AQUARIUS TRAINING
TABLE OF CONTENTS
Lesson 1. Introduction to Aquarius Platinum
Lesson 2. Preparing Aquarius
Lesson 3. Treatment Mode
Lesson 4. Pressures, Alarms & Fluid Balance
These materials were created by Edwards Liesciences Corporation as part o the CriticalCare Nephrology Alliance. Seeking to provide CRRT proessionals with the highest caliber
o multidisciplinary expertise, products and services, the Alliance brings together the uniquestrengths o Baxter Healthcare Corporation, an expert in renal therapies, and Edwards, aleader in critical care.
Module 2: Aquarius Training
-
8/6/2019 Mod 2 Edwards Aquarius
4/69
(page intentionally blank)
-
8/6/2019 Mod 2 Edwards Aquarius
5/69
Notes:
Module 2: Learning Objectives
The material presented in this program is based on the Aquarius
Operators Manual, Platinum Sotware version 6 US and the
Aquarius v. 6 screens. The Operators Manual and the screens
provide the instructions necessary or the proper operation
o the Aquarius System. This training program is designed to
complement the Aquarius Operators Manual and screens. It
is not intended as a replacement. Please reer to the Operators
Manual and the screens or more detail.
The Aquarius System is indicated or continuous solute and/oruid removal in patients with acute renal ailure or uid overload.
It may also be used in Therapeutic Plasma Exchange (TPE)
therapies.
THE AQUARIUS SYSTEM
Understandtheset-up,primingandoperationoftheAquarius Hemofltration system
Understandpressuremonitoringandalarmtroubleshooting
Completethereturndemonstrationportionofthistrainingprogram
LEARNING OBJECTIVES
-
8/6/2019 Mod 2 Edwards Aquarius
6/69
(page intentionally blank)
-
8/6/2019 Mod 2 Edwards Aquarius
7/69
Lesson 1: Introduction to Aquarius Platinum
-
8/6/2019 Mod 2 Edwards Aquarius
8/69
(page intentionally blank)
-
8/6/2019 Mod 2 Edwards Aquarius
9/69
Notes:
AQUARIUS PLATINUM DESIGN CONCEPTS
Patient saety
Innovation
Flexibility
Easeofuse
Consistentmessaging
Upgradableasnewneeds/featuresarise
INTERACTING WITH AQUARIUS PLATINUMThe operator interaces with Aquarius by turning andpushing the Main Selector button. This button is a rotaryswitch located below the display screen. It is used toselect and confrm dierent unctions and to modiytreatment parameters.
USING THE SELECTOR BUTTON:
Turn: to highlight the parameter
Push once: to select the parameter
Turn right: to
Turn left: to
Push once: to confrm new values
Main Selector button
Module 2: Lesson 1: Introduction to Aquarius Platinum
-
8/6/2019 Mod 2 Edwards Aquarius
10/69
Notes:
8
Help screens will providedetailed inormation aboutindividual screens and allowdirect access to History
NAVIGATING BETWEEN SCREENS
The operator moves between screens by selecting a parameter and confrming the choice: Areyou sure messages indicate that the operator will need to confrm this choice beore movingorward. The deault choice is No.
Moves to the nextscreen
Moves back toprevious screen
Confrms theselected choice
Cancels the selectedchoice and movesback to previousscreen
Module 2: Lesson 1: Introduction to Aquarius Platinum
-
8/6/2019 Mod 2 Edwards Aquarius
11/69
Notes:
9
CONSISTENT MESSAGING
Aquarius uses clear and consistent messaging, delivered in yellow or red message boxes, toinorm and guide the operator through the screens. Example:
KEY FUNCTIONS OF THE AQUARIUS INTERFACE
Messages in a red
background describe the
alarm condition that is
currently underway and
needs the immediate
attention o the user.
Messages in a yellow box
may describe a set up
instruction, a parameter
that is being programmed
or an alarm that reminds
the user that something
needs to be done.
Blood Pump keyMute key
Clamp keyMain Selector button
Balance Start/Stop key
Module 2: Lesson 1: Introduction to Aquarius Platinum
-
8/6/2019 Mod 2 Edwards Aquarius
12/69
Notes:
10
INDIVIDUAL FUNCTION KEYS
The individual unction keys and their displayed unctions are explained below.
1. Operation status display
Three status lights indicate the dierent operation modes.
A. An illuminated red status light indicates:
Warning! Alarm or system error.
B. An illuminated green and yellow ashing status light indicates:
Caution! Heater sel test is running. Priming mode is not accessible.
C. An illuminated green and yellow status light indicates:
Caution! Treatment time has run down or Treatment was stopped.
Caution! Bag change required.
Caution! Anticoagulant syringe is empty.
Caution! Aquarius is in Preparation or Recirculation or Connection mode.
D. An illuminated green status light indicates:
Treatment is running. No alarms are active.
E. All three status lights illuminating one ater the other indicate:
Machine is perorming system test.
Module 2: Lesson 1: Introduction to Aquarius Platinum
-
8/6/2019 Mod 2 Edwards Aquarius
13/69
Notes:
1
2. Mute unction key
Pressing the Mute key enables the operator to silence the alarm signal or a period otwo minutes. The LED integrated into the key ashes. I the cause o the alarm cannot beremoved within the given period, the audible alarm is reactivated. I another alarm occursduring this period, the audible alarm is generated at once.
3. Clamp unction key
Pressing the Clamp key opens the return line clamp during air alarm to allow removal o air
bubbles rom within the tubing set. The LED integrated into the key ashes. When the air isremoved, treatment may be resumed by pressing the Blood Pump key. The return line clamp isautomatically reactivated ater one minute.
4. Main Selector button
Themainselectorbuttonisamulti-functionrotarybutton.Thesefunctionsinclude:
a) Selecting unction windows by turning the selector button
b) Confrming selected unctions by pressing the selector button
c) Selecting input parameters by turning the selector button
d) Opening the input window or the selected parameter by pressing the selector button
e) Raising the parameter input or the selected parameter by turning the selector button tothe right
) Lowering the parameter input or the selected parameter by turning the selector button tothe let
g) Confrming the entered parameter by pressing the selector button. The modifed parameteris displayed on the screen
Module 2: Lesson 1: Introduction to Aquarius Platinum
-
8/6/2019 Mod 2 Edwards Aquarius
14/6912
Notes:
12
Notes:
5. Balance Start/Stop unction key
Pressing the Balance Start/Stop key stops the fltrate and substitution/dialysate pumps. Thiscan be used to temporarily stop treatment, or example, to exchange bags. I an alarm occurs inthe fltrate and/or the substitution/dialysate circuits, the pumps stop and the LED integrated intothe key starts ashing. Ater correcting the cause o the alarm the pumps may be restarted bypressing the Balance Start/Stop key.
Balance Start/Stop key is also called Start Treatment key.
6. Blood Pump unction key
Pressing the Blood Pump key starts or stops the ow o blood through the blood circuit. I analarm occurs in the blood circuit, all pumps stop and the LED integrated in the blood pump keystarts ashing. Ater correcting the cause o the alarm the system is restarted by pressing theBloodPumpkeyagain.Filtratepump,pre-dilutionandpost-dilutionpumpsstartwithdelayinrelation to the blood pump.
The Blood Pump key is also used to immediately stop all pumps in case o an unpredictable
occurrence.
Module 2: Lesson 1: Introduction to Aquarius Platinum
-
8/6/2019 Mod 2 Edwards Aquarius
15/69
(page intentionally blank)
-
8/6/2019 Mod 2 Edwards Aquarius
16/69
(page intentionally blank)
-
8/6/2019 Mod 2 Edwards Aquarius
17/69
Lesson 2: Preparing Aquarius
-
8/6/2019 Mod 2 Edwards Aquarius
18/69
(page intentionally blank)
-
8/6/2019 Mod 2 Edwards Aquarius
19/69
Notes:
1
POWER SUPPLYAquarius has two power supply switches.The Main switch, on the let side o the machineshould be set to position 1, the ON/OFF sot keyswitch is on the right side o the monitor displayscreen.
SYSTEM TEST
The system test takes approximately our minutes. When it is fnished, an audible alarm isgenerated and the green and yellow status lights are illuminated. The pumps stop in the correct
position or the tubing set to be installed. The operator can now proceed to set up.
Main Switch ON/OFF sot key
CAUTION: The Aquarius system test must be perormed beore tubing, bags, pressuresensors and solutions are placed on the machine. The pump doors must be closed.
Module 2: Lesson 2: Preparing Aquarius
-
8/6/2019 Mod 2 Edwards Aquarius
20/69
Notes:
18
SETTING UP AQUARIUS
Select the therapy to be used and confrm1.
SCUF
CVVH
CVVHD
CVVHDF
Therapeutic plasma exchange(TPE)
2. Select the type o Aqualinetubing set to be used and confrm Aqualinetubing set or regular treatment-Bloodowratesfrom30-450mL/min-Extracorporealvolume100mL
AqualineStubing set or low blood ow treatment
-Bloodowratesfrom10-200mL/min-Extracorporealvolume61mL
Module 2: Lesson 2: Preparing Aquarius
-
8/6/2019 Mod 2 Edwards Aquarius
21/69
Notes:
1
3. Place selected Aqualine on Aquarius
Follow the detailed instructions listed below to place the lines and flter on the machine
A. Place colored pump indicators in pumps. The colored pump holder segments are always atthe bottom o the pumps
NOTE:Alternatively,step-by-step,on-screeninstructionsor the placement o the linesand flters can be accessedin the Zoom graphic screenso Aquarius
Blood pump (red):
top let pump
Filtrate pump (yellow):
bottom let pump
Post-dilutionpump
(green): top right
pump
Pre-dilution/dialysate
pump (green): bottom
right pump
Module 2: Lesson 2: Preparing Aquarius
-
8/6/2019 Mod 2 Edwards Aquarius
22/69
Notes:
20
B. Place automatic degassing chamber(ADU) in holder under green pumps.Ensure that the shortest line with theclamp is towards the back and thechamber cover is closed securelyaround the tubing
Chamber cover closed (B)
Line attached to ADU sensor (C)
C. Attach the line with the luer lock,hydrophobic flter to the ADU sensor.
The clamp should remain open
Module 2: Lesson 2: Preparing Aquarius
-
8/6/2019 Mod 2 Edwards Aquarius
23/69
Notes:
2
E. Locate the return line chamber(also reerred to as the dripchamber) on the Aqualine,then insert the line underneath itthrough air detector channel andline clamp. Push and lock the airdetector
D. Insert heater coil into heater onthe right side o Aquarius (abovethe green scale) and close door
Return linechamber
Air detector
Return line clamp
Module 2: Lesson 2: Preparing Aquarius
-
8/6/2019 Mod 2 Edwards Aquarius
24/69
Notes:
22
G. Insert blood leak chamber intothe blood leak detector on the let
side o Aquarius (above the yellowscale)
F. Attach the pressure domes to theour pressure sensors (Preflter,Filtrate, Return, Access) and closeclamps
Filtrate
Return
Access
Preflter
Module 2: Lesson 2: Preparing Aquarius
-
8/6/2019 Mod 2 Edwards Aquarius
25/69
Notes:
2Module 2: Lesson 2: Preparing Aquarius
I. Attach fltrate line (short line romthe blood leak detector) to theclear luer lock fltrate port on thetop side o the hemoflter
H. Place prescribed hemoflter inflter holder and attach the redhemoflter connector to the redtop o the hemoflter and the bluehemoflter connector to the blue
bottom o the hemoflter
-
8/6/2019 Mod 2 Edwards Aquarius
26/69
Notes:
24 Module 2: Lesson 2: Preparing Aquarius
CVVH:
J. Attach ree line (located on the Aqualineasitexitsthepre-dilution/dialysatepump)
SCUF,CVVHDorCVVHDF:
attach ree line to the port locatedabove the red end o the hemoflter
attach ree line to the dialysate port onthe bottom side o the hemoflter
NOTE: The line exiting thepre-dilution pump (bottom greenpump) is called the ree line. The reeline connection point determineswhat the therapy will be
Free line
Free line
connection
Free line
connection
-
8/6/2019 Mod 2 Edwards Aquarius
27/69
Notes:
2Module 2: Lesson 2: Preparing Aquarius
L. Connect the red end o theAqualine to the prime collectionbag
NOTE: Aquarius primes with areverse ow through the bloodcircuit
K. HangemptyprimecollectionbagontheIVpolelocatedontherighttopsideoftheAquarius
M.Hangaone-literbagofprimingsolutionontheIVpole(usuallyheparinizednormalsaline)
InsertthespikedY-connectorprovidedwiththeAqualine into the priming solution bag
N. Remove the blue spiked end rom the
bloodline and connect it to one luerlockportoftheY-connector.(IfyouarenotusingaY-connector, connect theblue spiked end o the bloodline to thepriming solution.)
-
8/6/2019 Mod 2 Edwards Aquarius
28/69
Notes:
26 Module 2: Lesson 2: Preparing Aquarius
O. Hang empty fltrate bag(s) on
yellowfltrate scale and connectthe fltrate line (long line out oblood leak detector) to the bag(s)
P. Hang prepared solution(substitution/dialysate) on greensubstitution scale and connect the
substitution/dialysate line
NOTE: Aquarius uses a singlesolution source that is split bythe machine and delivered asboth dialysate and substitution
solution, depending upon thetherapy chosen. Up to our bagso solution may be used at onetime. I using more than one bag, amultiport maniold will be required.
To ensure proper, simultaneousemptying o multiple bags, all
clamps must be open.
NOTE: Ensure that an equalnumber o bags is used on thesubstitution/dialysate scale andthe fltrate scale. Use empty5-liter bags on the ltrate scale.For example, if TWO 2.5-literbags are used on the substitution/dialysate scale, use TWO 5-liter
empty bags on the fltrate scale
-
8/6/2019 Mod 2 Edwards Aquarius
29/69
Notes:
2
4. Prepare anticoagulant syringe
B. Select the volume in the syringe andconirm. Wait until the syringe drivermoves to the correct position
A. Aquarius may use a 50 mL or 60 mLsyringe flled to a maximum o 50 mL.Only use the heparin syringe typethat the Aquarius System has beencalibrated to use (or example, a BD50 cc syringe). The Aquarius Systemmust be calibrated or the particulartype o syringe you are using by acertifed technician. A list o approvedsyringes is available in section 6.5o Aquarius Operators Manual. The
syringe size that should be used isshown at the right o the Preparesyringe screen
Module 2: Lesson 2: Preparing Aquarius
WARNING: Clamp anticoagulant linei no anticoagulant is used!
-
8/6/2019 Mod 2 Edwards Aquarius
30/69
Notes:
28
D Install syringe into pump
E. Prime anticoagulant line by selecting and confrming Prime anticoagulant line. Each press o
the selector button delivers one mL
F. Program heparin inusion rate
Syringe attached
Syringe in pump
C. Attach syringe to the anticoagulantline on the tubing set
Module 2: Lesson 2: Preparing Aquarius
-
8/6/2019 Mod 2 Edwards Aquarius
31/69
Notes:
2
PRIMING AQUARIUS
1. Select Next to go to Priming
A. Check all lines or proper connection and open all clamps
WARNING: When No anticoagulant is selected, the clamp on the anticoagulant line MUSTremain closed!
B. Confrm that the correct solutions are used
WARNING: The solutions used or CRRT with Aquarius are prescription uids. To ensure
patient saety, please check that the prescribed solutions have been selected and properlyprepared
2. Select Start priming
A. The priming procedure requiresabout 800 mL o saline and takesapprox. 9 min (Aquarius displaysthe message Priming. Pleasewait and the clock counts downtozero)
B. Thepre-andpost-dilutionlinesare primed with uid rom thesubstitution/dialysate bag(s).
The blood circuit and the fltratelines are primed with uid romthe priming solution bag (usuallyheparinizednormalsaline)
Module 2: Lesson 2: Preparing Aquarius
-
8/6/2019 Mod 2 Edwards Aquarius
32/69
Notes:
30
C. Lines being primed are highlightedand Access, Return and TMP
pressures are displayed
D. Priming starts at 80 mL/min orthe frst 6 minutes (approximately500 mL) then increases to 150mL/min
.
E. When priming is completed, youwill hear an alarm and see themessage Priming completed
F. I needed, options are available toreprime parts, or all o the circuit
G. I the priming procedure issatisactory, select Next andconfrm
Module 2: Lesson 2: Preparing Aquarius
-
8/6/2019 Mod 2 Edwards Aquarius
33/69
Notes:
3
CLAMP & PRESSURE TEST
The Clamp & pressure test checks the clamp, the air detector, and the access, return and
pre-lterpressures
It MUST be done beore you can proceed to connection mode. Connect access and return
linestoasinglebagofnormalsalineusingtheY-connectorprovidedaspartoftheAqualine,
or any other appropriate connector
The test takes just a ew moments. I the test ails, ollow the instructions displayed on the
screen to determine the point o ailure. Check connections and select Retest
When Clamp & pressure test is completed, Aquarius switches to Start connection mode
Select Go to recirculationAquarius switches to Recirculation mode
Press the Blood Pump key to start the ow o saline within the tubing set. You are now
recirculating saline inside the circuit
Module 2: Lesson 2: Preparing Aquarius
-
8/6/2019 Mod 2 Edwards Aquarius
34/69
Notes:
32
RECIRCULATION MODE
RecirculationmodecanbeenteredimmediatelyafterClamp&pressuretestorduringa
treatment when theres the need to disconnect a patient temporarily. Temporary disconnection
will be discussed in the Disconnection and End Treatment sections
Recirculation o normal saline through the circuit prior to patient connection may acilitate the
removal o air trapped in the circuit1
Recirculation may be continued through the blood circuit until connection is needed
Programming o prescribed parameters may be done during recirculation (recommended)
Select Go to programming at any time during recirculation to enter treatment parameters
Module 2: Lesson 2: Preparing Aquarius
1 Dr. Patrick Honore and Dr. Didier Journois
-
8/6/2019 Mod 2 Edwards Aquarius
35/69
Notes:
3
PROGRAMMING AQUARIUS
Allparametersareprogrammed in thisscreen
Eachparametershows a yellowscreen that providesdetailed inormationabout the selected
parameter only
Programmingcanbedone ater priming,during recirculation,or immediately beoreconnection to thepatient. Parametersmay also be adjustedat any time duringtreatment
Select Previous toreturn to the mainscreen
SelectGotoconnection whenready to connect thepatient
USING THE SELECTOR BUTTON:
Turn: to highlight the parameter
Push once: to select the parameter
Turn right: to
Turn left: to
Push once: to confrm new values
Module 2: Lesson 2: Preparing Aquarius
-
8/6/2019 Mod 2 Edwards Aquarius
36/69
Notes:
34
PROGRAMMING PARAMETERS
The parameters available include:
Description
This parameter is optional and may be left at 0:00 h:min, unless the fluid removal rate is setto 0 mL/hour. When set, the Aquarius will run for the indicated length of time and thendisplays the message 'Therapy target achieved by time'. The operator may now programmore time to continue the treatment or move to the disconnection phase
Enter the amount of fluid that is prescribed for the patient to lose per hour
Enter the amount of fluid to be removed. After this goal is achieved, the Aquarius will displaythe message 'Therapy target achieved by fluid loss'. If there is a fluid loss rate set, then thisvalue must be programmed
Available in CVVH and CVVHDF modes only, this is the amount of fluid that will be deliveredpost-filter
Available in CVVHD and CVVHDF modes only, this is the amount of fluid that will be deliveredas dialysate
Available in CVVH mode only, this is the amount of fluid that will be delivered pre-filter
The number of bags hanging from the substitution/dialysate scale. An equal number ofempty bags must also be hung from the filtrate scale
This is the infusion rate from the integrated anticoagulation syringe pump in mL/hour
If a bolus of heparin is prescribed, enter it here for immediate administration when in treat-ment mode
When set at 0, the fluid warmer will be off. It is adjustable from 35 to 39 C (95 to102.2 F). The warmer heats both substitution and dialysate fluid
Parameter
Module 2: Lesson 2: Preparing Aquarius
-
8/6/2019 Mod 2 Edwards Aquarius
37/69
Notes:
3
CONNECTION MODE
SelectSingleConnectionorDoubleConnectionandfollowtheinstructionsonthescreen
PressBloodPumpkeytostarttreatment
Only the access line is attached to patients access port, initially, and theblood pump is started to discard the saline rom the circuit, up to the airdetector. A blood sensor, located inside the air detector, will automaticallystop the pump
Both access and return lines are simultaneously attached to the patientsaccess and return ports
Module 2: Lesson 2: Preparing Aquarius
-
8/6/2019 Mod 2 Edwards Aquarius
38/69
(page intentionally blank)
-
8/6/2019 Mod 2 Edwards Aquarius
39/69
Lesson 3: Treatment Mode
-
8/6/2019 Mod 2 Edwards Aquarius
40/69
(page intentionally blank)
-
8/6/2019 Mod 2 Edwards Aquarius
41/69
Notes:
Module 2: Lesson 3: Treatment Mode 3
TREATMENT MODE
The Treatment screen displays pertinent inormation and allows the operator to monitor severalparameters throughout the therapy:
Blood ow
Cumulative totals o uid loss and substitution/dialysate since last total reset (heparin is notreset)
Pressures (Access, Return, TMP and Pressure Drop)
-SeedetailedinformationonPressuresinModule2Lesson4
-
8/6/2019 Mod 2 Edwards Aquarius
42/69
Notes:
Module 2: Lesson 3: Treatment Mode40
Allows the operator to returnto the programming screenany time during treatment tochange parameters
Allows the operator toview additional treatmentinormation, with no ability tomake changes
Provides the ollowing options:-Viewhistory-Gotorecirculation
-Endtreatment-Changesyringe-Changetherapy
Screen selection choices:
-
8/6/2019 Mod 2 Edwards Aquarius
43/69
Notes:
Module 2: Lesson 3: Treatment Mode 4
The history o the last three treatments is available. Data is visible as a list or a graph. Pressures,
programmed parameters, patient data and events are stored.
-
8/6/2019 Mod 2 Edwards Aquarius
44/69
Notes:
Module 2: Lesson 3: Treatment Mode42
DISCONNECTING PATIENT FROM AQUARIUS
The patient may be disconnected rom the Aquarius by selecting and confrming Recirculation orEnd treatment rom the Options screen.
Disconnecting the patient rom the Recirculation screen allows temporary disconnection ora limited period o time. The flter, lines, bags and solutions all remain on the machine andthe blood pump runs in recirculation mode until the patient is ready to be reconnected to thesame circuit
Disconnecting the patient rom the End treatment screen will mean the patient treatment willend. The flter lines, bags and all solutions must be removed rom the machine and Aquarius mustbe turned o. A new set up will be needed i the patient therapy is to be continued
REMINDER: When disconnecting patients rom Aquarius, there may be circumstancesin which blood should NOT be returned to the patient. These circumstances mayinclude, but are not limited to: cardiac arrest, uid overload, major blood leak, a circuitthat was primed with blood or albumin, hemolysis or a clotted circuit. Always checkwith the prescribing physician.2
NOTE: The ollowing supplies may be needed:
1. One bag o normal saline or rinsing blood back to patient
2. Two saline syringes and the prescribed anticoagulant to lock the vascular
catheter
3. Caps or the vascular catheter
CAUTION: Always ollow hospital policy or standard precautions. Gloves, mask and aace shield should be worn when connecting or disconnecting blood lines rom patientsand removing lines rom Aquarius
2 Dr. Noel Gibney
-
8/6/2019 Mod 2 Edwards Aquarius
45/69
Notes:
Module 2: Lesson 3: Treatment Mode 4
RECIRCULATION MODE FOR TEMPORARY DISCONNECTION
Go to the Options screen and select
Thisoptionstartsthetemporarydisconnectionofthepatient.Thescreenmenuguidestheoperator through the disconnection procedure
On-screendirectionsguidetheoperatorthroughtheaccessandreturndisconnectionprocess
Volumeofreinfusionuidisdisplayed
AlltreatmentparametersarestoredinHistory
-
8/6/2019 Mod 2 Edwards Aquarius
46/69
Notes:
Module 2: Lesson 3: Treatment Mode44
END TREATMENT MODE
Beore ending treatment,please record all pertinentdata rom the Treatmentscreen
REMINDER: Once youconrm the End treat-ment selection, you willnot be able to go back
To end the treatment anddisconnect the patient romthe Aquarius, go to theOptions screen and select
Follow instructions on thescreens to disconnect thepatient rom the Aquarius
The volume o fuid used toreturn blood to the patientwill be recorded on thedisconnection screens as thereinusion volume in milliliters
-
8/6/2019 Mod 2 Edwards Aquarius
47/69
Notes:
Module 2: Lesson 3: Treatment Mode 4
Ifsystemhasclottedandanycircuitpressureisgreaterthan400mmHg,followon-screen
instructions or relieving pressure
Remove Aqualine and bags rom Aquarius
WARNING: Never switch o the Aquarius beore the tubing set is completely removed
is highlighted. Push the selector button to turn the Aquarius o
WARNING: Removing Aqualine under high pressure conditions may lead to a blood leak omembrane rupture in pressure domes
-
8/6/2019 Mod 2 Edwards Aquarius
48/69
(page intentionally blank)
-
8/6/2019 Mod 2 Edwards Aquarius
49/69
Lesson 4: Pressures, Alarms and Fluid Balance
-
8/6/2019 Mod 2 Edwards Aquarius
50/69
-
8/6/2019 Mod 2 Edwards Aquarius
51/69
Module 2: Lesson 4: Pressures, Alarms & Fluid Balance 4
AQUARIUS PRESSURE MONITORING
All extracorporeal circuits are run by pressures. Aquarius uses our pressure readings to monitorand control the system. Three pressures (Access, Return and Pre-lter) monitor the blood sideand one pressure (Filtrate) monitors the fuid side. All our pressures are used in combination tohelp you manage the treatment saely and eectively.
ACCESS PRESSURE
Typically monitors the negative pressure
generated to remove blood rom the patient
Access pressure may be positive under
certain conditions (example: ECMO)The patients catheter and Aquarius
blood fow rates aect this pressure
Measuringrange:-250to+100mmHg
Upperalarmlimit:automaticsetting
between-50and+100mmHg
Lower alarm limit: automatic setting
between-250and-100mmHg
Alarm window size during treatment:
+/-100
mmHg
around
the
actual
value
Displayed
on
the
Treatment
screen
RETURN PRESSURE
Monitorsthepositive pressure generated
to return the blood to the patient
The patients catheter and Aquarius
blood fow rates aect this pressure
Clottingincircuitaffectsthispressure
Measuringrange:0to+350mmHg
Upperalarmlimit:automaticsettingbetween110and350mmHg
Loweralarmlimit:automaticsettingbetween10and250mmHg
Alarmwindowsizeduringtreatment:+/-100mmHg
DisplayedontheTreatmentscreen
-
8/6/2019 Mod 2 Edwards Aquarius
52/69
Notes:
Module 2: Lesson 4: Pressures, Alarms & Fluid Balance50
PRE-FILTER PRESSURE
Monitorsthepositive pressure
generated in the circuit immediately
beore the blood enters the
hemolter
Bloodowrateandreturnpressure
aect this pressure
Highestpressurepointinthecircuit
Arisingpre-lterpressuremaybe
indicative o lter clotting
Upperlimit:+400mmHg
Lowerlimit:-100mmHg
DisplayedontheMorescreen
FILTRATE PRESSURE
Monitorsthenegativeand/orpositive
pressure generated in the ltrate
compartment o the hemolter. Thisaects the movement o fuid across
the membrane
Arisingltratepressuremaybe
indicative o lter clotting
Bloodowratesanduidowrates
aect this pressure
Upperlimit:+400mmHg
Lowerlimit:-400mmHg
DisplayedontheMorescreen
-
8/6/2019 Mod 2 Edwards Aquarius
53/69
Notes:
Module 2: Lesson 4: Pressures, Alarms & Fluid Balance 5
CalculatedusingReturn,Pre-lterandFiltratepressures
Calculatedas:ReturnPressure+Pre-lterPressure _
2
TMPisacombinationofthetotal
pressure exerted on the membrane
(positive pressure inside the bers and
negative pressure outside the lters)
Upperlimit:+400mmHg(CRRT)
Upperlimit:+100mmHg(TPE)
Lowerlimit:-30mmHgDisplayedontheTreatmentscreen
PD (PRESSURE DROP)
Calculatedusingpressurelossfromtop
to bottom o hemolter
Lengthofhemolteraffectspressure
Calculatedas:
Pre-lter Pressure Return Pressure
DisplayedontheTreatmentscreen
TMP (TRANSMEMBRANE PRESSURE)
Filtrate Pressure
-
8/6/2019 Mod 2 Edwards Aquarius
54/69
Notes:
Module 2: Lesson 4: Pressures, Alarms & Fluid Balance52
AQUARIUS ALARMS
The Aquarius saety mode is determined by the nature o occurring alarms:
Alarm type Aquarius safety mode
Alarms in blood circuit Stops all pumps
Closes return line clamp
Generates visual and audible alarms
Alarms in filtrate / dialysate circuit Stops filtrate pump
Stops pre-dilution pump and post-dilution pump
Generates visual and audible alarmsSystem error Stops all pumps
Closes return line clamp
Generates visual and audible alarms
Alarmsignalsareaudible and visual
Thestatuslightschangefromgreentored,yellow,oracombinationdependingonthetypeof
alarm
Cause o the alarm is displayed in a message box in the lower let corner o the screen AlarmsmaybesilencedfortwominutesbypressingtheMutekey
BloodcircuitalarmsareclearedbypressingtheBloodPumpkey.Oncethebloodpumphas
beenrestarted,presstheBalanceStart/Stopkeytorestarttheuidpumps
FluidcircuitalarmsareclearedbypressingtheBalanceStart/Stopkey
-
8/6/2019 Mod 2 Edwards Aquarius
55/69
Notes:
Module 2: Lesson 4: Pressures, Alarms & Fluid Balance 5
BLOOD CIRCUIT ALARMS
Generatesaudibleandvisualalarm
Stopsallpumps
LEDinBloodPumpandBalanceStart/Stopkeyashes
Closesreturnlineclamp(ifairdetectedorreturnpressuredropsbelowlowerlimitof
+10mmHg)
FLUID CIRCUIT ALARMS
Generatesaudibleandvisualalarm
Stopsalluidpumps LEDinBalanceStart/Stopkeyashes
BloodPumpcontinuestorun
SYSTEM ALARMS
Generatesaudibleandvisualalarms
Stopsallpumps
LEDinBloodPumpandBalanceStart/Stopkeyashes
Closesreturnlineclamp(ifairdetectedorreturnpressuredropsbelowlowerlimitof
+10mmHg)
NOTE: ABalancealarmoccurswhen a50mL (20mLwith useof AqualineSdierence is detected between the fuid volume pumped and the scalemeasurement o the substitution or ltrate solutions. This volume may result in anadditional50mL(20mL)infusiontothepatientor50mL(20mL)ofuidremovedrom the patient i a tubing line is kinked or blocked. Correct the cause o the alarmandpresstheBalanceStart/Stopkeytoresumetreatment
WARNING: I repeated balance alarms occur, stop treatment to protect the patient rom
excessive fuid removal or overload. Call technical support to resolve the problemDonotusethemachineuntilithasbeenrepaired
-
8/6/2019 Mod 2 Edwards Aquarius
56/69
TROUBLESHOOTING SYSTEM ERRORS, ALARMS AND MESSAGES
System Errors: Red indication
I the ollowing system errors are not corrected ater system test and i during treatment nocorrection is possible, please call technical service to repair the Aquarius.
During the system test (after switching the system on) andalsoduringoperation theAquarussystem automatically perorms tests to check the saety-critical components. I errors occurduring these tests, the system switches to the saety mode and generates a system errormessage.ThiserrormessageappearswiththeabbreviationCPU1(controlprocessorunit1)orCPU2.
Display: Cause: Options for error removal:
Accesssensor Access pressure values deviatefrom limits.
Control system and protectivesystem values deviate fromeach other (outside of limits)
During clamp/pressure test nopressure increase is detected
Restart system (no tubing must be installed). If theerror cannot be reset, put system out of operationand notify technical support.
Access pressure dome may not be connectedcorrectly (ensure by pressing the dome firmlytowards the access transducer)
ADC/Voltage Voltage supply or AD-converterfailure
Restart system (no tubing must be installed). If theerror cannot be reset, put system out of operationand notify technical support.
Air detector Air detector test failed. Press blood pump key. If error cannot be reset, putsystem out of operation and notify technical support.
Backup
No dates at the backup Restart system (no tubing may be installed). If theerror cannot be reset, put system out of operationand notify technical support.
Balance filtration Check filtration scale
Values between protective andcontrol system deviate fromeach other (outside of limits).
Actual values are outside oflimits.
Restart system (no tubing may be installed). If theerror cannot be reset, put system out of operationand notify technical support.
Balance substitution Check substitution scale
Values between protective andcontrol system deviate fromeach other (outside of limits).
Actual values are outside oflimits.
Restart system (no tubing may be installed). If theerror cannot be reset, put system out of operationand notify technical support.
BLD (Blood leak detector) Blood leak detector does notwork properly.
Press the Blood pump key. If the system errorcannot be reset, put system out of operation andnotify technical support.
*Rev. A - 12/2007
Notes:
Module 2: Lesson 4: Pressures, Alarms & Fluid Balance54
-
8/6/2019 Mod 2 Edwards Aquarius
57/69
Notes:
Module 2: Lesson 4: Pressures, Alarms & Fluid Balance 5
Display: Cause: Options for error removal:
Clamp doesnt closeClamp doesnt open
Clamp test failed.
Clamp does not close.
Clamp does not open.
Correct tubing set position in clamp.
Press blood pump key. If the error cannot be reset,put system out of operation and notify technicalsupport.
CODE Controller CPU program code
test failed Master CPU program code test
failed
Restart system (no tubing may be installed). If the
error cannot be reset, put system out of operationand notify technical support
Commu front system Data transfer to display systemdefective.
Set up safety mode for the patient.
Switch system off and on again after approx.1 minute (no tubing must be installed). If the errorcannot be reset, put system out of operation andnotify technical support.
Commu protection system Error during data transfer.
Power supply for protectivesystem is defective.
Press blood pump key or Balance Start/Stop key.
Switch system off and on again after approx.1 minute. If the error cannot be reset, put systemout of operation and notify technical support.
EEPROM Master CPU calibration datatest failed
Controller CPU calibration datatest failed
Restart system (no tubing may be installed). If theerror cannot be reset, put system out of operation
and notify technical support
Error1 CPU Master CPU Register test failed
Controller CPU Register testfailed
Restart system (no tubing may be installed). If theerror cannot be reset, put system out of operationand notify technical support
Error2 CPU Master CPU RAM test failed
Controller CPU RAM test failed
Restart system (no tubing may be installed). If theerror cannot be reset, put system out of operationand notify technical support
Error3 CPU Master CPU jump test failed
Controller CPU jump test failed
Restart system (no tubing may be installed). If theerror cannot be reset, put system out of operationand notify technical support
Blood pump Flow rate test failed.
Blood pump drive defective.
Blood pump is not stopped.
Actual value of number ofrevolutions deviates from setvalue outside of limits.
Switch system off and on again after approx. 1minute (no tubing must be installed).
Press blood pump key.
If the error cannot be removed, notify technicalsupport.
Error blood detection Blood at optic sensor(airdetector)
Restart system (no tubing may be installed). If theerror cannot be reset, put system out of operationand notify technical support.
-
8/6/2019 Mod 2 Edwards Aquarius
58/69
Notes:
Module 2: Lesson 4: Pressures, Alarms & Fluid Balance56
Display: Cause: Options for error removal:
Heater Master hasnt pass the systemtest
Restart system (no tubing may be installed). If theerror cannot be reset, put system out of operationand notify technical support.
Operation mode The data transfer valuesregarding the operation modedeviate between master- andcontroller CPU.
Press blood pump key to start a new check.
If the alarm appears again, call technical support.
Postdilution pump Flow rate test failed. Postdilution pump drive
defective.
Postdilution pump is notstopped.
Actual value of number ofrevolutions deviates from setvalue outside of limits.
Switch system off and on again after approx.1 minute (no tubing may be installed).
Press Balance Start/Stop key.
If the error cannot be reset, put system out ofoperation and notify technical support.
Program run Program failure Master CPU
Program failure Controller CPU
Restart system (no tubing may be installed). If theerror cannot be reset, put system out of operationand notify technical support.
Filtration pump Flow rate test failed.
Filtrate pump drive defective.
Filtrate pump is not stopped.
Actual value of number ofrevolutions deviates from setvalue outside of limits.
Switch system off and on again after approx. 1minute (no tubing must be installed).
Press Balance Start/Stop key.
If error cannot be reset, notify technical support.
Heparin pump Actual values of control andprotective systems deviate fromeach other (outside of limits)
Actual values deviate fromlimits
After selftest: Restart system (no tubing must beinstalled). If the error cannot be reset, put systemout of operation and notify technical support.
During treatment:
1. Check the Heparin line is not clamped.
2. Go to "Options" and "Change Syringe", followingthe on-screen text. Please Note: You DO NOT haveto remove the syringe during this process
3. If problem persists, program the pump to 0,clamp the line and remove the syringe. If problempersists, end treatment and call technical service.
-
8/6/2019 Mod 2 Edwards Aquarius
59/69
Notes:
Module 2: Lesson 4: Pressures, Alarms & Fluid Balance 5
Display: Cause: Options for error removal:
Predilution pump Flow rate test failed.
Predilution pump drivedefective.
Predilution pump is notstopped.
Actual value of number ofrevolutions deviates from setvalue outside limits.
Switch system off and on again after approx.1 minute (no tubing must be installed).
Press Balance Start/Stop key.
If the error cannot be reset, put system out ofoperation and notify technical support.
Return sensor Return pressure values deviatefrom limits.
Values between protectivesystem and control system
deviate from each other(outside of limits).
During clamp/pressure test nopressure increase is detected
Restart system (no tubing may be installed). If theerror cannot be reset, put system out of operationand notify technical support.
Clamp pressure test:o Press domes firmly towards the transducer.
Return and pre filter domes must be connectedcorrectly
o Insert tube into return line clampo Ensure that there is no leakage between blood
pump and return line clamp
Sensor Voltage Voltage supply or AD-converterfailure
Restart system (no tubing may be installed). If theerror cannot be reset, put system out of operationand notify technical support.
Software version Different Software versiondetected at master andcontroller
Restart system (no tubing may be installed). If theerror cannot be reset, put system out of operationand notify technical support.
Timer Timer difference betweenmaster and controller > 20%
Restart system (no tubing may be installed). If theerror cannot be reset, put system out of operation
and notify technical support.TMP sensor Filtrate and/or prefilter pressure
values deviate from limits.
Values between protectivesystem and control systemdeviate from each other(outside of limits)
Restart system (no tubing may be installed). If theerror cannot be reset, put system out of operationand notify technical support.
Vcc Master/communication RAM, EPROM or EEPROM aredefective.
Switch system off and on again after approx.1 minute. If the system error cannot be reset, putsystem out of operation and notify technical support.
XRAM Master CPU extern RAM failed
Controller CPU external RAMtest failed
Restart system (no tubing may be installed). If theerror cannot be reset, put system out of operationand notify technical support.
-
8/6/2019 Mod 2 Edwards Aquarius
60/69
Notes:
Module 2: Lesson 4: Pressures, Alarms & Fluid Balance58
Alarms: Red IndicationI the Aquarius system detects an out-o-range condition during the system test or duringoperation or i parameters exceed or drop below the respective limits, an alarm message isgenerated and the Aquarius system switches to the saety mode.
Display:
Cause:
Options for error removal:
High accesspressure
Low accesspressure
Access pressure has exceededthe upper alarm limit.
Access pressure has droppedbelow the lower alarm limit.
Check access blood line, including access and pre-filter sensors, for kinks or occlusions.
Check position of patient access.
Check blood flow rate.
Air Detected Return tubing line contains air ormicro-foam.
Make sure that tubing line does not contain air.
Check access and filter connections for sources of airleaks.
When you clear the Air Detected alarm, make surethere is no air or foam trapped in the line between thereturn detector chamber and the patient end.
To remove air from the tubing line:Step1. Open the return line clamp.Step 2. Remove all air from the tubing set.Step 3. Close the return line clamp.Step 4. Resume treatment by starting the bloodpump.
Warning! If the air detected alarm does notclear and air is visible in the venous de-aerationchamber, disconnect the patient from the instrumentand recirculate per your centers procedure.
Note: You may see micro-bubbles smaller than the airdetector sensitivity.
Balance system off The balance system has been offfor 5 minutes.
Switch on
-
8/6/2019 Mod 2 Edwards Aquarius
61/69
Display:
Cause:
Options for error removal:
15 seconds.
Clamp heparin line The Heparin syringe has beenremoved
Clamp Heparin line
Degassing chamber missing The substitution degassingchamber is not inserted or thesensors are defective.
Insert the substitution chamber properly. Ensure thechamber is in contact with the sensor of the holder. Ifthis doesnt help switch off heater to end the
treatment. Then call technical service.High filtrate pressure
Low filtrate pressure
Filtrate pressure exceeds or fallsbelow the alarm limits.
Check pressure sensor.
Check tubing set for narrow sections.
Check filter and exchange if required.
Check blood flow-to-fluid removal or blood flow-toturnover ratio.
Filtrate flow failure The number of revolutions of thefiltrate pump exceeded or fell
below the alarm limits by 5%.
Restart treatment
If problem persist call a repair technician
Heparin syringe missing A Heparin rate has beenprogrammed and no syringe hasbeen inserted in the plunger.
The Heparin syringe is not
inserted properly
Insert Heparin syringe if Heparin is needed
Set Heparin rate to zero ( No Heparin required)
Balance Alarm
Check substitution line
Or
Balance alarm
Check filtration line
Balancing deviates from the setvalues entered by the operator.
Check flow rates of the pumps.
Check fluid removal and turnover input parameters.Note: Potentially pumps can't deliver program withvery high volume because of pressure peaks.
Check bag hanging on scale.(Filtration or substitution)
Check tubing set for narrow sections.(Filtration or substitution lines) .
Blood leak Filtrate / plasma contains blood. Check filter for blood leak and exchange filter ifrequired.
Check pressure conditions.
Blood flow failure The number of revolutions of thepump exceeds or falls below the
alarm limits by 5%.
Restart blood pump.
If problem persist call technician
Blood pump off The blood pump was manuallyswitched off for 1 minute.
Press blood pump key to switch blood pump onagain.
Check transducer connections The pressure domes have notdetected any pressure change for
Ensure the domes are properly connected.
Notes:
Module 2: Lesson 4: Pressures, Alarms & Fluid Balance 5
-
8/6/2019 Mod 2 Edwards Aquarius
62/69
Notes:
Module 2: Lesson 4: Pressures, Alarms & Fluid Balance60
Display: Cause:
Options for error removal:
Indication battery high At the power fail indication batteryhigh voltage is detected
Change battery
Control charging voltage
Control AD-converter/CPU
Keyboard failure A key press longer than60 seconds. detected by master-cpu.
The keyboard needs repair
Line/substitution failure A substitution deviation is detectedthat influences the patient balance
Check lines
Check clamps
Check bags
Check for leakages
Master key transfer A key press longer than 60
seconds. detected by master-cpu.
The keyboard needs repair
The communication cable is defective
Main battery high At the main battery high voltage isdetected
Control charging voltage/unit
Control/change battery
Control AD-converter/CPU
Post-dilution failure The number of revolutions of thePost-dilution pump exceeded or
fell below alarm limits by 5%.
Restart treatment
If problem persist call technician.
Pre-dilution failure The number of revolutions of thePre-dilution pump exceeded or fell
below alarm limits by 5%.
Restart treatment
If problem persist call technician.
Pump door One of the pump doors open Check if one of the pump doors is open
Close the door
High pre-filter pressure
Low Pre-filter pressure
The pre-filter pressure exceedsupper alarm limit. Check pressure sensor. Check filter and exchange if required.
Check blood flow.
Check tubing set for narrow sections.
High return pressure Return pressure exceeds or falls Check return line for connection, kinks or occlusions.
Low return pressure below the alarm limits. Check return pressure transducer.
Check blood flow rate.
Check position of patient access.
Syringe removed The Heparin syringe has beenremoved
Check the heparin syringe
-
8/6/2019 Mod 2 Edwards Aquarius
63/69
Notes:
Module 2: Lesson 4: Pressures, Alarms & Fluid Balance 6
Display:
Cause:
Options for error removal:
High Temperature
Temperature controller high (Fluidpumps will not run)
A high temperature is detected inthe substitution degassingchamber
The substitution solutiontemperature is above40 C
The plate temperature of theheater is above 57C
Check the substitution line for kinks
Open heater door to cool the heater down.
If the alarm disappears the pump will startautomatically
Low temperature The substitution fluid temperaturehas been below 33C for morethan 10 minutes.
Check the heater bag
Check substitution temperature
Temperature controller High temperature of the heater atthe controller
Open heater and cool down the plate temperature
Restart treatment
Turn over failure The substitution volumecalculated by pump revolutionsdeviate more than 350g from thesubstitution volume measured onthe scale
Check lines
Check clamps
Check bags
Check for leakages
High TMP
Low TMP
TMP exceeds or drops belowrespective alarm limits.
Check pressure sensor.
Check tubing set for narrow line sections.
Check blood flow-to-fluid removal or blood flow-to-turnover ratio.
-
8/6/2019 Mod 2 Edwards Aquarius
64/69
Notes:
Module 2: Lesson 4: Pressures, Alarms & Fluid Balance62
Messages: Yellow Indication
I the Aquarius detects out-o-range conditions or reminders that do not conorm to the intendeduse o the system, the operator gets detailed inormation dened or the individual conditions andthe system switches to the saety mode.
Display:
Cause:
Options for error removal:
Ensure the return tubing does not contain air.
Ensure the return tubing is installed properly in theair detector sensors clamping mechanism and isnot scratched or damaged.
After correcting the error, Select and confirmPrevious, then perform the clamp/pressure testagain to proceed.
Air detected, Clamp/Pressuretest disabled
The air detector detected air in thereturn line.
Warning!If the air detected alarm cannot becleared, discontinue treatment and donot return the extracorporeal blood tothe patient.
Anticoagulant off The Heparin rate is programmed tozero.
If Heparin is not required proceed to next screen.
If Heparin is required, insert a syringe containingHeparin and program the desired Heparin rate.
Balance initializing Scales and fluid pumps initializewhen the balance system is started.
This is just a reminder
Occurs each time the balance system is turned ON
Balance system off The balance system is off, all fluid
pumps are stopped.
Correct the cause and switch the balance system
on again.Blood detected
(Blood pump stops)
During the connection orrecirculation phase blood isdetected in the return line
The blood pump stops and starts now only for 5seconds again. Switch to treatment mode
Blood pump off The blood pump was manuallyswitched off.
Press blood pump key to switch blood pump onagain.
Change substitution bag
Change filtrate bag
The filtrate bag has reachedmaximum permissible weight or thesubstitution solution bags do notcontain solution.
Replace empty substitution solution bags with fornew bags filled with solution
Replace full filtrate bags with empty bags
Check access transducer The access/ return dome is notproperly connected.
During clamp and pressure test no pressureincrease is found when the clamp is closed
-
8/6/2019 Mod 2 Edwards Aquarius
65/69
Notes:
Module 2: Lesson 4: Pressures, Alarms & Fluid Balance 6
Display:
Cause:
Options for error removal:
Exchange filter and set The assumed filter life time of 72hours has been exceeded or thetotal using time for the actual tubingset is more than 80 hours.(Including priming, connection,recirculation and treatment time)
Exchange filter and tubing set
Disconnect and start a new treatment with new filterand a new tube system
Function not available During treatment the Off key isused
To switch off the machine, select End treatment,perform the disconnection program until theAquarius off mode
Filt./effluent bag change soon Filtrate/effluent bag change in lessthan 10 minutes
Prepare the Filtrate/effluent bag change
Heater cools down If balance system has stopped for aprolonged period, the heater platetemperature is above 43C.
The pumps stop until the temperature is in a safecondition (below 42 C). This may take up to 10min. The pumps will start automatically.
Heater self test running.
! Wait
Heater Selftest is in progress when
the start-priming screen is reach.
Wait until heater Selftest is completed.
When heater Selftest is finished, the yellow lightupper the screen become fixed (green diode stillfixed during and after heater Selftest)
Heparin syringe missing The Heparin syringe is not in theHeparin pump.
nsert a Heparin syringe in Heparin pump or program
the Heparin rate to zero.
Check degassing chamber The degassing unit system hasdetected a pressure issue.
- After Selftest: Degassing pressure sensor ordegassing module defect. Don't use the Aquariusmachine and call technician.
- During use: Pressure sensor detects less than -300 mmHg. Degassing Filter is wet. Clamp thepressure line, disconnect filter from sensor and usea syringe to dry it. Reconnect the line and unclampit.
Check lines The postdilution pump has beenstopped for longer than 3 min toregulate the fluid loss.
Check the substitution line and the filtration line ifthe bags are opened, all clamps are removed andthe tubes and bag inputs are not kinked.
End of Treatment End of Treatment Disconnect patient or reprogram patient parameter
-
8/6/2019 Mod 2 Edwards Aquarius
66/69
Notes:
Module 2: Lesson 4: Pressures, Alarms & Fluid Balance64
Display: Cause:
Options for error removal:
No fluid chamber detected The substitution degassingchamber is not inserted.
Insert the chamber correctly
Power failure The Aquarius machine hasdetected a power failure
Reconnect the Aquarius to main power supply
Return the blood to the patient and start a newtreatment if power from main power supply isavailable
Please program Hourly fluid loss or fluid loss total isnot programmed
Select programming mode and program both hourlyfluid loss and fluid loss total.
High filtration ratio 1. Fluid removal rate exceeds 33%of the blood flow rate.
2. Exchange of fluid or plasmaacross the membrane is too high incomparison to the blood flow rate.
3. The post dilution substitution rateis higher than what is acceptable forthe current blood flow rate.
Decrease fluid removal or plasma exchange rate
Increase blood flow rate
Evaluate ratio of pre vs. post dilution substitution
solutions.
Indication battery low Power failure indication (beep)battery is low
The battery is automatically charged if you go onwith this treatment.
This message indicates that in the event of a power
failure the Aquarius will beep less than 2 min
Insert BLD chamber The blood leak detector has notbeen inserted into the blood leakchamber.
Insert the chamber
Reprime to correctly fill up the chamber.
Ensure no scratches or marks are present on theAqualine chamber
Insert tube
Insert tube to air detector
The air detection system is notoperational after priming
Insert correctly return line into the air detectionsystem.
Make sure that the air detection system is wellinserted, if not push it back firmly.
Ensure the green light of the clamp key is lighting.
Main battery low After a power failure, the mainpower supply battery must becharged.
The battery is automatically charged if you go onwith this treatment.
This message indicates that in the event of a powerfailure the Aquarius will run less than 2 min
Negative UF A negative UF is programmed This is just a reminder
No bag No bag was hung on the scale Hang empty collection bag and/or substitutionsolution bag on the scale
-
8/6/2019 Mod 2 Edwards Aquarius
67/69
Notes:
Module 2: Lesson 4: Pressures, Alarms & Fluid Balance 6
Display: Cause:
Options for error removal:
Please wait The balance system is stopped This is an indication that the system will startautomatically after some minutes.
Program dialysate In CVVHD dialysate rate is notprogrammed.
Select programming and program a dialysate rate.
Program goal Treatment goal is not programmed. Select programming and program either time, fluidloss and fluid loss total, or both.
Program treatment pumps In CVVH/CVVHDF pre- and post-dilution/post-dilution and dialysateare not programmed.
Select programming and program pre- and post-dilution or post-dilution and dialysate rate.
Pump door open One of the pump doors is open. Close pump door(s).
Read error help instructions To solve the alarm furtherinformation is needed Further information is available from the helpscreens.
Return pressure low The return pressure is below 10mmHg
During the first minute of treatment this is areminder.
Syringe empty. Go to OPTION tochange
The syringe located in the Heparinpump is empty.
Clamp Heparin line.
Take syringe out of pump and disconnect from line.
Fill new syringe with Heparin.
Enter syringe volume and confirm.
Place syringe in pump and connect line.
Confirm
Subst./dialysate bag changesoon
Substitution/dialysate bag changein less than 10 minutes
Prepare Subst./dialysate bag change
Too much weight Too many bags are on the scale. Ensure that the same number of substitution
solution and filtrate bags hang on the scale hooks. Note: The maximum number of bags on the scale is
4 substitution solution and 4 filtrate bags.
-
8/6/2019 Mod 2 Edwards Aquarius
68/69Module 2: Lesson 4: Pressures,Alarms & Fluid Balance66
Notes:
Module 2: Lesson 4: Pressures, Alarms & Fluid Balance66
Notes:
AQUARIUS FLUID BALANCE
ProgrammedpatientuidremovalwillberecordedhourlyontheTreatmentscreenasacumulative value
Non-Aquariusinsandoutsforthepatientmustbecalculatedandcontrolledbytheuser
UFvariationismonitoredanddisplayedintheMorescreen.Theaccuracyis50mLforregularcircuits,20mLforlowvolumecircuits
ResettotalsintheProgrammingscreenresetsallcumulativetotalsforuidloss,anddialysate and substitution fuid. Treatment clocks are also reset. These values are located ontheTreatmentscreen.
Values are programmed in theProgrammingscreen
Dialysateand/orsubstitutionuidare automatically removed by the
Aquarius
WhenprogrammingHourlypatientuidloss,aTotaluid
lossMUSTalsobeprogrammed.Programming the time is optional
Treatmentproceedsuntilaprogrammed target is achieved
WhenPatientuidlossissetatzero, time must be programmed
NOTE: Aquarius records all fuids as a cumulative number rom the start o treatment
NOTE: When the target
is reached, the message
Therapytargetachieved
by fuid loss (or time) is
displayed. The target can
be reprogrammed and
the treatment continued
NOTE: Anticoagulant volumes are NOT reset
-
8/6/2019 Mod 2 Edwards Aquarius
69/69
Caution:Federal(U.S.)lawrestrictsthisdevicetosalebyoronthe
order o a physician or other licensed practitioner. For sae and
proper use o this device reer to the Operators Manual.