RATIONALE FOR THIS WORK
RESULTS
Our two research groups used metallic porous frits to mechanically support PDMS membranes, allowing the deployment of the UMSs to ocean depths of 2000 meters
MEMBRANE INTERFACE EVALUATIONS FOR UNDERWATER MASS SPECTROMETERS A. M. Cardenas-Valencia1, T. Gentz2, M. Schlueter2, S. K. Toler1, R. T. Short1
1 Marine and Space Sensing Program, SRI International, U.S.A,2 Alfred Wegener Institute for Polar and Marine Research (AWI), Bremerhaven, Germany
The multifold applications of the underwater mass spectrometer (UMS) include energy exploration, ocean acidification, and ecosystem studies (1,2). A mechanically supported membrane interface assembly allows for deepwater UMS applications. The UMS membrane interface assembly consists of a polydimethyl siloxane (PDMS) capillary mechanically supported by a
fabricated frit. The fabrication of the frit requires using a Dremel tool with a diamond-coated wheel to shape larger Hastalloy C porous
frits to the specified size. This frit fabrication procedure is time-consuming and cumbersome, and the porosity of the final frits is very difficult, if not impossible, to replicate. We report on new porous metallic structures that facilitate the fabrication of the membrane assembly.
The limit of detection (methane concentration in the tens of nanomolars) and sensitivity (on the order of 10-1 pico-amps/nanomole of methane) were comparable with those obtained with the previously fabricated Hastalloy C frits. The calibration parameters for the supported
membrane interface assembly depend on flow rate. We could achieve further optimization of the new frits by changing the dimensions of the channel through which the sample flows.
SELECTED REFERENCES 1. Bell, R.J., et al. (2011), Limnol. Oceanogr.-Meth.
9: pp. 164-175. 2. Cardenas-Valencia, A.M., et al. (2013), Rapid
Commun. in Mass Sp. 27: pp. 635-642.
SUMMARY
Microcontroller/ Peripheral board
Roughing pump
Pump
Membrane assembly
Electronics
Methane calibration curves with German frit 1200 (32.5% Porosity) (~ 1/8” diam)
y = 4.25E-14x + 1.58E-14 R² = 0.9997
0
2E-13
4E-13
6E-13
8E-13
1E-12
1.2E-12
1.4E-12
1.6E-12
1.8E-12
0 10 20 30 40
Mas
s spe
c sig
nal (
m/z
=15)
Concentration µmol/kg
y = 1.93E-13x - 6.67E-14 R² = 9.98E-01
y = 2.29E-13x - 1.31E-14 R² = 9.99E-01
y = 2.51E-13x - 5.55E-14 R² = 9.98E-01
y = 2.66E-13x - 3.89E-14 R² = 9.99E-01
0
2E-12
4E-12
6E-12
8E-12
1E-11
1.2E-11
0 10 20 30 40 50
8101520
Mas
s spe
c sig
nal (
m/z
=15)
Concentration µmol/kg
Detection limit for 10 mL/min (using 1 σ)= 20 nM
Flow, mL/min
Methane calibration curves with German frit 1200 (48.3% Porosity) (~ 1/8” diam)
CONTACT INFORMATION [email protected]
(727) 498-6721
Detection limit (using 1 σ)= 80 nM
y = 2.73E-13x + 1.54E-13 R = 0.9977
0
2E-12
4E-12
6E-12
8E-12
1E-11
1.2E-11
0 10 20 30 40
Methane calibration curves with home-made frit (~1/8” diam)
Mas
s spe
c sig
nal (
m/z
=15)
Concentration µmol/kg
Detection limit (using 1 σ)= 30 nM
Heater cartridges
High Pressure Membrane Assembly enabled the development of
underwater mass spectrometry
PDMS membrane
PEEK cap
Vacuum
Sample inlet
Heater block
Thermocouple
Epoxy
Sample outlet
Frit
The Fraunhofer Institute in Dresden, Germany, used powder metallurgical processes to manufacture frits with diameters of approximately 3.0 mm (1/8”) and known porosities of 48.3 % and 32.5% These frits were used to fabricate new membrane interface assemblies. Using a custom-heated membrane probe with the new porous frits, we
were able to calibrate dissolved methane concentrations to mass spectrometer responses (m/z 15) using linear least-squares fitting procedures.
METHOD