Absorbance Based Light Emitting Diode Optical Sensors and ...
Meat Color Measurement Three approaches --- all very different in what they measure 1. Light...
-
Upload
ursula-oliver -
Category
Documents
-
view
216 -
download
0
Transcript of Meat Color Measurement Three approaches --- all very different in what they measure 1. Light...
Meat Color Measurement
Three approaches --- all very different in what they measure
1. Light absorbance methods
2. Light reflectance measurements
3. Human panel assessments
Light absorbance methods for meat color
Pigment quantitation --- how much?
a. extract the pigments from the meat sample
-fresh meat, use water or weak salt
solution
-this is for for heme AND globin
(intact myoglobin)
- cured meat- acetone for heme alone-the pigment is not water soluble
b. measure light absorbed at a wavelength appropriate for the pigment being measured
maximum absorbance:oxymyoglobin 580 nm
myoglobin 550 nm
metmyoglobin 505 nm
nitrosylhemochrome 545nm
c. Can determine total amount of pigment
i.e. 250 ppm of oxymyoglobin:
but………………?
color is not obvious – usually from a mixture of pigments
Amount of pigment is used for color values in calculating formulas ---
because it shows the potential intensity of color
– works well for cured meat pigment but difficult to do with fresh meat pigment because of rapid changes while handling
– also valuable in cured meat because it is related to effectiveness of curing reactionMore complete conversion for NO2
- NO gives greater amounts of nitrosylhemochrome
Light reflectance measurements
– Truly objective quantitation of color– widely used– measures light reflected from surfaces ---
– consequently measures what we see
– objective values (color numbers) allow communication of color and exact reproduction of color– i.e. paint industry
All reflectance systems are based on the original 1931 CIE system (International
Commission on Illumination)- established use of standardized light sources
i.e. Source A - incandescent lamp
B - noon sunlight
C - average daylight
– more recently these have become defined by lamp temperature
i.e. lamp at 6500 Ko is D65
– also may include flourescent lamps, etc.– must also consider aperture size and sample locations,
multiple measurements, i.e., 460 lab
Human vision is sensitive to light in the range of ~380 nm (blue) to 700+ nm (red)
– Lighter color reflect more total light over all wavelengths
– darker colors reflect less total light
There are three characteristics of any color that can be measured and reproduced
1. color perceived i.e. red, green, blue– due to dominant wavelength reflected usually called “hue”
2. lightness/darkness of the color– due to total amount of light reflected over the range of
380 - 700 nm
– called lightness or “value”
3. intensity of the color– due to the proportion of the total light reflected that occurs
at the dominant wave length
– called chroma or “purity”, “strength”, “intensity”, vividness”
All color reflectance measurement systems use 3 numbers and most can be converted
to others
CIE uses X,Y, Z
Hunter L, a, b
Minolta can be set to any of several systems
Hunter has been largely adopted for meat color measurement because
L = lightness
a = redness (+) green (-)
b = yellow (+) blue (-)
L measures Lightness 100 - perfect whiteness 0 - Black
a Redness + Gray 0 Greeness -
b Yellowness + Gray 0 Blueness -
Example of pigments in solution for reflectance values
myoglobin 8.0 13.7 7.6
metmyoglobin 8.7 4.1 10.6
nitrosylhemochrome 26.0 19.5 15.1
meat samples typically will give L values of 35-60 and a values of 2 -30.
L a boxymyoglobin 8.0 28.6 13.6
Human panels --- visual appraisal
– much more difficult than it seems due to individual perceptions and biases
– try to remove “human” factor as much as possible
– can include “liking” or “preference” for a specific color which the absorbance or reflectance values cannot do