    Module 2: Aquarius Training

  • 8/6/2019 Mod 2 Edwards Aquarius

    2/69

  • 8/6/2019 Mod 2 Edwards Aquarius

    3/69

    Notes:

    AQUARIUS TRAINING

    TABLE OF CONTENTS

    Lesson 1. Introduction to Aquarius Platinum

    Lesson 2. Preparing Aquarius

    Lesson 3. Treatment Mode

    Lesson 4. Pressures, Alarms & Fluid Balance

    These materials were created by Edwards Liesciences Corporation as part o the CriticalCare Nephrology Alliance. Seeking to provide CRRT proessionals with the highest caliber

    o multidisciplinary expertise, products and services, the Alliance brings together the uniquestrengths o Baxter Healthcare Corporation, an expert in renal therapies, and Edwards, aleader in critical care.

    Module 2: Aquarius Training

  • 8/6/2019 Mod 2 Edwards Aquarius

    4/69

    (page intentionally blank)

  • 8/6/2019 Mod 2 Edwards Aquarius

    5/69

    Notes:

    Module 2: Learning Objectives

    The material presented in this program is based on the Aquarius

    Operators Manual, Platinum Sotware version 6 US and the

    Aquarius v. 6 screens. The Operators Manual and the screens

    provide the instructions necessary or the proper operation

    o the Aquarius System. This training program is designed to

    complement the Aquarius Operators Manual and screens. It

    is not intended as a replacement. Please reer to the Operators

    Manual and the screens or more detail.

    The Aquarius System is indicated or continuous solute and/oruid removal in patients with acute renal ailure or uid overload.

    It may also be used in Therapeutic Plasma Exchange (TPE)

    therapies.

    THE AQUARIUS SYSTEM

    Understandtheset-up,primingandoperationoftheAquarius Hemofltration system

    Understandpressuremonitoringandalarmtroubleshooting

    Completethereturndemonstrationportionofthistrainingprogram

    LEARNING OBJECTIVES

  • 8/6/2019 Mod 2 Edwards Aquarius

    6/69

    (page intentionally blank)

  • 8/6/2019 Mod 2 Edwards Aquarius

    7/69

    Lesson 1: Introduction to Aquarius Platinum

  • 8/6/2019 Mod 2 Edwards Aquarius

    8/69

    (page intentionally blank)

  • 8/6/2019 Mod 2 Edwards Aquarius

    9/69

    Notes:

    AQUARIUS PLATINUM DESIGN CONCEPTS

    Patient saety

    Innovation

    Flexibility

    Easeofuse

    Consistentmessaging

    Upgradableasnewneeds/featuresarise

    INTERACTING WITH AQUARIUS PLATINUMThe operator interaces with Aquarius by turning andpushing the Main Selector button. This button is a rotaryswitch located below the display screen. It is used toselect and confrm dierent unctions and to modiytreatment parameters.

    USING THE SELECTOR BUTTON:

    Turn: to highlight the parameter

    Push once: to select the parameter

    Turn right: to

    Turn left: to

    Push once: to confrm new values

    Main Selector button

    Module 2: Lesson 1: Introduction to Aquarius Platinum

  • 8/6/2019 Mod 2 Edwards Aquarius

    10/69

    Notes:

    8

    Help screens will providedetailed inormation aboutindividual screens and allowdirect access to History

    NAVIGATING BETWEEN SCREENS

    The operator moves between screens by selecting a parameter and confrming the choice: Areyou sure messages indicate that the operator will need to confrm this choice beore movingorward. The deault choice is No.

    Moves to the nextscreen

    Moves back toprevious screen

    Confrms theselected choice

    Cancels the selectedchoice and movesback to previousscreen

    Module 2: Lesson 1: Introduction to Aquarius Platinum

  • 8/6/2019 Mod 2 Edwards Aquarius

    11/69

    Notes:

    9

    CONSISTENT MESSAGING

    Aquarius uses clear and consistent messaging, delivered in yellow or red message boxes, toinorm and guide the operator through the screens. Example:

    KEY FUNCTIONS OF THE AQUARIUS INTERFACE

    Messages in a red

    background describe the

    alarm condition that is

    currently underway and

    needs the immediate

    attention o the user.

    Messages in a yellow box

    may describe a set up

    instruction, a parameter

    that is being programmed

    or an alarm that reminds

    the user that something

    needs to be done.

    Blood Pump keyMute key

    Clamp keyMain Selector button

    Balance Start/Stop key

    Module 2: Lesson 1: Introduction to Aquarius Platinum

  • 8/6/2019 Mod 2 Edwards Aquarius

    12/69

    Notes:

    10

    INDIVIDUAL FUNCTION KEYS

    The individual unction keys and their displayed unctions are explained below.

    1. Operation status display

    Three status lights indicate the dierent operation modes.

    A. An illuminated red status light indicates:

    Warning! Alarm or system error.

    B. An illuminated green and yellow ashing status light indicates:

    Caution! Heater sel test is running. Priming mode is not accessible.

    C. An illuminated green and yellow status light indicates:

    Caution! Treatment time has run down or Treatment was stopped.

    Caution! Bag change required.

    Caution! Anticoagulant syringe is empty.

    Caution! Aquarius is in Preparation or Recirculation or Connection mode.

    D. An illuminated green status light indicates:

    Treatment is running. No alarms are active.

    E. All three status lights illuminating one ater the other indicate:

    Machine is perorming system test.

    Module 2: Lesson 1: Introduction to Aquarius Platinum

  • 8/6/2019 Mod 2 Edwards Aquarius

    13/69

    Notes:

    1

    2. Mute unction key

    Pressing the Mute key enables the operator to silence the alarm signal or a period otwo minutes. The LED integrated into the key ashes. I the cause o the alarm cannot beremoved within the given period, the audible alarm is reactivated. I another alarm occursduring this period, the audible alarm is generated at once.

    3. Clamp unction key

    Pressing the Clamp key opens the return line clamp during air alarm to allow removal o air

    bubbles rom within the tubing set. The LED integrated into the key ashes. When the air isremoved, treatment may be resumed by pressing the Blood Pump key. The return line clamp isautomatically reactivated ater one minute.

    4. Main Selector button

    Themainselectorbuttonisamulti-functionrotarybutton.Thesefunctionsinclude:

    a) Selecting unction windows by turning the selector button

    b) Confrming selected unctions by pressing the selector button

    c) Selecting input parameters by turning the selector button

    d) Opening the input window or the selected parameter by pressing the selector button

    e) Raising the parameter input or the selected parameter by turning the selector button tothe right

    ) Lowering the parameter input or the selected parameter by turning the selector button tothe let

    g) Confrming the entered parameter by pressing the selector button. The modifed parameteris displayed on the screen

    Module 2: Lesson 1: Introduction to Aquarius Platinum

  • 8/6/2019 Mod 2 Edwards Aquarius

    14/6912

    Notes:

    12

    Notes:

    5. Balance Start/Stop unction key

    Pressing the Balance Start/Stop key stops the fltrate and substitution/dialysate pumps. Thiscan be used to temporarily stop treatment, or example, to exchange bags. I an alarm occurs inthe fltrate and/or the substitution/dialysate circuits, the pumps stop and the LED integrated intothe key starts ashing. Ater correcting the cause o the alarm the pumps may be restarted bypressing the Balance Start/Stop key.

    Balance Start/Stop key is also called Start Treatment key.

    6. Blood Pump unction key

    Pressing the Blood Pump key starts or stops the ow o blood through the blood circuit. I analarm occurs in the blood circuit, all pumps stop and the LED integrated in the blood pump keystarts ashing. Ater correcting the cause o the alarm the system is restarted by pressing theBloodPumpkeyagain.Filtratepump,pre-dilutionandpost-dilutionpumpsstartwithdelayinrelation to the blood pump.

    The Blood Pump key is also used to immediately stop all pumps in case o an unpredictable

    occurrence.

    Module 2: Lesson 1: Introduction to Aquarius Platinum

  • 8/6/2019 Mod 2 Edwards Aquarius

    15/69

    (page intentionally blank)

  • 8/6/2019 Mod 2 Edwards Aquarius

    16/69

    (page intentionally blank)

  • 8/6/2019 Mod 2 Edwards Aquarius

    17/69

    Lesson 2: Preparing Aquarius

  • 8/6/2019 Mod 2 Edwards Aquarius

    18/69

    (page intentionally blank)

  • 8/6/2019 Mod 2 Edwards Aquarius

    19/69

    Notes:

    1

    POWER SUPPLYAquarius has two power supply switches.The Main switch, on the let side o the machineshould be set to position 1, the ON/OFF sot keyswitch is on the right side o the monitor displayscreen.

    SYSTEM TEST

    The system test takes approximately our minutes. When it is fnished, an audible alarm isgenerated and the green and yellow status lights are illuminated. The pumps stop in the correct

    position or the tubing set to be installed. The operator can now proceed to set up.

    Main Switch ON/OFF sot key

    CAUTION: The Aquarius system test must be perormed beore tubing, bags, pressuresensors and solutions are placed on the machine. The pump doors must be closed.

    Module 2: Lesson 2: Preparing Aquarius

  • 8/6/2019 Mod 2 Edwards Aquarius

    20/69

    Notes:

    18

    SETTING UP AQUARIUS

    Select the therapy to be used and confrm1.

    SCUF

    CVVH

    CVVHD

    CVVHDF

    Therapeutic plasma exchange(TPE)

    2. Select the type o Aqualinetubing set to be used and confrm Aqualinetubing set or regular treatment-Bloodowratesfrom30-450mL/min-Extracorporealvolume100mL

    AqualineStubing set or low blood ow treatment

    -Bloodowratesfrom10-200mL/min-Extracorporealvolume61mL

    Module 2: Lesson 2: Preparing Aquarius

  • 8/6/2019 Mod 2 Edwards Aquarius

    21/69

    Notes:

    1

    3. Place selected Aqualine on Aquarius

    Follow the detailed instructions listed below to place the lines and flter on the machine

    A. Place colored pump indicators in pumps. The colored pump holder segments are always atthe bottom o the pumps

    NOTE:Alternatively,step-by-step,on-screeninstructionsor the placement o the linesand flters can be accessedin the Zoom graphic screenso Aquarius

    Blood pump (red):

    top let pump

    Filtrate pump (yellow):

    bottom let pump

    Post-dilutionpump

    (green): top right

    pump

    Pre-dilution/dialysate

    pump (green): bottom

    right pump

    Module 2: Lesson 2: Preparing Aquarius

  • 8/6/2019 Mod 2 Edwards Aquarius

    22/69

    Notes:

    20

    B. Place automatic degassing chamber(ADU) in holder under green pumps.Ensure that the shortest line with theclamp is towards the back and thechamber cover is closed securelyaround the tubing

    Chamber cover closed (B)

    Line attached to ADU sensor (C)

    C. Attach the line with the luer lock,hydrophobic flter to the ADU sensor.

    The clamp should remain open

    Module 2: Lesson 2: Preparing Aquarius

  • 8/6/2019 Mod 2 Edwards Aquarius

    23/69

    Notes:

    2

    E. Locate the return line chamber(also reerred to as the dripchamber) on the Aqualine,then insert the line underneath itthrough air detector channel andline clamp. Push and lock the airdetector

    D. Insert heater coil into heater onthe right side o Aquarius (abovethe green scale) and close door

    Return linechamber

    Air detector

    Return line clamp

    Module 2: Lesson 2: Preparing Aquarius

  • 8/6/2019 Mod 2 Edwards Aquarius

    24/69

    Notes:

    22

    G. Insert blood leak chamber intothe blood leak detector on the let

    side o Aquarius (above the yellowscale)

    F. Attach the pressure domes to theour pressure sensors (Preflter,Filtrate, Return, Access) and closeclamps

    Filtrate

    Return

    Access

    Preflter

    Module 2: Lesson 2: Preparing Aquarius

  • 8/6/2019 Mod 2 Edwards Aquarius

    25/69

    Notes:

    2Module 2: Lesson 2: Preparing Aquarius

    I. Attach fltrate line (short line romthe blood leak detector) to theclear luer lock fltrate port on thetop side o the hemoflter

    H. Place prescribed hemoflter inflter holder and attach the redhemoflter connector to the redtop o the hemoflter and the bluehemoflter connector to the blue

    bottom o the hemoflter

  • 8/6/2019 Mod 2 Edwards Aquarius

    26/69

    Notes:

    24 Module 2: Lesson 2: Preparing Aquarius

    CVVH:

    J. Attach ree line (located on the Aqualineasitexitsthepre-dilution/dialysatepump)

    SCUF,CVVHDorCVVHDF:

    attach ree line to the port locatedabove the red end o the hemoflter

    attach ree line to the dialysate port onthe bottom side o the hemoflter

    NOTE: The line exiting thepre-dilution pump (bottom greenpump) is called the ree line. The reeline connection point determineswhat the therapy will be

    Free line

    Free line

    connection

    Free line

    connection

  • 8/6/2019 Mod 2 Edwards Aquarius

    27/69

    Notes:

    2Module 2: Lesson 2: Preparing Aquarius

    L. Connect the red end o theAqualine to the prime collectionbag

    NOTE: Aquarius primes with areverse ow through the bloodcircuit

    K. HangemptyprimecollectionbagontheIVpolelocatedontherighttopsideoftheAquarius

    M.Hangaone-literbagofprimingsolutionontheIVpole(usuallyheparinizednormalsaline)

    InsertthespikedY-connectorprovidedwiththeAqualine into the priming solution bag

    N. Remove the blue spiked end rom the

    bloodline and connect it to one luerlockportoftheY-connector.(IfyouarenotusingaY-connector, connect theblue spiked end o the bloodline to thepriming solution.)

  • 8/6/2019 Mod 2 Edwards Aquarius

    28/69

    Notes:

    26 Module 2: Lesson 2: Preparing Aquarius

    O. Hang empty fltrate bag(s) on

    yellowfltrate scale and connectthe fltrate line (long line out oblood leak detector) to the bag(s)

    P. Hang prepared solution(substitution/dialysate) on greensubstitution scale and connect the

    substitution/dialysate line

    NOTE: Aquarius uses a singlesolution source that is split bythe machine and delivered asboth dialysate and substitution

    solution, depending upon thetherapy chosen. Up to our bagso solution may be used at onetime. I using more than one bag, amultiport maniold will be required.

    To ensure proper, simultaneousemptying o multiple bags, all

    clamps must be open.

    NOTE: Ensure that an equalnumber o bags is used on thesubstitution/dialysate scale andthe fltrate scale. Use empty5-liter bags on the ltrate scale.For example, if TWO 2.5-literbags are used on the substitution/dialysate scale, use TWO 5-liter

    empty bags on the fltrate scale

  • 8/6/2019 Mod 2 Edwards Aquarius

    29/69

    Notes:

    2

    4. Prepare anticoagulant syringe

    B. Select the volume in the syringe andconirm. Wait until the syringe drivermoves to the correct position

    A. Aquarius may use a 50 mL or 60 mLsyringe flled to a maximum o 50 mL.Only use the heparin syringe typethat the Aquarius System has beencalibrated to use (or example, a BD50 cc syringe). The Aquarius Systemmust be calibrated or the particulartype o syringe you are using by acertifed technician. A list o approvedsyringes is available in section 6.5o Aquarius Operators Manual. The

    syringe size that should be used isshown at the right o the Preparesyringe screen

    Module 2: Lesson 2: Preparing Aquarius

    WARNING: Clamp anticoagulant linei no anticoagulant is used!

  • 8/6/2019 Mod 2 Edwards Aquarius

    30/69

    Notes:

    28

    D Install syringe into pump

    E. Prime anticoagulant line by selecting and confrming Prime anticoagulant line. Each press o

    the selector button delivers one mL

    F. Program heparin inusion rate

    Syringe attached

    Syringe in pump

    C. Attach syringe to the anticoagulantline on the tubing set

    Module 2: Lesson 2: Preparing Aquarius

  • 8/6/2019 Mod 2 Edwards Aquarius

    31/69

    Notes:

    2

    PRIMING AQUARIUS

    1. Select Next to go to Priming

    A. Check all lines or proper connection and open all clamps

    WARNING: When No anticoagulant is selected, the clamp on the anticoagulant line MUSTremain closed!

    B. Confrm that the correct solutions are used

    WARNING: The solutions used or CRRT with Aquarius are prescription uids. To ensure

    patient saety, please check that the prescribed solutions have been selected and properlyprepared

    2. Select Start priming

    A. The priming procedure requiresabout 800 mL o saline and takesapprox. 9 min (Aquarius displaysthe message Priming. Pleasewait and the clock counts downtozero)

    B. Thepre-andpost-dilutionlinesare primed with uid rom thesubstitution/dialysate bag(s).

    The blood circuit and the fltratelines are primed with uid romthe priming solution bag (usuallyheparinizednormalsaline)

    Module 2: Lesson 2: Preparing Aquarius

  • 8/6/2019 Mod 2 Edwards Aquarius

    32/69

    Notes:

    30

    C. Lines being primed are highlightedand Access, Return and TMP

    pressures are displayed

    D. Priming starts at 80 mL/min orthe frst 6 minutes (approximately500 mL) then increases to 150mL/min

    .

    E. When priming is completed, youwill hear an alarm and see themessage Priming completed

    F. I needed, options are available toreprime parts, or all o the circuit

    G. I the priming procedure issatisactory, select Next andconfrm

    Module 2: Lesson 2: Preparing Aquarius

  • 8/6/2019 Mod 2 Edwards Aquarius

    33/69

    Notes:

    3

    CLAMP & PRESSURE TEST

    The Clamp & pressure test checks the clamp, the air detector, and the access, return and

    pre-lterpressures

    It MUST be done beore you can proceed to connection mode. Connect access and return

    linestoasinglebagofnormalsalineusingtheY-connectorprovidedaspartoftheAqualine,

    or any other appropriate connector

    The test takes just a ew moments. I the test ails, ollow the instructions displayed on the

    screen to determine the point o ailure. Check connections and select Retest

    When Clamp & pressure test is completed, Aquarius switches to Start connection mode

    Select Go to recirculationAquarius switches to Recirculation mode

    Press the Blood Pump key to start the ow o saline within the tubing set. You are now

    recirculating saline inside the circuit

    Module 2: Lesson 2: Preparing Aquarius

  • 8/6/2019 Mod 2 Edwards Aquarius

    34/69

    Notes:

    32

    RECIRCULATION MODE

    RecirculationmodecanbeenteredimmediatelyafterClamp&pressuretestorduringa

    treatment when theres the need to disconnect a patient temporarily. Temporary disconnection

    will be discussed in the Disconnection and End Treatment sections

    Recirculation o normal saline through the circuit prior to patient connection may acilitate the

    removal o air trapped in the circuit1

    Recirculation may be continued through the blood circuit until connection is needed

    Programming o prescribed parameters may be done during recirculation (recommended)

    Select Go to programming at any time during recirculation to enter treatment parameters

    Module 2: Lesson 2: Preparing Aquarius

    1 Dr. Patrick Honore and Dr. Didier Journois

  • 8/6/2019 Mod 2 Edwards Aquarius

    35/69

    Notes:

    3

    PROGRAMMING AQUARIUS

    Allparametersareprogrammed in thisscreen

    Eachparametershows a yellowscreen that providesdetailed inormationabout the selected

    parameter only

    Programmingcanbedone ater priming,during recirculation,or immediately beoreconnection to thepatient. Parametersmay also be adjustedat any time duringtreatment

    Select Previous toreturn to the mainscreen

    SelectGotoconnection whenready to connect thepatient

    USING THE SELECTOR BUTTON:

    Turn: to highlight the parameter

    Push once: to select the parameter

    Turn right: to

    Turn left: to

    Push once: to confrm new values

    Module 2: Lesson 2: Preparing Aquarius

  • 8/6/2019 Mod 2 Edwards Aquarius

    36/69

    Notes:

    34

    PROGRAMMING PARAMETERS

    The parameters available include:

    Description

    This parameter is optional and may be left at 0:00 h:min, unless the fluid removal rate is setto 0 mL/hour. When set, the Aquarius will run for the indicated length of time and thendisplays the message 'Therapy target achieved by time'. The operator may now programmore time to continue the treatment or move to the disconnection phase

    Enter the amount of fluid that is prescribed for the patient to lose per hour

    Enter the amount of fluid to be removed. After this goal is achieved, the Aquarius will displaythe message 'Therapy target achieved by fluid loss'. If there is a fluid loss rate set, then thisvalue must be programmed

    Available in CVVH and CVVHDF modes only, this is the amount of fluid that will be deliveredpost-filter

    Available in CVVHD and CVVHDF modes only, this is the amount of fluid that will be deliveredas dialysate

    Available in CVVH mode only, this is the amount of fluid that will be delivered pre-filter

    The number of bags hanging from the substitution/dialysate scale. An equal number ofempty bags must also be hung from the filtrate scale

    This is the infusion rate from the integrated anticoagulation syringe pump in mL/hour

    If a bolus of heparin is prescribed, enter it here for immediate administration when in treat-ment mode

    When set at 0, the fluid warmer will be off. It is adjustable from 35 to 39 C (95 to102.2 F). The warmer heats both substitution and dialysate fluid

    Parameter

    Module 2: Lesson 2: Preparing Aquarius

  • 8/6/2019 Mod 2 Edwards Aquarius

    37/69

    Notes:

    3

    CONNECTION MODE

    SelectSingleConnectionorDoubleConnectionandfollowtheinstructionsonthescreen

    PressBloodPumpkeytostarttreatment

    Only the access line is attached to patients access port, initially, and theblood pump is started to discard the saline rom the circuit, up to the airdetector. A blood sensor, located inside the air detector, will automaticallystop the pump

    Both access and return lines are simultaneously attached to the patientsaccess and return ports

    Module 2: Lesson 2: Preparing Aquarius

  • 8/6/2019 Mod 2 Edwards Aquarius

    38/69

    (page intentionally blank)

  • 8/6/2019 Mod 2 Edwards Aquarius

    39/69

    Lesson 3: Treatment Mode

  • 8/6/2019 Mod 2 Edwards Aquarius

    40/69

    (page intentionally blank)

  • 8/6/2019 Mod 2 Edwards Aquarius

    41/69

    Notes:

    Module 2: Lesson 3: Treatment Mode 3

    TREATMENT MODE

    The Treatment screen displays pertinent inormation and allows the operator to monitor severalparameters throughout the therapy:

    Blood ow

    Cumulative totals o uid loss and substitution/dialysate since last total reset (heparin is notreset)

    Pressures (Access, Return, TMP and Pressure Drop)

    -SeedetailedinformationonPressuresinModule2Lesson4

  • 8/6/2019 Mod 2 Edwards Aquarius

    42/69

    Notes:

    Module 2: Lesson 3: Treatment Mode40

    Allows the operator to returnto the programming screenany time during treatment tochange parameters

    Allows the operator toview additional treatmentinormation, with no ability tomake changes

    Provides the ollowing options:-Viewhistory-Gotorecirculation

    -Endtreatment-Changesyringe-Changetherapy

    Screen selection choices:

  • 8/6/2019 Mod 2 Edwards Aquarius

    43/69

    Notes:

    Module 2: Lesson 3: Treatment Mode 4

    The history o the last three treatments is available. Data is visible as a list or a graph. Pressures,

    programmed parameters, patient data and events are stored.

  • 8/6/2019 Mod 2 Edwards Aquarius

    44/69

    Notes:

    Module 2: Lesson 3: Treatment Mode42

    DISCONNECTING PATIENT FROM AQUARIUS

    The patient may be disconnected rom the Aquarius by selecting and confrming Recirculation orEnd treatment rom the Options screen.

    Disconnecting the patient rom the Recirculation screen allows temporary disconnection ora limited period o time. The flter, lines, bags and solutions all remain on the machine andthe blood pump runs in recirculation mode until the patient is ready to be reconnected to thesame circuit

    Disconnecting the patient rom the End treatment screen will mean the patient treatment willend. The flter lines, bags and all solutions must be removed rom the machine and Aquarius mustbe turned o. A new set up will be needed i the patient therapy is to be continued

    REMINDER: When disconnecting patients rom Aquarius, there may be circumstancesin which blood should NOT be returned to the patient. These circumstances mayinclude, but are not limited to: cardiac arrest, uid overload, major blood leak, a circuitthat was primed with blood or albumin, hemolysis or a clotted circuit. Always checkwith the prescribing physician.2

    NOTE: The ollowing supplies may be needed:

    1. One bag o normal saline or rinsing blood back to patient

    2. Two saline syringes and the prescribed anticoagulant to lock the vascular

    catheter

    3. Caps or the vascular catheter

    CAUTION: Always ollow hospital policy or standard precautions. Gloves, mask and aace shield should be worn when connecting or disconnecting blood lines rom patientsand removing lines rom Aquarius

    2 Dr. Noel Gibney

  • 8/6/2019 Mod 2 Edwards Aquarius

    45/69

    Notes:

    Module 2: Lesson 3: Treatment Mode 4

    RECIRCULATION MODE FOR TEMPORARY DISCONNECTION

    Go to the Options screen and select

    Thisoptionstartsthetemporarydisconnectionofthepatient.Thescreenmenuguidestheoperator through the disconnection procedure

    On-screendirectionsguidetheoperatorthroughtheaccessandreturndisconnectionprocess

    Volumeofreinfusionuidisdisplayed

    AlltreatmentparametersarestoredinHistory

  • 8/6/2019 Mod 2 Edwards Aquarius

    46/69

    Notes:

    Module 2: Lesson 3: Treatment Mode44

    END TREATMENT MODE

    Beore ending treatment,please record all pertinentdata rom the Treatmentscreen

    REMINDER: Once youconrm the End treat-ment selection, you willnot be able to go back

    To end the treatment anddisconnect the patient romthe Aquarius, go to theOptions screen and select

    Follow instructions on thescreens to disconnect thepatient rom the Aquarius

    The volume o fuid used toreturn blood to the patientwill be recorded on thedisconnection screens as thereinusion volume in milliliters

  • 8/6/2019 Mod 2 Edwards Aquarius

    47/69

    Notes:

    Module 2: Lesson 3: Treatment Mode 4

    Ifsystemhasclottedandanycircuitpressureisgreaterthan400mmHg,followon-screen

    instructions or relieving pressure

    Remove Aqualine and bags rom Aquarius

    WARNING: Never switch o the Aquarius beore the tubing set is completely removed

    is highlighted. Push the selector button to turn the Aquarius o

    WARNING: Removing Aqualine under high pressure conditions may lead to a blood leak omembrane rupture in pressure domes

  • 8/6/2019 Mod 2 Edwards Aquarius

    48/69

    (page intentionally blank)

  • 8/6/2019 Mod 2 Edwards Aquarius

    49/69

    Lesson 4: Pressures, Alarms and Fluid Balance

  • 8/6/2019 Mod 2 Edwards Aquarius

    50/69

  • 8/6/2019 Mod 2 Edwards Aquarius

    51/69

    Module 2: Lesson 4: Pressures, Alarms & Fluid Balance 4

    AQUARIUS PRESSURE MONITORING

    All extracorporeal circuits are run by pressures. Aquarius uses our pressure readings to monitorand control the system. Three pressures (Access, Return and Pre-lter) monitor the blood sideand one pressure (Filtrate) monitors the fuid side. All our pressures are used in combination tohelp you manage the treatment saely and eectively.

    ACCESS PRESSURE

    Typically monitors the negative pressure

    generated to remove blood rom the patient

    Access pressure may be positive under

    certain conditions (example: ECMO)The patients catheter and Aquarius

    blood fow rates aect this pressure

    Measuringrange:-250to+100mmHg

    Upperalarmlimit:automaticsetting

    between-50and+100mmHg

    Lower alarm limit: automatic setting

    between-250and-100mmHg

    Alarm window size during treatment:

    +/-100

    mmHg

    around

    the

    actual

    value

    Displayed

    on

    the

    Treatment

    screen

    RETURN PRESSURE

    Monitorsthepositive pressure generated

    to return the blood to the patient

    The patients catheter and Aquarius

    blood fow rates aect this pressure

    Clottingincircuitaffectsthispressure

    Measuringrange:0to+350mmHg

    Upperalarmlimit:automaticsettingbetween110and350mmHg

    Loweralarmlimit:automaticsettingbetween10and250mmHg

    Alarmwindowsizeduringtreatment:+/-100mmHg

    DisplayedontheTreatmentscreen

  • 8/6/2019 Mod 2 Edwards Aquarius

    52/69

    Notes:

    Module 2: Lesson 4: Pressures, Alarms & Fluid Balance50

    PRE-FILTER PRESSURE

    Monitorsthepositive pressure

    generated in the circuit immediately

    beore the blood enters the

    hemolter

    Bloodowrateandreturnpressure

    aect this pressure

    Highestpressurepointinthecircuit

    Arisingpre-lterpressuremaybe

    indicative o lter clotting

    Upperlimit:+400mmHg

    Lowerlimit:-100mmHg

    DisplayedontheMorescreen

    FILTRATE PRESSURE

    Monitorsthenegativeand/orpositive

    pressure generated in the ltrate

    compartment o the hemolter. Thisaects the movement o fuid across

    the membrane

    Arisingltratepressuremaybe

    indicative o lter clotting

    Bloodowratesanduidowrates

    aect this pressure

    Upperlimit:+400mmHg

    Lowerlimit:-400mmHg

    DisplayedontheMorescreen

  • 8/6/2019 Mod 2 Edwards Aquarius

    53/69

    Notes:

    Module 2: Lesson 4: Pressures, Alarms & Fluid Balance 5

    CalculatedusingReturn,Pre-lterandFiltratepressures

    Calculatedas:ReturnPressure+Pre-lterPressure _

    2

    TMPisacombinationofthetotal

    pressure exerted on the membrane

    (positive pressure inside the bers and

    negative pressure outside the lters)

    Upperlimit:+400mmHg(CRRT)

    Upperlimit:+100mmHg(TPE)

    Lowerlimit:-30mmHgDisplayedontheTreatmentscreen

    PD (PRESSURE DROP)

    Calculatedusingpressurelossfromtop

    to bottom o hemolter

    Lengthofhemolteraffectspressure

    Calculatedas:

    Pre-lter Pressure Return Pressure

    DisplayedontheTreatmentscreen

    TMP (TRANSMEMBRANE PRESSURE)

    Filtrate Pressure

  • 8/6/2019 Mod 2 Edwards Aquarius

    54/69

    Notes:

    Module 2: Lesson 4: Pressures, Alarms & Fluid Balance52

    AQUARIUS ALARMS

    The Aquarius saety mode is determined by the nature o occurring alarms:

    Alarm type Aquarius safety mode

    Alarms in blood circuit Stops all pumps

    Closes return line clamp

    Generates visual and audible alarms

    Alarms in filtrate / dialysate circuit Stops filtrate pump

    Stops pre-dilution pump and post-dilution pump

    Generates visual and audible alarmsSystem error Stops all pumps

    Closes return line clamp

    Generates visual and audible alarms

    Alarmsignalsareaudible and visual

    Thestatuslightschangefromgreentored,yellow,oracombinationdependingonthetypeof

    alarm

    Cause o the alarm is displayed in a message box in the lower let corner o the screen AlarmsmaybesilencedfortwominutesbypressingtheMutekey

    BloodcircuitalarmsareclearedbypressingtheBloodPumpkey.Oncethebloodpumphas

    beenrestarted,presstheBalanceStart/Stopkeytorestarttheuidpumps

    FluidcircuitalarmsareclearedbypressingtheBalanceStart/Stopkey

  • 8/6/2019 Mod 2 Edwards Aquarius

    55/69

    Notes:

    Module 2: Lesson 4: Pressures, Alarms & Fluid Balance 5

    BLOOD CIRCUIT ALARMS

    Generatesaudibleandvisualalarm

    Stopsallpumps

    LEDinBloodPumpandBalanceStart/Stopkeyashes

    Closesreturnlineclamp(ifairdetectedorreturnpressuredropsbelowlowerlimitof

    +10mmHg)

    FLUID CIRCUIT ALARMS

    Generatesaudibleandvisualalarm

    Stopsalluidpumps LEDinBalanceStart/Stopkeyashes

    BloodPumpcontinuestorun

    SYSTEM ALARMS

    Generatesaudibleandvisualalarms

    Stopsallpumps

    LEDinBloodPumpandBalanceStart/Stopkeyashes

    Closesreturnlineclamp(ifairdetectedorreturnpressuredropsbelowlowerlimitof

    +10mmHg)

    NOTE: ABalancealarmoccurswhen a50mL (20mLwith useof AqualineSdierence is detected between the fuid volume pumped and the scalemeasurement o the substitution or ltrate solutions. This volume may result in anadditional50mL(20mL)infusiontothepatientor50mL(20mL)ofuidremovedrom the patient i a tubing line is kinked or blocked. Correct the cause o the alarmandpresstheBalanceStart/Stopkeytoresumetreatment

    WARNING: I repeated balance alarms occur, stop treatment to protect the patient rom

    excessive fuid removal or overload. Call technical support to resolve the problemDonotusethemachineuntilithasbeenrepaired

  • 8/6/2019 Mod 2 Edwards Aquarius

    56/69

    TROUBLESHOOTING SYSTEM ERRORS, ALARMS AND MESSAGES

    System Errors: Red indication

    I the ollowing system errors are not corrected ater system test and i during treatment nocorrection is possible, please call technical service to repair the Aquarius.

    During the system test (after switching the system on) andalsoduringoperation theAquarussystem automatically perorms tests to check the saety-critical components. I errors occurduring these tests, the system switches to the saety mode and generates a system errormessage.ThiserrormessageappearswiththeabbreviationCPU1(controlprocessorunit1)orCPU2.

    Display: Cause: Options for error removal:

    Accesssensor Access pressure values deviatefrom limits.

    Control system and protectivesystem values deviate fromeach other (outside of limits)

    During clamp/pressure test nopressure increase is detected

    Restart system (no tubing must be installed). If theerror cannot be reset, put system out of operationand notify technical support.

    Access pressure dome may not be connectedcorrectly (ensure by pressing the dome firmlytowards the access transducer)

    ADC/Voltage Voltage supply or AD-converterfailure

    Restart system (no tubing must be installed). If theerror cannot be reset, put system out of operationand notify technical support.

    Air detector Air detector test failed. Press blood pump key. If error cannot be reset, putsystem out of operation and notify technical support.

    Backup

    No dates at the backup Restart system (no tubing may be installed). If theerror cannot be reset, put system out of operationand notify technical support.

    Balance filtration Check filtration scale

    Values between protective andcontrol system deviate fromeach other (outside of limits).

    Actual values are outside oflimits.

    Restart system (no tubing may be installed). If theerror cannot be reset, put system out of operationand notify technical support.

    Balance substitution Check substitution scale

    Values between protective andcontrol system deviate fromeach other (outside of limits).

    Actual values are outside oflimits.

    Restart system (no tubing may be installed). If theerror cannot be reset, put system out of operationand notify technical support.

    BLD (Blood leak detector) Blood leak detector does notwork properly.

    Press the Blood pump key. If the system errorcannot be reset, put system out of operation andnotify technical support.

    *Rev. A - 12/2007

    Notes:

    Module 2: Lesson 4: Pressures, Alarms & Fluid Balance54

  • 8/6/2019 Mod 2 Edwards Aquarius

    57/69

    Notes:

    Module 2: Lesson 4: Pressures, Alarms & Fluid Balance 5

    Display: Cause: Options for error removal:

    Clamp doesnt closeClamp doesnt open

    Clamp test failed.

    Clamp does not close.

    Clamp does not open.

    Correct tubing set position in clamp.

    Press blood pump key. If the error cannot be reset,put system out of operation and notify technicalsupport.

    CODE Controller CPU program code

    test failed Master CPU program code test

    failed

    Restart system (no tubing may be installed). If the

    error cannot be reset, put system out of operationand notify technical support

    Commu front system Data transfer to display systemdefective.

    Set up safety mode for the patient.

    Switch system off and on again after approx.1 minute (no tubing must be installed). If the errorcannot be reset, put system out of operation andnotify technical support.

    Commu protection system Error during data transfer.

    Power supply for protectivesystem is defective.

    Press blood pump key or Balance Start/Stop key.

    Switch system off and on again after approx.1 minute. If the error cannot be reset, put systemout of operation and notify technical support.

    EEPROM Master CPU calibration datatest failed

    Controller CPU calibration datatest failed

    Restart system (no tubing may be installed). If theerror cannot be reset, put system out of operation

    and notify technical support

    Error1 CPU Master CPU Register test failed

    Controller CPU Register testfailed

    Restart system (no tubing may be installed). If theerror cannot be reset, put system out of operationand notify technical support

    Error2 CPU Master CPU RAM test failed

    Controller CPU RAM test failed

    Restart system (no tubing may be installed). If theerror cannot be reset, put system out of operationand notify technical support

    Error3 CPU Master CPU jump test failed

    Controller CPU jump test failed

    Restart system (no tubing may be installed). If theerror cannot be reset, put system out of operationand notify technical support

    Blood pump Flow rate test failed.

    Blood pump drive defective.

    Blood pump is not stopped.

    Actual value of number ofrevolutions deviates from setvalue outside of limits.

    Switch system off and on again after approx. 1minute (no tubing must be installed).

    Press blood pump key.

    If the error cannot be removed, notify technicalsupport.

    Error blood detection Blood at optic sensor(airdetector)

    Restart system (no tubing may be installed). If theerror cannot be reset, put system out of operationand notify technical support.

  • 8/6/2019 Mod 2 Edwards Aquarius

    58/69

    Notes:

    Module 2: Lesson 4: Pressures, Alarms & Fluid Balance56

    Display: Cause: Options for error removal:

    Heater Master hasnt pass the systemtest

    Restart system (no tubing may be installed). If theerror cannot be reset, put system out of operationand notify technical support.

    Operation mode The data transfer valuesregarding the operation modedeviate between master- andcontroller CPU.

    Press blood pump key to start a new check.

    If the alarm appears again, call technical support.

    Postdilution pump Flow rate test failed. Postdilution pump drive

    defective.

    Postdilution pump is notstopped.

    Actual value of number ofrevolutions deviates from setvalue outside of limits.

    Switch system off and on again after approx.1 minute (no tubing may be installed).

    Press Balance Start/Stop key.

    If the error cannot be reset, put system out ofoperation and notify technical support.

    Program run Program failure Master CPU

    Program failure Controller CPU

    Restart system (no tubing may be installed). If theerror cannot be reset, put system out of operationand notify technical support.

    Filtration pump Flow rate test failed.

    Filtrate pump drive defective.

    Filtrate pump is not stopped.

    Actual value of number ofrevolutions deviates from setvalue outside of limits.

    Switch system off and on again after approx. 1minute (no tubing must be installed).

    Press Balance Start/Stop key.

    If error cannot be reset, notify technical support.

    Heparin pump Actual values of control andprotective systems deviate fromeach other (outside of limits)

    Actual values deviate fromlimits

    After selftest: Restart system (no tubing must beinstalled). If the error cannot be reset, put systemout of operation and notify technical support.

    During treatment:

    1. Check the Heparin line is not clamped.

    2. Go to "Options" and "Change Syringe", followingthe on-screen text. Please Note: You DO NOT haveto remove the syringe during this process

    3. If problem persists, program the pump to 0,clamp the line and remove the syringe. If problempersists, end treatment and call technical service.

  • 8/6/2019 Mod 2 Edwards Aquarius

    59/69

    Notes:

    Module 2: Lesson 4: Pressures, Alarms & Fluid Balance 5

    Display: Cause: Options for error removal:

    Predilution pump Flow rate test failed.

    Predilution pump drivedefective.

    Predilution pump is notstopped.

    Actual value of number ofrevolutions deviates from setvalue outside limits.

    Switch system off and on again after approx.1 minute (no tubing must be installed).

    Press Balance Start/Stop key.

    If the error cannot be reset, put system out ofoperation and notify technical support.

    Return sensor Return pressure values deviatefrom limits.

    Values between protectivesystem and control system

    deviate from each other(outside of limits).

    During clamp/pressure test nopressure increase is detected

    Restart system (no tubing may be installed). If theerror cannot be reset, put system out of operationand notify technical support.

    Clamp pressure test:o Press domes firmly towards the transducer.

    Return and pre filter domes must be connectedcorrectly

    o Insert tube into return line clampo Ensure that there is no leakage between blood

    pump and return line clamp

    Sensor Voltage Voltage supply or AD-converterfailure

    Restart system (no tubing may be installed). If theerror cannot be reset, put system out of operationand notify technical support.

    Software version Different Software versiondetected at master andcontroller

    Restart system (no tubing may be installed). If theerror cannot be reset, put system out of operationand notify technical support.

    Timer Timer difference betweenmaster and controller > 20%

    Restart system (no tubing may be installed). If theerror cannot be reset, put system out of operation

    and notify technical support.TMP sensor Filtrate and/or prefilter pressure

    values deviate from limits.

    Values between protectivesystem and control systemdeviate from each other(outside of limits)

    Restart system (no tubing may be installed). If theerror cannot be reset, put system out of operationand notify technical support.

    Vcc Master/communication RAM, EPROM or EEPROM aredefective.

    Switch system off and on again after approx.1 minute. If the system error cannot be reset, putsystem out of operation and notify technical support.

    XRAM Master CPU extern RAM failed

    Controller CPU external RAMtest failed

    Restart system (no tubing may be installed). If theerror cannot be reset, put system out of operationand notify technical support.

  • 8/6/2019 Mod 2 Edwards Aquarius

    60/69

    Notes:

    Module 2: Lesson 4: Pressures, Alarms & Fluid Balance58

    Alarms: Red IndicationI the Aquarius system detects an out-o-range condition during the system test or duringoperation or i parameters exceed or drop below the respective limits, an alarm message isgenerated and the Aquarius system switches to the saety mode.

    Display:

    Cause:

    Options for error removal:

    High accesspressure

    Low accesspressure

    Access pressure has exceededthe upper alarm limit.

    Access pressure has droppedbelow the lower alarm limit.

    Check access blood line, including access and pre-filter sensors, for kinks or occlusions.

    Check position of patient access.

    Check blood flow rate.

    Air Detected Return tubing line contains air ormicro-foam.

    Make sure that tubing line does not contain air.

    Check access and filter connections for sources of airleaks.

    When you clear the Air Detected alarm, make surethere is no air or foam trapped in the line between thereturn detector chamber and the patient end.

    To remove air from the tubing line:Step1. Open the return line clamp.Step 2. Remove all air from the tubing set.Step 3. Close the return line clamp.Step 4. Resume treatment by starting the bloodpump.

    Warning! If the air detected alarm does notclear and air is visible in the venous de-aerationchamber, disconnect the patient from the instrumentand recirculate per your centers procedure.

    Note: You may see micro-bubbles smaller than the airdetector sensitivity.

    Balance system off The balance system has been offfor 5 minutes.

    Switch on

  • 8/6/2019 Mod 2 Edwards Aquarius

    61/69

    Display:

    Cause:

    Options for error removal:

    15 seconds.

    Clamp heparin line The Heparin syringe has beenremoved

    Clamp Heparin line

    Degassing chamber missing The substitution degassingchamber is not inserted or thesensors are defective.

    Insert the substitution chamber properly. Ensure thechamber is in contact with the sensor of the holder. Ifthis doesnt help switch off heater to end the

    treatment. Then call technical service.High filtrate pressure

    Low filtrate pressure

    Filtrate pressure exceeds or fallsbelow the alarm limits.

    Check pressure sensor.

    Check tubing set for narrow sections.

    Check filter and exchange if required.

    Check blood flow-to-fluid removal or blood flow-toturnover ratio.

    Filtrate flow failure The number of revolutions of thefiltrate pump exceeded or fell

    below the alarm limits by 5%.

    Restart treatment

    If problem persist call a repair technician

    Heparin syringe missing A Heparin rate has beenprogrammed and no syringe hasbeen inserted in the plunger.

    The Heparin syringe is not

    inserted properly

    Insert Heparin syringe if Heparin is needed

    Set Heparin rate to zero ( No Heparin required)

    Balance Alarm

    Check substitution line

    Or

    Balance alarm

    Check filtration line

    Balancing deviates from the setvalues entered by the operator.

    Check flow rates of the pumps.

    Check fluid removal and turnover input parameters.Note: Potentially pumps can't deliver program withvery high volume because of pressure peaks.

    Check bag hanging on scale.(Filtration or substitution)

    Check tubing set for narrow sections.(Filtration or substitution lines) .

    Blood leak Filtrate / plasma contains blood. Check filter for blood leak and exchange filter ifrequired.

    Check pressure conditions.

    Blood flow failure The number of revolutions of thepump exceeds or falls below the

    alarm limits by 5%.

    Restart blood pump.

    If problem persist call technician

    Blood pump off The blood pump was manuallyswitched off for 1 minute.

    Press blood pump key to switch blood pump onagain.

    Check transducer connections The pressure domes have notdetected any pressure change for

    Ensure the domes are properly connected.

    Notes:

    Module 2: Lesson 4: Pressures, Alarms & Fluid Balance 5

  • 8/6/2019 Mod 2 Edwards Aquarius

    62/69

    Notes:

    Module 2: Lesson 4: Pressures, Alarms & Fluid Balance60

    Display: Cause:

    Options for error removal:

    Indication battery high At the power fail indication batteryhigh voltage is detected

    Change battery

    Control charging voltage

    Control AD-converter/CPU

    Keyboard failure A key press longer than60 seconds. detected by master-cpu.

    The keyboard needs repair

    Line/substitution failure A substitution deviation is detectedthat influences the patient balance

    Check lines

    Check clamps

    Check bags

    Check for leakages

    Master key transfer A key press longer than 60

    seconds. detected by master-cpu.

    The keyboard needs repair

    The communication cable is defective

    Main battery high At the main battery high voltage isdetected

    Control charging voltage/unit

    Control/change battery

    Control AD-converter/CPU

    Post-dilution failure The number of revolutions of thePost-dilution pump exceeded or

    fell below alarm limits by 5%.

    Restart treatment

    If problem persist call technician.

    Pre-dilution failure The number of revolutions of thePre-dilution pump exceeded or fell

    below alarm limits by 5%.

    Restart treatment

    If problem persist call technician.

    Pump door One of the pump doors open Check if one of the pump doors is open

    Close the door

    High pre-filter pressure

    Low Pre-filter pressure

    The pre-filter pressure exceedsupper alarm limit. Check pressure sensor. Check filter and exchange if required.

    Check blood flow.

    Check tubing set for narrow sections.

    High return pressure Return pressure exceeds or falls Check return line for connection, kinks or occlusions.

    Low return pressure below the alarm limits. Check return pressure transducer.

    Check blood flow rate.

    Check position of patient access.

    Syringe removed The Heparin syringe has beenremoved

    Check the heparin syringe

  • 8/6/2019 Mod 2 Edwards Aquarius

    63/69

    Notes:

    Module 2: Lesson 4: Pressures, Alarms & Fluid Balance 6

    Display:

    Cause:

    Options for error removal:

    High Temperature

    Temperature controller high (Fluidpumps will not run)

    A high temperature is detected inthe substitution degassingchamber

    The substitution solutiontemperature is above40 C

    The plate temperature of theheater is above 57C

    Check the substitution line for kinks

    Open heater door to cool the heater down.

    If the alarm disappears the pump will startautomatically

    Low temperature The substitution fluid temperaturehas been below 33C for morethan 10 minutes.

    Check the heater bag

    Check substitution temperature

    Temperature controller High temperature of the heater atthe controller

    Open heater and cool down the plate temperature

    Restart treatment

    Turn over failure The substitution volumecalculated by pump revolutionsdeviate more than 350g from thesubstitution volume measured onthe scale

    Check lines

    Check clamps

    Check bags

    Check for leakages

    High TMP

    Low TMP

    TMP exceeds or drops belowrespective alarm limits.

    Check pressure sensor.

    Check tubing set for narrow line sections.

    Check blood flow-to-fluid removal or blood flow-to-turnover ratio.

  • 8/6/2019 Mod 2 Edwards Aquarius

    64/69

    Notes:

    Module 2: Lesson 4: Pressures, Alarms & Fluid Balance62

    Messages: Yellow Indication

    I the Aquarius detects out-o-range conditions or reminders that do not conorm to the intendeduse o the system, the operator gets detailed inormation dened or the individual conditions andthe system switches to the saety mode.

    Display:

    Cause:

    Options for error removal:

    Ensure the return tubing does not contain air.

    Ensure the return tubing is installed properly in theair detector sensors clamping mechanism and isnot scratched or damaged.

    After correcting the error, Select and confirmPrevious, then perform the clamp/pressure testagain to proceed.

    Air detected, Clamp/Pressuretest disabled

    The air detector detected air in thereturn line.

    Warning!If the air detected alarm cannot becleared, discontinue treatment and donot return the extracorporeal blood tothe patient.

    Anticoagulant off The Heparin rate is programmed tozero.

    If Heparin is not required proceed to next screen.

    If Heparin is required, insert a syringe containingHeparin and program the desired Heparin rate.

    Balance initializing Scales and fluid pumps initializewhen the balance system is started.

    This is just a reminder

    Occurs each time the balance system is turned ON

    Balance system off The balance system is off, all fluid

    pumps are stopped.

    Correct the cause and switch the balance system

    on again.Blood detected

    (Blood pump stops)

    During the connection orrecirculation phase blood isdetected in the return line

    The blood pump stops and starts now only for 5seconds again. Switch to treatment mode

    Blood pump off The blood pump was manuallyswitched off.

    Press blood pump key to switch blood pump onagain.

    Change substitution bag

    Change filtrate bag

    The filtrate bag has reachedmaximum permissible weight or thesubstitution solution bags do notcontain solution.

    Replace empty substitution solution bags with fornew bags filled with solution

    Replace full filtrate bags with empty bags

    Check access transducer The access/ return dome is notproperly connected.

    During clamp and pressure test no pressureincrease is found when the clamp is closed

  • 8/6/2019 Mod 2 Edwards Aquarius

    65/69

    Notes:

    Module 2: Lesson 4: Pressures, Alarms & Fluid Balance 6

    Display:

    Cause:

    Options for error removal:

    Exchange filter and set The assumed filter life time of 72hours has been exceeded or thetotal using time for the actual tubingset is more than 80 hours.(Including priming, connection,recirculation and treatment time)

    Exchange filter and tubing set

    Disconnect and start a new treatment with new filterand a new tube system

    Function not available During treatment the Off key isused

    To switch off the machine, select End treatment,perform the disconnection program until theAquarius off mode

    Filt./effluent bag change soon Filtrate/effluent bag change in lessthan 10 minutes

    Prepare the Filtrate/effluent bag change

    Heater cools down If balance system has stopped for aprolonged period, the heater platetemperature is above 43C.

    The pumps stop until the temperature is in a safecondition (below 42 C). This may take up to 10min. The pumps will start automatically.

    Heater self test running.

    ! Wait

    Heater Selftest is in progress when

    the start-priming screen is reach.

    Wait until heater Selftest is completed.

    When heater Selftest is finished, the yellow lightupper the screen become fixed (green diode stillfixed during and after heater Selftest)

    Heparin syringe missing The Heparin syringe is not in theHeparin pump.

    nsert a Heparin syringe in Heparin pump or program

    the Heparin rate to zero.

    Check degassing chamber The degassing unit system hasdetected a pressure issue.

    - After Selftest: Degassing pressure sensor ordegassing module defect. Don't use the Aquariusmachine and call technician.

    - During use: Pressure sensor detects less than -300 mmHg. Degassing Filter is wet. Clamp thepressure line, disconnect filter from sensor and usea syringe to dry it. Reconnect the line and unclampit.

    Check lines The postdilution pump has beenstopped for longer than 3 min toregulate the fluid loss.

    Check the substitution line and the filtration line ifthe bags are opened, all clamps are removed andthe tubes and bag inputs are not kinked.

    End of Treatment End of Treatment Disconnect patient or reprogram patient parameter

  • 8/6/2019 Mod 2 Edwards Aquarius

    66/69

    Notes:

    Module 2: Lesson 4: Pressures, Alarms & Fluid Balance64

    Display: Cause:

    Options for error removal:

    No fluid chamber detected The substitution degassingchamber is not inserted.

    Insert the chamber correctly

    Power failure The Aquarius machine hasdetected a power failure

    Reconnect the Aquarius to main power supply

    Return the blood to the patient and start a newtreatment if power from main power supply isavailable

    Please program Hourly fluid loss or fluid loss total isnot programmed

    Select programming mode and program both hourlyfluid loss and fluid loss total.

    High filtration ratio 1. Fluid removal rate exceeds 33%of the blood flow rate.

    2. Exchange of fluid or plasmaacross the membrane is too high incomparison to the blood flow rate.

    3. The post dilution substitution rateis higher than what is acceptable forthe current blood flow rate.

    Decrease fluid removal or plasma exchange rate

    Increase blood flow rate

    Evaluate ratio of pre vs. post dilution substitution

    solutions.

    Indication battery low Power failure indication (beep)battery is low

    The battery is automatically charged if you go onwith this treatment.

    This message indicates that in the event of a power

    failure the Aquarius will beep less than 2 min

    Insert BLD chamber The blood leak detector has notbeen inserted into the blood leakchamber.

    Insert the chamber

    Reprime to correctly fill up the chamber.

    Ensure no scratches or marks are present on theAqualine chamber

    Insert tube

    Insert tube to air detector

    The air detection system is notoperational after priming

    Insert correctly return line into the air detectionsystem.

    Make sure that the air detection system is wellinserted, if not push it back firmly.

    Ensure the green light of the clamp key is lighting.

    Main battery low After a power failure, the mainpower supply battery must becharged.

    The battery is automatically charged if you go onwith this treatment.

    This message indicates that in the event of a powerfailure the Aquarius will run less than 2 min

    Negative UF A negative UF is programmed This is just a reminder

    No bag No bag was hung on the scale Hang empty collection bag and/or substitutionsolution bag on the scale

  • 8/6/2019 Mod 2 Edwards Aquarius

    67/69

    Notes:

    Module 2: Lesson 4: Pressures, Alarms & Fluid Balance 6

    Display: Cause:

    Options for error removal:

    Please wait The balance system is stopped This is an indication that the system will startautomatically after some minutes.

    Program dialysate In CVVHD dialysate rate is notprogrammed.

    Select programming and program a dialysate rate.

    Program goal Treatment goal is not programmed. Select programming and program either time, fluidloss and fluid loss total, or both.

    Program treatment pumps In CVVH/CVVHDF pre- and post-dilution/post-dilution and dialysateare not programmed.

    Select programming and program pre- and post-dilution or post-dilution and dialysate rate.

    Pump door open One of the pump doors is open. Close pump door(s).

    Read error help instructions To solve the alarm furtherinformation is needed Further information is available from the helpscreens.

    Return pressure low The return pressure is below 10mmHg

    During the first minute of treatment this is areminder.

    Syringe empty. Go to OPTION tochange

    The syringe located in the Heparinpump is empty.

    Clamp Heparin line.

    Take syringe out of pump and disconnect from line.

    Fill new syringe with Heparin.

    Enter syringe volume and confirm.

    Place syringe in pump and connect line.

    Confirm

    Subst./dialysate bag changesoon

    Substitution/dialysate bag changein less than 10 minutes

    Prepare Subst./dialysate bag change

    Too much weight Too many bags are on the scale. Ensure that the same number of substitution

    solution and filtrate bags hang on the scale hooks. Note: The maximum number of bags on the scale is

    4 substitution solution and 4 filtrate bags.

  • 8/6/2019 Mod 2 Edwards Aquarius

    68/69Module 2: Lesson 4: Pressures,Alarms & Fluid Balance66

    Notes:

    Module 2: Lesson 4: Pressures, Alarms & Fluid Balance66

    Notes:

    AQUARIUS FLUID BALANCE

    ProgrammedpatientuidremovalwillberecordedhourlyontheTreatmentscreenasacumulative value

    Non-Aquariusinsandoutsforthepatientmustbecalculatedandcontrolledbytheuser

    UFvariationismonitoredanddisplayedintheMorescreen.Theaccuracyis50mLforregularcircuits,20mLforlowvolumecircuits

    ResettotalsintheProgrammingscreenresetsallcumulativetotalsforuidloss,anddialysate and substitution fuid. Treatment clocks are also reset. These values are located ontheTreatmentscreen.

    Values are programmed in theProgrammingscreen

    Dialysateand/orsubstitutionuidare automatically removed by the

    Aquarius

    WhenprogrammingHourlypatientuidloss,aTotaluid

    lossMUSTalsobeprogrammed.Programming the time is optional

    Treatmentproceedsuntilaprogrammed target is achieved

    WhenPatientuidlossissetatzero, time must be programmed

    NOTE: Aquarius records all fuids as a cumulative number rom the start o treatment

    NOTE: When the target

    is reached, the message

    Therapytargetachieved

    by fuid loss (or time) is

    displayed. The target can

    be reprogrammed and

    the treatment continued

    NOTE: Anticoagulant volumes are NOT reset

  • 8/6/2019 Mod 2 Edwards Aquarius

    69/69

    Caution:Federal(U.S.)lawrestrictsthisdevicetosalebyoronthe

    order o a physician or other licensed practitioner. For sae and

    proper use o this device reer to the Operators Manual.