[email protected] Assistant Professor Mike … · Spiral CT: Single and Multiple Detector...

42
Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999 Spiral CT: Single and Multiple Detector Systems AAPM Refresher Course Nashville, TN July 28,1999 Mike McNitt-Gray, PhD, DABR Assistant Professor UCLA Radiological Sciences [email protected]

Transcript of [email protected] Assistant Professor Mike … · Spiral CT: Single and Multiple Detector...

Page 1: mmcnittgray@mednet.ucla.edu Assistant Professor Mike … · Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999 X-Ray CT Systems

Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999

Spiral CT:Single and Multiple Detector Systems

AAPM Refresher CourseNashville, TNJuly 28,1999

Mike McNitt-Gray, PhD, DABRAssistant Professor

UCLA Radiological [email protected]

Page 2: mmcnittgray@mednet.ucla.edu Assistant Professor Mike … · Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999 X-Ray CT Systems

Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999

X-Ray CT Systems All Use:

• Thinly Collimated X-Ray Beam• Multiple Views• Detectors to Collect X-Ray Photons• Data Acquisition System• Image Reconstruction Algorithm (Filtered Back Projection)• Film/Monitors To Display Axial Slices

Page 3: mmcnittgray@mednet.ucla.edu Assistant Professor Mike … · Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999 X-Ray CT Systems

Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999

3rd generationrotate-rotate

4th generationrotate-stationary

Page 4: mmcnittgray@mednet.ucla.edu Assistant Professor Mike … · Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999 X-Ray CT Systems

Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999

Conventional CT Systems(GE Hi-Lite Advantage) :

• Tube Rotates Around Stationary Patient (Table is Incremented Between Acquisitions)• All Views in a Slice are at Same Table Position• Power to X-Ray Tube via Cord• Scan CW and CCW to Wind/Unwind Cord• Interscan Delays:

3.5 Seconds Between Slices

Page 5: mmcnittgray@mednet.ucla.edu Assistant Professor Mike … · Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999 X-Ray CT Systems

Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999

Kalender’s 1990 article was titled : “Spiral volumetric CTwith single-breath-hold technique, continuoustransport, and continuous scanner rotation.”

Differences from Conventional:• Continuous Tube Rotation - No Interscan Delays (Power to X-ray Tube via Slip Ring)• Continuous Table Motion as Tube Rotates• Each View is at a DIFFERENT Table Position Form Images by Synthesizing Projection Data via

Interpolation

Spiral (Helical) CT Systems

Page 6: mmcnittgray@mednet.ucla.edu Assistant Professor Mike … · Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999 X-Ray CT Systems

Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999

Data Acquisition and Image Formation

• Some Data Acquisition Parameters Behave asAlways (kVp, mAs, Recon. Filter)

• Others (Collimation, Table Speed/Pitch, andInterpolation Algorithm) Interact to InfluenceImage Formation and Image Quality

• However, Image Formation has its ownindependent parameter- reconstruction interval

Page 7: mmcnittgray@mednet.ucla.edu Assistant Professor Mike … · Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999 X-Ray CT Systems

Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999

Data Acquisition

Pitch = Table Movement per Rotation Slice Collimation(Essentially a measure of Relative Speed)• Contiguous Spiral

Pitch = 1 (10 mm / 10 mm)• Extended (Non-Contiguous) Spiral

Pitch = 2 (20 mm/ 10 mm)• Overlapping Spiral

Pitch = 1/2 ( 5 mm / 10 mm)

Page 8: mmcnittgray@mednet.ucla.edu Assistant Professor Mike … · Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999 X-Ray CT Systems

Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,19990

z, mmt, sec

Direction of Continuous Patient Transport

Center of Spiral PathWidth of Spiral Path (From Collimation)

Pitch=1; Contiguous Spiral

Page 9: mmcnittgray@mednet.ucla.edu Assistant Professor Mike … · Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999 X-Ray CT Systems

Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,19990

z, mmt, sec

Direction of Continuous Patient Transport

Center of Spiral PathWidth of Spiral Path (From Collimation)

Pitch=2; Extended Spiral

Page 10: mmcnittgray@mednet.ucla.edu Assistant Professor Mike … · Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999 X-Ray CT Systems

Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999

Image Formation

◆ Collect Projection Data◆ These Data are NOT all at same table position, so:◆ Synthesize a set of Planar Projection Data via

Interpolation (interpolate between views taken atsame projection angle, but different table positions),then

◆ Use Filtered Back Projection on Synthesized PlanarData

◆ Because a VOLUME of data has been acquired,IMAGE LOCATION IS ARBITRARY

Page 11: mmcnittgray@mednet.ucla.edu Assistant Professor Mike … · Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999 X-Ray CT Systems

Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999

Selected Image Plane (Arbitrary)

Path of Continuously Rotating X-ray Tube (and Projection Data)

0

z, mmt, sec

Direction of Continuous Patient Transport

Page 12: mmcnittgray@mednet.ucla.edu Assistant Professor Mike … · Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999 X-Ray CT Systems

Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999

Selected Image Plane (Arbitrary)

Path of Continuously Rotating X-ray Tube (and Projection Data)

0

z, mmt, sec

Direction of Continuous Patient Transport

Page 13: mmcnittgray@mednet.ucla.edu Assistant Professor Mike … · Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999 X-Ray CT Systems

Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999

Arbitrarily selected planar section at position z

zz1 z1 + d

Planar data for arbitrary positions are calculated from spiral data by interpolation

d

360 degree algorithm

Page 14: mmcnittgray@mednet.ucla.edu Assistant Professor Mike … · Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999 X-Ray CT Systems

Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999

Arbitrarily selected planar section at position z

Planar data for arbitrary positions are calculated from spiral data by interpolation

zz1 z1+d/2

180 Degree Algorithm

measured spiralcalculated spiral

d

Page 15: mmcnittgray@mednet.ucla.edu Assistant Professor Mike … · Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999 X-Ray CT Systems

Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999

Arbitrarily selected planar section at position z

Planar data for arbitrary positions are calculated from spiral data by interpolation

zz1 z1+d/2

180 Degree Algorithm

measured spiralcalculated spiral

d

Page 16: mmcnittgray@mednet.ucla.edu Assistant Professor Mike … · Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999 X-Ray CT Systems

Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999

Image Formation

◆ Independent of IMAGE ACQUISITION PARAMETERS(Because a VOLUME of Data is Acquired andInterpolation is used)Images can be formed ANYWHERE -->

Slice Location is Arbitrary

◆ Overlapping Images can be createdFROM A SINGLE VOLUMETRIC SCAN– (e.g. 10 mm thick, 5 mm apart).

◆ Effective Slice Thickness Is Determined by AcquisitionParameters (Collimation, Pitch) and Interpolation

Page 17: mmcnittgray@mednet.ucla.edu Assistant Professor Mike … · Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999 X-Ray CT Systems

Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,19990

z, mmt, sec

Center of ImageWidth of Image (From Collimation)

Contiguous Reconstruction - No Overlap

Image Number 1 2 3

Page 18: mmcnittgray@mednet.ucla.edu Assistant Professor Mike … · Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999 X-Ray CT Systems

Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,19990

z, mmt, sec

Overlapping Reconstruction - 50% Overlap

Image Number 1 2 3

Page 19: mmcnittgray@mednet.ucla.edu Assistant Professor Mike … · Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999 X-Ray CT Systems

Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999

Contiguous reconstruction - object aligned with slices

Image Number 1 2 3 4

Page 20: mmcnittgray@mednet.ucla.edu Assistant Professor Mike … · Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999 X-Ray CT Systems

Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999

Contiguous reconstruction - object aligned in between slices

Image Number 1 2 3 4

Page 21: mmcnittgray@mednet.ucla.edu Assistant Professor Mike … · Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999 X-Ray CT Systems

Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999

Overlapping reconstruction - object aligned in between slices

2 4 6 8 Image Number 1 3 5 7

Note: Object is Volume Averaged in Slices 3 and 5, but contained completely within slice 4

Page 22: mmcnittgray@mednet.ucla.edu Assistant Professor Mike … · Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999 X-Ray CT Systems

Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999

Spiral CT - Differences From Axial

Image Quality (Assuming 180 interp. algorithm):

• Effective Slice Thickness Increases with Pitch(~10% increase for pitch 1; ~ 30% for pitch 2;

~ 70% for pitch 3)• Hence volume averaging increases with pitch• Noise is > Conventional Axial under identicalacquisition conditions, but does not change w/ pitch

Page 23: mmcnittgray@mednet.ucla.edu Assistant Professor Mike … · Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999 X-Ray CT Systems

Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999

Spiral CT - Differences From Axial

Radiation Dose:

For Pitch of 1 ~ Same as Comparable ContiguousConventional Scans

For Pitch 1.5 - Approximately 2/3 that of ContiguousScans

For Pitch 2 - Approximately 1/2 of ContiguousScans

• Radiation Dose Proportional to 1/pitch

Page 24: mmcnittgray@mednet.ucla.edu Assistant Professor Mike … · Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999 X-Ray CT Systems

Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999

• Faster Tube Rotation Times• Greater Heat Capacity Tubes• More Detector Planes• Faster Image Reconstruction

What’s Next ?What’s Next ?

Faster, Thinner Spirals with LOTS of ImagesFaster, Thinner Spirals with LOTS of Images

Page 25: mmcnittgray@mednet.ucla.edu Assistant Professor Mike … · Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999 X-Ray CT Systems

Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999

Faster Tube Rotation Times

Most Major CT Mfrs now offer subsecond scannersbetween .5 - .8 s for a full rotation

However, to get same photon statistics with shorterexposure time means tube current must increase

need higher kW generators and higher HeatCapacity Tubes to sustain tube currents.

Page 26: mmcnittgray@mednet.ucla.edu Assistant Professor Mike … · Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999 X-Ray CT Systems

Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999

Tubes with Greater Heat Capacity

Current Tubes are Typically 2-3 MHU(Axial Scanners used < 1 MHU)

Top of Line NOW are > 6 MHU

This allows longer spirals( less tube cooling delays)

Also allows increase in tube current to offsetreduced scan times

Page 27: mmcnittgray@mednet.ucla.edu Assistant Professor Mike … · Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999 X-Ray CT Systems

Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999

More Detector Planes

Elscint CT-Twin in 1994 had two adjacent detectors

This allowed simultaneous acquisition of 2 images -2 X coverage for same scan time

Now GE, Siemens, Picker and Toshiba have, or haveannounced, Multidetector systems

Page 28: mmcnittgray@mednet.ucla.edu Assistant Professor Mike … · Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999 X-Ray CT Systems

Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999

Collect Multiple Slices Simultaneously to:

Cover More Patient in the Same TimeCover Same Amount of Patient, but w/ Thinner SlicesScan a Volume at Arterial Phase, Venous Phase,

Parenchymal Phase after Contrast Injection

Page 29: mmcnittgray@mednet.ucla.edu Assistant Professor Mike … · Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999 X-Ray CT Systems

Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999

GE- 4 Row Detector

16 rows of detectors, use 4 at a timeAllows collection of 4x1.25 mm slices (or 4x2.5, or 4x5)

SIMULTANEOUSLYWill go to 16 rows simultaneous when data transfer is

quick enough

Page 30: mmcnittgray@mednet.ucla.edu Assistant Professor Mike … · Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999 X-Ray CT Systems

Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999

LightSpeed Matrix Detector

HiLightHiLight Detector Detector

852 channels in 49852 channels in 49oo degree arc degree arc

Each channel is 1 mm wideby 20 mm tall

LightSpeedLightSpeed Matrix Detector Matrix Detector

14,592 14,592 elementselements in 55 in 55oo degree arc degree arc

Each ELEMENT is 1 mm wideby 1.25 mm tall

16 Rows x 912 Channels

Page 31: mmcnittgray@mednet.ucla.edu Assistant Professor Mike … · Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999 X-Ray CT Systems

Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999

How Is It Used?

Off

Off

Active

4 x 1.25 mm Mode4 x 1.25 mm Mode5 mm per rotation5 mm per rotation

4 x 2.5 mm Mode4 x 2.5 mm Mode10 mm per rotation10 mm per rotation

Off

Off

Active

4 x 3.75 mm Mode4 x 3.75 mm Mode15 mm per rotation15 mm per rotation

4 x 5 mm Mode4 x 5 mm Mode20 mm per rotation20 mm per rotation

3 Rows = One DASChannel

Page 32: mmcnittgray@mednet.ucla.edu Assistant Professor Mike … · Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999 X-Ray CT Systems

Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999

Helical Imaging Modes

DetectorDetectorConfigurationsConfigurations

4 x 1.25 mm4 x 1.25 mm

4 x 2.5 mm4 x 2.5 mm

4 x 3.75 mm4 x 3.75 mm

4 x 5 mm4 x 5 mm

ImageImage Thicknesses Thicknesses

1.25, 2.5 mm1.25, 2.5 mm

2.5, 3.75, 5.0 mm2.5, 3.75, 5.0 mm

3.75*, 5.0, 7.5 mm3.75*, 5.0, 7.5 mm

5.0, 7.5, 10.0 mm5.0, 7.5, 10.0 mm

3.75 mm3.75 mm

7.5 mm7.5 mm

11.25 mm11.25 mm

15 mm15 mm

7.5 mm7.5 mm

15 mm15 mm

22.5 mm22.5 mm

30 mm30 mm

Table Travel Per RotationTable Travel Per Rotation

Note: * 3.75 mm thickness not available at 22.5 mm/rotNote: * 3.75 mm thickness not available at 22.5 mm/rot

HQHQ HSHS

Page 33: mmcnittgray@mednet.ucla.edu Assistant Professor Mike … · Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999 X-Ray CT Systems

Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999

Acquisition & Reconstruction

Multiple RowData Acquisition

Multiple RowRecon Algorithms 4 x 2.5 mm

Images

2 x 5 mmImages

1 x 10 mmImage

Matrix Detector4 x 2.5 mm Mode

More Complex Than Single-SliceMore Complex Than Single-Slice

Page 34: mmcnittgray@mednet.ucla.edu Assistant Professor Mike … · Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999 X-Ray CT Systems

Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999

180180oo00oo

360360oo

4:1 Pitch4:1 Pitch

180180oo00oo

360360oo

3:1 Pitch3:1 PitchSignificantSignificant

Data OverlapData Overlapw/o IQ gainw/o IQ gain

Limited OverlapLimited OverlapOptimum IQOptimum IQ

Choice Of Pitch Critical to IQChoice Of Pitch Critical to IQChoice Of Pitch Critical to IQ

Page 35: mmcnittgray@mednet.ucla.edu Assistant Professor Mike … · Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999 X-Ray CT Systems

Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999

Faster Reconstructions

Currently about 2-3 seconds per imageIn some commercial packages (options)

~ 1 secondFaster Computers and Dedicated Processors will reduce

this < 1 sec/image

Page 36: mmcnittgray@mednet.ucla.edu Assistant Professor Mike … · Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999 X-Ray CT Systems

Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999

What will the Current/Near Future SystemLook Like?

Fast Tube/Gantry Rotation (���� .5 second)Tube Heat Capacity > 6 MHUMultiDetector System (4 now with pathways to 16 and up

to 34 depending on manufacturer)Faster Reconstruction Times (subsecond)

Hundreds (Thousands?) of Images PossibleHow to review all these?

Page 37: mmcnittgray@mednet.ucla.edu Assistant Professor Mike … · Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999 X-Ray CT Systems

Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999

What Will You Be Able to Do?What Will You Be Able to Do?

Single Breath Thoracic Scan -Single Breath Thoracic Scan -Cover 35 cm w/ 1.25 mm slices <10 sCover 35 cm w/ 1.25 mm slices <10 s

Page 38: mmcnittgray@mednet.ucla.edu Assistant Professor Mike … · Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999 X-Ray CT Systems

Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999

What Will You Be Able to Do?What Will You Be Able to Do?

Two Phase Liver with Contrast -Two Phase Liver with Contrast -Cover 12 cm w/1.25 mm slices 2XCover 12 cm w/1.25 mm slices 2X (up and back) in <(8s + scan delay of ~5s (up and back) in <(8s + scan delay of ~5s))

Page 39: mmcnittgray@mednet.ucla.edu Assistant Professor Mike … · Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999 X-Ray CT Systems

Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999

What Will You Be Able to Do?What Will You Be Able to Do?

ER Trauma Case ER Trauma Case Cover 180 cm (head to toe) in a single Cover 180 cm (head to toe) in a single 48 s scan with 5mm (soon with 1.25mm)48 s scan with 5mm (soon with 1.25mm)

Page 40: mmcnittgray@mednet.ucla.edu Assistant Professor Mike … · Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999 X-Ray CT Systems

Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999

ReferencesBasic CT Physics Overview

Newton TH and Potts DG, eds, Radiology of the Skull and Brain:Technical Aspects of Computed Tomography, The C.V. Mosby Co.,St. Louis, 1981.

Principles of Spiral CTKalender WA; Seissler W; Klotz E; Vock P. Spiral volumetric CT with

single-breath-hold technique, continuous transport, and continuousscanner rotation. Radiology, 1990 Jul, 176(1):181-3.

Crawford CR, King KF. Computed Tomography Scanning withsimultaneous patient translation. Medical Physics 1990, 17:967-982.

Remy-Jardin M, Remy J, eds. Spiral CT of the Chest, Springer, Berlin,1996. Chapter 1- Principles of Spiral CT by M.W. Vannier and G.Wang.

Page 41: mmcnittgray@mednet.ucla.edu Assistant Professor Mike … · Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999 X-Ray CT Systems

Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999

References

Physics of Spiral CT

Kalender WA; Polacin A. Physical performancecharacteristics of spiral CT scanning. Medical Physics,1991 Sep-Oct, 18(5):910-5.

Polacin A; Kalender WA; Marchal G. Evaluation of sectionsensitivity profiles and image noise in spiral CT.Radiology, 1992 Oct, 185(1):29-35.

Polacin A; Kalender WA; Brink J; Vannier MA.Measurement of slice sensitivity profiles in spiral CT.Medical Physics, 1994 Jan, 21(1):133-40.

Page 42: mmcnittgray@mednet.ucla.edu Assistant Professor Mike … · Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999 X-Ray CT Systems

Spiral CT: Single and Multiple Detector Systems - Michael McNitt-Gray , AAPM Refresher Course July 27,1999

ReferencesPhysics of Spiral CT (cont’d)

Wang G; Vannier MW. Stair-step artifacts in three-dimensional helical CT: an experimental study.Radiology, 1994 Apr, 191(1):79-83.

Wang G; Vannier MW. Longitudinal resolution involumetric X-ray computerized tomography--analyticalcomparison between conventional and helicalcomputerized tomography. Medical Physics, 1994 Mar,21(3):429-33.

Kasales CJ; Hopper KD; Ariola DN; TenHave TR; MeilstrupJW; Mahraj RP; Van Hook D; Westacott S; Sefczek RJ;Barr JD. Reconstructed helical CT scans: improvementin z-axis resolution compared with overlapped andnonoverlapped conventional CT scans. AJR. AmericanJournal of Roentgenology, 1995 May, 164(5):1281-4.