RATIONALE FOR THIS WORK RESULTS Our two research groups used metallic porous frits to mechanically support PDMS membranes, allowing the deployment of the UMSs to ocean depths of 2000 meters MEMBRANE INTERFACE EVALUATIONS FOR UNDERWATER MASS SPECTROMETERS A. M. Cardenas-Valencia 1 , T. Gentz 2 , M. Schlueter 2 , S. K. Toler 1 , R. T. Short 1 1 Marine and Space Sensing Program, SRI International, U.S.A, 2 Alfred Wegener Institute for Polar and Marine Research (AWI), Bremerhaven, Germany The multifold applications of the underwater mass spectrometer (UMS) include energy exploration, ocean acidification, and ecosystem studies (1,2). A mechanically supported membrane interface assembly allows for deepwater UMS applications. The UMS membrane interface assembly consists of a polydimethyl siloxane (PDMS) capillary mechanically supported by a fabricated frit. The fabrication of the frit requires using a Dremel tool with a diamond-coated wheel to shape larger Hastalloy C porous frits to the specified size. This frit fabrication procedure is time-consuming and cumbersome, and the porosity of the final frits is very difficult, if not impossible, to replicate. We report on new porous metallic structures that facilitate the fabrication of the membrane assembly. The limit of detection (methane concentration in the tens of nanomolars) and sensitivity (on the order of 10 -1 pico-amps/nanomole of methane) were comparable with those obtained with the previously fabricated Hastalloy C frits. The calibration parameters for the supported membrane interface assembly depend on flow rate. We could achieve further optimization of the new frits by changing the dimensions of the channel through which the sample flows. SELECTED REFERENCES 1. Bell, R.J., et al. (2011), Limnol. Oceanogr.-Meth. 9: pp. 164-175. 2. Cardenas-Valencia, A.M., et al. (2013), Rapid Commun. in Mass Sp. 27: pp. 635-642. SUMMARY Microcontroller/ Peripheral board Roughing pump Pump Membrane assembly Electronics Methane calibration curves with German frit 1200 (32.5% Porosity) (~ 1/8” diam) y = 4.25E-14x + 1.58E-14 R² = 0.9997 0 2E-13 4E-13 6E-13 8E-13 1E-12 1.2E-12 1.4E-12 1.6E-12 1.8E-12 0 10 20 30 40 Mass spec signal (m/z=15) Concentration µmol/kg y = 1.93E-13x - 6.67E-14 R² = 9.98E-01 y = 2.29E-13x - 1.31E-14 R² = 9.99E-01 y = 2.51E-13x - 5.55E-14 R² = 9.98E-01 y = 2.66E-13x - 3.89E-14 R² = 9.99E-01 0 2E-12 4E-12 6E-12 8E-12 1E-11 1.2E-11 0 10 20 30 40 50 8 10 15 20 Mass spec signal (m/z=15) Concentration µmol/kg Detection limit for 10 mL/min (using 1 σ)= 20 nM Flow, mL/min Methane calibration curves with German frit 1200 (48.3% Porosity) (~ 1/8” diam) CONTACT INFORMATION [email protected] (727) 498-6721 Detection limit (using 1 σ)= 80 nM y = 2.73E-13x + 1.54E-13 R = 0.9977 0 2E-12 4E-12 6E-12 8E-12 1E-11 1.2E-11 0 10 20 30 40 Methane calibration curves with home-made frit (~1/8” diam) Mass spec signal (m/z=15) Concentration µmol/kg Detection limit (using 1 σ)= 30 nM Heater cartridges High Pressure Membrane Assembly enabled the development of underwater mass spectrometry PDMS membrane PEEK cap Vacuum Sample inlet Heater block Thermocouple Epoxy Sample outlet Frit The Fraunhofer Institute in Dresden, Germany, used powder metallurgical processes to manufacture frits with diameters of approximately 3.0 mm (1/8”) and known porosities of 48.3 % and 32.5% These frits were used to fabricate new membrane interface assemblies. Using a custom-heated membrane probe with the new porous frits, we were able to calibrate dissolved methane concentrations to mass spectrometer responses (m/z 15) using linear least-squares fitting procedures. METHOD

Transcript of MEMBRANE INTERFACE EVALUATIONS FOR …epic.awi.de/34009/1/Poster_HEMS_2013.pdf · Our two research...

Page 1: MEMBRANE INTERFACE EVALUATIONS FOR …epic.awi.de/34009/1/Poster_HEMS_2013.pdf · Our two research groups used metallic porous frits to mechanically support PDMS membranes, ... 1.4E-12.

RATIONALE FOR THIS WORK

RESULTS

Our two research groups used metallic porous frits to mechanically support PDMS membranes, allowing the deployment of the UMSs to ocean depths of 2000 meters

MEMBRANE INTERFACE EVALUATIONS FOR UNDERWATER MASS SPECTROMETERS A. M. Cardenas-Valencia1, T. Gentz2, M. Schlueter2, S. K. Toler1, R. T. Short1

1 Marine and Space Sensing Program, SRI International, U.S.A,2 Alfred Wegener Institute for Polar and Marine Research (AWI), Bremerhaven, Germany

The multifold applications of the underwater mass spectrometer (UMS) include energy exploration, ocean acidification, and ecosystem studies (1,2). A mechanically supported membrane interface assembly allows for deepwater UMS applications. The UMS membrane interface assembly consists of a polydimethyl siloxane (PDMS) capillary mechanically supported by a

fabricated frit. The fabrication of the frit requires using a Dremel tool with a diamond-coated wheel to shape larger Hastalloy C porous

frits to the specified size. This frit fabrication procedure is time-consuming and cumbersome, and the porosity of the final frits is very difficult, if not impossible, to replicate. We report on new porous metallic structures that facilitate the fabrication of the membrane assembly.

The limit of detection (methane concentration in the tens of nanomolars) and sensitivity (on the order of 10-1 pico-amps/nanomole of methane) were comparable with those obtained with the previously fabricated Hastalloy C frits. The calibration parameters for the supported

membrane interface assembly depend on flow rate. We could achieve further optimization of the new frits by changing the dimensions of the channel through which the sample flows.

SELECTED REFERENCES 1. Bell, R.J., et al. (2011), Limnol. Oceanogr.-Meth.

9: pp. 164-175. 2. Cardenas-Valencia, A.M., et al. (2013), Rapid

Commun. in Mass Sp. 27: pp. 635-642.

SUMMARY

Microcontroller/ Peripheral board

Roughing pump

Pump

Membrane assembly

Electronics

Methane calibration curves with German frit 1200 (32.5% Porosity) (~ 1/8” diam)

y = 4.25E-14x + 1.58E-14 R² = 0.9997

0

2E-13

4E-13

6E-13

8E-13

1E-12

1.2E-12

1.4E-12

1.6E-12

1.8E-12

0 10 20 30 40

Mas

s spe

c sig

nal (

m/z

=15)

Concentration µmol/kg

y = 1.93E-13x - 6.67E-14 R² = 9.98E-01

y = 2.29E-13x - 1.31E-14 R² = 9.99E-01

y = 2.51E-13x - 5.55E-14 R² = 9.98E-01

y = 2.66E-13x - 3.89E-14 R² = 9.99E-01

0

2E-12

4E-12

6E-12

8E-12

1E-11

1.2E-11

0 10 20 30 40 50

8101520

Mas

s spe

c sig

nal (

m/z

=15)

Concentration µmol/kg

Detection limit for 10 mL/min (using 1 σ)= 20 nM

Flow, mL/min

Methane calibration curves with German frit 1200 (48.3% Porosity) (~ 1/8” diam)

CONTACT INFORMATION [email protected]

(727) 498-6721

Detection limit (using 1 σ)= 80 nM

y = 2.73E-13x + 1.54E-13 R = 0.9977

0

2E-12

4E-12

6E-12

8E-12

1E-11

1.2E-11

0 10 20 30 40

Methane calibration curves with home-made frit (~1/8” diam)

Mas

s spe

c sig

nal (

m/z

=15)

Concentration µmol/kg

Detection limit (using 1 σ)= 30 nM

Heater cartridges

High Pressure Membrane Assembly enabled the development of

underwater mass spectrometry

PDMS membrane

PEEK cap

Vacuum

Sample inlet

Heater block

Thermocouple

Epoxy

Sample outlet

Frit

The Fraunhofer Institute in Dresden, Germany, used powder metallurgical processes to manufacture frits with diameters of approximately 3.0 mm (1/8”) and known porosities of 48.3 % and 32.5% These frits were used to fabricate new membrane interface assemblies. Using a custom-heated membrane probe with the new porous frits, we

were able to calibrate dissolved methane concentrations to mass spectrometer responses (m/z 15) using linear least-squares fitting procedures.

METHOD