,please

contact our sales representative or your local distributor.

n Warning and DisclaimerThe product is provided on an “as is” basis. YOKOGAWA shall have neither liability nor responsibility to any person or entity with respect to any direct or indirect loss or damage arising from using the product or any defect of the product that YOKOGAWA can not predict in advance.

Media No. IM 11M10B01-01E 9thEdition:Nov.2019(YK)AllRightsReservedCopyright©2009,YokogawaElectricCorporation

9thEdition:Nov.20,2019-00

Page 3: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

ii

IM 11M10B01-01E

<INTRODUCTION>

n Notes on Hardware

l Appearance and AccessoriesCheckthefollowingwhenyoureceivetheproduct:

• Appearance• Standard accessories

Contactoursalesrepresentativeoryourlocaldistributoriftheproduct’scoatinghascomeoff,ithas been damaged, or there is shortage of required accessories.

l ModelandSuffixCodesThenameplateontheproductcontainsthemodelandsuffixcodes.Comparethemwiththoseinthegeneralspecificationtomakesuretheproductisthecorrectone.Ifyouhaveanyquestions,contact our sales representative or your local distributor.

n Symbol MarksThroughoutthisuser’smanual,youwillfindseveraldifferenttypesofsymbolsareusedtoidentifydifferentsectionsoftext.Thissectiondescribestheseicons.

WARNINGIdentifiesinstructionsthatmustbeobservedinordertoavoidphysicalinjuryandelectricshockordeath of the operator.

CAUTIONIdentifiesinstructionsthatmustbeobservedinordertopreventthesoftwareorhardwarefrombeing damaged or the system from becoming faulty.

NOTEIdentifiesimportantinformationrequiredtounderstandoperationsorfunctions.

9thEdition:Nov.20,2019-00

Page 4: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

iii

IM 11M10B01-01E

<INTRODUCTION>

n AboutUniqueRepresentationsUsedinthisOperationManualWhenoperationkeys,contentsdisplayedonthedisplay,andlampdisplaysarespecificallydescribed in the text or anywhere else in this operation manual, in principle, they are represented in the following ways.

• Operation key Representedby[]. Example:[ENT]key• Display contents on the display Representedby“”. Example:Maindisplay--> “HEAt” Example:Sub-display--> “CAL” Example:Datadisplay-->“10.00”(intheONstate) Example LED lamp --> PPM (intheONstate), PPM (intheOFFstate)• Graphicalrepresentationoftheflashingstate Represented by italic characters or the mark.

Flashingstate 1.000 Flashing state of decimal point

1.000• Representation of digital characters TheOX400employsa7-segmentalphanumericdigitaldisplay.Forthecorrespondence

between the display characters on the display and alphanumeric characters for explanation, seeFigure5.2.

n Operation ParametersThe OX400 is shipped with default parameters as shown in Table 8.1. Change them according to thepurposeofuse.Forhowtochangeparameters,seeChapter6,“OperationandParameters.”Itis recommended to write down the changed operation parameters as user set values as shown in Table 8.1.

n Notes on UseThe OX400 is a product that conforms to the general safety requirements of the IEC standard. Be sure to observe the following precautions when you operate it.

HandlingPrecautions• Installation location

The OX400 is structurally non-explosion proof so you cannot use it in an explosive atmosphere. Also, see 3.1, “Installation Location.”

• Power supply BesuretocheckthatthepowersupplyvoltagespecificationoftheOX400matchesthevoltage of the power supply before turning on the power.

• Protective ground Be sure to connect the power plug of the OX400 to the 3P socket with a protective ground pin in order to prevent electric shock.

• Fuse Besuretouseadesignatedfuseinordertopreventafire.Besuretoturnoffthepowerbefore replacing the fuse. Never use a fuse holder other than a designated one.

• Removing cover ThereisaheatedareainsidetheOX400,andtouchingitdirectlymaycauseaburninjury.Never remove the cover except to replace the sensor.

• Whenopeningthecovertoreplacethesensor,turnoffthepowerswitch,removethepowerplugfromthesocket,andwaitfor1.5hoursormorebeforeopeningthecover.

9thEdition:Nov.20,2019-00

Page 5: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

iv

IM 11M10B01-01E

<INTRODUCTION>

• Before connecting wires to the terminal block on the rear panel of the OX400, remove the powerplugfromthesocketinordertopreventelectricshock.Afterfinishingthewiring,secure the removed special terminal cover with screws.

• The OX400 is a measuring instrument intended to be installed indoors, so do not install it in a location that is exposed to direct sunlight, rain, and wind.

• DonotusetheOX400inalocationthatmaybesubjecttovibrationsandimpact.Doingsomay cause destruction and damage to the internal sensor.

• Do not apply a pressure of 30 kPaG or more to the gas inlet. Doing so may cause destruction of the suction pump and damage to the sensor. Be sure to reduce the sample gas pressure to the atmospheric pressure level before introducing the gas.

• Be sure to use metal piping for the gas piping; particularly when handling oxygen concentration in 1 vol % or less. Use of piping materials such as polyethylene, vinyl, rubber, andplasticmaycausesignificanterrorsinmeasurementsbecauseoftheirlargeoxygentransmission and absorption rates.

• Thepresenceofcorrosivegascomponents(H2S, SOx, HCl, NH3,HF,andthelike)orpotentiallytoxicelements(Si,Sn,Cd,Te,As,P,andthelike)inthesamplegasmaycausedeteriorationofthesensor.Besuretoremovethemwithanactivatedcarbonfilterorthelikein the previous step of the OX400 before introducing the gas.

• The presence of combustible gas in the sample gas may cause errors in measurements because oxygen in the sample gas will be consumed by combustion. Be sure to remove it withafilterorthelikeinthestepbeforeintroducingthesamplegasintotheOX400.

• Makesurethetemperatureofthesamplegasis50°Corless.• Be careful because the presence of water droplets in the sample gas may cause damage to

the sensor.• Keepthesupplygasflowrateandpressureasconstantaspossiblewhileintroducingthe

gas.• Keep the sample gas outlet open to the atmosphere during operation. If a gas line must

be used to discharge gas, use a gas line with a connector that has the largest possible diameter in order to prevent back pressure.

• Do not use the supplied power cord with another device. • Powercablesuffixcodesof“-D”canbeusedinJapan,becauseanothercabledoesn’t

conform to PSE marking.

n ProductDisposalThe instrument should be disposed of in accordance with local and national legislation/

regulations.

9thEdition:Nov.20,2019-00

Page 6: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

v

IM 11M10B01-01E

<INTRODUCTION>

u After-sales Warrantyn Donotmodifytheproduct.

n Duringthewarrantyperiod,forrepairunderwarrantycarryorsendtheproducttothelocalsalesrepresentativeorserviceoffice.Yokogawawillreplaceorrepairanydamagedpartsandreturntheproducttoyou.Beforereturningaproductforrepairunderwarranty,provideuswiththemodelnameandserialnumberandadescriptionoftheproblem.Anydiagramsordataexplainingtheproblemwouldalsobeappreciated.l If we replace the product with a new one, we won’t provide you with a repair report.l Yokogawa warrants the product for the period stated in the pre-purchase quotation

Yokogawashallconductdefinedwarrantyservicebasedonitsstandard.Whenthecustomer site is located outside of the service area, a fee for dispatching the maintenance engineer will be charged to the customer.

n Inthefollowingcases,customerwillbechargedrepairfeeregardlessofwarrantyperiod.• Failureofcomponentswhichareoutofscopeofwarrantystatedininstructionmanual.• Failurecausedbyusageofsoftware,hardwareorauxiliaryequipment,whichYokogawa

Electric did not supply.• Failureduetoimproperorinsufficientmaintenancebyuser.• Failureduetomodification,misuseoroutside-of-specificationsoperationwhichYokogawa

does not authorize.• Failureduetopowersupply(voltage,frequency)beingoutsidespecificationsorabnormal.• Failurecausedbyanyusageoutofscopeofrecommendedusage.• Anydamagefromfire,earthquake,stormsandfloods,lightning,disturbances,riots,warfare,

radiation and other natural changes.

n Yokogawadoesnotwarrantconformancewiththespecificapplicationattheusersite.Yokogawawillnotbeardirect/indirectresponsibilityfordamageduetoaspecificapplication.

n YokogawaElectricwillnotbearresponsibilitywhentheuserconfigurestheproductintosystemsorresellstheproduct.

n Maintenanceserviceandsupplyingrepairpartswillbecoveredforfiveyearsaftertheproductionends.Forrepairforthisproduct,pleasecontactthenearestsalesofficedescribedinthisinstructionmanual.

9thEdition:Nov.20,2019-00

Page 7: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

vi

IM 11M10B01-01E

<INTRODUCTION>

u CEmarkingproductsn AuthorizedRepresentativeinEEA

TheAuthorizedRepresentativeforthisproductinEEAisYokogawaEuropeB.V.(Euroweg2,3825HDAmersfoort,TheNetherlands).

n IdentificationTagThismanualandtheidentificationtagattachedonpackingboxareessentialpartsoftheproduct.Keep them together in a safe place for future reference.

n UsersThis product is designed to be used by a person with specialized knowledge.

9thEdition:Nov.20,2019-00

Page 8: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

vii

IM 11M10B01-01E

<CONTENTS>

Model OX400Low Concentration Zirconia Oxygen Analyzer[Style: S2]

CONTENTS

IM11M10B01-01E9thEdition

INTRODUCTION.......................................................................................................iu After-sales Warranty....................................................................................vu CEmarkingproducts..................................................................................vi1. Outline....................................................................................................... 1-12. Specifications........................................................................................... 2-1

2.1 StandardSpecifications...................................................................................2-12.2 Characteristics..................................................................................................2-32.3 ModelandSuffixCodes...................................................................................2-42.4 StandardAccessories......................................................................................2-42.5 ExternalDimensions........................................................................................2-52.6 PipingDiagram................................................................................................2-102.7 WiringDiagram................................................................................................2-11

3. Installation................................................................................................. 3-13.1 InstallationLocation.........................................................................................3-13.2 HowtoInstall.....................................................................................................3-2

3.2.1 InstallingDesktopType ......................................................................3-23.2.2 InstallingPanelMountType ...............................................................3-33.2.3 MountingActivatedCarbonFilter.......................................................3-4

4. PipingandWiring..................................................................................... 4-14.1 Piping.................................................................................................................4-14.2 Wirings...............................................................................................................4-2

4.2.1 PowerSupplyWiring..........................................................................4-24.2.2 SignalWiring ......................................................................................4-34.2.3 Communication ..................................................................................4-6

5. NamesandFunctions.............................................................................. 5-15.1 FrontPanel........................................................................................................5-15.2 RearPanel..........................................................................................................5-3

6. OperationandParameters...................................................................... 6-16.1 StartupandSettings.........................................................................................6-2

6.1.1 SettingOutputRange“rnG” ...............................................................6-26.1.2 SettingSecondaryOutput“oUt2” ......................................................6-46.1.3 SettingHOLDFunction“HoLd” ..........................................................6-46.1.4 SettingBurnoutFunction“nAMU” ......................................................6-5

9thEdition:Nov.20,2019-00

Page 9: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

viii

IM 11M10B01-01E

<CONTENTS>

6.1.5 SettingAlarms(OxygenConcentrationHigh/LowAlarms)“ALM” .....6-66.1.6 SettingCalibrationGasConcentration“SEt.C” .................................6-76.1.7 SettingSensorConstant“SEnS” .......................................................6-76.1.8 SettingOutputSmoothing“SMoo” .....................................................6-86.1.9 SettingMulti-Selector“MLS” ..............................................................6-86.1.10 CheckingCalibrationCoefficient“CoEF” ...........................................6-86.1.11 DisplayingCellResistanceValue“CEL.r” .........................................6-96.1.12 DisplayingSoftwareRevision“rEV” ...................................................6-9

6.2 Calibration“CAL”...........................................................................................6-106.3 Communication...............................................................................................6-126.4 MeasurementGasSamplingUsingAspirator.............................................6-146.5 SwitchingMeasurementFlowPathUsingMulti-Selector..........................6-156.6 NumericalValue,DecimalPoint,andUnitInputOperation.......................6-16

7. InspectionandMaintenance................................................................... 7-17.1 RoutineInspectionandMaintenance.............................................................7-17.2 InspectionintheEventofanError.................................................................7-1

7.2.1 InspectionintheEventofanAlarm(ALM) ........................................7-27.2.2 InspectionintheEventofanError(ERR)..........................................7-2

7.3 HowtoReplaceSensor....................................................................................7-37.3.1 Removing Sensor .............................................................................. 7-37.3.2 InstallingSensor ................................................................................7-5

7.4 HowtoReplaceActivatedCarbonFilter........................................................7-77.5 HowtoReplacetheLineFilter.........................................................................7-8

8. ParameterSettings................................................................................... 8-1CustomerMaintenancePartsList......................................CMPL11M10B01-02ERevision Information...............................................................................................i

9thEdition:Nov.20,2019-00

Page 10: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

<1. Outline> 1-1

IM 11M10B01-01E

1. OutlineThe OX400 is a highly accurate and reliable low concentration zirconia oxygen analyzer that is capable of measuring a wide range of concentrations, from 0-10 ppm up to 0-100 vol% O2. This is the latest oxygen analyzer from Yokogawa, and its development was based on the company’s long experience and strong track record with this technology.Aproprietarynewthin-filmdepositiontechnologywasusedinthezirconiasensorthatcreatesamolecular bond between the zirconia element and the platinum layer. This prevents separation, enables a reduction in sensor size and ensures a high-speed response and long life.The OX400 can be used to control and monitor various semiconductor applications, and to control environment, air leakage into inert gas, and other processes.

Features

Longlifeandhigh-speedresponse• ThankstotheuseofYokogawa’sproprietarynewthin-filmdepositiontechnology,thesensor

has three times the lifespan of those used in our earlier products.• A cylindrical sensor design facilitates the replacement of measurement gases, thereby

helping to assure a high-speed response.

Highperformanceandhighreliability• Superior repeatability and linearity even at low oxygen concentrations• Either pump or aspirator sampling can be selected, depending on the application.

Built-infunctionsandavarietyofself-diagnosisfunctions• Comes with multi selector, auto range, partial range,andpumpon/offfunctions• A variety of self-diagnosis functions are provided that detect malfunctions such as heater

temperature error, temperature sensor burnout, and sensor resistance value error.

Superiormaintainability• The sensor can be replaced on-site.• Compact and lightweight for easy installation.

Applications• Oxygenconcentrationcontrolinsemiconductor-relateddiffusionanddryingfurnacesandin

LCD manufacturing processes• Oxygenconcentrationcontrolinsolderpotflowandre-flowovens,andgloveboxesusedin

electronics manufacturing, and in gas production processes• Oxygen concentration measurements to prevent dust explosions during powder transfer

9thEdition:Nov.20,2019-00

Page 11: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

Blank Page

Page 12: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

<2.Specifications> 2-1

IM 11M10B01-01E

2. Specifications2.1 StandardSpecificationsMeasurementobject

:Oxygenconcentrationsininertgasescontainingnoflammablegas,silica,corrosivegas,orliquid(includingwatervapor).

Measurementsystem:Zirconiasystem

Samplingmethod: Pump,aspirator,ornosuctiondevice. Forthesamplinggaspressurein each sampling method, refer to the sampling gas conditions.

Pumpandaspiratorsuctionflowrate:Approx.1.0L/min.Aspiratorsuctionconditions:AirorN2,supplypressure65to100kPaG,totaldischargeflow10L/minmax.(whengasinletandoutletareatambientatmosphericpressure).

Sample gas conditionsFlowrate :200±25mL/min(onlyappliestosensor)Temperature :0-50°C(non-condensing)Humidity :Non-condensingPressure :Pumpandaspirator: 0-300PaG

Nosuctiondevice: 40kPaGmax

Measurementrange: 0-10ppmO2 to 0-100 vol% O2.

Resolution: 0.01ppmO2.

Display:4digitLED.Maindisplay:O2concentration(autoswitching).Subdisplay:Parameteroralarm/errornumber

Unit :%,ppm.

OutputrangeAuto :0-10ppm,0-100ppm,0-1000ppm,0-1%,0-10%,0-100%(default) Other:0-0 ppm, 0-00 ppm, 0-000 ppm, 0-%, 0-0% can be set, isanintegerfrom2to9.Fixed :Setto0-10ppm,0-100ppm,0-1000ppm,0-1%,0-10%,or0-100%.Partial :Lowervalueoruppervalueofrangecanbeset. Note:Span(uppervalue-lowervalue)is20%FSormoreofabovefixedrange. Example:200-400ppmwhenfixedrangeis0-1000ppm,20-40ppmwhenfixedrangeis

0-100 ppm.Analogoutput: 2outputs

Primary :4to20mADC(maximumloadresistance:550Ω) Secondary :Selectfrom0-1,0-5,0-10VDC(loadresistance:10kΩorgreater)

Contactoutput: 3 outputs

Error contact, O2 concentration high/low alarm contact, range marker output

9thEdition:Nov.20,2019-00

Page 13: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

<2.Specifications> 2-2

IM 11M10B01-01E

Multiselector(optional) :Contactoutputforswitchingsamplegasflow,measumentflowinformationcontactoutput. Note:Fordetails,seeexternaldimensions.

ContactoutputspecificationsGeneral :RelayNominalcontactcapacity: 2A30VDC,2A240VAC(120VACfor100Vpowersupply)forresistiveloadMaximumpower:60W,480VAMaximumvoltage:30VDC,264VAC(132VACfor100Vpowersupply)Maximumcurrent:2ADC/AC

Contactinput:Voltage-freecontact,1point:RemoteswitchingforsamplegassuctionpumpON/OFF:Inputsignal: Opensignal: 100kΩormore

Closesignal: 200Ωorless

Self-diagnostics Error(failure) :Sensoruniterror,heatertemperatureerror,temperaturesensordisconnection,

device temperature error, CPU error, fan stop.Alarm(warning) :Heaterunstable,sensordefect,electromotiveforceabnormal,

asymmetry voltage error, calibration error, sensor resistance error, O2 concentration upper/lower, over range.

SerialcommunicationComm.signal :RS-232,oneway Baudrate :38,400bps Data(ASCII) :O2 concentration, unit, alarm/error

Calibrationmethods:(1)3point: 10ppm,1000ppm,Air(2)2point: zeroandspancalibrationmaybesetfreely(3)1point:(4)Aircalibration

Warm-uptime:Within20min

Powersupply:Powersupply :100-120VAC/200-240VAC,50/60HzAcceptablerange:100to120VAC±10%,50/60Hz 200to240VAC±10%,50/60Hz

Powerconsumption:100to120VAC,200VAmax.200to240VAC,400VAmax.

Dimensions :213(W)x132(H)xapprox.375(D)mm

Weight:Approx.5kg

Finish:Polyestercoating

Line connectionGasinlet:Rc1/4or1/4NPTfemale

9thEdition:Nov.20,2019-00

Page 14: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

<2.Specifications> 2-3

IM 11M10B01-01E

Gasoutlet:Rc1/4or1/4NPTfemaleElectrical connection

Externaloutputterminals:M3screwContactinput/outputterminals:M3screwSerialcommunication:D-sub9pinconnectorGround:withinpowercordconnector

Environment and operational conditionsInstallationconditions: Indoors,panelorwallmounting,nonexplosionarea Ambienttemperature: 0to40°C,non-condensing Ambienthumidity: 5to85%RH Storagetemperature: -5to50°C

Safety, EMC, and RoHS conformity standardsSafety:EN61010-1 CAN/CSA-C22.2No.61010-01,ULStd.No.61010-1EMC: EN61326-1ClassA,Table2(*1) EN61326-2-3,EN61000-3-2,EN61000-3-3 EMCRegulatoryArrangementinAustraliaandNewZealand(RCM)EN61326-1ClassA KCMarking:KoreaElectromagneticConformityStandard*1: Influenceofimmunityenvironment(CriteriaA):±20%ofF.S.Note1:ThisinstrumentisaClassAproduct,anditisdesignedforuseintheindustrialenvironment.Pleaseusethisinstrumentinthe

industrial environment only.Note2:Thecurrentorvoltagesignaloutputcablelengthandcontactinputcablelengthmustbenolongerthan30mforCEmarking.

RS232connectioncablelengthmustbenolongerthan3m.

Installationaltitude: 2000morlessCategorybasedonIEC61010: II(Note)PollutiondegreebasedonIEC61010: 2(Note)Note:Installationcategory,calledover-voltagecategory,specifiesimpulsewithstandvoltage.CategoryIIisforelectricalequipment. Pollution degree indicates the degree of existence of solid, liquid, gas or other inclusions which may reduce dielectric strength.Pollutiondegree2isthenormalindoorenvironment.

RoHS: EN50581

2.2 CharacteristicsRepeatability

:±1%FS(Hereafter,either10,100,1000ppm,1%,10%,or100%O2isFS)Linearity

:±2%FS ±3%FS(0-100ppmorless)

Response time :90%response:Within10sec(0-1%ormore) :Within30sec(lessthan0-1%)

Drift :±2%FS/week

9thEdition:Nov.20,2019-00

Page 15: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

<2.Specifications> 2-4

IM 11M10B01-01E

2.3 ModelandSuffixCodes [Style: S2]

Model SuffixCode Option Code Description

OX400 .............................. ................... Low Concentration Zirconia Oxygen AnalyzerPower supply(*1)

-5-3

...................

...................100-120VAC200-240VAC

Sampling method -P-A-N

...................

...................

...................

Built-in pumpWith aspiratorNo suction device

Line connection RT

...................

...................Rc 1/41/4 NPT

User’s manual -J-E

...................

...................JapaneseEnglish

Powercable(*1) -D-F-H-Q-R-T

...................

...................

...................

...................

..................

...................

UL/CSAcable(2m)VDEcable(2.5m)GBcable(2.5m)BScable(2m)SAAcable(2.5m)BSMIcable(2m)

Option Mounting hardwareMulti selector function

Filter

/P/MS/A

Panel mountMulti selector functionActivatedcarbonfilter(*2)

*1: Powercableoftwo-polewithearthingplugisattached. Suffixcode“-D”and“-T”ofpowercablecannotbespecifiedwhen“-3”ofpowersupplyisspecified. Powercableof“-D”canbeusedinJapan,becauseanothercabledoesn’tconformtoPSEmarking. Powercableof“-F”canbeusedinKorea,becauseanothercabledoesn’tconformtoKCmarking. Power cable of “-T” can be used in Taiwan, because another cable doesn’t conform to BSMI marking.*2: When“R”oflineconnectionisspecified,K9643KHfilter(Rc1/4)isattached,when“T”oflineconnectionisspecified,K9643KJ

filter(1/4NPT)isattached. “/A”isspecifiedwhenitisusedforflowfurnaceorreflowfurnace.

2.4 StandardAccessoriesCheck the standard and optional accessories when you receive the product.Accessories

Item Partno.andrating Qt’yFuse A1113EF:Time-lagT3.15conformedtoIEC60127 1User’s manual IM11M10B01-01E(English),-01(Japanese) 1Aspirator kit K9643KA(Rc1/4),K9643KB(1/4NPT) OptionalPanel mount kit K9643KC OptionalActivatedcarbonfilter K9643KH(Rc1/4),K9643KJ(1/4NPT) Optional

ConsumablesItem Partno.andrating Qt’y

Activatedfilterelementkit(15timesreplacement) K9643KK 1Filterkit(5timesreplacement) K9643KL 1Sensorassembly(includingO-ring) K9643KG 1Snapring(retainer) Y9011EV 1(*1)Plate K9213FB 1Filter K9643FB 1Snap ring plier K9643ZE 1

*1: Qt’yof10piecesormorecanbepurchased.

9thEdition:Nov.20,2019-00

Page 16: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

<2.Specifications> 2-5

IM 11M10B01-01E

2.5 ExternalDimensionsPanelmounttypewithbuilt-inpumpornosuctiondevice(OX400----/P)

135

+2 0

101.

6±0.

3

240 ±1

225 ±1

19.33302010 to 20

66

132

101.

6

260

238

Seal

Ventilationholes

FrameFrame

Screw

4xM5 screw

Panel Cutout

Mount

Note: Pump ON/OFF switch; when no suction device [-N] is specified for the sampling method, this switch is not installed.

For the notes on panel mounting, see page 6.

Unit:mm

9thEdition:Nov.20,2019-00

Page 17: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

<2.Specifications> 2-6

IM 11M10B01-01E

Panelmounttypewithaspirator(OX400--A--/P)

40 to 50

(Line connection : in case of Rc1/4)

(Line connection : in case of Rc1/4)

(Line connection : in case of 1/4NPT)

(Line connection : in case of 1/4NPT)

Mount

Ventilationholes

Bushing

Bushing

Aspirator

Aspirator

Connector

4xM5 screw

Panel Cutout

60 to 70

101.

6±0.

3

135+2 0

240 ±1

225 ±1

19.33302010 to 20

FrameFrame

Screw

Seal

66

132

101.

6

260

238

For the notes on panel mounting, see page 6.

Unit:mm

9thEdition:Nov.20,2019-00

Page 18: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

<2.Specifications> 2-7

IM 11M10B01-01E

Desktoptypewithbuilt-inpumpornosuctiondevice(OX400----)

10 to 20213

132

20

350 19.3

Rear

Note: Pump ON/OFF switch; when no suction device [-N] is specified for the sampling method, this switch is not installed.

For the notes on panel mounting, see page 6.

With multi selector function35.3

40

Sample gas inlet

Sample gas outlet

46

Unit:mm

9thEdition:Nov.20,2019-00

Page 19: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

<2.Specifications> 2-8

IM 11M10B01-01E

Desktoptypewithaspirator(OX400--A--)

With multi selector function

Rear

40 to 5060 to 70

(Line connection : in case of 1/4NPT)

(Line connection : in case of 1/4NPT)

(Line connection : in case of Rc1/4)

(Line connection : in case of Rc1/4)

BushingBushing

AspiratorAspirator

Connector

10 to 20213

132

20

350 19.3

Sample gas inlet

Aspirator gas inlet (Φ6)Aspirator and sample gas outlet (Φ8)

35.3

4640

For the notes on panel mounting, see page 6.

Unit:mm

9thEdition:Nov.20,2019-00

Page 20: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

<2.Specifications> 2-9

IM 11M10B01-01E

Activatedcarbonfilter

(Rc1/4 or 1/4NPT)

(Rc1/4 or 1/4NPT)

110

K9643KH: Rc1/4K9643KJ: 1/4NPT

± 0.5 ± 0.5

Unit: mm

Sample gas outlet

2-Ф6.5 holes

2-M5 screws

Mark “A”

In case of K9643KJ

Activated carbon filter of option code “/A”

Sample gas inlet

Fixing holes

more than 200

Maintenance spaceThe filter holder including piping connection can be moved upward to replace a filter element.

more than200

Maintenance spaceThe filter holder including piping connection can be moved downward to replace a filter element.

110 150 130

47

29.5

2515

10

(157)

35

35

60

47

9thEdition:Nov.20,2019-00

Page 21: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

<2.Specifications> 2-10

IM 11M10B01-01E

2.6 PipingDiagram

PP

Flow meter

Suction pump

Heater

Oxygensensor

IN

OUT

IN

OUT

Oxygensensor

With aspirator (Sampling method [-A])

Built-in pump (Sampling method [-P])

Filter1

Filter2

RestrictionFilter1

Filter2

Aspirator

Sample gas

Sample gas

Pipe connection

Rc1/4 or

1/4NPT

Air or N2 supply gas

inlet ø6 connector

ø8 connector

(Operable from the front panel)

Note: If no suction device [-N] is specified, the aspirator is removed from this diagram.

Flow meter

Heater

Restriction(Operable from the front panel)

Throttlevalve

Throttlevalve

9thEdition:Nov.20,2019-00

Page 22: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

<2.Specifications> 2-11

IM 11M10B01-01E

2.7 WiringDiagram

MS1

MS2

MS3

MSI1

MSI2

MSICOM

MSCOM

1st flow path

2nd flow path

3rd flow path

OX400 Rear Terminal (M3 screw)

The following terminals are added for the multi-selector “/MS” option. The customer needs to supply a switching device and carry out the necessary wiring.

Contact outputfor switching ofmeasurement flowpath

Contact output for measurement flow path data

Common

Common

Common

PUMPOFF

mA

G

V

RC2

RC3

RCCOM

RC1

FAIL

DOPump ON/OFFContact input(available when suffix code of with built-in pump “-P” specified)

Secondry output

(0-1/0-5/0-10 V DC)

Power cord (two-pole plug with earthing contact)

Primary output

(4 to 20 mA DC)

+

-

+

-

Alarm(Oxygen concentration high/low)Contact output

Error (Fail)Contact output

Range marker Contact output

(*2) (*3)

(*1)

*1: Use the earthing contact of power cord to ground to earth. Use the supplied power cord only.*2: Ground the measurement output signal line shield on the receiving side. The G-terminal is connected to a ground pin. Use this if the line shield cannot be grounded on the receiving side. Be very careful not to ground the line at two points.*3: The signal output and contact input cable must be no longer than 30 m for CE marking. RS232 cable must be no longer than 3 m.

(NOT using)

9thEdition:Nov.20,2019-00

Page 23: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

Blank Page

Page 24: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

<3. Installation> 3-1

IM 11M10B01-01E

3. Installation

3.1 InstallationLocationThe OX400 is a measuring instrument intended to be installed indoors. Install and operate it in a location that meets the following conditions in order to ensure the best performance. (1) Locationwherethereisnocorrosivegas.(2) Indoorlocationwherethereisnomechanicalvibration.(3) Locationthatisnotexposedtodirectsunlightandradiantheat.(4) Locationthatisfreeofdustanddirtparticles.(5) Locationwitharoomtemperatureof0to40°Candnocondensation.

WARNINGNeverusetheOX400inanexplosionhazardousarea.Doingsomaycauseafire,explosion,orthe like.

• The OX400 is a measuring instrument intended to be used indoors. Do not install it in a location that is exposed to direct sunlight, wind, and rain.

• DonotusetheOX400inalocationthatmaybesubjecttovibrationsandimpact.Doingsomay cause destruction and damage to the internal sensor.

• Do not apply a pressure of 30 kPaG or more to the gas inlet. Doing so may cause destruction of the suction pump and damage to the sensor. Be sure to reduce the sample gas pressure to the atmospheric pressure level before introducing it.

• Be sure to use metal piping for the gas piping; particularly when handling oxygen concentration in 1 vol % or less. Use of piping materials such as polyethylene, vinyl, rubber, andplasticmaycausesignificanterrorsinmeasurementsbecauseoftheirlargeoxygentransmission and absorption rates.

• Thepresenceofcorrosivegascomponents(H2S, SOx, HCl, NH3,HF,andthelike)orpotentiallytoxicelements(Si,Sn,Cd,Te,As,P,andthelike)inthesamplegasmaycausedeteriorationofthesensor.Besuretoremovethemwithanactivatedcarbonfilterorthelikein the previous step of the OX400 before introducing the gas.

• The presence of combustible gas in the sample gas may cause errors in measurements because oxygen in the sample gas will be consumed by combustion. Be sure to remove it withafilterorthelikeinthepreviousstepofintroducingthesamplegasintotheOX400.

• Makesurethetemperatureofthesamplegasis50°Corless.• Be careful because the presence of water droplets in the sample gas may cause damage to

the sensor. • Keepthesupplygasflowrateandpressureasconstantaspossiblewhileintroducingthe

gas. • Keep the sample gas outlet open to the atmosphere during operation. If a gas line must

be used for discharge gas, use a gas line with a connector that has the largest possible diameter in order to prevent back pressure.

9thEdition:Nov.20,2019-00

Page 25: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

<3. Installation> 3-2

IM 11M10B01-01E

3.2 HowtoInstall3.2.1 InstallingDesktopType

PlaceandoperatetheOX400onalevelsurfaceasshowninFigure3.1.(1) Provideadistanceof100mmormorebehindtheOX400inordertonotblocktheoutletof

the cooling fan on the rear panel. (2) TheairinletofthecoolingfanislocatedonthebottompaneloftheOX400.Besureto

provideadistancefromtheinstallationsurface(desktop)largerthanthatoftheheightofthelegs of the OX400.

20

Hot air is discharged from the air outlet on the rear panel of the OX400. Maintain at least 100 mm of free space around the OX400 to ensure adequate ventilation.

Notes on installation

Cooling fan

Unit: mm

Do not block the air inlet.

Distance of 20 or more

Air is discharged from the cooling fan

Distance of 100 or more

Figure3.1 NotesonInstallingDesktopType

9thEdition:Nov.20,2019-00

Page 26: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

<3. Installation> 3-3

IM 11M10B01-01E

3.2.2 InstallingPanelMountType(1) AttachapanelmountframetothesideoftheOX400,insertitintothepanel,andsecurely

screw it to the panel. (2) Provideadistanceof100mmormorebehindtheOX400inordertonotblocktheoutletof

the cooling fan on the rear panel. (3) TheairinletofthecoolingfanislocatedonthebottompaneloftheOX400.Besureto

provideadistanceof25mmormorefromthebottompanelinordertonotblockthebottompanel of the OX400.

135

+2 0

101.

6±0

.3

240 ±1

225±1

19.33302010 to 20

66

132

101.

6

260

238

Cooling fan

Unit: mm

Do not block

the air inlet.

Air is discharged from the cooling fan

Distance of 100 or more

Seal

FrameFrame

Screw

4xM5 screwPanel Cutout

Mount

3. Make sure the panel is at least 2 mm thick.

2. Maintain at least 100 mm of free space around the measuring instrument in order to ensure adequate ventilation.

1. Make sure the bottom supports do not block the ventilation outlet on the bottom panel of the measuring instrument.

Notes on mounting

Distance of 25 or more

Figure3.2 NotesonMountingPanelMountType

9thEdition:Nov.20,2019-00

Page 27: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

<3. Installation> 3-4

IM 11M10B01-01E

3.2.3 MountingActivatedCarbonFilterMountthefilteronapanelorwall.TofixthefilterwithM5screws.Whenreplaceafilterpack,200mm or more of maintenance space is necessary. If necessary, the piping connection should be removable.

(Rc1/4 or 1/4NPT)

(Rc1/4 or 1/4NPT)

110

K9643KH: Rc1/4K9643KJ: 1/4NPT

± 0.5 ± 0.5

Unit: mm

Sample gas outlet

2-ø6.5 holes

2-M5 screws

Mark “A”

In case of K9643KJ

Activated carbon filter of option code “/A”

Sample gas inlet

Fixing holes

more than 200

Maintenance spaceThe filter holder including piping connection can be moved upward to replace a filter element.

more than200

Maintenance spaceThe filter holder including piping connection can be moved downward to replace a filter element.

110 150 130

47

29.5

2515

10

(157)

35

35

60

47

Figure3.3 MountingActivatedCarbonFilter

NOTETheactivatedcarbonfilterusedinthisequipmenthavelimitedlife,andperiodicmaintenanceor replacement will be required. Install this equipment in an accessible location to facilitate maintenance.Maintenance conditions vary depending on the gas conditions. Replacement intervals for the filterandactivatedcarbonshouldbedeterminedinaccordancewiththeoperatingconditions.•Activatedcarbonfilterlifeisabout50hours.(ifmeasuredgascontainsisopropylalcoholofconcentration500ppm,andflowis500mL/min.)•Incaseofusingthebuilt-inpumporwiththeaspirator,theactivatedcarbonfilterlifeisabout25

hours.

9thEdition:Nov.20,2019-00

Page 28: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

<4. Piping and Wiring> 4-1

IM 11M10B01-01E

4. PipingandWiring

4.1 PipingBe sure to observe the following precautions when connecting the gas pipe to the OX400.(1) TheconnectionsforboththegasinletandoutletareRc1/4or1/4NPT.Usethespecified

thread and securely connect the gas pipe so that no leakage will occur.

WARNINGWhen screwing in the pipe, be sure to hold the inlet hexagonal part in place with a wrench or the like. Not doing so and using a strong force when screwing in the pipe may cause the thread on the OX400 to rotate, resulting in damage to the internal pipe.

(2) Withrespecttopiping,useametalpipe.Useofmaterialssuchasplastic,vinyl,rubber,andthe like may result in inaccurate measurements due to the transmission of oxygen from the air and absorption onto the inside surface of the pipe. Particularly with respect to silicon tube, be careful because due to its large oxygen transmission rate, accurate measurements cannot be performed in the low concentration range.

(3) Becarefulofleakagefromthepipebecauseitmaycausemeasurementerrors.Particularlyin the low concentration range, take great care because even though the pressure inside thepipeispositive,oxygenmayflowfromtheairduetodiffusion,resultinginalargeerror.

(4) Fluctuationsofbackpressureatthepipeoutletmaycausemeasurementerrors,soalwayskeep the pressure at the atmosphere pressure level as much as possible during operation.

PP

Flow meter

Suction pump

Heater

Oxygensensor

IN

OUT

IN

OUT

Oxygensensor

With aspirator (Sampling method [-A])

Built-in pump (Sampling method [-P])

Filter1

Filter2

RestrictionFilter1

Filter2

Aspirator

Sample gas

Sample gas

Pipe connection

Rc1/4 or

1/4NPT

Air or N2 supply gas

inlet ø6 connector

ø8 connector

(Operable from the front panel)

Note: If no suction device [-N] is specified, the aspirator is removed from this diagram.

Flow meter

Heater

Restriction(Operable from the front panel)

Throttlevalve

Throttlevalve

Figure4.1 PipingFlowDiagram

9thEdition:Nov.20,2019-00

Page 29: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

<4. Piping and Wiring> 4-2

IM 11M10B01-01E

!INLET

OUTLET

N200

C US

R

172608

MS1

RC1

RC2

RC3

DO

FAIL

MS2

MS3

MSI1

MSI2

MSICOM

SERIAL(RS232)

FUSE250V T3.15A

PUMPOFF

MSCOM RCCOM

+

+

-

-

mA

V

Terminal screws are M3

Primary output terminal

Secondary output terminal

Range marker contact output terminal

Contact input(with pump “-P” option)

G

NOT using

Power cord connector fortwo-pole plug with earthing contact

Error (Fail) contact output terminal

Alarm (DO) contactoutput terminal(high/low oxygen concentration alarm)

Multi-selector terminal(with “/MS” option)Contact output for switching of measurement flow path

Contact output for measurement flowpath data

RS232 communication connector

Figure4.2 PipingforSampleGas

4.2 Wirings4.2.1 PowerSupplyWiring

CAUTIONForthepowersupply,besuretousetheratedvoltageshownontherearpanel.ConnectiontoapowersupplywithadifferentvoltagemaycausedamagetotheOX400.Powercordconnectorwith a protective ground pin is used for power supply wiring for the OX400. Be sure to securely insert the connector of supplied power supply cord to the OX400, and connect the plug to a 3P socket with a protective ground pin.

WARNINGBefore connecting wires, be sure to remove the power plug from the socket and check that the poweristurnedoffinordertopreventelectricshock.Furthermore,afterfinishingthewiring,besure to secure the special terminal cover with screws.

• Do not use the supplied power cord with another device.

9thEdition:Nov.20,2019-00

Page 30: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

<4. Piping and Wiring> 4-3

IM 11M10B01-01E

MS1

MS2

MS3

MSI1

MSI2

MSICOM

MSCOM

1st flow path

2nd flow path

3rd flow path

OX400 Rear Terminal (M3 screw)

The following terminals are added for the multi-selector “/MS” option. The customer needs to supply a switching device and carry out the necessary wiring.

Contact outputfor switching ofmeasurement flowpath

Contact output for measurement flow path data

Common

Common

Common

PUMPOFF

mA

G

V

RC2

RC3

RCCOM

RC1

FAIL

DOPump ON/OFFContact input(available when suffix code of with built-in pump “-P” specified)

Secondry output

(0-1/0-5/0-10 V DC)

Power cord (two-pole plug with earthing contact)

Primary output

(4 to 20 mA DC)

+

-

+

-

Alarm(Oxygen concentration high/low)Contact output

Error (Fail)Contact output

Range marker Contact output

(*2) (*3)

(*1)

*1: Use the earthing contact of power cord to ground to earth. Use the supplied power cord only.*2: Ground the measurement output signal line shield on the receiving side. The G-terminal is connected to a ground pin. Use this if the line shield cannot be grounded on the receiving side. Be very careful not to ground the line at two points.*3: The signal output and contact input cable must be no longer than 30 m for CE marking. RS232 cable must be no longer than 3 m.

(NOT using)

Figure4.3 ExternalWiringDiagram

4.2.2 SignalWiringTable 4.1 shows the signal assignments of terminals on the rear panel. Perform wiring as necessary. Itisrecommendedtouseaninsulationsleevecrimpterminal(forM3screws)forcabletermination.

9thEdition:Nov.20,2019-00

Page 31: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

<4. Piping and Wiring> 4-4

IM 11M10B01-01E

Table4.1 ListofInputandOutputSignals

Type of Signal Terminal Marking Description Screw Wiring Limit

Primary output terminal mA + - 4-20mAcurrentoutput M3 (*1)Secondary output terminal V + - 0-1,0-5,0-10VDCvoltageoutput M3 (*1)Range marker terminal RC1,RC2,RC3,

RCCOMForautorangerecognition M3

Contact input terminal PUMPOFF ForpumpremoteON/OFF M3 (*1)Error contact output terminal FAIL Outputintheeventofanerror(ERR) M3Alarm contact output terminal

DO Output in the event of an oxygen concentrationalarm(ALM7)

M3

Contact output terminal for multi-selector

MS1,MS2,MS3,MSCOMMSI1,MSI2,MSICOM

Formeasurementflowpathswitching(1to3)Forinformationonflowpathundermeasurement

M3

*1: Thecurrentorvoltagesignaloutputcablelengthmustbenolongerthan30mforCEuse.

(1) Primaryoutputterminal Thisisaterminaltooutputoxygenconcentrationata4-20mADCcurrent.Useitataload

resistanceof550Ωorless.Thewireshallbeshielded,andtheshieldshallbeconnectedtothe ground on the receive side and open on the OX400 side.

(2) Secondaryoutputterminal This is a terminal to output oxygen concentration in terms of voltage. Specify in advance

oneof0-1,0-5,and0-10VDC.Useitataloadresistanceof10kΩormore.Thewireshallbe shielded, and the shield shall be connected to the ground on the receive side and open on the OX400 side.

(3) Rangemarkeroutput Range recognition information such as the auto range is expressed as a 3-bit code. This is a

non-voltage contact output signal.• ”1”whenlinesbetweenRC1,RC2,andRC3terminalsandRCCOMareclosed,and“0”

whenopen.RangerecognitioninformationisasshowninTable4.2.Thecontactcapacityis2A/30VDC.

Table4.2 OutputCorrespondenceTableofMeasurementRangeandMarker

MeasurementRange RC3 RC2 RC10-10ppm(Note) 0 0 10-100 ppm(Note) 0 1 00-1000 ppm(Note) 0 1 10-1%(Note) 1 0 00-10%(Note) 1 0 10-100% 1 1 0(MANUALfixedrange) 1 1 1(Partialrange) 0 0 0

(4) Contactinputterminal Thisisaterminaltoremotelyturnonandofftheinternalpumpusinganexternalcontact.

This is enabled only when the pump switch on the front panel of the OX400 is ON. The pumpisOFFwhenthecontactisclosed,andONwhenthecontactisopen.

Nothing happens if the OX400 is not equipped with a pump.

(Note)Rangeofautorangecanbechangedasfollows.

0-20ppm 0-30 ppm ····· 0-90 ppm

0-200ppm 0-300 ppm ····· 0-900 ppm

0-2000ppm 0-3000 ppm ····· 0-9000 ppm

0-2% 0-3% ····· 0-9%

0-20% 0-30% ····· 0-90%

9thEdition:Nov.20,2019-00

Page 32: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

<4. Piping and Wiring> 4-5

IM 11M10B01-01E 9thEdition:Nov.20,2019-00

(5) Errorcontactoutputterminal Ifanerror(ERRx)occurs,thecontactcloses.(6) Alarmcontactoutputterminal Ifanoxygenconcentrationhigh/lowlimitalarm(ALM7)occurs,thecontactcloses.(7) Contactoutputterminalforflowpathswitchingbymulti-selector(option) Thisterminalisaddedonlywhentheoption“/MS”isspecified.• Outputterminalformeasurementpathflowswitching:Thisisacontactoutputtoswitch

valvesforsamplingflowpathsbypaneloperationoftheOX400.Youcanselectandturnononeofthethreeflowpaths.“1”whenlinesbetweenMS1,MS2,MS3terminalsandMSCOMareclosed,and“0”whenopen.Fordetails,seeSection6.5.

Table4.3 MeasurementFlowPathSwitchingOutputbyMulti-selector

MS1-MSCOM MS2-MSCOM MS3-MSCOMMSOFF 0 0 0MS1ON(Selectflowpath1) 1 0 0MS2ON(Selectflowpath2) 0 1 0MS3ON(Selectflowpath3) 0 0 1

• Outputofinformationonflowpathundermeasurementbymulti-selector:Answer-backoutputoftheflowpaththatisbeingmeasured.”1”whenlinesbetweenterminalsMSI1/MSI2and MSICOM are closed, and “0” when open.

Table4.4 OutputofInformationonFlowPathunderMeasurementbyMulti-selector

MSI1-MSICOM MSI2-MSICOMMSOFF 0 0MS1(Flowpath1undermeasurement) 1 0MS2(Flowpath2undermeasurement) 0 1MS3(Flowpath3undermeasurement) 1 1

CAUTIONBecarefulbecausethecontactoutputisopenwhenthepoweroftheOX400isOFF.

Page 33: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

<4. Piping and Wiring> 4-6

IM 11M10B01-01E

!INLET

OUTLET

N200

C US

R

172608

MS1

RC1

RC2

RC3

DO

FAIL

MS2

MS3

MSI1

MSI2

MSICOM

SERIAL(RS232)

FUSE250V T3.15A

PUMPOFF

MSCOM RCCOM

+

+

-

-

mA

V

Terminal screws are M3

Primary output terminal

Secondary output terminal

Range marker contact output terminal

Contact input(with pump “-P” option)

G

NOT using

Power cord connector fortwo-pole plug with earthing contact

Error (Fail) contact output terminal

Alarm (DO) contactoutput terminal(high/low oxygen concentration alarm)

Multi-selector terminal(with “/MS” option)Contact output for switching of measurement flow path

Contact output for measurement flowpath data

RS232 communication connector

Figure4.4 WiringTerminals

4.2.3 CommunicationAD-sub9-pinconnectorforRS232communicationislocatedontherearpaneloftheOX400.RS232connectioncablelengthmustbenolongerthan3mforCEmarking.

Page 34: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

<5. Names and Functions> 5-1

IM 11M10B01-01E

5. NamesandFunctions

5.1 FrontPanel1

2

3

4

56

89

10

11 127

Figure5.1 DescriptionofFrontPanel

1 Main Display The main display digitally displays the oxygen concentration and set value in four digits with

adecimalpoint.Thedisplayflashesatone-secondintervalswhilewaitingforkeyinput.2 Oxygen Concentration Unit LED The unit of displayed oxygen concentration is displayed by LED. The LED is not on when a

non-oxygenconcentrationvalueisdisplayed.Theunityouselectedflashesuntilyouconfirmit.

3 MODE LED The LED turns on when you push the MODE switch to enable the maintenance mode. The

LEDisoffinthemeasurementmode.4 ERR/ALM LED The event of an error causes the LED to turn on. The event of an alarm causes the LED to

flash.Ifanerrorandalarmoccuratthesametime,theerrorisgivenpriorityandtheLEDturnson.TheLEDisoffduringnormaloperation.

5 Sub-Display The sub-display displays the description of the currently displayed item or menu item in four

digits with a decimal point. If an error or alarm occurs, the error or alarm code corresponding to the content is displayed. “HEAt” is displayed during the warm-up period.

6 Flowmeter Theflowmeterusesafloatertoindicatetheflowrateofgasflowingintothesensor.Always

setthevalueto200±25mL/min(afloatershouldbebetweenupperandlowerbarofthe200mL/minbaronflowmeter).

7 MODE Key Holddownthekeyfor2secondstoswitchfromthemeasurementmodetothemaintenance

mode(orfromthemaintenancemodetothemeasurementmode).8 Arrow Keys

9thEdition:Nov.20,2019-00

Page 35: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

<5. Names and Functions> 5-2

IM 11M10B01-01E

[◄],[►]:Pressthekeystomovethroughthedigitsofthenumbertobeset.[▲],[▼]:Pressthe keys to scroll through the numbers or items to be set.

9 SETTING/ENTER Key Pressthekeytoconfirmtheitemornumbertobeset.Press[ENT]todisplaytheoperation

description. 10 GasFlowRateAdjustmentKnob Usetheknobtoadjustthegasflowrateofgasflowingintothesensor.

11 PUMP Switch Turntheswitchontousethebuilt-inpump.ThisswitchiseffectiveonlywhenthePOWER

switch is on. 12 POWER Switch

This is the power switch of the OX400. TheLEDdisplayemploysa7-segmentalphanumericdisplay.Figure5.2showsthedisplaycharacters and alphanumeric characters in the display.

LED Display LED Display LED Display

9

8

7

6

5

4

3

2

1

0

Z

Y

X

W

V

U

T

S

R

Q

P

O

N

M

L

K

J

I

H

G

F

E

D

C

B

A

Alphanumerics Alphanumerics Alphanumerics

Figure5.2 LEDDisplayandAlphanumericCharacters

9thEdition:Nov.20,2019-00

Page 36: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

<5. Names and Functions> 5-3

IM 11M10B01-01E

5.2 RearPanel

!INLET

OUTLET

N200

C US

R

172608 SERIAL(RS-232)

FUSE250V T3.15A

!

RCCOM

RC3

RC2

RC1

FAIL

DOPUMP OFF

G

V-

+

mA-

+

MS1

MS2

MS3

MSCOM

MSICOM

MSI2

MSI1

1

122

3

4

5

6

7

8

9

10

11

13

NOT using

Figure5.3 DescriptionofRearPanel

1 Gas Inlet This is an inlet for introducing sample gas. The connector is Rc1/4 or 1/4NPT.

2 Gas Outlet This is an outlet for discharging sample gas. The connector is Rc1/4 or 1/4NPT.

3 Fan This is a cooling fan inside the OX400. Make sure the outlet of the fan is not blocked

4 Power Plug This is a 3P power plug with a ground terminal. A fuse is included. Use the supplied power

cord. Do not use the power cord with another device. 5 ContactOutputTerminal(DOOutput) The event of a high or low oxygen concentration alarm “ALM7” causes output at the

terminal. It does not work without a concentration alarm. 6 ContactOutputTerminal(FAIL) This is a contact output terminal for errors. The event of an error causes output.

7 Contact Output Terminal for Range Output In the auto range, the current range is output at the contact output terminal

8 FuseRatingDisplay The rating of the power fuse is displayed.

9 Contact Input Terminal Thiscontactinputterminalisusedtoturnonandoffthesuctionpumpfromtheoutside.

10 ExternalOutputTerminal(PrimaryOutput) Measuredvaluesareoutputat4-20mADCinthesetmeasurementrange.

11 ExternalOutputTerminal(SecondaryOutput) Measuredvaluesareoutputatthesetvoltage(0-1,0-5,and0-10VDC)

12 Multi-Selector Contact Output Terminal

9thEdition:Nov.20,2019-00

Page 37: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

<5. Names and Functions> 5-4

IM 11M10B01-01E

Thisisthecontactoutputtoswitchmeasurementflowpath.Furthermore,measurement flowpath data is output at this contact output terminal.

13 RS232Connector A D-sub 9-pin connector is connected to this connector when using serial communication.

9thEdition:Nov.20,2019-00

Page 38: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

<6. Operation and Parameters> 6-1

IM 11M10B01-01E

6. OperationandParametersThe OX400 has two modes, “MeasurementMode” and “Maintenance Mode.” Oxygen concentration is displayed in the measurement mode. The setting of operation parameters and calibrationoperationareperformedinthemaintenancemode.Furthermore,theMODELEDonthe front panel is on in the maintenance mode. To enter the maintenance mode, or to return to the measurement mode from the maintenance mode, hold down the [MODE] key for two seconds. TheparametersshowninFigure6.1areavailableinthemaintenancemode.Parameterscanbedisplayedbyscrollingwiththe[▼],[▲]keys.

MODE

▼↓↑▲

CALへ

Hold down for 2 seconds

Hold down for 2 seconds

CAL

▼↓↑▲

[ENT]

rnG

oUt2

HoLd

nAMU

ALM

SEt.C

CoEF

rEV

SEnS

▼↓↑▲

▼↓↑▲

▼↓↑▲

▼↓↑▲

▼↓↑▲

▼↓↑▲

▼↓↑▲

▼↓↑▲

▼↓↑▲MODE

[ENT]

[ENT]

[ENT]

[ENT]

[ENT]

[ENT]

[ENT]

[ENT]

[ENT]

CEL.r

▼↓↑▲

[ENT]

Performs calibration. (3-point / 2-point / 1-point / Air)

Destination (reference section)

Maintenance mode

“0-1 V DC”

To “rEV”

To measurement mode

Description Default

To Calibration “CAL”(Section 6.2)

To Displaying Cell Resistance Value (Section 6.1.11)To Setting Output Range(Section 6.1.1)

To Setting Secondary Output (Section 6.1.2)

To Setting Hold(Section 6.1.3)

To Setting Burnout(Section 6.1.4)

To Setting Alarms(Section 6.1.5)

To Setting Calibration Gas Concentration(Section 6.1.6)

To Displaying Calibration Coefficient (Section 6.1.10)To Setting Sensor Constant(Section 6.1.7)To Displaying Software Revision “rEV”(Section 6.1.12)

For maintenance and diagnosis.

For maintenance and diagnosis.

Sets the measurement range. (Auto / Fixed / Partial)

Selects the voltage output. (0-1 / 0-5 / 0-10 V DC)Specifies the output hold during maintenance. (None/Previous value)Specifies the burnout function in the event of a failure (None / Burn-up / Burn-down)

Sets the oxygen concentration alarm (DO). (None / HL&LL / Only HL / Only LL)Configures the setting at the time of calibration. (Note 1) (See 6.2 Calibration “CAL” when configuring the setting while performing calibration)

The setting must be configured when replacing sensors. (ZE1 / SP1 / ZE4 / SP4 / CON4)

Displays the revision of the software for maintenance.

“All”(3-point)

“Auto”

“None”

“None”

“None”

Note 1: When “Air” is selected in Calibration “Cal,” the “SEt.C” menu is not displayed.Note 2: Some settings are omitted in the above flow chart. For setting output smoothing function of “SMoo” menu, see Section 6.1.8. When switching of measurement flow path using the Multi-selector, set the measurement flow path (1 to 3) in “MLS.” For the operation of the “MLS” menu, see Section 6.1.9.

Figure6.1 ListofParametersinMaintenanceMode

9thEdition:Nov.20,2019-00

Page 39: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

<6. Operation and Parameters> 6-2

IM 11M10B01-01E

6.1 StartupandSettingsTurningonthePower

Turn on the POWER switch on the front panel. After all displays turn on, the warm-up screen “HEAt” appears, and the OX400 automatically enters the warm-up mode. Theremainingwarm-uptimeisdisplayedandcounteddownfrom20,andwhenthewarm-up time ends, the measurement screen appears automatically. The warm-up time is about 20minutes.ThemAoutputis4mAduringthewarm-uptime,andthevoltageoutputis0V.Furthermore,itisrecommendedtocheckandsettheparametersduringthewarm-uptime. The following shows the operating procedure. 1.Settheparametersrequiredforoperation(inthemaintenancemode). 2.Checkandadjustthesamplegasflowrate(200±25mL/min,afloatershouldbebetweenupperand lowerbarofthe200mL/minbaronflowmeter) 3.Checkthemeasuredvalues(inthemeasurementmode),andperformcalibration,ifnecessary(inthemaintenancemode).

6.1.1 SettingOutputRange“rnG”Operation:Holddownthe[MODE]keyfor2seconds,select“rnG”withthe[▲],[▼]keys,andpressthe[ENT]key.Roughly three types of measurement range are available, “Auto Range,” “MANUAL Range,” and “Partial Range.” (1)AutoRange“AUto”(autorange)

With respect to this range, the range is switched automatically depending on the oxygen concentrationvalue.Enterrangecode“1”(default)ifthefullscaleisalways10x10n, as in thecaseof0to10ppm,0to100ppm,0to1000ppm,0to10%,0to100%.Furthermore,enter range code()“2”to“9”,asinthecaseof0to0 ppm, 0 to 00 ppm, 0 to 000 ppm, 0 to 0%.Forexample,enterrangecode“2”if20x10n applies, as in the case of 0 to 20ppm,0to200ppm,0to2000ppm,0to2%,0to20%.

(2)MANUALRange“MAn”(fixedrange) Thisrangeisafixedrange.Selectonefromthefollowingsixranges:0to10ppm,0to100

ppm, 0 to 1000 ppm, 0 to 1%, 0 to 10%, and 0 to 100%. The range is always the same and is independent from the oxygen concentration.

(3)PartialRange“FrEE”–“Fr.HI”–“Fr.Lo”(freerange) Withrespecttothisrange,anyrangecanbesetandfixed.However,thesmallestspanof

therangemustbemorethan20%FSoftheaboveMANUALrange. Examples:2to4ppmiftheMANUALrangeis0to10ppm

60to80ppmiftheMANUALrangeis0to100ppmNOTE:Flashingitaliccharacterssuchas“1.000”usedinthesubsequentoperationflowcharts

meanwaitingforkeyinput.Furthermore,thelargerboxmeansthemaindisplay(largedigitaldisplay)onthefrontpanelandthesmallerboxmeansthesub-display(smalldigitaldisplay).

AUto

rnG

Main displayMODE LEDThe MODE LED turns on in the maintenance mode.

The event of an error causes the LED to turn on.The event of an alarm causes the LED to flash.

Sub-display

ERR/ALM LED

The respective unit turns on.

%PPM

Unit LED

MODEERR/ALM

Operation in the maintenance mode

Press the key to return to the previous operation.

Hold down the key for 2 seconds to return to the measurement mode.

You may need to enter a numerical value, decimal point, and unit during the operation. The following operation flow charts may omit this operation. For details of the numerical value, decimal point, and unit operation, see Section 6.6.

MODE

MODE

9thEdition:Nov.20,2019-00

Page 40: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

<6. Operation and Parameters> 6-3

IM 11M10B01-01E

%

%

%

%

%

% %

%

%10 0.0

Fr.HI

PPM Pressing the [MODE] key in the event of an input error returns to previous operation.

The main display flashes when an input error occurs (Note 3).

▼ ▲CAL

rnG

▼↓↑▲

▼↓↑▲

[ENT]

AUto

rnG

[ENT] 1

AUto

1

AUto

[ENT] [ENT]

MAn

rnG

[ENT]

1 0 0

MAn

1 0

MAn

1

MAn

▼↓↑▲

▼↓↑▲

▼↓↑▲

PPM

PPM

PPM

1 0 0 0

MAn

10 0 0

MAn

10 0 0

0 0 0

0

0

0

MAn

▼↓↑▲

▼↓↑▲

▼↓↑▲

PPM

PPM

PPM

▼↓↑▲

[ENT] 100 00

MAn

PPM [ENT]

FrEE

rnG

[ENT] 50. 00

Fr.Lo

PPM 50 .00

Fr.Lo

PPM [ENT] [ENT]

[ENT]

MODE

▼ ▲

% %9 0 .00

Fr.HI

PPM [ENT] 90. 00

Fr.HI

PPM [ENT]

Hold down for 2 seconds.

To “rEV”

Setting the MANUAL Range

Setting the Auto Range(Example of setting the range of 10 x 10n)

(Note 1)Auto RangeEnter codes 1 to 9.(Note 1)

Return to “rnG.”

Return to “rnG.”

To 100% (Note 1): Auto Range CodesCode “1” allows setting the range in which the full scale is always 10 X 10n, as in the case of 0 to 10 ppm, 100 ppm, 1000 ppm, 1%, and so on. Set code “2” if 20 X 10n applies, as in the case of 0 to 20 ppm, 0 to 200 ppm, 0 to 2000 ppm, 0 to 2%, and so on. Likewise, codes “1” to “9” can be set.

Range of 0 to 100 ppm is confirmed

The above is an example of setting the fixed range of 0 to 100 ppm.

(Note 2) How to set the manual fixed rangeSelect one of the six ranges as follows: 0 to 10 ppm, 0 to 100 ppm, 0 to 1000 ppm, 0 to 1%, 0 to 10%, 0 to 100%.

To set the partial range, set the MANUAL Range and then the Hi/Low values of Partial Range (Note 3) .For details of the numerical value, decimal point, and unit operation in the partial range, see Section 6.6.

To 10 ppm

(Example of setting 50 to 90 ppm)(Note 3) To set a partial range, select “FrEE” with the ▼ key. Then, set the Hi/Low values. However, if the Hi and Low values are reversed, or the span (the smallest width) is not larger than 20%FS of the MANUAL Range, an input error occurs (the display flashes).

Enter the high value in the patial range.Enter the numerical value, decimal point, and unit.

The high value is confirmed.(the range of 50 to 90 ppm is confirmrd).

Enter the low value in the patial range.Enter the numerical value, decimal point, and unit.

The Low value is confirmed (the range of 50 to 90 ppm is setting).

Return to “rnG.”

MANUAL Range(Note 2)

%

Setting the Partial Range

Press the key to return to the previous operation. Hold down the key for 2 seconds to return to the measurement mode.

MODE MODE

Figure6.2 SettingRange“rnG”

9thEdition:Nov.20,2019-00

Page 41: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

<6. Operation and Parameters> 6-4

IM 11M10B01-01E

6.1.2 SettingSecondaryOutput“oUt2”Operation:Holddownthe[MODE]keyfor2seconds,select“oUt2”withthe[▲],[▼]keys,andpressthe[ENT]key.Withrespecttothesecondaryoutput,selectoneofthethreetypesofvoltageoutput:0-1Vis“1,”0-5Vis“5,”and0-10VDCis“10.”Figure6.3showsanexampleofsetting0-1VDC“1.”

CAL

rnG

oUt2

HoLd

▼↓↑▲

▼↓↑▲

▼↓↑▲

▼↓↑▲

[ENT] 1

oUt2

PPM

5

oUt2

10

oUt2

▼↓↑▲

▼↓↑▲

▼↓↑▲

▼↓↑▲

[ENT] 1

oUt2

PPM [ENT]

MODE

(0-10 V D C )

(0-5 V D C )

% %

To “rEV”

Hold down for 2 seconds.

To “10”0-1 V DC is confirmed.

Return to “oUt2.”

To “1”

Figure6.3 SettingSecondaryOutput“oUt2”

6.1.3 SettingHOLDFunction“HoLd”Operation:Pressthe[MODE]keyfor2seconds,selectHoLdwiththe[▲],[▼]keys,andpressthe[ENT]key.Set the output state in the maintenance mode. Select either “non” or “PrEV.” The former option means no hold, and the latter means hold the previous value. The following is an example of no hold.

HoLd

▼↓↑▲

CAL

▼↓↑▲

▼↓↑▲

[ENT]

HoLd

PPM

HoLd

▼↓↑▲

▼↓↑▲

▼↓↑▲

[ENT]

HoLd

PPM [ENT]

MODE

non

PrEV

non

To “rEV”

Hold down for 2 seconds. To “PrEV”

No hold is confirmed.

Return to “HoLd.”

(Hold the previous value.)

To “non”

% %

Figure6.4 SettingHOLD“HoLd”

9thEdition:Nov.20,2019-00

Page 42: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

<6. Operation and Parameters> 6-5

IM 11M10B01-01E

6.1.4 SettingBurnoutFunction“nAMU”Operation:Holddownthe[MODE]keyfor2seconds,select“nAMU”withthe[▲],[▼]keys,andpressthe[ENT]key.Settheburnoutfunctionfortheprimarycurrentoutputof4-20mADC(compliantwithNAMUR).Selectoneoftheoptions:Noburnoutfunctionis“non,”burn-upis“Er.HI,”andburn-downis“Er.Lo.” Figure6.5showsanexampleofsettingburn-up.NOTE:BurnoutisafunctioncompliantwithNAMURthatallowsincreasing/decreasingthe

current output to the high/low limits in the event of a failure. Burn-up allows increasing the outputtothehighlimitof21.0mAandburn-downallowsdecreasingtheoutputtothelowlimitof3.6mA.Theburnoutfunctionworksintheeventofanerror(whentheFAILcontactoutputcloses).

nAMU

▼↓↑▲

CAL

▼↓↑▲

▼↓↑▲

[ENT]

nAMU

PPM

nAMU

▼↓↑▲

▼↓↑▲

▼↓↑▲

[ENT] EEr. HHI

nAMU

PPM [ENT]

nAMU

▼↓↑▲

▼↓↑▲

ALM

▼↓↑▲

MODE

To “Er.Lo” burn-down

Burn-up is confirmed.

Return to “nAMU.”

To “non”

nnoonn

EErr. HHI

EErr. LLoo

%

To “rEV”

Hold down for 2 seconds.

%

Figure6.5 SettingBurnout“nAMU”

9thEdition:Nov.20,2019-00

Page 43: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

<6. Operation and Parameters> 6-6

IM 11M10B01-01E

6.1.5 SettingAlarms(OxygenConcentrationHigh/LowAlarms)“ALM”

Operation:Holddownthe[MODE]keyfor2seconds,select“ALM”withthe[▲],[▼]keys,andpressthe[ENT]key.High and low limit alarms can be set to the measured oxygen concentration values. With respect tothealarmsetting,thefollowingfouroptionsareavailable:Noalarm“oFF,”high/lowlimits“ALL,”high limit alarm “AL.HI,” and low limit alarm “AL.Lo.” If the high and low limit values are reversed or the set value is larger than 100%, an input error occurs and the setting is not accepted. Furthermore,theDOcontactoutputclosesintheeventofanoxygenconcentrationalarm(ALM7).Inotherwords,analarmcontactisactivated.Figure6.6showsasettingexample.Withrespecttotheinputmethodofanumericalvalue,decimalpoint,andunit,seeSection6.6.

CAL

ALM

▼↓↑▲

▼↓↑▲

[ENT]

oFF

ALM

[ENT] [ENT]

▼↓↑▲

oFF

ALM

▼↓↑▲

▼↓↑▲

▼ ▲

▼ ▲

100. 0

AL.HI

PPM [ENT]

1100 00 .11

AL.Lo

PPM

[ENT] [ENT] 00. 000

AL.Lo

PPM [ENT] 0. 000

AL.Lo

PPM [ENT] ALL

ALM

11 00. 0

AL.HI

PPM

HI

ALM

▼ ▲

Lo

ALM

[ENT] 00. 000

AL.Lo

PPM [ENT] 0. 000

AL.Lo

PPM [ENT]

[ENT] 11 00. 0

AL.HI

PPM [ENT] 100. 0

AL.HI

PPM [ENT]

MODE

% % %

%

%

% %

% %

To “rEV”

Hold down for 2 seconds.

To “Lo”No concentration alarm is confirmed.

Return to “ALM.”

Return to “ALM.”

Return to “ALM.”

Return to “ALM.”

Enter the concentration alarm low limit value.Enter the numerical value, decimal point, and unit.

Enter the numerical value, decimal point, and unit.

Enter the numerical value, decimal point, and unit.

Enter the numerical value, decimal point, and unit.

Low limit value is confirmed.

Enter the concentration alarm high limit value.

High limit value is confirmed.

The main display flashes in the event of an input error.

Examples of input errorHigh and low limit values are reversed.Input value is larger than 100%.

Pressing the [MODE] key in the event of an input error returns to the previous operation.

Enter the concentration alarm high limit value.

High limit value is confirmed.

Enter the concentration alarm low limit value.

Low limit value is confirmed.

To “oFF”

Figure6.6 SettingAlarm(OxygenConcentrationAlarm)“ALM”

9thEdition:Nov.20,2019-00

Page 44: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

<6. Operation and Parameters> 6-7

IM 11M10B01-01E

6.1.6 SettingCalibrationGasConcentration“SEt.C”Operation:Holddownthe[MODE]keyfor2seconds,select“SEt.C”withthe[▲],[▼]keys,andpressthe[ENT]key.Set the O2 concentration of gas used for calibration. Alternatively, this setting can be performed while performing calibration “CAL.” The gas concentration that can be set is restricted depending onwhichcalibrationisperformed(whichdependsonthespecificationincalibration“CAL”).Theoxygenconcentrationofairisfixedto20.6%O2,soitcannotbeset.Figure6.7showsasettingexample in 3-point calibration “ALL.” With respect to the input method of a numerical value, decimalpoint,andunit,seeSection6.6.

SEt.C

▼↓↑▲

MODE CAL

▼↓↑▲

▼↓↑▲

10.00

PPM

10.00PPM

1000 1000

PPM PPM

%

%

%

%

10. 00

1000

10. 00

1000

[ENT] [ENT] [ENT]

[ENT] [ENT]

To “rEV”

Hold down for 2 seconds.

Return to “SEt.C.”

10 ppm gas concentration is confirmed.

1000 ppm gas concentration is confirmed.

Figure6.7 SettingCalibrationGasConcentration“SEt.C”

6.1.7 SettingSensorConstant“SEnS”Operation:Holddownthe[MODE]keyfor2seconds,select“SEnS”withthe[▲],[▼]keys,andpressthe[ENT]key.Be sure to enter the sensor constant when replacing sensors. With respect to the sensor constant,thefollowingfiveparametersareavailable:ZE1,SP1,ZE4,SP4,andCON4,whichareontheconstanttagofthesensor.Besuretoentertherightparameter.Figure6.8showsasettingexample.Theseconstantsarespecifictothesensorandremainthesameforalloperations.Withrespecttotheinputmethodofanumericalvalue,decimalpoint,andunit,seeSection6.6.

ZE1

00AbZE1 SP1

AE0ESP1

CON4PPM

1021CON4

PPM

ZE4

FF79ZE4

SP4

AC75SP4

[ENT]

[ENT] [ENT] [ENT] [ENT] [ENT] [ENT]

[ENT] [ENT] [ENT] [ENT] [ENT] % %

SEnS

0

0

A

1

F

A

000

0AC E0F F7C

C77 038

Entering password

Enter “0330” and press the [ENT] key.

Enter the zero sensor constant 1 (in hexadecimal). Input is confirmed. Input is confirmed.

Input is confirmed. Input is confirmed.

Input is confirmed.Enter the span sensor constant 1 (in hexadecimal).

Enter the zero sensor constant 4 (in hexadecimal).

Enter the span sensor constant 4 (in hexadecimal).

Enter the gas concentration of CON4 (1000 ppm).Enter the numerical value only.

Return to “SEnS.”

*The entered sensor constants are updated altogether when the input of CON4 is confirmed. If the input is interrupted halfway, it returns to the sensor constant before the input.

SEnS

Figure6.8 SettingSensorConstants“SEnS”

9thEdition:Nov.20,2019-00

Page 45: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

<6. Operation and Parameters> 6-8

IM 11M10B01-01E

6.1.8 SettingOutputSmoothing“SMoo”If the oxygen concentration of sample gas rapidly changes, and the measured value is used for control,harmfulresultsmayoccursuchasfrequentonandoffswitching.In such a case, signal changes can be smoothed by giving an appropriate time constant and performingcalculation.Smoothingfactorfrom0to60secondscanbeset.Whentheoutputsmoothing is set, the smoothed value is an instantaneous value, which is displayed and output in analog form. The smoothed value is also used as the instantaneous value for checking the concentrationalarmandforRS232communication.Theoutputsmoothingis“non“0secondbydefault. Operation: Holddownthe[MODE]keyfor2seconds,select“SMoo”withthe[▲],[▼]

keys,andpressthe[ENT]key.Figure6.9showsanexampleofsettingthesmoothingfactorto30seconds.

SMoo SMoo

[ENT] [ENT] SMoo 00 30

Entering 30 sec. of smoothing factor.Enter “30” and press the [ENT] key.

Return to “SMoo.”

MODE Hold down for 2 seconds.

Figure6.9 SettingSmoothingFactor“SMoo”

6.1.9 SettingMulti-Selector“MLS”Thisitemisanoption,soyoudonotneedtosetitunlessyouspecifiedtheoptioncode“/MS.”Operation:Holddownthe[MODE]keyfor2seconds,select“MLS”withthe[▲],[▼]keys,andpressthe[ENT]key.Settherelaycontacttoswitchmeasurementflowforpath.Uptothreesamplingflowscanbeselected. Even though this setting is performed for an OX400 for which the option code “/MS” wasnotspecified,nothingwillhappen.Figure6.10showsasettingexample(setmeasurementflowpath1to3).Withrespecttotheusageexample,seeSection6.5.

MLS

▼↓↑▲

MODE CAL

▼↓↑▲

▼↓↑▲

[ENT] 3

MLS

PPM

0

MLS

▼↓↑▲

▼↓↑▲

▼↓↑▲

[ENT] 0

MLS

PPM [ENT]

▼↓↑▲▼↓↑▲

% %

To “rEV”

Hold down for 2 seconds. To “0”

0 is confirmed.

Return to “MLS.”

To “3”

Figure6.10 SettingMulti-Selector“MLS”

6.1.10 CheckingCalibrationCoefficient“CoEF”This item is for checking the sensor state and is not a setting. Operation:Holddownthe[MODE]keyfor2seconds,select“CoEF”withthe[▲],[▼]keys,andpressthe[ENT]key.

9thEdition:Nov.20,2019-00

Page 46: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

<6. Operation and Parameters> 6-9

IM 11M10B01-01E

Checkthecurrentcalibrationcoefficient.Thesecoefficientsareupdatedforeachcalibration.Figure6.11showsanexampleofcheckingthecoefficient.

CoEF

▼↓↑▲

CAL

▼↓↑▲

▼↓↑▲

[ENT] 1. 011

dS1

PPM[ENT] - 0. 23

dZ1

PPM [ENT]

▼↓↑▲

1. 013

dS4

1. 35

dZ4

[ENT] PPM [ENT] PPM

MODE

% %

%%

To “rEV”

Hold down for 2 seconds.

Return to “CoEF.”

Figure6.11 CheckingCalibrationCoefficient“CoEF”

6.1.11 DisplayingCellResistanceValue“CEL.r”This item is for checking the sensor state and is not a setting. Operation:Holddownthe[MODE]keyfor2seconds,select“CEL.r”withthe[▲],[▼]keys,andpressthe[ENT]key.Thesensorcellresistancevalueisdisplayed.TheunitisΩ.Ifthevaluebecomeslargerthan1050Ω,“oVER”appears.

NOTEWhen the sensor is replaced, the cell resistance value is not displayed. It is displayed when calibration of something other than air is performed.

Figure6.12showsanexampleofdisplayingthecellresistancevalue.

CEL.r

MODE CAL

▼↓↑▲

▼↓↑▲

[ENT] 56. 1

CEL.r

PPM[ENT]

▼↓↑▲

%

To “rEV”

Hold down for 2 seconds.

Return to “CEL.r.”

Figure6.12 DisplayingCellResistanceValue“CEL.r”

6.1.12 DisplayingSoftwareRevision“rEV”This item is for checking the software revision of the OX400 and is not a setting. Operation:Holddownthe[MODE]keyfor2seconds,select“rEV”withthe[▲],[▼]keys,

andpressthe[ENT]key.Thecurrentsoftwarerevisionisdisplayed.Figure6.13showsadisplayexample.

9thEdition:Nov.20,2019-00

Page 47: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

<6. Operation and Parameters> 6-10

IM 11M10B01-01E

rEV

▼↓↑▲

MODE CAL

▼↓↑▲

▼↓↑▲

[ENT] 1. 8

rEV

PPM[ENT]

▼↓↑▲

%

To “rEV”

Hold down for 2 seconds.

Return to “rEV.”

Figure6.13 DisplaySoftwareRevision“rEV”

6.2 Calibration“CAL”Be sure to perform calibration in the measurement mode after the warm-up operation ends. Calibration cannot be performed during the warm-up operation. If an error occurs during calibration, that calibration will be invalid, and keys other than the [MODE] key become invalid. Holddownthe[MODE]keyfor2secondstoreturntothebeginningofthecalibration“CAL”andpress the [MODE] key again to return to the measurement state. With respect to the calibration methods of the OX400, there are four types of method. (1)3-point(All)calibration“ALL”

By performing calibration at the three points of 10 ppm, 1000 ppm, and Air, the linearity for the entire zone between 0-10 ppm and 100% O2 is guaranteed. Gas used for calibration is restricted to the ranges of O2 concentration as follows.

(1)8ppm≤10ppmgas≤20ppm (2)800ppm≤1000ppmgas≤2000ppm (3)O2concentrationinAir:20.6%(Normalairiswithinthisrange)

Note: Besuretousegasthatmeetsconditions(1),(2),and(3)forcalibration.(2)2-pointcalibration“2Pnt”

Select zero and span calibration points in the measurement range you will use to perform calibration. However, gas used for calibration is restricted to the ranges of O2 concentration as follows.

<When the range is 1000 ppm or less.> (1)Zerogasis8%FSormore.

(2)Spangasislessthan120%FS. (3) Betweenzeroandspanofthesmallestrangemustbemorethan20%FSoftheMANUAL

range. Example1:Twopointsof1ppmand3ppminthe10ppmrange Example2:Twopointsof20ppmand40ppminthe100ppmrange

< When the range is more than 1000 ppm.> (1)Zerogasis800ppmormore.

(2)Spangasis100%FSorless. (3)Betweenzeroandspanofthesmallestrangemustbemorethan2000ppm.

Example3:Twopointsof0.6%and0.8%inthe1%range Example4:Twopointsof20.6%and90%inthe100%range

(3)1-pointcalibration“1Pnt”Principally, 1-point calibration allows obtaining an accurate concentration in an area near the calibration point, but the error may become greater as the distance from the calibration point increases. This is a useful calibration method when it is used in some limited areas. Select one point in an area as close to the concentration you want to measure as possible to perform calibration. However, there are the following limits to the O2 concentration. Be sure to perform calibration in that range of the O2 concentration. O2concentrationlimits:0.9ppmormoreand12%orless,orbetween35%and100%.

(4)Aircalibration“AIR”This is air-only calibration. Ues the cleanest air possible to perform calibration.

Note:Ifyoupressthe[MODE]keyduringcalibrationtoexitthecalibrationmode,thatcalibrationbecomesinvalid.

9thEdition:Nov.20,2019-00

Page 48: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

<6. Operation and Parameters> 6-11

IM 11M10B01-01E

If an error occurs during calibration, the ERR/ALM lamp turns on, and an error/alarm code is displayed on the sub-display. That calibration may become invalid depending on the error factor. Figure6.14showsthecalibrationmethod(includingthecalibrationgasconcentrationsetting).Withrespecttotheinputmethodofanumericalvalue,decimalpoint,andunit,seeSection6.6.

C AL

AIIRCAL

[ENT]

ALLCAL

[ENT]

AIRPPM [ENT] 20.60

OKPPM [ENT]

1 0.3810.00

PPM [ENT]

10.38PPM

[ENT][ENT] 0 989

1000PPM [ENT] 1001

989PPM

AIRPPM [ENT]

AIRPPM [ENT]

[ENT]

1PntCAL

[ENT] 5 0.251Pnt

PPM [ENT]

50.25PPM [ENT] 50.25

OKPPM [ENT]

2PntCAL

[ENT] 10.00ZEro

PPM [ENT]

10.24PPM [ENT] 9 0.25

SPAnPPM [ENT] 90.18

90.25PPM

[ENT]90.25OK

PPM

[ENT]

M O D E

OK

[ENT]PPM

50.07

10.00

20.78

10.21

20.60 20.95 20.60

%

%

%

%

%

% % %

% % %

%

% % %

% %

Hold down for 2 seconds.

If an error occurs, and “ERR” appears in the sub-display, calibration may be not accepted.

Execute 1-point calibration

Execute 2-point calibration

Execute Air calibration

Execute 3-point (All) calibration

To “ALL” Enter the gas concentration at any calibration point.Enter the numerical value, decimal point, and unit.

Enter the numerical value, decimal point, and unit.

Enter the numerical value, decimal point, and unit.

Wait for stabilization after sampling calibration gas.

Display the gas concentration at the calibration point.

Calibration successful

Return to “CAL.”

(*)

Enter the gas concentration at zero point.Wait for stabilization after sampling zero gas.

Display the gas concentration at zero point.

Enter the gas concentration at span point.Wait for stabilization after sampling span gas.

Display the gas concentration at span point.

Calibration successful.

Return to “CAL.”

Wait for stabilization after sampling air. Calibration successful

Return to “CAL”

Enter the 10-ppm gas concentration. Enter the 1000-ppm

gas concentration.Enter the numerical value only. Enter the numerical value only.

Wait for stabilization after sampling 10-ppm gas.

Wait for stabilization after sampling 1000-ppm gas.

Display the 10-ppm gas concentration.

Display the 1000-ppm gasconcentration.

Return to “CAL.”

Check the air concentration (setting is not needed).

Wait for stabilization after sampling air. Calibration successful.

Note : If an error occurs during calibration, that calibration may become invalid. With respect to the action in the event of an error, see Section 7.2. (*) After `OK` is displayed, wait 10 seconds before pressing ENT key.

To “1Pnt”

Figure6.14 Calibration“CAL”

9thEdition:Nov.20,2019-00

Page 49: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

<6. Operation and Parameters> 6-12

IM 11M10B01-01E

6.3 CommunicationTheOX400hasRS232serialcommunicationasstandard.Oxygenconcentration,alarm,anderror information are transmitted via this communication. The following shows the communication specifications.• CommunicationSpecifications

Table6.1 CommunicationSpecifications

Item DescriptionCommunication method One-way(transmissiononly),asynchronousData format ASCIIBaud rate 38400 bpsData length 8-bitParity NoneStop bit 1-bitFlowcontrol None

• Data Logging Packet

A A A A . A A A , B B B , C C C C , D D D D CR

Data delimiter “,” Terminate code

Figure6.15 ConfigurationofDataLoggingPacket

When communication terminal sends “TS” command to the OX400, OX400 will feed back data loggingpacketperiodically(approx.200ms.)Whenyouwanttostopcomm,sendthestopcommand “CR”. The content of the data logging packet has the following meanings.

Table6.2 ContentofData

Data ContentAAAA.AAA Oxygen concentration valueBBB UnitCCCC Error(hexadecimaldisplay)DDDD Alarm(hexadecimaldisplay)

(1)Oxygenconcentrationvalue The oxygen concentration value is displayed in the AAAA.AAA format and up to the three

decimal places are output. Note: Whentheoutputsmoothingisset,thesmoothedvaluebecomesanoxygenconcentrationvalue.(2)Unit

The unit is displayed in the BBB format. The output character is “%” or “ppm.”(3)Error The error is displayed in the CCCC format. The output characters are displayed in

hexadecimalformat,andTable6.3showstheerrorfactorsandoutputcharacters.

Rear side of the apparatus.

12345

6789

* Please use crossing cable.

4-40 UNC

Serial(RS-232)

D-sub (male)

1 : N.C2 : RXD3 : TXD4 : N.C5 : GND6 : N.C7 : N.C8 : N.C9 : N.C

9thEdition:Nov.20,2019-00

Page 50: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

<6. Operation and Parameters> 6-13

IM 11M10B01-01E

Table6.3 ErrorFactorandOutputCharacterList

Outputcharacters Error factor0001 Err1:Sensorerror0002 Err2:Heatertemperatureerror0004 Err3:Temperaturesensorburnout0008 Err4:Devicetemperatureerror0010 Err5:CPUerror0020 Err6:FANstop

If multiple errors occur, the logical sum of output characters is output. Example1:0003;Err1andErr2areoccurring

Example2:0017;ThefourerrorsErr1,Err2,Err3,andErr5areoccurring.MSB LSB

Err1: Sensor errorErr2: Heater temperature errorErr3: Temperature sensor errorErr4: Device temperature burnout

Err5: CPU error

Err6: FAN stop

01

1111

23456789101112131415

Figure6.16 ConfigurationofErrorData(0017)

(4)AlarmThe alarm is displayed in the DDDD format. The output characters are displayed in hexadecimal format,andTable6.4showstheerrorfactorsandoutputcharacters.

Table6.4 AlarmFactorandOutputCharacterList

OutputCharacters AlarmFactors0001 ALM1:Heaterunstable(temperatureunstable)0002 ALM2:Sensorfailure0004 ALM3:Electromotiveforce(EMF)error0008 ALM4:Asymmetricalvoltageerror0010 ALM5:Calibrationerror0020 ALM6:Sensorresistancevalueerror0040 ALM7:OxygenconcentrationHi/Lowlimiterror0080 ALM8:Overrangeerror

If multiple alarms occur, the logical sum of output characters is output. Example1:00C0:ALM7andALM8areoccurring.

ALM1: Heater unstable (temperature unstable)ALM2: Sensor failureALM3: Electromotive force (EMF) errorALM4: Asymmetrical voltage error

ALM5: Calibration error

ALM6: Sensor resistance value error

ALM7: Oxygen concentration Hi/Low limit error

ALM8: Over range error

MSB LSB0123456789101112131415

11

Figure6.17 ConfigurationofAlarmData(00C0)

9thEdition:Nov.20,2019-00

Page 51: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

<6. Operation and Parameters> 6-14

IM 11M10B01-01E

6.4 MeasurementGasSamplingUsingAspirator

Ifyouselected“-A”:withanaspiratorasthesamplingmethod,anaspiratorisincludedintheOX400 package. Connect the aspirator to the sample gas outlet on the rear panel of the OX400. Ifyouselected“T”:1/4NPTasthepipingconnector,connectanadapterbetweenthesamplegasoutlet and the aspirator. Supply clean air or N2 gas to the aspirator. The external diameter of the gas supply inlet of the one-touchconnectortypeaspiratorisø6,andthegasoutletisø8.Donotconfusethesupplyinletand gas outlet when you connect the aspirator.

Connect the adapter if the piping connector is 1/4NPT.

Gas outlet (for external diameter Φ8)

Supply gas inlet (for external diameter Φ6)

Gas outlet

Air or N2 gas

Unit: mm

Rear panel of the OX400

Figure6.18 ExampleofConnectingAspiratorandGasPiping

Figure6.19showstherelationshipbetweenthesupplypressure,suctiongasflowrate,andtotaldischargegasflowrateoftheaspirator.Theairsupplypressuretotheaspiratorshallbeintherangebetween65and100kPaG.Also,maintaintheairsupplypressureatacertainvaluetopreventthepressurefromfluctuating.Ifthepressurefluctuates,thesuctionflowratefluctuatesresultinginafluctuation of measured data.Theallowablepressurefluctuationrangeisthesetpressure±2kPaG.Furthermore,theoutlet pressure shall be equal to the atmospheric pressure. Also make sure back pressure is not applied. Checkthesamplegasflowrate,andadjustitto200±25mL/min(afloatershouldbeonthebarofthe200mL/minbaronflowmeter)usingthethrottleknobonthefrontpanelafteradjustingtheairsupply pressure to the aspirator.

0

200

400

600

800

1000

1200

1400

60 65 70 75 80 85 90 95 100 1056

7

8

9

10

11

12

13Aspirator’s suction characteristics

Suc

tion

flow

rate

(mL/

min

)

Aspirator’s total discharge flow rateSuction flow rate

Asp

irato

r’s to

tal d

isch

arge

flow

rate

(L/m

in)

Air supply pressure range

Air supply pressure (kPaG)

Figure6.19 Aspirator’sSuctionCharacteristics

9thEdition:Nov.20,2019-00

Page 52: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

<6. Operation and Parameters> 6-15

IM 11M10B01-01E

6.5 SwitchingMeasurementFlowPathUsingMulti-Selector

This item is an option, so if you did not specify option “/MS,” this function does not work. When oxygen concentration at multiple locations is measured, the multi-selector allows switching ofmeasurementflowpathusingtherelaycontactoutput.Threegasflowscanbeswitchedfromthe panel. SpecifytheflowpathNo.(“1”through“3”)inthemaintenancemode“MLS.”Thedefaultin“MLS”is“0.”Fordetails,seeSection6.1.9,“SettingMulti-Selector“MLS”.”Furthermore,outputofmeasurementflowpathdatacanbeperformedbycombiningtwocontacts.

MS1

MS2

MS3

MSI1

MSI2

MSICOM

MSCOM

MSI1-MSICOM MSI2-MSICOM

MS OFF OPEN OPEN

MS1 (1st flow) CLOSE OPEN

MS2 (2nd flow) OPEN CLOSE

MS3 (3rd flow) CLOSE CLOSE

Solenoid valve

Contact output for switching of measurement flow path

1st flow path

2nd flow path

3rd flow pathPower supply

OX400 terminal markings

Contact output to the customer

Output for measurement flow path data. (Note)

(Note) The following table shows the relationships of flow no. contact signals. The flow set in “MLS” closes.

Figure6.20 SwitchingMeasurementFlowPathUsingMulti-Selector (ExampleofThreeGasFlows)

Whenthemulti-selectorisused,theflowpath“MLS1”through“MLS3”undermeasurementare displayed on the sub-display on the front panel. If an error or alarm occurs while the multi-selectorisused,theflowpathNo. and error/alarm are displayed sequentially.

9thEdition:Nov.20,2019-00

Page 53: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

<6. Operation and Parameters> 6-16

IM 11M10B01-01E

6.6 NumericalValue,DecimalPoint,andUnitInputOperation

Figure6.21showsanexampleofoperationtoinputanumericalvalue,decimalpoint,andunit.Italiccharactersindicatetheflashingstate.

ZEroPPM

ZEroPPM

ZEroPPM

ZEroPPM

ZEroPPM

ZEroPPM

ZEroPPM

ZEroPPM

ZEroPPM

ZEroPPM

%%%%

%%%

% % %

10.00 10.00 10.0 0 10.00

[ENT] [MODE]

[ENT]

100.0 1000.

100.01.000

(0→1→ • • • →9→0) (9→8→• • • →0→9)

10.00

10.00

Numerical value input

Increment by pressing the ▲key. Decrement by pressing the ▼ key.

The 4th digit flashes. The 3rd digit flashes. The 2nd digit flashes. The 1st digit flashes.

Decimal point and unit input

*The decimal point and the selected unit LED flash during input.

To 100%

To 10 ppm

Input is confirmed.Figure6.21 ExampleofOperationtoInputNumericalValue,DecimalPoint,andUnit

9thEdition:Nov.20,2019-00

Page 54: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

<7. Inspection and Maintenance> 7-1

IM 11M10B01-01E

7. InspectionandMaintenanceRoutine inspection and maintenance is important to ensure operation of the OX400 in a good condition. Perform regular inspection and maintenance in accordance with the following instructions.

7.1 RoutineInspectionandMaintenance(1)Checkingreadings

Measure calibration gas about once every two to three months and check the readings. If an error is found with the calibration gas concentration, perform zero-span calibration in the range you use.

(2)Checkinggasflowrate Regularlycheckthesensorgasflowrateinordertomakesurethattheflowrateis200

±25mL/min(afloatershouldbebetweenupperandlowerbarofthe200mL/minbaronflowmeter).

(3)Other Regularly check for signs of malfunction such as strange noises of the pump and fan, or

unusually high temperature of the case. (4)Replacingthefuse BesuretoturnoffthepoweroftheOX400andremove

the power plug from the socket before replacing the fuse. The fuse is installed in the lower part of the power cordplugontherearpaneloftheOX400(seerightfigure).Removethepowercordfromtheplug,pullthefuse holder out of the lower part of the plug, replace the fuse with a new one, and push the fuse holder back in place. Be sure to replace the old fuse with a correctly rated one.

If a replaced fuse burns out soon, the circuit is likely to be defective. Please contact our service department.

7.2 InspectionintheEventofanErrorIf an error occurs, the ERR/ALM lamp on the front panel turns on. An error/alarm code is displayed on the sub-display. If multiple errors occur, codes are displayed in the order of occurrence. Anerrormeansamalfunction,sotheheateristurnedoffandthemeasurementisstopped.When an error occurs, repair is required. On the other hand, an alarm means a warning, so measurement is continued. An alarm occurring during calibration may invalidate the calibration depending on the content of the alarm.The following shows the display and output in the event of an error/alarm.

ERR/ALM

ERR/ALM

Error occurs

LED lamp

Alarm occurs

On

Flashing

Sub-display

Err code

ALM code

[ErrX]

[ALMX]

FAIL contact DO contact

CLOSE

Closed only in the event of “ALM7”

OPEN

OPEN

mA output

Scale-out by burnout

Measurement output

Front Panel Rear Panel

EventNotice

(Note) When burnout function is disabled, the mA output in the event of an error (fail) is “4 mA.” Furthermore, the mA output during warm-up is also “4 mA.” FAIL (error) contact and DO (alarm) contact are “OPEN” when the power is OFF.

Fuse Holder

Fuse

9thEdition:Nov.20,2019-00

Page 55: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

<7. Inspection and Maintenance> 7-2

IM 11M10B01-01E

7.2.1 InspectionintheEventofanAlarm(ALM)If an error is detected by the self-diagnosis function of the OX400 during operation, the ERR/ALM lampflashesandanalarm(ALM)codeisdisplayedonthesub-display.Ifanerrorisdetected,take action according to the Table 7.1. Table7.1 Alarm(ALM)ConditionsandActions

Display Content Diagnosis Condition ActionALM1 Unstable heater

temperature The heatertemperaturechangedby±5°Cormore from the standard temperature.

Check to make sure the sensorgasflowrateorpower supply is not unstable

ALM2 Sensor error The sensor electromotive force during calibrationbecamegreaterthan±30%relativeto the standard.

Check the calibration gas O2 concentration valueRe-calibrateReplace the sensor

ALM3 Electromotive force error Asensorelectromotiveforceerrorof±15to±30%relativetothestandardoccurredduringcalibration.

Check the calibration gas O2 concentration valueRe-calibrateReplace the sensor

ALM4 Asymmetric voltage error The air electromotive force became greater than±10mVduringcalibration.Calibrationupto±30mVispossible.Calibrationbeyond±30mVisnotpossible.

Check the air concentrationRe-calibrateReplace the sensor

ALM5 Calibration error Thecalibrationcoefficientbecamegreaterthanthespecifiedvalues(zero:±50,span:1±0.2).Whenzerois±50to±75andspanis1±0.2to1±0.4, calibration is possibleWhenzeroexceeds±75andspanexceeds1±0.4,calibrationisnotpossible

Check the calibration gas O2 concentrationRe-calibrateReplace the sensor

ALM6 Sensor resistance value error

Thesensorimpedance(cellresistancevalue)becamegreaterthan1kΩduringcalibration.

Prepare a replacement sensor

ALM7 Oxygen concentrationhigh/low limit error

The oxygen concentration became larger than the high/low limits.

Change the high/low limits as needed

ALM8 Over range error The oxygen concentration became greater than100%FS,orbecameanegativevalue.

Change the rangeRe-calibrate

7.2.2 InspectionintheEventofanError(ERR)Ifanerrorisdetectedduringoperation,theERR/ALMlampturnsonandanerror(ERR)codeisdisplayedonthesub-display.Ifanerrorisdetected,takeactionaccordingtoTable7.2.

Table7.2 Error(ERR)ConditionsandActions

Display Factor Diagnosis Condition ActionErr1 Sensor failure Sensor burnout. The sensor

electromotive force became lessthan-50mV.

Replace the sensor

Err2 Heater temperature error The heater temperature becamegreaterby±30°Crelative to the standard temperature.

Replace the heaterCheckthetemperatureadjustmentcircuit(Contactourservicedepartment)

Err3 Temperature sensor burnout

The temperature sensor has burned out.

Replace the heater(Contactourservicedepartment)

Err4 OX400 temperature error The temperature inside the OX400 became greater than70°C.

Check to make sure the ventilation holes, cooling fan outlet, and line are not blocked, lines are not blocked, and so on. Turnthepoweroffandthenon.

Err5 CPU failure A CPU failure occurred. Contact our service department. Err6 Fanstop The cooling fan stopped. Contact our service department.

Note: Ifanerror(ERR)occurs,theheateristurnedoff.Tocleartheerror,youneedtoturnthepoweroffandthenon.Itisrecommendedtoturnthepoweroffandthenontoseewhethertheerroroccursagain.BecarefulbecauseifthepoweroftheOX400isturnedoff,thecontactoutput“OPEN.”

9thEdition:Nov.20,2019-00

Page 56: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

<7. Inspection and Maintenance> 7-3

IM 11M10B01-01E

7.3 HowtoReplaceSensorWhen the sensor has deteriorated, replace it with a new sensor. As for the part number, see CMPL at the end of this manual. Be sure to perform calibration after replacing the sensor. Replace the sensor according to the following instructions.

7.3.1 RemovingSensor(1) Turnoffthepower,andbesuretopullthepowercordoutofthesocket.(2) Waitforabout1.5hoursuntiltheheatertemperaturefallstotheambienttemperaturelevel.

WARNINGBesuretoreplacethesensoraftertheheatertemperaturehasfallensufficiently.Notdoingsomaycauseburninjuries.

(3) Toremovetheuppercover,removesetscrewsatthreepointsontherearpaneloftheOX400.

(4) Slidetheuppercoverbackwardsandremoveitupwards.

WARNINGAVERTISSEMENT

Rear panel Fan

Sensor

Heater

Side panel

When the upper cover has been removed, you can see the heater sensor.

Remove the three screws and slide the upper cover backwards.

!INLET

OUTLET

N200

C US

R

172608 SERIAL(RS-232)

FUSE250V T3.15A

!

RCCOM

RC3

RC2

RC1

FAIL

DOPUMP OFF

G

V-

+

mA-

+

MS1

MS2

MS3

MSCOM

MSICOM

MSI2

MSI1

Figure7.1 RemovingCover

(5) RemovetheleadwireconnectorofthesensorfromthePCBboard.(6) Loosenthejointring 1 on the sensor outlet side and remove it towards the heater side. (7) Pulloutthejoint 2 while holding the sensor.

9thEdition:Nov.20,2019-00

Page 57: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

<7. Inspection and Maintenance> 7-4

IM 11M10B01-01E

(8) Rotatetheinlet-sidejointnut 3 counter clockwise seen from the heater side and remove it.

O-ring

Sensor

Lead wire connector

Lead wire connector

Heater

Heater

1

1

2

2

3

3

(9) Pullouttheholdingplate 4 , remove the ring 1 , and pull the sensor out of the heater straightforward.

O-ring

Lead wire connector

Heater

1 23

4

Figure7.2 RemovingSensor

9thEdition:Nov.20,2019-00

Page 58: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

<7. Inspection and Maintenance> 7-5

IM 11M10B01-01E

7.3.2 InstallingSensor(1) Passanewsensorthroughtheheater.

Sensor

Lead wire connector

Heater

(2) Insertthenut 3 , holder 5 , and O-ring 6 in this order from the front end of the sensor. Atthispoint,settheO-ringinthepositionabout3mmfromthesensorfrontend(ItisrecommendedtoreplacetheO-ringwithanewone).

O-ring

Lead wire connector

Heater

5 3

6

Approx. 3 mm

Sensor

Marking

13 mm

(3) Insertthesensorintotheconnectorandtightenthenut 3 firmlywithfingers.Atthispoint,make sure that the lead wire of the sensor is located in the upper right position.

Lead wire connector

SensorHeter

Move the sensor back and forth to align the nut with the sensor marking line.

3

Figure7.3 InstallingSensor

9thEdition:Nov.20,2019-00

Page 59: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

<7. Inspection and Maintenance> 7-6

IM 11M10B01-01E

(4) Insertthering 1 and holding plate 4 in this order into the sensor. Hold the sensor with fingersandrotatethejoint 2 right and left to push it into the sensor slowly.

Lead wire connector

3

Heater

Sensor

1 4

2

(5) Rotateandtightenthering 1 intothejoint 2 . Secure the holding plate 4 with two screws(7).

Connect the lead wire connector of the sensor to the PCB board.

Lead wire connector

Heater

Sensor

7

Sensor constant tag

3 1 4

2

(6) Whentheinstallationofthesensorisfinished,takeanoteofthesensorconstantonthesensor constant tag attached to the sensor lead wire, and then close the cover.

(7) Turnonthepower,andwhen“HEAt”appearsandcountdownstarts,entersensorconstant“SEnS.”Asforhowtoenterit,see6.1.7“SettingSensorConstant.”

Be sure to perform calibration after replacing the sensor.

9thEdition:Nov.20,2019-00

Page 60: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

<7. Inspection and Maintenance> 7-7

IM 11M10B01-01E

7.4 HowtoReplaceActivatedCarbonFilterActivatedcharcoalfilterwillbeprovidedwhen“/A”optionisspecified.FiltershouldbeinstalledbetweengassamplinglineandgasinletofOX400.Filterhasalifetimeandneedtoreplaceperiodically, its replacement interval vary depending on the application. As for the part number, see CMPL at the end of this manual.(1) Loosenthefourscrewswithwashers B of holder A and remove holder A .(Figure7.4)(2) Removecottonfilter C ,feltfilter D , and O-ring E .(3) Disposeofactivatedcarbon F . Remove the bottom side holder same as the upper side holder and pull out the activated

carbon.Referringitem(5)to(7),setthebottomsideholder.(4) Fillthecasewithapproximately9gramsofnewactivatedcarbonuptoapproximately8

mmfromtheuppersurfaceofthecase.Shakethecasegentlytofilltheintersticesoftheactivated carbon.

(5) Iftheremovedfeltfilter D isdirty,replaceitwithanewone.Setfeltfilter D onto the activated carbon F so that the laminate surface faces activated carbon F .

(6) Ifcottonfilter C isdirty,replaceitwithanewone.Seta0.2gramcottonfilterontofeltfilter D .

(7) AlignO-ring E with groove 1 and secure holder A with screws with washers B .

CAUTIONTightly close the lid of the replacement activated carbon case and store it in an indoor location whereitwillnotbesubjecttohumidityanddirectsunlight.

Cotton filter (*1)

Cotton filter

Felt filterLaminate surface faces activated carbon

Felt filterLaminate surface faces activated carbon

O-ring

O-ring

Activated carbon

*1: When fixing the holder, please do not get a cotton yarn of the cotton filter between the O-ring and the holder. This causes a leak.

Figure7.4 HowtoReplaceActivatedCarbon

9thEdition:Nov.20,2019-00

Page 61: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

<7. Inspection and Maintenance> 7-8

IM 11M10B01-01E

7.5 HowtoReplacetheLineFilterThelinefilterisinsidethegasinletconnectorontherearpanel.Thelinefilteriscoveredbyaplate, and both these parts are held in place by a C-shaped snap ring. The following explains how toreplacethelinefilter.Thisrequiresapairofsnapringpliersandsometweezers.(1) Insertthetipsofthesnapringpliersintheholesonthesnapringandapplypressureto

release and remove the snap ring.(2) Withthetweezers,removefirsttheplateandthenthefilter.(3) Attachthenewfilter,plate,andsnapringinthisorder.• Insertthefilterintotheconnectorwiththesmoothsidefacingup.• Placetheplateoveritwiththefinemeshsidefacingdown.• With the snap ring pliers, apply pressure to insert the snap ring, and hold down and secure

thefilterandplateinplacewiththesnapring.Atthispoint,makesureyouattachthesnapringfirmlyinplacesothattheplateissecuredenoughthatitcannotbemovedevenifanattempt to move it with tweezers is made.

CAUTIONThesnapringmustbesecurelyattachedtofirmlyholdthefilterandplateinplace.Otherwise,thefilterwillnotfunctioncorrectly.

Gas Inlet

Filter(K9643FB)The smooth side facing up

The fine mesh side facing down

Plate(K9213FB)

C-shaped snap ring(Y9011EV)

Up

Dwon

Figure7.6 HowtoReplacetheLineFilter

9thEdition:Nov.20,2019-00

Page 62: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

<8. Parameter Settings> 8-1

IM 11M10B01-01E

8. ParameterSettingsTable8.1

ParameterSettin

gList(C

ustomerSettin

gs)

Menu

Para

met

er 1

Para

met

er 2

Para

met

er 3

Functio

ndetail

Defaults

Desc

riptio

nCu

stom

er

Setti

ngs

Refe

renc

e se

ctio

n

Outpu

trange

(4-20mADC

)“rnG

“Auto”

Auto

rang

e“1

”(Note1)

6.1.1

“MAn

”M

anua

l ran

ge0-

1000

ppm

“FrEE”

Parti

al ra

nge

0-10

00 p

pm“Fr.L

o”Lo

w lim

it0

ppm

“Fr.H

I”H

igh

limit

1000

ppm

Second

aryou

tput

“oUt2”

“1”/“5”/“10”

Volta

ge o

utpu

t“1

”0-

1V D

C6.1.2

Holdfunctio

n“H

oLd”

“non

”/“PrEV

”Pr

evio

us o

r non

non

Non

hol

d6.1.3

Burnou

tfun

ction

“nAM

U”“non

”/“Er.Lo”/“Er.H

I”H

i/Low

/non

non

Non

bur

nout

6.1.4

O2con

c.Hi/Low

al

arm

“ALM

“oFF

”N

on a

larm

oFF

Non

ala

rm6.1.5

“ALL

”“A

L.Lo

”“A

L.HI”

Low

limit

Hi li

mit

“AL.HI”

Hi li

mit

only

“AL.Lo

”Lo

w lim

it on

ly

Calib

ratio

n ga

s co

ncen

tratio

n“SEt.C”

“1Pn

t”1

poin

t10

00 p

pmC

al-g

as c

onc.

(Note2)

6.1.6

“2Pn

t”2points

“Zero”

Zero

-gas

con

c.10

0.0

ppm

(Note3)

“SPA

n”Sp

an-g

as c

onc.

1000

ppm

(Note4)

“ALL

”3

poin

ts“10”

10 p

pm g

as c

onc.

10.0

0 pp

m(Note5)

“1000”

1000

ppm

gas

con

c.10

00 p

pm(Note6)

Outpu

tsmoo

thing

“SMoo

”“0to60”

Smoo

thin

g fa

ctor

.0

sec

Non(N

ote7)

6.1.8

Multi-selector

“MLS

”“0”/“1”/“2”/“3”

Flow

pathNo.1to3

“0”

Non

use

6.1.9

(Note1)AutoRangebetween0-10ppm

,0-100ppm

,0-1000ppm,0-1%,0-10%

,and0-100%.

(Note2)Calibrationgasconcentrationinputrange:0.90ppmto12.00%,35.00%to100.0%.G

asconcentrationcanalsobesetfrom“C

AL”calibration.

(Note3)“ZEro”inputrange:1.00ppmto100.0%.G

asconcentrationcanalsobesetfrom“C

AL”calibration,

(Note4)“S

PAn”inputrange;1.00ppmto100.0%.Inputisnotpossibleifthesm

allestspanislessthan20%

oftherangeafterinputting“SPA

n.”G

asconcentrationcan

also

be

set f

rom

“CAL

” cal

ibra

tion,

(Note5)Calibrationgasconcentrationinputrange:8.00to20.00ppm

.Gasconcentrationcanalsobesetfrom“C

AL”calibration,

(Note6)Calibrationgasconcentrationinputrange:800to2000ppm.G

asconcentrationcanalsobesetfrom“C

AL”calibration,

(Note7)Iftheoutputsmoothingisset,thesmoothedvaluebecom

esaninstantaneousvaluewhichisdisplayedandoutputinanalogformat.Thesmoothedvalueisalso

usedastheinstantaneousvalue(measuredvalue)forcheckingconcentrationalarmsandRS2

32com

munication.

(Note8)Setparam

eterswillnotbeerasedbyturningthepoweroffandon.

9thEdition:Nov.20,2019-00

Page 63: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

Blank Page

Page 64: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

CustomerMaintenanceParts List

OX400Low Concentration Zirconia Oxygen Analyzer[Style: S2]

CMPL 11M10B01-02E2nd Edition : Apr. 2011 (YK)

© Copyright Mar. 2010 (YK). 1st Edition : Oct. 2010 (YK)

7

8

3

1 2

Aspirator (Rc1/4)

Panel Mounting Hardware

Sensor related parts

Aspirator (1/4NPT)

6 5 4

Item Part No. Qty Description 1 K9643KA Aspirator Assembly (Rc1/4) K9643WK 1 Aspirator G7010XA 1 Bushing

2 K9643KB Aspirator Assembly (1/4NPT) K9643WK 1 Aspirator G7010XA 1 Bushing G9612CK 1 Connector

3 K9643KC Panel Mounting Hardware Assembly K9643GM 1 Frame (left-side) K9643GN 1 Frame (right-side) Y9514EU 4 Screw

*4 Heater Assembly 5 K9643KG 1 Sensor Assembly with O-ring 6 K9643WR 1 O-ring

7 A1113EF 1 Fuse (250V T3.15A)

8 Line Filter K9643FB 1 Filter K9213FB 1 Plate Y9011EV 1 Snap Ring (Retainer) K9643ZE 1 Snap Ring Plier (Tool to remove snap ring)

* Do not exchange these parts. Call service personnel.

(internal)

(internal)

Heater

!INLET

OUTLET

N200

C US

R

172608 SERIAL(RS-232)

FUSE250V T3.15A

!

RCCOM

RC3

RC2

RC1

FAIL

DOPUMP OFF

G

V-

+

mA-

+

MS1

MS2

MS3

MSCOM

MSICOM

MSI2

MSI1

Page 65: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

2

CMPL 11M10B01-02E2nd Edition : Apr. 2011 (YK)

Item Part No. Qty Description 1 K9643KH 1 Filter Assembly (Rc1/4) K9643KJ 1 Filter Assembly (1/4NPT)

2 K9643KK 1 Activated Carbon Filter Element Kit (15 times replacement) 3 K9643KL 1 Filter Kit (5 times replacement)

1

2

3

Cotton filter

Felt filter

ActivatedCarbonFilter(Optioncode"/A")

Page 66: Model OX400 Low Concentration Zirconia Oxygen Analyzer [Style: … · 2019-11-20 · User’s Manual: IM11M10B01-01E(this manual) * the “E” in the document number is the language

i

IM 11M10B01-01E

Revision InformationlTitle : ModelOX400LowConcentrationZirconiaOxygenAnalyzer[Style:S2]lManualNo. : IM11M10B01-01E

Nov.2019/9thEdition Correctedsomeerros.(page6-9)

Jan.2018/8thEdition Correctedsomeerros.(pages2-1to2-8)

Aug.2017/7thEdition Correctedsomeerros.(pages2-3,6-11)

May2017/6thEdition AdditionRoHS(pagesi,iv,vi,2-3)

Feb.2016/5thEdition Revised EMC standards, etc.

Aug.2015/4thEdition P2-3,Section2.1,“StandardSpecifications”ConformancetoSafetyandEMCstandards: AddedofKCMarkingandsomespellerrorcorrectionsandadditionofspecificationdescription. P2-4,Section2.3,“ModelandSuffixCodes”Note2:Additionofthedescription. P2-5,Section2.4“Consumables”K9643FB:Deletedtheasterisk. P6-1,Section6,“OperationandParameters”Figure6.1:Changedofparameterscord. P6-7,Section6.1.7,“SettingSensorConstant“SEnS””:Changedtheparameterscordof descriptionandfigure6.8. P6-9,Section6.1.9,“SettingMulti-Selector“MLS”Figure6.11:Changedofparameterscode. P6-12,Section6.3,“Communication”•CommunicationSpecifications:Addedoffigure. P 7-7, Some coments added.

May.2011/3rdEdition P2-1,Somerevisionofoutputrangedescription(autorangemodified);P2-4,SomerevisionofMS-

codedescription("/F"changedto"/A"foractivatedcarbonfilter);P2-5,SomerevisionofConsumable(Partsfor"/A"activatedcarbonfilteradded);P2-9,SomerevisionofExternalDimensions(for"/A"activatedcarbonfilter);P3-4,SomerevisionofdimensionsonFigure3.3(for"/A"activatedcarbonfilter);P4-4,SomerevisionofTable4.2RangeMarker(autorangemodified);P6-2,SomerevisionofAutoRangedescriptiononsubsection6.1.1(Howtosetrangecodemodified);P7-7SomerevisionofSubsection7.4"HowtoReplaceActivatedCarbon,FeltFilter,andCottonFilter"(for"/A"activatedcarbonfilter);CMPL11M10B01-02E,revisedto2ndedition(for"/A"activatedcarbonfilter).

Nov2010/2ndEdition(StylechangedtoS2) P i INTRODUCTION, Caution mark added to Safety explanation;P2-1,Somerevisionofaspirator

suctioncondition;P2-2,Somerevisionofpowersupplyandconsumption;P2-3,SomerevisionofConformancetoSafetydescription(CSAcertified);P2-4,SomerevisionofMS-codedescription;P2-5,SomerevisionofConsumable(PartsforLineFilteradded);P2-6,SomerevisionofExternalDimensions(Ventirationholesmodified);P2-7toP2-8,Somerevisionofexternaldimensions(rearpanelmodified);P2-11,SomerevisionofWiringDiagram(Cautionforpowercordmodified);P3-2to3-3,SomerevisionofdimensionsonFigure3.1to3.2(Ventirationholesmodified);P4-2,SomerevisionofFigure4.2(rearpanelmodified);P4-3,SomerevisionofFigure4.3WiringDiagram(Cautionforpowercordmodified);P4-5,SomerevisionofFigure4.4(rearpanelmodified);P5-1,SomerevisionofFlowmeterdescription(Howtoadjustmodified);P5-3,SomerevisionofFigure5.3(rearpanelmodified);P6-2,SomerevisionofFlowmeterdescriptiononsubsection6.1(Howtoadjustflowratemodified);P6-14,SomerevisionofFlowmeterdescriptiononsubsection6.4(Howtoadjustflowratemodified,somerevisionofFigure6.19);P7-1SomerevisionofCheckinggasflowratedescription on subsection 7.1;P7-3,SomerevisionofFigure7.1(rearpanelmodified);P7-5,Somerevisionofsec.7.3.2InstallingSensor(rmarkedlineaddedtosensorassembly); P 7-8, Subsection 7.5"HowtoReplacetheLineFilter"added;CMPL11M10B01-02EforStyleS2 newly published.

Jun.2009/1stEdition Newly published