Method Transfer Crossing Multiple Borders
-
Upload
institute-of-validation-technology -
Category
Health & Medicine
-
view
329 -
download
1
description
Transcript of Method Transfer Crossing Multiple Borders
8/2/2010
1
Method Transfer: Crossing Multiple Borders
Xiande (Andy) Wang, Ph.D.
Analytical Research and Development Cordis Corporation,
A Johnson & Johnson Company,Warren, NJ 07059
IVT Annual Method Validation 2010
Outline
1. Lifecycle of Analytical Methods
2. Border I: Between Methods
– Case study: combination of 3 methods into one
3. Border Between Instrumentation/Technique
– Case study: validation of an HPLC and UPLC method side by side
4. Border Between Groups
– Case study: Troubleshooting cross functions during method transfer
5. Conclusions
8/2/2010
2
Life Cycle of an Analytical method
Method Development
Method validation
Method transfer
Life Cycle of Analytical methods
Development
validationtransfer
8/2/2010
3
Life Cycle of Analytical Methods - Examples
Development
validationtransfer
Life Cycle of Analytical Methods - Examples
Development
validationtransfer
HPLC
UPLC HPLC
UPLC
HPLCUPLC
HPLC
HPLC
HPLC
UPLC
UPLC
UPLC UPLC
UPLC
UPLCHPLC
HPLC
HPLC
8/2/2010
4
Borders to Cross in Lifecycle of Method
Development
validationtransfer
HPLC
UPLC HPLC
UPLC
HPLCUPLC
HPLC
HPLC
HPLC
UPLC
UPLC
UPLC UPLC
UPLC
UPLCHPLC
HPLC
HPLC
Border II
Border I
Border III
Border I: Between Methods
UPLC
UPLC
UPLCHPLC
HPLC
HPLC
8/2/2010
5
Questions Around Border I
• Can the method be used interchangeably?
• Methods require the same instrumentation, column, reagents, materials?
• Solution (standard, sample, mobile phase) storage and stability
• Timing of validation, transfer; effective date, versions
• Will methods be run in the same or different lab?
• Sample shipping, storage at different environment
• Consistency in any other areas
How is HPLC Assay Method Developed?Method Objectives
– Stability indicating• Peak purity
• Resolution of all species
• LC-MS compatible
• Elution of all species/compounds
– Long gradient method
– Orthogonal
– Robust
– Sensitivity
8/2/2010
6
11
Systematic HPLC Assay Method Development
Samples are stressed under
different conditions
Stressed samples are analyzed with a generic method
Representative stressed samples are
chosen
Method screening is conducted with selected samples
A primary method is identified
The method is optimized
Stressed samples are analyzed with
optimized method
The primary method is ready for further optimization /
validation
12
General guidelines for HPLC Column Selection
• Select high-purity silica-based columns• C18 and C8– Hydrophobic, retentive and stable• Phenyl – medium polarity components; unique selectivity for
aromatics• Hydrophilic end-capped phases (retentive for water soluble
compounds)• Polar-embedded phases (amide, carbamate, ether, sulfonamide)
– Less tailing for basic analytes– “Orthogonal” to C8/C18, No phase collapse
• Explore selectivity differences between C18, polar-embedded or phenyl bonded phases – Consult column selectivity chart – For low pH Applications, select column resistant to hydrolytic cleavage
(e.g., StableBond, X-Bridge C18)
– For high-pH application, select columns stable at high pH (e.g., Gemini, X-Bridge, Extend, Luna)
M.W. Dong, Modern HPLC for Practicing Scientists, Wiley, 2006, Chap. 3.
8/2/2010
7
13
• Waters: Symmetry, SunFire, XTerra *, ACQUITY*, X-Bridge*, Atlantis, NovaPak, m-Bondapak, Spherisorb
• Agilent: Zorbax StableBond, Eclipse XDB, Extend C18*, Bonus
• Phenomenex: Luna*, Prodigy, Synergi*, Gemini*
• Supelco: Discovery, Ascentis, Supelcosil
• Varian: Inertsil, Polaris*
• Thermo: HyPURITY, Hypersil, Prism, Hypersil Gold *
• MacMod: ProntoSIL, ACE (Adv. Chrom. Tech.)
• YMC: YMCbasic, Pack Pro
• Eka Chemicals: Kromasil
• GL Sciences: Inertsil
• Macherey Nagel: Nucleosil
• Merck KGaA: Chromolith (Monolith)
• Bischoff: ProntoSIL*
• Grace: Vydac, Platinum (Alltech)
• Dionex: Acclaim, Acclaim PA, Acclaim PA2*
Columns based on high-purity silica are underlined. Hybrid particles are in bold. Phases stable in high pH are italicized and marked with *.
Some Popular HPLC Columns
M.W. Dong, Modern HPLC for Practicing Scientists, Wiley, 2006, Chap. 3.
HPLC Assay Method Development: Column Screening
E. F. Hewitt, P. Lukulay, and S. Galushko, J Chromatography A, 1107, 79.
8/2/2010
8
15
HPLC Column Screen Set: An Example
Orthogonal Screening – Columns Stationary Phase Column pH Rangea Manufacturer Part Number
C18 – Twin Technology Gemini C18, 5 μm, 110A, 4.6 x 150 mm 1-12 Phenomenex 00F-4435-E0 Phenyl with Hexyl (C6) linker,
endcapped Luna Phenyl-Hexyl, 3 μm, 4.6 x 150 mm 1.5-10 Phenomenex 00F-4256-E0
C18-20% C loading Discovery HS-C18, 3μm, 4.6 x 150 mm 2-8 Supelco 569252-U C18 – polar embedded, hybrid
particle with Shield Technology XTerra RP18, 3.5 μm, 4.6 x 150 mm 1-12 Waters 186000442
C18– silica Sunfire C18, 3.5 μm, 4.6 x 150 mm 2.8 Waters 186002554 Pentafluorophenyl Curosil PFP, 3 μm, 4.6 x 150 mm 2-7.5 Phenomenex 00F-4122-E0
aColumns were screened only against mobiles phases within their compatible pH range.
Slide courtesy of H. Rasmussen et al
16
HPLC Screening Conditions: An Example
Orthogonal Screening Method Description Time (min) %Water %Acetonitrile % Modifiera Flow Rate (ml/min)
0 85 10 5 1.0 40 10 85 5 1.0 45 10 85 5 1.0
45.10 85 10 5 1.0 60 85 10 5 1.0
Injection Volume 5 μL Detection 280 nm; DAD (190 – 400 nm)
Column Temperature Ambient Sample Temperature 5oC
aModifier stock solutions are prepared at a concentration 20 times higher than the desired mobile phase concentration since mobile phases are prepared at time of use with the HPLC quaternary pump.
Modifier Mobile Phase Concentration
Approximate pH
Trifluoroacetic Acid (TFA) 0.05% 2 Formic Acid 0.1% 2.8
Ammonium Acetate + Acetic Acid 8 mM + 0.1% 4 Ammonium Acetate 8 mM 7
Ammonium Acetate + Ammonium Hydroxide
8 mM + 0.05% 10.2
Ammonium Hydroxide 0.05% 10.8
Slide courtesy of H. Rasmussen et al
8/2/2010
9
Summary of HPLC Assay Method Development
• There is a systematic approach for assay method development.
• Column screening is an effective tool.
• It will enhance efficiency by narrowing down the list of columns for screening/selection.
• It is critical to seek input from other labs (QA, QC) during method development.
• It’s important to look at the big picture across methods.
What is Important in Developing HPLC CU Method?Method Objectives
– Short
– Specific
– Isocratic preferred
– Robust
– Compatible with assay method
– Can be used for dissolution testing
8/2/2010
10
19
Systematic HPLC CU Method Development
samples are analyzed with a
generic method
Specificity, retention, peak shape etc.
acceptable
Method screening is conducted
A primary method is identified
The method is optimized
The primary method is ready for further optimization / validation
Yes
No
HPLC CU Method Development: Column Screening
E. F. Hewitt, P. Lukulay, and S. Galushko, J Chromatography A, 1107, 79.
8/2/2010
11
What is Important in Developing HPLC Method for Dissolution Testing?
– Short
– Specific
– Isocratic preferred
– Robust
– Compatible with CU method
– Sensitive
How to Minimize Border I (Between Methods)
• Use same materials/chemicals: grade, vendor
• Adopt similar standard/sample solution: procedure of preparation, concentration, pH, storage, expiry
• Use same column: vendor, stationary phase, dimension, particle size; alternative column
• Use same Instrumentation
• Be consistent in write-up: same product description, formulation number, etc.
8/2/2010
12
Case Study: Combination of 3 methods into 1
Method number
Method title
N1 Identification, Assay and Content Uniformity Determination of … by RP-HPLC-UV
O1 Identification and Assay of … by RP-HPLC-UV
O2 Determination of Content Uniformity of … by RP-HPLC-UV
O3 Identification of … by UV Spectroscopy
Method O1: Identification and Assay of … by RP-HPLC-UV
Parameter Value
HPLC Column Agilent, Zorbax Eclipse XDB-C18, 150 mm x 4.6 mm x 3.5 µm
Flow Rate 1.0 ± 0.1 mL/min Injection Volume 25 µL
Column Temperature 40 ± 2 °C
Detection Wavelength
278 nm.Note: If using an Agilent DAD detector, set the bandwidth to 4
nm and the reference wavelength off.
Mobile Phases A: 80:20% (v/v) 0.02% formic acid:THF
B: 75:20:5:0.02% (v/v/v/v), acetonitrile: THF : water: formic acid
Gradient Program
(Linear Gradient Profile)
Time (minutes) % Mobile Phase A % Mobile Phase B
0.0 55 45 3 55 45 7 42 58 13 10 90 15 55 45
AU
0.000
0.005
0.010
0.015
0.020
Minutes1.00 2.00 3.00 4.00 5.00 6.00 7.00 8.00 9.00 10.00 11.00 12.00 13.00 14.00 15.00
8/2/2010
13
Method O2: Determination of Content Uniformity of and Dissolution … by RP-HPLC-UV
Parameter Value
Column
Phenomenex Luna C18 (2), 4.6 x 50 mm, 3 μm or Phenomenex Gemini C18, 4.6 x 50 mm, 3 μm
HPLC Column Column
Temperature 35oC ± 2oC
Autosampler Temperature
Ambient (20 to 25 °C if temperature control is
available)
Mobile Phase 55:45, 0.02% v/v Formic Acid:THF
Flow Rate 1.2 mL/minute
Detector
278 nmNote: If using Agilent PDA
detector, set bandwidth to 4 nm and reference wavelength off.
Injection Volume 25 μL Run Time 10 minutes
Method O3: Identification of … by UV Spectroscopy
8/2/2010
14
Method N1: Identification, Assay and Content Uniformity Determination of … by RP-HPLC-UV
Parameter Value
HPLC Column Agilent Zorbax Eclipse XDB-C18, 100 mm x 4.6 mm, 3.5 µm
Flow Rate 1.0 mL/min Injection Volume 20 µL
Column Temperature 45 oC ± 2°C
Detection Wavelength
Assay Identification278 nm
Note: If using an Agilent DAD detector, set the bandwidth to 4 nm and
the reference wavelength off.
200 to 400 nm
Mobile Phases A: (20:80)THF: Formate Buffer B: (75:20:5)ACN:THF:Formate Buffer
Gradient Program (Linear
Gradient)
Time (min) % Mobile Phase A
% Mobile Phase B
0.0 47 53 3.0 47 53 5.0 20 80 6.0 2 98 7.5 2 98 7.6 47 53 10 47 53
AU
-0.004
-0.002
0.000
0.002
0.004
0.006
0.008
0.010
0.012
0.014
0.016
0.018
0.020
0.022
0.024
0.026
0.028
Minutes1.00 2.00 3.00 4.00 5.00 6.00 7.00 8.00 9.00
2.18
1
2.57
7
3.05
93.
390
3.59
5
RAPA
MYC
IN -
4.05
5
4.97
9
BHT
- 7.1
96
AU
0.00
0.05
0.10
0.15
0.20
0.25
0.30
0.35
0.40
0.45
0.50
nm200.00 220.00 240.00 260.00 280.00 300.00 320.00 340.00 360.00 380.00 400.00
208.0
269.0
279.0
291.0
349.0
Benefits of a Combined Method
Benefit Details
Shortened project timelines
One, instead of three, set of test methods, validation protocols, validation reports
Enhancedthroughput
Run time cut from 18 to 10min; one instrument/set-up for both assay/CU; eliminate requirement of a UV spectrometer.
Reduce cost One , instead of two, set of HPLC column is neededLess reference standard, organic solvent Less sample shipment cost
Eliminated causes for bias between assay and CU
When two methods are used, possible bias could come from different instrument set up, different way of peak integration, different separation capacity, etc. One method eliminate all these possible causes.
8/2/2010
15
Benefits of Elimination of Border I
Assay/CU/dissolution
Assay/CU/dissolutionAssay/CU/dissolution
• Same columns
• Same reagents, vendor, expiry dates
• Same materials such as glassware, HPLC vials, etc.
• No discrepancy (bias) of results due to integration, separation capacity, etc.
• Same solution (standard, sample, mobile phase)storage and stability
• Transfer at the same time
• Same method, same version
• One HPLC system needed
• Minimize sample shipping, storage at different environment
Summary of Strategy to Minimize/Eliminate Border I
• Assay, CU, and dissolution (HPLC) methods have different objectives and hence the strategy for method development may be different.
• However, it is important to keep the other method in mind during method development.
• Column screening is an effective tool for assay, CU or dissolution methods.
• It is effective to narrow down the list of columns for screening/selection.
• It is critical to seek input from other labs (QA, QC) during method development.
• It’s important to leverage the knowledge, as much as possible, of one method and apply it to another.
• It offers a lot of benefit to use similar or the same procedures, instrumentation and materials for all methods, and transfer them at the same time.
8/2/2010
16
Border II: HPLC vs. UPLC
UPLC
UPLC
UPLCHPLC
HPLC
HPLC
UPLC vs. HPLC: Example of Assay Method
Ref: L. Pereira, Poster at Pittcon 2007, Chicago, Illinois, February 2007.
8/2/2010
17
UPLC vs. HPLC: Example of CU Method
http://www.waters.com/waters
Issues Around Border II (HPLC vs. UPLC)
• Availability of instruments at all sites
• Different UHPLC systems have different design, not a direct transfer
• Training/knowledge and technical challenges on the new instruments
• Limited availability and technical challenges of columns
• Project history
• Cost
8/2/2010
18
How to Minimize Border II (HPLV vs. UPLC)
• Use the same stationary phase with different dimension and particle size
• Same mobile phase but slight different gradient
• Same protocol template
• Use the same solutions and validate the methods side by side
Case study: Validation of HPLC and UPLC Assay Method Side by SideChromatograms of the Final Methods
HPLC
UPLC
AU
-0.004
-0.002
0.000
0.002
0.004
0.006
0.008
0.010
0.012
0.014
0.016
0.018
0.020
0.022
0.024
0.026
0.028
Minutes1.00 2.00 3.00 4.00 5.00 6.00 7.00 8.00 9.00
2.18
1
2.57
7
3.05
9
3.39
03.
595
RA
PAM
YC
IN -
4.05
5
4.97
9
BHT
- 7.
196
AU
-0.002
0.000
0.002
0.004
0.006
0.008
0.010
0.012
Minutes0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50 5.00 5.50 6.00
RAPA
MYC
IN -
2.35
8
BHT
- 4.2
02
8/2/2010
19
Case study: Validation of HPLC and UPLC Assay Method Side by SideFinal Method Parameters
HPLC UPLCParameter Value
HPLC Column Agilent Zorbax Eclipse XDB-C18, 100 mm x 4.6 mm, 3.5 µm
Flow Rate 1.0 mL/min Injection Volume 20 µL
Column Temperature 45 oC ± 2°C
Detection Wavelength
Assay Identification278 nm
Note: If using an Agilent DAD detector, set the bandwidth to 4 nm and
the reference wavelength off.
200 to 400 nm
Mobile Phases A: (20:80)THF: Formate Buffer B: (75:20:5)ACN:THF:Formate Buffer
Gradient Program (Linear
Gradient)
Time (min) % Mobile Phase A
% Mobile Phase B
0.0 47 53 3.0 47 53 5.0 20 80 6.0 2 98 7.5 2 98 7.6 47 53 10 47 53
Parameter Value UPLC
Column AgilentZorbax Eclipse XDB-C18, 100
mm x 3.0 mm, 1.8 µm Flow Rate 0.7 mL/min Injection Volume 5 µL
Autosampler Temperature
10 oC
Column Temperature 45 oC ± 2 °C
Detection Wavelength
Assay Sampling Rate
278 nm Note: If using an Agilent
DAD detector, set the bandwidth to 4 nm and
the reference wavelength off.
20 (points/sec)
Mobile Phases
A: (20:80)THF: Formate Buffer B: (75:20:5)ACN:THF:Formate
Buffer
Gradient Program (Linear
Gradient)
Time (min) % Mobile Phase A
% Mobile Phase B
0.0 47 53 2.0 47 53 3.5 20 80 4.2 2 98 5.2 2 98 5.3 47 53 6.5 47 53
Run Time 6.5 minutes (Retention time is ~2.5 min for sirolimus, ~4.3 min for BHT.
Seal Wash Acetonitrile
Weak Wash Acetonitrile/water: 50/50, 600 µL volume
Case study: Validation of HPLC and UPLC Assay Method Side by SideAccuracy (spiked recovery)
~Level Actual
Concentration (μg/mL)
Individual % Recovery
Mean % Recovery
% RSD
40% 67.4 99.9
99.9 0.1 99.7 99.9
100% 151.7 99.5
99.6 0.2 99.8 99.5
160% 242.7 100.6
100.5 0.2 100.6 100.3
~Level Actual
Concentration (μg/mL)
Individual % Recovery
Mean % Recovery
% RSD
40% 67.4 100.9
101.0 0.1 101.0 101.1
100% 151.7 99.4
99.6 0.2 99.6 99.8
160% 242.7 99.3
99.0 0.2 98.9 98.9
HPLC
UPLC
8/2/2010
20
Validation of HPLC and UPLC CU Method Side by Side:Accuracy (spiked recovery)
HPLC
UPLC
~Level Actual
Concentration (μg/mL)
Individual %
Recovery
Mean % Recovery % RSD
40% 12.1 100.6
100.4 0.4 100.7 100.0
100% 30.3 98.9
98.9 0.1 98.8 99.0
160% 48.5 99.6
99.8 0.2 100.0 99.6
~Level Actual
Concentration (μg/mL)
Individual %
Recovery
Mean % Recovery % RSD
40% 12.1 100.9
101.2 0.3 101.6 101.3
100% 30.3 100.4
100.0 0.4 100.0 99.5
160% 48.5 100.4
101.3 0.9 102.2 101.5
Validation of HPLC and UPLC Assay Method Side by Side:Intermediate Precision
HPLC
UPLC
Determination %LC Analyst 1 %LC Analyst 2
1 103.1 103.42 101.9 102.63 101.4 103.04 101.6 103.05 102.4 103.46 102.2 104.0
Mean (n=6) 102.1 103.2%RSD (n=12) 0.8
Determination %LC Analyst 1 %LC Analyst 2
1 103.8 101.32 101.8 102.13 100.9 102.34 100.6 102.25 101.8 102.76 102.2 102.9
Mean (n=6) 101.9 102.2%RSD (n=12) 0.5
8/2/2010
21
Validation of HPLC and UPLC CU Method Side by Side:Intermediate Precision
HPLC
UPLC
Determination %LC Analyst 1 %LC Analyst 21 101.5 102.72 102.0 104.23 101.0 102.34 103.4 102.35 105.1 102.36 101.3 101.77 103.1 102.68 103.1 102.79 102.1 102.7
10 102.9 102.0Mean (n=10) 102.6 102.6
%RSD (n=20) 0.9
Determination %LC Analyst 1 %LC Analyst 21 102.0 103.42 101.6 103.73 101.3 103.44 100.4 102.75 102.6 102.26 101.0 101.77 101.7 103.68 100.8 102.69 101.6 102.9
10 101.4 102.2Mean (n=10) 101.4 102.7
%RSD (n=20) 0.9
Validation of HPLC and UPLC Assay Method Side by Side:Linearity
HPLC
UPLC
R=0.9997
8/2/2010
22
Validation of HPLC and UPLC CU Method Side by Side:Linearity
HPLC
UPLC
R=0.9997
R=1.0000
Validation of HPLC and UPLC CU Method Side by Side:Method Equivalency
HPLC UPLC
Determi-nation
%LC with new
method
%LC old CU
method1 98.4 98.72 99.2 99.63 98.7 99.94 100.1 100.35 100.0 99.96 100.4 100.67 100.4 100.28 100.1 100.59 100.7 100.210 99.9 100.4
Mean (n=10) 99.8 100.0
%RSD (n=10) 0.8 0.6
Absolute Difference
(%)0.2
Determi-nation %LC with
new method
%LC with old
CU method
1 102.0 101.32 101.6 100.63 101.3 100.84 100.4 100.45 102.6 102.26 101.0 100.77 101.7 101.18 100.8 100.69 101.6 101.410 101.4 101.0
Mean (n=10)
101.4 101.0
%RSD 0.6 0.5Absolute
difference (%)
0.4
8/2/2010
23
Summary of Case Study to Minimize Border II (HPLC vs. UPLC)
• Use the same stationary phase with different dimension and particle size
• Same mobile phase but slight different gradient
• Same protocol template
• Use the same solutions and validate the method side by side
• The knowledge of both methods were exchanged during the process
• HPLC and UPLC method proved to be equivalent and can be used interchangeably
Border III. Between Functions
Development
validationtransfer
HPLC
UPLC HPLC
UPLC
HPLCUPLC
HPLC
HPLC
HPLC
UPLC
UPLC
UPLC UPLC
UPLC
UPLCHPLC
HPLC
HPLC
Border II
Border I
Border III
8/2/2010
24
III-a: Method Development vs. Validation
• Often reside in the same functional group
• Natural redundancy in terms of robustness, forced degradation, alternative columns
Method Development
Method validation
Method transfer
What Is Analytical Method Transfer
• Protocol driven study with pre-defined acceptance criteria
• Transfer of validated analytical procedures to a new laboratory
• Verification of a method’s suitability for its intended use
• Demonstration of a laboratory’s proficiency in running a particular method
• No official guidelines
8/2/2010
25
Options for Method Transfer
• Comparative testing:A set of samples are tested in both labs and resulting data are compared with predetermined acceptance criteria.
• Co-validation between two labs:The receiving laboratory is involved in method validation but have to identify which validation parameters are to be generated or challenged by the two labs.
• Complete or partial method validation:A repeat of method validation either completely or partially.
• Transfer waiver (omission of formal validation):Needs justification as to why method transfer was not needed. For example, lab is already testing the product.
Border III-b: Method Transfer and Validation
• Method transfers are closely related to validation
• Method transfer is more challenging because multiple laboratories and companies are involved– Different approaches to Validation and Transfer– Different expectations of what is an acceptable
validation– Different instruments and facilities
8/2/2010
26
Method Transfer and Validation
Method transfer• Can be part of the validation
• Protocol driven study with pre-defined acceptance criteria
• Transfer of validated analytical procedures to a new laboratory
• Verification of a method’s suitability for its intended use in a new laboratory
• Demonstration of a laboratory’s proficiency in running a particular method
• No official guidelines
Method validation• Protocol driven study with pre-defined
acceptance criteria
• Validation of analytical procedures in a laboratory
• Verification of a method’s suitability for its intended use
• Demonstration of a laboratory’s proficiency in running a particular method
• http://www.ich.org/LOB/media/MEDIA417.pdf; http://www.fda.gov/cder/guidance/2396dft.pdf
Objectives of Method Transfer
• Maintain the validated state of the method and meet all regulatory requirements
• Minimize surprises!
– Open and responsive communication
– Pre-determined expectations
– Clearly documented and communicated technical details
– Pre-transfer evaluation by experienced technical staff at receiving site
– Technical contact available for troubleshooting at transferring site
8/2/2010
27
Typical Method Transfer Steps
• Discussions Initiated
• Review of Method and Validation
• Laboratory Evaluation
• Protocol (Transfer or validation) Written
• Protocol Approved
• Experimental evidence from a transfer study generated
• Report (Transfer or validation) Written
• Report Approved
• Transfer Complete
Preparation for a Method Transfer
• Method – Details about method– Specific instrument
• Method development history report• Training/discussion on the method • Materials
– Reference Standard– Samples for Evaluation– Difficult to purchase supplies
• Specifications• Technical Contact• Details about product
8/2/2010
28
Prior to Formal Method Transfer
• Receiving laboratory should perform the method– Helps to determine where there are differences
and gaps in documentation• Lack of detailed test method instructions
– Assay Conditions– Calculations– System Suitability
– Differences with instrumentation or reagents
Prior to Formal Method Transfer
• Training of Personnel– Review of relevant SOPs
– Observation of test procedure
– Performing test procedure
• Helpful to include development, qualification and validation reports to recipient laboratory
8/2/2010
29
Method Validation and Transfer
• Method transfers are closely related to validation
• Method transfer is usually part of validation
Method Development
Method validation
Method transfer
Border III-c: Method Transfer and Method Development:Before or During Method Development
• Define goals of end method
• Dynamic platform of communication
• Exchange of knowledge
8/2/2010
30
Border III-c: Method Transfer and Development:
• Is the method inadequate by today’s scientific standard or regulatory requirement?
• Is sufficient data available to permit simplification of the method?
• Does monitoring of laboratory deviation suggest a need for method improvement ?
• Do newer method for similar products significantly outperform?
• Is the volume of testing justify further method optimization or automation?
After method Transfer
Border III-c: Method Transfer and Development:
– Concerns?
– Observations?
– Investigations/troubleshooting: Must involve multiple labs
During Method Transfer
8/2/2010
31
Case Study: Extraneous peaks associated with HPLC vials
AU
0.000
0.001
0.002
0.003
0.004
0.005
0.006
0.007
0.008
0.009
0.010
Minutes0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50 5.00 5.50 6.00 6.50 7.00
3.10
8
3.67
0
RA
PA
MY
CIN
- 4.
133
Polypropylene
Deactivated glass vials
Case Study: Extraneous peaks associated with Glass Pipettes
AU
-0.002
-0.001
0.000
0.001
0.002
0.003
0.004
0.005
0.006
0.007
0.008
0.009
0.010
0.011
Minutes0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50 5.00
Unk
now
n1 -
1.69
0
Unk
now
n2 -
2.52
6
Unk
now
n3 -
2.97
7
Unk
now
n4 -
3.49
6
Rap
amyc
in -
3.94
7
Above: chromatograms of two different lots of glass pipettes .Plastic transfer pipettes, or no pipettes, were recommended.
8/2/2010
32
Case Study: Extraneous peaks associated with Extraction Vials
Extraction Vial Volume
(mL)
Number per case
Price ( $/case)
Price ($/vial) Comments
Nalgene-Teflon 10 10 224.70 22.47 Re usable. VWR Cat# 210030 10 333.30 33.33 Re usable. VWR Cat# 2100
BD Falcon-Poly Propylene, snap cap
5 500 160.00 0.32 VWR Cat # 60819-706
14 500 195.30 0.39 VWR Cat # 60819-740
BD Falcon-Poly Propylene, screw cap
15 500 220.89 0.44 VWR Cat# 21008-918
Starplex-Polypropylene
screw cap
5 1500 186.18 0.12 only 2 sizes available VWR14216-262
10 1000 135.96 0.14 only 2 sizes available VWR14216-266
15 500 192.00 0.38 Sigma Aldrich Cat# Z7204NUNC Poly propylene snap cap
Supelco pre-cleaned clear glass
7 100 92.50 0.93 Pre-cleaned* Sigma Aldrich27341
15 100 102.50 1.03 Pre-cleaned* Sigma Aldrich27342
22 100 116.00 1.16 Pre-cleaned* Sigma Aldrich27343
Supelco Silanized Clear Glass 4 1000 267.00 0.27
Larger sizes are available urequest, caps not included. Sigma Aldrich Cat# 27114
Kimble-glass 8 144 152.49 1.06 Currently being used VWR
66009-984
16 144 193.44 1.34 Currently being used VWR66009-986
Treatment to Glass Extraction Vials
• Rinse with 0.02% formic acid in acetonitrile
• Rinse with acetonitrile
• Wash with a regular washing cycle for other glassware
• Soak with 0.02% acid (formic, acetic or nitric) in water and rinse with DI water
• Soak/sonicate in DI water
• Details are critical to ensure accurate comparison cross labs.
8/2/2010
33
Standard Solution in Unwashed Glass Vials
0.00
0.50
1.00
1.50
2.00
2.50
3.00
3.50
0 1 2 3 4 5 6 7 8
Tota
l Im
purit
y A
rea%
Day
Standard Solution in Pre-cleaned Glass Vials
0
0.5
1
1.5
2
2.5
3
3.5
4
4.5
5
0 1 2 3 4 5 6 7 8
Tota
l Im
purit
y A
rea
%
Day
2ml
3ml
4ml
5ml
6ml
7ml
9ml
10ml
8/2/2010
34
Standard Solution in Silanized Glass Vials
Standard Solutions in Glass Vials Soaked and Sonicated in DI Water
8/2/2010
35
Standard Solution in Glass Vials Rinsed with MeCN 3 Times
0
0.5
1
1.5
2
2.5
3
0 1 2 3 4 5 6 7 8
Tota
l Im
purit
y ar
ea %
Day
Direct pour
8ml- std soln-1
8ml- std soln-2
8ml- std soln-3
8ml- std soln-4
8ml- std soln-5
Standard Solution in Glass Vials Rinsed with 0.02% Formic Acid in MeCN 3 Times
0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1
0 1 2 3 4 5 6 7 8
Tota
l Im
purit
y A
rea%
Day
Direct pour
8ml- std soln-1
8ml- std soln-2
8ml- std soln-3
8ml- std soln-4
8ml- std soln-5
8/2/2010
36
Standard Solutions in Glass Vials Treated with 0.02% Acid (Nitric, Acetic or formic)
Standard Solution in Washed Glass Vials
0
0.02
0.04
0.06
0.08
0.1
0 1 2 3 4 5 6 7
To
tal i
mp
uri
ty %
Day
Standard in Washed Extraction Vial
2
3
4
5
6
7
9
10
8/2/2010
37
Standard Solution in BD Falcon Polypropylene Vials
0.00
0.02
0.04
0.06
0.08
0.10
0.12
0.14
0.16
0 1 2 3 4 5 6 7
Tota
l Im
puri
ty a
rea%
Day
small-1
small-2
small-3
small-4
small-5
small-6
small-7
small-8
small-9
Summary to Case Study: Extraneous peaks Associate with Glassware
• Pay attention to the glass grade, vendor and treatment.
• Glass vials vary within the same lot/box.
• Rinse with 0.02% acid (formic, acetic, nitric) acid in water is effective for this method.
• Wash the glass vials with acidic detergent is effective.
• Use of polypropylene vials eliminates the problem.
• It is critical for multiple labs to be involved, to carry our experiments and share data with details.
8/2/2010
38
Method transfer vs. method development
• Start from the beginning of the lifecycle (define goals, communicate limitations)
• Maintain a dynamic and continuous process
• Build strong partnership and co-ownership through regular meetings, visits, design and execute experiments together
• Transfer knowledge, not just method
Method Transfer vs. validation Vs. Development
Method Development
Method validation
Method transfer
• Dynamic platform of communication
• Design and execute experiments (AMERT, Investigations)
• Share ownership• Exchange of knowledge
8/2/2010
39
Conclusions
• Analytical method lifecycle is a dynamic and continuous process
• Transfer of knowledge, instead of method, is desired in every stage of the life cycle
• Scientists/managers need to zoom in and zoom out to consider needs of other projects, methods or labs.
• It enhances efficiency to eliminate borders between methods, techniques and functional groups, as much as possible
• Strong partnership and co-ownership is the key to successful methods