Medical ultrasound imaging
J.Hozman, E.Dove, J.Kybic
2008–2020
![Page 2: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/2.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detection
Part I
Introduction to medical ultrasound
![Page 3: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/3.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detection
Introduction
Ultrasound acousticsWavesWave equationReflection and refractionInterface reflectionAttenuation
Medical ultrasoundDevicesCardiologic USIntravascular USArtefacts
Generation/detectionGenerationSteering/BeamformingFocusingProcessing and control
![Page 4: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/4.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detection
Medical ultrasound basics
• Acoustic waves, frequency 2 ∼ 50MHz• Measure the time and intensity of the echo• Harmless• Stopped by air and dense tissues (bone)
![Page 5: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/5.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detection
Ultrasound Principle
![Page 6: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/6.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionWaves Wave equation Reflection and refraction Interface reflection Attenuation
Introduction
Ultrasound acousticsWavesWave equationReflection and refractionInterface reflectionAttenuation
Medical ultrasoundDevicesCardiologic USIntravascular USArtefacts
Generation/detectionGenerationSteering/BeamformingFocusingProcessing and control
![Page 7: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/7.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionWaves Wave equation Reflection and refraction Interface reflection Attenuation
Sinusoidal pressure source
![Page 8: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/8.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionWaves Wave equation Reflection and refraction Interface reflection Attenuation
Fyzikální základy � rozsahy veliþin
M��ená veliþina Symbol Jednotka (rozm�r)Rozsah obvyklých hodnotm��ené veliþiny v klinické
praxiRychlost c m.s
-11540 m.s
-1 (m�kká tká�)
Vlnová délka O mm 0,6 a� 0,15 mm (m�kká tká�)
Kmitoþet f hertz 2,5 a� 10 MHz
Modul pru�nosti E pascal 25 GPa (kost)
Akustická impedance Z kg.m-2
.s-1
1,63.106 kg.m
-2.s
-1
Hustota U kg.m-3
1000 kg.m-3
(voda)
Intenzita I W.cm-2
typicky 1 a� 10 mW. cm-2
Tlak p pascal nebo bar 0,6 baru nebo 0,06 MPa
![Page 9: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/9.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionWaves Wave equation Reflection and refraction Interface reflection Attenuation
Elementary volume
Speed u, pressure p, density %, area A, mass m.
![Page 10: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/10.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionWaves Wave equation Reflection and refraction Interface reflection Attenuation
Newton’s law
Motion along z :
F = ma = mdudt
= m
(∂u
∂t+∂u
∂z
∂z
∂t
)≈ m
∂u
∂t
force F = pA:
(p(z)− p(z + ∆z))A = m∂u
∂t
for ∆z � z :
−∂p∂z
∆z A = m∂u
∂t
as m = ρA∆z
−∂p∂z
= ρ∂u
∂t
![Page 11: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/11.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionWaves Wave equation Reflection and refraction Interface reflection Attenuation
Conservation of mass law
Difference of entering and exiting mass, density change:
A(u(z + ∆z)ρ(z + ∆z)− u(z)ρ(z)
)= −A∆z
∂ρ
∂t
for ∆z � z :
∂ρu
∂z= −∂ρ
∂t
density ρ = ρ0 + ρ1, ρ0 = const, ρ1 � ρ0:
ρ0∂u
∂z+∂ρ1
∂t= 0
Compressibility (stlačitelnost) ρ1ρ0
= Kp, K = 1/E :
∂u
∂z+ K
∂p
∂t= 0
![Page 12: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/12.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionWaves Wave equation Reflection and refraction Interface reflection Attenuation
1D wave equation
ρ∂u
∂t+∂p
∂z= 0 derive by z
∂u
∂z+ K
∂p
∂t= 0 derive by t
ρ∂2u
∂t∂z+∂2p
∂z2 = 0
∂2u
∂z∂t+ K
∂2p
∂t2= 0
subtract
∂2p
∂z2 − Kρ∂2p
∂t2= 0
similarly
∂2u
∂z2 − Kρ∂2u
∂t2= 0
![Page 13: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/13.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionWaves Wave equation Reflection and refraction Interface reflection Attenuation
Wave equation solution
Harmonic wave:
p = p+ cos(ωt − kz︸ ︷︷ ︸φ
)
where k is the wave number (vlnové číslo) [rad/m].
Wave speed (phase velocity):
φ0 = ωt − kz → z =ω
kt − φ0
k
c = ω/k
c = λf because ω = 2πf , k =2πλ
![Page 14: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/14.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionWaves Wave equation Reflection and refraction Interface reflection Attenuation
Wave speed
p = p+ cos(ωt − kz︸ ︷︷ ︸φ
)
∂2p
∂z2 = −p+k2 cos(ωt − kz)
∂2p
∂t2= −p+ω2 cos(ωt − kz)
The wave equation
∂2p
∂z2 = Kρ∂2p
∂t2
holds if
k2 = ρKω2 → c =1√ρK
=
√E
ρbecause c =
ω
k
![Page 15: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/15.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionWaves Wave equation Reflection and refraction Interface reflection Attenuation
Speed of Sound in Tissue
• The speed of sound in a human tissue depends on the average density U(kg·m3) and the compressibility K(m2·N-1) of the tissue. 0
1c
KU
![Page 16: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/16.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionWaves Wave equation Reflection and refraction Interface reflection Attenuation
Other wave equation solution
p = p− cos(ωt + kz)
Any forward or backward wave (by linearity and harmonicdecomposition).
p = f+(z + ct) + f−(z − ct)
Forward and backward wave combination:
p = p′(
cos(ωt − kz) + cos(ωt + kz))
Standing wave:
p = 2p′ cos(ωt) cos(kz)
![Page 17: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/17.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionWaves Wave equation Reflection and refraction Interface reflection Attenuation
Tissue Characteristics
• Engineers and scientists working in ultrasound have found that a convenient way of expressing relevant tissue properties is to use characteristic (or acoustic) impedance Z (kg·m-2
·s-1)
0Z cU
![Page 18: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/18.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionWaves Wave equation Reflection and refraction Interface reflection Attenuation
Acoustic impedance
Za =p (pressure)I (flow)
[Pa · s/m3]
“acoustic Ohm”.
For an infinite tube:
Za =ρ0c
S
Z = ρ0c is a specific acoustic impedance.
Unit [kg/s ·m2]=1 Rayl.
![Page 19: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/19.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionWaves Wave equation Reflection and refraction Interface reflection Attenuation
Fyzikální základy - veliþiny
MateriálRychlost zvuku
cm.s-1
HustotaU
kg.m-3
Akustická impedanceZ
kg.m-2.s-1 (x 106)
Koeficient absorpceD
dB.cm-1.MHz-1
Vzduch 330 1,3 0,00043
Tuk 1470 970 1,42 0,6
Ricínový olej 1500 933 1,40
Voda 1492 1000 1,48
M�kká tká� 1500 <1000 ~1,45 1,0
Mozek 1530 1020 1,56 0,85
Krev 1570 1020 1,60 0,18
Ledviny 1561 1030 1,61
Játra 1549 1060 1,64 0,9
Sval 1568 1040 1,63
Sval (podélná vlákna) 1,2 1,65
Sval (p�íþná vlákna) 3,3 1,65
Oþní þoþky 1620 1130 1,83 2,0
Kost 4080 1700 6,12 6,1
Plast 3,2 2,0
![Page 20: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/20.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionWaves Wave equation Reflection and refraction Interface reflection Attenuation
Specular Reflection
• The first, specular echoes, originate from relatively large, strongly reflective, regularly shaped objects with smooth surfaces. These reflections are angle dependent, and are described by reflectivity equation . This type of reflection is called specular reflection.
![Page 21: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/21.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionWaves Wave equation Reflection and refraction Interface reflection Attenuation
Primární parametrické pole amodulace ultrazvukového signálu
- odraz a lom UZV vln,
- útlum UZV energie,
![Page 22: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/22.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionWaves Wave equation Reflection and refraction Interface reflection Attenuation
Snell’s law
sin θ2sin θ1
=v2
v1=
c2c1
![Page 23: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/23.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionWaves Wave equation Reflection and refraction Interface reflection Attenuation
Reflectivity
2 1
2 1
_t ir
i
t i
Z Zcos cosp
RZ Zp
cos cos
T T
T T
�
2 1
2 1
Z ZR
Z Z�
�
At normal incidence, Ti = Tt = 0 and
![Page 24: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/24.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionWaves Wave equation Reflection and refraction Interface reflection Attenuation
Reflectivity for Various Tissues
Materials at Interface Reflectivity Brain-skull bone 0.66 Fat-muscle 0.10 Fat-kidney 0.08 Muscle-blood 0.03 Soft tissue-water 0.05 Soft tissue-air 0.9995
![Page 25: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/25.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionWaves Wave equation Reflection and refraction Interface reflection Attenuation
Scattered Reflection
• The second type of echoes are scattered that originate from small, weakly reflective, irregularly shaped objects, and are less angle-dependent and less intense. The mathematical treatment of non-specular reflection (sometimes called “speckle”) involves the Rayleigh probability density function. This type of reflection, however, sometimes dominates medical images, as you will see in the laboratory demonstrations.
![Page 26: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/26.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionWaves Wave equation Reflection and refraction Interface reflection Attenuation
![Page 27: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/27.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionWaves Wave equation Reflection and refraction Interface reflection Attenuation
Echoes from Two Interfaces
![Page 28: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/28.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionWaves Wave equation Reflection and refraction Interface reflection Attenuation
Echoes from Internal Organ
![Page 29: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/29.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionWaves Wave equation Reflection and refraction Interface reflection Attenuation
Attenuation
• Most engineers and scientists working in the ultrasound characterize attenuation as the “half-value layer,” or the “half-power distance.” These terms refer to the distance that ultrasound will travel in a particular tissue before its amplitude or energy is attenuated to half its original value.
![Page 30: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/30.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionWaves Wave equation Reflection and refraction Interface reflection Attenuation
Attenuation
• Divergence of the wavefront• Elastic reflection of wave energy• Elastic scattering of wave energy• Absorption of wave energy
![Page 31: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/31.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionWaves Wave equation Reflection and refraction Interface reflection Attenuation
Ultrasound Attenuation
Material Half–power distance (cm) Water 380 Blood 15 Soft tissue 5 to 1 except muscle 1 to 0.6 Bone 0.7 to 0.2 Air 0.08 Lung 0.05
![Page 32: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/32.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionWaves Wave equation Reflection and refraction Interface reflection Attenuation
Attenuation
• As a general rule, the attenuation coefficient is doubled when the frequency is doubled.
^ `0 exp 2avgI I zD �
![Page 33: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/33.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionWaves Wave equation Reflection and refraction Interface reflection Attenuation
Fyzikální základy - veliþiny
MateriálRychlost zvuku
cm.s-1
HustotaU
kg.m-3
Akustická impedanceZ
kg.m-2.s-1 (x 106)
Koeficient absorpceD
dB.cm-1.MHz-1
Vzduch 330 1,3 0,00043
Tuk 1470 970 1,42 0,6
Ricínový olej 1500 933 1,40
Voda 1492 1000 1,48
M�kká tká� 1500 <1000 ~1,45 1,0
Mozek 1530 1020 1,56 0,85
Krev 1570 1020 1,60 0,18
Ledviny 1561 1030 1,61
Játra 1549 1060 1,64 0,9
Sval 1568 1040 1,63
Sval (podélná vlákna) 1,2 1,65
Sval (p�íþná vlákna) 3,3 1,65
Oþní þoþky 1620 1130 1,83 2,0
Kost 4080 1700 6,12 6,1
Plast 3,2 2,0
![Page 34: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/34.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionDevices Cardiologic US Intravascular US Artefacts
Introduction
Ultrasound acousticsWavesWave equationReflection and refractionInterface reflectionAttenuation
Medical ultrasoundDevicesCardiologic USIntravascular USArtefacts
Generation/detectionGenerationSteering/BeamformingFocusingProcessing and control
![Page 35: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/35.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionDevices Cardiologic US Intravascular US Artefacts
Medical ultrasound devices
![Page 36: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/36.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionDevices Cardiologic US Intravascular US Artefacts
Medical ultrasound devices
![Page 37: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/37.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionDevices Cardiologic US Intravascular US Artefacts
Medical ultrasound devices
![Page 38: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/38.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionDevices Cardiologic US Intravascular US Artefacts
Medical applications of ultrasound imaging
• Cardiology (heart)• Gynecology: breast, fetus (pregnancy)• Internal organs: liver, kidney, thyroid gland• Intravascular ultrasound• Therapeutic ultrasound: shock wave (kidney stone), thermal
effects (rehabilitation)
![Page 39: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/39.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionDevices Cardiologic US Intravascular US Artefacts
Imaging modes
A osciloscopic, intensity/timeB 2D in the probe planeC 2D perpendicular
TM 1D+timeQ Doppler (speed)
![Page 40: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/40.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionDevices Cardiologic US Intravascular US Artefacts
![Page 41: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/41.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionDevices Cardiologic US Intravascular US Artefacts
Introduction
Ultrasound acousticsWavesWave equationReflection and refractionInterface reflectionAttenuation
Medical ultrasoundDevicesCardiologic USIntravascular USArtefacts
Generation/detectionGenerationSteering/BeamformingFocusingProcessing and control
![Page 42: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/42.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionDevices Cardiologic US Intravascular US Artefacts
Conventional Cardiac 2D Ultrasound
![Page 43: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/43.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionDevices Cardiologic US Intravascular US Artefacts
Heart
![Page 44: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/44.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionDevices Cardiologic US Intravascular US Artefacts
Heart
![Page 45: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/45.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionDevices Cardiologic US Intravascular US Artefacts
Heart
![Page 46: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/46.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionDevices Cardiologic US Intravascular US Artefacts
B-mode Image of Heart
![Page 47: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/47.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionDevices Cardiologic US Intravascular US Artefacts
Traditional Ultrasound Images
End-diastole End-systole
![Page 48: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/48.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionDevices Cardiologic US Intravascular US Artefacts
![Page 49: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/49.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionDevices Cardiologic US Intravascular US Artefacts
Introduction
Ultrasound acousticsWavesWave equationReflection and refractionInterface reflectionAttenuation
Medical ultrasoundDevicesCardiologic USIntravascular USArtefacts
Generation/detectionGenerationSteering/BeamformingFocusingProcessing and control
![Page 50: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/50.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionDevices Cardiologic US Intravascular US Artefacts
Progression of Vascular Disease
![Page 51: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/51.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionDevices Cardiologic US Intravascular US Artefacts
IVUS Catheter
• 1 - Rotating shaft• 2 - Acoustic window• 3 - Ultrasound crystal• 4 - Rotating beveled acoustic
mirror
![Page 52: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/52.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionDevices Cardiologic US Intravascular US Artefacts
Slightly Diseased Artery in Cross-section
PlaqueCatheter
![Page 53: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/53.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionDevices Cardiologic US Intravascular US Artefacts
An array of Images
![Page 54: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/54.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionDevices Cardiologic US Intravascular US Artefacts
3D IVUS
![Page 55: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/55.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionDevices Cardiologic US Intravascular US Artefacts
Other ultrasound examples
![Page 56: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/56.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionDevices Cardiologic US Intravascular US Artefacts
Other ultrasound examples
![Page 57: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/57.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionDevices Cardiologic US Intravascular US Artefacts
Other ultrasound examples
![Page 58: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/58.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionDevices Cardiologic US Intravascular US Artefacts
Introduction
Ultrasound acousticsWavesWave equationReflection and refractionInterface reflectionAttenuation
Medical ultrasoundDevicesCardiologic USIntravascular USArtefacts
Generation/detectionGenerationSteering/BeamformingFocusingProcessing and control
![Page 59: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/59.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionDevices Cardiologic US Intravascular US Artefacts
Geometrická distorze UZV zobrazení
- zm�nou rychlosti �í�ení UZV vlny,
![Page 60: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/60.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionDevices Cardiologic US Intravascular US Artefacts
Geometrická distorze UZV zobrazení
- skladbou tkání,
![Page 61: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/61.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionDevices Cardiologic US Intravascular US Artefacts
Geometrická distorze UZV zobrazení
- násobnou reflexí,
![Page 62: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/62.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionDevices Cardiologic US Intravascular US Artefacts
Geometrická distorze UZV zobrazení
- vlivem koneþné �í�ky UZV svazku,
![Page 63: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/63.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionDevices Cardiologic US Intravascular US Artefacts
Geometrická distorze UZV zobrazení
- pohybem tká�ových struktur,
![Page 64: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/64.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionGeneration Steering/Beamforming Focusing Processing and control
Introduction
Ultrasound acousticsWavesWave equationReflection and refractionInterface reflectionAttenuation
Medical ultrasoundDevicesCardiologic USIntravascular USArtefacts
Generation/detectionGenerationSteering/BeamformingFocusingProcessing and control
![Page 65: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/65.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionGeneration Steering/Beamforming Focusing Processing and control
Generace a detekce UZV signálu
- generace UZV impulsu 10 a� 100 mW/cm2,- vysoký dosa�ený odstup S/�,- malá akustická vazba mezi jednotlivými m�niþi,
- po�adavky na konstrukci systému,
- krátký generovaný impuls ~ 2Ps,
- lehká, snadno manipulovatelná konstrukce,
- vysoká úþinnost p�enosu energie mezi m�niþi,
- tlumení zp�tné akustické vlny,
- dosa�ení �irokého úhlového krytí snímaného pole,
- potlaþení vibrací,
- princip vstupní jednotky digitálního sonografu,
![Page 66: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/66.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionGeneration Steering/Beamforming Focusing Processing and control
Zdroje ultrazvukového vln�ní
- zdrojem UZV vln�ní UZV m�niþ v sond�,
- p�ímý a nep�ímý piezoelektrický jev,
- charakteristickým parametrem sondy je rezonanþnífrekvence, urþená tl. m�niþe,
- co nejkrat�í impuls p�i vysílání x velká citlivost,
![Page 67: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/67.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionGeneration Steering/Beamforming Focusing Processing and control
Pressure Generation
• Piezoelectric crystal • ‘piezo’ means pressure, so piezoelectric
means – pressure generated when electric field is
applied– electric energy generated when pressure is
applied
![Page 68: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/68.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionGeneration Steering/Beamforming Focusing Processing and control
Charged Piezoelectric Molecules
Highly simplified effect of E field
![Page 69: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/69.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionGeneration Steering/Beamforming Focusing Processing and control
Piezoelectric Effect
![Page 70: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/70.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionGeneration Steering/Beamforming Focusing Processing and control
Zpracování UZV signálu
- p�izp$sobení akustických impedancí
![Page 71: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/71.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionGeneration Steering/Beamforming Focusing Processing and control
Transducer
![Page 72: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/72.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionGeneration Steering/Beamforming Focusing Processing and control
Pressure Radiated by Sharp Pulse
![Page 73: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/73.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionGeneration Steering/Beamforming Focusing Processing and control
Zdroje ultrazvukového vln�ní
- ultrazvukové pole,
- blízké pole (blízká Fresnelova oblast),
- vzdálené pole (vzdálená Fraunhoferova oblast),
OO
4
22 �
DL
![Page 74: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/74.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionGeneration Steering/Beamforming Focusing Processing and control
Zdroje ultrazvukového vln�ní
- vyza�ovací diagram sondy,
- postranní laloky - tlumení x akustická vazba,
- významnou úlohu sehrává pom�r ,OD
DO
- 22,1sin
bO
- sin
- Fraunhoferova formule,
![Page 75: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/75.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionGeneration Steering/Beamforming Focusing Processing and control
Rozli�ovací schopnost
![Page 76: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/76.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionGeneration Steering/Beamforming Focusing Processing and control
Introduction
Ultrasound acousticsWavesWave equationReflection and refractionInterface reflectionAttenuation
Medical ultrasoundDevicesCardiologic USIntravascular USArtefacts
Generation/detectionGenerationSteering/BeamformingFocusingProcessing and control
![Page 77: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/77.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionGeneration Steering/Beamforming Focusing Processing and control
Zpracování UZV signálu
- vychylování UZV svazku - poziþní jednotka
- mech. systémy s lineárním snímáním,
- mech. systémy se sektorovým snímáním,
- elektronické systémy s lineárním snímáním,
- elektronické systémy se sektorovým snímáním,
- rotaþní systém,
- systém s kývající sondou,
- fokuzace UZV svazku
![Page 78: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/78.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionGeneration Steering/Beamforming Focusing Processing and control
UZV sonda s mech. rozkladem - Siemens
![Page 79: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/79.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionGeneration Steering/Beamforming Focusing Processing and control
UZV sonda s mech. rozkladem - Siemens
![Page 80: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/80.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionGeneration Steering/Beamforming Focusing Processing and control
El. systémy s lineárním snímáním
![Page 81: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/81.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionGeneration Steering/Beamforming Focusing Processing and control
El. systémy se sektorovým snímáním
![Page 82: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/82.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionGeneration Steering/Beamforming Focusing Processing and control
Introduction
Ultrasound acousticsWavesWave equationReflection and refractionInterface reflectionAttenuation
Medical ultrasoundDevicesCardiologic USIntravascular USArtefacts
Generation/detectionGenerationSteering/BeamformingFocusingProcessing and control
![Page 83: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/83.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionGeneration Steering/Beamforming Focusing Processing and control
Fokuzace svazku UZV signálu - typy
- fokuzace zrcadly,
- elektronická fokuzace,
- statická,
- dynamická,
- fokuzace UZV þoþkou,
- s lineární �adou m�niþ$,
- s anulární sondou,
- v re�imu vysílání,
- v re�imu p�íjmu,
- velikostí apertury,
- v re�imu vysílání,
- elektronicko-optická,
![Page 84: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/84.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionGeneration Steering/Beamforming Focusing Processing and control
Fokuzace þoþkou
![Page 85: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/85.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionGeneration Steering/Beamforming Focusing Processing and control
Fokuzace þoþkou
![Page 86: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/86.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionGeneration Steering/Beamforming Focusing Processing and control
Fokuzace þoþkou
O..44,2 ¸̧¹
·¨̈©
§
D
ld f
![Page 87: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/87.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionGeneration Steering/Beamforming Focusing Processing and control
El. fok. stat. s lin. �ad. m�n. v r. vysílání
![Page 88: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/88.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionGeneration Steering/Beamforming Focusing Processing and control
El. fok. stat. s lin. �ad. m�n. v r. p�íjmu
![Page 89: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/89.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionGeneration Steering/Beamforming Focusing Processing and control
Phased Linear Array
![Page 90: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/90.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionGeneration Steering/Beamforming Focusing Processing and control
Beam Direction
![Page 91: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/91.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionGeneration Steering/Beamforming Focusing Processing and control
El.-optická fok. stat. s lin. �ad. m�n.
![Page 92: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/92.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionGeneration Steering/Beamforming Focusing Processing and control
Introduction
Ultrasound acousticsWavesWave equationReflection and refractionInterface reflectionAttenuation
Medical ultrasoundDevicesCardiologic USIntravascular USArtefacts
Generation/detectionGenerationSteering/BeamformingFocusingProcessing and control
![Page 93: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/93.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionGeneration Steering/Beamforming Focusing Processing and control
Zpracování elektrického sign. - B mód
![Page 94: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/94.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionGeneration Steering/Beamforming Focusing Processing and control
Zesilovaþe s þasov� �ízeným zesílením
![Page 95: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/95.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionGeneration Steering/Beamforming Focusing Processing and control
Amplitudov� �ízené zesilovaþe
![Page 96: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/96.jpg)
Introduction Ultrasound acoustics Medical ultrasound Generation/detectionGeneration Steering/Beamforming Focusing Processing and control
Geom. vztah sekt. sním. a TV zobr. rastru
![Page 97: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/97.jpg)
Doppler ultrasound US contrast agents Harmonic imaging 3D US imaging
Part II
Modern ultrasound imaging
![Page 98: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/98.jpg)
Doppler ultrasound US contrast agents Harmonic imaging 3D US imaging
Doppler ultrasound
US contrast agents
Harmonic imaging
3D US imaging
![Page 99: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/99.jpg)
Doppler ultrasound US contrast agents Harmonic imaging 3D US imaging
�������������� ������ �������������� ������ �� !�� !����"
��#$�%� ��&�'()�
*+,��-�,�.�)'�/)�������0(��/�102) ��/��- 3$*$-443
![Page 100: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/100.jpg)
Doppler ultrasound US contrast agents Harmonic imaging 3D US imaging
*+,��-�,�.�)'�/)�������0(��/�102) ��/��- 3$*$-443
356-�7�8��(91)�:�����1:��/)��������9
356+�7�2��:����.��9�/���)':
356;�7�:<�:��(:��=1����/ :��
35*;�7���>=�:2���.��9�/���)':
?�-@$33$354*��)1'.9�#!� 9����)A�3+$*$35;*��:���:!���)1�
�B�����)�� �C�:)������1:���)2�9�2D�8�'�2�)�()�:()��2"
![Page 101: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/101.jpg)
Doppler ultrasound US contrast agents Harmonic imaging 3D US imaging
*+,��-�,�.�)'�/)�������0(��/�102) ��/��- 3$*$-443
3$���)����=����'C����)���B�.9������:�� ���()>
����1:��/)�8�:2/:��:
�
��
���
��
�
�
��� �
�
��
��� �
�
���
������
��
� ��
�� �
�
���
��
����1)> � ���()>
��
��
![Page 102: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/102.jpg)
Doppler ultrasound US contrast agents Harmonic imaging 3D US imaging
*+,��-�,�.�)'�/)�������0(��/�102) ��/��- 3$*$-443
-$���)����=����� ���()>�)���B�.9������:�'C���
�
��
���
���
����
��
�
��
���
���
�
�
���
�
�
�
�
�� ��� ���
����1)> � ���()>
��
��
��
�� ��
�� ���
![Page 103: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/103.jpg)
Doppler ultrasound US contrast agents Harmonic imaging 3D US imaging
*+,��-�,�.�)'�/)�������0(��/�102) ��/��- 3$*$-443
��9F���(���'/��:�C���)�1���/�� )C�
�G:�B���>1:�����<-�)�/�GG��(������9�')�:C.=(:�)�����F:�������� �( F:(:���=�
����1:��/)8�:2/:��:
��
�
�
� �
�
��
��
��
�
�
�
�
����
��
� �
����
��
�
��
� ��
�� �
�
���
� �� � ��
� ��
�� �
�
���
��
![Page 104: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/104.jpg)
Doppler ultrasound US contrast agents Harmonic imaging 3D US imaging
Blood flow speed measurement
• Doppler effect: Frequence changes if the source moves withrespect to the receiver.• Reflection from red blood cells• Red blood cells
• Moving receiver• Moving source
• Doppler shiftfr = ft + fd fd ≈ 2
v
cfc
• We only measure the projection along the ray.
![Page 105: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/105.jpg)
Doppler ultrasound US contrast agents Harmonic imaging 3D US imaging
*+,��-�,�.�)'�/)�������0(��/�102) ��/��- 3$*$-443
� ��)�D���#�=1/���1)�D���#�=1
�� ����� � �� � ���� ��� �� �� ������
/���1)>
� ���()>
����1=���
���=��.�>
':��1�/)>
�:(�C91)�:�����1:��/)���#�=19
![Page 106: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/106.jpg)
Doppler ultrasound US contrast agents Harmonic imaging 3D US imaging
*+,��-�,�.�)'�/)�������0(��/�102) ��/��- 3$*$-443
%:C���1�/0��1�F2��/:�/DG:�9/:C:�0(�/'�)B9�'�)(:�)��H
�D�1:C2:(��=��.:���/�C:(�C91=���9��:
� � � ��� �� ����� ����� ������������ �� ��
� �� � � �� ���
�
���� ���� ��������
��������
����
���
�� � ���
���
��
� �������
� �� �� �����
� ���
� � �� �� � ������
C/���=��.:2�/���1)�0�8�:2/:��:
C/���=��.:2�/���1)�0�8�:2/:��:
��:����( ��=��1�F2)
����1:� /���#�=1!��:1':�9�>����( �
���1)>:��
���"���1)>:��
���"���1)>:��
�&�"
![Page 107: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/107.jpg)
Doppler ultrasound US contrast agents Harmonic imaging 3D US imaging
*+,��-�,�.�)'�/)�������0(��/�102) ��/��- ;$*$-443
����/���1)��9�8�:2/:����;�K&'�)�8�:2/:��������1:��/)��#�=19�;!5�2&'�(9��(:��C1�G���21)C�D��( ����29�2�/:!���$;!44;5�K&'�)�'=����D��( ����29�2�/:!���$�6!@@6-�K&'$
� :/0����( ��/�9���8��()��!�C)��9�'�)(0�2:(8�:2/:�>��B�����9/9��)����D���C�2=�����( �9!�2�:�D�.����C:(�C91)���' ��)1�')�B�/=�
,=21)C���(�G1:�2)
� �21)C
�( ��/0�C:(�C91)>�������0(�
![Page 108: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/108.jpg)
Doppler ultrasound US contrast agents Harmonic imaging 3D US imaging
*+,��-�,�.�)'�/)�������0(��/�102) ��/��- ;$*$-443
/���1)>
� ���()>
����1=���
���=��.�>
':��1�/)>
�=��.�> ��
� ����
��� �� ���
� � � ��� ��� �� � ������
�
��
,=21)C���9��� =C=����( ��/0B��C:(�C91=���9
![Page 109: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/109.jpg)
Doppler ultrasound US contrast agents Harmonic imaging 3D US imaging
*+,��-�,�.�)'�/)�������0(��/�102) ��/��- ;$*$-443
���� � �����
�� ��� � ��� �� ���
�
�
8='�/D����9/� ���������������)�������2/)C�)�9������#�=1���� �
� � ����� � �� ����
��29C�')�:C.=(:�)(�1��9C�!��)2�C���)�:(:
8��Q)�C�81�Q��C�� :C�D���2"�7��C����C�!���$ �� �� �
�:/:��:�81�Q��'� ��D���2"�7�2����C !���$ �� �� �
� �� �� ������ �� � ��� �� ���� �
�='�/0�'��)��/=���2/)C�)�9����B���#�=1 �$%�)�$&
![Page 110: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/110.jpg)
Doppler ultrasound US contrast agents Harmonic imaging 3D US imaging
*+,��-�,�.�)'�/)�������0(��/�102) ��/��- ;$*$-443
8='�/D���9/
8='�/D���9/ �
��
��
�
��
�
C�� :C�D��2
'� ��D��2��
��
�� �� ������ �
�� �� ����
��
�� �
�
�
�
�� �� ������ � �6"
�� �� ���� �*"
�� �� ���� � �-"
�� �� ������ � �3"
����� ���� �� ����������� ��
![Page 111: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/111.jpg)
Doppler ultrasound US contrast agents Harmonic imaging 3D US imaging
*+,��-�,�.�)'�/)�������0(��/�102) ��/��- ;$*$-443
����1=���3
���=��.�>
':��1�/)>
�=��.�> ��
� �� �� �����
�� �����
�
��
����1=���-
�=��.�>
�=��.�>
��
�
��
��
��
� �� �� �� ���
�� �� �� �
����1:� /��#�=1
R�1�D�����0(�2��:�)�)�����#�=1�/0��1�F2���C��/�C)����C�� :C�0(9�)�'� ��0(9���29�/:�8�:2/:�>����.1)���
![Page 112: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/112.jpg)
Doppler ultrasound US contrast agents Harmonic imaging 3D US imaging
*+,��-�,�.�)'�/)�������0(��/�102) ��/��- ;$*$-443
&1)/���(�G1:�2)�7����9���91�/0�8�:2/:��:!���$��91�/0���B1������)�����9�8�:2/:��� ��
� �������)�������� :C��)/9����B�.9�/:�8='�/0(���'C�19
� �� � ���� ��� �� �� ������
��#�=1�����0 ����1:� /���#�=1
��� � ��� � ����
��
���� �� ����
���
� � ����� �� ���� �� ��� �������
� �� ��������� ���� ���� ��� ���� �������
�
![Page 113: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/113.jpg)
Doppler ultrasound US contrast agents Harmonic imaging 3D US imaging
*+,��-�,�.�)'�/)�������0(��/�102) ��/��- 3-$*$-443
�����9=1�������1:��/�20�����0(���
7�2�����9=1���/���1=������ ��:(!
7�/����C �C/)���:'�7:1:(:���!
7�/'��2�/)����.�:(�C=��G� 2�9��,��/)'29!
7�(�F�����C:�:2�/)��/:120����B1������� ��29!
7��:��'1�G��B1�9.29�7�(��F��/����:2��=1���B��1�F:2$
![Page 114: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/114.jpg)
Doppler ultrasound US contrast agents Harmonic imaging 3D US imaging
*+,��-�,�.�)'�/)�������0(��/�102) ��/��- 3-$*$-443
�91'�������1:��/�20�����0(��� ���91�:C� )/:"
7��91'���/���1=������ ��:(!
7�/����C ��:C:����:'�7:1:(:��!
7�/'��2�/)����.�:(�C=��C012�9��91'9!
7��(:':�����'�)B9�C:�:2�/)�D�B����B1������� ��29!
7���'1�G��B1�9.29�7�(0� ���:2��=1���B��1�F:2$
![Page 115: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/115.jpg)
Doppler ultrasound US contrast agents Harmonic imaging 3D US imaging
*+,��-�,�.�)'�/)�������0(��/�102) ��/��- 3-$*$-443
�(:':�����'�)B9�C:�:2�/)�D�B����B1������� ��29�9��
�����
��
��
� ���
�
��
�� �
�
�
�
�
� �
�
��
�
� � ���
��
��
�
�
��
��
��� ��
�
�
�
�
�� � �
���
�
�
�
�
����
��
�
�� ��
�����
��
� ��
�
��
�� ���
����
�
�
���
�
�
������� �
��� �
�C=��
�������
![Page 116: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/116.jpg)
Doppler ultrasound US contrast agents Harmonic imaging 3D US imaging
*+,��-�,�.�)'�/)�������0(��/�102) ��/��- 3-$*$-443
�����2D�� �21)C
���
������
�
� ���� ����!"#� ��� ����
�:�/'C=1:������34�(�( F:(:��)( ���()<�(=1������B1������� �
��������
����
�!"
�!�
�!�
�!
�
!"
!�
!�
!
!� !� !�� ! ! � !�
��#$%��������
&%���'��(��
��#$%��������
&�%&'�)��
%%�*+� �
�
�"
&��)������%%�%��
����
�
���� ��
�� �
![Page 117: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/117.jpg)
Doppler ultrasound US contrast agents Harmonic imaging 3D US imaging
Doppler US — examples
![Page 118: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/118.jpg)
Doppler ultrasound US contrast agents Harmonic imaging 3D US imaging
Doppler US — examples
![Page 119: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/119.jpg)
Doppler ultrasound US contrast agents Harmonic imaging 3D US imaging
Doppler US — examples
![Page 120: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/120.jpg)
Doppler ultrasound US contrast agents Harmonic imaging 3D US imaging
Doppler ultrasound
US contrast agents
Harmonic imaging
3D US imaging
![Page 121: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/121.jpg)
Doppler ultrasound US contrast agents Harmonic imaging 3D US imaging
Contrast agents
• 1968, Gramiak, saline injection• Mikrobubbles (2 ∼ 5µm)• Asymetric compression/expansion• Stabilization (synthetic polymers), up to 5− 10min.• Injection.• Albunex, Optison, Echovist, Levovist. . .
![Page 122: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/122.jpg)
Doppler ultrasound US contrast agents Harmonic imaging 3D US imaging
Flash Contrast Imaging
US bubble destabilization.
normal
![Page 123: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/123.jpg)
Doppler ultrasound US contrast agents Harmonic imaging 3D US imaging
Flash Contrast Imaging
US bubble destabilization.
flash, bubbles broken
![Page 124: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/124.jpg)
Doppler ultrasound US contrast agents Harmonic imaging 3D US imaging
Flash Contrast ImagingUS bubble destabilization.
filling up
Myocardial perfusion evaluation.
![Page 125: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/125.jpg)
Doppler ultrasound US contrast agents Harmonic imaging 3D US imaging
Doppler ultrasound
US contrast agents
Harmonic imaging
3D US imaging
![Page 126: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/126.jpg)
Doppler ultrasound US contrast agents Harmonic imaging 3D US imaging
Nonlinear response
Assymetric bubble compression
![Page 127: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/127.jpg)
Doppler ultrasound US contrast agents Harmonic imaging 3D US imaging
Harmonic imaging• Transmit f0, receive 2f0
• Bandwith limitation• Bubbles not needed, tissue nonlinearity
standard US 2nd harmonic
![Page 128: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/128.jpg)
Doppler ultrasound US contrast agents Harmonic imaging 3D US imaging
Harmonic imaging• Transmit f0, receive 2f0• Bandwith limitation
• Bubbles not needed, tissue nonlinearity
standard US 2nd harmonic
![Page 129: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/129.jpg)
Doppler ultrasound US contrast agents Harmonic imaging 3D US imaging
Harmonic imaging• Transmit f0, receive 2f0• Bandwith limitation• Bubbles not needed, tissue nonlinearity
standard US 2nd harmonic
![Page 130: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/130.jpg)
Doppler ultrasound US contrast agents Harmonic imaging 3D US imaging
Pulse Inversion Harmonic Imaging• Two pulses, second inverted• Responses summed• Filtration not needed
• Several pulses (Power Pulse Inversion)
![Page 131: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/131.jpg)
Doppler ultrasound US contrast agents Harmonic imaging 3D US imaging
Pulse Inversion Harmonic Imaging• Two pulses, second inverted• Responses summed• Filtration not needed
• Several pulses (Power Pulse Inversion)
standard image (liver) pulse inversion
![Page 132: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/132.jpg)
Doppler ultrasound US contrast agents Harmonic imaging 3D US imaging
Pulse Inversion Harmonic Imaging• Two pulses, second inverted• Responses summed• Filtration not needed• Several pulses (Power Pulse Inversion)
standard image (liver) pulse inversion
![Page 133: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/133.jpg)
Doppler ultrasound US contrast agents Harmonic imaging 3D US imaging
Doppler ultrasound
US contrast agents
Harmonic imaging
3D US imaging
![Page 134: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/134.jpg)
Doppler ultrasound US contrast agents Harmonic imaging 3D US imaging
3D Reconstruction
![Page 135: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/135.jpg)
Doppler ultrasound US contrast agents Harmonic imaging 3D US imaging
3D Ultrasound
Traditional 2D New 3D
![Page 136: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/136.jpg)
Doppler ultrasound US contrast agents Harmonic imaging 3D US imaging
Real-time 3D Ultrasound
![Page 137: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/137.jpg)
Doppler ultrasound US contrast agents Harmonic imaging 3D US imaging
![Page 138: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/138.jpg)
Doppler ultrasound US contrast agents Harmonic imaging 3D US imaging
Velocity of Contraction
Normal Abnormal
![Page 139: Medical ultrasound imaging - cw.fel.cvut.cz](https://reader030.fdocuments.us/reader030/viewer/2022020621/61e826865b7c2e73625b72b9/html5/thumbnails/139.jpg)
Doppler ultrasound US contrast agents Harmonic imaging 3D US imaging
Conclusions
• Non-invasive, affordable and portable imaging technique• Excellent soft tissue imaging• Lower image quality (wrt CT or MRI) due to speckle but
improving• Low penetration depth versus resolution• Does not pass through air or gas• Does not pass through bones, shadows• Modern techniques — 3D, contract agents, Doppler