ELECTRICAL SAFETY
MEDICAL EQUIPMENT
By
Ismail Yusof
For
![Page 2: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/2.jpg)
Electrical Safety – Medical Equipment
Contents:
1 Various Test Equipment used in Biomedical
Engineering
2 Introduction to Electrical Safety
3 Electricity - Physiological Effects
4 General Electrical Safety
5 Electricity – Leakage Current
6 Electrical Equipment – Classes and Types
7 Electrical Safety Tests
![Page 3: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/3.jpg)
INTRODUCTION TO VARIOUS TEST
EQUIPMENT USED IN BIOMEDICAL
ENGINEERING
![Page 4: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/4.jpg)
Workshop Tool & Equipment
Toolkit for general repair
![Page 5: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/5.jpg)
Insulation and Resistance Measuring Instrument, 1000 V, Digital
Workshop Tool & Equipment
![Page 6: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/6.jpg)
Soldering Station
with temperature controller
Workshop Tool & Equipment
![Page 7: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/7.jpg)
Digital MultimeterAnalog Multimeter
Workshop Tool & Equipment
![Page 8: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/8.jpg)
Electrical Safety Analyser
Test Instrument
![Page 9: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/9.jpg)
Electrical Safety Analyser
Test Instrument
![Page 10: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/10.jpg)
Test Instrument
![Page 11: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/11.jpg)
It measures delivered energy on its internal load.
The defibrillator pulse is stored and it can be viewed
via ECG output leads, paddles or scope output.
Test Instrument
![Page 12: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/12.jpg)
Blood pressure simulator
that provides simulation
to test blood pressure monitors.
Test Instrument
![Page 13: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/13.jpg)
The pulse oximeter testing method to simulate
the light absorption and arterial blood flow
of a human finger
Test Instrument
![Page 14: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/14.jpg)
Power meter to service
the 1 MHz and 3 MHz therapy level
ultrasound generators.
Test Instrument
![Page 15: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/15.jpg)
External pacemaker tester designed for
accurately testing all types of
external pacemaker.
Test Instrument
![Page 16: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/16.jpg)
Test Instrument
![Page 17: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/17.jpg)
Measure output power and RF leakage
ESU Analyser
Test Instrument
![Page 18: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/18.jpg)
Infusion Pump Analyzer
Test Instrument
![Page 19: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/19.jpg)
SAFETY TESTERS
Safety Check
– Electrical Safety Tests (IEC 601)
– The basic purpose of safety testing in medical electrical
equipment is to be sure that a device is safe for the patient
and user. Such as:
• Grounding resistance of portable medical equipment measured between equipment chassis and ground pin of power plug shall not exceed 0.50 ohms.
![Page 21: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/21.jpg)
• Biomedical Engineers assist in struggle
against illness and disease by providing
support for materials and tools that can be
utilized for diagnosis and treatment by
health care professionals.
• Biomedical Engineer have achieved this
position by assuming responsibility for
managing the safe use of medical
equipment within the hospital environment
Biomedical Engineer-definition
![Page 22: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/22.jpg)
A Historical Perspective
• Engineer’s first enter to clinical scene : late
1960s in response to concerns about
electrical safety of hospital patient
• Ralph Nader (consumer activist) : “at the very
least, 1200 Americans are electrocuted
annually during routine diagnostic and
therapeutic procedures in hospitals” Ladies
Home Journal, April 24, 1970
• Raise the level of concern about safe use of
medical device
![Page 23: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/23.jpg)
A Historical Perspective• Overnight growth of a new industry : hospital electrical
safety
• Joint Commission on the Accreditation of Hospital : “hospital must inspect all equipment used on or near a patient for electrical safety at least every 6 months”
• Hospital administrator’s options
– Paying medical device manufacturer
– Service contracting
– In-house staff
![Page 24: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/24.jpg)
Safety Issue
• In USA, about 10,000 device-related
patient injuries each year
• Most injuries are attributable to improper
use of a device as a result of inadequate
training and lack of experience
• Medical personnel rarely read user
manuals until a problem occurred
• “Everything that can go wrong eventually
will go wrong” Murphy’s law
![Page 25: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/25.jpg)
Electrical Shock
• Electric Shock : unwanted or
unnecessary physiological response to
current.
• Three phenomena :
– (1) electrical stimulation of excitable tissue
(nerve & muscle)
– (2) resistive heating of tissue, and
– (3) electrochemical burns and tissue damage
![Page 26: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/26.jpg)
Electrical Safety
• Electric Safety is the best possible
limitation of hazardous electrical Macro –
and/or Microshocks, sustained by
patients, as well as explosion, fire or
damage to equipments and buildings.
![Page 27: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/27.jpg)
Single Phase Power Supply
• The application of the safety testing on this training will
only apply to the Single Phase ( 3 wires ) 100 to 240 Volt
power supply.
• Any medical equipment use either 3 or 2 pins plug which
connected to the wall socket shall follow this electrical
safety procedure.
![Page 28: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/28.jpg)
The IEC 601.1 Standard
Several standard serve as ruling authority in determine how medical equipment are to be tested such UL, CSA, CE, TUV …
The International Electro Technical Commission’s IEC 601-1 standard is accepted and implemented around the world.
All this explanation will be based on IEC 601-1 Standard
![Page 29: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/29.jpg)
The IEC 601.1 Standard
![Page 30: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/30.jpg)
Medical Electrical Equipment
Medical Electrical Equipment provided with no more than one connection to a particular supply mains and intended to;
1. Diagnose the patient,
2. Treat the patient,
3. Monitor the patient under medical supervision,
4. Makes physical or electrical contact with the patient ,
5. Transfers energy to or from the patient and/or detects such energy transfer to or from patient.
The equipment includes those accessories as defined by the manufacturer which are necessary to enable the normal use of the equipment.
![Page 31: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/31.jpg)
Safety of Medical Equipment
The purpose of safety testing medical electronic
equipment is to ensure that a device is safe from
electrical hazards to patients, maintenance personnel's
and users.
Electric shock are caused by electricity flowing through
the body after touching a damaged electrical device and
results muscle spasms, burns, cardiac and respiratory
arrest and Ventricular Fibrillation
![Page 32: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/32.jpg)
Electrical Safety
• Electric safety in hospital is a shared
responsibility between several parties, in
addition to the physician, including:
• The nurses
• All engineers (electrical, biomedical, facility,
etc)
• Manufacturers
• The hospital
![Page 33: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/33.jpg)
Electrical Safety – Critical Points
• The electrical installation, no matter how safe, is only
part of the safety requirements.
• Plugs and cords must be checked and rejected if
defective.
• Only devices tested for safety should be used.
• Electrical compatibility of the entire electrical system
must be tested regularly.
• Patients leads must be attached and connected
properly.
• Radio-frequency devices (including mobile
telephones) must be excluded.
![Page 34: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/34.jpg)
Basic safety should be performed on line powered
before installation and after every repairs are;
– Ground wire integrity ( Resistance )
– Ground wire leakage
The basic electrical characteristic usually cause the
most leakage currents in modern equipment is
Capacitive Reactance Coupling in power cord.
The typical range of human body resistance with the
skin intact is 500 Ohm to 1,000 Ohm.
The “let go” current of a shock is 14 mA in most
people.
Basic Safety
![Page 35: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/35.jpg)
Electrical safety NOT dependent on voltage but on Leakage Current.
At low voltage, leakage current flow through body may be fatal to us.
Patient may connected to several device simultaneously . (ICU.)
Patient may connected conductively with electronic circuit. (ECG.)
Contact directly to internal tissue. ( natural orifices or break in the skin.)
Why Electrical Safety?
![Page 36: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/36.jpg)
Electricity - Physiological Effects
Burns
When an electric current passes through
any substance having electrical
resistance, heat is produced. The amount
of heat depends on the power-dissipated
(I2R or VI). Whether or not the heat
produces a burn depends on the current
density.
![Page 37: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/37.jpg)
Burns
![Page 38: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/38.jpg)
Electricity - Physiological Effects
Muscle Cramps
When an electrical stimulus is applied to a
motor nerve or muscle, the muscle does exactly
what it is designed to do in the presence of such
a stimulus i.e. it contracts. The prolonged
involuntary contraction of muscles (tetanus)
caused by external electrical stimulus is
responsible for the phenomenon where a
person who is holding an electrically live object
can be unable to let go.
![Page 39: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/39.jpg)
A muscle cramp is a painful, involuntary muscle contraction.
![Page 40: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/40.jpg)
Electricity - Physiological Effects
Respiratory Arrest
The muscles between the ribs (intercostal
muscles) need to repeatedly contract and relax
in order in order to facilitate breathing.
Prolonged tetanus of these muscles can
therefore prevent breathing.
![Page 41: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/41.jpg)
Respiratory Arrest
![Page 42: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/42.jpg)
Electricity - Physiological Effects
Cardiac Arrest
The heart is a muscular organ which needs to
able to contract and relax repetitively in order to
perform its function as a pump for the blood.
Tetanus of the heart musculature will prevent
the pumping process.
![Page 43: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/43.jpg)
Cardiac Arrest
![Page 44: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/44.jpg)
Electricity - Physiological Effects
Ventricular Fibrillation
The ventricles of the heart are the chambers
responsible for pumping blood out of the heart.
When the heart is in ventricular fibrillation, the
musculature of the ventricles undergoes irregular,
uncoordinated twitching resulting in no net blood
flow. The condition proves fatal if not corrected in
a very short space of time.
![Page 45: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/45.jpg)
Ventricular fibrillation (VF) is a deadly arrhythmia.
In VF, the electrical signals that trigger the
heartbeat become very fast and chaotic in the
lower chambers of the heart. The heart no longer
can pump blood to the brain or body.
![Page 46: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/46.jpg)
Electricity - Physiological Effects
![Page 47: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/47.jpg)
Electricity - Physiological Effects
![Page 48: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/48.jpg)
Electricity - Physiological Effects
Electrolysis
The movement of ions of opposite polarities in
opposite directions through a medium is called
electrolysis and can be made to occur be made to
occur by passing DC currents through body
tissues or fluids.
![Page 49: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/49.jpg)
electrode(metal)
wire(metal)
electrode(metal)
wire(metal)
e-Charge carrier
TissueElectrolyte/Dielectric
cell
skin
ionsCharge carrier
Electricity Conduction Mechanism
Electrochemical Reactions
e-Charge carrier
![Page 50: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/50.jpg)
Effect of entry point on current distribution
For dog: 20A, human : 80~600 A
Safety limit : 10 A
Important Susceptibility Parameters
![Page 51: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/51.jpg)
Macroshock
Hazards
Macroshock is the most
common type of shock
received and occurs when the
human body becomes a
conductor of electric current
passing by means other than
directly through the heart.
![Page 52: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/52.jpg)
Microshock
Hazards
Electrically conducting devices
unintentionally make direct
contact with heart muscles,
ventricular fibrillation may be
induced by minute current (e.g.
200 uA) well below the threshold
of feeling.
![Page 53: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/53.jpg)
General Electrical Safety
![Page 54: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/54.jpg)
General Electrical Safety
![Page 55: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/55.jpg)
General Electrical Safety
![Page 56: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/56.jpg)
General Electrical Safety
![Page 57: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/57.jpg)
General Electrical Safety
![Page 58: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/58.jpg)
General Electrical Safety
![Page 59: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/59.jpg)
General Electrical Safety
![Page 60: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/60.jpg)
General Electrical Safety
![Page 61: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/61.jpg)
Electricity - Leakage Currents
Leakage Currents
Current that is not functional. The following
leakage currents are defined: Earth Leakage
Current , Enclosure Leakage Current , and
Patient Leakage Current .
Earth
Leakage
Current
Enclosure
Leakage
Current
Patient
Leakage
Current
![Page 62: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/62.jpg)
Electricity - Leakage Currents
Causes of leakage currents
If any conductor is raised to a potential above earth
potential, then some current is bound to flow from that
conductor to earth. This is true even of conductors that are
well insulated from earth, since there is no such thing as
perfect insulation or infinite resistance. The amount of
current that flows depends on:
• The voltage on the conductor
• The capacitive reactance between the conductor and
earth
• The resistance between the conductor and earth
![Page 63: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/63.jpg)
Leakage
Current
Leakage current flows between
power supply wires and earth
ground within the rated value. In
case of the multiple operation,
however, the total leakage current
will be the sum of the leakage
current flows from each power
supply.
![Page 64: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/64.jpg)
Electricity - Leakage Currents
Earth leakage current
Earth leakage current is the current which
normally flows in the earth conductor of a
protectively earthed piece of equipment.
![Page 65: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/65.jpg)
Electricity - Leakage Currents
![Page 66: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/66.jpg)
Enclosure leakage current
Enclosure leakage current is described as
the current that flows from an exposed
conductive part of the conductor to earth
through a conductor other than the
protective earth conductor.
Electricity –
Enclosure Leakage Currents
![Page 67: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/67.jpg)
Electricity –
Enclosure Leakage Currents
![Page 68: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/68.jpg)
Patient leakage current
Patient leakage current is the leakage current that
flows through a patient connected to an applied
part or parts.
Electricity –
Patient Leakage Current
![Page 69: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/69.jpg)
Electricity –
Patient Leakage Current
![Page 70: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/70.jpg)
Patient auxiliary current
The patient auxiliary current is defined as the
current which normally flows between parts of
the applied part through the patient which is not
intended to produce a physiological effect.
Electricity –
Patient Auxiliary Current
![Page 71: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/71.jpg)
Electricity –
Patient Auxiliary Current
![Page 72: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/72.jpg)
Classes of Equipment
All electrical equipment is categorized into
classes according to the method of protection
against electric shock that is used.
Types of Equipment
The degree of protection for medical electrical
equipment is defined by the type designation.
Electricity –
Class and Types of Equipment
![Page 73: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/73.jpg)
Classes of Equipment
All electrical equipment is categorized into
classes according to the method of protection
against electric shock that is used.
Electricity –
Class of Equipment
CLASS I CLASS II
CLASS III
![Page 74: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/74.jpg)
Class I equipment
Class 1 equipment has a protective earth. The
basic means of protection is the insulation
between live parts and exposed conductive
parts such as the metal enclosure. In the event
of a fault which would otherwise cause an
exposed conductive part to become live, the
supplementary protection (i.e. protective earth)
comes into effect.
Electricity –
Class of Equipment
![Page 75: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/75.jpg)
General Electrical Safety
![Page 76: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/76.jpg)
CLASS I EQUIPMENT
![Page 77: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/77.jpg)
Class II equipment
The method of protection against electric shock
in the case of class II equipment is either double
insulation or reinforced insulation. In double
insulated equipment the basic protection is
afforded by the first layer of insulation. If basic
protection fails then supplementary protection is
afforded by a second layer of insulation
preventing contact with live parts.
Electricity –
Class of Equipment
![Page 78: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/78.jpg)
Electricity –
Class of Equipment
Class II equipment
The symbol for class II equipment is 2
concentric squares indicating double
insulation as shown below.
![Page 79: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/79.jpg)
General Electrical Safety
![Page 80: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/80.jpg)
CLASS II EQUIPMENT
![Page 81: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/81.jpg)
Class III equipment
Class III equipment is defined as that in which
protection against electric shock relies on the
fact that no voltages higher than safety extra
low voltage (SELV) are present.SELV is defined
in turn in the relevant standard as a voltage not
exceeding 25V ac or 60V dc.
Electricity –
Class of Equipment
![Page 82: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/82.jpg)
Types of Equipment
The degree of protection for medical
electrical equipment is defined by the type
designation.
Electricity –
Types of Equipment
Type B Type BF Type CF
![Page 83: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/83.jpg)
Type Symbol Definition
B Equipment providing a particular degree of protection against electric shock, particularly regarding allowable leakage currents and reliability of the protective earth connection (if present).
BF As type B but with isolated or floating (F type) applied part or parts
CF Equipment providing a higher degree of protection against electric shock than type BF, particularly with regard to allowable leakage currents, and having floating applied parts.
Electricity –
Types of Equipment
![Page 84: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/84.jpg)
Normal Condition
A basic principle behind the philosophy of
electrical safety is that in the event of a
single abnormal external condition arising
or of the failure of a single means of
protection against a hazard, no safety
hazard should arise.
Electrical Safety Tests
![Page 85: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/85.jpg)
Single Fault Condition
Condition in which a single means of
electrical safety protection is defective or
an abnormal condition is present.
Examples of a single fault condition would
be interruption of the ground conductor on
a Class I equipment or opening of the
neutral supply conductor to the
equipment.
Electrical Safety Tests
![Page 86: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/86.jpg)
Single Fault Condition
Electrical Safety Tests
![Page 87: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/87.jpg)
Protective Earth Continuity
Applicable to Class1, all types
Limit 0.2 ohms
![Page 88: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/88.jpg)
Insulation Tests
Applicable to Class 1, all types
Limit Not less than 50Mohms
![Page 89: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/89.jpg)
Insulation Tests
Applicable to Class II, all types having applied parts
Limit : > 50 MOhms
![Page 90: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/90.jpg)
Earth Leakage Current
Applicable to: Class 1 equipment all types
Limits: 0.5mA in NC, 1mA in SFC or 5mA and 10mA respectively for permanently installed equipment
![Page 91: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/91.jpg)
Enclosure Leakage Current
Applicable to Class 1 and class II equipment , all types.
Limit0.1mA in NC, 0.5mA in SFC
![Page 92: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/92.jpg)
Patient Leakage Current
Applicable to All Classes B and BF equipment having applied parts
Limits 0.1mA in NC, 0.5mA in SFC
![Page 93: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/93.jpg)
Patient Auxiliary Current
Applicable to All equipment having applied parts.
Limits 0.1mA in NC, 0.5mA in SFC
![Page 94: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/94.jpg)
Mains on Applied Parts
Applicable to Class 1 and class II equipment , types BF&CF equipment having applied parts.
Limits BF 5mA; CF 0.05mA per electrode
![Page 95: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/95.jpg)
Safety Test as done by
BEMS
![Page 96: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/96.jpg)
Electrical Safety and the Patients Depends on
Three Things: -
• An alert, caring, knowledgeable person
• Properly maintained, and applied equipment
• Proper grounding or double insulation of line-
powered equipment
Electrical Safety Review
![Page 97: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/97.jpg)
Report, tag, and do not use equipment with:
• Frayed wires or cracked insulation
• Damaged plugs or missing ground prongs
Electrical Safety Review
![Page 98: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/98.jpg)
Report, tag, and do not use receptacles with:
• Only two slots
• Missing cover plates
• Loose mountings
• Weak gripping force
Electrical Safety Review
![Page 99: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/99.jpg)
Inspections of Equipment: -
• Check of Strain-Relief
Electrical Safety
Review
![Page 100: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/100.jpg)
Avoid:
• Using cheaters (three-prong adapters)
• Using extension cords
• Placing liquids on electrical equipment
• Pulling plugs out of receptacles by the cord
• Rolling equipment over power cords
• Using defective equipment or receptacles
• Using equipment that sparks, smokes, or shocks
Electrical Safety Review
![Page 101: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/101.jpg)
Be sure that users know how to use equipment
properly. Also be sure to:
• Check that line-powered equipment has three-
wire grounding cords, unless device is double
insulated
• Ask for help and instruction when needed
• Report defective equipment and receptacles
Electrical Safety Review
![Page 102: Medical Electrical Safety](https://reader030.fdocuments.us/reader030/viewer/2022020115/554e22afb4c90511778b4f5c/html5/thumbnails/102.jpg)
Always use your common sense
Do not end up like this person!
Electrical Safety Review